| Class 2 Device Recall MacLab | |
Date Initiated by Firm | February 16, 2007 |
Date Posted | August 07, 2007 |
Recall Status1 |
Terminated 3 on December 11, 2011 |
Recall Number | Z-1132-2007 |
Recall Event ID |
37640 |
510(K)Number | K050093 |
Product Classification |
cardiac catheterization data system - Product Code DQK
|
Product | GE Medical Systems Mac-Lab System, cardiac catheterization data system. Software Version 6.5 2021423-015 Revision D GE Medical Systems, Information Technologies, Inc., 8200 West Tower Avenue, Milwaukee, WI 53223 USA. |
Code Information |
Serial Number SBT05460156GA SBT05460162GA SBT05460163GA AA505242210GA SBT06511750GA 2003 acq1 2003 acq2 AA504130408FA SBT06120684GA AA505242190GA AA504300892GA SBT05520430GA SBT05520431GA SBT05520432GA SBT05520433GA SBT06341246GA AA504391102GA AA505091801GA AA504110305FA SBT05440090GA SBT05470225GA SBT05480245GA AA504451274GA AA504451275GA AA504451276GA SBT06100629GA SBT06230966GA SBT06100626GA SBT06100627GA SBT06110661GA SBT06311146GA AA505322445GA SBT06491674GA SBT06501680GA AA504210563FA AA504210564FA AA504210565FA AA504210562FA AA505212125GA AA505342501GA AA505242213GA SBT05500362GA SBT06331210GA SBT06511743GA SBT06511730GA SBT06451560GA SBT06401398GA SBT06401418GA SBT05500378GA SBT05480241GA SBT05480244GA SBT06501688GA SBT06331187GA SBT06311166GA SBT06311167GA SBT06331213GA SBT06341216GA SBT06341217GA SBT06341215GA SBT05500390GA SBT05500389GA AA504150487FA SBT06481627GA AA505232177GA AA505222144GA AA505242217GA AA505252219GA AA505141929GA AA505091803GA AA505161970GA SBT06371293GA SBT07021831GA SBT07021830GA AA504240694GA AA504280854GA AA504531544GA AA504361066GA SBT06190820GA SBT06190823GA AA505101827GA SBT05500393GA SBT06190829GA SBT05430065GA SBT06020452GA AA504110303FA AA504110304FA AA504120333FA AA504120334FA AA504120335FA AA504120336FA AA504120337FA AA504130433FA AA505372563GA AA504240693GA AA504130405FA AA504130407FA SBT06401390GA SBT06401419GA AA505091794GA AA505091792GA AA505091807GA AA505101816GA SBT06441536GA SBT06441537GA SBT06441540GA 2003 acq-tbd 2003 acq-tbd AA505111846GA AA505081716GA SBT06511732GA 2003 acq-tbd AA505342490GA AA505081735GA SBT06481611GA AA505342510GA AA505322402GA AA504250706GA AA504351024GA AA504110293FA AA505332482GA SBT06511733GA SBT06150737GA SBT05450128GA SBT05450146GA SBT05450151GA SBT05460152GA SBT06501693GA AA504200526FA AA504200527FA SBT07061898GA SBT07061899GA SBT05500408GA SBT06160772GA SBT06020466GA SBT05480271GA AA505262257GA AA505262258GA SBT06060578GA AA505322446GA SBT06150745GA SBT06230965GA SBT06240981GA SBT06190819GA SBT06371293GA SBT06511745GA SBT06501715GA AA505332479GA AA505342496GA AA505342497GA AA505342505GA AA505422646GA SBT06371308GA SBT06371318GA SBT05460179GA AA504441241GA AA504441257GA AA505081769GA SBT06220947GA AA505212115GA AA505232155GA AA505232159GA AA505372575GA SBT06501710GA SBT06511734GA SBT06511736GA AA505302383GA AA505372597GA AA504521483GA AA504020173FA AA504020174FA AA505332458GA AA505332469GA AA505342491GA SBT05480251GA SBT05480266GA SBT06291080GA SBT06291090GA AA504240702GA AA505081721GA AA505081725GA SBT05480264GA SBT06381351GA AA504200529FA AA505151940GA AA505332456GA AA505332457GA AA505332466GA AA505071691GA AA504120315FA