Date Initiated by Firm |
February 02, 2016 |
Date Posted |
August 08, 2016 |
Recall Status1 |
Terminated 3 on June 12, 2017 |
Recall Number |
Z-2353-2016 |
Recall Event ID |
73255 |
Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
Product |
S3 MedSurg Bed, Model 3005. Stryker Medical |
Code Information |
MDL#: E341062; 3005S3EX; 3005S3Patriot; 3005S3Signature; 3005S3PX1; 3005S3PX2; 3005S3PX3; 3005S3PX4; 3005S3EX; Serial Numbers: 141115067 through 150615027; Serial Numbers: 141115067 141115068 141115069 141115070 141115071 141115072 141115073 141115074 141115075 141115076 141115077 141115078 141115079 141115080 141115081 141115082 141115083 141115084 141115085 141115086 141115087 141115088 141115089 141115090 141115091 141115092 141115093 141115094 141115095 141115096 141115097 141115098 141115099 141115100 141115101 141115102 141115103 141115104 141115105 141115106 141115107 141115108 141115109 141115110 141115111 141115112 141115113 141115114 141115115 141115116 141115117 141115118 141115119 141115120 141115121 141115122 141115123 141115124 141115125 141115126 141115127 141115128 141115129 141115130 141115131 141115132 141115133 141115134 141115135 141115152 141115153 141115154 141115155 141115156 141115157 141115158 141115159 141115160 141115161 141115162 141115163 141115164 141115165 141115166 141115167 141115168 141115169 141115170 141115171 141115172 141115173 141115174 141115175 141115176 141115177 141115178 141115179 141115180 141115181 141115182 141115183 141115184 141115185 141115186 141115187 141115188 141115189 141115190 141115191 141115192 141115193 141115194 141115195 141115196 141115197 141115198 141115199 141115200 141115201 141115202 141115203 141115204 141115205 141115206 141115207 141115208 141115209 141115210 141115211 141115212 141115213 141115214 141115215 141115216 141115217 141115218 141115219 141115220 141115221 141115222 141115223 141115224 141115225 141115226 141115227 141115228 141115229 141115230 141115231 141115232 141115233 141115234 141115235 141115236 141115237 141115238 141115239 141115240 141115241 141115242 141115243 141115244 141115245 141115246 141115247 141115248 141115249 141115250 141115251 141115252 141115253 141115254 141115255 141115256 141115257 141115258 141115259 141115260 141115261 141115262 141115263 141115264 141115265 141115266 141115267 141115268 141115269 141115270 141115271 141115272 141115273 141115274 141115275 141115276 141115277 141115278 141115279 141115280 141115281 141115282 141115283 141115284 141115285 141115286 141115287 141115288 141115289 141115290 141115291 141115292 141115293 141115294 141115295 141115296 141115297 141115298 141115299 141115300 141115301 141115302 141115303 141115304 141115305 141115306 141115307 141115308 141115309 141115310 141115311 141115312 141115313 141115314 141115315 141115316 141115317 141115318 141115319 141115320 141115321 141115322 141115323 141115324 141115325 141115326 141115327 141115328 141115329 141115330 141115331 141115332 141115333 141115334 141115335 141115336 141115337 141115338 141115339 141115340 141115341 141115342 141115343 141115344 141115345 141115346 141115347 141115348 141115349 141115350 141115351 141115352 141115353 141115354 141115355 141115356 141115357 141115358 141115359 141115360 141115361 141115362 141115363 141115364 141115365 141115366 141115367 141115368 141115369 141115370 141115371 141115372 141115373 141115374 141115375 141115376 141115377 141115378 141115379 141115380 141115381 141115382 141115383 141115384 141115385 141115386 141115387 141115388 141115389 141115390 141115391 141115392 141115393 141115394 141115395 141115396 141115397 141115398 141115399 141115400 141115401 141115402 141115403 141115404 141115405 141115406 141115407 141115408 141115409 141115410 141115411 141115412 141115413 141115414 141115415 141115416 141115417 141115418 141115419 141115420 141115421 141115422 141115423 141115424 141115425 141115426 141115427 141115428 141115429 141115430 141115431 141115432 141115433 141115434 141115435 141115436 141115437 141115438 141115439 141115440 141115441 141115442 141115443 141115444 141115445 141115446 141115447 141115448 141115449 141115450 141115451 141115452 141115453 141115454 141115455 141115456 141115457 141115458 141115459 141115460 141115461 141115462 141115463 141115464 141115465 141115466 141115467 141115468 141115469 141115470 141115501 141115502 141115503 141115504 141115505 141115506 141115507 141115508 141115509 141115510 141115511 141115512 141115513 141115514 141115515 141115516 141115517 141115518 141115519 141115520 141115521 141115522 141115523 141115524 141115525 141115526 141115527 141115528 141115529 141115530 141115531 141115532 141115533 141115534 141115535 141115536 141115537 141115538 141115539 141115540 141115541 141115542 141115543 141115544 141115545 141115546 141115547 141115548 141115549 141115550 141115551 141115552 141115553 141115554 141115555 141115556 141115557 141115558 141115559 141115560 141115561 141115562 141115563 141115564 141115565 141115566 141115567 141115568 141115569 141115570 141115571 141115572 141115573 141115574 141115575 141115576 141115577 141115578 141115579 141115580 141115581 141115582 141115583 141115584 141115585 141115586 141115587 141115588 141115589 141115590 141115591 141115592 141115593 141115594 141115595 141115596 141115597 141115598 141115599 141115600 141115601 141115602 141115603 141115604 141115605 141115606 141115607 141115608 141115609 141115610 141115611 141115612 141115613 141115614 141115615 141115616 141115617 141115618 141115619 141115620 141115621 141115622 141115623 141115624 141115625 141115626 141115627 141115628 141115629 141115630 141115631 141115632 141115633 141115634 141115635 141115636 141115637 141115638 141115639 141115640 141115641 141115642 141115643 141115644 141115645 141115646 141115647 141115648 141115649 141115650 141215009 141215010 141215011 141215012 141215013 141215014 141215015 141215016 141215019 141215020 141215021 141215022 141215023 141215024 141215025 141215026 141215027 141215028 141215029 141215030 141215031 141215032 141215033 141215034 141215035 141215036 141215037 141215038 141215039 141215040 141215041 141215042 141215043 141215044 141215045 141215046 141215047 141215048 141215049 141215050 141215051 141215052 141215053 141215054 141215055 141215056 141215057 141215058 141215059 141215060 141215061 141215062 141215063 141215064 141215065 141215066 141215067 141215068 141215069 141215070 141215071 141215072 141215073 141215074 141215075 141215076 141215077 141215078 141215079 141215080 141215081 141215082 141215083 141215084 141215085 141215086 141215087 141215088 141215089 141215090 141215091 141215092 141215093 141215094 141215095 141215096 141215097 141215098 141215099 141215100 141215101 141215102 141215103 141215104 141215105 141215106 141215107 141215108 141215109 141215110 141215111 141215112 141215113 141215114 141215115 141215116 141215117 141215118 141215119 141215120 141215121 141215122 141215123 141215124 141215125 141215126 141215127 141215128 141215129 141215130 141215131 141215132 141215133 141215134 141215135 141215136 141215137 141215138 141215139 141215140 141215141 141215142 141215143 141215144 141215145 141215501 141215502 141215503 141215504 141215505 141215506 141215507 141215508 141215509 141215510 141215511 141215512 141215513 141215514 141215515 141215516 141215517 141215518 141215519 141215520 141215521 141215522 141215523 141215524 141215525 141215526 141215527 141215528 141215529 141215530 141215531 141215532 141215533 141215534 141215535 141215536 141215537 141215538 141215539 141215540 141215541 141215542 141215543 141215544 141215545 141215546 141215547 141215548 141215549 141215550 141215551 141215552 141215553 141215554 141215555 141215556 141215558 141215559 141215560 141215561 141215562 141215563 141215564 141215565 141215566 141215567 141215568 141215569 141215570 141215571 141215572 141215573 141215574 141215575 141215576 141215577 141215578 141215579 141215580 141215581 141215582 141215583 141215584 141215585 141215586 141215587 141215588 141215589 141215590 141215591 141215592 141215593 141215594 141215595 141215596 141215597 141215598 141215599 141215600 141215601 141215602 141215603 141215604 141215605 141215606 141215607 141215608 141215609 141215610 141215611 141215612 141215613 141215614 141215615 141215616 141215617 141215618 141215619 141215620 141215621 141215622 141215623 141215624 141215625 141215626 141215627 141215628 141215629 141215630 141215631 141215632 141215633 141215634 141215635 141215636 141215637 141215638 141215639 141215640 141215641 141215642 141215643 141215644 141215645 141215646 141215647 141215648 141215649 141215650 141215651 141215652 141215653 141215654 141215655 141215656 141215657 141215658 141215659 141215660 141215661 141215662 141215663 141215664 141215665 141215666 141215667 141215668 141215669 141215670 141215671 141215672 141215673 141215674 141215675 141215676 141215677 141215678 141215679 141215680 141215681 141215682 141215683 141215684 141215685 141215686 141215687 141215688 141215689 141215690 141215691 141215692 141215693 141215694 141215695 141215696 141215697 141215698 141215699 141215700 141215701 141215702 141215703 141215704 141215705 141215706 141215707 141215708 141215709 141215710 141215711 141215712 141215713 141215714 141215715 141215716 141215717 141215718 141215719 141215720 141215721 141215722 141215723 141215724 141215725 141215726 141215727 141215728 141215729 141215730 141215731 141215732 141215733 141215734 141215735 141215736 141215737 141215738 141215739 141215740 141215741 141215742 141215743 141215744 141215745 141215746 141215747 141215748 141215749 141215750 141215751 141215752 141215753 141215754 141215755 141215756 141215757 141215758 141215759 141215760 141215761 141215762 141215763 141215764 141215765 141215766 141215767 141215768 141215769 141215770 141215771 141215772 141215773 141215774 141215775 141215776 141215777 141215778 141215779 141215780 141215781 141215782 141215783 141215784 141215785 141215786 141215787 141215788 141215789 141215790 141215791 141215792 141215793 141215794 141215795 141215796 141215797 141215798 141215799 141215800 141215801 141215802 141215803 141215804 141215805 141215806 141215807 141215808 141215809 141215810 141215811 141215812 141215813 141215814 141215815 141215816 141215817 141215818 141215819 141215820 141215821 141215822 141215823 141215824 141215825 141215826 141215827 141215828 141215829 141215830 141215831 141215832 141215833 141215834 141215835 141215836 141215837 141215838 141215839 141215840 141215841 141215842 141215843 141215844 141215845 141215846 141215847 141215848 141215849 141215850 141215851 141215852 141215853 141215854 141215855 141215856 141215857 141215858 141215859 141215860 141215861 141215862 141215863 141215864 141215865 141215866 141215867 141215868 141215869 141215870 141215871 141215872 141215873 141215874 141215875 141215876 141215877 141215878 141215879 141215880 141215881 141215882 141215883 141215884 141215885 141215886 141215887 141215888 141215889 141215890 141215891 141215892 141215893 141215894 141215895 141215896 141215897 141215898 141215899 141215900 141215901 141215902 141215903 141215904 141215905 141215906 141215907 141215908 141215909 141215910 141215911 141215912 141215913 141215914 141215915 141215916 141215917 141215918 141215919 141215920 141215921 141215922 141215923 141215924 141215925 141215926 141215927 141215928 141215929 141215930 141215931 141215932 141215933 141215934 141215935 141215936 141215937 141215938 141215939 141215940 141215942 141215943 141215944 141215945 141215946 141215947 141215948 141215949 141215950 