|
Class 2 Device Recall Ultraview Digital Telemetry Transmitter |
|
Date Initiated by Firm |
July 18, 2008 |
Date Posted |
September 18, 2008 |
Recall Status1 |
Terminated 3 on March 21, 2011 |
Recall Number |
Z-1652-2008 |
Recall Event ID |
48985 |
510(K)Number |
k925510
|
Product Classification |
Telemetry Reciever - Product Code DSI
|
Product |
Ultraview Digital Telemetry Transmitter, Model 90478 The Spacelabs Medical Digital Telemetry system provides continuous electrocardiographic (ECG) monitoring to detect ST segment changes, abnormal cardiac rhythms, including asystole, ventricular fibrillation and ventricular tachycardia. |
Code Information |
Serial Numbers 478-111508, 478-111509, 478-111510, 478-111511, 478-111962, 478-111963, 478-111964, 478-111965
478-114831, 478-117016, 478-117395, 478-117602, 478-122332, 478-122333, 478-122535, 478-124039, 478-124040, 478-124042, 478-124055, 478-108923, 478-108924, 478-108925, 478-108926, 478-108927, 478-108928, 478-108929, 478-108930, 478-112419, 478-112420, 478-112421, 478-112422, 478-114943, 478-114944, 478-114945, 478-114946, 478-114947, 478-114948, 478-114949, 478-114950, 478-114951, 478-114952, 478-122698, 478-122704, 478-122711, 478-137602, 478-137603, 478-144927
478-122680, 478-122680, 478-122680, 478-122680, 478-128481, 478-128483, 478-128515, 478-128525, 478-112052, 478-113089, 478-113097, 478-113704, 470-110122, 478-110126, 478-110939, 478-110940, 478-113098, 478-113099, 478-113101, 478-120415, 478-122158, 478-122159, 478-122163, 478-124049, 478-124057, 478-124058, 478-124570, 478-124571, 478-124994, 478-124996, 478-113527, 478-113528, 478-113529, 478-113530, 478-113531, 478-113532, 478-113533, 478-113761, 478-111405, 478-111803, 478-111827, 478-111828, 478-111829, 478-111836, 478-111837, 478-111838, 478-111839, 478-111840, 478-111841, 478-111842, 478-111843, 478-111844, 478-111845, 478-111846, 478-111847, 478-111848, 478-111849, 478-111850, 478-111851, 478-111852, 478-111853, 478-111854, 478-111887, 478-111888, 478-111996, 478-112013, 478-112014, 478-112016, 478-118020, 478-118367, 478-119474, 478-119475, 478-122679, 478-122685, 478-122687, 478-122688, 478-122691, 478-122692, 478-124272, 478-124274, 478-124723, 478-124724, 478-124725, 478-124726, 478-124732, 478-124733, 478-124734, 478-124735, 478-124736, 478-124739, 478-124740, 478-124741, 478-124742, 478-124743, 478-124748, 478-124750, 478-124751, 478-124762, 478-127124, 478-127129, 478-127155, 478-127473, 478-127997, 478-128063, 478-128067, 478-128528, 478-114841, 478-114866, 478-114868, 478-114869, 478-114870, 478-114871, 478-114872, 478-114873, 478-105101, 478-105145, 478-105269, 478-105481, 478-105531, 478-105617, 478-105634, 478-106017, 478-110011, 478-110012, 478-110013, 478-110024, 478-110025, 478-111180, 478-115421, 478-113006, 478-113007, 478-113008, 478-113010, 478-115358, 478-117023, 478-117034, 478-117074, 478-118351, 478-118352, 478-122327, 478-122329, 478-123849, 478-123854, 478-123941, 478-123949, 478-128001
478-129153
478-120034, 478-120040, 478-120041, 478-120047, 478-106809, 478-106810, 478-105975, 478-110117, 478-110123, 478-115390, 478-115426, 478-116810, 478-116817, 478-116824, 478-117122, 478-117126, 478-117127, 478-118041, 478-118256, 478-124126, 478-124127, 478-124128, 478-124331, 478-125680, 478-125681, 478-125682, 478-128661, 478-128662, 478-128665, 478-128667, 478-129167, 478-129684, 478-129687, 478-129691
478-129949
478-132784, 478-132785, 478-127983, 478-127985, 478-127990, 478-127991, 478-127993, 478-127994, 478-127995, 478-128020, 478-107614, 478-107615, 478-107616, 478-107617, 478-107618, 478-107613, 478-100222, 478-100501, 478-105173, 478-105174, 478-108214, 478-108353, 478-108527, 478-108648, 478-128336, 478-128337, 478-128548, 478-128551, 478-128552, 478-128555, 478-128556, 478-128557, 478-127070, 478-127083, 478-128450, 478-128487, 478-110116, 478-110192, 478-110193, 478-110194, 478-110238, 478-130761, 478-131067, 478-131068, 478-131070, 478-131071, 478-131072, 478-130563, 478-130773, 478-130774, 478-130775, 478-130776, 478-130777, 478-130778, 478-130779, 478-130780, 478-130781, 478-130782, 478-131692, 478-131693, 478-132052, 478-132236, 478-132779, 478-132795, 478-132796, 478-132797, 478-132798, 478-132799, 478-132800, 478-132801, 478-132802, 478-134083, 478-134084, 478-134085, 478-134086, 478-134087, 478-134088, 478-134089, 478-134090, 478-134091, 478-131671, 478-132283, 478-132287, 478-132320, 478-132321, 478-132695, 478-132696, 478-132697, 478-132698, 478-132699, 478-132700, 478-132701, 478-132702, 478-132703, 478-132704, 478-132705, 478-132706, 478-132707, 478-132708, 478-132709, 478-132710, 478-132711, 478-132712, 478-132713, 478-132715, 478-132716, 478-132717, 478-132718, 478-132719, 478-132720, 478-132721, 478-132722, 478-132723, 478-132724, 478-132725, 478-132726, 478-132727, 478-132728, 478-132729, 478-133278, 478-133279, 478-133280, 478-133281, 478-133282, 478-133283, 478-133331, 478-133335, 478-133370, 478-133371, 478-133914, 478-133915, 478-133916, 478-133918, 478-133919, 478-133920, 478-133921, 478-133922, 478-133923, 478-133924, 478-133925, 478-133926, 478-133927, 478-133928, 478-133929, 478-133930, 478-133931, 478-133932, 478-133933, 478-133934, 478-133935, 478-133936, 478-133937, 478-133938, 478-133939, 478-133940, 478-133941, 478-133942, 478-133943, 478-133944, 478-133945, 478-133946, 478-133947, 478-133949, 478-134006, 478-134007, 478-134254, 478-134255, 478-130756, 478-130757, 478-130762, 478-130763, 478-130764, 478-133332, 478-133333, 478-133334, 478-133917, 478-133978, 478-133979, 478-133980, 478-133982, 478-130229, 478-130515, 478-130516, 478-130517, 478-130519, 478-132787, 478-132788, 478-132789, 478-139434, 478-132280, 478-132281, 478-132284, 478-132286, 478-132362, 478-132365, 478-132367, 478-132369, 478-132410, 478-132411, 478-132413, 478-132416, 478-132931, 478-132933, 478-132999, 478-133000, 478-133001, 478-133002, 478-133003, 478-133004, 478-133005, 478-133006, 478-133013, 478-133023, 478-133328, 478-133329, 478-133330, 478-134234, 478-134235, 478-134236, 478-134237, 478-134239, 478-134240, 478-134295, 478-134296, 478-134297, 478-134298, 478-134299, 478-134300, 478-134301, 478-134302, 478-134585, 478-134586, 478-134595, 478-134596, 478-134597, 478-134602, 478-134603, 478-134606, 478-136373, 478-136374, 478-136375, 478-136376, 478-136377, 478-136378, 478-136379, 478-136380, 478-136388, 478-136389, 478-136390, 478-136391, 478-136392, 478-136393, 478-136395, 478-136396, 478-136401, 478-136402, 478-136403, 478-136404, 478-136405, 478-136406, 478-136407, 478-136408, 478-136633, 478-136634, 478-136637, 478-136638, 478-109950, 478-124986, 478-124989, 478-125008, 478-125631, 478-125632, 478-141804, 478-141805, 478-141806, 478-141807, 478-141808, 478-141809, 478-141810, 478-141811, 478-141812, 478-141813, 478-141814, 478-141816, 478-143399, 478-143400, 478-143401, 478-143402, 478-146839, 478-146840, 478-146841, 478-146842, 478-146844, 478-146845, 478-146846, 478-146847, 478-146848, 478-146849, 478-146850, 478-146851
