Class 2 Device Recall Nuclear Gamma Cameras |
|
Date Initiated by Firm |
March 10, 2014 |
Date Posted |
April 26, 2014 |
Recall Status1 |
Terminated 3 on August 04, 2015 |
Recall Number |
Z-1510-2014 |
Recall Event ID |
67901 |
510(K)Number |
K922080 K952684
|
Product Classification |
System, tomography, computed, emission - Product Code KPS
|
Product |
Nuclear gamma cameras (ADAC Vertex Plus, ADAC Solus, ADAC Vertex V60 & ADAC Vertex Classic) Philips Medical System, Cleveland, OH 44143.
Used to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel. |
Code Information |
ADAC Vertex Plus & ADAC Solus System Code: 88297, Serial #'s: 11, 1006, 1008, 1027, 1092, 1097, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 6052, 6261, 8002, 10252, 10630, 11090, 85247, 513519, 513520, 940384, 9205345, 9408136, 9503084, 9506016, 9511123, 9512009, 9512018, 9601012, 9601015, 9601016, 9601024, 9601036, 9602161, 9602254, 9603336, 9604016, 9604021, 9605024, 9605026, 9605037, 9606211, 9607015, 9607024, 9608031, 9608046, 9608049, 9608052, 9609063, 9609081, 9610058, 9611035, 9611077, 9611082, 9612072, 9702095, 9702150, 9703154, 9703165, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708076, 9708078, 9708084, 9708088, 9708227, 9709022, 9709061, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711271, 9711291, 9801047, 9801227, 9801232, 9801297, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803303, 9803312, 9803316, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9805376, 9806014, 9806363, 9806377, 9806388, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901162, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905003, 9905012, 9905659, 9905667, 9906670, 9907016, 9907019, 9909011, 9909161, 9910015, 9910017, 9910031, 9911458, 9911698, 9911702, 9911711, 9912181, 9912463, 98066002, 802080322, 0002032, 0007001M, 0008009B, 0011002B, 0101002C, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 130234-01NVERTEX, 5934V, 6090VM, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 9054VM/V, 9056VM/V, 9130VM, 9806373A, 9810428ADAC#7973VM, 9894K, AD1261, AMT00848VM, B0006129, B0007273, B0008001, B0008003, B0008012, B0008021, B0008030, B001200212, B0102114, B0104125, B0104128, B0105138, B0106169, B01070191, B01070192, B0107179, B01080199, B01080204, B01080207, B01090227, B01090228, B01100231, B01100232, B01100233, B01100236, B01100237, B01100238, B01100241, B01100242, B01110251, B01110252, B02050071, B02050262, B02050263, B02050265, B02060167, B02060270, B02060279, B0206027S, B02070289, B02070290, B02070291, B02070297, B02070300, B0207296, B02080318, B0208312, B02090327, B02090334, B02100345, B02110348, B02120352, B03010355, B03010359, B03020048, B03080380, C0006050, C0006051, C0007060, C0008070, C0009075, C0101099, C0102103, C0102108, C0104118, C0107145, C01080152, C01080153, C01080154, C02100218, C02110220, C02120223, C03010227, C03050230, C03060193, CNF, M@10458, M@10482, M@10547, M@105561, M@10573, M@10582, M@10589, M@10591, M@10676, M@10772, M@10837, M@10839, M@10848, M@10850, M@10871, M@10878, M@10882, M@10927, M@10937, M@10967, M@10979, M@11015, M@11020, M@11036, M@11103, M@11109, M@11140, M@11178, M@11184, M@11381, M@11398, M@11423, M@11498, M@11525, M@11571, M@11590, M@11609, M@11644, M@11881, M@11939, M@11949, M@11957, M@11972, M@12060, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13663, M@13665, M@13688, M@13691, M@13748, M@13813