|
Class 2 Device Recall 5X250 MM Curved Cannula, Arm 1 |
|
Date Initiated by Firm |
March 13, 2014 |
Date Posted |
April 04, 2014 |
Recall Status1 |
Terminated 3 on September 03, 2014 |
Recall Number |
Z-1334-2014 |
Recall Event ID |
67758 |
510(K)Number |
K112208
|
Product Classification |
System,surgical,computer controlled instrument - Product Code NAY
|
Product |
5X250 MM Curved Cannula, Arm 1 for use with daVinci SI Surgical System, model number IS3000.
Intuitive Surgical 1266 Kifer Road Sunnyvale, CA 94086
The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation,clip- ligation, electrocautery and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators,monopolar cautery, bipolar cautery, 5mm curved cannulae,5mm and 10 mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port. |
Code Information |
Part number 428071-03: Lot numbers: VE113001 VE121004 VE121205 VE122004 VE122604 VE123014 VE113401 VE121007 VE121407 VE122005 VE122901 VE123107 VE114001 VE121103 VE121410 VE122104 VE122904 VE123202 VE114401 VE121105 VE121604 VE122204 VE122908 VE120602 VE121109 VE121608 VE122302 VE122911 VE120701 VE121110 VE121804 VE122403 VE123007. |
Recalling Firm/ Manufacturer |
Intuitive Surgical, Inc. 1266 Kifer Rd Bldg 100 Sunnyvale CA 94086-5304
|
For Additional Information Contact |
Mark Johnson 408-523-2100
|
Manufacturer Reason for Recall |
Intuitive Surgical has become aware of the potential for certain lots of the Single-Site 5mm Curved Cannulae to become damaged during use. The second part of this action is to update Single-Site labeling. These labeling changes instruct users how to inspect cannulae prior to use and to appropriately handle cannulae during use.
|
FDA Determined Cause 2 |
Device Design |
Action |
Intuitive sent an Urgent Medical Device letter dated March 25, 2014, to all users of cannulas that have the potential to be damaged during use as well as a second letter to all single site cannula users that provided clarity to instruction for use for inspections and handling of the product.
Certain lots manufactured prior to August of 2012, and users were asked to use Attachment A to help identify affected lots numbers. Phase 1 - Cannulas should then be inspected with information provided in Attachment B. Any damaged devices should be quarantined and returned to intuitive. If inspection does not find damage, the device may continue to be used. Any other cannulae and any non damaged product should be used according to Attachment C's instructions for use.
Phase 2 - Once replacement cannulae are available, the undamaged cannulae should be returned.
The Acknowledgement form should be completed and returned to Intuitive. Intuitive will contact consignees when replacement cannulae are available.
Questions should be directed to Customer Service at 800-876-1310.
Those consignees that use the 5 mm canulae but do not have affected product were provided with a letter that only includes the Single Site Cannula inspection and Use addendum. |
Quantity in Commerce |
5217 total cannulae |
Distribution |
Worldwide Distribution - USA and the countries of Australia, Belgium, Brazil, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Lebanon, Monaco, Netherlands, Norway, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = NAY and Original Applicant = INTUITIVE SURGICAL, INC.
|
|
|
|