Date Initiated by Firm |
March 25, 2021 |
Create Date |
June 01, 2021 |
Recall Status1 |
Terminated 3 on March 20, 2023 |
Recall Number |
Z-1719-2021 |
Recall Event ID |
87661 |
Product Classification |
Cement, bone, vertebroplasty - Product Code NDN
|
Product |
11G DIRECT WORKING CANNULA DIAMOND TIP MINI, CODE: INTVMN-DWCD. For use in Orthopedic / spinal procedures. |
Code Information |
Lot: 2002437 |
Recalling Firm/ Manufacturer |
Zavation 220 Lakeland Pkwy Flowood MS 39232-9552
|
For Additional Information Contact |
Quality Department 601-919-1119
|
Manufacturer Reason for Recall |
Products distributed as sterile may not have been adequately sterilized
|
FDA Determined Cause 2 |
Vendor change control |
Action |
Zavation initially notified their consignees of the problem by email beginning on 03/24/2021 and requested they quarantine the product. The firm sent a subsequent email beginning on 04/13/2021 requesting the return of the product. The recall is being conducted to the user level. |
Quantity in Commerce |
58 units |
Distribution |
US nationwide distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|