| | Class 2 Recall Cook Antimicrobial Catheter |  |
| Date Posted |
March 16, 2007 |
| Recall Number |
Z-0637-2007 |
| Product |
COOK Triple Lumen Central Venous Catheter Tray with Cook Spectrum® Glide® Antimicrobial Catheter with EZ--Pass® Hydrophilic Coating: Reorder Number C-UTLMY-701J-RSC-ABRM-HC-FST-A. |
| Code Information |
Lots 1794998 and 1793611.
|
Recalling Firm/ Manufacturer |
Cook, Inc. 750 Daniels Way Bloomington, Indiana 47404-9120 |
| Consumer Instructions |
Contact the recalling firm for information |
Reason for Recall |
These catheters do not include the proximal sidehole as required by the specification.
|
| Action |
Consignees were notified by phone on 1/2/07 and instructed to return the product. |
| Quantity in Commerce |
100 |
| Distribution |
Nationwide, including the states of Florida, Louisiana, Maine, Massachusetts, Minnesota, Missouri, New Jersey and North Carolina. |
| |
|
|
|
|