|
Class 2 Device Recall GTIC Globe Trotter IC |
|
Date Initiated by Firm |
January 31, 2008 |
Date Posted |
March 18, 2008 |
Recall Status1 |
Terminated 3 on May 08, 2008 |
Recall Number |
Z-1053-2008 |
Recall Event ID |
46742 |
510(K)Number |
K001019
|
Product Classification |
Neonatal Incubator - Product Code FMZ
|
Product |
Draeger Globe Trotter IC, Catalog Number: MU20509, Draeger Medical, Inc., Telford, PA 18969 |
Code Information |
Serial Numbers: QT05675, QT05676, ST05708, and ST05704 |
Recalling Firm/ Manufacturer |
Draeger Medical, Inc. 3135 Quarry Rd Telford PA 18969-1042
|
For Additional Information Contact |
Kathy Anderson 215-721-5400
|
Manufacturer Reason for Recall |
Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com. |
Quantity in Commerce |
4 units |
Distribution |
Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FMZ and Original Applicant = HILL-ROM AIR-SHIELDS
|
|
|
|