| | Class 2 Recall Stryker stretcher |  |
| Date Posted |
September 21, 2007 |
| Recall Number |
Z-1009-2007 |
| Product |
Stryker Medical -Wheeled Electric Stretcher, Model 1550 (with and without scales) |
| Code Information |
Serial numbers 0508083666, 0509069531, 0509069532, 0509069533, 0509069534, 0509069535, 0509069542, 0509069564, 0509069565, 0509069566, 0509069567, 0509069568, 0509069569, 0509069588, 0509069589, 0509069590, 0509069591, 0509069662, 0509069663, 0509069664, 0509069665, 0509069693, 0509069694, 0509069695, 0509069696, 0509069697, 0509069698, 0509069699, 0509069700, 0509069701, 0509069770, 0509069771, 0509069772, 0509069773, 0509069774, 0509069775, 0509069776, 0509069777, 0509069778, 0509069795, 0509069796, 0509069806, 0509069816, 0509069817, 0509069818, 0509069819, 0509069832, 0509069835, 0509069836, 0509069837, 0509069838, 0509069839, 0509069840, 0509069841, 0509069848, 0509069849, 0509069873, 0509069874, 0509069875, 0509069876, 0509069877, 0510071024, 0510071025,
0510071026, 0510071032, 0510071033, 0510071034, 0510071035, 0510071036, 0510071037, 0510071038, 0510071039, 0510071040, 0510071041, 0510071042, 0510071043, 0510071154, 0510071155, 0510071156, 0510071157, 0510071158, 0510071159, 0510071160, 0510071205, 0510071206, 0510071207, 0510071208, 0510071209, 0510071210, 0510071231, 0510071232, 0510071233, 0510071234, 0510071235, 0510071263, 0510071264, 0510071265, 0510071266, 0510071267, 0510071308, 0510071309, 0510071310, 0510071311, 0510071312, 0510071327, 0510071328, 0510071329, 0510071349, 0510071350, 0510071351, 0510071352, 0510071353, 0510071354, 0510071355, 0510071356, 0511071505, 0511071506, 0511071507, 0511071508, 0511071509, 0511071510, 0511071511, 0511071512, 0511071513, 0511071514, 0511071515, 0511071516, 0511071572, 0511071573, 0511071574, 0511071575, 0511071576, 0511071577, 0511071578, 0511071579, 0511071583, 0511071584, 0511071585, 0511071593, 0511071594, 0511071595, 0511071596, 0511071597, 0511071598, 0511071599, 0511071600, 0511071601, 0511071602, 0511071603, 0511071604, 0511071605, 0511071653, 0511071654, 0511071655, 0511071656, 0511071687, 0511071688, 0511071689, 0511071690, 0511071691, 0511071692, 0511071693, 0511071694, 0511071695, 0511071696, 0511071697, 0511071698, 0511071699, 0511071725, 0511071726, 0512070508, 0512070509, 0512070510, 0512070511, 0512070517, 0512070518, 0512070519, 0512070520, 0512070521, 0512070522, 0512070523, 0512070524, 0512070535, 0512070536, 0512070537, 0512070538, 0512070539, 0512070540, 0512070541, 0512070542, 0512070544, 0512070597, 0512070598, 0512070599, 0512070600, 0512070607, 0512070608, 0512070609, 0512070610, 0512070611, 0512070612, 0512070613, 0512070614, 0512070648, 0512070649, 0512070650, 0512070651, 0512070656, 0512070657, 0512070658, 0512070659, 0512070660, 0512070661, 0512070701, 0512070702, 0512070703, 0512073509, 0512073510, 0512073511, 0512073518, 0512073519, 0512073529, 0512073530, 0512073531, 0512073532, 0512073533, 0512073587, 0512073588, 0512073589, 0512073590, 0512073591, 0512073592, 0512073593, 0512073594, 0512073609, 0512073610, 0512073611, 0512073612, 0512073613, 0512073614, 0512073634, 0512073635, 0512073636, 0601084021, 0601084022, 0601084023, 0601084024, 0601084025, 0601084026, 0601084027, 0601084028, 0601084029, 0601084030, 0601084080, 0601084081, 0601084082, 0601084104, 0601084107, 0601084108, 0601084109, 0601084110, 0601084111, 0601084135, 0601084136, 0601084141, 0601084142, 