Class 2 Device Recall Medtronic Sofamor Danek, Satellite System Sphere PEEK |
|
Date Initiated by Firm |
September 13, 2007 |
Date Posted |
December 22, 2007 |
Recall Status1 |
Terminated 3 on August 18, 2009 |
Recall Number |
Z-0203-2008 |
Recall Event ID |
44967 |
Product Classification |
Spinal System - Product Code NVR
|
Product |
Medtronic Sofamor Danek, Satellite System Sphere PEEK, Size 11 mm, Material PEEK-Optima, quantity 1, Sterile, Rx only, REF 9000211, Manufactured at: Warsaw, Indiana, USA |
Code Information |
ALL LOTS to include: LX53, LX59,MG73, MY47, MY49, NC36, ND61, NE27 |
Recalling Firm/ Manufacturer |
Medtronic Sofamor Danek USA Inc 1800 Pyramid Place Memphis TN 38132-1719
|
For Additional Information Contact |
Bert Kelly 901-396-3133
|
Manufacturer Reason for Recall |
Labeling, including indications for use, lack 510 (K) approval
|
FDA Determined Cause 2 |
Labeling design |
Action |
On September 12, 2007, the recalling firm's representatives were instructed by electronic mail to remove any manuals in the field and contact surgeons who have evaluated, implanted or been trained to implant the Satellite System and to collect any guides they may have on site. The sales representatives that have accounts where the Satellite product was shipped directly were mailed the information and instructed to hand deliver the 9/13/07, Urgent Device Recall notices to each consignees risk management office. The recall was then expanded on 10/31/2007, to include the surgical manual, the implants and associated instruments. The delivery of the second Urgent Device Recall Notices, dated 10/25/07, were handled in three ways: 1) The letter was delivered to all initial consignees that received satellite products along with a questionnaire (response card) to be completed by the Risk Manager or equivalent administrative function., 2) The letter without the questionnaire was sent directly to target surgeons who are known to have implanted or have been trained to implant Satellite Product 3) a field action confirmation form was sent to Sales Representative with surgeon customers that are receiving the recall communication. The form will be completed by the Sales Representative to confirm the surgeons have received the recall communication and have returned any of the surgical techniques and instruments. |
Quantity in Commerce |
198 units |
Distribution |
Nationwide |