|
Class 1 Device Recall Animas IR1250 Insulin Pump |
|
Date Initiated by Firm |
August 04, 2008 |
Date Posted |
November 25, 2008 |
Recall Status1 |
Terminated 3 on February 04, 2010 |
Recall Number |
Z-0294-2009 |
Recall Event ID |
49476 |
510(K)Number |
K042873
|
Product Classification |
Insulin infusion pump - Product Code LZG
|
Product |
Battery Caps used with Animas IR1250 Insulin Pump.
The pump is used to help maintain blood glucose targets. |
Code Information |
Part Numbers: 100-250-00, 100-251-00, 100-252-00, and 100-399-00. Serial Numbers: 06-42218-11, 07-42240-11, 07-42240-11, 07-51400-11, 08-42308-11, 08-42308-11, 08-43206-11, 10-43310-11, 10-43310-11, 11-52221-11, 12-51236-11, 13-43320-11, 13-44028-11, 14-42328-11, 14-50178-11, 14-51421-11, 15-45009-11, 15-55000-11, 16-42270-11, 16-53050-11, 16-55001-11, 17-42419-11, 17-43162-11, 17-44039-11, 17-44039-11, 18-41624-11, 18-55003-11, 18-55010-11, 19-50359-11, 19-55011-11, 20-54040-11, 20-55005-11, 20-55100-11, 21-44307-11, 21-54203-11, 21-55013-11, 21-55020-11, 21-55020-11, 22-45044-11, 22-55007-11, 22-55014-11, 23-54212-11, 23-55008-11, 23-55103-11, 23-55103-11, 23-56001-11, 24-36312-11, 24-55009-11, 24-55023-11, 24-55030-11, 24-55104-11, 24-55111-11, 25-55017-11, 25-55112-11, 25-55200-11, 25-56003-11, 25-56010-11, 26-41653-11, 26-43435-11, 26-55018-11, 26-55032-11, 26-55113-11, 26-55120-11, 26-56004-11, 26-56011-11, 27-42545-11, 27-42545-11, 27-51286-11, 27-54223-11, 27-55019-11, 27-55033-11, 27-55114-11, 27-55121-11, 27-55202-11, 27-56005-11, 27-56100-11, 28-42289-11, 28-45219-11, 28-52509-11, 28-55122-11, 28-55203-11, 28-55210-11, 28-56006-11, 28-56013-11, 28-56101-11, 29-55035-11, 29-55211-11, 29-56007-11, 29-56021-11, 29-56102-11, 30-45080-11, 30-55205-11, 30-56008-11, 30-56103-11, 31-52364-11, 31-54153-11, 31-55206-11, 31-55213-11, 31-55301-11, 31-56009-11, 31-56111-11, 32-44346-11, 32-52703-11, 32-55038-11, 32-55119-11, 32-55207-11, 32-56017-11, 32-56024-11, 32-56031-11, 33-44428-11, 33-44429-11, 33-51299-11, 33-51806-11, 33-53440-11, 33-55046-11, 33-55208-11, 33-55303-11, 33-56018-11, 33-56025-11, 33-56032-11, 33-56120-11, 34-45422-11, 34-55054-11, 34-55128-11, 34-56019-11, 34-56026-11, 34-56040-11, 34-56114-11, 34-56121-11, 34-56202-11, 35-42824-11, 35-51639-11, 35-52706-11, 35-55305-11, 35-55400-11, 35-56027-11, 35-56041-11, 35-56108-11, 35-56115-11, 35-56203-11, 35-56210-11, 36-42649-11, 36-43392-11, 36-44195-11, 36-54510-11, 36-55306-11, 36-55313-11, 36-55320-11, 36-55401-11, 36-56028-11, 36-56035-11, 36-56042-11, 36-56116-11, 36-56123-11, 36-56130-11, 36-56211-11, 37-55307-11, 37-55314-11, 37-55321-11, 37-55402-11, 37-56029-11, 37-56036-11, 37-56043-11, 37-56050-11, 37-56117-11, 37-56131-11, 37-56205-11, 37-56300-11, 38-44711-11, 38-45419-11, 38-47141-11, 38-52635-11, 38-54336-11, 38-55072-11, 38-55153-11, 38-55234-11, 38-55315-11, 38-55322-11, 38-55403-11, 38-56037-11, 38-56051-11, 38-56118-11, 38-56132-11, 38-56206-11, 38-56220-11, 39-42592-11, 