|
Class 2 Device Recall GE Healthcare Carescape Patient Data Module |
|
Date Initiated by Firm |
November 27, 2013 |
Date Posted |
January 14, 2014 |
Recall Status1 |
Terminated 3 on September 07, 2016 |
Recall Number |
Z-0747-2014 |
Recall Event ID |
67115 |
510(K)Number |
K071073
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
GE Healthcare Carescape Patient Data Module
The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. |
Code Information |
Mfg Lot or Serial # SA307140598GA SA307140599GA SA307140600GA SA307140601GA SA307140605GA SA307140606GA SA307140607GA SA307140610GA SA307140611GA SA307140612GA SA307140614GA SA307140615GA SA307140618GA SA307140619GA SA307140622GA SA307140623GA SA307140627GA SA307140628GA SA307140631GA SA307140632GA SA307140636GA SA307140637GA SA307140639GA SA307140640GA SA307140641GA SA307140642GA SA307140644GA SA307150645GA SA307150646GA SA307150655GA SA307150656GA SA307020391GA SA307030405GA SA307030424GA SA307070446GA SA307070447GA SA307070448GA SA307080462GA SA307080463GA SA307080484GA SA307080508GA SA307080523GA SA307110548GA SA307110556GA SA307110571GA SA307190716GA SA307190737GA SA307190739GA SA307190750GA SA307190768GA SA307020379GA SA307020384GA SA307020386GA SA307030407GA SA307030409GA SA307030412GA SA307030414GA SA307060432GA SA307070436GA SA307070440GA SA307070441GA SA307070442GA SA307070450GA SA307070453GA SA307070454GA SA307070455GA SA307080472GA SA307080477GA SA307080478GA SA307080504GA SA307080520GA SA307080526GA SA307110559GA SA307110560GA SA307110572GA SA307110575GA SA307110579GA SA307120587GA SA307120588GA SA307150681GA SA307180694GA SA307180696GA SA307180697GA SA307190717GA SA307190718GA SA307190720GA SA307190722GA SA307190724GA SA307190727GA SA307190728GA SA307190729GA SA307190730GA SA307190732GA SA307190733GA SA307190734GA SA307190738GA SA307190741GA SA307190742GA SA307190743GA SA307190744GA SA307190745GA SA307190749GA SA307190764GA SA307190766GA SA307190769GA SA307190770GA SA307190771GA SA307020383GA SA307030411GA SA307030413GA SA307050429GA SA307070443GA SA307080459GA SA307080467GA SA307080521GA SA307080522GA SA307080524GA SA307150651GA SA307180699GA SA307190736GA SA307190755GA SA307190773GA SA307190774GA SA307200784GA SA307220983GA SA307221081GA SA307221089GA SA307020393GA SA307020396GA SA307020397GA SA307020399GA SA307030417GA SA307030419GA SA307030423GA SA307070438GA SA307080473GA SA307080493GA SA307080498GA SA307080499GA SA307080509GA SA307080510GA SA307080533GA SA307150671GA SA307180698GA SA307180701GA SA307180702GA SA307180704GA SA307180705GA SA307180706GA SA307190731GA SA307190751GA SA307190752GA SA307190754GA SA307190756GA SA307190757GA SA307190772GA SA307200776GA SA307200777GA SA307210815GA SA307210819GA SA307210863GA SA307220869GA SA307220870GA SA307220872GA SA307220875GA SA307220877GA SA307220878GA SA307220879GA SA307220880GA SA307220888GA SA307220890GA SA307220892GA SA307220893GA SA307220898GA SA307220900GA SA307220902GA SA307220903GA SA307220904GA SA307220923GA SA307220931GA SA307220932GA SA307220963GA SA307220986GA SA307220999GA SA307221016GA SA307221026GA SA307221027GA SA307221028GA SA307221092GA SA307221084GA SA307221109GA SA307221179GA SA307231201GA SA307231375GA SA307231402GA SA307231416GA SA307231475GA SA307080502GA SA307231199GA SA307231208GA SA307231374GA SA307080466GA SA307080469GA SA307080488GA SA307080489GA SA307080491GA SA307080497GA SA307140621GA SA307180707GA SA307180708GA SA307190753GA SA307190759GA SA307190760GA SA307190761GA SA307190762GA SA307200792GA SA307200793GA SA307200794GA SA307200795GA SA307200796GA SA307200797GA SA307200798GA SA307200799GA SA307200801GA SA307200802GA SA307200803GA SA307200804GA SA307210829GA SA307210835GA SA307220936GA SA307220945GA SA307221000GA SA307221001GA SA307221003GA SA307221009GA SA307221043GA SA307221045GA SA307221047GA SA307221061GA SA307221064GA SA307221070GA SA307231264GA SA307231351GA SA307220949GA SA307210832GA SA307210833GA SA307210838GA SA307210843GA SA307220938GA SA307220942GA SA307220954GA SA307220955GA SA307220959GA SA307221039GA SA307221040GA SA307221041GA SA307221048GA SA307221056GA SA307221059GA SA307221068GA SA307220908GA SA307221100GA SA307221101GA SA307221110GA SA307221113GA SA307221152GA SA307221158GA SA307221160GA SA307221164GA SA307221170GA SA307221175GA SA307231211GA SA307231218GA SA307231255GA SA307231268GA SA307231365GA SA307231469GA SA307150675GA SA307221030GA SA307221069GA SA307221096GA SA307221097GA SA307221104GA SA307221106GA SA307221108GA SA307221119GA SA307221144GA SA307221145GA SA307221154GA SA307221156GA SA307221162GA SA307221168GA SA307221169GA SA307221171GA SA307221172GA SA307221173GA SA307221177GA SA307221181GA SA307221183GA SA307221186GA SA307231189GA SA307231194GA SA307231196GA SA307231209GA SA307231215GA SA307231217GA SA307231392GA SA307221096GA SA307221106GA SA307221154GA SA307221156GA SA307221168GA SA307221171GA SA307221172GA SA307221173GA SA307221183GA SA307231209GA SA307200805GA SA307210837GA SA307220937GA SA307220941GA SA307220946GA SA307220948GA SA307220950GA SA307220952GA SA307220956GA SA307220958GA SA307220961GA SA307221004GA SA307221011GA SA307221012GA SA307221037GA SA307221052GA SA307221067GA SA307221076GA SA307221102GA SA307221178GA SA307231200GA SA307231202GA SA307231207GA SA307231387GA SA307231391GA SA307231393GA SA307231395GA SA307231396GA SA307231426GA SA307231438GA SA307231467GA SA307231438GA SA307220947GA SA307221072GA SA307231390GA SA307231433GA SA307220939GA SA307220943GA SA307220957GA SA307220960GA SA307221060GA SA307221062GA SA307210826GA SA307210828GA SA307210831GA SA307210840GA SA307210841GA SA307220940GA SA307220944GA SA307220976GA SA307220984GA SA307220991GA SA307221002GA SA307221005GA SA307221006GA SA307221044GA SA307221046GA SA307221050GA SA307221057GA SA307221058GA SA307221066GA SA307221074GA SA307221120GA SA307221125GA SA307221126GA SA307221129GA SA307221130GA SA307221132GA SA307221136GA SA307221137GA SA307231204GA SA307231257GA SA307231262GA SA307231263GA SA307231352GA SA307231366GA SA307231369GA SA307231378GA SA307231379GA SA307231413GA SA307231414GA SA307231419GA SA307231428GA SA307231429GA SA307231444GA SA307231445GA SA307241510GA SA307241512GA SA307241513GA SA307241514GA SA307241512GA SA307220897GA SA307231205GA SA307231376GA SA307231385GA SA307231474GA SA307210821GA SA307210836GA SA307221007GA SA307221051GA SA307221105GA SA307221176GA SA307221185GA SA307231210GA SA307231275GA SA307231362GA SA307231423GA SA307231463GA SA307241582GA SA307241676GA SA307241682GA SA307231358GA SA307231359GA SA307231364GA SA307231479GA SA307241519GA SA307241577GA SA307241583GA SA307241674GA SA307241677GA SA307241690GA SA307241706GA SA307251724GA SA307251726GA SA307251746GA SA307251747GA SA307251757GA SA307251727GA SA307221038GA SA307221079GA SA307221124GA SA307221134GA SA307221139GA SA307221140GA SA307221142GA SA307231220GA SA307231223GA SA307231224GA SA307231225GA SA307231226GA SA307231228GA SA307231229GA SA307231231GA SA307231232GA SA307231234GA SA307231236GA SA307231243GA SA307231245GA SA307231271GA SA307231277GA SA307231279GA SA307231286GA SA307231291GA SA307231293GA SA307231294GA SA307231295GA SA307231299GA SA307231300GA SA307231304GA SA307231306GA SA307231310GA SA307231312GA SA307231313GA SA307231314GA SA307231318GA SA307231324GA SA307231325GA SA307231293GA SA307231312GA SA307231222GA SA307231247GA SA307231278GA SA307231284GA SA307231317GA SA307231321GA SA307231322GA SA307241562GA SA307220906GA SA307221115GA SA307241581GA SA307221008GA SA307231315GA SA307241686GA SA307241673GA SA307241708GA SA307251710GA SA307251713GA SA307251735GA SA307210824GA SA307210825GA SA307220953GA SA307231248GA SA307231276GA SA307231305GA SA307231329GA SA307231334GA SA307231335GA SA307231336GA SA307231340GA SA307231344GA SA307231346GA SA307231347GA SA307231348GA SA307231448GA SA307231449GA SA307231451GA SA307231454GA SA307231456GA SA307231458GA SA307231493GA SA307231494GA SA307231495GA SA307231496GA SA307231497GA SA307231498GA
SA307231500GA SA307231502GA SA307241525GA SA307241527GA SA307241533GA SA307241534GA SA307241539GA SA307241540GA SA307241546GA SA307241549GA SA307241556GA SA307241557GA SA307241558GA SA307241560GA SA307241561GA SA307241565GA SA307241568GA SA307241589GA SA307241590GA SA307241605GA SA307241610GA SA307241611GA SA307241612GA SA307241615GA SA307241621GA SA307241622GA SA307241623GA SA307241626GA SA307241628GA SA307241631GA SA307241632GA SA307241633GA SA307261816GA SA307261825GA SA307221033GA SA307221042GA SA307231301GA SA307231338GA SA307291973GA SA307291974GA SA307291977GA SA307291978GA SA307291979GA SA307291980GA SA307291981GA SA307291982GA SA307080483GA SA307080501GA SA307190775GA SA307221034GA SA307221075GA SA307231221GA SA307231238GA SA307231239GA SA307231273GA SA307231274GA SA307231290GA SA307231297GA SA307231320GA SA307231326GA SA307231327GA SA307231460GA SA307241531GA SA307241535GA SA307241543GA SA307241545GA SA307241548GA SA307241553GA SA307241564GA SA307241566GA SA307241569GA SA307241570GA SA307241571GA SA307241572GA SA307241593GA SA307241595GA SA307241597GA SA307241598GA SA307241603GA SA307241606GA SA307241607GA SA307241609GA SA307241613GA SA307241614GA SA307241616GA SA307241617GA SA307241618GA SA307241619GA SA307241620GA SA307241624GA SA307241627GA SA307241634GA SA307241635GA SA307241636GA SA307241637GA SA307241638GA SA307291976GA SA307220901GA SA307221013GA SA307251715GA SA307251719GA SA307251721GA SA307251729GA SA307251743GA SA307251750GA SA307302002GA SA307302050GA SA307302066GA SA307200800GA SA307210839GA SA307231394GA SA307241668GA SA307241691GA SA307251748GA SA307251769GA SA307291983GA SA307302047GA SA307302055GA SA307200788GA SA307220914GA SA307231431GA SA307231457GA SA307231462GA SA307241515GA SA307241530GA SA307241544GA SA307241547GA SA307241554GA SA307241578GA SA307241592GA SA307241602GA SA307241625GA SA307241640GA SA307241701GA SA307251711GA SA307251712GA SA307251716GA SA307251737GA SA307251740GA SA307251741GA SA307251778GA SA307251782GA SA307261824GA SA307261827GA SA307241516GA SA307080513GA SA307210827GA SA307220935GA SA307221010GA SA307221049GA SA307231230GA SA307231233GA SA307231235GA SA307231240GA SA307231246GA SA307231282GA SA307231303GA SA307231311GA SA307231316GA SA307231319GA SA307231323GA SA307231330GA SA307231333GA SA307231337GA SA307231341GA SA307231343GA SA307231345GA SA307231350GA SA307231453GA SA307231455GA SA307231459GA SA307231499GA SA307231501GA SA307241523GA SA307241528GA SA307241529GA SA307241536GA SA307241537GA SA307241538GA SA307241541GA SA307241542GA SA307241550GA SA307241551GA SA307241552GA SA307241563GA SA307241567GA SA307241588GA SA307241591GA SA307241594GA SA307241600GA SA307241604GA SA307241608GA SA307241630GA SA307251736GA SA307251738GA SA307251739GA SA307251755GA SA307261817GA SA307261818GA SA307261822GA SA307261826GA SA307291877GA SA307291878GA SA307291879GA SA307291880GA SA307291881GA SA307291882GA SA307291884GA SA307291886GA SA307291887GA SA307291926GA SA307291928GA SA307302003GA SA307302046GA SA307302051GA SA307302052GA SA307302053GA SA307302054GA SA307302056GA SA307302059GA SA307302062GA SA307302063GA SA307302064GA SA307302068GA SA307302069GA SA307302070GA SA307302071GA SA307302072GA SA307302074GA SA307302076GA SA307302077GA SA307302078GA SA307302079GA SA307302080GA SA307302081GA SA307302082GA SA307302083GA SA307302084GA SA307312087GA SA307312089GA SA307312093GA SA307312094GA SA307312095GA SA307312096GA SA307312097GA SA307312099GA SA307312101GA SA307312103GA SA307312106GA SA307312109GA SA307322147GA SA307251736GA SA307251738GA SA307251739GA SA307251755GA SA307302054GA SA307220935GA SA307221049GA SA307231230GA SA307231303GA SA307231311GA SA307231323GA SA307261822GA SA307291877GA SA307291926GA SA307302003GA SA307302051GA SA307302052GA SA307302053GA SA307302056GA SA307302059GA SA307302062GA SA307302063GA SA307302068GA SA307302069GA SA307302071GA SA307302074GA SA307302076GA SA307302077GA SA307302079GA SA307302080GA SA307302081GA SA307302082GA SA307302084GA SA307312089GA SA307312094GA SA307312097GA SA307312106GA SA307322147GA SA307231240GA SA307231282GA SA307231319GA SA307231343GA SA307231499GA SA307261818GA SA307261826GA SA307291878GA SA307291879GA SA307291880GA SA307291881GA SA307291882GA SA307291884GA SA307291886GA SA307291887GA SA307291928GA SA307302046GA SA307302064GA SA307302070GA SA307302072GA SA307302078GA SA307302083GA SA307312087GA SA307312093GA SA307312095GA SA307312096GA SA307312099GA SA307312101GA SA307312103GA SA307312109GA SA307251764GA SA307312108GA SA307322111GA SA307322113GA SA307322146GA SA307322160GA SA307322165GA SA307221116GA SA307221161GA SA307221163GA SA307221180GA SA307231411GA SA307241518GA SA307241575GA SA307241579GA SA307241585GA SA307241587GA SA307241670GA SA307241681GA SA307241684GA SA307241689GA SA307241705GA SA307251718GA SA307251722GA SA307251753GA SA307251754GA SA307251756GA SA307251758GA SA307251760GA SA307251762GA SA307251763GA SA307251765GA SA307251766GA SA307251767GA SA307251768GA SA307251770GA SA307251771GA SA307251772GA SA307251773GA SA307251774GA SA307251776GA SA307251777GA SA307251779GA SA307251784GA SA307251786GA SA307220988GA SA307221018GA SA307221053GA SA307221103GA SA307231332GA SA307231477GA SA307241680GA SA307241683GA SA307251734GA SA307251742GA SA307251744GA SA307251745GA SA307251751GA SA307251752GA SA307251775GA SA307251780GA SA307251781GA SA307251783GA SA307291961GA SA307291963GA SA307291964GA SA307291965GA SA307291966GA SA307291970GA SA307291971GA SA307220996GA SA307221107GA SA307231212GA SA307231397GA SA307231404GA SA307231406GA SA307231424GA SA307241695GA SA307251785GA SA307251788GA SA307251790GA SA307251791GA SA307251794GA SA307251795GA SA307251796GA SA307251797GA SA307251798GA SA307251800GA SA307251802GA SA307251805GA SA307261812GA SA307261814GA SA307261835GA SA307291897GA SA307291899GA SA307291905GA SA307291907GA SA307291910GA SA307291911GA SA307291913GA SA307291914GAMKE SA307291918GA SA307291920GAMKE SA307291935GA SA307322197GA SA307261829GA SA307261831GA SA307261833GA SA307231370GA SA307231473GA SA307251789GA SA307261828GA SA307261830GA SA307261836GA SA307261842GA SA307261847GA SA307291951GA SA307291952GA SA307291955GA SA307302016GA SA307291959GA SA307302057GA SA307322115GA SA307322189GA SA307322200GA SA307302075GA SA307332231GA SA307332233GAMKE SA307332234GA SA307332236GA SA307332237GA SA307332240GA SA307332271GA SA307231280GA SA307241526GA SA307261821GA SA307291885GA SA307291923GA SA307291924GA SA307291925GA SA307291929GA SA307291931GA SA307291932GA SA307291933GA SA307291934GA SA307291962GA SA307291967GA SA307291968GA SA307291969GA SA307291972GA SA307302001GA SA307302004GA SA307302048GA SA307200791GA SA307231287GA SA307231298GA SA307231331GA SA307231339GA SA307231342GA SA307231349GA SA307231450GA SA307231452GA SA307231461GA SA307241599GA SA307241601GA SA307241629GA SA307241639GA SA307261819GA SA307261823GA SA307291876GA SA307291883GA SA307291927GA SA307291930GA SA307301992GA SA307221135GA SA307231296GA SA307312090GA SA307332232GA SA307332238GA SA307332266GA SA307332267GA SA307332268GA SA307332269GA SA307332270GA SA307332272GA SA307332273GA SA307332274GA SA307332275GA SA307332277GA SA307332279GA SA307332280GA SA307332282GA SA307332283GA SA307332284GA SA307332285GA SA307332286GA SA307332287GA SA307332288GA SA307332290GA SA307332291GA SA307332292GA SA307342338GA SA307322123GA SA307322156GA SA307322157GA SA307342331GA SA307241702GA SA307291949GA SA307221036GAMKE SA307231237GA SA307231281GA SA307231292GA SA307241702GA SA307291949GA SA307302049GA SA307302067GA SA307302085GA SA307312086GA SA307312088GA SA307312091GA SA307312092GA SA307312098GA SA307312102GA SA307322158GA SA307322183GA SA307332235GA SA307332239GA SA307332281GA SA307332289GA SA307342333GA SA307342334GA SA307342335GAMKE SA307342337GA SA307342339GA SA307342340GA SA307342341GA SA307342342GA SA307342344GA SA307342345GA SA307342346GA SA307342348GA SA307342349GA SA307342350GA SA307241702GA SA307291949GA SA307231354GA SA307231377GA SA307231381GA SA307231427GA SA307231434GA SA307231443GA SA307231476GA SA307241508GA SA307241669GA SA307241672GA SA307291888GA SA307291948GA SA307291953GA SA307291957GA SA307291958GA SA307301986GA SA307301987GA SA307301989GA SA307020398GA SA307020400GA SA307030420GA SA307220951GA SA307231219GA SA307231241GA SA307231307GA SA307231354GA SA307231377GA SA307231381GA SA307231427GA SA307231434GA SA307231443GA SA307231476GA SA307241508GA SA307241669GA SA307241672GA SA307261820GA SA307291888GA SA307291948GA SA307291953GA SA307291957GA SA307291958GA SA307301986GA SA307301987GA SA307301989GA SA307312100GA SA307322153GA SA307322155GA SA307322184GA SA307322185GA SA307322188GA SA307322194GA SA307322195GA SA307322196GA SA307332278GA SA307342332GA SA307342336GA SA307342343GA SA307342347GA SA307241508GA SA307241669GA SA307241672GA SA307291888GA SA307291948GA SA307291953GA SA307291957GA SA307291958GA SA307301986GA SA307301987GA SA307301989GA SA307231384GA SA307231398GA SA307231439GA SA307231470GA SA307231483GA SA307231484GA SA307231486GA SA307231488GA SA307231489GA SA307241504GA SA307241506GA SA307241507GA SA307231384GA SA307231398GA SA307231439GA SA307231470GA SA307231483GA SA307231484GA SA307231486GA SA307231488GA SA307231489GA SA307241504GA SA307241506GA SA307241507GA SA307291946GA SA307291943GA SA307301988GA SA307301996GA SA307302006GA SA307302008GA SA307302010GA SA307302011GA SA307302017GA SA307382627GA SA307382628GA SA307382629GA SA307382630GA SA307382631GA SA307382632GA SA307382633GA SA307382634GA SA307382635GA SA307382636GA SA307382637GA SA307382638GA SA307382639GA SA307382640GA SA307382641GA SA307382642GA SA307382643GA SA307382644GA SA307382645GA SA307382646GA SA307382647GA SA307382648GA SA307382649GA SA307382650GA SA307382651GA SA307382652GA SA307382653GA SA307382655GA SA307382656GA SA307382659GA SA307382660GA SA307382661GA SA307382662GA SA307382663GA SA307382664GA SA307382665GA SA307382666GA SA307392667GA SA307392668GA SA307392669GA SA307392671GA SA307392672GA SA307392673GA SA307392675GA SA307392676GA SA307392677GA SA307392678GA SA307392679GA SA307392682GA SA307392699GA SA307392700GA SA307392701GA SA307392702GA SA307392703GA SA307392705GA SA307392707GA SA307392710GA SA307392713GA SA307392714GA SA307231260GA SA307231355GA SA307231383GA SA307231260GA SA307231355GA SA307231383GA SA307392683GA SA307392684GA SA307392688GA SA307392690GA SA307392692GA SA307392694GA SA307392695GA SA307392696GA SA307392697GA SA307392698GA SA307392704GA SA307392706GA SA307392709GA SA307392711GA SA307392715GA SA307392716GA SA307392718GA SA307392719GA SA307392747GA SA307221118GA SA307231203GA SA307231250GA SA307231251GA SA307231265GA SA307231266GA SA307231382GA SA307231388GA SA307231399GA SA307231405GA SA307231410GA SA307231412GA SA307231418GA SA307231420GA SA307231432GA SA307231435GA SA307231437GA SA307231490GA SA307241503GA SA307241505GA SA307221118GA SA307231203GA SA307231250GA SA307231251GA SA307231265GA SA307231266GA SA307231382GA SA307231388GA SA307231399GA SA307231405GA SA307231410GA SA307231412GA SA307231418GA SA307231420GA SA307231432GA SA307231435GA SA307231437GA SA307231490GA SA307241503GA SA307241505GA SA307322150GA SA307322152GAMKE SA307332276GA SA307382658GA SA307392691GA SA307392708GA SA307392736GA SA307392746GA SA307392753GA SA307392761GA SA307382654GA SA307392680GA SA307392681GA SA307392685GA SA307392686GA SA307392717GA SA307392720GA SA307392721GA SA307392722GA SA307392724GA SA307392726GA SA307392727GA SA307392728GA SA307392729GA SA307392730GA SA307392731GA SA307392732GA SA307392733GA SA307392734GA SA307392735GA SA307392737GA SA307392738GA SA307392739GA SA307392740GA SA307392741GA SA307392742GA SA307392743GA SA307392744GA SA307392745GA SA307392749GA SA307392750GA SA307392751GA SA307392752GA SA307392754GA SA307392755GA SA307392757GA SA307392758GA SA307392759GA SA307392760GA SA307392762GA SA307392763GA SA307392764GA SA307392765GA SA307392767GA SA307392768GA SA307392769GA SA307392820GA SA307392821GA SA307392823GA SA307392824GA SA307392825GA SA307392827GA SA307392828GA SA307392829GA SA307392832GA SA307392833GA SA307392834GA SA307392835GA SA307392836GA SA307392837GA SA307392838GA SA307392839GA SA307392840GA SA307392841GA SA307392842GA SA307392843GA SA307392844GA SA307392845GA SA307392846GA SA307392849GA SA307392850GA SA307392851GA SA307392852GA SA307392853GA SA307392854GA SA307392856GA SA307392858GA SA307392859GA SA307392860GA SA307392861GA SA307392862GA SA307392863GA SA307392864GA SA307392865GA SA307392866GA SA307392867GA SA307392868GA SA307392869GA SA307392932GA SA307392933GA SA307392934GA SA307392935GA SA307392936GA SA307392937GA SA307392938GA SA307392939GA SA307392941GA SA307392942GA SA307392943GA SA307392944GA SA307392945GA SA307392947GA SA307392948GA SA307392949GA SA307392951GA SA307392953GA SA307392954GA SA307392955GA SA307392956GA SA307392957GA SA307392958GA SA307392959GA SA307392963GA SA307392965GA SA307392967GA SA307392969GA SA307392972GA SA307392973GA SA307392974GA SA307392975GA SA307392976GA SA307392977GA SA307392980GA SA307392982GA SA307392984GA SA307392986GA SA307392988GA SA307392989GA SA307392991GA SA307392994GA SA307392997GA SA307392998GA SA307393001GA SA307393003GA SA307393004GA SA307393005GA SA307393007GA SA307393009GA SA307393010GA SA307393011GA SA307393012GA SA307393013GA SA307393014GA SA307393016GA SA307393018GA SA307393019GA SA307393020GA SA307393021GA SA307393023GA SA307393024GA SA307393025GA SA307393026GA SA307392987GA SA307231425GA SA307231440GA SA307231425GA SA307231440GA SA307302012GA SA307241697GA SA307251759GA SA307261811GA SA307261813GA SA307261839GA SA307261841GA SA307261845GA SA307291862GA SA307291863GA SA307291864GA SA307291872GA SA307291900GA SA307291940GA SA307301993GA SA307301994GA SA307301998GA SA307302000GA SA307382518GA SA307382545GA SA307392971GA SA307392978GA SA307393000GA SA307221014GA SA307231191GA SA307231267GA SA307231368GA SA307261837GA SA307261843GA SA307261844GA SA307291950GA SA307302018GA SA307302022GA SA307210818GA SA307221114GA SA307231363GA SA307231372GA SA307231373GA SA307231386GA SA307231403GA SA307231408GA SA307231422GA SA307231436GA SA307231442GA SA307231471GA SA307231485GA SA307231492GA SA307210818GA SA307221114GA SA307231363GA SA307231372GA SA307231373GA SA307231386GA SA307231403GA SA307231408GA SA307231422GA SA307231436GA SA307231442GA SA307231471GA SA307231485GA SA307231492GA SA307392748GA SA307392822GA SA307392826GA SA307392830GA SA307392855GA SA307392857GA SA307392950GA SA307392960GA SA307392961GA SA307392962GA SA307392964GA SA307392979GA SA307393031GA SA307393032GA SA307403106GA SA307403108GA SA307221153GA SA307221165GA SA307221182GA SA307291861GA SA307291869GA SA307291871GA SA307291874GA SA307291875GA SA307291891GA SA307291893GA SA307291894GA SA307291895GA SA307291903GA SA307291904GA SA307302039GA SA307322179GA SA307332228GA SA307362396GA SA307241586GA SA307231356GA SA307231389GA SA307231430GA SA307291901GA SA307302023GA SA307302043GA SA307322125GA SA307322126GA SA307322133GA SA307322176GA SA307332209GA SA307332210GA SA307332213GA SA307332215GA SA307332216GA SA307332221GA SA307332224GA SA307332227GA SA307332229GA SA307332241GA SA307332243GA SA307332245GA SA307332246GA SA307332247GA SA307332250GA SA307382558GA SA307392983GA SA307302021GA SA307332249GA SA307332252GA SA307332253GA SA307332258GA SA307332260GA SA307332264GA SA307362407GA SA307332261GA SA307342297GA SA307342301GA SA307342306GA SA307342307GA SA307342311GA SA307352367GA SA307352368GA SA307362383GA SA307362404GA SA307362406GA SA307392996GA SA307403104GA SA307392996GA SA307403104GA SA307332251GA SA307332254GA SA307332256GA SA307332257GA SA307332262GA SA307342302GA SA307342303GA SA307342309GA SA307342316GA SA307342318GA SA307342319GA SA307342320GA SA307342325GA SA307342329GA SA307352351GA SA307342312GA SA307342317GA SA307423170GA SA307403129GA SA307423156GA SA307423161GA SA307342328GA SA307352353GA SA307382568GA SA307382577GAMKE SA307382591GAMKE SA307382594GAMKE SA307382621GAMKE SA307382622GA SA307332225GA SA307332230GA SA307332244GA SA307342308GA SA307352355GA SA307352360GA SA307352374GA SA307362379GA SA307362395GA SA307362398GA SA307362399GA SA307362405GA SA307220978GA SA307342296GA SA307362378GA SA307382459GA SA307382463GA SA307382486GA SA307382488GA SA307382494GA SA307382504GA SA307382510GA SA307382516GA SA307382544GA SA307382551GA SA307382552GA SA307382563GA SA307382589GA SA307382595GA SA307382609GA SA307382623GA SA307382624GA SA307443241GA SA307443367GA SA307110544GA SA307342324GA SA307362375GA SA307382479GA SA307382484GA SA307382537GA SA307382549GA SA307231197GA SA307231464GA SA307231466GA SA307231468GA SA307180695GA SA307190765GA SA307220987GA SA307231197GA SA307231259GA SA307231417GA SA307231464GA SA307231466GA SA307231468GA SA307231472GA SA307231491GA SA307291898GA SA307291912GA SA307291936GA SA307302028GA SA307302035GA SA307322172GA SA307332223GA SA307332255GA SA307332259GA SA307332263GA SA307342298GA SA307342299GA SA307342300GA SA307342305GA SA307342321GA SA307342322GA SA307342323GA SA307352352GA SA307352354GA SA307352356GA SA307352359GA SA307352363GA SA307352364GA SA307352369GA SA307352370GA SA307352373GA SA307362377GA SA307362380GA SA307362381GA SA307362393GA SA307362394GA SA307362400GA SA307362415GA SA307362426GA SA307362429GA SA307382458GA SA307382469GA SA307382470GA SA307382482GA SA307382492GA SA307382502GA SA307382506GA SA307382534GA SA307382535GA SA307382550GA SA307382555GA SA307382566GA SA307210820GA SA307210858GA SA307231465GA SA307251761GA SA307332265GA SA307342330GA SA307352357GA SA307362387GA SA307362403GA SA307362414GA SA307362421GA SA307362423GA SA307362430GA SA307362440GA SA307382530GA SA307382542GA SA307382547GA SA307080525GA SA307241532GA SA307392970GA SA307403048GA SA307403057GA SA307403074GA SA307403075GA SA307403079GA SA307403082GA SA307403095GA SA307403103GA SA307403115GA SA307403116GA SA307403120GA SA307403136GA SA307403139GA SA307403140GA SA307413100GA SA307413112GA SA307413117GA SA307413123GA SA307413124GA SA307221117GA SA307362397GA SA307362401GA SA307362420GA SA307362424GA SA307362431GA SA307362434GA SA307362438GA SA307362441GA SA307382444GA SA307382447GA SA307382449GA SA307382450GA SA307382457GA SA307382460GA SA307382465GA SA307382466GA SA307382467GA SA307382472GA SA307382474GA SA307382475GA SA307382480GA SA307382481GA SA307382483GA SA307382485GA SA307382489GA SA307382491GA SA307382497GA SA307382498GA SA307382500GA SA307382501GA SA307382508GA SA307382509GA SA307382513GA SA307382514GA SA307382515GA SA307382517GA SA307382520GA SA307382521GA SA307382523GA SA307382524GA SA307382525GA SA307382529GA SA307382543GA SA307382546GA SA307382553GA SA307382554GA SA307382556GA SA307382561GA SA307382565GA SA307382592GA SA307382493GA SA307382511GA SA307443388GA SA307382531GA SA307382574GA SA307382579GA SA307382601GA SA307382612GA SA307392766GA SA307392847GA SA307393022GA SA307403040GA SA307403041GA SA307403042GA SA307403044GA SA307403045GA SA307403047GA SA307403049GA SA307403051GA SA307403052GA SA307403053GA SA307403055GA SA307403058GA SA307403059GA SA307403062GA SA307403063GA SA307403065GA SA307403067GA SA307403068GA SA307403071GA SA307403072GA SA307403073GA SA307403124GA SA307403127GA SA307403128GA SA307403131GA SA307403135GA SA307403137GA SA307403138GA SA307403141GA SA307403143GA SA307403146GA SA307403149GA SA307403150GA SA307403152GA SA307413101GA SA307413102GA SA307413104GA SA307413105GA SA307413108GA SA307413111GA SA307413113GA SA307413114GA SA307413115GA SA307413116GA SA307413118GA SA307413119GA SA307413120GA SA307413122GA SA307423126GA SA307423127GA SA307423128GA SA307423129GA SA307423130GA SA307423131GA SA307423133GA SA307423134GA SA307423135GA SA307423136GA SA307423137GA SA307423138GA SA307423140GA SA307423141GA SA307423142GA SA307423144GA SA307423145GA SA307423146GA SA307423147GA SA307423148GA SA307423149GA SA307423150GA SA307423152GA SA307423153GA SA307423154GA SA307423157GA SA307423162GA SA307423165GA SA307423167GA SA307423168GA SA307423171GA SA307423175GA SA307423176GA SA307423178GA SA307423181GA SA307443185GAMKE SA307443186GA SA307443187GA SA307443190GA SA307443191GA SA307443192GAMKE SA307443193GA SA307443194GA SA307443196GA SA307443197GA SA307443198GAMKE SA307443199GA SA307443200GA SA307443201GA SA307443203GA SA307443204GA SA307443205GA SA307443209GA SA307443210GA SA307443212GA SA307443214GA SA307443215GA SA307443216GAMKE SA307443217GA SA307443218GA SA307443221GA SA307443222GA SA307443223GA SA307443224GA SA307443225GA SA307443235GA SA307443239GA SA307443246GA SA307443250GA SA307443253GA SA307443270GA SA307443305GA SA307443307GA SA307443309GA SA307443310GA SA307443313GA SA307443315GA SA307443316GA SA307443318GA SA307443320GA SA307443324GA SA307443326GA SA307443327GA SA307443330GA SA307443335GAMKE SA307443340GA SA307443342GA SA307443343GAMKE SA307443344GA SA307443346GA SA307443347GA SA307443355GA SA307453391GA SA307453411GA SA307453414GA SA307453418GA SA307453529GA SA307453542GA SA307453551GA SA307453558GA SA307453564GA SA307453604GA SA307453606GA SA307342313GA SA307362391GA SA307362392GA SA307362408GA SA307362412GA SA307362413GA SA307362419GA SA307362439GA SA307382448GA SA307382451GA SA307382453GA SA307382454GA SA307382461GA SA307382464GA SA307382468GA SA307382471GA SA307382478GA SA307382490GA SA307382495GA SA307382499GA SA307382505GA SA307382507GA SA307382522GA SA307382526GA SA307382528GA SA307382532GA SA307382536GA SA307382538GA SA307382539GA SA307382540GA SA307382557GA SA307382559GA SA307382560GA SA307382564GA SA307382569GA SA307382587GA SA307382600GA SA307382610GA SA307443248GA SA307443273GA SA307443308GA SA307443314GA SA307443341GA SA307443385GA SA307443387GA SA307453413GA SA307453549GA SA307453589GA SA307443314GA SA307443208GA SA307443245GA SA307382575GA SA307382581GA SA307382582GA SA307382584GA SA307382586GA SA307382593GA SA307382597GA SA307382613GA SA307382618GA SA307382625GA SA307382626GA SA307403043GA SA307403050GA SA307403060GA SA307403061GA SA307403087GA SA307403101GA SA307403110GA SA307403123GA SA307403134GA SA307413103GA SA307413106GA SA307443230GA SA307443232GA SA307443236GA SA307443240GA SA307443244GA SA307443252GA SA307443260GA SA307443269GA SA307443359GA SA307443362GA SA307443383GA SA307453392GA SA307453401GA SA307453404GA SA307453405GA SA307453407GA SA307453423GA SA307453433GA SA307443358GA SA307453409GA SA307453416GA SA307453437GA SA307453596GA SA307453603GA SA307443206GA SA307443207GA SA307443211GA SA307443328GA SA307443331GA SA307443325GA SA307443332GA SA307443349GA SA307443353GA SA307453402GA SA307453421GA SA307453480GA SA307453486GA SA307453526GA SA307453531GA SA307453532GA SA307453533GA SA307453534GA SA307453537GA SA307453538GA SA307453543GA SA307453562GA SA307453563GA SA307453565GA SA307453568GA SA307443227GA SA307443231GA SA307443251GA SA307443360GA SA307443375GA SA307443381GA SA307453390GA SA307453399GA SA307453424GA SA307453425GA SA307453427GA SA307453429GA SA307453435GA SA307453556GA SA307453586GA SA307453592GA SA307342314GA SA307231478GA SA307231478GA SA307302015GA SA307332206GA SA307342315GA SA307362384GA SA307362402GA SA307362417GA SA307362435GA SA307362442GA SA307382446GA SA307382473GA SA307382477GA SA307382541GA SA307382567GA SA307382573GA SA307382583GA SA307382611GA SA307382615GA SA307382616GA SA307443228GA SA307443254GA SA307453593GA SA307453594GA SA307382476GA SA307443262GA SA307443361GA SA307443371GA SA307443386GA SA307453400GA SA307453417GA SA307453420GA SA307453428GA SA307453430GA SA307453438GA SA307453555GA SA307453557GA SA307453588GA SA307453597GA SA307453599GA SA307453600GA SA307453602GA SA307453613GA SA307453614GA SA307453616GA SA307362376GA SA307382599GA SA307362385GA SA307453572GA SA307453579GA SA307453585GA SA307453601GA SA307453608GA SA307453610GA SA307483661GA SA307483666GA SA307483667GA SA307483669GA SA307483671GA SA307483673GA SA307483676GA SA307483680GA SA307483681GA SA307483682GA SA307483684GA SA307483685GA SA307483686GA SA307483691GA SA307483696GA SA307483713GA SA307483726GA SA307483730GA SA307483732GA SA307453584GA SA307453587GA SA307453609GA SA307483689GA SA307483694GA SA307483703GA SA307483712GA SA307483715GA SA307483720GA SA307483722GA SA307483739GA SA307483743GA SA307483744GA SA307483752GA SA307483761GA SA307251801GA SA307382527GA SA307382576GA SA307382578GA SA307423151GA SA307443229GA SA307453442GA SA307453463GA SA307453541GA SA307453569GA SA307453570GA SA307453571GA SA307453573GA SA307453574GA SA307453575GA SA307453578GA SA307453582GA SA307453583GA SA307453611GA SA307453615GA SA307453617GA SA307453618GA SA307483659GA SA307483660GA SA307483662GA SA307483668GA SA307483670GA SA307483672GA SA307483677GA SA307483678GA SA307483679GA SA307483683GA SA307483687GA SA307483688GA SA307483690GA SA307483692GA SA307483697GA SA307483698GA SA307483699GA SA307483700GA SA307483701GA SA307483702GA SA307483704GA SA307483705GA SA307483706GA SA307483707GA SA307483711GA SA307483714GA SA307483721GA SA307483728GA SA307483729GA SA307483733GA SA307483734GA SA307483736GA SA307483737GA SA307483738GA SA307483740GA SA307483741GA SA307483742GA SA307483745GA SA307483746GA SA307483747GA SA307483749GA SA307483750GA SA307483751GA SA307483753GA SA307483754GA SA307483756GA SA307483757GA SA307483758GA SA307483759GA SA307483760GA SA307483762GA SA307483765GA SA307483766GA SA307483767GA SA307483771GA SA307483775GA SA307483777GA SA307483782GA SA307483783GA SA307483789GA SA307483794GA SA307483795GA SA307483796GA SA307483797GA SA307483798GA SA307483800GA SA307483801GA SA307483803GA SA307483804GA SA307483805GA SA307483806GA SA307483809GA SA307483810GA SA307483812GA SA307483814GA SA307483816GA SA307483817GA SA307483818GA SA307483819GA SA307483821GA SA307483823GA SA307483825GA SA307483826GA SA307483832GA SA307483834GA SA307483841GA SA307483843GA SA307504090GA SA307504092GA SA307504093GA SA307504094GA SA307504136GA SA307504138GA SA307504143GA SA307504144GA SA307504145GA SA307504151GA SA307504152GA SA307504153GA SA307504186GA SA307504217GA SA307504219GA SA307382620GA SA307443226GA SA307443233GA SA307443234GA SA307443237GA SA307443255GA SA307443256GA SA307443258GA SA307443259GA SA307443261GA SA307443263GA SA307443264GA SA307443265GA SA307443266GA SA307443268GA SA307443271GA SA307443272GA SA307443366GA SA307443368GA SA307443369GA SA307443372GA SA307443373GA SA307443376GA SA307443377GA SA307443379GA SA307443380GA SA307443382GA SA307443384GA SA307453398GA SA307453415GA SA307453419GA SA307453426GA SA307453436GA SA307453524GA SA307453580GA SA307453581GA SA307453612GA SA307483663GA SA307483664GA SA307483674GA SA307483675GA SA307483693GA SA307483695GA SA307483709GA SA307483710GA SA307483716GA SA307483717GA SA307483719GA SA307483723GA SA307483724GA SA307483725GA SA307483727GA SA307483731GA SA307483735GA SA307483755GA SA307483763GA SA307483764GA SA307483769GA SA307483770GA SA307483772GA SA307483773GA SA307483776GA SA307483778GA SA307483780GA SA307483784GA SA307483788GA SA307483790GA SA307483792GA SA307483808GA SA307483813GA SA307483822GA SA307483824GA SA307483827GA SA307483828GA SA307483829GA SA307483835GA SA307483838GA SA307483840GA SA307493845GA SA307493846GA SA307493847GA SA307493848GA SA307493849GA SA307493850GA SA307493851GA SA307493854GA SA307493863GA SA307493933GA SA307493936GA SA307493943GA SA307493947GA SA307493949GA SA307493951GA SA307493992GA SA307493994GA SA307493995GA SA307493996GA SA307493997GA SA307493999GA SA307494000GA SA307494001GA SA307494002GA SA307494003GA SA307494004GA SA307494005GA SA307494006GA SA307494007GAMKE SA307494008GA SA307494009GA SA307494010GA SA307494011GA SA307494012GA SA307494013GA SA307494014GA SA307494016GA SA307494017GA SA307494019GA SA307494020GA SA307494021GA SA307504067GA SA307504068GA SA307504069GA SA307504070GA SA307504071GA SA307504072GA SA307504073GA SA307504074GA SA307504075GA SA307504076GA SA307504077GA SA307504078GA SA307504079GA SA307504080GA SA307504081GA SA307504082GA SA307504083GA SA307504084GA SA307504085GA SA307504086GA SA307504087GA SA307504088GA SA307504089GA SA307504091GA SA307504095GA SA307504096GA SA307504104GA SA307504105GA SA307504106GA SA307504107GA SA307504108GA SA307504109GA SA307504111GA SA307504112GA SA307504114GA SA307504119GA SA307504125GA SA307504129GA SA307504130GA SA307504131GA SA307504132GA SA307504137GA SA307504139GA SA307504140GA SA307504141GA SA307504142GA SA307504148GA SA307504149GA SA307504150GA SA307504154GA SA307504155GA SA307504156GA SA307504157GA SA307504159GA SA307504160GA SA307504161GA SA307504162GA SA307504163GA SA307504164GA SA307504166GA SA307504168GA SA307504170GA SA307504171GA SA307504173GA SA307504175GA SA307504177GA SA307504178GA SA307504181GA SA307504182GA SA307504183GA SA307504184GA SA307504185GA SA307504193GA SA307504213GA SA307504215GA SA307504216GA SA307504221GA SA307504224GA SA307221141GA SA307291975GA SA307291984GA SA307382657GA SA307392670GA SA307392674GA SA307392689GA SA307392725GA SA307403112GA SA307443312GA SA307443333GA SA307443345GA SA307453445GA SA307453448GA SA307453457GA SA307453461GA SA307453465GA SA307453489GA SA307453498GA SA307453521GA SA307453544GA SA307453546GA SA307453566GA SA307463650GA SA307463654GA SA307483830GA SA307483831GA SA307483837GA SA307493852GA SA307493853GA SA307493855GA SA307493856GA SA307493857GA SA307493858GA SA307493859GA SA307493860GA SA307493861GA SA307493864GA SA307493866GA SA307493867GA SA307493868GA SA307493869GA SA307493870GA SA307493872GA SA307493873GA SA307493874GA SA307493875GA SA307493876GA SA307493877GA SA307493879GA SA307493881GA SA307493888GA SA307493890GA SA307493893GA SA307493894GA SA307493895GA SA307493896GA SA307493897GA SA307493904GA SA307493915GA SA307493920GA SA307493924GA SA307493934GA SA307493935GA SA307493937GA SA307493940GA SA307493941GA SA307493942GA SA307493944GA SA307493948GA SA307443348GA SA307453441GA SA307453447GA SA307453449GA SA307453454GA SA307453459GA SA307453467GA SA307453470GA SA307453481GA SA307453482GA SA307453487GA SA307453490GA SA307453494GA SA307453495GA SA307453519GA SA307453523GA SA307453540GA SA307453545GA SA307453547GA SA307453548GA SA307453567GA SA307463649GA SA307463651GA SA307463652GA SA307463653GA SA307463655GA SA307463656GA SA307463657GA SA307463658GA SA307504192GA SA307514320GA SA307514326GA SA307504192GA SA307453439GA SA307453450GA SA307453452GA SA307453455GA SA307453460GA SA307453468GA SA307453471GA SA307453473GA SA307453478GA SA307453484GA SA307453496GA SA307453497GA SA307453559GA SA307493865GA SA307493882GA SA307493883GA SA307493884GA SA307493899GA SA307493900GA SA307493905GA SA307493910GA SA307493911GA SA307493913GA SA307493919GA SA307493922GA SA307504174GA SA307504199GA SA307514332GA SA307514336GA SA307514342GA SA307514352GA SA307413107GA SA307453440GA SA307453462GA SA307453476GA SA307453477GA SA307453479GA SA307453488GA SA307453491GA SA307453492GA SA307453493GA SA307453522GA SA307453527GA SA307453528GA SA307453535GA SA307453536GA SA307453561GA SA307493878GA SA307493892GA SA307493902GA SA307493903GA SA307493906GA SA307493916GA SA307493918GA SA307493921GA SA307493928GA SA307493958GA SA307493965GA SA307493966GA SA307493967GA SA307493969GA SA307493982GA SA307493983GA SA307493984GA SA307493988GA SA307493991GA SA307494022GA SA307494023GA SA307494024GA SA307494025GA SA307494026GA SA307494028GA SA307494029GA SA307494031GA SA307494033GA SA307494035GA SA307494038GA SA307494039GA SA307494040GA SA307494042GA SA307494043GA SA307494044GA SA307494050GA SA307494051GA SA307393035GA SA307443202GA SA307453446GA SA307453456GA SA307453485GA SA307453525GA SA307493871GA SA307493880GA SA307493885GA SA307493887GA SA307493891GA SA307493898GA SA307493901GA SA307493907GA SA307493908GA SA307493909GA SA307493912GA SA307493914GA SA307493917GA SA307493923GA SA307493925GA SA307493926GA SA307493927GA SA307493929GA SA307493930GA SA307493931GA SA307493952GA SA307493953GA SA307493954GA SA307493955GA SA307493956GA SA307493957GA SA307493959GA SA307493960GA SA307493961GA SA307493962GA SA307493963GA SA307493964GA SA307493970GA SA307493971GA SA307493972GA SA307493973GA SA307493974GA SA307493975GA SA307493976GA SA307493977GA SA307493980GA SA307493981GA SA307493985GA SA307493987GA SA307493989GA SA307493990GA SA307494027GA SA307494030GA SA307494032GA SA307494034GA SA307494037GA SA307494045GA SA307494047GA SA307494048GA SA307494049GA SA307504066GA SA307494015GA SA307494018GA SA307443370GA SA307504206GA SA307423143GA SA307423143GA SA307493945GA SA307493946GA SA307504147GA SA307251803GA SA307443363GA SA307504102GA SA307504118GA SA307504122GA SA307504126GA SA307504134GA SA307504135GA SA307504146GA SA307504176GA SA307504180GA SA307514334GA SA307514358GA SA307514414GA SA307514415GA SA307514423GA SA307514424GA SA307483768GA SA307493932GA SA307493938GA SA307493950GA SA307504110GA SA307504115GA SA307504117GA SA307504128GA SA307504133GA SA307504172GA SA307504207GA SA307504208GA SA307504223GA SA307514337GA SA307514366GA SA307514368GA SA307514366GA SA307514368GA SA307362382GA SA307443365GA SA307453598GA SA307483774GA SA307483811GA SA307493939GA SA307504165GA SA307504196GA SA307504197GA SA307504201GA SA307504218GA SA307504220GA SA307514330GA SA307514333GA SA307504120GA SA307504123GA SA307504127GA SA307504179GA SA307514348GA SA307514434GA SA307504233GA SA307514291GA SA307514319GA SA307514355GA SA307514356GA SA307514359GA SA307514363GA SA307514367GA SA307514370GA SA307514372GA SA307514384GA SA307514387GA SA307514388GA SA307514389GA SA307514418GA SA308044513GA SA308044517GA SA308044520GA SA308044523GA SA308044527GA SA307514391GA SA307514410GA SA308074554GA SA308044521GA SA308074610GA SA308084663GA SA308084736GA SA308084737GA SA308084738GA SA308084740GA SA308084742GA SA308084743GA SA308084749GA SA308084750GA SA308084751GA SA308084769GA SA308084776GA SA308084778GA SA307483836GA SA308074603GA SA308074606GA SA308084622GA SA308084753GA SA308084754GA SA308084761GA SA308084762GA SA308084766GA SA308084777GA SA308084779GA SA308084784GA SA307514308GA SA307514360GA SA307514361GA SA307514371GA SA307514376GA SA307514377GA SA307514378GA SA307514386GA SA308074538GA SA308074539GA SA308074547GA SA308074548GA SA308074549GA SA308074550GA SA308074552GA SA308074555GA SA308074556GA SA308074564GA SA308074577GA SA308074578GA SA308074580GA SA308074585GA SA308074587GA SA308074588GA SA308074589GA SA308074592GA SA308074597GA SA308074598GA SA308074604GA SA308084667GA SA308084670GA SA308084674GA SA308084679GA SA308084682GA SA308084684GA SA308084685GA SA308084690GA SA308084692GA SA308084693GA SA308084699GA SA308084701GA SA308084702GA SA308084706GA SA308084707GA SA308084710GA SA308084717GA SA308084786GA SA308084612GA SA308084619GA SA308084646GA SA308084652GA SA308084653GA SA308084658GA SA308084661GA SA308084662GA SA308084664GA SA308084712GA SA308084729GA SA308084756GA SA308084758GA SA308084759GA SA308094796GA SA308094797GA SA308094801GA SA308094811GA SA308094814GA SA308094815GA SA308094816GA SA308094819GA SA308094820GA SA308094878GA SA308074567GA SA308074591GA SA308074595GA SA308084671GA SA308084673GA SA308084677GA SA308084688GA SA308084689GA SA308084694GA SA308084695GA SA308084696GA SA308084697GA SA308084705GA SA308084708GA SA308084711GA SA308084718GA SA308084719GA SA308084724GA SA308084681GA SA308084715GA SA308084721GA SA308084720GA SA308084725GA SA308094843GA SA308094849GA SA308084615GA SA308084616GA SA308084620GA SA308084621GA SA308084623GA SA308084643GA SA308084654GA SA308084770GA SA308094800GA SA308094803GA SA308094807GA SA308094809GA SA308094813GA SA308094818GA SA308094867GA SA308094869GA SA308094871GA SA308094872GA SA308094875GA SA308094880GA SA308094925GA SA308094927GA SA308094929GA SA308094938GA SA308094940GA SA308094958GA SA308094960GA SA308084675GA SA308084680GA SA308084686GA SA308084704GA SA308084722GA SA308084787GA SA308084791GA SA308084792GA SA308084795GA SA308094836GA SA308094838GA SA308094839GA SA308094841GA SA308094863GA SA308094891GA SA308094894GA SA308094841GA SA307514357GA SA308074537GA SA308074540GA SA308074542GA SA308074545GA SA308074551GA SA308074553GA SA308074559GA SA308074562GA SA308074569GA SA308074570GA SA308074571GA SA308074575GA SA308074579GA SA308074593GA SA308074594GA SA308074596GA SA308084691GA SA308084788GA SA308084789GA SA308084790GA SA308084793GA SA308084794GA SA308094837GA SA308094842GA SA308094844GA SA308094847GA SA308094850GA SA308094851GA SA308094852GA SA308094853GA SA308094855GA SA308094858GA SA308094860GA SA308094861GA SA308094862GA SA308094864GA SA308094865GA SA307453451GA SA307453466GA SA308084735GA SA308094877GA SA308094943GA SA308094955GA SA308094959GA SA308094984GA SA308094989GA SA308094991GA SA308094999GA SA308095014GA SA308094934GA SA308094942GA SA308094949GA SA308094965GA SA308094966GA SA308094978GA SA308095000GA SA308095013GA SA308094979GA SA308095055GA SA307514364GA SA307514365GA SA307514373GA SA307514374GA SA307514375GA SA307514380GA SA307514383GA SA307514385GA SA308074561GA SA308074576GA SA308094968GA SA307514379GA SA307514381GA SA307514382GA SA308074541GA SA308074543GA SA308074557GA SA308074565GA SA308094884GA SA308105077GA SA308105081GA SA308105086GA SA308105088GA SA308074605GA SA308084703GA SA308094859GA SA308094915GA SA308095016GA SA308095017GA SA308095018GA SA308095019GA SA308095020GA SA308095025GA SA308095026GA SA308095028GA SA308095029GA SA308095030GA SA308095032GA SA308095033GA SA308095035GA SA308095038GA SA308095040GA SA308095041GA SA308095042GA SA308095043GA SA308095044GA SA308095054GA SA308095058GA SA308095063GA SA308095069GA SA308095070GA SA308105075GA SA308105076GA SA308105078GA SA308105080GA SA308095021GA SA308094954GA SA308095002GA SA307483786GA SA307504190GA SA307504198GA SA307504200GA SA307504203GA SA307504209GA SA307514325GA SA307514328GA SA307514329GA SA307514331GA SA307514340GA SA307483786GA SA307504190GA SA307504198GA SA307504200GA SA307504203GA SA307504209GA SA307514325GA SA307514328GA SA307514329GA SA307514331GA SA307514340GA SA308084774GA SA308084783GA SA308094944GA SA308095015GA SA308095074GA SA308105173GA SA308105156GA SA308105193GA SA308105185GA SA308105196GA SA308105198GA SA308105203GA SA308105165GA SA308105079GA SA308105143GA SA308105145GA SA308105138GA SA308094903GA SA308105119GA SA308105108GA SA308105109GA SA308105116GA SA308105120GA SA308105137GA SA308105110GA SA308105130GA SA308105134GA SA308095022GA SA308094921GA SA308095023GA SA308095027GA SA308095031GA SA308095034GA SA308095036GA SA308095037GA SA308095039GA SA308095060GA SA308095071GA SA308095072GA SA308095073GA SA308105128GA SA308105131GA SA308105132GA SA308105133GA SA308105136GA SA308105141GA SA308094921GA SA308105128GA SA308105131GA SA308105132GA SA308105133GA SA308105136GA SA308105091GA SA308105094GA SA308105095GA SA308105097GA SA308105101GA SA308105107GA SA308105111GA SA308105115GA SA308105139GA SA308105144GA SA308105094GA SA308105097GA SA308105107GA SA308105144GA SA308105091GA SA308105095GA SA308105115GA SA308105139GA SA308105101GA SA308105140GA SA308105082GA SA308105102GA SA308115265GA SA308115276GA SA308125312GA SA308125325GA SA308125326GA SA308115296GA SA308105100GA SA308105127GA SA308115288GA SA308115289GA SA308115292GA SA308115297GA SA308115300GA SA308125375GA SA308125376GA SA308125379GA SA308125380GA SA308125382GA SA308125384GA SA308125386GA SA308125394GA SA308135518GA SA308135527GA SA308115306GA SA308094832GA SA308105191GA SA308115294GA SA308115303GA SA308125344GA SA308135521GA SA308135541GA SA308145655GA SA308145657GA SA308145671GA SA308145684GA SA308145688GA SA308145692GA SA308145693GA SA308145694GA SA308145695GA SA308145696GA SA308145697GA SA308094857GA SA308115287GA SA308115291GA SA308115298GA SA308115301GA SA308125335GA SA308125355GA SA308125374GA SA308135519GA SA308145668GA SA308145690GA SA308155875GA SA308155876GA SA308155878GA SA308155970GA SA308105121GA SA308105123GA SA308105125GA SA308105126GA SA308105220GA SA308125393GA SA308125413GA SA308125416GA SA308125418GA SA308125424GA SA308125425GA SA308125426GA SA308125427GA SA308125428GA SA308135482GA SA308135488GA SA308135489GA SA308125390GA SA308125409GA SA308125412GA SA308135466GA SA308135486GA SA308115305GA SA308125327GA SA308135444GA SA308135446GA SA308135454GA SA308135468GA SA308135494GA SA308135533GA SA308145686GA SA308145687GA SA308145818GA SA308155867GA SA308155869GA SA308155969GA SA308155975GA SA308155979GA SA308166005GA SA308166007GA SA308166008GA SA308166010GA SA308166013GA SA308166015GA SA308166017GA SA308166018GA SA308166019GA SA308166020GA SA308166022GA SA308166023GA SA308166024GA SA308166025GA SA308166026GA SA308166028GA SA308166029GA SA308166030GA SA308166031GA SA308166033GA SA308166034GA SA308166040GA SA308166042GA SA308166073GA SA308166081GA SA308166082GA SA308166130GA SA308166165GA SA308186310GA SA307514321GA SA308125341GA SA308125348GA SA308125350GA SA308125354GA SA308125356GA SA308125360GA SA308125364GA SA308125369GA SA308074583GA SA308084713GA SA308094845GA SA308105092GA SA308105093GA SA308105098GA SA308105103GA SA308105122GA SA308105124GA SA308074583GA SA308084713GA SA308094845GA SA308105092GA SA308105093GA SA308105098GA SA308105103GA SA308105122GA SA308105124GA SA308125395GA SA308125396GA SA308125397GA SA308125399GA SA308125403GA SA308125404GA SA308125405GA SA308125406GA SA308125408GA SA308125419GA SA308125423GA SA308135431GA SA308135432GA SA308135433GA SA308135436GA SA308135439GA SA308135440GA SA308135442GA SA308135445GA SA308135447GA SA308135449GA SA308135450GA SA308135451GA SA308135452GA SA308135456GA SA308135457GA SA308135458GA SA308135459GA SA308135460GA SA308135462GA SA308135463GA SA308135464GA SA308135465GA SA308135467GA SA308135472GA SA308135475GA SA308135481GA SA308135484GA SA308135485GA SA308135492GA SA308135493GA SA308135495GA SA308135496GA SA308135497GA SA308135498GA SA308135499GA SA308135500GA SA308135501GA SA308135502GA SA308135503GA SA308135507GA SA308135508GA SA308135513GA SA308135547GA SA308135551GA SA308145675GA SA308145685GA SA308155845GA SA308155847GA SA308155850GA SA308155851GA SA308155858GA SA308155864GA SA308155956GA SA308155968GA SA308155973GA SA308155974GA SA308155980GA SA308155981GA SA308166037GA SA308166041GA SA308166043GA SA308166050GA SA308166052GA SA308166053GA SA308166054GA SA308166057GA SA308166058GA SA308166059GA SA308166060GA SA308166061GA SA308166062GA SA308166063GA SA308166068GA SA308166069GA SA308166072GA SA308166075GA SA308166077GA SA308166079GA SA308166080GA SA308166086GA SA308166113GA SA308135504GA SA307403070GA SA308115290GA SA308125400GA SA308125402GA SA308125410GA SA308125414GA SA308125420GA SA308125421GA SA308125429GA SA308135430GA SA308135434GA SA308135435GA SA308135437GA SA308135438GA SA308135441GA SA308135453GA SA308135455GA SA308135471GA SA308135474GA SA308135476GA SA308135477GA SA308135479GA SA308135480GA SA308135490GA SA308135491GA SA308135505GA SA308135510GA SA308135511GA SA308135514GA SA308135546GA SA308135548GA SA308135549GA SA308135550GA SA308135552GA SA308135553GA SA308135555GA SA308135556GA SA308135557GA SA308135559GA SA308135561GA SA308135566GA SA308135567GA SA308135568GA SA308135569GA SA308135571GA SA308135572GA SA308135573GA SA308135576GA SA308135577GA SA308135579GA SA308135581GA SA308135582GA SA308135584GA SA308135587GA SA308135588GA SA308135592GA SA308135593GA SA308135594GA SA308135595GA SA308135597GA SA308135598GA SA308135599GA SA308135601GA SA308135602GA SA308135603GA SA308135605GA SA308135606GA SA308135616GA SA308135617GA SA308135625GA SA308135630GA SA308166045GA SA308206594GA SA308196568GA SA308155866GA SA308155900GA SA308166090GA SA308166095GA SA308166111GA SA308166112GA SA308166115GA SA308166119GA SA308196585GA SA308135443GA SA308135448GA SA308135469GA SA308135506GA SA308135565GA SA308135586GA SA308135590GA SA308135611GA SA308135613GA SA308135614GA SA308135621GA SA308135629GA SA308135633GA SA308135591GA SA308135618GA SA308145699GA SA308155840GA SA308155863GA SA308155865GA SA308155872GA SA308155931GA SA308166091GA SA308166096GA SA308166098GA SA308166100GA SA308166103GA SA308166108GA SA308166118GA SA308135512GA SA308135563GA SA308135596GA SA308135615GA SA308145729GA SA308135509GA SA308135608GA SA308135609GA SA308135622GA SA308135627GA SA308155886GA SA308155904GA SA308155908GA SA308155916GA SA308155917GA SA308155919GA SA308155929GA SA308155932GA SA308155938GA SA308155955GA SA308155959GA SA308155960GA SA308155965GA SA308155966GA SA308165991GA SA308165993GA SA308165996GA SA308166000GA SA308166001GA SA308166002GA SA308166004GA SA308166009GA SA308166011GA SA308166085GA SA308166088GA SA308166101GA SA308094808GA SA308155843GA SA308155846GA SA308155853GA SA308155854GA SA308155868GA SA308155874GA SA308166066GA SA308166067GA SA308166074GA SA308166083GA SA308166107GA SA308125401GA SA308145661GA SA308145754GA SA308145786GA SA308196486GA SA308196491GA SA308196492GA SA308196511GA SA308145781GA SA308155961GA SA308155964GA SA308166114GA SA308135478GA SA308135545GA SA308135619GA SA308145698GA SA308145700GA SA308145701GA SA308145704GA SA308145708GA SA308145717GA SA308145723GA SA308145726GA SA308145727GA SA308145731GA SA308145734GA SA308145738GA SA308145743GA SA308145744GA SA308145748GA SA308145760GA SA308145763GA SA308145766GA SA308145768GA SA308145770GA SA308145774GA SA308145775GA SA308145787GA SA308145803GA SA308145831GA SA308145835GA SA308155983GA SA308155984GA SA308155989GA SA308186382GA SA308166123GA SA308186376GA SA308186386GA SA308186395GA SA308206593GA SA308155918GA SA308166120GA SA308166122GA SA308186352GA SA308186353GA SA308186354GA SA308186357GA SA308186359GA SA308186362GA SA308186364GA SA308186368GA SA308186370GA SA308186371GA SA308186373GA SA308186378GA SA308186383GA SA308186387GA SA308186398GA SA308186403GA SA308186406GA SA308186407GA SA308125377GA SA308135483GA SA308135639GA SA308145705GA SA308145707GA SA308145724GA SA308145741GA SA308145745GA SA308145746GA SA308145753GA SA308145761GA SA308145817GA SA308145827GA SA308155860GA SA308155873GA SA308155882GA SA308155885GA SA308155888GA SA308155892GA SA308155897GA SA308155898GA SA308155899GA SA308155903GA SA308155906GA SA308155909GA SA308155914GA SA308155915GA SA308155920GA SA308155925GA SA308155928GA SA308155933GA SA308155936GA SA308155937GA SA308155939GA SA308155942GA SA308155944GA SA308155946GA SA308155947GA SA308155949GA SA308155950GA SA308155953GA SA308155967GA SA308155986GA SA308155990GA SA308165994GA SA308165997GA SA308165999GA SA308166006GA SA308166012GA SA308166014GA SA308166097GA SA308166124GA SA308186365GA SA308186369GA SA308186402GA SA308186416GA SA308186417GA SA308186431GA SA308186432GA SA308155954GA SA308155957GA SA308166121GA SA308186350GA SA308186360GA SA308186366GA SA308186374GA SA308186375GA SA308186377GA SA308186380GA SA308186381GA SA308186385GA SA308186388GA SA308186390GA SA308186391GA SA308186396GA SA308186397GA SA308186401GA SA308186404GA SA308186405GA SA308186410GA SA308186444GA SA308186445GA SA308155893GA SA308166102GA SA308186372GA SA308186422GA SA308186429GA SA308186433GA SA308196454GA SA308196455GA SA308196457GA SA308196459GA SA308196463GA SA308196464GA SA308196468GA SA308196471GA SA308196474GA SA308145815GA SA308166192GA SA308186321GA SA308105135GA SA308135648GA SA308135649GA SA308145703GA SA308145714GA SA308145722GA SA308145725GA SA308145730GA SA308145732GA SA308145739GA SA308145742GA SA308145747GA SA308145750GA SA308145755GA SA308145756GA SA308145758GA SA308145759GA SA308145762GA SA308145764GA SA308145765GA SA308145771GA SA308145773GA SA308145777GA SA308145778GA SA308145779GA SA308145805GA SA308145808GA SA308145812GA SA308145813GA SA308145814GA SA308145819GA SA308145822GA SA308145824GA SA308145826GA SA308145829GA SA308145832GA SA308145834GA SA308145836GA SA308145837GA SA308155857GA SA308155985GA SA308166188GA SA308166200GA SA308166202GA SA308166205GA SA308166206GA SA308176225GA SA308176226GA SA308176228GA SA308176230GA SA308176231GA SA308176233GA SA308176241GA SA308176242GA SA308176294GA SA308186312GA SA308186313GA SA308186315GA SA308186316GA SA308186322GA SA308186327GA SA308186329GA SA308186330GA SA308145757GA SA308145830GA SA308145838GA SA308155945GA SA308166138GA SA308166139GA SA308166143GA SA308166159GA SA308166177GA SA308166216GA SA308176219GA SA308176221GA SA308176222GA SA308176229GA SA308176234GA SA308176235GA SA308176249GA SA308176255GA SA308176259GA SA308176271GA SA308176275GA SA308176287GA SA308176297GA SA308176301GA SA308186314GA SA308186317GA SA308186318GA SA308186324GA SA308186325GA SA308186326GA SA308186328GA SA308115295GA SA308125349GA SA308145740GA SA308145749GA SA308145772GA SA308145806GA SA308145807GA SA308145816GA SA308145820GA SA308145823GA SA308145828GA SA308145833GA SA308155921GA SA308155923GA SA308155987GA SA308155988GA SA308166126GA SA308166128GA SA308166129GA SA308166131GA SA308166132GA SA308166133GA SA308166134GA SA308166135GA SA308166137GA SA308166141GA SA308166142GA SA308166145GA SA308166149GA SA308166153GA SA308166160GA SA308166163GA SA308166166GA SA308166167GA SA308166169GA SA308166173GA SA308166174GA SA308166176GA SA308166179GA SA308166180GA SA308166184GA SA308166187GA SA308166189GA SA308166194GA SA308166195GA SA308166197GA SA308166207GA SA308166208GA SA308166209GA SA308166210GA SA308176218GA SA308176237GA SA308176238GA SA308176239GA SA308176243GA SA308176244GA SA308176246GA SA308176247GA SA308176253GA SA308176254GA SA308176260GA SA308176261GA SA308176262GA SA308176263GA SA308176264GA SA308176265GA SA308176266GA SA308176267GA SA308176268GA SA308176269GA SA308176272GA SA308176273GA SA308176274GA SA308176276GA SA308176279GA SA308176280GA SA308176282GA SA308176283GA SA308176286GA SA308176289GA SA308176290GA SA308176295GA SA308176296GA SA308176298GA SA308176299GA SA308176300GA SA308176302GA SA308176303GA SA308176304GA SA308176305GA SA308176306GA SA308186392GA SA308186394GA SA308186424GA SA308186426GA SA308186427GA SA308186437GA SA308196446GA SA308196447GA SA308196448GA SA308196449GA SA308196450GA SA308196451GA SA308196452GA SA308196453GA SA308196456GA SA308196466GA SA308196467GA SA308196472GA SA308196473GA SA308196475GA SA308196477GA SA308196478GA SA308196480GA SA308196485GA SA308196488GA SA308196493GA SA308196494GA SA308196495GA SA308196497GA SA308196498GA SA308196499GA SA308196500GA SA308196501GA SA308196502GA SA308196503GA SA308196505GA SA308196506GA SA308196510GA SA308196512GA SA308196513GA SA308196514GA SA308196515GA SA308196516GA SA308196517GA SA308196518GA SA308196520GA SA308196521GA SA308196522GA SA308196523GA SA308196526GA SA308196527GA SA308196528GA SA308196529GA SA308196530GA SA308196531GA SA308196532GA SA308196533GA SA308196535GA SA308196536GA SA308196537GA SA308196542GA SA308196545GA SA308196546GA SA308196547GA SA308196549GA SA308196550GA SA308196551GA SA308196553GA SA308196555GA SA308196557GA SA308196559GA SA308196561GA SA308196562GA SA308196563GA SA308196564GA SA308196581GA SA308206592GA SA308206597GA SA308206598GA SA308206605GA SA308206694GA SA308206695GA SA308206706GA SA308166144GA SA308196538GA SA308166136GA SA308166147GA SA308166148GA SA308166175GA SA308166182GA SA308166190GA SA308166201GA SA308166211GA SA308166213GA SA308176223GA SA308176227GA SA308176285GA SA308176308GA SA308186319GA SA308186320GA SA308186323GA SA308196584GA SA308145767GA SA308145810GA SA308166127GA SA308166161GA SA308166172GA SA308166186GA SA308176277GA SA308176278GA SA308176307GA SA308266784GA SA308166146GA SA308166150GA SA308166151GA SA308166154GA SA308166155GA SA308166156GA SA308166157GA SA308166168GA SA308166178GA SA308166183GA SA308166196GA SA308166215GA SA308176256GA SA308145821GA SA308166152GA SA308166158GA SA308166162GA SA308166164GA SA308166170GA SA308166171GA SA308166181GA SA308166185GA SA308166191GA SA308166193GA SA308166199GA SA308166203GA SA308166204GA SA308166212GA SA308166214GA SA308176224GA SA308176232GA SA308176245GA SA308176248GA SA308176250GA SA308176251GA SA308176257GA SA308176258GA SA308176270GA SA308176281GA SA308176284GA SA308176288GA SA308176292GA SA308186331GA SA308186333GA SA308186335GA SA308206588GA SA308206595GA SA308206604GA SA308266829GA SA308266857GA SA308176236GA SA308176240GA SA308186309GA SA308186311GA SA308266811GA SA308266813GA SA308266814GA SA308266817GA SA308266818GA SA308266819GA SA308266820GA SA308266821GA SA308266824GA SA308266825GA SA308266826GA SA308266830GA SA308266835GA SA308266837GA SA308266839GA SA308266842GA SA308266845GA SA308266846GA SA308266856GA SA308266875GA SA308266877GA SA308266884GA SA308286920GA SA308286929GA SA308135600GA SA308145712GA SA308145797GA SA308266815GA SA308266816GA SA308266822GA SA308266831GA SA308266832GA SA308266833GA SA308266834GA SA308266836GA SA308266838GA SA308266840GA SA308266841GA SA308266843GA SA308266844GA SA308266847GA SA308266848GA SA308266849GA SA308266850GA SA308266851GA SA308266853GA SA308266854GA SA308266855GA SA308266859GA SA308266860GA SA308266862GA SA308266863GA SA308266864GA SA308266865GA SA308266866GA SA308266867GA SA308266868GA SA308266869GA SA308266870GA SA308266871GA SA308266872GA SA308266873GA SA308266874GA SA308266876GA SA308266878GA SA308266882GA SA308266885GA SA308266886GA SA308266887GA SA308266890GA SA308266893GA SA308266894GA SA308266895GA SA308266896GA SA308266898GA SA308266899GA SA308266900GA SA308266901GA SA308266902GA SA308286903GA SA308286905GA SA308286906GA SA308286907GA SA308286909GA SA308286911GA SA308286913GA SA308286915GA SA308286916GA SA308286919GA SA308286921GA SA308286923GA SA308286924GA SA308286925GA SA308286928GA SA308286931GA SA308286940GA SA308286950GA SA308286954GA SA308286955GA SA308286958GA SA308286967GA SA308286968GA SA308286971GA SA308286973GA SA308286974GA SA308286978GA SA308287000GA SA308287001GA SA308287011GA SA308287012GA SA308287013GA SA308186337GA SA308196560GA SA308145780GA SA308166092GA SA308166140GA SA308206596GA SA308286936GA SA308286942GA SA308135652GA SA308186332GA SA308186336GA SA308186338GA SA308186340GA SA308186341GA SA308186342GA SA308206615GA SA308206629GA SA308266798GA SA308266888GA SA308286963GA SA308286964GA SA308286965GA SA308287003GA SA308287009GA SA308166110GA SA308286951GA SA308286953GA SA308286961GA SA308286984GA SA308286985GA SA308286993GA SA308286994GA SA308287006GA SA308206587GA SA308206660GA SA308266858GA SA308266880GA SA308286933GA SA308286939GA SA308286949GA SA308286959GA SA308286960GA SA308286962GA SA308286966GA SA308286975GA SA308286977GA SA308286979GA SA308286981GA SA308286982GA SA308286988GA SA308287007GA SA308287008GA SA308287010GA SA308266881GA SA308286945GA SA308286969GA SA308286996GA SA308287016GA SA308166198GA SA308196574GA SA308206630GA SA308176220GA SA308186334GA SA308266789GA SA308266897GA SA308286910GA SA308286912GA SA308286914GA SA308286918GA SA308286932GA SA308286938GA SA308286941GA SA308286948GA SA308286972GA SA308286980GA SA308286983GA SA308286987GA SA308286989GA SA308286995GA SA308287004GA SA308287005GA SA308287017GA SA308287018GA SA308287019GA SA308287020GA SA308287021GA SA308287022GA SA308287023GA SA308287024GA SA308317065GA SA308317066GA SA308317067GA SA308317068GA SA308317069GA SA308317070GA SA308317071GA SA308317072GA SA308317073GA SA308317074GA SA308317075GA SA308317076GA SA308317077GA SA308317078GA SA308317079GA SA308317080GA SA308317081GA SA308317082GA SA308317083GA SA308317084GA SA308317085GA SA308317086GA SA308317087GA SA308317088GA SA308317089GA SA308317090GA SA308317091GA SA308317092GA SA308317093GA SA308317094GA SA308317095GA SA308317097GA SA308317101GA SA308317102GA SA308317103GA SA308317104GA SA308317106GA SA308317107GA SA308317108GA SA308317109GA SA308317110GA SA308317111GA SA308317112GA SA308317113GA SA308317114GA SA308317115GA SA308317116GA SA308317117GA SA308317118GA SA308317119GA SA308317120GA SA308317121GA SA308317122GA SA308317123GA SA308317124GA SA308317125GA SA308317126GA SA308317127GA SA308317128GA SA308317129GA SA308317130GA SA308317131GA SA308317132GA SA308317133GA SA308317134GA SA308317135GA SA308317136GA SA308317137GA SA308317138GA SA308317139GA SA308317140GA SA308317141GA SA308317142GA SA308317143GA SA308317144GA SA308317145GA SA308317146GA SA308317147GA SA308317148GA SA308317149GA SA308317150GA SA308317152GA SA308317154GA SA308317155GA SA308317156GA SA308317157GA SA308317158GA SA308317159GA SA308317160GA SA308317161GA SA308317162GA SA308317163GA SA308317164GA SA308317165GA SA308317183GA SA308317184GA SA308317185GA SA308317186GA SA308317187GA SA308317188GA SA308317189GA SA308317190GA SA308317191GA SA308317193GA SA308317196GA SA308317197GA SA308317198GA SA308317199GA SA308317200GA SA308317204GA SA308317206GA SA308317207GA SA308317208GA SA308317209GA SA308317210GA SA308317211GA SA308317212GA SA308317213GA SA308317214GA SA308186389GA SA308206612GA SA308206673GA SA308206677GA SA308206687GA SA308206693GA SA308206714GA SA308206715GA SA308206716GA SA308206717GA SA308206726GA SA308206727GA SA308206728GA SA308206730GA SA308206731GA SA308206735GA SA308206741GA SA308206745GA SA308206746GA SA308206748GA SA308206749GA SA308206752GA SA308206754GA SA308206756GA SA308206758GA SA308216765GA SA308216766GA SA308216770GA SA308216773GA SA308216774GA SA308216777GA SA308286990GA SA308317098GA SA308317105GA SA308317166GA SA308317167GA SA308317169GA SA308317170GA SA308317171GA SA308317173GA SA308317175GA SA308317176GA SA308317177GA SA308317179GA SA308317192GA SA308317194GA SA308317195GA SA308317225GA SA308317227GA SA308317232GA SA308317233GA SA308317234GA SA308317236GA SA308317244GA SA308317245GA SA308317251GA SA308317254GA SA308317258GA SA308317259GA SA308317262GA SA308317264GA SA308317265GA SA308317268GA SA308317270GA SA308317271GA SA308145681GA SA308155887GA SA308186351GA SA308196569GA SA308196575GA SA308196582GA SA308206613GA SA308206614GA SA308206618GA SA308206634GA SA308206647GA SA308206650GA SA308206651GA SA308206652GA SA308206653GA SA308206654GA SA308206655GA SA308206656GA SA308206657GA SA308206658GA SA308206659GA SA308206662GA SA308206666GA SA308206667GA SA308206668GA SA308206671GA SA308206672GA SA308206675GA SA308206676GA SA308206678GA SA308206680GA SA308206682GA SA308206684GA SA308206685GA SA308206686GA SA308206688GA SA308206689GA SA308206690GA SA308206691GA SA308206692GA SA308206696GA SA308206698GA SA308206700GA SA308206701GA SA308206703GA SA308206704GA SA308206705GA SA308206707GA SA308206710GA SA308206711GA SA308206720GA SA308206722GA SA308206725GA SA308286937GA SA308286986GA SA308317099GA SA308317100GA SA308317168GA SA308317180GA SA308317181GA SA308317182GA SA308317228GA SA308317230GA SA308317231GA SA308317237GA SA308317238GA SA308317239GA SA308317240GA SA308317241GA SA308317243GA SA308317250GA SA308317255GA SA308317256GA SA308317257GA SA308317260GA SA308317261GA SA308317263GA SA308317266GA SA308317267GA SA308317269GA SA308317272GA SA308317274GA SA308317278GA SA308317303GA SA308317311GA SA308176217GA SA308206702GA SA308206708GA SA308206712GA SA308206713GA SA308206719GA SA308206724GA SA308206733GA SA308206739GA SA308206744GA SA308206750GA SA308317151GA SA308317229GA SA308317246GA SA308317248GA SA308317249GA SA308317273GA SA308317276GA SA308317277GA SA308317280GA SA308317282GA SA308317283GA SA308317284GA SA308317289GA SA308317291GA SA308317295GA SA308317296GA SA308317305GA SA308317306GA SA308317307GA SA308317308GA SA308317309GA SA308327340GA SA308327341GA SA308327345GA SA308327351GA SA308327352GA SA308327353GA SA308327354GA SA308327356GA SA308327357GA SA308327358GA SA308266861GA SA308317096GA SA308317153GA SA308317172GA SA308317201GA SA308317202GA SA308317203GA SA308317205GA SA308317223GA SA308317224GA SA308317247GA SA308317252GA SA308317253GA SA308317275GA SA308317279GA SA308317281GA SA308317287GA SA308317292GA SA308317293GA SA308317294GA SA308317297GA SA308317298GA SA308317299GA SA308317301GA SA308317304GA SA308327312GA SA308327315GA SA308327316GA SA308327318GA SA308327321GA SA308327322GA SA308327323GA SA308327326GA SA308327327GA SA308327328GA SA308327329GA SA308327330GA SA308327331GA SA308327332GA SA308327333GA SA308327334GA SA308327339GA SA308327342GA SA308327343GA SA308327347GA SA308327348GA SA308327349GA SA308327350GA SA308327355GA SA308327360GA SA308327361GA SA308327362GA SA308327364GA SA308327365GA SA308327367GA SA308327370GA SA308327371GA SA308327372GA SA308327378GA SA308327379GA SA308327385GA SA308327386GA SA308327401GA SA308327402GA SA308327403GA SA308327421GA SA308327424GA SA308327427GA SA308327428GA SA308327430GA SA308327432GA SA308327437GA SA308327438GA SA308327440GA SA308327441GA SA308327442GA SA308327448GA SA308327450GA SA308327454GA SA308327457GA SA308327459GA SA308327462GA SA308327463GA SA308327466GA SA308327467GA SA308327470GA SA308327474GA SA308327483GA SA308327493GA SA308327496GA SA308327501GA SA308327502GA SA308327509GA SA308327510GA SA308327524GA SA308327319GA SA308327335GA SA308327363GA SA308327373GA SA308327381GA SA308327382GA SA308327384GA SA308327445GA SA308327447GA SA308327452GA SA308327453GA SA308327455GA SA308327456GA SA308327468GA SA308327469GA SA308327473GA SA308206718GA SA308206736GA SA308206740GA SA308206747GA SA308317235GA SA308317285GA SA308317286GA SA308317288GA SA308327344GA SA308327359GA SA308327451GA SA308196540GA SA308196579GA SA308206591GA SA308206732GA SA308206755GA SA308206761GA SA308266892GA SA308297041GA SA308317174GA SA308317178GA SA308206709GAMKE SA308206738GA SA308206729GA SA308286922GA SA308286970GA SA308327368GA SA308327380GA SA308327383GA SA308327420GA SA308327426GA SA308327429GA SA308327434GA SA308327436GA SA308327439GA SA308327444GA SA308327458GA SA308327461GA SA308327465GA SA308327482GA SA308327492GA SA308327495GA SA308327503GA SA308327505GA SA308327513GA SA308327515GA SA308327525GA SA308327551GA SA308337629GA SA308166039GA SA308327536GA SA308327537GA SA308327325GA SA308327337GA SA308327366GA SA308327377GA SA308327431GA SA308327460GA SA308327477GA SA308327485GA SA308327491GA SA308327497GA SA308327500GA SA308327508GA SA308327514GA SA308327516GA SA308327519GA SA308327531GA SA308327532GA SA308327533GA SA308337563GA SA308337564GA SA308337565GA SA308337567GA SA308337568GA SA308337569GA SA308337571GA SA308337572GA SA308337573GA SA308337574GA SA308337576GA SA308337577GA SA308337578GA SA308337579GA SA308337580GA SA308337581GA SA308337582GA SA308337583GA SA308337586GA SA308337589GA SA308337590GA SA308337591GA SA308337606GA SA308337607GA SA308337608GA SA308337609GA SA308337611GA SA308337612GA SA308337613GA SA308337614GA SA308337615GA SA308337616GA SA308337628GA SA308337630GA SA308337633GA SA308327317GA SA308327346GA SA308327472GA SA308327480GA SA308327486GA SA308327520GA SA308327530GA SA308327542GA SA308327543GA SA308327544GA SA308327548GA SA308327556GA SA308337587GA SA308337592GA SA308337595GA SA308337596GA SA308337601GA SA308337610GA SA308337637GA SA308337684GA SA308317242GA SA308317310GA SA308327425GA SA308327433GA SA308327435GA SA308327449GA SA308327475GA SA308327478GA SA308327479GA SA308327481GA SA308327488GA SA308327499GA SA308327506GA SA308327507GA SA308327517GA SA308327518GA SA308327523GA SA308327527GA SA308327529GA SA308327554GA SA308327555GA SA308327557GA SA308327558GA SA308337584GA SA308337585GA SA308337603GA SA308337604GA SA308337618GA SA308337643GA SA308337662GA SA308337682GA SA308337689GA SA308337690GA SA308337695GA SA308337698GA SA308337699GA SA308337700GA SA308347726GA SA308357947GA SA308357948GA SA308357949GA SA308357950GA SA308357952GA SA308357953GA SA308357954GA SA308357961GA SA308327324GA SA308327369GA SA308327375GA SA308327494GA SA308327528GA SA308327547GA SA308327549GA SA308337588GA SA308337594GA SA308337598GA SA308337599GA SA308337602GA SA308337621GAMKE SA308337622GA SA308337624GA SA308337625GA SA308337641GA SA308337651GA SA308337663GA SA308337669GA SA308337680GA SA308337681GA SA308337687GA SA308327374GA SA308317290GA SA308327446GA SA308337597GA SA308337617GA SA308337619GA SA308337620GA SA308337623GA SA308337640GA SA308337642GA SA308337679GA SA308337696GA SA308337701GA SA308347723GA SA308347730GA SA308347743GA SA308347752GA SA308347817GA SA308297059GA SA308297061GA SA308297062GA SA308337626GA SA308337639GA SA308337644GA SA308337645GA SA308337649GA SA308337650GA SA308337653GA SA308337654GA SA308337655GA SA308337656GA SA308337657GA SA308337659GA SA308337660GA SA308337664GA SA308337665GA SA308337670GA SA308337671GA SA308337678GA SA308337683GA SA308337685GA SA308337692GA SA308337693GA SA308337694GA SA308357951GA SA308347728GA SA308347729GA SA308347742GA SA308347744GA SA308347852GA SA308357942GA SA308327464GA SA308347748GA SA308347763GA SA308347807GA SA308347810GA SA308347814GA SA308347820GA SA308347825GA SA308347828GA SA308347832GA SA308347833GA SA308347834GA SA308347835GA SA308347837GA SA308347838GA SA308347842GA SA308347843GA SA308186423GA SA308196525GA SA308206625GA SA308206759GA SA308206760GA SA308216762GA SA308216763GA SA308216764GA SA308216768GA SA308216772GA SA308216776GA SA308216780GA SA308347781GA SA308347786GA SA308347788GA SA308347794GA SA308317226GA SA308327511GA SA308337566GA SA308337635GA SA308337691GA SA308347727GA SA308347749GA SA308347819GA SA308347853GA SA308357955GA SA308357992GA SA308368127GA SA308347724GA SA308347731GA SA308347818GA SA308347822GA SA308357959GA SA308357962GA SA308347757GA SA308357940GA SA308327512GA SA308347732GA SA308347765GA SA308347766GA SA308347775GA SA308347776GA SA308347777GA SA308347780GA SA308347782GA SA308347784GA SA308347795GA SA308347797GA SA308347860GA SA308327487GA SA308327521GA SA308337705GA SA308337708GA SA308347762GA SA308347764GA SA308347771GA SA308347778GA SA308347793GA SA308347802GA SA308347803GA SA308347806GA SA308347826GA SA308347854GA SA308347857GA SA308357862GA SA308357864GA SA308357868GA SA308357879GA SA308357881GA SA308357882GA SA308357885GA SA308357899GA SA308357906GA SA308357914GA SA308357921GA SA308357926GA SA308357956GA SA308357957GA SA308357958GA SA308357960GA SA308357963GA SA308357964GA SA308357965GA SA308357966GA SA308357968GA SA308357969GA SA308357970GA SA308357971GA SA308357972GA SA308357973GA SA308357974GA SA308357978GA SA308357979GA SA308357980GA SA308357981GA SA308357982GA SA308357983GA SA308357984GA SA308357985GA SA308357986GA SA308357987GA SA308357988GA SA308357989GA SA308357994GA SA308357996GA SA308357998GA SA308357999GA SA308358000GA SA308358001GA SA308358002GA SA308358003GA SA308358004GA SA308358005GA SA308358006GA SA308358007GA SA308358008GA SA308358009GA SA308358011GA SA308358012GA SA308358014GA SA308358015GA SA308358016GA SA308358017GA SA308358018GA SA308358019GA SA308358020GA SA308358021GA SA308358024GA SA308358029GA SA308358045GA SA308358046GA SA308368048GA SA308368049GA SA308368051GA SA308368053GA SA308368054GA SA308368055GA SA308368056GA SA308368059GA SA308368063GA SA308368069GA SA308368072GA SA308368073GA SA308368075GA SA308368080GA SA308368081GA SA308368084GA SA308368086GA SA308368090GA SA308368091GA SA308368099GA SA308368100GA SA308368101GA SA308368106GA SA308368108GA SA308368114GA SA308368123GA SA308368130GA SA308368133GA SA308368134GA SA308368135GA SA308368138GA SA308368140GA SA308368142GA SA308368147GA SA308368160GA SA308368163GA SA308368164GA SA308368170GA SA308368174GA SA308368182GA SA308368183GA SA308368184GA SA308368186GA SA308368197GA SA308368200GA SA308368201GA SA308368209GA SA308368212GA SA308368229GA SA308368235GA SA308378283GA SA308347725GA SA308347741GA SA308347808GA SA308347813GA SA308347815GA SA308347821GA SA308347823GA SA308347827GA SA308347831GA SA308347839GA SA308347841GA SA308357865GA SA308357874GA SA308357875GA SA308357876GA SA308357877GA SA308357883GA SA308357886GA SA308357887GA SA308357891GA SA308357892GA SA308357894GA SA308357896GA SA308357898GA SA308357911GA SA308357913GA SA308357916GA SA308357917GA SA308357918GA SA308357919GA SA308357920GA SA308357923GA SA308357925GA SA308357927GA SA308357933GA SA308357934GA SA308357935GA SA308357936GA SA308357937GA SA308357938GA SA308357939GA SA308358013GA SA308358022GA SA308358026GA SA308358032GA SA308358033GA SA308358034GA SA308358037GA SA308358038GA SA308358043GA SA308368066GA SA308368067GA SA308368079GA SA308368082GA SA308368083GA SA308368085GA SA308368088GA SA308368089GA SA308368102GA SA308368104GA SA308368105GA SA308368118GA SA308368126GA SA308368128GA SA308368129GA SA308368143GA SA308368153GA SA308368158GA SA308368165GA SA308368166GA SA308368171GA SA308368180GA SA308368187GA SA308368196GA SA308368199GA SA308368203GA SA308368227GA SA308368233GA SA308368244GA SA308368248GA SA308378254GA SA308378278GA SA308378284GA SA308378292GA SA308347856GA SA308357930GA SA308358044GA SA308368057GA SA308368058GA SA308368087GA SA308368095GA SA308368103GA SA308368107GA SA308368109GA SA308368110GA SA308368112GA SA308368125GA SA308368131GA SA308368132GA SA308368137GA SA308368139GA SA308368145GA SA308368146GA SA308368148GA SA308368149GA SA308368155GA SA308368156GA SA308368157GA SA308368159GA SA308368162GA SA308368172GA SA308368185GA SA308368188GA SA308368189GA SA308368190GA SA308368206GA SA308368207GA SA308368208GA SA308368225GA SA308368226GA SA308368237GA SA308368238GA SA308368245GA SA308378277GA SA308378279GA SA308378282GA SA308378287GA SA308347770GA SA308357866GA SA308357922GA SA308358023GA SA308358025GA SA308358028GA SA308358031GA SA308358036GA SA308358039GA SA308358040GA SA308358041GA SA308358042GA SA308368047GA SA308368061GA SA308368064GA SA308368065GA SA308368068GA SA308368071GA SA308368074GA SA308368076GA SA308368078GA SA308368096GA SA308368097GA SA308368098GA SA308368136GA SA308368144GA SA308368151GA SA308368154GA SA308368168GA SA308368173GA SA308368176GA SA308368192GA SA308368193GA SA308368194GA SA308368198GA SA308368205GA SA308368210GA SA308368214GA SA308368215GA SA308368218GA SA308368219GAMKE SA308368221GA SA308368222GA SA308378251GA SA308378253GA SA308378255GA SA308378261GA SA308378262GA SA308378267GA SA308378314GA SA308378320GA SA308378323GA SA308196487GA SA308216769GA SA308216779GA SA308358027GA SA308368092GA SA308368094GA SA308368169GA SA308368175GA SA308368177GA SA308368178GA SA308368179GA SA308368181GA SA308368191GA SA308368213GA SA308368217GA SA308368220GA SA308368223GA SA308368224GA SA308368228GA SA308368230GA SA308368241GA SA308378275GA SA308378285GA SA308378293GA SA308378299GA SA308378300GA SA308378301GA SA308378303GA SA308378305GA SA308378307GA SA308378308GA SA308378309GA SA308378311GA SA308378313GA SA308378315GA SA308378316GA SA308378317GA SA308378318GA SA308378319GA SA308378322GA SA308378324GA SA308378325GA SA308378326GA SA308378327GA SA308378328GA SA308378329GA SA308388412GA SA308388414GA SA308388419GA SA308388421GA SA308388422GA SA308388424GA SA308388425GA SA308388426GA SA308388427GA SA308388428GA SA308388430GA SA308388431GA SA308388433GA SA308388435GA SA308388437GA SA308388439GA SA308388444GA SA308337575GA SA308347737GA SA308347768GA SA308347773GA SA308347774GA SA308347783GA SA308347787GA SA308347799GA SA308347804GA SA308347840GA SA308347855GA SA308347858GA SA308347859GA SA308357871GA SA308357889GA SA308357897GA SA308357909GA SA308357910GA SA308357993GA SA308357995GA SA308368052GA SA308368150GA SA308368161GA SA308378256GA SA308378258GA SA308378259GA SA308378260GA SA308378264GA SA308378265GA SA308378266GA SA308378268GA SA308378269GA SA308378270GA SA308378271GA SA308378272GA SA308378273GA SA308388365GA SA308388366GA SA308388409GA SA308388410GA SA308388413GA SA308388417GA SA308388423GA SA308388432GA SA308388434GA SA308388436GA SA308388442GA SA308388451GA SA308388453GA SA308388454GA SA308388488GA SA308388490GA SA308388491GA SA308388492GA SA308388493GA SA308388496GA SA308388497GA SA308388499GA SA308388500GA SA308388502GA SA308388504GA SA308388505GA SA308388507GA SA308388508GA SA308388509GA SA308388510GA SA308388511GA SA308388512GA SA308388513GA SA308388514GA SA308388515GA SA308388516GA SA308388517GA SA308388518GA SA308388519GA SA308388520GA SA308388521GA SA308388522GA SA308388523GA SA308388524GA SA308388525GA SA308388526GA SA308388527GA SA308388529GA SA308388530GA SA308388533GA SA308388534GA SA308388535GA SA308388536GA SA308388537GA SA308388538GA SA308388539GA SA308388541GA SA308388542GA SA308388543GA SA308388547GA SA308398639GA SA308398641GA SA308398642GA SA308398643GA SA308398645GA SA308398646GA SA308398647GA SA308378294GA SA308378295GA SA308378298GA SA308378302GA SA308378330GA SA308378331GA SA308378332GA SA308378333GA SA308378334GA SA308378335GA SA308388407GA SA308388408GA SA308388411GA SA308388420GA SA308388429GA SA308388438GA SA308388440GA SA308388441GA SA308388445GA SA308388447GA SA308388450GA SA308388457GA SA308388461GA SA308388463GA SA308388464GA SA308388465GA SA308388466GA SA308388467GA SA308388469GA SA308388470GA SA308388471GA SA308388472GA SA308388473GA SA308388474GA SA308388475GA SA308388478GA SA308388480GA SA308388482GA SA308388483GA SA308388485GA SA308388486GA SA308388487GA SA308388498GA SA308388528GA SA308388531GA SA308388532GA SA308388545GA SA308388549GA SA308388551GA SA308388552GA SA308388553GA SA308388555GA SA308388556GA SA308388557GA SA308388559GA SA308388560GA SA308388562GA SA308388565GA SA308388570GA SA308388578GA SA308388583GA SA308388584GA SA308388585GA SA308388587GA SA308388590GA SA308388600GA SA308388602GA SA308388604GA SA308388605GA SA308388608GA SA308388609GA SA308388610GA SA308388616GA SA308388625GA SA308388626GA SA308388628GA SA308388635GA SA308398640GA SA308398644GA SA308388468GA SA308388484GA SA308388567GA SA308388568GA SA308388579GA SA308388582GA SA308388586GA SA308388588GA SA308388606GA SA308388479GA SA308388613GA SA308358035GA SA308378304GA SA308388607GA SA308388615GA SA308388618GA SA308388619GA SA308388620GA SA308388622GA SA308398676GA SA308398677GA SA308398681GA SA308398694GA SA308398696GA SA308398697GA SA308398698GA SA308398699GA SA308398700GA SA308398701GA SA308398702GA SA308398703GA SA308398704GA SA308398705GA SA308398707GA SA308398708GA SA308388614GA SA308398745GA SA308398648GA SA308398651GA SA308398654GA SA308398657GA SA308398661GA SA308398720GA SA308398730GA SA308398733GA SA308398740GA SA308398742GA SA308398749GA SA308398772GA SA308398773GA SA308398781GA SA308398785GA SA308398817GA SA308398866GA SA308398874GA SA308408910GA SA308378291GA SA308378312GA SA308388577GA SA308398695GA SA308398741GA SA308398821GA SA308398824GA SA308398848GA SA308408930GA SA308378310GA SA308398652GA SA308398665GA SA308408906GA SA308408916GA SA308408925GA SA308398717GA SA308398743GA SA308398753GA SA308398755GA SA308398769GA SA308398649GA SA308398655GA SA308398683GA SA308398731GA SA308398795GA SA308398802GA SA308398803GA SA308398810GA SA308398812GA SA308398825GA SA308398826GA SA308398828GA SA308398830GA SA308398855GA SA308398857GA SA308398858GA SA308398903GA SA308408915GA SA308398746GA SA308398758GA SA308398771GA SA308398778GA SA308398871GA SA308297063GA SA308337658GA SA308337661GA SA308337668GA SA308337672GA SA308388561GA SA308388595GA SA308398841GA SA308398847GA SA308398853GA SA308398860GA SA308408907GA SA308408912GA SA308398714GA SA308398721GA SA308398722GA SA308398725GA SA308398729GA SA308398732GA SA308398752GA SA308398754GA SA308398756GA SA308398762GA SA308398763GA SA308398770GA SA308409010GA SA308409016GA SA308409022GA SA308409027GA SA308409034GA SA308398670GA SA308408911GA SA308408924GA SA308408934GA SA308408946GA SA308408948GA SA308408949GA SA308408957GA SA308409012GA SA308409013GA SA308409019GA SA308409025GA SA308409030GA SA308409039GA SA308409045GA SA308409047GA SA308409050GA SA308409057GA SA308409060GA SA308409063GA SA308409064GA SA308409074GA SA308409075GA SA308409082GA SA308409107GA SA308337702GA SA308347769GA SA308347779GA SA308347785GA SA308347790GA SA308347796GA SA308368093GA SA308388580GA SA308388593GA SA308409032GA SA308409036GA SA308409078GA SA308409081GA SA308409090GA SA308409092GA SA308409163GA SA308409166GA SA308409172GA SA308409175GA SA308409180GA SA308409186GA SA308409202GA SA308409203GA SA308409204GA SA308409207GA SA308419215GA SA308419220GA SA308419238GA SA308419244GA SA308398896GA SA308409037GA SA308419248GA SA308378289GA SA308388415GA SA308388416GA SA308388418GA SA308388448GA SA308388452GA SA308388455GA SA308388456GA SA308388458GA SA308388459GA SA308388460GA SA308388462GA SA308388477GA SA308398653GA SA308398711GA SA308398779GA SA308398882GA SA308398892GA SA308398893GA SA308398895GA SA308398897GA SA308398898GA SA308398901GA SA308408967GA SA308408968GA SA308409073GA SA308409079GA SA308388546GA SA308398877GA SA308398887GA SA308409076GA SA308327471GA SA308357975GA SA308368119GA SA308378257GA SA308388494GA SA308388495GA SA308388501GA SA308388544GA SA308398719GA SA308398765GA SA308398868GA SA308398872GA SA308398875GA SA308398876GA SA308398879GA SA308398883GA SA308398884GA SA308398891GA SA308398894GA SA308408965GA SA308408966GA SA308408970GA SA308408971GA SA308408975GA SA308408976GA SA308408979GA SA308408980GA SA308408981GA SA308408982GA SA308408983GA SA308408985GA SA308408988GA SA308378263GA SA308408984GA SA308408969GA SA308409083GA SA308419265GA SA308357990GA SA308368236GA SA308388449GA SA308398660GA SA308398738GA SA308398783GA SA308398784GA SA308398878GA SA308398880GA SA308398881GA SA308398885GA SA308398886GA SA308398888GA SA308398889GA SA308398899GA SA308398900GA SA308408960GA SA308409017GA SA308409033GA SA308409058GA SA308409071GA SA308409091GA SA308409094GA SA308409128GA SA308409167GA SA308409181GA SA308409205GA SA308409206GA SA308409208GA SA308409210GA SA308409211GA SA308419214GA SA308419217GA SA308419223GA SA308419224GA SA308419225GA SA308419228GA SA308419235GA SA308419252GA SA308419254GA SA308419269GA SA308419270GA SA308419279GA SA308419241GA SA308408917GA SA308409020GA SA308409041GA SA308409042GA SA308409043GA SA308409044GA SA308409046GA SA308409056GA SA308409059GA SA308409069GA SA308409087GA SA308409171GA SA308409179GA SA308409185GA SA308419219GA SA308419227GA SA308419234GA SA308419253GA SA308419261GA SA308368167GA SA308398813GA SA308409001GA SA308409176GA SA308409190GA SA308409191GA SA308409192GA SA308409193GA SA308409195GA SA308409199GA SA308419213GA SA308419221GA SA308419226GA SA308429323GA SA308429324GA SA308429330GA SA308429332GA SA308429334GA SA308429335GA SA308429336GA SA308429338GA SA308429343GA SA308429344GA SA308429348GA SA308429358GA SA308206669GA SA308206674GA SA308206681GA SA308206683GA SA308378290GA SA308419256GA SA308429322GA SA308429326GA SA308429333GA SA308429337GA SA308429339GA SA308429345GA SA308429361GA SA308368152GA SA308398865GA SA308408972GA SA308408977GA SA308408991GA SA308408993GA SA308408996GA SA308409139GA SA308409141GA SA308409146GA SA308409147GA SA308409148GA SA308409149GA SA308409151GA SA308419280GA SA308419282GA SA308419284GA SA308419291GA SA308429381GA SA308429382GA SA308357997GA SA308388489GA SA308398658GA SA308408974GA SA308408986GA SA308408987GA SA308409113GA SA308409132GA SA308409133GA SA308419287GA SA308419290GA SA308419296GA SA308419302GA SA308419304GA SA308429365GA SA308429370GA SA308429371GA SA308429372GA SA308429373GA SA308429374GA SA308429376GA SA308429377GA SA308429378GA SA308429388GA SA308429390GA SA308429406GA SA308337632GA SA308357991GA SA308378252GA SA308398716GA SA308398761GA SA308398777GA SA308398861GA SA308408990GA SA308409002GA SA308409004GA SA308409093GA SA308409096GA SA308409099GA SA308409108GA SA308409109GA SA308409110GA SA308409111GA SA308409114GA SA308409115GA SA308409116GA SA308409117GA SA308409118GA SA308409119GA SA308409121GA SA308409122GA SA308409123GA SA308409124GA SA308409126GA
