• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall lnvacare Precise Rx Pediatric Flowmeter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall lnvacare Precise Rx Pediatric Flowmeter see related information
Date Initiated by Firm June 21, 2016
Date Posted August 08, 2016
Recall Status1 Terminated 3 on February 07, 2017
Recall Number Z-2365-2016
Recall Event ID 74581
510(K)Number K914488  
Product Classification Conserver, oxygen - Product Code NFB
Product Invacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16 AW. Packed one per carton.
The lnvacare Precise Rx" Pediatric Flowmeter adapts to lnvacare Concentrators for low-flow usage.The gauge has a scale printed on the front side and the back side. The front scale is the primary source for monitoring oxygen flow. The rear scale is an aide for monitoring oxygen flow.
Code Information Serial numbers for Model IRCPF16AW:   16CF024370, 16CF024373, 16CF024376, 16CF024379, 16CF024382,16CF024368,16CF024369, 16CF024371,16CF024372, 16CF024374,16CF024375, 16CF024377,16CF024378, 16CF024380,16CF024381, 16CF037743.   
Recalling Firm/
Manufacturer
Invacare Corporation
1 Invacare Way
Elyria OH 44035-4190
For Additional Information Contact
877-413-6008
Manufacturer Reason
for Recall
The rear gradation scale was screen printed upside down on the flowmeter gauge.
FDA Determined
Cause 2
Process control
Action lnvacare Corporation sent an Urgent Medical Device Recall letter dated June 27, 2016, to all affected consignees with return receipt to notify them of the recall. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees with questions were instructed to call 877-413-6008.
Quantity in Commerce 395 units
Distribution Worldwide Distribution - US to CA, CO, CT, FL, GA, IA, ID, IL,IN,KS,KY,LA,MA,MD,MI,MS,MT,NC,NE,NJ,NM,NY,OH,PA,SC, TN,TX,UT,VA,WA and Internationally to Canada, Germany and New Zealand.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = NFB and Original Applicant = INVACARE CORP.
-
-