141215951 141215952 141215953 141215954 141215955 141215956 141215957 141215958 141215959 141215960 141215961 141215962 141215963 141215964 141215965 141215966 141215967 141215968 141215969 141215970 141215971 141215972 141215973 141215974 141215975 141215976 141215977 141215978 141215979 141215980 141215981 141215982 141215983 141215984 141215985 141215986 141215987 141215988 141215989 141215990 141215991 141215992 141215993 141215994 141215995 141215996 141215997 141215998 141216000 141216001 141216002 141216003 141216004 141216005 141216006 141216007 141216008 141216009 141216010 141216011 141216012 141216013 141216014 141216015 141216016 141216017 141216018 141216019 141216020 141216021 141216022 141216023 141216024 141216025 141216026 141216027 141216028 141216029 141216030 141216031 141216032 141216033 141216034 141216035 141216036 141216037 141216038 141216039 141216040 141216041 141216042 141216043 141216044 141216045 141216046 141216047 141216048 141216049 141216050 141216051 141216052 141216053 141216054 141216055 141216056 141216057 141216058 141216059 141216060 141216061 141216062 141216063 141216064 141216065 141216066 141216067 141216068 141216069 141216070 141216071 141216072 141216073 141216074 141216075 141216076 141216077 141216078 141216079 141216080 141216081 141216082 141216083 141216084 141216085 141216086 141216087 141216088 141216089 141216090 141216091 141216092 141216093 141216094 141216095 141216096 141216097 141216098 141216099 141216100 141216101 141216102 141216103 141216104 141216105 141216106 141216107 141216108 141216109 141216110 141216111 141216112 141216113 141216114 141216115 141216116 141216117 141216118 141216119 141216120 141216121 141216122 141216123 141216124 141216125 141216126 141216127 141216128 141216129 141216130 141216131 141216132 141216133 141216134 141216135 141216136 141216137 141216138 141216139 141216140 141216141 141216142 141216143 141216144 141216145 141216146 141216147 141216148 141216149 141216150 141216151 141216152 141216153 141216154 141216155 141216156 141216157 141216158 141216159 141216160 141216161 141216162 141216163 141216164 141216165 141216166 141216167 141216168 141216169 141216170 141216171 141216172 141216173 141216174 141216175 141216176 141216177 141216178 141216179 141216180 141216181 141216182 141216183 141216184 141216185 141216186 141216187 141216188 141216189 141216190 141216191 141216192 141216193 141216194 141216195 141216196 141216197 141216198 141216199 141216200 141216201 141216202 141216203 141216204 141216205 141216206 141216207 141216208 141216209 141216210 141216211 141216212 141216213 141216214 141216215 141216216 141216217 141216218 141216219 141216220 141216221 141216222 141216223 141216224 141216225 141216226 141216227 141216228 141216229 141216230 141216231 141216232 141216233 141216234 141216235 141216236 141216237 141216238 141216239 141216240 141216241 141216242 141216243 141216244 141216245 141216246 141216247 141216248 141216249 141216250 141216251 141216252 141216253 141216254 141216255 141216256 141216257 141216258 141216259 141216260 141216261 141216262 141216263 141216264 141216265 141216266 141216267 141216268 141216269 141216270 141216271 141216272 141216273 141216274 141216275 141216276 141216277 141216278 141216279 141216280 141216281 141216282 141216283 141216284 141216285 141216286 141216287 141216288 141216289 141216290 141216291 141216292 141216293 141216294 141216295 141216296 141216297 141216298 141216299 141216300 141216301 141216302 141216303 141216304 141216305 141216306 141216307 141216308 141216309 141216310 141216311 141216312 141216313 141216314 141216315 141216316 141216317 141216318 141216319 141216320 141216321 141216322 141216323 141216324 141216325 141216326 141216327 141216328 141216329 141216330 141216331 141216332 141216333 141216334 141216335 141216336 141216337 141216338 141216339 141216340 141216341 141216342 141216343 141216344 141216345 141216346 141216347 141216348 141216349 141216350 141216351 141216352 141216353 141216354 141216355 141216356 141216357 141216358 141216359 141216360 141216361 141216362 141216363 141216364 141216365 141216366 141216367 141216368 141216369 141216370 141216371 141216372 141216373 141216374 141216375 141216376 141216377 141216378 141216379 141216380 141216381 141216382 141216383 141216384 141216385 141216386 141216387 141216388 141216389 141216390 141216391 141216392 141216393 141216394 141216395 141216396 141216397 141216398 141216399 141216400 141216401 141216402 141216403 141216404 141216405 141216406 141216407 141216408 141216409 141216410 141216411 141216412 141216413 141216414 141216415 141216416 141216417 141216418 141216419 141216420 141216421 141216422 141216423 141216424 141216425 141216426 141216427 141216428 141216429 141216430 141216431 141216432 141216433 141216434 141216435 141216436 141216437 141216438 141216439 141216440 141216441 141216442 141216443 141216444 141216445 141216446 141216447 141216448 141216449 141216450 141216451 141216452 141216453 141216454 141216455 141216456 141216457 141216458 141216459 141216460 141216461 141216462 141216463 141216464 141216465 141216466 141216467 141216468 141216469 141216470 141216471 141216472 141216473 141216474 141216475 141216476 141216477 141216478 141216479 141216480 141216481 141216482 141216483 141216484 141216485 141216486 141216487 141216488 141216489 141216490 141216491 141216492 141216493 141216494 141216495 141216496 141216497 141216498 141216499 141216501 141216502 141216503 141216504 141216505 141216506 141216507 141216508 141216509 141216510 141216511 141216512 141216513 141216514 141216515 141216516 141216517 141216518 141216519 141216520 141216521 141216522 141216523 141216524 141216525 141216526 141216527 141216528 141216529 141216530 141216531 141216532 141216533 141216534 141216535 141216536 141216537 141216538 141216539 141216540 141216541 141216542 141216543 141216544 141216545 141216546 141216547 141216548 141216549 141216550 141216551 141216552 141216553 141216554 141216555 141216556 141216557 141216558 141216559 141216560 141216561 141216562 141216563 141216564 141216565 141216566 141216567 141216568 141216569 141216570 141216571 141216572 141216573 141216574 141216575 141216576 141216577 141216578 141216579 141216580 141216581 141216582 141216583 141216584 141216585 141216586 141216587 141216588 141216589 141216590 141216591 141216592 141216593 141216594 141216595 141216596 141216597 141216598 141216599 141216600 141216601 141216602 141216603 141216604 141216605 141216606 141216607 141216608 141216609 141216610 141216611 141216612 141216613 141216614 141216615 141216616 141216617 141216618 141216619 141216620 141216621 141216622 141216623 141216624 141216625 141216626 141216627 141216628 141216629 141216630 141216631 141216632 141216633 141216634 141216635 141216636 141216637 141216638 141216639 141216640 141216641 141216642 141216643 141216644 141216645 141216646 141216647 141216648 141216649 141216650 141216651 141216652 141216653 141216654 141216655 141216656 141216657 141216658 141216659 141216660 141216661 141216662 141216663 141216664 141216665 141216666 141216667 141216668 141216669 141216670 141216671 141216672 141216673 141216674 141216675 141216676 141216677 141216678 141216679 141216680 141216681 141216682 141216683 141216684 141216685 141216686 141216687 141216688 141216689 141216690 141216691 141216692 141216693 141216694 141216695 141216696 141216697 141216698 141216699 141216700 141216701 141216702 141216703 141216704 141216705 141216706 141216707 141216708 141216709 141216710 141216711 141216712 141216713 141216714 141216715 141216716 141216717 141216718 141216719 141216720 141216722 141216723 141216724 141216725 141216726 141216727 141216728 141216729 141216730 141216731 141216732 141216733 141216734 141216735 141216736 141216737 141216738 141216739 141216740 141216741 141216742 141216743 141216744 141216745 141216746 141216747 141216748 141216749 141216750 141216751 141216752 141216753 141216754 141216755 141216756 141216757 141216758 141216759 141216760 141216761 141216762 141216763 141216764 141216765 141216766 141216767 141216768 141216769 141216770 141216771 141216772 141216773 141216774 141216775 141216776 141216777 141216778 141216779 141216780 141216781 141216782 141216783 141216784 141216785 141216786 141216787 141216788 141216789 141216790 141216791 141216792 141216793 141216794 141216795 141216796 141216797 141216798 141216799 141216800 141216801 141216802 141216803 141216804 141216805 141216806 141216807 141216808 141216809 141216810 141216811 141216812 141216813 141216814 141216815 141216816 141216817 141216818 141216819 141216820 141216821 141216822 141216823 141216824 141216825 141216826 141216827 141216828 141216829 141216830 141216831 141216832 141216833 141216834 141216835 141216836 141216837 141216838 141216839 141216840 141216841 141216842 141216843 141216844 141216845 141216846 141216847 141216848 141216849 141216850 141216851 141216852 141216853 141216854 141216855 141216856 141216857 141216858 141216859 141216860 141216861 141216862 141216863 141216864 141216865 141216866 141216867 141216868 141216869 141216870 141216871 141216872 141216873 141216874 141216875 141216876 141216877 141216878 141216879 141216880 141216881 141216882 141216883 141216884 141216885 141216886 141216887 141216888 141216889 141216890 141216891 141216892 141216893 141216894 141216895 141216896 141216897 141216898 141216899 141216900 141216901 141216902 141216903 141216904 141216905 141216906 141216907 141216908 141216909 141216910 141216911 141216912 141216913 141216914 141216915 141216916 141216917 141216918 141216919 141216920 141216921 141216922 141216923 141216924 141216925 141216926 141216927 141216928 141216929 141216930 141216931 141216932 141216933 141216934 141216935 141216936 141216937 141216938 141216939 141216940 141216941 141216942 141216943 141216944 141216945 141216946 141216947 141216948 141216949 141216950 141216951 141216952 141216953 141216954 141216955 141216956 141216957 141216958 141216959 141216960 141216961 141216962 141216963 141216964 141216965 141216966 141216967 141216968 141216969 141216970 141216971 141216972 141216973 141216974 141216975 141216976 141216977 141216978 141216979 141216980 141216981 141216982 141216983 141216984 141216985 141216986 141216987 141216988 141216989 141216990 141216991 141216992 141216993 141216994 141216995 141216996 141216997 141216998 141216999 141217000 141217001 141217002 141217003 141217004 141217005 141217006 141217007 141217008 141217009 141217010 141217011 150115501 150115502 150115503 150115504 150115505 150115506 150115507 150115508 150115509 150115510 150115511 150115512 150115513 150115514 150115515 150115516 150115517 150115518 150115519 150115520 150115521 150115522 150115523 150115524 150115525 150115526 150115527 150115528 150115529 150115530 150115531 150115532 150115533 150115534 150115535 150115536 150115537 150115538 150115539 150115540 150115541 150115542 150115543 150115544 150115545 150115546 150115547 150115548 150115549 150115550 150115551 150115552 150115553 150115554 150115555 150115556 150115557 150115558 150115559 150115560 150115561 150115562 150115563 150115564 150115565 150115566 150115567 150115568 150115569 150115570 150115571 150115572 150115573 150115574 150115575 150115576 150115577 150115578 150115579 150115580 150115581 150115582 150115583 150115584 150115585 150115586 150115587 150115588 150115589 150115590 150115591 150115592 150115593 150115594 150115595 150115596 150115597 150115598 150115599 150115600 150115601 150115602 150115603 150115604 150115605 150115606 150115607 150115608 150115609 150115610 150115611 150115612 150115613 150115614 150115615 150115616 150115617 150115618 150115619 150115620 150115621 150115622 150115623 150115624 150115625 150115626 150115627 150115628 150115629 150115630 150115631 150115632 150115633 150115634 150115635 150115636 150115637 150115638 150115639 150115640 150115641 150115642 150115643 150115644 150115645 150115646 150115647 150115648 150115649 150115650 150115651 150115652 150115653 150115654 150115655 150115656 150115657 150115658 150115659 150115660 150115661 150115662 150115663 150115664 150115665 150115666 150115667 150115668 150115669 150115670 150115671 150115672 150115673 150115674 150115675 150115676 150115677 150115678 150115679 150115680 150115681 150115682 150115683 150115684 150115685
