| | Class 3 Recall Hill-Rom Stretcher; Model P8000F |  |
| Date Posted |
September 16, 2008 |
| Recall Number |
Z-1655-2008 |
| Product |
Hill-Rom Procedural Stretcher with auto-contour option; Model P8000F. Wheeled hospital stretcher consisting of a platform mounted on a wheeled frame, designed to transport patients and to allow for patient care before, during and after transport. |
| Code Information |
All units with serial numbers H352AN3566 through I222AN9588 and shipped prior to 8/13/07.
|
Recalling Firm/ Manufacturer |
Hill-Rom, Inc. 125 E Pearl St Batesville, Indiana 47006 |
Reason for Recall |
The auto contour function on/off handle may be inadvertently activated while the head of the stretcher is raised, increasing the possibility of the head section to become jammed and preventing it from lowering. This may result in a delay to the procedure.
|
| Action |
Consignees will be visited to correct the problem and notified via a hand delivered Device Modification letter dated 4/3/08 at the time of the visit. If you have any questions please contact Jeffrey C. Shaul, Director, RA/QA Acute Care at (812) 931-2604. |
| Quantity in Commerce |
564 |
| Distribution |
Worldwide Distribution --- including USA and countries of Australia, Belgium, Canada, France, Germany, Israel, Italy, Netherlands, Spain, Switzerland and United Kingdom. |
| |
|
|
|
|