• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Medical & Radiation Emitting Device Recalls

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Recall
Maxi Move Patient Lift with Lock & Load System (Combi Hanger)
see related information
Date Posted September 30, 2008
Recall Number Z-2470-2008
Product Maxi Move Patient Lift with Lock & Load System (Combi Hanger); Manufactured by Arjo Hospital Equipment AB, Eslov, Sweden; Models KMBB4CLU2FUS, KMBB4MLU2FUS, KMBB4MSU2FUS, KMBB4NLX2FUS, KMBB4NSX2FUS, KMBB4OLU2FUS, KMBB4OSU2FUS, KMBB4PLX2FUS, KMBB4PSX2FUS, KMBB4QLU2FUS, and KMBB4RLX2FUS. The lifter is intended to be used under the direct supervision of trained personnel for the transfer of residents from one location to another.
Code Information Model KMBB4CLU2FUS, serial #SEE0741325, SEE0741326, SEE0741327, SEE0741328, SEE0741329, SEE0741330, SEE0741331, SEE0741332, SEE0741333; Model KMBB4MLU2FUS, serial #SEE0612303, SEE0612304, SEE0612305, SEE0612306, SEE0612307, SEE0612308, SEE0612309, SEE0612310, SEE0612311, SEE0612312, SEE0612313, SEE0612314, SEE0612315, SEE0612316, SEE0612321, SEE0612322, SEE0612323, SEE0612324, SEE0612325, SEE0612326, SEE0612327, SEE0612328, SEE0612329, SEE0612330, SEE0612331, SEE0612332, SEE0612333, SEE0612334, SEE0622295, SEE0622296, SEE0622297, SEE0622298, SEE0622299, SEE0622300, SEE0622301, SEE0622302, SEE0622303, SEE0622304, SEE0622305, SEE0622306, SEE0622307, SEE0622308, SEE0622309, SEE0622310, SEE0622311, SEE0622312, SEE0622313, SEE0622314, SEE0622315, SEE0622316, SEE0622317, SEE0622318, SEE0622319, SEE0622320, SEE0622431, SEE0622432, SEE0622433, SEE0622434, SEE0622436, SEE0622437, SEE0622438, SEE0622439, SEE0622441, SEE0622442, SEE0622443, SEE0622444, SEE0622445, SEE0622446, SEE0622447, SEE0622448, SEE0622455, SEE0622456, SEE0622457, SEE0622459, SEE0622460, SEE0622461, SEE0622462, SEE0622463, SEE0622464, SEE0622465, SEE0622466, SEE0622467, SEE0622468, SEE0622469, SEE0622470, SEE0622471, SEE0622472, SEE0622473, SEE0704574, SEE0704575, SEE0716297, SEE0717173, SEE0717174, SEE0727181, SEE0727182, SEE0729323, SEE0729325, SEE0729326, SEE0733361, SEE0733362, SEE0733363, SEE0733364, SEE0736390, SEE0736391, SEE0736392, SEE0736393, SEE0736394, SEE0739387, SEE0739388, SEE0739389, SEE0743119, SEE0743120, SEE0745220, SEE0745221, SEE0745222, SEE0746268, SEE0746269, SEE0746270, SEE0746271, SEE0746272, SEE0746274, SEE0747593, SEE0747594, SEE0747595, SEE0747596, SEE0747597, SEE0747598, SEE0803280, SEE0803281; Model KMBB4MSU2FUS, serial # SEE0614128, SEE0614129, SEE0614130, SEE0614131, SEE0614132, SEE0614133, SEE0614134, SEE0614135, SEE0614136, SEE0614137, SEE0614138, SEE0614139, SEE0614140, SEE0614141, SEE0614142, SEE0614143, SEE0614144, SEE0622249, SEE0622250, SEE0622252, SEE0622253, SEE0622254, SEE0622255, SEE0622257, SEE0622259, SEE0622260, SEE0622261, SEE0622263, SEE0622264, SEE0622265, SEE0622267, SEE0622268, SEE0622270, SEE0622271, SEE0622272, SEE0622273, SEE0628311, SEE0628312, SEE0628314, SEE0628315, SEE0628319, SEE0628320, SEE0628322, SEE0628323, SEE0628324, SEE0628326, SEE0629134, SEE0629140, SEE0631183, SEE0631193, SEE0631197, SEE0702534, SEE0702535, SEE0702536, SEE0702537, SEE0702538, SEE0702540, SEE0702541, SEE0711494, SEE0711495, SEE0711498, SEE0712679, SEE0712680, SEE0717171, SEE0717172, SEE0719092, SEE0719093, SEE0719094, SEE0719095, SEE0719096, SEE0728221, SEE0733296, SEE0733297, SEE0739293, SEE0739294, SEE0739295, SEE0739362, SEE0739363, SEE0739364, SEE0742802, SEE0742803, SEE0742804, SEE0742805, SEE0745185, SEE0745186, SEE0806536, SEE0806537; Model KMBB4NLX2FUS, serial # SEE0615013, SEE0615014, SEE0615015, SEE0615016, SEE0615017, SEE0615018, SEE0615019, SEE0615020, SEE0615021, SEE0615022, SEE0615023, SEE0622286, SEE0622287, SEE0622288, SEE0622289, SEE0622290, SEE0622291, SEE0622449, SEE0622450, SEE0622451, SEE0622452, SEE0622453, SEE0622454, SEE0629144, SEE0629145, SEE0629146, SEE0629147, SEE0629148, SEE0629149, SEE0629150, SEE0629151, SEE0629152, SEE0629153, SEE0631198, SEE0631199, SEE0631200, SEE0631201, SEE0631202, SEE0631203, SEE0631204, SEE0631205, SEE0631206, SEE0631207, SEE0645370, SEE0645371, SEE0645372, SEE0645373, SEE0645374, SEE0647500, SEE0647501, SEE0647502, SEE0647503, SEE0647504, SEE0647505, SEE0647506, SEE0647507, SEE0650032, SEE0650033, SEE0650034, SEE0650035, SEE0650036, SEE0723150, SEE0723152, SEE0724576, SEE0724577, SEE0724578, SEE0724579, SEE0727179, SEE0727180, SEE0728164, SEE0728165, SEE0728166, SEE0729320, SEE0729321, SEE0729322, SEE0733298, SEE0733299, SEE0733300, SEE0745187, SEE0745188; Model KMBB4NSX2FUS, serial #SEE0612335, SEE0612336, SEE0614304, SEE0614306, SEE0614307, SEE0614308, SEE0614309, SEE0614310, SEE0614311, SEE0614312, SEE0614313, SEE0614314, SEE0614315, SEE0614316, SEE0614317, SEE0614318, SEE0614319, SEE0614320, SEE0614321, SEE0614322, SEE0614323, SEE0614324, SEE0614325, SEE0614326, SEE0622323, SEE0622324, SEE0622325, SEE0622326, SEE0622327, SEE0622328, SEE0622329, SEE0622330, SEE0622331, SEE0622332, SEE0622333, SEE0622334, SEE0622335, SEE0622336, SEE0622337, SEE0622338, SEE0622339, SEE0622340, SEE0622341, SEE0622342, SEE0622343, SEE0622344, SEE0622345, SEE0622346, SEE0622347, SEE0629154, SEE0629155, SEE0629156, SEE0629157, SEE0629158, SEE0629159, SEE0629160, SEE0629161, SEE0629162, SEE0629163, SEE0629164, SEE0629165, SEE0629166, SEE0629167, SEE0629168, SEE0629169, SEE0629170, SEE0629171, SEE0629172, SEE0632024, SEE0632026, SEE0632027, SEE0632028, SEE0632029, SEE0632030, SEE0632031, SEE0632032, SEE0632033, SEE0650037, SEE0650038, SEE0650039, SEE0652124, SEE0652125, SEE0652126, SEE0716021, SEE0716022, SEE0716023, SEE0716024, SEE0716025, SEE0716026, SEE0716027, SEE0716028, SEE0716029, SEE0716030, SEE0719097, SEE0719098, SEE0719099, SEE0719100, SEE0719101, SEE0719102, SEE0719601, SEE0719602, SEE0727339, SEE0727340, SEE0727341, SEE0727342, SEE0729176, SEE0729177, SEE0729178, SEE0729179, SEE0733301, SEE0733302, SEE0733303, SEE0733304, SEE0733305, SEE0742806, SEE0742807; Model KMBB4OLU2FUS, serial #SEE0551014, SEE0551015, SEE0551016, SEE0551017, SEE0551018, SEE0551019, SEE0551020, SEE0551021, SEE0551028, SEE0551029, SEE0551031, SEE0614452, SEE0614453, SEE0614454, SEE0614455, SEE0614456, SEE0614457, SEE0614458, SEE0614459, SEE0614460, SEE0614461, SEE0614462, SEE0614463, SEE0620535, SEE0620536, SEE0620537, SEE0620538, SEE0620539, SEE0620540, SEE0620541, SEE0620542, SEE0620543, SEE0620544, SEE0620545, SEE0620546, SEE0620547, SEE0620548, SEE0620549, SEE0620550, SEE0620551, SEE0620552, SEE0620553, SEE0620554, SEE0620555, SEE0620556, SEE0620557, SEE0620558, SEE0620559, SEE0620560, SEE0620561, SEE0620562, SEE0620563, SEE0620564, SEE0620565, SEE0620566, SEE0620567, SEE0620568, SEE0620569, SEE0620570, SEE0620571, SEE0620572, SEE0620573, SEE0620574, SEE0620575, SEE0620576 , SEE0620577, SEE0620578, SEE0620579, SEE0620580, SEE0620581, SEE0620582, SEE0620583, SEE0620584, SEE0620585, SEE0620586, SEE0620587, SEE0620588, SEE0620589, SEE0622598, SEE0622599, SEE0622600, SEE0622601, SEE0622602, SEE0622603, SEE0622604, SEE0622605, SEE0622606, SEE0622607, SEE0622608, SEE0622609, SEE0622610, SEE0622611, SEE0622612, SEE0622613, SEE0622614, SEE0622615, SEE0622616, SEE0622617, SEE0622618, SEE0622619, SEE0622620, SEE0622621, SEE0624160, SEE0624161, SEE0624162, SEE0624163, SEE0624164, SEE0624165, SEE0624166, SEE0624167, SEE0624168, SEE0624169, SEE0624170, SEE0624171, SEE0624172, SEE0624173, SEE0624174, SEE0624175, SEE0624177, SEE0624178, SEE0624179, SEE0624180, SEE0624181, SEE0624182, SEE0624183, SEE0624184, SEE0624185, SEE0624186, SEE0624187, SEE0624188, SEE0624189, SEE0624190, SEE0624191, SEE0624192, SEE0624193, SEE0624194, SEE0624195, SEE0624196, SEE0624197, SEE0624198, SEE0624199, SEE0624200, SEE0624201, SEE0624202, SEE0624203, SEE0624204, SEE0624205, SEE0624206, SEE0624207, SEE0624208, SEE0624209, SEE0624210, SEE0624211, SEE0624212, SEE0624213, SEE0624214, SEE0624215, SEE0624216, SEE0624217, SEE0624218, SEE0624219, SEE0624220, SEE0624221, SEE0624222, SEE0624223, SEE0624224, SEE0624225, SEE0624226, SEE0624227, SEE0624228, SEE0624229, SEE0624230, SEE0624231, SEE0624232, SEE0624233, SEE0624234, SEE0624235, SEE0624236, SEE0624237, SEE0624238, SEE0624239, SEE0624240, SEE0624567, SEE0624568, SEE0624569, SEE0624570, SEE0624571, SEE0624572, SEE0624573, SEE0624574, SEE0624575, SEE0624576, SEE0624577, SEE0624578, SEE0624579, SEE0624580, SEE0624581, SEE0624582, SEE0624583, SEE0624584, SEE0624585, SEE0624586, SEE0624587, SEE0624588, SEE0624589, SEE0624590, SEE0624591, SEE0624592, SEE0624593, SEE0624594, SEE0624595, SEE0624596, SEE0624597, SEE0624598, SEE0624599, SEE0624600, SEE0624601, SEE0624602, SEE0624603, SEE0624604, SEE0624605, SEE0624606, SEE0624607, SEE0624608, SEE0624609, SEE0624610, SEE0624611, SEE0624612, SEE0624613, SEE0624614, SEE0624615, SEE0624616, SEE0624617, SEE0624618, SEE0624619, SEE0624620, SEE0624621, SEE0624622, SEE0624623, SEE0624624, SEE0624625, SEE0624626, SEE0624627, SEE0624628, SEE0624629, SEE0624630, SEE0624631, SEE0624632, SEE0624633, SEE0624634, SEE0624635, SEE0624636, SEE0631208, SEE0631209, SEE0631210, SEE0631211, SEE0631212, SEE0631213, SEE0631214, SEE0631215, SEE0631216, SEE0631217, SEE0631218, SEE0631219, SEE0631220, SEE0631233, SEE0631234, SEE0631235, SEE0631236, SEE0631237, SEE0631238, SEE0631239, SEE0631240, SEE0631241, SEE0631242, SEE0631243, SEE0631244, SEE0631245, SEE0631246, SEE0631247, SEE0631248, SEE0631249, SEE0631250, SEE0631251, SEE0631252, SEE0631253, SEE0631254, SEE0631255, SEE0631256, SEE0631257, SEE0631258, SEE0631259, SEE0631260, SEE0631261, SEE0631262, SEE0631263, SEE0631264, SEE0631265, SEE0631266, SEE0631267, SEE0631268, SEE0631269, SEE0631270, SEE0631271, SEE0631272, SEE0631273, SEE0631274, SEE0631275, SEE0631276, SEE0631277, SEE0631278, SEE0631279, SEE0631280, SEE0631281, SEE0631282, SEE0631283, SEE0631284, SEE0631285, SEE0631286, SEE0631287, SEE0631288, SEE0631289, SEE0631290, SEE0631291, SEE0631292, SEE0631293, SEE0631294, SEE0631295, SEE0631296, SEE0631297, SEE0631298, SEE0631299, SEE0632065, SEE0632066, SEE0632067, SEE0632068, SEE0632069, SEE0632070, SEE0632071, SEE0632072, SEE0632073, SEE0632074, SEE0632075, SEE0632076, SEE0632077, SEE0632078, SEE0632079, SEE0632080, SEE0632081, SEE0632082, SEE0632083, SEE0632084, SEE0633073, SEE0633074, SEE0633075, SEE0633076, SEE0633077, SEE0633078, SEE0633079, SEE0633080, SEE0633081, SEE0633082, SEE0633083, SEE0633084, SEE0633085, SEE0633086, SEE0633087, SEE0633088, SEE0633089, SEE0633090, SEE0633091, SEE0633092, SEE0633093, SEE0633094, SEE0633095, SEE0633096, SEE0638090, SEE0638091, SEE0638092, SEE0638093, SEE0638094, SEE0638095, SEE0638096, SEE0638097, SEE0638098, SEE0638099, SEE0638100, SEE0638101, SEE0638102, SEE0638103, SEE0638104, SEE0638105, SEE0638106, SEE0638107, SEE0638108, SEE0638109, SEE0638110, SEE0638111, SEE0638112, SEE0638113, SEE0638114, SEE0638115, SEE0638116, SEE0638117, SEE0638118, SEE0638119, SEE0638120, SEE0638121, SEE0638122, SEE0638123, SEE0638124, SEE0638125, SEE0639165, SEE0639166, SEE0639167, SEE0639168, SEE0639169, SEE0639170, SEE0639171, SEE0639173, SEE0639175, SEE0639176, SEE0639177, SEE0639178, SEE0639179, SEE0639180, SEE0639181, SEE0639182, SEE0639183, SEE0639184, SEE0640339, SEE0640341, SEE0640342, SEE0640343, SEE0640344, SEE0640345, SEE0640346, SEE0640347, SEE0640348, SEE0640349, SEE0640350, SEE0640351, SEE0640352, SEE0640353, SEE0640354, SEE0640355, SEE0640356, SEE0640357, SEE0640358, SEE0643467, SEE0643468, SEE0643469, SEE0643470, SEE0643471, SEE0643473, SEE0643474, SEE0643475, SEE0643476, SEE0643477, SEE0643478, SEE0643479, SEE0643480, SEE0643481, SEE0643482, SEE0643483, SEE0643484, SEE0643485, SEE0643486, SEE0643487, SEE0643488, SEE0643489, SEE0643490, SEE0643491, SEE0643492, SEE0643493, SEE0643494, SEE0643495, SEE0643496, SEE0645375, SEE0645376, SEE0645377, SEE0645378, SEE0645379, SEE0645380, SEE0645381, SEE0645382, SEE0645383, SEE0645384, SEE0645385, SEE0645386, SEE0645387, SEE0645388, SEE0645389, SEE0645390, SEE0645391, SEE0645392, SEE0645393, SEE0645394, SEE0645395, SEE0645396, SEE0645397, SEE0645398, SEE0645399, SEE0645400, SEE0645401, SEE0645402, SEE0645403, SEE0645404, SEE0701516, SEE0701517, SEE0701518, SEE0701519, SEE0701520, SEE0701521, SEE0701522, SEE0701523, SEE0703111, SEE0703112, SEE0703113, SEE0703114, SEE0703116, SEE0703117, SEE0703118, SEE0703119, SEE0703120, SEE0703121, SEE0703122, SEE0703123, SEE0703124, SEE0703125, SEE0703126, SEE0703127, SEE0703128, SEE0703129, SEE0703130, SEE0703131, SEE0703132, SEE0703133, SEE0703134, SEE0703135, SEE0703136, SEE0703137, SEE0703138, SEE0703139, SEE0703140, SEE0703141, SEE0703142, SEE0703143, SEE0703144, SEE0703145, SEE0703146, SEE0703147, SEE0704548, SEE0704549, SEE0704550, SEE0704551, SEE0704552, SEE0704553, SEE0704554, SEE0704555, SEE0704556, SEE0704557, SEE0704558, SEE0704559, SEE0704560, SEE0704561, SEE0704562, SEE0704563, SEE0704564, SEE0704565, SEE0704566, SEE0704567, SEE0704568, SEE0704569, SEE0704570, SEE0704571, SEE0704572, SEE0708014, SEE0708015, SEE0708016, SEE0708017, SEE0708018, SEE0708019, SEE0708020, SEE0708021, SEE0708022, SEE0708023, SEE0708024, SEE0708025, SEE0708026, SEE0708027, SEE0708028, SEE0708029, SEE0708030, SEE0708031, SEE0708032, SEE0708033, SEE0708239, SEE0708240, SEE0708241, SEE0708242, SEE0708243, SEE0708244, SEE0708245, SEE0708246, SEE0708247, SEE0708248, SEE0708249, SEE0708250, SEE0708252, SEE0708253, SEE0712121, SEE0712122, SEE0712123, SEE0712124, SEE0712125, SEE0712126, SEE0712127, SEE0712128, SEE0712129, SEE0712130, SEE0712131, SEE0712132, SEE0712133, SEE0712134, SEE0712135, SEE0712681, SEE0712682, SEE0712683, SEE0712684, SEE0712685, SEE0712686, SEE0712687, SEE0712688, SEE0712689, SEE0712690, SEE0713268, SEE0713269, SEE0713270, SEE0713271, SEE0713272, SEE0713273, SEE0713274, SEE0713276, SEE0713277, SEE0713278, SEE0715246, SEE0715247, SEE0715248, SEE0715249, SEE0715250, SEE0715251, SEE0715252, SEE0715253, SEE0715254, SEE0716282, SEE0716283, SEE0716284, SEE0716285