|
Class 2 Device Recall BHM Medical Scales |
|
Date Initiated by Firm |
December 19, 2007 |
Date Posted |
February 26, 2008 |
Recall Status1 |
Terminated 3 on November 18, 2009 |
Recall Number |
Z-1135-2008 |
Recall Event ID |
46210 |
Product Classification |
Medical Scales - Product Code FRW
|
Product |
BHM Medical Scales, 800 lb. scale weight limit; Part Number 700.00510, Manufactured by BHM Medical Inc., Magog, QC, Canada J1X 5Y5;
These scales were sold for installation on or as a component of the following patient lifts: a) Manufactured by BHM Medical Inc., Magog, QC, Canada J1X 5Y5: Part Number: 88570101, Ergolift 600 BAL. Wide Opening Patient Lift; Part Number : ERGOLIFT-600+SC, Ergolift 600 Lbs with Scale Patient Lift; Part Number: ERGOLIFT-2 + SC, Ergolift 400 Lbs with Scale Patient Lift; b) Manufactured for Sunrise Medical, Longmont, CO 80503: Part Number : 700.05610.06, Hoyer-600 Spreader Bar with Scale; Part Number: 9600102.06, Hoyer-600 Sunrise Patient Lifter with Scale, Model HPL-600WBSC; Part Number: 9600132.06, Hoyer-600 Wide Opening Patient Lifter with Scale
|
Code Information |
Part Number 700.00510, Serial Numbers :ERSC-3044, ERSC-3060, ERSC-5318, ERSC-5323, ERSC-5324, ERSC-5907, ERSC-5967, ERSC-6032, ERSC-6073, ERSC-6232, ERSC-6316, ERSC-6326, ERSC-6354, ERSC-6355, ERSC-6357, ERSC-6365, ERSC-6372, ERSC-6532, ERSC-6703, ERSC-6716, ERSC-6717, ERSC-6719, ERSC-7024, ERSC-7025, ERSC-7060, ERSC-7064, ERSC-7347, ERSC-7351, ERSC-7532, ERSC-7551, ERSC-7553, ERSC-7554, ERSC-7555, ERSC-7556, ERSC-7559, ERSC-7560, ERSC-7561, ERSC-7562, ERSC-7565, ERSC-7566, ERSC-7567, ERSC-7614, ERSC-7617, ERSC-7623, ERSC-7625, ERSC-7665, ERSC-7709, ERSC-8487, ERSC-8488, ERSC-8492, ERSC-8498, ERSC-8500, ERSC-8501, ERSC-8503, ERSC-8509, ERSC-8561, ERSC-8565, ERSC-8632, ERSC-8634, ERSC-8635, ERSC-8638, ERSC-8644, ERSC-8648, ERSC-8658, ERSC-8675, ERSC-8834, ERSC-8848, ERSC-8874, ERSC-8883, ERSC-8886, ERSC-8887, ERSC-8890, ERSC-8894, ERSC-8896, ERSC-8899, ERSC-8900, ERSC-8918, ERSC-8920, ERSC-8921, ERSC-8923, ERSC-8924, ERSC-8925, ERSC-8927, ERSC-8992, ERSC-9348, and ERSC-9361----- Part Number 88570101, Serial Numbers: 885-0100. 885-0104, 885-0105, 885-0139, 885-0210, 885-0211, 885-0218, 885-0219, 885-0220, 885-0266, 885-0267, 885-0269, 885-0273, 885-0275, and 885-2676---- Part Number ERGOLIFT-600+SC, Serial Numbers: ER600-1402, ER600-1406, ER600-1435, ER600-1436, ER600-1437, ER600-1438, ER600-1460, ER600-1475, ER600-1476, ER600-1477, ER600-1478, ER600-1479, ER600-1480, ER600-1481, ER600-1483, ER600-1484, ER600-1485, ER600-1486, ER600-1515, ER600-1517, ER600-1594, ER600-1671, ER600-1672, ER600-1673, ER600-1674, ER600-1675, ER600-1676, ER600-1677, ER600-1721, ER600-1722, ER600-1724, ER600-1803, ER600-1988, ER600-1989, ER600-1990, ER600-1991, ER600-1992, and ER600-2070---- Part Number ERGOLIFT-2 + SC, Serial Numbers: ERLI-1471, ERLI-1472, ERLI-1473, ERLI-1474, ERLI-3046, ERLI-3144, ERLI-3145, ERLI-3146, ERLI-3232, ERLI-3271, ERLI-3390, ERLI-3391, ERLI-3392, ERLI-3398, ERLI-3436, ERLI-3466, ERLI-3471, ERLI-3496, ERLI-3508, and ERLI-3526---- Part Number 700.05610.06, Serial numbers: 5427, 5428, 5639, 5640, 5641, 5642, 5644, 5645, 5646, 5647, 5657, 5658, 5913, 5915, 5916, 5917, 5918, 5919, 5920, 5962, 5985, 5986, 5987, 5988, 5989, 5990, 5991, 5992, 5993, 5994, 5995, 5996, 5997, 5998, 5999, 6000, 6001, 6002, 6003, 6004, 6005, 6006, 6007, 6008, 6009, 6010, 6011, 6013, 6014, 6015, 6016, 6017, 6020, 6042, 6045, 6046, 6050, 6053, 6054, 6056, 6099, 6100, 6101, 6102, 6103, 6104, 6105, 6106, 6107, 6108, 6109, 6110, 6111, 6112, 6113, 6114, 6115, 6116, 6119, 6120, 6123, 6129, 6135, 6136, 6137, 6138, 6139, 6140, 6141, 6142, 6143, 6144, 6145, 6146, 6147, 6148, 6149, 6150, 6151, 6152, 8424, 8425, 8426, 8427, 8428, 8429, 8433, 8434, 8435, 8436 8437, 8438, 8439, 8440, 8441, 8460, 8461, 8462, 8463, 8464, 8465, 8466, 8467, 8468, and 8469---- Part Number 9600102.06, Serial Numbers: ER600-1330, ER600-1331, ER600-1332, ER600-1333, ER600-1334, ER600-1335, ER600-1336, ER600-1337, ER600-1338, ER600-1339, ER600-1377, ER600-1379, ER600-1380, ER600-1381, ER600-1382, ER600-1383, ER600-1384, ER600-1561, ER600-1562, ER600-1563, ER600-1564, ER600-1565, ER600-1566, ER600-1567, ER600-1568, ER600-1569, ER600-1570, ER600-1571, ER600-1572, ER600-1573, ER600-1574, ER600-1575, ER600-1576, ER600-1577, ER600-1578, ER600-1579, ER600-1580, ER600-1581, ER600-1582, ER600-1583, ER600-1584, ER600-1585, ER600-1586, ER600-1587, ER600-1588, ER600-1589, ER600-1590, ER600-1641, ER600-1642, ER600-1643, ER600-1644, ER600-1645, ER600-1646, ER600-1647, ER600-1648, ER600-1649, ER600-1650, ER600-1651, ER600-1652, ER600-1653, ER600-1654, ER600-1655, ER600-1656, ER600-1657, ER600-1658, ER600-1659, ER600-1660, ER600-1755, ER600-1756, ER600-1757, ER600-1758, ER600-1759, ER600-1760, ER600-1761, ER600-1762, ER600-1763, ER600-1764, ER600-1765, ER600-1766, ER600-1767, ER600-1768, ER600-1769, ER600-1770, ER600-1771, ER600-1772, ER600-1773, ER600-1774, ER600-1854, ER600-1855, ER600-1856, ER600-1857, ER600-1858, ER600-1859, ER600-1860, ER600-1861, ER600-1862, ER600-1863, ER600-1864, ER600-1865, ER600-1866, ER600-1867, ER600-1868, ER600-1869, ER600-1870, ER600-1871, ER600-1872, ER600-1873, ER600-1874, ER600-1875, ER600-1876, ER600-1877, ER600-1878, ER600-1879, ER600-1880, ER600-1881, ER600-1882, ER600-1883, ER600-1948, ER600-1949, ER600-1950, ER600-1951, ER600-1952, ER600-1953, ER600-1954,ER600-1955 , ER600-1956, ER600-1957, ER600-1958, ER600-1959, ER600-1960, ER600-1961, ER600-1962, ER600-1963, ER600-1964, ER600-1965, ER600-1966, ER600-1967, ER600-2014, ER600-2015, ER600-2016, ER600-2017, ER600-2018, ER600-2019, ER600-2020, ER600-2021, ER600-2022, ER600-2023, ER600-2024, ER600-2025, ER600-2026, ER600-2027, ER600-2028, ER600-2029, ER600-2030, ER600-2031, ER600-2032, and