39-53358-11, 39-53372-11, 39-55073-11, 39-55080-11, 39-55323-11, 39-55404-11, 39-55411-11, 39-56038-11, 39-56045-11, 39-56052-11, 39-56119-11, 39-56126-11, 39-56133-11, 39-56140-11, 39-56302-11, 40-44470-11, 40-52637-11, 40-52637-11, 40-55074-11, 40-55081-11, 40-55317-11, 40-55324-11, 40-55331-11, 40-55405-11, 40-56046-11, 40-56053-11, 40-56060-11, 40-56127-11, 40-56134-11, 40-56141-11, 40-56208-11, 40-56208-11, 40-56215-11, 40-56222-11, 40-56303-11, 40-56310-11, 41-43742-11, 41-54353-11, 41-54610-11, 41-55075-11, 41-55075-11, 41-55082-11, 41-55082-11, 41-55318-11, 41-55325-11, 41-55332-11, 41-55406-11, 41-55420-11, 41-56047-11, 41-56054-11, 41-56061-11, 41-56128-11, 41-56135-11, 41-56142-11, 41-56209-11, 41-56216-11, 41-56223-11, 41-56230-11, 41-56311-11, 42-44296-11, 42-52741-11, 42-54280-11, 42-55076-11, 42-55083-11, 42-55319-11, 42-55326-11, 42-55333-11, 42-55407-11, 42-55414-11, 42-55421-11, 42-55502-11, 42-56048-11, 42-56055-11, 42-56062-11, 42-56129-11, 42-56136-11, 42-56143-11, 42-56217-11, 42-56224-11, 42-56231-11, 42-56305-11, 43-55077-11, 43-55084-11, 43-55327-11, 43-55334-11, 43-55341-11, 43-55408-11, 43-55415-11, 43-55422-11, 43-55510-11, 43-56049-11, 43-56056-11, 43-56063-11, 43-56070-11, 43-56137-11, 43-56225-11, 43-56232-11, 43-56401-11, 44-44562-11, 44-53458-11, 44-55078-11, 44-55085-11, 44-55092-11, 44-55328-11, 44-55416-11, 44-55423-11, 44-55430-11, 44-55511-11, 44-56057-11, 44-56064-11, 44-56071-11, 44-56138-11, 44-56145-11, 44-56152-11, 44-56226-11, 44-56233-11, 44-56240-11, 44-56307-11, 44-56314-11, 45-44732-11, 45-47500-11, 45-49113-11, 45-54371-11, 45-54614-11, 45-55079-11, 45-55086-11, 45-55093-11, 45-55329-11, 45-55336-11, 45-55343-11, 45-55350-11, 45-55417-11, 45-55424-11, 45-55431-11, 45-55512-11, 45-55600-11, 45-56058-11, 45-56065-11, 45-56139-11, 45-56146-11, 45-56153-11, 45-56160-11, 45-56227-11, 45-56234-11, 45-56241-11, 45-56308-11, 45-56315-11, 45-56322-11, 45-56403-11, 45-56410-11, 46-45550-11, 46-52840-11, 46-55094-11, 46-55337-11, 46-55344-11, 46-55352-11, 46-55418-11, 46-55425-11, 46-55432-11, 46-55513-11, 46-55520-11, 55-55601-11, 46-56059-11, 46-56066-11, 46-56073-11, 46-56080-11, 46-56228-11, 46-56235-11, 46-56242-11, 46-56309-11, 46-56316-11, 46-56323-11, 46-56330-11, 46-56404-11, 46-56411-11, 47-37620-11, 47-44558-11, 47-44660-11, 47-52915-11, 47-54380-11, 47-55095-11, 47-55095-11, 47-55190-11, 47-55338-11, 47-55345-11, 47-55352-11, 47-55419-11, 47-55426-11, 47-55514-11, 47-55521-11, 47-55602-11, 47-56067-11, 47-56074-11, 47-56081-11, 47-56148-11, 47-56155-11, 47-56229-11, 47-56236-11, 47-56243-11, 47-56250-11, 47-56317-11, 47-56324-11, 47-56331-11, 47-56405-11, 47-56412-11, 47-56500-11, 48-53469-11, 48-54550-11, 48-55089-11, 48-55272-11, 48-55339-11, 48-55346-11, 48-55353-11, 48-55360-11, 48-55427-11, 48-55434-11, 48-55441-11, 48-55515-11, 48-55522-11, 48-55603-11, 48-56068-11, 48-56075-11, 48-56149-11, 48-56163-11, 48-56237-11, 48-56251-11, 48-56325-11, 48-56332-11, 