 Class 2 Device Recall M/DN Intramedullary Fixation see related information
Date Posted February 20, 2014
Recall Status1 Open
Recall Number Z-1010-2014
Recall Event ID 66500
Product Classification Screw, fixation, bone, non-spinal, metallic - Product Code NDJ
Product INTERLOCKING IM SCREW 4.5 and INTERLOCKING IM SCREW LG

Some items starting with codes 0022530
Code Information 60625043
60678699
60690742
60728338
60747676
60757045
60813551
60864020
60864021
60915935
60915936
61003973
60806395
60829584
60864023
60867711
60867767
60550320
60569951
60597681
60653347
60690746
60735651
60757047
60757048
60808005
60817585
60829586
60829587
60864037
60864038
60864040
60867771
61161394
61190864
61194691
61215177
61319429
61379510
60518291
60606752
60620346
60711145
60728343
60757050
60806397
60806398
60829592
60829594
60867782
60867785
61071422
61150075
61173568
61319435
61319436
61379514
60748952
60762088
60806399
60876899
60876901
60876902
61248174
61325065
60123303
60130220
60135214
60142918
60150073
60218846
60243102
60279663
60309141
60323311
60338439
60349093
60387528
60393392
60407659
60422612
60422615
60422617
60422618
60505967
60505968
60569960
60597683
60597684
60602462
60620369
60620371
60690750
60690751
60714161
60714162
60735662
60735663
60757051
60757052
60808010
60808011
60814567
60814568
60829596
60829597
60867787
60867788
60867789
60867790
60867792
60956540
60956547
61003994
61004003
61004006
61004010
61035979
61035980
61064854
61066610
61066611
61071425
61071426
61084123
61100815
61158640
61211625
61211626
61221436
61249859
61271267
61286461
61421225
61496669
61509881
61535667
61564743
61664934
61729550
61729552
61746802
61746803
62262065
60123304
60132558
60142920
60152971
60253769
60253770
60280195
60288490
60310563
60310564
60310565
60323343
60338521
60367544
60393393
60393395
60407661
60410614
60438323
60451198
60472688
60487155
60487156
60491717
60491718
60505969
60602464
60602465
60602466
60602467
60620387
60653365
60690752
60690753
60714165
60735665
60735666
60757057
60757058
60757059
60817613
60829604
60867807
60867809
60867811
60867813
60956553
60979622
60985458
60985459
60985460
61024754
61024757
61031015
61031016
61035968
61035972
61041707
61041709
61041713
61064855
61066614
61066615
61067796
61073739
61091734
61096524
61104693
61133860
61133862
61150093
61153543
61199894
61210272
61224276
61269415
61295067
61362462
61382019
61489978
61504080
61504081
61564747
61650565
61657650
61664935
60099580
60103529
60109578
60109579
60123308
60123309
60127535
60132562
60132563
60135218
60140394
60142924
60142925
60150077
60154195
60154196
60154197
60155336
60167149
60169750
60186875
60218855
60229996
60242017
60243123
60243124
60253778
60253779
60253780
60265723
60265724
60265725
60288499
60288501
60288502
60288503
60288504
60288505
60310578
60310582
60323351
60323352
60338534
60338537
60338539
60338540
60358127
60367556
60367560
60380468
60385469
60393402
60393403
60407668
60407669
60410616
60422634
60425160
60425161
60444819
60444820
60452048
60455028
60472693
60472694
60479134
60479135
60487157
60491724
60491725
60505973
60505974
60505975
60505976
60505977
60569969
60569970
60575726
60602498
60602499
60602501
60602502
60602504
60602505
60602506
60620394
60620395
60620396
60630485
60653377
60653381
60674212
60690760
60690761
60714182
60714183
60714184
60714186
60735674
60735676
60757069
60757070
60757072
60757073
60757074
60817603
60817604
60817605
60817607
60817608
60817609
60876980
60876984
60876985
60876987
60926474
60926475
60926476
60926478
60926479
60926480
60926481
60926482
60966586
60966587
60966588
60966590
60966591
60979628
60985467
61009884
61024666
61030980
61030983
61035951
61041617
61041618
61041698
61064857
61066619
61066621
61067798
61071429
61071440
61075712
61075713
61084125
61084126
61096534
61096536
61133873
61133877
61150110
61150112
61150114
61150116
61161408
61199901
61202928
61202931
61228843
61269436
61269439
61307114
61329372
61343035
61343037
61356918
61356922
61356926
Recalling Firm/
Manufacturer
Zimmer, Inc.
345 E Main St
Warsaw IN 46580-2746
800-613-6131
For Additional Information Contact
800-613-6131
Manufacturer Reason
for Recall
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged pro
FDA Determined
Cause 2
Packaging change control
Action Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Quantity in Commerce 7,044,680 total
Distribution Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55