0601084143, 0601084144, 0601084145, 0601084146, 0601084147, 0601084199, 0601084200, 0601084201, 0601084202, 0601084203, 0601084204, 0601084214, 0601084215, 0601084227, 0601084228, 0601084229, 0601084230, 0601084231, 0601084232, 0601084233, 0601084234, 0601084241, 0601084242, 0601084243, 0601084244, 0601084245, 0601084252, 0601084253, 0601084254, 0601084255, 0601084261, 0601084262, 0601084263, 0601084264, 0601084267, 0601084268, 0601084269, 0601084270, 0601084271, 0601084272, 0601084273, 0601084274, 0601084275, 0601084276, 0601084277, 0601084278, 0601084279, 0601084280, 0601084325, 0601084326, 0601084327, 0601084328, 0601084329, 0601084330, 0601084359, 0601084360, 0601084361, 0601084362, 0601084363, 0601084364, 0601084382, 0601084383, 0601084384, 0601084385, 0601084386, 0602079501, 0602079502, 0602079503, 0602079504, 0602079505, 0602079506, 0602079524, 0602079525, 0602079526, 0602079527, 0602079528, 0602079529, 0602079557, 0602079558, 0602079559, 0602079560, 0602079561, 0602079562, 0602079563, 0602079591, 0602079592, 0602079593, 0602079594, 0602079595, 0602079596, 0602079597, 0602079598, 0602079599, 0602079648, 0602079649, 0602079650, 0602079651, 0602079661, 0602079662, 0602079663, 0602079664, 0602079665, 0602079666, 0602079667, 0602079668, 0602079685, 0602079686, 0602079687, 0602079688, 0602079695, 0602079696, 0602079697, 0602079698, 0602079699, 0602079716, 0602079718, 0602079719, 0602079720, 0602079721, 0602079722, 0602079723, 0602079724, 0602079725, 0602079726, 0602079727, 0602079728, 0602079729, 0602079730, 0602079731, 0602079732, 0602079733, 0602079734, 0602079735, 0602079802, 0602079803, 0602079804, 0602079805, 0602079806, 0602079807, 0602079808, 0602079809, 0602079810, 0602079835, 0602079836, 0602079844, 0602079845, 0602079846, 0602079847, 0602079848, 0602079849, 0602079883, 0602079886, 0602079890, 0602079891, 0602079892, 0602079893, 0603084508, 0603084509, 0603084510, 0603084511, 0603084515, 0603084537, 0603084541, 0603084542, 0603084543, 0603084544, 0603084545, 0603084546, 0603084547, 0603084548, 0603084549, 0603084566, 0603084567, 0603084568, 0603084569, 0603084570, 0603084571, 0603084572, 0603084589, 0603084590, 0603084591, 0603084592, 0603084593, 0603084594, 0603084632, 0603084633, 0603084634, 0603084635, 0603084636, 0603084637, 0603084638, 0603084644, 0603084645, 0603084647, 0603084648, 0603084649, 0603084689, 0603084742, 0603084743, 0603084744, 0603084745, 0603084746, 0603084747, 0603084773, 0603084774, 0603084775, 0603084776, 0603084777, 0603084778, 0603084801, 0603084802, 0603084803, 0603084804, 0603084805, 0603084812, 0603084813, 0603084814, 0603084815, 0604079501, 0604079502, 0604079503, 0604079524, 0604079525, 0604079639, 0604079640, 0604079641, 0604079649, 0604079691, 0604079706, 0604079707, 0604079708, 0604079725, 0604079726, 0604079727, 0604079734, 0604079735, 0604079736, 0604079800, 0604079801, 0604079803, 0604079805, 0604079806, 0604079807, 0604079808, 0604079809, 0604079810, 0604079854, 0604079859, 0604079860, 0604079861, 0604079862, 0604079863, 0604079864, 0604079865, 0604079866, 0604079867, 0604079868, 0604079869, 0604079870, 0605087570, 0605087588, 0605087589, 0605087590, 0605087591, 0605087592, 0605087593, 0605087652, 0605087653, 0605087660, 0605087661, 0605087662, 0605087663, 0605087721, 0605087722, 0605087723, 0605087727, 0605087728, 0605087729, 0605087730, 0605087731, 0605087788, 0605087789, 0605087790, 0605087791, 0605087792, 0605087793, 0605087810, 