478-110624, 478-110968, 478-113052, 478-113053, 478-113090, 478-113091, 478-113092, 478-113093, 478-113094, 478-113111, 478-113112, 478-115377, 478-115395, 478-115397, 478-115400, 478-115402, 478-115404, 478-115405, 478-115417, 478-115418, 478-115419, 478-115420, 478-115422
478-113599, 478-113600, 478-117388, 478-117391, 478-118780, 478-118781, 478-118784, 478-118785, 478-122162, 478-122164
478-108544, 478-108545, 478-111549, 478-111550, 478-111552, 478-111555, 478-111557, 478-111558, 478-111559, 478-111560, 478-144585
478-117158, 478-117804, 478-117805, 478-117806, 478-117807, 478-117808, 478-117809, 478-117810, 478-117811, 478-122542, 478-122543
478-130293
478-111709, 478-111858, 478-111859, 478-111860, 478-111861, 478-111862, 478-115005, 478-115334, 478-116340, 478-117533, 478-117534, 478-117535, 478-117536, 478-117537, 478-117538, 478-117764
478-117134, 478-118259, 478-118260, 478-119686, 478-122224, 478-122225, 478-122226, 478-122231
478-125078, 478-128335, 478-128340
478-130301, 478-130882, 478-132238, 478-132248, 478-132268, 478-132282, 478-132285
478-109902, 478-109904, 478-109905, 478-109906, 478-109935, 478-109936, 478-110120, 478-110121, 478-110318, 478-110319, 478-111042, 478-111043
478-129946, 478-129947, 478-129951, 478-130520
478-128333, 478-128545, 478-128546, 478-128547, 478-129165, 478-129168, 478-129169, 478-129172
478-108772, 478-108777, 478-108789, 478-108931, 478-108932, 478-108933, 478-108935, 478-108936, 478-108937, 478-108938, 478-108939, 478-111823, 478-111824, 478-111825, 478-111826, 478-115437, 478-115438, 478-111894, 478-111895, 478-111896, 478-111897, 478-142240, 478-142241, 478-142242, 478-142243, 478-143973, 478-143978, 478-146906, 478-146907, 478-146908, 478-146910
478-106050, 478-106063
478-110203, 478-110450, 478-110458, 478-110598, 478-115844, 478-116162, 478-116163, 478-116271, 478-122862, 478-122873, 478-122875, 478-122876, 478-136422, 478-136423, 478-136424, 478-136425, 478-136426, 478-136427, 478-136428, 478-136431, 478-145278, 478-145322, 478-145323, 478-145552, 478-146101, 478-146167, 478-146952, 478-147043, 478-147070, 478-147153, 478-147164, 478-147165
478-112023, 478-113451, 478-113452, 478-113453, 478-113455, 478-113456, 478-113457, 478-113458, 478-113459, 478-119358, 478-119359, 478-122700, 478-122701
478-128338, 478-128339, 478-128341, 478-128342
478-112793, 478-112794, 478-116378, 478-124853, 478-126549, 478-126551, 478-126553, 478-126556, 478-126557, 478-126856, 478-126862, 478-126870, 478-126894
478-115231, 478-115234, 478-115287, 478-115293, 478-115294, 478-115295, 478-115296, 478-115351, 478-117720, 478-132714, 478-132732, 478-133975, 478-133976
478-125684, 478-125687
478-116797, 478-116798
478-145297, 478-145298, 478-145299, 478-145300, 478-145301
478-110493, 478-112048, 478-113096, 478-114216, 478-121458, 478-121459, 478-122161, 478-122165, 478-132730, 478-132731, 478-132733, 478-132734, 478-133326, 478-133327
478-141815, 478-142590, 478-142602
478-113727, 478-113795, 478-113796, 478-113798, 478-129381, 478-129796, 478-129797, 478-129883, 478-129884, 478-129885, 478-129886, 478-129887, 478-129888, 478-129889, 478-129890, 478-129892, 478-129355, 478-129444, 478-105999, 478-130223, 478-109741, 478-109763, 478-109773, 478-109774, 478-109775, 478-109776, 478-109777, 478-109778, 478-111498, 478-111612, 478-111613, 478-111614, 478-111615, 478-111617, 478-111657, 478-111702, 478-111705, 478-111733, 478-112260, 478-112262, 478-113287, 478-113535, 478-113536, 478-113537, 478-113538, 478-113539, 478-113540, 478-113541, 478-113542, 478-113543, 478-113544, 478-113545, 478-113546, 478-113547, 478-113549, 478-113550, 478-113551, 478-113552, 478-113553, 478-113554, 478-113555, 478-113556, 478-113557, 478-113558, 478-113559, 478-113560, 478-113561, 478-113562, 478-113563, 478-113564, 478-113565, 478-113566, 478-113567, 478-113568, 478-113569, 478-113570, 478-113571, 478-113572, 478-115448, 478-115471, 478-115473, 478-115474, 478-115475, 478-115476, 478-115477, 478-115478, 478-115479, 478-115480, 478-115481, 478-115482, 478-115484, 478-115485, 478-115486, 478-115487, 478-116892, 478-117203, 478-117204, 478-117205, 478-117264, 478-117306, 478-117307, 478-117308, 478-117309, 478-117542, 478-117913, 478-117914, 478-117915, 478-117916, 478-117917, 478-117918, 478-117919, 478-117920, 478-118563, 478-118564, 478-118565, 478-118566, 478-118567, 478-118568, 478-118569, 478-118570, 478-118571, 478-118572, 478-118573, 478-118574, 478-118575, 478-118576, 478-118577, 478-118578, 478-118579, 478-118580, 478-118581, 478-118582, 478-118855, 478-118948, 478-118949, 478-118950, 478-118951, 478-119098, 478-119317, 478-119335, 478-119336, 478-119341, 478-119342, 478-119355, 478-120247, 478-120248, 478-120249, 478-120279, 478-120512, 478-120515, 478-120516, 478-120517, 478-120518, 478-120519, 478-120521, 478-120522, 478-120523, 478-120524, 478-120606, 478-120673, 478-120676, 478-120739, 478-120740, 478-120741, 478-120745, 478-120806, 478-120840, 478-120861, 478-120879, 478-120883, 478-120904, 478-120905, 478-120921, 478-122061, 478-122117, 478-122308, 478-122557, 478-122569, 478-126539, 478-126543, 478-126545, 478-126602, 478-126608, 478-126619, 478-126635, 478-126642, 478-126655, 478-126659, 478-126844, 478-126886, 478-126945, 478-127025, 478-127096, 478-127111, 478-127112, 478-127120, 478-127136, 478-127432, 478-127442, 478-127486, 478-127487, 478-127488, 478-127489, 478-127490, 478-127492, 478-127493, 478-127494, 478-127495, 478-127496, 478-127497, 478-132775, 478-132776, 478-132777, 478-132778, 478-132792, 478-132793, 478-132794, 478-133203, 478-133204, 478-133205, 478-133206, 478-133207, 478-133208, 478-133209, 478-133210, 478-133211, 478-133212, 478-133213, 478-133214, 478-133215, 