, SEE0716286, SEE0716287, SEE0716288, SEE0716289, SEE0716290, SEE0716291, SEE0716292, SEE0716293, SEE0716294, SEE0716295, SEE0716296, SEE0717156, SEE0717157, SEE0717158, SEE0717159, SEE0717160, SEE0717161, SEE0717162, SEE0717163, SEE0717164, SEE0717165, SEE0717166, SEE0717167, SEE0717168, SEE0717169, SEE0717170, SEE0719561, SEE0719562, SEE0719563, SEE0719564, SEE0719565, SEE0719566, SEE0719567, SEE0719568, SEE0719569, SEE0719570, SEE0719571, SEE0719572, SEE0719573, SEE0719575, SEE0719576, SEE0719577, SEE0719578, SEE0719579, SEE0719580, SEE0720189, SEE0720190, SEE0720191, SEE0720192, SEE0720193, SEE0720194, SEE0720195, SEE0720196, SEE0720197, SEE0720198, SEE0720200, SEE0720201, SEE0720202, SEE0720203, SEE0720204, SEE0720205, SEE0720206, SEE0720207, SEE0720208, SEE0720209, SEE0720210, SEE0720211, SEE0720212, SEE0720213, SEE0721345, SEE0721346, SEE0721347, SEE0721348, SEE0721349, SEE0721350, SEE0721351, SEE0721352, SEE0721353, SEE0721354, SEE0721355, SEE0721356, SEE0721357, SEE0721358, SEE0721359, SEE0721360, SEE0721361, SEE0721362, SEE0721363, SEE0721364, SEE0721365, SEE0721366, SEE0721367, SEE0721368, SEE0721369, SEE0721370, SEE0721371, SEE0721372, SEE0721373, SEE0721374, SEE0721375, SEE0721376, SEE0721377, SEE0721378, SEE0721379, SEE0721380, SEE0721381, SEE0721382, SEE0721383, SEE0721384, SEE0721385, SEE0721386, SEE0721387, SEE0721388, SEE0721389, SEE0721390, SEE0721391, SEE0721392, SEE0721393, SEE0721394, SEE0721395, SEE0721396, SEE0721397, SEE0721398, SEE0721399, SEE0721400, SEE0721401, SEE0721402, SEE0721403, SEE0721404, SEE0721405, SEE0721406, SEE0721407, SEE0721408, SEE0721409, SEE0721410, SEE0721411, SEE0721412, SEE0721413, SEE0721414, SEE0721415, SEE0721416, SEE0721417, SEE0721418, SEE0721419, SEE0721420, SEE0721421, SEE0721422, SEE0727343, SEE0727344, SEE0727345, SEE0727346, SEE0727347, SEE0727348, SEE0727349, SEE0727350, SEE0727351, SEE0727352, SEE0727353, SEE0727354, SEE0727355, SEE0727356, SEE0727357, SEE0727358, SEE0727359, SEE0727360, SEE0727361, SEE0727362, SEE0727363, SEE0727364, SEE0727365, SEE0727366, SEE0727367, SEE0727368, SEE0727369, SEE0727370, SEE0727371, SEE0727372, SEE0727373, SEE0727374, SEE0727375, SEE0727376, SEE0727377, SEE0727378, SEE0727379, SEE0729334, SEE0729335, SEE0729336, SEE0729337, SEE0729338, SEE0729339, SEE0729340, SEE0729341, SEE0729342, SEE0729343, SEE0729344, SEE0729345, SEE0729346, SEE0729347, SEE0729348, SEE0729349, SEE0729350, SEE0729351, SEE0729352, SEE0729353, SEE0729354, SEE0729355, SEE0729356, SEE0729357, SEE0729358, SEE0729359, SEE0729360, SEE0729361, SEE0729362, SEE0729363, SEE0731254, SEE0731255, SEE0731256, SEE0731257, SEE0731258, SEE0731259, SEE0731260, SEE0731261, SEE0731262, SEE0731263, SEE0732004, SEE0732005, SEE0732006, SEE0732007, SEE0732008, SEE0732009, SEE0732010, SEE0732011, SEE0732012, SEE0732013, SEE0732014, SEE0732015, SEE0732016, SEE0732017, SEE0732018, SEE0732019, SEE0732020, SEE0732021, SEE0732022, SEE0732023, SEE0733306, SEE0733307, SEE0733308, SEE0733309, SEE0733310, SEE0733311, SEE0733312, SEE0733313, SEE0733314, SEE0733315, SEE0733316, SEE0733317, SEE0733318, SEE0733319, SEE0733320, SEE0734347, SEE0734348, SEE0734349, SEE0734350, SEE0734351, SEE0734352, SEE0734353, SEE0734354, SEE0734355, SEE0734356, SEE0734357, SEE0734358, SEE0734359, SEE0734360, SEE0734361, SEE0735406, SEE0735407, SEE0735408, SEE0735409, SEE0735410, SEE0735411, SEE0735412, SEE0735413, SEE0735414, SEE0735415, SEE0735416, SEE0735417, SEE0735418, SEE0735419, SEE0735420, SEE0735421, SEE0735422, SEE0735423, SEE0735424, SEE0735425, SEE0736494, SEE0736495, SEE0736497, SEE0736498, SEE0736499, SEE0736500, SEE0736501, SEE0736502, SEE0736503, SEE0736504, SEE0736505, SEE0736506, SEE0736507, SEE0736508, SEE0736509, SEE0736510, SEE0736511, SEE0736512, SEE0736513, SEE0736514, SEE0736515, SEE0737210, SEE0737211, SEE0737212, SEE0737213, SEE0737214, SEE0737215, SEE0737216, SEE0737217, SEE0737218, SEE0737219, SEE0737220, SEE0737221, SEE0737222, SEE0737223, SEE0737224, SEE0737225, SEE0737226, SEE0737227, SEE0737228, SEE0737229, SEE0737230, SEE0737231, SEE0737232, SEE0737235, SEE0737236, SEE0737238, SEE0737239, SEE0737240, SEE0737241, SEE0737242, SEE0737243, SEE0737244, SEE0737245, SEE0737246, SEE0737247, SEE0737248, SEE0737502, SEE0738281, SEE0738282, SEE0738283, SEE0738284, SEE0738285, SEE0738286, SEE0738287, SEE0738288, SEE0738289, SEE0738290, SEE0738291, SEE0738292, SEE0738293, SEE0738294, SEE0738295, SEE0738296, SEE0738297, SEE0738298, SEE0738299, SEE0738300, SEE0738301, SEE0738302, SEE0738303, SEE0738304, SEE0738305, SEE0738306, SEE0738307, SEE0738308, SEE0738309, SEE0738310, SEE0738311, SEE0738312, SEE0738313, SEE0738314, SEE0738315, SEE0738316, SEE0738317, SEE0738318, SEE0738319, SEE0738320, SEE0738321, SEE0738322, SEE0739123, SEE0739124, SEE0739125, SEE0739126, SEE0739127, SEE0739128, SEE0739129, SEE0739130, SEE0739131, SEE0739132, SEE0739133, SEE0739134, SEE0739135, SEE0739136, SEE0739137, SEE0739138, SEE0739139, SEE0739140, SEE0739141, SEE0739142, SEE0739143, SEE0739144, SEE0739145, SEE0739146, SEE0739147, SEE0739148, SEE0739365, SEE0739366, SEE0739367, SEE0739368, SEE0739369, SEE0739370, SEE0739372, SEE0739373, SEE0739374, SEE0739375, SEE0739376, SEE0739377, SEE0739378, SEE0739379, SEE0739381, SEE0739382, SEE0739384, SEE0739385, SEE0739386, SEE0740636, SEE0740637, SEE0740639, SEE0740641, SEE0740642, SEE0740643, SEE0740644, SEE0740645, SEE0740646, SEE0740647, SEE0740648, SEE0740649, SEE0740650, SEE0740651, SEE0740652, SEE0740653, SEE0740654, SEE0740655, SEE0740656, SEE0740657, SEE0740658, SEE0740659, SEE0740660, SEE0740661, SEE0740662, SEE0740663, SEE0740664, SEE0740665, SEE0741280, SEE0741281, SEE0741282, SEE0741283, SEE0741284, SEE0741285, SEE0741286, SEE0741287, SEE0741288, SEE0741289, SEE0741290, SEE0741291, SEE0741292, SEE0741293, SEE0741294, SEE0741295, SEE0741296, SEE0741297, SEE0741298, SEE0741299, SEE0742820, SEE0742822, SEE0742823, SEE0742824, SEE0742825, SEE0742826, SEE0742827, SEE0742828, SEE0742829, SEE0742830, SEE0742831, SEE0742832, SEE0742833, SEE0742834, SEE0742835, SEE0742836, SEE0742837, SEE0742838, SEE0742839, SEE0742840, SEE0742841, SEE0742842, SEE0742843, SEE0742844, SEE0742845, SEE0742846, SEE0742847, SEE0742848, SEE0742849, SEE0742850, SEE0742851, SEE0742852, SEE0742853, SEE0742854, SEE0744163, SEE0744164, SEE0744165, SEE0744166, SEE0744168, SEE0744169, SEE0744170, SEE0744171, SEE0744172, SEE0744173, SEE0744174, SEE0744175, SEE0744176, SEE0744177, SEE0744178, SEE0744179, SEE0744180, SEE0744181, SEE0744182, SEE0744183, SEE0744184, SEE0744185, SEE0744186, SEE0744187, SEE0744188, SEE0744190, SEE0744191, SEE0744192, SEE0744193, SEE0744194, SEE0744195, SEE0744196, SEE0744197, SEE0745189, SEE0745190, SEE0745191, SEE0745192, SEE0745193, SEE0745194, SEE0745196, SEE0746253, SEE0746254, SEE0746255, SEE0746256, SEE0746257, SEE0746258, SEE0746259, SEE0746260, SEE0746261, SEE0746262, SEE0746263, SEE0746264, SEE0746265, SEE0746266, SEE0746267, SEE0747599, SEE0747600, SEE