• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Zimmer Hemovac Wound Drainage Device

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Zimmer Hemovac Wound Drainage Device see related information
Date Posted September 19, 2008
Recall Status1 Terminated on November 12, 2009
Recall Number Z-1487-2008
Recall Event ID 47233
Product Classification non-powered, portable suction appartus for a single patient - Product Code GCY
Product Zimmer Hemovac Wound Drainage Device, 400ml Comp Evac 3/16" PVC w/trocar, 4,8mm, 15Fr, PVC, 107cm, soft, sterile, latex free, Rx only; Zimmer, Swindon, Wiltshire, United Kingdom; Zimmer, Dover, Ohio; REF 00-2505-000-10.
Code Information All lots manufactured on or before 1/18/08. Product has a 5 year expiration period.
Recalling Firm/
Manufacturer
Zimmer Inc.
345 E Main St
Warsaw IN 46580-2746
For Additional Information Contact
330-365-3793
Manufacturer Reason
for Recall
The sterility of the device may be compromised due to the possible presence of slits or pinholes in the packaging.
FDA Determined
Cause 2
Packaging process control
Action Consignees were notified via recall letter dated 1/29/08 to locate and return the affected product. The recall action was expanded to additional products/lots via recall letter dated 3/17/08. Custom kit packers were instructed to notify their customers.
Distribution Nationwide, Australia, Canada, Chile, China, Czech Republic, Dubai, Ecuador, France, Germany, Greece, Hungary, Italy, Japan, Korea, Poland, Romania, Russia, Saudi Arabia, Serbia & Montenegro, Singapore, Slovakia, South Africa, Spain, and Switzerland.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
-
-