|
Class 2 Device Recall MEDTRONIC PERFUSION TUBING PACKS |
|
Date Initiated by Firm |
August 18, 2015 |
Date Posted |
September 04, 2015 |
Recall Status1 |
Terminated 3 on May 09, 2016 |
Recall Number |
Z-2736-2015 |
Recall Event ID |
72025 |
510(K)Number |
K800178 K012538 K891687 K122811
|
Product Classification |
Tubing, pump, cardiopulmonary bypass - Product Code DWE
|
Product |
MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3, TL7M46R7, TL2Q68R17, 7U65R2, M074704B, M230002A, BB7M38Rll, M450013B, BB8N86R4, HY2U43R9, 8C72R4, M650123A, HY8R55R ,TL7M46R7, M971406C, M450013B, 7U65R2, M394415A, M450013B. Enables the perfusionist to manage the patient's hematocrit and fluid status.
|
Code Information |
209050007 209087876 209107465 209135076 209162502 209195930 209241496 209241613 209270018 209270092 209297994 209298091 209298094 209310403 209326025 209373545 209388464 209396017 209401269 209401312 209415313 209431890 209464644 209512240 209556873 209621268 209644548 209653827 |
Recalling Firm/ Manufacturer |
Medtronic Perfusion Systems 7611 Northland Dr N Brooklyn Park MN 55428-1088
|
For Additional Information Contact |
Medtronic Lifeline Technical Services 877-526-7890
|
Manufacturer Reason for Recall |
Hemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification. These potentially affected units have been distributed as stand-alone devices and in specific lots of Medtronic Perfusion Tubing packs.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Consignees were sent a Medtronic "Urgent Medical Device Recall" letter dated August 2015. The letter was addressed to "Risk Manager". The letter described the problem and the product involved in the recall. Informed consignees that no action is needed for product that has been used, but patients should continue to be managed in accordance with their standard patient management protocol. For unused product they recommend to immediately quarantine and return the affected product to Medtronic. Contact Customer Service at 800-854-3570 and reference RADAR #253277 to initiated a return and credit of product. Requested consignees to complete the Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality. If the product has been forwarded to another facility, consignees are advised to notify that facility. For questions contact Medtronic Lifeline Technical Services at 877-526-7890 or your Medtronic representative. |
Quantity in Commerce |
Total 1594 (1306 US, 288 OUS) |
Distribution |
Worldwide Distribution-US (nationwide) including the states of AL, AR, CA, CO, FL, HI, ID, IN, KS, LA, MA, MS, MO, NY, NJ, OH, PA, TN, TX, UT, and WI, and the countries of Bolivia, Denmark, Dominican Republic, Italy, Latvia, Lebanon, Mexico, Panama, Singapore, Venezuela, and Zambia. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DWE and Original Applicant = EXTRACORPOREAL MEDICAL SPECIALITIES, INC. 510(K)s with Product Code = DWE and Original Applicant = MEDTRONIC BLOOD SYSTEMS, INC. 510(K)s with Product Code = DWE and Original Applicant = MEDTRONIC INC. 510(K)s with Product Code = DWE and Original Applicant = MEDTRONIC VASCULAR
|
|
|
|