AA504120316FA AA505322408GA SBT06110650GA SBT06110659GA SBT06110674GA AA505372572GA AA505372573GA AA505372577GA AA505282288GA AA505282290GA SBT06491663GA AA504280859GA AA504280860GA SBT06150732GA AA504070242FA SBT06230961GA SBT06251016GA AA505081773GA AA505081775GA AA505081777GA AA504220599FA AA504220609FA AA504220606FA AA504220605FA AA504220608FA AA504220610FA AA504220607FA SBT06491672GA 2003 acq1 2003 acq2 2003 acq3 AA504351027GA SBT06451559GA SBT06451554GA AA504391090GA AA505071659GA SBT06251034GA AA505212132GA AA505372618GA SBT06481634GA SBT06381346GA AA505161989GA AA505172000GA AA505192057GA AA505312397GA AA505312399GA SBT06180794GA AA504391104GA AA504391105GA AA504391107GA AA504391088GA AA504451269GA AA504491390GA AA504421200GA AA504491389GA AA504531548GA AA505021596GA AA505031628GA AA505292320GA AA505292321GA AA505292324GA AA504361064GA AA504361070GA SBT06070593GA SBT06421462GA SBT06421461GA AA504351058GA AA504451278GA AA504340982GA AA504391095GA AA504391096GA SBT06020479GA SBT06020484GA AA505151958GA AA504130432FA SBT06180791GA AA505091797GA AA504351045GA AA504351050GA SBT05500363GA AA504330966GA SBT06220944GA SBT06230967GA SBT06230970GA SBT06230971GA SBT06230974GA SBT06331209GA AA504391113GA SBT05430054GA SBT05500403GA AA505172002GA SBT06511748GA SBT06281063GA SBT06281070GA SBT06291079GA SBT05450116GA SBT05470223GA AA504451282GA AA504451283GA AA504180515FA AA504431209GA AA504431210GA AA504431211GA AA505312398GA AA505362536GA SBT05440098GA SBT05450123GA SBT05460172GA SBT05440099GA SBT06060576GA AA504451281GA AA505222143GA AA504391117GA AA504240685GA AA504240686GA AA505452664GA AA505302376GA AA505302353GA AA505302356GA AA505302377GA AA505302378GA SBT06511731GA SBT06511747GA SBT06180805GA AA505242187GA AA504531541GA AA504531542GA SBT06251002GA AA505222145GA SBT06391364GA AA505031614GA SBT06050551GA AA505141925GA SBT06501720GA SBT06501721GA SBT06451556GA AA505081758GA AA505081759GA AA504110306FA AA505091798GA SBT06491669GA SBT06491675GA AA504260758GA AA505081772GA AA505081784GA AA504020166FA 2003 acq-tbd SBT06481609GA SBT06100620GA SBT06100631GA SBT06120687GA SBT06251039GA AA505322447GA SBT06150753GA AA504531545GA AA504180515FA AA504210546FA AA504210547FA SBT05460176GA SBT05440091GA SBT05440092GA SBT05430064GA AA504120348FA 2003 acq-tbd AA505081781GA AA505141936GA AA505141937GA SBT06251013GA SBT06190828GA AA505292323GA SBT06120685GA AA504280857GA SBT06291083GA AA504451284GA AA504531524GA AA505322414GA AA504411154GA AA504250715GA SBT06040524GA AA504401123GA AA504411162GA AA504401122GA AA504411161GA AA504421199GA AA505151960GA AA505171995GA AA505202092GA AA504250707GA AA504330960GA SBT05440107GA SBT05470233GA SBT05470234GA AA505011568GA AA505292313GA 2003 acq1 2003 acq2 2003 acq3 SBT06060582GA 2003 acq1 2003 acq2 2003 acq3 2003 acq4 SBT06501704GA SBT05480261GA SBT05480262GA SBT05340007GA SBT05350029GA SBT05350030GA AA504501425GA AA504501432GA AA504501433GA AA504501429GA AA505322412GA AA505292309GA AA505292310GA AA505292311GA AA505282283GA SBT06331183GA SBT06331201GA AA504340999GA SBT06281046GA SBT06331208GA AA504140463FA SBT06220935GA SBT06371284GA AA504210550FA SBT06371278GA AA504391080GA AA504391081GA AA504140465FA AA505262238GA AA505262240GA AA505332475GA