• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall VERTIER Surgical Table

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall VERTIER Surgical Table see related information
Date Initiated by Firm July 02, 2008
Date Posted September 18, 2008
Recall Status1 Terminated 3 on January 07, 2010
Recall Number Z-2378-2008
Recall Event ID 49021
Product Classification Operating-Room Table - Product Code FQO
Product VERTIER Surgical Table, Catalog #0788-100-000S (SIDNE-enabled standard table); distributed by Stryker Communications, Flower Mound, TX.

Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.
Code Information All lot codes/serial numbers.
Recalling Firm/
Manufacturer
Stryker Communications Corp
1410 Lakeside Pkwy Ste 100
Flower Mound TX 75028-4026
For Additional Information Contact
972-410-7310
Manufacturer Reason
for Recall
Hydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly resulting in the unexpected and rapid movement of the table.
FDA Determined
Cause 2
Device Design
Action Urgent: Device Recall letters were sent to consignees on 07/02/08 via FedEx with return receipt. Consignees were asked to discontinue use of the tables until a Stryker representative has come to the facility and updated the table. The firm has identified and validated replacement parts that prevent this occurrence.
Quantity in Commerce 18 units.
Distribution Nationwide Distribution --- including states of AR, CA, FL, GA, ID, IL, IN, KS, KY, MN, MO, ND, NJ, NY, OH, OK, OR, SD, TX, UT, VA and Washington DC.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-