ER600-2033---- Part Number 9600132.06, Serial Numbers: ER600-2146, ER600-2147, ER600-2148, ER600-2149, ER600-2150, ER600-2151, ER600-2154, ER600-2155, ER600-2156, ER600-2157, ER600-2158, ER600-2159, ER600-2160, ER600-2161, ER600-2162, ER600-2163, ER600-2164, ER600-2165, ER600-2166, ER600-2167, ER600-2168, ER600-2169, ER600-2170, ER600-2320, ER600-2321, ER600-2322, ER600-2323, ER600-2324, ER600-2325, ER600-2326, ER600-2327, ER600-2328, ER600-2329, ER600-2330, ER600-2331, ER600-2332, ER600-2333, ER600-2334, ER600-2335, ER600-2336, ER600-2337, ER600-2338, ER600-2339, ER600-2340, ER600-2341, ER600-2342, ER600-2343, ER600-2344, ER600-2345, ER600-2346, ER600-2347, ER600-2348, ER600-2349, ER600-2350, ER600-2351, ER600-2352, ER600-2353, ER600-2354, ER600-2355, ER600-2356, ER600-2357, ER600-2358, ER600-2359, ER600-2421, ER600-2422, ER600-2423, ER600-2424, ER600-2425, ER600-2426, ER600-2427, ER600-2428, ER600-2429, ER600-2430, ER600-2431, ER600-2432, ER600-2433, ER600-2434, ER600-2435, ER600-2436, ER600-2437, ER600-2438, ER600-2448, ER600-2449, ER600-2450, ER600-2451, ER600-2452, ER600-2453, ER600-2454, ER600-2455, ER600-2456, ER600-2457, ER600-2458, ER600-2459, ER600-2460, ER600-2461, ER600-2462, ER600-2463, ER600-2464, ER600-2465, ER600-2466, ER600-2467, ER600-2468, ER600-2469, ER600-2492, ER600-2493, ER600-2494, ER600-2495, ER600-2497, ER600-2498, ER600-2499, ER600-2500, ER600-2501, ER600-2502, ER600-2503, ER600-2504, ER600-2505, ER600-2506, ER600-2507, ER600-2508, ER600-2509, ER600-2510, ER600-2511, ER600-2512, ER600-2513, ER600-2514, ER600-2515, ER600-2516, ER600-2517, ER600-2518, ER600-2519, ER600-2520, ER600-2521, ER600-2522, ER600-2523, ER600-2524, ER600-2525, ER600-2526, ER600-2527, ER600-2528, ER600-2529, ER600-2530, ER600-2531, ER600-2532, ER600-2533, ER600-2534, ER600-2535, ER600-2536, ER600-2537, ER600-2538, ER600-2539, ER600-2540, ER600-2541, ER600-2542, ER600-2543, ER600-2544, ER600-2545, ER600-2546, ER600-2547, ER600-2548, ER600-2549, ER600-2550, ER600-2551, ER600-2552, ER600-2553, ER600-2554, ER600-2555, ER600-2556, ER600-2557, ER600-2558, ER600-2559, ER600-2560, ER600-2561, ER600-2562, ER600-2563, ER600-2564, ER600-2565, ER600-2566, ER600-2567, ER600-2568, ER600-2569, ER600-2570, ER600-2571, ER600-2572, ER600-2573, ER600-2574, ER600-2575, ER600-2576, ER600-2577, ER600-2578, ER600-2579, ER600-2580, ER600-2581, ER600-2582, ER600-2583, ER600-2584, ER600-2585, ER600-2586, ER600-2587, ER600-2588, ER600-2589, ER600-2590, ER600-2591, ER600-2592, ER600-2593, ER600-2594, ER600-2595, ER600-2596, ER600-2597, ER600-2598, ER600-2599, ER600-2600, ER600-2601, ER600-2602, ER600-2603, ER600-2604, ER600-2605, ER600-2606, ER600-2607, ER600-2608, ER600-2609, ER600-2610, ER600-2611, ER600-2612, ER600-2613, ER600-2614, ER600-2615, ER600-2616, ER600-2617, ER600-2618, ER600-2619, ER600-2620 ,ER600-2621, ER600-2622, ER600-2623, ER600-2624, ER600-2625, ER600-2626, ER600-2627, ER600-2628, ER600-2629, ER600-2630, ER600-2631, ER600-2632, ER600-2633, ER600-2634, ER600-2635, ER600-2636, ER600-2764, ER600-2765, ER600-2766, ER600-2767, ER600-2768, ER600-2769, ER600-2770, ER600-2771, ER600-2772, ER600-2773, ER600-2774, ER600-2775, ER600-2776, ER600-2777, ER600-2778, ER600-2779, ER600-2780, ER600-2781, ER600-2782, ER600-2783, ER600-2784, ER600-2785, ER600-2786, ER600-2787, ER600-2788, ER600-2789, ER600-2790, ER600-2791, ER600-2792, ER600-2793, ER600-2794, ER600-2795, ER600-2818, ER600-2819, ER600-2820, ER600-2821, ER600-2822, ER600-2823, ER600-2824, ER600-2825, ER600-2826, ER600-2827, ER600-2838, ER600-2839, ER600-2840, ER600-2841, ER600-2842, ER600-2843, ER600-2844, ER600-2845, ER600-2846, ER600-2847, ER600-2848, ER600-2849, ER600-2850, ER600-2851, ER600-2852, ER600-2853, ER600-2854, ER600-2855, ER600-2856, ER600-2857, ER600-2858, ER600-2859, ER600-2882, ER600-2883, ER600-2884, ER600-2885, ER600-2886, ER600-2887, ER600-2888, ER600-2889, ER600-2910, ER600-2911, ER600-2912, ER600-2913, ER600-2914, ER600-2915, ER600-2916, ER600-2917, ER600-2918, ER600-2919, ER600-2920, and ER600-2921 |
Recalling Firm/ Manufacturer |
Arjo, Inc. 50 Gary Ave Ste A Roselle IL 60172-1684
|
For Additional Information Contact |
Ms. Traci Giovenco 630-307-2756
|
Manufacturer Reason for Recall |
Unintended Hanger Bar Detachment: The locknut on the scales may loosen during use and allow the hanger bar assembly of the patient lift to detach and fall.
|
FDA Determined Cause 2 |
Device Design |
Action |
Arjo sent Urgent Device Field Correction Notification letters dated 12/19/07 to the end user accounts who received the affected scales, advising them of the potential for the hanger bar assembly to detach. The accounts were given the option immediately stopping use of the lifts with the affected scales or removing the scales from the lift so the lift can continue to be used. Instructions for removal of the scales were provided with the letter. The accounts were requested to complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter, and indicating whether they will complete the correction upon receipt of the correction parts or if they require the assistance of an Arjo Service Technician. |
Quantity in Commerce |
772 scales |
Distribution |
Nationwide including the states of Alabama, Arizona, California, Delaware, Florida, Hawaii, Iowa, Illinois, Indiana, Massachusetts, Michigan, Minnesota, Mississippi, North Carolina, North Dakota, New Hampshire, New Jersey, New York, Ohio, Pennsylvania, South Dakota, Tennessee, Washington and Wisconsin. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|