Class 2 Device Recall Mindray |
|
Date Initiated by Firm |
June 15, 2016 |
Create Date |
August 30, 2016 |
Recall Status1 |
Terminated 3 on March 28, 2017 |
Recall Number |
Z-2689-2016 |
Recall Event ID |
74758 |
510(K)Number |
K151954
|
Product Classification |
Gas-machine, anesthesia - Product Code BSZ
|
Product |
A7 Anesthesia Delivery System, a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. Part numbers 0632F-PA0000X (US domestic) and 0632B-00014 (international) |
Code Information |
P/N 0632F-PA0000X - (Domestic) and 0632B-PA00014 - (International) Serial numbers FR-5B000341, FR-5B000342, FR-5B000343, FR-5B000348, FR-5B000349, FR-51000148, FR-51000149, FR-51000150, FR-51000151, FR-51000152, FR-51000153, FR-51000154, FR-51000159 , FR-51000161, FR-51000162, FR-51000163, FR-51000164, FR-51000165, FR-51000166, FR-51000167, FR-51000168, FR-51000170, FR-51000171, FR-52000172, FR-52000173, FR-52000175, FR-52000176, FR-52000177, FR-52000178, FR-52000180, FR-52000181, FR-52000182, FR-52000183, FR-52000184, FR-52000185, FR-52000186, FR-53000187, FR-53000188, FR-53000189, FR-53000190, FR-53000191, FR-53000192, FR-53000193, FR-53000194, FR-53000197, FR-53000198, FR-53000199, FR-53000200, FR-53000203, FR-53000205, FR-53000206, FR-53000207, FR-53000209, FR-53000211, FR-53000212, FR-53000213, FR-53000214, FR-53000215, FR-53000216, FR-53000217, FR-54000218, FR-54000220, FR-54000221, FR-54000223, FR-54000224, FR-54000228, FR-54000229, FR-54000231, FR-54000233, FR-54000234, FR-54000235, FR-54000236, FR-54000237, FR-54000238, FR-54000239, FR-54000240, FR-54000241, FR-54000242, FR-54000243, FR-54000244, FR-56000245, FR-56000246, FR-56000247, FR-56000248, FR-56000249, FR-56000250, FR-56000251, FR-56000252, FR-56000253, FR-56000254, FR-56000255, FR-56000256, FR-56000257, FR-56000258, FR-56000259, FR-56000260, FR-56000261, FR-56000262, FR-56000265, FR-56000266, FR-56000267, FR-56000268, FR-57000269, FR-57000270, FR-57000271, FR-57000272, FR-57000274, FR-57000275, FR-57000276, FR-57000277, FR-57000278, FR-57000279, FR-57000280, FR-57000281, FR-57000282, FR-57000283, FR-57000284, FR-57000285, FR-57000286, FR-57000287, FR-57000288, FR-58000289, FR-58000290, FR-58000291, FR-58000292, FR-58000293, FR-58000294, FR-58000295, FR-58000296, FR-58000297, FR-58000298, FR-58000299, FR-58000300, FR-58000301, FR-58000302, FR-58000303, FR-58000304, FR-58000305, FR-58000306, FR-58000307, FR-58000308, FR-58000312, FR-58000313, FR-58000314, FR-58000315, FR-58000316, FR-58000317, FR-58000318, FR-58000319, FR-58000320, FR-58000321, FR-58000322, FR-58000323, FR-58000325, FR-59000330, FR-59000331, FR-59000333, FR-59000334, FR-59000335, FR-59000336, FR-59000338, and FR-59000339. |
Recalling Firm/ Manufacturer |
Mindray DS USA, Inc. dba Mindray North America 800 Macarthur Blvd Mahwah NJ 07430-2001
|
For Additional Information Contact |
Ms. Diane Arpino 201-995-8407
|
Manufacturer Reason for Recall |
Potential for a leak to occur on the back-up O2 and air e-size cylinder yokes on the A7 Anesthesia Delivery System.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Mindray sent via certified mail with return receipt a recall letter dated June 15, 2016 to their affected customers. Mindray will replace the gasket on all e-size cylinder yokes on the affected A7 systems. Customer was instructed to contact their local Mindray Service Representative to arrange for this replacement. Customers can continue to use their A7 system while awaiting the replacement of the gasket(s). Customers can contact Ms. Diane Arpino, Director, Quality Operations and Regulatory Affairs via email to: d.arpino@mindray.com or via telephone to: (201)995-8407 |
Quantity in Commerce |
167 units (165 units - US) and (2 units - International) |
Distribution |
US Nationwide Distribution to AL, MA, MD, MN, MO, NE, NJ, OK, OR, PA and VA; and Canada |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = BSZ and Original Applicant = SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
|
|
|
|