478-133948, 478-134001, 478-134002, 478-134005, 478-134256, 478-134257, 478-134258, 478-134259, 478-134261, 478-134262, 478-135146, 478-135147, 478-135148, 478-135182, 478-135183, 478-135184, 478-135191, 478-135192, 478-135213, 478-135214, 478-136965, 478-136966, 478-136967, 478-136968, 478-136969, 478-137544, 478-137548, 478-137556, 478-137557, 478-137558, 478-137559, 478-137560, 478-137630, 478-138697, 478-138698, 478-138699, 478-138700, 478-140092, 478-140110, 478-140122, 478-140123, 478-140124, 478-140125, 478-140126, 478-140127, 478-140128, 478-140129, 478-140132, 478-140133, 478-140134, 478-140135, 478-140137, 478-140138, 478-140139, 478-140140, 478-140141, 478-140142, 478-140143, 478-140144, 478-140145, 478-140146, 478-140540, 478-140541, 478-140542, 478-140543, 478-140544, 478-140545, 478-140546, 478-140547, 478-140548, 478-140549, 478-140550, 478-140551, 478-143935, 478-144477, 478-144583, 478-144584, 478-144586, 478-144587, 478-144588, 478-144592, 478-144596, 478-144609, 478-144616, 478-144617, 478-144618, 478-144619, 478-144620, 478-144621, 478-144622, 478-144624, 478-144625, 478-144626, 478-144627, 478-144629, 478-144630, 478-144631, 478-144632, 478-144633, 478-144634, 478-144635, 478-144636, 478-144637, 478-144638, 478-144639, 478-144640, 478-144641, 478-144642, 478-144643, 478-144644, 478-144645, 478-144646, 478-144647, 478-144648, 478-144649, 478-144650, 478-144671, 478-144672, 478-144673, 478-144674, 478-144675, 478-144676, 478-144677, 478-144678, 478-144679, 478-144680, 478-144681, 478-144682, 478-144683, 478-144684, 478-144685, 478-144686, 478-144687, 478-144688, 478-144689, 478-144690, 478-144691, 478-144692, 478-144693, 478-144695, 478-144709, 478-144710, 478-144711, 478-144712, 478-144832, 478-144843, 478-144844, 478-144845, 478-144846, 478-144847, 478-144848, 478-144849, 478-146254, 478-146255, 478-146888, 478-146889, 478-146890, 478-146891, 478-146892, 478-146893, 478-146894, 478-146895, 478-146896, 478-146897, 478-146898, 478-146899, 478-146900, 478-146901, 478-147191, 478-147192, 478-147197, 478-147198, 478-147212, 478-147261, 478-147262, 478-147265, 478-147266, 478-147267, 478-147268, 478-147269, 478-147270, 478-147271
478-112404
478-129943, 478-129944, 478-129945, 478-129948, 478-129952, 478-130231, 478-130513, 478-130518
478-145035, 478-145363, 478-145616, 478-145617, 478-146203
478-106144, 478-110402, 478-110403, 478-110404, 478-110405, 478-112022, 478-112025, 478-112491
478-110920, 478-110921, 478-110975, 478-110977, 478-110978, 478-110979, 478-110983, 478-111004, 478-113700, 478-113701, 478-113702, 478-113703, 478-115398, 478-115399, 478-115401, 478-115407, 478-115408, 478-115409, 478-115410, 478-115411, 478-115412, 478-115413, 478-115414, 478-115415, 478-115416, 478-124997, 478-124998, 478-125001, 478-125004, 478-143367, 478-143368, 478-143369, 478-143370
478-110494, 478-110934, 478-110935, 478-127065
478-136878, 478-136879, 478-136880, 478-136881, 478-137222, 478-137223, 478-137224, 478-137561, 478-137562, 478-137563, 478-137564, 478-137565, 478-137566, 478-137567, 478-137568, 478-137569, 478-137570, 478-137571, 478-137572, 478-137573, 478-137574, 478-137575, 478-138415, 478-138717, 478-138718, 478-138719, 478-139025, 478-139026, 478-139901, 478-139902, 478-140272, 478-140285, 478-140288, 478-140289, 478-140290, 478-140291, 478-140292, 478-140293, 478-140294, 478-140295, 478-140296, 478-140297, 478-140298, 478-140299, 478-140300, 478-140301, 478-140302, 478-140303, 478-140304, 478-140305, 478-140306, 478-140307, 478-140308, 478-140309, 478-140310, 478-140311, 478-140349, 478-140350, 478-140351, 478-140352, 478-140353, 478-140355, 478-140358, 478-140359, 478-140360, 478-140361, 478-140362, 478-140363, 478-140364, 478-140410, 478-140414, 478-140418, 478-140432, 478-140460, 478-140464, 478-140469, 478-140471, 478-140476, 478-140477, 478-140478, 478-140479, 478-140480, 478-140482, 478-140483, 478-140484, 478-140485, 478-140486, 478-140488, 478-140489, 478-140490, 478-140492, 478-140493, 478-140494, 478-140495, 478-140709, 478-142624, 478-142659, 478-142747, 478-143259, 478-143260, 478-143261, 478-143262, 478-143263, 478-143264, 478-143265, 478-143266, 478-143267, 478-143268, 478-143269, 478-143270, 478-143685, 478-143686, 478-143930, 478-143931, 478-143932, 478-143933, 478-143934, 478-144262, 478-144472, 478-144666, 478-144667, 478-144970, 478-144994, 478-144995, 478-144996, 478-144997, 478-144998, 478-144999, 478-145000, 478-145001, 478-145002, 478-145003, 478-145004, 478-145048, 478-145207, 478-145357, 478-145358, 478-145359, 478-145364, 478-145365, 478-145366, 478-145367, 478-145368, 478-145369, 478-145370, 478-145533, 478-145534, 478-145535, 478-145536, 478-145538, 478-145539, 478-145540, 478-145541, 478-147188, 478-147189
478-146936
478-118777, 478-118778, 478-118900, 478-118903, 478-118904, 478-118905, 478-118906, 478-118907
478-108843
478-127071, 478-127466
478-122541, 478-122669, 478-122670, 478-123084, 478-123086, 478-123088, 478-123846, 478-123850, 478-123851, 478-123852, 478-123853, 478-123943
478-129143, 478-129144, 478-129145, 478-129146, 478-129148, 478-129149, 478-129155, 478-129156, 478-129158, 478-129162, 478-130567, 478-130570, 478-130571, 478-130572
478-126528
478-128769, 478-128770, 478-128771, 478-128772
478-130566
478-105175, 478-107427, 478-107428, 478-107429, 478-107430, 478-109603, 478-109604, 478-110336, 478-110347, 478-110348
478-145321, 478-145360, 478-145361, 478-145362
478-142405
478-125627, 478-125628, 478-125629, 478-125630
478-129548, 478-129550
478-147585, 478-147586, 478-147587, 478-147588
478-120422, 478-124054, 478-124121, 478-124122, 478-124123, 478-124124, 478-124125, 478-124332, 478-124333, 478-124334, 478-124470, 478-124471, 478-124472, 478-124473, 478-124474, 478-124476, 478-124563, 478-124567, 478-124569, 478-128550, 478-128553, 478-128554
478-111119
478-100466, 478-107417, 478-107418, 478-107420, 478-107426, 478-107790, 478-107791, 478-107792, 478-107793, 478-107843, 478-108790, 478-108794, 478-108799, 478-108814, 478-108815, 478-108840, 478-108841, 478-108876, 478-108877, 478-108878, 478-119096, 478-119333, 478-120423, 478-121450, 478-121454, 478-122166, 478-124475, 478-124477, 478-124479, 478-124565, 478-124566, 478-128024, 478-128334, 478-129689