150115686 150115687 150115688 150115689 150115690 150115691 150115692 150115693 150115694 150115695 150115696 150115697 150115698 150115699 150115700 150115701 150115702 150115703 150115704 150115705 150115706 150115707 150115708 150115709 150115710 150115711 150115712 150115713 150115714 150115715 150115716 150115717 150115718 150115719 150115720 150115721 150115722 150115723 150115724 150115725 150115726 150115727 150115728 150115729 150115730 150115731 150115732 150115733 150115734 150115735 150115736 150115737 150115738 150115739 150115740 150115741 150115742 150115743 150115744 150115745 150115746 150115747 150115748 150115749 150115750 150115751 150115752 150115753 150115754 150115755 150115756 150115757 150115758 150115759 150115760 150115761 150115762 150115763 150115764 150115765 150115766 150115767 150115768 150115769 150115770 150115771 150115772 150115773 150115774 150115775 150115776 150115777 150115778 150115779 150115780 150115781 150115782 150115783 150115784 150115785 150115786 150115787 150115788 150115789 150115790 150115791 150115792 150115793 150115794 150115795 150115796 150115797 150115798 150115799 150115800 150115801 150115802 150115803 150115804 150115805 150115806 150115807 150115808 150115809 150115810 150115811 150115812 150115813 150115814 150115815 150115816 150115817 150115818 150115819 150115820 150115821 150115822 150115823 150115824 150115825 150115826 150115827 150115828 150115829 150115830 150115831 150115832 150115833 150115834 150115835 150115836 150115837 150115838 150115839 150115840 150115841 150115842 150115843 150115844 150115845 150115846 150115847 150115848 150115849 150115850 150115851 150115852 150115853 150115854 150115855 150115856 150115857 150115858 150115859 150115860 150115861 150115862 150115863 150115864 150115865 150115866 150115867 150115868 150115869 150115870 150115871 150115872 150115873 150115874 150115875 150115876 150115877 150115878 150115879 150115880 150115881 150115882 150115883 150115884 150115885 150115886 150115887 150115888 150115889 150115890 150115891 150115892 150115893 150115894 150115895 150115896 150115897 150115898 150115899 150115900 150115901 150115902 150115903 150115904 150115905 150115906 150115907 150115908 150115909 150115910 150115911 150115912 150115913 150115914 150115915 150115916 150115917 150115918 150115919 150115920 150115921 150115922 150115923 150115924 150115925 150115926 150115927 150115928 150115929 150115930 150115931 150115932 150115933 150115934 150115935 150115936 150115937 150115938 150115939 150115940 150115941 150115942 150115943 150115944 150115945 150115946 150115947 150115948 150115949 150115950 150115951 150115952 150115953 150115954 150115955 150115956 150115957 150115958 150115959 150115960 150115961 150116001 150116002 150116003 150116004 150116005 150116006 150116007 150116008 150116009 150116010 150116011 150116012 150116013 150116014 150116015 150116016 150116017 150116018 150116019 150116020 150116021 150116022 150116023 150116024 150116025 150116026 150116027 150116028 150116029 150116030 150116031 150116032 150116033 150116034 150116035 150116036 150116037 150116038 150116039 150116040 150116041 150116042 150116043 150116044 150116045 150116046 150116047 150116048 150116049 150116050 150116051 150116052 150116053 150116054 150116055 150116056 150116057 150116058 150116059 150116060 150116061 150116062 150116063 150116064 150116065 150116066 150116067 150116068 150116069 150116070 150116071 150116072 150116073 150116074 150116075 150116076 150116077 150116078 150116079 150116080 150116081 150116082 150116083 150116084 150116085 150116086 150116087 150116088 150116089 150116090 150116091 150116092 150116093 150116094 150116095 150116096 150116097 150116098 150116099 150116100 150116101 150116102 150116103 150116104 150116105 150116106 150116107 150116108 150116109 150116110 150116111 150116112 150116113 150116114 150116115 150116116 150116117 150116118 150116119 150116120 150116121 150116122 150116123 150116124 150116125 150116126 150116127 150116128 150116129 150116130 150116131 150116132 150116133 150116134 150116135 150116136 150116137 150116138 150116139 150116140 150116141 150116142 150116143 150116144 150116145 150116146 150116147 150116148 150116149 150116150 150116151 150116152 150116153 150116154 150116155 150116156 150116157 150116158 150116159 150116160 150116161 150116162 150116163 150116164 150116165 150116166 150116167 150116168 150116169 150116170 150116171 150116172 150116173 150116174 150116175 150116176 150116177 150116178 150116179 150116180 150116181 150116182 150116183 150116184 150116185 150116186 150116187 150116188 150116189 150116190 150116191 150116192 150116193 150116194 150116195 150116196 150116197 150116198 150116199 150116200 150116201 150116202 150116203 150116204 150116205 150116206 150116207 150116208 150116209 150116210 150116211 150116212 150116213 150116214 150116215 150116216 150116217 150116218 150116219 150116220 150116221 150116222 150116223 150116224 150116225 150116226 150116227 150116228 150116229 150116230 150116231 150116232 150116233 150116234 150116235 150116236 150116237 150116238 150116239 150116240 150116241 150116242 150116243 150116244 150116245 150116246 150116247 150116248 150116249 150116250 150116251 150116252 150116253 150116254 150116255 150116256 150116257 150116258 150116259 150116260 150116261 150116262 150116263 150116264 150116265 150116266 150116267 150116268 150116269 150116270 150116271 150116272 150116273 150116274 150116275 150116276 150116277 150116278 150116279 150116280 150116281 150116282 150116283 150116284 150116285 150116286 150116287 150116288 150116289 150116290 150116291 150116292 150116293 150116294 150116295 150116296 150116297 150116298 150116299 150116300 150116301 150116302 150116303 150116304 150116305 150116306 150116307 150116308 150116309 150116310 150116311 150116312 150116313 150116314 150116315 150116316 150116317 150116318 150116319 150116320 150116321 150116322 150116323 150116324 150116325 150116326 150116327 150116328 150116329 150116330 150116331 150116332 150116333 150116334 150116335 150116336 150116337 150116338 150116339 150116340 150116341 150116342 150116343 150116344 150116345 150116346 150116347 150116348 150116349 150116350 150116351 150116352 150116353 150116354 150116355 150116356 150116357 150116358 150116359 150116360 150116361 150116362 150116363 150116364 150116365 150116366 150116367 150116368 150116369 150116370 150116371 150116372 150116373 150116374 150116375 150116376 150116377 150116378 150116379 150116380 150116381 150116382 150116383 150116384 150116385 150116386 150116387 150116388 150116389 150116390 150116391 150116392 150116393 150116394 150116395 150116396 150116397 150116398 150116399 150116400 150116401 150116402 150116403 150116404 150116405 150116406 150116407 150116408 150116409 150116410 150116411 150116412 150116413 150116414 150116415 150116416 150116417 150116418 150116419 150116420 150116421 150116422 150116423 150116424 150116425 150116426 150116427 150116428 150116429 150116430 150116431 150116432 150116433 150116434 150116435 150116436 150116437 150116438 150116439 150116440 150116441 150116442 150116443 150116444 150116445 150116446 150116447 150116448 150116449 150116450 150116451 150116452 150116453 150116454 150116455 150116456 150116457 150116458 150116459 150116460 150116461 150116462 150116463 150116464 150116465 150116466 150116467 150116468 150116469 150116470 150116471 150116472 150116473 150116474 150116475 150116476 150116477 150116478 150116479 150116480 150116481 150116482 150116483 150116484 150116485 150116486 150116487 150116488 150116489 150116490 150116491 150116492 150116493 150116494 150116495 150116496 150116497 150116498 150116499 150116500 150116501 150116502 150116503 150116504 150116505 150116506 150116507 150116508 150116509 150116510 150116511 150116512 150116513 150116514 150116515 150116516 150116517 150116518 150116519 150116520 150116521 150116522 150116523 150116524 150116525 150116526 150116527 150116528 150116529 150116530 150116531 150116532 150116533 150116534 150116535 150116536 150116537 150116538 150116539 150116540 150116541 150116542 150116543 150116544 150116545 150116546 150116547 150116548 150116549 150116550 150116551 150116552 150116553 150116554 150116555 150116556 150116557 150116558 150116559 150116560 150116561 150116562 150116563 150116564 150116565 150116566 150116567 150116568 150116569 150116570 150116571 150116572 150116573 150116574 150116575 150116576 150116577 150116578 150116579 150116580 150116581 150116582 150116583 150116584 150116585 150116586 150116587 150116588 150116589 150116590 150116591 150116592 150116593 150116594 150116595 150116596 150116597 150116598 150116599 150116600 150116602 150116603 150116604 150116605 150116606 150116607 150116608 150116609 150116610 150116611 150116612 150116613 150116614 150116615 150116616 150116617 150116618 150116619 150116620 150116621 150116622 150116623 150116624 150116625 150116626 150116627 150116628 150116629 150116630 150116631 150116632 150116633 150116634 150116635 150116636 150116637 150116638 150116639 150116640 150116641 150116642 150116643 150116644 150116645 150116646 150116647 150116648 150116649 150116650 150116651 150116652 150116653 150116654 150116655 150116656 150116657 150116658 150116659 150116660 150116661 150116662 150116663 150116664 150116665 150116666 150116667 150116668 150116669 150116670 150116671 150116672 150116673 150116674 150116675 150116676 150116677 150116678 150116679 150116680 150116681 150116682 150116683 150116684 150116685 150116686 150116687 150116688 150116689 150116690 150116691 150116692 150116693 150116694 150116695 150116696 150116697 150116698 150116699 150116700 150116701 150116702 150116703 150116704 150116705 150116706 150116707 150116708 150116709 150116710 150116711 150116712 150116713 150116714 150116715 150116716 150116717 150116718 150116719 150116720 150116721 150116722 150116723 150116724 150116725 150116726 150116727 150116728 150116729 150116730 150116731 150116732 150116733 150116734 150116735 150116736 150116737 150116738 150116739 150116740 150116741 150116742 150116743 150116744 150116745 150116746 150116747 150116748 150116749 150116750 150116751 150116752 150116753 150116754 150116755 150116756 150116757 150116758 150116759 150116760 150116761 150116762 150116763 150116764 150116765 150116766 150116767 150116768 150116769 150116770 150116771 150116772 150116773 150116774 150116775 150116776 150116777 150116778 150116779 150116780 150116781 150116782 150116783 150116784 150116785 150116786 150116787 150116788 150116789 150116790 150116791 150116792 150116793 150116794 150116795 150116796 150116797 150116798 150116799 150116800 150116801 150116802 150116803 150116804 150116805 150116806 150116807 150116808 150116809 150116810 150116811 150116812 150116813 150116814 150116815 150116816 150116817 150116818 150116819 150116820 