SA308409127GA SA308409131GA SA308409134GA SA308409135GA SA308409137GA SA308409140GA SA308409142GA SA308409143GA SA308409145GA SA308409150GA SA308409152GA SA308419283GA SA308419285GA SA308419286GA SA308419288GA SA308419289GA SA308419292GA SA308419293GA SA308419294GA SA308419297GA SA308419298GA SA308419299GA SA308419301GA SA308419303GA SA308419305GA SA308419306GA SA308419307GA SA308429366GA SA308429367GA SA308429368GA SA308429369GA SA308429375GA SA308429379GA SA308429380GA SA308429383GA SA308429384GA SA308429385GA SA308429386GA SA308429387GA SA308429389GA SA308429391GA SA308429392GA SA308429393GA SA308429394GA SA308429396GA SA308429397GA SA308429400GA SA308429401GA SA308429402GA SA308429403GA SA308429405GA SA308378288GA SA308398822GA SA308409038GA SA308409065GA SA308409130GA SA308409154GA SA308409182GA SA308409209GA SA308419222GA SA308419247GA SA308429346GA SA308429356GA SA308429398GA SA308409112GA SA308409136GA SA308409144GA SA308419295GA SA308429318GA SA308429362GA SA308429419GA SA308449537GA SA308449605GA SA308449612GA SA308449613GA SA308449614GA SA308449630GA SA308459724GA SA308459742GA SA308459745GA SA308459746GA SA308459747GA SA308459748GA SA308459749GA SA308459751GA SA308459753GA SA308449539GA SA308388629GA SA308398687GA SA308419275GA SA308419276GA SA308419277GA SA308419278GA SA308429309GA SA308429310GA SA308429311GA SA308429313GA SA308429314GA SA308429316GA SA308429317GA SA308429319GA SA308429320GA SA308429364GA SA308439425GA SA308439426GA SA308439427GA SA308439431GA SA308439433GA SA308439434GA SA308439435GA SA308439439GA SA308439442GA SA308439444GA SA308439445GA SA308439460GA SA308439461GA SA308439462GA SA308459843GA SA308368216GA SA308429315GA SA308439438GA SA308439446GA SA308449491GA SA308449528GA SA308449565GA SA308449576GA SA308449579GA SA308459644GA SA308459648GA SA308459649GA SA308459650GA SA308459652GA SA308459653GA SA308459654GA SA308459655GA SA308459656GA SA308459657GA SA308459658GA SA308459661GA SA308459665GA SA308459666GA SA308459679GA SA308459707GA SA308449621GA SA308439457GA SA308449483GA SA308449485GA SA308449488GA SA308449489GA SA308449498GA SA308449553GA SA308459708GA SA308459737GA SA308459739GA SA308459744GA SA308459752GA SA308459744GA SA308459737GA SA308459739GA SA308459752GA SA308459690GA SA308459691GA SA308459692GA SA308459696GA SA308459772GA SA308459796GA SA308459845GA SA308459847GA SA308459850GA SA308459862GA SA308459863GA SA308459865GA SA308459869GA SA308459876GA SA308449636GA SA308459703GA SA308459729GA SA308459754GA SA308459755GA SA308459761GA SA308459766GA SA308459793GA SA308459871GA SA308449536GA SA308449617GA SA308429412GA SA308429414GA SA308429420GA SA308449543GA SA308449549GA SA308449608GA SA308449637GA SA308459712GA SA308459767GA SA308408938GA SA308409201GA SA308439429GA SA308439452GA SA308439458GA SA308449484GA SA308449490GA SA308449500GA SA308449501GA SA308449507GA SA308449515GA SA308449519GA SA308449523GA SA308449524GA SA308449526GA SA308449529GA SA308449530GA SA308449531GA SA308449532GA SA308449569GA SA308449570GA SA308449571GA SA308449574GA SA308449581GA SA308449583GA SA308449585GA SA308449589GA SA308449590GA SA308449591GA SA308449593GA SA308449597GA SA308459643GA SA308459645GA SA308459674GA SA308459693GA SA308459694GA SA308459695GA SA308459797GA SA308459839GA SA308459841GA SA308459844GA SA308459864GA SA308459866GA SA308469879GA SA308429409GA SA308429410GA SA308429418GA SA308429421GA SA308429422GA SA308429423GA SA308439468GA SA308439477GA SA308449533GA SA308449534GA SA308449540GA SA308449542GA SA308449544GA SA308449545GA SA308449546GA SA308449548GA SA308449599GA SA308449609GA SA308449610GA SA308449615GA SA308449627GA SA308449634GA SA308459701GA SA308459702GA SA308459709GA SA308459710GA SA308459714GA SA308459720GA SA308459721GA SA308459722GA SA308459735GA SA308459740GA SA308459741GA SA308459743GA SA308459757GA SA308459758GA SA308459769GA SA308459792GA SA308439455GA SA308439466GA SA308439470GA SA308449502GA SA308449511GA SA308449598GA SA308459838GA SA308429417GA SA308459750GA SA308459814GA SA308459784GA SA308459790GA SA308439474GA SA308439476GA SA308449551GA SA308449602GA SA308449622GA SA308449624GA SA308459718GA SA308459719GA SA308459723GA SA308459730GA SA308459738GA SA308459759GA SA308459762GA SA308459783GA SA308459785GA SA308459789GA SA308459815GA SA308459699GA SA308479900GA SA308489917GA SA308489922GA SA308489927GA SA308489948GA SA308489962GA SA308489983GA SA308176291GA SA308398650GA SA308439475GA SA308449625GA SA308459781GA SA308459798GA SA308459800GA SA308459802GA SA308459803GA SA308459804GA SA308459806GA SA308459810GA SA308459811GA SA308459812GA SA308459816GA SA308459820GA SA308459821GA SA308459827GA SA308459834GA SA308459836GA SA308489921GA SA308489994GA SA308409168GA SA308429408GA SA308429416GA SA308439465GA SA308439478GA SA308449538GA SA308449555GA SA308449556GA SA308449560GA SA308449561GA SA308449600GA SA308449601GA SA308449626GA SA308449631GA SA308459711GA SA308459717GA SA308459733GA SA308459756GA SA308459765GA SA308459770GA SA308459773GA SA308459774GA SA308459782GA SA308459791GA SA308489911GA SA308489912GA SA308491108GA SA308491115GA SA308491116GA SA308491117GA SA308491122GA SA308491124GA SA308491125GA SA308491126GA SA308491129GA SA308501322GA SA308429407GA SA308429411GA SA308429424GA SA308439472GA SA308439473GA SA308449541GA SA308449547GA SA308449550GA SA308449552GA SA308449554GA SA308449557GA SA308449558GA SA308449559GA SA308449562GA SA308449596GA SA308449603GA SA308449607GA SA308449620GA SA308449633GA SA308449635GA SA308459639GA SA308459670GA SA308459704GA SA308459705GA SA308459713GA SA308459715GA SA308459726GA SA308459728GA SA308459734GA SA308459763GA SA308459768GA SA308459776GA SA308459779GA SA308459780GA SA308459787GA SA308459805GA SA308459807GA SA308459813GA SA308459817GA SA308459818GA SA308459819GA SA308459822GA SA308459823GA SA308459824GA SA308459825GA SA308459826GA SA308459829GA SA308459832GA SA308459833GA SA308459835GA SA308459837GA SA308481053GA SA308481054GA SA308481055GA SA308481056GA SA308481057GA SA308481058GA SA308481059GA SA308481060GA SA308491072GA SA308491073GA SA308491074GA SA308491077GA SA308491078GA SA308491079GA SA308491080GA SA308491082GA SA308491083GA SA308491085GA SA308491086GA SA308491088GA SA308491089GA SA308491090GA SA308491093GA SA308491097GA SA308501252GA SA308501254GA SA308501255GA SA308501257GA SA308501258GA SA308501259GA SA308501263GA SA308501264GA SA308449629GA SA308491076GA SA308491081GA SA308491084GA SA308491091GA SA308491094GA SA308491095GA SA308491098GA SA308491099GA SA308491100GA SA308491101GA SA308491212GA SA308501261GA SA308501266GA SA308501271GA SA308501272GA SA308501273GA SA308501274GA SA308501275GA SA308501276GA SA308501277GA SA308501278GA SA308501279GA SA308501280GA SA308501281GA SA308501282GA SA308501283GA SA308501284GA SA308501285GA SA308501286GA SA308501287GA SA308501288GA SA308501289GA SA308501290GA SA308501291GA SA308501294GA SA308501295GA SA308501296GA SA308501297GA SA308501298GA SA308501302GA SA308501304GA SA308501305GA SA308501306GA SA308501307GA SA308501309GA SA308501311GA SA308501312GA SA308501313GA SA308501314GA SA308501315GA SA308501316GA SA308501317GA SA308501319GA SA308501320GA SA308501321GA SA308501323GA SA308501324GA SA308501325GA SA308501326GA SA308501331GA SA308501334GA SA308501335GA SA308501339GA SA308501351GA SA308501373GA SA308501382GA SA308398873GA SA308449616GA SA308459777GA SA308459801GA SA308459809GA SA308459831GA SA308481052GA SA308491071GA SA308501253GA SA308501256GA SA308501260GA SA308501262GA SA308501265GA SA308501267GA SA308501268GA SA308501269GA SA308501270GA SA308511562GA SA307231285GA SA308388503GA SA308398728GA SA308408973GA SA308429404GA SA308429413GA SA308429415GA SA308439471GA SA308439479GA SA308449516GA SA308449535GA SA308449563GA SA308449564GA SA308449572GA SA308449604GA SA308449611GA SA308449619GA SA308449623GA SA308459706GA SA308459725GA SA308459727GA SA308459736GA SA308459760GA SA308459771GA SA308491102GA SA308491103GA SA308491104GA SA308491106GA SA308491107GA SA308491109GA SA308491110GA SA308491112GA SA308491113GA SA308491131GA SA308491134GA SA308491144GA SA308491167GA SA308491170GA SA308491235GA SA308481014GA SA308481035GA SA308481036GA SA308481041GA SA308481044GA SA308481046GA SA308481063GA SA308481068GA SA308489918GA SA308489960GA SA308489964GA SA308489979GA SA308491105GA SA308491118GA SA308491119GA SA308491120GA SA308491123GA SA308491132GA SA308491133GA SA308491135GA SA308491138GA SA308491139GA SA308491140GA SA308491141GA SA308491142GA SA308491173GA SA308491180GA SA308491182GA SA308491184GA SA308491185GA SA308491197GA SA308491209GA SA308501366GA SA308501369GA SA308449513GA SA308449577GA SA308459642GA SA308459678GA SA308459808GA SA308491130GA SA308491192GA SA308491194GA SA308491225GA SA308491228GA SA308501327GA SA308501328GA SA308501332GA SA308501333GA SA308501336GA SA308501340GA SA308501341GA SA308501342GA SA308501343GA SA308501344GA SA308501350GA SA308501353GA SA308501354GA SA308501364GA SA308501374GA SA308501386GA SA308501388GA SA308501425GA SA308511596GA SA308511604GA SA308388443GA SA308388476GA SA308398815GA SA308429325GA SA308449525GA SA308449566GA SA308449568GA SA308459638GA SA308459676GA SA308459794GA SA308459877GA SA308469885GA SA308469888GA SA308479890GA SA308479892GA SA308479895GA SA308388443GA SA308388476GA SA308398815GA SA308429325GA SA308449525GA SA308449566GA SA308449568GA SA308449628GA SA308459638GA SA308459676GA SA308459794GA SA308459799GA SA308459828GA SA308459877GA SA308469885GA SA308469888GA SA308479890GA SA308479892GA SA308479895GA SA308491075GA SA308491087GA SA308491114GA SA308491128GA SA308491143GA SA308491145GA SA308491146GA SA308491147GA SA308491148GA SA308491149GA SA308491150GA SA308491151GA SA308491159GA SA308491164GA SA308491165GA SA308491166GA SA308491171GA SA308491172GA SA308491176GA SA308491179GA SA308491183GA SA308491186GA SA308491187GA SA308491188GA SA308491189GA SA308491190GA SA308491191GA SA308491193GA SA308491195GA SA308491196GA SA308491198GA SA308491199GA SA308491200GA SA308491201GA SA308491202GA SA308491203GA SA308491204GA SA308491205GA SA308491206GA SA308491207GA SA308491208GA SA308491210GA SA308491211GA SA308491214GA SA308491215GA SA308491217GA SA308491218GA SA308491220GA SA308491221GA SA308491222GA SA308491223GA SA308491224GA SA308491226GA SA308491231GA SA308491232GA SA308491241GA SA308501292GA SA308501308GA SA308501357GA SA308501358GA SA308501359GA SA308501360GA SA308501361GA SA308501362GA SA308501363GA SA308501367GA SA308501368GA SA308501371GA SA308501379GA SA308501381GA SA308501384GA SA308501390GA SA308501394GA SA308501412GA SA308511555GA SA308511579GA SA308511580GA SA308511590GA SA308511593GA SA308511598GA SA308511599GA SA308511600GA SA308511606GA SA308511612GA SA308511626GA SA308511637GA SA308491159GA SA308491207GA SA308491241GA SA308481005GA SA308489971GA SA308489976GA SA308491111GA SA308491127GA SA308491219GA SA308491244GA SA308491245GA SA308491246GA SA308491247GA SA308491251GA SA308501398GA SA308501399GA SA308501402GA SA308501404GA SA308501405GA SA308501406GA SA308501409GA SA308501415GA SA308501422GA SA308501423GA SA308501424GA SA308501427GA SA308501428GA SA308511435GA SA308511436GA SA308511437GA SA308511438GA SA308511439GA SA308511442GA SA308511448GA SA308511449GA SA308511454GA SA308511478GA
SA308511480GA SA308511485GA SA308511486GA SA308511490GA SA308511491GA SA308511492GA SA308511493GA SA308511495GA SA308511530GA SA308511534GA SA308511535GA SA308511536GA SA308511630GA SA308266879GA SA308459698GA SA308459731GA SA308481038GA SA308489973GA SA308491096GA SA308491136GA SA308491169GA SA308501303GA SA308501310GA SA308501329GA SA308501337GA SA308501349GA SA308501355GA SA308501370GA SA308501376GA SA308501377GA SA308501378GA SA308501380GA SA308501387GA SA308501389GA SA308501391GA SA308501416GA SA308501430GA SA308501432GA SA308501433GA SA308511446GA SA308511563GA SA308511571GA SA308511573GA SA308511574GA SA308511575GA SA308511581GA SA308511582GA SA308511583GA SA308511592GA SA308511597GA SA308511601GA SA308511602GA SA308511605GA SA308511607GA SA308511608GA SA308511611GA SA308511617GA SA308511618GA SA308439453GA SA308449487GA SA308459659GA SA308491121GA SA308491137GA SA308491174GA SA308491181GA SA308491227GA SA308491230GA SA308491233GA SA308491234GA SA308491236GA SA308491237GA SA308491238GA SA308491239GA SA308491240GA SA308491243GA SA308481061GA SA308491248GA SA308501401GA SA308501411GA SA308501419GA SA308501431GA SA308511440GA SA308511441GA SA308511444GA SA308511477GA SA308449496GA SA308459732GA SA308489954GA SA308511559GA SA308439454GA SA308439456GA SA308439463GA SA308439464GA SA308449480GA SA308449486GA SA308449494GA SA308449495GA SA308449499GA SA308449505GA SA308511464GA SA308419216GA SA308439459GA SA308449481GA SA308449517GA SA308449606GA SA308449618GA SA308449632GA SA308459788GA SA308491178GA SA308501420GA SA308511499GA SA308511526GA SA308511527GA SA308511585GA SA308511587GA SA308511624GA SA308358010GA SA308409125GA SA308419308GA SA308501392GA SA308501413GA SA308511578GA SA308511594GA SA308511595GA SA308511609GA SA308511614GA SA308511616GA SA308511619GA SA308511621GA SA308511622GA SA308511623GA SA308511648GA SA308511649GA SA308511650GA SA308511651GA SA308511652GA SA308511653GA SA308511654GA SA308511656GA SA308511657GA SA308511658GA SA308511659GA SA308511660GA SA308511661GA SA308511662GA SA308511663GA SA308511664GA SA308511667GA SA308511668GA SA308511672GA SA308511675GA SA308491161GA SA308491177GA SA308491249GA SA308501408GA SA308511502GA SA308511505GA SA308511517GA SA308511519GA SA308511524GA SA308511525GA SA308511533GA SA308459700GA SA308459830GA SA308501375GA SA308501395GA SA308511457GA SA308511565GA SA308511566GA SA308511568GA SA308511569GA SA308511572GA SA308511584GA SA308511586GA SA308511588GA SA308511589GA SA308511610GA SA308511613GA SA308511615GA SA308511625GA SA308511627GA SA308521810GA SA308501293GA SA308501393GA SA308501396GA SA308501397GA SA308511554GA SA308511556GA SA308511557GA SA308511558GA SA308511560GA SA308511564GA SA308511567GA SA308511576GA SA308511577GA SA308511603GA SA308511674GA SA308511676GA SA308511678GA SA308511679GA SA308511681GA SA308511682GA SA308511683GA SA308511686GA SA308521783GA SA308521784GA SA308521785GA SA308521786GA SA308521787GA SA308521788GA SA308521789GA SA308521790GA SA308521791GA SA308521792GA SA308521794GA SA308521798GA SA308521806GA SA308511591GA SA308511671GA SA308511677GA SA308521809GA SA308511472GA SA308511504GA SA308511689GA SA308521811GA SA308449504GA SA308501421GA SA308511447GA SA308511508GA SA308491250GA SA308501407GA SA308501414GA SA308511451GA SA308511456GA SA308511461GA SA308511470GA SA308511476GA SA308511496GA SA308511506GA SA308511523GA SA308511529GA SA308511532GA SA308511550GA SA308511639GA SA308511641GA SA308511693GA SA308511695GA SA308511696GA SA308511697GA SA308511698GA SA308511699GA SA308511702GA SA308511703GA SA308511704GA SA308511705GA SA308511708GA SA308511710GA SA308511711GA SA308511715GA SA308511717GA SA308511719GA SA308511723GA SA308511724GA SA308511725GA SA308511727GA SA308511728GA SA308521780GA SA308521781GA SA308521800GA SA308521807GA SA308511453GA SA308511460GA SA308511462GA SA308511471GA SA308511484GA SA308511488GA SA308511500GA SA308511509GA SA308511514GA SA308511542GA SA308511547GA SA308511548GA SA308511552GA SA308511629GA SA308511633GA SA308511634GA SA308511636GA SA308511638GA SA308511644GA SA308511645GA SA308521782GA SA308511455GA SA308511640GA SA308521776GA SA308521777GA SA308521779GA SA308521804GA SA308155978GA SA308429329GA SA308489957GA SA308511465GA SA308511466GA SA308511467GA SA308511487GA SA308511489GA SA308511501GA SA308511510GA SA308511513GA SA308511646GA SA308511687GA SA308511757GA SA308511764GA SA308521774GA SA309011869GA SA309011870GA SA309011871GA SA309011872GA SA309011873GA SA309011876GA SA309011877GA SA309011878GA SA309011881GA SA309011883GA SA308511452GA SA308511474GA SA308511483GA SA308511522GA SA308511537GA SA308511643GA SA308511647GA SA308511709GA SA308511771GA SA308521773GA SA308521775GA SA309011875GA SA308511521GA SA308511540GA SA308286930GA SA308501318GA SA308511561GA SA308511665GA SA308511746GA SA308511754GA SA308511763GA SA308521778GA SA308521799GA SA308521812GA SA308521813GA SA308521814GA SA308521815GA SA308521817GA SA308521818GA SA308521819GA SA308521820GA SA308521821GA SA308521822GA SA308521824GA SA308521826GA SA308521828GA SA308521830GA SA309011879GA SA309021929GA SA309021931GA SA309021934GA SA309021935GA SA309021940GA SA309021941GA SA309021942GA SA309021943GA SA309021944GA SA309021945GA SA309021951GA SA308429352GA SA308459840GA SA308511468GA SA308511479GA SA308511481GA SA308511503GA SA308511511GA SA308511520GA SA308511531GA SA308511545GA SA308511551GA SA308521831GA SA309021932GA SA309021946GA SA309021956GA SA309021959GA SA309021960GA SA309021967GA SA309021968GA SA309021969GA SA309021970GA SA309021971GA SA309021972GA SA309021973GA SA309031989GA SA309031990GA SA309031991GA SA309031992GA SA309031993GA SA309031995GA SA309031997GA SA309031998GA SA309031999GA SA309032001GA SA309032002GA SA309032003GA SA309032004GA SA309032005GA SA309032006GA SA309032008GA SA309032022GA SA309032024GA SA309032027GA SA309032028GA SA309032029GA SA309032030GA SA309032032GA SA309032033GA SA309032037GA SA309032041GA SA309032042GA SA309032043GA SA309032048GA SA309032050GA SA309032052GA SA308511541GA SA308511620GA SA308511680GA SA308511684GA SA308511685GA SA308511750GA SA308511752GA SA308511753GA SA308511755GA SA308511765GA SA308511769GA SA308521795GA SA308521796GA SA308521801GA SA308521802GA SA308521803GA SA308521805GA SA308521823GA SA308521825GA SA308521829GA SA309011834GA SA309011835GA SA309011836GA SA309011837GA SA309011838GA SA309011839GA SA309011840GA SA309011841GA SA309011845GA SA309021926GA SA309021927GA SA309021930GA SA309021947GA SA309021950GA SA309021952GA SA309021954GA SA309021955GA SA309021958GA SA309021961GA SA309021962GA SA309021963GA SA309021964GA SA309021965GA SA309021966GA SA309011844GA SA309032083GA SA309011858GA SA309011860GA SA309011861GA SA309011886GA SA309011889GA SA309011897GA SA309011900GA SA309011903GA SA309021918GA SA309021920GA SA309021922GA SA308337605GA SA308429395GA SA308491092GA SA308501301GA SA308501365GA SA308511666GA SA308511669GA SA308521827GA SA308521832GA SA308521833GA SA309011842GA SA309011846GA SA309011849GA SA309011850GA SA309011851GA SA309011852GA SA309011853GA SA309011854GA SA309011855GA SA309011856GA SA309011857GA SA309011859GA SA309011862GA SA309011863GA SA309011864GA SA309011865GA SA309011866GA SA309011884GA SA309011885GA SA309011887GA SA309011888GA SA309011890GA SA309011891GA SA309011892GA SA309011894GA SA309011895GA SA309011896GA SA309011898GA SA309011899GA SA309011901GA SA309011902GA SA309021917GA SA309021921GA SA309021923GA SA309032087GA SA309021974GA SA309021986GA SA308511458GA SA308511459GA SA308511469GA SA308511473GA SA308511475GA SA308511538GA SA308511628GA SA308511690GA SA309021912GA SA309032092GA SA309032105GA SA309032106GA SA309032108GA SA309032110GA SA309032113GA SA309042116GA SA309042118GA SA309042122GA SA309042123GA SA309042150GA SA309042169GA SA309042170GA SA309042173GA SA308491168GA SA308511539GA SA308511631GA SA309011843GA SA309032010GA SA309032015GA SA309032018GA SA309032019GA SA309032020GA SA309042208GA SA309032066GA SA309042213GA SA309042214GA SA309032070GA SA309032071GA SA309032072GA SA309042131GA SA309021949GA SA309032000GA SA309032007GA SA309032025GA SA309032026GA SA309032031GA SA309032034GA SA309032039GA SA309032040GA SA309032078GA SA309032080GA SA309032086GA SA309032091GA SA309042133GA SA309042195GA SA309042201GA SA309042207GA SA309062266GA SA309062269GA SA309062277GA SA309032075GA SA309032082GA SA309042127GA SA309042128GA SA309032073GA SA309032079GA SA309032081GA SA309062292GA SA309032076GA SA309032077GA SA309032074GA SA309052229GA SA309062274GA SA309021981GA SA309032064GA SA309052220GA SA309052222GA SA309052224GA SA309052225GA SA309052227GA SA309052228GA SA309052236GA SA309052239GA SA309052232GA SA309052233GA SA309032014GA SA309042186GA SA309042189GA SA309062267GA SA309062271GA SA309062273GA SA309062278GA SA309062279GA SA309062281GA SA309062282GA SA309062283GA SA309062284GA SA309062285GA SA309062298GA SA309062299GA SA309062302GA SA309062303GA SA309062312GA SA309092496GA SA309042145GA SA309042180GA SA309042183GA SA309042184GA SA309042185GA SA309042187GA SA309052216GA SA309052219GA SA309072348GA SA309072352GA SA309072362GA SA309072363GA SA309072366GA SA309072383GA SA309072384GA SA309072385GA SA309092500GA SA309092502GA SA309092527GA SA309092531GA SA309092532GA SA309062325GA SA309062330GA SA309062336GA SA309072351GA SA309072356GA SA309072368GA SA309072375GA SA309072387GA SA309082443GA SA309082449GA SA309082452GA SA309082454GA SA309082456GA SA309082457GA SA309092485GA SA309092487GA SA309092489GA SA309092491GA SA309092492GA SA309092493GA SA309082429GA SA309092503GA SA309092505GA SA309092506GA SA309092507GA SA309092508GA SA309092510GA SA309092511GA SA309092512GA SA309092513GA SA309092514GA SA309092515GA SA309092517GA SA309092518GA SA309092519GA SA309092520GA SA309092521GA SA309092522GA SA309092524GA SA309092525GA SA309092526GA SA309092528GA SA309092529GA SA309092530GA SA309092533GA SA309092534GA SA309092535GA SA309092536GA SA309092537GA SA309092538GA SA309092539GA SA309092540GA SA309092541GA SA309092542GA SA309092543GA SA309112594GA SA309112595GA SA309112596GA SA309112605GA SA309072343GA SA309072344GA SA309072347GA SA309072350GA SA309072355GA SA309072357GA SA309072364GA SA309072365GA SA309072370GA SA309072372GA SA309072374GA SA309072377GA SA309072379GA SA309072380GA SA309072382GA SA309072391GA SA309092494GA SA309092509GA SA309112566GA SA309112569GA SA309112571GA SA309112572GA SA309112574GA SA309112576GA SA309112577GA SA309112578GA SA309112579GA SA309112580GA SA309112582GA SA309112583GA SA309112584GA SA309112585GA SA309112586GA SA309112587GA SA309112588GA SA309112589GA SA309112590GA SA309112591GA SA309112592GA SA309112597GA SA309112598GA SA309112599GA SA309112600GA SA309112602GA SA309112603GA SA309112604GA SA309112606GA SA309112607GA SA309112608GA SA309112611GA SA309112613GA SA309112614GA SA309112615GA SA309112616GA SA309112617GA SA309112618GA SA309122666GA SA309122687GA SA309122690GA SA309122691GA SA309062296GA SA309062300GA SA309062304GA SA309062305GA SA309062308GA SA309062309GA SA309062310GA SA309062313GA SA309062315GA SA309062318GA SA309082462GA SA309092474GA SA309092477GA SA309092478GA SA309092479GA SA309092480GA SA309092481GA SA309092482GA SA309092483GA SA309092484GA SA309092486GA SA309092488GA SA309092490GA SA309112544GA SA309112545GA
SA309112546GA SA309112547GA SA309112548GA SA309112549GA SA309112550GA SA309112551GA SA309112552GA SA309112553GA SA309112554GA SA309112555GA SA309112556GA SA309112557GA SA309112558GA SA309112559GA SA309112560GA SA309112562GA SA309112563GA SA309112565GA SA309112567GA SA309112568GA SA309112573GA SA309112581GA SA309112593GA SA309122619GA SA309122645GA SA309122646GA SA309122647GA SA309122648GA SA309122649GA SA309122650GA SA309122652GA SA309122653GA SA309122655GA SA309122656GA SA309122657GA SA309122658GA SA309122659GA SA309122661GA SA309122663GA SA309122665GA SA309122667GA SA309122669GA SA309122670GA SA309122671GA SA308459786GA SA309011848GA SA309021904GA SA309021905GA SA309021906GA SA309021907GA SA309021908GA SA309021909GA SA309021910GA SA309021911GA SA309021913GA SA309021914GA SA309021915GA SA309021916GA SA309021919GA SA309021976GA SA309021977GA SA309021978GA SA309021979GA SA309021980GA SA309021982GA SA309021983GA SA309021984GA SA309021985GA SA309021988GA SA309122620GA SA309122623GA SA309122627GA SA309122628GA SA309122629GA SA309122630GA SA309122632GA SA309122633GA SA309122634GA SA309122635GA SA309122638GA SA309122639GA SA309122640GA SA309122674GA SA309122676GA SA309122677GA SA309122678GA SA309122681GA SA309122683GA SA309122685GA SA309122692GA SA309122696GA SA309122697GA SA309122699GA SA307342313GR SA307362391GR SA307362412GR SA307362413GR SA307382448GR SA307382454GR SA307382461GR SA307382476GR SA307382490GR SA307382506GR SA307382600GR SA307453417GR SA309122664GA SA309122672GA SA309122673GA SA309122680GA SA309122682GA SA309122686GA SA309122689GA SA309122694GA SA309092523GA SA309112609GA SA309112610GA SA309112612GA SA309122621GA SA309122622GA SA309122624GA SA309122625GA SA309122626GA SA309122643GA SA309122721GA SA309122724GA SA309122726GA SA309122733GA SA308511512GA SA309122695GA SA309122712GA SA309122713GA SA309122715GA SA309122719GA SA309122720GA SA309122729GA SA309122730GA SA309122732GA SA309122735GA SA309082423GA SA309132858GA SA309132859GA SA309132860GA SA309132862GA SA309132863GA SA309132866GA SA309132867GA SA309132869GA SA307322146GR SA309122636GA SA309122688GA SA309122717GA SA309122782GA SA309122773GA SA309122779GA SA309122766GA SA309132856GA SA309132868GA SA309142890GA SA309122644GA SA309122775GA SA309122770GA SA309122774GA SA309122769GA SA309122776GA SA309122752GA SA309122764GA SA309122765GA SA309122771GA SA309122778GA SA309122784GA SA309122791GA SA309132802GA SA309132794GA SA309132795GA SA309132808GA SA309132811GA SA309142893GA SA309132792GA SA309132798GA SA307020391GR SA307080462GR SA307080463GR SA307190775GR SA307221051GR SA307231274GR SA307241618GR SA307312108GR SA307322165GR SA307392829GR SA307392864GR SA307423156GR SA307453476GR SA307493885GR SA307493989GR SA308074559GR SA309132796GA SA309132799GA SA309132804GA SA309132807GA SA309142872GA SA309142876GA SA309142879GA SA309142881GA SA309142884GA SA309142896GA SA309142914GA SA309142915GA SA309132810GA SA309062286GA SA309132797GA SA309132813GA SA309132826GA SA309132831GA SA309132832GA SA309132833GA SA309132835GA SA309132836GA SA309122631GA SA309122684GA SA309132861GA SA309142882GA SA309142885GA SA309142886GA SA309142887GA SA309142888GA SA309142891GA SA309142894GA SA309142910GA SA309142917GA SA309142918GA SA309142926GA SA309122654GA SA309122662GA SA309122727GA SA309122757GA SA309132815GA SA309132816GA SA309132819GA SA309132824GA SA309132825GA SA309132834GA SA309132844GA SA309132846GA SA309142873GA SA309142875GA SA309142878GA SA309142911GA SA309142912GA SA309142913GA SA309142929GA SA309142930GA SA309142931GA SA309142934GA SA309142939GA SA309142940GA SA309142941GA SA309142942GA SA309142951GA SA309142953GA SA309142962GA SA309153029GA SA309153035GA SA309153045GA SA309153055GA SA309153056GA SA309153058GA SA309153059GA SA309153061GA SA309153064GA SA309153065GA SA309153066GA SA309153067GA SA309153068GA SA309153069GA SA309153075GA SA309153079GA SA309153081GA SA309153082GA SA309153083GA SA309153086GA SA309153087GA SA309153088GA SA309173100GA SA309173105GA SA309233579GA SA309142956GA SA309142969GA SA309142973GA SA309142980GA SA309183225GA SA309193302GA SA309233561GA SA309233563GA SA309233564GA SA309233565GA SA309233566GA SA309233567GA SA309233568GA SA309233569GA SA309233571GA SA309233572GA SA309233573GA SA309233574GA SA309233575GA SA309233576GA SA309233577GA SA309233578GA SA309233580GA SA309233582GA SA309233583GA SA309233584GA SA309233585GA SA309233586GA SA309233587GA SA309233588GA SA309233589GA SA309233590GA SA309233591GA SA309243593GA SA309243596GA SA309243597GA SA309243601GA SA309243602GA SA309243603GA SA309243604GA SA309243605GA SA309243606GA SA309243607GA SA309243608GA SA309243609GA SA309243610GA SA309243611GA SA309243612GA SA309243613GA SA309243614GA SA309243615GA SA309243616GA SA309243617GA SA309243618GA SA309243619GA SA309243620GA SA309243622GA SA309243624GA SA309243626GA SA309243628GA SA309243629GA SA309243630GA SA309243631GA SA309142903GA SA309142919GA SA309142928GA SA309142959GA SA309142978GA SA309173180GA SA309193300GA SA309243634GA SA309183198GA SA309183208GA SA309183214GA SA309183224GA SA309183229GA SA309132793GA SA309132818GA SA309132820GA SA309132821GA SA309132827GA SA309132828GA SA309132829GA SA309132830GA SA309132837GA SA309132838GA SA309132839GA SA309142883GA SA309142958GA SA309142961GA SA309173103GA SA309173104GA SA309173107GA SA309183195GA SA309183215GA SA309183217GA SA309183218GA SA309183219GA SA309183220GA SA309183223GA SA309193267GA SA309193280GA SA309243594GA SA309243599GA SA309243600GA SA309243621GA SA309173101GA SA309173110GA SA309173111GA SA309153084GA SA309173117GA SA309173126GA SA309173127GA SA309173128GA SA309173140GA SA309173145GA SA309173146GA SA309173147GA SA309173149GA SA309173155GA SA309173157GA SA309173160GA SA309173163GA SA309173165GA SA309173170GA SA309173171GA SA309173174GA SA309173175GA SA309173178GA SA309173184GA SA309223503GA SA309223508GA SA309223516GA SA309173133GA SA309233517GA SA309233520GA SA309233523GA SA309233526GA SA309233527GA SA309233529GA SA309233533GA SA309142946GA SA309142955GA SA309142994GA SA309183236GA SA309142904GA SA309142932GA SA309142944GA SA309142947GA SA309142948GA SA309142950GA SA309142960GA SA309142964GA SA309142966GA SA309142974GA SA309142993GA SA309142995GA SA309142996GA SA309142998GA SA309143001GA SA309143002GA SA309143005GA SA309143007GA SA309143009GA SA309143017GA SA309143022GA SA309183193GA SA309183194GA SA309183205GA SA309183206GA SA309183213GA SA309183221GA SA309183231GA SA309183232GA SA309183239GA SA309183241GA SA309183244GA SA309193254GA SA309193268GA SA309193276GA SA309203379GA SA309203388GA SA309203394GA SA309203405GA SA309203409GA SA309203425GA SA309203428GA SA309203429GA SA309203433GA SA309203434GA SA309223485GA SA309243627GA SA309243695GA SA309142943GA SA309142945GA SA309142971GA SA309143008GA SA309183190GA SA309183191GA SA309183228GA SA309183243GA SA309193322GA SA309203390GA SA309203407GA SA309203427GA SA309223456GA SA309223458GA SA309223462GA SA309223466GA SA309223471GA SA309223482GA SA309233518GA SA309233522GA SA309233525GA SA309233528GA SA309233530GA SA309233531GA SA309233534GA SA309233536GA SA309233539GA SA309233547GA SA309233550GA SA309243595GA SA309243690GA SA308439469GA SA309142936GA SA309142938GA SA309142954GA SA309142957GA SA309142967GA SA309142970GA SA309142972GA SA309142975GA SA309142976GA SA309142977GA SA309142979GA SA309142997GA SA309142999GA SA309143000GA SA309143003GA SA309143004GA SA309143006GA SA309143011GA SA309143014GA SA309143015GA SA309143018GA SA309143021GA SA309143023GA SA309153091GA SA309153096GA SA309183227GA SA309183230GA SA309183233GA SA309183235GA SA309183238GA SA309183245GA SA309183246GA SA309183252GA SA309193255GA SA309193258GA SA309193259GA SA309193262GA SA309193263GA SA309193264GA SA309193285GA SA309193292GA SA309193306GA SA309193308GA SA309193309GA SA309193310GA SA309193311GA SA309193315GA SA309203364GA SA309203376GA SA309203384GA SA309203391GA SA309203392GA SA309203396GA SA309203403GA SA309203406GA SA309203410GA SA309203412GA SA309203417GA SA309203424GA SA309203430GA SA309223435GA SA309223443GA SA309223445GA SA309223449GA SA309223450GA SA309223455GA SA309223469GA SA309223472GA SA309223481GA SA309233524GA SA309243670GA SA309243673GA SA309243674GA SA309243675GA SA309243700GA SA309243701GA SA309243710GA SA309253714GA SA309253716GA SA309253717GA SA309253718GA SA309253719GA SA309253720GA SA309253723GA SA309253726GA SA309253728GA SA309253766GA SA309253767GA SA309253771GA SA309253772GA SA309253773GA SA309253774GA SA309253775GA SA309253777GA SA309253778GA SA309253790GA SA309253792GA SA309253837GA SA309253839GA SA309253845GA SA309253846GA SA309253863GA SA307020383GR SA307070453GR SA307080520GR SA307120588GR SA307221105GR SA307221185GR SA307221188GR SA307241673GR SA307342317GR SA307382471GR SA307382569GR SA308206738GR SA308327336GA SA308337638GA SA309153099GA SA309183248GA SA309193273GA SA309193281GA SA309193282GA SA309193283GA SA309193286GA SA309193287GA SA309193288GA SA309193289GA SA309193290GA SA309193291GA SA309193293GA SA309193295GA SA309193314GA SA309193318GA SA309193329GA SA309193331GA SA309193336GA SA309193337GA SA309193346GA SA309193352GA SA309193356GA SA309193357GA SA309193362GA SA309203421GA SA309243625GA SA309243681GA SA309243698GA SA309253730GA SA309253732GA SA309253782GA SA309253784GA SA309253833GA SA309253834GA SA309253848GA SA309253849GA SA309253850GA SA309253852GA SA309253855GA SA309253856GA SA309253861GA SA309253866GA SA309253867GA SA309253869GA SA309253871GA SA309253873GA SA309253874GA SA309253878GA SA309253880GA SA309253881GA SA309253882GA SA309253885GA SA309253886GA SA309253890GA SA309253891GA SA309253894GA SA309253895GA SA309253896GA SA309253898GA SA309253899GA SA309253900GA SA309253908GA SA309253911GA SA308317300GA SA308327320GA SA308327550GA SA308347746GA SA308347747GA SA308347750GA SA309143010GA SA309143012GA SA309143013GA SA309143016GA SA309143019GA SA309143020GA SA309153090GA SA309153092GA SA309153093GA SA309153094GA SA309153095GA SA309153097GA SA309153098GA SA309173102GA SA309183192GA SA309183197GA SA309183211GA SA309183222GA SA309183226GA SA309183234GA SA309183247GA SA309183251GA SA309193253GA SA309193256GA SA309193260GA SA309193265GA SA309193269GA SA309193272GA SA309193278GA SA309233562GA SA309233570GA SA309233581GA SA309243592GA SA309263918GA SA309263919GA SA307110559GR SA307190769GR SA307382478GR SA307443262GR SA307453428GR SA307453599GR SA308084784GR SA308135527GR SA308196568GR SA308206709GR SA309263946GA SA309263951GA SA309263958GA SA309283983GA SA309284011GA SA309283983GA SA309284011GA SA309263917GA SA309253909GA SA309263943GA SA309263945GA SA309263948GA SA309263949GA SA308327313GA SA309263940GA SA309263935GA SA309263952GA SA309263953GA SA309263954GA SA309243648GA SA309243658GA SA309243663GA SA309243665GA SA309243667GA SA309243668GA SA309243671GA SA309243692GA SA309243694GA SA309243696GA SA309243703GA SA309243706GA SA309243708GA SA309253731GA SA309253743GA SA309253753GA SA309253755GA SA309253756GA SA309253757GA SA309253760GA SA309253761GA SA309253763GA SA309253765GA SA309253781GA SA309253807GA SA309253812GA SA309253887GA SA309253906GA SA309263923GA SA309263924GA SA309263925GA SA309263929GA SA309263930GA SA309263933GA SA309263937GA SA309263938GA SA309263939GA SA309263941GA SA309263957GA SA309294028GA
SA309294029GA SA309294047GA SA309294050GA SA309294051GA SA309294053GA SA309294055GA SA309294057GA SA309294058GA SA309294060GA SA309294061GA SA309294065GA SA309294067GA SA309294068GA SA309294071GA SA309294089GA SA309294090GA SA309223489GA SA309294030GA SA309294035GA SA309294037GA SA309294044GA SA309294048GA SA309294064GA SA309294077GA SA309294079GA SA309294084GA SA309294085GA SA309294087GA SA309294092GA SA309294093GA SA309294103GA SA309294112GA SA309294113GA SA309253744GA SA309304234GA SA309304240GA SA309304245GA SA307221058GR SA307241559GR SA307332280GR SA307392989GR SA307403060GR SA307403140GR SA307413106GR SA308074537GR SA308074551GR SA308074569GR SA308074593GR SA308074596GR SA308094850GR SA308105101GR SA308135486GR SA308135615GR SA308135618GR SA308145776GR SA309283989GA SA309283992GA SA309283995GA SA309304220GA SA309324303GA SA309284004GA SA309304280GA SA309284007GA SA309304229GA SA309304248GA SA309324342GA SA309324395GA SA309324400GA SA309324411GA SA309324417GA SA309183201GA SA309263914GA SA309283976GA SA309283977GA SA309284001GA SA309284005GA SA309284012GA SA309304284GA SA309324481GA SA309324482GA SA309324519GA SA309324520GA SA309324527GA SA309324532GA SA309324533GA SA309324534GA SA309324537GA SA309324544GA SA309324545GA SA309324546GA SA309324547GA SA309324551GA SA309334756GA SA309334760GA SA309334767GA SA309334768GA SA309334776GA SA309334777GA SA309334779GA SA309334780GA SA309334781GA SA309334782GA SA309334797GA SA309334800GA SA309334801GA SA309334802GA SA309334803GA SA309334804GA SA309334807GA SA309334810GA SA309334811GA SA309334831GA SA309304218GA SA309324355GA SA309324358GA SA309324359GA SA309324361GA SA309324362GA SA309324363GA SA309324364GA SA309324366GA SA309324367GA SA309324370GA SA309324372GA SA309324374GA SA309324378GA SA309324379GA SA309324381GA SA309324388GA SA309324390GA SA309324397GA SA309324398GA SA309324403GA SA309324405GA SA309324406GA SA309324407GA SA309324419GA SA309324420GA SA309324421GA SA309324422GA SA309324423GA SA309324435GA SA309324436GA SA309334669GA SA309253840GA SA309283999GA SA309324317GA SA309324321GA SA309324413GA SA309324425GA SA309324437GA SA309324438GA SA309324442GA SA309324443GA SA309324445GA SA309324515GA SA309324516GA SA309324538GA SA309324542GA SA309324543GA SA309334557GA SA309334561GA SA309334562GA SA309334563GA SA309334564GA SA309334566GA SA309334569GA SA309334575GA SA309334577GA SA309334583GA SA309334587GA SA309334588GA SA309334591GA SA309334592GA SA309334593GA SA309334596GA SA309334597GA SA309334598GA SA309334604GA SA309334650GA SA309334763GA SA309334766GA SA309334770GA SA309334774GA SA309334799GA SA309334830GA SA309304283GA SA309334558GA SA309334706GA SA309324539GA SA309324541GA SA309324549GA SA309334571GA SA309334574GA SA309334579GA SA309334642GA SA309334652GA SA309334664GA SA309334665GA SA309334701GA SA309334705GA SA309334713GA SA309334721GA SA309334747GA SA309334753GA SA309334764GA SA309324394GA SA309334560GA SA309193312GA SA309193339GA SA309193340GA SA309193343GA SA309203393GA SA309243669GA SA309243677GA SA309243678GA SA309243679GA SA309243680GA SA309243682GA SA309243684GA SA309243685GA SA309243686GA SA309243687GA SA309243689GA SA309243693GA SA309243697GA SA309304161GA SA309324424GA SA309334581GA SA309334582GA SA309334590GA SA309334602GA SA309334612GA SA309334639GA SA309334681GA SA309334699GA SA309344850GA SA309344857GA SA309344862GA SA309344869GA SA309354896GA SA309354919GA SA309354945GA SA309354972GA SA309365110GA SA309365171GA SA309365173GA SA309365214GA SA307190734GR SA307190771GR SA307231479GR SA307453438GR SA307453557GR SA307453588GR SA307453597GR SA307453602GR SA308196560GR SA309253713GA SA309253715GA SA309253722GA SA309253750GA SA309334573GA SA309334576GA SA309334586GA SA309334722GA SA309334728GA SA309334732GA SA309334734GA SA309334735GA SA309334739GA SA309334771GA SA309334772GA SA309344832GA SA309344833GA SA309344836GA SA309344837GA SA309344852GA SA309344853GA SA309344854GA SA309344855GA SA309354913GA SA309354982GA SA309365113GA SA309365118GA SA309365134GA SA309365141GA SA309365150GA SA309365154GA SA309365157GA SA309365158GA SA309365179GA SA309365193GA SA309365198GA SA309365206GA SA309365217GA SA309365221GA SA309344874GA SA309344875GA SA309344877GA SA309344879GA SA309354884GA SA309354888GA SA309354889GA SA309354891GA SA309354893GA SA309354897GA SA309354905GA SA309354910GA SA309354924GA SA309354928GA SA309354948GA SA309354952GA SA309354953GA SA309354955GA SA309354964GA SA309354965GA SA309354968GA SA309354976GA SA309354987GA SA309354988GA SA309354993GA SA309355000GA SA309355003GA SA309355004GA SA309355006GA SA309365076GA SA309334638GA SA309344872GA SA309354933GA SA309354949GA SA309354956GA SA309354957GA SA309354959GA SA309354974GA SA309354980GA SA309354985GA SA309354989GA SA309354995GA SA309355007GA SA309344871GA SA309344873GA SA309344876GA SA309344878GA SA309344880GA SA309354882GA SA309354885GA SA309354886GA SA309354904GA SA309354909GA SA309354914GA SA309354922GA SA309354936GA SA309354939GA SA309354941GA SA309354950GA SA309354962GA SA309354963GA SA309354969GA SA309354977GA SA309354986GA SA309355002GA SA309354990GA SA309354998GA SA309334814GA SA309334816GA SA309334826GA SA309365203GA SA309284008GA SA309304179GA SA309334675GA SA309334757GA SA309334758GA SA309334762GA SA309334783GA SA309334784GA SA309334794GA SA309334795GA SA309334815GA SA309334823GA SA309334828GA SA309354991GA SA309355008GA SA309355010GA SA309355012GA SA309355013GA SA309355014GA SA309355016GA SA309355017GA SA309365029GA SA309365039GA SA309365047GA SA309365055GA SA309365059GA SA309365064GA SA309365075GA SA309365084GA SA309365091GA SA309365095GA SA309365098GA SA309365099GA SA309365105GA SA309365108GA SA309365111GA SA309365112GA SA309365122GA SA309365131GA SA309365147GA SA309365151GA SA309365155GA SA309365163GA SA309365174GA SA309365189GA SA309365204GA SA309365216GA SA309365294GA SA309365305GA SA309365306GA SA309365307GA SA309365319GA SA309365324GA SA309365341GA SA309365342GA SA309365346GA SA309365349GA SA309365355GA SA309365361GA SA309375379GA SA309375420GA SA309375435GA SA309375454GA SA309375456GA SA309375457GA SA309375460GA SA309375461GA SA309375463GA SA309375465GA SA309375466GA SA309375468GA SA309375469GA SA309375471GA SA309375472GA SA309375473GA SA309375475GA SA309375482GA SA309375485GA SA309375506GA SA309385528GA SA309334813GA SA309334829GA SA309354970GA SA309354971GA SA309354979GA SA309355019GA SA309355022GA SA309355025GA SA309355026GA SA309355027GA SA309365028GA SA309365033GA SA309365054GA SA309365060GA SA309365062GA SA309365067GA SA309365068GA SA309365078GA SA309365079GA SA309365082GA SA309365086GA SA309365241GA SA309365250GA SA309365251GA SA309365257GA SA309365270GA SA309365332GA SA309365335GA SA309375369GA SA309344864GA SA309355024GA SA309365032GA SA309365034GA SA309365036GA SA309365037GA SA309365038GA SA309365041GA SA309365042GA SA309365043GA SA309365046GA SA309365050GA SA309365051GA SA309365053GA SA309365056GA SA309365057GA SA309365069GA SA309365071GA SA309365073GA SA309365074GA SA309365077GA SA309365240GA SA309365245GA SA309365248GA SA309365249GA SA309365268GA SA309365272GA SA309365273GA SA309365274GA SA309365275GA SA309375428GA SA309365340GA SA309375367GA SA309375377GA SA307020379GR SA307020386GR SA307030405GR SA307030412GR SA307070442GR SA307070447GR SA307070449GR SA307070454GR SA307080478GR SA307080504GR SA307080525GR SA307080526GR SA307110572GR SA307110575GR SA307150681GR SA307180694GR SA307180696GR SA307190717GR SA307190729GR SA307190732GR SA307190736GR SA307190738GR SA307190743GR SA307190744GR SA307190749GR SA307190750GR SA307190764GR SA307190770GR SA307220990GR SA307231262GR SA307231463GR SA307241676GR SA307342312GR SA307362408GR SA307382451GR SA307382465GR SA307443361GR SA307443371GR SA307443386GR SA307453400GR SA307453420GR SA307453430GR SA307453600GR SA307453613GR SA307453614GR SA307453616GR SA308135518GR SA308196584GR SA309334621GA SA309334700GA SA309344834GA SA309344859GA SA309355015GA SA309365061GA SA309365063GA SA309365065GA SA309365260GA SA309365261GA SA309365262GA SA309365263GA SA309365264GA SA309365266GA SA309365279GA SA309365280GA SA309365283GA SA309365285GA SA309365287GA SA309365288GA SA309365289GA SA309365291GA SA309365292GA SA309365295GA SA309365296GA SA309365297GA SA309365299GA SA309365301GA SA309365302GA SA309365310GA SA309365312GA SA309375424GA SA309375434GA SA309334785GA SA309334789GA SA309334790GA SA309334793GA SA309354978GA SA309355020GA SA309365045GA SA309365072GA SA309365093GA SA309365094GA SA309365123GA SA309365129GA SA309365130GA SA309365168GA SA309365187GA SA309365199GA SA309365205GA SA309365207GA SA309365215GA SA309365243GA SA309365276GA SA309365293GA SA309365315GA SA309365316GA SA309365321GA SA309365322GA SA309365325GA SA309365327GA SA309365331GA SA309365333GA SA309365336GA SA309365337GA SA309365338GA SA309365343GA SA309365350GA SA309365351GA SA309365352GA SA309365353GA SA309365356GA SA309365357GA SA309365358GA SA309365359GA SA309375368GA SA309375371GA SA309375372GA SA309375373GA SA309375375GA SA309375380GA SA309375381GA SA309375483GA SA309375492GA SA309375507GA SA309385544GA SA309385546GA SA309385547GA SA309385552GA SA309385553GA SA309385556GA SA309385567GA SA309385579GA SA309385605GA SA309385607GA SA309385608GA SA309385640GA SA309193257GA SA309193277GA SA309263921GA SA309263969GA SA309283978GA SA309284009GA SA309324540GA SA309334759GA SA309334786GA SA309354942GA SA309365092GA SA309365120GA SA309365143GA SA309365162GA SA309365165GA SA309365166GA SA309365167GA SA309365180GA SA309365182GA SA309365185GA SA309365197GA SA309365277GA SA309365278GA SA309365286GA SA309365298GA SA309365300GA SA309365303GA SA309365304GA SA309365308GA SA309365309GA SA309365311GA SA309365344GA SA309365354GA SA309365362GA SA309375363GA SA309375366GA SA309375370GA SA309375378GA SA309375383GA SA309375384GA SA309375418GA SA309375419GA SA309375422GA SA309375423GA SA309375426GA SA309375427GA SA309375429GA SA309375430GA SA309375432GA SA309375433GA SA309375436GA SA309375440GA SA309375442GA SA309375443GA SA309375444GA SA309375447GA SA309375448GA SA309375449GA SA309375451GA SA309375455GA SA309375458GA SA309375459GA SA309375464GA SA309375467GA SA309375470GA SA309375494GA SA309385524GA SA309385538GA SA309385574GA SA309385584GA SA309385627GA SA309385628GA SA309385637GA SA309385510GA SA309385554GA SA307514357GR SA309354883GA SA309354892GA SA309354898GA SA309354900GA SA309354903GA SA309354906GA SA309354907GA SA309354961GA SA309354966GA SA309355023GA SA309365083GA SA309365239GA SA309365252GA SA309365254GA SA309365258GA SA309375425GA SA309375452GA SA309375453GA SA309375477GA SA309375488GA SA309375495GA SA309375502GA SA309385550GA SA308511559GR SA309283997GA SA309294069GA SA309294116GA SA309304201GA SA309304244GA SA309304262GA SA309304272GA SA309324548GA SA309334809GA SA309344863GA SA309344865GA SA309354916GA SA309354926GA SA309354967GA SA309365145GA SA309365184GA SA309365223GA SA309365224GA SA309365225GA SA309365226GA SA309365227GA SA309365228GA SA309365229GA SA309365230GA SA309365231GA SA309365232GA SA309365233GA SA309365234GA SA309365235GA SA309365236GA SA309365237GA SA309365271GA SA309365347GA SA309375364GA SA309375406GA SA309375441GA SA309385561GA SA309395705GA SA309395710GA SA309395717GA SA309395746GA SA309395754GA SA309395755GA SA309395788GA SA309395792GA
SA309395794GA SA309405821GA SA309405825GA SA309405830GA SA309405845GA SA309405851GA SA309193328GA SA309193360GA SA309223442GA SA309223444GA SA309223457GA SA309223465GA SA309223468GA SA309223470GA SA309223476GA SA309223484GA SA309243688GA SA309243691GA SA308074575GR SA309365246GA SA309365348GA SA309385514GA SA309395777GA SA309395793GA SA309395812GA SA309405827GA SA309405832GA SA309405836GA SA309405848GA SA309405865GA SA309405866GA SA309405917GA SA307020399GR SA307080498GR SA307080510GR SA307150671GR SA307180704GR SA307180705GR SA307210839GR SA307241532GR SA307241556GR SA307302066GR SA307382662GR SA307443348GR SA307453462GR SA307463652GR SA308074540GR SA308074542GR SA308074553GR SA308074570GR SA308074571GR SA308074579GR SA308074594GR SA308166130GR SA308266849GR SA308459699GR SA309032014GR SA309233579GR SA309284013GA SA309324470GA SA309324504GA SA309324521GA SA309334806GA SA309365103GA SA309365181GA SA309365183GA SA309365186GA SA309365323GA SA309365345GA SA309375392GA SA309375393GA SA309375394GA SA309375396GA SA309375397GA SA309375399GA SA309375401GA SA309375403GA SA309375407GA SA309375409GA SA309375411GA SA309375414GA SA309375416GA SA309405867GA SA309405871GA SA309405873GA SA309405875GA SA309405876GA SA309405877GA SA309405883GA SA309405884GA SA309405885GA SA309405886GA SA309405887GA SA309405889GA SA309405890GA SA309405891GA SA309405892GA SA309405893GA SA309405894GA SA309405895GA SA309405896GA SA309405897GA SA309405898GA SA309405899GA SA309405900GA SA309405901GA SA309405902GA SA309405903GA SA309405904GA SA309405906GA SA309405912GA SA309405914GA SA309405915GA SA309405918GA SA309405919GA SA309405921GA SA309415922GA SA309415924GA SA309415925GA SA309415926GA SA309415927GA SA309415929GA SA309415931GA SA309415933GA SA309415934GA SA309415935GA SA309415936GA SA309415937GA SA309415938GA SA309415939GA SA309415949GA SA309415951GA SA309415952GA SA309415954GA SA309415955GA SA309415993GA SA309375410GA SA309405868GA SA309405869GA SA309405879GA SA307080493GR SA307200800GR SA308266899GR SA308317256GR SA308327451GR SA308501281GR SA309142893GR SA309415950GA SA309324485GA SA309334778GA SA309375400GA SA309375404GA SA309375412GA SA309405878GA SA309405880GA SA309405881GA SA309405911GA SA309405913GA SA309426102GA SA309426103GA SA309426104GA SA307070438GR SA307080508GR SA307080533GR SA307180702GR SA307180706GR SA307190756GR SA307190757GR SA307251741GR SA307251746GR SA307514373GR SA308266820GR SA308266836GR SA308317066GR SA308327509GR SA308388535GR SA308398891GR SA308398896GR SA308501322GR SA309032016GR SA309365326GA SA309375382GA SA309375474GA SA309405857GA SA309405888GA SA309415959GA SA309426023GA SA309426047GA SA309426049GA SA309426056GA SA309426058GA SA309426132GA SA309426133GA SA309426140GA SA309426141GA SA309426159GA SA307190768GR SA307210836GR SA307231394GR SA308074567GR SA308084671GR SA308084694GR SA308084695GR SA308084708GR SA308084718GR SA308084719GR SA308084720GR SA308266865GR SA308266878GR SA309283991GA SA309365209GA SA309365318GA SA309375395GA SA309375408GA SA309375413GA SA309375415GA SA309405813GA SA309405856GA SA309405859GA SA309405860GA SA309405864GA SA309426070GA SA309426084GA SA309334808GA SA309405858GA SA309405862GA SA309405909GA SA309415953GA SA309415956GA SA309415967GA SA309426026GA SA309426027GA SA309426029GA SA309426041GA SA309426129GA SA309426139GA SA309426144GA SA309426146GA SA309426158GA SA309426160GA SA309426161GA SA309426165GA SA309405834GA SA309365329GA SA309375402GA SA309405837GA SA309405838GA SA309405844GA SA309405852GA SA309415957GA SA309415960GA SA309415961GA SA309415962GA SA309415964GA SA309415966GA SA309415968GA SA309415970GA SA309415987GA SA309415988GA SA309415990GA SA309415991GA SA309415994GA SA309415995GA SA309415996GA SA309415997GA SA309415998GA SA309415999GA SA309416001GA SA309416002GA SA309416004GA SA309416005GA SA309416006GA SA309416007GA SA309416008GA SA309416009GA SA309416011GA SA309416012GA SA309416013GA SA309416014GA SA309416015GA SA309416016GA SA309416017GA SA309416018GA SA309416020GA SA309426025GA SA309426031GA SA309426032GA SA309426034GA SA309426035GA SA309426036GA SA309426042GA SA309426043GA SA309426045GA SA309426046GA SA309426051GA SA309426052GA SA309426054GA SA309426057GA SA309426061GA SA309426063GA SA309426064GA SA309426075GA SA309426077GA SA309426079GA SA309426081GA SA309426083GA SA309426085GA SA309426089GA SA309426090GA SA309426091GA SA309426093GA SA309426096GA SA309426120GA SA309365127GA SA309365178GA SA309385611GA SA309354887GA SA309354937GA SA309436271GA SA309436300GA SA309436305GA SA309436310GA SA309436334GA SA309436355GA SA309426105GA SA309426109GA SA309426110GA SA309426117GA SA309426122GA SA309436269GA SA309436274GA SA309436275GA SA309436280GA SA309436284GA SA309436288GA SA309436289GA SA309436290GA SA309436291GA SA309436298GA SA309436301GA SA309436303GA SA309436308GA SA309436313GA SA309436335GA SA309446426GA SA309446444GA SA309426114GA SA309426116GA SA309426125GA SA309426127GA SA309426149GA SA309426150GA SA309436191GA SA309436196GA SA309436199GA SA309436201GA SA309436206GA SA309436207GA SA309436226GA SA309446433GA SA309456639GA SA309426038GA SA309426148GA SA309446376GA SA309446378GA SA309446380GA SA309446387GA SA309446388GA SA309446411GA SA309456604GA SA309456608GA SA309456611GA SA309456612GA SA309456621GA SA309456630GA SA309456636GA SA309456641GA SA309456642GA SA309456651GA SA309456654GA SA309456658GA SA309456662GA SA309456667GA SA309456668GA SA309456669GA SA309456670GA SA309456671GA SA309456672GA SA309456673GA SA309456674GA SA309456677GA SA309456680GA SA309456681GA SA309456683GA SA309456684GA SA309456685GA SA309456687GA SA309456689GA SA309456690GA SA309456691GA SA309456693GA SA309456694GA SA309456696GA SA309456698GA SA309456703GA SA309456704GA SA309456705GA SA309456706GA SA309456707GA SA309456708GA SA309456710GA SA309456714GA SA309456716GA SA309456717GA SA309466801GA SA309466803GA SA309466804GA SA309466805GA SA309466807GA SA309466808GA SA309466809GA SA309466810GA SA309466812GA SA309466813GA SA309466814GA SA309466815GA SA309466816GA SA309466817GA SA309466818GA SA309466819GA SA309466820GA SA309466821GA SA309466822GA SA309466823GA SA309466824GA SA309466825GA SA309466826GA SA309466827GA SA309466828GA SA309466829GA SA309466830GA SA309466832GA SA309466833GA SA309466834GA SA309466835GA SA309466836GA SA309466837GA SA309466840GA SA309466842GA SA309466843GA SA309466844GA SA309466846GA SA309466849GA SA309466850GA SA309466851GA SA309466852GA SA309466853GA SA309466854GA SA309466856GA SA309466857GA SA309466859GA SA309466860GA SA309476868GA SA309476875GA SA309476878GA SA309476886GA SA309476897GA SA309476901GA SA309476902GA SA309476903GA SA309476904GA SA309476906GA SA309476907GA SA309476908GA SA309476910GA SA309476912GA SA309476913GA SA309476914GA SA309476919GA SA309476923GA SA309476928GA SA309476932GA SA309426128GA SA309426131GA SA309426134GA SA309426136GA SA309426152GA SA309426156GA SA309426157GA SA309426162GA SA309426163GA SA309436217GA SA307080473GR SA307110579GR SA307120587GR SA307220920GR SA307241519GR SA307241583GR SA307251715GR SA307251784GR SA307483780GR SA307483832GR SA308074611GR SA308094800GR SA308105176GR SA308105178GR SA308125330GR SA308337656GR SA308368201GR SA308378311GR SA308378326GR SA308388618GR SA308398702GR SA308408915GR SA308439458GR SA308449484GR SA308449500GR SA308449515GR SA308449519GR SA308449531GR SA308449570GR SA308449593GR SA308459846GR SA308459847GR SA308489921GR SA308489954GR SA308491241GR SA308511447GR SA308511459GR SA308511510GR SA308511521GR SA308511545GR SA309062312GR SA309122664GR SA309436187GA SA309436188GA SA309436190GA SA309436192GA SA309436195GA SA309436197GA SA309436202GA SA309436203GA SA309436204GA SA309436205GA SA309436208GA SA309436223GA SA309436311GA SA309436351GA SA309436352GA SA309436354GA SA309436358GA SA309436368GA SA309436371GA SA309426113GA SA309436273GA SA309436293GA SA309436294GA SA309436304GA SA309436306GA SA309436309GA SA309436312GA SA309436314GA SA309436332GA SA309436336GA SA309436338GA SA309436341GA SA309436342GA SA309436344GA SA309436345GA SA309436346GA SA309436347GA SA309436357GA SA309436364GA SA309436365GA SA309436366GA SA309436367GA SA309436369GA SA309436372GA SA309436375GA SA309436320GA SA309436330GA SA309436356GA SA309436359GA SA309446427GA SA309446429GA SA309446430GA SA309446431GA SA309446439GA SA309446447GA SA309446454GA SA309446458GA SA309446477GA SA309446479GA SA309456564GA SA309466741GA SA309476994GA SA309415989GA SA309416003GA SA309426028GA SA309426033GA SA309426055GA SA309436167GA SA309436170GA SA309436179GA SA309436182GA SA309436189GA SA309436198GA SA309436210GA SA309436212GA SA309436215GA SA309436218GA SA309436219GA SA309436220GA SA309436323GA SA309446391GA SA309446395GA SA309446404GA SA309446418GA SA309446425GA SA309446468GA SA309446472GA SA309446487GA SA309446488GA SA309446503GA SA309446505GA SA309446520GA SA309446532GA SA309456540GA SA309456553GA SA309456601GA SA309456626GA SA309466831GA SA309466845GA SA309476874GA SA309476879GA SA309476885GA SA309476894GA SA309477006GA SA309477012GA SA309477022GA SA307241577GR SA308084688GR SA308166026GR SA308449611GR SA309436361GA SA309446469GA SA309476909GA SA309354930GA SA309395741GA SA309436337GA SA309436373GA SA309446440GA SA309446451GA SA309446462GA SA309446467GA SA309446475GA SA309446496GA SA309446502GA SA309446507GA SA309446512GA SA309446523GA SA309446531GA SA309456536GA SA309456542GA SA309456556GA SA309456721GA SA309456727GA SA309476934GA SA309426121GA SA309436270GA SA309436297GA SA309446377GA SA309446382GA SA309446384GA SA309446393GA SA309446397GA SA309446402GA SA309446405GA SA309446406GA SA309446409GA SA309446410GA SA309446414GA SA309446416GA SA309446417GA SA309446422GA SA309446432GA SA309446441GA SA309446459GA SA309456599GA SA309456602GA SA309456605GA SA309456613GA SA309456616GA SA309456617GA SA309456618GA SA309456619GA SA309456620GA SA309456625GA SA309456627GA SA309456628GA SA309456631GA SA309456632GA SA309497301GA SA309426115GA SA309426130GA SA309436176GA SA309436181GA SA309436194GA SA309436224GA SA309446398GA SA309446413GA SA309456606GA SA309456607GA SA309456609GA SA309456629GA SA309304237GA SA309466858GA SA309476918GA SA309477004GA SA309477010GA SA309477031GA SA309477039GA SA309477044GA SA309477069GA SA309477097GA SA309477110GA SA309477111GA SA309477121GA SA309477135GA SA309446465GA SA309446478GA SA309446506GA SA309446517GA SA309446519GA SA309456539GA SA309456541GA SA309456559GA SA309456560GA SA309456644GA SA309456645GA SA309456726GA SA309456732GA SA309426106GA SA309426107GA SA309426111GA SA309426124GA SA309436268GA SA309436277GA SA309436282GA SA309436283GA SA309436287GA SA309436296GA SA309436307GA SA309436325GA SA309436343GA SA309446457GA SA309446498GA SA309446501GA SA309446521GA SA309446524GA SA309253888GA SA309446446GA SA309446526GA SA309446533GA SA309446534GA SA309456557GA SA309456723GA SA309476940GA SA309476947GA SA309476949GA SA309476953GA SA309517464GA SA309517476GA SA309446396GA SA309446415GA SA309446420GA SA309456600GA SA309456603GA SA309456610GA SA309456622GA SA309456624GA SA309477000GA SA309477002GA SA309477025GA SA309477030GA SA309477056GA SA309477061GA SA309477062GA SA309477079GA SA309477080GA SA309477081GA SA309477086GA SA309477087GA SA309477089GA SA309477106GA SA309477107GA SA309477109GA SA309477117GA SA309477122GA SA309477142GA
SA309497158GA SA309466752GA SA309466755GA SA309466756GA SA309466768GA SA309466771GA SA309466791GA SA309466792GA SA309507382GA SA309517467GA SA309517491GA SA309517495GA SA309517514GA SA309517518GA SA309517520GA SA309517534GA SA309476930GA SA309476998GA SA309477016GA SA309477018GA SA309477027GA SA309477034GA SA309477035GA SA309477036GA SA309477040GA SA309477041GA SA309477043GA SA309477047GA SA309477048GA SA309477057GA SA309477065GA SA309477070GA SA309477071GA SA309477072GA SA309477075GA SA309477076GA SA309477078GA SA309477082GA SA309477084GA SA309477088GA SA309477099GA SA309477101GA SA309477104GA SA309477112GA SA309477113GA SA309477115GA SA309477123GA SA309477125GA SA309477128GA SA309477129GA SA309477132GA SA309477134GA SA309477138GA SA309477139GA SA309477140GA SA309477141GA SA309436317GA SA309436326GA SA309436333GA SA309436340GA SA309436349GA SA309436374GA SA309446434GA SA309446435GA SA309446436GA SA309446471GA SA309477124GA SA309507356GA SA309507361GA SA309507366GA SA309507371GA SA309507377GA SA309507383GA SA309507385GA SA309507392GA SA309507396GA SA309507421GA SA309507426GA SA309507430GA SA309507435GA SA309507440GA SA309507445GA SA309507448GA SA309517454GA SA309517458GA SA309517460GA SA309517461GA SA309517465GA SA309517472GA SA309517481GA SA309517488GA SA309517490GA SA309517492GA SA309517497GA SA309517498GA SA309517499GA SA309517500GA SA309517502GA SA309517503GA SA309517504GA SA309517506GA SA309517507GA SA309517508GA SA309517510GA SA309517515GA SA309517519GA SA309517522GA SA309517530GA SA309517531GA SA309517532GA SA309436178GA SA309507312GA SA309517547GA SA309517549GA SA309517552GA SA309517553GA SA309517554GA SA309517562GA SA309517567GA SA309517568GA SA309517569GA SA309517570GA SA309517573GA SA309517574GA SA309517575GA SA309517584GA SA309517608GA SA309517623GA SA309446486GA SA309456562GA SA309466770GA SA309466772GA SA309477083GA SA309477095GA SA309477114GA SA309497144GA SA309497145GA SA309497147GA SA309497148GA SA309497149GA SA309497150GA SA309497164GA SA309497165GA SA309497170GA SA309497172GA SA309497174GA SA309497177GA SA309517456GA SA309517478GA SA309517556GA SA309517558GA SA309517571GA SA309517589GA SA309517611GA SA309415971GA SA309456614GA SA309456692GA SA309476920GA SA309476924GA SA309477023GA SA309477026GA SA309477037GA SA309477042GA SA309477046GA SA309477049GA SA309477052GA SA309477054GA SA309477055GA SA309477059GA SA309477060GA SA309477063GA SA309477066GA SA309477067GA SA309477068GA SA309477073GA SA309477077GA SA309477090GA SA309477100GA SA309477102GA SA309477119GA SA309477120GA SA309477126GA SA309477127GA SA309477137GA SA309497143GA SA309497151GA SA309497153GA SA309497154GA SA309497157GA SA309497160GA SA309497163GA SA309497166GA SA309497167GA SA309497169GA SA309497173GA SA309497176GA SA309517814GA SA309416000GA SA309517546GA SA309517561GA SA309517578GA SA309517581GA SA309517582GA SA309517586GA SA309517588GA SA309517591GA SA309517593GA SA309517598GA SA309517602GA SA309426137GA SA309476921GA SA309476997GA SA309477007GA SA309477009GA SA309477013GA SA309477015GA SA309477028GA SA309477029GA SA309477130GA SA309507420GA SA309517452GA SA309517470GA SA309517471GA SA309517480GA SA309517487GA SA309517496GA SA309517528GA SA309517630GA SA309517632GA SA309517635GA SA309517637GA SA309517639GA SA309517640GA SA309517648GA SA309517652GA SA309517653GA SA309517659GA SA307220880GR SA307220892GR SA307231205GR SA307231257GR SA307231352GR SA307231385GR SA307231414GR SA307231475GR SA307241510GR SA307241513GR SA307382453GR SA307382587GR SA307453433GR SA308044522GR SA308145655GR SA308166052GR SA308337671GR SA308409036GR SA308419237GR SA309092488GR SA309112571GR SA309334675GR SA309375468GR SA309375506GR SA309415928GA SA309415969GA SA309415992GA SA309426048GA SA309426059GA SA309426060GA SA309426069GA SA309426099GA SA309426166GA SA309446445GA SA309446452GA SA309446456GA SA309446464GA SA309507432GA SA309507434GA SA309507442GA SA309517469GA SA309517473GA SA309517474GA SA309517477GA SA309517483GA SA309517489GA SA309517493GA SA309517494GA SA309517505GA SA309517511GA SA309517513GA SA309517517GA SA309517523GA SA309517529GA SA309517538GA SA309517541GA SA309517545GA SA309517560GA SA309517564GA SA309517576GA SA309517579GA SA309517580GA SA309517583GA SA309517585GA SA309517587GA SA309517628GA SA309517638GA SA309517644GA SA309517647GA SA309517651GA SA309517662GA SA309517663GA SA309436183GA SA307070436GR SA307080522GR SA307210820GR SA307221115GR SA307332228GR SA307494014GR SA309112566GR SA309426039GA SA309436183GA SA309446476GA SA309446484GA SA309456546GA SA309456552GA SA309456554GA SA309466841GA SA309476893GA SA309476905GA SA309476916GA SA309476941GA SA309476943GA SA309476956GA SA309476957GA SA309477011GA SA309476961GA SA309476962GA SA309476970GA SA309476971GA SA309476973GA SA309476975GA SA309476976GA SA309497156GA SA309497159GA SA309497161GA SA309517763GA SA309517789GA SA309446407GA SA309446514GA SA309456550GA SA309466847GA SA309476862GA SA309476972GA SA309476978GA SA309476990GA SA309477005GA SA309477014GA SA309497175GA SA309507370GA SA309517484GA SA309517521GA SA309517533GA SA309517537GA SA309517540GA SA309517543GA SA309517629GA SA309517661GA SA309517707GA SA309517735GA SA309517775GA SA309517790GA SA309517795GA SA309517797GA SA309517804GA SA309517813GA SA309517820GA SA309517869GA SA309517870GA SA309477094GA SA309517610GA SA309517620GA SA309476929GA SA309476939GA SA309477001GA SA309477020GA SA309477053GA SA309477058GA SA309477096GA SA309497152GA SA309497168GA SA309507314GA SA309517563GA SA309517592GA SA309517594GA SA309517599GA SA309517675GA SA309517687GA SA309517704GA SA309517838GA SA309517848GA SA309517849GA SA309517856GA SA309517859GA SA309517862GA SA309517863GA SA309517867GA SA309528004GA SA309528008GA SA309528011GA SA309528013GA SA309528015GA SA309517676GA SA309517679GA SA309517686GA SA309517705GA SA309517781GA SA309517796GA SA309517816GA SA309517817GA SA309517823GA SA309517830GA SA309517852GA SA309517853GA SA309527923GA SA309527993GA SA309527994GA SA309527999GA SA309528000GA SA309528005GA SA309528007GA SA309528010GA SA309476948GA SA309476955GA SA309476958GA SA309476974GA SA309517800GA SA309517875GA SA309507429GA SA309517509GA SA309517525GA SA309517527GA SA309517535GA SA309517626GA SA309517646GA SA309517649GA SA309517650GA SA309517658GA SA309517666GA SA309517669GA SA309517671GA SA309517673GA SA309517678GA SA309517681GA SA309517684GA SA309517688GA SA309517689GA SA309517694GA SA309517698GA SA309517700GA SA309517709GA SA309517765GA SA309517776GA SA309517787GA SA310028142GA SA309436172GA SA309477017GA SA309517559GA SA309517572GA SA309517595GA SA309517851GA SA309517865GA SA309527998GA SA309528002GA SA309528012GA SA309528014GA SA310028114GA SA310028140GA SA310028149GA SA309476954GA SA309507343GA SA309517516GA SA309517885GA SA309517892GA SA309517898GA SA309517899GA SA309517902GA SA309517906GA SA309527911GA SA309527939GA SA309527943GA SA309527956GA SA309527987GA SA309538103GA SA310028130GA SA309304283GR SA309365122GR SA309365155GR SA309365354GR SA309517847GA SA310028126GA SA310028129GA SA310028134GA SA310028138GA SA310028315GA SA310028317GA SA310028319GA SA310028326GA SA310028331GA SA310028337GA SA310038343GA SA310028196GA SA310028201GA SA310028218GA SA310028252GA SA310028272GA SA310028294GA SA309507348GA SA309538026GA SA309538031GA SA309538045GA SA309538048GA SA309538049GA SA309538087GA SA309538090GA SA309538092GA SA309538096GA SA309538104GA SA310028154GA SA309538029GA SA309538034GA SA309538043GA SA309538046GA SA309538050GA SA309538093GA SA310028189GA SA310038380GA SA310038441GA SA310048620GA SA310068809GA SA310068824GA SA309426022GA SA309538058GA SA309538063GA SA309538081GA SA310048579GA SA309517822GA SA309517826GA SA310038530GA SA310048640GA SA309538098GA SA309538106GA SA310028262GA SA310048567GA SA310048643GA SA310038352GA SA310038427GA SA310048564GA SA310048668GA SA310058690GA SA310058691GA SA310058717GA SA310068833GA SA310038537GA SA310058716GA SA310058720GA SA310058730GA SA310058743GA SA309365147GR SA309517544GA SA309517877GA SA309538095GA SA310028290GA SA310038359GA SA310038391GA SA310038408GA SA310038444GA SA310038449GA SA310038348GA SA310038497GA SA310038513GA SA310038534GA SA310048542GA SA310048553GA SA310048647GA SA310048649GA SA309517542GA SA309527997GA SA309538062GA SA309538085GA SA309426040GA SA309507358GA SA309517660GA SA309517664GA SA309517668GA SA309517672GA SA309517674GA SA309517690GA SA309517691GA SA309517703GA SA309517731GA SA309527912GA SA310028155GA SA310028160GA SA310028182GA SA310028186GA SA310028191GA SA310028199GA SA310028213GA SA310028214GA SA310028223GA SA310028265GA SA310028266GA SA310028275GA SA309405874GA SA309426154GA SA309517550GA SA309517555GA SA309517557GA SA309517565GA SA309517566GA SA309517577GA SA309517590GA SA309517597GA SA309517600GA SA309517601GA SA309517603GA SA309517604GA SA309517605GA SA309517606GA SA309517607GA SA309517609GA SA309517613GA SA309517616GA SA309517617GA SA309517618GA SA309517619GA SA309517621GA SA309517622GA SA309517624GA SA309517625GA SA309517842GA SA309517854GA SA309517855GA SA309527992GA SA309527995GA SA309527996GA SA309528003GA SA309538052GA SA309538056GA SA309538059GA SA309538060GA SA309538061GA SA309538070GA SA309538079GA SA309538080GA SA310028148GA SA310048663GA SA309476872GA SA309477051GA SA309517819GA SA309517824GA SA309517831GA SA309517871GA SA309517878GA SA309517880GA SA309517890GA SA309517900GA SA309517901GA SA309527914GA SA309527948GA SA310038510GA SA310048563GA SA310048660GA SA310048676GA SA310048682GA SA310048683GA SA310048686GA SA310058699GA SA310058705GA SA310058707GA SA310058718GA SA310058721GA SA310058722GA SA310058724GA SA310058726GA SA310058733GA SA310058735GA SA310058738GA SA310058739GA SA310058740GA SA310058744GA SA310058745GA SA310058749GA SA310058750GA SA310058753GA SA310058754GA SA310058755GA SA310058756GA SA310058757GA SA310058758GA SA310058759GA SA310058760GA SA310058762GA SA310058763GA SA310058764GA SA310058765GA SA310068797GA SA310068798GA SA310068799GA SA310068800GA SA310068801GA SA310068803GA SA310068804GA SA310068805GA SA310068806GA SA310068808GA SA310068810GA SA310068812GA SA310068814GA SA310068818GA SA310068819GA SA310068823GA SA310068825GA SA310068831GA SA310068832GA SA310068838GA SA310068839GA SA310068844GA SA310118990GA SA310118995GA SA310119000GA SA310118978GA SA309507407GA SA310038430GA SA310048619GA SA310068925GA SA310068938GA SA310068941GA SA309395708GA SA309395756GA SA309405824GA SA309405841GA SA309436295GA SA309436299GA SA309436316GA SA309436339GA SA309436353GA SA309436370GA SA309446442GA SA309446449GA SA309446461GA SA309446470GA SA309446473GA SA309446480GA SA309446491GA SA309446492GA SA309446527GA SA309527916GA SA309538094GA SA310028187GA SA310028197GA SA310028203GA SA310028211GA SA310028269GA SA310028274GA SA310028291GA SA310028306GA SA310038395GA SA310068848GA SA310068852GA SA310068870GA SA310068875GA SA310068886GA SA310068887GA SA310068892GA SA310068893GA SA310068895GA SA310068898GA SA310068899GA SA310068902GA SA310068914GA SA310068919GA SA310068921GA SA310068922GA SA310068926GA SA310028112GA SA310028115GA SA310028120GA SA310028122GA SA310028127GA SA310028132GA SA310028135GA SA310028145GA SA310028180GA SA310028181GA SA310028184GA SA310028195GA SA310028217GA SA310028221GA SA310028222GA SA310028268GA SA310028284GA