0747602, SEE0747603, SEE0747604, SEE0747606, SEE0747607, SEE0747608, SEE0747609, SEE0747610, SEE0747611, SEE0747612, SEE0747613, SEE0747614, SEE0748512, SEE0748513, SEE0748514, SEE0748515, SEE0748516, SEE0748517, SEE0748518, SEE0748519, SEE0748521, SEE0748522, SEE0748523, SEE0748524, SEE0748525, SEE0748526, SEE0748527, SEE0748528, SEE0748529, SEE0748530, SEE0748531, SEE0748532, SEE0748533, SEE0748534, SEE0748535, SEE0748536, SEE0748538, SEE0748539, SEE0748540, SEE0748541, SEE0751461, SEE0751462, SEE0751463, SEE0751464, SEE0751465, SEE0751466, SEE0751467, SEE0751468, SEE0751469, SEE0751470, SEE0751471, SEE0751472, SEE0751473, SEE0751474, SEE0751475, SEE0802644, SEE0802645, SEE0802646, SEE0802647, SEE0802648, SEE0802649, SEE0802650, SEE0802652, SEE0802653, SEE0802654, SEE0802655, SEE0802656, SEE0802657, SEE0804156, SEE0804158, SEE0804161, SEE0804162, SEE0804163, SEE0804164, SEE0804165, SEE0804166, SEE0804167, SEE0804168, SEE0804169, SEE0804170, SEE0804171, SEE0804172, SEE0804174, SEE0804175, SEE0804176, SEE0804177, SEE0804178, SEE0804179, SEE0804180, SEE0804182, SEE0804183, SEE0804184, SEE0804185, SEE0804186, SEE0804187, SEE0804188, SEE0804189, SEE0804190, SEE0806118, SEE0806119, SEE0806120, SEE0806121, SEE0806122, SEE0806123, SEE0806124, SEE0806125, SEE0806126, SEE0806128, SEE0806129, SEE0806130, SEE0806131, SEE0806132, SEE0806133, SEE0806135, SEE0806136, SEE0806139, SEE0806140, SEE0806141, SEE0806142, SEE0808062, SEE0808063, SEE0808065, SEE0808066, SEE0808067, SEE0808068, SEE0808069, SEE0808070, SEE0808071, SEE0808072, SEE0808073, SEE0808074, SEE0808075, SEE0808076, SEE0808478, SEE0808479, SEE0808480, SEE0808481, SEE0808482, SEE0808483, SEE0808484, SEE0808485, SEE0808486, SEE0808487, SEE0808488, SEE0808489, SEE0808490, SEE0808491, SEE0808492, SEE0810125, SEE0810126, SEE0810129, SEE0810130, SEE0810131, SEE0810132, SEE0810133, SEE0810134, SEE0810135, SEE0810136, SEE0810137, SEE0810138, SEE0810139; Model KMBB4OSU2FUS, serial #SEE0551022, SEE0551023, SEE0551024, SEE0551025, SEE0551026, SEE0551027, SEE0551036, SEE0551037, SEE0551038, SEE0551039, SEE0551040, SEE0551041, SEE0611218, SEE0611219, SEE0611220, SEE0611221, SEE0611222, SEE0611223, SEE0611224, SEE0611225, SEE0611226, SEE0611227, SEE0611228, SEE0611229, SEE0611230, SEE0611231, SEE0611232, SEE0611233, SEE0611234, SEE0611235, SEE0611236, SEE0611237, SEE0611238, SEE0611239, SEE0611240, SEE0611241, SEE0611242, SEE0611243, SEE0611244, SEE0611245, SEE0611246, SEE0611247, SEE0611248, SEE0611249, SEE0611250, SEE0611251, SEE0611252, SEE0611253, SEE0611254, SEE0611255, SEE0611256, SEE0611257, SEE0611258, SEE0611259, SEE0611260, SEE0611261, SEE0611262, SEE0611263, SEE0611264, SEE0611265, SEE0611266, SEE0611267, SEE0611268, SEE0611269, SEE0611270, SEE0611271, SEE0611272, SEE0611273, SEE0611274, SEE0611275, SEE0611276, SEE0611277, SEE0611278, SEE0611279, SEE0611280, SEE0611281, SEE0611282, SEE0611283, SEE0611284, SEE0611285, SEE0611286, SEE0611287, SEE0611288, SEE0611289, SEE0611290, SEE0611291, SEE0611292, SEE0611293, SEE0611294, SEE0611295, SEE0611296, SEE0611297, SEE0611298, SEE0611299, SEE0611300, SEE0611301, SEE0611302, SEE0611303, SEE0611304, SEE0611305, SEE0611306, SEE0611307, SEE0611308, SEE0611309, SEE0611310, SEE0611311, SEE0611312, SEE0611313, SEE0611314, SEE0611316, SEE0611317, SEE0611318, SEE0611319, SEE0611320, SEE0611321, SEE0613038, SEE0613039, SEE0613040, SEE0613041, SEE0613042, SEE0613043, SEE0613044, SEE0613045, SEE0613046, SEE0613047, SEE0613048, SEE0613049, SEE0613050, SEE0613051, SEE0613052, SEE0613053, SEE0613054, SEE0613055, SEE0613056, SEE0613057, SEE0613058, SEE0613059, SEE0613060, SEE0613061, SEE0613062, SEE0613063, SEE0613064, SEE0613065, SEE0613066, SEE0613067, SEE0613068, SEE0613069, SEE0613070, SEE0613071, SEE0613072, SEE0613073, SEE0613074, SEE0613075, SEE0613076, SEE0613077, SEE0613078, SEE0613079, SEE0613080, SEE0613081, SEE0613082, SEE0613083, SEE0613084, SEE0613085, SEE0613086, SEE0613087, SEE0613088, SEE0613089, SEE0613090, SEE0613091, SEE0613092, SEE0613093, SEE0613308, SEE0613309, SEE0613310, SEE0613311, SEE0613312, SEE0613313, SEE0613314, SEE0613315, SEE0613316, SEE0613317, SEE0613318, SEE0613319, SEE0613320, SEE0613321, SEE0613322, SEE0613323, SEE0613324, SEE0613325, SEE0613326, SEE0613327, SEE0613328, SEE0613329, SEE0613330, SEE0613331, SEE0613332, SEE0613333, SEE0613334, SEE0613335, SEE0613336, SEE0613337, SEE0613338, SEE0613339, SEE0613340, SEE0613341, SEE0613342, SEE0613343, SEE0613344, SEE0613345, SEE0613378, SEE0613379, SEE0613380, SEE0613381, SEE0613382, SEE0613383, SEE0613384, SEE0613385, SEE0613386, SEE0613387, SEE0613388, SEE0613389, SEE0613390, SEE0613391, SEE0613392, SEE0613393, SEE0613394, SEE0613395, SEE0613396, SEE0613397, SEE0613398, SEE0613399, SEE0613400, SEE0613401, SEE0613402, SEE0613403, SEE0613404, SEE0613405, SEE0613406, SEE0613407, SEE0613408, SEE0613409, SEE0613410, SEE0613411, SEE0613412, SEE0613413, SEE0613414, SEE0613415, SEE0613416, SEE0613417, SEE0613418, SEE0613419, SEE0613420, SEE0613421, SEE0613422, SEE0613423, SEE0613424, SEE0613425, SEE0613426, SEE0613427, SEE0613428, SEE0613429, SEE0613430, SEE0613431, SEE0613432, SEE0613433, SEE0613434, SEE0613435, SEE0613436, SEE0613437, SEE0613438, SEE0613439, SEE0613440, SEE0613441, SEE0613442, SEE0613443, SEE0613444, SEE0613445, SEE0613446, SEE0613447, SEE0613448, SEE0613449, SEE0613450, SEE0613451, SEE0613452, SEE0613453, SEE0613454, SEE0613455, SEE0613456, SEE0613457, SEE0613458, SEE0613459, SEE0613460, SEE0613461, SEE0613462, SEE0613463, SEE0613464, SEE0613465, SEE0613466, SEE0613467, SEE0613468, SEE0613469, SEE0613470, SEE0613471, SEE0613472, SEE0613473, SEE0613474, SEE0613475, SEE0613476, SEE0613477, SEE0613478, SEE0613479, SEE0613480, SEE0613481, SEE0613482, SEE0613483, SEE0613484, SEE0613485, SEE0613486, SEE0613487, SEE0613488, SEE0613489, SEE0613490, SEE0613491, SEE0613492, SEE0613493, SEE0613494, SEE0613495, SEE0613496, SEE0613497, SEE0616104, SEE0616105, SEE0616106, SEE0616107, SEE0616108, SEE0616109, SEE0616110, SEE0616111, SEE0616112, SEE0616113, SEE0616114, SEE0616115, SEE0616116, SEE0616117, SEE0616118, SEE0616119, SEE0616120, SEE0616121, SEE0616122, SEE0616123, SEE0616124, SEE0616125, SEE0616126, SEE0616127, SEE0616128, SEE0616129, SEE0616130, SEE0616131, SEE0616132, SEE0616133, SEE0616134, SEE0616135, SEE0616136, SEE0616137, SEE0616138, SEE0616139, SEE0616140, SEE0616141, SEE0616142, SEE0616143, SEE0616144, SEE0616145, SEE0616146, SEE0616147, SEE0616148, SEE0616149, SEE0618111, SEE0618112, SEE0618113, SEE0618114, SEE0618115, SEE0618116, SEE0618117, SEE0618118, SEE0618119, SEE0618120, SEE0618121, SEE0618122, SEE0618123, SEE0618124, SEE0618125, SEE0618126, SEE0618127, SEE0618128, SEE0618129, SEE0618130, SEE0618131, SEE0618132, SEE0618133, SEE0619132, SEE0619133, SEE0619134, SEE0619135, SEE0619136, SEE0619137, SEE0619138, SEE0619139, SEE0619140, SEE0619141, SEE0619142, SEE0619143, SEE0619144, SEE0619145, SEE0619146, SEE0619147, SEE0619148, SEE0619