  Class 2 Device Recall GTIC Globe Trotter IC see related information
Date Initiated by Firm January 31, 2008
Date Posted March 18, 2008
Recall Status1 Terminated 3 on May 08, 2008
Recall Number Z-1053-2008
Recall Event ID 46742
510(K)Number K001019  
Product Classification Neonatal Incubator - Product Code FMZ
Product Draeger Globe Trotter IC, Catalog Number: MU20509, Draeger Medical, Inc., Telford, PA 18969
Code Information Serial Numbers: QT05675, QT05676, ST05708, and ST05704
Recalling Firm/
Manufacturer
Draeger Medical, Inc.
3135 Quarry Rd
Telford PA 18969-1042
For Additional Information Contact Kathy Anderson
215-721-5400
Manufacturer Reason
for Recall
Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
FDA Determined
Cause 2
Nonconforming Material/Component
Action The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.
Quantity in Commerce 4 units
Distribution Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = FMZ and Original Applicant = HILL-ROM AIR-SHIELDS
-
-