SA310028300GA SA310028307GA SA310038476GA SA310038503GA SA310038514GA SA310038524GA SA310038525GA SA310048541GA SA310048658GA SA310048664GA SA310048665GA SA310048666GA SA310048667GA SA310048670GA SA310048680GA SA310048681GA SA310048684GA SA310048687GA SA310058692GA SA310058694GA SA310058696GA SA310058697GA SA310058698GA SA310058700GA SA310058701GA SA310058704GA SA310058708GA SA310058709GA SA310068853GA SA310068868GA SA310118960GA SA310118961GA SA310118964GA SA310118974GA SA310118976GA SA310118985GA SA310118988GA SA310118983GA SA310118968GA SA310118991GA SA310118998GA SA310119003GA SA310119007GA SA310139041GA SA310139055GA SA310139058GA SA310139062GA SA309517883GA SA309517903GA SA309527955GA SA309527957GA SA309517883GA SA309517903GA SA309527955GA SA309527957GA SA310118972GA SA310068943GA SA310118982GA SA310058737GA SA310068927GA SA310139069GA SA310139210GA SA310139250GA SA310139255GA SA310139266GA SA310139271GA SA310139273GA SA310139281GA SA310139285GA SA309538084GA SA310028171GA SA310038520GA SA310038527GA SA310139093GA SA310139094GA SA310139098GA SA310139154GA SA310139158GA SA310139162GA SA310139166GA SA310139173GA SA310139174GA SA310139175GA SA310139183GA SA310068858GA SA310068880GA SA310139057GA SA310139061GA SA310139066GA SA310139068GA SA310139087GA SA310139244GA SA310038440GA SA310139071GA SA310139086GA SA310139088GA SA310139089GA SA310139248GA SA310139249GA SA310139251GA SA310139257GA SA310139259GA SA310139262GA SA310139263GA SA310139274GA SA310139275GA SA310139276GA SA310139277GA SA310139282GA SA310048558GA SA310139212GA SA310139234GA SA310159368GA SA310159380GA SA310169529GA SA309517837GA SA310038483GA SA310038487GA SA310038492GA SA310038492GAMKE SA310038507GA SA310048633GA SA310048634GA SA310048641GA SA310048673GA SA310048679GA SA310058761GA SA310068929GA SA310139044GA SA310139045GA SA310139046GA SA310139054GA SA310139056GA SA310139059GA SA310139076GA SA310139077GA SA310139278GA SA310149316GA SA310149319GA SA310159373GA SA310159375GA SA310159378GA SA310159383GA SA310159386GA SA310139163GA SA310149348GA SA310149351GA SA310149352GA SA310159430GA SA310169480GA SA310139120GA SA310159423GA SA310159424GA SA310159427GA SA310159429GA SA310159431GA SA310159433GA SA310159436GA SA310159439GA SA310159440GA SA310159441GA SA310159449GA SA307080477GR SA307190724GR SA307261847GR SA308115247GR SA308429362GR SA308511455GR SA308511643GR SA309354898GR SA310139119GA SA310139133GA SA310139070GA SA310139073GA SA310139075GA SA310139085GA SA310139156GA SA310139245GA SA310139254GA SA310139256GA SA310139260GA SA310139264GA SA310139269GA SA310139280GA SA310139283GA SA310139287GA SA310159454GA SA309517656GA SA309517657GA SA309517701GA SA309517706GA SA309517711GA SA309517713GA SA309517715GA SA309517722GA SA309517728GA SA309517736GA SA309517737GA SA309517772GA SA309517773GA SA309517778GA SA309517788GA SA309517793GA SA309517798GA SA309517799GA SA309517803GA SA309517805GA SA309517810GA SA309517811GA SA309517815GA SA309517821GA SA309517882GA SA309517891GA SA309517904GA SA309527910GA SA309527913GA SA309527920GA SA309527927GA SA309527930GA SA309527933GA SA309527937GA SA309527945GA SA309527965GA SA309527967GA SA309527969GA SA309527986GA SA309517656GA SA309517657GA SA309517701GA SA309517706GA SA309517711GA SA309517713GA SA309517715GA SA309517722GA SA309517728GA SA309517736GA SA309517737GA SA309517772GA SA309517773GA SA309517778GA SA309517788GA SA309517793GA SA309517798GA SA309517799GA SA309517803GA SA309517805GA SA309517810GA SA309517811GA SA309517815GA SA309517821GA SA309517882GA SA309517891GA SA309517904GA SA309527910GA SA309527913GA SA309527920GA SA309527927GA SA309527930GA SA309527933GA SA309527937GA SA309527945GA SA309527965GA SA309527967GA SA309527969GA SA309527986GA SA310139051GA SA310139063GA SA309517656GA SA309517657GA SA309517701GA SA309517706GA SA309517711GA SA309517713GA SA309517715GA SA309517722GA SA309517728GA SA309517736GA SA309517737GA SA309517772GA SA309517773GA SA309517778GA SA309517788GA SA309517793GA SA309517798GA SA309517799GA SA309517803GA SA309517805GA SA309517810GA SA309517811GA SA309517815GA SA309517821GA SA309517882GA SA309517891GA SA309517904GA SA309527910GA SA309527913GA SA309527920GA SA309527927GA SA309527930GA SA309527933GA SA309527937GA SA309527945GA SA309527965GA SA309527967GA SA309527969GA SA309527986GA SA310139051GA SA310139063GA SA310139291GA SA310139293GA SA310139298GA SA310139305GA SA308105091GR SA308105095GR SA308105097GR SA308135466GR SA308135504GR SA309436187GR SA309436188GR SA309436190GR SA309436192GR SA309436195GR SA309436197GR SA309436202GR SA309436203GR SA309436204GR SA309436205GR SA309436208GR SA309436223GR SA310139106GA SA310139188GA SA310139194GA SA310149339GA SA310149341GA SA310149342GA SA310149344GA SA310149345GA SA310149349GA SA310149350GA SA310149354GA SA310149355GA SA310149356GA SA310159359GA SA310139185GA SA310139213GA SA310048562GA SA310118966GA SA310118973GA SA310118984GA SA310118997GA SA310139092GA SA310139095GA SA310139096GA SA310139104GA SA310139107GA SA310139114GA SA310139115GA SA310139122GA SA310139127GA SA310139134GA SA310139135GA SA310139141GA SA310139142GA SA310139143GA SA310139144GA SA310139145GA SA310139146GA SA310139148GA SA310139149GA SA310139150GA SA310139151GA SA310139152GA SA310139153GA SA310139155GA SA310139157GA SA310139159GA SA310139160GA SA310139161GA SA310139164GA SA310139165GA SA310139167GA SA310139168GA SA310139172GA SA310139177GA SA310139178GA SA310139179GA SA310139180GA SA310139184GA SA310139186GA SA310139190GA SA310139191GA SA310139193GA SA310139198GA SA310139200GA SA310139201GA SA310139202GA SA310139203GA SA310139205GA SA310139206GA SA310139207GA SA310139209GA SA310139215GA SA310139218GA SA310139219GA SA310139231GA SA310139290GA SA310139294GA SA310139296GA SA310149340GA SA310149346GA SA310159374GA SA310139303GA SA310149357GA SA310159435GA SA310159437GA SA310159437GAMKE SA310159438GA SA310159450GA SA310169481GA SA310169483GA SA310169491GA SA310169498GA SA310199581GA SA310199590GA SA310199596GA SA310199598GA SA310199599GA SA310199600GA SA310199601GA SA310199602GA SA310199603GA SA310199604GA SA310199605GA SA310199606GA SA310199607GA SA310199608GA SA310199609GA SA310199610GA SA310199612GA SA310199613GA SA310199615GA SA310199616GA SA310199617GA SA310199618GA SA310199619GA SA310199620GA SA310199621GA SA310199622GA SA310199626GA SA310199627GA SA310199640GA SA310169523GA SA310199565GA SA310139074GA SA310139080GA SA310139082GA SA310139083GA SA310139265GA SA310139267GA SA310139268GA SA310139270GA SA310139284GA SA310139288GA SA310169499GA SA310169502GA SA310169503GA SA310199579GA SA310199597GA SA310219802GA SA310219803GA SA310219804GA SA310219807GA SA310219808GA SA310219810GA SA310219814GA SA310219815GA SA310219817GA SA310219819GA SA310219821GA SA310219823GA SA310219826GA SA310219828GA SA310219840GA SA310219841GA SA310219843GA SA310219844GA SA310219846GA SA310219847GA SA310219854GA SA310139047GA SA310139065GA SA310149309GA SA310149321GA SA310149333GA SA310149336GA SA310159367GA SA310219816GA SA310219820GA SA310219822GA SA309527962GA SA310028273GA SA310038386GA SA310038462GA SA310048604GA SA310048613GA SA310048616GA SA310169485GA SA310169505GA SA310139272GA SA310219805GA SA310219809GA SA310219813GA SA310219825GA SA310139052GA SA310139222GA SA310139286GA SA310149328GA SA310159455GA SA310199576GA SA310219788GA SA310219790GA SA310219792GA SA310219801GA SA310219835GA SA310199614GA SA310259860GA SA310259861GA SA310259869GA SA310259870GA SA310199594GA SA310199623GA SA310199683GA SA310199723GA SA310199745GA SA310259855GA SA310259856GA SA310259858GA SA310259859GA SA310259862GA SA310259863GA SA310259864GA SA310259865GA SA310259866GA SA310259867GA SA310259868GA SA310259871GA SA310259872GA SA310259873GA SA310259875GA SA310259876GA SA310259879GA SA310259881GA SA310259885GA SA310259886GA SA310259887GA SA310259888GA SA310259890GA SA310259891GA SA310259892GA SA310259893GA SA310259894GA SA310199567GA SA310199572GA SA310199667GA SA310199730GA SA310219782GA SA310219783GA SA310219784GA SA310219785GA SA310219786GA SA310219789GA SA310219791GA SA310219793GA SA310219794GA SA310219795GA SA310219797GA SA310219798GA SA310219799GA SA310219800GA SA310219832GA SA310219836GA SA310219837GA SA310219839GA SA310259857GA SA310259874GA SA310259880GA SA310259889GA SA310219827GA SA310219811GA SA310219818GA SA310219830GA SA310199717GA SA310199719GA SA310199735GA SA310199738GA SA310199739GA SA310219812GA SA310219829GA SA310219845GA SA310259877GA SA310259878GA SA310259883GA SA310259911GA SA310259912GA SA310259913GA SA310259914GA SA310259923GA SA310259928GA SA310259960GA SA310259980GA SA310259985GA SA310259987GA SA310259989GA SA310260004GA SA310260012GA SA310169500GA SA310169501GA SA310169506GA SA310169507GA SA310169532GA SA310199562GA SA310199564GA SA310199566GA SA310199570GA SA310199573GA SA310199575GA SA310199578GA SA310199580GA SA310199583GA SA310199588GA SA310199592GA SA310199593GA SA310199645GA SA310199647GA SA310199648GA SA310199661GA SA310199663GA SA310199665GA SA310199666GA SA310199669GA SA310199672GA SA310199689GA SA310199702GA SA310199710GA SA310199715GA SA310199727GA SA310199737GA SA310219787GA SA310219796GA SA310219833GA SA310219834GA SA310219838GA SA310259973GA SA310259975GA SA310259976GA SA310259988GA SA310260001GA SA310260003GA SA310260006GA SA310260007GA SA310260008GA SA310260009GA SA310260010GA SA310260011GA SA310260016GA SA310260017GA SA310260018GA SA310260019GA SA310260020GA SA310260021GA SA310260022GA SA310260023GA SA310260024GA SA310260025GA SA310260026GA SA310260027GA SA310260028GA SA310260030GA SA310260034GA SA310260042GA SA310260047GA SA310260049GA SA310260053GA SA310260057GA SA310260067GA SA310260072GA SA310260078GA SA310260079GA SA310260082GA SA310118970GA SA310118989GA SA310118996GA SA310119001GA SA310119006GA SA310119013GA SA310119014GA SA310119017GA SA310139121GA SA310139126GA SA310139306GA SA310139308GA SA310199716GA SA310260216GA SA310260217GA SA310260219GA SA310260221GA SA310260225GA SA310260229GA SA310260230GA SA310260233GA SA310260241GA SA310260250GA SA310260297GA SA310260305GA SA310260307GA SA310260308GA SA310260309GA SA310260310GA SA310260311GA SA310260313GA SA310260314GA SA310260322GA SA310260324GA SA310260325GA SA310260327GA SA310260328GA SA310260330GA SA310260331GA SA310260332GA SA310260334GA SA310260335GA SA310260336GA SA310260337GA SA310260339GA SA310260341GA SA310260342GA SA310260347GA SA310260348GA SA310260351GA SA310260352GA SA310259882GA SA310259974GA SA310259979GA SA310259981GA SA310259982GA SA310260014GA SA310260015GA SA310260031GA SA310260033GA SA310260035GA SA310260036GA SA310260038GA SA310260039GA SA310260044GA SA310260050GA SA310260056GA SA310260060GA SA310260061GA SA310260063GA SA310260064GA SA310260065GA SA310260066GA SA310260068GA SA310260069GA SA310260071GA SA310260073GA SA310260074GA SA310260075GA SA310260080GA SA310260215GA SA310260226GA SA310260231GA SA310260239GA SA310260240GA SA310260244GA SA310260245GA SA310260246GA SA310260247GA SA310260248GA SA310260249GA SA310260254GA SA310260255GA SA310260257GA SA310260385GA SA310260401GA SA310260404GA SA310260406GA SA310260407GA SA310260408GA SA310260410GA SA310260411GA SA310260414GA SA310260421GA SA310260423GA SA310260469GA SA310260470GA SA310260471GA SA310260472GA SA310260473GA SA310260486GA SA310260490GA SA310260493GA SA310260494GA SA310260495GA SA310260499GA SA310068884GA SA310068915GA SA310260043GA SA310260059GA SA310260218GA SA310260224GA SA310260392GA SA310260396GA SA310260417GA SA310260422GA SA310260424GA SA310260428GA SA310260437GA SA310260488GA SA310260491GA SA310260496GA SA310260312GA SA310260532GA SA310199649GA SA310199652GA SA310199654GA SA310199657GA SA310199658GA SA310199660GA SA310199670GA SA310199674GA SA310199675GA SA310199676GA SA310199677GA SA310199678GA SA310199686GA SA310199587GA SA310199589GA SA310199591GA SA310199595GA SA310199642GA SA310199643GA SA310199644GA SA310199646GA SA310199650GA SA310199651GA SA310199653GA SA310199655GA SA310199656GA SA310199664GA SA310199668GA SA310199673GA SA310199679GA SA310199680GA SA310199703GA SA310199704GA SA310199713GA SA310199741GA SA310199752GA SA310199753GA SA309476968GA SA310159476GA SA310169526GA SA310260546GA SA310118967GA SA310118993GA SA310260362GA SA310260372GA SA310260412GA SA310149334GA SA310260046GA SA310260232GA SA310260343GA SA310260346GA SA310260350GA SA310260367GA SA310260390GA SA310260393GA SA310260415GA SA309446394GA SA309446403GA SA310068816GA SA310118971GA SA310118977GA SA310118979GA SA310290746GA SA310300791GA SA310300795GA SA310300798GA SA310300806GA SA310270585GA SA310270586GA SA310270587GA SA310290658GA SA310290704GA SA310290705GA SA310290707GA SA310290708GA SA310290710GA SA310290711GA SA310290718GA SA310290723GA SA310290724GA SA310290729GA SA310290731GA SA310300774GA SA310260041GA SA310260051GA SA310260054GA SA310260055GA SA310260062GA SA310260081GA SA310290714GA SA310331323GA SA310351534GA SA310351543GA SA310351549GA SA310361759GA SA310361764GA SA310361769GA SA310372025GA SA310372059GA SA310372063GA SA310372144GA SA310372152GA SA310372158GA SA310372275GA SA310372280GA SA310372304GA SA310372310GA SA310361644GA SA310361809GA SA310372204GA SA310361794GA SA310382671GA SA310290660GA SA310361874GA SA310361876GA SA310361897GA SA310372097GA SA310372104GA SA310382552GA SA310382581GA SA310382584GA SA310382686GA SA310393028GA SA310259977GA SA310260365GA SA310392998GA SA310393014GA SA310290653GA SA310290655GA SA310290657GA SA310290659GA SA310290661GA SA310290748GA SA310290749GA SA310290750GA SA310290754GA SA310290755GA SA310290756GA SA310290757GA SA310290758GA SA310290759GA SA310361647GA SA310361877GA SA310372108GA SA310393017GA SA310393052GA SA310382648GA SA310382626GA SA310382631GA SA310382729GA SA310382761GA SA310393037GA SA310393054GA SA310393071GA SA310393073GA SA310351531GA SA310361792GA SA310361798GA SA310361801GA SA310361803GA SA310361804GA SA310361805GA SA310361807GA SA310361808GA SA310361815GA SA310361820GA SA310361838GA SA310361840GA SA310361841GA SA310393150GA SA310393155GA SA310393162GA SA310393165GA SA310393166GA SA310393167GA SA310393170GA SA310393171GA SA310393173GA SA310393175GA SA310321055GA SA310321062GA SA310361949GA SA310382418GA SA310382450GA SA310382499GA SA310382513GA SA310382608GA SA310382647GA SA310382649GA SA310382663GA SA310393156GA SA310393158GA SA310393164GA SA310393169GA SA310393177GA SA310393279GA SA310259944GA SA310331337GA SA310331342GA SA310331343GA SA310331345GA SA310331349GA SA310331351GA SA310341373GA SA310341376GA SA310341387GA SA310341388GA SA310341399GA SA310341400GA SA310341401GA SA310341402GA SA310341403GA SA310341406GA SA310382386GA SA310382391GA SA310382472GA SA310382476GA SA310382477GA SA310382479GA SA310382484GA SA310382494GA SA310382618GA SA310382619GA SA310382634GA SA310382636GA SA310382645GA SA310382378GA SA310382449GA SA310382511GA SA310382606GA SA310382607GA SA310382610GA SA310382612GA SA310382613GA SA310382615GA SA310382616GA SA310382642GA SA310382652GA SA310382656GA SA310382664GA SA310423572GA SA310423573GA SA310310959GA SA310290745GA SA310290753GA SA310310959GA SA310413471GA SA310413476GA SA310423587GA SA310423607GA SA310413493GA SA310413498GA SA310413500GA SA310413507GA SA310413514GA SA310413524GA SA310413535GA SA310423604GA SA310423606GA SA310423611GA SA310423617GA SA310423619GA SA310423621GA SA310423626GA SA310423640GA SA310423649GA SA310423650GA SA310423730GA SA310423821GA SA310423825GA SA310423836GA SA310423838GA SA310423842GA SA310423864GA SA310423929GA SA310423935GA SA310423943GA SA310423959GA SA310423962GA SA310424000GA SA310424001GA SA310424017GA SA310321010GA SA310321010GA SA310413491GA SA310413499GA SA310413509GA SA310413511GA SA310413538GA SA310423613GA SA310423614GA SA310423618GA SA310423622GA SA310423628GA SA310423637GA SA310423639GA SA310423663GA SA310423709GA SA310423933GA SA310423937GA SA310423946GA SA310423950GA SA310423953GA SA310423956GA SA310423957GA SA310423960GA SA310423961GA SA310423967GA SA310423971GA SA310424002GA SA310424003GA SA310424004GA SA310424006GA SA310424008GA SA310424025GA SA310424027GAMKE SA310424033GA SA310424034GAMKE SA310424039GA SA310424046GA SA310434199GAMKE SA310434200GA SA310434204GAMKE SA310434205GA SA310434207GAMKE SA310434208GA SA310434210GA SA310434211GAMKE SA310434212GA SA310434213GA SA310434227GAMKE SA310434238GA SA310444247GAMKE SA310444253GAMKE SA310372179GA SA310382540GA SA310382541GA SA310382542GA SA310382544GA SA310382550GA SA310382553GA SA310382554GA SA310382557GA SA310382558GA SA310382562GA SA310382563GA SA310382575GA SA310382580GA SA310382593GA SA310382599GA SA310382602GA SA310382604GA SA310382693GA SA310403363GA SA310403364GA SA310403372GA SA310403433GA SA310423758GA SA310423793GA SA310423896GA SA310423914GA SA310423917GA SA310423919GA SA310423979GA SA310423994GA SA310434070GA SA310423777GA SA310423779GA SA310423787GA SA310423819GA SA310423871GA SA310444296GA SA310310864GA SA310382832GA SA310392903GA SA310392904GA SA310392943GA SA310392946GA SA310392997GA SA310392999GA SA310393006GA SA310393007GA SA310393008GA SA310393010GA SA310393011GA SA310393012GA SA310393015GA SA310393023GA SA310393026GA SA310393029GA SA310393035GA SA310393036GA SA310393048GA SA310393059GA SA310393061GA SA310393064GA SA310393065GA SA310393074GA SA310393075GA SA310393188GA SA310393190GA SA310393197GA SA310393200GA SA310393213GA SA310393219GA SA310393229GA SA310393249GA SA310393250GA SA310393252GA SA310393254GA SA310393260GA SA310393261GA SA310393268GA SA310393269GA SA310393274GA SA310393281GA SA310393283GA SA310393294GA SA310393304GA SA310423888GA SA310423989GA SA310434119GA SA310434120GA SA310434132GA SA310434136GA SA310434140GA SA310434151GA SA310434160GA SA310434170GA SA310434185GA SA310434186GA SA310434188GA SA310341368GA SA310341391GA SA310392912GA SA310393212GA SA310393216GA SA310321060GA SA310321060GA SA310393149GA SA310393151GA SA310393152GA SA310393154GA SA310393168GA SA310393172GA SA310393174GA SA310393178GA SA310403341GA SA310403348GA SA310403379GA SA310403384GA SA310403390GA SA310403398GA SA310403405GA SA310403414GA SA310403415GA SA310403418GA SA310403436GA SA310403442GA SA310423780GA SA310423785GA SA310423800GA SA310423811GA SA310423916GA SA310423974GA SA310434063GA SA310434064GA SA310434066GA SA310434071GA SA310434077GA SA310434084GA SA310434090GA SA310434091GA SA310434094GA SA310434097GA SA310434106GA SA310434107GA SA310434108GA SA310434110GA SA310434114GA SA310434118GA SA310434125GA SA310434134GA SA310434148GA SA310434173GA SA310434194GA SA310434195GA SA310444266GA SA310434089GA SA310260359GA SA310331162GA SA310351499GA SA310361741GA SA310361742GA SA310361743GA SA310361745GA SA310361746GA SA310361747GA SA310361748GA SA310361749GA SA310361750GA SA310361751GA SA310361753GA SA310361758GA SA310361763GA SA310361766GA SA310361770GA SA310361776GA SA310361810GA SA310444496GA SA310444504GA SA310455031GA SA310393080GA SA310393270GA SA310393280GA SA310393295GA SA310393309GA SA310393310GA SA310423766GA SA310341398GA SA310341404GA SA310455048GA SA310455056GA SA310455170GA SA310444369GA SA310444503GA SA310444513GA SA310444531GA SA310424044GA SA310424031GA SA310454982GA SA310423862GA SA310424010GA SA310454976GA SA310454981GA SA310454986GA SA310454996GA SA310455000GA SA310455004GA SA310455016GA SA310455018GA SA310455030GA SA310455058GA SA310455061GA SA310455062GA SA310455156GA SA310455161GA SA310455164GA SA310465352GA SA310465366GA SA310465394GA SA310423651GA SA310423694GA SA310423600GA SA310455006GA SA310455026GA SA310455057GA SA310382387GA SA310382413GA SA310382415GA SA310310974GA SA310310978GA SA310321007GA SA310321019GA SA310321041GA SA310321066GA SA310310974GA SA310310978GA SA310321007GA SA310321019GA SA310321041GA SA310321066GA SA310403396GA SA310403420GA SA310423683GA SA310423738GA SA310423921GA SA310423986GA SA310423988GA SA310455256GA SA310455275GA SA310465314GA SA310475525GA SA310475537GA SA310475611GA SA310485808GA SA310444499GA SA310444509GA SA310444839GA SA310444840GA SA310444842GA SA310444856GA SA310444859GA SA310444893GA SA310454928GA SA310454939GA SA310455259GA SA310403435GA SA310485856GA SA310444259GA SA310392895GA SA310393106GA SA310485867GA SA310475566GA SA310475585GA SA310475610GA SA310475628GA SA310475629GA SA310475630GA SA310475634GA SA310475637GA SA310475684GA SA310475685GA SA310475686GA SA310475696GA SA310475697GA SA310475701GA SA310475707GA SA310475728GA SA310475734GA SA310485786GA SA310485794GA SA310485800GA SA310485801GA SA310485804GA SA310485807GA SA310485814GA SA310485822GA SA310485833GA SA310423799GA SA310495953GA SA310495955GA SA310475682GA SA310485788GA SA310485792GA SA310485795GA SA310485811GA SA310392897GA SA310392911GA SA310393086GA SA310393094GA SA310393108GA SA310393109GA SA310393114GA SA310393131GA SA310393222GA SA310393233GA SA310393262GA SA310444364GA SA310455055GA SA310455091GA SA310465367GA SA310321046GA SA310290679GA SA310321046GA SA310361952GA SA310392949GA SA310393138GA SA310393139GA SA310393140GA SA310393142GA SA310393143GA SA310393145GA SA310423680GA SA310423684GA SA310423882GA SA310423886GA SA310423906GA SA310434081GA SA310434082GA SA310434115GA SA310485851GA SA310485881GA SA310393237GA SA310393255GA SA310393296GA SA310403319GA SA310403320GA SA310403321GA SA310403322GA SA310403325GA SA310403332GA SA310413451GA SA310423666GA SA310423667GA SA310423670GA SA310423671GA SA310423672GA SA310423674GA SA310423675GA SA310423677GA SA310423681GA SA310423688GA SA310423696GA SA310423698GA SA310423701GA SA310423704GA SA310423708GA SA310423712GA SA310423713GA SA310423715GA SA310423720GA SA310423721GA SA310423741GA SA310423745GA SA310423753GA SA310423769GA SA310423822GA SA310423839GA SA310423841GA SA310423940GA SA310423949GA SA310423951GA SA310444851GA SA310444863GA SA310454938GA SA310454989GA SA310455015GA SA310455154GA SA310475481GA SA310475485GA SA310475489GA SA310485882GA SA310495919GA SA310495970GA SA310331081GA SA310393083GA SA310393130GA SA310403311GA SA310423748GA SA310423759GA SA310423866GA SA310423883GA SA310423887GA SA310423904GA SA310423912GA SA310423985GA SA310434048GA SA310434059GA SA310434069GA SA310434098GA SA310434139GA SA310434180GA SA310434183GA SA310454915GA SA310454946GA SA310454984GA SA310455009GA SA310455034GA SA310455065GA SA310455066GA SA310455069GA SA310455118GA SA310455120GA SA310455122GA SA310455125GA SA310455127GA SA310455131GA SA310455142GA SA310455227GA SA310455228GA SA310455232GA SA310455233GA SA310455248GA SA310455251GA SA310455257GA SA310455264GA SA310455266GA SA310455268GA SA310455272GA SA310455278GA SA310455282GA SA310455286GA SA310455289GA SA310465298GA SA310465310GA SA310465312GA SA310465317GA SA310372302GA SA310423697GA SA310506841GA SA310507174GA SA310507176GA SA310507183GA SA310507201GA SA310507205GA SA310507237GA SA310507292GA SA310507298GA SA310507301GA SA310507304GA SA310507309GA SA310507310GA SA310507319GA SA310507326GA SA310507328GA SA310507329GA SA310507332GA SA310507336GA SA310507337GA SA310507356GA SA310507366GA SA310507367GA SA310507371GA SA310507372GA SA310507377GA SA310507379GA SA310507381GA SA310507382GA SA310507383GA SA310507384GA SA310507385GA SA310507386GA SA310507387GA SA310507499GA SA310517583GA SA310361707GA SA310372294GA SA310403328GA SA310403426GA SA310413544GA SA310423668GA SA310423736GA SA310423755GA SA310423778GA SA310423818GA SA310423873GA SA310423874GA SA310423893GA SA310423895GA SA310423907GA SA310423910GA SA310423958GA SA310423978GA SA310423980GA SA310423987GA SA310454918GA SA310454925GA SA310454927GA SA310454944GA SA310454947GA SA310454948GA SA310454949GA SA310454950GA SA310454952GA SA310454953GA SA310454990GA SA310454991GA SA310454992GA SA310454995GA SA310455028GA SA310455047GA SA310455146GA SA310455147GA SA310455149GA SA310455150GA SA310455152GA SA310455157GA SA310475464GA SA310475536GA SA310475575GA SA310475579GA SA310475581GA SA310475582GA SA310475592GA SA310475598GA SA310475600GA SA310475608GA SA310475649GA SA310475663GA SA310475709GA SA310475719GA SA310475730GA SA310485785GA SA310485819GA SA310485824GA SA310485828GA SA310485843GA SA310485848GA SA310485853GA SA310485859GA SA310506749GA SA310507219GA SA310517588GA SA310517706GA SA310517789GA SA310517790GA SA310518161GA SA310518162GA SA310518205GA SA310351558GA SA310423652GA SA310423657GA SA310423903GA SA310423973GA SA310434074GA SA310434076GA SA310434078GA SA310434092GA SA310434124GA SA310434157GA SA310434182GA SA310455084GA SA310455123GA SA310455130GA SA310465350GA SA310465359GA SA310465383GA SA310465396GA SA310475501GA SA310496650GA SA310506747GA SA310517641GA SA310518223GA SA310403409GA SA310495945GA SA310496375GA SA310496376GA SA310496383GA SA310496394GA SA310496415GA SA310496418GA SA310496450GA SA310496624GA SA310506711GA SA310506725GA SA310506784GA SA310506785GA SA310506787GA SA310506788GA SA310506801GA SA310506808GA SA310506810GA SA310506813GA SA310506814GA SA310506821GA SA310506822GA SA310506838GA SA310506860GA SA310506869GA SA310506870GA SA310506882GA SA310506895GA SA310506989GA SA310507036GA SA310507038GA SA310507067GA SA310507082GA SA310507149GA SA310507156GA SA310507157GA SA310507163GA SA310507196GA SA310507197GA SA310507199GA SA310507200GA SA310507202GA SA310507203GA SA310507212GA SA310507216GA SA310507230GA SA310507262GA SA310507307GA SA310507357GA SA310507380GA SA310517708GA SA310517721GA SA310517987GA SA310518014GA SA310518152GA SA310518160GA SA310518164GA SA310518217GA SA310518225GA SA310518232GA SA310518237GA SA310403383GA SA310413556GA SA310423593GA SA310423845GA SA310423942GA SA310423944GA SA310454929GA SA310454935GA SA310454936GA SA310454937GA SA310454940GA SA310454975GA SA310454987GA SA310454988GA SA310454994GA SA310455012GA SA310455024GA SA310455025GA SA310455046GA SA310455052GA SA310455053GA SA310455148GA SA310455151GA SA310455153GA SA310455155GA SA310455159GA SA310455160GA SA310455163GA SA310455173GA SA310455174GA SA310496156GA SA310496470GA SA310496483GA SA310496513GA SA310496517GA SA310496630GA SA310496638GA SA310506665GA SA310506692GA SA310506732GA SA310506859GA SA310506874GA SA310506881GA SA310507062GA SA310507092GA SA310507093GA SA310507095GA SA310507107GA SA310507114GA SA310507116GA SA310507122GA SA310507406GA SA310507413GA SA310507417GA SA310507456GA SA310507481GA SA310507482GA SA310507483GA SA310507485GA SA310507486GA SA310507490GA SA310507492GA SA310507493GA SA310507494GA SA310507497GA SA310507504GA SA310507507GA SA310507509GA SA310507511GA SA310507512GA SA310507516GA SA310517585GA SA310517591GA SA310423690GA SA310506843GA SA310507215GA SA310507242GA SA310507317GA SA310517645GA SA310517646GA SA310517651GA SA310517652GA SA310517664GA SA310517678GA SA310517680GA SA310517683GA SA310518168GA SA310518169GA SA310518170GA SA310518176GA SA310518319GA SA310392914GA SA310393088GA SA310393192GA SA310393196GA SA310393230GA SA310393235GA SA310393291GA SA310393303GA SA310475710GA SA310495890GA SA310496368GA SA310496378GA SA310496384GA SA310496386GA SA310496387GA SA310496388GA SA310496393GA SA310496401GA SA310496428GA SA310496469GA SA310496529GA SA310496618GA SA310496626GA SA310496631GA SA310506658GA SA310506682GA SA310506691GA SA310506706GA SA310506812GA SA310506836GA SA310506853GA SA310506877GA SA310507000GA SA310507012GA SA310507016GA SA310507024GA SA310507025GA SA310507027GA SA310507050GA SA310507069GA SA310507075GA SA310507127GA SA310507154GA SA310507162GA SA310507171GA SA310507267GA SA310507270GA SA310507272GA SA310507276GA SA310507288GA SA310507296GA SA310507338GA SA310507344GA SA310507346GA SA310507375GA SA310507399GA SA310507464GA SA310507519GA SA310517560GA SA310517866GA SA310260482GA SA310361684GA SA310361708GA SA310393077GA SA310393284GA SA310434053GA SA310434085GA SA310434138GA SA310434181GA SA310455230GA SA310506654GA SA310506672GA SA310506703GA SA310506708GA SA310506717GA SA310506726GA SA310506729GA SA310506990GA SA310506999GA SA310507005GA SA310507089GA SA310507090GA SA310507104GA SA310507123GA SA310507140GA SA310507408GA SA310507409GA SA310507429GA SA310507430GA SA310507432GA SA310507443GA SA310507445GA SA310507449GA SA310507452GA SA310507461GA SA310507463GA SA310507469GA SA310507473GA SA310507476GA SA310507477GA SA310507478GA SA310507488GA SA310507489GA SA310507508GA SA310507518GA SA310507523GA SA310507529GA SA310507533GA SA310507534GA SA310507535GA SA310507536GA SA310507537GA SA310507539GA SA310507547GA SA310507550GA SA310507551GA SA310507552GA SA310507555GA SA310507556GA SA310517558GA SA310517559GA SA310517563GA SA310517566GA SA310517567GA SA310517568GA SA310517570GA SA310517574GA SA310517581GA SA310517589GA SA310517596GA SA310517745GA SA310517817GA SA310517869GA SA310517875GA SA310518242GA SA310518275GA SA310518335GA SA310518337GA SA310518350GA SA310518361GA SA310518414GA SA310393234GA SA310403350GA SA310403413GA SA310403423GA SA310403427GA SA310403440GA SA310423742GA SA310423744GA SA310423756GA SA310423761GA SA310423770GA SA310423771GA SA310423773GA SA310423794GA SA310423806GA SA310423807GA SA310423809GA SA310423812GA SA310423814GA SA310423816GA SA310423872GA SA310423879GA SA310423900GA SA310424037GA SA310455126GA SA310455254GA SA310465293GA SA310465294GA SA310465305GA SA310465307GA SA310465313GA SA310465315GA SA310465318GA SA310475510GA SA310475557GA SA310475609GA SA310475626GA SA310475631GA SA310475639GA SA310475695GA SA310475732GA SA310475733GA SA310485768GA SA310485782GA SA310485789GA SA310485790GA SA310485793GA SA310485796GA SA310485806GA SA310485810GA SA310485812GA SA310485816GA SA310485818GA SA310485834GA SA310485836GA SA310485847GA SA310485863GA SA310506674GA SA310506683GA SA310506713GA SA310506736GA SA310506741GA SA310507080GA SA310507083GA SA310507084GA SA310507110GA SA310507131GA SA310507141GA SA310507146GA SA310507161GA SA310507238GA SA310507303GA SA310507308GA SA310507320GA SA310507330GA SA310507396GA SA310507400GA SA310507416GA SA310507421GA SA310507426GA SA310507441GA SA310507447GA SA310507455GA SA310507491GA SA310507505GA SA310507506GA SA310507517GA SA310517587GA SA310517602GA SA310517605GA SA310517606GA SA310517607GA SA310517608GA SA310517612GA SA310517613GA SA310517614GA SA310517616GA SA310517619GA SA310517633GA SA310517635GA SA310517636GA SA310517638GA SA310517650GA SA310517657GA SA310517660GA SA310517661GA SA310517670GA SA310517697GA SA310517704GA SA310517711GA SA310517712GA SA310517717GA SA310517719GA SA310517728GA SA310517729GA SA310517731GA SA310517738GA SA310517742GA SA310517744GA SA310517751GA SA310517754GA SA310517758GA SA310517761GA SA310517768GA SA310517769GA SA310517770GA SA310517774GA SA310517776GA SA310517777GA SA310517778GA SA310517780GA SA310517783GA SA310517785GA SA310517786GA SA310517792GA SA310518147GA SA310518149GA SA310518156GA SA310518157GA SA310518163GA SA310518179GA SA310518185GA SA310518187GA SA310518190GA SA310518204GA SA310518209GA SA310518210GA SA310518214GA SA310518230GA SA310518241GA SA310518252GA SA310518259GA SA310518263GA SA310518265GA SA310518266GA SA310518269GA SA310518270GA SA310518280GA SA310518282GA SA310518286GA SA310518289GA SA310518304GA SA310518305GA SA310423705GA SA310454924GA SA310454926GA SA310454934GA SA310454943GA SA310454945GA SA310423685GA SA310424019GA SA310444277GA SA310444304GA SA310454983GA SA310496331GA SA310496615GA SA310507043GA SA310507395GA SA310507553GA SA310507554GA SA310517696GA SA310518261GA SA310518291GA SA310518299GA SA310518332GA SA310518352GA SA310518356GA SA310518364GA SA310518365GA SA310518384GA SA310518394GA SA310518466GA SA310518706GA SA310518715GA SA310518719GA SA310518720GA SA310518721GA SA310310918GA SA310310918GA SA310496514GA SA310496653GA SA310506746GA SA310506831GA SA310506832GA SA310507022GA SA310507045GA SA310507228GA SA310507410GA SA310507415GA SA310507419GA SA310507420GA SA310507427GA SA310507431GA SA310507433GA SA310507439GA SA310507442GA SA310507454GA SA310507521GA SA310507525GA SA310507526GA SA310507543GA SA310507546GA SA310517564GA SA310517565GA SA310517595GA SA310517597GA SA310517656GA SA310517760GAMKE SA310517762GA SA310517791GA SA310518154GA SA310518194GA SA310518228GA SA310518235GA SA310518256GA SA310518271GA SA310518274GA SA310518292GA SA310518294GA SA310518303GA SA310518307GA SA310518312GA SA310518320GA SA310518343GA SA310518345GA SA310518366GA SA310518415GA SA310518437GA SA310518444GA SA310518457GA SA310518458GA SA310518537GA SA310518554GA SA310518617GA SA310518618GA SA310518633GA SA310518639GA SA310518640GA SA310518649GA SA310518651GAMKE SA310518659GA SA310518672GA SA310518690GA SA310518717GA SA310518718GA SA310529071GA SA310529081GA SA310529096GA SA310393186GA SA310393267GA SA310423673GA SA310423678GA SA310423682GA SA310423947GA SA310424032GA SA310444251GA SA310444258GA SA310444269GA SA310444270GA SA310444275GA SA310444278GA SA310444291GA SA310444292GA SA310444294GA SA310444301GA SA310444306GA SA310444312GA SA310444324GA SA310444325GA SA310444326GA SA310496643GA SA310506868GA SA310507128GA SA310507145GA SA310507392GA SA310507402GA SA310507422GA SA310507428GA SA310507460GA SA310507522GA SA310517610GA SA310517624GA SA310517634GA SA310517723GA SA310517724GA SA310517736GA SA310517746GA SA310517747GA SA310517752GA SA310517756GA SA310517763GA SA310517764GA SA310517765GA SA310517766GA SA310517767GA SA310517771GA SA310517772GA SA310517773GA SA310517782GA SA310517823GA SA310517830GA SA310517877GA SA310517982GA SA310517983GA SA310517985GA SA310517990GA SA310517991GA SA310517993GA SA310517994GA SA310517995GA SA310517996GA SA310518001GA SA310518002GA SA310518007GA SA310518010GA SA310518159GA SA310518247GA SA310518251GA SA310518255GA SA310518257GA SA310518258GA SA310518268GA SA310518296GA SA310518297GA SA310518324GA SA310518325GA SA310518326GA SA310518330GA SA310518333GA SA310518336GA SA310518341GA SA310518342GA SA310518349GA SA310518351GA SA310518353GA SA310518358GA SA310518359GA SA310518360GA SA310518367GA SA310518368GA SA310518370GA SA310518371GA SA310518376GA SA310518385GA SA310518386GA SA310518393GA SA310518397GA SA310518398GA SA310518401GA SA310518402GA SA310518403GA SA310518408GA SA310518409GA SA310518410GA SA310518416GA SA310518417GA SA310518439GA SA310518446GA SA310518447GA SA310518449GA SA310518459GA SA310518461GA SA310518464GA SA310518472GA SA310518475GA SA310518485GA SA310518499GA SA310518523GA SA310518527GA SA310518530GA SA310518535GA SA310518539GA SA310518541GA SA310518542GA SA310518551GA SA310518552GA SA310518553GA SA310518555GA SA310518557GA SA310518615GA SA310518621GA SA310518624GA SA310518625GA SA310518626GA SA310518627GA SA310518631GA SA310518632GA SA310518634GA SA310518636GA SA310518637GA SA310518638GA SA310518642GA SA310518643GA SA310518645GA SA310518648GA SA310518652GA SA310518657GA SA310518658GA SA310518662GA SA310518670GA SA310518674GA SA310518687GA SA310518688GA SA310518691GA SA310518692GA SA310518693GA SA310518695GA SA310518696GA SA310518699GA SA310518700GA SA310518702GA SA310518703GA SA310518704GA SA310518710GA SA310518712GA SA310518713GA SA310518716GA SA310518723GA SA310528895GA SA310528899GA SA310528962GA SA310528963GA SA310528979GA SA310528985GA SA310528992GA SA310528998GA SA310529007GA SA310529010GA SA310529011GA SA310529073GA SA310529079GA SA310529086GA SA310529090GA SA310529098GA SA310529099GA SA310529101GA SA310529113GA SA310361850GA SA310517753GA SA310517755GA SA310517757GA SA310518630GA SA310518635GA SA310518654GA SA310518655GA SA310518660GA SA310518661GA SA310518665GA SA310518666GA SA310518697GA SA310518698GA SA310528961GA SA310528981GA SA310529088GA SA310529093GA SA310529103GA SA310529106GA SA310529112GA SA310518433GA SA310518487GA SA310528943GA SA310290668GA SA310361887GA SA310382754GA SA310424028GA SA310424038GA SA310455017GA SA310506820GA SA310518465GA SA310518694GA SA310528901GA SA310528906GA SA310528936GA SA310528940GA SA310528944GA SA310529163GA SA310529222GA SA310518391GA SA310528849GA SA310528932GA SA310528947GA SA310529060GA SA310529087GA SA310529095GA SA310529097GA SA310529114GA SA310529122GA SA310434088GA SA310496324GA SA310496340GA SA310496349GA SA310506655GA SA310518009GA SA310518276GA SA310518369GA SA310518424GA SA310518434GA SA310518436GA SA310518451GA SA310518473GA SA310518503GA SA310518528GA SA310518543GA SA310518548GA SA310518561GA SA310518673GA SA310518686GA SA310528865GA SA310528870GA SA310528878GA SA310528911GA SA310528949GA SA310528951GA SA310528953GA SA310528957GA SA310529001GA SA310529003GA SA310529031GA SA310529032GA SA310529033GA SA310529042GA SA310529043GA SA310529157GA SA310529161GA SA310529164GA SA310529199GA SA310529200GA SA310529202GA SA310529203GA SA310529206GA SA310529211GA SA310529214GA SA310529219GA SA311029364GA SA311029365GA SA310507480GA SA310528749GA SA310529156GA SA310382839GA SA310518390GA SA310518423GA SA310518477GA SA310518676GA SA310528937GA SA310528939GA SA310529034GA SA310529041GA SA310529160GA SA310529218GA SA310528937GA SA310528939GA SA310518452GA SA310518453GA SA310518463GA SA310518479GA SA310518492GA SA310518493GA SA310518501GA SA310518521GA SA310528775GA SA310528778GA SA310528792GA SA310528813GA SA310528842GA SA310528891GA SA310528914GA SA310528915GA SA310528918GA SA310528925GA SA310528952GA SA310528959GA SA310528960GA SA310260320GA SA310260383GA SA310260402GA SA310341381GA SA310361785GA SA310382838GA SA310392886GA SA310393202GA SA310434122GA SA310455119GA SA310455121GA SA310507218GA SA310518198GA SA310518203GA SA310518208GA SA310518262GA SA310518396GA SA310518399GA SA310518483GA SA310518495GA SA310518509GA SA310518533GA SA310528733GA SA310528795GA SA310528928GA SA310528955GA SA310529061GA SA310529069GA SA310529102GA SA310539239GA SA310539311GA SA310361695GA SA310393144GA SA310496211GA SA310496212GA SA310496622GA SA310507014GA SA310507302GA SA310507311GA SA310518348GA SA310518373GA SA310518388GA SA310518407GA SA310518411GA SA310518412GA SA310518450GA SA310518476GA SA310528801GA SA310528802GA SA310528803GA SA310528912GA SA310539306GA SA311039518GA SA311039529GA SA311039551GA SA310393246GA SA310403349GA SA310413495GA SA310518430GA SA310518460GA SA310518462GA SA310518471GA SA310518474GA SA310518480GA SA310518482GA SA310518484GA SA310518491GA SA310528736GA SA310528744GA SA310528773GA SA310528787GA SA310528922GA SA310528930GA SA311039382GA SA310413458GA SA310475591GA SA310518678GA SA310518679GA SA310518685GA SA310518724GA SA310528856GA SA310528858GA SA310528861GA SA310528863GA SA310528864GA SA310528869GA SA310528967GA SA310528994GA SA310529006GA SA310529013GA SA310529028GA SA310529030GA SA310529035GA SA310529038GA SA310529040GA SA310529044GA SA310529047GA SA310529150GA SA310529159GA SA310529197GA SA310529208GA SA310529216GA SA310529220GA SA310413545GA SA310496329GA SA310496614GA SA310518675GA SA310518711GA SA310518714GA SA310528862GA SA310528868GA SA310528879GA SA311039387GA SA311039389GA SA311039393GA SA311039409GA SA311039416GA SA311039432GA SA311039449GA SA311039453GA SA311039458GA SA311039459GA SA311039463GA SA311039472GA SA311039473GA SA311039475GA SA310351554GA SA310434099GA SA310528854GA SA310528898GA SA310528999GA SA310529027GA SA311039370GA SA311039372GA SA311039442GA SA310507064GA SA310507074GA SA310507078GA SA310507094GA SA310507096GA SA310507113GA SA310507125GA SA310507135GA SA310507143GA SA310507389GA SA310507398GA SA310507401GA SA310507404GA SA310507407GA SA310507414GA SA310507458GA SA310507475GA SA310517732GA SA310528859GA SA310528897GA SA310529029GA SA310529046GA SA310529165GA SA310529201GA SA310423583GA SA310444494GA SA310444864GA SA310496214GA SA310506656GA SA310506707GA SA310507021GA SA310528972GA SA311039391GA SA311039392GA SA311039488GA SA310518663GA SA310528968GA SA310260553GA SA310310926GA SA310382635GA SA310507040GA SA310507072GA SA310507073GA SA310507108GA SA310507129GA SA310507132GA SA310507397GA SA310507436GA SA310507484GA SA310517662GA SA310517828GA SA310517839GA SA310517986GA SA310517998GA SA310518003GA SA310528969GA SA311059870GA SA311059909GA SA310259884GA SA310310883GA SA310331249GA SA310382657GA SA310413473GA SA310413559GA SA310413560GA SA310423635GA SA310423662GA SA310424012GA SA310434242GA SA310444272GA SA310455033GA SA310475540GA SA310475681GA SA310496226GA SA310496471GA SA310496472GA SA310507070GA SA310507077GA SA310507100GA SA310507106GA SA310507111GA SA310507121GA SA310507388GA SA310507412GA SA310507444GA SA310507448GA SA310507451GA SA310507496GA SA310507503GA SA310507527GA SA310507548GA SA310517720GA SA310517737GA SA310517739GA SA310517818GA SA310517821GA SA310517838GA SA310517997GA SA310518677GA SA310528883GA SA310528887GA SA310528888GA SA310528889GA SA310528975GA SA310528980GA SA310528990GA SA310528997GA SA310529004GA SA310529005GA SA310529008GA SA310529048GA SA310529049GA SA310529050GA SA310529053GA SA310529057GA SA310529158GA SA310539321GA SA311059871GA SA311059872GA SA311059873GA SA311059874GA SA311059875GA SA311059876GA SA311059877GA SA311059878GA SA311059879GA SA311059880GA SA311059881GA SA311059882GA SA311059883GA SA311059884GA SA311059885GA SA311059886GA SA311059887GA SA311059888GA SA311059889GA SA311059890GA SA311059891GA SA311059892GA SA311059893GA SA311059894GA SA311059897GA SA311059898GA SA311059900GA SA311059901GA SA311059903GA SA311059904GA SA311059905GA SA311059906GA SA311059907GA SA311059908GA SA311069606GA SA311069726GA SA310423577GA SA310496623GA SA310506854GA SA310507138GA SA310517825GA SA310518004GA SA310528852GA SA310528987GA SA310529000GA SA310539349GA SA311059895GA SA311059896GA SA311059899GA SA311059902GA SA311060134GA SA311060135GA SA310260040GA SA310321001GA SA310351501GA SA310444289GA SA310444521GA SA310475613GA SA310507056GA SA310507541GA SA310517865GA SA310517988GA SA310518628GA SA310518653GA SA310518656GA SA310518664GA SA310518671GA SA311049576GA SA311060108GA SA311069619GA SA311069709GA SA311069710GA SA311069711GA SA311060129GA SA311060151GA SA311060217GA SA310413534GA SA310444248GA SA310444849GA SA310475703GA SA310496148GA SA310496288GA SA310496295GA SA310496310GA SA310496462GA SA310517841GA SA310517878GA SA310528871GA SA310539347GA SA310539351GA SA311039377GA SA311039450GA SA311039464GA SA311039467GA SA311039483GA SA311039485GA SA311060143GA SA311060153GA SA311060160GA SA311060169GA SA311069677GA SA311039377GA SA311060075GA SA311060220GA SA310310993GA SA310518400GA SA310529204GA SA310529207GA SA311059592GA SA311060165GA SA311060194GA SA311060209GA SA311060215GA SA311060218GA SA311060309GA SA311069780GA SA310496306GA SA310496314GA SA310496322GA SA310507018GA SA311049578GA SA311060070GA SA311060112GA SA311060115GA SA311060133GA SA311060137GA SA311060234GA SA311060239GA SA311060241GA SA311060242GA SA311060244GA SA311060247GA SA311060248GA SA311060249GA SA311060251GA SA311060261GA SA311060263GA SA311060264GA SA311060268GA SA311060275GA SA311060301GA SA311069730GA SA310351562GA SA310528923GA SA310528933GA SA310528935GA SA311059601GA SA311069657GA SA311069694GA SA311069697GA SA311069745GA SA311060142GA SA311060179GA SA311060184GA SA311060210GA SA311060211GA SA311069721GA SA310361698GA SA310496215GA SA310496220GA SA310496221GA SA310496232GA SA310496296GA SA310496371GA SA310496456GA SA310496457GA SA310496467GA SA310496474GA SA310496489GA SA310496490GA SA310496492GA SA310496499GA SA311060084GA SA311060096GA SA311060098GA SA311060381GA SA311070547GA SA311059581GA SA311060072GA SA311080605GA SA310518272GA SA310518278GA SA310518288GA SA310528880GA SA311060119GA SA311060124GA SA311060139GA SA311060141GA SA311060150GA SA311060152GA SA311060158GA SA311060176GA SA311060182GA SA311060199GA SA311060206GA SA311060231GA SA311060278GA SA311060285GA SA311060296GA SA311060346GA SA311060358GA SA311060363GA SA311069714GA SA311069729GA SA311070523GA SA311070537GA SA311070542GA SA311070574GA SA311070588GA SA311070593GA SA311080599GA SA311080632GA SA311080639GA SA311080704GA SA311090814GA SA311090826GA SA311090832GA SA310528777GA SA311059585GA SA311059588GA SA311059593GA SA311059596GA SA311060077GA SA311060085GA SA311060086GA SA311060104GA SA311060284GA SA311069819GA SA311070474GA SA311070538GA SA311070575GA SA311070579GA SA311070597GA SA311090893GA SA311070521GA SA311070576GA SA311070595GA SA311090834GA SA310518681GA SA311060166GA SA311060192GA SA311060283GA SA311060293GA SA311060294GA SA311060300GA SA311060304GA SA311060318GA SA311060325GA SA311060337GA SA311060349GA SA311060351GA SA311060354GA SA311060359GA SA311060368GA SA311060370GA SA311060389GA SA311080634GA SA311080636GA SA311080640GA SA311080643GA SA311080662GA SA311080710GA SA311080723GA SA311080726GA SA311080741GA SA311090813GA SA311090815GA SA311090819GA SA311090820GA SA310361945GA SA310496276GA SA310496317GA SA311060122GA SA311060127GA SA311060128GA SA311060130GA SA311060131GA SA311060132GA SA311060175GA SA311060177GA SA311060178GA SA311060187GA SA311060190GA SA311060191GA SA311060222GA SA311060223GA SA311060226GA SA311060272GA SA311060276GA SA311060286GA SA311060287GA SA311060288GA SA311060289GA SA311060290GA SA311060292GA SA311060306GA SA311060307GA SA311060312GA SA311060313GA SA311060314GA SA311060315GA SA311060319GA SA311060321GA SA311060344GA SA311060348GA SA311060357GA SA311060361GA SA311060362GA SA311060367GA SA311060390GA SA311069712GA SA311069727GA SA311070540GA SA311070544GA SA311070545GA SA311070546GA SA311070550GA SA311070551GA SA311070552GA SA311070554GA SA311070561GA SA311070563GA SA311070565GA SA311070566GA SA311070570GA SA311070582GA SA311070594GA SA311080600GA SA311080601GA SA311080603GA SA311080608GA SA311080610GA SA311080611GA SA311080614GA SA311080617GA SA311080618GA SA311080619GA SA311080620GA SA311080621GA SA311080622GA SA311080623GA SA311080624GA SA311080625GA SA311080637GA SA311080641GA SA311080649GA SA311080655GA SA311090837GA SA310496140GA SA311039510GA SA311039540GA SA311039547GA SA311059597GA SA311060069GA SA311060103GA SA311060188GA SA311060195GA SA311060200GA SA311060203GA SA311060216GA SA311060279GA SA311060282GA SA311060295GA SA311070524GA SA311070596GA SA311080645GA SA311080646GA SA311090775GA SA310528855GA SA311039520GA SA311039526GA SA311039528GA SA311039539GA SA311039545GA SA311059582GA SA311060004GA SA311060007GA SA311060023GA SA311060024GA SA311060028GA SA311060035GA SA311060041GA SA311060097GA SA311060236GA SA311060302GA SA311069640GA SA311069789GA SA311069807GA SA311069816GA SA311069817GA SA311069836GA SA311069837GA SA311069850GA SA311070518GA SA311070527GA SA311070529GA SA311070532GA SA311070533GA SA311070555GA SA311070572GA SA311070577GA SA311070578GA SA311070581GA SA311080627GA SA311080631GA SA311080656GA SA311080657GA SA311080712GA SA311080733GA SA311080739GA SA311080740GA SA311080742GA SA311090759GA SA311090766GA SA311090771GA SA311090774GA SA311090782GA SA311090785GA SA311090788GA SA311090791GA SA311090796GA SA311090801GA SA311090811GA SA311090812GA SA311090817GA SA311090818GA SA311090825GA SA311090829GA SA311090833GA SA311090836GA SA311090839GA SA311090840GA SA311090841GA SA311090844GA SA311090847GA SA311090848GA SA311090858GA SA311090867GA SA311090880GA SA311090881GA SA310507305GA SA311039537GA SA311060015GA SA311060037GA SA311060067GA SA311060088GA SA311060092GA SA311060106GA SA311070443GA SA311080671GA SA311080699GA SA311080702GA SA311090776GA SA311090781GA SA311070528GA SA311070534GA SA311070539GA SA311070569GA SA311090787GA SA311090795GA SA311090822GA SA311090824GA SA311101216GA SA310403374GA SA310423869GA SA310434153GA SA310506762GA SA310507164GA SA310507266GA SA310518181GA SA310528746GA SA311039494GA SA311049577GA SA311059580GA SA311059589GA SA311059600GA SA311060027GA SA311060074GA SA311069627GA SA311069638GA SA311069649GA SA311069658GA SA311069758GA SA311080667GA SA311080668GA SA311080688GA SA311080689GA SA311080692GA SA311090753GA SA311090763GA SA311090790GA SA311090794GA SA311090883GA SA311090885GA SA311101174GA SA311101175GA SA311101179GA SA311101192GA SA311101195GA SA311101198GA SA311101199GA SA311101202GA SA311101203GA SA311101214GA SA311101218GA SA311101236GA SA311101237GA SA311101238GA SA311101239GA SA311101242GA SA311101269GA SA310423723GA SA310496283GA SA310496366GA SA310507015GA SA311090806GA SA310434146GA SA311090986GA SA310506771GA SA311059587GA SA311059602GA SA311060017GA SA311060038GA SA311069625GA SA311069639GA SA311069641GA SA311069648GA SA311069651GA SA311069655GA SA311069656GA SA311069659GA SA311069674GA SA311069678GA SA311069785GA SA311069814GA SA311069820GA SA311069828GA SA311069835GA SA311069841GA SA311070476GA SA311080664GA SA311080665GA SA311080666GA SA311080669GA SA311080680GA SA311080684GA SA311080690GA SA311080698GA SA311080700GA SA311080701GA SA311090895GA SA311090897GA SA311090902GA SA311091009GA SA311091023GA SA311091026GA SA311091030GA SA311091037GA SA311091043GA SA311091045GA SA311091050GA SA311091054GA SA311091055GA SA311091058GA SA311091059GA SA311091060GA SA311101063GA SA311101070GA SA311101083GA SA311101280GA SA311101308GA SA311101319GA SA311101321GA SA311101340GA SA311101351GA SA311101359GA SA311101374GA SA311101376GA SA311101381GA SA310382849GA SA310495916GA SA310506786GA SA311039552GA SA311060008GA SA311060011GA SA311060012GA SA311060013GA SA311069653GA SA311069679GA SA311069692GA SA311069699GA SA311069753GA SA311069755GA SA311069762GA SA311069763GA SA311069765GA SA311069767GA SA311069769GA SA311069771GA SA311069772GA SA311069774GA SA311069777GA SA311069781GA SA311069782GA SA311069786GA SA311069787GA SA311069788GA SA311069803GA SA311069827GA SA311069840GA SA311069857GA SA311080663GA SA311080683GA SA311080685GA SA311080694GA SA311080719GA SA311080720GA SA311080730GA SA311080738GA SA311090747GA SA311090750GA SA311090757GA SA311090758GA SA311090772GA SA311090875GA SA311090884GA SA311090890GA SA311090937GA SA311090977GA SA311090978GA SA311090982GA SA311090983GA SA311090984GA SA311090985GA SA311091001GA SA311091018GA SA311091019GA SA311091022GA SA311091024GA SA311091057GA SA311101064GA SA311101065GA SA311101067GA SA311101069GA SA311101071GA SA311101082GA SA311101124GA SA311101311GA SA311101316GA SA311101317GA SA311101320GA SA311101323GA SA311101324GA SA311101325GA SA311101326GA SA311101327GA SA311101328GA SA311101330GA SA311101331GA SA311101337GA SA311101341GA SA311101348GA SA311101353GA SA311101356GA SA311101361GA SA311101365GA SA311101366GA SA311101369GA SA311101370GA SA311101375GA SA311101379GA SA311101430GA SA311101469GA SA311101475GA SA311060323GA SA311060353GA SA311060377GA SA311069665GA SA311090764GA SA311090808GA SA311090874GA SA311101204GA SA311101240GA SA311101253GA SA311101256GA SA311101284GA SA311101285GA SA311101289GA SA311101291GA SA311101298GA SA310403312GA SA310403346GA SA310529054GA SA310539316GA SA311069637GA SA311069644GA SA311090966GA SA311091014GA SA311091015GA SA311090852GA SA311090945GA SA310423772GA SA310518441GA SA311090969GA SA311091011GA SA311091013GA SA311101102GA SA311101119GA SA311101261GA SA311101262GA SA311101263GA SA311090786GA SA311091041GA SA311091051GA SA311132384GA SA311132386GA SA311132435GA SA311132465GA SA311132470GA SA311132507GA SA310517659GA SA310517667GA SA310517679GA SA311090865GA SA311090871GA SA311090889GA SA311101387GA SA311101411GA SA311101477GA SA311101480GA SA311101483GA SA311101494GA SA311101496GA SA311101502GA SA311101503GA SA311101517GA SA311101520GA SA311101521GA SA311101523GA SA311101556GA SA310413497GA SA310413537GA SA310496229GA SA310496303GA SA310507119GA SA311060116GA SA311060117GA SA311060118GA SA311060120GA SA311060123GA SA311060125GA SA311060144GA SA311060145GA SA311060159GA SA311060245GA SA311060253GA SA311060257GA SA311060258GA SA311060259GA SA311060262GA SA311069713GA SA311069715GA SA311069716GA SA311069718GA SA311069720GA SA311069725GA SA311069733GA SA311069735GA SA311059598GA SA311059599GA SA311069744GA SA311080613GA SA311090873GA SA311090963GA SA311091002GA SA311091056GA SA311101066GA SA311101068GA SA311101096GA SA311101104GA SA311101116GA SA311101117GA SA311060393GA SA311070520GA SA311070568GA SA311080652GA SA311080731GA SA311101111GA SA311101180GA SA311101187GA SA311101190GA SA311101228GA SA311101229GA SA311101234GA SA311101244GA SA311101245GA SA311101250GA SA311101264GA SA311101281GA SA311101419GA SA311101421GA SA311070522GA SA311080628GA SA311080711GA SA311080718GA SA311080721GA SA311080724GA SA311080725GA SA311080729GA SA311101304GA SA311101425GA SA311090882GA SA311101388GA SA311101401GA SA311101402GA SA311101391GA SA311101488GA SA311101495GA SA311101506GA SA311101507GA SA311101513GA SA311101515GA SA311101527GA SA311101528GA SA311101531GA SA311101535GA SA311101536GA SA311101544GA SA311111559GA SA311111562GA SA311111568GA SA311111574GA SA311111604GA SA311111610GA SA311101522GA SA311090762GA SA311090765GA SA311090854GA SA311101184GA SA311101217GA SA311101482GA SA311101499GA SA311121714GA SA311121718GA SA311070567GA SA311080635GA SA311080717GA SA311090748GA SA311090755GA SA311090777GA SA311090784GA SA311090807GA SA311090810GA SA311090816GA SA311090868GA SA311101212GA SA311101448GA SA311090894GA SA311090899GA SA311090904GA SA311090959GA SA311090993GA SA311090994GA SA311101074GA SA311101079GA SA311101107GA SA311101163GA SA311101428GA SA311101436GA SA311101446GA SA311101454GA SA311101478GA SA311101505GA SA311111623GA SA311111626GA SA311111629GA SA311121700GA SA311121703GA SA311121749GA SA311132422GA SA311132425GA SA311132455GA SA311132461GA SA311132464GA SA311132490GA SA311132495GA SA311132502GA SA311132505GA SA311132515GA SA311132517GA SA311070549GA SA311090954GA SA311090991GA SA311090996GA SA311091000GA SA311101125GA SA311101130GA SA311101168GA SA311101392GA SA311101398GA SA311101399GA SA311101406GA SA311101409GA SA311101413GA SA311101414GA SA311101423GA SA311101550GA SA311101552GA SA311111557GA SA311111561GA SA311111567GA SA311111573GA SA311111577GA SA311111600GA SA311111609GA SA311121698GA SA311121740GA SA311121741GA SA311121746GA SA311132278GA SA311132441GA SA311132444GA SA311132452GA SA311132460GA SA311132463GA SA311132488GA SA311132491GA SA311132494GA SA311132500GA SA311132508GA SA311132510GA SA311132522GA SA311132527GA SA311060329GA SA311060366GA SA311060374GA SA311080714GA SA311080735GA SA311080743GA SA311080744GA SA311090752GA SA311090756GA SA311090809GA SA311090866GA SA311101519GA SA311101543GA SA311111572GA SA311111575GA SA311111580GA SA311111582GA SA311111594GA SA311111603GA SA311122216GA SA311101383GA SA311101405GA SA311101486GA SA311101532GA SA311101537GA SA311101546GA SA311101547GA SA311111566GA SA311111581GA SA311111593GA SA311070473GA SA311090953GA SA311090971GA SA311101429GA SA311101435GA SA311101473GA SA311101479GA SA311101497GA SA311101514GA SA311101529GA SA311101540GA SA311111601GA SA311121708GA SA311121725GA SA311121742GA SA311121743GA SA311122127GA SA311122128GA SA311132413GA SA311132430GA SA311132438GA SA311132440GA SA311132449GA SA311132454GA SA311132499GA SA311101134GA SA311101334GA SA311101427GA SA311101459GA SA311101498GA SA311101504GA SA311101525GA SA311101538GA SA311101542GA SA311101545GA SA311101549GA SA311101554GA SA311111560GA SA311111571GA SA311111578GA SA311132496GA SA311090908GA SA311090958GA SA311132327GA SA311132521GA SA311132523GA SA311132421GA SA311132509GA SA310507139GA SA311060392GA SA311070477GA SA311070480GA SA311090926GA SA311101177GA SA311101201GA SA311101230GA SA311101303GA SA311101306GA SA311101358GA SA311101372GA SA311101382GA SA311101407GA SA311101412GA SA311101415GA SA311101491GA SA311101493GA SA311101511GA SA311101512GA SA311101516GA SA311101541GA SA311101555GA SA311111558GA SA311111564GA SA311122204GA SA310372076GA SA310382364GA SA310382814GA SA311059584GA SA311060395GA SA311069630GA SA311069645GA SA311069652GA SA311069660GA SA311069663GA SA311069666GA SA311069667GA SA311069688GA SA311069695GA SA311069696GA SA311069701GA SA311069704GA SA311069705GA SA311069706GA SA311090769GA SA311101182GA SA311101200GA SA311101219GA SA311101232GA SA311101233GA SA311101246GA SA311101277GA SA311101290GA SA311101301GA SA311101307GA SA311101416GA SA311101417GA SA311101476GA SA311101485GA SA311101487GA SA311101489GA SA311101510GA SA311101526GA SA311101534GA SA311101539GA SA311111563GA SA311121744GA SA311070509GA SA311090909GA SA311090919GA SA311090923GA SA311090925GA SA311090927GA SA311090939GA SA311090946GA SA311090947GA SA311090950GA SA311090951GA SA311090952GA SA311090967GA SA311090970GA SA311090989GA SA311091038GA SA311101076GA SA311101108GA SA311101115GA SA311101118GA SA311101120GA SA311101349GA SA311101350GA SA311101367GA SA311101371GA SA311101377GA SA311121729GA SA311121730GA SA311122168GA SA311122214GA SA311122215GA SA311132282GA SA311122170GA SA311122176GA SA311122191GA SA311122194GA SA311122197GA SA311122232GA SA311122274GA SA311132281GA SA311132294GA SA311132302GA SA311132324GA SA311090910GA SA311090931GA SA311090936GA SA311090956GA SA311090957GA SA311090974GA SA311101081GA SA311101122GA SA311101128GA SA311101145GA SA311101149GA SA311101150GA SA311101167GA SA311101171GA SA311101332GA SA311101433GA SA311111624GA SA311111633GA SA311122109GA SA311122117GA SA311122161GA SA311132417GA SA311132492GA SA311080626GA SA311122012GA SA311122074GA SA311122080GA SA311122136GA SA311122153GA SA311122171GA SA311122172GA SA311122182GA SA311122184GA SA311122201GA SA311122208GA SA311122209GA SA311122225GA SA311122239GA SA311122256GA SA311122260GA SA311132303GA SA311132308GA SA311132311GA SA311132312GA SA311132323GA SA311132328GA SA311132332GA SA311132333GA SA311132336GA SA311132338GA SA311132339GA SA311132341GA SA311132346GA SA311132347GA SA311132355GA SA311132356GA SA311132357GA SA311132359GA SA311132361GA SA311132364GA SA311132365GA SA311132390GA SA311132394GA SA311132403GA SA311069770GA SA311069830GA SA311090988GA SA311090990GA SA311091005GA SA311101148GA SA311101346GA SA311121707GA SA311121727GA SA311121728GA SA311121733GA SA311121735GA SA311121738GA SA311121851GA SA311121876GA SA311122056GA SA311122061GA SA311122067GA SA311122124GA SA311122159GA SA311122163GA SA311122178GA SA311122181GA SA311122183GA SA311122202GA SA311122210GA SA311122213GA SA311122217GA SA311122218GA SA311122223GA SA311122228GA SA311122233GA SA311122243GA SA311122246GA SA311122250GA SA311122253GA SA311122255GA SA311122261GA SA311122263GA SA311122264GA SA311122266GA SA311122268GA SA311122269GA SA311122273GA SA311132285GA SA311132289GA SA311132292GA SA311132293GA SA311132304GA SA311132305GA SA311132306GA SA311132329GA SA311132337GA SA311132342GA SA311132345GA SA311132351GA SA311132353GA SA311132354GA SA311132362GA SA311132392GA SA311132400GA SA311132506GA SA311132526GA SA311069654GA SA311069668GA SA311069675GA SA311069683GA SA311069689GA SA311069702GA SA311069708GA SA311069759GA SA311069839GA SA311069849GA SA311069851GA SA311069854GA SA311069856GA SA311069860GA SA311090903GA SA311090980GA SA311091021GA SA311091028GA SA311091031GA SA311091032GA SA311091042GA SA311091044GA SA311091048GA SA311091061GA SA311091062GA SA311101131GA SA311101309GA SA311101310GA SA311101318GA SA311101322GA SA311101335GA SA311101339GA SA311101343GA SA311101344GA SA311101352GA SA311101363GA SA311101380GA SA311101426GA SA311101437GA SA311101438GA SA311101447GA SA311101462GA SA311101471GA SA311101472GA SA311101474GA SA311121798GA SA311121804GA SA311121847GA SA311122108GA SA311122132GA SA311122138GA SA311122155GA SA311122167GA SA311132382GA SA311132557GA SA311132589GA SA311070478GA SA311070479GA SA311070482GA SA311090965GA SA311090997GA SA311091008GA SA311091010GA SA311091016GA SA311091017GA SA311091020GA SA311091025GA SA311101313GA SA311101333GA SA311101338GA SA311101342GA SA311101345GA SA311101347GA SA311101354GA SA311101355GA SA311101357GA SA311101360GA SA311101364GA SA311111608GA SA311122231GA SA311122237GA SA311122240GA SA311122241GA SA311122242GA SA311122247GA SA311122249GA SA311122254GA SA311132300GA SA311132309GA SA311132310GA SA311132315GA SA311132316GA SA311132317GA SA311132319GA SA311132320GA SA311132334GA SA311132335GA SA311132349GA SA311132350GA SA311132363GA SA311132368GA SA311132402GA SA311132407GA SA311132411GA SA311059595GA SA311069822GA SA311069824GA SA311069842GA SA311069847GA SA311090928GA SA311111687GA SA311121712GA SA311121723GA SA311121734GA SA311121755GA SA311121758GA SA311121766GA SA311121767GA SA311121800GA SA311121802GA SA311121803GA SA311121807GA SA311121808GA SA311121821GA SA311121828GA SA311121831GA SA311121835GA SA311121848GA SA311121946GA SA311121999GA SA311122001GA SA311122060GA SA311122113GA SA311122131GA SA311122185GA SA311122234GA SA311122245GA SA311122252GA SA311122265GA SA311122271GA SA311122272GA SA311132276GA SA311132296GA SA311132298GA SA311132301GA SA311132307GA SA311132331GA SA311132343GA SA311132344GA SA311132358GA SA311132360GA SA311132369GA SA311132370GA SA311132374GA SA311132375GA SA311132376GA SA311132378GA SA311132381GA SA311132383GA SA311132385GA SA311132391GA SA311132393GA SA311132395GA SA311132396GA SA311132397GA SA311132401GA SA311132405GA SA311132408GA SA311132409GA SA311132410GA SA311132434GA SA311132457GA SA311132458GA SA311132459GA SA311132467GA SA311132468GA SA311132472GA SA311132476GA SA311132479GA SA311132481GA SA311132482GA SA311132533GA SA311132534GA SA311132537GA SA311132549GA SA311132550GA SA311132552GA SA311132553GA SA311132555GA SA311132570GA SA311132581GA SA311132582GA SA311132585GA SA311132587GA SA311132591GA SA311132595GA SA311132597GA SA311070475GA SA311070481GA SA311090896GA SA311090905GA SA311090906GA SA311090912GA SA311090917GA SA311090929GA SA311090930GA SA311090942GA SA311090943GA SA311090961GA SA311090962GA SA311090987GA SA311101095GA SA311101110GA SA311101112GA SA311111689GA SA311121768GA SA311121773GA SA311121777GA SA311121782GA SA311121783GA SA311121792GA SA311121806GA SA311121809GA SA311121813GA SA311121816GA SA311121817GA SA311121827GA SA311122020GA SA311122024GA SA311132418GA SA311132433GA SA311132439GA SA311132447GA SA311132448GA SA310392898GA SA310434058GA SA310434123GA SA310434137GA SA310465295GA SA310496410GA SA310496411GA SA310507316GA SA310528810GA SA310528825GA SA310529115GA SA311060020GA SA311069676GA SA311069681GA SA311069682GA SA311070409GA SA311080682GA SA311080687GA SA311080695GA SA311090940GA SA311090941GA SA311090960GA SA311122118GA SA311122119GA SA311122123GA SA311101302GA SA311122034GA SA311142754GA SA311142758GA SA311142760GA SA311142763GA SA311142765GA SA311142768GA SA311142769GA SA311142775GA SA311142779GA SA311142780GA SA311142782GA SA311142784GA SA311142786GA SA311142788GA SA311142759GA SA311142767GA SA311101283GA SA311121867GA SA311121877GA SA311122017GA SA311122051GA SA311122211GA SA311142973GA SA311101432GA SA311142748GA SA311142749GA SA311142756GA SA311142783GA SA311142791GA SA311142793GA SA311142941GA SA311142979GA SA311101143GA SA311101155GA SA311101441GA SA311101442GA SA311101451GA SA311101456GA SA311101461GA SA311101463GA SA311101464GA SA311101465GA SA311111659GA SA311101078GA SA311101091GA SA311101113GA SA311101126GA SA311101136GA SA311101158GA SA311101164GA SA311101165GA SA311101450GA SA311101467GA SA311101468GA SA311111631GA SA311111680GA SA311111681GA SA311121717GA SA311121748GA SA311121765GA SA311122014GA SA311122022GA SA311122032GA SA311122033GA SA311122035GA SA311122050GA SA311122133GA SA311122135GA SA311122149GA SA311069844GA SA311069845GA SA311090901GA SA311090932GA SA311090975GA SA311090981GA SA311111688GA SA311111694GA SA311121775GA SA311121780GA SA311121822GA SA311121823GA SA311121830GA SA311121899GA SA311122104GA SA311132498GA SA311132504GA SA311132512GA SA311132514GA SA311132558GA SA311132561GA SA311132563GA SA311132565GA SA311132573GA SA311132574GA SA311132577GA SA311132586GA SA311142764GA SA311142792GA SA311142873GA SA311142876GA SA311142877GA SA311142882GA SA311142892GA SA311142896GA SA311142896GA SA311142898GA SA311142901GA SA311142904GA SA311142922GA SA311142944GA SA311142945GA SA311142969GA SA311142976GA SA311111662GA SA311111663GA SA311111667GA SA311111671GA SA311111674GA SA311111676GA SA311111677GA SA311111679GA SA311111685GA SA311111686GA SA311111695GA SA311111696GA SA311101086GA SA311101087GA SA311101109GA SA311101458GA SA311111588GA SA311111589GA SA311111597GA SA311111598GA SA311111605GA SA311111613GA SA311111625GA SA311111637GA SA311121721GA SA311122065GA SA311122079GA SA311122082GA SA311142890GA SA311101147GA SA311111570GA SA311111576GA SA311111585GA SA311111587GA SA311111590GA SA311111591GA SA311111592GA SA311111596GA SA311111607GA SA311111612GA SA311121858GA SA311121868GA SA311121880GA SA311121892GA SA311122025GA SA311132578GA SA311142874GA SA311142883GA SA311142886GA SA311142900GA SA311142905GA SA311101431GA SA311121724GA SA311121820GA SA311121992GA SA311122003GA SA311122027GA SA311122042GA SA311122049GA SA311122084GA SA311122088GA SA311122098GA SA311122114GA SA311122116GA SA311142921GA SA311142943GA SA311153110GA SA311121890GA SA311132415GA SA311132572GA SA311132579GA SA311132592GA SA311132594GA SA311142789GA SA311142879GA SA311142880GA SA311142891GA SA311153124GA SA311142700GA SA311153101GA SA311153149GA SA311153161GA SA311153164GA SA311153206GA SA311153209GA SA311153210GA SA311101101GA SA311142707GA SA311142823GA SA311142910GA SA311142913GA SA311142954GA SA311142963GA SA311142971GA SA311142972GA SA311142982GA SA311153112GA SA311153172GA SA311090892GA SA311090898GA SA311121751GA SA311121770GA SA311121776GA SA311121833GA SA311122085GA SA311122095GA SA311122105GA SA311122120GA SA311122150GA SA311122156GA SA311142802GA SA311060386GA SA311090878GA SA311090888GA SA311121856GA SA311121885GA SA311121891GA SA311121897GA SA311121909GA SA311122011GA SA311122175GA SA311122186GA SA311122200GA SA311122205GA SA311132325GA SA311132453GA SA311142790GA SA311142931GA SA311142974GA SA311153080GA SA311153144GA SA311153154GA SA311153157GA SA311153163GA SA311153192GA SA311153213GA SA311153217GA SA311153223GA SA311153225GA SA310475517GA SA310506675GA SA310506696GA SA310506700GA SA310506731GA SA310506733GA SA310506734GA SA310506739GA SA310506744GA SA310506754GA SA311132562GA SA311142785GA SA311142920GA SA311153084GA SA311153090GA SA311153115GA SA311153165GA SA311153169GA SA311153176GA SA311142679GA SA311142689GA SA311142696GA SA311132416GA SA311132427GA SA311069684GA SA311111654GA SA311111655GA SA311111657GA SA311121981GA SA311142984GA SA311101453GA SA311101470GA SA311090838GA SA311132606GA SA311132612GA SA311132623GA SA311132636GA SA311142781GA SA311142795GA SA311132590GA SA311069796GA SA311153068GA SA311153230GA SA311153236GA SA311153238GA SA311153240GA SA311153243GA SA311153245GA SA311153248GA SA311153249GA SA311153250GA SA311153275GA SA311153276GA SA311153281GA SA311153287GA SA311153288GA SA311153293GA SA311153294GA SA311153295GA SA311153297GA SA311153300GA SA311039495GA SA311059590GA SA311059603GA SA311069650GA SA311069747GA SA311069760GA SA311069764GA SA311069768GA SA311069779GA SA311069792GA SA311069793GA SA311069794GA SA311069795GA SA311069798GA SA311069799GA SA311069801GA SA311069804GA SA311069813GA SA311069833GA SA311069843GA SA311069853GA SA311121799GA SA311121843GA SA311153043GA SA311060396GA SA311153216GA SA311163393GA SA311142719GA SA311142907GA SA311142911GA SA311153096GA SA311153104GA SA311153111GA SA311153118GA SA311153136GA SA311153170GA SA311153188GA SA311153190GA SA311153194GA SA311153195GA SA311153196GA SA311153197GA SA311153219GA SA311153266GA SA308398742GR SA308511625GR SA309092523GR SA309143016GR SA309283997GR SA309283999GR SA309304262GR SA309304272GR SA309365227GR SA309365232GR SA309365347GR SA309517572GR SA310028114GR SA310028140GR SA310149355GR SA310259951GR SA310260386GR SA310372203GR SA310382325GR SA310382326GR SA310393186GR SA310423753GR SA310423978GR SA310507320GR SA311111692GA SA311121788GA SA311121790GA SA311121796GA SA311121801GA SA311121814GA SA311121815GA SA311121819GA SA311121829GA SA311121832GA SA311121838GA SA311121839GA SA311121840GA SA311121844GA SA311121845GA SA311132372GA SA311132389GA SA311132529GA SA311132616GA SA311132630GA SA311142713GA SA311142716GA SA311142718GA SA311142745GA SA311142752GA SA311142798GA SA311142804GA SA311142806GA SA311142903GA SA311142929GA SA311142964GA SA311142966GA SA311153088GA SA311153089GA SA311153094GA SA311153108GA SA311153114GA SA311153117GA SA311153119GA SA311153120GA SA311153129GA SA311153132GA SA311153138GA SA311153139GA SA311153146GA SA311153150GA SA311153162GA SA311153181GA SA311153189GA SA311153226GA SA311153256GA SA311153350GA SA311153351GA SA311153352GA SA311163376GA SA311163398GA SA311163884GA SA311163899GA SA311163981GA SA311101273GA SA311111569GA SA311121701GA SA311122016GA SA311122023GA SA311122039GA SA311122041GA SA311122048GA SA311142885GA SA311142894GA SA311142935GA SA311142936GA SA311142937GA SA311142938GA SA311142940GA SA311142942GA SA311142946GA SA311142958GA SA311163412GA SA311163873GA SA311163900GA SA311163923GA SA311163926GA SA311163928GA SA311163941GA SA311163948GA SA311163950GA SA311163977GA SA311163989GA SA311163995GA SA311163996GA SA311142720GA SA311153317GA SA311163783GA SA311153036GA SA311111665GA SA311111666GA SA311111668GA SA311111670GA SA311111678GA SA311111682GA SA311111683GA SA311111684GA SA311111690GA SA311111693GA SA311121752GA SA311121759GA SA311121763GA SA311121769GA SA311121772GA SA311121785GA SA311121791GA SA311121793GA SA311121794GA SA311121795GA SA311121824GA SA311121825GA SA311132483GA SA311132610GA SA311132620GA SA311153065GA SA311163872GA SA311163905GA SA311163968GA SA311142678GA SA311142717GA SA311142727GA SA311142739GA SA311142824GA SA311142852GA SA311153241GA SA311153247GA SA311153257GA SA311153264GA SA311153272GA SA311153299GA SA311163901GA SA311163902GA SA311163903GA SA311163904GA SA311163916GA SA311163919GA SA311163949GA SA311163970GA SA311163986GA SA311163991GA SA311163992GA SA309283983GR SA309324548GR SA309334756GR SA309365225GR SA309365229GR SA309365237GR SA309375364GR SA309416005GR SA309476874GR SA309476893GR SA309497149GR SA309517559GR SA309517853GR SA310028149GR SA310260303GR SA310260308GR SA310260359GR SA310372167GR SA310382347GR SA310393263GR SA310517610GR SA310518235GR SA310518390GR SA310518423GR SA311163390GA SA311163862GA SA311132604GA SA311142721GA SA311153032GA SA311153047GA SA311153048GA SA311153052GA SA311153060GA SA311163403GA SA311163416GA SA311163858GA SA311163878GA SA311163891GA SA311163896GA SA311163906GA SA311163907GA SA311163927GA SA311163957GA SA311163959GA SA311195079GA SA311205456GA SA311205486GA SA311205501GA SA311091027GA SA311163401GA SA311163410GA SA311163857GA SA311163860GA SA311163868GA SA311163871GA SA311163876GA SA311163880GA SA311163881GA SA311163882GA SA311163886GA SA311163894GA SA311163911GA SA311163913GA SA311163924GA SA311163938GA SA311163942GA SA311163952GA SA311163953GA SA311163954GA SA311163963GA SA311163965GA SA311163966GA SA311163967GA SA311163969GA SA311163982GA SA311163985GA SA311163990GA SA311256694GA SA311163369GA SA311163406GA SA311195056GA SA311205463GA SA311205480GA SA311163628GA SA311163639GA SA311184454GA SA311163782GA SA311163846GA SA311163936GA SA311163937GA SA311174166GA SA311174181GA SA311184355GA SA311184437GA SA311163936GA SA311174166GA SA311163937GA SA311174181GA SA311163915GA SA311163922GA SA311163932GA SA311174158GA SA311174165GA SA311174180GA SA311174186GA SA311174187GA SA311174189GA SA311174195GA SA311174196GA SA311174200GA SA311174205GA SA311174208GA SA311174213GA SA311174214GA SA311174215GA SA311174216GA SA311174217GA SA311174219GA SA311174225GA SA311174226GA SA311174227GA SA311174229GA SA311174231GA SA311174232GA SA311174233GA SA311174246GA SA311174247GA SA311174248GA SA311174249GA SA311174250GA SA311174251GA SA311174253GA SA311184336GA SA311184343GA SA311184346GA SA311184350GA SA311184351GA SA311184352GA SA311184354GA SA311184356GA SA311184357GA SA311184358GA SA311184361GA SA311184365GA SA311184368GA SA311184369GA SA311184374GA SA311184378GA SA311184381GA SA311184382GA SA311184385GA SA311184389GA SA311184395GA SA311184403GA SA311184404GA SA311184405GA SA311184408GA SA311184409GA SA311184411GA SA311184540GA SA311184550GA SA311184555GA SA311184561GA SA311184580GA SA311184586GA SA311184596GA SA311287202GA SA311153055GA SA311153319GA SA311153333GA SA311153342GA SA311153347GA SA311163430GA SA311163435GA SA311163436GA SA311163459GA SA311163460GA SA311163466GA SA311163469GA SA311163483GA SA311163490GA SA311163567GA SA311163696GA SA311163710GA SA311163723GA SA311163724GA SA311163726GA SA311163784GA SA311163799GA SA311163800GA SA311163917GA SA311174004GA SA311174016GA SA311174019GA SA311174035GA SA311174073GA SA311174080GA SA311174191GA SA311174244GA SA311184407GA SA311184415GA SA311184416GA SA311184418GA SA311184420GA SA311184425GA SA311184426GA SA311184432GA SA311184434GA SA311184435GA SA311184436GA SA311184441GA SA311184444GA SA311184446GA SA311184447GA SA311184448GA SA311184452GA SA311184455GA SA311184458GA SA311184459GA SA311184462GA SA311184522GA SA311184523GA SA311184531GA SA311184548GA SA311184549GA SA311184563GA SA311184601GA SA311184602GA SA311184605GA SA311184609GA SA311121995GA SA311142722GA SA311142848GA SA311143015GA SA311153056GA SA311153072GA SA311153237GA SA311153318GA SA311153335GA SA311153340GA SA311163446GA SA311163453GA SA311163606GA SA311194950GA SA311194961GA SA311194965GA SA311194977GA SA311194986GA SA311194996GA SA311194998GA SA311195000GA SA311195003GA SA311195016GA SA311195037GA SA311195038GA SA311195040GA SA311205290GA SA311205291GA SA311205292GA SA311205294GA SA311205295GA SA311205296GA SA311205297GA SA311205298GA SA311205299GA SA311205300GA SA311205301GA SA311205302GA SA311205303GA SA311205307GA SA311205308GA SA311205309GA SA311205312GA SA311205314GA SA311205316GA SA311205319GA SA311205320GA SA311205321GA SA311205324GA SA311205326GA SA311205327GA SA311205330GA SA311205332GA SA311205333GA SA311205334GA SA311205335GA SA311205336GA SA311205338GA SA311132639GA SA311142714GA SA311142821GA SA311142854GA SA311142990GA SA311143001GA SA311143004GA SA311153044GA SA311153050GA SA311153059GA SA311153063GA SA311174133GA SA311174164GA SA311174168GA SA311174172GA SA311174194GA SA311174221GA SA311174222GA SA311184359GA SA311184360GA SA311184364GA SA311184366GA SA311184367GA SA311184392GA SA311184394GA SA311184396GA SA311184414GA SA311225993GA SA311256687GA SA311256695GA SA311256788GA