48-56406-11, 48-56420-11, 48-56501-11, 49-38168-11, 49-44574-11, 49-47430-11, 49-52762-11, 49-55097-11, 49-55347-11, 49-55354-11, 40-55361-11, 49-55428-11, 49-55442-11, 49-55509-11, 49-55516-11, 49-55530-11, 49-55604-11, 49-55611-11, 49-56069-11, 49-56076-11, 40-56090-11, 49-56164-11, 49-56171-11, 49-56238-11, 49-56245-11, 49-56252-11, 49-56319-11, 49-56326-11, 49-56333-11, 49-56407-11, 49-56414-11, 49-56421-11, 49-56502-11, 50-52925-11, 50-54619-11, 56-54626-11, 50-54721-11, 50-55098-11, 50-55281-11, 50-55348-11, 50-55355-11, 50-55362-11, 50-55429-11, 50-55436-11, 50-55450-11, 50-55517-11, 50-55524-11, 50-55531-11, 50-55612-11, 50-55700-11, 50-56084-11, 50-56091-11, 50-56165-11, 50-56172-11, 50-56239-11, 50-56346-11, 50-56253-11, 50-56327-11, 50-56324-11, 50-56341-11, 50-56408-11, 50-56415-11, 50-56422-11, 50-56503-11, 50-56510-11, 51-52764-11, 51-52771-11, 51-54722-11, 51-55099-11, 51-55275-11, 51-55282-11, 51-55349-11, 51-55356-11, 51-55363-11, 51-55370-11, 51-55437-11, 51-55444-11, 51-55451-11, 51-55518-11, 51-55606-11, 51-55613-11, 51-55620-11, 51-55701-11, 51-56085-11, 51-56092-11, 51-56159-11, 51-56247-11, 51-56254-11, 51-56261-11, 51-56328-11, 51-56335-11, 51-56342-11, 51-56416-11, 51-56430-11, 51-56504-11, 51-56511-11, 52-37618-11, 52-37618-11, 52-44739-11, 52-52846-11, 52-53751-11, 52-54628-11, 52-55276-11, 52-55283-11, 52-55290-11, 52-55371-11, 52-55438-11, 52-55445-11, 52-55452-11, 52-55519-11, 52-55526-11, 52-55533-11, 52-55540-11, 52-55621-11, 52-55702-11, 52-56079-11, 52-56093-11, 52-56167-11, 52-56174-11, 52-56181-11, 52-56248-11, 52-56255-11, 52-56262-11, 52-56329-11, 52-56336-11, 52-56343-11, 52-56417-11, 52-56431-11, 52-56505-11, 52-56512-11, 52-56600-11, 53-43863-11, 53-43863-11, 53-53914-11, 53-54812-11, 53-55277-11, 53-55284-11, 53-55291-11, 53-55358-11, 53-55365-11, 53-55372-11, 53-55439-11, 53-55446-11, 53-55453-11, 53-55460-11, 53-55527-11, 53-55534-11, 53-55541-11, 53-55608-11, 53-55615-11, 53-55622-11, 53-55703-11, 53-55710-11, 53-56094-11, 53-56168-11, 53-56182-11, 53-56249-11, 53-56253-11, 53-56270-11, 53-56337-11, 53-56344-11, 53-56351-11, 53-56425-11, 53-56432-11, 53-56506-11, 53-56513-11, 53-56601-11, 54-52774-11, 54-54644-11, 54-55285-11, 54-55359-11, 54-55366-11, 54-55373-11, 54-55380-11, 54-55447-11, 54-55454-11, 54-55461-11, 54-55528-11, 54-55535-11, 54-55542-11, 54-55616-11, 54-55623-11, 54-55630-11, 54-55704-11, 54-55711-11, 54-56095-11, 54-56169-11, 54-56176-11, 54-56183-11, 54-56190-11, 54-56257-11, 54-56264-11, 54-56271-11, 54-56338-11, 54-56345-11, 54-56352-11, 54-56419-11, 54-56426-11, 54-56433-11, 54-56440-11, 54-56507-11, 54-56514-11, 54-56602-11, 55-43791-11, 55-49172-11, 55-49151-11, 55-54395-11, 55-54733-11, 55-55286-11, 55-55293-11, 55-55367-11, 55-55374-11, 55-55382-11, 55-55448-11, 55-55455-11, 55-55462-11, 55-55529-11, 55-55536-11, 55-55543-11, 55-55550-11, 55-55617-11, 55-55624-11, 55-55631-11, 55-55705-11, 55-55712-11, 55-55712-11, 55-55800-11, 55-56177-11, 55-56184-11, 