150116821 150116822 150116823 150116824 150116825 150116826 150116827 150116828 150116829 150116830 150116831 150116832 150116833 150116834 150116835 150116836 150116837 150116838 150116839 150116840 150116841 150116842 150116843 150116844 150116845 150116846 150116847 150116848 150116849 150116850 150116851 150116852 150116853 150116854 150116855 150116856 150116857 150116858 150116859 150116860 150116861 150116862 150116863 150116864 150116865 150116866 150116867 150116868 150116869 150116870 150116871 150116872 150116873 150116874 150116875 150116876 150116877 150116878 150116879 150116880 150116881 150116882 150116883 150116884 150116885 150116886 150116887 150116888 150116889 150116890 150116891 150116892 150116893 150116894 150116895 150116896 150116897 150116898 150116899 150116900 150116901 150116902 150116903 150116904 150116905 150116906 150116907 150116908 150116909 150116910 150116911 150116912 150116913 150116914 150116915 150116916 150116917 150116918 150116919 150116920 150116921 150116922 150116923 150116924 150116925 150116926 150116927 150116928 150116929 150116930 150116931 150116932 150116933 150116934 150116935 150116936 150116937 150116938 150116939 150116940 150116941 150116942 150116943 150116944 150116945 150116946 150116947 150116948 150116949 150116950 150116951 150116952 150116953 150116954 150116955 150116956 150116957 150116958 150116959 150116960 150116961 150116962 150116963 150116964 150116965 150116966 150116967 150116968 150116970 150116971 150116972 150116973 150116974 150116975 150116976 150116977 150116978 150116979 150116980 150116981 150116982 150116983 150116984 150116985 150116986 150116987 150116988 150116989 150116990 150215001 150215002 150215003 150215004 150215005 150215006 150215007 150215008 150215009 150215010 150215011 150215012 150215013 150215014 150215015 150215016 150215017 150215018 150215019 150215020 150215021 150215022 150215023 150215024 150215025 150215026 150215027 150215028 150215029 150215030 150215031 150215032 150215033 150215034 150215035 150215036 150215037 150215038 150215039 150215040 150215041 150215042 150215043 150215044 150215045 150215046 150215047 150215048 150215049 150215050 150215051 150215052 150215053 150215054 150215055 150215056 150215057 150215058 150215059 150215060 150215061 150215062 150215063 150215064 150215065 150215066 150215067 150215068 150215069 150215070 150215071 150215072 150215073 150215074 150215075 150215076 150215085 150215086 150215087 150215088 150215089 150215090 150215091 150215092 150215093 150215094 150215095 150215096 150215097 150215098 150215099 150215100 150215101 150215102 150215103 150215104 150215105 150215106 150215107 150215108 150215109 150215110 150215111 150215112 150215113 150215114 150215115 150215116 150215117 150215118 150215119 150215120 150215121 150215122 150215123 150215124 150215125 150215126 150215127 150215128 150215129 150215130 150215131 150215132 150215133 150215134 150215139 150215141 150215142 150215143 150215144 150215145 150215146 150215147 150215148 150215149 150215150 150215151 150215152 150215153 150215154 150215155 150215156 150215157 150215158 150215159 150215160 150215161 150215162 150215163 150215164 150215165 150215166 150215167 150215168 150215169 150215170 150215171 150215172 150215173 150215174 150215175 150215176 150215177 150215178 150215179 150215180 150215181 150215182 150215183 150215184 150215185 150215186 150215187 150215188 150215189 150215190 150215191 150215192 150215193 150215194 150215195 150215196 150215197 150215198 150215199 150215200 150215201 150215202 150215203 150215204 150215205 150215206 150215207 150215208 150215209 150215210 150215211 150215212 150215213 150215214 150215215 150215216 150215217 150215218 150215219 150215220 150215221 150215222 150215223 150215224 150215225 150215226 150215227 150215228 150215229 150215230 150215231 150215232 150215233 150215234 150215235 150215236 150215237 150215238 150215239 150215240 150215241 150215242 150215243 150215244 150215245 150215246 150215247 150215248 150215249 150215250 150215251 150215252 150215253 150215254 150215255 150215256 150215257 150215258 150215259 150215260 150215261 150215262 150215263 150215264 150215265 150215266 150215267 150215268 150215269 150215270 150215271 150215272 150215273 150215274 150215275 150215276 150215277 150215278 150215279 150215280 150215281 150215282 150215283 150215284 150215285 150215286 150215287 150215288 150215289 150215290 150215291 150215292 150215293 150215294 150215295 150215296 150215297 150215298 150215299 150215300 150215301 150215302 150215303 150215304 150215305 150215306 150215307 150215308 150215309 150215310 150215311 150215312 150215313 150215314 150215315 150215316 150215317 150215318 150215319 150215320 150215321 150215322 150215323 150215324 150215325 150215326 150215327 150215328 150215329 150215330 150215331 150215332 150215333 150215334 150215335 150215336 150215337 150215338 150215339 150215340 150215341 150215342 150215343 150215344 150215345 150215346 150215347 150215348 150215349 150215350 150215351 150215352 150215353 150215354 150215355 150215356 150215357 150215358 150215359 150215360 150215361 150215362 150215363 150215364 150215365 150215366 150215367 150215368 150215369 150215370 150215371 150215372 150215373 150215374 150215375 150215376 150215377 150215378 150215379 150215380 150215381 150215382 150215383 150215384 150215385 150215386 150215387 150215388 150215389 150215390 150215391 150215392 150215393 150215394 150215395 150215396 150215397 150215398 150215399 150215400 150215401 150215402 150215403 150215404 150215405 150215406 150215407 150215408 150215409 150215410 150215411 150215412 150215413 150215414 150215415 150215416 150215417 150215418 150215419 150215420 150215421 150215422 150215423 150215424 150215425 150215426 150215427 150215428 150215429 150215430 150215431 150215432 150215433 150215434 150215435 150215436 150215437 150215438 150215439 150215440 150215441 150215442 150215443 150215444 150215445 150215446 150215447 150215448 150215449 150215450 150215451 150215452 150215453 150215454 150215455 150215456 150215457 150215458 150215459 150215460 150215461 150215462 150215463 150215464 150215465 150215466 150215467 150215468 150215469 150215470 150215471 150215472 150215473 150215474 150215475 150215476 150215477 150215478 150215479 150215480 150215481 150215482 150215483 150215484 150215485 150215486 150215487 150215501 150215502 150215503 150215504 150215505 150215506 150215507 150215508 150215509 150215510 150215511 150215512 150215513 150215514 150215515 150215516 150215517 150215518 150215519 150215520 150215521 150215522 150215523 150215524 150215525 150215526 150215527 150215528 150215529 150215530 150215531 150215532 150215533 150215534 150215535 150215536 150215537 150215538 150215539 150215540 150215541 150215542 150215543 150215544 150215545 150215546 150215547 150215548 150215549 150215550 150215584 150215585 150215586 150215587 150215588 150215589 150215590 150215591 150215592 150215593 150215594 150215595 150215596 150215597 150215598 150215599 150215600 150215601 150215602 150215603 150215604 150215605 150215606 150215607 150215608 150215609 150215610 150215611 150215612 150215613 150215614 150215615 150215616 150215617 150215618 150215619 150215620 150215621 150215622 150215623 150215624 150215625 150215626 150215627 150215628 150215629 150215630 150215631 150215632 150215633 150215634 150215635 150215636 150215637 150215638 150215639 150215640 150215641 150215642 150215643 150215644 150215645 150215646 150215647 150215648 150215649 150215650 150215651 150215652 150215653 150215654 150215655 150215656 150215657 150215658 150215659 150215660 150215661 150215662 150215663 150215664 150215665 150215666 150215667 150215668 150215669 150215670 150215671 150215672 150215673 150215674 150215675 150215676 150215677 150215678 150215679 150215680 150215681 150215682 150215683 150215684 150215685 150215686 150215687 150215688 150215689 150215690 150215691 150215692 150215693 150215694 150215695 150215696 150215697 150215698 150215699 150215700 150215701 150215702 150215703 150215704 150215705 150215706 150215707 150215708 150215709 150215710 150215711 150215712 150215713 150215714 150215715 150215716 150215717 150215718 150215719 150215720 150215721 150215722 150215723 150215724 150215725 150215726 150215727 150215728 150215729 150215730 150215731 150215732 150215733 150215734 150215735 150215736 150215737 150215738 150215741 150215742 150215743 150215744 150215745 150215746 150215747 150215748 150215749 150215750 150215751 150215752 150215753 150215754 150215755 150215756 150215757 150215758 150215759 150215760 150215761 150215762 150215763 150215764 150215765 150215766 150215767 150215768 150215769 150215770 150215771 150215772 150215773 150215774 150215775 150215776 150215777 150215778 150215779 150215780 150215781 150215782 150215783 150215784 150215785 150215786 150215787 150215788 150215789 150215790 150215791 150215792 150215793 150215794 150215795 150215796 150215797 150215798 150215799 150215800 150215801 150215802 150215803 150215804 150215805 150215806 150215807 150215808 150215809 150215810 150215811 150215812 150215813 150215814 150215815 150215816 150215817 150215818 150215819 150215820 150215821 150215822 150215823 150215824 150215825 150215826 150215827 150215828 150215829 150215830 150215831 150215832 150215833 150215834 150215835 150215836 150215837 150215838 150215839 150215840 150215841 150215842 150215843 150215844 150215845 150215846 150215847 150215848 150215849 150215850 150215851 150215852 150215853 150215854 150215855 150215856 150215857 150215858 150215859 150215860 150215861 150215862 150215863 150215864 150215865 150215866 150215867 150215868 150215869 150215870 150215871 150215872 150215873 150215874 150215875 150215876 150215877 150215878 150215879 150215880 150215881 150215882 150215883 150215884 150215885 150215886 150215887 150215888 150215889 150215890 150215891 150215892 150215893 150215894 150215895 150215896 150215897 150215898 150215899 150215900 150215901 150215902 150215903 150215904 150215905 150215906 150215907 150215908 150215909 150215910 150215911 150215912 150215913 150215914 150215915 150215916 150215917 150215918 150215919 150215920 150215921 150215922 150215923 150215924 150215925 150215926 150215927 150215928 150215929 150215930 150215931 150215932 150215933 150215934 150215935 150215936 150215937 150215938 150215939 150215940 150215941 150215942 150215943 150215944 150215945 150215946 150215947 150215948 150215949 150215950 150215951 150215952 150215953 150215954 150215955 150215956 150215957 150215958 150215959 150215960 150215961 150215962 150215963 150215964 150215965 150215966 150215967 150215968 150215969 150215970 150215971 150215972 150215973 150215974 150215975 150215976 150215977 150215978 150215979 150215980 150215981 150215982 150215983 150215984 150215985 150215986 150215987 150215988 150215989 150216001 150216002 150216003 150216004 150216005 150216006 150216007 150216008 150216009 150216010 150216011 150216012 150216013 150216014 150216015 150216016 150216017 150216018 150216019 150216020 150216021 150216022 150216023 150216024 150216025 150216026 150216027 150216028 150216029 150216030 150216031 150216032 150216033 150216034 150216035 150216036 150216037 150216038 150216039 150216040 150216041 150216042 150216043 150216044 150216045 150216046 150216047 150216048 150216049 150216050 150216051 150216052 150216053 150216054 150216055 150216056 150216057 150216058 150216059 150216060 150216061 150216062 150216063 150216064 150216065 150216066 150216067 150216068 150216069 150216070 150216071 150216072 