SBT05500404GA SBT05450124GA SBT05450125GA SBT05450129GA SBT05450130GA SBT05460154GA SBT05460160GA SBT05470226GA SBT05470232GA AA505192060GA AA505222142GA AA504481355GA AA504481356GA AA504481357GA AA504260802GA AA505141912GA SBT06381343GA AA505192048GA AA505242194GA AA505262237GA AA505262241GA AA505342508GA SBT06401399GA SBT05440088GA SBT05440096GA SBT05440106GA AA505081738GA AA505081739GA AA505081744GA AA505081745GA SBT05500380GA SBT05460168GA SBT06401386GA AA504260796GA AA504260775GA AA504260774GA AA504260777GA AA504260797GA AA505212117GA AA504391108GA AA505242196GA AA505282292GA AA504411140GA AA504441255GA SBT06070602GA SBT06100609GA SBT06501712GA SBT06511724GA SBT06511735GA SBT06501700GA AA505021578GA SBT07021833GA AA504330965GA AA505202088GA AA504471345GA AA504471346GA AA504471318GA AA505091800GA AA504130435FA AA504130436FA AA504130439FA AA504130437FA AA504130441FA AA504130442FA AA505071694GA AA505071715GA AA505071688GA AA505071698GA SBT06160763GA AA504260740GA AA504260742GA AA504260743GA AA504260741GA AA504240694GA AA505362540GA SBT05430063GA SBT06160761GA SBT07051885GA SBT07051886GA SBT07051889GA SBT06010434GA AA504120338FA AA504120339FA AA504120340FA AA504260745GA AA504260746GA SBT06140711GA SBT06281065GA SBT07041865GA SBT07041867GA SBT07041870GA AA505282282GA AA504501416GA AA505222144GA AA505322448GA SBT06140713GA SBT05500381GA AA505111850GA AA505111856GA AA505111857GA AA504110290FA AA504130350FA SBT06301134GA AA505081764GA AA505081766GA AA505081770GA SBT05480242GA AA505011569GA SBT06371307GA SBT05440103GA AA504180510FA AA504180511FA AA504180512FA SBT06311145GA AA504521478GA AA504521479GA AA504411163GA AA504340992GA SBT06251003GA SBT06251007GA SBT06251008GA AA504110302FA AA505342518GA AA505232171GA AA505232167GA AA505232168GA AA505232172GA AA505161988GA SBT06251004GA SBT06251012GA AA504491383GA SBT05480279GA SBT05500382GA AA505212113GA AA505292307GA AA505292312GA AA505292319GA AA505242200GA SBT06050555GA SBT06040529GA SBT06050554GA SBT06110639GA SBT06100611GA SBT06100633GA SBT06110636GA AA505292317GA AA505292318GA AA504471317GA AA505282291GA AA505262254GA AA505262255GA SBT06120682GA AA504250718GA AA504250720GA AA504250722GA AA504250723GA AA504250721GA SBT05430068GA SBT05430071GA SBT06100607GA AA505212120GA AA504070243FA AA504070244FA AA504070249FA AA505322419GA AA504120307FA AA504120308FA AA504401131GA AA504140483FA AA505091795GA AA504260764GA AA504260765GA AA504260766GA AA504250726GA SBT05500385GA SBT05500399GA SBT05500400GA SBT06501685GA SBT06501687GA SBT06511727GA SBT06160773GA AA504531528GA AA504531529GA AA504451277GA AA505242193GA SBT06281052GA SBT06110642GA AA504411166GA SBT06251021GA SBT06240986GA SBT06251016GA SBT06381327GA SBT06331207GA AA504300894GA BetaSSNtbd BetaSSNtbd BetaSSNtbd BetaSSNtbd BetaSSNtbd BetaSSNtbd BetaSSNtbd SBT05480254GA SBT06240987GA SBT06240990GA AA504030200FA AA504030201FA AA504030202FA SBT06401420GA SBT05500407GA SBT06501705GA AA504461294GA AA504471329GA AA504471318GA AA504330978GA AA504340990GA AA504340994GA AA505222141GA SBT05500370GA SBT05500410GA SBT05500411GA SBT05500417GA SBT05500419GA SBT05500421GA SBT05500386GA SBT05500402GA SBT05500409GA SBT06040526GA SBT06060565GA SBT05460157GA SBT05460158GA SBT07051893GA SBT06511733GA