, M@14062, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@6002J, M@6012J, M@6075VM, M@6076K, M@6077K, M@6108VA, M@6123VA, M@6143VA, M@6160VM, M@6183VA, M@6244VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6344VM, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6366K, M@6368J, M@6370VA, M@6371VM, M@6375J, M@6376J, M@6460VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6769K, M@6771VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6912K, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6946VM, M@6947J, M@6985J, M@7003VA, M@7004VA, M@7005VA, M@7017VM, M@7024VM, M@7034K, M@7042VM, M@7053VM, M@7070VM, M@7073VM, M@7085K, M@7086J, M@7095VM, M@7096VM, M@7099K, M@7100VM, M@7101VM, M@7102VM, M@7108VM, M@7109VM, M@7151VM, M@7166K, M@7171VM, M@7172K, M@7181J, M@7190VM, M@7198VA, M@7205K, M@7215K, M@7216VM, M@7217VM, M@7221VA, M@7222VA, M@7229VM, M@7233VM, M@7237VM, M@7245J, M@7246VM, M@7247VM, M@7250VM, M@7253K, M@7300J, M@7308VM, M@7358VA, M@7376VM, M@7380VM, M@7384VM, M@7390VM, M@7393VM, M@7394VM, M@7410VM, M@7420VM, M@7463VM, M@7464VA, M@7470VM, M@7490VM, M@7494K, M@7501VM, M@7503VM-FU1, M@7504VM, M@7505VM1, M@7514VM, M@7515VM, M@7536VM, M@7537VM, M@7538VM, M@7556VM, M@7557VM, M@7569VM, M@7572VM, M@7591VM, M@7592VM, M@7603K, M@7605VM, M@7606VM, M@7623K, M@7624K, M@7626K, M@7627K, M@7628VA, M@7638VM, M@7643VM, M@7654VM, M@7664VM, M@7665K, M@7673VM, M@7674K, M@7679VM, M@7682K, M@7685VM, M@7691VM, M@7695VM, M@7697VM, M@7723VM, M@7729VM, M@7733VM, M@7739K, M@7740VM, M@7760VM, M@7761VM, M@7770VM, M@7776VM, M@7789K, M@7801VM, M@7811VM, M@7812VM, M@7819VA, M@7823K, M@7824VM, M@7830K, M@7855J, M@7857VM, M@7860VM, M@7861VM, M@7874VM, M@7877VM, M@7918K, M@7924K, M@7925K, M@7929VM, M@7930VM, M@7932VM, M@7946J, M@7957VM, M@7961VM, M@7962VM, M@7963K, M@7965VM, M@7968VM, M@7975K, M@7987VM, M@7995VM, M@7997VM, M@8004VM, M@8018J, M@8021VM, M@8044VM, M@8064VM, M@8076K, M@8089VM, M@8096VM, M@8105VM, M@8124VA, M@8146VM, M@8155VM, M@8178VM, M@8203VM, M@8219VA, M@8240K, M@8241VM, M@8247K, M@8248VM, M@8252VM, M@8284K, M@8301K, M@8302VM, M@8306K, M@8337VM, M@8356K, M@8380VM, M@8394K, M@8399VM, M@8436K, M@8462VM, M@8484VM, M@8491K, M@8510VM, M@8528VM, M@8537VM, M@8540K, M@8544K, M@8551VM, M@8561K, M@8606K, M@8609VM, M@8651VM, M@8683VM, M@8686K, M@8692VM, M@8693VM, M@8726VM, M@8738K, M@8759K, M@8771VM, M@8781VM, M@8797K, M@8799RV, M@8800VM, M@8803K, M@8812VM, M@8814VM, M@8817RK, M@8876K, M@8882K, M@8884K, M@8888K, M@8897K, M@8903K, M@8906RK, M@8920VM, M@8922VM, M@8935K, M@8936K, M@8942K, M@8943K, M@8948RV, M@8955K, M@8966K, M@8967K, M@8990K, M@8996VM, M@900K, M@9032K, M@9041K, M@9046K, M@9049VM, M@9063K, M@9080K, M@9081K, M@9085K, M@9091VM, M@9107RV, M@9118K, M@9129K, M@9131K, M@9140K, M@9166VM, M@9185K, M@9186K, M@9191K, M@9194VM, M@9221VM, M@9222VM, M@9231K, M@9247RV, M@9252VM, M@9267VM, M@9283K, M@9293VM, M@9308K, M@9309RV, M@9316VM, M@9327K, M@9337K, M@9338VM, M@9340VM, M@9344K, M@9349K, M@9360VM, M@9418VM, M@9426K, M@9434K, M@9442K, M@9452K, M@9471RV, M@9476K, M@9479K, M@9479K, M@9480K, M@9502K, M@9515VM, M@9522K, M@9525K, M@9532VM, M@9542VM, M@9543K, M@9562K, M@9563K, M@9566K, M@9583VM, M@9590K, M@9594K, M@9598K, M@9607VM, M@9609K, M@9622K, M@9639K, M@9646K, M@9653VM, M@9672VM, M@9712K, M@9724VM, M@9730K, M@9765K, M@9769K, M@9799K, M@9816K, M@9831VM, M@9834K, M@9837K, M@AMT1258, M@MCD6416VM, M@MCD7036VM, M@O1317, M@R372820, M@S5195VA, M@S5365-V, M@S5820VM, M@S5907V11, M@S6155VA, M@S7115VM, M@S7156VM, M@S7223K, M@S7559VA, M@S7571VM, M@S7817K, M@S7856K, M@S7864K, M@S7904VM, M@S7941VA, M@S8687K, N@9369VM, R03010363, S0011003, S6235VM, UNK-6597VM, UNK-7590VM, UNK-8097J1, UNK-8533VM1, UNK-8860VM, UNK-9868K, V0007070, V0012099, V0104109, V0104110, V0105112, V0106121, V0106123, V0106124, V01070133, V01090146, V01100151, V01100156, V02050166, V02070174, V02070179, V02080190, V02080194, V02080196, V02080200, V03060224, & V03070229. ADAC