SA311153228GA SA311153233GA SA311153246GA SA311153251GA SA311153252GA SA311153253GA SA311153261GA SA311153263GA SA311153267GA SA311153269GA SA311153270GA SA311153274GA SA311153277GA SA311153279GA SA311153285GA SA311153291GA SA311153296GA SA311153304GA SA311163698GA SA311184417GA SA311163479GA SA311174122GA SA311184421GA SA311184424GA SA311184428GA SA311184431GA SA311184440GA SA311184450GA SA311184479GA SA311184495GA SA311184663GA SA311195070GA SA311195081GA SA311195082GA SA311195083GA SA311195086GA SA311195088GA SA311195090GA SA311195097GA SA311195099GA SA311195100GA SA311195104GA SA311195105GA SA311195107GA SA311195110GA SA311195111GA SA311195112GA SA311195113GA SA311195119GA SA311195120GA SA311195122GA SA311195259GA SA311195260GA SA311195262GA SA311195263GA SA311195279GA SA311195282GA SA311205310GA SA311236076GA SA311236085GA SA311236092GA SA311256709GA SA311256766GA SA311256767GA SA311256768GA SA311256771GA SA311256773GA SA311256775GA SA311256778GA SA311256783GA SA311256789GA SA311256791GA SA311256792GA SA311256795GA SA311256796GA SA311256800GA SA311256801GA SA311266899GA SA311266952GA SA311266963GA SA311266977GA SA311266980GA SA311266991GA SA311267009GA SA311060328GA SA311060355GA SA311142996GA SA311163417GA SA311174075GA SA311174240GA SA311174254GA SA311184397GA SA311184412GA SA311184487GA SA311184665GA SA311184666GA SA311184667GA SA311195053GA SA311195055GA SA311195057GA SA311195058GA SA311195059GA SA311195073GA SA311195077GA SA311246546GA SA311246562GA SA311256786GA SA311266959GA SA311174066GA SA311184456GA SA311184463GA SA311184467GA SA311184469GA SA311184472GA SA311184475GA SA311184478GA SA311184481GA SA311184499GA SA311184503GA SA311184509GA SA311195092GA SA311195096GA SA311195108GA SA311195115GA SA311195117GA SA311195151GA SA311205448GA SA311205450GA SA311205451GA SA311205455GA SA311205461GA SA311205465GA SA311205471GA SA311205473GA SA311205477GA SA311205478GA SA311205483GA SA311205484GA SA311205499GA SA311205500GA SA311121983GA SA311122008GA SA311142997GA SA311143007GA SA311153301GA SA311153346GA SA311163440GA SA311163561GA SA311163585GA SA311174072GA SA311174084GA SA311174144GA SA311184429GA SA311184439GA SA311184466GA SA311184483GA SA311184494GA SA311184496GA SA311246517GA SA311246529GA SA311246534GA SA311246542GA SA311266955GA SA311266972GA SA311266973GA SA311266974GA SA311266976GA SA311266978GA SA311266979GA SA311266982GA SA311266983GA SA311266985GA SA311266988GA SA311266990GA SA311266997GA SA311267011GA SA310382768GA SA311060376GA SA311070519GA SA311121837GA SA311122072GA SA311142856GA SA311153069GA SA311153231GA SA311153234GA SA311153280GA SA311153290GA SA311153292GA SA311153298GA SA311153303GA SA311153326GA SA311153357GA SA311153365GA SA311163428GA SA311163431GA SA311163437GA SA311163439GA SA311163442GA SA311163447GA SA311163449GA SA311163464GA SA311163468GA SA311163473GA SA311163480GA SA311163487GA SA311163545GA SA311163552GA SA311163560GA SA311163580GA SA311163613GA SA311163623GA SA311163624GA SA311163645GA SA311184390GA SA311184430GA SA311184445GA SA311184465GA SA311184489GA SA311194976GA SA311195008GA SA311195050GA SA311195098GA SA311195102GA SA311195125GA SA311195126GA SA311195128GA SA311195129GA SA311195132GA SA311195137GA SA311195139GA SA311195141GA SA311195143GA SA311195144GA SA311195149GA SA311195162GA SA311195163GA SA311195166GA SA311195185GA SA311195217GA SA311195220GA SA311195222GA SA311195224GA SA311195225GA SA311195226GA SA311195228GA SA311195229GA SA311195232GA SA311195237GA SA311195239GA SA311195241GA SA311195246GA SA311195253GA SA311195273GA SA311195280GA SA311195281GA SA311195286GA SA311205341GA SA311205343GA SA311205348GA SA311205353GA SA311205355GA SA311205356GA SA311205357GA SA311205359GA SA311205363GA SA311205364GA SA311205449GA SA311205452GA SA311205453GA SA311205457GA SA311205458GA SA311205469GA SA311205474GA SA311205479GA SA311205481GA SA311205482GA SA311205487GA SA311205488GA SA311205490GA SA311205494GA SA311205495GA SA311225976GA SA311225987GA SA311226002GA SA311236082GA SA311236098GA SA311236104GA SA311236116GA SA311236286GA SA311236288GA SA311236293GA SA311236337GA SA311246443GA SA311246516GA SA311246518GA SA311246520GA SA311246522GA SA311246523GA SA311246524GA SA311246525GA SA311246527GA SA311246528GA SA311246531GA SA311246537GA SA311246543GA SA311246545GA SA311246547GA SA311246551GA SA311246552GA SA311246559GA SA311246560GA SA311246566GA SA311246568GA SA311246570GA SA311246571GA SA311246573GA SA311246576GA SA311246581GA SA311256769GA SA311256787GA SA311256797GA SA311256799GA SA311266953GA SA311266954GA SA311266956GA SA311266957GA SA311266958GA SA311266960GA SA311266965GA SA311266967GA SA311266968GA SA311266969GA SA311266970GA SA311266971GA SA311266975GA SA311266981GA SA311266984GA SA311266987GA SA311266989GA SA311267000GA SA311267003GA SA311267004GA SA311267006GA SA311267007GA SA311267010GA SA311142816GA SA311153037GA SA311153039GA SA311153054GA SA311153071GA SA311153255GA SA311153282GA SA311153309GA SA311153311GA SA311153356GA SA311153358GA SA311163433GA SA311163441GA SA311163443GA SA311163445GA SA311163470GA SA311163484GA SA311163584GA SA311163587GA SA311163592GA SA311163617GA SA311163649GA SA311184493GA SA311184688GA SA311194940GA SA311194966GA SA311194971GA SA311194997GA SA311195011GA SA311195012GA SA311195013GA SA311195014GA SA311195039GA SA311195041GA SA311195173GA SA311195176GA SA311195192GA SA311195193GA SA311195195GA SA311195197GA SA311195202GA SA311195216GA SA311195219GA SA311195221GA SA311195234GA SA311195244GA SA311236296GA SA311236297GA SA311236312GA SA311236319GA SA311236334GA SA311246398GA SA311246418GA SA311246526GA SA311246530GA SA311246549GA SA311246567GA SA311163448GA SA311163729GA SA311163801GA SA311174053GA SA311174060GA SA311174062GA SA311184419GA SA311184427GA SA311184433GA SA311184443GA SA311184451GA SA311184484GA SA311184484GAMKE SA311184492GA SA311184500GA SA311184502GA SA311184506GA SA311184671GA SA311184674GA SA311184680GA SA311184683GA SA311184685GA SA311184687GA SA311184703GA SA311184704GA SA311184705GA SA311184707GA SA311184708GA SA311184712GA SA311184713GA SA311184715GA SA311195227GA SA311195230GA SA311195231GA SA311195233GA SA311195242GA SA311205443GA SA311205446GA SA311205460GA SA311205467GA SA311205492GA SA311236292GA SA311236323GA SA311236326GA SA311236332GA SA311236338GA SA311163489GA SA311163542GA SA311174149GA SA311195064GA SA311195089GA SA311195091GA SA311195101GA SA311195106GA SA311195118GA SA311195123GA SA311195124GA SA311195133GA SA311195135GA SA311195140GA SA311195142GA SA311195145GA SA311195148GA SA311195157GA SA311195160GA SA311195161GA SA311195165GA SA311195168GA SA311195169GA SA311205445GA SA311205464GA SA311236079GA SA311236187GA SA311236244GA SA311236276GA SA311236344GA SA311256721GA SA311142686GA SA311142730GA SA311142801GA SA311142815GA SA311142853GA SA311174085GA SA311174131GA SA311184480GA SA311184507GA SA311184513GA SA311184643GA SA311184675GA SA311184676GA SA311184698GA SA311184702GA SA311184709GA SA311184710GA SA311205362GA SA311225980GA SA311226000GA SA311236075GA SA311236081GA SA311236087GA SA311236090GA SA311236094GA SA311236097GA SA311236100GA SA311236101GA SA311236102GA SA311236110GA SA311236111GA SA311236294GA SA311236341GA SA311111638GA SA311111640GA SA311111641GA SA311111669GA SA311111675GA SA311121747GA SA311121750GA SA311121760GA SA311121761GA SA311121764GA SA311121789GA SA311122069GA SA311122125GA SA311122130GA SA311142723GA SA311142746GA SA311163971GA SA311174176GA SA311184485GA SA311184628GA SA311184630GA SA311184631GA SA311184640GA SA311184657GA SA311184662GA SA311195067GA SA311195080GA SA311236077GA SA311236285GA SA311236290GA SA311236291GA SA311236322GA SA311236327GA SA311236335GA SA311236340GA SA311163815GA SA311174156GA SA311184529GA SA311184535GA SA311184624GA SA311184627GA SA311184629GA SA311184644GA SA311184654GA SA311184659GA SA311184695GA SA311184714GA SA311195215GA SA311153244GA SA311153278GA SA311153302GA SA311153320GA SA311153323GA SA311153327GA SA311153329GA SA311153339GA SA311153341GA SA311153343GA SA311153353GA SA311153359GA SA311153362GA SA311163562GA SA311163586GA SA311163625GA SA311163640GA SA311163641GA SA311174059GA SA311174064GA SA311174126GA SA311184468GA SA311184504GA SA311184505GA SA311184679GA SA311184681GA SA311184692GA SA311184717GA SA311184718GA SA311184740GA SA311184746GA SA311184748GA SA311184781GA SA311184801GA SA311184821GA SA311236325GA SA311163934GA SA311174159GA SA311184658GA SA311236330GA SA311236336GA SA311246411GA SA311184588GA SA311184697GA SA311184796GA SA311205408GA SA311205429GA SA311205434GA SA311153334GA SA311184795GA SA311184815GA SA311184820GA SA311184827GA SA311184828GA SA311184862GA SA311236035GA SA311236233GA SA311236234GA SA311236235GA SA311236239GA SA311236240GA SA311236241GA SA311236242GA SA311236245GA SA311236246GA SA311236248GA SA311236260GA SA311236261GA SA311236262GA SA311236263GA SA311236264GA SA311236265GA SA311236269GA SA311236270GA SA311236273GA SA311236274GA SA311236275GA SA311236277GA SA311236278GA SA311236279GA SA311163935GA SA311184626GA SA311184642GA SA311184656GA SA311184771GA SA311184778GA SA311184783GA SA311184818GA SA311195054GA SA311195094GA SA311266962GA SA308511487GR SA309304179GR SA309395794GR SA309395812GR SA309517830GR SA309527946GR SA310260522GR SA310454991GR SA310475637GR SA310518637GR SA311060244GR SA311060295GR SA311195164GA SA311153242GA SA311153273GA SA311153321GA SA311163559GA SA311163620GA SA311163817GA SA311163818GA SA311163824GA SA311163840GA SA311163943GA SA311174002GA SA311174007GA SA311174024GA SA311174027GA SA311174031GA SA311174038GA SA311174040GA SA311174049GA SA311174051GA SA311174052GA SA311174076GA SA311174078GA SA311174081GA SA311174083GA SA311174087GA SA311174088GA SA311174089GA SA311174091GA SA311174093GA SA311174095GA SA311174096GA SA311174097GA SA311174098GA SA311174100GA SA311174103GA SA311174107GA SA311174119GA SA311174120GA SA311174121GA SA311174123GA SA311174130GA SA311174140GA SA311174141GA SA311174142GA SA311174143GA SA311174147GA SA311174150GA SA311174151GA SA311174152GA SA311174153GA SA311184460GA SA311184749GA SA311195066GA SA311163574GA SA311163630GA SA311174071GA SA311184461GA SA311184491GA SA311184574GA SA311184668GA SA311184669GA SA311184706GA SA311184734GA SA311184756GA SA311195206GA SA311205391GA SA311205392GA SA311205427GA SA309375454GR SA310028280GR SA311184616GA SA311184655GA SA311236315GA SA311236317GA SA311184637GA SA311195175GA SA311195203GA SA311184647GA SA311195261GA SA311195269GA SA311184744GA SA311194955GA SA311194960GA SA311194969GA SA311195010GA SA311195019GA SA311195174GA SA311195178GA SA311195183GA SA311195186GA SA311195187GA SA311195189GA SA311195207GA SA311205440GA SA311256774GA SA311266992GA SA311287109GA SA311287113GA SA311287123GA SA311287134GAMKE SA311287171GA SA311287197GA SA311287216GA SA311287217GA SA311184638GA SA311184652GA SA311205373GA SA311205376GA SA311205379GA SA311205491GA SA311236064GA SA311287112GA SA311287115GA SA311287131GA SA311287132GA SA311287166GA SA311287170GA SA311287178GA SA311287182GA SA311287211GA SA311297260GA SA311297348GA SA311297356GA SA311297364GA SA311317709GA SA311153073GA SA311184422GA SA311184453GA SA311205304GA SA311205305GA SA311205318GA SA311205409GA SA311184569GA SA311184579GA SA311184633GA SA311163533GA SA311163543GA SA311163565GA SA311163579GA SA311163627GA SA311163629GA SA311163633GA SA311163790GA SA311174009GA SA311184423GA SA311184793GA SA311184811GA SA311184876GA SA311184877GA SA311194880GA SA311194891GA SA311194926GA SA311195002GA SA311195005GA SA311195020GA SA311246679GA SA311287120GA SA311287126GA SA311287128GA SA311287139GA SA311287142GA SA311287148GA SA311287155GA SA311287156GA SA311287162GA SA311287167GA SA311287168GA SA311287169GA SA311287174GA SA311287175GA SA311287177GA SA311287180GA SA311287183GA SA311287184GA SA311287185GA SA311287186GA SA311287187GA SA311287188GA SA311287189GA SA311287191GA SA311287192GA SA311287193GA SA311287194GA SA311287196GA SA311287198GA SA311287199GA SA311287200GA SA311287201GA SA311287203GA SA311287205GA SA311287206GA SA311287207GA SA311287208GA SA311287209GA SA311287210GA SA311287212GA SA311287213GA SA311287214GA SA311287215GA SA311297255GA SA311297256GA SA311297257GA SA311297258GA SA311297263GA SA311297265GA SA311297267GA SA311297270GA SA311297276GA SA311297277GA SA311297278GA SA311297280GA SA311297281GA SA311297285GA SA311297286GA SA311297287GA SA311297289GA SA311297329GA SA311297333GA SA311297336GA SA311297337GA SA311297338GA SA311297339GA SA311297343GA SA311297344GA SA311297345GA SA311297347GA SA311297355GA SA311297358GA SA311297359GA SA311297361GA SA311297363GA SA309355023GR SA309365271GR SA309375473GR SA309385528GR SA309395746GR SA309436297GR SA309527923GR SA310139066GR SA310139244GR SA310517721GR SA311153316GA SA311153328GA SA311153330GA SA311153344GA SA311163734GA SA311205390GA SA311236343GA SA311246384GA SA311246553GA SA311246558GA SA311246564GA SA311246572GA SA311256706GA SA311256708GA SA311256714GA SA311256716GA SA311256717GA SA311256719GA SA311256720GA SA311266886GA SA311287106GA SA311287114GA SA311287116GA SA311287117GA SA311287118GA SA311287119GA SA311287130GA SA311287135GA SA311287140GA SA311287141GA SA311287143GA SA311287144GA SA311132651GA SA311163612GA SA311163776GA SA311163830GA SA311163838GA SA311174032GA SA311174041GA SA311184477GA SA311184672GA SA311184673GA SA311184677GA SA311184678GA SA311184682GA SA311184689GA SA311184691GA SA311184693GA SA311184694GA SA311184699GA SA311184700GA SA311184701GA SA311184720GA SA311184726GA SA311184727GA SA311184728GA SA311184731GA SA311184741GA SA311184742GA SA311184743GA SA311184745GA SA311184755GA SA311184757GA SA311184759GA SA311184766GA SA311184767GA SA311184772GA SA311184773GA SA311184774GA SA311184775GA SA311184777GA SA311184780GA SA311184785GA SA311184787GA SA311184794GA SA311184802GA SA311184804GA SA311184806GA SA311184807GA SA311184814GA SA311184817GA SA311184860GA SA311184878GA SA311194892GA SA311194895GA SA311194898GA SA311194900GA SA311194902GA SA311194903GA SA311194908GA SA311194910GA SA311194912GA SA311194914GA SA311194919GA SA311194928GA SA311194936GA SA311194945GA SA311194949GA SA311194951GA SA311194963GA SA311194964GA SA311194967GA SA311194970GA SA311194982GA SA311194988GA SA311195179GA SA311195180GA SA311195188GA SA311195190GA SA311195194GA SA311195196GA SA311195200GA SA311195201GA SA311195205GA SA311195210GA SA311195211GA SA311195212GA SA311195213GA SA311195214GA SA311195240GA SA311195247GA SA311205397GA SA311205398GA SA311205402GA SA311205412GA SA311205414GA SA311205424GA SA311205425GA SA311205432GA SA311215845GA SA311215853GA SA311256755GA SA311266925GA SA311184686GA SA311184782GA SA311184812GA SA311184834GA SA311184836GA SA311184837GA SA311194901GA SA311194906GA SA311194907GA SA311194909GA SA311194947GA SA311194953GA SA311194979GA SA311194981GA SA311194992GA SA311205395GA SA311205399GA SA311205400GA SA311205410GA SA311205411GA SA311205439GA SA311205573GA SA311215835GA SA311215836GA SA311215839GA SA311215861GA SA311225906GA SA311256751GA SA311297351GA SA311297353GA SA311327836GA SA311337973GA SA311236295GA SA311225900GA SA311142740GA SA311184516GA SA311184819GA SA311184822GA SA311184839GA SA311184840GA SA311194879GA SA311194881GA SA311194887GA SA311194894GA SA311194922GA SA311194937GA SA311194946GA SA311194954GA SA311194958GA SA311194968GA SA311194972GA SA311194975GA SA311194985GA SA311194990GA SA311194999GA SA311195018GA SA311195191GA SA311205350GA SA311215724GA SA311215725GA SA311215735GA SA311215737GA SA311215738GA SA311215741GA SA311236066GA SA311236067GA SA311236068GA SA311236069GA SA311236074GA SA311236078GA SA311266878GA SA311266879GA SA311266880GA SA311266881GA SA311266882GA SA311266883GA SA311266889GA SA311266890GA SA311266894GA SA311266895GA SA311266896GA SA311266898GA SA311266901GA SA311266903GA SA311266906GA SA311266907GA SA311317656GA SA311317658GA SA311317664GA SA311317670GA SA311101132GA SA311142712GA SA311153035GA SA311163461GA SA311163598GA SA311163789GA SA311174128GA SA311184747GA SA311184758GA SA311184789GA SA311184800GA SA311184809GA SA311184813GA SA311184824GA SA311184846GA SA311194882GA SA311194884GA SA311194886GA SA311194888GA SA311194905GA SA311194915GA SA311194916GA SA311194917GA SA311194920GA SA311194925GA SA311194929GA SA311194930GA SA311194943GA SA311194952GA SA311194956GA SA311194974GA SA311194983GA SA311194989GA SA311194991GA SA311205360GA SA311205381GA SA311205466GA SA311215691GA
SA311225909GA SA311184723GA SA311184830GA SA311184835GA SA311184859GA SA311194883GA SA311194889GA SA311194897GA SA311194913GA SA311194921GA SA311194941GA SA311195204GA SA311205574GA SA311205597GA SA311205607GA SA311205617GA SA311205618GA SA311205619GA SA311194885GA SA311163576GA SA311163647GA SA311163650GA SA311163718GA SA311163832GA SA311163844GA SA311174125GA SA311184867GA SA311195045GA SA311195046GA SA311195181GA SA311195209GA SA311205592GA SA311205601GA SA311205615GA SA311215684GA SA311215687GA SA311215688GA SA311215692GA SA311215697GA SA311215718GA SA311215719GA SA311215721GA SA311215728GA SA311236106GA SA311184738GA SA311184826GA SA311184829GA SA311184833GA SA311184850GA SA311184864GA SA311184870GA SA311184875GA SA311194893GA SA311194899GA SA311194927GA SA311194938GA SA311194942GA SA311194980GA SA311194994GA SA311194995GA SA311195001GA SA311195006GA SA311195017GA SA311195021GA SA311195182GA SA311205345GA SA311205367GA SA311205580GA SA311205614GA SA311215802GA SA311215803GA SA311215805GA SA311215807GA SA311215808GA SA311215810GA SA311215811GA SA311215812GA SA311215814GA SA311215834GA SA311215837GA SA311215850GA SA311215851GA SA311215852GA SA311215854GA SA311215855GA SA311215856GA SA311215859GA SA311215860GA SA311225908GA SA311225911GA SA311225912GA SA311225945GA SA311287111GA SA311287179GA SA311287195GA SA311297327GA SA311297330GA SA311297335GA SA311297341GA SA311297362GA SA311307440GA SA311307446GA SA311307450GA SA311215713GA SA311236016GA SA311236032GA SA311236036GA SA311236043GA SA311236047GA SA311236236GA SA311246450GA SA311256804GA SA311266922GA SA310518623GA SA311039486GA SA311236023GA SA311236025GA SA311236050GA SA311236060GA SA311236363GA SA311297273GA SA311327878GA SA311317662GA SA311317671GA SA311317672GA SA311327807GA SA311327809GA SA311327824GA SA311327881GA SA311327882GA SA311039451GA SA311215736GA SA311225966GA SA311236015GA SA311236028GA SA311236228GA SA311297279GA SA311297334GA SA311236040GA SA311163601GA SA311163607GA SA311163638GA SA311163823GA SA311236369GA SA311266938GA SA310528965GA SA310529058GA SA311029363GA SA311039423GA SA311039424GA SA311039425GA SA311039441GA SA311039452GA SA311184536GA SA311184620GA SA311215682GA SA311215689GA SA311215703GA SA311215704GA SA311215707GA SA311215708GA SA311215709GA SA311215710GA SA311215711GA SA311215712GA SA311215716GA SA311215720GA SA311215722GA SA311215726GA SA311215730GA SA311215732GA SA311287190GA SA311297254GA SA311307410GA SA311307411GA SA311307412GA SA311307418GA SA311307421GA SA311307424GA SA311307427GA SA311307428GA SA311307437GA SA311307438GA SA311307441GA SA311307443GA SA311307448GA SA311307454GA SA311307512GA SA311307529GA SA309344865GR SA309375471GR SA309375482GR SA309436270GR SA309446441GR SA309446459GR SA309517639GR SA310290663GR SA310351552GR SA310392937GR SA311070517GA SA311070585GA SA311070586GA SA311070587GA SA311080606GA SA311080642GA SA311080658GA SA311080661GA SA311080703GA SA311080705GA SA311080706GA SA311080707GA SA311080709GA SA311080716GA SA311080722GA SA311090793GA SA311101191GA SA311101193GA SA311101197GA SA311101206GA SA311101210GA SA311101222GA SA311101231GA SA311101247GA SA311101249GA SA311101265GA SA311101266GA SA311101267GA SA311101268GA SA311101276GA SA311236185GA SA311236186GA SA311236281GA SA311236282GA SA311256822GA SA311256829GA SA311256840GA SA311266964GA SA311287021GA SA311287042GA SA311287046GA SA311287051GA SA311287057GA SA311287100GA SA311287103GA SA310454954GA SA311184724GA SA311184733GA SA311184786GA SA311184841GA SA311184842GA SA311184861GA SA311184873GA SA311184874GA SA311194932GA SA311194939GA SA311194944GA SA311194948GA SA311195184GA SA311205420GA SA311205423GA SA311205562GA SA311205587GA SA311205588GA SA311205604GA SA311205605GA SA311205608GA SA311205609GA SA311205611GA SA311205613GA SA311205616GA SA311215727GA SA311215806GA SA311215857GA SA311225892GA SA311225896GA SA311225902GA SA311225903GA SA311225904GA SA311225905GA SA311225919GA SA311225921GA SA311246485GA SA311317707GA SA311327884GA SA311327888GA SA311327889GA SA311327893GA SA311327895GA SA311327898GA SA311428162GA SA311458224GA SA311458241GA SA311468409GA SA311205475GA SA311236206GA SA311246639GA SA311246660GA SA311256805GA SA311256819GA SA311256847GA SA311256856GA SA311287012GA SA311287020GA SA311287023GA SA311287028GA SA311287033GA SA311287037GA SA311287038GA SA311287044GA SA311287045GA SA311287056GA SA311287058GA SA311287059GA SA311287061GA SA311287077GA SA311287097GA SA311297259GA SA311307419GA SA311307425GA SA311307436GA SA311307444GA SA311307445GA SA311307449GA SA311205315GA SA311225893GA SA311225942GA SA311225952GA SA311236029GA SA311236216GA SA311236355GA SA311236376GA SA311246452GA SA311246461GA SA311246470GA SA311287016GA SA311307456GA SA311327816GA SA311338000GA SA311468377GA SA311287025GA SA311287071GA SA311287078GA SA311287102GA SA311287154GA SA311287158GA SA311287161GA SA311458217GA SA311458221GA SA311458226GA SA311458229GA SA311458354GA SA311458374GA SA311468375GA SA311039419GA SA311184449GA SA311297366GA SA311428157GA SA311458223GA SA311458240GA SA311458244GA SA311458245GA SA311458253GA SA311458341GA SA311458344GA SA311458348GA SA311468402GA SA311468404GA SA311236231GA SA311246618GA SA311256727GA SA311256729GA SA311256739GA SA311256744GA SA311256747GA SA311256750GA SA311256761GA SA311287019GA SA311287093GA SA311287099GA SA311327875GA SA311337983GA SA311428017GA SA311458222GA SA311458237GA SA311458239GA SA311458350GA SA311468396GA SA311468397GA SA310529108GA SA311039413GA SA311039414GA SA311039469GA SA311039479GA SA311039484GA SA311060299GA SA311132560GA SA311142909GA SA311205472GA SA311236210GA SA311236249GA SA311236256GA SA311256726GA SA311256731GA SA311256732GA SA311256734GA SA311256737GA SA311256740GA SA311256756GA SA311256758GA SA311256759GA SA311256760GA SA311256763GA SA311448177GA SA311458345GA SA311458351GA SA311458358GA SA311458362GA SA311468380GA SA311468383GA SA311468388GA SA310496651GA SA310528875GA SA310529009GA SA311236088GA SA311236227GA SA311236271GA SA311236329GA SA311246383GA SA311246386GA SA311246388GA SA311246390GA SA311246392GA SA311246393GA SA311246397GA SA311246399GA SA311246412GA SA311246419GA SA311246425GA SA311246426GA SA311246429GA SA311246436GA SA311246440GA SA311246442GA SA311246444GA SA311246509GA SA311246586GA SA311246597GA SA311246603GA SA311246611GA SA311246620GA SA311246621GA SA311246638GA SA311246664GA SA311256748GA SA311317718GA SA311317736GA SA311327808GA SA311327963GA SA311428093GA SA311428095GA SA311428097GA SA311428098GA SA311428099GA SA311428100GA SA311428101GA SA311428102GA SA311428103GA SA311428104GA SA311428106GA SA311428107GA SA311428108GA SA311428109GA SA311428110GA SA311428111GA SA311428113GA SA311428114GA SA311428117GA SA311428118GA SA311428120GA SA311428122GA SA311428123GA SA311428126GA SA311428127GA SA311428132GA SA311428133GA SA311428135GA SA311428138GA SA311428140GA SA311428141GA SA311428143GA SA311428147GA SA311428149GA SA311428150GA SA311428152GA SA311428153GA SA311428155GA SA311428156GA SA311428160GA SA311428161GA SA311428163GA SA311428165GA SA311428166GA SA311428167GA SA311428168GA SA311428169GA SA311428170GA SA311448189GA SA311448190GA SA311458352GA SA311327808GA SA311184858GA SA311256718GA SA311287022GA SA311287030GA SA311287053GA SA311287104GA SA311458272GA SA311458282GA SA311458292GA SA311039480GA SA311101208GA SA311215701GA SA311236348GA SA311236349GA SA311246463GA SA311246466GA SA311246493GA SA311246565GA SA311256723GA SA311256742GA SA311256776GA SA311256790GA SA311256831GA SA311266902GA SA311297299GA SA311297373GA SA311297382GA SA311297399GA SA311297400GA SA311307423GA SA311307453GA SA311307501GA SA311307507GA SA311307582GA SA311307586GA SA311307588GA SA311307590GA SA311307592GA SA311307593GA SA311307596GA SA311307597GA SA311317648GA SA311317651GA SA311317733GA SA311327814GA SA311327830GA SA311327887GA SA311327891GA SA311327945GA SA311327946GA SA311327947GA SA311428092GA SA311428116GA SA311428128GA SA311428146GA SA311428151GA SA311458258GA SA311458260GA SA311468395GA SA310413561GA SA310518705GA SA310518725GA SA310528896GA SA310528984GA SA311029366GA SA311039405GA SA311039412GA SA311039417GA SA311039418GA SA311039421GA SA311039422GA SA311039430GA SA311039439GA SA311256785GA SA311307500GA SA311307502GA SA311307503GA SA311307506GA SA311307508GA SA311307510GA SA311307511GA SA311307581GA SA311307585GA SA311307587GA SA311307589GA SA311317643GA SA311317647GA SA311317652GA SA311317657GA SA311317663GA SA311317667GA SA311317703GA SA311317704GA SA311317708GA SA311317716GA SA311317720GA SA311317724GA SA311317729GA SA311317734GA SA311317735GA SA311317740GA SA311327810GA SA311327827GA SA311327866GA SA311327879GA SA311327896GA SA311327897GA SA311327935GA SA311327939GA SA311327941GA SA311327944GA SA311327948GA SA311327949GA SA311327952GA SA311327954GA SA311327955GA SA311327957GA SA311327961GA SA311327965GA SA311327967GA SA310393097GA SA310528844GA SA310528917GA SA310528920GA SA310529107GA SA311060076GA SA311060102GA SA311246622GA SA311246650GA SA311246651GA SA311246655GA SA311246662GA SA311246663GA SA311246670GA SA311256733GA SA311256738GA SA311256752GA SA311317730GA SA311317732GA SA310393243GA SA310496370GA SA310528845GA SA310528902GA SA310528977GA SA310529116GA SA310529120GA SA311039468GA SA311195238GA SA311205569GA SA311236061GA SA311236182GA SA311236212GA SA311236222GA SA311236238GA SA311236247GA SA311236346GA SA311236356GA SA311236360GA SA311236361GA SA311236371GA SA311236380GA SA311246472GA SA311246492GA SA311246497GA SA311246500GA SA311246515GA SA311246595GA SA311246598GA SA311246631GA SA311256728GA SA311256757GA SA311256762GA SA311256807GA SA311256826GA SA311256870GA SA311287024GA SA311287027GA SA311287029GA SA311287036GA SA311287048GA SA311287050GA SA311287072GA SA311287082GA SA311287084GA SA311287085GA SA311287087GA SA311287101GA SA311287151GA SA311297354GA SA311307544GA SA311317600GA SA311327817GA SA311327819GA SA311327829GA SA311327943GA SA311508588GA SA311297293GA SA311236347GA SA311246462GA SA311246583GA SA311246585GA SA311246587GA SA311246594GA SA311246600GA SA311246600GAMKE SA311246602GA SA311246615GA SA311246619GA SA311246625GA SA311246628GA SA311246632GA SA311246642GA SA311246645GA SA311246653GA SA311246656GA SA311246666GA SA311246674GA SA311246675GA SA311246677GA SA311287026GA SA311287034GA SA311287108GA SA311297293GA SA311327820GA SA311327821GA SA311327832GA SA311468407GA SA310444287GA SA310539322GA SA311039410GA SA311039477GA SA311163427GA SA311184639GA SA311205476GA SA311215699GA SA311215705GA SA311215706GA SA311246404GA SA311266872GA SA311266911GA SA311287136GA SA311307514GA SA311307531GA SA311327956GA SA311327960GA SA311327964GA SA311327968GA SA311327970GA SA311448188GA SA311458230GA SA311508572GA SA311508589GA SA311246480GA SA311246588GA SA311246606GA SA311246624GA SA311246640GA SA311246649GA SA311246652GA SA311246654GA SA311246658GA SA311246659GA SA311246665GA SA311246669GA SA311246673GA SA311246681GA SA311256725GA SA311256730GA SA311256745GA SA311256823GA SA311256825GA SA311256832GA SA311307561GA SA311327909GA SA311327933GA SA311508561GA SA311508590GA SA311508591GA SA311039385GA SA311039429GA SA311163563GA SA311174077GA SA311236251GA SA311236372GA SA311246456GA SA311246460GA SA311246465GA SA311246468GA SA311246483GA SA311246484GA SA311246490GA SA311246499GA SA311246502GA SA311246506GA SA311246507GA SA311246513GA SA311246591GA SA311246592GA SA311246593GA SA311246596GA SA311246599GA SA311246601GA SA311246604GA SA311246605GA SA311246607GA SA311246608GA SA311246609GA SA311246613GA SA311246614GA SA311246616GA SA311246617GA SA311246626GA SA311246627GA SA311246633GA SA311246634GA SA311246635GA SA311246637GA SA311246641GA SA311246643GA SA311246644GA SA311246646GA SA311246647GA SA311246648GA SA311246672GA SA311246676GA SA311256809GA SA311256814GA SA311256815GA SA311256830GA SA311256833GA SA311256834GAMKE SA311256835GA SA311256838GA SA311256839GA SA311256841GA SA311256855GA SA311256868GA SA311287013GA SA311287015GA SA311287031GA SA311287039GA SA311287040GA SA311287049GA SA311287054GA SA311287055GA SA311287075GA SA311287079GA SA311287083GA SA311287088GA SA311287094GA SA311297223GA SA311297228GA SA311297250GA SA311297292GA SA311297298GA SA311297301GA SA311297325GA SA311297340GA SA311297368GA SA311297370GA SA311297371GA SA311297375GA SA311297376GA SA311297377GA SA311297378GA SA311297379GA SA311297380GA SA311297381GA SA311297383GA SA311297384GA SA311297385GA SA311297386GA SA311297387GA SA311297393GA SA311297394GA SA311297395GA SA311297396GA SA311297397GA SA311297398GA SA311297402GA SA311297403GA SA311297404GA SA311297405GA SA311297407GA SA311307463GA SA311307464GA SA311307466GA SA311307471GA SA311307472GA SA311307475GA SA311307479GA SA311307481GA SA311307482GA SA311307484GA SA311307491GA SA311307492GA SA311307493GA SA311307494GA SA311307495GA SA311307496GA SA311307499GA SA311307541GA SA311307542GA SA311307543GA SA311307545GA SA311307548GA SA311307549GA SA311307550GA SA311307551GA SA311307552GA SA311307555GA SA311307556GA SA311307557GA SA311307559GA SA311307560GA SA311307563GA SA311307564GA SA311307565GA SA311307566GA SA311307567GA SA311307571GA SA311307572GA SA311307575GA SA311307577GA SA311307578GA SA311307579GA SA311317598GA SA311317606GA SA311317608GA SA311317613GA SA311317618GA SA311317620GA SA311317621GA SA311317627GA SA311317628GA SA311317635GA SA311317638GA SA311317640GA SA311317690GA SA311327837GA SA311327840GA SA311327841GA SA311327846GA SA311327851GA SA311327901GA SA311327902GA SA311327903GA SA311327904GA SA311327906GA SA311327907GA SA311327908GA SA311327912GA SA311327916GA SA311327917GA SA311327918GA SA311327921GA SA311327922GA SA311327923GA SA311327924GA SA311327926GA SA311327928GA SA311327929GA SA311327930GA SA311327931GA SA311337971GA SA311337972GA SA311337974GA SA311337977GA SA311337979GA SA311337982GA SA311337984GA SA311337987GA SA311337988GA SA311337990GA SA311337991GA SA311337992GA SA311337994GA SA311337995GA SA311338001GA SA311338002GA SA311338005GA SA311508566GA SA311508593GA SA311236184GA SA311317755GA SA311317772GA SA311317773GA SA311327927GA SA311327937GA SA311327950GA SA311327951GA SA311508640GA SA311508643GA SA311508646GA SA311508648GA SA311508650GA SA311508652GA SA311508655GA SA311508656GA SA311508664GA SA311508665GA SA311508668GA SA311508673GA SA311508674GA SA311508675GA SA311236367GA SA311256743GA SA311256753GA SA311468394GA SA311508560GA SA311508563GA SA311508564GA SA311508565GA SA311508567GA SA311508568GA SA311508569GA SA311508570GA SA311508573GA SA311508574GA SA311508575GA SA311508578GA SA311508581GA SA311508582GA SA311508584GA SA311508585GA SA311508586GA SA311508592GA SA311508596GA SA311508597GA SA311508598GA SA311508599GA SA311508614GA SA311508623GA SA311508624GA SA311508663GA SA311518703GA SA311518707GA SA311518710GA SA311518712GA SA311518713GA SA311518714GA SA311518715GA SA311518716GA SA311518718GA SA311246474GA SA311287065GA SA311287069GA SA311287070GA SA311287073GA SA311287089GA SA311287150GA SA311297219GA SA311317614GA SA311317693GA SA311317765GA SA311518683GA SA311518695GA SA311518701GA SA311518702GA SA311518727GA SA311518736GA SA311317614GA SA311287047GA SA311287060GA SA311287064GA SA311287066GA SA311287090GA SA311287153GA SA311307476GA SA311307574GA SA311327843GA SA311518699GA SA309365234GR SA309477028GR SA311195060GA SA311195078GA SA311195109GA SA311195121GA SA311195127GA SA311287159GA SA311287165GA SA311287218GA SA311297221GA SA311297222GA SA311297225GA SA311297226GA SA311297227GA SA311297229GA SA311297232GA SA311297234GA SA311297236GA SA311297238GA SA311297243GA SA311297246GA SA311297249GA SA311518681GA SA311518686GA SA311518704GA SA311518705GA SA311518706GA SA311518709GA SA311518719GA SA311518721GA SA311518722GA SA311518723GA SA311518724GA SA311518730GA SA311518731GA SA311518732GA SA311518734GA SA311528737GA SA311528738GA SA311528739GA SA311528741GA SA311528742GA SA311528743GA SA311528744GA SA311039478GA SA311039490GA SA311132545GA SA311132600GA SA311184784GA SA311184823GA SA311184851GA SA311184865GA SA311236309GA SA311287124GA SA311297220GA SA311297231GA SA311297237GA SA311297239GA SA311297240GA SA311297242GA SA311297245GA SA311297253GA SA311297275GA SA311297302GA SA311297303GA SA311297306GA SA311297314GA SA311297316GA SA311297326GA SA311297367GA SA311297369GA SA311297372GA SA311297374GA SA311297388GA SA311297389GA SA311297391GA SA311297392GA SA311297401GA SA311297406GA SA311297408GA SA311297409GA SA311307468GA SA311317599GA SA311317605GA SA311317619GA SA311317679GA SA311317681GA SA311317697GA SA311317699GA SA311317700GA SA311317701GA SA311317702GA SA311428130GA SA311428136GA SA311428137GA SA311428139GA SA311428164GA SA311428171GA SA311508576GA SA311518687GA SA311518689GA SA311518693GA SA311518697GA SA311518700GA SA311287074GA SA311287080GA SA311184527GA SA311195069GA SA311236303GA SA311236307GA SA311236314GA SA311236328GA SA311236339GA SA311236342GA SA311327873GA SA311327899GA SA311518708GA SA312018746GA SA312018754GA SA312018755GA SA311508654GA SA312018851GA SA312018856GA SA312018857GA SA312018861GA SA312018862GA SA312018863GA SA312018865GA SA312018867GA SA312018868GA SA312018869GA SA312018870GA SA312018871GA SA311287076GA SA311297304GA SA311297311GA SA311297324GA SA311317754GA SA311488477GA SA311488478GA SA311297322GA SA312018847GA SA311317692GA SA311246678GA SA311508647GA SA311518735GA SA312018841GA SA312018844GA SA312018845GA SA312018846GA SA312018848GA SA312018866GA SA312018897GA SA312028910GA SA312028914GAMKE SA312028915GA SA312028916GA SA312028917GA SA312028919GA SA312028921GA SA312028922GA SA312028926GA SA312028928GA SA312028929GA SA312028932GA SA312028933GA SA312028936GA SA308511483GR SA308511540GR SA309395705GR SA312018850GA SA312018852GA SA312018887GA SA312018898GA SA312018902GA SA311518733GA SA308388468GR SA308398817GR SA308409032GR SA308491168GR SA308501421GR SA308511464GR SA308511501GR SA308511512GR SA308511551GR SA309294116GR SA309304201GR SA309324342GR SA309354967GR SA309375456GR SA309395717GR SA309395754GR SA309405851GR SA309426121GR SA309446472GR SA309446487GR SA309446503GR SA309446505GR SA309476957GR SA309476976GR SA309517679GR SA309517707GR SA309517796GR SA309517816GR SA309517817GR SA309517823GR SA310139061GR SA310199565GR SA310219839GR SA310393139GR SA310413473GR SA310423680GR SA310423684GR SA311090776GR SA311142984GR SA312029014GA SA312029069GA SA308459843GR SA308511481GR SA309324411GR SA309354926GR SA309385561GR SA309395792GR SA309405845GR SA309446532GR SA309456540GR SA310485793GR SA311090781GR SA311468382GR SA312028979GA SA312028991GA SA312029020GA SA312029023GA SA312029034GA SA312029034GA SA311317685GA SA311458252GA SA311518717GA SA311528740GA SA312028948GA SA312028982GA SA312028984GA SA312028986GA SA312028987GA SA312028988GA SA312028989GA SA312028992GA SA312028993GA SA312028994GA SA312028995GA SA312029001GA SA312029005GA SA312029007GA SA312029009GA SA312029011GA SA312029026GA SA312029039GA SA312029062GA SA312029065GA SA311508636GA SA311518690GA SA311518720GA SA312028966GA SA312028997GA SA312028998GA SA312029003GA SA312029004GA SA312029006GA SA312029010GA SA312029012GA SA312029017GA SA312029028GA SA312029032GA SA312029035GA SA312029036GA SA312029038GA SA312029040GA SA312029043GA SA312029045GA SA312029047GA SA312029050GA SA312029052GA SA312029055GA SA312029058GA SA312029063GA SA312029066GA SA312039084GA SA312039086GA SA312039093GA SA310517646GR SA311468433GA SA311468445GA SA311468446GA SA311488468GA SA311488469GA SA311488470GA SA311488485GA SA312018808GA SA312029057GA SA311236218GA SA311297294GA SA311297390GA SA311317676GA SA311317678GA SA311317680GA SA311317683GA SA311317687GA SA311317696GA SA311327835GA SA311327838GA SA311327845GA SA311327849GA SA311327850GA SA311327855GA SA311327858GA SA311327861GA SA311327863GA SA311327864GA SA311337998GA SA311468420GA SA311468428GA SA311468438GA SA309446468GR SA309456546GR SA309456552GR SA310259884GR SA310260553GR SA310331249GR SA310496653GR SA310507400GR SA310517824GR SA311090882GR SA311132350GR SA311205492GR SA311317602GA SA311317615GA SA311317641GA SA311458367GA SA312039140GA SA311327842GA SA311327853GA SA311327857GA SA311428058GA SA311428071GA SA311468424GA SA311468432GA SA311468443GA SA311488456GA SA311488458GA SA311488462GA SA311488464GA SA311488471GA SA311488472GA SA311488473GA SA311488475GA SA311488476GA SA311488480GA SA311488482GA SA311488484GA SA311488486GA SA311488487GA SA311488488GA SA311184711GA SA311195177GA SA311195218GA SA311195223GA SA311195236GA SA311195243GA SA311195245GA SA311236013GA SA311256821GA SA311266947GA SA311287063GA SA311287067GA SA311297321GA SA311317629GA SA311327839GA SA311428006GA SA311428007GA SA311428008GA SA311428009GA SA311428010GA SA311428014GA SA311428016GA SA311428019GA SA311428021GA SA311428022GA SA311428024GA SA311428027GA SA311428029GA SA311428030GA SA311428034GA SA311428036GA SA311428037GA SA311428038GA SA311428039GA SA311428040GA SA311428041GA SA311428042GA SA311428044GA SA311428045GA SA311428046GA SA311428047GA SA311428048GA SA311428049GA SA311428051GA SA311428052GA SA311428053GA SA311428054GA SA311428067GA SA311428074GA SA311428075GA SA311428076GA SA311428080GA SA311428082GA SA311428083GA SA311428085GA SA311428087GA SA311428089GA SA311458369GA SA311468414GA SA311468419GA SA311488459GA SA312028980GA SA312039083GA SA312039087GA SA312039089GA SA312039102GA SA312039103GA SA311428025GA SA311428055GA SA311458370GA SA311458372GA SA311458373GA SA311468408GA SA311468448GA SA311488474GA SA312028978GA SA312029002GA SA312029031GA SA312039076GA SA312039077GA SA312039080GA SA312039081GA SA312039085GA SA312039090GA SA312039094GA SA312039098GA SA312039101GA SA312039107GA SA312039131GA SA312039133GA SA312039136GA SA312039179GA SA312039205GA SA312039220GA SA312039221GA SA312049241GA SA312049244GA SA312049246GA SA312049247GA SA312049248GA SA312049250GA SA312049251GA SA312049252GA SA312049253GA SA312049256GA SA312049258GA SA312049262GA SA312049264GA SA312049268GA SA311468415GA SA311488465GA SA311488481GA SA311458347GA SA311468384GA SA312028950GA SA312039108GA SA312039109GA SA312039111GA SA312039112GA SA312039113GA SA312039118GA SA312039120GA SA312039122GA SA312039128GA SA312039129GA SA312039142GA SA312039143GA SA312039144GA SA312039149GA SA312039151GA SA312039152GA SA312039154GA SA312039157GA SA312039158GA SA312039159GA SA312039162GA SA312039164GA SA312039168GA SA312039180GA SA312039223GA SA312039225GA SA312039228GA SA312039230GA SA312039231GA SA312039232GA SA312049235GA SA312049236GA SA312049238GA SA312049249GA SA312049254GA SA312049255GA SA312049259GA SA312049261GA SA312049263GA SA312049267GA SA312049272GA SA312049276GA SA312049287GA SA312049293GA SA312049299GA SA312049370GA SA312049378GA SA312049391GA SA312049396GA SA312049400GA SA312059507GA SA312059510GA SA312059512GA SA312059522GA SA312059525GA SA311317623GA SA311317637GA SA311317639GA SA311458249GA SA311508600GA SA312029024GA SA312029027GA SA312029033GA SA312029060GA SA312039078GA SA312039082GA SA312039092GA SA312039097GA SA312039100GA SA312039105GA SA312039106GA SA312039114GA SA312039117GA SA312039125GA SA312039130GA SA312039137GA SA312049242GA SA312049243GA SA312049260GA SA312049265GA SA311458333GA SA311488454GA SA311488455GA SA311488460GA SA311488479GA SA312018809GA SA312018810GA SA312018811GA SA312018812GA SA312018813GA SA312018814GA SA312018815GA SA312018817GA SA312018819GA SA312018820GA SA312018822GA SA312018823GA SA312018824GA SA312018826GA SA312018827GA SA312018828GA SA312018830GA SA312018831GA SA312018835GA SA312018836GA SA312018837GA SA312018839GA SA312018840GA SA312018843GA SA312018881GA SA312018886GA SA312018889GA SA312018893GA SA312018895GA SA312018904GA SA312029016GA SA312029021GA SA312029022GA SA312029025GA SA312029030GA SA312029046GA SA312029054GA SA312029056GA SA312029059GA SA312029061GA SA312029064GA SA312039190GA SA312049270GA SA312049271GA SA312049279GA SA312049281GA SA312049289GA SA312049292GA SA312049294GA SA312049296GA SA312049297GA SA312049298GA SA312049300GA SA312049302GA SA312049305GA SA312049307GA SA312049309GA SA312049312GA SA312049315GA SA312049316GA SA312049317GA SA312049426GA SA312049432GA SA312049459GA SA312039196GA SA312049342GA SA312049371GA SA312049390GA SA312049430GA SA312049454GA SA312049468GA SA312059471GA SA312059528GA SA312059530GA SA312059533GA SA312059534GA SA312049372GA SA312049373GA SA312049374GA SA312049377GA SA312049383GA SA312049385GA SA312049388GA SA312049389GA SA312049392GA SA312049393GA SA312049395GA SA312049399GA SA312049405GA SA312049407GA SA312049408GA SA312049411GA SA312049414GA SA312049418GA SA312049419GA SA312049437GA SA312059500GA SA312059502GA SA312059503GA SA312059515GA SA312059516GA SA312059520GA SA312049318GA SA312049321GA SA312049329GA SA312049331GA SA312049332GA SA312049339GA SA312049341GA SA312049346GA SA312049348GA SA312049350GA SA312049353GA SA312049354GA SA312049359GA SA312049362GA SA312049363GA SA312049423GA SA312049436GA SA312049449GA SA312049452GA SA312059489GA SA312079899GA SA312079900GA SA312079902GA SA312079918GA SA312079919GA SA312079922GA SA312079924GA SA312079938GA SA312049275GA SA312049386GA SA312069818GA SA312069822GA SA312069837GA SA312069850GA SA312069855GA SA312069857GA SA312069860GA SA312069861GA SA312069862GA SA312069863GA SA312069864GA SA312069865GA SA312069868GA SA312069869GA SA312069881GA SA312069865GA SA312049386GA SA312069818GA SA312069861GA SA312069862GA SA312069881GA SA312049275GA SA312069855GA SA312069857GA SA312069860GA SA312069864GA SA312069868GA SA312069837GA SA312069850GA SA312069863GA SA312069869GA SA312069822GA SA307220890GR SA307220904GR SA307220931GR SA307221013GR SA307221026GR SA307221168GR SA307241668GR SA307241677GR SA307241691GR SA307241708GR SA307251710GR SA307251729GR SA308186433GR SA308511496GR SA308511520GR SA309375441GR SA309405821GR SA309446484GR SA309446488GR SA309476970GR SA309517533GR SA310310883GR SA310382657GR SA310393073GR SA310423635GR SA310424012GR SA310444272GR SA310507110GR SA310507131GR SA310507396GR SA310507416GR SA310507426GR SA310507441GR SA310507447GR SA310517707GR SA311101219GR SA311205467GR SA312049278GA SA312049283GA SA312049288GA SA312049368GA SA312049387GA SA312049394GA SA312049401GA SA312049402GA SA312049403GA SA312049409GA SA312049410GA SA312049416GA SA312059513GA SA312059514GA SA312059517GA SA312059518GA SA312059519GA SA312059521GA SA312059523GA SA312059524GA SA312059526GA SA312059638GA SA312069646GA SA312069652GA SA312069654GA SA312069657GA SA312069659GA SA312069834GA SA312069875GA SA312079887GA SA312079908GA SA312079915GA SA312079917GA SA312079926GA SA312079927GA SA312079928GA SA312079933GA SA312079934GA SA312079935GA SA312079936GA SA312079937GA SA312079944GA SA312079948GA SA312079966GA SA312079970GA SA312079972GA SA312079974GA SA312079975GA SA312079992GA SA312079995GA SA312079998GA SA312059609GA SA312059619GA SA312059620GA SA312070006GA SA312049347GA SA312049365GA SA312049427GA SA312049429GA SA312049433GA SA312049442GA SA312049443GA SA312049447GA SA312049461GA SA312049466GA SA312059480GA SA312059483GA SA312059494GA SA312059496GA SA312079940GA SA312079941GA SA312079950GA SA312079953GA SA312079956GA SA312079958GA SA312079961GA SA312079977GA SA312079980GA SA312079987GA SA312079988GA SA312079994GA SA312079996GA SA311458371GA SA312018875GA SA312039123GA SA312039126GA SA312039134GA SA312039138GA SA312039147GA SA312039148GA SA312039153GA SA312039161GA SA312039165GA SA312039166GA SA312039169GA SA312039172GA SA312039173GA SA312039174GA SA312039175GA SA312039177GA SA312039183GA SA312039193GA SA312059474GA SA312059537GA SA312059539GA SA312059543GA SA312059552GA SA312059553GA SA312059559GA SA312059562GA SA312059563GA SA312059612GA SA312059615GA SA312059631GA SA312059636GA SA312039091GA SA312039127GA SA312039163GA SA312039181GA SA312039185GA SA312039194GA SA312039197GA SA312039198GA SA312039199GA SA312039201GA SA312039207GA SA312039211GA SA312069698GA SA312080398GA SA312080401GA SA309426039GR SA312069715GA SA307251757GR SA310507146GR SA312080402GA SA312080467GA SA312080512GA SA312080523GA SA312059608GA SA312059610GA SA312059611GA SA312059613GA SA312059614GA SA312059616GA SA312059617GA SA312059618GA SA312059625GA SA312059632GA SA312059633GA SA312059635GA SA312059637GA SA312059639GA SA312069640GA SA312069643GA SA312069656GA SA312069663GA SA312069667GA SA312069668GA SA312069669GA SA312069672GA SA312069675GA SA312069688GA SA312069707GA SA312069720GA SA312080397GA SA312080528GA SA307190745GR SA307220898GR SA307220901GR SA307220983GR SA307221096GR SA307221114GR SA307221154GR SA307221183GR SA307231366GR SA307231476GR SA307241674GR SA307251713GR SA307251782GR SA307362392GR SA308084673GR SA308186372GR SA310506731GR SA310506734GR SA311111566GR SA311142752GR SA311142903GR SA311153188GR SA312049324GA SA312049336GA SA312049361GA SA312059557GA SA312069648GA SA312069749GA SA312069760GA SA312080386GA SA312080390GA SA312080416GA SA312080417GA SA312080419GA SA312080422GA SA312080423GA SA312080425GA SA312080432GA SA312080433GA SA312080465GA SA312080488GA SA312080494GA SA312080496GA SA312080498GA SA312080500GA SA312080505GA SA312080506GA SA312080509GA SA312080514GA SA312080515GA SA312080526GA SA312080532GA SA312080497GA SA312080510GA SA312059511GA SA312070024GA SA312079943GA SA312079945GA SA312079963GA SA312079985GA SA312080462GA SA312080420GA SA312080489GA SA312080531GA SA312080533GA SA312080539GA SA312080555GA SA312080556GA SA312080558GA SA312080561GA SA312080571GA SA312080577GA SA312080584GA SA312080587GA SA312080590GA SA312080596GA SA312080604GA SA312080608GA SA312080617GA SA312080622GA SA312080623GA SA312080624GA SA312080625GA SA312080626GA SA312080631GA SA312080645GA SA312080673GA SA312080674GA SA312080679GA SA312080686GA SA312080688GA SA312080690GA SA312080692GA SA312049358GA SA312049360GA SA312049428GA SA312059475GA SA312059481GA SA312059535GA SA312059536GA SA312059538GA SA312059540GA SA312059542GA SA312059547GA SA312059549GA SA312059550GA SA312059551GA SA312059554GA SA312059555GA SA312059556GA SA312059558GA SA312059561GA SA312059564GA SA312059566GA SA312059569GA SA312080389GA SA312080403GA SA312080410GA SA312080415GA SA312080424GA SA312080427GA SA312080429GA SA312080431GA SA312080434GA SA312080463GA SA312080464GA SA312080466GA SA312080471GA SA312080472GA SA312080473GA SA312080476GA SA312080477GA SA312080478GA SA312080479GA SA312080486GA SA312080487GA SA312080508GA SA312080513GA SA312080517GA SA312080520GA SA312080522GA SA312080525GA SA312080535GA SA312080536GA SA312080537GA SA312080538GA SA312080540GA SA312080541GA SA312080543GA SA312080545GA SA312080548GA SA312080549GA SA312080552GA SA312080559GA SA312080560GA SA312080566GA SA312080573GA SA312080574GA SA312080575GA SA312080579GA SA312080594GA SA312070002GA SA312070023GA SA312070055GA SA312070058GA SA312070060GA SA312070085GA SA312080388GA SA312080396GA SA312080399GA SA312080405GA SA312080407GA SA312080418GA SA312080428GA SA312080568GA SA312080582GA SA312080585GA SA312080600GA SA312080605GA SA312080607GA SA312080613GA SA312080616GA SA312080620GA SA312080635GA SA312080638GA SA312080639GA SA312080640GA SA312080652GA SA312080657GA SA312080660GA SA312080664GA SA312080696GA SA312080710GA SA312080733GA SA312080734GA SA312080735GA SA312080430GA SA312080599GA SA312080601GA SA312070005GA SA312070016GA SA312070020GA SA312070029GA SA312070032GA SA312070033GA SA312070056GA SA312070061GA SA312070090GA SA312070092GA SA312070096GA SA312070111GA SA312070136GA SA312080534GA SA312080649GA SA312080656GA SA312080659GA SA312080661GA SA312080662GA SA312080667GA SA312080678GA SA312080695GA SA312080701GA SA312080706GA SA312080708GA SA312080719GA SA312080744GA SA312080763GA SA312080769GA SA312080779GA SA312080783GA SA312080791GA SA312080792GA SA312080797GA SA312080802GA SA312080814GA SA312080840GA SA312080903GA SA312080909GA SA312049438GA SA312059472GA SA312070086GA SA312070097GA SA312070221GA SA312070232GA SA312080663GA SA312080677GA SA312080707GA SA312080724GA SA312080761GA SA312080765GA SA312080775GA SA312069693GA SA312069694GA SA312069717GA SA312069725GA SA312069787GA SA312070017GA SA312070053GA SA312070082GA SA312070083GA SA312070087GA SA312070091GA SA312070093GA SA312070094GA SA312080670GA SA312080830GA SA312080848GA SA312080852GA SA312080871GA SA312080873GA SA312080885GA SA312080900GA SA312080901GA SA312080905GA SA312070062GA SA312070084GA SA312070095GA SA312080409GA SA312080412GA SA312080653GA SA312080691GA SA312049320GA SA312049326GA SA312049344GA SA312049434GA SA312059479GA SA312059486GA SA312059490GA SA312059493GA SA312070219GA SA312070004GA SA312070007GA SA312070008GA SA312070009GA SA312070010GA SA312070014GA SA312070018GA SA312070019GA SA312070026GA SA312070027GA SA312070035GA SA312070037GA SA312070038GA SA312070040GA SA312079979GA SA312079982GA SA312079991GA SA312080550GA SA312080591GA SA312080629GA SA312080648GA SA312080655GA SA312080683GA SA312080700GA SA312069713GA SA312070043GA SA312070069GA SA312070074GA SA312070076GA SA312070080GA SA312070081GA SA312070124GA SA312070132GA SA312070135GA SA312080542GA SA312080610GA SA312080666GA SA312080689GA SA312080693GA SA312080720GA SA312080729GA SA312080731GA SA312080749GA SA312080751GA SA312080781GA SA312080800GA SA312080803GA SA312080824GA SA312080826GA SA312080833GA SA312080872GA SA312080874GA SA307030411GR SA307110560GR SA307241690GR SA307251735GR SA308459744GR SA309122640GR SA309304244GR SA309365226GR SA309365230GR SA309365233GR SA309365235GR SA309365348GR SA309476885GR SA309476916GR SA309477022GR SA310372168GR SA310382331GR SA310434088GR SA310507275GR SA310517645GR SA310518168GR SA310518169GR SA310518170GR SA310518179GR SA310518207GR SA310518225GR SA310518477GR SA310529069GR SA311142802GR SA311163435GR SA311163460GR SA311163466GR SA311163726GR SA311184444GR SA312070001GA SA312070247GA SA312070294GA SA312080570GA SA312080681GA SA312080817GA SA312080819GA SA312080860GA SA312080876GA SA312080884GA SA312070064GA SA312070079GA SA312070100GA SA312070106GA SA312070110GA SA312070115GA SA312070123GA SA312070155GA SA312070163GA SA312070175GA SA312070182GA SA312070184GA SA312070191GA SA312070216GA SA312070220GA SA312070258GA SA312070259GA SA312070262GA SA312070264GA SA312070266GA SA312070269GA SA312070272GA SA312070275GA SA312070276GA SA312070278GA SA312070279GA SA312070281GA SA312070285GA SA312070288GA SA312070290GA SA312070314GA SA312070317GAMKE SA312080563GA SA312080564GA SA312080597GA SA312080612GA SA312080647GA SA312080711GA SA312080717GA SA312080721GA SA312080727GA SA312080753GA SA312080764GA SA312080796GA SA312080798GA SA312080804GA SA312080806GA SA312080808GA SA312080810GA SA312080811GA SA312080813GA SA312080878GA SA312080882GA SA312080894GA SA312070187GA SA312070193GA SA312070195GA SA312070200GA SA312070209GA SA312070280GA SA312070287GA SA312070291GA SA312070318GA SA312070321GA SA312080684GA SA312080716GA SA312080725GA SA312080754GA SA312080758GA SA312080771GA SA312080777GA SA312080827GA SA312080834GA SA312080839GA SA312080858GA SA312080892GA SA312080893GA SA307210819GR SA307220900GR SA307231376GR SA307231444GR SA307241586GR SA307251727GR SA307251743GR SA307291955GR SA312049280GA SA312049369GA SA312070034GA SA312070046GA SA312070048GA SA312070052GA SA312070054GA SA312070070GA SA312070071GA SA312070073GA SA312070101GA SA312070102GA SA312070107GA SA312070109GA SA312070113GA SA312070114GA SA312070116GA SA312070120GA SA312070126GA SA312070133GA SA312070151GA SA312079952GA SA312079957GA SA312079959GA SA312079962GA SA312079967GA SA312079968GA SA312079989GA SA312080426GA SA312080646GA SA312080704GA SA312080714GA SA312080759GA SA312080794GA SA312080820GA SA312080845GA SA312080877GA SA312080881GA SA312080890GA SA312080912GA SA312080916GA SA312091019GA SA312091020GA SA312091021GA SA312091022GA SA312091027GA SA312091033GA SA312091042GA SA312091053GA SA312091042GA SA312070044GA SA312070065GA SA312070197GA SA312070204GA SA312070211GA SA312070231GA SA312070243GA SA312070248GA SA312070250GA SA312070253GA SA312070271GA SA312070283GA SA312070307GA SA312070308GA SA312070320GA SA312080529GA SA312080544GA SA312080553GA SA312080554GA SA312080598GA SA312080630GA SA312080705GA SA312080713GA SA312080748GA SA312080770GA SA312080787GA SA312080801GA SA312080805GA SA312080809GA SA312080812GA SA312080837GA SA312080886GA SA312080899GA SA312080914GA SA312091018GA SA312091024GA SA312091025GA SA312091028GA SA312091030GA SA312091031GA SA312091032GA SA312091034GA SA312091035GA SA312091041GA SA312091044GA SA312091045GA SA312091048GA SA312091058GA SA312091060GA SA312091062GA SA312091067GA SA312091070GA SA312091071GA SA312091074GA SA312091075GA SA312091076GA SA312091082GA SA312091109GA SA312091116GA SA312049356GA SA312049441GA SA312049446GA SA312049451GA SA312049453GA SA312049457GA SA312049462GA SA312049463GA SA312049467GA SA312049469GA SA312059473GA SA312069642GA SA312069651GA SA312069653GA SA312069661GA SA312069664GA SA312069665GA SA312070042GA SA312070049GA SA312070059GA SA312070067GA SA312070075GA SA312070112GA SA312070131GA SA312070142GA SA312070150GA SA312070152GA SA312070156GA SA312070158GA SA312070199GA SA312070217GA SA312070226GA SA312070324GA SA312070347GA SA312070354GA SA312070361GA SA312070373GA SA312080668GA SA312080671GA SA312080675GA SA312080728GA SA312080732GA SA312091017GA SA312091023GA SA312091026GA SA312091037GA SA312091038GA SA312091039GA SA312091040GA SA312091050GA SA312091051GA SA312091052GA SA312091054GA SA312091057GA SA312091061GA SA312091064GA SA312091069GA SA312091077GA SA312091079GA SA312091083GA SA312070030GA SA312070057GA SA312070068GA SA312070139GA SA312070143GA SA312070164GA SA312070198GA SA312070201GA SA312070215GA SA312079942GA SA312091043GA SA312091059GA SA312069804GA SA312070039GA SA312070130GA SA312070141GA SA312070145GA SA312070146GA SA312070147GA SA312070148GA SA312070149GA SA312070153GA SA312070154GA SA312070157GA SA312070159GA SA312070161GA SA312070162GA SA312070165GA SA312070166GA SA312070167GA SA312070168GA SA312070169GA SA312070170GA SA312070171GA SA312070172GA SA312070173GA SA312070176GA SA312070177GA SA312070178GA SA312070179GA SA312070183GA SA312070188GA SA312070190GA SA312070196GA SA312070202GA SA312070206GA SA312070208GA SA312070210GA SA312070212GA SA312070225GA SA312070227GA SA312070230GA SA312070233GA SA312070236GA SA312070237GA SA312070241GA SA312070242GA SA312070244GA SA312070245GA SA312070246GA SA312070249GA SA312070254GA SA312070255GA SA312070256GA SA312070257GA SA312070260GA SA312070261GA SA312070263GA SA312070265GA SA312070267GA SA312070270GA SA312070273GA SA312070274GA SA312070282GA SA312070284GA SA312070286GA SA312070289GA SA312070292GA SA312070293GA SA312070295GA SA312070296GA SA312070297GA SA312070298GA SA312070299GA SA312070300GA SA312070301GA SA312070303GA SA312070305GA SA312070306GA SA312070309GA SA312070310GA SA312070311GA SA312070312GA SA312070313GA SA312070315GA SA312070319GA SA312070322GA SA312070323GA SA312070325GA SA312070326GA SA312070327GA SA312070328GA SA312070329GA SA312070330GA SA312070332GA SA312070333GA SA312070334GA SA312070335GA SA312070336GA SA312070338GA SA312070339GA SA312070342GA SA312070343GA SA312070346GA SA312070349GA SA312070350GA SA312070353GA SA312070355GA SA312070356GA SA312070357GA SA312070359GA SA312070363GA SA312070364GA SA312070365GA SA312070366GA SA312070367GA SA312070368GA SA312070369GA SA312070370GA SA312070372GA SA312070374GA SA312070375GA SA312070376GA SA312070377GA SA312070378GA SA312070380GA SA312070381GA SA312070384GA SA312070385GA SA312090917GA SA312090921GA SA312090934GA SA312090942GA SA312090948GA SA312090966GA SA312090967GA SA312090969GA SA312090970GA SA312090972GA SA312090974GA SA312090981GA SA312090982GA SA312090983GA SA312090985GA SA312090987GA SA312090988GA SA312090989GA SA312090990GA SA312090993GA SA312090994GA SA312090995GA SA312090996GA SA312090997GA SA312090998GA SA312090999GA SA312091000GA SA312091001GA SA312091002GA SA312091003GA SA312091004GA SA312091006GA SA312091011GA SA312091046GA SA312091066GA SA312091094GA SA312091095GA SA312091104GA SA312091108GA SA312070304GA SA312070331GA SA312070351GA SA312070371GA SA312070382GA SA312091007GA SA312101406GA SA312091085GA SA312091113GA SA312069743GA SA312101453GA SA312101455GA SA312111529GA SA312091036GA SA312091063GA SA312091072GA SA312091078GA SA312091081GA SA312091084GA SA312091086GA SA312091087GA SA312091088GA SA312091090GA SA312091091GA SA312091096GA SA312091110GA SA312091111GA SA312080491GA SA312101389GA SA312101448GA SA312111510GA SA312111517GA SA312111518GA SA312111558GA SA312111560GA SA312111566GA SA312111572GA SA312111582GA SA312111584GA SA312111587GA SA312111591GA SA312069754GA SA312069766GA SA312090927GA SA312091055GA SA312091073GA SA312091092GA SA312091103GA SA312091107GA SA312091245GA SA312101446GA SA312101449GA SA312101450GA SA312101452GA SA312111521GA SA312111540GA SA312111549GA SA312111552GA SA312111565GA SA312111569GA SA312111575GA SA312111580GA SA312111585GA SA312111586GA SA312111588GA SA312111590GA SA312069678GA SA312069734GA SA312069742GA SA312069755GA SA312069761GA SA312069767GA SA312091080GA SA312091114GA SA312101445GA SA312101454GA SA312101456GA SA312101457GA SA312111546GA SA312049237GA SA312111524GA SA312111536GA SA312111567GA SA312101447GA SA312111515GA SA312111523GA SA312111528GA SA312111545GA SA312111551GA SA312111561GA SA312111562GA SA312111570GA SA312111573GA SA312111657GA SA312111658GA SA312111659GA SA312111661GA SA312111662GA SA312111665GA SA312111668GA SA312111670GA SA312111671GA SA312111672GA SA312111674GA SA312111675GA SA312111676GA SA312111678GA SA312111679GA SA312111680GA SA312111681GA SA312111683GA SA312111684GA SA312111686GA SA312111687GA SA312111691GA SA312121692GA SA312121693GA SA312121696GA SA312121697GA SA312121699GA SA312121710GA SA312121714GA SA312121720GA SA312121722GA SA312121723GA SA312121724GA SA312121725GA SA312121728GA SA312121730GA SA312121731GA SA312121733GA SA312121734GA SA312121735GA SA312121737GA SA312121738GA SA312121743GA SA312121745GA SA312121747GA SA312121753GA SA312091190GA SA312091202GA SA312091219GA SA312091244GA SA312091246GA SA312111512GA SA312111514GA SA312111519GA SA312111522GA SA312111525GA SA312111530GA SA312111531GA SA312111533GA SA312111534GA SA312111535GA SA312111537GA SA312111538GA SA312111541GA SA312111542GA SA312111543GA SA312111544GA SA312111548GA SA312111554GA SA312111556GA SA312111557GA SA312111559GA SA312111563GA SA312111564GA SA312111568GA SA312111571GA SA312111577GA SA312111578GA SA312111581GA SA310518715GR SA311236085GR SA311236092GR SA311266955GR SA311266976GR SA311266979GR SA311266982GR SA311266983GR SA311266991GR SA311267010GR SA312111513GA SA312069775GA SA312069779GA SA312069790GA SA312069792GA SA312069793GA SA312069794GA SA312069795GA SA312069796GA SA312069799GA SA312069801GA SA312069802GA SA312069803GA SA312069810GA SA312069823GA SA312069844GA SA312069846GA SA312069849GA SA312090978GA SA312091125GA SA312091145GA SA312091146GA SA312091148GA SA312091154GA SA312091166GA SA312091178GA SA312091196GA SA312091201GA SA312091204GA SA312091206GA SA312091207GA SA312091211GA SA312091218GA SA312091220GA SA312091222GA SA312091232GA SA312091234GA SA312091241GA SA312091243GA SA312091254GA SA312091257GA SA312111652GA SA312111666GA SA312111685GA SA312111688GA SA312111690GA SA312121695GA SA312121698GA SA312121703GA SA312121709GA SA312121713GA SA312121739GA SA312121740GA SA312121751GA SA312121755GA SA312121794GA SA312121798GA SA312091226GA SA312121700GA SA312121706GA SA312121707GA SA312121716GA SA312121729GA SA312121757GA SA312121759GA SA312121767GA SA312121784GA SA312121786GA SA312121787GA SA312121796GA SA312121797GA SA312121800GA SA312121804GA SA312121721GA SA312091252GA SA312091253GA SA312091269GA SA312091277GA SA312091278GA SA312091280GA SA312091282GA SA312091283GA SA312091284GA SA312111654GA SA312121813GA SA312080551GA SA312091239GA SA312091275GA SA312091287GA SA312091289GA SA312091293GA SA312091294GA SA312091297GA SA312091335GA SA312121708GA SA312121768GA SA312121771GA SA312121772GA SA312121778GA SA312121790GA SA312121811GA SA312121812GA SA312121814GA SA312121817GA SA312121821GA SA312121827GA SA312121831GA SA312121833GA SA312121913GA SA312091265GA SA312091298GA SA312091302GA SA312091355GA SA312101428GA SA312111486GA SA312111669GA SA312121705GA SA312121715GA SA312121717GA SA312121718GA SA312121719GA SA312121799GA SA312121818GA SA312091274GA SA312091279GA SA312091304GA SA312091306GA SA312091309GA SA312091311GA SA312091319GA SA312091322GA SA312091330GA SA312091336GA SA312091340GA SA312091345GA SA312091347GA SA312091349GA SA312121829GA SA312121834GA SA312121873GA SA312121887GA SA312121892GA SA312121903GA SA312121997GA SA312091366GA SA312121837GA SA312121838GA SA312121849GA SA312121897GA SA312121910GA SA312121920GA SA312121926GA SA312121928GA SA312121939GA SA312121953GA SA312121955GA SA312090928GA SA312091197GA SA312091215GA SA312091221GA SA312091223GA SA312091224GA SA312091225GA SA312091227GA SA312091229GA SA312091230GA SA312091231GA SA312091233GA SA312091236GA SA312091237GA SA312091238GA SA312091240GA SA312091242GA SA312091247GA SA312091248GA SA312091258GA SA312091259GA SA312091260GA SA312091261GA SA312091266GA SA312091268GA SA312091271GA SA312091272GA SA312091276GA SA312121908GA SA312069816GA SA312069826GA SA312069838GA SA312070041GA SA312070105GA SA312090923GA SA312091009GA SA312091010GA SA312091013GA SA312091014GA SA312091137GA SA312091153GA SA312091310GA SA312091332GA SA312091337GA SA312091365GA SA312101375GA SA312101386GA SA312101416GA SA312101418GA SA312101434GA SA312101441GA SA312101443GA SA312122026GA SA312070089GA SA312121762GA SA312121855GA SA312121859GA SA312121929GA SA312121936GA SA312121980GA SA312121982GA SA312121985GA SA312101451GA SA312121839GA SA312121894GA SA312121933GA SA312121940GA SA312121941GA SA312121944GA SA312121946GA SA312121951GA SA312121952GA SA312121960GA SA312121973GA SA312121976GA SA312121984GA SA312122014GA SA312122055GA SA312122057GA SA312069842GA SA312070063GA SA312070088GA SA312070137GA SA312121930GA SA312121932GA SA312121934GA SA312121935GA SA312121958GA SA312121966GA SA312121988GA SA312121990GA SA312122003GA SA312122004GA SA312122007GA SA312122016GA SA312122037GA SA312122038GA SA312122041GA SA312122043GA SA312122052GA SA312122054GA SA312122059GA SA312122062GA SA312122064GA SA312122069GA SA312122071GA SA312132090GA SA312132093GA SA312132095GA SA312132102GA SA312132109GA SA312132110GA SA312132113GA SA312132130GA SA312132132GA SA312132138GA SA312132140GA SA312132142GA SA312132173GA SA312132177GA SA312132178GA SA312069666GA SA312069674GA SA312069677GA SA312069679GA SA312069683GA SA312069685GA SA312069689GA SA312069709GA SA312069712GA SA312122018GA SA312101436GA SA312111461GA SA312111463GA SA312111465GA SA312111471GA SA312111506GA SA312111507GA SA312121841GA SA312121957GA SA312122025GA SA312122028GA SA312132089GA SA312132100GA SA312132103GA SA312132118GA SA312132136GA SA312132175GA SA312152417GA SA312069692GA SA312069700GA SA312069716GA SA312069721GA SA312069722GA SA312070022GA SA312132101GA SA312121891GA SA312121899GA SA312121956GA SA312121962GA SA312121967GA SA312121974GA SA312121983GA SA312122031GA SA312122079GA SA312122080GA SA312122087GA SA312132107GA SA312132126GA SA312132135GA SA312152371GA SA312152372GA SA312152373GA SA312152374GA SA312152375GA SA312152376GA SA312152377GA SA312152378GA SA312152379GA SA312152380GA SA312152382GA SA312152386GA SA312152387GA SA312152388GA SA312152390GA SA312152395GA SA312152399GA SA312152400GA SA312152404GA SA312152406GA SA312152409GA SA312152413GA SA312152415GA SA312152418GA SA312152419GA SA312152422GA SA312152424GA SA312152425GA SA312152431GA SA312152433GA SA312152434GA SA312152436GA SA310382839GR SA310507215GR SA311184448GR SA311184452GR SA311184455GR SA312070214GA SA312091250GA SA312091251GA SA312091256GA SA312091262GA SA312111509GA SA312121732GA SA312121805GA SA312121959GA SA312121964GA SA312122042GA SA312122051GA SA312122058GA SA312122063GA SA312122076GA SA312122082GA SA312122084GA SA312122085GA SA312132105GA SA312132106GA SA312132114GA SA312132115GA SA312132116GA SA312132119GA SA312132123GA SA312132124GA SA312132128GA SA312132139GA SA312132141GA SA312132143GA SA312132144GA SA312132145GA SA312132146GA SA312132147GA SA312132150GA SA312132164GA SA312132166GA SA312132168GA SA312132172GA SA312152311GA SA312152313GA SA312152341GA SA312152353GA SA312152363GA SA312152366GA SA312152414GA SA312152432GA SA312152444GA SA312152448GA SA312152451GA SA312152452GA SA312152461GA SA312152463GA SA312152466GA SA312152467GAMKE SA312192565GA SA312192566GA SA312192567GA SA312192568GA SA312192569GA SA312192570GA SA312192571GA SA312192572GA SA312192573GA SA312192574GA SA312192575GA SA312192576GA SA312192577GA SA312192578GA SA312192579GA SA312192580GAMKE SA312192581GA SA312192582GA SA312192583GA SA312192584GA SA312192585GA SA312192586GA SA312192587GA SA312192589GA SA312192591GA SA312192592GA SA312192593GA SA312192594GA SA312192596GA SA312192598GA SA312192601GA SA312192602GA SA312192606GA SA312192607GA SA312192608GA SA312192609GA SA312192610GA SA312192611GA SA312192612GA SA312192613GA SA312192614GA SA312192615GA SA312192616GA SA312192618GA SA312192619GA SA312192621GA SA312192622GA SA312192624GA SA312192625GA SA312192626GA SA312192627GA SA312192628GA SA312192629GA SA312192630GA SA312192634GA SA312192635GA SA312192636GA SA312192638GA SA312192640GA SA312192641GA SA312192642GAMKE SA312192643GA SA312192647GA SA312192649GA SA312152391GA SA312111682GA SA312121802GA SA312121938GA SA312121949GA SA312122030GA SA312122036GA SA312122039GA SA312122040GA SA312122048GA SA312122050GA SA312122060GA SA312122065GA SA312122067GA SA312122068GA SA312132134GA SA312132167GA SA312132170GA SA312132174GA SA312132176GA SA312152323GA SA312152338GA SA312152340GA SA312152342GA SA312152348GA SA312152354GA SA312152357GA SA312152408GA SA312152411GA SA312152429GA SA312059591GA SA312069714GA SA312121774GA SA312121937GA SA312121994GA SA312122009GA SA312152334GA SA312152355GA SA312152456GA SA312152470GA SA312152483GA SA312152484GA SA312192599GA SA312192604GA SA312192617GA SA312192645GA SA312192650GA SA312192652GA SA312192653GA SA312192655GA SA312192658GA SA312192659GA SA312192667GA SA312122083GA SA312132091GA SA312132148GA SA312132149GA SA312132151GA SA312132165GA SA312192654GA SA312192663GA SA312192669GA SA312192687GA SA307070450GR SA307200777GR SA307221081GR SA307221104GR SA307231370GR SA307251721GR SA307251747GR SA308166102GR SA309395710GR SA310393145GR SA310475537GR SA311266957GR SA311266974GR SA311266984GR SA311267003GR SA312079960GA SA312091208GA SA312142250GA SA312152327GA SA312152398GA SA312152410GA SA312152416GA SA312152420GA SA312152421GA SA312152426GA SA312152427GA SA312152430GA SA312152439GA SA312152440GA SA312152441GA SA312152445GA SA312152446GA SA312152447GA SA312152449GA SA312152450GA SA312152453GA SA312152454GA SA312152455GA SA312152457GA SA312152458GA SA312152459GA SA312152462GA SA312152464GA SA312152468GA SA307070450GR SA312122001GA SA312152471GA SA312069697GA SA312069711GA SA312070011GA SA312070025GA SA312070031GA SA312070036GA SA312079920GA SA312079946GA SA312079971GA SA312079983GA SA312079993GA SA312111673GA SA312121977GA SA312132137GA SA312152312GA SA312152316GA SA312152318GA SA312152329GA SA312152356GA SA312152360GA SA312152364GA SA312152368GA SA312152472GA SA312152481GA SA312152496GA SA312152508GA SA312152515GA SA312152525GA SA312152533GA SA312152543GA SA312152544GA SA312152550GA SA312152559GA SA312192656GA SA312192657GA SA312192664GA SA312192665GA SA312192668GA SA312192673GA SA312192675GA SA312192677GA SA312192685GA SA312192688GA SA312192689GA SA312192690GA SA312192691GA SA312192692GA SA312192693GA SA312192694GA SA312192699GA SA312192700GA SA312192702GA SA312192703GA SA312192704GA SA312192705GA SA312192706GA SA312192707GA SA312192708GA SA312192709GA SA312192710GA SA312192711GA SA312192712GA SA312192715GA SA312192716GA SA312192718GA SA312192719GA SA312192720GA SA312192721GA SA312192722GA SA312192725GA SA312192727GA SA312192728GA SA312192729GA SA312192731GA SA307221156GR SA307251726GR SA308084724GR SA309426041GR SA309446404GR SA309517567GR SA310506843GR SA310517652GR SA311132356GR SA311508572GR SA311508589GR SA312091068GA SA312152478GA SA312152479GA SA312152488GA SA312152505GA SA312152511GA SA312152512GA SA312152513GA SA312152519GA SA312152520GA SA312152523GA SA312152526GA SA312152529GA SA312152530GA SA312152536GA SA312152538GA SA312152539GA SA312152545GA SA312152546GA SA312152554GA SA312192661GA SA312192662GA SA312192666GA SA312192672GA SA312192676GA SA312192678GA SA312192679GA SA312232826GA SA312232831GA SA312232834GA SA312232835GA SA312121711GA SA312121902GA SA312121902GA SA307231210GR SA312152331GA SA312253251GA SA312253104GA SA312253361GA SA312263785GA SA312069839GA SA312091286GA SA312294770GA SA312294776GA SA312355402GA SA312294497GA SA312376217GA SA312469178GA SA312468536GA SA312470025GA SA307221035GA SA307221055GA SA307221065GA SA307221071GA SA307221077GA SA307221078GA SA307221080GA SA307221121GA SA307221122GA SA307221123GA SA307221127GA SA307221128GA SA307221131GA SA307221133GA SA307221138GA SA307231227GA SA307231244GA SA307231270GA SA307231302GA SA307231252GA SA307231360GA SA307241521GA SA307241573GA SA307241574GA SA307241685GA SA308501356GAMKE SA308501372GAMKE SA310310854GA SA310310861GA SA310310925GA SA310260253GA SA310290706GA SA310290712GA SA310300768GA SA310300769GA SA310300771GA SA310300773GA SA310300776GA SA310300778GA SA310300780GA SA310300781GA SA310300783GA SA310300784GA SA310300785GA SA310300786GA SA310310866GA SA310310867GA SA310310868GA SA310310871GA SA310260376GA SA310300770GA SA310300777GA SA310300779GA SA310310840GA SA310310845GA SA310310848GA SA310310852GA SA310310857GA SA310310859GA SA310310862GA SA310310872GA SA310310875GA SA310310876GA SA310310885GA SA310310898GA SA310310902GA SA310310904GA SA310310907GA SA310310908GA SA310310916GA SA310310920GA SA310310921GA SA310310922GA SA310310924GA SA310260522GA SA310270582GA SA310270584GA SA310290739GA SA310300763GA SA310300766GA SA310300788GA SA310290716GA SA310290717GA SA310290725GA SA310068947GA SA310260479GAMKE SA310260480GA SA310260521GAMKE SA310260523GA SA310260524GA SA310260528GA SA310260534GA SA310260539GA SA310290760GA SA310331145GA SA310260374GA SA310331080GA SA310331092GA SA310331149GA SA310331331GA SA310331332GA SA310331335GA SA310331341GA SA310331347GA SA310331348GA SA310331352GA SA310331356GA SA310321014GA SA310321020GA SA310321024GA SA310331218GA SA310331219GA SA310331223GA SA310331228GA SA310331230GA SA310331234GA SA310331255GA SA310331261GA SA310331266GA SA310321048GA SA310321061GA SA310270588GA SA310290715GA SA310290719GA SA310290747GA SA310290752GA SA310300789GA SA310300793GA SA310300794GA SA310300797GA SA310300799GA SA310300801GA SA310300802GA SA310300804GA SA310300805GA SA310331109GA SA310331139GA SA310331145GA SA310331154GA SA310310895GA SA310259929GA SA310260243GA SA310290686GA SA310310865GA SA310310956GA SA310310973GA SA310331220GA SA310331221GA SA310331222GA SA310331224GA SA310331231GA SA310331238GA SA310331242GA SA310331243GA SA310331245GA SA310331248GA SA310331251GA SA310331253GA SA310331256GA SA310331259GA SA310331264GA SA310331265GA SA310331270GA SA310331271GA SA310331273GA SA310331274GA SA310331276GA SA310331277GA SA310331278GA SA310331279GA SA310331281GA SA310331283GA SA310331284GA SA310331289GA SA310331293GA SA310331297GA SA310331300GA SA310331301GA SA310331304GA SA310331305GA SA310331322GA SA310341412GA SA310351500GA SA310351510GA SA310351513GA SA310351514GA SA310351516GA SA310351550GA SA310351563GA SA310351564GA SA310351588GA SA310351532GA SA310351541GA SA310351546GA SA310351547GA SA310351548GA SA310351551GA SA310351552GA SA310361752GA SA310361755GA SA310361756GA SA310361757GA SA310361761GA SA310361767GA SA310361768GA SA310361797GA SA310300790GA SA310351530GA SA310361594GA SA310361622GA SA310361624GA SA310361679GA SA310361691GA SA310361693GA SA310361694GA SA310361696GA SA310361702GA SA310361703GA SA310361709GA SA310361844GA SA310361932GA SA310361957GA SA307251778GR SA310331327GA SA310341357GA SA310341362GA SA310341363GA SA310341364GA SA310341366GA SA310341367GA SA310341369GA SA310341370GA SA310341385GA SA310341419GA SA310341424GA SA310331115GA SA310341372GA SA310341417GA SA310341422GA SA310361596GA SA310361598GA SA310361603GA SA310361604GA SA310361605GA SA310361606GA SA310361610GA SA310361611GA SA310361613GA SA310361614GA SA310361616GA SA310361618GA SA310361627GA SA310259920GA SA310259971GA SA310260237GA SA310260242GA SA310361589GA SA310361601GA SA310361621GA SA310361633GAMKE SA310361641GA SA310361655GA SA310361666GAMKE SA310361668GA SA310361669GAMKE SA310361671GAMKE SA310361673GAMKE SA310361674GA SA310361688GA SA310361690GAMKE SA310361692GA SA310361697GA SA310361699GA SA310361700GA SA310361701GA SA310361704GA SA310361705GA SA310361706GAMKE SA310361920GA SA310361954GA SA310361955GA SA310371988GA SA310371989GA SA310371990GA SA310371991GA SA310372007GA SA310058725GA SA310058752GA SA310331138GA SA310341379GA SA310341380GA SA310351469GA SA310351472GA SA310351473GA SA310351474GA SA310351475GA SA310351486GA SA310351487GA SA310351491GA SA310351492GA SA310351493GA SA310351497GA SA310351503GA SA310351504GA SA310351507GA SA310351509GA SA310351524GA SA310351559GA SA310351560GA SA310351561GA SA310351565GA SA310351566GA SA310351574GA SA310351584GA SA310351587GA SA310372027GA SA310372054GA SA310372065GA SA310372067GA SA310372069GA SA310372139GA SA310372145GA SA310372146GA SA310372149GA SA310372306GA SA310118963GA SA310118986GA SA310118999GA SA310119002GA SA310341451GA SA310351476GA SA310351477GA SA310351478GA SA310351479GA SA310351482GA SA310351483GA SA310351485GA SA310351533GA SA310351535GAMKE SA310351537GA SA310351538GA SA310351542GA SA310351545GA SA310351555GAMKE SA310351556GAMKE SA310341374GA SA307220932GR SA307241686GR SA310259964GA SA310321047GA SA310321049GA SA310321056GA SA310321057GA SA310321063GA SA310321068GA SA310321073GA SA310331098GA SA310331128GA SA310331136GA SA310331159GA SA310331160GA SA310331329GA SA310331350GA SA310361591GA SA310361592GA SA310361602GA SA310361612GA SA310361615GA SA310361617GA SA310361648GA SA310259916GA SA310260236GA SA310260252GA SA310260326GA SA310260529GA SA310260531GA SA310260533GA SA310260536GA SA310260541GA SA310260542GA SA310260543GA SA310260544GA SA310260547GA SA310260554GA SA310290619GA SA310290699GA SA310341427GA SA310361630GA SA310361678GA SA310361710GA SA310361914GA SA310361922GA SA310361923GA SA310361925GA SA310361927GA SA310361928GA SA310361929GA SA310361937GA SA310361956GA SA310361958GA SA310371966GA SA310371967GA SA310371968GA SA310371969GA SA310371971GA SA310371972GA SA310371975GA SA310371976GA SA310371977GA SA310371978GA SA310371980GA SA310371983GA SA310371992GA SA310371993GA SA310371995GA SA310371996GA SA310371998GA SA310371999GA SA310372001GA SA310372012GA SA310372015GA SA310372018GA SA310372035GA SA310372039GA SA310310901GA SA310331288GA SA310331292GA SA310331294GA SA310331295GA SA310331296GA SA310331298GA SA310331316GA SA310331326GA SA310341410GA SA310341415GA SA310341425GA SA310341426GA SA310341428GA SA310341429GA SA310341430GA SA310341431GA SA310341432GA SA310341434GA SA310341435GA SA310341436GA SA310341458GA SA310361593GA SA310361771GA SA310361772GA SA310361773GA SA310361774GA SA310361777GA SA310361778GA SA310361779GA SA310361783GA SA310361784GA SA310361787GA SA310361788GA SA310361789GA SA310361791GA SA310361796GA SA310361811GA SA310361930GA SA310361946GA SA310361948GA SA310371970GA SA310371979GA SA310371981GA SA310371984GA SA310371986GA SA310371997GA SA310372004GA SA310372005GA SA310372024GA SA310372029GA SA310372036GA SA310372037GA SA310372040GA SA310372049GA SA310372199GA SA310260214GA SA310260353GA SA310260354GA SA310260355GA SA310260373GA SA310260375GA SA310260377GA SA310260379GA SA310260382GA SA310260384GA