Date Initiated by Firm |
December 07, 2022 |
Date Posted |
February 10, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1113-2023 |
Recall Event ID |
91483 |
Product Classification |
Central venous catheter tray - Product Code OFF
|
Product |
Surgical convenience kits labeled as: CENTURION MEDICAL PRODUCTS NEONATAL PICC TRAY, Reorder Number CVI4685 |
Code Information |
GTIN 40653160990384, Lot # 22BBJ259
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
For Additional Information Contact |
Karin Johnson 886-359-1704
|
Manufacturer Reason for Recall |
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
The firm sent a correction notice to its direct consignees on 12/07/2022 by US mail and email. The notice explained the issue and requested the following:
"Please check you stock for affected kits and over-labeling will be sent to place on kits to ensure that all labeling is accurate. No kits are to be returned back to Medline. All components are not affected. Our records indicate that these were used as a component in the list of kit(s) purchased from Medline Industries, LP."
"If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP."
"If you have any questions, please contact the Recall Department at 866-359-1704." |
Quantity in Commerce |
60 units |
Distribution |
Nationwide |