AA505332470GA SBT06180808GA SBT06180812GA SBT06190814GA SBT06110644GA SBT06110662GA AA505212108GA SBT06461579GA SBT06461580GA SBT06461581GA SBT06281042GA AA505081741GA SBT06020469GA SBT06020467GA SBT05450118GA SBT05450131GA SBT06020474GA AA505161982GA AA505182022GA AA504210577FA AA504210578FA AA504471348GA AA504471349GA AA504471350GA SBT06501716GA AA505141931GA AA505212102GA AA505212101GA AA505212103GA AA505141930GA AA505141935GA AA505372562GA SBT06501708GA SBT06501714GA AA504240699GA AA504240700GA AA504240701GA SBT06281075GA SBT05460171GA SBT05460178GA AA505332477GA AA505372616GA SBT06501718GA AA504501431GA SBT06150751GA SBT06160757GA SBT06180806GA SBT06220932GA AA505372583GA AA505372579GA AA505372581GA AA505372586GA SBT05500375GA SBT05500388GA SBT05500397GA SBT06301137GA AA505362538GA AA505342512GA AA504260800GA AA504130431FA AA504130434FA AA504130446FA AA504260763GA AA504341004GA AA504120309FA SBT06020482GA SBT06020483GA SBT06070590GA SBT06070592GA SBT06070596GA AA505182028GA AA505212100GA SBT06501703GA SBT06501711GA AA504260753GA SBT06281071GA SBT06281074GA SBT05440089GA 2003 acq-tbd 2003 acq-tbd SBT06481644GA AA505172014GA AA504030199FA AA505452658GA SBT06020464GA SBT06371277GA AA505081780GA AA504421195GA AA504300895GA AA504290875GA AA505332462GA 2003 acq1 2003 acq2 2003 acq3 SBT05460159GA AA505091803GA AA505161970GA AA505332480GA SBT06491652GA SBT06491655GA SBT06281076GA SBT06281078GA SBT06331189GA SBT06501686GA SBT06511728GA SBT06511742GA AA504120346FA AA504120345FA AA504130353FA AA504130354FA AA504491374GA AA503510088FA AA504030191FA AA504030192FA AA504030193FA SBT06190819GA AA505081768GA AA505101840GA AA505101841GA AA505101842GA AA505101843GA AA504280864GA AA504280865GA AA505131903GA AA505131906GA 2003 acq-tbd 2003 acq-tbd AA504120311FA AA504521480GA AA504521481GA AA504130398FA AA504130399FA AA504130400FA AA504130401FA AA504130402FA AA504130403FA SBT05480273GA AA504140466FA AA504140467FA AA504140468FA AA504140469FA AA504140470FA AA504020186FA AA504020187FA AA504020188FA AA505302368GA AA505302369GA AA505101829GA SBT06371314GA AA504411164GA AA504411168GA SBT05460170GA SBT06251019GA SBT06491654GA SBT06491659GA SBT06491657GA SBT06491664GA SBT06491668GA AA505372582GA AA505242189GA AA505151946GA AA505151947GA AA505151957GA AA505151964GA AA504471312GA SBT06501694GA AA504491385GA AA505081723GA AA504351020GA AA504290873GA AA504290876GA 2003 acq1 2003 acq2 2003 acq3 2003 acq4 AA504391087GA SBT05470219GA SBT06160764GA SBT06160765GA AA505011567GA AA505021585GA AA504491375GA AA505372570GA AA505372580GA AA505171997GA AA505172001GA AA505172003GA AA505242214GA SBT06160771GA SBT06301132GA SBT06301141GA AA504130428FA 2003 acq-tbd AA504260757GA SBT06150752GA SBT06311171GA SBT06311175GA SBT06321176GA SBT06371315GA SBT06381324GA SBT06381328GA SBT06381345GA SBT06381348GA AA504020167FA AA504020168FA AA504020171FA SBT06251017GA SBT06251025GA SBT06180801GA SBT06190813GA SBT06190815GA SBT06180803GA SBT06190818GA SBT06190816GA SBT06331205GA SBT06331206GA AA505322406GA AA505322409GA SBT06511749GA SBT06511752GA SBT06511756GA AA504180519FA AA505081717GA SBT05430067GA SBT05430072GA SBT05480243GA AA505192055GA AA504351051GA AA504351056GA AA505362530GA AA504130351FA SBT06060579GA SBT06070605GA