SA310260387GA SA310260389GA SA310260395GA SA310260397GA SA310260399GA SA310260405GA SA310260409GA SA310260420GA SA310260429GA SA310260430GA SA310260434GA SA310260436GA SA310260438GA SA310260439GA SA310331272GA SA310341411GA SA310361607GA SA310361620GA SA310361623GA SA310361625GA SA310361626GA SA310361640GA SA310361652GA SA310361667GA SA310361890GA SA310361891GA SA310361892GA SA310361893GA SA310361894GA SA310361895GA SA310361898GA SA310361899GA SA310361900GA SA310361901GA SA310361902GA SA310361906GA SA310361907GA SA310361912GA SA310361915GA SA310361916GA SA310361918GA SA310361919GA SA310361921GA SA310361924GA SA310361926GA SA310361933GA SA310361934GA SA310361935GA SA310361936GA SA310361938GA SA310361939GA SA310361961GA SA310371965GA SA310371973GA SA310371974GA SA310371982GA SA310371985GA SA310371987GA SA310372003GA SA310372010GA SA310372016GA SA310372017GA SA310372021GA SA310372022GA SA310372026GA SA310372030GA SA310372043GA SA310372048GA SA310372053GA SA310372055GA SA310372058GA SA310372064GA SA310372151GA SA310372155GA SA310372156GA SA310372159GA SA310372163GA SA310372307GA SA310382530GA SA310382532GA SA310382545GA SA310382568GA SA310382577GA SA310382587GA SA310382596GA SA310382633GA SA310382659GA SA310382662GA SA310382667GA SA310382696GA SA310382739GA SA310382769GA SA310382782GA SA310382813GA SA310382821GA SA310382822GA SA310382830GA SA310260303GA SA310260386GA SA310331313GA SA310341365GA SA310372105GA SA310382408GA SA310382429GA SA310382504GA SA310382605GA SA310382611GA SA310382630GA SA310382637GA SA310382639GA SA310382706GA SA310382707GA SA310382708GA SA310382709GA SA310382710GA SA310382712GA SA310382713GA SA310382714GA SA310382715GA SA310382716GA SA310382717GA SA310382718GA SA310382719GA SA310382721GA SA310382722GA SA310382723GA SA310382724GA SA310382726GA SA310382733GA SA310382734GA SA310382736GA SA310382737GA SA310382740GA SA310382741GA SA310382748GA SA310382755GA SA310382757GA SA310382766GA SA310382805GA SA310290649GA SA310290650GA SA310321070GA SA310321072GA SA310331232GA SA310351467GA SA310351468GA SA310351471GA SA310351480GA SA310351481GA SA310351490GA SA310351494GA SA310351496GA SA310351502GA SA310351505GA SA310351506GA SA310351508GA SA310351517GA SA310351519GA SA310351523GA SA310351529GA SA310361590GA SA310361595GA SA310361597GA SA310361599GA SA310361619GA SA310361634GA SA310361637GA SA310361639GA SA310361643GA SA310361649GA SA310361650GA SA310361653GA SA310361654GA SA310361659GA SA310361672GA SA310361680GA SA310361681GA SA310361760GA SA310361762GA SA310361765GA SA310361780GA SA310361781GA SA310361795GA SA310361799GA SA310361800GA SA310361802GA SA310361826GA SA310361831GA SA310361832GA SA310361833GA SA310361834GA SA310361839GA SA310372303GA SA310382385GA SA310382420GA SA310382421GA SA310382424GA SA310382425GA SA310382427GA SA310382428GA SA310382432GA SA310382438GA SA310382442GA SA310382446GA SA310382454GA SA310382458GA SA310382465GA SA310382466GA SA310382468GA SA310382469GA SA310382470GA SA310382473GA SA310382474GA SA310382478GA SA310382483GA SA310382485GA SA310382492GA SA310382497GA SA310382501GA SA310382508GA SA310382516GA SA310382517GA SA310382524GA SA310382547GA SA310382571GA SA310382592GA SA310382609GA SA310382614GA SA310382617GA SA310382621GA SA310382627GA SA310382638GA SA310382640GA SA310382643GA SA310382644GA SA310382646GA SA310382650GA SA310382654GA SA310382655GA SA310382658GA SA310382660GA SA310382661GA SA310382670GA SA310382674GA SA310382695GA SA310382762GA SA310382777GA SA310382791GA SA310382800GA SA310382836GA SA310382843GA SA310392866GA SA310392934GA SA310392935GA SA310392935GAMKE SA310392936GA SA310392937GA SA310392938GA SA310392939GA SA310392942GA SA310392944GA SA310392947GA SA310392950GA SA310392951GA SA310392952GA SA310392953GA SA310392954GA SA310392955GA SA310392957GA SA310392959GA SA310392962GA SA310392993GA SA310393020GA SA310393025GA SA310393032GA SA310393039GA SA310393053GA SA310259972GA SA310260526GA SA310260555GA SA310321040GA SA310321042GA SA310321069GA SA310331334GA SA310341358GA SA310341360GA SA310341371GA SA310341377GA SA310341408GA SA310351484GA SA310351495GA SA310351518GA SA310351520GA SA310351521GA SA310351525GA SA310351526GA SA310361651GA SA310361812GA SA310361814GA SA310361817GA SA310361821GA SA310361822GA SA310361823GA SA310361824GA SA310361825GA SA310361828GA SA310361829GA SA310361830GA SA310361835GA SA310361843GA SA310361852GA SA310361853GA SA310361857GA SA310361859GA SA310361863GA SA310361864GA SA310361873GA SA310361880GA SA310361886GA SA310372072GA SA310372077GA SA310372079GA SA310372082GA SA310372084GA SA310372088GA SA310372096GA SA310372098GA SA310372099GA SA310372101GA SA310372107GA SA310372118GA SA310372189GA SA310372193GA SA310372202GA SA310372289GA SA310382329GA SA310382330GA SA310382332GA SA310382333GA SA310382334GA SA310382335GA SA310382343GA SA310382359GA SA310382371GA SA310382447GA SA310382531GA SA310382534GA SA310382535GA SA310382538GA SA310382543GA SA310382548GA SA310382555GA SA310382556GA SA310382561GA SA310382565GA SA310382566GA SA310382570GA SA310382572GA SA310382576GA SA310382578GA SA310382583GA SA310382590GA SA310382591GA SA310382598GA SA310382601GA SA310382603GA SA310382665GA SA310382673GA SA310382682GA SA310382684GA SA310382685GA SA310382689GA SA310382690GA SA310382700GA SA310382704GA SA310382720GA SA310382784GA SA310382786GA SA310382790GA SA310382792GA SA310382798GA SA310382847GA SA310392887GA SA310393009GA SA310393024GA SA310393027GA SA310393030GA SA310393033GA SA310393034GA SA310393049GA SA310393051GA SA310393055GA SA310393056GA SA310393057GA SA310393060GA SA310393062GA SA310393063GA SA310393066GA SA310393067GA SA310393068GA SA310393072GA SA310393076GA SA310393146GA SA310393147GA SA310393148GA SA310290720GA SA310382622GA SA310382628GA SA310382632GA SA310382725GA SA310382727GA SA310382732GA SA310382743GA SA310382744GA SA310382745GA SA310382746GA SA310382747GA SA310382750GA SA310382751GA SA310382752GA SA310382756GA SA310382759GA SA310382764GA SA310382765GA SA310393050GA SA310393058GA SA310393070GA SA310219831GA SA310260032GA SA310260360GA SA310290622GA SA310290709GA SA310290713GA SA310290721GA SA310290722GA SA310290726GA SA310290727GA SA310290728GA SA310290730GA SA310290732GA SA310290733GA SA310300792GA SA310300796GA SA310300800GA SA310300803GA SA310331217GA SA310341382GA SA310341386GA SA310372023GA SA310382527GA SA310382783GA SA310392994GA SA310392995GA SA310392996GA SA310393000GA SA310393001GA SA310393003GA SA310393004GA SA310393013GA SA310393019GA SA310393031GA SA310393038GA SA310393040GA SA310393041GA SA310393042GA SA310393044GA SA310393045GA SA312111667GA SA312121704GA SA312121712GA SA312121746GA SA312121758GA SA312121781GA SA312121792GA SA312121793GA SA312121840GA SA312121847GA SA312121851GA SA312121853GA SA312121877GA SA312121886GA SA312121889GA SA312121893GA SA312121904GA SA312121917GA SA312152317GA SA312152321GA SA312152339GA SA312059565GA SA312069778GA SA312070028GA SA312111655GA SA312121694GA SA312121742GA SA312121754GA SA312121788GA SA312121810GA SA312121828GA SA312121835GA SA312121836GA SA312121845GA SA312121854GA SA312121860GA SA312121861GA SA312121862GA SA312121863GA SA312121864GA SA312121869GA SA312121870GA SA312121880GA SA312121881GA SA312121882GA SA312121883GA SA312121884GA SA312121885GA SA312121888GA SA312121905GA SA312121906GA SA312121907GA SA312121909GA SA312121911GA SA312121912GA SA312121915GA SA312121918GA SA312121921GA SA312121923GA SA312121992GA SA312121998GA SA312152314GA SA312152315GA SA312152343GA SA312152359GA SA312152491GA SA312152499GA SA312152500GA SA312152552GA SA312152557GA SA312232803GA SA312232811GA SA312232822GA SA312232852GA SA312232855GA SA312232871GA SA312242956GA SA312242965GA SA312242978GA SA312242982GA SA312242997GA SA312243002GA SA312243003GA SA307070441GR SA307190742GR SA307220903GR SA307221106GR SA308155961GR SA308186422GR SA308196454GR SA308196455GR SA308196468GR SA308511473GR SA310518642GR SA311225976GR SA311226002GR SA311236082GR SA311236116GR SA311256797GR SA311256799GR SA311266954GR SA311266960GR SA311266963GR SA311266965GR SA311266968GR SA311266972GR SA311266973GR SA311266975GR SA311266977GR SA311266978GR SA311266980GR SA311266981GR SA311266985GR SA311266987GR SA311266988GR SA311266989GR SA311266990GR SA311266997GR SA311267006GR SA311267009GR SA311267011GR SA312070013GA SA312111656GA SA312121752GA SA312121777GA SA312121779GA SA312121782GA SA312121783GA SA312121785GA SA312121789GA SA312121856GA SA312121867GA SA312152507GA SA312152522GA SA312152551GA SA312152553GA SA312192595GA SA312192646GA SA312232792GA SA312232793GA SA312232794GA SA312232804GA SA312232816GA SA312232817GA SA312232818GA SA312232867GA SA312242953GA SA312242957GA SA312242958GA SA312242959GA SA312242960GA SA312242961GA SA312242962GA SA312242963GA SA312242964GA SA312242966GA SA312242967GA SA312242970GA SA312242972GA SA312242975GA SA312242976GA SA312242979GA SA312242980GA SA312242986GA SA312242987GA SA312242990GA SA312242992GA SA312242998GA SA312242999GA SA312243004GA SA312243007GA SA312243008GA SA312243010GA SA312243011GA SA312243012GA SA312243017GA SA312243021GA SA312243026GA SA312111589GA SA312121874GA SA312132131GA SA312152309GA SA312152310GA SA312152320GA SA312152328GA SA312152333GA SA312152336GA SA312152337GA SA312152352GA SA312152358GA SA312152477GA SA312152502GA SA312152548GA SA312152549GA SA312242955GA SA312242974GA SA312242977GA SA312242981GA SA312242988GA SA312242989GA SA312243000GA SA312243001GA SA312243005GA SA312243014GA SA312243015GA SA312243018GA SA312243020GA SA312243022GA SA312243023GA SA312243025GA SA312243030GA SA312243031GA SA312243032GA SA312243033GA SA312243034GA SA312243036GA SA312243037GA SA312243038GA SA312243041GA SA312243044GA SA312243045GA SA312243046GA SA312243047GA SA312243048GA SA312243049GA SA312243050GA SA312243051GA SA312243053GA SA312243054GA SA312243056GA SA312243057GA SA312253063GA SA312253064GA SA312253065GA SA312253067GA SA312253068GA SA312253070GA SA312253072GA SA312253074GA SA312253091GA SA312253092GA SA312253094GA SA312253096GA SA312253099GA SA312253100GA SA312253101GA SA312253111GA SA307221173GR SA307231428GR SA312242971GA SA312242991GA SA312242994GA SA312243013GA SA312253061GA SA312253062GA SA312253073GA SA312253075GA SA312253076GA SA312253078GA SA312253080GA SA312253084GA SA312253085GA SA312253086GA SA312253087GA SA312253088GA SA312253089GA SA312253105GA SA312253107GA SA312253108GA SA312253109GA SA312253110GA SA312253115GA SA312253118GA SA312253123GA SA312253124GA SA312253128GA SA312253130GA SA312253133GA SA312253134GA SA312253135GA SA312253136GA SA312253138GA SA312253139GA SA312253141GA SA312253144GA SA312253146GA SA312253147GA SA312253148GA SA312253149GA SA312253151GA SA312253153GA SA312253159GA SA312253164GA SA312253172GA SA312253183GA SA312253191GA SA312253207GA SA312192713GA SA312192724GA SA312232872GA SA312232907GA SA312253097GA SA312253154GA SA312253158GA SA312253160GA SA312253186GA SA312253200GA SA307251719GR SA312232819GA SA312253060GA SA312253095GA SA312253098GA SA312253106GA SA312253113GA SA312253122GA SA312253125GA SA312253129GA SA312253132GA SA312253152GA SA312253167GA SA312253171GA SA312253234GA SA312253265GA SA312253175GA SA312253180GA SA312253277GA SA312253289GA SA312253304GA SA312253306GA SA312253316GA SA312253326GA SA312253328GA SA312253331GA SA312253346GA SA312253348GA SA312253350GA SA312253353GA SA312253357GA SA312253359GA SA312253360GA SA312253366GA SA312253370GA SA312253373GA SA312253382GA SA312253407GA SA312253408GA SA312253409GA SA312253412GA SA312253415GA SA312253419GA SA312253420GA SA312253421GA SA312253433GA SA312253436GA SA312253443GA SA312253449GA SA312253466GA SA312253479GA SA312253506GA SA312253508GA SA312253509GA SA312253531GA SA312253533GA SA312253534GA SA312253544GA SA312253548GA SA312253551GA SA312122013GA SA312152325GA SA312152474GA SA312152492GA SA312152498GA SA312152501GA SA312152504GA SA312152506GA SA312152510GA SA312152518GA SA312152524GA SA312152535GA SA312152562GA SA312232824GA SA312232825GA SA312232827GA SA312232832GA SA312232837GA SA312232839GA SA312232842GA SA312232843GA SA312232845GAMKE SA312243028GA SA312253081GA SA312253126GA SA312253137GA SA312253155GA SA312253187GA SA312253226GA SA312253236GA SA312253290GA SA312253302GA SA312253322GA SA312253323GA SA312253324GA SA312253329GA SA312253333GA SA312253338GA SA312253340GA SA312253342GA SA312253347GA SA312253356GA SA312253358GA SA312253365GA SA312253372GA SA312253374GA SA312253376GA SA312253378GA SA312253380GA SA312253383GA SA312253386GA SA312253388GA SA312253390GA SA312253391GA SA312253396GA SA312253398GA SA312253400GA SA312253401GA SA312253402GA SA312253404GA SA312253410GA SA312253413GA SA312253414GA SA312253418GA SA312253424GA SA312253434GA SA312253435GA SA312253438GA SA312253442GA SA312253444GA SA312253447GA SA312253450GA SA312253461GA SA312253468GA SA312253470GA SA312253472GA SA312253474GA SA312253477GA SA312253478GA SA312253481GA SA312253483GA SA312253484GA SA312253485GA SA312253486GA SA312253490GA SA312253491GA SA312253492GA SA312253494GA SA312253495GA SA312253498GA SA312253500GA SA312253501GA SA312253503GA SA312253504GA SA312253507GA SA312253510GA SA312253511GA SA312253514GA SA312253515GA SA312253518GA SA312253528GA SA312253529GA SA312253532GA SA312253536GA SA312253537GA SA312253538GA SA312253540GA SA312253541GA SA312253542GA SA312253543GA SA312253545GA SA312253546GA SA312253547GA SA312253549GA SA312253550GA SA312253552GA SA312253579GA SA312253583GA SA312253595GA SA312253599GA SA312253600GA SA312253606GA SA312263614GA SA312263662GA SA312111462GA SA312111476GA SA312111491GA SA312111495GA SA312111502GA SA312111504GA SA312253209GA SA312253212GA SA312253224GA SA312253225GA SA312253248GA SA312253258GA SA312253259GA SA312253260GA SA312253343GA SA312253439GA SA312253496GA SA312263616GA SA312263620GA SA312263627GA SA312263635GA SA312263644GA SA312263661GA SA312263664GA SA312263685GA SA312263703GA SA312263704GA SA312263706GA SA312263707GA SA312263709GA SA312263716GA SA312263719GA SA312263726GA SA312263731GA SA312263733GA SA312263749GA SA312263761GA SA312263762GA SA312263797GA SA312243024GA SA312253165GA SA312253166GA SA312253168GA SA312253169GA SA312253170GA SA312253173GA SA312253174GA SA312253176GA SA312253178GA SA312253182GA SA312253188GA SA312253190GA SA312253244GA SA312253245GA SA312253252GA SA312253261GA SA312253270GA SA312253275GA SA312253280GA SA312253283GA SA312253303GA SA312253337GA SA312253345GA SA312253482GA SA312253497GA SA312263643GA SA312263649GA SA312263651GA SA312263652GA SA312263673GA SA312263680GA SA312263682GA SA312263688GA SA312263689GA SA312263690GA SA312263691GA SA312263692GA SA312263694GA SA312263695GA SA312263699GA SA312263701GA SA312263708GA SA312263711GA SA312263712GA SA312263713GA SA312263715GA SA312263717GA SA312263718GA SA312263720GA SA312263723GA SA312263728GA SA312263729GA SA312263730GA SA312263732GA SA312263734GA SA312263738GA SA312263744GA SA312263745GA SA312263746GA SA312263747GA SA312263753GA SA312263754GA SA312263755GA SA312263756GA SA312263757GA SA312263759GA SA312263760GA SA312263769GA SA312263770GA SA312263771GA SA312263772GA SA312263774GA SA312263776GA SA312263777GA SA312263778GA SA312263779GA SA312263780GA SA312263781GA SA312263783GA SA312263784GA SA312263786GA SA312263789GA SA312263790GA SA312263792GA SA312263795GA SA312263798GA SA312263804GA SA312263805GA SA312263810GA SA312263829GA SA312263830GA SA312263831GA SA312263832GA SA312263833GA SA312263834GA SA312263836GA SA312263837GA SA312263838GA SA312263839GA SA312263840GA SA312263841GA SA312263842GA SA312263843GA SA312263844GA SA312263845GA SA312263846GA SA312263847GA SA312263849GA SA312263850GA SA312263851GA SA312263852GA SA312263854GA SA312263855GA SA312263856GA SA312263857GA SA312263858GA SA312263859GA SA312263860GA SA312263861GA SA312263864GA SA312263865GA SA312263867GA SA312263868GA SA312263869GA SA312263871GA SA312263872GA SA312263875GA SA312263876GA SA312263877GA SA312263878GA SA312263879GA SA312263880GA SA312263881GA SA312263882GA SA312263883GA SA312263884GA SA312263885GA SA312263886GA SA312263888GA SA312263889GA SA312263890GA SA312263893GA SA312263894GA SA312263895GA SA312263896GA SA312263897GA SA312263899GA SA312263900GA SA312263901GA SA312263902GA SA312263903GA SA312263906GA SA312263907GA SA312263912GA SA312263915GA SA312263916GA SA312263917GA SA312263918GA SA312263930GA SA312253336GA SA312263679GA SA312263681GA SA312263684GA SA312263698GA SA312263722GA SA312263724GA SA312263725GA SA312263736GA SA312263739GA SA312263740GA SA312263741GA SA312263742GA SA312263758GA SA312263775GA SA312263807GA SA312263812GA SA312263870GA SA312263873GA SA312263913GA SA312263919GA SA312263924GA SA312263927GA SA312263955GA SA312263964GA SA312059565GA SA312111655GA SA312121694GA SA312121742GA SA312121754GA SA312121810GA SA312121828GA SA312121835GA SA312121836GA SA312121845GA SA312121854GA SA312121860GA SA312121861GA SA312121862GA SA312121863GA SA312121864GA SA312121869GA SA312121880GA SA312121881GA SA312121882GA SA312121883GA SA312121884GA SA312121885GA SA312121888GA SA312121905GA SA312121906GA SA312121907GA SA312121909GA SA312121911GA SA312121912GA SA312121915GA SA312121921GA SA312121923GA SA312121992GA SA312121998GA SA312242956GA SA312242965GA SA312242978GA SA312242997GA SA312243002GA SA312263743GA SA312263748GA SA312263767GA SA312263782GA SA312263794GA SA312263809GA SA312263816GA SA312263822GA SA312263956GA SA312263958GA SA312263962GA SA312263963GA SA312263965GA SA312263969GA SA312263970GA SA312263971GA SA312264010GA SA312264027GA SA312264029GA SA312264040GA SA312264042GA SA312264044GA SA312264046GA SA312264049GA SA312264050GA SA312264053GA SA312264059GA SA312264061GA SA312263892GA SA312253502GA SA312253520GA SA312253588GA SA312253592GA SA312263618GA SA312263624GA SA312263645GA SA312242982GA SA312253179GA SA312263977GA SA312284299GA SA312284305GA SA312284316GA SA312284328GA SA312284329GA SA312284335GA SA312284165GA SA312284269GA SA312284278GA SA312264048GA SA312284155GA SA312284180GA SA312284245GA SA312253291GA SA312253300GA SA312253314GA SA312253317GA SA312253327GA SA312253330GA SA312253352GA SA312263908GA SA312263920GA SA312263923GA SA312263990GA SA312264000GA SA312264002GA SA312284139GA SA312284185GA SA312284279GA SA312284284GA SA312284285GA SA312284292GA SA312284301GA SA312284302GA SA312284303GA SA312284304GA SA312284313GA SA312284314GA SA312284315GA SA312284317GA SA312284318GA SA312284319GA SA312284320GA SA312284321GA SA312284322GA SA312284324GA SA312284325GA SA312284326GA SA312284327GA SA312284330GA SA312284331GA SA312284332GA SA312284333GA SA312284334GA SA312284338GA SA312284339GA SA312284340GA SA312284341GA SA312284342GA SA312294344GA SA312294349GA SA312294350GA SA312294353GA SA312284298GA SA312101397GA SA312263817GA SA312263914GA SA312264007GA SA312284223GA SA312284234GA SA312284246GA SA312284247GA SA312284248GA SA312284250GA SA312284251GA SA312284252GA SA312284255GA SA312284256GA SA312284257GA SA312284258GA SA312284268GA SA312284273GA SA312284297GA SA312294343GA SA312294345GA SA312294354GA SA312294355GA SA312294356GA SA312294357GA SA312294360GA SA312294361GA SA312294362GA SA312294363GA SA312294365GA SA312294367GA SA312294368GA SA312294369GA SA312294371GA SA312294372GA SA312294375GA SA312294376GA SA312294377GA SA312294378GA SA312294379GA SA312294381GA SA312294384GA SA312294385GA SA312294386GA SA312294388GA SA312294391GA SA312294392GA SA312294393GA SA312294394GA SA312294395GA SA312294397GA SA312294399GA SA312294400GA SA312294402GA SA312294404GA SA312294405GA SA312294406GA SA312294407GA SA312294408GA SA312294409GA SA312294410GA SA312294412GA SA312294414GA SA312294416GA SA312294420GA SA312294422GA SA312294424GA SA312294425GA SA312294429GA SA312294432GA SA312294437GA SA312294438GA SA312294440GA SA312294443GA SA312294450GA SA312294451GA SA312294454GA SA312294456GA SA312294457GA SA312294461GA SA312294463GA SA312294464GA SA312294467GA SA312294471GA SA312294473GA SA312294485GA SA312294517GA SA312294556GA SA312294427GA SA312294439GA SA312294447GA SA312294476GA SA312294482GA SA312294484GA SA312294491GA SA312294426GA SA312294433GA SA312294481GA SA312294546GA SA312294551GA SA312294613GA SA312294614GA SA312294616GA SA312294618GA SA312294632GA SA312294656GA SA312294657GA SA312294662GA SA312294663GA SA312294665GA SA312294679GA SA312294597GA SA312294641GA SA312294644GA SA312294654GA SA312294661GA SA312294666GA SA312294667GA SA312294669GA SA312294670GA SA312294673GA SA312294674GA SA312294677GA SA312294382GA SA312294445GA SA312294502GA SA312294548GA SA312294589GA SA312294593GA SA312294594GA SA312294595GA SA312294598GA SA312294600GA SA312294601GA SA312294603GA SA312294604GA SA312294606GA SA312294609GA SA312294612GA SA312294623GA SA312294626GA SA312294630GA SA312294642GA SA312232853GA SA312232854GA SA312232862GA SA312232863GA SA312232864GA SA312232866GA SA312232874GA SA312232875GA SA312232876GA SA312232877GA SA312232880GA SA312253232GA SA312253557GA SA312253561GA SA312253562GA SA312253564GA SA312253567GA SA312253568GA SA312253570GA SA312253572GAMKE SA312253587GA SA312253596GA SA312253598GA SA312263613GA SA312263622GA SA312263623GA SA312263625GA SA312263629GA SA312263631GA SA312263634GA SA312263640GA SA312263641GA SA312263642GA SA312263646GA SA312263650GA SA312263654GA SA312263656GA SA312263657GA SA312263669GA SA312263670GA SA312263672GA SA312263674GA SA312263750GA SA312263799GA SA312263800GA SA312263802GA SA312263813GA SA312263951GA SA312263954GA SA312263960GA SA312263961GA SA312263967GA SA312263968GA SA312263972GA SA312263974GA SA312264014GA SA312264015GA SA312264016GA SA312264019GA SA312264020GA SA312264022GA SA312264024GA SA312264025GA SA312264028GA SA312264034GA SA312264035GA SA312264037GA SA312264051GA SA312264052GA SA312264055GA SA312264056GA SA312264058GA SA312264062GA SA312264063GA SA312264065GA SA312264071GA SA312264089GA SA312264093GA SA312264094GA SA312264096GA SA312264100GA SA312284116GA SA312284123GA SA312284126GA SA312284217GA SA312284238GA SA312284240GA SA312284253GA SA312284263GA SA312284270GA SA312294413GA SA312294535GA SA312294560GA SA312294566GA SA312294568GA SA312294571GA SA312294578GA SA312294579GA SA312294581GA SA312294584GA SA312294586GA SA312294588GA SA312294592GA SA312294605GA SA312294607GA SA312294608GA SA312294634GA SA312294658GA SA312111526GA SA312192590GA SA312192603GA SA312192605GA SA312192632GA SA312192633GA SA312192639GA SA312192651GA SA312192674GA SA312192726GA SA312232856GA SA312232857GA SA312232858GA SA312232882GA SA312232886GA SA312232895GA SA312232897GA SA312232905GA SA312253268GA SA312253274GA SA312253375GA SA312253523GA SA312253525GA SA312253526GA SA312253527GA SA312253530GA SA312253535GA SA312253539GA SA312253553GA SA312253554GA SA312253555GA SA312253556GA SA312253558GA SA312253565GA SA312253566GA SA312253569GA SA312253571GA SA312253573GA SA312253574GA SA312253575GA SA312253576GA SA312253577GA SA312253580GA SA312253581GA SA312253582GA SA312253584GA SA312253585GA SA312253586GA SA312253590GA SA312253597GA SA312253601GA SA312253602GA SA312253603GA SA312253605GA SA312263615GA SA312263619GA SA312263658GA SA312263660GA SA312263675GA SA312263806GA SA312263821GA SA312264033GA SA312284219GA SA312284220GA SA312284226GA SA312284229GA SA312284236GA SA312284237GA SA312284239GA SA312284241GA SA312284242GA SA312284243GA SA312284249GA SA312284259GA SA312284261GA SA312284264GA SA312284265GA SA312284266GA SA312284267GA SA312284272GA SA312284274GA SA312284276GA SA312294591GA SA312294635GA SA312294680GA SA312294684GA SA312294690GA SA312294692GA SA312294716GA SA312294754GA SA312294764GA SA310393138GR SA312111469GA SA312192631GA SA312192681GA SA312232881GA SA312232888GA SA312232892GA SA312232898GA SA312232899GA SA312232900GA SA312232902GA SA312232910GA SA312253216GA SA312253235GA SA312253241GA SA312253254GA SA312263667GA SA312284222GA SA312294617GA SA312294436GA SA312294544GA SA312294582GA SA312152537GA SA312294430GA SA312294449GA SA312294487GA SA312294494GA SA312294602GA SA312294678GA SA312294685GA SA312294691GA SA308449621GR SA309466841GR SA309477006GRMKE SA309517537GR SA310434100GR SA310444851GR SA310507242GR SA310507317GR SA310517664GR SA311153342GR SA311287197GR SA312091333GA SA312101430GA SA312111482GA SA312111493GA SA312111496GA SA312111664GA SA312121726GA SA312121764GA SA312121981GA SA312152324GA SA312152516GA SA312152517GA SA312152521GA SA312152532GA SA312152534GA SA312152555GA SA312232800GA SA312232809GA SA312232861GA SA312232883GA SA312232904GA SA312243009GA SA312243039GA SA312243042GA SA312253103GA SA312253127GA SA312253354GA SA312263676GA SA312263863GA SA312263983GA SA312284159GA SA312284254GA SA312314798GA SA312314800GA SA312314811GA SA312314812GA SA312314813GA SA312314816GA SA312314818GA SA312314824GA SA312314828GA SA307030413GR SA312253292GA SA312263666GA SA312263687GA SA312263696GA SA312263808GA SA312263953GA SA312264017GA SA312264064GA SA312264087GA SA312284228GA SA312314796GA SA312314802GA SA312314809GA SA312314823GA SA312314825GA SA312314808GA SA312314827GA SA312152476GA SA312232812GA SA312243029GA SA312253079GA SA312253093GA SA312253112GA SA312253181GA SA312253285GA SA312253286GA SA312253288GA SA312253309GA SA312253310GA SA312253319GA SA312253321GA SA312253334GA SA312253335GA SA312253344GA SA312253371GA SA312263862GA SA312264005GA SA312284130GA SA312284179GA SA312284182GA SA312294348GA SA312294519GA SA312294675GA SA312294693GA SA312294694GA SA312294731GA SA311122166GA SA312101440GA SA312192682GA SA312232879GA SA312232901GA SA312253198GA SA312253201GA SA312253281GA SA312253489GA SA312253578GA SA312263628GA SA312263648GA SA312263653GA SA312263678GA SA312263697GA SA312263714GA SA312263764GA SA312263787GA SA312263952GA SA312263959GA SA312264018GA SA312264045GA SA312284124GA SA312284227GA SA311327882GR SA312263621GA SA312263638GA SA312263663GA SA312263768GA SA312263828GA SA312264095GA SA312284132GA SA312284279GA SA312284302GA SA312284326GA SA312294448GA SA312152322GA SA312253156GA SA312253332GA SA312324832GA SA312294590GA SA310424037GR SA310507366GR SA311163483GR SA311184416GR SA307110548GR SA308084776GR SA308186429GR SA308196471GR SA309365236GR SA311184390GR SA311256787GR SA311266952GR SA311266958GR SA311266970GR SA311468404GR SA312080822GA SA312111516GA SA312121822GA SA312121857GA SA312121876GA SA312121924GA SA312152326GA SA312152365GA SA312152486GA SA312152490GA SA312152503GA SA312152509GA SA312152527GA SA312152531GA SA312232820GA SA312243040GA SA312294704GA SA312294707GA SA312294711GA SA312294715GA SA312294762GA SA312294763GA SA312294768GA SA312294769GA SA312294771GA SA312294773GA SA312294775GA SA312294778GA SA312294779GA SA312294780GA SA312294781GA SA312294782GA SA312294785GA SA312294786GA SA312294787GA SA312294788GA SA312294789GA SA312294790GA SA312294791GA SA312344868GA SA312080756GA SA312243019GA SA312344908GA SA312344936GA SA312344941GA SA312344942GA SA312344945GA SA312344955GA SA312344976GA SA312344989GA SA312344995GA SA312345004GA SA312345007GA SA312345010GA SA312345012GA SA312345013GA SA312345015GA SA312345020GA SA312345036GA SA312345038GA SA312345040GA SA312345042GA SA312345050GA SA312345052GA SA312345060GA SA312345065GA SA312345071GA SA312345109GA SA312344847GA SA312344922GA SA312344946GA SA312344988GA SA312345002GA SA312345044GA SA312345047GA SA312345051GA SA312345053GA SA312345054GA SA312345055GA SA312345057GA SA312345058GA SA312345062GA SA312345069GA SA312345106GA SA312345108GA SA312345122GA SA312345125GA SA312345128GA SA312345129GA SA312345140GA SA312345151GA SA312345158GA SA312253464GA SA312344965GA SA312345045GA SA312345067GA SA312345070GA SA312345105GA SA312345116GA SA312345117GA SA312345119GA SA312345121GA SA312345123GA SA312345124GA SA312345126GA SA312345127GA SA312345132GA SA312345133GA SA312345144GA SA312355212GA SA312355218GA SA312355223GA SA312355224GA SA312253267GA SA312253377GA SA312345034GA SA312345076GA SA312345078GA SA312345081GA SA312345085GA SA312345087GA SA312345090GA SA312345092GA SA312345100GA SA312345134GA SA312345168GA SA312345169GA SA312345170GA SA312345176GA SA312345189GA SA312345190GA SA312345193GA SA312355195GA SA312355196GA SA312355201GA SA312355202GA SA312355205GA SA312355210GA SA312355211GA SA312355216GA SA312355221GA SA312355318GA SA312355319GA SA312355322GA SA312355324GA SA312355327GA SA312355329GA SA312355330GA SA312355331GA SA312355332GA SA312355335GA SA312355336GA SA312355339GA SA312355342GA SA312355343GA SA312355344GA SA312355407GA SA312355408GA SA312355409GA SA312355414GA SA312345048GA SA312355227GA SA312355231GA SA312355234GA SA312355235GA SA312355387GA SA312355389GA SA312355225GA SA312355449GA SA312355473GA SA312355517GA SA312344985GA SA312355228GA SA312355232GA SA312355237GA SA312355239GA SA312355240GA SA312355242GA SA312355243GA SA312355244GA SA312355245GA SA312355246GA SA312355249GA SA312355251GA SA312355252GA SA312355254GA SA312355439GA SA312355440GA SA312355446GA SA312355451GA SA312355468GA SA312355207GA SA312111487GA SA312253196GA SA312344940GA SA312344959GA SA312344968GA SA312344970GA SA312344973GA SA312355442GA SA312355504GA SA312355509GA SA312355518GA SA312355521GA SA312294488GA SA312294490GA SA312294500GA SA312294503GA SA312294506GA SA312294510GA SA312294525GA SA312294529GA SA312294621GA SA312294683GA SA312294687GA SA312294689GA SA312294699GA SA312294703GA SA312294705GA SA312294706GA SA312294708GA SA312294709GA SA312294712GA SA312294713GA SA312294714GA SA312294717GA SA312294718GA SA312294720GA SA312294721GA SA312294723GA SA312294725GA SA312294726GA SA312294727GA SA312294728GA SA312294730GA SA312294732GA SA312294753GA SA312294755GA SA312294757GA SA312294759GA SA312294760GA SA312294765GA SA312294766GA SA312294767GA SA312294772GA SA312294783GA SA312345049GA SA312345066GA SA312345107GA SA312345110GA SA312345112GA SA312345113GA SA312345114GA SA312345115GA SA312242969GA SA312243052GA SA312253114GA SA312253177GA SA312253320GA SA312253355GA SA312365774GA SA312365790GA SA312375895GA SA312375900GA SA312375901GA SA312375950GA SA312253305GA SA312284183GA SA312355375GA SA312355376GA SA312355378GA SA312355379GA SA312355381GA SA312355382GA SA312355383GA SA312355384GA SA312355388GA SA312355390GA SA312355391GA SA312355392GA SA312355393GA SA312355394GA SA312355396GA SA312355397GA SA312355398GA SA312355399GA SA312355400GA SA312355401GA SA312355403GA SA312355404GA SA312355447GA SA312355448GA SA312355453GA SA312355456GA SA312355459GA SA312355460GA SA312355462GA SA312355463GA SA312355470GA SA312375913GA SA312375924GA SA312375932GA SA312375937GA SA312375938GA SA312375940GA SA312375941GA SA312375982GA SA312375988GA SA312375991GA SA312375994GA SA312375998GA SA312376000GA SA312376001GA
SA312376004GA SA312376008GA SA312376013GA SA312294431GA SA312294475GA SA312294688GA SA312365712GA SA312365804GA SA312365813GA SA312375952GA SA312375955GA SA312375956GA SA312375957GA SA312375958GA SA312375959GA SA312375960GA SA312375961GA SA312375963GA SA312375964GA SA312375965GA SA312375966GA SA312375968GA SA312375973GA SA312375979GA SA312376022GA SA312376027GA SA312376039GA SA312376057GA SA312376061GA SA312376064GA SA312376065GA SA312376066GA SA312376069GA SA312376070GA SA312101399GA SA312111489GA SA312284121GA SA312355502GA SA312355503GA SA312355512GA SA312355530GA SA312365618GA SA312365663GA SA312365664GA SA312365716GA SA312365722GA SA312365723GA SA312365724GA SA312365726GA SA312365727GA SA312365728GA SA312365729GA SA312365734GA SA312365736GA SA312365738GA SA312365739GA SA312365740GA SA312365745GA SA312365855GAMKE SA312375867GA SA312375869GA SA312375884GA SA312375886GA SA312375904GA SA312375935GA SA312375953GA SA312375954GA SA312375981GA SA312376011GA SA312376044GA SA311195081GR SA312152362GA SA312152475GA SA312152495GA SA312152528GA SA312232808GA SA312243027GA SA312253437GA SA312314794GA SA312314804GA SA312314820GA SA312365589GA SA312365595GA SA312365598GA SA312365602GA SA312365621GA SA312365625GA SA312365628GA SA312365634GA SA312365647GA SA312365648GA SA312365650GA SA312365678GA SA312365751GA SA312365758GA SA312365762GA SA312365763GA SA312365765GA SA312365832GA SA312365833GA SA312365837GA SA312365839GA SA312365841GA SA312365844GA SA312365845GA SA312365850GA SA312376087GA SA312376133GA SA312376196GA SA312376207GA SA312376295GA SA312386311GA SA312111483GA SA312192695GA SA312253194GA SA312253195GA SA312253385GA SA312253394GA SA312253560GA SA312263659GA SA312263710GA SA312263735GA SA312263752GA SA312264031GA SA312294722GA SA312294724GA SA312294729GA SA312294758GA SA312314795GA SA312314797GA SA312314799GA SA312314803GA SA312314805GA SA312314806GA SA312314807GA SA312314810GA SA312314815GA SA312314817GA SA312314819GA SA312314821GA SA312314822GA SA312314826GA SA312365653GA SA312365655GA SA312365670GA SA312365675GA SA312365680GA SA312365711GA SA312365713GA SA312365718GA SA312365719GA SA312365720GA SA312365721GA SA312365725GA SA312365730GA SA312365735GA SA312365737GA SA312365741GA SA312365742GA SA312365743GA SA312365744GA SA312365749GA SA312365759GA SA312365760GA SA312365831GA SA312365835GA SA312365838GA SA312365842GA SA312365847GA SA312365848GA SA312365849GA SA312375874GA SA312375876GA SA312375887GA SA312375888GA SA312375890GA SA312375920GA SA312375930GA SA312375939GA SA312375962GA SA312375972GA SA312375975GA SA312375987GA SA312375989GA SA312375990GA SA312376003GA SA312376102GA SA312376103GA SA312376106GA SA312376116GA SA312376120GA SA312376122GA SA312376126GA SA312376128GA SA312376129GA SA312376131GA SA312376137GA SA312376140GA SA312376141GA SA312376143GA SA312376146GA SA312376151GA SA312376152GA SA312376155GA SA312376156GA SA312376164GA SA312376174GA SA312376177GA SA312376179GA SA312376182GA SA312376188GA SA312376208GA SA312378036GA SA312378038GA SA312378040GA SA312378043GA SA312378059GA SA312378063GA SA312111473GA SA312111499GA SA312232813GA SA312232847GA SA312242954GA SA312294417GA SA312294428GA SA312294460GA SA312294477GA SA312294478GA SA312294483GA SA312294489GA SA312294492GA SA312294496GA SA312294499GA SA312294501GA SA312294505GA SA312294507GA SA312294518GA SA312294520GA SA312294522GA SA312294524GA SA312294527GA SA312294587GA SA312294625GA SA312294646GA SA312294660GA SA312294681GA SA312294682GA SA312294686GA SA312294695GA SA312294697GA SA312294698GA SA312294700GA SA312294701GA SA312294702GA SA312294756GA SA312294777GA SA312294784GA SA312314793GA SA312314801GA SA312314814GA SA312375898GA SA312375907GA SA312375916GA SA312375923GA SA312375928GA SA312375942GA SA312375944GA SA312375947GA SA312375996GA SA312376010GA SA312376016GA SA312376018GA SA312376019GA SA312376020GA SA312376024GA SA312376026GA SA312376029GA SA312376030GA SA312376031GA SA312376032GA SA312376033GA SA312376034GA SA312376035GA SA312376036GA SA312376038GA SA312376041GA SA312376042GA SA312376043GA SA312376045GA SA312376046GA SA312376048GA SA312376049GA SA312376051GA SA312376053GA SA312376056GA SA312378017GA SA312344919GA SA312345028GA SA312345147GA SA312365531GA SA312365731GA SA312365802GA SA312365810GA SA312365811GA SA312376021GA SA312376040GA SA312376059GA SA312376060GA SA312376063GA SA312376067GA SA312376068GA SA312376072GA SA312376074GA SA312376083GA SA312376085GA SA312376086GA SA312376123GA SA312376130GA SA312376134GA SA312376135GA SA312376136GA SA312376139GA SA312376142GA SA312376145GA SA312376147GA SA312376149GA SA312376150GA SA312376153GA SA312376154GA SA312376157GA SA312376158GA SA312376159GA SA312376160GA SA312376161GA SA312376165GA SA312376167GA SA312376169GA SA312376170GA SA312376171GA SA312376172GA SA312376173GA SA312376175GA SA312376176GA SA312376181GA SA312376183GA SA312376185GA SA312376186GA SA312376187GA SA312376189GA SA312376192GA SA312376193GA SA312376195GA SA312376197GA SA312376201GA SA312376204GA SA312376205GA SA312376209GA SA312376214GA SA312376233GA SA312376235GA SA312376236GA SA312376237GA SA312376240GA SA312376241GA SA312376245GA SA312376246GA SA312376247GA SA312376248GA SA312376249GA SA312376250GA SA312376287GA SA312376292GA SA312376297GA SA312378006GA SA312378008GA SA312378009GA SA312378021GA SA312378024GA SA312378030GA SA307190730GR SA307190774GR SA311327818GR SA312018874GR SA312018883GR SA312018884GR SA312018905GR SA312028927GR SA312028938GR SA312028940GR SA312028941GR SA312080790GR SA312080856GR SA312080864GR SA312345014GA SA312355241GA SA312376047GA SA312376062GA SA312376132GA SA312376184GA SA312376198GA SA312376202GA SA312376206GA SA312376210GA SA312376211GA SA312376212GA SA312376215GA SA312376216GA SA312376219GA SA312376220GA SA312376221GA SA312376222GA SA312376223GA SA312376224GA SA312376225GA SA312376226GA SA312376228GA SA312376229GA SA312376230GA SA312376234GA SA312376239GA SA312376242GA SA312376243GA SA312376244GA SA312376300GA SA312378029GA SA312378032GA SA312378033GA SA312378034GA SA312386350GA SA312386372GA SA312386373GA SA312386377GA SA312386378GA SA312386380GA SA312386381GA SA312386382GA SA312386383GA SA312386385GA SA312386387GA SA312386390GA SA312386393GA SA312386394GA SA312386396GA SA312386399GA SA312386406GA SA312386413GA SA312386415GA SA312386421GA SA312386423GA SA312386428GA SA312344881GA SA312355197GA SA312355199GA SA312355204GA SA312355213GA SA312355214GA SA312355220GA SA312386388GA SA312386389GA SA312386401GA SA312386416GA SA312386417GA SA312386420GA SA312386431GA SA312386433GA SA312386438GA SA312386440GA SA312386441GA SA312386442GA SA312386448GA SA312386450GA SA312386454GA SA312386455GA SA312386458GA SA312386460GA SA312386484GA SA312386485GA SA312386487GA SA312386488GA SA312386490GA SA312355215GA SA312365573GA SA312378048GA SA312386326GA SA312386333GA SA312386347GA SA312386353GA SA312386356GA SA312386360GA SA312386361GA SA312386362GA SA312386363GA SA312386367GA SA312386368GA SA312386369GA SA312386405GA SA312386434GA SA312386435GA SA312386445GA SA312386447GA SA312386463GA SA312386465GA SA312386466GA SA312386471GA SA312386472GA SA312386475GA SA312386476GA SA312386510GA SA312386512GA SA312386514GA SA312386537GA SA312386546GA SA312386549GA SA312386553GA SA312386566GA SA312386568GA SA312386569GA SA312386570GA SA312386577GA SA312386579GA SA312386586GA SA312386607GA SA312386637GA SA307190739GR SA307220999GR SA307221171GR SA307231423GR SA309517807GR SA310392949GR SA310393140GR SA311122061GR SA311317733GR SA311327814GR SA311327881GR SA311428116GR SA311468402GR SA312049448GA SA312122000GR SA312122005GR SA312355377GA SA312365605GA SA312375926GA SA312375993GA SA312376089GA SA312378051GA SA312386330GA SA312386337GA SA312386359GA SA312386364GA SA312386376GA SA312386384GA SA312386400GA SA312386407GA SA312386411GA SA312386414GA SA312386418GA SA312386419GA SA312386425GA SA312386439GA SA312386452GA SA312386453GA SA312386456GA SA312386459GA SA312386461GA SA312386464GA SA312386468GA SA312386470GA SA312386473GA SA312386479GA SA312386486GA SA312386493GA SA312386496GA SA312386504GA SA312386513GA SA312386520GA SA312386525GA SA312386527GA SA312386536GA SA312386538GA SA312386540GA SA312386542GA SA312386544GA SA312386548GA SA312386551GA SA312386557GA SA312386558GA SA312386560GA SA312386563GA SA312386571GA SA312386581GA SA312386585GA SA312386592GA SA312386593GA SA312386598GA SA312386599GA SA312386609GA SA312386610GA SA312386613GA SA312386614GA SA312386617GA SA312386620GA SA312386623GA SA312386626GA SA312386653GA SA312386654GA SA312386655GA SA312386656GA SA312386657GA SA312386659GA SA312386660GA SA312386662GA SA312386664GA SA312386666GA SA312386670GA SA312386673GA SA312386681GA SA312386682GA SA312386695GA SA312386700GA SA312396702GA SA312396703GA SA312396721GA SA312396750GA SA312396752GA SA309477011GR SA309477012GR SA309497175GR SA310423816GR SA310517651GR SA310517678GR SA311184425GR SA311184441GR SA312355406GA SA312378019GA SA312386408GA SA312386436GA SA312386443GA SA312386444GA SA312386446GA SA312386449GA SA312386451GA SA312386474GA SA312386478GA SA312386500GA SA312386518GA SA312386522GA SA312386528GA SA312386531GA SA312386533GA SA312386543GA SA312386550GA SA312386603GA SA312386605GA SA312386627GA SA312386630GA SA312386631GA SA312386642GA SA312386644GA SA312386649GA SA312386652GA SA312386669GA SA312386676GA SA312386678GA SA312386683GA SA312386684GA SA312386686GA SA312386687GA SA312386689GA SA312386692GA SA312386694GA SA312386696GA SA312396701GA SA312396707GA SA312396708GA SA312396709GA SA312396710GA SA312396711GA SA312396712GA SA312396713GA SA312396714GA SA312396715GA SA312396716GA SA312396717GA SA312396719GA SA312396720GA SA312396723GA SA312396724GA SA312396725GA SA312396726GA SA312396727GA SA312396728GA SA312396729GA SA312396730GA SA312396731GA SA312396735GA SA312396743GA SA312396745GA SA312396749GA SA312396751GA SA312396754GA SA312396777GA SA312396783GA SA312396793GA SA312396795GA SA312386587GA SA312386606GA SA312386628GA SA312386629GA SA312386633GA SA312386646GA SA312386663GA SA312386667GA SA312386677GA SA312396913GA SA312396915GA SA312396916GA SA312396918GA SA312396919GA SA312396920GA SA312396921GA SA312396924GA SA312396926GA SA312396927GA SA312396929GA SA312396931GA SA312396932GA SA312396933GA SA312396934GA SA312396935GA SA312396936GA SA312396937GA SA312396939GA SA312396940GA SA312396988GA SA312396989GA SA307302016GR SA309517813GR SA309527918GR SA310038352GR SA310038427GR SA311174229GR SA311184382GR SA311195067GR SA311266992GR SA311287109GR SA312080748GR SA312080787GR SA312378035GA SA312386492GA SA312386494GA SA312386541GA SA312386565GA SA312386596GA SA312396848GA SA312396901GA SA312396909GA SA312396938GA SA312396961GA SA312396994GA SA312397014GA SA312397082GA SA312355461GA SA312376281GA SA312386498GA SA312386499GA SA312386507GA SA312386516GA SA312386539GA SA312386545GA SA312386554GA SA312386555GA SA312386556GA SA312386559GA SA312386564GA SA312386578GA SA312386582GA SA312386583GA SA312386584GA SA312386588GA SA312386589GA SA312386591GA SA312386594GA SA312386595GA SA312386597GA SA312386601GA SA312386602GA SA312386604GA SA312386608GA SA312386615GA SA312386622GA SA312386625GA SA312386641GA SA312386647GA SA312386648GA SA312386650GA SA312386658GA SA312396911GA SA312396914GA SA312396922GA SA312396925GA SA312396971GA SA312396973GA SA312396974GA SA312396983GA SA312396985GA SA312396986GA SA312396997GA SA312397000GA SA312397234GA SA312386495GA SA312386508GA SA312386572GA SA312386575GA SA312397161GA SA312397165GA SA312397173GA SA312397175GA SA312397180GA SA312397196GA SA312396977GA SA312396981GA SA312397204GA SA309517775GR SA309517786GR SA310413495GR SA311163915GR SA311184352GR SA311184354GR SA311184385GR SA312080553GR SA312080598GR SA312080630GR SA312080805GR SA312396789GA SA312396887GA SA312386668GA SA312396975GA SA312396987GA SA312396990GA SA312396993GA SA312396995GA SA312397070GA SA312397071GA SA312397171GA SA312397197GA SA312397210GA SA312407278GA SA312407288GA SA312344954GA SA312355477GA SA312397042GA SA312397155GA SA312397157GA SA312397159GA SA312397160GA SA312397168GA SA312397174GA SA312397176GA SA312397177GA SA312397178GA SA312397181GA SA312397182GA SA312397183GA SA312397184GA SA312397185GA SA312397186GA SA312397187GA SA312397191GA SA312397195GA SA312397202GA SA312397205GA SA312397206GA SA312397207GA SA312397209GA SA312397273GA SA312417340GA SA312417341GA SA312417359GA SA312417369GA SA312396801GA SA312396812GA SA312396820GA SA312396821GA SA312396841GA SA312396843GA SA312396854GA SA312396869GA SA312396890GA SA312396891GA SA312396893GA SA312396896GA SA312396992GA SA312396996GA SA312397001GA SA312397003GA SA312397004GA SA312397006GA SA312397008GA SA312397009GA SA312397012GA SA312397019GA SA312397021GA SA312397024GA SA312397027GA SA312397028GA SA312397029GA SA312397030GA SA312397062GA SA312397064GA SA312397065GA SA312397066GA SA312397067GA SA312397068GA SA312397069GA SA312397072GA SA312397075GA SA312397076GA SA312397078GA SA312397080GA SA312397084GA SA312397085GA SA312397086GA SA312397089GA SA312397252GA SA312397254GA SA312417367GA SA312417372GA SA312417375GA SA312417379GA SA312417385GA SA312345049GA SA312345066GA SA312345107GA SA312345110GA SA312345112GA SA312345113GA SA312345114GA SA312345115GA SA312397224GA SA312407289GA SA312407300GA SA312396704GA SA312396755GA SA312396758GA SA312396759GA SA312396772GA SA312396773GA SA312396790GA SA312396799GA SA312396804GA SA312396806GA SA312396811GA SA312396829GA SA312396832GA SA312396840GA SA312396846GA SA312396851GA SA312396866GA SA312396867GA SA312396872GA SA312396874GA SA312396876GA SA312396880GA SA312397246GA SA312376125GA SA312397272GA SA312407294GA SA312407297GA SA312417301GA SA312417313GA SA312417320GA SA312417321GA SA312417330GA SA312417334GA SA312417335GA SA312417336GA SA312417343GA SA312417347GA SA312417348GA SA312417355GA SA312417373GA SA312397002GA SA312397026GA SA312397169GA SA312407281GA SA312407292GA SA312407296GA SA312386437GA SA312397201GA SA312386524GA SA312397005GA SA312407282GA SA309517629GR SA309517727GR SA309517797GR SA309517832GR SA310423573GR SA311174253GR SA311184336GR SA311215701GR SA311287171GR SA311287217GR SA312132115GR SA310423806GR SA311163430GR SA312396718GA SA312396733GA SA312396734GA SA312396737GA SA312396738GA SA312396742GA SA312396746GA SA312396753GA SA312396757GA SA312396761GA SA312396764GA SA312396765GA SA312396767GA SA312396768GA SA312396769GA SA312396770GA SA312396771GA SA312396774GA SA312396778GA SA312396779GA SA312396784GA SA312396785GA SA312396786GA SA312396788GA SA312396792GA SA312396794GA SA312396797GA SA312396798GA SA312396803GA SA312396813GA SA312396814GA SA312396816GA SA312396830GA SA312396831GA SA312396833GA SA312396835GA SA312396836GA SA312396845GA SA312396849GA SA312396850GA SA312253349GA SA312396942GA SA312396763GA SA312396815GA SA312396817GA SA312396826GA SA312396855GA SA312396863GA SA312396865GA SA312396877GA SA312396878GA SA312396882GA SA312397100GA SA312397101GA SA312397103GA SA312397138GA SA312397140GA SA312397147GA SA312080888GA SA312152540GA SA312386680GA SA312396741GA SA312396744GA SA312396809GA SA312396822GA SA312396825GA SA312396828GA SA312396838GA SA312396839GA SA312396885GA SA312396907GA SA312396950GA SA312396956GA SA312396963GA SA312396965GA SA312397039GA SA312397044GA SA312397045GA SA312397054GA SA312397055GA SA312397093GA SA312417337GA SA312417339GA SA312417342GA SA312417351GA SA312417371GA SA312417378GA SA312417380GA SA312417384GA SA312417387GA SA312457740GA SA312457768GA SA312457878GA SA312457891GA SA312457910GA SA312457915GA SA312458015GA SA312458018GA SA312458020GA SA312386365GA SA312386671GA SA312386693GA SA312396732GA SA312396747GA SA312396748GA SA312396756GA SA312396760GA SA312396762GA SA312396776GA SA312396780GA SA312396782GA SA312396791GA SA312396796GA SA312396808GA SA312396844GA SA312396852GA SA312396900GA SA307200784GR SA309517718GR SA309517735GR SA311317730GR SA312111496GA SA312457810GA SA312457812GA SA312468802GA SA312468815GA SA312468821GA SA312468842GA SA312468854GA SA312468856GA SA309405825GR SA309517734GR SA311256769GR SA312386331GA SA312407290GA SA312407291GA SA312417391GA SA312417412GA SA312457683GA SA312457685GA SA312457692GA SA312468785GA SA312468790GA SA312468809GA SA312468810GA SA312468811GA SA312468830GA SA312468832GA SA312468834GA SA312468836GA SA312468843GA SA312468847GA SA312468855GA SA312468860GA SA312468863GA SA312468869GA SA312468880GA SA312468922GA SA312468923GA SA312468926GA SA312468931GA SA312468941GA SA312468945GA SA312468947GA SA312468950GA SA312468951GA SA312468958GA SA312457980GA SA312458099GA SA312458104GA SA312458195GA SA312458228GA SA312458232GA SA312458455GA SA312468586GA SA312468600GA SA312468955GA SA312468961GA SA312468965GA SA312468997GA SA312468998GA SA312469043GA SA312469045GA SA312469048GA SA312469053GA SA312469056GA SA312469057GA SA312469058GA SA312469065GA SA312469232GA SA312458233GA SA312458264GA SA312458397GA SA312458467GA SA312458468GA SA312458469GA SA312468539GA SA312468544GA SA312468554GA SA312468556GA SA312468558GA SA312468560GA SA312468561GA SA312468578GA SA312468590GA SA312469112GA SA312469159GA SA312469161GA SA312469180GA SA312469208GA SA312469212GA SA312469214GA SA312469216GA SA312469219GA SA312469220GA SA312469310GA SA312469224GA SA312469226GA SA312458160GA SA312468596GA SA312468597GA SA312468601GA SA312468604GA SA312468608GA SA312468609GA SA312468610GA SA312468611GA SA312468612GA SA312468614GA SA312468615GA SA312468616GA SA312468664GA SA312468665GA SA312468666GA SA312468668GA SA312468672GA SA312468675GA SA312468679GA SA312468686GA SA312468695GA SA312468696GA SA312468704GA SA312468707GA SA312468710GA SA312468711GA SA312468725GA SA312469014GA SA312469020GA SA312469024GA SA312469025GA SA312469026GA SA312469027GA SA312469037GA SA312469080GA SA312469089GA SA312469093GA SA312469094GA SA312469096GA SA312469098GA SA312469099GA SA312469100GA SA312469101GA SA312469227GA SA312469261GA SA312469417GA SA312469420GA SA312469428GA SA312469442GA SA312469508GA SA312469521GA SA312479532GA SA312479579GA SA312479580GA SA312468595GA SA312468599GA SA312468617GA SA312468652GA SA312468687GA SA312468689GA SA312468701GA SA312468702GA SA312468715GA SA312468724GA SA312468726GA SA312468731GA SA312468734GA SA312468738GA SA312469013GA SA312469015GA SA312469016GA SA312469017GA SA312469018GA SA312469019GA SA312469021GA SA312469022GA SA312469023GA SA312469028GA SA312469033GA SA312469073GA SA312469075GA SA312469076GA SA312469079GA SA312469082GA SA312469084GA SA312469087GA SA312469088GA SA312469091GA SA312469092GA SA312469095GA SA312469097GA SA312469223GA SA312469229GA SA312469280GA SA312479672GA SA312479673GA SA312479674GA SA312479676GA SA312479682GA SA312479697GA SA312479737GA SA312479741GA SA312479743GA SA312479747GA SA312479748GA SA312479775GA SA312479778GA SA312479800GA SA312479802GA SA312479805GA SA312479806GA SA312479813GA SA312479891GA SA312469377GA SA312469393GA SA312469396GA SA312469397GA SA312469398GA SA312469464GA SA312469473GA SA312469474GA SA312469476GA SA312469477GA SA312469479GA SA312469481GA SA312469486GA SA312469487GA SA312469491GA SA312470049GA SA312470068GA SA312479698GA SA312479761GA SA312479773GA SA312479779GA SA312479792GA SA312479822GA SA312479823GA SA312479826GA SA312479827GA SA312479831GA SA312479834GA SA312479836GA SA312479846GA SA312479847GA SA312479848GA SA312479849GA SA312479858GA SA311195112GR SA311297239GR SA311297240GR SA311297253GR SA311297306GR SA311297316GR SA311297326GR SA311297372GR SA311297391GR SA311297408GR SA311307468GR SA311317599GR SA311317697GR SA311317699GR SA311317702GR SA312458007GA SA312458047GA SA312458147GA SA312458176GA SA312468619GA SA312468656GA SA312468709GA SA312469413GA SA312470013GA SA312470021GA SA312470023GA SA312470027GA SA312470029GA SA312470030GA SA312470036GA SA312470040GA SA312470052GA SA312470053GA SA312470061GA SA312470063GA SA312470075GA SA312479675GA SA312479678GA SA312479684GA SA312479706GA SA312479736GA SA312479738GA SA312479745GA SA312479756GA SA312479757GA SA312479760GA SA312479763GA SA312479764GA SA312479774GA SA312479776GA SA312479780GA SA312479782GA SA312479786GA SA312479876GA SA312479897GA SA312479903GA SA312479904GA SA312479905GA SA312479906GA SA312479907GA SA312479909GA SA312479910GA SA312479912GA SA312479915GA SA312479931GA SA312479936GA SA312479940GA SA312397007GA SA312397011GA SA312397013GA SA312397016GA SA312397017GA SA312397020GA SA312397022GA SA312397023GA SA312458246GA SA312458256GA SA312468546GA SA312468576GA SA312468621GA SA312468727GA SA312479733GA SA312480234GA SA312480285GA SA312480294GA SA312480298GA SA312480301GA SA312480314GA SA312447448GA SA312447475GA SA312447517GA SA312447518GA SA312447523GA SA312447528GA SA312447530GA SA312447531GA SA312447557GA SA312468593GA SA312468653GA SA312468654GA SA312468683GA SA312468684GA SA312468685GA SA312468690GA SA312468693GA SA312468698GA SA312468719GA SA312468722GA SA312468723GA SA312468732GA SA312468736GA SA312468741GA SA312469248GA SA312469249GA SA312469263GA SA312469265GA SA312469268GA SA312469387GA SA312469440GA SA312469461GA SA312469471GA SA312469475GA SA312469485GA SA312469488GA SA312479538GA SA312479540GA SA312479732GA SA312479735GA SA312479749GA SA312479771GA SA312479772GA SA312479892GA SA312479893GA SA312479896GA SA312479914GA SA312479919GA SA312479920GA SA312479921GA SA312479922GA SA312479924GA SA312479934GA SA312479939GA SA312479947GA SA312479948GA SA312480128GA SA312480161GA SA312480177GA SA312480178GA SA312480185GA SA312480187GA SA312480204GA SA312480207GA SA312480458GA SA312480480GA SA312480504GA SA312480518GA SA312480522GA SA312480523GA SA312480524GA SA312480525GA SA312480526GA SA312480530GA SA312480531GA SA312480532GA SA312480536GA SA312480537GA SA312480539GA SA312480541GA SA312480542GA SA312480554GA SA312480561GA SA312480565GA SA312480566GA SA312480568GA SA312480570GA SA312480573GA SA312480612GA SA312447512GA SA312447521GA SA312447524GA SA312447527GA SA312447529GA SA312447532GA SA312447533GA SA312447534GA SA312447535GA SA312447537GA SA312447543GA SA312447545GA SA312447547GA SA312447551GA SA312447552GA SA312447553GA SA312447555GA SA312447569GA SA312469388GA SA312469389GA SA312469392GA SA312469394GA SA312469402GA SA312469421GA SA312469438GA SA312469456GA SA312469457GA SA312469460GA SA312469462GA SA312469468GA SA312469469GA SA312469470GA SA312469478GA SA312469482GA SA312469490GA SA312469492GA SA312469509GA SA312469511GA SA312469512GA SA312479577GA SA312479633GA SA312479659GA SA312480098GA SA312457759GA SA312458039GA SA312458157GA SA312468606GA SA312469209GA SA312469354GA SA312469483GA SA312479692GA SA312479900GA SA312480121GA SA312480122GA SA312480123GA SA312480167GA SA312480176GA SA312480192GA SA312480211GA SA312480215GA SA312480216GA SA312480218GA SA312480222GA SA312480233GA SA312480236GA SA312480237GA SA312480239GA SA312480240GA SA312480245GA SA312480246GA SA312480247GA SA312480250GA SA312480251GA SA312480252GA SA312480254GA SA312480256GA SA312480259GA SA312480260GA SA312480262GA SA312480263GA SA312480264GA SA312480267GA SA312480268GA SA312480269GA SA312480273GA SA312480274GA SA312480276GA SA312480278GA SA312480281GA SA312480282GA SA312480283GA SA312480286GA SA312480288GA SA312480289GA SA312480290GA SA312480291GA SA312480292GA SA312480295GA SA312480296GA SA312480299GA SA312480300GA SA312480302GA SA312480305GA SA312480306GA SA312480309GA SA312480311GA SA312480316GA SA312480318GA SA312480319GA SA312480320GA SA312480322GA SA312480323GA SA312480326GA SA312480328GA SA312480330GA SA312480356GA SA312480368GA SA312480380GA SA312480397GA SA312480398GA SA312480408GA SA312480416GA SA312480417GA SA312480423GA SA312480427GA SA312480432GA SA312480440GA SA312480442GA SA312480449GA SA312480451GA SA312480452GA SA312480456GA SA312480457GA SA312480460GA SA312480462GA SA312480465GA SA312480466GA SA312480471GA SA312480479GA SA312480481GA SA312480485GA SA312480487GA SA312480488GA SA312480490GA SA312480493GA SA312480494GA SA312480499GA SA312480510GA SA312480511GA SA312480513GA SA312480515GA SA312480516GA SA312480519GA SA312480520GA SA312480528GA SA312480533GA SA312480535GA SA312480538GA SA312480540GA SA312480560GA SA312480563GA SA312480575GA SA312480596GA SA312480598GA SA312480665GA SA312480726GA SA312480736GA SA312480746GA SA312480786GA SA312417368GA SA312447452GA SA312447454GA SA312447455GA SA312447457GA SA312447464GA SA312447465GA SA312447467GA SA312447471GA SA312447476GA SA312447478GA SA312447513GA SA312447544GA SA312447573GA SA312447574GA SA312447579GA SA312457626GA SA312458130GA SA312458139GA SA312458145GA SA312458151GA SA312458156GA SA312468602GA SA312468657GA SA312468677GA SA312468682GA SA312468708GA SA312468712GA SA312468713GA SA312468714GA SA312468716GA SA312468721GA SA312468728GA SA312468729GA SA312468730GA SA312468733GA SA312468735GA SA312468737GA SA312468739GA SA312468740GA SA312469165GA SA312469304GA SA312469385GA SA312469404GA SA312469411GA SA312469415GA SA312469425GA SA312469426GA SA312469434GA SA312469439GA SA312469446GA SA312469447GA SA312469452GA SA312469453GA SA312469495GAMKE SA312469504GA SA312469506GA SA312469519GA SA312479522GA SA312479537GA SA312479546GA SA312479547GA SA312479556GA SA312479578GA SA312479582GA SA312479583GA SA312479585GA SA312479593GA SA312479595GA SA312479606GA SA312479607GA SA312479612GA SA312479614GA SA312479615GA SA312479622GA SA312479623GA SA312479624GA SA312479625GA SA312479628GA SA312479629GA SA312479631GA SA312479650GA SA312479652GA SA312479653GA SA312479656GA SA312479660GA SA312479663GA SA312479664GA SA312479668GA SA312479669GA SA312479854GA SA312480217GA SA312480332GA SA312480335GA SA312480339GA SA312480355GA SA312480360GA SA312480364GA SA312480366GA SA312480372GA SA312480374GA SA312480385GA SA312480393GA SA312480395GA SA312480396GA SA312480638GA SA312480661GA SA312480662GA SA312480663GA SA312480664GA SA312480666GA SA312480667GA SA312480668GA SA312480669GA SA312480670GA SA312480671GA SA312480676GA SA312480678GA SA312480680GA SA312480683GA SA312480684GA SA312480687GA SA312480688GA SA312480690GA SA312480691GA SA312480705GA SA312480721GA SA312480729GA SA312480734GA SA312480745GA SA312468804GA SA312468894GA SA312468900GA SA312468927GA SA312468929GA SA312468985GA SA312479923GA SA312480819GA SA312480859GA SA312490873GA SA312490877GA SA312490888GA SA312447422GA SA312447515GA SA312447539GA SA312447541GA SA312447578GA SA312468773GA SA312468877GA SA312480340GA SA312480381GA SA312480390GA SA312480735GA SA312480784GA SA312480842GAMKE SA312480844GA SA312480846GA SA312480847GA SA312480848GA SA312480850GA SA312480852GA SA312480854GA SA312480855GA SA312480858GA SA312480862GA SA312480863GA SA312480869GA SA312490872GA SA312490881GA SA312490885GA SA312490891GA SA312490894GA SA312490913GA SA312490983GA SA312491150GA SA312491316GA SA312491317GA SA312491323GA SA312491324GA SA312491329GA SA312491333GA SA312491334GA SA312491336GA SA312491337GA SA312491339GA SA312491345GA SA312491357GA SA312491362GA SA312491365GA SA312396739GA SA312396802GA SA312396819GA SA312396847GA SA312396862GA SA312396875GA SA312396879GA SA312396906GA SA312469347GA SA312469480GA SA312469484GA SA312469489GA SA312470031GA SA312470039GA SA312470081GA SA312479746GA SA312479855GA SA312479872GA SA312479874GA SA312479875GA SA312479899GA SA312479938GA SA312479942GA SA312479968GA SA312480092GA SA312480095GA SA312480101GA SA312480102GA SA312480105GA SA312480106GA SA312480146GA SA312480208GA SA312480270GA SA312480644GA SA312480644GA SA312480709GA SA312480860GA SA312480865GA SA312480866GA SA312490880GA SA312491018GA SA312491047GA SA312491058GA SA312491063GA SA312491068GA SA312491070GA SA312491077GA SA312491084GA SA312491105GA SA312491109GA SA312491130GA SA312491136GA SA312491148GA SA312491155GA SA312491184GA SA309032112GA SA309042156GA SA309042158GA SA309042166GA SA309042204GA SA309042209GA SA309042212GA SA309527940GA SA309527940GA SA311090830GA SA311153186GA SA311195136GA SA311236331GA SA311246391GA SA311246394GA SA311246396GA SA311246401GA SA311246402GA SA311246403GA SA311246405GA SA311246406GA SA311246407GA SA311246413GA SA311246414GA SA311246415GA SA311246428GA SA311246431GA SA311246434GA SA311246448GA SA311297331GA SA311297332GA SA311307420GA SA311307433GA SA311307442GA SA311428096GA SA311428105GA SA311428112GA SA311428115GA SA311428119GA SA311428121GA SA311428124GA SA311428125GA SA311428129GA SA311428131GA SA311428134GA SA311428142GA SA311428144GA SA311428145GA SA311428148GA SA311428154GA SA311428158GA SA311428159GA SA312386503GA SA312049234GA SA307140603GA SA307140616GA SA307140620GA SA307140625GA SA307150647GA SA307150648GA SA307150649GA SA307150652GA SA307221111GA SA307221148GA SA307221167GA SA307231193GA SA307231198GA SA307231357GA SA307231367GA SA307231401GA SA307231409GA SA307220966GA SA307220970GA SA307220997GA SA307221019GA SA307221022GA SA307221032GA SA307221090GA SA307221093GA SA307221147GA SA307221188GA SA307231190GA SA307231371GA SA307231446GA SA307231447GA SA307241517GA SA307241675GA SA307241700GA SA307241520GA SA307241580GA SA307220871GA SA307220891GA SA307221166GA SA307241584GA SA307241693GA SA307241704GA SA307251709GA SA307251717GA SA307251723GA SA307251728GA SA307251749GA SA307312105GA SA307312107GA SA307312110GA SA307322112GA SA307322114GA SA307322116GA SA307322117GA SA307322118GA SA307322119GA SA307322120GA SA307322121GA SA307392723GA SA307393017GA SA307413109GAMKE SA307413110GA SA307423155GA SA307423158GA SA307423159GA SA307423163GA SA307423164GA SA307423169GA SA307423172GA SA307423173GA SA307423174GA SA307423177GA SA307423180GA SA307423182GA SA307423183GA SA307423184GA SA307443220GA SA307200780GA SA307362416GA SA307362418GA SA307362427GA SA307362436GA SA307504205GAMKE SA307514341GA SA308084617GA SA308084644GA SA308084773GA SA308094964GA SA308084760GA SA308105182GA SA308105183GA SA308105184GA SA308105192GA SA308105200GA SA308105206GA SA308105208GA SA308105215GA SA308105104GA SA308105105GA SA308105112GA SA308105146GA SA308105148GA SA308105149GA SA308105150GA SA308125362GA SA308135543GA SA308145659GA SA308145677GA SA308196504GA SA308206734GA SA308206737GA SA308206742GA SA308206743GA SA308398744GA SA308398768GA SA308449482GA SA308449492GA SA308449493GA SA308449497GA SA309132800GA SA309142916GA SA309153071GA SA309153073GA SA309153076GA SA309153080GA SA309153089GA SA309142965GA SA309142968GA SA309142982GA SA309304222GA SA309324434GA SA309334684GA SA309334745GA SA309344861GA SA309344866GA SA309354921GA SA309354940GA SA309334610GA SA309354927GA SA309395729GA SA309395780GA SA309426112GA SA309436186GA SA309436278GA SA309436285GA SA309507357GA SA309507368GA SA309507369GA SA309507372GA SA309507375GA SA309507399GA SA309507412GA SA309507414GA SA309507416GA SA309507419GA SA309507425GA SA309507433GA SA309507436GA SA309507437GA SA309507439GA SA309507443GA SA309507444GA SA309507447GA SA309507449GA SA309517450GA SA309517451GA SA309517455GA SA309517457GA SA310028121GA SA310028143GA SA310028313GA SA309517645GA SA309538019GA SA310028168GA SA310028172GA SA310028237GA SA310028242GA SA310028281GA SA310028299GA SA310028301GA SA310038356GA SA310038362GA SA310038366GA SA310038378GA SA310038417GA SA310038420GA SA310038421GA SA310038424GA SA310038464GA SA310038465GA SA310038467GA SA310038470GA SA310038475GA SA310048623GA SA310068869GA SA310068949GA SA309528006GA SA310028113GA SA310028119GA SA310038508GA SA310038523GA SA310048637GA SA310048653GA SA310048661GA SA310048662GA SA310048671GA SA310048674GA SA310048685GA SA310048688GA SA310058702GA SA310058706GA SA310119008GA SA310119010GA SA310119011GA SA310119012GA SA310119019GA SA310393176GA SA310260540GA SA310290637GA SA310423596GA SA310444256GA SA310455011GA SA310455022GA SA310475465GA SA310475531GA SA310475553GA SA310475555GA SA310475558GA SA310475562GA SA310475564GA SA310475578GA SA310475602GA SA310475616GA SA310475618GA SA310475620GA SA310475687GA SA310475700GA SA310475704GA SA310475705GA SA310475706GA SA310475714GA SA310475717GA SA310475718GA SA310485761GA SA310485763GA SA310485766GA SA310485770GA SA310485772GA SA310485791GA SA310485813GA SA310485820GA SA310485832GA SA310485839GA SA310485842GA SA310485845GA SA310485858GA SA310485862GA SA310485873GA SA310485880GA SA310496236GA SA310496315GA SA310496319GA SA310496325GA SA310496350GA SA310496369GA SA310496466GA SA310496475GA SA310496478GA SA310496486GA SA310496488GA SA310496493GA SA310506816GA SA310506842GA SA310507124GA SA310507213GA SA310507249GA SA310507252GA SA310507253GA SA310507339GA SA310507340GA SA310507347GA SA310507418GA SA310507424GA SA310507437GA SA310507459GA SA310507462GA SA310507470GA SA310507474GA SA310496389GA SA310496440GA SA310506755GA SA310506768GA SA310506774GA SA310506783GA SA310506799GA SA310506803GA SA310506807GA SA310507178GA SA310507195GA SA310507204GA SA310507222GA SA310507226GA SA310507233GA SA310507255GA SA310507257GA SA310507258GA SA310507280GA SA310507284GA SA310507289GA SA310507293GA SA310507294GA SA310507297GA SA310507299GA SA310507300GA SA310507306GA SA310507312GA SA310507315GA SA310507318GA SA310507321GA SA310507322GA SA310507323GA SA310507327GA SA310507331GA SA310507333GA SA310507334GA SA310507335GA SA310507342GA SA310507343GA SA310507345GA SA310507349GA SA310507352GA SA310507354GA SA310507355GA SA310507360GA SA310507361GA SA310507362GA SA310507363GA SA310507378GA SA310517599GA SA310517601GA SA310517611GA SA310517617GA SA310517623GA SA310517631GA SA310517639GA SA310517649GA SA310517675GA SA310517677GA SA310517694GA SA310518158GA SA310518165GA SA310518189GA SA310518206GA SA310518211GA SA310382840GA SA311039509GA SA311059583GA SA311069740GA SA311069757GA SA311080676GA SA311090914GA SA311090915GA SA311090916GA SA311090921GA SA311090922GA SA311090924GA SA311091036GA SA311091046GA SA311091052GA SA311090746GA SA311090749GA SA311121731GA SA311121732GA SA311121878GA SA311121881GA SA311122169GA SA311122173GA SA311122174GA SA311122177GA SA311122179GA SA311122187GA SA311122188GA SA311122190GA SA311122192GA
SA311122193GA SA311122196GA SA311122206GA SA311122207GA SA311122212GA SA311122221GA SA311122224GA SA311122226GA SA311122227GA SA311122235GA SA311122236GA SA311122238GA SA311122257GA SA311122258GA SA311122259GA SA311122262GA SA311122267GA SA311132277GA SA311132280GA SA311132290GA SA311132291GA SA311132295GA SA311132297GA SA311132318GA SA311132321GA SA311132330GA SA311132340GA SA311132348GA SA311132367GA SA311132399GA SA311132404GA SA311132406GA SA311122143GA SA311132423GA SA311132432GA SA311132436GA SA311132451GA SA311142914GA SA311142923GA SA311142928GA SA311142956GA SA311153081GA SA311153085GA SA311153087GA SA311153103GA SA311153107GA SA311153109GA SA311153122GA SA311153123GA SA311153130GA SA311153133GA SA311153137GA SA311153142GA SA311153160GA SA311153173GA SA311153182GA SA311153207GA SA311153214GA SA311153218GA SA311153220GA SA311153222GA SA311132380GA SA311142881GA SA311205454GA SA311184542GA SA311142888GA SA311215690GA SA311215695GA SA311215698GA SA311215715GA SA311215731GA SA311215740GA SA311307524GA SA311307538GA SA311307539GA SA311468376GA SA311468389GA SA311468406GA SA311468410GA SA311256808GA SA311256837GA SA311256858GA SA311246487GA SA310321043GA SA310321067GA SA310341414GA SA310341420GA SA310341457GA SA310341461GA SA310341463GA SA310341464GA SA310351527GA SA310351540GA SA310351544GA SA310351553GA SA310361609GA SA310361628GA SA310361629GA SA310361635GA SA310361642GA SA310361645GA SA310361661GA SA310361662GA SA310361663GA SA310361670GA SA310361675GA SA310361682GA SA310361687GA SA310361631GA SA310361656GA SA310361658GA SA310361664GA SA310361665GA SA310361676GA SA310361683GA SA310361685GA SA310361686GA SA312121825GA SA312232814GA SA312242993GA SA312242983GA SA312242985GA SA312243016GA SA312243043GA SA312152469GA SA312152487GA SA312253082GA SA312284177GA SA312284271GA SA312294441GA SA312294619GA SA312294569GA SA312355226GA SA312355229GA SA312355233GA SA312355247GA SA312355248GA SA312376199GA SA312376213GA SA312407293GA SA312407298GA SA312396805GA SA312397216GA SA312397251GA SA312079951GA SA312080392GA SA312080685GA SA312080702GA SA312080715GA SA312080730GA SA312080743GA SA312080757GA SA312080760GA SA312080723GA SA312080736GA SA312080739GA SA312080740GA SA312080741GA SA312080746GA SA312080750GA SA312080752GA SA312080755GA SA312080762GA SA312080780GA SA312080785GA SA312080788GA SA312080795GA SA312069645GA SA312069658GA SA312069681GA SA312069727GA SA312069735GA SA312069738GA SA312069741GA SA312069747GA SA312069748GA SA312069752GA SA312069756GA SA312069763GA SA312069774GA SA312091093GA SA312091101GA SA312069744GA SA312069771GA SA312091264GA SA307241671GA SA307241699GA SA307251787GA SA307251799GA SA307524463GA SA308044485GA SA308044508GA SA308044514GA SA308074609GA SA308084626GA SA308084628GA SA308084632GA SA308084633GA SA308084651GA SA308439447GA SA309385535GA SA309385564GA SA309405814GA SA309405815GA SA309405822GA SA309405823GA SA309405828GA SA309405846GA SA309405847GA SA309405855GA SA309446511GA SA309405816GA SA309446493GA SA309456547GA SA309466748GA SA309466773GA SA309466781GA SA309466783GA SA309517683GA SA310038522GA SA310331216GA SA310331235GA SA310331237GA SA310038429GA SA311184614GA SA312469059GA SA311317763GA SA311297241GA SA312028911GA SA312080562GA SA312080651GA SA312080774GA SA312080784GA SA312080799GA SA312080828GA SA312080838GA SA312080853GA SA312080857GA SA312080862GA SA312080863GA SA312080897GA SA312152482GA SA312152556GA SA307423166GA SA308044528GA SA308481003GA SA308481039GA SA308481040GA SA308481042GA SA308481062GA SA308481066GA SA308481069GA SA308489934GA SA308489938GA SA308489947GA SA308489949GA SA308489969GA SA309365320GA SA309405853GA SA309507424GA SA310159364GA SA310159366GA SA310159382GA SA310159459GA SA310199690GA SA310199712GA SA310199720GA SA312365777GA SA312365784GA SA312365799GA SA312365824GA SA310028250GA SA310028261GA SA310028297GA SA310028303GA SA310038383GA SA310038468GA SA310068864GA SA310068879GA SA311101295GA SA311111650GA SA311121762GA SA311297288GA SA311246420GA SA311246427GA SA311246432GA SA311246435GA SA311246439GA SA311246441GA SA311297274GA SA311307413GA SA311307414GA SA311307422GA SA311307426GA SA311307429GA SA311307430GA SA311307434GA SA311307435GA SA311307583GA SA311307584GA SA311307591GA SA311458214GA SA311458218GA SA311458219GA SA311458225GA SA311458231GA SA311458243GA SA311458256GA SA311458257GA SA311458262GA SA311458263GA SA311458343GA SA311458355GA SA311458356GA SA311458360GA SA311458364GA SA311458365GA SA311458368GA SA311518728GA SA311518729GA SA312263904GA SA312263905GA SA312263910GA SA312263921GA SA312263922GA SA312263925GA SA312263926GA SA312263933GA SA312263934GA SA312263937GA SA312263938GA SA312263940GA SA312263943GA SA312263944GA SA312263946GA SA312263948GA SA312263980GA SA312263981GA SA312263984GA SA312263985GA SA312263986GA SA312263987GA SA312263988GA SA312263989GA SA312263991GA SA312263992GA SA312263993GA SA312263994GA SA312263995GA SA312263996GA SA312263997GA SA312263998GA SA312263999GA SA312264001GA SA312264003GA SA312264004GA SA312264006GA SA312264008GA SA312091089GA SA312152489GA SA312152514GA SA312152541GA SA312152558GA SA312253325GA SA312294611GA SA312294380GA SA312294419GA SA312294453GA SA312294458GA SA312294532GA SA312294559GA SA312294561GA SA312294564GA SA312386379GA SA312386412GA SA312458258GA SA312458420GA SA312458425GA SA312458471GA SA312468548GA SA312468549GA SA312468557GA SA312111527GA SA312111660GA SA312121727GA SA312121736GA SA312080490GA SA312080634GA SA312080867GA SA312080868GA SA312080869GA SA312080883GA SA312080910GA SA312080546GA SA312080557GA SA312080578GA SA311488466GA SA311488467GA SA312069821GA SA312069859GA SA307080485GA SA307080537GA SA307221150GA SA307231195GA SA307231213GA SA307231256GA SA307231361GA SA307231407GA SA307221112GA SA307251793GA SA307291866GA SA307291868GA SA307393015GA SA307393038GA SA307443308GA SA307210823GA SA307241555GA SA307241559GA SA307241596GA SA307423125GA SA307443189GA SA307443219GA SA307443354GA SA307453443GA SA307453469GA SA307453472GA SA307453520GA SA307453530GA SA307453560GA SA307493886GA SA307493889GA SA307493968GA SA307493978GA SA307493979GA SA307493986GA SA307494036GA SA307494041GA SA307494046GA SA307504052GA SA307504053GA SA307504054GA SA307504055GA SA307504056GA SA307504057GA SA307504058GA SA307504059GA SA307504060GA SA307504061GA SA307504062GA SA307504063GA SA307504064GA SA307504065GA SA307504225GA SA307504226GA SA307504227GA SA307504228GA SA307504229GA SA307504230GA SA307504231GA SA307504234GA SA307514236GA SA307514237GA SA307514241GA SA307514242GA SA307514243GA SA307514244GA SA307514246GA SA307514249GA SA307514250GA SA307514251GA SA307514252GA SA307514253GA SA307514254GA SA307514255GA SA307514257GA SA307514258GA SA307514259GA SA307514260GA SA307514261GA SA307514262GA SA307514263GA SA307514264GA SA307514265GA SA307514266GA SA307514267GA SA307514268GA SA307514269GA SA307514270GA SA307514271GA SA307514272GA SA307514273GA SA307514274GA SA307514275GA SA307514278GA SA307514279GA SA307514280GA SA307514281GA SA307514282GA SA307514284GA SA307514285GA SA307514286GA SA307514287GA SA307514288GA SA307514292GA SA307514293GA SA307514294GA SA307514295GA SA307514296GA SA307514297GA SA307514300GA SA307514301GA SA307514305GA SA307514306GA SA307514307GA SA307514309GA SA307514310GA SA307514317GA SA307514362GA SA307483802GA SA307504103GA SA307504113GA SA307504116GA SA307504121GA SA307504187GA SA307504222GA SA307514413GA SA307524460GA SA307524461GA SA307524464GA SA307524465GA SA307524466GA SA307524467GA SA307524469GA SA307524471GA SA307524472GA SA308044486GA SA308044492GA SA308044493GA SA308044498GA SA308044499GA SA308044524GA SA308044531GA SA308044532GA SA307504101GA SA307514409GA SA307524462GA SA307524468GA SA307524476GA SA307524477GA SA307524479GA SA308044484GA SA308044487GA SA308044488GA SA308044489GA SA308044490GA SA308044494GA SA308044495GA SA308044497GA SA308044500GA SA308044502GA SA308044503GA SA308044504GA SA308044505GA SA308044506GA SA308044507GA SA308044509GA SA308044510GA SA308044512GA SA308044515GA SA308044516GA SA308044518GA SA308044519GA SA308044522GA SA308044526GA SA308044529GA SA308044530GA SA308044533GA SA308044534GA SA308074607GA SA308074581GA SA308074600GA SA308074601GA SA308084669GA SA308084672GA SA308084698GA SA308084709GA SA307301997GA SA308074608GA SA308084745GA SA308084748GA SA308084757GA SA308084771GA SA308094802GA SA308094812GA SA308094817GA SA308094821GA SA308094822GA SA308094823GA SA308094825GA SA308094826GA SA308094827GA SA308094828GA SA308094829GA SA308094830GA SA308094831GA SA308084614GA SA308084656GA SA308084730GA SA308084765GA SA308094833GA SA308094866GA SA308094881GA SA308094923GA SA308094926GA SA308094932GA SA308094945GA SA308094956GA SA308094961GA SA308094962GA SA308094963GA SA308094969GA SA308094971GA SA308094972GA SA308094974GA SA308094975GA SA308094976GA SA308094977GA SA308094982GA SA308094983GA SA308094985GA SA308094986GA SA308094987GA SA308094988GA SA308094990GA SA308094992GA SA308094993GA SA308094994GA SA308094995GA SA308094996GA SA308094997GA SA308095001GA SA308095003GA SA308095004GA SA308095005GA SA308095006GA SA308095007GA SA308095008GA SA308095009GA SA308095010GA SA308095011GA SA308095012GA SA308095045GA SA308095046GA SA308095049GA SA308095050GA SA308095052GA SA308105216GA SA308105218GA SA308105219GA SA308105221GA SA308105222GA SA308105225GA SA308105226GA SA308115231GA SA308115232GA SA308115233GA SA308115241GA SA308115243GA SA308115244GA SA308115246GA SA308115247GA SA308115249GA SA308115251GA SA308115253GA SA308044491GA SA308084785GA SA308094798GA SA308094799GA SA308094824GA SA308094835GA SA308094924GA SA308094928GA SA308094936GA SA308095053GA SA308105151GA SA308105152GA SA308105153GA SA308105154GA SA308105155GA SA308105157GA SA308105158GA SA308105159GA SA308105160GA SA308105161GA SA308105162GA SA308105163GA SA308105164GA SA308105166GA SA308105167GA SA308105168GA SA308105169GA SA308105170GA SA308105171GA SA308105172GA SA308105174GA SA308105175GA SA308105176GA SA308105177GA SA308105178GA SA308105179GA SA308105180GA SA308105181GA SA308105186GA SA308105187GA SA308105188GA SA308105189GA SA308105190GA SA308105194GA SA308105195GA SA308105197GA SA308105199GA SA308105201GA SA308105202GA SA308105204GA SA308105205GA SA308105207GA SA308105209GA SA308105211GA SA308105212GA SA308105213GA SA308105230GA SA308155855GA SA308155895GA SA308155844GA SA308155861GA SA308155862GA SA308155870GA SA308155910GA SA308155927GA SA308155976GA SA308166038GA SA308166087GA SA308166089GA SA308166093GA SA308166099GA SA308166104GA SA308166105GA SA308166106GA SA308105210GA SA308105227GA SA308105228GA SA308105229GA SA308115236GA SA308115238GA SA308115239GA SA308115240GA SA308115245GA SA308115254GA SA308115259GA SA308115260GA SA308125316GA SA308125317GA SA308125318GA SA308125319GA SA308155856GA SA308155890GA SA308155891GA SA308155894GA SA308155896GA SA308155902GA SA308155907GA SA308155911GA SA308155912GA SA308155913GA SA308155922GA SA308155930GA SA308155934GA SA308155952GA SA308165992GA SA308196476GA SA308125383GA SA308145751GA SA308155889GA SA308155926GA SA308166065GA SA308166117GA SA308186339GA SA308266782GA SA308266783GA SA308266785GA SA308266786GA SA308266787GA SA308266790GA SA308266791GA SA308266792GA SA308266793GA SA308266794GA SA308266795GA SA308266796GA SA308266797GA