0605087811, 0605087812, 0605087813, 0605087814, 0605087815, 0605087867, 0605087868, 0605087869, 0605087870, 0605087872, 0605087873, 0605087876, 0605087877, 0605087878, 0605087879, 0605087880, 0605087881, 0605087928, 0605087929, 0605087930, 0605087931, 0605087932, 0605087933, 0605087938, 0605087939, 0605087940, 0605087941, 0605087942, 0606079001, 0606079003, 0606079004, 0606079047, 0606079048, 0606079049, 0606079050, 0606079051, 0606079052, 0606079053, 0606079054, 0606079055, 0606079056, 0606079129, 0606079130, 0606079136, 0606079137, 0606079138, 0606079289, 0606079290, 0606079307, 0606079308, 0606079309, 0606079310, 0606079311, 0606079312, 0606079313, 0606079314, 0606079315, 0606079316, 0606079337, 0606079359, 0606079360, 0606079361, 0607080526, 0607080527, 0607080528, 0607080660, 0607080664, 0607080665, 0607080666, 0607080667, 0607080668, 0607080669, 0607080670, 0607080671, 0607080717, 0607080718, 0607080719, 0607080720, 0607080721, 0607080722, 0607080723, 0607080724, 0607080725, 0607080726, 0607080727, 0607080728, 0607080729, 0607080786, 0607080806, 0607080811, 0607080812, 0607080813, 0607080814, 0607080840, 0608080501, 0608080502, 0608080503, 0608080538, 0608080539, 0608080540, 0608080541, 0608080542, 0608080543, 0608080544, 0608080545, 0608080546, 0608080547, 0608080572, 0608080573, 0608080574, 0608080575, 0608080576, 0608080577, 0608080578, 0608080648, 0608080666, 0608080667, 0608080688, 0608080736, 0608080737, 0608080738, 0608080739, 0608080772, 0608080773, 0608080774, 0608080793, 0608080794, 0608080795, 0608080796, 0608080813, 0608080814, 0608080815, 0608080904, 0608080905, 0608080906, 0608080907, 0608080934, 0608080935, 0608080994, 0608080995, 0608084591, 0608084592, 0608084593, 0608084594, 0608084595, 0608084596, 0608084597, 0608084598, 0609081739, 0609081740, 0609081741, 0609081797, 0609081798, 0609081799, 0609081800, 0609081801, 0609081810, 0609081811, 0609081823, 0609081824, 0609081841, 0609081842, 0610082025, 0610082026, 0610082027, 0610082028, 0610082029, 0610082030, 0610082031, 0610082032, 0610082033, 0610082034, 0610082035, 0610082036, 0610082037, 0610082038, 0610082039, 0610082110, 0610082111, 0610082112, 0610082113, 0610082114, 0610082115, 0610082154, 0610082155, 0610082156, 0610082157, 0610082158, 0610082159, 0610082228, 0610082229, 0610082230, 0610082276, 0610082277, 0610082280, 0610082310, 0610082311, 0610082312, 0610082313, 0610082377, 0610082378, 0610082379, 0610082390, 0610082391, 0610082392, 0610082400, 0611080561, 0611080578, 0611080746, 0611080747, 0611080748, 0611080749, 0611080775, 0611080776, 0611080821, 0611080822, 0611080862, 0611080863, 0611080864, 0611080865, 0611080866, 0611080867, 0611080868, 0611080869, 0611080870, 0611080871, 0611080872, 0611080873, 0611080933, 0611080955, 0611080956, 0611080957, 0611080958, 0611080959, 0612081517, 0612081518, 0612081519, 0612081520, 0612081561, 0612081562, 0612081563, 0612081564, 0612081565, 0612081566, 0612081628, 0612081629, 0612081630, 0612081631, 0612081632, 0612081633, 0612081650, 0612081651, 0612081652, 0612081653, 0612081654, 0612081655, 0612081683, 0612081684, 0612081685, 0612081686, 0612081719, 0612081720, 0612081721, 0612081724, 0612081743, 0612081774, 0612081786, 0612081787, 0612081796, 0612081797, 0612081798, 0612081799, 0612081800, 0612081801, 0612081868, 0612081869, 0612081870, 0612081871, 0612081872, 0612081873, 0612081921, 0612081922, 0701089012, 0701089013, 0701089014, 0701089015, 0701089016, 