478-130759, 478-140374, 478-140710, 478-140731, 478-140732, 478-141143, 478-141144, 478-141145, 478-141146, 478-141578, 478-141579, 478-141580, 478-141581, 478-141582, 478-141583, 478-141584, 478-141585, 478-143624, 478-143625, 478-143626, 478-143627, 478-143628, 478-143629, 478-143630, 478-143631, 478-143632, 478-143633, 478-145637, 478-145638, 478-145639, 478-145640, 478-145641, 478-145642, 478-145643, 478-145644, 478-145645, 478-145646, 478-146919, 478-146920, 478-146921, 478-146922, 478-146923, 478-146924, 478-146925, 478-146926, 478-146927, 478-146928
478-111041, 478-125079, 478-125083, 478-125084, 478-125666, 478-125683, 478-125685, 478-125686, 478-125688, 478-125689, 478-126674, 478-126675, 478-126678, 478-128549, 478-128658, 478-128659, 478-128660, 478-128663, 478-128664, 478-129163, 478-129164, 478-129166, 478-129170, 478-129171, 478-129683, 478-129685, 478-129686, 478-129688, 478-129690, 478-129692
478-122109, 478-126251, 478-126262, 478-126270, 478-126271, 478-126275, 478-126280, 478-126282, 478-126284, 478-126285
478-128768, 478-129147, 478-129150, 478-129151
478-130224, 478-130225
478-144421, 478-144422, 478-145026, 478-145028, 478-145029, 478-145030, 478-145031, 478-145032, 478-145033, 478-145034, 478-145036, 478-145037, 478-145038, 478-145039, 478-145040
478-125011, 478-125012, 478-125013, 478-125014, 478-128873, 478-128874, 478-128875, 478-128876, 478-128877, 478-128878, 478-128879, 478-128880, 478-128881, 478-128882, 478-128883, 478-128884, 478-128885, 478-128886, 478-128887, 478-128888, 478-128889, 478-128890, 478-128891, 478-128892, 478-129314, 478-129679, 478-129814, 478-129821, 478-133989, 478-133990, 478-133991, 478-133992, 478-133993, 478-133994, 478-133995, 478-133996, 478-133997, 478-133998, 478-133999, 478-134000, 478-138407, 478-138408, 478-138409, 478-138410, 478-138411, 478-138412, 478-138413, 478-138414, 478-139033, 478-139035, 478-139037, 478-139038, 478-139039, 478-139040, 478-139041, 478-139042, 478-139056, 478-139057, 478-139157, 478-139158, 478-139160, 478-139161, 478-139162, 478-139163, 478-139165, 478-139166, 478-139167, 478-139168, 478-139169, 478-139170, 478-139171, 478-139172, 478-139173, 478-139174, 478-139175, 478-139176, 478-139178, 478-139179, 478-139180, 478-139181, 478-139190, 478-139191, 478-139192, 478-139193, 478-139194, 478-139195, 478-139196, 478-139197, 478-139198, 478-139199, 478-139200, 478-139201, 478-139202, 478-139203, 478-139204, 478-139205, 478-139206, 478-139207, 478-139208, 478-139209, 478-139210, 478-139211, 478-139212, 478-139213, 478-139214, 478-139215, 478-139216, 478-139217, 478-139218, 478-139219, 478-139220, 478-139221, 478-141989, 478-142793, 478-142806, 478-142807, 478-142808, 478-142816, 478-142831, 478-143576, 478-143579, 478-143580, 478-143583, 478-143714, 478-143715, 478-143716, 478-143717, 478-143718, 478-143719, 478-143720, 478-143721, 478-143722, 478-144786, 478-144787, 478-144788, 478-144790, 478-144791, 478-144792, 478-144793, 478-144796
478-125077, 478-125633, 478-125634, 478-125635, 478-125636
478-110118, 478-110119
478-112429, 478-112809, 478-112810, 478-112811, 478-112812, 478-112884, 478-112984, 478-112985, 478-112986, 478-112987, 478-112988, 478-112989, 478-116703, 478-118097, 478-118098
478-108601, 478-108695, 478-108696, 478-108697, 478-110931, 478-110957, 478-111007, 478-111008, 478-114058, 478-114059, 478-114060, 478-114061, 478-114095, 478-114096, 478-114097, 478-114098, 478-115216, 478-115218, 478-115222, 478-115224, 478-115227, 478-115228, 478-115275, 478-115938, 478-115939, 478-115940, 478-116075, 478-116076, 478-116077, 478-116242, 478-116649, 478-116722, 478-116897, 478-117437, 478-117628, 478-117629, 478-117630, 478-117631, 478-117632, 478-117633, 478-117634, 478-117635, 478-117636, 478-118839, 478-118844, 478-118845, 478-119976, 478-119979, 478-119980, 478-120667, 478-120668, 478-120675, 478-120677, 478-120678, 478-122686, 478-122694
478-126537, 478-126541, 478-126542, 478-126544
478-100256, 478-100261, 478-100349, 478-100361, 478-100364, 478-101426, 478-101445, 478-105355, 478-106408, 478-106413, 478-106414, 478-106419, 478-106852, 478-106891, 478-106896, 478-106897, 478-107050, 478-107083, 478-107084, 478-107085, 478-107133, 478-107134, 478-107379, 478-107380, 478-107381, 478-107382, 478-107385, 478-107386, 478-107387, 478-107389, 478-108140, 478-108593, 478-108614, 478-108616, 478-108771, 478-109362, 478-109364, 478-109365, 478-109383, 478-109384, 478-109385, 478-109927, 478-109928, 478-110663, 478-110693, 478-110743, 478-110748, 478-110786, 478-110787, 478-110788, 478-110791, 478-110793, 478-110794, 478-110795, 478-110796, 478-110824, 478-110825, 478-110829, 478-110830, 478-110901, 478-110902, 478-110903, 478-110904, 478-111065, 478-111066, 478-111067, 478-111068, 478-111139, 478-111140, 478-111141, 478-111142, 478-111262, 478-111263, 478-111286, 478-111296, 478-111611, 478-111921, 478-111922, 478-111923, 478-112423, 478-112424, 478-112425, 478-112523, 478-112528, 478-112529, 478-112530, 478-112531, 478-112532, 478-112533, 478-112534, 478-112535, 478-112655, 478-112683, 478-112684, 478-112685, 478-112686, 478-112687, 478-112795, 478-112796, 478-113212, 478-113213, 478-113214, 478-113215, 478-113293, 478-113294, 478-113295, 478-113296, 478-113393, 478-113394, 478-113395, 478-113396, 478-114377, 478-114378, 478-114510, 478-114511, 478-114779, 478-114780, 478-114781, 478-114782, 478-114783, 478-114784, 478-114786, 478-114787, 478-114899, 478-114903, 478-114905, 478-114906, 478-114922, 478-114924, 478-114926, 478-114927, 478-114928, 478-114929, 478-114930, 478-114931, 478-114932, 478-114933, 478-114934, 478-114935, 478-115289, 478-115290, 478-115291, 478-115292, 478-115428, 478-115454, 478-115455, 478-115456, 478-115457, 478-115458, 478-115461, 478-115462, 478-115604, 478-115605, 478-115629, 478-115630, 478-115763, 478-115783, 478-116093, 478-116094, 478-116095, 478-116390, 478-116457, 478-116546, 478-116847, 478-116854, 478-116937, 478-116940, 478-116976, 478-116981, 478-117206, 478-117207, 478-117208, 478-117263, 478-117275, 478-117279, 478-117280, 478-117281, 478-117282, 478-117283, 478-117284, 478-117285, 478-117286, 478-117693, 478-117694, 478-117700, 478-117701, 478-117702, 478-117703, 478-117711, 478-117879, 478-117929, 478-117933, 478-117934, 478-117935, 478-117936, 478-117952, 478-118000, 478-118001, 478-118002, 478-118003, 478-118004, 478-118006, 478-118009, 478-118067, 478-118068, 478-118069, 478-118070, 478-118085, 478-118086, 478-118087, 478-118088, 478-118466, 