149, SEE0619150, SEE0619151, SEE0619152, SEE0619153, SEE0619154, SEE0619155, SEE0619156, SEE0619157, SEE0619158, SEE0619159, SEE0619160, SEE0619161, SEE0619162, SEE0619163, SEE0619164, SEE0619165, SEE0619166, SEE0619167, SEE0619168, SEE0619169, SEE0619171, SEE0619172, SEE0619173, SEE0619174, SEE0619175, SEE0619176, SEE0619177, SEE0619178, SEE0619179, SEE0619180, SEE0619181, SEE0619182, SEE0619183, SEE0619184, SEE0619185, SEE0619186, SEE0619187, SEE0619188, SEE0619189, SEE0619190, SEE0619191, SEE0619192, SEE0619193, SEE0619194, SEE0619195, SEE0619196, SEE0619197, SEE0619198, SEE0619199, SEE0619200, SEE0619201, SEE0619202, SEE0619203, SEE0619204, SEE0619206, SEE0619207, SEE0619208, SEE0619209, SEE0619210, SEE0619211, SEE0619212, SEE0619213, SEE0619214, SEE0619215, SEE0619216, SEE0619217, SEE0619218, SEE0619219, SEE0619220, SEE0619221, SEE0619222, SEE0619223, SEE0619224, SEE0621313, SEE0621314, SEE0621315, SEE0621316, SEE0621317, SEE0621318, SEE0621320, SEE0621321, SEE0621322, SEE0621323, SEE0621324, SEE0621325, SEE0621326, SEE0621327, SEE0621328, SEE0621329, SEE0621330, SEE0621331, SEE0621332, SEE0621333, SEE0621334, SEE0621335, SEE0621336, SEE0621337, SEE0621338, SEE0621339, SEE0621340, SEE0621341, SEE0621342, SEE0621343, SEE0621344, SEE0621345, SEE0621346, SEE0621347, SEE0621348, SEE0621349, SEE0621350, SEE0621351, SEE0621352, SEE0621353, SEE0621354, SEE0621355, SEE0621356, SEE0621357, SEE0621358, SEE0621359, SEE0621361, SEE0621362, SEE0621363, SEE0621364, SEE0621365, SEE0621366, SEE0621367, SEE0621368, SEE0621369, SEE0621370, SEE0621371, SEE0621372, SEE0621373, SEE0621374, SEE0621375, SEE0621376, SEE0621377, SEE0622256, SEE0622262, SEE0622266, SEE0622269, SEE0623090, SEE0623091, SEE0623092, SEE0623093, SEE0623094, SEE0623095, SEE0623096, SEE0623097, SEE0623098, SEE0623100, SEE0623101, SEE0623102, SEE0623103, SEE0623104, SEE0623105, SEE0623106, SEE0623107, SEE0623108, SEE0623109, SEE0623110, SEE0623111, SEE0623112, SEE0623113, SEE0627469, SEE0627470, SEE0627471, SEE0627472, SEE0627473, SEE0627474, SEE0627475, SEE0627476, SEE0627477, SEE0627478, SEE0627479, SEE0627480, SEE0627481, SEE0627482, SEE0627483, SEE0627484, SEE0627485, SEE0627486, SEE0627487, SEE0627488, SEE0627489, SEE0627490, SEE0627491, SEE0627492, SEE0627493, SEE0627494, SEE0627495, SEE0627496, SEE0627498, SEE0627499, SEE0627500, SEE0627501, SEE0627502, SEE0627503, SEE0627504, SEE0627505, SEE0627506, SEE0627507, SEE0627508, SEE0627509, SEE0627510, SEE0627511, SEE0627512, SEE0627513, SEE0627514, SEE0627515, SEE0627516, SEE0627517, SEE0627518, SEE0627519, SEE0627520, SEE0627521, SEE0627522, SEE0627523, SEE0627524, SEE0627525, SEE0627526, SEE0627527, SEE0627528, SEE0627529, SEE0627530, SEE0627531, SEE0627532, SEE0627533, SEE0627534, SEE0627535, SEE0627536, SEE0627537, SEE0627538, SEE0627539, SEE0627540, SEE0627541, SEE0627542, SEE0627543, SEE0627544, SEE0628306, SEE0628307, SEE0628308, SEE0628309, SEE0628310, SEE0628313, SEE0628316, SEE0628317, SEE0628318, SEE0628321, SEE0628325, SEE0628327, SEE0628328, SEE0628329, SEE0628330, SEE0628331, SEE0628332, SEE0628333, SEE0628342, SEE0628343, SEE0628344, SEE0628345, SEE0628346, SEE0628347, SEE0628348, SEE0628349, SEE0628350, SEE0628351, SEE0628352, SEE0628353, SEE0628354, SEE0628355, SEE0628356, SEE0628357, SEE0628358, SEE0628359, SEE0628360, SEE0628361, SEE0628362, SEE0628363, SEE0628364, SEE0628365, SEE0628366, SEE0628367, SEE0628368, SEE0628369, SEE0628370, SEE0628371, SEE0628372, SEE0628373, SEE0628374, SEE0628375, SEE0628376, SEE0628377, SEE0628378, SEE0628379, SEE0628381, SEE0628382, SEE0628383, SEE0628386, SEE0628388, SEE0628391, SEE0628399, SEE0628401, SEE0628402, SEE0628403, SEE0628404, SEE0629128, SEE0629129, SEE0629130, SEE0629131, SEE0629132, SEE0629135, SEE0629136, SEE0629137, SEE0629138, SEE0629139, SEE0629141, SEE0629142, SEE0629143, SEE0629173, SEE0629174, SEE0629175, SEE0629177, SEE0631178, SEE0631179, SEE0631180, SEE0631181, SEE0631184, SEE0631185, SEE0631186, SEE0631187, SEE0631188, SEE0631189, SEE0631190, SEE0631191, SEE0631192, SEE0631194, SEE0631195, SEE0631196, SEE0639007, SEE0639008, SEE0639009, SEE0639010, SEE0639011, SEE0639012, SEE0639013, SEE0639014, SEE0639015, SEE0639016, SEE0639017, SEE0639018, SEE0639019, SEE0639020, SEE0639021, SEE0639022, SEE0639023, SEE0639024, SEE0639025, SEE0639026, SEE0639027, SEE0639028, SEE0639029, SEE0639030, SEE0639031, SEE0639032, SEE0639033, SEE0639034, SEE0639035, SEE0640152, SEE0640153, SEE0640154, SEE0640155, SEE0640156, SEE0640157, SEE0640158, SEE0640159, SEE0640160, SEE0640161, SEE0640162, SEE0640163, SEE0640164, SEE0640165, SEE0640166, SEE0640167, SEE0640168, SEE0640169, SEE0640170, SEE0640171, SEE0640172, SEE0640173, SEE0640174, SEE0640175, SEE0640176, SEE0640177, SEE0640178, SEE0640179, SEE0640180, SEE0640181, SEE0640182, SEE0640183, SEE0640184, SEE0640185, SEE0640186, SEE0642587, SEE0642588, SEE0642589, SEE0642590, SEE0642591, SEE0642592, SEE0642593, SEE0642594, SEE0642595, SEE0642596, SEE0642597, SEE0642598, SEE0642599, SEE0642600, SEE0642601, SEE0642602, SEE0642603, SEE0642604, SEE0642605, SEE0642606, SEE0644139, SEE0644140, SEE0644141, SEE0644142, SEE0644143, SEE0644144, SEE0644145, SEE0644146, SEE0644147, SEE0644148, SEE0644149, SEE0644150, SEE0644151, SEE0644152, SEE0644153, SEE0644155, SEE0644156, SEE0644157, SEE0644158, SEE0644159, SEE0644160, SEE0644161, SEE0644162, SEE0644163, SEE0644164, SEE0644165, SEE0644166, SEE0644167, SEE0644168, SEE0644169, SEE0644170, SEE0644171, SEE0644172, SEE0644173, SEE0644174, SEE0644175, SEE0644176, SEE0644177, SEE0644178, SEE0644179, SEE0645263, SEE0645264, SEE0645265, SEE0645266, SEE0645267, SEE0645268, SEE0645269, SEE0645270, SEE0645271, SEE0645272, SEE0645273, SEE0645274, SEE0645275, SEE0645276, SEE0645277, SEE0647328, SEE0647329, SEE0647330, SEE0647331, SEE0647332, SEE0647334, SEE0647335, SEE0647336, SEE0647337, SEE0647338, SEE0647339, SEE0647340, SEE0647341, SEE0647342, SEE0647343, SEE0647344, SEE0647345, SEE0647346, SEE0647347, SEE0647348, SEE0647349, SEE0647350, SEE0647351, SEE0647352, SEE0647353, SEE0647354, SEE0647355, SEE0647356, SEE0647357, SEE0647358, SEE0647359, SEE0647360, SEE0647361, SEE0647362, SEE0647363, SEE0647364, SEE0647365, SEE0647366, SEE0647367, SEE0647368, SEE0647369, SEE0647370, SEE0647371, SEE0647372, SEE0647373, SEE0647374, SEE0647375, SEE0647376, SEE0647377, SEE0702061, SEE0702062, SEE0702063, SEE0702064, SEE0702065, SEE0702066, SEE0702067, SEE0702068, SEE0702069, SEE0702070, SEE0702071, SEE0702072, SEE0702073, SEE0702074, SEE0702075, SEE0703241, SEE0703242, SEE0703243, SEE0703244, SEE0703245, SEE0703246, SEE0703247, SEE0703248, SEE0703249, SEE0703250, SEE0703251, SEE0703252, SEE0703253, SEE0703254, SEE0703255, SEE0704523, SEE0704524, SEE0704525, SEE0704526, SEE0704527, SEE0704528, SEE0704529, SEE0704530, SEE0704531, SEE0704532, SEE0704533, SEE0704534, SEE0704535, SEE0704536, SEE0704537, SEE0704538, SEE0704539, SEE0704540, SEE0704541, SEE0704542, SEE0704543, SEE0704544, SEE0704545, SEE0704546, SEE0704547, SEE0707099, SEE0707100, SEE0707101, SEE0707102, SEE0707103, SEE0707104, SEE0707105, SEE0707106, SEE0707107, SEE0707108, SEE0707109, SEE0707110, SEE0707111, SEE0707112, SEE0707113, SEE0707114, SEE0707115, SEE0707116, SEE0707117, SEE0707118, SEE0707119, SEE0707120, SEE0707121, SEE0707122, SEE0707123, SEE0707124, SEE0707125, SEE0707126, SEE0707127, SEE0707128, SEE0712691, SEE0712692, SEE0712693, SEE0712694, SEE0712695, SEE0714101, SEE0714102, SEE0714103, SEE0714104, SEE0714105, SEE0714106, SEE0714107, SEE0714108, SEE0714109, SEE0714110, SEE0714111, SEE0714112, SEE0714113, SEE0714114, SEE0714115, SEE0714116, SEE0714117, SEE0714118, SEE0714119, SEE0714120, SEE0714121, SEE0714122, SEE0714123, SEE0714124, SEE0714125, SEE0715201, SEE0715202, SEE0715203, SEE0715204, SEE0715205, SEE0715206, SEE0715207, SEE0715208, SEE0715209, SEE0715210, SEE0715211, SEE0715212, SEE0715213, SEE0715214, SEE0715215, SEE0715216, SEE0715217, SEE0715218, SEE0715219, SEE0715220, SEE0715221, SEE0715222, SEE0715223, SEE0715224, SEE0715225, SEE0715226, SEE0715227, SEE0715228, SEE0715229, SEE0715230, SEE0715231, SEE0715232, SEE0715233, SEE0715234, SEE0715235, SEE0715236, SEE0715237, SEE0715238, SEE0715239, SEE0715240, SEE0715241, SEE0715242, SEE0715243, SEE0715244, SEE0715245, SEE0716318, SEE0716319, SEE0716320, SEE0716321, SEE0716322, SEE0716323, SEE0716324, SEE0716325, SEE0716326, SEE0716327, SEE0716328, SEE0716329, SEE0716330, SEE0716331, SEE0716332, SEE0716333, SEE0716334, SEE0716335, SEE0716336, SEE0716337, SEE0716338, SEE0716339, SEE0716340, SEE0716341, SEE0716342, SEE0716343, SEE0716344, SEE0716345, SEE0716346, SEE0716347, SEE0716348, SEE0716349, SEE0716350, SEE0716351, SEE0716352, SEE0716353, SEE0716354, SEE0716355, SEE0716356, SEE0716357, SEE0716358, SEE0716359, SEE0717117, SEE0717118, SEE0717119, SEE0717120, SEE0717121, SEE0717122, SEE0717123, SEE0717124, SEE0717125, SEE0717126, SEE0717127, SEE0717128, SEE0717129, SEE0717130, SEE0717131, SEE0717132, SEE0717133, SEE0717134, SEE0717135, SEE0717136, SEE0717137, SEE0717138, SEE0717139, SEE0717140, SEE0717141, SEE0717142, SEE0717143, SEE0717144, SEE0717146, SEE0717147, SEE0717148, SEE0717149, SEE0717150, SEE0717151, SEE0717152, SEE0717153, SEE0717154, SEE0717155, SEE0718036, SEE0718037, SEE0718038, SEE0718039, SEE0718040, SEE0718041, SEE0718042, SEE0718043, SEE0718044, SEE0718045, SEE0718046, SEE0718047, SEE0718048, SEE0718049, SEE0718050, SEE0718051, SEE0718052, SEE0718053, SEE0718054, SEE0718055, SEE0718056, SEE0718057, SEE0718058, SEE0718059, SEE0718060, SEE0719072, SEE0719073, SEE0719074, SEE0719075, SEE0719076, SEE0719077, SEE0719078, SEE0719079, SEE0719080, SEE0719081, SEE0719082, SEE0719083, SEE0719084, SEE0719085, SEE0719086, SEE0719087, SEE0719088, SEE0719089, SEE0719090, SEE0719091, SEE0719581, SEE0719582, SEE0719583, SEE0719584, SEE0719585, SEE0719586, SEE0719587, SEE0719588, SEE0719589, SEE0719590, SEE0719591, SEE0719592, SEE0719593, SEE0719594, SEE0719595, SEE0719596, SEE0719597, SEE0719598, SEE0719599, SEE0719600, SEE0722021, SEE0722022, SEE0722023, SEE0722024, SEE0722025, SEE0722026, SEE0722027, SEE0722028, SEE0722029, SEE0722030, SEE0722031, SEE0722032, SEE0722033, SEE0722034, SEE0722035, SEE0722036, SEE0722037, SEE0722038, SEE0722039, SEE0722040, SEE0722247, SEE0722248, SEE0722249, SEE0722250, SEE0722251, SEE0722252, SEE0722253, SEE0722254, SEE0722255, SEE0722256, SEE0722257, SEE0722258, SEE0722259, SEE0722260, SEE0722261, SEE0722262, SEE0722263, SEE0722264, SEE0722265, SEE0722266, SEE0722267, SEE0722268, SEE0722269, SEE0722270, SEE0722271, SEE0722272, SEE0722273, SEE0722274, SEE0722275, SEE0722276, SEE0722277, SEE0722278, SEE0722279, SEE0722280, SEE0722281, SEE0722282, SEE0722283, SEE0722284, SEE0722285, SEE0722286, SEE0722287, SEE0722288, SEE0722289, SEE0722290, SEE0722291, SEE0722292, SEE0722293, SEE0722294, SEE0722295, SEE0722296, SEE0722297, SEE0722298, SEE0722299, SEE0722300, SEE0722301, SEE0722302, SEE0722303, SEE0722304, SEE0722305, SEE0722306, SEE0722307, SEE0722308, SEE0722309, SEE0722310, SEE0722311, SEE0722312, SEE0722313, SEE0722314, SEE0722315, SEE0722316, SEE0722317, SEE0722318, SEE0722319, SEE0722320, SEE0722321, SEE0722322, SEE0722323, SEE0722324, SEE0722325, SEE0722326, SEE0722327, SEE0722328, SEE0722329, SEE0722336, SEE0722337, SEE0722338, SEE0722339, SEE0722340, SEE0722341, SEE0722342, SEE0722343, SEE0722344, SEE0722345, SEE0722346, SEE0722347, SEE0722348, SEE0722349, SEE0722350, SEE0722351, SEE0722352, SEE0722353, SEE0722354, SEE0722355, SEE0722356, SEE0722357, SEE0722358, SEE0722359, SEE0722360, SEE0722361, SEE0722362, SEE0722363, SEE0722364, SEE0722365, SEE0722366, SEE0722367, SEE0722368, SEE0722369, SEE0722370, SEE0722372, SEE0722373, SEE0722374, SEE0722375, SEE0727183, SEE0727184, SEE0727185, SEE0727186, SEE0727187, SEE0727189, SEE0727190, SEE0727192, SEE0727193, SEE0727194, SEE0727246, SEE0727247, SEE0727248, SEE0727249, SEE0727250, SEE0727251, SEE0727252, SEE0727253, SEE0727254, SEE0727255, SEE0727256, SEE0727257, SEE0727258, SEE0727259, SEE0729389, SEE0729390, SEE0729391, SEE0729392, SEE0729393, SEE0729394, SEE0729395, SEE0729396, SEE0729397, SEE0729398, SEE0729399, SEE0729400, SEE0729401, SEE0729402, SEE0729403, SEE0729404, SEE0729405, SEE0729406, SEE0729407, SEE0729408, SEE0729409, SEE0729410, SEE0729411, SEE0729412, SEE0729413, SEE0729414, SEE0729415, SEE0729416, SEE0729417, SEE0729418, SEE0729419, SEE0729420, SEE0729421, SEE0729422, SEE0729423, SEE0732076, SEE0732077, SEE0732078, SEE0732079, SEE0732080, SEE0732081, SEE0732082, SEE0732083, SEE0732084, SEE0732085, SEE0732086, SEE0732087, SEE0732088, SEE0732089, SEE0732090, SEE0732091, SEE0732092, SEE0732093, SEE0732094, SEE0732095, SEE0732096, SEE0732097, SEE0732098, SEE0732099, SEE0732100, SEE0732156, SEE0732157, SEE0732158, SEE0732159, SEE0732160, SEE0732161, SEE0732162, SEE0732163, SEE0732164, SEE0732165, SEE0732166, SEE0732167, SEE0732168, SEE0732169, SEE0732170, SEE0732171, SEE0732172, SEE0732173, SEE0732174, SEE0732175, SEE0732176, SEE0732177, SEE0732178, SEE0732179, SEE0732180, SEE0732181, SEE0732182, SEE0732183, SEE0732184, SEE0732185, SEE0733321, SEE0733322, SEE0733323, SEE0733324, SEE0733325, SEE0733326, SEE0733327, SEE0733328, SEE0733329, SEE0733330, SEE0733331, SEE0733332, SEE0733333, SEE0733334, SEE0733335, SEE0733336, SEE0733337, SEE0733338, SEE0733339, SEE0733340, SEE0734362, SEE0734363, SEE0734364, SEE0734365, SEE0734366, SEE0734367, SEE0734368, SEE0734369, SEE0734370, SEE0734371, SEE0734372, SEE0734373, SEE0734374, SEE0734375, SEE0734376, SEE0734377, SEE0734378, SEE0734379, SEE0734380, SEE0734381, SEE0737585, SEE0737586, SEE0737587, SEE0737588, SEE0737589, SEE0737590, SEE0737591, SEE0737592, SEE0737593, SEE0737594, SEE0737595, SEE0737596, SEE0737597, SEE0737598, SEE0737599, SEE0737601, SEE0737602, SEE0737603, SEE0737604, SEE0737605, SEE0737606, SEE0739273, SEE0739274, SEE0739275, SEE0739276, SEE0739277, SEE0739278, SEE0739279, SEE0739280, SEE0739