• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Virtual Imaging RadPRO 40 kW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Virtual Imaging RadPRO 40 kW see related information
Date Initiated by Firm September 22, 2014
Date Posted February 09, 2015
Recall Status1 Terminated 3 on February 20, 2018
Recall Number Z-1064-2015
Recall Event ID 69388
510(K)Number K101517  
Product Classification System, x-ray, mobile - Product Code IZL
Product Digital Mobile X-ray Systems used to take diagnostic radiographic exposures or patients.
Code Information 1) Model: MOBILE 32KW ANALOG; Serial Numbers: G29393 , G29395 , G29404 , G30236 , G30258 , G33309 , G33344 , G38069 , G38082 , G38711 , G38712 , G40196 , G40590 , G40989 , G41143 , G41163 , G60166 , G60169 , G60461 , G62020;  2) Model: MOBILE 40KW; Serial Numbers: G33126 , G35630 , G35641 , G35718 , G37357 , G31259;  3) Model: MOBILE 40KW 50G; Serial Numbers: G30684;  4) Model: MOBILE 40KW 55C; Serial Numbers: G37528;  5) Model: MOBILE 40KW 70C; Serial Numbers: G36211 , G36227 , G36281 , G36573 , G36581 , G36626 , G37168 , G38613 , G39518 , G40190 , G40227 , G60432 , G60435 , G60518 , G61367 , G61368 , G61720 , G61725 , G61726;  6) Model: MOBILE 40KW ANALOG; Serial Numbers: G30240;  7) Model: MOBILE2 40KW 50C; Serial Numbers: G31414 , G32487;  8) Model: MOBILE2 40KW 50G; Serial Numbers: G29416 , G29426 , G29824 , G29873 , G29878 , G30025 , G30155 , G30156 , G30160 , G30165 , G30173 , G30570 , G30685 , G30689 , G30761 , G30762 , G30767 , G30808 , G30952 , G31065 , G31067 , G31077 , G31403 , G31761 , G32577 , G32587 , G33004 , G33144 , G33581 , G33723 , G35450 , G35620;  9) Model: MOBILE2 40KW 55C; Serial Numbers: G32753 , G32764 , G33102 , G33111 , G33721 , G35462 , G35480;  10) Model: MOBILE2 40KW 55C/60C; Serial Numbers: G33585 , G37536;  11) Model: MOBILE2 40KW 55G; Serial Numbers: G32584 , G32949 , G33103 , G33127 , G33591 , G33646 , G33657 , G35422 , G35504;  12) Model: MOBILE2 40KW 55G/60G; Serial Numbers: G33711;  13) Model: MOBILE3 40KW; Serial Numbers: G41234 , G36566 , G38272;  14) Model: MOBILE3 40KW 55C; Serial Numbers: G37494 , G37501 , G37508 , G37587;  15) Model: MOBILE3 40KW 55C/60C; Serial Numbers: G35611 , G37490 , G37528 , G37588;  16) Model: MOBILE3 40KW 70C; Serial Numbers: 60597 , G36137 , G36148 , G36206 , G36217 , G36222 , G36244 , G36272 , G36287 , G36468 , G36492 , G36526 , G36537 , G36542 , G36601 , G36643 , G36681 , G36689 , G36732 , G36751 , G36761 , G37108 , G37111 , G37119 , G37129 , G37155 , G37158 , G37188 , G37196 , G37209 , G37234 , G37239 , G37293 , G37359 , G37413 , G37422 , G37430 , G37442 , G37442 , G37456 , G37466 , G37475 , G38058 , G38076 , G38173 , G38178 , G38190 , G38198 , G38279 , G38292 , G38293 , G38385 , G38391 , G38629 , G38635 , G38652 , G38657 , G39465 , G39467 , G39533 , G39551 , G39588 , G39589 , G39639 , G39650 , G39657 , G39782 , G39791 , G39800 , G39801 , G39808 , G39809 , G39818 , G40016 , G40024 , G40025 , G40116 , G40120 , G40124 , G40126 , G40185 , G40186 , G40199 , G40203 , G40211 , G40227 , G41058 , G41066 , G41083 , G41098 , G41098 , G41100 , G41104 , G41108 , G41109 , G41114 , G41119 , G41192 , G41203 , G41213 , G41218 , G41222 , G41244 , G41245 , G41256 , G60058 , G60061 , G60062 , G60063 , G60064 , G60065 , G60066 , G60069 , G-60069 , G60070 , G60406 , G60415 , G60418 , G60429 , G60430 , G60431 , G60520 , G60576 , G60577 , G60582 , G60606 , G60631 , G60635 , G60652 , G61307 , G61311 , G61333 , G61337 , G61339 , G61356 , G61393 , G61394 , G61698 , G61706 , G61715 , G61716 , G61786 , G61795 , G61812 , G61814 , G61818 , G61867 , G61872 , G61873 , G61880 , G62011 , G62012 , G62015 , G62029 , G62030 , G62037 , G62037 , G62053 , G62059 , G62062 , G62066 , G36042 , G37280 , G60400;  1)7 Model: MOBILE3 40KW 80C; Serial Numbers: G38060 , G39822 , G60067 , G61702;  18) Model: MOBILE3 40KW W/501C; Serial Numbers: G38666;  19) Model: RADPRO MOBILE; Serial Numbers: G29864 , G30250 , G30254 , G30567 , G30771 , G30902 , G30960 , G31252 , G31256 , G31271 , G31285 , G31287 , G31412 , G31480 , G31526 , G32218 , G32498 , G32514 , G32572 , G32581 , G32657 , G32671 , G33004 , G33103 , G33112 , G33126 , G33212 , G33238 , G33309 , G33324 , G33329 , G33333 , G-33634 , G33708 . G35389 , G35422 , G-35429 , G35430 , G35442 , G35524 , G35665
Recalling Firm/
Manufacturer
Virtual Imaging, Inc.
5600 Broken Sound Blvd.
2nd Flr.
Boca Raton FL 33487
For Additional Information Contact Technical Support
561-893-8500
Manufacturer Reason
for Recall
There is a potential for unintended and erratic motion.
FDA Determined
Cause 2
Component design/selection
Action Virtual Imaging sent an FIELD CORRECTION NOTIFICATION letter dated September 22, 2014, to all affected customers. The letters advised direct accounts that a Virtual Imaging Field Engineer with contact them to schedule an appointment for the replacement of the Digital Motion Control Kit. Customers with any concerns about this situation can contact Virtual Imaging's Technical Support at 561-893-8500 or via e-mail at technicalsupport@vifla.com.
Quantity in Commerce 316 (US) 6 (Foreign)
Distribution USA (nationwide) to the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV) + US Army facility in South Korea; and the country of Canada.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = IZL and Original Applicant = SEDECAL S.A.
-
-