478-118467, 478-118468, 478-118469, 478-118470, 478-118471, 478-118846, 478-118849, 478-118850, 478-118851, 478-118852, 478-118853, 478-119049, 478-119055, 478-119067, 478-119068, 478-119200, 478-119201, 478-119204, 478-119205, 478-119206, 478-119207, 478-119208, 478-119209, 478-119210, 478-119212, 478-119220, 478-119223, 478-119421, 478-119422, 478-119457, 478-119462, 478-119463, 478-119478, 478-119480, 478-119483, 478-119485, 478-119486, 478-119487, 478-119488, 478-119489, 478-119493, 478-119712, 478-119739, 478-119888, 478-119897, 478-119917, 478-119928, 478-119929, 478-119931, 478-119932, 478-119933, 478-119936, 478-119939, 478-119940, 478-119941, 478-119951, 478-119952, 478-119953, 478-119954, 478-119955, 478-119956, 478-119957, 478-119958, 478-119968, 478-119969, 478-119970, 478-119971, 478-119972, 478-119973, 478-119974, 478-119975, 478-119977, 478-119978, 478-119981, 478-119983, 478-119994, 478-119995, 478-119996, 478-120008, 478-120012, 478-120013, 478-120016, 478-120017, 478-120018, 478-120021, 478-120030, 478-120456, 478-120461, 478-120467, 478-120480, 478-120481, 478-120483, 478-120484, 478-120485, 478-120497, 478-120498, 478-120500, 478-120572, 478-120578, 478-120582, 478-120583, 478-120584, 478-120585, 478-120586, 478-120587, 478-120588, 478-120589, 478-120590, 478-120591, 478-120592, 478-120593, 478-120594, 478-120595, 478-120596, 478-120597, 478-120622, 478-120623, 478-120624, 478-120625, 478-120626, 478-120627, 478-120634, 478-120635, 478-120636, 478-120637, 478-120644, 478-120725, 478-120755, 478-120759, 478-120772, 478-120773, 478-120797, 478-120803, 478-120808, 478-121637, 478-121705, 478-121706, 478-121754, 478-121763, 478-121773, 478-121909, 478-121912, 478-121913, 478-121922, 478-121925, 478-122003, 478-122004, 478-122008, 478-122009, 478-122010, 478-122065, 478-122066, 478-122113, 478-122248, 478-122270, 478-122272, 478-122293, 478-122336, 478-122349, 478-122383, 478-122578, 478-122723, 478-122724, 478-122725, 478-122736, 478-122755, 478-123303, 478-123325, 478-123326, 478-123328, 478-123329, 478-123330, 478-123366, 478-123466, 478-123467, 478-123483, 478-124924, 478-125099, 478-125109, 478-125110, 478-125116, 478-125126, 478-125142, 478-125170, 478-125193, 478-125227, 478-125228, 478-125249, 478-125524, 478-125533, 478-125534, 478-125538, 478-125553, 478-125600, 478-125604, 478-125617, 478-125618, 478-125620, 478-125626, 478-125651, 478-126036, 478-126044, 478-126051, 478-126081, 478-126096, 478-126097, 478-126098, 478-126099, 478-126687, 478-126690, 478-126691, 478-127632, 478-134003, 478-134004, 478-134476, 478-136501, 478-136502, 478-136503, 478-136504, 478-136505, 478-136506, 478-136507, 478-136508, 478-136509, 478-136510, 478-136511, 478-136512, 478-136513, 478-136514, 478-136515, 478-136516, 478-136517, 478-136518, 478-136519, 478-136520, 478-136521, 478-136522, 478-136523, 478-136524, 478-136563, 478-136564, 478-137414, 478-139044, 478-139045, 478-139046, 478-139047, 478-139048, 478-139049, 478-139050, 478-139051, 478-139052, 478-139053, 478-139054, 478-139055, 478-139368, 478-139373, 478-139410, 478-139412, 478-139609, 478-139611, 478-139612, 478-139613, 478-139688, 478-139690, 478-139691, 478-139692, 478-139862, 478-139864, 478-139865, 478-139866, 478-139867, 478-139868, 478-139869, 478-139870, 478-139903, 478-139904, 478-139905, 478-139906, 478-139907, 478-139908, 478-139909, 478-139910, 478-139911, 478-139912, 478-139913, 478-139914, 478-139915, 478-139916, 478-139917, 478-139918, 478-140189, 478-140199, 478-140241, 478-140255, 478-140256, 478-140257, 478-140258, 478-140259, 478-140260, 478-140261, 478-140262, 478-140263, 478-140264, 478-140265, 478-140266, 478-140267, 478-140268, 478-140269, 478-140270, 478-140271, 478-140275, 478-140276, 478-140277, 478-140278, 478-140279, 478-140280, 478-140281, 478-140283, 478-140284, 478-140287, 478-140312, 478-140313, 478-140314, 478-140315, 478-140316, 478-140317, 478-140318, 478-140319, 478-140320, 478-140366, 478-140685, 478-140686, 478-140687, 478-140688, 478-141498, 478-141676, 478-141745, 478-141746, 478-141747, 478-141748, 478-141749, 478-141750, 478-141752, 478-141753, 478-141754, 478-141755, 478-141757, 478-141759, 478-141760, 478-141761, 478-141762, 478-141763, 478-141764, 478-141765, 478-141766, 478-141767, 478-141768, 478-141769, 478-141770, 478-141771, 478-141772, 478-141773, 478-141774, 478-141775, 478-142080, 478-142098, 478-142112, 478-142187, 478-142198, 478-142199, 478-142200, 478-142201, 478-142202, 478-142203, 478-142204, 478-142205, 478-142206, 478-142207, 478-142208, 478-142209, 478-142218, 478-142221, 478-142222, 478-142237, 478-142238, 478-142251, 478-142252, 478-142253, 478-142254, 478-142255, 478-142257, 478-142258, 478-142259, 478-142260, 478-142261, 478-142262, 478-142263, 478-142264, 478-142320, 478-142321, 478-142323, 478-142325, 478-142326, 478-142336, 478-142338, 478-142339, 478-142340, 478-142341, 478-142342, 478-142343, 478-142346, 478-142347, 478-142348, 478-142349, 478-142350, 478-142351, 478-142352, 478-142353, 478-142354, 478-142355, 478-142356, 478-142357, 478-142358, 478-142359, 478-142360, 478-142361, 478-142362, 478-142363, 478-142364, 478-142365, 478-142366, 478-142367, 478-142404, 478-142406, 478-142407, 478-142408, 478-142409, 478-142410, 478-142411, 478-142412, 478-142413, 478-142414, 478-142415, 478-142416, 478-142417, 478-142418, 478-142419, 478-142420, 478-142421, 478-142540, 478-142541, 478-142542, 478-142543, 478-142544, 478-142545, 478-142546, 478-142804, 478-142817, 478-142818, 478-142820, 478-142821, 478-142836, 478-142861, 478-142899, 478-142920, 478-142921, 478-142922, 478-142923, 478-143045, 478-143046, 478-143047, 478-143048, 478-143049, 478-143050, 478-143054, 478-143056, 478-143057, 478-143058, 478-143059, 478-143077, 478-143078, 478-143079, 478-143080, 478-143144, 478-143145, 478-143146, 478-143147, 478-143148, 478-143359, 478-143360, 478-143361, 478-143362, 478-143363, 478-143364, 478-143365, 478-143366, 478-143371, 478-143372, 478-143420, 478-143421, 478-143422, 478-143423, 478-143424, 478-143425, 478-143426, 478-143427, 478-143428, 478-143429, 478-143430, 478-143431, 478-143725, 478-143726, 478-143727, 478-143728, 478-143983, 478-143986, 478-144001, 478-144008, 478-144009, 478-144010, 478-144012, 478-144013, 478-144185, 478-144205, 478-144224, 478-144469, 478-144568, 478-144570, 478-144612, 478-144613, 478-144614, 478-144716, 478-144717, 478-144718, 478-144719, 478-144764, 478-144765, 478-144766, 478-144767, 478-144885, 478-144886, 478-144887, 478-144888, 478-144889, 478-144890, 478-144891, 478-144892, 478-144901, 478-145117, 478-145118, 478-145119, 478-145120, 478-145121, 478-145818, 478-145819, 478-145828, 478-145831, 478-145861, 478-145862, 478-145863, 478-145864, 478-145865, 478-145866, 478-145867, 478-145868, 478-145869, 478-145870, 478-145871, 478-145872, 478-145873, 478-145876, 478-145989, 478-146133, 478-146134, 478-146135, 478-146136, 478-146248, 478-146326, 478-146327, 478-146328, 478-146329, 478-146330, 478-146331, 478-146332, 478-146333, 478-146334, 478-146335, 478-146336, 478-146337, 478-146338, 478-146339, 478-146736, 478-146787, 478-147334, 478-147394, 478-147395, 478-147396
