| | Class 2 Recall C1160 Universal Flexible Processing Tray |  |
| Date Posted |
August 05, 2008 |
| Recall Number |
Z-1190-2008 |
| Product |
Steris C1160 Universal Flexible Processing Tray, for use with STERIS SYSTEM 1 Processor, used for sterilization of flexible endoscopes. |
| Code Information |
All codes
|
Recalling Firm/ Manufacturer |
Steris Corp 5960 Heisley Rd Mentor, Ohio 44060-1834 |
| Consumer Instructions |
Contact the recalling firm for information |
| For Additional Information Contact |
Holly Wright Lee 440-392-7019
|
Reason for Recall |
Failure to effectively sterilize: Under certain conditions, the design/operation of the C1160 Tray may compromise the ability of the SYSTEM 1 processor to correctly assess the state of the high pressure pump during the Sterile Cycle. The L6 sensor fails to alarm when the high pressure pump is not working properly.
|
| Action |
Steris sent an Urgent Safety Alert/Recall letter dated 2/8/2008, to all customers. The letter stated that any device sterilized in the C1160 Tray should not be used clinically prior to a completion of a successful diagnostic cycle. Until further notice, you must run a diagnostic cycle after each sterile cycle involving a C1160 tray or alternately you must discontinue use of these trays. Steris service technicians will schedule visits to correct the problem. |
| Quantity in Commerce |
28,766 units |
| Distribution |
Worldwide Distribution, including USA, Puerto Rico, Canada, and Japan. |
| |
|
|
|
|