  Class 2 Device Recall IMMULITE 2000 Progesterone see related information
Date Initiated by Firm January 04, 2017
Create Date March 09, 2017
Recall Status1 Terminated 3 on July 18, 2019
Recall Number Z-1421-2017
Recall Event ID 76212
510(K)Number K932955  
Product Classification Radioimmunoassay, progesterone - Product Code JLS
Product IMMULITE 2000 Progesterone; Catalog Number: L2KPW2
L2KPW6; Siemens Material Number: 10381181, 10381170

Product Usage:
For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers - for the quantitative measurement of progesterone in serum as an aid in the diagnosis and treatment of disorders of the ovaries or placenta.
Code Information SMN: 10381181: L2KPW2 kit lot 453 L2KPW2 kit lot 454 L2KPW2 kit lot 455 L2KPW2 kit lot 456 L2KPW2 kit lot 457 L2KPW2 kit lot 458 L2KPW2 kit lot 459 L2KPW2 kit lot 460 L2KPW2 kit lot 462 L2KPW2 kit lot 463 L2KPW2 kit lot 464 L2KPW2 kit lot 465 L2KPW2 kit lot 467 L2KPW2 kit lot 468 L2KPW2 kit lot 470 L2KPW2 kit lot 471 L2KPW2 kit lot 472 L2KPW2 kit lot 474; SMN: 10381170: L2KPW6 kit lot 453 L2KPW6 kit lot 454 L2KPW6 kit lot 455 L2KPW6 kit lot 456 L2KPW6 kit lot 457 L2KPW6 kit lot 458 L2KPW6 kit lot 459 L2KPW6 kit lot 460 L2KPW6 kit lot 461 L2KPW6 kit lot 463 L2KPW6 kit lot 464 L2KPW6 kit lot 465 L2KPW6 kit lot 466 L2KPW6 kit lot 467 L2KPW6 kit lot 469 L2KPW6 kit lot 470 L2KPW6 kit lot 471 L2KPW6 kit lot 472 L2KPW6 kit lot 474 
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc
333 Coney St
East Walpole MA 02032-1516
For Additional Information Contact
508-668-5000
Manufacturer Reason
for Recall
DHEA-S causes falsely elevated progesterone results.
FDA Determined
Cause 2
Nonconforming Material/Component
Action Siemens sent an Urgent Medical Device Correction (UMDC) letter dated January 4, 2017 to all affected Siemens Healthcare Diagnostics customers in the United States dated . An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 4, 2017. These letters identified the affected product problem and actions to be taken. For questions contact your local Siemens Customer Care Center or your local Siemens Technical support.
Quantity in Commerce 11963
Distribution US Nationwide distribution including Puerto Rico
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JLS and Original Applicant = CIBA CORNING DIAGNOSTICS CORP.
-
-