SBT06120704GA SBT06120686GA AA505011573GA AA505081732GA AA505081733GA AA505081767GA AA504260748GA AA504260749GA AA504260750GA AA504260751GA AA505312395GA AA505342494GA SBT06100606GA AA505031617GA AA505081717GA SBT06150744GA AA505071656GA AA505071657GA AA505071660GA AA504491382GA SBT06301125GA SBT06371304GA SBT06371320GA SBT06381323GA SBT06301125GA AA505151959GA AA504330974GA AA504330975GA AA504330976GA AA504330979GA AA504330980GA AA504481354GA SBT06010445GA AA504310911GA AA504310913GA SBT07041866GA AA505212126GA AA505081761GA SBT06301121GA AA504120338FA AA504120339FA AA504120340FA AA504140472FA AA505372610GA AA505081726GA AA505081728GA AA505081729GA AA505081727GA SBT06411427GA AA504210572FA AA504210573FA AA504210574FA AA504120319FA AA504120320FA AA504120321FA AA504120322FA SBT05460164GA SBT06281056GA SBT06251040GA SBT06281054GA AA504120324FA AA504120323FA AA504501428GA AA504260767GA AA504260768GA AA504260769GA AA504260770GA SBT06281072GA SBT06281045GA SBT06281055GA AA504340993GA SBT05500395GA SBT05500387GA SBT06100613GA AA505322439GA SBT06381341GA SBT06010447GA SBT06451558GA AA505011562GA AA504441253GA AA504451272GA SBT06371289GA SBT06481636GA AA504280850GA SBT06421460GA AA505091803GA SBT06281051GA SBT06281053GA AA505422644GA AA504300906GA SBT06491650GA SBT06371276GA SBT06341218GA AA505111870GA AA505302355GA AA504411173GA AA504411175GA SBT06431498GA AA504451280GA AA504451279GA SBT07051891GA SBT06170780GA SBT06170782GA AA505242197GA AA505282294GA AA505292315GA AA505362529GA SBT05450120GA SBT06371294GA SBT06100619GA AA504130438FA AA504130440FA AA505161985GA AA505161986GA AA505161983GA AA505161984GA SBT06020472GA AA504421187GA AA504340995GA AA504401124GA AA504401125GA AA504401126GA AA504401130GA AA504401132GA AA504401134GA AA504401135GA AA504441240GA AA504461293GA SBT05430052GA AA504521485GA AA505101844GA AA505101845GA AA505111847GA AA504491378GA AA504491379GA AA504491380GA AA504391088GA AA504391104GA AA504391105GA AA504501423GA AA505021594GA AA505021595GA AA505071662GA AA505071664GA AA505111851GA SBT06240989GA SBT06230969GA AA504100285FA AA504100286FA SBT06251015GA SBT06251026GA AA504531531GA AA504531532GA AA504531533GA SBT06511757GA SBT07021823GA SBT07031847GA SBT07041856GA AA504260756GA SBT05500418GA SBT06060581GA SBT07061897GA AA505242186GA SBT06251023GA SBT06251029GA SBT06100622GA SBT06100628GA AA504230682GA AA505071661GA AA505322407GA SBT06371297GA SBT06381330GA SBT06160776GA SBT06230959GA AA505141909GA SBT06491648GA SBT06491656GA SBT06491662GA SBT06501702GA AA505071693GA AA504070229FA AA504100278FA AA504100280FA AA504100281FA AA504100282FA AA504110287FA AA504120332FA AA504130349FA AA504150486FA AA504160491FA AA504160496FA AA504160497FA AA504170499FA AA504180520FA AA504210555FA AA504210556FA AA504210569FA AA504220625FA AA504220662GA AA504240687GA AA504240688GA AA504240703GA AA504250732GA AA504250733GA AA504250736GA AA504260747GA AA504280848GA AA504280849GA AA504280851GA AA504280853GA AA504290866GA AA504290867GA AA504290869GA AA504300890GA AA504300892GA AA504330963GA AA504341001GA AA504341002GA AA504341003GA AA504341007GA AA504341008GA AA504341009GA AA504351022GA AA504351026GA AA504351029GA AA504351034GA AA504351047GA AA504351052GA AA504351053GA AA504351054GA