SA308266799GA SA308266800GA SA308266801GA SA308266802GA SA308266803GA SA308266804GA SA308266805GA SA308266806GA SA308266807GA SA308266808GA SA308266809GA SA308266810GA SA308266812GA SA308196524GA SA308206648GA SA308206665GA SA308266883GAMKE SA308337686GA SA308347754GA SA308347756GA SA308347829GA SA308347844GA SA308347845GAMKE SA308347846GA SA308347847GAMKE SA308347848GAMKE SA308347850GA SA308347851GAMKE SA308357943GAMKE SA308357944GA SA308357945GAMKE SA308357946GA SA308337704GA SA308347767GA SA308347789GA SA308347791GA SA308347800GA SA308206723GA SA308368070GA SA308388627GA SA308398712GA SA308409085GA SA308429347GA SA308378286GA SA308409161GA SA308409173GA SA308409189GA SA308419237GA SA308419250GA SA308419267GA SA308419272GA SA308419274GA SA307393008GA SA308459849GA SA308459857GA SA308459874GA SA308491229GA SA308491152GA SA308491153GA SA308491154GA SA308491155GA SA308491156GA SA308491157GA SA308491158GA SA308491160GA SA308491162GA SA308491163GA SA308511770GA SA308521816GA SA309021975GA SA309021987GA SA309032009GA SA309032012GA SA309032013GA SA309032016GA SA309032017GA SA309032021GA SA309032053GA SA309032054GA SA309032055GA SA309032056GA SA309032057GA SA309032058GA SA309032059GA SA309032060GA SA309032061GA SA309032062GA SA309032063GA SA309032065GA SA309032067GA SA309032068GA SA309032069GA SA309042132GA SA309052223GA SA309112570GA SA309122693GA SA309132865GA SA309142895GA SA309142901GA SA309142906GA SA309132803GA SA309142877GA SA309142907GA SA309132854GA SA309142989GA SA309153034GA SA309153038GA SA309153046GA SA309153048GA SA309153053GA SA308378297GA SA309193332GA SA309193333GA SA309193335GA SA309193338GA SA309193342GA SA309193344GA SA309193345GA SA309193347GA SA309193350GA SA309193355GA SA309193359GA SA309193361GA SA309203367GA SA309203373GA SA309203374GA SA309203377GA SA309203381GA SA309203385GA SA309203386GA SA309203389GA SA309203397GA SA309203398GA SA309203399GA SA309203408GA SA309203411GA SA309203413GA SA309203414GA SA309203415GA SA309203419GA SA309203422GA SA309203426GA SA309203431GA SA309223436GA SA309223437GA SA309223438GA SA309223440GA SA309223441GA SA309223446GA SA309223447GA SA309223448GA SA309223451GA SA309223452GA SA309223454GA SA309223459GA SA309223460GA SA309223461GA SA309223463GA SA309223464GA SA309223467GA SA309223473GA SA309223474GA SA309223475GA SA309223483GA SA309223486GA SA309223487GA SA309223488GA SA309223490GA SA309223491GA SA309223492GA SA309223493GA SA309223494GA SA309223495GA SA309223496GA SA309223497GA SA309223498GA SA309223499GA SA309223500GA SA309223501GA SA309223502GA SA309223504GA SA309223505GA SA309223506GA SA309223509GA SA309223510GA SA309223511GA SA309223512GA SA309223513GA SA309233519GA SA309233521GA SA309233532GA SA309233535GA SA309233537GA SA309233538GA SA309233540GA SA309233541GA SA309233542GA SA309233543GA SA309233545GA SA309233546GA SA309233548GA SA309233549GA SA309233551GA SA309243683GA SA309243699GA SA309243702GA SA309243705GA SA309243707GA SA309243711GA SA309243712GA SA309253721GA SA309253727GA SA309253745GA SA309253746GA SA309253754GA SA309365109GA SA309365115GA SA309365117GA SA309365164GA SA309365172GA SA309365177GA SA309365191GA SA309365192GA SA309365194GA SA309365196GA SA309365201GA SA309365202GA SA309365222GA SA309375503GA SA309446515GA SA309446528GA SA309456718GA SA309456720GA SA309456731GA SA309466735GA SA309466739GA SA309466746GA SA309466754GA SA309466759GA SA309466769GA SA309466780GA SA309466795GA SA309466797GA SA309466799GA SA309426155GA SA309446392GA SA309476861GA SA309476866GA SA309476867GA SA309476869GA SA309476870GA SA309476881GA SA309476883GA SA309476884GA SA309476887GA SA309476888GA SA309476889GA SA309476890GA SA309476891GA SA309476898GA SA309476993GA SA309476995GA SA309476996GA SA309477003GA SA309477008GA SA309477032GA SA309477136GA SA309365080GA SA309405819GA SA309436327GA SA309436331GA SA309446437GA SA309446482GA SA309446516GA SA309476950GA SA309517710GA SA309527915GA SA309527931GA SA309527938GA SA310038354GA SA310038372GA SA310038377GA SA310038389GA SA310038406GA SA310038410GA SA310038413GA SA310038414GA SA310038415GA SA310038419GA SA310038425GA SA310038446GA SA310038457GA SA310038459GA SA310058741GA SA310058747GA SA310068820GA SA310068828GA SA309538038GA SA310028193GA SA310028200GA SA310028219GA SA310028224GA SA310028271GA SA310028276GA SA310038426GA SA310038435GA SA310038438GA SA310038439GA SA310038443GA SA310038453GA SA310038472GA SA310038473GA SA310038474GA SA310048584GA SA310048585GA SA310048588GA SA310048589GA SA310048590GA SA310048593GA SA310048596GA SA310048597GA SA310048622GA SA310260413GA SA310260416GA SA310260364GA SA310392930GA SA310393112GA SA310393136GA SA310393263GA SA310393278GA SA310393293GA SA310393300GA SA310393308GA SA310392893GA SA310392901GA SA310392907GA SA310392909GA SA310392917GA SA310393089GA SA310393096GA SA310393185GA SA310393287GA SA310393297GA SA310393298GA SA310393302GA SA310393305GA SA310424042GA SA310413505GA SA310423576GA SA310423702GA SA310423829GA SA310423834GA SA310423952GA SA310423970GA SA310444506GA SA310444510GA SA310454997GA SA310455078GA SA310465371GA SA310465376GA SA310465379GA SA310444846GA SA310496159GA SA310496326GA SA310496330GA SA310496341GA SA310496358GA SA310496362GA SA310496363GA SA310496372GA SA310496454GA SA310496461GA SA310496479GA SA310496481GA SA310496485GA SA310496496GA SA310496500GA SA310496501GA SA310496509GA SA310496511GA SA310496518GA SA310496519GA SA310496520GA SA310496522GA SA310496524GA SA310496649GA SA310506849GA SA310506852GA SA310506867GA SA310507041GA SA310507047GA SA310507076GA SA310507115GA SA310434145GA SA310496617GA SA310496636GA SA310507061GA SA310507513GA SA310507514GA SA310517718GA SA310506776GA SA310506833GA SA310506839GA SA310507165GA SA310507167GA SA310507211GA SA310507214GA SA310507251GA SA310507261GA SA310507264GA SA310507265GA SA310507269GA SA310507273GA SA310507274GA SA310507275GA SA310507277GA SA310507282GA SA310507285GA SA310507341GA SA310507350GA SA310507358GA SA310507359GA SA310507364GA SA310507365GA SA310507369GA SA310517598GA SA310517603GA SA310517609GA SA310517615GA SA310517620GA SA310517628GA SA310517630GA SA310517655GA SA310518151GA SA310518155GA SA310518183GA SA310518188GA SA310518199GA SA310518207GA SA310506767GA SA310506798GA SA310506802GA SA310507147GA SA310507150GA SA310507192GA SA310507207GA SA310496321GA SA310507542GA SA310517561GA SA310517571GA SA310517579GA SA310517580GA SA310517582GA SA310517584GA SA310517586GA SA310517716GA SA310517748GA SA310517749GA SA310517750GA SA310517779GA SA310517816GA SA310517824GA SA310506793GA SA310506815GA SA310506823GA SA310506825GA SA310506829GA SA310507172GA SA310507173GA SA310507191GA SA310507209GA SA310507210GA SA310507229GA SA310507234GA SA310507245GA SA310507283GA SA310517647GA SA310517658GA SA310518173GA SA310518239GA SA310413522GA SA310496127GA SA310496309GA SA310528966GA SA310528988GA SA310528993GA SA311039481GA SA311039489GA SA311039474GA SA310260005GA SA310361657GA SA310423851GA SA310539350GA SA311039435GA SA311039455GA SA311039465GA SA310321023GA SA310382456GA SA310413479GA SA310413481GA SA310413518GA SA310423658GA SA310475726GA SA310507054GA SA310507394GA SA310528860GA SA310528876GA SA310528882GA SA310528964GA SA310528974GA SA310528991GA SA310528996GA SA311039371GA SA310423831GA SA310444311GA SA310455166GA SA310475617GA SA311039404GA SA310382846GA SA311060014GA SA311060045GA SA311069673GA SA310496143GA SA310496218GA SA310496278GA SA310496282GA SA310496285GA SA310496291GA SA311060126GA SA311060154GA SA311060157GA SA311060163GA SA311060198GA SA311069723GA SA311060172GA SA311070536GA SA311060148GA SA310393285GA SA311059586GA SA311059594GA SA311060033GA SA311069626GA SA311069628GA SA311069629GA SA311069631GA SA311069632GA SA311069633GA SA311069634GA SA311069635GA SA311069642GA SA311069643GA SA311069647GA SA311069664GA SA311069686GA SA311069698GA SA311069834GA SA311080672GA SA311080673GA SA311080675GA SA311080681GA SA311080686GA SA311080691GA SA311080693GA SA311080696GA SA311080697GA SA311060010GA SA311101220GA SA311101274GA SA311101300GA SA311101272GA SA311101279GA SA311101424GA SA311090938GA SA311101098GA SA311090792GA SA311122141GA SA311121850GA SA311121854GA SA311121863GA SA311122018GA SA311122031GA SA311122046GA SA311122054GA SA311122229GA SA311122230GA SA311122248GA SA311122251GA SA311122270GA SA311122275GA SA311132287GA SA311132313GA SA311132314GA SA311132322GA SA311132326GA SA311132352GA SA311132398GA SA311132556GA SA311132588GA SA311142750GA SA311142753GA SA311142755GA SA311142761GA SA311142762GA SA311142766GA SA311142770GA SA311142771GA SA311142774GA SA311142776GA SA311142777GA SA311142778GA SA311142794GA SA311142796GA SA311090934GA SA311090998GA SA311091007GA SA311101403GA SA311121737GA SA311121779GA SA311121849GA SA311121853GA SA311121855GA SA311121857GA SA311121862GA SA311121870GA SA311121879GA SA311121894GA SA311121912GA SA311122013GA SA311122015GA SA311122021GA SA311122026GA SA311122028GA SA311122029GA SA311122044GA SA311122047GA SA311122052GA SA311122053GA SA311122055GA SA311122058GA SA311132593GA SA311132598GA SA311142757GA SA311142773GA SA311111614GA SA311121852GA SA311121859GA SA311121860GA SA311121864GA SA311121865GA SA311121866GA SA311121871GA SA311121873GA SA311121874GA SA311121875GA SA311121882GA SA311121883GA SA311121884GA SA311121886GA SA311121887GA SA311121893GA SA311121896GA SA311121898GA SA311121915GA SA311121916GA SA311132554GA SA311142947GA SA311132377GA SA311132387GA SA311121943GA SA311121975GA SA311122007GA SA311122076GA SA311122087GA SA311122089GA SA311122129GA SA311132371GA SA311132518GA SA311142724GA SA311153046GA SA311153067GA SA311153147GA SA311153185GA SA311142918GA SA311153141GA SA311153202GA SA311163944GA SA311163383GA SA311163404GA SA311163856GA SA311163865GA SA311163869GA SA311163888GA SA311163893GA SA311163947GA SA311163951GA SA311163955GA SA311163962GA SA311163980GA SA311163890GA SA311163914GA SA311195052GA SA311195062GA SA311195074GA SA311195076GA SA311195085GA SA311195087GA SA311163930GA SA311174161GA SA311174170GA SA311184363GA SA311184373GA SA311184376GA SA311184379GA SA311184383GA SA311174134GA SA311184457GA SA311184470GA SA311184476GA SA311184486GA SA311184488GA SA311184490GA SA311184501GA SA311184508GA SA311184510GA SA311163809GA SA311174045GA SA311174127GA SA311184473GA SA311184482GA SA311184512GA SA311195007GA SA311195042GA SA311195043GA SA311195047GA SA311195048GA SA311195049GA SA311195051GA SA311195170GA SA311195171GA SA311195172GA SA311246519GA SA311246532GA SA311246535GA SA311246536GA SA311246538GA SA311246539GA SA311246540GA SA311246548GA SA311246550GA SA311246554GA SA311246555GA SA311246556GA SA311246557GA SA311246563GA SA311246579GA SA311153033GA SA311153077GA SA311153258GA SA311163438GA SA311163531GA SA311163535GA SA311163550GA SA311163566GA SA311163568GA SA311163588GA SA311163597GA SA311163699GA SA311184464GA SA311184471GA SA311184511GA SA311163450GA SA311163452GA SA311184521GA SA311153260GA SA311153348GA SA311163610GA SA311163835GA SA311174056GA SA311163648GA SA311163792GA SA311163794GA SA311163814GA SA311163821GA SA311163833GA SA311163854GA SA311174047GA SA311174065GA SA311184722GA SA311184730GA SA311184735GA SA311184736GA SA311184739GA SA311184751GA SA311184752GA SA311184762GA SA311184764GA SA311184768GA SA311184769GA SA311184770GA SA311184779GA SA311184788GA SA311184790GA SA311184791GA SA311184792GA SA311184797GA SA311184805GA SA311184808GA SA311184869GA SA311184871GA SA311205313GA SA311205317GA SA311205322GA SA311205331GA SA311205393GA SA311205404GA SA311205405GA SA311205406GA SA311205415GA SA311205417GA SA311205421GA SA311205422GA SA311205428GA SA311205430GA SA311205431GA SA311205435GA SA311205438GA SA311256811GA SA311256846GA SA311256851GA SA311256863GA SA311163805GA SA311184719GA SA311184725GA SA311184729GA SA311184732GA SA311184753GA SA311184776GA SA311184798GA SA311184799GA SA311132546GA SA311153307GA SA311174058GA SA311184690GA SA311184721GA SA311184737GA SA311184763GA SA311184816GA SA311184825GA SA311194962GA SA311194987GA SA311205328GA SA311205339GA SA311205403GA SA311205407GA SA311205413GA SA311205419GA SA311205433GA SA311205570GA SA311205586GA SA311205594GA SA311205595GA SA311205596GA SA311205598GA SA311205599GA SA311205600GA SA311205603GA SA311205612GA SA311205620GA SA311225897GA SA311307452GA SA311317650GA SA311039373GA SA311060232GA SA311060274GA SA311266877GA SA311266885GA SA311266887GA SA311266888GA SA311266904GA SA311266905GA SA311266908GA SA311060274GA SA311205311GA SA311225928GA SA311225929GA SA311225930GA SA311225951GA SA311225957GA SA311236041GA SA311236044GA SA311225923GA SA311225924GA SA311225927GA SA311225933GA SA311225944GA SA311225947GA SA311225955GA SA311225960GA SA311225970GA SA311236054GA SA311287014GA SA311287041GA SA311297272GA SA311307415GA SA311307417GA SA311307432GA SA311307532GA SA311236373GA SA311246451GA SA311246455GA SA311246458GA SA311246464GA SA311246475GA SA311246505GA SA311246508GA SA311246590GA SA311246667GA SA311256722GA SA311256724GA SA311256735GA SA311256736GA SA311256741GA SA311256820GA SA311256824GA SA311256853GA SA311297307GA SA311297310GA SA311184838GA SA311236217GA SA311236353GA SA311236354GA SA311236370GA SA311236375GA SA311236378GA SA311236381GA SA311246491GA SA311518726GA SA310310877GA SA310310886GA SA310310896GA SA310310903GA SA310310914GA SA310310917GA SA310260013GA SA310310890GA SA310310915GA SA310310930GA SA310310933GA SA310310944GA SA310310951GA SA310310952GA SA310310953GA SA310310954GA SA310310957GA SA310310961GA SA310310964GA SA310310987GA SA310331250GA SA310331267GA SA310331268GA SA310331312GA SA310331314GA SA310331315GA SA310331319GA SA310331320GA SA310331321GA SA310331325GA SA310361879GA SA310372075GA SA310372095GA SA310372100GA SA310372173GA SA310331226GA SA310331257GA SA310331260GA SA310331317GA SA310331355GA SA310341383GA SA310361836GA SA310372080GA SA310372110GA SA310372115GA SA310372117GA SA310372120GA SA310372130GA SA310372136GA SA310372174GA SA310372176GA SA310372197GA SA310372198GA SA310372201GA SA310372207GA SA310382344GA SA310382360GA SA310382539GA SA310382692GA SA310382749GA SA312253140GA SA312253145GA SA312253150GA SA312253157GA SA312253163GA SA312253318GA SA312253364GA SA312253454GA SA312263911GA SA312284300GA SA312294474GA SA312253298GA SA312294435GA SA312294479GA SA312294528GA SA312294774GA SA312232893GA SA312253264GA SA312232865GA SA312232903GA SA312263632GA SA312263677GA SA312263819GA SA312284230GA SA312111520GA SA312152494GA SA312253294GA SA312253363GA SA312232909GA SA312253203GA SA312253210GA SA312253440GA SA312253476GA SA312253591GA SA312263639GA SA312263668GA SA312263693GA SA312263763GA SA312263796GA SA312263815GA SA312264036GA SA312284117GA SA312344896GA SA312344910GA SA312344916GA SA312344949GA SA312344950GA SA312344978GA SA312345019GA SA312345021GA SA312345025GA SA312345027GA SA312345032GA SA312345035GA SA312345096GA SA312345101GA SA312345161GA SA312345179GA SA312344873GA SA312344923GA SA312344930GA SA312344937GA SA312344943GA SA312344961GA SA312344975GA SA312345023GA SA312345024GA SA312345029GA SA312365613GA SA312365622GA SA312365623GA SA312365631GA SA312365635GA SA312365639GA SA312365640GA SA312365644GA SA312365668GA SA312365689GA SA312365732GA SA312365834GA SA312365843GA SA312375967GA SA312375969GA SA312375970GA SA312375971GA SA312375974GA SA312375976GA SA312375977GA SA312375978GA SA312375980GA SA312375986GA SA312376006GA SA312376023GA SA312376104GA SA312376108GA SA312376111GA SA312376112GA SA312376113GA SA312376114GA SA312376115GA SA312376117GA SA312376118GA SA312376121GA SA312376124GA SA312378023GA SA312378041GA SA312378044GA SA312378045GA SA312378054GA SA312378065GA SA312376107GA SA312376238GA SA312378039GA SA312376105GA SA312365554GA SA312365854GA SA312376127GA SA312376166GA SA312376168GA SA312376191GA SA312376227GA SA312376293GA SA312376299GA SA312376303GA SA312376307GA SA312376310GA SA312378050GA SA312386313GA SA312386318GA SA312386319GA SA312386321GA SA312386324GA SA312386325GA SA312386327GA SA312386328GA SA312386336GA SA312386340GA SA312386345GA SA312386348GA SA312386349GA SA312386351GA SA312386352GA SA312386354GA SA312386355GA SA312386357GA SA312386358GA SA312386322GA SA312386341GA SA312386344GA SA312386371GA SA312386375GA SA312386392GA SA312386395GA SA312386398GA SA312386483GA SA312386501GA SA312386517GA SA312386529GA SA312386552GA SA312386562GA SA312386567GA SA312386580GA SA312396856GA SA312396967GA SA312397037GA SA312397040GA SA312397043GA SA312397048GA SA312397053GA SA312397057GA SA312397058GA SA312397059GA SA312397091GA SA312397094GA SA312397095GA SA312397096GA SA312397097GA SA312397099GA SA312397102GA SA312397105GA SA312397107GA SA312397109GA SA312397110GA SA312396775GA SA312396781GA SA312396853GA SA312396873GA SA312396884GA SA312396888GA SA312396897GA SA312396908GA SA312396943GA SA312396944GA SA312396946GA SA312396949GA SA312396953GA SA312396954GA SA312396955GA SA312396957GA SA312396960GA SA312396962GA SA312396964GA SA312396968GA SA312417323GA SA312417344GA SA312417346GA SA312417349GA SA312417362GA SA312417364GA SA312417365GA SA312417366GA SA312417374GA SA312417382GA SA312417386GA SA312397257GA SA312417306GA SA312417311GA SA312417316GA SA312417317GA SA312417318GA SA312417328GA SA312417329GA SA312417332GA SA312468598GA SA312468620GA SA312468658GA SA312468660GA SA312468669GA SA312469262GA SA312469414GA SA312469441GA SA312479531GA SA312479535GA SA312479549GA SA312479610GA SA312479617GA SA312479621GA SA312479626GA SA312479627GA SA312479649GA SA312468659GA SA312468661GA SA312468703GA SA312479544GA SA312479572GA SA312479573GA SA312479803GA SA312479804GA SA312479814GA SA312479819GA SA312479825GA SA312479837GA SA312479838GA SA312479839GA SA312479842GA SA312479845GA SA312479850GA SA312469400GA SA312469401GA SA312469403GA SA312469405GA SA312469407GA SA312469408GA SA312469410GA SA312469412GA SA312469423GA SA312469427GA SA312469430GA SA312469493GA SA312469494GA SA312469496GA SA312469497GA SA312469498GA SA312469503GA SA312469516GA SA312469517GA SA312479528GA SA312479529GA SA312479530GA SA312479533GA SA312479534GA SA312479536GA SA312479539GA SA312479541GA SA312480198GA SA312480232GA SA312480242GA SA312480244GA SA312480266GA SA312479695GA SA312480097GA SA312468720GA SA312469432GA SA312480124GA SA312480125GA SA312480126GA SA312480129GA SA312480130GA SA312480132GA SA312480135GA SA312480136GA SA312480138GA SA312480143GA SA312480148GA SA312480150GA SA312480151GA SA312480153GA SA312480155GA SA312480156GA SA312480157GA SA312480158GA SA312480159GA SA312480163GA SA312480164GA SA312480166GA SA312480168GA SA312480169GA SA312480170GA SA312480171GA SA312480172GA SA312480173GA SA312480174GA SA312480179GA SA312480180GA SA312480181GA SA312480182GA SA312480186GA SA312480188GA SA312480195GA SA312480196GA SA312480210GA SA312480212GA SA312480214GA SA312480225GA SA312480230GA SA312480265GA SA312480461GA SA312480475GA SA312480506GA SA312480508GA SA312480564GA SA312480613GA SA312480697GA SA312479581GA SA312480428GA SA312480429GA SA312480433GA SA312480435GA SA312480444GA SA312480521GA SA312480527GA SA312480529GA SA312480534GA SA312480543GA SA312480545GA SA312480549GA SA312480550GA SA312480556GA SA312480557GA SA312480567GA SA312480569GA SA312480574GA SA312480580GA SA312480584GA SA312480586GA SA312480588GA SA312480589GA SA312480590GA SA312480592GA SA312480593GA SA312480594GA SA312480599GA SA312480605GA SA312480606GA SA312480608GA SA312480611GA SA312480617GA SA312480621GA SA312480623GA SA312480625GA SA312480626GA SA312480627GA SA312480634GA SA312480635GA SA312480636GA SA312480637GA SA312480639GA SA312480640GA SA312480641GA SA312480643GA SA312480647GA SA312480652GA SA312480659GA SA312480692GA SA312480696GA SA312480698GA SA312480699GA SA312480701GA SA312480702GA SA312480706GA SA312480707GA SA312480710GA SA312491246GA SA312491342GA SA312018906GA SA311195235GA SA311195248GA SA311297297GA SA312028949GA SA312028967GA SA312028990GA SA312028996GA SA312029015GA SA312029053GA SA312029068GA SA312029072GA SA311468412GA SA311468426GA SA311468439GA SA311194931GA SA311194973GA SA311205576GA SA311205583GA SA311205606GA SA311307490GA SA311458307GA SA311458308GA SA311458309GA SA311458311GA SA311458312GA SA311458313GA SA311458314GA SA311458316GA SA311458319GA SA311458331GA SA311458334GA SA311468413GA SA311468416GA SA311468418GA SA311468421GA SA311468422GA SA311468425GA SA311468427GA SA311468429GA SA311468431GA SA311468436GA SA311468437GA SA311468440GA SA311468441GA SA311468442GA SA311468444GA SA311468449GA SA311488463GA SA312039115GA SA312039188GA SA312049422GA SA312049424GA SA312049444GA SA312069824GA SA312069835GA SA312079889GA SA312079893GA SA312079896GA SA312079898GA SA312079901GA SA312079904GA SA312079905GA SA312079906GA SA312079909GA SA312079910GA SA312079911GA SA312079925GA SA312069827GA SA312049323GA SA312049325GA SA312049333GA SA312049334GA SA312049345GA SA312049357GA SA312049366GA SA312059478GA SA312049458GA SA312059476GA SA312059629GA SA312059630GA SA312079949GA SA312080468GA SA312080470GA SA312080475GA SA312080518GA SA312080519GA SA312080527GA SA312080480GA SA312080643GA SA312080699GA SA312080786GA SA312080789GA SA312080825GA SA312080841GA SA312080850GA SA312080854GA SA312080889GA SA312080895GA SA312080906GA SA312080907GA SA312080887GA SA312080861GA SA312080902GA SA312080904GA SA312080908GA SA312070268GA SA312091049GA SA312070078GA SA312091029GA SA312091047GA SA312091065GA SA312069819GA SA312069788GA SA312069791GA SA312069800GA SA312069830GA SA312069840GA SA312069845GA SA312069854GA SA312091138GA SA312091140GA SA312091150GA SA312111677GA SA312121744GA SA312091270GA SA312091281GA SA312091290GAMKE SA312091300GA SA312091307GA SA312091308GA SA312091312GA SA312091315GA SA312091316GA SA312091320GA SA312091323GA SA312091325GA SA312091326GA SA312091329GA SA312091331GA SA312091334GA SA312091338GA SA312091339GA SA312091341GA SA312091342GA SA312091343GA SA312091346GA SA312091348GA SA312091356GA SA312091358GA SA312091359GA SA312091360GA SA312091362GA SA312091364GA SA312121999GA SA312132108GA SA311488466GA SA311488467GA SA312069705GA SA312091318GA SA312091351GA SA312101367GA SA312101368GA SA312101371GA SA312101373GA SA312101381GA SA312101382GA SA312101383GA SA312101384GA SA312101385GA SA312101387GA SA312101388GA SA312101391GA SA312101392GA SA312101393GA SA312101394GA SA312069821GA SA312069859GA SA312121942GA SA312121927GA SA312121947GA SA312121995GA SA312122029GA SA312122046GA SA312122000GA SA312122005GA SA311236189GA SA311256854GA SA311468430GA SA311468435GA SA311468447GA SA311488453GA SA312049367GA SA312049375GA SA312049376GA SA312049379GA SA312049381GA SA312049384GA SA312049413GA SA312049420GA SA312079885GA SA312079886GA SA312079913GA SA312079931GA SA312079947GA SA312079976GA SA312079981GA SA312079984GA SA312101429GA SA312111500GA SA312111505GA SA308449588GA SA308449594GA SA308459640GA SA308459647GA SA308459671GA SA308459677GA SA308459684GA SA308459859GA SA308469887GA SA308469889GA SA308479908GA SA308481000GA SA308481004GA SA308481019GA SA308481025GA SA308481045GA SA308481047GA SA308481049GA SA308481051GA SA308489920GA SA308489925GA SA308489926GA SA308489929GA SA308489930GA SA308489931GA SA308489939GA SA308489940GA SA308489944GA SA308489945GA SA308489952GA SA308489953GA SA308489955GA SA308489956GA SA308489966GA SA308489968GA SA308489970GA SA308489972GA SA308489977GA SA308489978GA SA308489984GA SA308489986GA SA308489987GA SA308489988GA SA308489989GA SA308489992GA SA308489996GA SA308489999GA SA310270583GA SA310290741GA SA310290743GA SA310290744GA SA310444286GA SA310506715GA SA310507112GA SA310507403GA SA310507453GA SA310507457GA SA310507520GA SA310507544GA SA311090964GA SA311101088GA SA311101090GA SA311101135GA SA311101139GA SA311101157GA SA311101457GA SA311101466GA SA311111620GA SA311111634GA SA311111635GA SA311111639GA SA311111643GA SA311111644GA SA311111645GA SA311111646GA SA311111647GA SA311111648GA SA311111651GA SA311111652GA SA311111653GA SA311111656GA SA311111658GA SA311111660GA SA311111664GA SA311121980GA SA311122068GA SA311122070GA SA311122075GA SA311122078GA SA311122081GA SA311122083GA SA311122090GA SA311122091GA SA311122092GA SA311122097GA SA311122110GA SA311122121GA SA311122122GA SA311122137GA SA311122139GA SA311122144GA SA311122146GA SA311122151GA SA311122152GA SA311122157GA SA311122158GA SA311122162GA SA311122165GA SA311132428GA SA311132493GA SA311132538GA SA311132542GA SA311132544GA SA311132601GA SA311132602GA SA311132608GA SA311132609GA SA311132611GA SA311132621GA SA311132622GA SA311132624GA SA311132625GA SA311132626GA SA311132627GA SA311132628GA SA311132629GA SA311132632GA SA311132633GA SA311132637GA SA311132638GA SA311132640GA SA311132642GA SA311132643GA SA311132644GA SA311132645GA SA311132647GA SA311132654GA SA311132656GA SA311132657GA SA311132661GA SA311132664GA SA311132665GA SA311132670GA SA311132671GA SA311132672GA SA311132673GA SA311142687GA SA311142699GA SA311142843GA SA311163402GA SA311163861GA SA311163864GA SA311163866GA SA311163875GA SA311163879GA SA311163895GA SA311163910GA SA311205293GA SA311225953GA SA311236006GA SA311236007GA SA311236008GA SA311236014GA SA311236020GA SA311236021GA SA311236024GA SA311236042GA SA311236046GA SA311236052GA SA311236056GA SA311236165GA SA311236188GA SA311236200GA SA311236202GA SA311236204GA SA311236205GA SA311236208GA SA311236213GA SA311236214GA SA311236220GA SA311236224GA SA311236226GA SA311236230GA SA311236232GA SA311236253GA SA311236255GA SA311236258GA SA311236280GA SA311236364GA SA311246449GA SA311246476GA SA311246478GA SA311246495GA SA311246498GA SA311246501GA SA311256802GA SA311256806GA SA311256813GA SA311256849GA SA311256861GA SA311256864GA SA311256867GA SA311266913GA SA311266916GA SA311266917GA SA311266918GA SA311266920GA SA311266921GA SA311266924GA SA311266926GA SA311266927GA SA311266929GA SA311266930GA SA311266931GA SA311266932GA SA311266933GA SA311266934GA SA311266936GA SA311266937GA SA311266939GA SA311266940GA SA311266943GA SA311266944GA SA311266945GA SA311266948GA SA311266949GA SA311266950GA SA311307576GA SA311327915GA SA311184866GA SA312192588GA SA312192597GA SA312192600GA SA312192723GA SA312253197GA SA312264039GA SA312470071GA SA312470082GA SA312470083GA SA312470089GA SA312480103GA SA312480104GA SA312490969GA SA312490971GA SA312491012GA SA312491048GA SA312018864GA SA309436321GA SA309436329GA SA309436363GA SA309446428GA SA309446438GA SA309446443GA SA309446485GA SA309446522GA SA309466767GA SA311195061GA SA311195068GA SA311236065GA SA311256798GA SA308266823GA SA308266828GA SA308266889GA SA308266891GA SA308286904GA SA308286908GA SA308286917GA SA308286926GA SA308286927GA SA308286934GA SA308286946GA SA308286947GA SA308286952GA SA308286956GA SA308286957GA SA308286991GA SA308286992GA SA308286997GA SA308286998GA SA308286999GA SA308287002GA SA308287014GA SA308287015GA SA308327376GA SA309263970GA SA310118994GA SA310139214GA SA310139217GA SA310139220GA SA310139223GA SA310139224GA SA310139228GA SA310139233GA SA310139235GA SA310139289GA SA310139307GA SA310149343GA SA310423750GA SA311069707GA SA311132540GA SA311195199GA SA311428072GA SA311132299GA SA312294373GA SA312294415GA SA312294468GA SA312294469GA SA312294516GA SA312294526GA SA312294531GA SA312294558GA SA312365776GA SA312365780GA SA312365791GA SA312365801GA SA312365807GA SA307322203GA SA307514339GA SA307514345GA SA308094805GA SA308094922GA SA308094930GA SA308094931GA SA308094939GA SA308094941GA SA308094950GA SA308094951GA SA308459856GA SA308479901GA SA308489919GA SA308489924GA SA308489932GA SA308489933GA SA308489936GA SA308489941GA SA308489942GA SA308489958GA SA308489975GA SA308489980GA SA308489981GA SA308489985GA SA308449567GA SA309072393GA SA309072395GA SA309072396GA SA309072398GA SA309072400GA SA309072401GA SA309082415GA SA309173108GA SA309173109GA SA309173112GA SA309173114GA SA309385522GA SA309385525GA SA309385664GA SA309456538GA SA309456543GA SA309456551GA SA309456555GA SA309456558GA SA310496633GA SA310496634GA SA310506676GA SA310506704GA SA310506710GA SA310506719GA SA310507495GA SA310528921GA SA310475483GA SA311132617GA SA311132631GA SA311174199GA SA311174224GA SA311163711GA SA311163716GA SA311256812GA SA311256816GA SA311256869GA SA307312104GA SA307453539GA SA310038454GA SA309334613GA SA309334640GA SA309334666GA SA309334667GA SA309334670GA SA309334686GA SA309334704GA SA309334754GA SA309344856GA SA309344860GA SA309344867GA SA309344870GA SA312018759GA SA307231242GA SA308388481GA SA308388592GA SA308388597GA SA308368211GA SA308368231GA SA308388558GA SA308388572GA SA308388581GA SA308388621GA SA308388623GA SA308388624GA SA308388633GA SA308398664GA SA308398669GA SA308398678GA SA308398692GA SA308398706GA SA308398709GA SA308398710GA SA308398786GA SA308398787GA SA308398788GA SA308398789GA SA308398790GA SA308398792GA SA308398793GA SA308398796GA SA308398797GA SA308398798GA SA308398799GA SA308398800GA SA308398801GA SA308398804GA SA308398806GA SA308398808GA SA308398809GA SA308398811GA SA308398816GA SA308398819GA SA308398820GA SA308398823GA SA308398829GA SA308398831GA SA308398832GA SA308398833GA SA308398834GA SA308398835GA SA308398837GA SA308398838GA SA308398839GA SA308398840GA SA308398842GA SA308398843GA SA308398844GA SA308398846GA SA308398849GA SA308398852GA SA308398904GA SA308357888GA SA308388566GA SA308388571GA SA308388574GA SA308388599GA SA308388630GA SA308398671GA SA308398672GA SA308398673GA SA308398674GA SA308398675GA SA308398679GA SA308398682GA SA308398684GA SA308398686GA SA308398791GA SA308398794GA SA308398805GA SA308378281GA SA308378306GA SA308398836GA SA308408936GA SA308408939GA SA308408942GA SA308408947GA SA308408950GA SA308408953GA SA308408954GA SA308408955GA SA308408958GA SA308408959GA SA308408961GA SA308408962GA SA308408963GA SA308409009GA SA308409011GA SA308409014GA SA308409018GA SA308409021GA SA308409029GA SA308409051GA SA308409052GA SA308409062GA SA308409066GA SA308409067GA SA308388554GA SA308388573GA SA308388575GA SA308388589GA SA308388598GA SA308388601GA SA308388617GA SA308388636GA SA308388637GA SA308388638GA SA308398666GA SA308398667GA SA308398668GA SA308398680GA SA308398685GA SA308398688GA SA308398690GA SA308398691GA SA308388550GA SA308388563GA SA308388564GA SA308388591GA SA308388594GA SA308388631GA SA308388634GA SA308358030GA SA308368060GA SA308368232GA SA308378280GA SA308398850GA SA308398851GA SA308398859GA SA308408918GA SA308408919GA SA308408920GA SA308408922GA SA308408926GA SA308408928GA SA308408933GA SA308408940GA SA308408943GA SA308408944GA SA308408951GA SA308419233GA SA308419239GA SA308419243GA SA308419258GA SA308419264GA SA308429363GA SA308409188GA SA308429327GA SA308429331GA SA308429342GA SA308429349GA SA308429350GA SA308429351GA SA308429353GA SA308429354GA SA308398807GA SA308398845GA SA308409170GA SA308409174GA SA308409177GA SA308409200GA SA308409212GA SA308419218GA SA308419229GA SA308419231GA SA308419232GA SA308419236GA SA308419240GA SA308419242GA SA308419257GA SA308419259GA SA308419260GA SA308419262GA SA308419266GA SA308419268GA SA308501400GA SA308501418GA SA308511434GA SA308511494GA SA308511497GA SA308511498GA SA308511507GA SA308511549GA SA308511445GA SA308511463GA SA308511544GA SA308511553GA SA308511635GA SA308511443GA SA308511516GA SA308511543GA SA308511546GA SA308511688GA SA308511691GA SA308511692GA SA308511694GA SA308511700GA SA308511701GA SA308511706GA SA308511712GA SA308511716GA SA308511718GA SA308511720GA SA308511721GA SA308511722GA SA308501403GA SA308511482GA SA308511528GA SA308511632GA SA308511642GA SA308511707GA SA308511714GA SA308511726GA SA308511729GA SA308511730GA SA308511731GA SA308511732GA SA308511733GA SA308511734GA SA308511735GA SA308511736GA SA308511737GA SA308511738GA SA308511739GA SA308511740GA SA308511741GA SA308511742GA SA308511743GA SA308511744GA SA308511745GA SA308511747GA SA308511748GA SA308511749GA SA308511768GA SA308511772GA SA309011867GA SA309011882GA SA309031996GA SA309042157GA SA309042162GA SA309042163GA SA309042164GA SA309042165GA SA309042168GA SA309042171GA SA309042172GA SA309042174GA SA309132801GA SA309132805GA SA309132812GA SA309132814GA SA309132817GA SA309132822GA SA309132823GA SA309173113GA SA309173115GA SA309173120GA SA309173164GA SA309173173GA SA309173176GA SA309173177GA SA309173179GA SA309173181GA SA309173182GA SA309173188GA SA309173189GA SA309243642GA SA309243643GA SA309243645GA SA309243646GA SA309243650GA SA309243651GA SA309243652GA SA309243653GA SA309243654GA SA309243655GA SA309253865GA SA309314291GA SA309324300GA SA309324304GA SA309324332GA SA309354923GA SA309354932GA SA309354934GA SA309354954GA SA309354960GA SA309354973GA SA309354984GA SA309354992GA SA309354996GA SA309354997GA SA309354999GA SA309355001GA SA309355005GA SA309355011GA SA309365066GA SA309365081GA SA309395696GA SA309395697GA SA309395699GA SA309395701GA SA309395711GA SA309395733GA SA309395744GA SA309395751GA SA309395791GA SA309395801GA SA309395808GA SA309405826GA SA309405840GA SA309446525GA SA309446529GA SA309446530GA SA309456548GA SA309466776GA SA310199699GA SA310260474GA SA311060021GA SA311060050GA SA307392995GA SA307080495GA SA307221159GA SA309324383GA SA309324502GA SA309324525GA SA309334761GA SA309334769GA SA309334773GA SA309334775GA SA309334787GA SA309334792GA SA309334796GA SA309334798GA SA309334805GA SA309334817GA SA309334818GA SA309334819GA SA309334820GA SA309334821GA SA309334822GA SA309334824GA SA309334827GA SA309354890GA SA309354929GA SA309354931GA SA309354935GA SA309354951GA SA309354958GA SA309354981GA SA309355009GA SA309355018GA SA309365030GA SA309365031GA SA309365040GA SA309365049GA SA309365052GA SA309365058GA SA309365070GA SA309365087GA SA309365089GA SA309365096GA SA309365100GA SA309365106GA SA309365107GA SA309365119GA SA309365121GA SA309365146GA SA309365148GA SA309365149GA SA309365153GA SA309365156GA SA309365159GA SA309365160GA SA309365161GA SA309365169GA SA309365170GA SA309365175GA SA309365176GA SA309365188GA SA309365195GA SA309365200GA SA309365208GA SA309365210GA SA309365211GA SA309365212GA SA309365213GA SA309365218GA SA309365219GA SA309365220GA SA309365238GA SA309365244GA SA309365247GA SA309365253GA SA309365255GA SA309365269GA SA309365313GA SA309365314GA SA309354899GA SA309354911GA SA309354946GA SA309365035GA SA309365242GA SA309365256GA SA309405817GA SA309375365GA SA309385617GA SA309385674GA SA309395759GA SA309405849GA SA309405863GA SA309405872GA SA309415923GA SA309415932GA SA309415940GA SA309405818GA SA309426066GA SA309426067GA SA309426071GA SA309426078GA SA309456719GA SA309456728GA SA309456729GA SA309466742GA SA309466775GA SA309466777GA SA309446401GA SA309466802GA SA309466806GA SA309466811GA SA309466838GA SA309466839GA SA309466855GA SA309476864GA SA309476873GA SA309476876GA SA309476877GA SA309476880GA SA309476882GA SA309476899GA SA309476900GA SA309507441GA SA309517839GA SA309507404GA SA309517501GA SA309517526GA SA309517627GA SA309517636GA SA309517641GA SA309517642GA SA309517643GA SA309517524GA SA309517539GA SA309517654GA SA310028109GA SA310028146GA SA310028156GA SA310028159GA SA310028194GA SA310028198GA SA310028204GA SA310028207GA SA310028210GA SA310028278GA SA310028302GA SA310038432GA SA310068854GA SA310068905GA SA310068910GA SA310068939GA SA310038374GA SA310068847GA SA310068850GA SA310068873GA SA310068883GA SA310068909GA SA310068928GA SA310068932GA SA310068937GA SA310139040GA SA310139050GA SA310068849GA SA310068878GA SA310139042GA SA310139043GA SA310139048GA SA310139049GA SA310139053GA SA310139060GA SA310139064GA SA310139072GA SA310139078GA SA310139079GA SA310259945GA SA310392884GA SA310393182GA SA310393183GA SA310393208GA SA310393221GA SA310423665GA SA310423669GA SA310423679GA SA310423692GA SA310423700GA SA310423703GA SA310423706GA SA310423707GA SA310423710GA SA310423711GA SA310423714GA SA310423716GA SA310423718GA SA310423719GA SA310423722GA SA310423727GA SA310423927GA SA310423941GA SA310423954GA SA310423964GA SA310423969GA SA310424014GA SA310434087GA SA310434121GA SA310434147GA SA310475449GA SA310475458GA SA310475468GA SA310475470GA SA310475473GA SA310475479GA SA310475484GA SA310475499GA SA310475504GA SA310475505GA SA310475508GA SA310475509GA SA310475511GA SA310475522GA SA310475524GA SA310475533GA SA310475480GA SA310475523GA SA310496328GA SA310496365GA SA310496459GA SA310496482GA SA310496338GA SA310496339GA SA310496455GA SA310496505GA SA310496521GA SA310496525GA SA310496621GA SA310496627GA SA310496637GA SA310496641GA SA310496642GA SA310496644GA SA310507109GA SA310423792GA SA310496224GA SA310496395GA SA310496407GA SA310496409GA SA310496420GA SA310496424GA SA310496425GA SA310496429GA SA310496444GA SA310496445GA SA310496449GA SA310496451GA SA310506727GA SA310506735GA SA310506763GA SA310506765GA SA310506770GA SA310507059GA SA310507063GA SA310507133GA SA310507137GA SA310507151GA SA310507155GA SA310507160GA SA310507193GA SA310507217GA SA310506760GA SA310506780GA SA310506795GA SA310506805GA SA310506834GA SA310507168GA SA310507169GA SA310507184GA SA310507186GA SA310507239GA SA310507248GA SA310507250GA SA310507256GA SA310507286GA SA310507295GA SA310507368GA SA310507370GA SA310506857GA SA310518180GA SA310507435GA SA310507531GA SA310506688GA SA310506690GA SA310507046GA SA310507144GA SA310507393GA SA310518285GA SA310518300GA SA310518308GA SA310518315GA SA310518317GA SA310331340GA SA310403380GA SA310423975GA SA310507177GA SA310507291GA SA310517622GA SA310517627GA SA310517640GA SA310517654GA SA310517665GA SA310517666GA SA310517668GA SA310517669GA SA310517671GA SA310517674GA SA310518148GA SA310518166GA SA310518196GA SA310518201GA SA310518215GA SA310518218GA SA310518224GA SA310518238GA SA310518243GA SA310518246GA SA310518249GA SA310518250GA SA310518264GA SA310518267GA SA310518273GA SA310518279GA SA310518293GA SA310518295GA SA310518301GA SA310518310GA SA310518313GA SA310518314GA SA310518318GA SA310518321GA SA310518327GA SA310518334GA SA310518619GA SA310518346GA SA310518387GA SA310518389GA SA310518429GA SA310518435GA SA310518438GA SA310528945GA SA310529152GA SA310529221GA SA311039388GA SA311039399GA SA311039408GA SA311039420GA SA311039437GA SA311039447GA SA311039454GA SA311070513GA SA311070541GA SA311070562GA SA311070564GA SA311080728GA SA311080732GA SA311080734GA SA311090773GA SA311090778GA SA311090779GA SA311090780GA SA311090800GA SA311090804GA SA311090835GA SA311090842GA SA311090850GA SA311090853GA SA311090856GA SA311090861GA SA310455168GA SA311090887GA SA311111661GA SA311111672GA SA311122134GA SA311132525GA SA311142978GA SA311142961GA SA311122057GA SA311142747GA SA311142934GA SA311153204GA SA311266994GA SA311142711GA SA311163377GA SA311163385GA SA311163396GA SA311195063GA SA311195065GA SA311195114GA SA311256701GA SA311184518GA SA311184519GA SA311184544GA SA311184554GA SA311184558GA SA311184559GA SA311184560GA SA311184564GA SA311184571GA SA311184572GA SA311184581GA SA311184582GA SA311184584GA SA311184590GA SA311184592GA SA311184608GA SA311184610GA SA311184611GA SA311184612GA SA311174238GA SA311174242GA SA311236093GA SA311236096GA SA311184362GA SA311184532GA SA311184545GA SA311184556GA SA311184565GA SA311184567GA SA311184568GA SA311184591GA SA311184597GA SA311184599GA SA311184603GA SA311184606GA SA311195198GA SA311205470GA SA311121969GA SA311121972GA SA311121974GA SA311121976GA SA311121988GA SA311122006GA SA311132539GA SA311132607GA SA311132668GA SA311142807GA SA311163458GA SA311163467GA SA311163485GA SA311163553GA SA311174067GA SA311184438GA SA311184474GA SA311184498GA SA311205459GA SA311205468GA SA311205485GA SA311205493GA SA311205496GA SA311205497GA SA311225985GA SA311236289GA SA311236311GA SA311236333GA SA311174167GA SA311184832GA SA311266884GA SA311205394GA SA311205418GA SA311236051GA SA311236352GA SA311236357GA SA311236359GA SA311236366GA SA311236368GA SA311236377GA SA311246469GA SA311246511GA SA311317644GA SA311327870GA SA311327871GA SA311327876GA SA311236019GA SA311236027GA SA311236033GA SA311236059GA SA311236173GA SA311236181GA SA311236266GA SA311236272GA SA311246504GA SA311266935GA SA311266941GA SA311327880GA SA311307439GA SA311458213GA SA311458259GA SA311458342GA SA311458349GA SA311458359GA SA311468386GA SA311468390GA SA311468400GA SA311246421GA SA311307431GA SA311307509GA SA311307580GA SA311307594GA SA311317666GA SA311317706GA SA311317711GA SA311317715GA SA311317717GA SA311317719GA SA311317727GA SA311317728GA SA311317739GA SA311317741GA SA311317742GA SA311256818GA SA310259896GA SA310259900GA SA310260366GA SA310260371GA SA310260391GA SA310290751GA SA310331082GA SA310331083GA SA310331084GA SA310331085GA SA310331086GA SA310331087GA SA310331089GA SA310331090GA SA310331091GA SA310331093GA SA310331094GA SA310331095GA SA310331096GA SA310331097GA SA310331099GA SA310331100GA SA310331101GA SA310331102GA SA310331103GA SA310331105GA SA310331106GA SA310331107GA SA310331108GA SA310331110GA SA310331111GA SA310331112GA SA310331113GA SA310331116GA SA310331119GA SA310331120GA SA310331121GA SA310331122GA SA310331124GA SA310331125GA SA310331129GA SA310331130GA SA310331131GA SA310331132GA SA310331133GA SA310331134GA SA310331135GA SA310331137GA SA310331140GA SA310331141GA SA310331142GA SA310331143GA SA310331144GA SA310331146GA SA310331147GA SA310331148GA SA310331150GA SA310331151GA SA310331152GA SA310331153GA SA310331155GA SA310331156GA SA310331157GA SA310331158GA SA310331161GA SA310331163GA SA310331164GA SA310331165GA SA310331166GA SA310331167GA SA310331168GA SA310331169GA SA310331170GA SA310331171GA SA310331172GA SA310331173GA SA310331174GA SA310331175GA SA310331176GA SA310331177GA SA310331178GA SA310331179GA SA310331180GA SA310331181GA SA310331182GA SA310331183GA SA310331184GA SA310331185GA SA310331186GA SA310331187GA SA310331188GA SA310331189GA SA310331190GA SA310331191GA SA310331192GA SA310331193GA SA310331194GA SA310331195GA SA310331196GA SA310331197GA SA310331198GA SA310331199GA SA310331200GA SA310331201GA SA310331202GA SA310331203GA SA310331204GA SA310331205GA SA310331206GA SA310331207GA SA310331208GA SA310331209GA SA310331210GA SA310331211GA SA310331212GA SA310331213GA SA312263887GA SA312263931GA SA312284262GA SA312284275GA SA312294530GA SA312294615GA SA312294622GA SA312294636GA SA312294649GA SA312294651GA SA312294655GA SA312294659GA SA312294664GA SA312294672GA SA312294676GA SA312294650GA SA312152349GA SA312152480GA SA312152497GA SA312152547GA SA312232797GA SA312232802GA SA312232815GA SA312232821GA SA312253077GA SA312253119GA SA312264009GA SA312344852GA SA312344933GA SA312376078GA SA312376286GA SA312386323GA SA312386335GA SA312386339GA SA312386342GA SA312386343GA SA312386346GA SA312386561GA SA312386576GA SA312386611GA SA312386612GA SA312386616GA SA312386618GA SA312386619GA SA312386621GA SA312386634GA SA312386635GA SA312386636GA SA312386638GA SA312386639GA SA312386640GA SA312386505GA SA312386509GA SA312386526GA SA312396787GA SA312396807GA SA312396861GA SA312396910GA SA312396941GA SA312396945GA SA312397035GA SA312397049GA SA312397050GA SA312397051GA SA312397056GA SA312397060GA SA312397092GA SA312397098GA SA312397119GA SA312386573GA SA312386574GA SA312397038GA SA312397137GA SA312397163GA SA312386397GA SA312386429GA SA312386497GA SA312386521GA SA312386590GA SA312396980GA SA312397167GA SA312397269GA SA312397275GA SA312407276GA SA312407277GA SA312407279GA SA312407284GA SA312407285GA SA312397018GA SA312397025GA SA312397088GA SA312397172GA SA312407287GA SA312407299GA SA312457602GA SA312457603GA SA312457604GA SA312457605GA SA312457606GA SA312457607GA SA312457608GA SA312457609GA SA312457611GA SA312457613GA SA312457615GA SA312457616GA SA312457617GA SA312457650GA SA312457651GA SA312397077GA SA312427418GA SA312457612GA SA312457614GA SA312457618GA SA312457620GA SA312457621GA SA312457627GA SA312457630GA SA312457688GA SA312457713GA SA312457601GA SA312457610GA SA312457619GA SA312457622GA SA312457624GA SA312457625GA SA312457628GA SA312457629GA SA312457631GA SA312457632GA SA312457634GA SA312457635GA SA312457636GA SA312457638GA SA312457640GA SA312457641GA SA312457642GA SA312457643GA SA312457646GA SA312457647GA SA312457648GA SA312457649GA SA312457652GA SA312457653GA SA312457654GA SA312457656GA SA312457658GA SA312457659GA SA312457660GA SA312457662GA SA312457663GA SA312457664GA SA312457669GA SA312457677GA SA312457678GA SA312457681GA SA312457689GA SA312457691GA SA312457710GA SA312457712GA SA312457715GA SA312457716GA SA312457720GA SA312457731GA SA312457733GA SA312457734GA SA312457735GA SA312457739GA SA312457743GA SA312457744GA SA312457745GA SA312457746GA SA312457747GA SA312457748GA SA312457749GA SA312457750GA SA312457752GA SA312457754GA SA312457755GA SA312457756GA SA312457757GA SA312457758GA SA312457760GA SA312457762GA SA312457763GA SA312457764GA SA312457765GA SA312457766GA SA312457767GA SA312457770GA SA312457772GA SA312457773GA SA312457774GA SA312457775GA SA312457776GA SA312457777GA SA312457778GA SA312457779GA SA312457780GA SA312457781GA SA312457782GA SA312457783GA SA312457784GA SA312457787GA SA312457788GA SA312457789GA SA312457790GA SA312457792GA SA312457793GA SA312457794GA SA312457796GA SA312457797GA SA312457798GA SA312457799GA SA312457802GA SA312457803GA SA312457804GA SA312457809GA SA312457813GA SA312457815GA SA312457816GA SA312457817GA SA312457818GA SA312457819GA SA312457821GA SA312457822GA SA312457823GA SA312457824GA SA312457825GA SA312457826GA SA312457827GA SA312457828GA SA312457829GA SA312457831GA SA312457832GA SA312457833GA SA312457834GA SA312457835GA SA312457837GA SA312457838GA SA312457839GA SA312457840GA SA312457841GA SA312457842GA SA312457843GA SA312457845GA SA312457846GA SA312457847GA SA312457849GA SA312457850GA SA312457851GA SA312457852GA SA312457853GA SA312457854GA SA312457855GA SA312457856GA SA312457857GA SA312457859GA SA312457860GA SA312457861GA SA312457862GA SA312457863GA SA312457864GA SA312457865GA SA312457866GA SA312457867GA SA312457868GA SA312457869GA SA312457870GA SA312457871GA SA312457872GA SA312457873GA SA312457874GA SA312457875GA SA312457876GA SA312457877GA SA312457879GA SA312457880GA SA312457881GA SA312457882GA SA312457883GA SA312457884GA SA312457885GA SA312457886GA SA312457887GA SA312457888GA SA312457890GA SA312457892GA SA312457893GA SA312457895GA SA312457896GA SA312457897GA SA312457898GA SA312457901GA SA312457902GA SA312457903GA SA312457905GA SA312457907GA SA312457911GA SA312457912GA SA312457913GA SA312457916GA SA312457917GA SA312457918GA SA312457920GA SA312457924GA SA312457926GA SA312457928GA SA312457929GA SA312457931GA SA312457932GA SA312457933GA SA312457934GA SA312457935GA SA312457936GA SA312457937GA SA312457938GA SA312457942GA SA312457946GA SA312457948GA SA312457949GA SA312457950GA SA312457956GA SA312457958GA SA312457963GA SA312457966GA SA312457967GA SA312457971GA SA312457972GA SA312457973GA SA312457974GA SA312457975GA SA312457977GA SA312457978GA SA312457982GA SA312457983GA SA312457986GA SA312457990GA SA312457997GA SA312457999GA SA312458003GA SA312458004GA SA312458005GA SA312458006GA SA312458011GA SA312458013GA SA312458016GA SA312458022GA SA312458023GA SA312458024GA SA312458028GA SA312458032GA SA312458038GA SA312458040GA SA312458041GA SA312458042GA SA312458044GA SA312458045GA SA312458048GA SA312458049GA SA312458052GA SA312458053GA SA312458054GA SA312458057GA SA312458058GA SA312458059GA SA312458061GA SA312458062GA SA312458063GA SA312458066GA SA312458067GA SA312458077GA SA312458082GA SA312458085GA SA312458088GA SA312458090GA SA312458095GA SA312458121GA SA312458124GA SA312458191GA SA312457769GA SA312457830GA SA312457844GA SA312457894GA SA312457899GA SA312457904GA SA312457906GA SA312457908GA SA312457909GA SA312457919GA SA312457921GA SA312457922GA SA312457925GA SA312457930GA SA312457939GA SA312457940GA SA312457941GA SA312457943GA SA312457944GA SA312457945GA SA312457947GA SA312457951GA SA312457952GA SA312457953GA SA312457954GA SA312457955GA SA312457957GA SA312457959GA SA312457960GA SA312457961GA SA312457962GA SA312457965GA SA312457970GA SA312457976GA SA312457979GA SA312457981GA SA312457994GA SA312457996GA SA312458000GA SA312458002GA SA312458008GA SA312458010GA SA312458012GA SA312458017GA SA312458019GA SA312458025GA SA312458026GA SA312458030GA SA312458034GA SA312458037GA SA312458050GA SA312458069GA SA312458074GA SA312458076GA SA312458079GA SA312458080GA SA312458084GA SA312458086GA SA312458087GA SA312458089GA SA312458093GA SA312458097GA SA312458098GA SA312458100GA SA312458101GA SA312458102GA SA312458106GA SA312458108GA SA312458110GA SA312458113GA SA312458116GA SA312458118GA SA312458120GA SA312458122GA SA312458125GA SA312458127GA SA312458182GA SA312458184GA SA312458196GA SA312458198GA SA312458202GA SA312458204GA SA312458206GA SA312458207GA SA312458208GA SA312458209GA SA312458210GA SA312458212GA SA312458213GA SA312458215GA SA312458216GA SA312458217GA SA312458218GA SA312458220GA SA312458221GA SA312458222GA SA312458235GA SA312458236GA SA312458238GA SA312458239GA SA312458240GA SA312458243GA SA312458248GA SA312458402GA SA312458413GA SA312458429GA SA312458430GA SA312458434GA SA312458435GA SA312458456GA SA312458457GA SA312458458GA SA312458459GA SA312458461GA SA312458462GA SA312458463GA SA312458465GA SA312457722GA SA312457900GA SA312458071GA SA312458200GA SA312458205GA SA312458211GA SA312458214GA SA312458224GA SA312458227GA SA312458229GA SA312458234GA SA312458237GA SA312458242GA SA312458253GA SA312458259GA SA312458260GA SA312458262GA SA312458263GA SA312458266GA SA312458394GA SA312458396GA SA312458399GA SA312458406GA SA312458409GA SA312458410GA SA312458418GA SA312458419GA SA312458422GA SA312458423GA SA312458432GA SA312468532GA SA312468533GA SA312468537GA SA312468538GA SA312468540GA SA312468542GA SA312468545GA SA312468552GA SA312468553GA SA312468572GA SA312468575GA SA312468579GA SA312468581GA SA312468591GA SA312457645GA SA312458056GA SA312458068GA SA312458091GA SA312458105GA SA312458112GA SA312458193GA SA312458197GA SA312458201GA SA312458219GA SA312458230GA SA312458231GA SA312458241GA SA312458247GA SA312458249GA SA312458250GA SA312458251GA SA312458254GA SA312458255GA SA312458257GA SA312458261GA SA312458265GA SA312458267GA SA312458395GA SA312458400GA SA312458401GA SA312458403GA SA312458405GA SA312458407GA SA312458408GA SA312458411GA SA312458412GA SA312458414GA SA312458415GA SA312458416GA SA312458417GA SA312458421GA SA312458424GA SA312458426GA SA312458428GA SA312458431GA SA312468534GA SA312468541GA SA312468547GA SA312468550GA SA312468551GA SA312468562GA SA312468563GA SA312468564GA SA312468565GA SA312468566GA SA312468567GA SA312468571GA SA312468577GA SA312468582GA SA312468583GA SA312468585GA SA312468587GA SA312468588GA SA312468589GA SA312468772GA SA312468774GA SA312468775GA SA312468777GA SA312468778GA SA312468779GA SA312468780GA SA312468781GA SA312468786GA SA312468787GA SA312468791GA SA312468792GA SA312468794GA SA312468796GA SA312468798GA SA312468800GA SA312468801GA SA312468805GA SA312468806GA SA312468807GA SA312468813GA SA312468814GA SA312468816GA SA312468817GA SA312468818GA SA312468819GA SA312468820GA SA312468826GA SA312468849GA SA312468850GA SA312468852GA SA312468853GA SA312468857GA SA312468858GA SA312468859GA SA312468861GA SA312468864GA SA312468867GA SA312457644GA SA312458043GA SA312458051GA SA312468784GA SA312468793GA SA312468831GA SA312468844GA SA312468851GA SA312468862GA SA312468924GA SA312468925GA SA312468946GA SA312468953GA SA312468962GA SA312468980GA SA312468984GA SA312468990GA SA312468996GA SA312469000GA SA312469003GA SA312469042GA SA312469052GA SA312469068GA SA312469107GA SA312469110GA SA312469115GA SA312469125GA SA312469127GA SA312469128GA SA312469135GA SA312469140GA SA312469154GA SA312469199GA SA312458009GA SA312458126GA SA312468568GA SA312468570GA SA312468838GA SA312468841GA SA312468907GA SA312468944GA SA312468968GA SA312468971GA SA312468974GA SA312468975GA SA312468979GA SA312468981GA SA312468987GA SA312468995GA SA312469002GA SA312469004GA SA312469005GA SA312469009GA SA312469046GA SA312469047GA SA312469049GA SA312469050GA SA312469051GA SA312469060GA SA312469061GA SA312469062GA SA312469064GA SA312469066GA SA312469067GA SA312469070GA SA312469071GA SA312469103GA SA312469104GA SA312469105GA SA312469108GA SA312469116GA SA312469117GA SA312469118GA SA312469119GA SA312469121GA SA312469123GA SA312469124GA SA312469129GA SA312469137GA SA312469139GA SA312469142GA SA312469157GA SA312469164GA SA312469172GA SA312469182GA SA312469184GA SA312469188GA SA312469189GA SA312469193GA SA312469194GA SA312469206GA SA312469207GA SA312469213GA SA312469217GA SA312469218GA SA312469221GA SA312469283GA SA312469284GA SA312469285GA SA312469286GA SA312469287GA SA312469291GA SA312469292GA SA312469293GA SA312469294GA SA312469302GA SA312469308GA SA312469309GA SA312469350GA SA312469351GA SA312469356GA SA312469358GA SA312355253GA SA312386386GA SA312386530GA SA312386665GA SA312457623GA SA312457633GA SA312457637GA SA312457661GA SA312457665GA SA312457673GA SA312457676GA SA312457680GA SA312457684GA SA312457695GA SA312457699GA SA312457708GA SA312457709GA SA312457714GA SA312457725GA SA312457729GA SA312457732GA SA312458183GA SA312458188GA SA312458190GA SA312469007GA SA312469215GA SA312469282GA SA312469296GA SA312469297GA SA312469299GA SA312469300GA SA312469307GA SA312469311GA SA312469342GA SA312469343GA SA312469344GA SA312469345GA SA312469346GA SA312469348GA SA312469352GA SA312469353GA SA312469355GA SA312469357GA SA312469359GA SA312469360GA SA312469362GA SA312469363GA SA312469365GA SA312469366GA SA312469368GA SA312469370GA SA312469371GA SA312457753GA SA312458109GA SA312468803GA SA312468902GA SA312468954GA SA312468959GA SA312468963GA SA312468964GA SA312468967GA SA312468969GA SA312468970GA SA312468978GA SA312468982GA SA312468988GA SA312468989GA SA312468993GA SA312468999GA SA312469006GA SA312469010GA SA312469011GA SA312469106GA SA312469111GA SA312469120GA SA312469132GA SA312469141GA SA312469143GA SA312469147GA SA312469149GA SA312469150GA SA312469151GA SA312469153GA SA312469156GA SA312469162GA SA312469163GA SA312469166GA SA312469167GA SA312469169GA SA312469170GA SA312469171GA SA312469174GA SA312469176GA SA312469177GA SA312469181GA SA312469186GA SA312469190GA SA312469192GA SA312469197GA SA312469198GA SA312469289GA SA312469290GA SA312469298GA SA312469301GA SA312469303GA SA312469349GA SA312479685GA SA312479696GA SA312479791GA SA312479793GA SA312479794GA SA312479797GA SA312479811GA SA312479817GA SA312479818GA SA312479754GA SA312479788GA SA312479795GA SA312479801GA SA312457639GA SA312457655GA SA312457666GA SA312457667GA SA312457668GA SA312457670GA SA312457671GA SA312457672GA SA312457674GA SA312457675GA SA312457682GA SA312457687GA SA312457690GA SA312457694GA SA312457696GA SA312457697GA SA312457698GA SA312457717GA SA312457718GA SA312457724GA SA312457730GA SA312457737GA SA312457738GA SA312457741GA SA312457742GA SA312457751GA SA312479553GA SA312479594GA SA312479742GA SA312480118GA SA312480120GA SA312458070GA SA312457761GA SA312469191GA SA312480333GA SA312480334GA SA312480336GA SA312480341GA SA312480344GA SA312480345GA SA312480350GA SA312480354GA SA312480357GA SA312480358GA SA312480359GA SA312480361GA SA312480362GA SA312480379GA SA312480389GA SA312091100GA SA312070118GA SA312070134GA SA312080865GA SA312070340GA SA311428032GA SA312039088GA SA312039132GA SA307030426GA SA307070445GA SA307080461GA SA307080464GA SA307080465GA SA307080470GA SA307080490GA SA307080512GA SA307080528GA SA307080532GA SA307080535GA SA307150663GA SA307180700GA SA307180703GA SA307180709GA SA307190763GA SA307302065GA SA307392931GA SA307443243GA SA307443364GA SA307443374GA SA307453395GA SA307453397GA SA307453550GA SA308084648GA SA308074590GA SA308105085GA SA308105090GA SA308398715GA SA308398727GA SA308398747GA SA308398750GA SA308398757GA SA308398759GA SA308398760GA SA308398766GA SA308398775GA SA308398782GA SA308368243GA SA308378321GA SA308388576GA SA308398814GA SA308398818GA SA308409089GA SA308409184GA SA308419230GA SA308419255GA SA308419263GA SA308429312GA SA308429359GA SA308439428GA SA308439430GA SA308439436GA SA308439437GA SA308439440GA SA308439441GA SA308439448GA SA308439449GA SA308459846GA SA308449512GA SA308449518GA SA308449521GA SA308449575GA SA308449595GA SA308459641GA SA308459669GA SA308459673GA SA308501299GA SA308501300GA SA308501338GA SA308501346GA SA308501347GA SA308501385GA SA308511570GA SA309052230GA SA309062276GA SA309173124GA SA309173130GA SA309173131GA SA309173132GA SA309173134GA SA309173135GA SA309173136GA SA309173137GA SA309173138GA SA309173139GA SA309173141GA SA309173143GA SA309173144GA SA309173148GA SA309173152GA SA309173153GA SA309173154GA SA309173156GA SA309173158GA SA309173168GA SA309304287GA SA309365085GA SA309365101GA SA309365267GA SA309385543GA SA309385551GA SA309385557GA SA309385624GA SA309375405GA SA309405870GA SA309405882GA SA309405908GA SA309405916GA SA309405920GA SA309365152GA SA309365317GA SA309365328GA SA309365330GA SA309365334GA SA309365339GA SA309365360GA SA309375374GA SA309375376GA SA309436362GA SA309456676GA SA309466848GA SA309476863GA SA309436200GA SA309446381GA SA309476925GA SA309365259GA SA309365265GA SA309365281GA SA309365282GA SA309365284GA SA309375437GA SA309375438GA SA309375439GA SA309375445GA SA309375446GA SA309375450GA SA309477074GA SA309446450GA SA309456545GA SA309456549GA SA309466744GA SA309466758GA SA309466760GA SA309466763GA SA309466782GA SA309466784GA SA309466793GA SA309466800GA SA309395730GA SA309395761GA SA309436318GA SA309436350GA SA309456535GA SA309476937GA SA309405854GA SA309476960GA SA309517670GA SA309517764GA SA309517767GA SA310038499GA SA310038512GA SA310048565GA SA310048644GA SA310048677GA SA310058703GA SA310058719GA SA310058731GA SA310058732GA SA310058734GA SA310058736GA SA310068802GA SA310068807GA SA310068811GA SA310068817GA SA310068821GA SA310068826GA SA310068829GA SA310068830GA SA310068835GA SA310068840GA SA310068841GA SA310068843GA SA309527909GA SA310048652GA SA310048672GA SA310038495GA SA310118965GA SA310118975GA SA310119004GA SA310139090GA SA310139091GA SA310139097GA SA310139099GA SA310139101GA SA310139102GA SA310139103GA SA310139105GA SA310139108GA SA310139109GA SA310139110GA SA310139111GA SA310139113GA SA310139116GA SA310139117GA SA310139118GA SA310139123GA SA310139124GA SA310139125GA SA310139128GA SA310139129GA SA310139130GA SA310139131GA SA310139132GA SA310139136GA SA310139137GA SA310139138GA SA310139139GA SA310139140GA SA310139147GA SA310139170GA SA310139171GA SA310139182GA SA310139196GA SA310038480GA SA310038481GA SA310139112GA SA310139292GA SA310139297GA SA310149347GA SA310159425GA SA310159445GA SA310159446GA SA310159448GA SA310159452GA SA310159447GA SA310159451GA SA310259895GA SA310259897GA SA310259899GA SA310259901GA SA310259902GA SA310259903GA SA310259905GA SA310259906GA SA310259907GA SA310259908GA SA310259909GA SA310259910GA SA310259931GA SA310259933GA SA310259934GA SA310259935GA SA310259936GA SA310259937GA SA310259940GA SA310259947GA SA310259952GA SA310259962GA SA310259968GA SA310259969GA SA310118962GA SA310118992GA SA310119005GA SA310119009GA SA310119018GA SA310260357GA SA310260358GA SA310260361GA SA310260363GA SA310260368GA SA310260369GA SA310260370GA SA310260378GA SA310260380GA SA310260381GA SA310260400GA SA310260403GA SA310260419GA SA310260425GA SA310260426GA SA310260427GA SA310260431GA SA310260432GA SA310260433GA SA310260435GA SA310260440GA SA310048689GA SA310068822GA SA310159389GA SA310290651GA SA310290734GA SA310290735GA SA310290736GA SA310290738GA SA310290740GA SA310290742GA SA310382574GA SA310361790GA SA310361872GA SA310372074GA SA310372081GA SA310372083GA SA310372089GA SA310372102GA SA310372129GA SA310372170GA SA310372172GA SA310372191GA SA310382361GA SA310382362GA SA310372092GA SA310372094GA SA310382546GA SA310382582GA SA310382589GA SA310382691GA SA310382697GA SA310382698GA SA310382651GA SA310403437GA SA310423813GA SA310403317GA SA310403385GA SA310403401GA SA310403438GA SA310403445GA SA310423995GA SA310423995GAMKE SA310434102GA SA310434168GA SA310455067GA SA310455134GA SA310475452GA SA310475471GA SA310475520GA SA310475535GA SA310475538GA SA310475541GA SA310475546GA SA310475547GA SA310475548GA SA310475549GA SA310475550GA SA310475552GA SA310475556GA SA310475565GA SA310475567GA SA310475568GA SA310475569GA SA310475571GA SA310475574GA SA310475576GA SA310475577GA SA310475580GA SA310475584GA SA310475587GA SA310475588GA SA310475590GA SA310475593GA SA310475594GA SA310475595GA SA310403344GA SA310403345GA SA310403376GA SA310403386GA SA310403400GA SA310403406GA SA310403425GA SA310403446GA SA310423760GA SA310423762GA SA310423764GA SA310423765GA SA310423775GA SA310423783GA SA310423915GA SA310423981GA SA310423983GA SA310331079GA SA310331088GA SA310361775GA SA310361782GA SA310361793GA SA310361806GA SA310361849GA SA310361856GA SA310361883GA SA310361885GA SA310372071GA SA310372086GA SA310372109GA SA310372119GA SA310372127GA SA310372128GA SA310372131GA SA310372181GA SA310372211GA SA310382318GA SA310382320GA SA310382324GA SA310382328GA SA310382337GA SA310382367GA SA310382373GA SA310382525GA SA310382526GA SA310382549GA SA310382551GA SA310382564GA SA310382567GA SA310382594GA SA310382595GA SA310382597GA SA310382666GA SA310382669GA SA310382676GA SA310382678GA SA310382683GA SA310382688GA SA310382694GA SA310382702GA SA310382703GA SA310382772GA SA310382781GA SA310382788GA SA310382789GA SA310382794GA SA310382799GA SA310382804GA SA310382807GA SA310382810GA SA310382811GA SA310382815GA SA310382841GA SA310392851GA SA310392852GA SA310392854GA SA310392856GA SA310392857GA SA310392860GA SA310392862GA SA310392865GA SA310392867GA SA310392875GA SA310392877GA SA310392878GA SA310392880GA SA310392881GA SA310392883GA SA310392885GA SA310392888GA SA310392890GA SA310392892GA SA310392896GA SA310392906GA SA310392908GA SA310392913GA SA310392916GA SA310392918GA SA310392920GA SA310392921GA SA310392922GA SA310392926GA SA310392927GA SA310392928GA SA310392929GA SA310392932GA SA310393078GA SA310393079GA SA310393081GA SA310393082GA SA310393085GA SA310393087GA SA310393091GA SA310393093GA SA310393098GA SA310393099GA SA310393100GA SA310393101GA SA310393102GA SA310393103GA SA310393104GA SA310393105GA SA310393107GA SA310393110GA SA310393111GA SA310393113GA SA310393115GA SA310393127GA SA310393132GA SA310393133GA SA310393134GA SA310393135GA SA310393137GA SA310393181GA SA310393184GA SA310393189GA SA310393199GA SA310393201GA SA310393204GA SA310393205GA SA310393206GA SA310393207GA SA310393209GA SA310393210GA SA310393211GA SA310393214GA SA310393218GA SA310393220GA SA310393223GA SA310393225GA SA310393226GA SA310393228GA SA310393232GA SA310393240GA SA310393241GA SA310393244GA SA310393245GA SA310393247GA SA310393253GA SA310393256GA SA310393257GA SA310393264GA SA310393266GA SA310393273GA SA310393286GA SA310393289GA SA310393301GA SA310393307GA SA310260291GA SA310403388GA SA310403392GA SA310403403GA SA310403419GA SA310423897GA SA310423901GA SA310423902GA SA310423918GA SA310423982GA SA310423996GA SA310434050GA SA310434051GA SA310496225GA SA310496458GA SA310496468GA SA310496497GA SA310496527GA SA310496528GA SA310506848GA SA310506878GA SA310506994GA SA310506995GA SA310507030GA SA310507034GA SA310507042GA SA310507052GA SA310507053GA SA310507058GA SA310507065GA SA310507126GA SA310382711GA SA310506996GA SA310507044GA SA310507060GA SA310507068GA SA310507098GA SA310507099GA SA310507103GA SA310507142GA SA310507411GA SA310496374GA SA310496385GA SA310506773GA SA310506809GA SA310506828GA SA310506835GA SA310506855GA SA310507224GA SA310506796GA SA310506817GA SA310506818GA SA310506824GA SA310506826GA SA310506837GA SA310506840GA SA310506844GA SA310507181GA SA310507225GA SA310507235GA SA310507240GA SA310507241GA SA310507243GA SA310507244GA SA310507246GA SA310507279GA SA310507281GA SA310507314GA SA310507351GA SA310507373GA SA310507374GA SA310434131GA SA310434174GA SA310434190GA SA310434192GA SA310506757GA SA310506772GA SA310506778GA SA310506782GA SA310507152GA SA310507153GA SA310507166GA SA310507170GA SA310507175GA SA310507179GA SA310507182GA SA310507247GA SA310506827GA SA310507188GA SA310507220GA SA310507271GA SA310507278GA SA310517663GA SA310518200GA SA310518213GA SA310506819GA SA310506830GA SA310507185GA SA310507190GA SA310507227GA SA310507231GA SA310507232GA SA310507236GA SA310507260GA SA310518454GA SA310528812GA SA310518550GA SA310528772GA SA310528822GA SA310528904GA SA310382529GA SA310392899GA SA310403394GA SA310495913GA SA310507223GA SA310518381GA SA310518419GA SA310518421GA SA310518443GA SA310518467GA SA310518498GA SA310518507GA SA310518508GA SA310518510GA SA310518511GA SA310518536GA SA310518544GA SA310518545GA SA310518558GA SA310528734GA SA310528765GA SA310528766GA SA310528790GA SA310528905GA SA310528907GA SA310528909GA SA310528910GA SA310528913GA SA310528919GA SA310528926GA SA310528929GA SA310528931GA SA310528934GA SA310528938GA SA310528946GA SA310528950GA SA310528958GA SA310529064GA SA310529068GA SA310529070GA SA310529072GA SA310529075GA SA310529076GA SA310529082GA SA310529084GA SA310529091GA SA310529092GA SA310529100GA SA310529117GA SA310529118GA SA310529119GA SA310518375GA SA310518413GA SA310518418GA SA310518456GA SA310518549GA SA310528726GA SA310528751GA SA310528754GA SA310528769GA SA310528806GA SA310528916GA SA310529062GA SA310529063GA SA310529066GA SA310529074GA SA310529080GA SA310529083GA SA310529085GA SA310529109GA SA310529110GA SA310260227GA SA310506876GA SA310507471GA SA310507540GA SA310528881GA SA310528982GA SA310528989GA SA310260317GA SA310260356GA SA310413563GA SA310424013GA SA310434201GA SA310475698GA SA310496279GA SA310496302GA SA310496648GA SA310507013GA SA310507532GA SA310518620GA SA310518680GA SA310518682GA SA310528853GA SA310529026GA SA310529055GA SA310529056GA SA310529147GA SA310529155GA SA311039375GA SA311039438GA SA311039440GA SA311039448GA SA311039470GA SA311060005GA SA311060066GA SA311060164GA SA311060207GA SA311060001GA SA311060003GA SA311060006GA SA311060040GA SA311060043GA SA311060044GA SA311060099GA SA311069636GA SA311069685GA SA311069687GA SA311069742GA SA311069746GA SA311069749GA SA311069750GA SA311069751GA SA311060185GA SA311060189GA SA311060205GA SA311060340GA SA311060347GA SA311060373GA SA311060388GA SA311070514GA SA311070543GA SA311070558GA SA311080604GA SA311080630GA SA311080650GA SA311080654GA SA311080659GA SA311060149GA SA311060227GA SA311060267GA SA311060291GA SA311080602GA SA311060016GA SA311060071GA SA311060093GA SA310382837GA SA310434072GA SA311060095GA SA311060110GA SA311091012GA SA310528829GA SA311069669GA SA311069680GA SA311069739GA SA311069741GA SA311069766GA SA311069815GA SA311069859GA SA311069862GA SA311069864GA SA311069865GA SA311069866GA SA311090948GA SA311090973GA SA311091029GA SA311101312GA SA311101315GA SA311101329GA SA311101373GA SA311101378GA SA311060111GA SA311090913GA SA310507528GA SA311069821GA SA311069823GA SA311069825GA SA311069855GA SA311069868GA SA311132466GA SA311132473GA SA311132524GA SA311121888GA SA311101400GA SA311142872GA SA311142949GA SA311142951GA SA311142952GA SA311142959GA SA311142960GA SA311142962GA SA311142967GA SA311142975GA SA311111599GA SA311111606GA SA311132284GA SA311142970GA SA311153082GA SA311153121GA SA311153128GA SA311153131GA SA311153140GA SA311153151GA SA311163368GA SA311163370GA SA311163371GA SA311163389GA SA311163394GA SA311163397GA SA311163420GA SA311163897GA SA311142826GA SA311121919GA SA311121949GA SA311121957GA SA311121968GA SA311122106GA SA311132414GA SA311142695GA SA311142697GA SA311142706GA SA311153312GA SA311153325GA SA311153336GA SA311153354GA SA311153355GA SA311153360GA SA311163547GA SA311163583GA SA311163642GA SA311163707GA SA311163713GA SA311163715GA SA311163728GA SA311163733GA SA311163781GA SA311163808GA SA311174160GA SA311174162GA SA311174178GA SA311174179GA SA311174182GA SA311174198GA SA311174206GA SA311122244GA SA311132596GA SA311174201GA SA311174203GA SA311174235GA SA311174236GA SA311174237GA SA311174239GA SA311174241GA SA311174243GA SA311174245GA SA311174252GA SA311184371GA SA311184387GA SA311184399GA SA311184400GA SA311184401GA SA311184402GA SA311184406GA SA311184410GA SA311184413GA SA311184524GA SA311195250GA SA311195256GA SA311195258GA SA311195264GA SA311195266GA SA311195267GA SA311195268GA SA311195270GA SA311195277GA SA311195278GA SA311195283GA SA311195284GA SA311195285GA SA311195287GA SA311195288GA SA311205358GA SA311225977GA SA311225979GA SA311225981GA SA311225986GA SA311225988GA SA311225989GA SA311225991GA SA311225992GA SA311225994GA SA311225995GA SA311225997GA SA311225999GA SA311236080GA SA311236084GA SA311236086GA SA311236115GA SA311236119GA SA311236121GA SA311236122GA SA311163407GA SA311163409GA SA311163415GA SA311246521GA SA311246561GA SA311256682GA SA311256683GA SA311256684GA SA311256685GA SA311256686GA SA311256688GA SA311256690GA SA311256691GA SA311256692GA SA311256693GA SA311256696GA SA311256697GA SA311256698GA SA311256699GA SA311256700GA SA311256702GA SA311256704GA SA311256710GA SA311256712GA SA311256715GA SA311256764GA SA311256772GA SA311256777GA SA311256779GA SA311256782GA SA311256784GA SA311256794GA SA311132516GA SA311184632GA SA311195130GA SA311195134GA SA311195147GA SA311195156GA SA311195254GA SA311184520GA SA311184618GA SA311184621GA SA311184623GA SA311184635GA SA311184497GA SA311205325GA SA311174108GA SA311184754GA SA311184803GA SA311205329GA SA311205337GA SA311205416GA SA311205436GA SA311205441GA SA311287173GA SA311297262GA SA311297264GA SA311195131GA SA311195255GA SA311195271GA SA311297266GA SA311297271GA SA311297291GA SA311297328GA SA311297342GA SA311297349GA SA311297350GA SA311297352GA SA311297365GA SA311307523GA SA311205621GA SA311215858GA SA311225895GA SA311215683GA SA311215685GA SA311215693GA SA311215696GA SA311215700GA SA311215702GA SA311215714GA SA311215717GA SA311215723GA SA311215739GA SA311236302GA SA311327883GA SA311317661GA SA311317669GA SA311184847GA SA311184868GA SA311194918GA SA311205563GA SA311205565GA SA311205567GA SA311205568GA SA311205572GA SA311205575GA SA311205577GA SA311205579GA SA311205581GA SA311205582GA SA311205584GA SA311205585GA SA311205591GA SA311205593GA SA311205602GA SA311215809GA SA311215838GA SA311215694GA SA311236038GA SA311246494GA SA311225938GA SA311225974GA SA311236003GA SA311236034GA SA311327934GA SA311428011GA SA311428057GA SA311448173GA SA311448179GA SA311448181GA SA311448192GA SA311448195GA SA311448202GA SA311184670GA SA311184831GA SA311194923GA SA311194935GA SA311194957GA SA311194959GA SA311195009GA SA311205610GA SA311246459GA SA311246477GA SA311246486GA SA311246488GA SA311246489GA SA311246514GA SA311256746GA SA311256749GA SA311256754GA SA311256810GA SA311256827GA SA311256828GA SA311256845GA SA311256848GA SA311256852GA SA311256859GA SA311256860GA SA311256862GA SA311256817GA SA311246589GA SA311246610GA SA311246612GA SA311246630GA SA311246636GA SA311246661GA SA311246671GA SA311307458GA SA311307473GA SA311428090GA SA311448194GA SA311246446GA SA311246584GA SA311246623GA SA311246657GA SA311246668GA SA311287152GA SA311317691GA SA311317698GA SA311317759GA SA311317766GA SA311317774GA SA311327910GA SA311327920GA SA311428073GA SA311428094GA SA311458212GA SA311458220GA SA311458236GA SA311458353GA SA311458361GA SA311458366GA SA311468399GA SA310331354GA SA310372073GA SA310372078GA SA310372087GA SA310372090GA SA310372093GA SA310372103GA SA310372106GA SA310372111GA SA310372112GA SA310372121GA SA310372124GA SA310372126GA SA310372133GA SA310372165GA SA310372166GA SA310372169GA SA310372175GA SA310372182GA SA310372184GA SA310372185GA SA310372196GA SA310372208GA SA310382315GA SA310382319GA SA310382321GA SA310382327GA SA310382338GA SA310382346GA SA310382350GA SA310382354GA SA310382356GA SA310382358GA SA310382365GA SA310382368GA SA310382370GA SA310382322GA SA310382372GA SA310372113GA SA310372134GA SA310372195GA SA310382433GA SA310382653GA SA310382536GA SA310382681GA SA310382773GA SA310382779GA SA310382803GA SA310382812GA SA310382845GA SA310392858GA SA310392861GA SA310392863GA SA310392869GA SA310392889GA SA312080831GA SA312132098GA SA312152473GA SA312232796GA SA312232799GA SA312232805GA SA312232810GA SA312242995GA SA312242996GA SA312243006GA SA312253069GA SA312253083GA SA312253090GA SA312253116GA SA312253117GA SA312253142GA SA312253161GA SA312253162GA SA312253233GA SA312253238GA SA312253247GA SA312253256GA SA312253263GA SA312253269GA SA312253272GA SA312253276GA SA312253278GA SA312253279GA SA312253369GA SA312253215GA SA312253221GA SA312253239GA SA312253425GA SA312253426GA SA312253456GA SA312253459GA SA312253463GA SA312253467GA SA312253475GA SA312263866GA SA312263935GA SA312263936GA SA312345056GA SA312345104GA SA312345118GA SA312345120GA SA312345130GA SA312386320GA SA312396991GA SA312396972GA SA312386624GA SA312470044GA SA312469463GA SA312480421GA SA312480441GA SA312480558GA SA312480559GA SA312480571GA SA312480578GA SA312480595GA SA312480597GA SA312480607GA SA312480703GA SA312480712GA SA312480713GA SA312480714GA SA312480716GA SA312480718GA SA312480720GA SA312469244GA SA312480577GA SA312480700GA SA312480807GA SA312490907GA SA312490908GA SA312490914GA SA311287160GA SA312018777GA SA312049273GA SA312049274GA SA312049380GA SA312049398GA SA312049404GA SA312049412GA SA312059501GA SA312059504GA SA312059505GA SA312059506GA SA312059508GA SA312059509GA SA312059529GA SA312059532GA SA312059578GA SA312059581GA SA312059584GA SA312059586GA SA312059587GA SA312059595GA SA312059599GA SA312069815GA SA312069817GA SA312069829GA SA312069847GA SA312069852GA SA312049382GA SA312069856GA SA312069858GA SA312069866GA SA312069867GA SA312069870GA SA312069871GA SA312069872GA SA312069873GA SA312069874GA SA312069876GA SA312069877GA SA312069878GA SA312069879GA SA312069880GA SA312069882GA SA312069883GA SA312069884GA SA312079888GA SA312079890GA SA312079891GA SA312079892GA SA312079894GA SA312079895GA SA312079897GA SA312079903GA SA312079907GA SA312079912GA SA312079914GA SA312079916GA SA312079921GA SA312079929GA SA312079930GA SA312079932GA SA312049425GA SA312049435GA SA312049439GA SA312049460GA SA312059477GA SA312059484GA SA312059487GA SA312059491GA SA312059492GA SA312059495GA SA312059497GA SA312059498GA SA312080658GA SA312080698GA SA312070302GA SA312070344GA SA312091056GA SA312069670GA SA312069680GA SA312069684GA SA312069686GA SA312069687GA SA312069726GA SA312069728GA SA312069730GA SA312069731GA SA312069732GA SA312069733GA SA312069736GA SA312069737GA SA312069739GA SA312069740GA SA312069745GA SA312069746GA SA312069750GA SA312069751GA SA312069753GA SA312069757GA SA312069758GA SA312069759GA SA312069762GA SA312069764GA SA312069765GA SA312069768GA SA312069769GA SA312069770GA SA312069772GA SA312091099GA SA312091102GA SA312091115GA SA312101458GA SA312069690GA SA312069696GA SA312069719GA SA312069776GA SA312069780GA SA312069782GA SA312069785GA SA312069805GA SA312069806GA SA312069808GA SA312069812GA SA312069814GA SA312090931GA SA312091216GA SA312111629GA SA312111640GA SA312111663GA SA312121701GA SA312121702GA SA312121741GA SA312121749GA SA312121750GA SA312121806GA SA312069832GA SA312101396GA SA312101401GA SA312101402GA SA312101404GA SA312101405GA SA312101407GA SA312101411GA SA312101412GA SA312101414GA SA312101421GA SA312101433GA SA312101438GA SA312121848GA SA312121809GA SA312121819GA SA312121871GA SA312121872GA SA312121875GA SA312121879GA SA312121895GA SA312121898GA SA312121963GA SA312121971GA SA312121978GA SA312121979GA SA312121986GA SA312121987GA SA312091291GA SA312091296GA SA312091299GA SA312091313GA SA312091314GA SA312091317GA SA312091321GA SA312091327GA SA312091328GA SA312091344GA SA312091350GA SA312091352GA SA312091353GA SA312091361GA SA312101369GA SA312101372GA SA312101374GA SA312101376GA SA312101377GA SA312101415GA SA312091305GA SA312101378GA SA312101379GA SA312101380GA SA312101390GA SA312101408GA SA312101410GA SA312111466GA SA312111470GA SA312111472GA SA312111484GA SA312111485GA SA312152319GA SA312152332GA SA312152335GA SA312152344GA SA312152347GA SA312152350GA SA312152361GA SA312152367GA SA307080539GA SA307080540GA SA307080539GA SA307080540GA SA307150658GA SA307150665GA SA307150666GA SA307080529GA SA307080475GA SA307080480GA SA307080482GA SA307080505GA SA307080542GA SA307110551GA SA307110552GA SA307110561GA SA307110568GA SA307110578GA SA307140617GA SA307140629GA SA307140630GA SA307140638GA SA307150650GA SA307150657GA SA307150662GA SA307150664GA SA307150668GA SA307150670GA SA307150673GA SA307150684GA SA307080460GA SA307110549GA SA307110553GA SA307110577GA SA307150667GA SA307150669GA SA307150674GA SA307150676GA SA307020390GA SA307070439GA SA307080471GA SA307080487GA SA307080496GA SA307080500GA SA307080507GA SA307080515GA SA307080534GA SA307130597GA SA307221054GA SA307221073GA SA307221054GA SA307221073GA SA307221143GA SA307231272GA SA307231288GA SA307231289GA SA307231308GA SA307231309GA SA307241524GA SA307231269GA SA307302073GA SA307322122GA SA307322124GA SA307322137GA SA307322138GA SA307322139GA SA307322140GA SA307322141GA SA307322142GA SA307322143GA SA307322144GA SA307322145GA SA307322148GA SA307322149GA SA307322151GA SA307322154GA SA307322159GA SA307322161GA SA307322162GA SA307322163GA SA307322164GA