150216073 150216074 150216075 150216076 150216077 150216078 150216079 150216080 150216081 150216082 150216083 150216084 150216085 150216086 150216087 150216088 150216089 150216090 150216091 150216092 150216093 150216094 150216095 150216096 150216097 150216098 150216099 150216100 150216101 150216102 150216103 150216104 150216105 150216106 150216107 150216108 150216109 150216110 150216111 150216112 150216113 150216114 150216115 150216116 150216117 150216118 150216119 150216120 150216121 150216122 150216123 150216124 150216125 150216126 150216127 150216128 150216129 150216130 150216131 150216132 150216133 150216134 150216135 150216136 150216137 150216138 150216139 150216140 150216141 150216142 150216143 150216144 150216145 150216146 150216147 150216148 150216149 150216150 150216151 150216152 150216153 150216154 150216155 150216156 150216157 150216158 150216159 150216160 150216161 150216162 150216163 150216164 150216165 150216166 150216167 150216168 150216169 150216170 150216171 150216172 150216173 150216174 150216175 150216176 150216177 150216178 150216179 150216180 150216181 150216182 150216183 150216184 150216185 150216186 150216187 150216188 150216189 150216190 150216191 150216192 150216193 150216194 150216195 150216196 150216197 150216198 150216199 150216200 150216201 150216202 150216203 150216204 150216205 150216206 150216207 150216208 150216209 150216210 150216211 150216212 150216213 150216214 150216215 150216216 150216217 150216218 150216219 150216222 150216223 150216224 150216225 150216226 150216227 150216228 150216229 150216230 150216231 150216232 150216233 150216234 150216235 150216236 150216237 150216238 150216239 150216240 150216241 150216242 150216243 150216244 150216245 150216246 150216247 150216248 150216249 150216250 150216251 150216252 150216253 150216254 150216255 150216256 150216257 150216258 150216259 150216260 150216261 150216262 150216263 150216264 150216265 150216266 150216267 150216268 150216269 150216270 150216271 150216272 150216273 150216274 150216275 150216276 150216277 150216278 150216279 150216280 150216281 150216282 150216283 150216284 150216285 150216286 150216287 150216288 150216289 150216290 150216291 150216292 150216293 150216294 150216295 150216296 150216297 150216298 150216299 150216300 150216301 150216302 150216303 150216304 150216305 150216306 150216307 150216308 150216309 150216310 150216311 150216312 150216313 150216315 150216316 150216317 150216318 150216319 150216320 150216321 150216322 150216323 150216324 150216325 150216326 150216327 150216328 150216329 150216330 150216331 150216332 150216333 150216334 150216335 150216336 150216337 150216338 150216339 150216340 150216341 150216342 150216343 150216344 150216345 150216346 150216347 150216348 150216349 150216350 150216351 150216352 150216353 150216354 150216355 150216356 150216357 150216358 150216359 150216360 150216361 150216362 150216363 150216364 150216365 150216366 150216367 150216368 150216369 150216370 150216371 150216372 150216375 150216376 150216377 150216378 150216379 150216380 150216381 150216382 150216383 150216384 150216385 150216386 150216387 150216388 150216389 150216390 150216391 150216392 150216393 150216394 150216395 150216396 150216397 150216398 150216399 150216400 150216401 150216402 150216403 150216404 150216405 150216406 150216407 150216408 150216409 150216410 150216411 150216412 150216413 150216414 150216415 150216416 150216417 150216418 150216419 150216420 150216421 150216422 150216423 150216424 150216425 150216426 150216427 150216428 150216429 150216430 150216431 150216432 150216433 150216434 150216435 150216436 150216437 150216438 150216439 150216440 150216441 150216442 150216443 150216444 150216445 150216446 150216447 150216448 150216449 150216450 150216451 150216452 150216453 150216454 150216455 150216456 150216457 150216458 150216459 150216460 150216461 150216462 150216463 150216464 150216465 150216466 150216467 150216468 150216469 150216470 150216488 150216489 150216490 150216491 150216492 150216493 150216494 150216495 150216496 150216497 150216498 150216499 150216500 150216501 150216502 150216503 150216504 150216505 150216506 150216507 150216508 150216509 150216510 150216511 150216513 150216514 150216515 150216516 150216517 150216518 150216519 150216520 150216521 150216522 150216523 150216524 150216525 150216526 150216527 150216528 150216529 150216530 150216531 150216532 150216533 150216534 150216535 150216536 150216537 150216538 150216539 150216540 150216541 150216542 150216543 150216544 150216545 150216546 150216547 150216548 150216549 150216550 150216551 150216552 150216553 150216554 150216555 150216556 150216557 150216558 150216559 150216560 150216561 150216562 150216563 150216564 150216565 150216566 150216567 150216568 150216569 150216570 150216571 150216572 150216573 150216574 150216575 150216576 150216577 150216578 150216579 150216580 150216581 150216582 150216583 150216584 150216585 150216586 150216587 150216588 150216589 150216590 150216591 150216592 150216593 150216594 150216595 150216596 150216597 150216598 150216599 150216600 150216601 150216602 150216603 150216604 150216605 150315001 150315002 150315003 150315004 150315005 150315006 150315007 150315008 150315009 150315010 150315011 150315012 150315013 150315014 150315015 150315016 150315017 150315018 150315019 150315020 150315021 150315022 150315023 150315024 150315025 150315026 150315027 150315028 150315029 150315030 150315031 150315032 150315033 150315034 150315035 150315036 150315037 150315039 150315040 150315041 150315042 150315043 150315044 150315045 150315046 150315047 150315048 150315049 150315050 150315051 150315052 150315053 150315054 150315055 150315056 150315057 150315058 150315059 150315060 150315061 150315062 150315063 150315064 150315065 150315066 150315067 150315068 150315069 150315070 150315071 150315072 150315073 150315074 150315075 150315076 150315077 150315078 150315079 150315080 150315081 150315082 150315083 150315084 150315085 150315086 150315087 150315088 150315089 150315090 150315091 150315092 150315093 150315094 150315095 150315096 150315097 150315098 150315099 150315100 150315101 150315102 150315103 150315104 150315105 150315106 150315107 150315108 150315109 150315110 150315111 150315112 150315113 150315114 150315115 150315116 150315117 150315118 150315119 150315120 150315121 150315122 150315123 150315124 150315125 150315126 150315127 150315128 150315129 150315130 150315131 150315132 150315133 150315134 150315135 150315136 150315137 150315138 150315139 150315140 150315141 150315142 150315143 150315144 150315145 150315146 150315147 150315148 150315149 150315150 150315151 150315152 150315153 150315154 150315155 150315156 150315157 150315158 150315159 150315160 150315161 150315162 150315163 150315164 150315165 150315166 150315167 150315168 150315169 150315170 150315171 150315172 150315173 150315174 150315175 150315176 150315177 150315178 150315179 150315180 150315181 150315182 150315183 150315184 150315185 150315186 150315187 150315188 150315189 150315190 150315191 150315192 150315193 150315194 150315195 150315196 150315197 150315198 150315199 150315200 150315201 150315202 150315203 150315204 150315205 150315206 150315207 150315208 150315209 150315210 150315211 150315212 150315213 150315214 150315215 150315216 150315217 150315218 150315219 150315220 150315221 150315222 150315223 150315224 150315225 150315226 150315227 150315228 150315229 150315230 150315231 150315232 150315233 150315234 150315235 150315236 150315237 150315238 150315239 150315240 150315241 150315242 150315243 150315244 150315245 150315246 150315247 150315248 150315249 150315250 150315251 150315252 150315253 150315254 150315255 150315256 150315257 150315258 150315259 150315260 150315261 150315262 150315263 150315264 150315266 150315308 150315309 150315310 150315311 150315312 150315313 150315314 150315315 150315316 150315317 150315318 150315319 150315320 150315321 150315322 150315323 150315324 150315325 150315326 150315327 150315328 150315329 150315330 150315331 150315332 150315333 150315334 150315335 150315336 150315337 150315338 150315339 150315340 150315341 150315342 150315343 150315344 150315345 150315346 150315347 150315348 150315349 150315350 150315351 150315352 150315353 150315354 150315355 150315356 150315357 150315358 150315359 150315360 150315361 150315362 150315363 150315364 150315365 150315366 150315367 150315368 150315369 150315370 150315371 150315372 150315373 150315374 150315375 150315376 150315377 150315378 150315379 150315380 150315381 150315382 150315383 150315384 150315385 150315386 150315387 150315388 150315389 150315390 150315391 150315392 150315393 150315394 150315395 150315396 150315397 150315398 150315399 150315400 150315401 150315402 150315403 150315404 150315405 150315406 150315407 150315408 150315409 150315410 150315411 150315412 150315413 150315414 150315415 150315416 150315417 150315418 150315419 150315420 150315421 150315422 150315423 150315424 150315425 150315426 150315427 150315428 150315429 150315430 150315431 150315432 150315433 150315434 150315435 150315436 150315437 150315438 150315439 150315440 150315441 150315442 150315443 150315444 150315445 150315446 150315447 150315448 150315449 150315450 150315451 150315452 150315453 150315454 150315455 150315456 150315457 150315458 150315459 150315460 150315461 150315462 150315463 150315464 150315465 150315466 150315467 150315468 150315469 150315470 150315471 150315472 150315473 150315474 150315475 150315476 150315477 150315478 150315479 150315480 150315481 150315482 150315483 150315484 150315485 150315486 150315487 150315488 150315489 150315490 150315491 150315492 150315493 150315494 150315495 150315496 150315497 150315498 150315499 150315500 150315501 150315502 150315503 150315504 150315505 150315506 150315507 150315508 150315509 150315510 150315511 150315512 150315513 150315514 150315515 150315516 150315517 150315518 150315519 150315520 150315521 150315522 150315523 150315524 150315525 150315526 150315527 150315528 150315529 150315530 150315531 150315532 150315533 150315534 150315535 150315536 150315537 150315538 150315539 150315540 150315541 150315542 150315543 150315544 150315545 150315546 150315547 150315548 150315549 150315550 150315551 150315552 150315553 150315554 150315555 150315556 150315557 150315558 150315559 150315560 150315561 150315562 150315563 150315564 150315565 150315566 150315567 150315568 150315569 150315570 150315571 150315572 150315573 150315574 150315575 150315576 150315577 150315578 150315579 150315580 150315581 150315582 150315583 150315584 150315585 150315586 150315587 150315588 150315589 150315590 150315591 150315592 150315593 150315594 150315595 150315596 150315597 150315598 150315599 150315600 150315601 150315602 150315603 150315604 150315605 150315606 150315607 150315608 150315609 150315610 150315611 150315612 150315613 150315614 150315615 150315616 150315617 150315618 150315619 150315620 150315621 150315622 150315623 150315624 150315625 150315626 150315627 150315628 150315629 150315630 150315631 150315632 150315633 150315634 150315635 150315636 150315637 150315638 150315639 150315640 150315641 150315642 150315643 150315644 150315645 150315646 150315647 150315648 150315649 150315650 150315651 150315652 150315653 150315654 150315655 150315656 150315657 150315658 150315659 150315660 150315661 150315662 150315663 150315664 150315665 150315666 150315667 150315668 150315669 150315670 150315671 150315672 150315673 150315674 150315675 150315676 150315677 150315678 150315679 150315680 150315681 150315682 150315683 150315684 150315685 150315686 150315687 150315688 150315689 150315690 150315691 150315692 150315693 150315694 150315695 150315696 150315697 150315698 150315699 150315700 150315701 150315702 