478-105280, 478-105976, 478-105977, 478-105996, 478-105997, 478-105998, 478-106003, 478-106004, 478-106074, 478-106075, 478-106505, 478-109829, 478-109830, 478-109831, 478-109832, 478-110626, 478-111175, 478-112024, 478-112037, 478-112039, 478-112041, 478-112494, 478-112495, 478-112500, 478-113109, 478-116877, 478-116888, 478-116889, 478-117079, 478-118040, 478-139752, 478-143282, 478-143283, 478-143284, 478-143285, 478-143286, 478-143287, 478-143693, 478-143694, 478-143695, 478-143696, 478-143697, 478-143698, 478-144037, 478-144038, 478-146207, 478-146210
478-105979, 478-105981, 478-105982, 478-105983, 478-105984, 478-105985, 478-105986, 478-105987, 478-106170, 478-106171, 478-106172, 478-106173, 478-106174, 478-106179, 478-107338, 478-107339, 478-107340, 478-107341, 478-107342, 478-107343, 478-107344, 478-107345, 478-107346, 478-107347, 478-107348, 478-107555, 478-107556, 478-107557, 478-107619, 478-107620, 478-107621, 478-107679, 478-107680, 478-107681, 478-107736, 478-108103, 478-108104, 478-108105, 478-108106, 478-108107, 478-108108, 478-109048, 478-109065, 478-109066, 478-109067, 478-110114, 478-110115, 478-110127, 478-110128, 478-110129, 478-110130, 478-110131, 478-110132, 478-110133, 478-110134, 478-110135, 478-110136, 478-110186, 478-110187, 478-110188, 478-110189, 478-110190, 478-110191, 478-110239, 478-110240, 478-110290, 478-110701, 478-110702, 478-110703, 478-110770, 478-110771, 478-110772, 478-110839, 478-110840, 478-110841, 478-111040, 478-111165, 478-111166, 478-111167, 478-111168, 478-111169, 478-111170, 478-111171, 478-111172, 478-111173, 478-111174, 478-111309, 478-111310, 478-111311, 478-111312, 478-111313, 478-111314, 478-111315, 478-111316, 478-111317, 478-111318, 478-111319, 478-111320, 478-111321, 478-111322, 478-111323, 478-111324, 478-111325, 478-111326, 478-111327, 478-111328, 478-111329, 478-111330, 478-111331, 478-111332, 478-111333, 478-111334, 478-111335, 478-111336, 478-111337, 478-111338, 478-113001, 478-113002, 478-113003, 478-113004, 478-113005, 478-113009, 478-113011, 478-113012, 478-113013, 478-113014, 478-113015, 478-113016, 478-113017, 478-113018, 478-113019, 478-113020, 478-113021, 478-113022, 478-113023, 478-113024, 478-113025, 478-113026, 478-113027, 478-113028, 478-113029, 478-113030, 478-113601, 478-113602, 478-113603, 478-113801, 478-113802, 478-113803, 478-113804, 478-113805, 478-113862, 478-113863, 478-113864, 478-113865, 478-113866, 478-113867, 478-113868, 478-113869, 478-113870, 478-113871, 478-113872, 478-113873, 478-113874, 478-113875, 478-113876, 478-113877, 478-113878, 478-113879, 478-115352, 478-115353, 478-115354, 478-115355, 478-115356, 478-115357, 478-115359, 478-115360, 478-115361, 478-115362, 478-115363, 478-115364, 478-115365, 478-115366, 478-115367, 478-115368, 478-115369, 478-115370, 478-115371, 478-115372, 478-115373, 478-115374, 478-115375, 478-115376, 478-116445, 478-116451, 478-116452, 478-116453, 478-117049, 478-117146, 478-117147, 478-117148, 478-117149, 478-117150, 478-117151, 478-117152, 478-117153, 478-117154, 478-117155, 478-117156, 478-117157, 478-117159, 478-117160, 478-117161, 478-117162, 478-117163, 478-117164, 478-117165, 478-117166, 478-117167, 478-117168, 478-117169, 478-117170, 478-118273, 478-118274, 478-118333, 478-118334, 478-118335, 478-118336, 478-118337, 478-118338, 478-118339, 478-118340, 478-118341, 478-118342, 478-118343, 478-118344, 478-118345, 478-118346, 478-118347, 478-118348, 478-118349, 478-118350, 478-118749, 478-118750, 478-118751, 478-118752, 478-118753, 478-118754, 478-118755, 478-118756, 478-118757, 478-118758, 478-118759, 478-118760, 478-118762, 478-118763, 478-118765, 478-118766, 478-118767, 478-118768, 478-118769, 478-118770, 478-118771, 478-118772, 478-118773, 478-118901, 478-118902, 478-118908, 478-118909, 478-122328, 478-122534, 478-122536, 478-122671, 478-122673, 478-122674, 478-122675, 478-122676, 478-122677, 478-122907, 478-122908, 478-122910, 478-122911, 478-122912, 478-123845, 478-123942, 478-124041, 478-124043, 478-124044, 478-124045, 478-124046, 478-124047, 478-124048, 478-134655, 478-134656, 478-134657, 478-134658, 478-134659, 478-134660, 478-134661, 478-134662, 478-141217, 478-141218, 478-141219, 478-141220, 478-141221, 478-141222, 478-141223, 478-141224, 478-141983, 478-141984, 478-142223, 478-142224, 478-142225, 478-142226, 478-142227, 478-142649, 478-142650, 478-142651, 478-143278, 478-143279, 478-143280, 478-143281, 478-143447, 478-143448, 478-143449, 478-143450, 478-143687, 478-143688, 478-143689, 478-143690, 478-143927, 478-143928, 478-143929, 478-143970, 478-143971, 478-143972, 478-143974, 478-143975, 478-143976, 478-143977, 478-143979, 478-147294, 478-147295
478-106825, 478-133977, 478-133981, 478-133983, 478-133984, 478-133985, 478-133986, 478-133987, 478-133988, 478-137692, 478-139189, 478-139614, 478-139615, 478-139616, 478-139617, 478-139618, 478-139619, 478-139620, 478-139621, 478-139622, 478-139623, 478-140408, 478-140552, 478-140733, 478-141450, 478-141451, 478-141452, 478-141453, 478-141454, 478-141455, 478-141557, 478-142565, 478-142566, 478-142567, 478-142568, 478-142569, 478-143691, 478-143692, 478-144928, 478-144929, 478-144930, 478-144931, 478-144932, 478-144933, 478-144934, 478-144935, 478-144936, 478-145112, 478-145113, 478-146515, 478-146516, 478-146517, 478-146518, 478-146519, 478-146520, 478-146719, 478-146720, 478-146721, 478-146929, 478-146930, 478-146931, 478-146932, 478-146933, 478-146934