281, SEE0739282, SEE0739283, SEE0739284, SEE0739285, SEE0739286, SEE0739287, SEE0739288, SEE0739289, SEE0739290, SEE0739291, SEE0739292, SEE0739406, SEE0739407, SEE0739408, SEE0739409, SEE0739410, SEE0740668, SEE0740670, SEE0740671, SEE0740672, SEE0740673, SEE0743099, SEE0743100, SEE0743101, SEE0743102, SEE0743103, SEE0743104, SEE0743105, SEE0743106, SEE0743107, SEE0745347, SEE0745348, SEE0745349, SEE0745350, SEE0745351, SEE0745352, SEE0745353, SEE0745354, SEE0745355, SEE0745356, SEE0745360, SEE0745362, SEE0745363, SEE0745364, SEE0745365, SEE0745366, SEE0745367, SEE0745368, SEE0745369, SEE0745370, SEE0745371, SEE0745372, SEE0745373, SEE0745374, SEE0745375, SEE0745376, SEE0745377, SEE0745378, SEE0745379, SEE0745380, SEE0745381, SEE0745382, SEE0745383, SEE0745384, SEE0745385, SEE0745386, SEE0745387, SEE0745388, SEE0745389, SEE0745390, SEE0745391, SEE0745392, SEE0746437, SEE0746438, SEE0746439, SEE0746440, SEE0746441, SEE0746442, SEE0746443, SEE0746444, SEE0746445, SEE0746446, SEE0746447, SEE0746448, SEE0746449, SEE0746450, SEE0746451, SEE0746452, SEE0746453, SEE0746454, SEE0746455, SEE0746456, SEE0746457, SEE0746458, SEE0746459, SEE0746460, SEE0746461, SEE0748542, SEE0748543, SEE0748544, SEE0748545, SEE0748546, SEE0748548, SEE0748549, SEE0748550, SEE0748551, SEE0748552, SEE0748554, SEE0748555, SEE0748556, SEE0748557, SEE0748558, SEE0748559, SEE0748560, SEE0748561, SEE0751476, SEE0751477, SEE0751478, SEE0751479, SEE0751480, SEE0751481, SEE0751482, SEE0751483, SEE0751484, SEE0751485, SEE0751486, SEE0751487, SEE0751488, SEE0751489, SEE0804452, SEE0804453, SEE0804454, SEE0804455, SEE0804456, SEE0804457, SEE0804458, SEE0804459, SEE0804460, SEE0804461, SEE0804462, SEE0804463, SEE0804464, SEE0804465, SEE0804466, SEE0806213, SEE0806214, SEE0806215, SEE0806216, SEE0806217, SEE0806218, SEE0806219, SEE0806220, SEE0806221, SEE0806222, SEE0808203, SEE0808204, SEE0808205, SEE0808206, SEE0808209, SEE0808210, SEE0808211, SEE0808212, SEE0809391, SEE0809392, SEE0809393, SEE0809394, SEE0809395, SEE0809396, SEE0809397, SEE0809398; Model KMBB4PLX2FUS, serial #SEE0548229, SEE0548230, SEE0548231, SEE0548232, SEE0550416, SEE0550418, SEE0621198, SEE0621199, SEE0621200, SEE0621201, SEE0621202, SEE0621203, SEE0621204, SEE0621205, SEE0621206, SEE0621207, SEE0621208, SEE0621209, SEE0621210, SEE0621211, SEE0621212, SEE0621213, SEE0621214, SEE0621215, SEE0621216, SEE0621217, SEE0621218, SEE0621219, SEE0621220, SEE0621221, SEE0621222, SEE0621223, SEE0621224, SEE0622403, SEE0622404, SEE0622405, SEE0622406, SEE0622407, SEE0622474, SEE0622475, SEE0622476, SEE0622477, SEE0622478, SEE0622479, SEE0622480, SEE0622481, SEE0622482, SEE0622483, SEE0622484, SEE0622485, SEE0622486, SEE0622487, SEE0622488, SEE0622489, SEE0622490, SEE0622491, SEE0622492, SEE0622493, SEE0622494, SEE0622495, SEE0622496, SEE0622497, SEE0622498, SEE0622499, SEE0622500, SEE0622501, SEE0622502, SEE0622503, SEE0622504, SEE0622505, SEE0622506, SEE0622507, SEE0622508, SEE0622509, SEE0622510, SEE0622511, SEE0622512, SEE0622513, SEE0622514, SEE0622515, SEE0622516, SEE0622517, SEE0622518, SEE0622519, SEE0622520, SEE0622521, SEE0622522, SEE0622523, SEE0622524, SEE0622525, SEE0622526, SEE0622527, SEE0622528, SEE0622529, SEE0622530, SEE0622531, SEE0622532, SEE0622533, SEE0622534, SEE0622535, SEE0622536, SEE0629191, SEE0629192, SEE0629193, SEE0629194, SEE0629195, SEE0629196, SEE0629197, SEE0629198, SEE0629199, SEE0629200, SEE0629201, SEE0629202, SEE0629203, SEE0630432, SEE0631162, SEE0631163, SEE0631164, SEE0631165, SEE0631166, SEE0631167, SEE0631168, SEE0631169, SEE0631170, SEE0631171, SEE0631172, SEE0631173, SEE0631174, SEE0631300, SEE0631301, SEE0631302, SEE0631303, SEE0631304, SEE0631305, SEE0631306, SEE0633097, SEE0633098, SEE0633099, SEE0633100, SEE0633101, SEE0633102, SEE0633103, SEE0633104, SEE0633105, SEE0633106, SEE0633107, SEE0633108, SEE0633109, SEE0633110, SEE0637256, SEE0637257, SEE0637258, SEE0637259, SEE0637260, SEE0637261, SEE0637262, SEE0637263, SEE0637264, SEE0637265, SEE0637266, SEE0637267, SEE0637268, SEE0637269, SEE0637270, SEE0637271, SEE0637272, SEE0637273, SEE0637275, SEE0637276, SEE0637277, SEE0637278, SEE0637279, SEE0637280, SEE0640187, SEE0640188, SEE0640189, SEE0640190, SEE0640191, SEE0640192, SEE0640193, SEE0640194, SEE0640195, SEE0640196, SEE0640197, SEE0640198, SEE0640199, SEE0640200, SEE0640201, SEE0643572, SEE0643573, SEE0643574, SEE0643575, SEE0643576, SEE0643577, SEE0643578, SEE0643579, SEE0643580, SEE0643581, SEE0645665, SEE0645666, SEE0645667, SEE0645668, SEE0645669, SEE0645670, SEE0645671, SEE0645672, SEE0645673, SEE0645674, SEE0645675, SEE0645676, SEE0645677, SEE0645678, SEE0645679, SEE0651283, SEE0651284, SEE0651285, SEE0651286, SEE0651287, SEE0651288, SEE0651289, SEE0651290, SEE0651291, SEE0651292, SEE0701508, SEE0701509, SEE0701510, SEE0701511, SEE0701512, SEE0701513, SEE0701514, SEE0701515, SEE0706051, SEE0706052, SEE0706053, SEE0706054, SEE0706055, SEE0706056, SEE0706057, SEE0706058, SEE0706059, SEE0706060, SEE0706061, SEE0706062, SEE0706063, SEE0706064, SEE0706065, SEE0706066, SEE0706067, SEE0706068, SEE0706069, SEE0706070, SEE0708254, SEE0708255, SEE0708256, SEE0708257, SEE0708258, SEE0708259, SEE0708260, SEE0708261, SEE0711271, SEE0711272, SEE0711273, SEE0711274, SEE0711275, SEE0711276, SEE0711277, SEE0711278, SEE0712696, SEE0712697, SEE0712698, SEE0712699, SEE0713279, SEE0713280, SEE0713281, SEE0713282, SEE0719054, SEE0719055, SEE0719056, SEE0719057, SEE0719058, SEE0719059, SEE0719060, SEE0719061, SEE0719062, SEE0719063, SEE0719064, SEE0719065, SEE0719066, SEE0719067, SEE0719068, SEE0719069, SEE0719070, SEE0719071, SEE0719603, SEE0719604, SEE0719605, SEE0719606, SEE0720214, SEE0720215, SEE0720216, SEE0720217, SEE0723065, SEE0723066, SEE0723067, SEE0723068, SEE0723069, SEE0723070, SEE0723071, SEE0723072, SEE0723073, SEE0723074, SEE0723075, SEE0723076, SEE0723077, SEE0723078, SEE0723079, SEE0723080, SEE0723081, SEE0723082, SEE0723083, SEE0723084, SEE0723085, SEE0723086, SEE0723087, SEE0723088, SEE0723089, SEE0723090, SEE0723091, SEE0723092, SEE0723093, SEE0723094, SEE0723095, SEE0723096, SEE0723097, SEE0723098, SEE0723099, SEE0723100, SEE0723101, SEE0723102, SEE0723103, SEE0723104, SEE0723105, SEE0723106, SEE0723107, SEE0723108, SEE0723109, SEE0723110, SEE0723111, SEE0723112, SEE0723113, SEE0727167, SEE0727168, SEE0727169, SEE0727170, SEE0727171, SEE0727172, SEE0727173, SEE0727174, SEE0727175, SEE0727176, SEE0727177, SEE0727178, SEE0728160, SEE0728161, SEE0728162, SEE0728163, SEE0731308, SEE0731309, SEE0731310, SEE0731311, SEE0731312, SEE0731313, SEE0731314, SEE0731315, SEE0733341, SEE0733342, SEE0733343, SEE0733344, SEE0733345, SEE0733346, SEE0733347, SEE0733348, SEE0733349, SEE0733350, SEE0733351, SEE0733352, SEE0734433, SEE0734434, SEE0734435, SEE0734436, SEE0734437, SEE0734438, SEE0734439, SEE0734440, SEE0734441, SEE0734442, SEE0736001, SEE0736002, SEE0736003, SEE0736004, SEE0736005, SEE0736006, SEE0736007, SEE0736008, SEE0736009, SEE0736010