150315703 150315704 150315705 150315706 150315707 150315708 150315709 150315710 150315711 150315712 150315713 150315714 150315715 150315716 150315717 150315718 150315719 150315720 150315721 150315722 150315723 150315724 150315759 150315760 150315761 150315762 150315763 150315764 150315765 150315766 150315767 150315768 150315769 150315770 150315771 150315772 150315773 150315774 150315775 150315776 150315777 150315778 150315779 150315780 150315781 150315782 150316001 150316002 150316003 150316004 150316005 150316006 150316007 150316008 150316009 150316010 150316011 150316012 150316013 150316014 150316015 150316016 150316017 150316018 150316019 150316020 150316021 150316022 150316023 150316024 150316025 150316026 150316027 150316028 150316029 150316030 150316031 150316032 150316033 150316034 150316035 150316036 150316037 150316038 150316039 150316040 150316041 150316042 150316043 150316044 150316045 150316046 150316047 150316048 150316049 150316050 150316051 150316052 150316053 150316054 150316055 150316056 150316057 150316058 150316059 150316060 150316061 150316062 150316063 150316064 150316065 150316066 150316067 150316068 150316069 150316070 150316071 150316072 150316073 150316074 150316075 150316076 150316077 150316078 150316079 150316080 150316081 150316082 150316083 150316084 150316085 150316086 150316087 150316088 150316089 150316090 150316091 150316092 150316093 150316094 150316095 150316096 150316097 150316098 150316099 150316100 150316101 150316102 150316103 150316104 150316105 150316106 150316107 150316108 150316109 150316110 150316111 150316112 150316113 150316114 150316115 150316116 150316117 150316118 150316119 150316120 150316121 150316122 150316123 150316124 150316125 150316126 150316127 150316128 150316129 150316130 150316131 150316132 150316133 150316134 150316135 150316136 150316137 150316138 150316139 150316140 150316141 150316142 150316143 150316144 150316145 150316146 150316147 150316148 150316149 150316150 150316151 150316152 150316153 150316154 150316155 150316156 150316157 150316158 150316159 150316160 150316161 150316162 150316163 150316164 150316165 150316166 150316167 150316168 150316169 150316170 150316171 150316172 150316173 150316174 150316175 150316176 150316177 150316178 150316179 150316180 150316181 150316182 150316183 150316184 150316185 150316186 150316187 150316188 150316189 150316190 150316191 150316192 150316193 150316194 150316195 150316196 150316197 150316198 150316199 150316200 150316201 150316202 150316203 150316204 150316205 150316206 150316207 150316208 150316209 150316210 150316211 150316212 150316213 150316214 150316215 150316216 150316217 150316218 150316219 150316220 150316221 150316222 150316223 150316224 150316225 150316226 150316227 150316228 150316229 150316230 150316231 150316232 150316233 150316234 150316235 150316236 150316237 150316238 150316239 150316240 150316241 150316242 150316243 150316244 150316245 150316246 150316247 150316248 150316249 150316250 150316251 150316252 150316253 150316254 150316255 150316256 150316257 150316258 150316259 150316260 150316261 150316262 150316263 150316264 150316265 150316266 150316267 150316268 150316269 150316270 150316271 150316272 150316273 150316274 150316275 150316276 150316277 150316278 150316279 150316280 150316281 150316282 150316283 150316284 150316285 150316286 150316287 150316288 150316289 150316290 150316291 150316292 150316293 150316294 150316295 150316296 150316297 150316298 150316299 150316300 150316301 150316302 150316303 150316304 150316305 150316306 150316307 150316308 150316309 150316310 150316311 150316312 150316313 150316314 150316315 150316316 150316317 150316318 150316319 150316320 150316321 150316322 150316323 150316324 150316325 150316326 150316327 150316328 150316329 150316330 150316331 150316332 150316333 150316334 150316335 150316336 150316337 150316338 150316339 150316340 150316341 150316342 150316343 150316344 150316345 150316346 150316347 150316348 150316349 150316350 150316351 150316352 150316353 150316354 150316355 150316356 150316357 150316358 150316359 150316360 150316361 150316362 150316363 150316364 150316365 150316366 150316367 150316368 150316369 150316370 150316371 150316372 150316373 150316374 150316375 150316376 150316377 150316378 150316379 150316380 150316381 150316382 150316383 150316384 150316385 150316386 150316387 150316388 150316389 150316390 150316391 150316392 150316393 150316394 150316395 150316396 150316397 150316398 150316399 150316400 150316401 150316402 150316403 150316404 150316405 150316406 150316407 150316408 150316409 150316410 150316411 150316412 150316413 150316414 150316415 150316416 150316417 150316418 150316419 150316420 150316421 150316422 150316423 150316424 150316425 150316426 150316427 150316428 150316429 150316430 150316431 150316432 150316433 150316434 150316435 150316436 150316437 150316438 150316439 150316440 150316441 150316442 150316443 150316444 150316445 150316446 150316447 150316448 150316449 150316450 150316451 150316452 150316453 150316454 150316455 150316456 150316457 150316458 150316459 150316460 150316461 150316462 150316463 150316464 150316465 150316466 150316467 150316468 150316469 150316470 150316471 150316472 150316473 150316474 150316475 150316476 150316477 150316478 150316479 150316480 150316481 150316482 150316483 150316484 150316485 150316486 150316487 150316488 150316489 150316490 150316491 150316492 150316493 150316494 150316495 150316496 150316497 150316498 150316499 150316500 150316501 150316502 150316503 150316504 150316505 150316506 150316507 150316508 150316509 150316510 150316511 150316512 150316513 150316514 150316515 150316516 150316517 150316518 150316519 150316520 150316521
150316522 150316523 150316524 150316525 150316526 150316527 150316528 150316529 150316530 150316531 150316532 150316533 150316534 150316535 150316536 150316537 150316538 150316539 150316540 150316541 150316542 150316543 150316544 150316545 150316546 150316547 150316548 150316549 150316550 150316552 150316553 150316554 150316555 150316556 150316557 150316558 150316559 150316560 150316561 150316562 150316563 150316564 150316565 150316566 150316567 150316568 150316569 150316570 150316571 150316572 150316573 150316574 150316575 150316576 150316577 150316578 150316579 150316580 150316581 150316582 150316583 150316584 150316585 150316586 150316587 150316588 150316589 150316590 150316591 150316592 150316593 150316594 150316595 150316596 150316597 150316598 150316599 150316600 150316601 150316602 150316603 150316604 150316605 150316606 150316607 150316608 150316609 150316610 150316611 150316612 150316613 150316614 150316615 150316616 150316617 150316618 150316619 150316620 150316621 150316622 150316623 150316624 150316625 150316626 150316627 150316628 150316629 150316630 150316631 150316632 150316633 150316634 150316635 150316636 150316637 150316638 150316639 150316640 150316641 150316642 150316643 150316644 150316645 150316646 150316647 150316648 150316649 150316650 150316651 150316652 150316653 150316654 150316655 150316656 150316657 150316658 150316659 150316660 150316661 150316662 150316663 150316664 150316665 150316666 150316667 150316668 150316669 150316670 150316671 150316672 150316673 150316674 150316675 150316676 150316677 150316678 150316679 150316680 150316681 150316682 150316683 150316684 150316685 150316686 150316687 150316688 150316689 150316690 150316691 150316692 150316693 150316694 150316695 150316696 150316697 150316698 150316699 150316700 150316701 150316702 150316703 150316704 150316705 150316706 150316707 150316708 150316709 150316710 150316711 150316712 150316713 150316714 150316715 150316716 150316717 150316718 150316719 150316720 150316721 150316722 150316723 150316724 150316725 150316726 150316727 150316728 150316729 150316730 150316731 150316732 150316733 150316734 150316735 150316736 150316737 150316738 150316739 150316740 150316741 150316742 150316743 150316744 150316745 150316746 150316747 150316748 150316749 150316750 150316751 150316752 150316753 150316754 150316755 150316756 150316757 150316758 150316759 150316760 150316761 150316762 150316763 150316764 150316765 150316766 150316767 150316768 150316769 150316770 150316771 150316772 150316773 150316774 150316775 150316776 150316777 150316778 150316779 150316780 150316781 150316782 150316783 150316784 150316785 150316786 150316787 150316788 150316789 150316790 150316791 150316792 150316793 150316794 150316795 150316796 150316797 150316798 150316799 150316800 150316801 150316802 150316803 150316804 150316805 150316806 150316807 150316808 150316809 150316810 150316811 150316812 150316813 150316814 150316815 150316816 150316817 150316818 150316819 150316820 150316821 150316822 150316823 150316824 150316825 150316826 150316827 150316828 150316829 150316830 150316831 150316832 150316833 150316834 150316835 150316836 150316837 150316838 150316839 150316840 150316841 150316842 150316843 150316844 150316845 150316846 150316847 150316848 150316849 150316850 150316851 150316852 150316853 150316854 150316855 150316856 150316857 150316858 150316859 150316860 150316861 150316862 150316863 150316864 150316865 150316866 150316867 150316868 150316869 150316870 150316871 150316872 150316873 150316874 150316875 150316876 150316877 150316878 150316879 150316880 150316881 150316882 150316883 150316884 150316885 150316886 150316887 150316888 150316889 150316890 150316891 150316892 150316893 150316894 150316895 150316896 150316897 150316898 150316899 150316900 150316901 150316902 150316903 150316904 150316905 150316906 150316907 150316908 150316909 150316910 150316911 150316912 150316913 150316914 150316915 150316916 150316917 150316918 150316919 150316920 150316921 150316922 150316923 150316924 150316925 150316926 150316927 150316928 150316929 150316930 150316931 150316932 150316933 150316934 150316935 150316936 150316937 150316938 150316939 150316940 150316941 150316942 150316943 150316944 150316945 150316946 150316947 150316948 150316949 150316950 150316951 150316952 150316953 150316954 150316955 150316956 150316957 150316958 150316959 150316960 150316961 150316962 150316963 150316964 150316965 150316966 150316967 150316968 150316969 150316970 150316971 150316972 150316973 150316974 150316975 150316976 150316977 150316978 150316979 150316980 150316981 150316982 150316983 150316984 150316985 150316986 150316987 150316988 150316989 150316990 150316991 150316992 150415003 150415004 150415005 150415006 150415007 150415008 150415009 150415010 150415011 150415012 150415013 150415014 150415015 150415016 150415017 150415018 150415019 150415020 150415021 150415022 150415023 150415024 150415025 150415026 150415027 150415028 150415029 150415030 150415031 150415032 150415033 150415034 150415035 150415036 150415037 150415038 150415039 150415040 150415041 150415042 150415043 150415044 150415045 150415046 150415047 150415048 150415049 150415050 150415051 150415052 150415053 150415054 150415055 150415056 150415057 150415058 150415059 150415060 150415061 150415062 150415063 150415064 150415065 150415066 150415067 150415068 150415069 150415070 150415071 150415072 150415073 150415074 150415075 150415076 150415077 150415078 150415079 150415080 150415081 150415082 150415083 150415084 150415085 150415086 150415087 150415088 150415089 150415090 150415091 150415092 150415093 150415094 150415095 150415096 150415097 150415098 150415099 150415100 150415101 150415102 150415103 150415104 150415105 150415106 150415107 150415108 150415109 150415110 150415111 150415112 150415113 150415114 150415115 150415116 150415117 150415118 150415119 150415120 150415121 150415122 150415123 150415124 150415125 150415126 150415127 150415128 150415129 150415130 150415131 150415132 150415133 150415134 