SA307322166GA SA307322186GA SA307322187GA SA307322190GA SA307322191GA SA307322192GA SA307322193GA SA307322198GA SA307322199GA SA307322201GA SA307322202GA SA307322204GA SA307050430GAMKE SA307231283GA SA307231328GA SA307291859GA SA307291860GA SA307291892GA SA307392831GA SA307392930GA SA307392946GA SA307392952GA SA307392966GA SA307392968GA SA307392981GA SA307392990GA SA307392992GA SA307392999GA SA307393002GA SA307393006GA SA307393027GA SA307393029GA SA307403066GA SA307403069GA SA307403080GA SA307403084GA SA307403130GA SA307403132GA SA307403133GA SA307403142GA SA307403144GA SA307403145GA SA307403147GA SA307403148GA SA307403151GA SA307403153GA SA307403154GA SA307392693GA SA307392712GA SA307392756GA SA307393030GA SA307393033GA SA307393034GA SA307393036GA SA307393037GA SA307403078GA SA307403081GA SA307403086GA SA307403088GA SA307403090GA SA307403091GA SA307403092GA SA307403093GA SA307403096GA SA307403097GA SA307403098GA SA307403099GA SA307403100GA SA307403102GA SA307403105GA SA307403107GA SA307403109GA SA307403111GA SA307403113GA SA307403114GA SA307403117GA SA307403118GA SA307403119GA SA307403121GA SA307403122GA SA307403125GA SA307403126GA SA307403155GA SA307403085GA SA307423160GA SA307382519GA SA307403076GA SA307403083GA SA307403094GA SA307443188GA SA307443213GA SA307443306GA SA307443321GA SA307443329GA SA307443334GA SA307443336GA SA307443337GA SA307443339GA SA307443350GA SA307443351GA SA307443352GA SA307443356GA SA307423139GA SA307443311GA SA307443319GA SA307443322GA SA307443338GA SA307443357GA SA307453422GA SA307453432GA SA307453444GA SA307453453GA SA307453464GA SA307453474GA SA307453483GA SA307453590GA SA307453458GA SA307453475GA SA307514445GA SA307514458GAMKE SA307514239GA SA307514240GA SA307514247GA SA307514283GA SA307514298GA SA307514299GA SA307514303GA SA307514304GA SA307514313GA SA307514314GA SA307514315GA SA307514318GA SA308084683GA SA308084676GA SA308084678GA SA308094935GAMKE SA308094946GA SA308094846GA SA308094885GA SA308094913GA SA308094919GA SA308094856GA SA308094882GA SA308094886GA SA308094901GA SA308094902GA SA308094904GA SA308094905GA SA308094906GA SA308094909GA SA308094910GA SA308094912GA SA308094917GA SA308094918GA SA308094920GA SA308105083GA SA308084668GA SA308084700GA SA308094892GA SA308094897GA SA308105084GA SA308105087GA SA308105089GA SA308105106GA SA308105113GA SA308105114GA SA308105117GA SA308105118GA SA308105142GA SA308105147GA SA308095062GA SA308135524GA SA307382445GA SA308115261GA SA308115282GA SA308125345GA SA308125351GA SA308135523GA SA308145796GA SA307443195GA SA308135461GA SA308135575GA SA308135632GA SA308135643GA SA308135644GA SA308135645GA SA308135650GA SA308135654GA SA308145710GA SA308145720GA SA308145736GA SA308145784GA SA308145791GA SA308145794GA SA308145802GA SA308145804GA SA308105096GA SA308105099GA SA308125407GA SA308125411GA SA308135470GA SA308135473GA SA308135564GA SA308135570GA SA308135578GA SA308135589GA SA308135610GA SA308135620GA SA308135623GA SA308135624GA SA308135628GA SA308135631GA SA308135634GA SA308135635GA SA308135636GA SA308135637GA SA308135638GA SA308135640GA SA308135642GA SA308135646GA SA308135647GA SA308135651GA SA308135653GA SA308145662GA SA308145702GA SA308145706GA SA308145709GA SA308145719GA SA308145721GA SA308145728GA SA308145733GA SA308145737GA SA308145752GA SA308145782GA SA308145783GA SA308145785GA SA308145788GA SA308145789GA SA308145790GA SA308145792GA SA308145793GA SA308145795GA SA308145798GA SA308145801GA SA308135554GA SA308135558GA SA308135560GA SA308135562GA SA308135583GA SA308135604GA SA308135607GA SA308135612GA SA308135626GA SA308135641GA SA308145713GA SA308145716GA SA308135574GA SA308145718GA SA308145735GA SA308145825GA SA308166055GA SA308186400GA SA308186421GA SA308186425GA SA308186441GA SA308196458GA SA308135515GA SA308145674GA SA308196548GA SA308196556GA SA308196567GA SA308196571GA SA308196509GA SA308357967GA SA308357976GA SA308357977GA SA308368111GA SA308368113GA SA308368115GA SA308368116GA SA308368117GA SA308368121GA SA308368122GA SA308368124GA SA308368141GA SA308145776GA SA308145811GA SA308378250GA SA308378274GA SA308388506GA SA308388540GA SA308398656GA SA308398663GA SA308398718GA SA308398739GA SA308398751GA SA308398764GA SA308398864GA SA308398869GA SA308398870GA SA308398713GAMKE SA308398662GA SA308398723GA SA308398735GA SA308398736GA SA308409197GA SA308409106GA SA308419300GA SA308429341GA SA308429357GA SA308155849GA SA308155935GA SA308449508GA SA308459682GA SA308459861GA SA308469886GA SA308479893GA SA308479898GA SA308479904GA SA308489951GA SA308489965GA SA308489982GA SA308481021GA SA308481024GA SA308481037GA SA308481064GA SA308481067GA SA308489935GA SA308489946GA SA308489967GA SA308491242GA SA308491175GA SA308501429GA SA309062326GA SA309062328GA SA309072354GA SA309072376GA SA308479903GA SA308479906GA SA308481001GA SA308481006GA SA308481007GA SA308481008GA SA308481013GA SA308481016GA SA308481017GA SA308481018GA SA308481022GA SA308481023GA SA308481027GA SA308481030GA SA308481033GA SA308481048GA SA308489928GA SA308489959GA SA308489961GA SA308489974GA SA308489990GA SA308489991GA SA309092473GA SA309092475GA SA309092476GA SA308084687GA SA308084716GA SA308084723GA SA308094899GA SA308094914GA SA309253857GA SA309263950GA SA309263955GA SA309294043GA SA309294052GA SA309294094GA SA309324472GA SA309324473GA SA309436319GA SA309476911GA SA309476915GA SA309476922GA SA309476926GA SA309476927GA SA309476931GA SA309446466GA SA309446489GA SA309446499GA SA309446518GA SA309456537GA SA309456544GA SA309456561GA SA309456722GA SA309456724GA SA309456725GA SA309456730GA SA309456734GA SA309466751GA SA309476936GA SA309476963GA SA309476987GA SA309497226GA SA309497274GA SA309497280GA SA309497281GA SA309497302GA SA309497303GA SA309497304GA SA309497305GA SA309497306GA SA309497307GA SA309497310GA SA309497311GA SA309477033GA SA309477045GA SA309477064GA SA309477085GA SA309477091GA SA309477092GA SA309477093GA SA309477098GA SA309477103GA SA309477116GA SA309477133GA SA309507336GA SA309507345GA SA309507347GA SA309507349GA SA309507351GA SA309507355GA SA309507359GA SA309507362GA SA309507363GA SA309507364GA SA309507374GA SA309507378GA SA309507384GA SA309507389GA SA309507391GA SA309507394GA SA309507398GA SA309507400GA SA309507401GA SA309507403GA SA309507405GA SA309507408GA SA309507409GA SA309507411GA SA309507415GA SA309507417GA SA309517468GA SA309517482GA SA309517485GA SA309517512GA SA309527953GA SA309446424GA SA309456648GA SA309456661GA SA309456678GA SA309456679GA SA309456686GA SA309517665GA SA309517680GA SA309517682GA SA309517699GA SA309517712GA SA309517717GA SA309517720GA SA309517725GA SA309517726GA SA309517729GA SA309517769GA SA309517779GA SA309517782GA SA309517783GA SA309517785GA SA309538064GA SA309538071GA SA309538077GA SA309538082GA SA309507337GA SA310038484GA SA310038488GA SA310038490GA SA310038493GA SA310038500GA SA310038506GA SA310048552GA SA310048559GA SA310048560GA SA310048572GA SA310048577GA SA310048580GA SA310048638GA SA310048645GA SA310048646GA SA310048678GA SA309497287GA SA310038466GA SA310028308GA SA310068845GA SA310068857GA SA310068872GA SA310068876GA SA310068881GA SA310068882GA SA310068890GA SA310068896GA SA310068897GA SA310068904GA SA310068908GA SA310068911GA SA310068930GA SA310068933GA SA310068945GA SA310068948GA SA310068950GA SA309517536GA SA309517774GA SA309517792GA SA309527990GA SA309538022GA SA310028150GA SA310028152GA SA310028157GA SA310028158GA SA310028161GA SA310028163GA SA310028164GA SA310028165GA SA310028167GA SA310028169GA SA310028177GA SA310028183GA SA310028202GA SA310028215GA SA310028216GA SA310028220GA SA310028240GA SA310028243GA SA310028255GA SA310028256GA SA310028263GA SA310028267GA SA310028286GA SA310028287GA SA310028298GA SA310028304GA SA310038350GA SA310038353GA SA310038355GA SA310038360GA SA310038363GA SA310038364GA SA310038365GA SA310038367GA SA310038368GA SA310038369GA SA310038376GA SA310038381GA SA310038394GA SA310038433GA SA310038434GA SA310038445GA SA310038448GA SA310038450GA SA310068846GA SA310068859GA SA310068863GA SA310068865GA SA310068871GA SA310068885GA SA310068900GA SA310068901GA SA310068916GA SA310068923GA SA310068936GA SA310068940GA SA310068942GA SA310139253GA SA310139081GA SA310149330GA SA310149331GA SA310149332GA SA310149337GA SA310139208GA SA310139221GA SA310068866GA SA310068889GA SA310068918GA SA310068934GA SA310219806GA SA310199682GA SA310199684GA SA310199691GA SA310199695GA SA310199701GA SA310199705GA SA310199707GA SA310199708GA SA310199709GA SA310199711GA SA310199721GA SA310199725GA SA310199728GA SA310199756GA SA310199757GA SA310199759GA SA310119016GA SA310199750GA SA310068827GA SA310139169GA SA310139187GA SA310139189GA SA310139192GA SA310139195GA SA310139197GA SA310139199GA SA310139211GA SA310139225GA SA310139226GA SA310139227GA SA310139229GA SA310139230GA SA310139236GA SA310139237GA SA310139239GA SA310139240GA SA310139241GA SA310159443GA SA310219842GA SA310259915GA SA310259917GA SA310259919GA SA310259921GA SA310259922GA SA310259924GA SA310259925GA SA310259926GA SA310259927GA SA310259930GA SA310259978GA SA310259983GA SA310259984GA SA310259990GA SA310260045GA SA310260048GA SA310260052GA SA310260058GA SA310260070GA SA310260076GA SA310260077GA SA310260220GA SA310260228GA SA310260234GA SA310260235GA SA310260238GA SA310260251GA SA310260476GA SA310260477GA SA310260478GA SA310260481GA SA310260483GA SA310260484GA SA310260525GA SA310260530GA SA310260535GA SA310260537GA SA310260538GA SA310260548GA SA310260549GA SA310260550GA SA310260551GA SA310169484GA SA310169504GA SA310260552GA SA310361689GA SA310372009GA SA310372028GA SA310372032GA SA310372034GA SA310361903GA SA310361913GA SA310382668GA SA310382677GA SA310382842GA SA310392855GA SA310392872GA SA310392891GA SA310403316GA SA310423774GA SA310475466GA SA310475482GA SA310475488GA SA310475519GA SA310475527GA SA310475529GA SA310475542GA SA310475543GA SA310475551GA SA310475586GA SA310475597GA SA310475612GA SA310475638GA SA310475644GA SA310475673GA SA310475702GA SA310485805GA SA310485809GA SA310485829GA SA310485838GA SA310485840GA SA310485861GA SA310485870GA SA310485883GA SA310485889GA SA310495910GA SA310495969GA SA310495971GA SA310496233GA SA310496237GA SA310496238GA SA310496484GA SA310403355GA SA310495895GA SA310495898GA SA310495902GA SA310495905GA SA310495907GA SA310495914GA SA310495915GA SA310495917GA SA310495937GA SA310495944GA SA310495947GA SA310495948GA SA310495949GA SA310495951GA SA310495952GA SA310495958GA SA310495960GA SA310495961GA SA310495972GA SA310495974GA SA310495975GA SA310496142GA SA310496274GA SA310496284GA SA310496290GA SA310496292GA SA310496294GA SA310496299GA SA310496304GA SA310496332GA SA310496347GA SA310496352GA SA310496353GA SA310496354GA SA310496360GA SA310496361GA SA310496364GA SA310496373GA SA310496464GA SA310496476GA SA310496504GA SA310496526GA SA310496531GA SA310496532GA SA310496533GA SA310434062GA SA310475477GA SA310475490GA SA310475526GA SA310475528GA SA310475589GA SA310475596GA SA310496132GA SA310496154GA SA310496208GA SA310496348GA SA310496367GA SA310496523GA SA310496289GA SA310496334GA SA310496377GA SA310496379GA SA310496400GA SA310496403GA SA310496413GA SA310496430GA SA310496507GA SA310496516GA SA310496619GA SA310496635GA SA310496639GA SA310496640GA SA310496647GA SA310496652GA SA310506660GA SA310506698GA SA310506720GA SA310506730GA SA310506758GA SA310506764GA SA310506769GA SA310506777GA SA310506779GA SA310506794GA SA310506797GA SA310506804GA SA310506811GA SA310506866GA SA310506898GA SA310507159GA SA310507180GA SA310507189GA SA310507194GA SA310507198GA SA310507208GA SA310507502GA SA310496147GA SA310496342GA SA310496359GA SA310496502GA SA310507026GA SA310507035GA SA310507039GA SA310507091GA SA310517734GA SA310517743GA SA310518221GA SA310518226GA SA310518227GA SA310518231GA SA310518309GA SA310506668GA SA310506721GA SA310506740GA SA310506751GA SA310506856GA SA310506858GA SA310517710GA SA310517787GA SA310517833GA SA310507434GA SA310507438GA SA310507467GA SA310507468GA SA310507557GA SA310517642GA SA310517643GA SA310517644GA SA310517648GA SA310517653GA SA310517684GA SA310518145GA SA310518153GA SA310518167GA SA310518171GA SA310518172GA SA310518174GA SA310518175GA SA310518178GA SA310506709GA SA310507087GA SA310517822GA SA310517834GA SA310517835GA SA310517843GA SA310517879GA SA310517984GA SA310517569GA SA310517733GA SA310517781GA SA310517819GA SA310517820GA SA310517829GA SA310517989GA SA310517992GA SA310517999GA SA310518006GA SA310518008GA SA310518011GA SA310528877GA SA310496327GA SA310496337GA SA310496345GA SA310518616GA SA310528851GA SA310528857GA SA310528873GA SA310528893GA SA310528900GA SA310528970GA SA310528971GA SA310528973GA SA310528978GA SA310528983GA SA310528986GA SA310529012GA SA310529146GA SA310529149GA SA311039369GA SA311039400GA SA311039411GA SA311039415GA SA310496335GA SA310528866GA SA310528874GA SA310528885GA SA311039394GA SA311039398GA SA311039407GA SA311039471GA SA310393069GA SA310444495GA SA310465377GA SA310260394GA SA310382363GA SA310382833GA SA310393215GA SA310403347GA SA310465309GA SA310465319GA SA310518260GA SA310518354GA SA310518425GA SA310518428GA SA310518497GA SA310518500GA SA310518504GA SA310518514GA SA310518517GA SA310518518GA SA310518520GA SA310518524GA SA310518562GA SA310528784GA SA310528786GA SA310528788GA SA310528789GA SA310528796GA SA310528797GA SA310528811GA SA310528841GA SA311039503GA SA311039506GA SA311039512GA SA311039514GA SA311039515GA SA311039517GA SA311039522GA SA310403314GA SA311060018GA SA311060039GA SA311060042GA SA311060078GA SA311060214GA SA311069646GA SA311069670GA SA311069671GA SA311069693GA SA311069743GA SA311069775GA SA311069776GA SA311069778GA SA311069791GA SA311069818GA SA311069852GA SA311060155GA SA311060171GA SA311060235GA SA311060238GA SA311060240GA SA311060254GA SA311060265GA SA311069719GA SA311060255GA SA311060270GA SA311060271GA SA311060297GA SA311060298GA SA311060303GA SA311060308GA SA311060343GA SA311060385GA SA311060397GA SA310528824GA SA311060087GA SA311060094GA SA311060109GA SA311060068GA SA311060089GA SA311060107GA SA311080678GA SA311080670GA SA311080679GA SA311101297GA SA311090849GA SA311090860GA SA311090870GA SA311090877GA SA311101241GA SA311101278GA SA311091033GA SA311091047GA SA311080615GA SA311101077GA SA311101080GA SA311101085GA SA311122199GA SA311101140GA SA311101141GA SA311132475GA SA311121913GA SA311142968GA SA311142953GA SA311121781GA SA311121786GA SA311121805GA SA311122142GA SA311122164GA SA311122100GA SA311069783GA SA311069809GA SA311069812GA SA311090944GA SA311091003GA SA311091006GA SA311101094GA SA311101105GA SA311101106GA SA311121753GA SA311121970GA SA311121977GA SA311121985GA SA311132373GA SA311132379GA SA311132388GA SA311132412GA SA311132474GA SA311132478GA SA311132484GA SA311132603GA SA311121797GA SA311121842GA SA311121924GA SA311121951GA SA311121978GA SA311121984GA SA311121991GA SA311121996GA SA311122000GA SA311122077GA SA311132424GA SA311132456GA SA311132469GA SA311132530GA SA311132532GA SA311132535GA SA311132536GA SA311132541GA SA311132547GA SA311132548GA SA311132615GA SA311132641GA SA311163993GA SA311163870GA SA311163908GA SA311163946GA SA311163961GA SA311163973GA SA311163974GA SA311163979GA SA311153051GA SA311174163GA SA311174197GA SA311163929GA SA311174171GA SA311174173GA SA311174174GA SA311174175GA SA311174183GA SA311174193GA SA311174204GA SA311184348GA SA311184546GA SA311184551GA SA311195075GA SA311174190GA SA311184375GA SA311184384GA SA311184388GA SA311184526GA SA311184573GA SA311184583GA SA311184587GA SA311184607GA SA311225982GA SA311226001GA SA311236071GA SA311236083GA SA311246389GA SA311246533GA SA311246541GA SA311246569GA SA311246574GA SA311246575GA SA311246577GA SA311246578GA SA311246580GA SA311246582GA SA311142694GA SA311142725GA SA311142731GA SA311142736GA SA311143010GA SA311143012GA SA311143022GA SA311153331GA SA311163429GA SA311163454GA SA311163455GA SA311163488GA SA311163532GA SA311163551GA SA311163594GA SA311163609GA SA311163697GA SA311163825GA SA311163836GA SA311163843GA SA311163850GA SA311184750GA SA311184765GA SA311184844GA SA311184857GA SA311205396GA SA311256844GA SA311256866GA SA311205306GA SA311205323GA SA311163556GA SA311163569GA SA311163571GA SA311163573GA SA311163575GA SA311163578GA SA311163590GA SA311163593GA SA311163600GA SA311163603GA SA311163604GA SA311163615GA SA311163616GA SA311163618GA SA311163619GA SA311163622GA SA311163631GA SA311163635GA SA311163637GA SA311163643GA SA311163644GA SA311163652GA SA311163654GA SA311163701GA SA311163725GA SA311163785GA SA311163786GA SA311163788GA SA311163793GA SA311163812GA SA311163813GA SA311163828GA SA311163851GA SA311163855GA SA311174001GA SA311174020GA SA311174030GA SA311174034GA SA311174039GA SA311174048GA SA311174050GA SA311174054GA SA311174061GA SA311174063GA SA311174068GA SA311174069GA SA311174070GA SA311174074GA SA311174079GA SA311174082GA SA311174086GA SA311174090GA SA311174092GA SA311174094GA SA311174099GA SA311174101GA SA311174102GA SA311174104GA SA311174105GA SA311174106GA SA311174109GA SA311174110GA SA311174111GA SA311174112GA SA311174113GA SA311174114GA SA311174115GA SA311174116GA SA311174117GA SA311174118GA SA311174124GA SA311174129GA SA311174135GA SA311174136GA SA311174137GA SA311174138GA SA311174139GA SA311174145GA SA311174146GA SA311174148GA SA311174154GA SA311174155GA SA311174157GA SA311184684GA SA311184760GA SA311184761GA SA311184810GA SA311266928GA SA311080677GA SA311132480GA SA311132503GA SA311163570GA SA311163596GA SA311163599GA SA311163605GA SA311163632GA SA311163634GA SA311163646GA SA311163653GA SA311163796GA SA311163807GA SA311266912GA SA311153041GA SA311163457GA SA311163539GA SA311163541GA SA311163557GA SA311163611GA SA311163651GA SA311163703GA SA311163706GA SA311163708GA SA311163709GA SA311163721GA SA311163727GA SA311163730GA SA311163731GA SA311163779GA SA311163787GA SA311163795GA SA311163797GA SA311163798GA SA311163802GA SA311163804GA SA311163810GA SA311163816GA SA311163819GA SA311163822GA SA311163826GA SA311163827GA SA311163831GA SA311163834GA SA311163842GA SA311173998GA SA311174008GA SA311174010GA SA311174011GA SA311174012GA SA311174013GA SA311174014GA SA311174022GA SA311174023GA SA311174025GA SA311174026GA SA311174028GA SA311174029GA SA311174036GA SA311174037GA SA311174042GA SA311174043GA SA311174046GA SA311174055GA SA311174057GA SA311317653GA SA311317654GA SA311317659GA SA311317714GA SA311163577GA SA311163608GA SA311163777GA SA311194904GA SA311194933GA SA311194934GA SA311194978GA SA311194984GA SA311194993GA SA311195004GA SA311205564GA SA311327826GA SA311327890GA SA310517784GA SA310529078GA SA311039368GA SA311039374GA SA311039376GA SA311039401GA SA311039406GA SA311039428GA SA311039431GA SA311039436GA SA311039443GA SA311039444GA SA311039444GAMKE SA311039445GA SA311039456GA SA311039462GA SA311039466GA SA311039476GA SA311039574GA SA311060228GA SA311142939GA SA311142977GA SA311215686GA SA311236070GA SA311225954GA SA311236004GA SA311236018GA SA311236022GA SA311236026GA SA311236031GA SA311236039GA SA311236045GA SA311236048GA SA311236053GA SA311236057GA SA311236123GA SA311236156GA SA311236157GA SA311236158GA SA311236162GA SA311236175GA SA311236176GA SA311236179GA SA311236180GA SA311236183GA SA311236192GA SA311236193GA SA311236194GA SA311236195GA SA311236196GA SA311236197GA SA311236198GA SA311236199GA SA311236201GA SA311236203GA SA311236207GA SA311236209GA SA311236215GA SA311236219GA SA311236221GA SA311236223GA SA311236225GA SA311236229GA SA311236257GA SA311236259GA SA311266914GA SA311266923GA SA311266942GA SA311287098GA SA311287176GA SA311307457GA SA311317682GA SA311317750GA SA311327847GA SA311337980GA SA311337985GA SA311337986GA SA311337989GA SA311337997GA SA311337999GA SA311338004GA SA311428013GA SA311428026GA SA311428028GA SA311448175GA SA311448176GA SA311448180GA SA311448196GA SA311448197GA SA311448198GA SA311448200GA SA311448203GA SA311448204GA SA311448206GA SA311448207GA SA311448208GA SA311448209GA SA311448210GA SA311448211GA SA311458267GA SA311458268GA SA311458269GA SA311458270GA SA311458271GA SA311458273GA SA311458276GA SA311458277GA SA311458278GA SA311458280GA SA311458281GA SA311458283GA SA311458284GA SA311458285GA SA311458286GA SA311458287GA SA311458288GA SA311458289GA SA311458290GA SA311458293GA SA311458294GA SA311458295GA SA311458318GA SA311458320GA SA311458323GA SA311458324GA SA311458325GA SA311458326GA SA311458327GA SA311458328GA SA311458330GA SA311458332GA SA311458337GA SA311458339GA SA311236345GA SA311246471GA SA311246473GA SA311287032GA SA311287092GA SA311297224GA SA310444298GA SA311205447GA SA311205489GA SA311307416GA SA311307451GA SA311317665GA SA311327894GA SA311142702GA SA311225948GA SA311236011GA SA311337975GA SA311337976GA SA311337981GA SA311428081GA SA311448172GA SA311448174GA SA311448193GA SA311448199GA SA311448201GA SA311458274GA SA311458291GA SA311458296GA SA311458315GA SA311458329GA SA311458335GA SA311458336GA SA311468378GA SA311468379GA SA311468385GA SA311468405GA SA311317721GA SA311317726GA SA311317731GA SA311256850GA SA311256842GA SA310528976GA SA311508571GA SA311508579GA SA311508580GA SA311508583GA SA311508594GA SA311508595GA SA311508601GA SA311508602GA SA311508603GA SA311508604GA SA311508605GA SA311508606GA SA311508607GA SA311508608GA SA311508609GA SA311508610GA SA311508611GA SA311508612GA SA311508613GA SA311508615GA SA311508617GA SA311508618GA SA311508620GA SA311508621GA SA311508622GA SA311508628GA SA311508629GA SA311508630GA SA311508631GA SA311508632GA SA311508633GA SA311508634GA SA311508635GA SA311508638GA SA311508639GA SA311508637GA SA311508653GA SA311508657GA SA311508637GA SA311508653GA SA311508657GA SA311327877GA SA311508619GA SA311518691GA SA311518696GA SA310310842GA SA310310843GA SA310310847GA SA310310850GA SA310310851GA SA310310856GA SA310310860GA SA310310863GA SA310310879GA SA310310889GA SA310310906GA SA310331227GA SA310331236GA SA310331241GA SA310331252GA SA310331254GA SA310331258GA SA310372000GA SA310372002GA SA310372006GA SA310372011GA SA310372013GA SA310372014GA SA310372019GA SA310372020GA SA310372033GA SA310372042GA SA310372045GA SA310372046GA SA310372047GA SA310372050GA SA310372051GA SA310331225GA SA310331229GA SA310331233GA SA310331239GA SA310331244GA SA310331246GA SA310331262GA SA310331269GA SA310331275GA SA310331282GA SA310331286GA SA310331290GA SA310331291GA SA310341407GA SA310341409GA SA310341416GA SA310341418GA SA310341421GA SA310341423GA SA310341433GA SA310341452GA SA310321052GA SA310361600GA SA310361646GA SA310361905GA SA310372031GA SA310372038GA SA310372041GA SA310372056GA SA310372057GA SA310372141GA SA310372143GA SA310372148GA SA310331104GA SA310372187GA SA310372200GA SA310382357GA SA310382445GA SA310382505GA SA312242984GA SA312284291GA SA312101437GA SA312253222GA SA312253230GA SA312344858GA SA312344867GA SA312344869GA SA312344876GA SA312344879GA SA312344880GA SA312344882GA SA312344885GA SA312344888GA SA312344889GA SA312344893GA SA312344899GA SA312344906GA SA312344907GA SA312344918GA SA312344920GA SA312344926GA SA312344927GA SA312344962GA SA312344966GA SA312355380GA SA312355385GA SA312355386GA SA312355395GA SA312355436GA SA312355441GA SA312355443GA SA312355444GA SA312355445GA SA312355450GA SA312365564GA SA312365584GA SA312365617GA SA312365624GA SA312365626GA SA312365629GA SA312365637GA SA312365645GA SA312365646GA SA312376284GA SA312376308GA SA312386312GA SA312386338GA SA312386366GA SA312345041GA SA312386315GA SA312386317GA SA312386329GA SA312386332GA SA312386334GA SA312386370GA SA312386672GA SA312386674GA SA312386675GA SA312386688GA SA312386690GA SA312386691GA SA312386697GA SA312386698GA SA312386699GA SA312396976GA SA312396978GA SA312396979GA SA312396982GA SA312397151GA SA312397152GA SA312397153GA SA312397154GA SA312397156GA SA312397162GA SA312397166GA SA312397179GA SA312397188GA SA312397189GA SA312397192GA SA312397193GA SA312397194GA SA312397199GA SA312397200GA SA312397208GA SA312397010GA SA312397015GA SA312397170GA SA312396984GA SA312396998GA SA312457771GA SA312457785GA SA312457801GA SA312457806GA SA312457808GA SA312469155GA SA312479829GA SA312479833GA SA312447459GA SA312457707GA SA312470077GA SA312479853GA SA312479967GA SA312480730GA SA312480731GA SA312491354GA SA312491359GA SA312491370GA SA312491406GA SA312491409GA SA312491412GA SA312491413GA SA312491414GA SA312491419GA SA312491421GA SA312491424GA SA312491428GA SA312491429GA SA312491430GA SA312491433GA SA312491438GA SA312491442GA SA312491444GA SA312491450GA SA312491452GA SA312491454GA SA312121901GA SA312121945GA SA312121954GA SA312122006GA SA312122019GA SA312122023GA SA312122024GA SA312122032GA SA312122035GA SA312122049GA SA312122061GA SA312122072GA SA312132120GA SA311297233GA SA311297235GA SA311317757GA SA311317762GA SA311142682GA SA311142692GA SA311142728GA SA311142734GA SA311142735GA SA311317604GA SA311317601GA SA311317636GA SA312029067GA SA312039075GA SA312039079GA SA312039095GA SA311246481GA SA311307573GA SA311327844GA SA311327854GA SA311327859GA SA311327860GA SA311327862GA SA311458275GA SA311458310GA SA311458317GA SA311458322GA SA311468423GA SA311468434GA SA311488452GA SA311488461GA SA311488489GA SA312018821GA SA311142685GA SA312049445GA SA312059624GA SA312059628GA SA312069729GA SA312069773GA SA312069699GA SA312069702GA SA312069703GA SA312069704GA SA312069708GA SA312069710GA SA312069811GA SA312070138GA SA312080632GA SA312080642GA SA312080387GA SA312080404GA SA312080406GA SA312080413GA SA312080414GA SA312080474GA SA312080586GA SA312080588GA SA312080619GA SA312080637GA SA312080644GA SA312080650GA SA312080654GA SA312080665GA SA312080669GA SA312080672GA SA312080676GA SA312080697GA SA312080703GA SA312080738GA SA312069789GA SA312070077GA SA312070127GA SA312070128GA SA312070144GA SA312070160GA SA312070181GA SA312070185GA SA312070189GA SA312070192GA SA312070203GA SA312070205GA SA312070207GA SA312070213GA SA312070218GA SA312070223GA SA312070224GA SA312070228GA SA312070229GA SA312070234GA SA312070235GA SA312070238GA SA312070239GA SA312070240GA SA312069706GA SA312069724GA SA312069781GA SA312069783GA SA312069786GA SA312069797GA SA312069798GA SA312069807GA SA312069809GA SA312069813GA SA312070194GA SA312070222GA SA312070316GA SA312070337GA SA312070352GA SA312070358GA SA312070362GA SA312070379GA SA312070383GA SA312079965GA SA312091255GA SA312121776GA SA312121991GA SA312121961GA SA312121965GA SA312121968GA SA312121969GA SA312121970GA SA312122034GA SA312122045GA SA312122047GA SA312122053GA SA312122066GA SA312122070GA SA312122073GA SA312122074GA SA312122075GA SA312122077GA SA312122078GA SA312122088GA SA312132096GA SA312132099GA SA312132111GA SA312132112GA SA312132169GA SA312121890GA SA307231353GA SA307030410GA SA307110555GA SA307030410GA SA307110555GA SA307221099GA SA307332220GA SA307332242GA SA307342310GA SA307342326GA SA307342327GA SA307352371GA SA307362388GA SA307382455GA SA308084726GA SA308388548GA SA308398890GA SA308206679GA SA308429360GA SA309122641GA SA309122642GA SA308115268GA SA312091324GA SA312396999GA SA309446504GA SA309446509GA SA309466789GA SA310444314GA SA310444498GA SA310517709GA SA310528892GA SA311039379GA SA311039461GA SA311070580GA SA311090767GA SA311090823GA SA311090828GA SA311090857GA SA311101252GA SA311101254GA SA311101257GA SA311101260GA SA311101248GA SA311132575GA SA311142751GA SA311132551GA SA311163885GA SA311184600GA SA311184398GA SA311184514GA SA311184575GA SA311184593GA SA311184595GA SA311163912GA SA311163874GA SA311163921GA SA311184372GA SA311184534GA SA311297357GA SA311468381GA SA311307536GA SA311468392GA SA311317722GA SA311508562GA SA310290620GA SA310341413GA SA312152542GA SA312253362GA SA312294493GA SA312376037GA SA312386661GA SA312386645GA SA312386651GA SA312458186GA SA312458194GA SA312458199GA SA312458223GA SA312458252GA SA312458454GA SA312469295GA SA312480853GA SA311246467GA SA312018872GA SA312049328GA SA312049351GA SA312049340GA SA312049355GA SA312059470GA SA312059545GA SA312080493GA SA312080516GA SA312080633GA SA312080682GA SA312121943GA SA307291937GA SA307291939GA SA307291954GA SA307301990GA SA307301995GA SA308074611GA SA308084642GA SA308084666GA SA308084727GA SA308084728GA SA308084731GA SA308084732GA SA308084733GA SA308084734GA SA308084746GA SA308094804GA SA308094806GA SA308084631GA SA308155877GA SA308155883GA SA308155924GA SA308155941GA SA308155962GA SA308165998GA SA308166003GA SA308459795GA SA309385675GA SA309395695GA SA309395702GA SA309395709GA SA309395740GA SA309395804GA SA309456563GA SA309466736GA SA309466743GA SA309466745GA SA309466747GA SA309466750GA SA309466774GA SA309466794GA SA310310963GA SA310506742GA SA310507498GA SA310517578GA SA310517700GA SA310517703GA SA310517715GA SA310517759GA SA310517775GA SA310517788GA SA312294465GA SA312294570GA SA312375912GA SA312375914GA SA312375927GA SA312375983GA SA312375984GA SA312375985GA SA312375995GA SA312375999GA SA312376007GA SA312376009GA SA312376012GA SA312376014GA SA312376028GA SA312376052GA SA312458107GA SA312470001GA SA312470002GA SA312470003GA SA312470007GA SA312470008GA SA312470016GA SA312470065GA SA312470069GA SA312470076GA SA312470090GA SA312479690GA SA312479693GA SA312479694GA SA312479750GA SA312479777GA SA312479949GA SA312479952GA SA312479958GA SA312479959GA SA312479970GA SA312480094GA SA312480100GA SA312480108GA SA312049330GA SA312049335GA SA312049364GA SA312080595GA SA312080611GA SA312121919GA SA307241688GA SA308388612GA SA308408909GA SA308408914GA SA308408923GA SA308408932GA SA308409026GA SA308408927GA SA308408931GA SA308408952GA SA308408956GA SA308409024GA SA308409035GA SA308409040GA SA308409061GA SA308084641GA SA308084647GA SA308084657GA SA308084764GA SA308084767GA SA308084768GA SA308084772GA SA308084780GA SA308084781GA SA308084782GA SA308094834GA SA307110546GA SA307110570GA SA307110573GA SA307150661GA SA307220885GA SA307231254GA SA307241687GA SA307241694GA SA307392687GA SA307392993GA SA307403056GA SA307403064GA SA307443317GA SA307443323GA SA308074582GA SA308074584GA SA308074599GA SA308084714GA SA308094810GA SA308094879GA SA308094840GA SA308094848GA SA308094854GA SA308094883GA SA308094887GA SA308094888GA SA308094889GA SA308094890GA SA308094893GA SA308094895GA SA308094896GA SA308094898GA SA308094900GA SA308094907GA SA308094916GA SA308115302GA SA308115304GA SA308115307GA SA308317302GA SA308327314GA SA308327526GA SA308327545GA SA308327546GA SA308337570GA SA308337600GA SA308337634GA SA308337688GA SA308337697GA SA308408998GA SA308408999GA SA308409101GA SA308409102GA SA308409105GA SA309072353GA SA307403077GA SA309042188GA SA309052215GA SA309052231GA SA309052234GA SA309052235GA SA309344881GA SA309446494GA SA310048557GA SA310048595GA SA310139247GA SA310199561GA SA310290621GA SA310290623GA SA310290625GA SA310290628GA SA310290629GA SA310290631GA SA310290633GA SA310290634GA SA310290638GA SA310290639GA SA310290640GA SA310290641GA SA310290645GA SA310290646GA SA310290647GA SA310290648GA SA310290662GA SA310290663GA SA310290664GA SA310290666GA SA310290669GA SA310290670GA SA310290671GA SA310290672GA SA310290674GA SA310290675GA SA310290676GA SA310290677GA SA310290678GA SA310290680GA SA310290681GA SA310290682GA SA310290683GA SA310290684GA SA310290685GA SA310290687GA SA310290688GA SA310290689GA SA310290690GA SA310290691GA SA310290692GA SA310290694GA SA310290695GA SA310290697GA SA310290698GA SA310290701GA SA310290652GA SA310423797GA SA310434083GA SA310434143GA SA310434162GA SA310361959GA SA310382398GA SA310382502GA SA310423595GA SA310423676GA SA310423695GA SA310434209GA SA310454985GA SA310455035GA SA310465321GA SA310465322GA SA310465326GA SA310465327GA SA310465328GA SA310465330GA SA310465331GA SA310465332GA SA310465333GA SA310465334GA SA310465338GA SA310465339GA SA310465342GA SA310465343GA SA310465346GA SA310465347GA SA310465380GA SA310465381GA SA310465397GA SA310465398GA SA310465403GA SA310465410GA SA310465411GA SA310465412GA SA310465420GA SA310465427GA SA310465431GA SA310465432GA SA310465433GA SA310465435GA SA310465438GA SA310465439GA SA310465441GA SA310465445GA SA310465446GA SA310465323GA SA310465325GA SA310465329GA SA310465341GA SA310465419GA SA310465443GA SA310475487GA SA310475515GA SA310475516GA SA310475518GA SA310475544GA SA310475545GA SA310475554GA SA310475572GA SA310475599GA SA310475601GA SA310475603GA SA310475604GA SA310475606GA SA310475607GA SA310475614GA SA310475619GA SA310475621GA SA310475622GA SA310475623GA SA310475624GA SA310475625GA SA310475627GA SA310475633GA SA310475635GA SA310475636GA SA310475640GA SA310475641GA SA310475642GA SA310475643GA SA310475645GA SA310475646GA SA310475647GA SA310475648GA SA310475650GA SA310475651GA SA310475652GA SA310475653GA SA310475654GA SA310475655GA SA310475656GA SA310475657GA SA310475658GA SA310475660GA SA310475661GA SA310475662GA SA310475664GA SA310475665GA SA310475666GA SA310475667GA SA310475668GA SA310475669GA SA310475670GA SA310475671GA SA310475672GA SA310475674GA SA310475675GA SA310475676GA SA310475677GA SA310475678GA SA310475679GA SA310475680GA SA310475683GA SA310475688GA SA310475689GA SA310475690GA SA310475691GA SA310475692GA SA310475693GA SA310475694GA SA310475699GA SA310475708GA SA310475713GA SA310475715GA SA310475716GA SA310475720GA SA310475721GA SA310475722GA SA310475723GA SA310475724GA SA310475725GA SA310475727GA SA310475729GA SA310475731GA SA310475735GA SA310475736GA SA310475737GA SA310475738GA SA310475739GA SA310475740GA SA310475741GA SA310475742GA SA310475743GA SA310475744GA SA310475745GA SA310475746GA SA310475747GA SA310475748GA SA310475749GA SA310475750GA SA310475751GA SA310475752GA SA310475753GA SA310475754GA SA310485755GA SA310485756GA SA310485757GA SA310485758GA SA310485759GA SA310485760GA SA310485762GA SA310485764GA SA310485765GA SA310485767GA SA310485771GA SA310485773GA SA310485774GA SA310485775GA SA310485776GA SA310485777GA SA310485778GA SA310485779GA SA310485780GA SA310485781GA SA310485783GA SA310485784GA SA310485787GA SA310485797GA SA310485798GA SA310485799GA SA310485802GA SA310485815GA SA310485821GA SA310485826GA SA310485835GA SA310485841GA SA310485844GA SA310485846GA SA310485850GA SA310485857GA SA310485860GA SA310485868GA SA310485869GA SA310485871GA SA310485874GA SA310485878GA SA310485884GA SA310485885GA SA310485888GA SA310403342GA SA310403373GA SA310403375GA SA310403397GA SA310403404GA SA310403408GA SA310403416GA SA310403417GA SA310403424GA SA310403444GA SA310423757GA SA310423776GA SA310423786GA SA310423789GA SA310423810GA SA310423815GA SA310423817GA SA310423890GA SA310434135GA SA310434156GA SA310434172GA SA310434187GA SA310496138GA SA310496149GA SA310496151GA SA310496206GA SA310496207GA SA310496357GA SA310496616GA SA310496620GA SA310496629GA SA310496300GA SA310496346GA SA310496355GA SA310496508GA SA310507134GA SA310507254GA SA310507465GA SA310507501GA SA310517714GA SA310517726GA SA310517730GA SA310517735GA SA310496155GA SA310496645GA SA310507105GA SA310507515GA SA310507545GA SA311039381GA SA311039386GA SA311039395GA SA311039396GA SA311039397GA SA311039402GA SA311039403GA SA311101282GA SA311101178GA SA311101181GA SA311101189GA SA311101213GA SA311101243GA SA311101386GA SA311101389GA SA311101408GA SA311101410GA SA311101418GA SA311101384GA SA311101393GA SA311101394GA SA311101396GA SA311101420GA SA311174169GA SA311174184GA SA311174188GA SA311174218GA SA311174220GA SA311174228GA SA311174234GA SA311184517GA SA311184538GA SA311184539GA SA311184547GA SA311184562GA SA311184552GA SA311184589GA SA311184615GA SA311184617GA SA311039491GA SA311060147GA SA311060173GA SA311060204GA SA311060280GA SA311327872GA SA311039460GA SA311039482GA SA311142948GA SA311327822GA SA311327868GA SA311327886GA SA311327892GA SA310361837GA SA310361854GA SA310361860GA SA310361861GA SA310361862GA SA310361878GA SA310361881GA SA310361882GA SA310361818GA SA310361819GA SA310361827GA SA310361842GA SA310361845GA SA310361846GA SA310361847GA SA310361848GA SA310361851GA SA310361855GA SA310361858GA SA310361865GA SA310361866GA SA310361867GA SA310361868GA SA310361869GA SA310361870GA SA310361871GA SA310361884GA SA310372114GA SA310372122GA SA310372125GA SA310372186GA SA310372188GA SA310372190GA SA310372192GA SA310372206GA SA310361943GA SA310372285GA SA310372292GA SA310372298GA SA310341389GA SA310361754GA SA310361786GA SA310361875GA SA310372091GA SA310372116GA SA310372123GA SA310372135GA SA310372137GA SA310372164GA SA310372177GA SA310372178GA SA310372183GA SA310382316GA SA310382317GA SA310382323GA SA310382339GA SA310382340GA SA310382342GA SA310382366GA SA310382369GA SA310382374GA SA310382528GA SA310382560GA SA310382569GA SA310382579GA SA310382586GA SA310382679GA SA310382680GA SA310382687GA SA310382699GA SA310382701GA SA310382771GA SA310382774GA SA310382775GA SA310382776GA SA310382778GA SA310382780GA SA310382787GA SA310382793GA SA310382795GA SA310382796GA SA310382797GA SA310382801GA SA310382802GA SA310382806GA SA310382808GA SA310382809GA SA310382816GA SA310382818GA SA310382819GA SA310382823GA SA310382824GA SA310382825GA SA310382827GA SA310382828GA SA310382829GA SA310382831GA SA310382834GA SA310382835GA SA310382844GA SA310382848GA SA310392859GA SA310392874GA SA310392876GA SA312294383GA SA312355455GA SA312355464GA SA312365778GA SA312365783GA SA312365785GA SA312365786GA SA312365788GA SA312365795GA SA312365808GA SA312365826GA SA312457693GA SA312457721GA SA312457726GA SA312468932GA SA312468933GA SA312468940GA SA312468957GA SA312468973GA SA312468983GA SA312468986GA SA312469113GA SA312469114GA SA312469122GA SA312469130GA SA312469136GA SA312469138GA SA312469144GA SA312469145GA SA312469146GA SA312469152GA SA312469158GA SA312469160GA SA312469175GA SA312469183GA SA312469185GA SA312469187GA SA312469195GA SA312469196GA SA312469200GA SA312469201GA SA312469202GA SA312469203GA SA312469204GA SA312469205GA SA312469210GA SA312469211GA SA312457848GA SA312458244GA SA312458404GA SA312468782GA SA312480727GA SA312491319GA SA312491320GA SA312491325GA SA312491327GA SA312491328GA SA312491332GA SA312491340GA SA312491344GA SA312491351GA SA312491360GA SA312491368GA SA312491374GA SA312491407GA SA312491408GA SA312491410GA SA312491422GA SA312491423GA SA312491425GA SA312491431GA SA312491432GA SA312491435GA SA312491436GA SA312491437GA SA312491439GA SA312018771GA SA311307540GA SA311307554GA SA311307562GA SA311317617GA SA311317622GA SA311317633GA SA311317675GA SA312039150GA SA312039167GA SA312039171GA SA312039182GA SA312039186GA SA312039187GA SA312039189GA SA312039192GA SA312039200GA SA312039203GA SA312039209GA SA312039210GA SA312039212GA SA309354975GA SA311317689GA SA311428015GA SA311428050GA SA311428084GA SA311428088GA SA312039178GA SA312039184GA SA312039191GA SA312039195GA SA312039206GA SA312039208GA SA312039215GA SA312039224GA SA312039233GA SA312049245GA SA312049257GA SA312049431GA SA312049450GA SA312049319GA SA312049440GA SA312059544GA SA312059546GA SA312059560GA SA312059567GA SA312059568GA SA312059605GA SA312059606GA SA312059607GA SA312059621GA SA312059622GA SA312059623GA SA312059626GA SA312059627GA SA312059634GA SA312049327GA SA312069644GA SA312069647GA SA312069649GA SA312069650GA SA312069660GA SA312069662GA SA312080521GA SA312080524GA SA312080569GA SA312080581GA SA312080593GA SA312080606GA SA312080609GA SA312080592GA SA312080602GA SA312080614GA SA312080615GA SA312091105GA SA312091098GA SA312091097GA SA312121791GA SA312121830GA SA312121866GA SA312121878GA SA312121914GA SA312121916GA SA312121993GA SA312091285GA SA312091295GA SA312091301GA SA312091303GA SA312091357GA SA312059531GA SA312070180GA SA312091015GA SA312091235GA SA312111539GA SA312122015GA SA312122017GA SA312122033GA SA312091354GA SA312121900GA SA312132129GA SA312152412GA SA312121931GA SA312132104GA SA312070252GA SA312111460GA SA312111467GA SA312091112GA SA312122044GA SA312152394GA SA312121826GA SA312192660GA SA312192714GA SA312192717GA SA308125387GA SA308196576GA SA309183249GA SA309193261GA SA309193270GA SA309193271GA SA309193316GA SA309193320GA SA309193326GA SA309517466GA SA309517486GA SA309517714GA SA309517716GA SA309517730GA SA309517766GA SA309517770GA SA310028208GA SA310028244GA SA310028270GA SA310028277GA SA310028282GA SA310028288GA SA310028293GA SA310038379GA SA310038382GA SA310038416GA SA310038471GA SA310068836GA SA310028133GA SA310038447GA SA310058723GA SA310058727GA SA310058728GA SA310058729GA SA310058742GA SA310058748GA SA310068815GA SA310068837GA SA310028305GA SA310038361GA SA310038390GA SA310038400GA SA310038409GA SA310038460GA SA310048586GA SA310068861GA SA310392900GA SA310392902GA SA310392924GA SA310393227GA SA310393242GA SA310393271GA SA310393282GA SA310393299GA SA310455234GA SA310413536GA SA311069754GA SA311069773GA SA311069784GA SA311070591GA SA311060327GA SA311060364GA SA311060378GA SA311070515GA SA310424036GA SA310455124GA SA310455139GA SA310528776GA SA310528780GA SA310528782GA SA310528783GA SA310528793GA SA310382817GA SA310393248GA SA310403351GA SA310403366GA SA310434079GA SA310507187GA SA310507259GA SA311039523GA SA311039527GA SA311069826GA SA311069858GA SA311069861GA SA311101314GA SA311101336GA SA311101362GA SA311101368GA SA311080612GA SA311080644GA SA311090851GA SA311090879GA SA311101251GA SA311101286GA SA311101293GA SA311122019GA SA311122030GA SA311122040GA SA311069846GA SA311111691GA SA311121754GA SA311121756GA SA311121757GA SA311121771GA SA311121774GA SA311121778GA SA311121784GA SA311121787GA SA311121810GA SA311121811GA SA311121818GA SA311121826GA SA311121834GA SA311121836GA SA311142772GA SA311101089GA SA311101093GA SA311101121GA SA311101169GA SA311101170GA SA311101173GA SA311101439GA SA311101443GA SA311101444GA SA311101452GA SA311111673GA SA311132569GA SA311132571GA SA311132580GA SA311132584GA SA311142875GA SA311142878GA SA311142887GA SA311111583GA SA311111584GA SA311111586GA SA311111595GA SA311111602GA SA311111615GA SA311163892GA SA311163719GA SA311163700GA SA311184442GA SA311184543GA SA311266966GA SA311267008GA SA311163626GA SA311163705GA SA311163778GA SA311205498GA SA311327867GA SA311468411GA SA311236211GA SA311327811GA SA311458357GA SA311468391GA SA310382537GA SA310382559GA SA310382573GA SA310382585GA SA312284287GA SA312284289GA SA312284232GA SA312284336GA SA312294346GA SA312294347GA SA312294351GA SA312294352GA SA312294358GA SA312294364GA SA312294370GA SA312294359GA SA312294366GA SA312294671GA SA312294761GA SA312284260GA SA312253389GA SA312253242GA SA312253189GA SA312294696GA SA312344931GA SA312253185GA SA312294710GA SA312345111GA SA312345131GA SA312365773GA SA312365779GA SA312365792GA SA312365793GA SA312365796GA SA312365800GA SA312365803GA SA312365805GA SA312365809GA SA312365814GA SA312365815GA SA312365817GA SA312365818GA SA312365819GA SA312365820GA SA312365822GA SA312365823GA SA312365825GA SA312365830GA SA312365771GA SA312365772GA SA312365775GA SA312365781GA SA312365782GA SA312365787GA SA312365789GA SA312365797GA SA312365806GA SA312365812GA SA312365821GA SA312365827GA SA312365828GA SA312365829GA SA312345157GA SA312365794GA SA312397221GA SA312397241GA SA312396912GA SA312457914GA SA312469173GA SA312480223GA SA312480472GA SA311184696GA SA311297244GA SA312018860GA SA307231192GA SA308196570GA SA308196572GA SA308196573GA SA308196578GA SA308196580GA SA308196583GA SA308206586GA SA308206589GA SA308206601GA SA308206602GA SA308206606GA SA308206607GA SA308206608GA SA308206609GA SA308206610GA SA308206616GA SA308206617GA SA308206619GA SA308206620GA SA308206621GA SA308206622GA SA308206623GA SA308206624GA SA308206626GA SA308206628GA SA308206631GA SA308206632GA SA308206633GA SA308206637GA SA308206638GA SA308206639GA SA308206640GA SA308206641GA SA308206643GA SA308206644GA SA308206645GA SA308206646GA SA308206661GA SA308206664GA SA308206670GA SA311225894GA SA311225899GA SA311225907GA SA311236351GA SA311246496GA SA311297251GA SA312049456GA SA312049465GA SA312069655GA SA312069676GA SA312080896GA SA307080531GA SA307150677GA SA309062331GA SA309203368GA SA309203378GA SA309517667GA SA309517677GA SA309517692GA SA309517695GA SA309517696GA SA309517697GA SA309517702GA SA309517718GA SA309517719GA SA309517721GA SA309517723GA SA309517724GA SA309517727GA SA309517733GA SA309517734GA SA309517771GA SA309517777GA SA309517780GA SA309517786GA SA309517791GA SA309517794GA SA309517801GA SA309517806GA SA309517807GA SA309517809GA SA309517818GA SA309517825GA SA309517832GA SA309517834GA SA309517836GA SA309517868GA SA309517873GA SA309517884GA SA309517886GA SA309517887GA SA309517894GA SA309517897GA SA309517905GA SA309527918GA SA309527919GA SA309527921GA SA309527922GA SA309527924GA SA309527925GA SA309527926GA SA309527932GA SA309527934GA SA309527935GA SA309527936GA SA309527941GA SA309527942GA SA309527944GA SA309527946GA SA309527947GA SA309527949GA SA309527950GA SA309527952GA SA309527960GA SA309527961GA SA309527963GA SA309527966GA SA309527971GA SA309527988GA SA309517677GA SA309517723GA SA309517791GA SA309517806GA SA309517825GA SA309517868GA SA309517884GA SA309517887GA SA309527925GA SA309527950GA SA309527966GA SA309527988GA SA309517719GA SA309527935GA SA309527949GA SA309517692GA SA309517695GA SA309517696GA SA309517718GA SA309517721GA SA309517724GA SA309517727GA SA309517733GA SA309517734GA SA309517777GA SA309517780GA SA309517786GA SA309517794GA SA309517801GA SA309517807GA SA309517809GA SA309517832GA SA309517834GA SA309517873GA SA309517886GA SA309517894GA SA309517897GA SA309517905GA SA309527918GA SA309527919GA SA309527921GA SA309527922GA SA309527924GA SA309527932GA SA309527934GA SA309527941GA SA309527942GA SA309527952GA SA309527961GA SA309517667GA SA309517697GA SA309517702GA SA309517771GA SA309517818GA SA309517836GA SA309527926GA SA309527936GA SA309527944GA SA309527946GA SA309527947GA SA309527960GA SA309527963GA SA309527971GA SA309507339GA SA309517768GA SA309517874GA SA309527970GA SA309538017GA SA309538023GA SA309538025GA SA309538027GA SA309538033GA SA309538040GA SA309538042GA SA309538044GA SA309538100GA SA309538108GA SA310028162GA SA310028166GA SA310028176GA SA310028192GA SA310028212GA SA310028226GA SA310413521GA SA310413526GA SA310413529GA SA310413530GA SA310413551GA SA310423598GA SA310423642GA SA310423643GA SA310423644GA SA310423648GA SA310423655GA SA310423717GA SA310423726GA SA310423733GA SA310423824GA SA310423826GA SA310423827GA SA310423835GA SA310423837GA SA310423840GA SA310423847GA SA310423850GA SA310423853GA SA310423922GA SA310423923GA SA310423924GA SA310423926GA SA310423930GA SA310423945GA SA310423963GA SA310423966GA SA310424009GA SA310413570GA SA310454993GA SA311060174GA SA311060183GA SA311060196GA SA311060320GA SA311060322GA SA311060339GA SA311070535GA SA311070559GA SA311070560GA SA311070589GA SA311080708GA SA311090768GA SA311090797GA SA311090821GA SA311090827GA SA311070531GA SA311060193GA SA311060186GA SA311060212GA SA311060219GA SA311060250GA SA311060256GA SA311060311GA SA311060316GA SA311060317GA SA311060341GA SA311060350GA SA311070516GA SA311070548GA SA311070592GA SA311121702GA SA311121736GA SA311121739GA SA311121745GA SA311121697GA SA311121719GA SA311121722GA SA310475467GA SA310506748GA SA311142950GA SA311142957GA SA311142965GA SA311142981GA SA311142983GA SA311142985GA SA311153153GA SA311153155GA SA311153167GA SA311153168GA SA311153174GA SA311153177GA SA311153178GA SA311153184GA SA311153193GA SA311153203GA SA311153183GA SA311153187GA SA311153198GA SA311153221GA SA311153224GA SA311215729GA SA311215733GA SA311215734GA SA312232807GA SA312386632GA SA312386643GA SA312396800GA SA312396824GA SA312397271GA SA312407286GA SA312386600GA SA312397135GA SA312397141GA SA308084763GA SA308094973GA SA308155880GA SA308155901GA SA308186379GA SA308186408GA SA308186411GA SA308186419GA SA308186428GA SA308186430GA SA308186434GA SA308186435GA SA308186436GA SA308186438GA SA308186439GA SA308186440GA SA308186443GA SA308196460GA SA308196461GA SA308196462GA SA308196479GA SA308196481GA SA308196482GA SA308196483GA SA308196484GA SA308196489GA SA308196490GA SA308196496GA SA308196534GA SA308196552GA SA308196554GA SA308196565GA SA308125388GA SA308125389GA SA308125391GA SA308125392GA SA308125398GA SA308125415GA SA308125417GA SA308125422GA SA308135487GA SA308429321GA SA308429328GA SA308429340GA SA308449503GA SA308449506GA SA308449510GA SA308449514GA SA308449520GA SA308459651GA SA308459675GA SA308459860GA SA308459868GA SA308479891GA SA308479894GA SA308479896GA SA308479899GA SA309072345GA SA309072358GA SA309132809GA SA311142787GA SA310528850GA SA311317624GA SA311317677GA SA311317684GA SA311317746GA SA311428031GA SA311448178GA SA311297313GA SA311317673GA SA311317688GA SA311327900GA SA311327911GA SA311327913GA SA311327932GA SA311337993GA SA311337996GA SA311338003GA SA307302029GA SA307352358GA SA307352362GA SA307352366GA SA307362390GA SA308206697GA SA308206721GA SA308388611GA SA308398854GA SA309153042GA SA309253738GA SA307231249GA SA308409049GA SA308409053GA SA308409070GA SA308409153GA SA308409155GA SA308409158GA SA308409159GA SA308409169GA SA309375462GA SA309385568GA SA309385595GA SA309385614GA SA309385630GA SA309385666GA SA309395719GA SA309395749GA SA309395771GA SA309395789GA SA312080842GA SA312080875GA SA307070444GA SA307190726GA SA307190767GA SA307120586GA SA307190721GA SA307190746GA SA307190758GA SA307231253GA SA307231415GA SA307231421GA SA307291873GA SA307302007GA SA307302019GA SA307302027GA SA307302030GA SA307302032GA SA307302033GA SA307302034GA SA307302038GA SA307302040GA SA307302041GA SA307302042GA SA307302044GA SA307302045GA SA307322127GA SA307322128GA SA307322129GA SA307322130GA SA307322131GA SA307322132GA SA307322134GA SA307322135GA SA307322136GA SA307322167GA SA307322168GA SA307322169GA SA307322170GA SA307322171GA SA307322173GA SA307322174GA SA307322177GA SA307322178GA SA307322180GA SA307322181GA SA307322182GA SA307332207GA SA307332208GA SA307332211GA SA307332212GA SA307332214GA SA307332217GA SA307332218GA SA307332219GA SA307332222GA SA307332226GA SA307332248GA SA307382548GA SA309062280GA SA309253725GA SA309385576GA SA309385658GA SA309385662GA SA309385667GA SA309395712GA SA309395716GA SA309395718GA SA309395723GA SA309395725GA SA309395731GA SA309395734GA SA309395750GA SA309517631GA SA309517633GA SA309517732GA SA310496137GA SA310496160GA SA310496287GA SA310496318GA SA310496344GA SA310496473GA SA310496487GA SA310496491GA SA310496494GA SA310496495GA SA310496463GA SA310506663GA SA310506745GA SA310506845GA SA310506851GA SA310507019GA SA310507029GA SA310507051GA SA310507057GA SA310507097GA SA310507117GA SA310507148GA SA311132583GA SA309032109GA SA309062268GA SA309062270GA SA309062272GA SA309062287GA SA309062289GA SA309062290GA SA309062291GA SA309062293GA SA309062294GA SA309062301GA SA309062306GA SA309062307GA SA309062311GA SA309062314GA SA309062316GA SA309072361GA SA312080395GA SA312080772GA SA311069748GA SA311069761GA SA311069838GA SA311132559GA SA311132567GA SA311132576GA SA311142889GA SA311142895GA SA311142899GA SA312294403GA SA312294444GA SA312294552GA SA312294631GA SA312294653GA SA308459672GA SA312080911GA SA311327818GA SA312018874GA SA312018883GA SA312018884GA SA312018888GA SA312018890GA SA312018896GA SA312018899GA SA312018905GA SA312028908GA SA312028909GA SA312028918GA SA312028927GA SA312028935GA SA312028938GA SA312028940GA SA312028941GA SA312039213GA SA312059548GA SA312049464GA SA312080687GA SA312080694GA SA312080782GA SA312080807GA SA312080567GA SA312080742GA SA312080778GA SA312080790GA SA312080835GA SA312080836GA SA312080843GA SA312080844GA SA312080856GA SA312080864GA SA312080880GA SA312080891GA SA312080898GA SA312080913GA SA312121865GA SA308125381GA SA308145689GA SA308176252GA SA309122768GA SA309426087GA SA309426100GA SA309466737GA SA309466738GA SA309466757GA SA309466761GA SA309466778GA SA309466790GA SA309476933GA SA309476938GA SA309476942GA SA309476944GA SA309476959GA SA309476964GA SA309476977GA SA309476981GA SA309476983GA SA309476989GA SA309476991GA SA309476992GA SA309436286GA SA310259938GA SA310259939GA SA310259941GA SA310259950GA SA310259953GA SA310259954GA SA310259955GA SA310259957GA SA310259958GA SA310259963GA SA310259965GA SA310259970GA SA310259951GA SA310259967GA SA310517831GA SA310517842GA SA310517876GA SA311060146GA SA311060181GA SA311195116GA SA311195150GA SA311236287GA SA311236316GA SA311236321GA SA311236324GA SA311039446GA SA311508659GA SA311508662GA SA311508672GA SA310331299GA SA310361813GA SA310361816GA SA310361908GA SA310372052GA SA310372066GA SA310372154GA SA310372157GA SA310372160GA SA310372277GA SA310372278GA SA310372279GA SA310372281GA SA310372282GA SA310372284GA SA310372286GA SA310372287GA SA310372293GA SA310372295GA SA310372297GA SA310372301GA SA310372313GA SA310382487GA SA310382488GA SA310382491GA SA310382498GA SA310382523GA SA312253184GA SA312253297GA SA312253312GA SA312253339GA SA312375934GA SA312375997GA SA312376002GA SA312376005GA SA312376015GA SA312376017GA SA312376054GA SA312376055GA SA312376058GA SA312396917GA SA312396923GA SA312397063GA SA312397073GA SA312397079GA SA312397090GA SA312397158GA SA312397190GA SA312397198GA SA312152460GA SA312152465GA SA308337627GA SA308347735GA SA308347738GA SA308347739GA SA308347740GA SA308347751GA SA308347755GA SA308347758GA SA308347759GA SA308347760GA SA308347761GA SA308347811GA SA308347812GA SA308347824GA SA308491070GA SA310259942GA SA310382820GA SA310392853GA SA310392864GA SA310392868GA SA310392870GA SA310392871GA SA310392873GA SA310392879GA SA310392915GA SA310392919GA SA310392923GA SA310403391GA SA310403393GA SA310403395GA SA310403399GA SA310423763GA SA310423781GA SA310423808GA SA310423875GA SA310423880GA SA310423913GA SA310434142GA SA310434150GA SA310434154GA SA310434155GA SA310434158GA SA310434166GA SA310434178GA SA310434189GA SA310403421GA SA310403429GA SA310403443GA SA310403447GA SA310423892GA SA310423908GA SA310424045GA SA310434056GA SA310434104GA SA310434109GA SA310434111GA SA310434113GA SA310434179GA SA310518432GA SA310528942GA SA310372085GA SA310393084GA SA310393292GA SA310403352GA SA310403410GA SA310403432GA SA310403448GA SA310403450GA SA310434054GA SA310518404GA SA310518470GA SA310518519GA SA310518526GA SA310518532GA SA310528731GA SA310528741GA SA310528774GA SA310529052GA SA311287062GA SA311287068GA SA312192730GA SA312253204GA SA312253205GA SA312253206GA SA312253211GA SA312253214GA SA312253217GA SA312253219GA SA312253227GA SA312253240GA SA312253253GA SA312253255GA SA312253257GA SA312253273GA SA312232860GA SA312344842GA SA312344857GA SA312344890GA SA312344974GA SA312355317GA SA312355413GA SA312355482GA SA312355507GA SA312355508GA SA312355515GA SA312375951GA SA312376119GA SA312376138GA SA312376144GA SA312376148GA SA312376163GA SA312376190GA SA312376162GA SA312376180GA SA312376194GA SA312376218GA SA312376231GA SA312376232GA SA312470035GA SA312470058GA SA312479943GA SA312479960GA SA307483748GA SA307483781GA SA307483785GA SA307483787GA SA307483842GA SA307493844GA SA307504188GA SA307504195GA SA307504212GA SA307504214GA SA307514338GA SA307514343GA SA307514346GA SA307514349GA SA307301985GA SA307483799GA SA307514390GA SA307514397GA SA307514398GA SA307514404GA SA307514405GA SA307514417GA SA307514431GA SA307514432GA SA307514435GA SA307514444GA SA307514446GA SA307514450GA SA307514451GA SA307524470GA SA308044501GA SA308044525GA SA307504232GA SA307514235GA SA307514238GA SA307514245GA SA307514248GA SA307514256GA SA307514276GA SA307514277GA SA307514289GA SA307514290GA SA307514302GA SA307514311GA SA307514312GA SA307514316GA SA308084649GA SA308115237GA SA308115248GA SA308115256GA SA308115257GA SA308115270GA SA308115284GA SA308115286GA SA308125313GA SA308125315GA SA308125320GA SA308125321GA SA308125322GA SA308125323GA SA308125328GA SA308084741GA SA308105223GA SA308115234GA SA308115235GA SA308115242GA SA308115252GA SA308115255GA SA308115258GA SA308115262GA SA308115263GA SA308115264GA SA308115266GA SA308115267GA SA308115272GA SA308115274GA SA308115275GA SA308115283GA SA308115273GA SA308115285GA SA308125357GA SA308125359GA SA308135520GA SA308135522GA SA308135525GA SA308135531GA SA308135536GA SA308135537GA SA308135538GA SA308135539GA SA308135540GA SA308145658GA SA308145660GA SA308145663GA SA308145666GA SA308145667GA SA308145669GA SA308145670GA SA308145672GA SA308145673GA SA307241679GA SA308105214GA SA308115250GA SA308115271GA SA308115279GA SA308115280GA SA308125309GA SA308125310GA SA308125311GA SA308125329GA SA308125330GA SA308125331GA SA308125334GA SA308125336GA SA308125337GA SA308125346GA SA308125347GA SA308125361GA SA308125367GA SA308125368GA SA308125370GA SA308125372GA SA308135529GA SA308135530GA SA308125353GA SA308125363GA SA308135516GA SA308135517GA SA308135526GA SA308135532GA SA308135535GA SA308135542GA SA308145656GA SA308145664GA SA308145665GA SA308125358GA SA308145676GA SA308145678GA SA308145679GA SA308145682GA SA308145683GA SA308145691GA SA308166056GA SA308166070GA SA308196558GA SA308125373GA SA308155971GA SA308206603GA SA308186393GA SA308216767GA SA308297032GA SA308409054GA SA308429355GA SA308439450GA SA308409048GA SA308409077GA SA308449509GA SA308449527GA SA308449573GA SA308449578GA SA308449580GA SA308449582GA SA308449584GA SA308449586GA SA308449587GA SA308449592GA SA308459646GA SA308459660GA SA308459662GA SA308459663GA SA308459664GA SA308459680GA SA308459681GA SA308491213GA SA308491216GA SA308511756GA SA309122722GA SA309042194GA SA309122660GA SA309142983GA SA309142987GA SA309142988GA SA309142990GA SA309142991GA SA309153026GA SA309153028GA SA309153030GA SA309153031GA SA309153036GA SA309153037GA SA309153039GA SA309153040GA SA309153041GA SA309153043GA SA309153044GA SA309153049GA SA309153050GA SA309153051GA SA309153052GA SA309153054GA SA309153057GA SA309153060GA SA309153062GA SA309153063GA SA309153070GA SA309153074GA SA309153077GA SA309153078GA SA309153085GA SA309173106GA SA309173162GA SA309193298GA SA309193299GA SA309263916GA SA310048669GA SA310058695GA SA310465361GA SA310496628GA SA310506657GA SA310506661GA SA310506678GA SA310506702GA SA310506728GA SA310506864GA SA310506865GA SA310506872GA SA310506879GA SA310506880GA SA310506892GA SA310506893GA SA310506894GA SA310506896GA SA310506897GA SA310506992GA SA310506998GA SA310507001GA SA310507002GA SA310507004GA SA310507006GA SA310507008GA SA310507009GA SA310507010GA SA310507017GA SA310507020GA SA310507023GA SA310507032GA SA310507048GA SA310506846GA SA310506862GA SA310506871GA SA310506993GA SA310506997GA SA310507007GA SA310507028GA SA310507031GA SA310507037GA SA310507049GA SA310507055GA SA310506875GA SA310506737GA SA311090976GA SA312284293GA SA312284294GA SA312284295GA SA312284323GA SA312386403GA SA312386410GA SA312386426GA SA312386432GA SA312386457GA SA312386477GA SA312447559GA SA312447564GA SA312480337GA SA312480342GA SA312480352GA SA312480382GA SA312480387GA SA312480402GA SA312480403GA SA312480411GA SA312480414GA SA312018748GA SA312018751GA SA312018758GA SA312018762GA SA312018768GA SA312018769GA SA312152330GA SA312152351GA SA312080621GA SA312029048GA SA310475451GA SA310475460GA SA310475530GA SA310475532GA SA310475534GA SA311184578GA SA312294401GA SA312294411GA SA312294442GA SA312294480GA SA312294486GA SA312294504GA SA312294533GA SA312294555GA SA312355437GA SA312355438GA SA312355452GA SA312355454GA SA312355457GA SA312355458GA SA312355465GA SA312355469GA SA312376077GA SA312376091GA SA312376282GA SA312376283GA SA312376285GA SA312376288GA SA312376290GA SA312376291GA SA312376296GA SA312376298GA SA312376301GA SA312376302GA SA312376304GA SA312376305GA SA312376306GA SA312376309GA SA312386491GA SA312386502GA SA312386506GA SA312386511GA SA312386515GA SA312386519GA SA312386523GA SA312386532GA SA312386534GA SA312386547GA SA312069671GA SA312069673GA SA312069682GA SA310260487GA SA310393002GA SA310424029GA SA310444262GA SA310455003GA SA310507086GA SA310507479GA SA310517577GA SA310518683GA SA310528894GA SA310529145GA SA311039390GA SA311060161GA SA311060162GA SA311060170GA SA311069722GA SA311069734GA SA311153086GA SA311153152GA SA311153156GA SA311153159GA SA311153166GA SA311153171GA SA311153179GA SA311153180GA SA311153191GA SA311153199GA SA311153200GA SA311153201GA SA311163372GA SA311163414GA SA311163422GA SA311163424GA SA311163425GA SA311163426GA SA311184594GA SA311184664GA SA311256705GA SA311184636GA SA311184528GA SA311266875GA SA311266891GA SA311266893GA SA311317645GA SA311297360GA SA311317646GA SA311317655GA SA311060197GA SA311060269GA SA311307513GA SA311307519GA SA311307521GA SA311307530GA SA311458340GA SA311458346GA SA311327936GA SA311327942GA SA311327812GA SA312232829GA SA312232833GA SA312232846GA SA312355230GA SA312386469GA SA312386480GA SA312386481GA SA312386482GA SA312386489GA SA312386535GA SA312386685GA SA312396928GA SA312396930GA SA312397274GA SA312470004GA SA312470014GA SA312470015GA SA312470017GA SA312470018GA SA312470033GA SA312470037GA SA312470070GA SA312470073GA SA312479542GA SA312479784GA SA312479926GA SA312479955GA SA312479957GA SA312479962GA SA312479965GA SA312480096GA SA312480113GA SA312480115GA SA312018749GA SA312018750GA SA312018760GA SA312018761GA SA312018770GA SA312018772GA SA312018766GA SA311458234GA SA312039074GA SA312018901GA SA312039202GA SA312039204GA SA308286943GA SA308388363GA SA308409129GA SA309042129GA SA309042144GA SA310372132GA SA310372171GA SA310372205GA SA310382533GA SA310382600GA SA310382785GA SA310382826GA SA310382850GA SA310403381GA SA310403389GA SA310403402GA SA310403407GA SA310423784GA SA310423788GA SA310423868GA SA310423878GA SA310434067GA SA310434096GA SA310434152GA SA310455115GA SA310455236GA SA310455260GA SA310455271GA SA310465297GA SA310465306GA SA310465311GA SA310465316GA SA311287017GA SA311287018GA SA311287035GA SA311287043GA SA311287052GA SA311287081GA SA311287086GA SA311287091GA SA311287096GA SA312263966GA SA312264011GA SA312264057GA SA312264099GA SA312397253GA SA312079923GA SA312079939GA SA312079955GA SA312070186GA SA312070251GA SA309477038GA SA310403343GA SA310403353GA SA310403431GA SA310403439GA SA310423795GA SA310423876GA SA310260029GA SA310290644GA SA310300772GA SA310310839GA SA310310881GA SA310310945GA SA309446497GA SA309466779GA SA309466786GA SA309466787GA SA309466796GA SA309466798GA SA310038397GA SA311184370GA SA311184525GA SA311184557GA SA311184613GA SA310260545GA SA310290693GA SA310321003GA SA310361608GA SA310372044GA SA310372070GA SA310372142GA SA310372150GA SA310372153GA SA310372162GA SA310372296GA SA310382375GA SA310382376GA SA310382377GA SA310382379GA SA310382380GA SA310382381GA SA310382383GA SA310382388GA SA310382395GA SA310382397GA SA310382399GA SA310382400GA SA310382401GA SA310382402GA SA310382403GA SA310382404GA SA310382405GA SA310382406GA SA310382407GA SA310382409GA SA310382410GA SA310382411GA SA310382412GA SA310382414GA SA310382416GA SA310382417GA SA310382419GA SA310382422GA SA310382423GA SA310382426GA SA310382430GA SA310382431GA SA310382434GA SA310382435GA SA310382437GA SA310382440GA SA310382443GA SA310382444GA SA310382451GA SA310382452GA SA310382453GA SA310382455GA SA310382459GA SA310382460GA SA310382461GA SA310382462GA SA310382463GA SA310382464GA SA312253387GA SA312253392GA SA312253397GA SA312253403GA SA312253405GA SA312253406GA SA312253411GA SA312253416GA SA312253417GA SA312253422GA SA312253423GA SA312253427GA SA312253428GA SA312253429GA SA312253430GA SA312253431GA SA312253432GA SA312253446GA SA312253448GA SA312253451GA SA312253452GA SA312253453GA SA312253455GA SA312253457GA SA312253458GA SA312253462GA SA312253471GA SA312253473GA SA312253487GA SA312253516GA SA312376203GA SA312386316GA SA308501330GA SA308501345GA SA308501348GA SA308501352GA SA308501383GA SA311246454GA SA311246482GA SA311246503GA SA311256803GA SA311256836GA SA311256857GA SA311256865GA SA311317761GA SA312121760GA SA312121761GA SA312121763GA SA312121766GA SA312121769GA SA312121770GA SA312121773GA SA312121775GA SA312121780GA SA312121795GA SA312121808GA SA312121815GA SA312121816GA SA312121820GA SA312121823GA SA312121850GA SA312121852GA SA307220887GA SA307220909GA SA307220924GA SA307221149GA SA307241696GA SA307251720GA SA307251731GA SA307251732GA SA307221063GA SA307382596GA SA307382608GA SA307382614GA SA307382617GA SA308095048GA SA308095057GA SA308196539GA SA308206635GA SA308206751GA SA308216775GA SA308206611GA SA308206636GA SA308206649GA SA308297026GA SA308297028GA SA308297031GA SA308297033GA SA308297054GA SA308297055GA SA308368120GA SA308408978GA SA308408992GA SA308408995GA SA308409000GA SA308409003GA SA308409005GA SA308409006GA SA308409007GA SA308409008GA SA308409095GA SA308409097GA SA308409098GA SA308409103GA SA308419281GA SA308347849GA SA308388364GA SA308408989GA SA308408994GA SA308408997GA SA308409100GA SA308409104GA SA308491128GA SA308491164GA SA308491165GA SA308491166GA SA308491176GA SA308491221GA SA308491222GA SA308491231GA SA308491232GA SA308501412GA SA309032023GA SA309032085GA SA309042114GA SA309042126GA SA309042149GA SA309042151GA SA309042152GA SA309042153GA SA309042154GA SA309042155GA SA309042159GA SA309042160GA SA309042161GA SA309042167GA SA309042175GA SA309042176GA SA309042177GA SA309042178GA SA309042179GA SA309042190GA SA309042196GA SA309042199GA SA308469884GA SA308511713GA SA308511751GA SA308511758GA SA308511759GA SA308511760GA SA308511761GA SA308511762GA SA308511766GA SA308511767GA SA309011868GA SA309011880GA SA309021924GA SA309021925GA SA309021928GA SA309021933GA SA309021936GA SA309021937GA SA309021938GA SA309021939GA SA309021948GA SA309021953GA SA309032090GA SA309032101GA SA309032102GA SA309032103GA SA309032104GA SA309032107GA SA309032111GA SA309042115GA SA309042117GA SA309042119GA SA309042120GA SA309042121GA SA309042124GA SA309042125GA SA309122637GA SA308511655GA SA308511670GA SA308511673GA SA308521793GA SA308521797GA SA308521808GA SA309032035GA SA309032036GA SA309032038GA SA309032044GA SA309032045GA SA309032046GA SA309032047GA SA309032049GA SA309032051GA SA309032084GA SA309032088GA SA309032089GA SA309032093GA SA309032094GA SA309032095GA SA309032096GA SA309032097GA SA309032098GA SA309032099GA SA309032100GA SA309082450GA SA309112575GA SA309122651GA SA309122708GA SA309122709GA SA309122710GA SA309122711GA SA309122714GA SA309122716GA SA309122725GA SA309122728GA SA309122731GA SA309122734GA SA309122736GA SA309122737GA SA309122738GA SA309122739GA SA309122740GA SA309122741GA SA309122742GA SA309122743GA SA309122744GA SA309122745GA SA309122746GA SA309122747GA SA309122748GA SA309122750GA SA309122751GA SA309122753GA SA309122754GA SA309122755GA SA309122756GA SA309122758GA SA309122759GA SA309122760GA SA309122761GA SA309122762GA SA309122763GA SA309122772GA SA309122777GA SA309122780GA SA309122781GA SA309122783GA SA309122785GA SA309122786GA SA309122787GA SA309122788GA SA309122789GA SA309122790GA SA309092504GA SA309142870GA SA309142871GA SA309142874GA SA309142889GA SA309142892GA SA309142897GA SA309142898GA SA309142899GA SA309142900GA SA309142902GA SA309142905GA SA309142908GA SA309142909GA SA309142920GA SA309223478GA SA309223514GA SA309233552GA SA309233553GA SA309233554GA SA309233555GA SA309233556GA SA309233557GA SA309233558GA SA309233559GA SA309233560GA SA309243633GA SA309243635GA SA309243636GA SA309243637GA SA309243638GA SA309243639GA SA309243640GA SA309243641GA SA309243647GA SA309243656GA SA309243657GA SA309243660GA SA309243661GA SA309243662GA SA309243664GA SA309243666GA SA309243672GA SA309112564GA SA309132842GA SA309132843GA SA309193301GA SA309193327GA SA309203432GA SA309253835GA SA309263931GA SA309193304GA SA309203420GA SA309223453GA SA309233544GA SA309243623GA SA309243644GA SA309243649GA SA309253724GA SA309253737GA SA309253739GA SA309253740GA SA309253748GA SA309263972GA SA309294108GA SA309304186GA SA309304208GA SA309304223GA SA309304231GA SA309304249GA SA309304252GA SA309304261GA SA309304264GA SA309304266GA SA309304269GA SA309324333GA SA309324341GA SA309324349GA SA309193330GA SA309193341GA SA309253768GA SA309253780GA SA309294015GA SA309294017GA SA309294018GA SA309294019GA SA309294020GA SA309294021GA SA309294022GA SA309294024GA SA309294026GA SA309294027GA SA309294032GA SA309294034GA SA309294036GA SA309294038GA SA309294040GA SA309294042GA SA309294046GA SA309294054GA SA309294056GA SA309294059GA SA309294062GA SA309294063GA SA309294066GA SA309294070GA SA309294072GA SA309294073GA SA309294078GA SA309294080GA SA309294081GA SA309294086GA SA309294088GA SA309294091GA SA309294107GA SA309294118GA SA309294119GA SA309294133GA SA309132840GA SA309132841GA SA309193363GA SA309294049GA SA309294095GA SA309294097GA SA309294098GA SA309294106GA SA309294110GA SA309294123GA SA309294125GA SA309294126GA SA309294127GA SA309294128GA SA309294130GA SA309304136GA SA309304138GA SA309304139GA SA309304146GA SA309304147GA SA309304150GA SA309304151GA SA309304152GA SA309304153GA SA309304154GA SA309304155GA SA309304156GA SA309304157GA SA309304158GA SA309304159GA SA309304160GA SA309304164GA SA309304165GA SA309304169GA SA309304171GA SA309304173GA SA309304178GA SA309304191GA SA309304203GA SA309304204GA SA309193349GA SA309203370GA SA309203371GA SA309203375GA SA309203404GA SA309223439GA SA309243676GA SA309253762GA SA309253764GA SA309253769GA SA309253770GA SA309253776GA SA309253783GA SA309253788GA SA309253791GA SA309253795GA SA309253799GA SA309253805GA SA309253811GA SA309253814GA SA309253818GA SA309253825GA SA309253827GA SA309253829GA SA309253831GA SA309253851GA SA309253859GA SA309253901GA SA309253902GA SA309253903GA SA309253904GA SA309253905GA SA309253907GA SA309263927GA SA309263928GA SA309263932GA SA309263934GA SA309263936GA SA309263942GA SA309263944GA SA309173159GA SA309173166GA SA309173186GA SA309173187GA SA309203366GA SA309253758GA SA309253779GA SA309253785GA SA309253786GA SA309253787GA SA309253793GA SA309253794GA SA309253796GA SA309253797GA SA309253798GA SA309253800GA SA309253801GA SA309253802GA SA309253804GA SA309253806GA SA309253809GA SA309253810GA SA309253813GA SA309253815GA SA309253816GA SA309253820GA SA309253822GA SA309253824GA SA309253826GA SA309253830GA SA309253832GA SA309253853GA SA309253854GA SA309253858GA SA309253860GA SA309253862GA SA309253872GA SA309263947GA SA309263956GA SA309263959GA SA309173151GA SA309253747GA SA309253789GA SA309253803GA SA309253879GA SA309263926GA SA309294045GA SA309294074GA SA309294075GA SA309294076GA SA309294082GA SA309294083GA SA309294096GA SA309294099GA SA309294100GA SA309294101GA SA309294102GA SA309294104GA SA309294105GA SA309294109GA SA309294111GA SA309294114GA SA309294115GA SA309294120GA SA309294121GA SA309294122GA SA309294129GA SA309294131GA SA309294132GA SA309294134GA SA309304135GA SA309304137GA SA309304140GA SA309304141GA SA309304142GA SA309304143GA SA309304144GA SA309304145GA SA309304148GA SA309304149GA SA309304162GA SA309304163GA SA309304166GA SA309304167GA SA309304168GA SA309304174GA SA309304175GA SA309304176GA SA309304177GA SA309304180GA SA309304181GA SA309304182GA SA309304183GA SA309304184GA SA309304185GA SA309304187GA SA309304188GA SA309304190GA SA309304192GA SA309304193GA SA309304194GA SA309304195GA SA309304196GA SA309304197GA SA309304198GA SA309304199GA SA309304200GA SA309304202GA SA309304205GA SA309304206GA SA309304207GA SA309304209GA SA309304210GA SA309304211GA SA309304212GA SA309304214GA SA309304215GA SA309304216GA SA309304217GA SA309304219GA SA309253817GA SA309294041GA SA309304189GA SA309324299GA SA309324327GA SA309324360GA SA309324380GA SA309324386GA SA309324427GA SA309324429GA SA309324430GA SA309324431GA SA309324432GA SA309324433GA SA309324439GA SA309324440GA SA309324441GA SA309324444GA SA309334556GA SA309334565GA SA309334567GA SA309334568GA SA309334570GA SA309334572GA SA309334578GA SA309334580GA SA309334584GA SA309334585GA SA309334589GA SA309334595GA SA309334599GA SA309334600GA SA309334601GA SA309334605GA SA309334606GA SA309334607GA SA309334608GA SA309334609GA SA309334611GA SA309334614GA SA309334615GA SA309334616GA SA309334617GA SA309334618GA SA309334619GA SA309334620GA SA309334622GA SA309334623GA SA309334625GA SA309334641GA SA309334643GA SA309334645GA SA309334646GA SA309334647GA SA309334648GA SA309334649GA SA309334651GA SA309334653GA SA309334654GA SA309334656GA SA309334657GA SA309334658GA SA309334659GA SA309334660GA SA309334661GA SA309334662GA SA309334663GA SA309334668GA SA309334672GA SA309334674GA SA309334676GA SA309334677GA SA309334678GA SA309334679GA SA309334680GA SA309334682GA SA309334683GA SA309334685GA SA309334687GA SA309334688GA SA309334689GA SA309334690GA SA309334692GA SA309334693GA SA309334694GA SA309334695GA SA309334696GA SA309334697GA SA309334702GA SA309334703GA SA309334707GA SA309334708GA SA309334710GA SA309334711GA SA309334712GA SA309334714GA SA309334715GA SA309334716GA SA309334717GA SA309334718GA SA309334719GA SA309334720GA SA309334723GA SA309334724GA SA309334725GA SA309334727GA SA309334729GA SA309334730GA SA309334737GA SA309334740GA SA309334741GA SA309334744GA SA309334748GA SA309334751GA SA309334752GA SA309334755GA SA309344838GA SA309344842GA SA309344843GA SA309344847GA SA309344851GA SA309314292GA SA309314293GA SA309314294GA SA309314296GA SA309314297GA SA309314298GA SA309324301GA SA309324302GA SA309324305GA SA309324306GA SA309324307GA SA309324308GA SA309324309GA SA309324311GA SA309324312GA SA309324313GA SA309324314GA SA309324315GA SA309324316GA SA309324319GA SA309324322GA SA309324323GA SA309324324GA SA309324325GA SA309324326GA SA309324329GA SA309324330GA SA309324331GA SA309324334GA SA309324335GA SA309324336GA SA309324337GA SA309324338GA SA309324340GA SA309324344GA SA309324345GA SA309324348GA SA309324350GA SA309324351GA SA309324353GA SA309304170GA SA309314290GA SA309314295GA SA309324318GA SA309324320GA SA309324347GA SA309324356GA SA309324357GA SA309324365GA SA309324368GA SA309324369GA SA309324371GA SA309324373GA SA309324375GA SA309324376GA SA309324377GA SA309324382GA SA309324384GA SA309324385GA SA309324387GA SA309324389GA SA309324391GA SA309324392GA SA309324393GA SA309324396GA SA309324399GA SA309324401GA SA309324402GA SA309324404GA SA309324408GA SA309324409GA SA309324410GA SA309324412GA SA309324414GA SA309324415GA SA309324416GA SA309324418GA SA309324426GA SA309324428GA SA309334559GA SA309354917GA SA309375478GA SA309375479GA SA309375480GA SA309375481GA SA309375486GA SA309375487GA SA309375489GA SA309375491GA SA309375496GA SA309375500GA SA309375501GA SA309375504GA SA309375505GA SA309385508GA SA309385511GA SA309385512GA SA309385515GA SA309385516GA SA309385517GA SA309385518GA SA309385519GA SA309385520GA SA309385527GA SA309385529GA SA309385530GA SA309385531GA SA309385532GA SA309385533GA SA309385536GA SA309385539GA SA309385540GA SA309385542GA SA309385549GA SA309385562GA SA309385563GA SA309385575GA SA309385578GA SA309385586GA SA309385593GA SA309354915GA SA309365048GA SA309385541GA SA309385560GA SA309385565GA SA309385566GA SA309385571GA SA309385572GA SA309385588GA SA309385592GA SA309385600GA SA309385601GA SA309385602GA SA309385604GA SA309385606GA SA309385613GA SA309385619GA SA309385620GA SA309385623GA SA309385629GA SA309385631GA SA309385635GA SA309385643GA SA309385646GA SA309385647GA SA309385648GA SA309385649GA SA309385650GA SA309385654GA SA309385657GA SA309385659GA SA309385660GA SA309385661GA SA309385663GA SA309385669GA SA309385680GA SA309385690GA SA309385691GA SA309385692GA SA309395722GA SA309304213GA SA309355021GA SA309253883GA SA309385671GA SA309385672GA SA309385673GA SA309385677GA SA309395726GA SA309395737GA SA309395738GA SA309395739GA SA309395742GA SA309395747GA SA309395760GA SA309395786GA SA309395787GA SA309395790GA SA309395798GA SA309395799GA SA309395802GA SA309395805GA SA309395806GA SA309395807GA SA309395810GA SA309334603GA SA309334624GA SA309334626GA SA309334627GA SA309334628GA SA309334629GA SA309334630GA SA309334631GA SA309334632GA SA309334633GA SA309334634GA SA309334635GA SA309334636GA SA309334637GA SA309334644GA SA309344839GA SA309375497GA SA309385526GA SA309385585GA SA309385587GA SA309385590GA SA309385599GA SA309385603GA SA309385615GA SA309385616GA SA309385639GA SA309385641GA SA309385645GA SA309385655GA SA309385656GA SA309385665GA SA309385668GA SA309385679GA SA309385681GA SA309385682GA SA309385683GA SA309385685GA SA309385686GA SA309385687GA SA309385688GA SA309446379GA SA309446383GA SA309446386GA SA309446389GA SA309446390GA SA309446408GA SA309446412GA SA309446421GA SA309456633GA SA309456637GA SA309456638GA SA309456640GA SA309456643GA SA309456647GA SA309456649GA SA309456650GA SA309456652GA SA309456656GA SA309456659GA SA309456660GA SA309456663GA SA309456664GA SA309456666GA SA309456682GA SA309456695GA SA309456697GA SA309456701GA SA309456702GA SA309456709GA SA309456711GA SA309456712GA SA309456713GA SA309456715GA SA309365044GA SA309375484GA SA309375493GA SA309375499GA SA309385513GA SA309385521GA SA309385523GA SA309385534GA SA309385537GA SA309385545GA SA309385548GA SA309385555GA SA309385558GA SA309385559GA SA309385570GA SA309385573GA SA309385577GA