, SEE0736011, SEE0736012, SEE0736013, SEE0736014, SEE0736368, SEE0736369, SEE0736370, SEE0736371, SEE0736372, SEE0737249, SEE0737250, SEE0737251, SEE0737252, SEE0737253, SEE0738323, SEE0738324, SEE0738325, SEE0738326, SEE0738327, SEE0739149, SEE0739150, SEE0739151, SEE0739152, SEE0739153, SEE0740524, SEE0740525, SEE0740526, SEE0740527, SEE0740528, SEE0740529, SEE0740530, SEE0740531, SEE0740532, SEE0740533, SEE0740534, SEE0740535, SEE0740536, SEE0740537, SEE0740538, SEE0741300, SEE0741301, SEE0741302, SEE0741303, SEE0741304, SEE0741305, SEE0741306, SEE0741307, SEE0741308, SEE0741309, SEE0741310, SEE0741311, SEE0741312, SEE0741313, SEE0741314, SEE0743109, SEE0743110, SEE0743111, SEE0743112, SEE0743113, SEE0743114, SEE0743115, SEE0743116, SEE0743117, SEE0743118, SEE0744038, SEE0744039, SEE0744040, SEE0744041, SEE0744042, SEE0744043, SEE0744044, SEE0744045, SEE0744046, SEE0744047, SEE0745223, SEE0745224, SEE0745225, SEE0745226, SEE0745227, SEE0745228, SEE0745229, SEE0745230, SEE0745231, SEE0749530, SEE0749531, SEE0749532, SEE0749533, SEE0749534, SEE0751081, SEE0751082, SEE0751083, SEE0751084, SEE0751085, SEE0751086, SEE0751087, SEE0751088, SEE0751089, SEE0751090, SEE0751091, SEE0751092, SEE0751093, SEE0751094, SEE0751095, SEE0802659, SEE0802660, SEE0802661, SEE0802662, SEE0802663, SEE0803272, SEE0803273, SEE0803274, SEE0803275, SEE0803276, SEE0803277, SEE0803278, SEE0803279, SEE0804414, SEE0804415, SEE0804416, SEE0804417, SEE0804418, SEE0804419, SEE0804420, SEE0804421, SEE0806143, SEE0806144, SEE0806145, SEE0806146, SEE0806147, SEE0806148, SEE0806149, SEE0806150, SEE0806151, SEE0806152, SEE0808213, SEE0808214, SEE0808215, SEE0808216, SEE0808217, SEE0808218, SEE0808219, SEE0808220, SEE0808531, SEE0808532, SEE0808533, SEE0808534, SEE0808535, SEE0808536, SEE0808537, SEE0808538, SEE0810117, SEE0810118, SEE0810119, SEE0810120, SEE0810121, SEE0810122, SEE0810123, SEE0810124; Model KMBB4PSX2FUS, serial #SEE0550414, SEE0550420, SEE0622622, SEE0622623, SEE0622624, SEE0622625, SEE0622626, SEE0622627, SEE0622628, SEE0622629, SEE0622630, SEE0622631, SEE0622632, SEE0622633, SEE0622634, SEE0622635, SEE0622636, SEE0622637, SEE0622638, SEE0622639, SEE0622640, SEE0622641, SEE0622642, SEE0622643, SEE0622644, SEE0622645, SEE0622646, SEE0622647, SEE0622648, SEE0622649, SEE0624523, SEE0624524, SEE0624525, SEE0624526, SEE0624527, SEE0624528, SEE0624530, SEE0624531, SEE0624532, SEE0624533, SEE0624534, SEE0624535, SEE0624536, SEE0624537, SEE0624538, SEE0624539, SEE0624540, SEE0624541, SEE0624542, SEE0624544, SEE0624545, SEE0624546, SEE0624547, SEE0624548, SEE0624549, SEE0624550, SEE0624551, SEE0624552, SEE0624553, SEE0624554, SEE0624555, SEE0624556, SEE0624557, SEE0624558, SEE0624559, SEE0624560, SEE0624561, SEE0624562, SEE0624563, SEE0624564, SEE0624565, SEE0624566, SEE0628336, SEE0628337, SEE0628338, SEE0628339, SEE0628340, SEE0628341, SEE0629204, SEE0629205, SEE0629206, SEE0629207, SEE0629208, SEE0629209, SEE0629210, SEE0629211, SEE0629212, SEE0629213, SEE0629214, SEE0629215, SEE0629216, SEE0629217, SEE0629218, SEE0629219, SEE0629220, SEE0629221, SEE0629222, SEE0629223, SEE0631116, SEE0631117, SEE0631118, SEE0631119, SEE0631120, SEE0631121, SEE0631122, SEE0631123, SEE0631124, SEE0631125, SEE0631126, SEE0631127, SEE0631128, SEE0631129, SEE0631130, SEE0631131, SEE0631132, SEE0631133, SEE0631134, SEE0631135, SEE0631136, SEE0631137, SEE0631138, SEE0631139, SEE0631140, SEE0631141, SEE0631142, SEE0631143, SEE0631144, SEE0631145, SEE0631146, SEE0631147, SEE0631148, SEE0631149, SEE0631150, SEE0631151, SEE0631307, SEE0631308, SEE0631309, SEE0631310, SEE0631311, SEE0631312, SEE0631313, SEE0647508, SEE0647509, SEE0647510, SEE0647511, SEE0647512, SEE0648644, SEE0648645, SEE0648646, SEE0648647, SEE0648648, SEE0648649, SEE0648650, SEE0648651, SEE0648652, SEE0648653, SEE0648654, SEE0648655, SEE0648656, SEE0648657, SEE0650041, SEE0650042, SEE0650043, SEE0650044, SEE0650045, SEE0650046, SEE0650047, SEE0650048, SEE0650049, SEE0650050, SEE0650051, SEE0650052, SEE0650053, SEE0650055, SEE0702269, SEE0702270, SEE0702271, SEE0702272, SEE0702273, SEE0702274, SEE0702275, SEE0706071, SEE0706072, SEE0706073, SEE0706074, SEE0706075, SEE0706076, SEE0706077, SEE0706078, SEE0706079, SEE0706080, SEE0720001, SEE0720002, SEE0720003, SEE0720004, SEE0720005, SEE0720006, SEE0720007, SEE0720008, SEE0720009, SEE0720010, SEE0720011, SEE0720012, SEE0720013, SEE0727380, SEE0727381, SEE0727382, SEE0727383, SEE0727385, SEE0727386, SEE0727387, SEE0727388, SEE0727389, SEE0731283, SEE0731285, SEE0731286, SEE0731287, SEE0731288, SEE0731289, SEE0733353, SEE0733354, SEE0733355, SEE0733356, SEE0733357, SEE0733358, SEE0733359, SEE0733360, SEE0734444, SEE0734445, SEE0734446, SEE0734447, SEE0734448, SEE0734449, SEE0735633, SEE0735634, SEE0735635, SEE0735636, SEE0735637, SEE0735638, SEE0735639, SEE0735640, SEE0736374, SEE0736375, SEE0736377, SEE0736378, SEE0736379, SEE0737312, SEE0737313, SEE0737314, SEE0738025, SEE0738026, SEE0738027, SEE0738588, SEE0738590, SEE0740676, SEE0740677, SEE0740679, SEE0740680, SEE0740681, SEE0740682, SEE0740683, SEE0745576, SEE0745577, SEE0745578, SEE0745579, SEE0745580, SEE0745581, SEE0745582; Model KMBB4QLU2FUS, serial #SEE0551030, SEE0551032, SEE0551033, SEE0551034, SEE0551035; and Model KMBB4RLX2FUS, serial #SEE0550419
Recalling Firm/
Manufacturer
Arjo, Inc.
50 Gary Ave
Roselle, Illinois 60172-1605
For Additional Information Contact Ms. Traci Giovenco
630-307-2756
Reason for
Recall
There is the potential for unintended dislocation of the Lock and Load hanger assembly (Combi Hanger) from the t-bar attached to the lifter jib of the Maxi Move patient lift.
Action Arjo sent the Field Safety Notice (FSN) dated 8/11/08 to all Maxi Move with Lock & Load system (Combi Hanger) customers on 9/8/08. The accounts were informed of the problem unintended dislocation of the hanger and possible detachment during patient transfer. The notice provided details, including serial number ranges, to identify the affected patient lifts affected, and instructed the accounts to re-train caregivers on how to ensure the hanger is properly attached to the T-bar before and during the initial phase of all lifting operations. Included with the FSN was a package containing a repair/replacement kit for the bushings in the T-bar with instructions, and updated copies of the Operating and Product Care instructions as well as a Preventative Maintenance Schedule manuals. The accounts were requested to complete and return to Arjo Inc. the enclosed customer response form indicating that they have received the notification package, that they installed the bushings kits in their affected Maxi Moves, or if they would like an Arjo Service Technician to perform the installation. Those locations needing a T-bar replacement will be contacted by an Arjo representative within 2 months to arrange a replacement of the T-bars.
Quantity in Commerce 4,400 patient lifts
Distribution Nationwide Distribution
 
-
-