150415135 150415136 150415137 150415138 150415139 150415140 150415141 150415142 150415143 150415144 150415145 150415146 150415147 150415148 150415149 150415150 150415151 150415152 150415153 150415154 150415155 150415156 150415157 150415158 150415159 150415160 150415161 150415162 150415163 150415164 150415165 150415166 150415167 150415168 150415169 150415170 150415171 150415172 150415173 150415174 150415175 150415176 150415177 150415178 150415179 150415180 150415181 150415182 150415183 150415184 150415185 150415186 150415187 150415188 150415189 150415190 150415191 150415192 150415193 150415194 150415195 150415196 150415197 150415198 150415199 150415200 150415201 150415202 150415203 150415204 150415205 150415206 150415207 150415208 150415209 150415210 150415211 150415212 150415213 150415214 150415215 150415216 150415217 150415218 150415219 150415220 150415221 150415222 150415223 150415224 150415225 150415226 150415227 150415228 150415229 150415230 150415231 150415232 150415233 150415234 150415235 150415236 150415237 150415238 150415239 150415240 150415241 150415242 150415243 150415244 150415245 150415246 150415247 150415248 150415249 150415250 150415251 150415252 150415253 150415254 150415255 150415256 150415257 150415258 150415259 150415260 150415261 150415262 150415263 150415264 150415265 150415266 150415267 150415268 150415269 150415270 150415271 150415272 150415273 150415274 150415275 150415276 150415277 150415278 150415279 150415280 150415281 150415282 150415283 150415284 150415285 150415286 150415287 150415288 150415289 150415290 150415291 150415292 150415293 150415294 150415295 150415296 150415297 150415298 150415299 150415300 150415301 150415302 150415303 150415304 150415305 150415306 150415307 150415308 150415309 150415310 150415311 150415312 150415313 150415314 150415315 150415316 150415317 150415318 150415319 150415320 150415321 150415322 150415323 150415324 150415325 150415326 150415327 150415328 150415329 150415330 150415331 150415332 150415333 150415334 150415335 150415336 150415337 150415338 150415339 150415340 150415341 150415342 150415343 150415344 150415345 150415346 150415347 150415348 150415349 150415350 150415351 150415352 150415353 150415354 150415355 150415356 150415357 150415358 150415359 150415360 150415361 150415362 150415363 150415364 150415365 150415366 150415367 150415368 150415369 150415370 150415371 150415372 150415373 150415374 150415375 150415376 150415377 150415378 150415379 150415380 150415381 150415382 150415383 150415384 150415385 150415386 150415387 150415388 150415389 150415390 150415391 150415392 150415393 150415394 150415395 150415396 150415397 150415398 150415399 150415400 150415401 150415402 150415403 150415404 150415405 150415406 150415407 150415408 150415409 150415410 150415411 150415412 150415413 150415414 150415415 150415416 150415417 150415418 150415419 150415420 150415421 150415422 150415423 150415424 150415425 150415426 150415427 150415428 150415429 150415430 150415431 150415432 150415433 150415434 150415435 150415436 150415437 150415438 150415439 150415440 150415441 150415442 150415443 150415444 150415445 150415446 150415447 150415448 150415449 150415450 150415451 150415452 150415453 150415454 150415455 150415456 150415457 150415458 150415459 150415460 150415461 150415462 150415463 150415464 150415465 150415466 150415467 150415468 150415469 150415470 150415471 150415472 150415473 150415474 150415475 150415476 150415477 150415478 150415479 150415480 150415481 150415482 150415483 150415484 150415485 150415486 150415487 150415488 150415489 150415490 150415491 150415492 150415493 150415494 150415495 150415496 150415497 150415498 150415499 150415500 150415501 150415502 150415503 150415504 150415505 150415506 150415507 150415508 150415509 150415510 150415511 150415512 150415513 150415514 150415515 150415516 150415517 150415518 150415519 150415520 150415521 150415522 150415523 150415524 150415525 150415526 150415527 150415528 150415529 150415530 150415531 150415532 150415533 150415534 150415535 150415536 150415537 150415538 150415539 150415540 150415541 150415542 150415543 150415544 150415545 150415546 150415547 150415548 150415549 150415550 150415551 150415552 150415553 150415554 150415555 150415556 150415557 150415558 150415559 150415560 150415561 150415562 150415563 150415564 150415565 150415566 150415567 150415568 150415569 150415570 150415571 150415572 150415580 150415581 150415582 150415583 150415584 150415585 150415586 150415587 150415588 150415589 150415590 150415591 150415592 150415593 150415594 150415595 150415596 150415597 150415598 150415599 150415600 150415601 150415602 150415603 150415604 150415605 150415606 150415607 150415608 150415609 150415610 150415611 150415612 150415613 150415614 150415615 150415616 150415617 150415618 150415619 150415620 150415621 150415622 150415623 150415624 150415625 150415626 150415627 150415628 150415629 150415630 150415631 150415632 150415633 150415634 150415635 150415636 150415637 150415638 150415639 150415640 150415641 150415642 150415643 150415644 150415645 150415646 150415647 150415648 150415649 150415650 150415651 150415652 150415653 150415654 150415655 150415656 150415657 150415658 150415659 150415660 150415661 150415662 150415663 150415664 150415665 150415666 150415667 150415668 150415669 150415670 150415671 150415672 150415673 150415674 150415675 150415676 150415677 150415678 150415679 150415680 150415681 150415682 150415683 150415684 150415685 150415686 150415687 150415688 150415689 150415690 150415691 150415692 150415693 150415694 150415695 150415696 150415697 150415698 150415699 150415700 150415701 150415702 150415703 150415704 150415705 150415706 150415707 150415708 150415709 150415710 150415711 150415712 150415713 150415714 150415715 150415716 150415717 150415718 150415719 150415720 150415721 150415722 150415723 150415724 150415725 150415726 150415727 150415728 150415729 150415730 150415731 150415732 150415733 150415734 150415735 150415736 150415737 150415738 150415739 150415740 150415741 150415742 150415743 150415744 150415745 150415746 150415747 150415748 150415749 150415750 150415751 150415752 150415753 150415754 150415755 150415756 150415757 150415758 150415759 150415760 150415761 150415762 150415763 150415764 150415765 150415766 150415767 150415768 150415769 150415770 150415771 150415772 150415773 150415774 150415775 150415776 150415777 150415778 150415779 150415780 150415781 150415782 150415783 150415784 150415785 150415786 150415787 150415788 150415789 150415790 150415791 150415792 150415793 150415794 150415795 150415796 150415797 150415798 150415799 150415800 150415801 150415802 150415803 150415804 150415805 150415806 150415807 150415808 150415809 150415810 150415811 150415812 150415813 150415814 150415815 150415816 150415817 150415818 150415819 150415820 150415821 150415822 150415823 150415824 150415825 150415826 150415827 150415828 150415829 150415830 150415831 150415832 150415833 150415834 150415835 150415836 150415837 150415838 150415839 150415840 150415841 150415842 150415843 150415844 150415845 150415846 150415847 150415848 150415849 150415850 150415851 150415852 150415853 150415854 150415855 150415856 150415857 150415858 150415859 150415860 150415861 150415862 150415863 150415864 150415865 150415866 150415867 150415868 150415869 150415870 150415871 150415872 150415873 150415874 150415875 150415876 150415877 150415878 150415879 150415880 150415881 150415882 150415883 150415884 150415885 150415886 150415887 150415888 150415889 150415890 150415891 150415892 150415893 150415894 150415895 150415896 150415897 150415898 150415899 150415900 150415901 150415902 150415903 150415904 150415905 150415906 150415907 150415908 150415909 150415910 150415911 150415912 150415913 150415914 150415915 150415916 150415918 150415919 150415920 150415921 150415922 150415923 150415924 150415925 150415926 150415927 150415928 150415929 150415930 150415931 150415932 150415933 150415934 150415935 150415936 150415937 150415938 150415939 150415940 150415941 150415942 150415943 150415944 150415945 150415946 150415947 150415948 150415949 150415950 150415951 150415952 150415953 150415954 150415955 150415956 150415957 150415958 150415959 150415960 150415961 150415962 150415963 150415964 150415965 150415966 150415967 150415968 150415969 150415970 150415971 150415972 150415973 150415974 150415975 150415976 150415977 150415978 150415979 150415980 150415981 150415982 150415983 150415984 150415985 150415986 150415987 150415988 150415989 150415990 150415991 150415992 150415993 150415994 150415995 150415996 150415997 150415998 150415999 150416000 150416001 150416002 150416003 150416004 150416005 150416006 150416007 150416008 150416009 150416010 150416011 150416012 150416013 150416014 150416015 150416016 150416017 150416018 150416019 150416020 150416021 150416022 150416023 150416024 150416025 150416026 150416027 150416028 150416029 150416030 150416031 150416032 150416033 150416034 150416035 150416036 150416037 150416038 150416039 150416040 150416041 150416042 150416043 150416044 150416045 150416046 150416047 150416048 150416049 150416050 150416051 150416052 150416053 150416054 150416055 150416056 150416057 150416058 150416059 150416060 150416061 150416062 150416063 150416064 150416065 150416066 150416067 150416068 150416069 150416070 150416071 150416072 150416073 150416074 150416075 150416076 150416077 150416078 150416079 150416080 150416081 150416082 150416083 150416084 150416085 150416086 150416087 150416088 150416089 150416090 150416091 150416092 150416093 150416094 150416095 150416096 150416097 150416098 150416099 150416100 150416101 150416102 150416103 150416104 150416105 150416106 150416107 150416108 150416109 150416110 150416111 150416112 150416113 150416114 150416115 150416116 150416117 150416118 150416119 150416120 150416121 150416122 150416123 150416124 150416125 150416126 150416127 150416128 150416129 150416130 150416131 150416132 150416133 150416134 150416135 150416136 150416137 150416138 150416139 150416140 150416141 150416142 150416143 150416144 150416145 150416146 150416147 150416148 150416149 150416150 150416151 150416152 150416153 150416154 150416155 150416156 150416157 150416158 150416159 150416160 150416161 150416162 150416163 150416164 150416165 150416166 150416167 150416168 150416169 150416170 150416171 150416172 150416173 150416174 150416175 150416176 150416177 150416178 150416179 150416180 150416181 150416182 150416183 150416185 150416186 150416187 150416188 150416189 150416190 150416200 150416201 150416202 150416203 150416204 150416205 150416206 150416207 150416208 150416209 150416210 150416211 150416212 150416213 150416214 150416215 150416216 150416217 150416218 150416219 150416220 150416221 150416222 150416223 150416224 150416225 150515001 150515002 150515003 150515004 150515005 150515006 150515007 150515008 150515009 150515010 150515011 150515012 150515013 150515014 150515015 150515016 150515017 150515018 150515019 150515020 150515021 150515022 150515023 150515024 150515025 150515026 150515027 150515028 150515029 150515030 150515031 150515032 150515033 150515034 150515035 150515036 150515037 150515038 150515039 150515040 150515041 150515042 150515071 150515072 150515073 150515074 150515075 150515076 150515077 150515078 150515079 150515080 150515081 150515082 150515083 150515084 150515085 150515086 150515087 150515088 150515089 150515090 150515091 150515092 150515093 150515094 150515095 150515096 150515097 150515098 150515099 150515100 150515101 150515102 150515103 150515104 150515105 150515106 150515107 150515108 150515109 150515110 150515111 150515112 150515113 150515114 150515115 150515116 150515117 150515118 150515119 150515120 150515121 150515122 150515123 150515124 