0701089017, 0701089067, 0701089068, 0701089069, 0701089070, 0701089071, 0701089072, 0701090501, 0701090502, 0701090503, 0701090504, 0701090505, 0701090506, 0701090548, 0701090569, 0701090570, 0701090571, 0701090572, 0701090573, 0701090574, 0701090628, 0701090629, 0701090630, 0701090631, 0701090632, 0701090633, 0701090634, 0701090635, 0701090636, 0701090637, 0701090638, 0701090639, 0701090640, 0701090641, 0701090642, 0701090643, 0701090644, 0701090645, 0701090811, 0701090812, 0701090813, 0701090836, 0701090837, 0701090838, 0701090839, 0701090840, 0701090841, 0701090842, 0701090843, 0701090898, 0701090899, 0701090900, 0701090901, 0701090902, 0701090903, 0701090904, 0701090905, 0701090906, 0701090907, 0701090965, 0702097501, 0702097502, 0702097503, 0702097504, 0702097505, 0702097506, 0702097533, 0702097534, 0702097535, 0702097536, 0702097537, 0702097538, 0702097600, 0702097660, 0702097661, 0702097662, 0702097663, 0702097664, 0702097665, 0702097706, 0702097707, 0702097708, 0702097709, 0702097710, 0702097741, 0702097742, 0702097743, 0702097744, 0702097745, 0702097746, 0702097790, 0702097791, 0702097792, 0702097793, 0702097794, 0702097795, 0702097817, 0702097818, 0702097819, 0702097820, 0702097821, 0702097822, 0702097834, 0702097835, 0702097836, 0702097837, 0702097838, 0702097839, 0702097888, 0702097889, 0702097890, 0702097891, 0702097892, 0702097893, 0702097917, 0702097918, 0702097919, 0702097920, 0702097921, 0702097922, 0703094502, 0703094503, 0703094505, 0703094506, 0703094507, 0703094508, 0703094543, 0703094545, 0703094546, 0703094598, 0703094599, 0703094600, 0703094601, 0703094602, 0703094603, 0703094604, 0703094605, 0703094606, 0703094607, 0703094608, 0703094609, 0703094610, 0703094648, 0703094677, 0703094678, 0703094679, 0703094680, 0703094681, 0703094682, 0703094755, 0703094756, 0703094757, 0703094758, 0703094759, 0703094760, 0703094862, 0703094863, 0703094864, 0703094865, 0703094866, 0703094867, 0703094888, 0703094889, 0703094897, 0703094898, 0703094899, 0703094900, 0703094901, 0703094903, 0703094923, 0703094924, 0703094925, 0703094926, 0703094927, 0703094928, 0703094943, 0703094944, 0703094945, 0703094946, 0703094947, 0703094980, 0703094981, 0703094982, 0703094996, 0703094997, 0703094998, 0703094999, 0704096516, 0704096517, 0704096518, 0704096519, 0704096529, 0704096530, 0704096531, 0704096532, 0704096554, 0704096555, 0704096556, 0704096557, 0704096574, 0704096575, 0704096576, 0704096583, 0704096584, 0704096585, 0704096586, 0704096623 and 0704096643.
|
Recalling Firm/ Manufacturer |
Stryker Medical Div. of Stryker Corporation 3800 E. Centre Ave. Portage, Michigan 49002 |
| For Additional Information Contact |
Mike Mazelin 800-669-4968 Ext. 6687
|
Reason for Recall |
The internal power cord may become damaged due to wear and, if this occurs, a shock hazard exists.
|
| Action |
Consignees were notified of the problem via letter dated 5/7/07 instructing them to immediately unplug the affected stretchers with scales, affix a warning label enclosed with the letter to the power cord of stretchers with scales, and that they will be contacted so that the firm can schedule a visit to upgrade all affected stretchers. A second notification letter was issued on August 7, 2007 to cover all 1550 stretchers with and without scales. |
| Quantity in Commerce |
1,048 (1009 in the U.S. plus 39 international) |
| Distribution |
Worldwide: USA, Latin America, Mexico and Spain. |
| |
|
|
|
|