AA504381079GA AA504391103GA AA504401120GA AA504401121GA AA504401136GA AA504411169GA AA504411170GA AA504411174GA AA504421196GA AA504421201GA AA504421202GA AA504421207GA AA504441237GA AA504441238GA AA504441242GA AA504441260GA AA504441263GA AA504441267GA AA504451270GA AA504451273GA AA504461295GA AA504471302GA AA504471306GA AA504471311GA AA504471316GA AA504471318GA AA504471323GA AA504471330GA AA504471347GA AA504491367GA AA504491371GA AA504491393GA AA504491395GA AA504491396GA AA504501415GA AA504501418GA AA504501420GA AA504501426GA AA504501430GA AA504521487GA AA504531522GA AA505011561GA AA505011564GA AA505021574GA AA505021579GA AA505021584GA AA505021597GA AA505031615GA AA505031634GA AA505041642GA AA505041643GA AA505041645GA AA505071655GA AA505071663GA AA505071665GA AA505071666GA AA505071667GA AA505071685GA AA505071696GA AA505071705GA AA505081720GA AA505081736GA AA505081746GA AA505081762GA AA505081765GA AA505081778GA AA505081783GA AA505091790GA AA505091804GA AA505101818GA AA505101826GA AA505111860GA AA505111868GA AA505141909GA AA505141934GA AA505151941GA AA505151955GA AA505151962GA AA505161968GA AA505171994GA AA505171998GA AA505171999GA AA505192053GA AA505192054GA AA505192058GA AA505192075GA AA505202086GA AA505202093GA AA505202096GA AA505202098GA AA505212104GA AA505212109GA AA505212112GA AA505232163GA AA505232173GA AA505232176GA AA505242184GA AA505242188GA AA505242195GA AA505242201GA AA505242205GA AA505242208GA AA505242210GA AA505242211GA AA505242212GA AA505282266GA AA505282274GA AA505282288GA AA505282290GA AA505282293GA AA505302350GA AA505302351GA AA505302359GA AA505302360GA AA505302382GA AA505312392GA AA505322410GA AA505322411GA AA505322416GA AA505322436GA AA505322437GA AA505322440GA AA505322450GA AA505332467GA AA505332483GA AA505342488GA AA505342492GA AA505342493GA AA505342499GA AA505342502GA AA505342509GA AA505362523GA AA505362528GA AA505362533GA AA505362535GA AA505372564GA AA505372565GA AA505372566GA AA505372567GA AA505372568GA AA505372591GA AA505372611GA AA505372612GA AA505372614GA AA505372615GA AA505422648GA AA505442656GA AA505452659GA AA505452662GA AA505452663GA AA505462665GA AA505462666GA AA505462671GA AA505462677GA AA505482687GA AA505482688GA AA505482695GA AA505512698GA AA505512701GA AA505512702GA AA505512703GA AA505522706GA AA505522707GA AA506022716GA AA506042722GA AA506052723GA AA506082725GA AA506082727GA RST07039958GA SBT05430051GA SBT05430055GA SBT05430058GA SBT05430066GA SBT05430069GA SBT05440097GA SBT05440104GA SBT05450111GA SBT05450113GA SBT05450117GA SBT05450127GA SBT05450132GA SBT05450133GA SBT05450133GA SBT05450142GA SBT05450143GA SBT05450149GA SBT05460182GA SBT05470217GA SBT05470224GA SBT05470230GA SBT05480239GA SBT05480246GA SBT05480249GA SBT05480252GA SBT05480253GA SBT05480256GA SBT05480259GA SBT05480269GA SBT05480274GA SBT05480275GA SBT05480276GA SBT05480277GA SBT05480278GA SBT05490323GA SBT05500366GA SBT05500368GA SBT05500373GA SBT05500374GA SBT05500383GA SBT05500412GA SBT06010435GA SBT06010436GA SBT06010437GA SBT06010437GA SBT06010438GA SBT06010444GA SBT06020450GA SBT06020480GA SBT06020485GA SBT06030505GA SBT06040528GA SBT06040529GA SBT06050553GA SBT06050554GA SBT06050555GA SBT06050559GA SBT06050561GA SBT06060564GA SBT06070588GA SBT06070589GA