478-105988, 478-105989, 478-105990, 478-105991, 478-106006, 478-106007, 478-106506, 478-106997, 478-106998, 478-107421, 478-107934, 478-108222, 478-108223, 478-108224, 478-108225, 478-108226, 478-109068, 478-109155, 478-109156, 478-109157, 478-109158, 478-109334, 478-109335, 478-109336, 478-109337, 478-109338, 478-109339, 478-109340, 478-109341, 478-109665, 478-109666, 478-109667, 478-109668, 478-110125, 478-110999, 478-111181, 478-111182, 478-112026, 478-112028, 478-112029, 478-112030, 478-112031, 478-112032, 478-112033, 478-112034, 478-112035, 478-112038, 478-112042, 478-112043, 478-112045, 478-112051, 478-112476, 478-112480, 478-112481, 478-112482, 478-112483, 478-112484, 478-112485, 478-112486, 478-112487, 478-112488, 478-112489, 478-112496, 478-112497, 478-112499, 478-113031, 478-113033, 478-113037, 478-113038, 478-113039, 478-113040, 478-113041, 478-113042, 478-113043, 478-113044, 478-113045, 478-113049, 478-113100, 478-113104, 478-113105, 478-113106, 478-113107, 478-113108, 478-113110, 478-113113, 478-114222, 478-115387, 478-115392, 478-115394, 478-115396, 478-116789, 478-116791, 478-116792, 478-116793, 478-116794, 478-116795, 478-116796, 478-116807, 478-116809, 478-116826, 478-116829, 478-117119, 478-117120, 478-117121, 478-117125, 478-117128, 478-117133, 478-117135, 478-117137, 478-117138, 478-117139, 478-117140, 478-117141, 478-117142, 478-117143, 478-119325, 478-119680, 478-119681, 478-119682, 478-119683, 478-119684, 478-119685, 478-119687, 478-119688, 478-119689, 478-120416, 478-120417, 478-120418, 478-120419, 478-120420, 478-120421, 478-120424, 478-121451, 478-121452, 478-121453, 478-121455, 478-121456, 478-121457, 478-124053, 478-124129, 478-124130, 478-124325, 478-124326, 478-124327, 478-124328, 478-124329, 478-124330, 478-124564, 478-124981, 478-124982, 478-124983, 478-124991, 478-124993, 478-124995, 478-125002, 478-125003, 478-125005, 478-125006, 478-125007, 478-125009, 478-125010, 478-125065, 478-125066, 478-125067, 478-125068, 478-125080, 478-126676, 478-126677, 478-128023, 478-128025, 478-128026, 478-128027, 478-129950, 478-130226, 478-130227, 478-130228, 478-130230, 478-130232, 478-130514, 478-130521, 478-130522, 478-130754, 478-130755, 478-130765, 478-131979, 478-132780, 478-132781, 478-132782, 478-140375, 478-140376, 478-140531, 478-140532, 478-140533, 478-140534, 478-140535, 478-140536, 478-140537, 478-140538, 478-140899, 478-140900, 478-141076, 478-141077, 478-141108, 478-141109, 478-141110, 478-141111, 478-141112, 478-141113, 478-141114, 478-141115, 478-141116, 478-141117, 478-141118, 478-141119, 478-141120, 478-141121, 478-141122, 478-143034, 478-143035, 478-143036, 478-143037, 478-143038, 478-143039, 478-143040, 478-143041, 478-143051, 478-143052, 478-143563, 478-143564, 478-143565, 478-143566, 478-144177, 478-144178, 478-144179, 478-144180, 478-144181, 478-144182, 478-144183, 478-144184, 478-144186, 478-144187, 478-144188, 478-144189, 478-144190, 478-144191, 478-144192, 478-144193, 478-144194, 478-144195, 478-144196, 478-144197, 478-144198, 478-144199, 478-144201, 478-144202, 478-144203, 478-144204, 478-144206, 478-144207, 478-144208, 478-144209, 478-144210, 478-144211, 478-144212, 478-144213, 478-144214, 478-144215, 478-144216, 478-144217, 478-144218, 478-144219, 478-144220, 478-144221, 478-144222, 478-144223, 478-144225, 478-144226, 478-144227, 478-144228, 478-144229, 478-144230, 478-144231, 478-144232, 478-144233, 478-144234, 478-144235, 478-144236, 478-144237, 478-144238, 478-144239, 478-144240, 478-144241, 478-144242, 478-144243, 478-144244, 478-144245, 478-144246, 478-144247, 478-144248, 478-144249, 478-144250, 478-144264, 478-145553, 478-145554, 478-145615, 478-145647, 478-145924, 478-146053, 478-146054, 478-146055, 478-146056, 478-146057, 478-146107, 478-146129, 478-146130, 478-146131, 478-146132, 478-146204, 478-146205, 478-146206, 478-146265, 478-146283, 478-146397, 478-146403, 478-146426, 478-146453, 478-146462, 478-1466911, 478-1466912, 478-1466913, 478-1466914, 478-1466915, 478-1466916, 478-146917, 478-146918, 478-100075, 478-100250, 478-111102, 478-111103, 478-112071, 478-112847, 478-112848, 478-112849, 478-112850, 478-112851, 478-112852, 478-112853, 478-112854, 478-112855, 478-112894, 478-112895, 478-112896, 478-112897, 478-112898, 478-112899, 478-112900, 478-112901, 478-112902, 478-112903, 478-112904, 478-112905, 478-112906, 478-112907, 478-114666, 478-116219, 478-122727, 478-125119
478-109954, 478-109955, 478-109956, 478-109957, 478-109958, 478-109959, 478-112631, 470-112426, 478-100366, 478-101160, 478-101556, 478-105117, 478-105237, 478-106022, 478-106031, 478-106090, 478-106097, 478-106194, 478-106263, 478-106422, 478-106426, 478-106583, 478-106593, 478-106602, 478-106614, 478-106615, 478-106622, 478-106624, 478-106625, 478-106630, 478-106634, 478-107101, 478-107103, 478-107469, 478-107470, 478-107471, 478-107472, 478-108020, 478-108021, 478-108022, 478-108023, 478-108024, 478-108025, 478-108958, 478-108959, 478-108960, 478-108967, 478-109027, 478-109934, 478-110041, 478-110042, 478-110043, 478-110044, 478-110045, 478-110046, 478-110047, 478-110048, 478-110049, 478-110050, 478-110051, 478-110576, 478-111071, 478-111121, 478-111122, 478-111123, 478-111124, 478-111125, 478-111127, 478-111129, 478-111711, 478-112015, 478-112147, 478-112427, 478-112559, 478-112569, 478-112570, 478-112571, 478-112572, 478-112611, 478-112612, 478-112613, 478-112621, 478-112622, 478-112623, 478-112638, 478-112639, 478-112756, 478-112757, 478-112758, 478-112759, 478-112760, 478-112761, 478-112762, 478-112763, 478-112765, 478-112766, 478-112767, 478-112781, 478-112782, 478-112783, 478-112784, 478-112785, 478-112786, 478-112787, 478-112788, 478-112789, 478-112790, 478-112791, 478-112792, 478-112797, 478-112798, 478-112799, 478-112800, 478-112808, 478-112826, 478-112827, 478-112828, 478-112845, 478-112846, 478-113210, 478-113211, 478-113582, 478-113669, 478-113762, 478-113763, 478-113765, 478-113766, 478-113767, 478-113768, 478-113769, 478-113770, 478-113771, 478-113779, 478-113785, 478-113788, 478-113790, 478-114023, 478-114025, 478-114134, 478-114135, 478-114220, 478-114364, 478-114365, 478-114366, 478-114368, 478-114369, 478-114370, 478-114371, 478-114372, 478-114373, 478-114374, 478-114375, 478-114399, 478-114400, 478-114401, 478-114402, 478-114443, 478-114506, 478-114507, 478-114508, 478-114509, 478-114585, 478-114587, 478-114588, 478-114595, 478-114628, 478-114633, 478-114635, 478-114664, 478-114686, 478-114687, 478-114688, 478-114689, 478-114691, 