• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Orthopedic Salvage System (OSS) 9cm

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Orthopedic Salvage System (OSS) 9cm see related information
Date Initiated by Firm December 16, 2016
Date Posted January 28, 2017
Recall Status1 Terminated 3 on July 21, 2017
Recall Number Z-1114-2017
Recall Event ID 76082
510(K)Number K083827  
Product Classification Prosthesis, knee, femorotibial, constrained, cemented, metal/polymer - Product Code KRO
Product Orthopedic Salvage System (OSS) 9cm
prosthesis, knee, femorotibial, constrained, cemented, metal/polymer
Knee System Tibial Bodies
Code Information Part # Lot # UDI  CP113462 646200 (01) 0 0880304 51008 1 (17) 250928 (10) 646200 CP113462 690040 (01) 0 0880304 51008 1 (17) 260714 (10) 690040 CP113462 003620 (01) 0 0880304 51008 1 (17) 260324 (10) 003620 CP113462 498380 (01) 0 0880304 51008 1 (17) 250929 (10) 498380 CP113462 690050 (01) 0 0880304 51008 1 (17) 260622 (10) 690050 CP113462 539760 (01) 0 0880304 51008 1 (17) 251007 (10) 539760 CP113462 409190 (01) 0 0880304 51008 1 (17) 260417 (10) 409190 CP113462 799300 (01) 0 0880304 51008 1 (17) 260802 (10) 799300 CP113462 460990 (01) 0 0880304 51008 1 (17) 260416 (10) 460990
Recalling Firm/
Manufacturer
Zimmer Biomet, Inc.
56 E Bell Dr
Warsaw IN 46582-6989
For Additional Information Contact Customer Service
574-371-3071
Manufacturer Reason
for Recall
labeling for specific lots of the Standard OSS Modular Proximal Tibia incorrectly identifies the components as AVL Modular Proximal Tibia. The mislabeling could have the potential risks: A delay in surgery less than 30 minutes. Patient undergoes a revision with increased risk of infection.
FDA Determined
Cause 2
Under Investigation by firm
Action On 12/16/2016 and 12/21/2016, Zimmer Biomet contacted distributors via telephone to notify them of the issue and requested that the product be placed in a quarantine location to prevent use. On 12/22/2016 URGENT MEDICAL DEVICE RECALL REMOVAL- LOT SPECIFIC notifications were sent to the affected distributors and Risk Managers / Surgeons via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Distributor responsibilities include: 1. Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1  Inventory Return Certification Form a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com within three (3) days. 4. Immediately return all affected product from your distributorship and affected hospitals and surgeons within your territory along with a completed Attachment 1  Inventory Return Certification Form to Zimmer Biomet. a. Request a Return Authorization Number via email to rgarequest@zimmerbiomet.com. Be sure to specify RECALL as the RGA type. b. For each return, send a copy of Attachment 1 to corporatequality.postmarket@zimmerbiomet.com. c. Include a hardcopy of Attachment 1 with your shipment for immediate processing. 5. Note that any hospitals and/or surgeons that received direct shipments of this product from Zimmer Biomet will be sent a copy of the Risk Manager and Surgeon Recall Notice directly. It is important that you review the list of hospitals and/or surgeons included with the email notification sent to your facility to identify additional accounts Zimmer Biomet has not notified. Using the Additional Accounts Form provided with the email notice sent to your facility, return contact information for a
Quantity in Commerce 16
Distribution WI NJ CA MD WA LA MA TX
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = KRO and Original Applicant = BIOMET MANUFACTURING CORP.
-
-