Vertex V60 System Code: 882030; Serial #'s: 10588, 9601004, V0005062, V03080235, V03080236, & V2070177. ADAC Vertex Classic System Code: 890122; Serial #'s: 64, 103, 500, 3334, 5005, 9308007, 9309020, 9310032, 9401059, 9401061, 9403076, 9404106, 9407122, 9410002, 9411006, 9411016, 9501210, 9502067, 9502070, 9503003, 9503079, 9503109, 9504019, 9504099, 9505021, 9505124, 9506140, 9506178, 9508171, 9508184, 9508193, 9509206, 9510222, 9510226, 9601146, 9602249, 9608042, 9609084, 9901004, 9901135, 0407V, 0682V, 0795V, 0822VP2, 20, 5405V, 93-10660-1292, 9408128, 9411007, 9502086, 9508169, 9509198, 9509199, B0106154, B02120033, M@10229AMT, M@207JD, M@4324V, M@4404V, M@4431V, M@4432V, M@4480V, M@4623V, M@4630V, M@4639V, M@4686V, M@4692V, M@4835V, M@4858V, M@4893V, M@4959V, M@5032V, M@5033V, M@5038V, M@5105V, M@5145V, M@5158V, M@5189V, M@5280V, M@5289V, M@5338V, M@5360V, M@5398V, M@5407V, M@5412V, M@5431V, M@5439V, M@5467V, M@5474V, M@5476V, M@5497V, M@5498V, M@5515V, M@5526V, M@5590V, M@5617V, M@5627V, M@5661V, M@5662V, M@5663V, M@5664V, M@5665V, M@5681V, M@5688V, M@5689V, M@5751V, M@5752V, M@5782V, M@5811V, M@5835V, M@5869V, M@5900V, M@5913V, M@5915V, M@5960V, M@5970V, M@6004V, M@6007V, M@6010V, M@6056V, M@6900V, M@9304RV, M@9374RV, M@9645RV, M@MVS5236V, M@MVS5852V, M@O1325 & UNK-AMT1238A. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland OH 44143-2131
|
For Additional Information Contact |
Mrs. Kumudini J. Carter 440-483-3032
|
Manufacturer Reason for Recall |
Slippage of the radius drive belt in the relative 180 degree orientation of the superior positioned detector head will allow the detector head to drift down and potentially come in direct contact with the patient.
|
FDA Determined Cause 2 |
Device Design |
Action |
On 03/10/2014, the firm sent Urgent - Medical Device Correction letters to their customers. The letter identified the affected product, the reason for recall, the hazard involved, and the actions that should be taken by customers. Customers are to cease use of the identified procedures until implementation of the Field Change Order. Additionally, customers should sign and return the attached Customer Response Form within 10 days of receipt, acknowledging that they have read and understood the Field Safety Notice and Addendum. The Field Safety Notice must be placed in customers' User Documentation. Customers are to contact their local Philips Healthcare representative or office, if they need further information or support. |
Quantity in Commerce |
1,043 cameras |
Distribution |
Worldwide Distribution -- USA, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Denmark, Egypt, France, Germany, Hong Kong, Ireland, Italy, Korea, Lebanon, Malaysia, Netherlands, Norway, Pakistan, Portugal, Russia, Singapore, Spain, Sweden, Thailand, & United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KPS and Original Applicant = ADAC LABORATORIES
|
|
|
|