478-115083, 478-115084, 478-115085, 478-115086, 478-116335, 478-116790, 478-116799, 478-116800, 478-116801, 478-116802, 478-116803, 478-116804, 478-116805, 478-116806, 478-116808, 478-116811, 478-116812, 478-116813, 478-116814, 478-116815, 478-116816, 478-116818, 478-116819, 478-116820, 478-116821, 478-116822, 478-116823, 478-116825, 478-117123, 478-117124, 478-117129, 478-117130, 478-117131, 478-117132, 478-117136, 478-117310, 478-117316, 478-118037, 478-118038, 478-118039, 478-118043, 478-118044, 478-118045, 478-118046, 478-118251, 478-118252, 478-118253, 478-118254, 478-118255, 478-118257, 478-118258, 478-118858, 478-119982, 478-120449, 478-120468, 478-120478, 478-120479, 478-122118, 478-122119, 478-122309, 478-122690, 478-122707, 478-122708, 478-122709, 478-122743, 478-124305, 478-124313, 478-124314, 478-124317, 478-124319, 478-124324, 478-127066, 478-128472, 478-128485, 478-128501, 478-128503, 478-128511, 478-128512, 478-128541, 478-129459, 478-129603, 478-129616, 478-129619, 478-129642, 478-129643, 478-130688, 478-130748, 478-125081, 478-125082, 478-109079, 478-110974, 478-112046, 478-113034, 478-113035, 478-113046, 478-113047, 478-113048, 478-113054, 478-113055, 478-113095, 478-113102, 478-113103, 478-114219, 478-114223, 478-114225, 478-115378, 478-115379, 478-115380, 478-115382, 478-115389, 478-116827, 478-116828, 478-116830, 478-116831, 478-118042, 478-122160, 478-122167, 478-124050, 478-124051, 478-124052, 478-124056, 478-124478, 478-124572, 478-124984, 478-124985, 478-124987, 478-124990, 478-124992, 478-124999, 478-125000, 478-125073, 478-125074, 478-125075, 478-125076, 478-106175, 478-106181, 478-106183, 478-128524, 478-129152, 478-127063, 478-127078, 478-128745, 478-130071, 478-130304, 478-130305, 478-130359, 478-146058, 478-146059, 478-130564, 478-130565, 478-130568, 478-130569, 478-120326, 478-120470, 478-120471, 478-120472, 478-120475, 478-120486, 478-120490, 478-120493, 478-120495, 478-122689, 478-118544, 478-119360, 478-122342, 478-124306, 478-124307, 478-124308, 478-124309, 478-124315, 478-124316, 478-124318, 478-124321, 478-124322, 478-124323, 478-126427, 478-126431, 478-126434, 478-126660, 478-127449, 478-127757, 478-127758, 478-127855, 478-128298, 478-128469, 478-128543, 478-128544, 478-109114, 478-116014, 478-116015, 478-116676, 478-118332, 478-122296, 478-122297, 478-122299, 478-122300, 478-122301, 478-122302, 478-122303, 478-122311, 478-122376, 478-122378, 478-128354, 478-128400, 478-128426, 478-128475, 478-127579, 478-128266, 478-110558, 478-122295, 478-122316, 478-130366, 478-132101, 478-132102, 478-132103, 478-132104, 478-127072
478-111194, 478-111211, 478-111266, 478-111267, 478-105379
478-108390, 478-108391, 478-110666, 478-110668, 478-110678, 478-110681, 478-110683, 478-110717, 478-110722, 478-110724, 478-110730, 478-110733, 478-110738, 478-110967, 478-111012
478-100055, 478-100377, 478-100437, 478-101115, 478-101144, 478-106394, 478-108846, 478-108856, 478-108857, 478-108869, 478-108870, 478-108874, 478-108875, 478-108879, 478-108882, 478-108885, 478-108891, 478-108892, 478-108893, 478-108895, 478-108921, 478-108922, 478-108941, 478-108942, 478-108944, 478-108946, 478-108947, 478-108955, 478-108956, 478-108957, 478-119332, 478-119349, 478-131047, 478-131073, 478-131074, 478-131098, 478-131102, 478-131118, 478-133446, 478-133447, 478-133448, 478-133449, 478-133450, 478-133451, 478-133452, 478-133453, 478-133454, 478-133455, 478-133456, 478-133457, 478-133458, 478-133459, 478-133460, 478-133479, 478-133480, 478-133481, 478-133482, 478-133483, 478-133484, 478-133485, 478-133486, 478-133487, 478-133488, 478-133489, 478-133490, 478-133491, 478-133492, 478-133493, 478-133494, 478-133495, 478-133496, 478-133497, 478-133498, 478-133499, 478-133500, 478-133501, 478-133503, 478-133504, 478-140502, 478-140503, 478-140504, 478-140505, 478-140506, 478-140507, 478-140508, 478-140509, 478-140510, 478-140511, 478-143339, 478-143340, 478-143341, 478-143342, 478-144082, 478-144089, 478-144090, 478-144091, 478-144092, 478-144093, 478-144094, 478-144095, 478-144097, 478-144099
478-117513, 478-117516, 478-117529, 478-117530, 478-117531, 478-117532, 478-117768, 478-117769, 478-117771, 478-117774, 478-117778, 478-117779
478-107858, 478-111152, 478-111339, 478-111340, 478-111351, 478-111353, 478-111354, 478-111355, 478-118463, 478-118464, 478-118465, 478-118472, 478-118473, 478-118475, 478-118476, 478-118477, 478-118481, 478-118482
478-110227, 478-110228, 478-110390, 478-118731, 478-118896, 478-118930, 478-119528, 478-119529, 478-121488, 478-121526, 478-121527, 478-121530, 478-121531, 478-121541, 478-121930, 478-121931, 478-121932, 478-137038, 478-137073, 478-137126, 478-137134, 478-137185, 478-137197, 478-137198, 478-137199, 478-137200, 478-137201, 478-137202, 478-137203, 478-137204, 478-137205, 478-137206, 478-137207, 478-137208, 478-137211, 478-137212, 478-137213, 478-137214, 478-137215, 478-137216, 478-137217, 478-137218, 478-137219, 478-137220, 478-137221, 478-137225, and 478-137226. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare, Incorporated 5150 220th Ave Se Issaquah WA 98029-6834
|
For Additional Information Contact |
Al Van Houdt 425-657-7200 Ext. 5970
|
Manufacturer Reason for Recall |
Telemetry Modules are failing to alarm for low heart rate and asystole.
|
FDA Determined Cause 2 |
Software design |
Action |
On 07/18/08 Spacelabs Healthcare issued a "URGENT-MEDICAL DEVICE CORRECTION" letter, dated July 16, 2008, to all customers who received the Spacelabs Medical Telemetry Receiver-Model 90478. The letter states circumstances in which the alarm fails, how to test each module for high rate alarm prior to patient use and a temporary work around if the alarm failure exists.` Letters have a receipt acknowledgement to ensure each customer receives the letter. Spacelabs Technical Support will personally install the new software, version V3.05.07EN, in all Spacelabs Medical Telemetry Receivers Model 90478. Technical Support will also be responsible for responding to customer technical questions and ensuring each customer is upgraded regarding this medical device recall. For assistance contact SpaceLabs Medical at 1-425-657-7200. |
Quantity in Commerce |
37,590 |
Distribution |
Nationwide and Internationally |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DSI and Original Applicant = SPACELABS MEDICAL, INC.
|
|
|
|