SA309385583GA SA309385591GA SA309385594GA SA309385596GA SA309385597GA SA309385598GA SA309385609GA SA309385610GA SA309385612GA SA309385618GA SA309385621GA SA309385622GA SA309385625GA SA309385626GA SA309385632GA SA309385633GA SA309385634GA SA309385636GA SA309385638GA SA309385642GA SA309385651GA SA309385652GA SA309385653GA SA309517548GA SA309517840GA SA309538065GA SA309538066GA SA309538072GA SA310028136GA SA310028141GA SA310028144GA SA310028309GA SA310028310GA SA310028311GA SA310028312GA SA310028314GA SA310028318GA SA310028322GA SA310028325GA SA310028330GA SA310028334GA SA310028339GA SA310038345GA SA310038346GA SA310038482GA SA310038485GA SA310038489GA SA310038491GA SA310038498GA SA310038502GA SA310038505GA SA310038509GA SA310038511GA SA310038515GA SA310038516GA SA310038518GA SA310038528GA SA310038529GA SA310038531GA SA310038532GA SA310038536GA SA310048539GA SA310048540GA SA310048543GA SA310048545GA SA310048546GA SA310048547GA SA310048549GA SA310048551GA SA310048561GA SA310048569GA SA310048570GA SA310048573GA SA310048574GA SA310048575GA SA310048578GA SA310048581GA SA310048582GA SA310048630GA SA310048631GA SA310048632GA SA310048636GA SA310048639GA SA310048642GA SA310048650GA SA310048654GA SA310048655GA SA310048657GA SA310048659GA SA310118969GA SA310118981GA SA310118987GA SA310139279GA SA310149312GA SA310149320GA SA310149329GA SA310149335GA SA310159361GA SA310159376GA SA310159377GA SA310159379GA SA310159457GA SA310159461GA SA310159477GA SA310260002GA SA310413452GA SA310413460GA SA310413461GA SA310413467GA SA310413468GA SA310413470GA SA310413472GA SA310413474GA SA310413475GA SA310413477GA SA310413478GA SA310413480GA SA310413482GA SA310413483GA SA310413484GA SA310413487GA SA310413488GA SA310413489GA SA310413490GA SA310413492GA SA310413494GA SA310413496GA SA310413501GA SA310413502GA SA310413503GA SA310413506GA SA310413510GA SA310413512GA SA310413513GA SA310413515GA SA310413519GA SA310413520GA SA310413525GA SA310413527GA SA310413528GA SA310413531GA SA310413532GA SA310413533GA SA310413539GA SA310413540GA SA310413546GA SA310413547GA SA310413550GA SA310413552GA SA310413553GA SA310413554GA SA310413562GA SA310413566GA SA310413567GA SA310413568GA SA310413569GA SA310423574GA SA310423575GA SA310423578GA SA310423579GA SA310423590GA SA310423594GA SA310423597GA SA310423599GA SA310423601GA SA310423603GA SA310423605GA SA310423612GA SA310423620GA SA310423624GA SA310423625GA SA310423629GA SA310423630GA SA310423631GA SA310423632GA SA310423633GA SA310423634GA SA310423636GA SA310423638GA SA310423641GA SA310423645GA SA310423647GA SA310423660GA SA310423724GA SA310423725GA SA310423728GA SA310423729GA SA310423731GA SA310423734GA SA310423735GA SA310423823GA SA310423828GA SA310423830GA SA310423832GA SA310423833GA SA310423844GA SA310423848GA SA310423849GA SA310423854GA SA310423855GA SA310423857GA SA310423858GA SA310423859GA SA310423860GA SA310423861GA SA310423920GA SA310423925GA SA310423931GA SA310423932GA SA310423934GA SA310423936GA SA310423938GA SA310423939GA SA310423999GA SA310423990GA SA310434133GA SA310434141GA SA310434171GA SA310403441GA SA310455283GA SA310465304GA SA310465308GA SA310517815GA SA310517873GA SA310518277GA SA310518283GA SA310423581GA SA310423582GA SA310423852GA SA310423863GA SA310423865GA SA310423997GA SA310434206GA SA310434214GA SA310434215GA SA310434216GA SA310434217GA SA310434218GA SA310434219GA SA310434220GA SA310434221GA SA310434222GA SA310434223GA SA310434224GA SA310434225GA SA310434228GA SA310434229GA SA310434231GA SA310434232GA SA310434233GA SA310434234GA SA310434235GA SA310434236GA SA310434237GA SA310434239GA SA310434240GA SA310434241GA SA310434243GA SA310434244GA SA310434245GA SA310444246GA SA310444249GA SA310444250GA SA310444252GA SA310444254GA SA310444255GA SA310444257GA SA310444260GA SA310444261GA SA310444263GA SA310444264GA SA310444265GA SA310444276GA SA310444281GA SA310444283GA SA310444284GA SA310444305GA SA310444308GA SA310455082GA SA310455087GA SA310455107GA SA310455145GA SA310455165GA SA310455172GA SA310455175GA SA310465324GA SA310465335GA SA310465337GA SA310465348GA SA310465349GA SA310465354GA SA310465368GA SA310465369GA SA310465370GA SA310465374GA SA310465385GA SA310465391GA SA310465400GA SA310465408GA SA310465414GA SA310465423GA SA310465437GA SA310518420GA SA310518426GA SA310518427GA SA310528809GA SA310518378GA SA310403449GA SA310434055GA SA310434075GA SA310434103GA SA311060002GA SA311060009GA SA310496126GA SA311101299GA SA311069690GA SA311069700GA SA311069703GA SA311090911GA SA311090949GA SA311090968GA SA311090979GA SA311090992GA SA311090999GA SA311101144GA SA311101156GA SA311101159GA SA311121997GA SA311122066GA SA311122111GA SA311132419GA SA311132420GA SA311132426GA SA311132429GA SA311132431GA SA311132437GA SA311132445GA SA311132446GA SA311132450GA SA311132462GA SA311132486GA SA311132487GA SA311132513GA SA311132520GA SA311132528GA SA311090935GA SA311090955GA SA311090972GA SA311090995GA SA311091040GA SA311101072GA SA311101073GA SA311101075GA SA311101092GA SA311101100GA SA311101103GA SA311101153GA SA311101160GA SA311111649GA SA311142677GA SA311142680GA SA311121841GA SA311121846GA SA311121971GA SA311121973GA SA311121979GA SA311121982GA SA311121986GA SA311121987GA SA311121989GA SA311121993GA SA311121994GA SA311121998GA SA311122002GA SA311122005GA SA311122009GA SA311122010GA SA311122071GA SA311122073GA SA311122093GA SA311122094GA SA311122101GA SA311122102GA SA311122112GA SA311122115GA SA311132485GA SA311132543GA SA311132605GA SA311132613GA SA311132614GA SA311132619GA SA311153232GA SA311153239GA SA311153268GA SA311153283GA SA311153284GA SA311153289GA SA311153314GA SA311163474GA SA311163481GA SA311163482GA SA311142683GA SA311142684GA SA311142688GA SA311142693GA SA311142698GA SA311142703GA SA311142729GA SA311142737GA SA311142741GA SA311142742GA SA311142744GA SA311142797GA SA311142855GA SA311143011GA SA311143017GA SA311143018GA SA311143020GA SA311153030GA SA311153038GA SA311153053GA SA311153306GA SA311153322GA SA311163432GA SA311163434GA SA311163444GA SA311163451GA SA311163456GA SA311163462GA SA311163463GA SA311163465GA SA311163471GA SA311163472GA SA311163475GA SA311163476GA SA311163477GA SA311163478GA SA311163486GA SA311163534GA SA311163536GA SA311163537GA SA311163538GA SA311163540GA SA311163544GA SA311163546GA SA311163548GA SA311163549GA SA311163554GA SA311163555GA SA311163558GA SA311163564GA SA311163581GA SA311163589GA SA311163591GA SA311163595GA SA311163614GA SA311163621GA SA311163636GA SA311163655GA SA311163702GA SA311163704GA SA311163712GA SA311163714GA SA311163717GA SA311163720GA SA311163722GA SA311163732GA SA311163780GA SA311163791GA SA311163803GA SA311163806GA SA311163811GA SA311163820GA SA311163829GA SA311163837GA SA311163839GA SA311163841GA SA311163848GA SA311163853GA SA311173999GA SA311174000GA SA311174003GA SA311174005GA SA311174006GA SA311174015GA SA311174017GA SA311174018GA SA311174021GA SA311174033GA SA311174044GA SA311205426GA SA311205437GA SA310529148GA SA310529153GA SA310539348GA SA311039433GA SA311060202GA SA311060221GA SA311060266GA SA311060273GA SA311060277GA SA311287204GA SA311297346GA SA311307515GA SA311307516GA SA311307517GA SA311307518GA SA311307520GA SA311307522GA SA311307525GA SA311307526GA SA311307527GA SA311307528GA SA311307533GA SA311307534GA SA311307535GA SA311307537GA SA311307595GA SA311317660GA SA311317668GA SA311317737GA SA311327823GA SA311327825GA SA311327874GA SA311448182GA SA311448183GA SA311448184GA SA311448185GA SA311448186GA SA311448187GA SA311448191GA SA311458215GA SA311458216GA SA311458227GA SA311458228GA SA311458232GA SA311458233GA SA311458235GA SA311458238GA SA311458242GA SA311458246GA SA311458247GA SA311458248GA SA311458250GA SA311458251GA SA311458254GA SA311458255GA SA311458261GA SA311458264GA SA311458265GA SA311458266GA SA311184843GA SA311184845GA SA311194890GA SA311194911GA SA311194924GA SA311246680GA SA310259918GA SA310259986GA SA310260037GA SA310260222GA SA310260333GA SA310260527GA SA310260556GA SA310290632GA SA310290642GA SA310290737GA SA310310846GA SA310310853GA SA310310855GA SA310310878GA SA310310880GA SA310310882GA SA310310887GA SA310310888GA SA310310897GA SA310310899GA SA310310900GA SA310310913GA SA310310919GA SA310310923GA SA310310927GA SA310310928GA SA310310929GA SA310310931GA SA310310932GA SA310310934GA SA310310935GA SA310310936GA SA310310937GA SA310310938GA SA310310946GA SA310310947GA SA310310948GA SA310310949GA SA310310950GA SA310310955GA SA310310958GA SA310310960GA SA310310962GA SA310310965GA SA310310966GA SA310310967GA SA310310968GA SA310310969GA SA310310970GA SA310310971GA SA310310972GA SA310310976GA SA310310977GA SA310310979GA SA310310980GA SA310310981GA SA310310982GA SA310310983GA SA310310984GA SA310310985GA SA310310986GA SA310310988GA SA310310989GA SA310310990GA SA310310991GA SA310310992GA SA310310994GA SA310310995GA SA310310996GA SA310310997GA SA310310998GA SA310310999GA SA310321000GA SA310321002GA SA310321004GA SA310321005GA SA310321006GA SA310321008GA SA310321009GA SA310321011GA SA310321012GA SA310321016GA SA310321021GA SA310321026GA SA310321027GA SA310321028GA SA310321029GA SA310321030GA SA310321031GA SA310321032GA SA310321033GA SA310321034GA SA310321035GA SA310321036GA SA310321037GA SA310321038GA SA310321039GA SA310321044GA SA310321051GA SA310321053GA SA310321054GA SA310259898GA SA310310841GA SA310310975GA SA310321013GA SA310321015GA SA310321017GA SA310321018GA SA310321022GA SA310321025GA SA310321045GA SA310321058GA SA310321059GA SA310321064GA SA310321074GA SA310321077GA SA310321078GA SA310331117GA SA310331118GA SA310331126GA SA310331127GA SA310361632GA SA310361636GA SA310361638GA SA310361888GA SA310361889GA SA310361904GA SA310361909GA SA310361910GA SA312294374GA SA312294387GA SA312263874GA SA312263891GA SA312263909GA SA312263929GA SA312263932GA SA312263941GA SA312263942GA SA312263945GA SA312263947GA SA312294515GA SA312294629GA SA312294633GA SA312294637GA SA312294639GA SA312294640GA SA312294647GA SA312294648GA SA312242968GA SA312253295GA SA312263898GA SA312294418GA SA312294421GA SA312294423GA SA312294434GA SA312294455GA SA312294466GA SA312294470GA SA312294472GA SA312294495GA SA312294508GA SA312294509GA SA312294511GA SA312294512GA SA312294513GA SA312294514GA SA312294521GA SA312294523GA SA312294534GA SA312294536GA SA312294537GA SA312294538GA SA312294539GA SA312294540GA SA312294541GA SA312294542GA SA312294543GA SA312294545GA SA312294547GA SA312294549GA SA312294550GA SA312294553GA SA312294554GA SA312294557GA SA312294562GA SA312294563GA SA312294565GA SA312294567GA SA312294572GA SA312294573GA SA312294574GA SA312294575GA SA312294576GA SA312294577GA SA312294580GA SA312294583GA SA312294585GA SA312294596GA SA312294610GA SA312294620GA SA312294624GA SA312294638GA SA312294643GA SA312294645GA SA312294652GA SA312294668GA SA312397061GA SA312397081GA SA312397203GA SA312407295GA SA312417305GA SA312417307GA SA312417309GA SA312417310GA SA312417312GA SA312417314GA SA312417319GA SA312417324GA SA312396818GA SA312397116GA SA312397117GA SA312397128GA SA312397134GA SA312397148GA SA312355479GA SA312396705GA SA312396736GA SA312396740GA SA312396766GA SA312396868GA SA312396958GA SA312397227GA SA312417302GA SA312417303GA SA312417304GA SA312417308GA SA312417322GA SA312417325GA SA312417327GA SA312397132GA SA312397150GA SA312397211GA SA312397212GA SA312397213GA SA312397214GA SA312397215GA SA312397217GA SA312397218GA SA312397219GA SA312397220GA SA312397222GA SA312397223GA SA312397225GA SA312397226GA SA312397228GA SA312397229GA SA312397230GA SA312397231GA SA312397232GA SA312397233GA SA312397235GA SA312397236GA SA312397237GA SA312397238GA SA312397239GA SA312397240GA SA312397242GA SA312397243GA SA312397244GA SA312397245GA SA312397247GA SA312397248GA SA312397249GA SA312397250GA SA312397255GA SA312397256GA SA312397258GA SA312397259GA SA312397260GA SA312397261GA SA312397262GA SA312397263GA SA312397264GA SA312397265GA SA312397266GA SA312397267GA SA312397268GA SA312397270GA SA312376200GA SA312396722GA SA312396895GA SA312397113GA SA312417315GA SA312417326GA SA312417331GA SA312417333GA SA312417338GA SA312417345GA SA312417350GA SA312417352GA SA312417353GA SA312417354GA SA312417356GA SA312417357GA SA312417358GA SA312417360GA SA312417361GA SA312417363GA SA312417370GA SA312417376GA SA312417377GA SA312417381GA SA312417383GA SA312417388GA SA312417389GA SA312417394GA SA312417396GA SA312417411GA SA312417413GA SA312417414GA SA312417415GA SA312417416GA SA312458128GA SA312458129GA SA312458131GA SA312458132GA SA312458133GA SA312458134GA SA312458135GA SA312458136GA SA312458137GA SA312458140GA SA312458141GA SA312458142GA SA312458143GA SA312458144GA SA312458146GA SA312458148GA SA312458149GA SA312458150GA SA312458152GA SA312458153GA SA312458154GA SA312458155GA SA312458158GA SA312458159GA SA312458161GA SA312458162GA SA312458163GA SA312458164GA SA312458165GA SA312458166GA SA312458169GA SA312458170GA SA312458171GA SA312458172GA SA312458173GA SA312458174GA SA312458175GA SA312458177GA SA312468605GA SA312468607GA SA312468613GA SA312468655GA SA312468662GA SA312468663GA SA312468670GA SA312468671GA SA312468674GA SA312468692GA SA312468697GA SA312468699GA SA312468700GA SA312468705GA SA312468706GA SA312468717GA SA312468718GA SA312468808GA SA312468812GA SA312468827GA SA312468829GA SA312468833GA SA312468837GA SA312468839GA SA312468840GA SA312468845GA SA312468846GA SA312468865GA SA312468868GA SA312468870GA SA312468873GA SA312468874GA SA312468875GA SA312468876GA SA312468879GA SA312468881GA SA312468882GA SA312468883GA SA312468884GA SA312468885GA SA312468886GA SA312468887GA SA312468888GA SA312468889GA SA312468892GA SA312468893GA SA312468895GA SA312468896GA SA312468897GA SA312468898GA SA312468899GA SA312468903GA SA312468904GA SA312468905GA SA312468906GA SA312468908GA SA312468909GA SA312468911GA SA312468912GA SA312468913GA SA312468914GA SA312468915GA SA312468916GA SA312468917GA SA312468918GA SA312468920GA SA312468921GA SA312468928GA SA312468939GA SA312468942GA SA312468943GA SA312468948GA SA312468949GA SA312469012GA SA312469074GA SA312469228GA SA312469230GA SA312469231GA SA312469233GA SA312469234GA SA312469235GA SA312469236GA SA312469237GA SA312469238GA SA312469239GA SA312469240GA SA312469241GA SA312469242GA SA312469245GA SA312469247GA SA312469255GA SA312469256GA SA312469259GA SA312469269GA SA312469270GA SA312469271GA SA312469272GA SA312469273GA SA312469275GA SA312469276GA SA312469277GA SA312469278GA SA312469372GA SA312469376GA SA312469378GA SA312469409GA SA312469458GA SA312469465GA SA312469472GA SA311184716GA SA311468387GA SA311508577GA SA311518698GA SA311518711GA SA311518725GA SA312018873GA SA312018876GA SA312018877GA SA312018878GA SA312018879GA SA312018880GA SA312018885GA SA312018894GA SA312028912GA SA312028913GA SA312028920GA SA312028923GA SA312028924GA SA312028925GA SA312028930GA SA312028931GA SA312028934GA SA312028937GA SA312028939GA SA312028942GA SA312028943GA SA312028944GA SA312028945GA SA312028946GA SA312028947GA SA312028951GA SA312028952GA SA312028953GA SA312028954GA SA312028955GA SA312028956GA SA312028957GA SA312028958GA SA312028959GA SA312028960GA SA312028961GA SA312028962GA SA312028963GA SA312028964GA SA312028965GA SA312028968GA SA312028969GA SA312028970GA SA312028971GA SA312028972GA SA312028973GA SA312028974GA SA312028975GA SA312028976GA SA312028977GA SA312028981GA SA312028983GA SA312028985GA SA312029000GA SA312029013GA SA312029029GA SA312029037GA SA312029041GA SA312029042GA SA312029044GA SA312029049GA SA312029070GA SA312080870GA SA312080879GA SA312069784GA SA312079978GA SA312090918GA SA312090919GA SA312090920GA SA312090922GA SA312090924GA SA312090925GA SA312090926GA SA312090929GA SA312090930GA SA312090932GA SA312090933GA SA312090935GA SA312090936GA SA312090937GA SA312090938GA SA312090939GA SA312090940GA SA312090941GA SA312090943GA SA312090944GA SA312090945GA SA312090946GA SA312090947GA SA312090949GA SA312090950GA SA312090951GA SA312090952GA SA312090953GA SA312090954GA SA312090955GA SA312090956GA SA312090957GA SA312090958GA SA312090959GA SA312090960GA SA312090961GA SA312090962GA SA312090963GA SA312090964GA SA312090965GA SA312090968GA SA312090971GA SA312090973GA SA312090975GA SA312090976GA SA312090977GA SA312090979GA SA312090980GA SA312090984GA SA312090986GA SA312090991GA SA312090992GA SA312091008GA SA312091012GA SA312091016GA SA312070277GA SA312091005GA SA312101432GA SA312111464GA SA312111474GA SA312111475GA SA312111477GA SA312111478GA SA312111480GA SA312111481GA SA312111490GA SA312111494GA SA312111497GA SA312111498GA SA312111501GA SA312111592GA SA312111593GA SA312111594GA SA312111595GA SA312111596GA SA312111597GA SA312111599GA SA312111600GA SA312111601GA SA312111602GA SA312111603GA SA312111604GA SA312111605GA SA312111606GA SA312111607GA SA312111608GA SA312111609GA SA312111610GA SA312111611GA SA312111612GA SA312111613GA SA312111614GA SA312111615GA SA312111616GA SA312111617GA SA312111618GA SA312111619GA SA312111620GA SA312111621GA SA312111622GA SA312111623GA SA312111624GA SA312111625GA SA312111626GA SA312111627GA SA312111628GA SA312111634GA SA312111635GA SA312111637GA SA312111638GA SA312111639GA SA312111641GA SA308297060GA SA308297064GA SA308337646GA SA308337647GA SA308337648GA SA308337652GA SA308337667GA SA308337673GA SA308337703GA SA308337706GA SA308337707GA SA308337709GA SA308337710GA SA308337711GA SA308337712GA SA308337713GA SA308337714GA SA308337715GA SA308337716GA SA308337717GA SA308337718GA SA308337719GA SA308337720GA SA308337721GA SA308337722GA SA308347772GA SA308347798GA SA309173169GA SA309193294GA SA309193296GA SA309193297GA SA309193317GA SA309193321GA SA309193323GA SA309193324GA SA309193325GA SA309173142GA SA309173167GA SA309173172GA SA309173183GA SA309173185GA SA309193303GA SA309193305GA SA309193307GA SA309193313GA SA309193319GA SA309193351GA SA309193353GA SA309203372GA SA309203416GA SA309203418GA SA309203423GA SA309243659GA SA309243704GA SA309243709GA SA309253729GA SA309253733GA SA309253734GA SA309253735GA SA309253736GA SA309253741GA SA309253742GA SA309253749GA SA309253751GA SA309253752GA SA309253821GA SA309253836GA SA309253838GA SA309253841GA SA309253842GA SA309253843GA SA309253844GA SA309253847GA SA309253864GA SA309253868GA SA309253870GA SA309253875GA SA309253876GA SA309253877GA SA309253884GA SA309253889GA SA309253892GA SA309253893GA SA309253897GA SA309253910GA SA309253912GA SA308327338GA SA309304230GA SA309304235GA SA309324449GA SA309324456GA SA309324466GA SA309324475GA SA309324483GA SA309324486GA SA309324490GA SA309324492GA SA309324506GA SA309354983GA SA309354994GA SA309375431GA SA309385684GA SA309395704GA SA309395753GA SA309395809GA SA309426062GA SA309426072GA SA309426073GA SA309426074GA SA309426076GA SA309426080GA SA309426082GA SA309426092GA SA309426098GA SA309426118GA SA309426123GA SA309436267GA SA309436276GA SA309405820GA SA309405833GA SA309405839GA SA309426024GA SA309426068GA SA309426088GA SA309426094GA SA309426097GA SA309426101GA SA309426126GA SA309426135GA SA309426138GA SA309426142GA SA309426143GA SA309426145GA SA309426151GA SA309426153GA SA309426164GA SA309436169GA SA309436173GA SA309436174GA SA309436175GA SA309436185GA SA309436193GA SA309436209GA SA309436211GA SA309436213GA SA309436214GA SA309436216GA SA309436221GA SA309436222GA SA309436225GA SA309375421GA SA309405829GA SA309446483GA SA309456733GA SA309466749GA SA309466762GA SA309466764GA SA309466765GA SA309466766GA SA309466788GA SA309476945GA SA309476951GA SA309476952GA SA309446453GA SA309446455GA SA309446495GA SA309446508GA SA310028170GA SA310028246GA SA310048598GA SA309538053GA SA309538073GA SA309517889GA SA309517893GA SA309517896GA SA309527929GA SA309527958GA SA310038351GA SA310038384GA SA310038392GA SA310038401GA SA310038404GA SA310038405GA SA310048607GA SA310048626GA SA310068867GA SA310068891GA SA309517784GA SA310048594GA SA310038370GA SA310038455GA SA310118980GA SA310068851GA SA310068855GA SA310068856GA SA310068860GA SA310068862GA SA310068874GA SA310068894GA SA310068903GA SA310068906GA SA310068912GA SA310068920GA SA310068935GA SA310149311GA SA310159370GA SA310159381GA SA310159463GA SA310159464GA SA310159465GA SA310159466GA SA310159467GA SA310159468GA SA310159469GA SA310260223GA SA310260349GA SA310139252GA SA310159460GA SA310159462GA SA310159474GA SA310169509GA SA310169514GA SA310169516GA SA310169518GA SA310169519GA SA310169521GA SA310169522GA SA310169524GA SA310169525GA SA310169528GA SA310169530GA SA310169531GA SA310169534GA SA310169535GA SA310169536GA SA310169537GA SA310199722GA SA310199748GA SA310260323GA SA310260329GA SA310260338GA SA310260340GA SA310260344GA SA310260345GA SA310260475GA SA310260485GA SA310260492GA SA310260497GA SA310260498GA SA310260500GA SA310290636GA SA310290700GA SA310413485GA SA310413486GA SA310423653GA SA310423843GA SA310423928GA SA310423948GA SA310423955GA SA310423965GA SA310423968GA SA310424007GA SA310424011GA SA310424015GA SA310424016GA SA310424021GA SA310424024GA SA310423687GA SA310423998GA SA310424005GA SA310424018GA SA310424020GA SA310424022GA SA310424023GA SA310260256GA SA310361677GA SA310372305GA SA310413523GA SA310434226GA SA310444267GA SA310444317GA SA310444318GA SA310444319GA SA310444320GA SA310444321GA SA310444356GA SA310444357GA SA310444359GA SA310444363GA SA310444488GA SA310444505GA SA310444508GA SA310444512GA SA310444515GA SA310444516GA SA310444525GA SA310444528GA SA310444529GA SA310444530GA SA310444535GA SA310444841GA SA310444855GA SA310444858GA SA310444866GA SA310475493GA SA310475500GA SA310475559GA SA310475560GA SA310475561GA SA310475563GA SA310475605GA SA310475659GA SA310475712GA SA310485769GA SA310485803GA SA310485817GA SA310485823GA SA310485831GA SA310485849GA SA310485852GA SA310485864GA SA310485865GA SA310485866GA SA310485872GA SA310485875GA SA310485887GA SA310454977GA SA310465344GA SA310475632GA SA310475711GA SA310485825GA SA310485827GA SA310485830GA SA310485837GA SA310485854GA SA310485855GA SA310485879GA SA310485886GA SA310496213GA SA310496217GA SA310496219GA SA310496234GA SA310496235GA SA310496239GA SA310496240GA SA310506659GA SA310506664GA SA310506666GA SA310506667GA SA310506671GA SA310506699GA SA310506705GA SA310506712GA SA310506847GA SA310506891GA SA310506669GA SA310506752GA SA310506850GA SA310506863GA SA310507033GA SA310496130GA SA310496133GA SA310496136GA SA310496150GA SA310496153GA SA310496158GA SA310496312GA SA310496320GA SA310506662GA SA310506677GA SA310506685GA SA310506686GA SA310506693GA SA310506701GA SA310506718GA SA310506723GA SA310507085GA SA310517562GA SA310517572GA SA310517575GA SA310517576GA SA310507510GA SA310507500GA SA310507120GA SA310507487GA SA310382624GA SA310434202GA SA310518622GA SA310518641GA SA310518644GA SA310529143GA SA310529144GA SA310529151GA SA310529154GA SA310529209GA SA310403377GA SA310403412GA SA310403430GA SA310423791GA SA310434060GA SA310434100GA SA310434101GA SA310434112GA SA310434117GA SA310259904GA SA310507348GA SA310518422GA SA310518489GA SA310518494GA SA310518496GA SA310518506GA SA310518525GA SA310539289GA SA310496298GA SA311060136GA SA311060138GA SA311060140GA SA311060167GA SA311060208GA SA311060230GA SA311069728GA SA311069731GA SA311069732GA SA311101275GA SA311069863GA SA311090900GA SA311090907GA SA311090918GA SA311090920GA SA311090933GA SA311091034GA SA311091035GA SA311091049GA SA311091053GA SA311101099GA SA311101114GA SA311101123GA SA311080633GA SA311091004GA SA311132288GA SA311122037GA SA311122038GA SA311090783GA SA311153208GA SA311163375GA SA311163379GA SA311163380GA SA311163381GA SA311163384GA SA311163391GA SA311163392GA SA311163395GA SA311163399GA SA311163413GA SA311163418GA SA311163419GA SA311163421GA SA311163887GA SA311163956GA SA311060156GA SA311060225GA SA311468393GA SA311468398GA SA311468401GA SA311468403GA SA310539359GA SA311327969GA SA311039434GA SA311458279GA SA310300775GA SA310300782GA SA310310844GA SA310310849GA SA310310858GA SA310310869GA SA310310870GA SA310310873GA SA310310874GA SA310310884GA SA310310891GA SA310310892GA SA310310893GA SA310310894GA SA310310911GA SA310321050GA SA310321065GA SA310321071GA SA312263617GA SA312263978GA SA312264013GA SA312264030GA SA312264038GA SA312264054GA SA312264060GA SA312294446GA SA312294452GA SA312294462GA SA312049310GA SA312059527GA SA312069695GA SA312069841GA SA312070050GA SA312091152GA SA312101395GA SA312101398GA SA312101400GA SA312101403GA SA312101413GA SA312101442GA SA312111459GA SA312111503GA SA312192684GA SA312192686GA SA312192701GA SA312192732GA SA312202740GA SA312232891GA SA312253202GA SA312253218GA SA312253460GA SA312253480GA SA312253524GA SA312253059GA SA312253066GA SA312253121GA SA312253131GA SA312253299GA SA312263928GA SA312263979GA SA312284337GA SA312294627GA SA312294792GA SA312365798GA SA312375891GA SA312375892GA SA312375893GA SA312375894GA SA312375896GA SA312375897GA SA312375899GA SA312375902GA SA312375903GA SA312375905GA SA312375906GA SA312375908GA SA312375909GA SA312375910GA SA312375911GA SA312375915GA SA312375917GA SA312375918GA SA312375919GA SA312375921GA SA312375922GA SA312375925GA SA312375929GA SA312375931GA SA312375933GA SA312375936GA SA312375943GA SA312375945GA SA312375946GA SA312375948GA SA312375949GA SA312376050GA SA312376289GA SA312376294GA SA312378025GA SA312378031GA SA312376178GA SA312386314GA SA312376025GA SA312378012GA SA312378014GA SA312378015GA SA312378018GA SA312378020GA SA312378022GA SA312378026GA SA312378028GA SA312397074GA SA312397083GA SA312397087GA SA312386679GA SA312396810GA SA312396827GA SA312396834GA SA312396837GA SA312396842GA SA312396857GA SA312396858GA SA312396864GA SA312396871GA SA312396883GA SA312396886GA SA312396889GA SA312396892GA SA312396894GA SA312396898GA SA312396902GA SA312396904GA SA312396905GA SA312417390GA SA312417417GA SA312470006GA SA312479883GA SA312479887GA SA312447514GA SA312480111GA SA312480112GA SA312447451GA SA312447458GA SA312447567GA SA312490875GA SA312491434GA SA312491441GA SA312491453GA SA309405831GA SA309405835GA SA311287095GA SA311287163GA SA311297230GA SA311297248GA SA311297252GA SA311297295GA SA311297296GA SA311297305GA SA311297308GA SA311297309GA SA311297312GA SA311297315GA SA311297317GA SA311297318GA SA311297319GA SA311297320GA SA311297323GA SA311307459GA SA311307477GA SA311307546GA SA311317609GA SA311317632GA SA311317674GA SA311317686GA SA311317744GA SA311317753GA SA311317756GA SA311317760GA SA311327856GA SA311327905GA SA311327914GA SA311327919GA SA311327925GA SA311337978GA SA311428023GA SA312039073GA SA312039116GA SA312039119GA SA312039124GA SA312039145GA SA312039146GA SA312039160GA SA312039170GA SA312039176GA SA312049417GA SA312070021GA SA312069723GA SA312079997GA SA312080722GA SA312080747GA SA312080766GA SA312080773GA SA312080793GA SA312080816GA SA312080832GA SA312070174GA SA312069828GA SA312091159GA SA312091175GA SA312091176GA SA312091177GA SA312091179GA SA312091188GA SA312091189GA SA312091193GA SA312091195GA SA312091198GA SA312091199GA SA312091205GA SA312091212GA SA312091214GA SA312091217GA SA312091228GA SA312070003GA SA312070012GA SA312070099GA SA312079954GA SA312079969GA SA312152346GA SA312192670GA SA312192680GA SA312192683GA SA312192697GA SA312192698GA SA312152493GA SA311090891GA SA311297300GA SA307504124GA SA309062288GA SA309072367GA SA309072369GA SA309072371GA SA309072373GA SA309072381GA SA309072386GA SA309072388GA SA309072389GA SA309072390GA SA309072392GA SA309072394GA SA309072397GA SA309072402GA SA309072403GA SA309072404GA SA309072405GA SA309072406GA SA309072407GA SA309072408GA SA309072409GA SA309072411GA SA309082412GA SA309082414GA SA309082416GA SA309082447GA SA309082455GA SA309082458GA SA309082461GA SA309082463GA SA309112561GA SA309122668GA SA309122718GA SA309122723GA SA309122749GA SA309122767GA SA309132845GA SA309132847GA SA309132848GA SA309132849GA SA309132850GA SA309132851GA SA309132852GA SA309132853GA SA309132855GA SA309142984GA SA309142985GA SA309142986GA SA309142992GA SA309153024GA SA309153025GA SA309153027GA SA309153032GA SA309153033GA SA309153047GA SA309294117GA SA309324354GA SA309334594GA SA309334673GA SA309334731GA SA309334733GA SA309334736GA SA309334738GA SA309334742GA SA309334746GA SA309334750GA SA309344835GA SA309344840GA SA309344841GA SA309344844GA SA309344846GA SA309344849GA SA309344858GA SA309344868GA SA309354894GA SA309354895GA SA309354902GA SA309354908GA SA309354912GA SA309354918GA SA309354920GA SA309354925GA SA309354938GA SA309354943GA SA309354944GA SA309354947GA SA307210814GA SA307210816GA SA307210830GA SA307210834GA SA307220964GA SA307221091GA SA307231206GA SA307231258GA SA307210808GA SA307210859GA SA307220979GA SA307221187GA SA307241511GA SA307241576GA SA307241692GA SA307221174GA SA307251804GA SA307261807GA SA307261808GA SA307261809GA SA307261834GA SA307291908GA SA307291909GA SA307291915GA SA307291916GA SA307291917GA SA307291919GA SA307291921GA SA307291922GA SA307261810GA SA307261815GA SA307301999GA SA307393028GA SA307231214GA SA307291890GA SA307291896GA SA307291902GA SA307302025GA SA307302031GA SA307291867GA SA307302020GA SA307261832GA SA307352361GA SA307352372GA SA307362409GA SA307362410GA SA307362411GA SA307362422GA SA307362425GA SA307362428GA SA307362432GA SA307362433GA SA307362437GA SA307382533GA SA307382571GA SA307382572GA SA307382580GA SA307382588GA SA307382590GA SA307382598GA SA307382602GA SA307382606GA SA307382607GA SA307382462GA SA307382487GA SA307382512GA SA307251725GA SA307291944GA SA307302013GA SA307342304GA SA307352365GA SA307382562GA SA307382570GA SA307382585GA SA307443257GA SA307443267GA SA307443378GA SA307453393GA SA307453431GA SA307453552GA SA307453595GA SA307504194GA SA307514322GA SA307514324GA SA307514327GA SA307514344GA SA308094868GA SA308094876GA SA308094948GA SA308094952GA SA308094957GA SA308094970GA SA308115308GA SA308155940GA SA308155948GA SA308327422GA SA308327423GA SA308327443GA SA308327476GA SA308327484GA SA308327489GA SA308327490GA SA308327498GA SA308327504GA SA308327522GA SA308327534GA SA308327538GA SA308327539GA SA308327540GA SA308327541GA SA308327552GA SA308327553GA SA308337559GA SA308337560GA SA308337561GA SA308337593GA SA308337631GA SA308337636GA SA307291906GA SA308135534GA SA308135544GA SA308166071GA SA308186356GA SA308186418GA SA308196507GA SA308196508GA SA308196519GA SA308196543GA SA308196544GA SA308196566GA SA308206590GA SA308206599GA SA308206600GA SA308206627GA SA308206642GA SA308206663GA SA308206699GA SA308206753GA SA308216771GA SA308216778GA SA308297029GA SA308297035GA SA308297037GA SA308297048GA SA308297051GA SA308297053GA SA308297058GA SA308297034GA SA308297038GA SA308297040GA SA308297042GA SA308297044GA SA308297045GA SA308297046GA SA308297047GA SA308297052GA SA308297056GA SA308297057GA SA308368062GA SA308378296GA SA308388446GA SA308408908GA SA308409072GA SA308409088GA SA308409160GA SA308409164GA SA308409165GA SA308409187GA SA308419245GA SA308419249GA SA308419251GA SA308419271GA SA308398659GA SA308398724GA SA308398726GA SA308398737GA SA308398748GA SA308398767GA SA308398774GA SA308398776GA SA308398862GA SA308398863GA SA308398867GA SA308368202GA SA308388632GA SA308398689GA SA308398902GA SA308409028GA SA308409055GA SA308409084GA SA308409156GA SA308409157GA SA308419246GA SA308368050GA SA308388596GA SA308398856GA SA308408905GA SA308408921GA SA308408937GA SA308408941GA SA308409015GA SA308409023GA SA308409031GA SA309082444GA SA309082445GA SA309082446GA SA309082448GA SA309082451GA SA309082460GA SA309082464GA SA309082465GA SA309082466GA SA309082467GA SA309112601GA SA309132857GA SA309142880GA SA309142922GA SA309142923GA SA309142924GA SA309142933GA SA309142937GA SA309142921GA SA309142925GA SA309142927GA SA309142935GA SA309142949GA SA309142952GA SA309142963GA SA308357878GA SA308511518GA SA309173150GA SA309193334GA SA309193354GA SA309193358GA SA309203365GA SA309203369GA SA309203380GA SA309203382GA SA309203402GA SA309223477GA SA309223479GA SA309223480GA SA309223507GA SA309223515GA SA309304239GA SA309304273GA SA309304279GA SA309283993GA SA309283994GA SA309283996GA SA309283998GA SA309284000GA SA309304225GA SA309304233GA SA309304246GA SA309304250GA SA309304254GA SA309304255GA SA309304256GA SA309304260GA SA309304265GA SA309304267GA SA309304270GA SA309304274GA SA309304276GA SA309304281GA SA309284006GA SA309304228GA SA309304241GA SA309304242GA SA309304247GA SA309304257GA SA309304268GA SA309304271GA SA309304285GA SA309304286GA SA309304288GA SA309324446GA SA309324447GA SA309324448GA SA309324450GA SA309324451GA SA309324452GA SA309324454GA SA309324457GA SA309324459GA SA309324460GA SA309324461GA SA309324462GA SA309324463GA SA309324464GA SA309324465GA SA309324467GA SA309324468GA SA309324471GA SA309324474GA SA309324476GA SA309324477GA SA309324478GA SA309324479GA SA309324484GA SA309324497GA SA309304253GA SA309324458GA SA309324469GA SA309324487GA SA309324488GA SA309324489GA SA309324491GA SA309324493GA SA309324495GA SA309324498GA SA309324499GA SA309324500GA SA309324501GA SA309324503GA SA309324505GA SA309324510GA SA309324511GA SA309324513GA SA309324514GA SA309324517GA SA309324518GA SA309324522GA SA309324523GA SA309324524GA SA309324526GA SA309324528GA SA309324529GA SA309324530GA SA309324531GA SA309324535GA SA309324536GA SA309324550GA SA309324552GA SA309324553GA SA309324554GA SA309324555GA SA309253828GA SA309334825GA SA309365088GA SA309365090GA SA309365097GA SA309365102GA SA309365104GA SA309365124GA SA309365125GA SA309365126GA SA309365128GA SA309365132GA SA309365133GA SA309365135GA SA309365136GA SA309365137GA SA309365138GA SA309365139GA SA309365140GA SA309365142GA SA309365144GA SA309334671GA SA309334691GA SA309334698GA SA309334709GA SA309334743GA SA309334749GA SA309344845GA SA309344848GA SA309365290GA SA309385582GA SA309395693GA SA309395694GA SA309395698GA SA309395706GA SA309395713GA SA309395745GA SA309395748GA SA309354901GA SA309385589GA SA309385676GA SA309385689GA SA309395700GA SA309395703GA SA309395707GA SA309395714GA SA309395715GA SA309395720GA SA309395727GA SA309395732GA SA309395735GA SA309395736GA SA309395752GA SA309395762GA SA309436272GA SA309436279GA SA309436281GA SA309436292GA SA309436315GA SA309436348GA SA309456635GA SA309456646GA SA309456653GA SA309456675GA SA309456688GA SA309497146GA SA309497162GA SA309497171GA SA309497230GA SA309497232GA SA309497282GA SA309497308GA SA309517835GA SA310038388GA SA310038393GA SA310038423GA SA310038436GA SA310048614GA SA310048621GA SA310068907GA SA310068931GA SA310139084GA SA310159453GA SA310199694GA SA310199742GA SA310199747GA SA310199751GA SA310199760GA SA310199659GA SA310199662GA SA310199687GA SA310199692GA SA310199693GA SA310199696GA SA310199697GA SA310199698GA SA310199714GA SA310199718GA SA310199726GA SA310199743GA SA310199746GA SA310199749GA SA310199754GA SA310199755GA SA310199761GA SA310199559GA SA310199560GA SA310199563GA SA310199568GA SA310199569GA SA310199571GA SA310199574GA SA310199577GA SA310199582GA SA310199584GA SA310199585GA SA310199586GA SA310199641GA SA310199671GA SA310199681GA SA310199685GA SA310199688GA SA310199700GA SA310199706GA SA310199724GA SA310199729GA SA310199731GA SA310199732GA SA310199733GA SA310199734GA SA310199736GA SA310199740GA SA310199758GA SA310270579GA SA310270580GA SA310270581GA SA310300761GA SA310300762GA SA310300764GA SA310300765GA SA310300767GA SA310300787GA SA310444282GA SA310444367GA SA310444487GA SA310444489GA SA310444491GA SA310444493GA SA310444497GA SA310444500GA SA310444501GA SA310444502GA SA310444517GA SA310444519GA SA310444526GA SA310444532GA SA310444534GA SA310444843GA SA310444844GA SA310444845GA SA310444852GA SA310444857GA SA310444860GA SA310444862GA SA310444865GA SA310444890GA SA310444891GA SA310423881GA SA310424040GA SA310434095GA SA310454931GA SA310454951GA SA310331330GA SA310393238GA SA310403315GA SA310403323GA SA310403329GA SA310403339GA SA310403354GA SA310403356GA SA310403358GA SA310403359GA SA310403360GA SA310403361GA SA310403362GA SA310403367GA SA310403370GA SA310403378GA SA310403434GA SA310423743GA SA310423747GA SA310423751GA SA310423802GA SA310423899GA SA310423977GA SA310423991GA SA310434049GA SA310434057GA SA310444268GA SA310444271GA SA310444273GA SA310444274GA SA310444279GA SA310444280GA SA310444285GA SA310444288GA SA310444290GA SA310444293GA SA310444295GA SA310444297GA SA310444299GA SA310444300GA SA310444303GA SA310444307GA SA310444309GA SA310444310GA SA310444313GA SA310444315GA SA310444316GA SA310444322GA SA310444323GA SA310444366GA SA310444368GA SA310444490GA SA310444492GA SA310444514GA SA310444520GA SA310444522GA SA310444848GA SA310444869GA SA310413516GA SA310423584GA SA310423856GA SA310434230GA SA310444302GA SA310444533GA SA310444853GA SA310444854GA SA310454978GA SA310454980GA SA310454999GA SA310455002GA SA310455008GA SA310455013GA SA310455014GA SA310455019GA SA310455023GA SA310455027GA SA310455049GA SA310455054GA SA310455059GA SA310455064GA SA310455158GA SA310455162GA SA310351539GA SA310423580GA SA310423846GA SA310444511GA SA310444847GA SA310444861GA SA310454930GA SA310454932GA SA310454941GA SA310454942GA SA310454979GA SA310454998GA SA310455005GA SA310455007GA SA310455010GA SA310455020GA SA310455050GA SA310455063GA SA310455171GA SA310413453GA SA310413543GA SA310423693GA SA310423699GA SA310423867GA SA310423877GA SA310423889GA SA310423909GA SA310423911GA SA310423976GA SA310423993GA SA310434105GA SA310455070GA SA310455071GA SA310455072GA SA310455073GA SA310455074GA SA310455117GA SA310455129GA SA310455135GA SA310455137GA SA310455141GA SA310455225GA SA310455226GA SA310455229GA SA310455239GA SA310455240GA SA310455241GA SA310455242GA SA310455243GA SA310455244GA SA310455246GA SA310455250GA SA310455252GA SA310455255GA SA310455258GA SA310455269GA SA310455276GA SA310455281GA SA310455285GA SA310455287GA SA310455290GA SA310465296GA SA310465320GA SA310465336GA SA310465340GA SA310465345GA SA310465353GA SA310465356GA SA310465357GA SA310465360GA SA310465362GA SA310465363GA SA310465365GA SA310465388GA SA310465395GA SA310465399GA SA310465401GA SA310465402GA SA310465404GA SA310465405GA SA310465406GA SA310465407GA SA310465409GA SA310465413GA SA310465415GA SA310465416GA SA310465417GA SA310465418GA SA310465421GA SA310465422GA SA310465424GA SA310465425GA SA310465426GA SA310465428GA SA310465429GA SA310465430GA SA310465434GA SA310465440GA SA310465442GA SA310465444GA SA310465447GA SA310465448GA SA310475450GA SA310475454GA SA310475455GA SA310475456GA SA310475457GA SA310475459GA SA310475462GA SA310475463GA SA310475469GA SA310475472GA SA310475475GA SA310475476GA SA310475478GA SA310475486GA SA310475491GA SA310475492GA SA310475502GA SA310475503GA SA310475506GA SA310475507GA SA310475512GA SA310475514GA SA310403382GA SA310403387GA SA310423732GA SA310424035GA SA310434068GA SA310455001GA SA310455021GA SA310455032GA SA310455051GA SA310455089GA SA310455167GA SA310455169GA SA310465351GA SA310465355GA SA310465358GA SA310465364GA SA310465372GA SA310465373GA SA310465375GA SA310465378GA SA310465382GA SA310465384GA SA310465387GA SA310465389GA SA310465390GA SA310465392GA SA310465393GA SA310424043GA SA310475461GA SA310475521GA SA310475453GA SA310506679GA SA310518244GA SA310518253GA SA310518254GA SA310518328GA SA310518331GA SA310518338GA SA310518339GA SA310518340GA SA310518344GA SA310518405GA SA310518469GA SA310506673GA SA310506680GA SA310506684GA SA310506687GA SA310506694GA SA310506695GA SA310506714GA SA310506716GA SA310506724GA SA310507071GA SA310507079GA SA310507390GA SA310507423GA SA310517573GA SA310517698GA SA310517699GA SA310517701GA SA310517702GA SA310517705GA SA310517707GA SA310517713GA SA310517722GA SA310517725GA SA310517827GA SA310517867GA SA310517868GA SA310518012GA SA310518646GA SA310518650GA SA310518667GA SA310518668GA SA310518669GA SA310518707GA SA310518708GA SA310518709GA SA310518722GA SA310496336GA SA310507066GA SA310507081GA SA310507101GA SA310507136GA SA310517590GA SA310517826GA SA310518005GA SA310517832GA SA310517874GA SA310518248GA SA310518281GA SA310518284GA SA310518287GA SA310518290GA SA310518302GA SA310518316GA SA310518329GA SA310507450GA SA310518701GA SA311039383GA SA311039426GA SA311039492GA SA311039427GA SA311039487GA SA310434065GA SA310529077GA SA311121812GA SA311132283GA SA311121713GA SA311121917GA SA311163405GA SA311163859GA SA311163863GA SA311163867GA SA311163877GA SA311163883GA SA311163889GA SA311163898GA SA311163909GA SA311163918GA SA311163920GA SA311163940GA SA311163945GA SA311163958GA SA311163960GA SA311163964GA SA311163972GA SA311163976GA SA311163978GA SA311163983GA SA311163984GA SA311163987GA SA311163988GA SA311163994GA SA311163997GA SA311153049GA SA311174177GA SA311174207GA SA311174210GA SA311174211GA SA311184625GA SA311184872GA SA312253284GA SA312264032GA SA312049421GA SA312111547GA SA312111550GA SA312111553GA SA312111555GA SA312111574GA SA312111576GA SA312111579GA SA312111583GA SA307140602GA SA307140609GA SA307140624GA SA307140626GA SA307140633GA SA307140634GA SA307140635GA SA307140643GA SA307301991GA SA307302014GA SA307423179GA SA307221157GA SA307231261GA SA307291858GA SA307291865GA SA307302024GA SA307291870GA SA307382452GA SA307504158GA SA307514347GA SA307514351GA SA307514392GA SA307514393GA SA307514394GA SA307514395GA SA307514396GA SA307514401GA SA307514403GA SA307514406GA SA307514408GA SA307514412GA SA307514429GA SA307514436GA SA307514437GA SA307514439GA SA307514440GA SA307514441GA SA307514442GA SA307514443GA SA307514447GA SA307514448GA SA307514449GA SA307514452GA SA307514454GA SA307514455GA SA307514457GA SA307524459GA SA307524473GA SA307524474GA SA307524475GA SA307524478GA SA307524481GA SA307524482GA SA307524483GA SA308044496GA SA308135528GA SA308368077GA SA308459854GA SA309375490GA SA310139243GA SA310149318GA SA310159365GA SA310159372GA SA310159384GA SA310159388GA SA310159390GA SA310159471GA SA310290656GA SA310434126GA SA310434128GA SA310434129GA SA310260398GA SA310341378GA SA310423891GA SA310424041GA SA310434061GA SA310434080GA SA310434116GA SA310434184GA SA310434191GA SA310455136GA SA310455238GA SA310455253GA SA310455270GA SA310455274GA SA310403318GA SA310423754GA SA310423782GA SA310423870GA SA310423884GA SA310423894GA SA310423898GA SA310423905GA SA310423972GA SA310423984GA SA310423992GA SA310434047GA SA310434073GA SA310434086GA SA310434127GA SA310434130GA SA310434144GA SA310434149GA SA310434159GA SA310434161GA SA310434163GA SA310434164GA SA310434165GA SA310434167GA SA310434169GA SA310434175GA SA310434176GA SA310434177GA SA310434193GA SA310455068GA SA310455116GA SA310455231GA SA310455288GA SA310506756GA SA310506759GA SA310506761GA SA310506775GA SA310506781GA SA310506791GA SA310506792GA SA310507158GA SA310507221GA SA310507263GA SA310507268GA SA310507287GA SA310507290GA SA310507088GA SA310507391GA SA310507405GA SA310507425GA SA310507440GA SA310507446GA SA310507466GA SA310507472GA SA310507530GA SA310517727GA SA310517740GA SA310517741GA SA310518455GA SA310518478GA SA310506806GA SA310507206GA SA310518322GA SA311059579GA SA311059591GA SA311059604GA SA311069662GA SA311069672GA SA311069691GA SA310475513GA SA310475570GA SA311060121GA SA311060246GA SA311039521GA SA311069752GA SA311069756GA SA311070526GA SA311070530GA SA311122189GA SA311122198GA SA311122203GA SA311122220GA SA311132279GA SA311132286GA SA311101133GA SA311101172GA SA311111628GA SA311122096GA SA311132471GA SA311132497GA SA311132501GA SA311121910GA SA311132564GA SA311132568GA SA311142884GA SA311142893GA SA311142897GA SA311142902GA SA311142906GA SA311184380GA SA311184391GA SA311184393GA SA311195071GA SA311256703GA SA311266909GA SA311194896GA SA311195208GA SA311184570GA SA311205378GA SA311205386GA SA311508616GA SA310351470GA SA310351488GA SA310351489GA SA310351498GA SA310351511GA SA310351512GA SA310351515GA SA310351522GA SA310351571GA SA310351586GA SA310361896GA SA310392940GA SA310392941GA SA310392945GA SA310392948GA SA310392956GA SA310392958GA SA310392960GA SA310392961GA SA310393016GA SA310393018GA SA310393021GA SA310393022GA SA310393043GA SA310393046GA SA310393047GA SA312253237GA SA312253266GA SA312253271GA SA312253282GA SA312294389GA SA312294390GA SA312294396GA SA312294398GA SA312458245GA SA312480367GA SA312480369GA SA312480732GA SA312480856GA SA312480857GA SA312480868GA SA312490871GA SA312018853GA SA312018854GA SA312018855GA SA312018858GA SA312018859GA SA312049239GA SA312049240GA SA312049266GA SA312080469GA SA312080530GA SA312080547GA SA312080572GA SA312080576GA SA312080580GA SA312080767GA SA312080829GA SA312121803GA SA307020375GA SA307020381GA SA307020385GA SA307020387GA SA307080481GA SA307190714GA SA307190719GA SA307190723GA SA307190735GA SA307190740GA SA307190747GA SA307190748GA SA307200779GA SA307200790GA SA307210807GA SA307210809GA SA307210811GA SA307220883GA SA307220889GA SA307220895GA SA307220899GA SA307220927GA SA307220965GA SA307020374GA SA307030404GA SA307030408GA SA307060435GA SA307190715GA SA307190725GA SA307210817GA SA307210860GA SA307210861GA SA307210862GA SA307220873GA SA307220874GA SA307220881GA SA307220882GA SA307220884GA SA307220886GA SA307220912GA SA307220918GA SA307220919GA SA307220926GA SA307220929GA SA307220930GA SA307220962GA SA307220971GA SA307220972GA SA307220973GA SA307220974GA SA307220977GA SA307220980GA SA307220994GA SA307221015GA SA307221025GA SA307221082GA SA307221083GA SA307221098GA SA307221151GA SA307221155GA SA307221184GA SA307231216GA SA307220922GA SA307220934GA SA307220995GA SA307220998GA SA307221029GA SA307221146GA SA307231380GA SA307251730GA SA307291889GA SA307291941GA SA307291942GA SA307291945GA SA307291947GA SA307291956GA SA307291960GA SA307302009GA SA307514354GA SA308155839GA SA308155842GA SA308155848GA SA308155859GA SA308155871GA SA308166109GA SA308166116GA SA308409080GA SA308409086GA SA308409178GA SA308419273GA SA308409068GA SA308511450GA SA309375498GA SA309385509GA SA309385569GA SA309385580GA SA309426108GA SA309436324GA SA309436360GA SA309334726GA SA309507342GA SA309517685GA SA309517693GA SA309517708GA SA309517833GA SA309517895GA SA309517907GA SA309527964GA SA309507376GA SA309507387GA SA309507413GA SA309517634GA SA309538028GA SA309538032GA SA309538035GA SA309538036GA SA309538037GA SA309538039GA SA309538041GA SA309538047GA SA309538088GA SA310028151GA SA310028153GA SA310028173GA SA310028175GA SA310028178GA SA310028179GA SA310028188GA SA310028190GA SA310028205GA SA310028209GA SA310028238GA SA310028239GA SA310028241GA SA310028245GA SA310028247GA SA310028248GA SA310028249GA SA310028251GA SA310028253GA SA310028254GA SA310028257GA SA310028259GA SA310028260GA SA310028264GA SA310028279GA SA310028280GA SA310028283GA SA310028295GA SA310028296GA SA310068888GA SA310068913GA SA310068917GA SA311184515GA SA311184533GA SA311448205GA SA310331263GA SA312069820GA SA312069825GA SA312069831GA SA312069833GA SA312069848GA SA312069851GA SA312069853GA SA311142681GA SA311142691GA SA311142701GA SA311142704GA SA311142715GA SA311153057GA SA312049322GA SA312049337GA SA312049338GA SA312049343GA SA312049349GA SA312049352GA SA312049455GA SA312059482GA SA312059485GA SA312059488GA SA312059499GA SA312080565GA SA312080737GA SA312080745GA SA312080776GA SA312080815GA SA312080866GA SA312080849GA SA312080823GA SA312080847GA SA312080851GA SA312080859GA SA312080915GA SA312122011GA SA312122021GA SA308511515GA SA311184634GA SA311287145GA SA311307504GA SA311317705GA SA311317710GA SA311317712GA SA311317713GA SA311317723GA SA311317738GA SA311327805GA SA311327806GA SA311327813GA SA311327815GA SA311327828GA SA311327831GA SA311508587GA SA309263915GA SA311080674GA SA312079964GA SA312079973GA SA312079986GA SA312079990GA SA309304258GA SA309304275GA SA309304277GA SA309324453GA SA309324455GA SA309324480GA SA309324494GA SA309324496GA SA309324507GA SA309324508GA SA309324509GA SA309324512GA SA309334788GA SA309334812GA SA309365114GA SA309365190GA SA309334765GA SA308105217GA SA309507427GA SA310393236GA SA310393259GA SA310393272GA SA310393275GA SA310393277GA SA310393288GA SA310403336GA SA310403337GA SA310413454GA SA310413455GA SA310413456GA SA310413457GA SA310413459GA SA310413462GA SA310413463GA SA310413464GA SA310413465GA SA310413466GA SA310413469GA SA310413508GA SA310413541GA SA310413542GA SA310413548GA SA310413549GA SA310413555GA SA310413557GA SA310413558GA SA310413564GA SA310413565GA SA310423585GA SA310423586GA SA310423588GA SA310423589GA SA310423591GA SA310423592GA SA310423602GA SA310423608GA SA310423609GA SA310423610GA SA310423615GA SA310423616GA SA310423623GA SA310423656GA SA310423659GA SA310423661GA SA310423664GA SA310423691GA SA310423740GA SA310423746GA SA310423752GA SA310423767GA SA310423768GA SA310496139GA SA310496141GA SA310496460GA SA310496465GA SA310496498GA SA310496515GA SA310518647GA SA310496157GA SA311101481GA SA311101484GA SA311101490GA SA311101492GA SA311101500GA SA311101501GA SA311101508GA SA311101509GA SA311101518GA SA311101524GA SA311101530GA SA311101533GA SA311101548GA SA311101551GA SA311111611GA SA311121704GA SA311121706GA SA311121710GA SA311121715GA SA311121716GA SA311069811GA SA311174209GA SA311174223GA SA311184377GA SA311184541GA SA311184577GA SA311184585GA SA311225978GA SA311225983GA SA311225984GA SA311225990GA SA311236063GA SA311236072GA SA311236073GA SA311236089GA SA311236091GA SA311236095GA SA311236099GA SA311236105GA SA311236107GA SA311236109GA SA311236112GA SA311236113GA SA311236118GA SA311266961GA SA311266986GA SA311184537GA SA311184645GA SA310361911GA SA310372062GA SA310382382GA SA310382384GA SA310382390GA SA310382392GA SA310382393GA SA310382394GA SA310382396GA SA310382436GA SA310382439GA SA310382441GA SA310382448GA SA310382457GA SA310382467GA SA310382471GA SA310382475GA SA310382480GA SA310382481GA SA310382482GA SA310382486GA SA310382489GA SA310382490GA SA310382493GA SA310382496GA SA310382500GA SA310382503GA SA310382506GA SA310382507GA SA310382509GA SA310382510GA SA310382512GA SA310382514GA SA310382515GA SA310382519GA SA310382520GA SA310382521GA SA310382522GA SA312253058GA SA312253192GA SA312253293GA SA312253296GA SA312253301GA SA312253315GA SA312253341GA SA312253351GA SA312345141GA SA312407280GA SA312407283GA SA312457969GA SA312457998GA SA312468555GA SA312479680GA SA312479683GA SA312479687GA SA312479689GA SA312479691GA SA312479699GA SA312479702GA SA312479703GA SA312479715GA SA312479722GA SA312479723GA SA312479728GA SA312479759GA SA312479770GA SA312479796GA SA312479798GA SA312479808GA SA312479812GA SA312479821GA SA312479828GA SA312479830GA SA312479832GA SA312479835GA SA312479843GA SA312479857GA SA312479862GA SA312479863GA SA312479864GA SA312479865GA SA312479867GA SA312479869GA SA312479870GA SA312479873GA SA312479877GA SA312479878GA SA312479879GA SA312479880GA SA312479881GA SA312479882GA SA312479885GA SA312479886GA SA312479888GA SA312479889GA SA312479890GA SA312479894GA SA312469443GA SA312469454GA SA312479559GA SA312479560GA SA312479562GA SA312479563GA SA312479564GA SA312479565GA SA312479567GA SA312479571GA SA312479574GA SA312479576GA SA312479587GA SA312479591GA SA312479597GA SA312479598GA SA312479601GA SA312479602GA SA312479604GA SA312479605GA SA312479609GA SA312479611GA SA312479630GA SA312479632GA SA312479634GA SA312479635GA SA312479636GA SA312479637GA SA312479639GA SA312479640GA SA312447511GA SA312447519GA SA312447520GA SA312447525GA SA312447536GA SA312447538GA SA312447550GA SA312447556GA SA312447561GA SA312447562GA SA312447563GA SA312447565GA SA312447571GA SA312447572GA SA312447575GA SA312447580GA SA312447582GA SA312447600GA SA312480109GA SA312480116GA SA312480117GA SA312059541GA SA312069641GA SA307220981GA SA307231400GA SA308409120GA SA308429399GA SA311428062GA SA307020377GA SA307030406GA SA307060433GA SA307070449GA SA307070451GA SA307080474GA SA307110543GA SA307110562GA SA307110563GA SA307110564GA SA307110565GA SA307110566GA SA307110567GA SA307110576GA SA307110580GA SA307110581GA SA307200778GA SA307200781GA SA307200782GA SA307200783GA SA307200785GA SA307200786GA SA307200787GA SA307200789GA SA307210806GA SA307210812GA SA307210813GA SA307220876GA SA307220894GA SA307220896GA SA307220905GA SA307220907GA SA307220911GA SA307220913GA SA307220915GA SA307220916GA SA307220917GA SA307220920GA SA307220921GA SA307220925GA SA307220928GA SA307220933GA SA307220967GA SA307220968GA SA307220969GA SA307220975GA SA307220985GA SA307220989GA SA307220990GA SA307220992GA SA307221017GA SA307221020GA SA307221021GA SA307221023GA SA307221024GA SA307221031GA SA307221085GA SA307221086GA SA307221087GA SA307221088GA SA307221094GA SA307210810GA SA307382603GA SA307382605GA SA307382619GA SA307443238GA SA307443242GA SA307443247GA SA307453389GA SA307453394GA SA307453396GA SA307453403GA SA307453406GA SA307453408GA SA307453410GA SA307453412GA SA307453553GA SA307453554GA SA307453591GA SA307453607GA SA307382604GA SA307443249GA SA307453605GA SA307220910GA SA307231487GA SA307251792GA SA307453576GA SA307453577GA SA307483665GA SA307483708GA SA307483718GA SA307483779GA SA307483791GA SA307483793GA SA307483807GA SA307483815GA SA307483833GA SA307483839GA SA307493862GA SA307493993GA SA307493998GA SA307504189GA SA307504191GA SA307504202GA SA307504204GA SA307504211GA SA307514323GA SA307514335GA SA307514350GA SA307514353GA SA307514411GA SA307514419GA SA307514420GA SA307514421GA SA307514422GA SA307514425GA SA307514426GA SA307514427GA SA307514428GA SA307514399GA SA307514402GA SA307514433GA SA307514453GA SA307514456GA SA307514400GA SA307514416GA SA307514430GA SA307514438GA SA307514458GA SA308074586GA SA308074602GA SA308084739GA SA308084744GA SA308084747GA SA308084752GA SA308084755GA SA308084775GA SA308084618GA SA308084624GA SA308084625GA SA308084635GA SA308084636GA SA308084640GA SA308084655GA SA308084659GA SA308084660GA SA308044511GA SA308094874GA SA308094933GA SA308094937GA SA308094967GA SA308095024GA SA308095047GA SA308095051GA SA308095056GA SA308095059GA SA308095061GA SA308095064GA SA308095065GA SA308095066GA SA308095067GA SA308095068GA SA308094870GA SA308094998GA SA308094953GA SA308115278GA SA308125324GA SA308125333GA SA308125338GA SA308125339GA SA308125340GA SA308125342GA SA308125343GA SA308125365GA SA308125371GA SA307483820GA SA308094873GA SA308094947GA SA308094980GA SA308094981GA SA308115277GA SA308115281GA SA308125332GA SA308125352GA SA308125366GA SA308145680GA SA308094911GA SA308115293GA SA308135580GA SA308135585GA SA308145711GA SA308145715GA SA308145769GA SA308145799GA SA308145800GA SA308166125GA SA308186349GA SA308186355GA SA308186358GA SA308186361GA SA308186363GA SA308186367GA SA308186384GA SA308186399GA SA308186409GA SA308186412GA SA308186413GA SA308186414GA SA308186415GA SA308186420GA SA308186442GA SA308196465GA SA308196469GA SA308196470GA SA308155881GA SA308155943GA SA308155951GA SA308155958GA SA308155963GA SA308165995GA SA308196541GA SA308166078GA SA308297025GA SA308297027GA SA308297030GA SA308297036GA SA308297039GA SA308297043GA SA308297049GA SA308266827GA SA308266852GA SA308286935GA SA308286944GA SA308286976GA SA308337666GA SA308347733GA SA308347734GA SA308347736GA SA308347745GA SA308347753GA SA308347805GA SA308347809GA SA308347816GA SA308347830GA SA308347836GA SA308357861GA SA308357867GA SA308357872GA SA308357880GA SA308357884GA SA308357893GA SA308357900GA SA308357903GA SA308357907GA SA308357908GA SA308357912GA SA308357915GA SA308357924GA SA308357928GA SA308357929GA SA308357931GA SA308357932GA SA308368195GA SA308368234GA SA308368239GA SA308368240GA SA308368242GA SA308368246GA SA308368247GA SA308368249GA SA308378276GA SA308357863GA SA308357869GA SA308357870GA SA308357873GA SA308357890GA SA308357901GA SA308357902GA SA308357904GA SA308357905GA SA308388569GA SA308408913GA SA308408929GA SA308408964GA SA308409162GA SA308409196GA SA308439443GA SA308439451GA SA308481009GA SA308481010GA SA308481011GA SA308481012GA SA308481050GA SA308489995GA SA308489997GA SA308489998GA SA308408935GA SA308408945GA SA308481034GA SA308481043GA SA308094908GA SA308145809GA SA308409138GA SA308439467GA SA308459716GA SA308459764GA SA308459775GA SA308459778GA SA307372443GA SA308155905GA SA308347801GA SA308449522GA SA308459668GA SA308459683GA SA308459685GA SA308459686GA SA308459687GA SA308459688GA SA308459689GA SA308459842GA SA308459848GA SA308459851GA SA308459852GA SA308459853GA SA308459855GA SA308459867GA SA308459870GA SA308459872GA SA308459873GA SA308459875GA SA308469878GA SA308469880GA SA308469881GA SA308469882GA SA308469883GA SA308481015GA SA308481020GA SA308481026GA SA308481028GA SA308481031GA SA308481032GA SA308489943GA SA308489950GA SA308489963GA SA308481029GA SA308501410GA SA308501417GA SA308501426GA SA308347792GA SA308459858GA SA308479897GA SA308479902GA SA308479905GA SA308479907GA SA308479909GA SA308489910GA SA308489913GA SA308489914GA SA308489915GA SA308489916GA SA308489923GA SA308489937GA SA308489993GA SA309062275GA SA309032011GA SA309092472GA SA309062295GA SA309062297GA SA309062319GA SA309062322GA SA309062324GA SA309062327GA SA309062332GA SA309062333GA SA309062334GA SA309062335GA SA309062337GA SA309062338GA SA309062339GA SA309062341GA SA309072342GA SA309072346GA SA309072349GA SA309072359GA SA309052221GA SA308459667GA SA309011874GA SA309021957GA SA309031994GA SA309042191GA SA309042192GA SA309042193GA SA309042197GA SA309042198GA SA309042200GA SA309042202GA SA309042203GA SA309042205GA SA309042206GA SA309042210GA SA309042211GA SA309011847GA SA309011893GA SA309042130GA SA309042134GA SA309042135GA SA309042136GA SA309042137GA SA309042138GA SA309042139GA SA309042140GA SA309042141GA SA309042142GA SA309042143GA SA309042146GA SA309042147GA SA309042148GA SA309042181GA SA309042182GA SA309052217GA SA309052218GA SA309052226GA SA309052237GA SA309052238GA SA309052240GA SA309052241GA SA309052242GA SA309052243GA SA309052244GA SA309052245GA SA309052246GA SA309052247GA SA309052248GA SA309052249GA SA309052250GA SA309052251GA SA309052252GA SA309052253GA SA309052254GA SA309052255GA SA309052256GA SA309052257GA SA309052258GA SA309052259GA SA309052260GA SA309052261GA SA309052262GA SA309052263GA SA309052264GA SA309052265GA SA309082418GA SA309082419GA SA309082420GA SA309082421GA SA309082422GA SA309082424GA SA309082425GA SA309082426GA SA309082427GA SA309082428GA SA309082430GA SA309082431GA SA309082432GA SA309082433GA SA309082434GA SA309082435GA SA309082436GA SA309082437GA SA309082438GA SA309082439GA SA309082440GA SA309082441GA SA309082442GA SA309092495GA SA309092497GA SA309092498GA SA309092499GA SA309092501GA SA309072378GA SA309072399GA SA309072410GA SA309082413GA SA309082417GA SA309082453GA SA309082459GA SA309082468GA SA309092469GA SA309092470GA SA309092471GA SA309062317GA SA309062321GA SA309062323GA SA309062329GA SA309062340GA SA309072360GA SA309122675GA SA309122679GA SA309122698GA SA309183237GA SA309193284GA SA308084637GA SA309173161GA SA309203383GA SA309203387GA SA309203395GA SA309203401GA SA309294124GA SA309324328GA SA309334791GA SA309446481GA SA309446500GA SA309446510GA SA309456634GA SA309456655GA SA309456657GA SA309466753GA SA309497229GA SA309497231GA SA309497268GA SA309497269GA SA309497271GA SA309497272GA SA309497273GA SA309497283GA SA309497284GA SA309497286GA SA309497288GA SA309476969GA SA309476982GA SA309497233GA SA309497234GA SA309497285GA SA309497301GA SA309497223GA SA309497224GA SA309497225GA SA309497227GA SA309497228GA SA309497270GA SA309497275GA SA309497276GA SA309497277GA SA309497278GA SA309476865GA SA309476917GA SA309476999GA SA309477021GA SA309477024GA SA309477108GA SA309497155GA SA309497279GA SA309507354GA SA309507360GA SA309507380GA SA309507381GA SA309507388GA SA309507390GA SA309507393GA SA309507395GA SA309507397GA SA309507402GA SA309507422GA SA309507406GA SA309507418GA SA309507428GA SA309517479GA SA309507373GA SA309507410GA SA309507423GA SA309517812GA SA309517876GA SA309517879GA SA309517888GA SA309527928GA SA309527951GA SA309527959GA SA309527968GA SA309062320GA SA309517802GA SA309517850GA SA309527908GA SA309538030GA SA310028174GA SA310028185GA SA310028206GA SA310028258GA SA310028289GA SA310028292GA SA310028327GA SA310028332GA SA310028340GA SA310038357GA SA310038358GA SA310038373GA SA310038375GA SA310038385GA SA310038396GA SA310038399GA SA310038402GA SA310038403GA SA310038412GA SA310038418GA SA310038422GA SA310038428GA SA310038437GA SA310038442GA SA310038452GA SA310038456GA SA310038458GA SA310038461GA SA310038463GA SA310038469GA SA310038478GA SA310038479GA SA310038486GA SA310038494GA SA310038496GA SA310038501GA SA310038504GA SA310038517GA SA310038519GA SA310038521GA SA310038526GA SA310038533GA SA310038535GA SA310038538GA SA310048544GA SA310048548GA SA310048550GA SA310048554GA SA310048555GA SA310048556GA SA310048566GA SA310048568GA SA310048571GA SA310048576GA SA310048583GA SA310048599GA SA310048600GA SA310048601GA SA310048602GA SA310048603GA SA310048605GA SA310048608GA SA310048609GA SA310048610GA SA310048611GA SA310048612GA SA310048615GA SA310048617GA SA310048625GA SA310048627GA SA310048629GA SA310048648GA SA310048651GA SA310048656GA SA310048675GA SA310058693GA SA310068834GA SA310068842GA SA309517872GA SA309538091GA SA309538107GA SA310028225GA SA310038371GA SA310038387GA SA310038398GA SA310038407GA SA310038411GA SA310038451GA SA310048591GA SA310048592GA SA310048606GA SA310048624GA SA309507379GA SA309507431GA SA309507438GA SA309507446GA SA309517453GA SA309517459GA SA309517475GA SA310068877GA SA310139067GA SA310139246GA SA310139261GA SA310149314GA SA310149327GA SA310159362GA SA310159363GA SA310159369GA SA310159371GA SA310139181GA SA310149353GA SA310149358GA SA310219824GA SA310259943GA SA310259946GA SA310259948GA SA310259959GA SA310260290GA SA310260292GA SA310260293GA SA310260294GA SA310260295GA SA310260296GA SA310260298GA SA310260299GA SA310260300GA SA310260301GA SA310260302GA SA310260304GA SA310260306GA SA310260316GA SA310260318GA SA310260319GA SA310260321GA SA309528001GA SA310058746GA SA310068813GA SA310119015GA SA310290624GA SA310290626GA SA310290627GA SA310290630GA SA310290635GA SA310290643GA SA310290665GA SA310290702GA SA310290703GA SA310382623GA SA310382641GA SA310382705GA SA310382731GA SA310382735GA SA310382738GA SA310382753GA SA310382758GA SA310382767GA SA310382770GA SA310393141GA SA310413517GA SA310423646GA SA310423654GA SA310424026GA SA310475539GA SA310475573GA SA310475583GA SA310475615GA SA310495892GA SA310495903GA SA310495909GA SA310495912GA SA310495933GA SA310495934GA SA310495941GA SA310495946GA SA310495950GA SA310495954GA SA310495957GA SA310495973GA SA310495896GA SA310495897GA SA310495901GA SA310495904GA SA310495906GA SA310495932GA SA310495943GA SA310341375GA SA310495908GA SA310496381GA SA310496382GA SA310496391GA SA310496392GA SA310496397GA SA310496406GA SA310496408GA SA310496414GA SA310496419GA SA310496421GA SA310496422GA SA310496423GA SA310496426GA SA310496432GA SA310496448GA SA310496452GA SA310495911GA SA310495956GA SA310495959GA SA310496131GA SA310496145GA SA310496152GA SA310496209GA SA310496216GA SA310496280GA SA310496286GA SA310496301GA SA310496316GA SA310496356GA SA310496477GA SA310496530GA SA310495891GA SA310495894GA SA310496128GA SA310496380GA SA310496390GA SA310496398GA SA310496412GA SA310496416GA SA310496417GA SA310496427GA SA310496431GA SA310496433GA SA310506670GA SA310506689GA SA310506697GA SA310506722GA SA310506873GA SA310507118GA SA310506738GA SA310496297GA SA310496305GA SA310496307GA SA310496308GA SA310496311GA SA310496313GA SA310496343GA SA310496506GA SA310517604GA SA310517618GA SA310517621GA SA310517625GA SA310517626GA SA310517629GA SA310517632GA SA310517637GA SA310496446GA SA310506766GA SA310506800GA SA310495893GA SA310495899GA SA310495900GA SA310495918GA SA310495942GA SA310507313GA SA310507324GA SA310507325GA SA310507353GA SA310517681GA SA310518146GA SA310518177GA SA310518182GA SA310518184GA SA310518186GA SA310518191GA SA310518192GA SA310518193GA SA310518195GA SA310518197GA SA310423805GA SA310517673GA SA310517676GA SA310517682GA SA310518306GA SA310518311GA SA310517844GA SA310517870GA SA310517871GA SA310518000GA SA310507376GA SA310518629GA SA310507524GA SA310507538GA SA310507549GA SA310528867GA SA310528872GA SA310528884GA SA311039378GA SA311039380GA SA311039384GA SA310496134GA SA310496135GA SA310496144GA SA310496146GA SA310496210GA SA310496277GA SA310496293GA SA310496351GA SA310496503GA SA310496510GA SA310496632GA SA310496646GA SA310506750GA SA310506753GA SA310506991GA SA310507003GA SA310507011GA SA310434052GA SA310528791GA SA311039533GA SA311039544GA SA311060305GA SA311060310GA SA310496399GA SA310518486GA SA310048587GA SA310048618GA SA310048628GA SA310382625GA SA310382629GA SA310382730GA SA310382760GA SA311080713GA SA311080715GA SA310454933GA SA311039532GA SA311060101GA SA311060105GA SA311060114GA SA311070557GA SA311070571GA SA311070573GA SA311070583GA SA311070584GA SA311070590GA SA311080647GA SA311080737GA SA311090843GA SA311090846GA SA311090855GA SA311060168GA SA311060213GA SA311060224GA SA311060229GA SA311060233GA SA311060252GA SA311060260GA SA311069724GA SA311069738GA SA311060326GA SA311060338GA SA311060345GA SA311060365GA SA311060369GA SA311060383GA SA311060398GA SA311060399GA SA311070553GA SA311080616GA SA311060331GA SA311060332GA SA311060335GA SA311060356GA SA311060371GA SA311060372GA SA311060375GA SA311060379GA SA311060380GA SA311060281GA SA311070556GA SA311080598GA SA311080651GA SA311080660GA SA311069790GA SA311069797GA SA311069805GA SA311069806GA SA311069808GA SA311069810GA SA311069829GA SA311069831GA SA311069832GA SA311069848GA SA311080727GA SA311080736GA SA311090799GA SA311090805GA SA311090831GA SA311090845GA SA311070525GA SA311101390GA SA311101395GA SA311101404GA SA311101422GA SA311101553GA SA311111565GA SA311111579GA SA311122036GA SA310507102GA SA311060334GA SA311080607GA SA311080609GA SA311080629GA SA311080638GA SA311080648GA SA311080653GA SA311080745GA SA311090751GA SA311090754GA SA311090760GA SA311090770GA SA311090789GA SA311090802GA SA311090803GA SA311090869GA SA311090872GA SA311090876GA SA311090886GA SA311101097GA SA311101176GA SA311101183GA SA311101185GA SA311101186GA SA311101188GA SA311101194GA SA311101196GA SA311101205GA SA311101207GA SA311101209GA SA311101211GA SA311101215GA SA311101223GA SA311101255GA SA311101258GA SA311101259GA SA311101271GA SA311101287GA SA311101288GA SA311101296GA SA311101305GA SA311101397GA SA311121869GA SA311121872GA SA311121895GA SA311122219GA SA311122222GA SA311142916GA SA311142932GA SA311153098GA SA311153102GA SA311153105GA SA311153125GA SA311153145GA SA311101270GA SA311101385GA SA311122043GA SA311122045GA SA311122059GA SA311122180GA SA311122195GA SA311132366GA SA311142930GA SA311153083GA SA311153106GA SA311153113GA SA311153116GA SA311153126GA SA311153127GA SA311153134GA SA311153143GA SA311153148GA SA311153158GA SA311153205GA SA311153211GA SA311153212GA SA311121699GA SA311121705GA SA311121709GA SA311121711GA SA311121720GA SA311121726GA SA311132566GA SA311142915GA SA311142919GA SA311142933GA SA311142955GA SA311142980GA SA311153135GA SA311121861GA SA311121889GA SA311121911GA SA311142908GA SA311153097GA SA311153099GA SA311153175GA SA311153227GA SA311163975GA SA307030403GA SA307080541GA SA307110547GA SA307110550GA SA307110558GA SA309507365GA SA311225998GA SA311246410GA SA311225996GA SA311236108GA SA311236114GA SA311236117GA SA311236120GA SA311205442GA SA311205444GA SA311205462GA SA311205340GA SA311205344GA SA311205352GA SA311205374GA SA311205377GA SA311205382GA SA311205383GA SA311205389GA SA311256781GA SA311256793GA SA311266873GA SA311266910GA SA311266993GA SA311266995GA SA311267005GA SA311256765GA SA311256770GA SA311287105GA SA311287107GA SA311287121GA SA311287122GA SA311287125GA SA311287127GA SA311287129GA SA311287133GA SA311287137GA SA311287138GA SA311287146GA SA311287147GA SA311287149GA SA311287172GA SA311287181GA SA311297261GA SA311297268GA SA311297269GA SA311297282GA SA311297283GA SA311297284GA SA311297290GA SA311174185GA SA311184604GA SA311184650GA SA311195095GA SA311287110GA SA311327833GA SA311195265GA SA311205346GA SA311205349GA SA311205351GA SA311205354GA SA311205361GA SA311205365GA SA311205368GA SA311205375GA SA311205380GA SA311205384GA SA311205385GA SA311205387GA SA311205388GA SA311307447GA SA311317649GA SA310068924GA SA311184386GA SA311184530GA SA311184553GA SA311184576GA SA311184641GA SA311184646GA SA311184648GA SA311184649GA SA311184651GA SA311184660GA SA311184661GA SA311195146GA SA311195153GA SA311195154GA SA311195158GA SA311195167GA SA311195252GA SA311195257GA SA311205366GA SA311266897GA SA311266900GA SA311307505GA SA311317725GA SA311195138GA SA311195152GA SA311195155GA SA311195159GA SA311195249GA SA311195251GA SA310518150GA SA310518202GA SA310518216GA SA310518219GA SA310518222GA SA310518233GA SA310518236GA SA310518240GA SA311205342GA SA311205347GA SA311256713GA SA311266876GA SA311266892GA SA311266996GA SA311266998GA SA311266999GA SA311267001GA SA311267002GA SA311205401GA SA311215804GA SA311215846GA SA311225898GA SA311225901GA SA311225910GA SA311225913GA SA311225914GA SA311225915GA SA311225916GA SA311225917GA SA311225918GA SA311225920GA SA311225922GA SA311225925GA SA311225926GA SA311225931GA SA311225932GA SA311225934GA SA311225935GA SA311225936GA SA311225937GA SA311225939GA SA311225940GA SA311225941GA SA311225943GA SA311225946GA SA311225949GA SA311225950GA SA311225956GA SA311225958GA SA311225959GA SA311225961GA SA311225962GA SA311225963GA SA311225964GA SA311225965GA SA311225967GA SA311225968GA SA311225969GA SA311225971GA SA311225972GA SA311225973GA SA311225975GA SA311236005GA SA311236009GA SA311236010GA SA311236012GA SA311236017GA SA311236030GA SA311236037GA SA311236049GA SA311236055GA SA311236058GA SA311236062GA SA311236190GA SA311236191GA SA311236237GA SA311236243GA SA311236250GA SA311236252GA SA311236254GA SA311236267GA SA311236268GA SA311236283GA SA311236284GA SA311236350GA SA311236358GA SA311236362GA SA311236365GA SA311236374GA SA311236379GA SA311246453GA SA311246457GA SA311246479GA SA311246510GA SA311246512GA SA311256871GA SA311266915GA SA311266919GA SA311266946GA SA311266951GA SA310518220GA SA310518229GA SA310518234GA SA311246544GA SA311256689GA SA311256707GA SA311256711GA SA311508641GA SA311508642GA SA311508644GA SA311508649GA SA311508651GA SA311508666GA SA311508671GA SA310159456GA SA310169511GA SA310259949GA SA310260315GA SA310260388GA SA310290654GA SA310331123GA SA310331328GA SA310331333GA SA310331336GA SA310331338GA SA310331339GA SA310331344GA SA310331346GA SA310331353GA SA310341359GA SA310341361GA SA310341384GA SA310372167GA SA310372168GA SA310372180GA SA310372194GA SA310372203GA SA310382325GA SA310382326GA SA310382331GA SA310382336GA SA310382347GA SA310371994GA SA310372068GA SA310372138GA SA310372140GA SA310372161GA SA310372288GA SA310372308GA SA310372309GA SA312253308GA SA312253311GA SA312253313GA SA312253367GA SA312253368GA SA312263939GA SA312284151GA SA312284169GA SA312284172GA SA312284173GA SA312284176GA SA312284184GA SA312284282GA SA312284296GA SA312284306GA SA312294498GA SA312294628GA SA312355475GA SA312355483GA SA312355505GA SA312355511GA SA312355523GAMKE SA312355525GA SA312365569GA SA312365570GA SA312365591GA SA312365592GA SA312365593GA SA312365594GA SA312365596GA SA312365597GA SA312365599GA SA312365600GA SA312365601GA SA312365603GA SA312365604GA SA312365606GA SA312365607GA SA312365608GA SA312365609GA SA312365610GA SA312365611GA SA312365612GA SA312365614GA SA312365615GA SA312365616GA SA312365619GA SA312365620GA SA312365627GA SA312365638GA SA312365641GA SA312365691GA SA312344921GA SA312355236GA SA312375992GA SA312386374GA SA312386391GA SA312386402GA SA312386404GA SA312386409GA SA312386422GA SA312386424GA SA312386427GA SA312386430GA SA312386462GA SA312386467GA SA312457686GA SA312457723GA SA312457727GA SA312457728GA SA312457736GA SA312457795GA SA312457811GA SA312457786GA SA312457800GA SA312457807GA SA312447554GA SA312469399GA SA312469406GA SA312469418GA SA312469419GA SA312469422GA SA312469429GA SA312469433GA SA312469435GA SA312469448GA SA312469450GA SA312469455GA SA312469467GA SA312469502GA SA312469505GA SA312469510GA SA312469513GA SA312469515GA SA312469520GA SA312479523GA SA312479525GA SA312479527GA SA312479543GA SA312479545GA SA312479548GA SA312479550GA SA312479551GA SA312479552GA SA312479554GA SA312479557GA SA312479558GA SA312479566GA SA312479568GA SA312479569GA SA312479570GA SA312479575GA SA312479584GA SA312479586GA SA312479588GA SA312479589GA SA312479590GA SA312479592GA SA312479599GA SA312479658GA SA312479671GA SA312480331GA SA312480338GA SA312480373GA SA312480383GA SA312480388GA SA312480413GA SA312480415GA SA312480420GA SA312447453GA SA312447466GA SA312447470GA SA312447472GA SA312447473GA SA312447477GA SA312447479GA SA312447480GA SA312447516GA SA312447542GA SA312447548GA SA312447568GA SA312447581GA SA312480377GA SA312480572GA SA312480581GA SA312480583GA SA312480761GA SA311508645GA SA311508658GA SA311508660GA SA311508661GA SA311508667GA SA311508670GA SA311508676GA SA311508677GA SA311508678GA SA311508679GA SA311508680GA SA311518682GA SA311518684GA SA311518685GA SA311518688GA SA311518692GA SA311518694GA SA312018745GA SA312018747GA SA312018752GA SA312018753GA SA312018756GA SA312018757GA SA312018763GA SA312018764GA SA312018765GA SA312018767GA SA312018773GA SA312018774GA SA312018775GA SA312018776GA SA312018838GA SA312018842GA SA312018849GA SA312018882GA SA312018891GA SA312018892GA SA312018900GA SA312018903GA SA312018907GA SA311287164GA SA311327848GA SA311327852GA SA311428018GA SA311428043GA SA312018816GA SA312018825GA SA312018833GA SA311508669GA SA312028999GA
SA312029008GA SA312029018GA SA312029019GA SA312029051GA SA312029071GA SA312039214GA SA312039216GA SA312039217GA SA312039218GA SA312039219GA SA312039222GA SA312039226GA SA312039227GA SA312039229GA SA312080391GA SA312080400GA SA312080408GA SA312080411GA SA312080421GA SA312080435GA SA312080492GA SA312080495GA SA312080499GA SA312080501GA SA312080502GA SA312080503GA SA312080504GA SA312080507GA SA312080511GA SA312080589GA SA312080680GA SA312080709GA SA312080712GA SA312080718GA SA312080726GA SA312080821GA SA312080846GA SA312080855GA SA312080583GA SA312080603GA SA312069701GA SA312080393GA SA312080618GA SA312080628GA SA312080636GA SA312080641GA SA312080394GA SA312080627GA SA312069718GA SA312069691GA SA312070045GA SA312070047GA SA312070051GA SA312070066GA SA312070072GA SA312070098GA SA312070103GA SA312070104GA SA312070108GA SA312070117GA SA312070119GA SA312070121GA SA312070122GA SA312070125GA SA312070129GA SA312070140GA SA312080818GA SA312069777GA SA312070341GA SA312070345GA SA312070348GA SA312070360GA SA312111511GA SA312111532GA SA312091267GA SA312091273GA SA312091292GA SA312111653GA SA312121756GA SA312121765GA SA312121801GA SA312121807GA SA312121832GA SA312121846GA SA312121858GA SA312121896GA SA312121996GA SA312122010GA SA312122012GA SA312122020GA SA312122022GA SA312122027GA SA312132117GA SA312132133GA SA312121868GA SA312152345GA SA312152485GA SA311069661GAMKE SA311069800GAMKE SA311069802GAMKE SA311101084GAMKE SA311101127GAMKE SA311101129GAMKE SA311101161GAMKE SA311101440GAMKE SA311101445GAMKE SA311101449GAMKE SA311101455GAMKE SA311122062GAMKE SA311122064GAMKE SA311122099GAMKE SA311122126GAMKE SA311122140GAMKE SA311122145GAMKE SA311122148GAMKE SA311122154GAMKE SA311122160GAMKE |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
800-558-7044
|
Manufacturer Reason for Recall |
GE Healthcare has recently become aware of a potential safety issue due to intermittent failure of the spring loaded
mounting latch of PDMs produced from April 2007 through December 2012.
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 26, 2013 to its consignees/customers. The letter was addressed to Healthcare Administrator / Risk Manger, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. The consignees/customers were instructed to please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions. GE Healthcare will provide a new latch at no charge once it is available. GE will contact you to arrange for installation of the latch.
If you have any questions regarding this notification, please contact Technical Support at 1-800-558-7044, or your local Service Representative. |
Quantity in Commerce |
43,033 (25,681 USA; 17,352 OUS) |
Distribution |
Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, ID, IL, IN, IA, KS,KY, LA, ME, MD, MA, MI, MN, MS, NO, NE, NV, NJ, NM, NY, NC, OH, OK,OR, PA, RI, SC, TN, TX, UT,VT, VA, WA, WI, and WY. Internationally to: Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Egypt,England, Estonia, Finland,France, Germany,Greece, Guatemala, Hungary,India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Kuwait, Lebanon, Lithuania, Malaysia, Mauritius, Mexico, Morocco, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates and United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
|
|
|
|