150515125 150515126 150515127 150515128 150515129 150515130 150515131 150515132 150515133 150515134 150515135 150515136 150515137 150515138 150515139 150515140 150515141 150515142 150515143 150515144 150515145 150515146 150515147 150515148 150515149 150515150 150515151 150515152 150515153 150515154 150515155 150515156 150515157 150515158 150515159 150515160 150515161 150515162 150515163 150515164 150515165 150515166 150515167 150515168 150515169 150515170 150515171 150515172 150515173 150515174 150515175 150515176 150515177 150515178 150515179 150515180 150515181 150515182 150515183 150515184 150515185 150515186 150515187 150515188 150515189 150515190 150515191 150515192 150515193 150515194 150515195 150515196 150515197 150515198 150515199 150515200 150515201 150515202 150515203 150515204 150515205 150515206 150515207 150515208 150515209 150515210 150515211 150515212 150515213 150515214 150515215 150515216 150515217 150515218 150515219 150515220 150515221 150515222 150515223 150515224 150515225 150515226 150515227 150515228 150515229 150515230 150515231 150515232 150515233 150515234 150515235 150515236 150515237 150515238 150515239 150515240 150515241 150515242 150515243 150515244 150515245 150515246 150515247 150515248 150515249 150515250 150515251 150515252 150515253 150515254 150515255 150515256 150515257 150515258 150515259 150515260 150515261 150515262 150515263 150515264 150515265 150515266 150515267 150515268 150515269 150515270 150515271 150515272 150515273 150515274 150515275 150515276 150515277 150515278 150515279 150515280 150515281 150515282 150515283 150515284 150515285 150515286 150515287 150515288 150515289 150515290 150515291 150515292 150515293 150515294 150515295 150515296 150515297 150515298 150515299 150515300 150515301 150515302 150515303 150515304 150515305 150515306 150515307 150515308 150515309 150515310 150515311 150515312 150515313 150515314 150515315 150515316 150515317 150515318 150515319 150515320 150515321 150515322 150515323 150515324 150515325 150515326 150515327 150515328 150515329 150515330 150515331 150515332 150515333 150515334 150515335 150515336 150515337 150515338 150515339 150515340 150515341 150515342 150515343 150515344 150515345 150515346 150515347 150515348 150515349 150515350 150515351 150515352 150515353 150515354 150515355 150515356 150515357 150515358 150515359 150515360 150515361 150515362 150515363 150515364 150515365 150515366 150515367 150515368 150515369 150515370 150515371 150515372 150515373 150515374 150515375 150515376 150515377 150515378 150515379 150515380 150515381 150515382 150515383 150515384 150515385 150515386 150515387 150515388 150515389 150515390 150515391 150515392 150515393 150515394 150515395 150515396 150515397 150515398 150515399 150515400 150515401 150515402 150515403 150515404 150515405 150515406 150515407 150515408 150515409 150515410 150515411 150515412 150515413 150515414 150515415 150515416 150515417 150515418 150515419 150515420 150515421 150515422 150515423 150515424 150515425 150515426 150515427 150515428 150515429 150515430 150515431 150515432 150515433 150515434 150515435 150515436 150515437 150515438 150515439 150515440 150515441 150515442 150515443 150515444 150515445 150515446 150515447 150515448 150515449 150515450 150515451 150515452 150515453 150515454 150515455 150515456 150515457 150515458 150515459 150515460 150515461 150515462 150515463 150515464 150515465 150515466 150515467 150515468 150515469 150515470 150515471 150515472 150515473 150515474 150515475 150515476 150515477 150515478 150515479 150515480 150515481 150515482 150515483 150515484 150515485 150515486 150515487 150515488 150515489 150515490 150515491 150515492 150515493 150515494 150515495 150515496 150515497 150515498 150515499 150515500 150515501 150515502 150515503 150515504 150515505 150515506 150515507 150515508 150515509 150515510 150515511 150515512 150515513 150515514 150515515 150515516 150515517 150515518 150515519 150515520 150515521 150515522 150515523 150515524 150515525 150515526 150515527 150515528 150515529 150515530 150515531 150515532 150515533 150515534 150515535 150515536 150515537 150515538 150515539 150515540 150515543 150515544 150515545 150515546 150515547 150515548 150515549 150515550 150515551 150515552 150515553 150515554 150515555 150515556 150515557 150515558 150515559 150515560 150515561 150515562 150515563 150515564 150515565 150515566 150515567 150515568 150515569 150515570 150515571 150515572 150515573 150515574 150515575 150515576 150515577 150515578 150515579 150515580 150515581 150515582 150515583 150515584 150515585 150515586 150515587 150515588 150515589 150515590 150515591 150515592 150515593 150515594 150515595 150515596 150515597 150515598 150515599 150515600 150515601 150515602 150515603 150515604 150515605 150515606 150515607 150515608 150515609 150515610 150515611 150515612 150515613 150515614 150515615 150515616 150515617 150515618 150515619 150515620 150515621 150515622 150515623 150515634 150515635 150515636 150515637 150515638 150515639 150515640 150515641 150515642 150515643 150515644 150515645 150515646 150515647 150516001 150516002 150516003 150516004 150516005 150516006 150516007 150516008 150516009 150516010 150516011 150516012 150516013 150516014 150516015 150516016 150516017 150516018 150516019 150516020 150516021 150516022 150516023 150516024 150516025 150516026 150516027 150516028 150516029 150516030 150516031 150516032 150516033 150516034 150516035 150516036 150516037 150516038 150516039 150516040 150516041 150516042 150516043 150516044 150516045 150516046 150516047 150516048 150516049 150516050 150516051 150516052 150516053 150516054 150516055 150516056 150516057 150516058 150516059 150516060 150516061 150516062 150516063 150516064 150516065 150516066 150516067 150516068 150516069 150516070 150516071 150516072 150516073 150516074 150516075 150516076 150516077 150516078 150516079 150516080 150516081 150516082 150516083 150516084 150516085 150516086 150516087 150516088 150516089 150516090 150516091 150516092 150516093 150516094 150516095 150516096 150516097 150516098 150516099 150516100 150516101 150516102 150516103 150516104 150516105 150516106 150516107 150516108 150516109 150516110 150516111 150516112 150516113 150516114 150516115 150516116 150516117 150516118 150516119 150516120 150516121 150516122 150516123 150516124 150516125 150516126 150516127 150516128 150516129 150516130 150516131 150516132 150516133 150516134 150516135 150516136 150516137 150516138 150516139 150516141 150516142 150516143 150516144 150516145 150516146 150516147 150516148 150516149 150516150 150516151 150516152 150516153 150516154 150516155 150516156 150516157 150516158 150516163 150516164 150516165 150516166 150516167 150516168 150516169 150516170 150516171 150516172 150516173 150516174 150516175 150516176 150516177 150516178 150516179 150516180 150516181 150516182 150516183 150516184 150516185 150516186 150516187 150516188 150516189 150516190 150516191 150516192 150516193 150516194 150516195 150516196 150516197 150516198 150516199 150516200 150516201 150516202 150516203 150516204 150516205 150516206 150516207 150516208 150516209 150516210 150516211 150516212 150516213 150516214 150516215 150516216 150516217 150516218 150516219 150516220 150516221 150516222 150516223 150516224 150516225 150516226 150516227 150516228 150516229 150516230 150516231 150516232 150516233 150516234 150516235 150516236 150516237 150516238 150516239 150516240 150516241 150516242 150516243 150516244 150516245 150516246 150516247 150516248 150516249 150516250 150516251 150516252 150516253 150516254 150516255 150516256 150516257 150516258 150516259 150516260 150516261 150516262 150516263 150516264 150516265 150516266 150516267 150516268 150516269 150516270 150516271 150516272 150516273 150516274 150516275 150516276 150516277 150516278 150516279 150516280 150516282 150516283 150516284 150516285 150516286 150516287 150516288 150516289 150516290 150516291 150516292 150516293 150516294 150516295 150516296 150516297 150516298 150516299 150516300 150516301 150516302 150516303 150516304 150516305 150516306 150516307 150516308 150516309 150516310 150516311 150516312 150516313 150516314 150516315 150516316 150516317 150516318 150516319 150516320 150516321 150516322 150516323 150516324 150516325 150516326 150516327 150516328 150516329 150516330 150516331 150516332 150516333 150516334 150516335 150516336 150516337 150516338 150516339 150516340 150516341 150516342 150516343 150516344 150516345 150516346 150516347 150516348 150516349 150516350 150516351 150516352 150516353 150516354 150516355 150516356 150516357 150516358 150516359 150516360 150516361 150516362 150516363 150516364 150516365 150516366 150516367 150516368 150516369 150516370 150516371 150516372 150516373 150516374 150516375 150516376 150516377 150516378 150516379 150516380 150516381 150516382 150516383 150516384 150516385 150516386 150516387 150516388 150516389 150516390 150516391 150516392 150516393 150516394 150516395 150516396 150516397 150516398 150516399 150516400 150516401 150516402 150516403 150516404 150516405 150516406 150516407 150516408 150516409 150516410 150516411 150516412 150516413 150516414 150516415 150516416 150516417 150516418 150516419 150516420 150516421 150516422 150516423 150516424 150516425 150516426 150516427 150516428 150516429 150516430 150516431 150516432 150516433 150516434 150516435 150516436 150516437 150516438 150516439 150516440 150516441 150516442 150516443 150516444 150516445 150516446 150516447 150516448 150516449 150516450 150516451 150516452 150516453 150516454 150516455 150516456 150516457 150516458 150516459 150516460 150516461 150516462 150516463 150516464 150516465 150516466 150516467 150516468 150516469 150516470 150516471 150516472 150516473 150516474 150516475 150516476 150516477 150516478 150516479 150516480 150516481 150516482 150516483 150516484 150516485 150516486 150615001 150615002 150615003 150615004 150615005 150615006 150615007 150615008 150615009 150615010 150615011 150615012 150615013 150615014 150615015 150615016 150615017 150615018 150615019 150615020 150615021 150615022 150615023 150615024 150615025 150615026 150615027 |
Recalling Firm/ Manufacturer |
Stryker Medical Division of Stryker Corporation 3800 E Centre Ave Portage MI 49002-5826
|
For Additional Information Contact |
Renata Sila 269-324-6609
|
Manufacturer Reason for Recall |
CPU board failures cause fowler (backrest) electronic controls to stop working.
|
FDA Determined Cause 2 |
Component change control |
Action |
Stryker Medical initiated a voluntary recall of the S3 MedSurg Bed, Model 3005, with notices sent on February 2, 2016 via certified mail. Stryker medical identified that Fowler does not raise up to desired angle; and the Fowler does not raise electronically, but manual function remains operational through the manual CPR release. Actions Needed for Customers:
1. Locate the bed(s) listed by serial number on the attached business reply form.
2. A Stryker ProCare Representative will contact your facility to schedule a time for repair.
3. Fax (269)488-8691 or email productfieldaction@stryker.com the enclosed business reply form to confirm receipt of this notification.
4. If you have further distributed the beds listed in this letter, please forward a copy of this notice to the new users and advise us of their new location. If you have any urgent questions or concerns, please, contact Stryker Customer Service at (1-800-327-0770). Normal business hours are Monday-Friday 8 a.m-5 p.m. (ET). |
Quantity in Commerce |
8490 |
Distribution |
Distribution US nationwide, Canada, and Mexico. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|