SBT06070595GA SBT06070597GA SBT06070601GA SBT06100612GA SBT06100614GA SBT06100616GA SBT06100630GA SBT06110637GA SBT06110647GA SBT06110655GA SBT06110663GA SBT06110666GA SBT06120683GA SBT06120702GA SBT06120703GA SBT06140708GA SBT06140710GA SBT06140712GA SBT06140719GA SBT06140726GA SBT06140727GA SBT06140728GA SBT06150733GA SBT06150736GA SBT06150741GA SBT06150742GA SBT06150754GA SBT06150756GA SBT06160777GA SBT06180784GA SBT06180792GA SBT06180793GA SBT06180796GA SBT06180800GA SBT06190827GA SBT06190830GA SBT06190832GA SBT06220931GA SBT06220933GA SBT06220936GA SBT06220946GA SBT06230961GA SBT06230973GA SBT06240986GA SBT06251006GA SBT06251036GA SBT06261041GA SBT06281050GA SBT06281077GA SBT06291081GA SBT06291087GA SBT06291088GA SBT06301124GA SBT06301126GA SBT06301128GA SBT06301129GA SBT06301133GA SBT06301135GA SBT06301138GA SBT06301140GA SBT06301143GA SBT06311147GA SBT06331188GA SBT06331190GA SBT06331191GA SBT06331203GA SBT06331204GA SBT06341246GA SBT06341248GA SBT06371281GA SBT06371293GA SBT06371312GA SBT06371313GA SBT06371322GA SBT06381331GA SBT06381333GA SBT06381335GA SBT06381353GA SBT06391375GA SBT06391376GA SBT06401387GA SBT06401400GA SBT06411423GA SBT06411428GA SBT06421456GA SBT06421465GA SBT06421466GA SBT06431499GA SBT06441523GA SBT06451549GA SBT06451550GA SBT06451551GA SBT06451552GA SBT06451555GA SBT06451557GA SBT06451561GA SBT06451563GA SBT06461576GA SBT06461577GA SBT06481612GA SBT06481613GA SBT06481617GA SBT06481619GA SBT06481623GA SBT06481628GA SBT06481632GA SBT06481641GA SBT06481642GA SBT06481643GA SBT06491647GA SBT06491658GA SBT06491666GA SBT06491667GA SBT06501690GA SBT06501692GA SBT06501698GA SBT06501701GA SBT06501707GA SBT06511723GA SBT06511738GA SBT06511741GA SBT06511744GA SBT06511751GA SBT06511755GA SBT06511759GA SBT06521761GA SBT07011806GA SBT07021821GA SBT07021826GA SBT07021832GA SBT07041854GA SBT07051888GA SBT07051890GA SBT07061896GA SBT07061904GA AA504070263FA AA504140481FA AA504150489FA AA504280861GA AA504411152GA AA504461296GA AA504471305GA AA504481359GA AA505021598GA AA505041646GA AA505071690GA AA505182029GA AA505182030GA AA505182038GA AA505182040GA AA505192047GA AA505192049GA AA505192050GA AA505202094GA AA505302352GA AA505302358GA AA505302361GA AA505302372GA AA505302375GA AA505452657GA SBT06140720GA |
Recalling Firm/ Manufacturer |
General Electric Medical Systems Information Technology 8200 W Tower Ave Milwaukee WI 53223-3219
|
Manufacturer Reason for Recall | GE Healthcare has identified conditions that could cause a potential loss of real-time monitoring with the CardioLab/Mac-Lab/ComboLab Electrophysiology and Hemodynamic Monitoring Systems. These issues are associated with an audio driver, and the use of the application's log and imaging windows. |
FDA Determined Cause 2 | Other |
Action | Consignees were sent a letter dated 2/16/07. The letter described the product and problem, made recommendations and gave information on software update. |
Quantity in Commerce | 1513 (1067 US and 446 OUS) |
Distribution | Worldwide, including USA, Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Pakistan, Panama, Puerto Rico, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Kingdom, and Venezuela. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DQK
|
|
|
|