• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Skytron Infinity Series surgical light

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Skytron Infinity Series surgical light see related information
Date Posted February 22, 2007
Recall Status1 Terminated on November 18, 2009
Recall Number Z-0531-2007
Recall Event ID 37120
510(K)Number K915285 
Product Classification surgical light - Product Code FTD
Product Skytron Infinity Series surgical light with dual lightheads, 5 bulbs in 1 - 22' lighthead and 8 bulbs in 1 - 30' diameter lighthead, Manufactured by Dai-ichi Shomei Co., Ltd., Made in Japan; Model IF3022B. (note only the 22'' lighthead is affected.)
Code Information Each lighthead bears a unique serial number. For units with model numbers containing IF and 22, all serial numbers with the following numbers in the third and fourth position of the numerical series: 95, 96, 97, 98, 99 or 00 (examples: xx97xxxx or xx00xxxB).
Recalling Firm/
Manufacturer
Skytron, Div. The KMW Group, Inc
5000 36th St Se
Grand Rapids MI 49512-2008
For Additional Information Contact Mary Preston-Murphy
800-759-8766 Ext. 3048
Manufacturer Reason
for Recall
If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.
FDA Determined
Cause 2
Other
Action The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.
Quantity in Commerce 5,000 for all models.
Distribution Nationwide and Canada.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
-
-