55-56258-11, 55-56265-11, 55-56272-11, 55-56339-11, 55-56346-11, 55-56353-11, 55-56360-11, 55-56427-11, 55-56441-11, 55-56508-11, 55-56515-11, 55-56603-11, 55-56610-11, 56-48254-11, 56-49057-11, 56-53843-11, 56-55287-11, 56-55294-11, 56-55368-11, 56-55382-11, 56-55449-11, 56-55456-11, 56-55463-11, 56-55470-11, 56-55537-11, 56-55544-11, 56-55551-11, 56-55618-11, 56-55625-11, 56-55713-11, 56-55720-11, 56-55801-11, 56-56097-11, 56-56178-11, 56-56185-11, 56-56259-11, 56-56266-11, 56-56273-11, 56-56280-11, 56-56347-11, 56-56354-11, 56-56435-11, 56-56442-11, 56-56516-11, 56-56604-11, 56-56611-11, 56-57252-11, 57-45582-11, 57-53953-11, 57-54397-11, 57-54830-11, 57-55288-11, 57-55295-11, 57-55369-11, 57-55376-11, 57-55383-11, 57-55390-11, 57-55457-11, 57-55464-11, 57-55538-11, 57-55545-11, 57-55552-11, 57-55619-11, 57-55626-11, 57-55640-11, 57-55707-11, 57-55714-11, 57-55721-11, 57-55802-11, 57-56098-11, 57-56179-11, 57-56186-11, 57-56193-11, 57-56257-11, 57-56281-11, 57-56348-11, 57-56355-11, 57-56362-11, 57-56429-11, 57-56517-11, 57-56531-11, 57-56605-11, 57-56612-11, 57-56700-11, 58-43794-11, 58-55289-11, 58-55377-11, 58-55384-11, 58-55391-11, 58-55465-11, 58-55472-11, 58-55539-11, 58-55546-11, 58-55553-11, 58-55560-11, 58-55627-11, 58-55641-11, 58-55708-11, 58-55715-11, 58-55722-11, 58-55803-11, 58-55810-11, 58-56099-11, 58-56187-11, 58-56194-11, 58-56268-11, 58-56275-11, 58-56349-11, 58-56356-11, 58-56363-11, 58-56370-11, 58-56444-11, 58-56518-11, 58-56532-11, 58-56606-11, 58-56613-11, 58-56620-11, 58-56701-11, 59-47711-11, 59-48183-11, 59-55297-11, 59-55378-11, 59-55385-11, 59-55392-11, 59-55459-11, 59-55466-11, 59-55473-11, 59-55480-11, 59-55547-11, 59-55554-11, 59-55561-11, 59-55628-11, 59-55635-11, 59-55642-11, 59-55709-11, 59-55716-11, 59-55723-11, 59-55730-11, 59-55804-11, 59-55811-11, 59-56188-11, 59-56195-11, 59-56269-11, 59-56276-11, 59-56283-11, 59-56290-11, 59-56357-11, 59-56364-11, 59-56371-11, 59-56452-11, 59-56533-11, 59-56540-11, 59-56607-11, 59-56614-11, 59-56621-11, 59-56702-11, 60-48191-11, 60-54573-11, 60-54583-11, 60-54664-11, 60-55379-11, 60-55386-11, 60-55467-11, 60-55474-11, 60-55481-11, 60-55555-11, 60-55643-11, 60-55650-11, 60-55717-11, 60-55724-11, 60-55731-11, 60-55805-11, 60-55812-11, 60-55900-11, 60-56189-11, 60-56196-11, 60-56284-11, 60-56291-11, 60-56358-11, 60-56372-11, 60-56439-11, 60-56446-11, 60-56453-11, 60-56460-11, 60-56534-11, 60-56541-11, 60-56608-11, 60-56622-11, 60-56703-11, 60-56710-11, 60-57182-11, 61-42994-11, 61-54584-11, 61-55299-11, 61-55387-11, 61-55394-11, 61-55468-11, 61-55475-11, 61-55482-11, 61-55549-11, 61-55556-11, 61-55563-11, 61-55570-11, 61-55637-11, 61-55651-11, 61-55718-11, 61-55725-11, 61-55732-11, 61-55806-11, 61-55813-11, 61-55901-11, 61-56197-11, 61-56278-11, 61-56292-11, 61-56359-11, 61-56366-11, 61-56373-11, 61-56380-11, 61-56447-11, 61-56454-11, 61-56461-11, 61-56528-11, 61-56535-11, 61-56542-11, 61-56609-11, 61-56630-11, 61-56704-11, 61-56711-11, 61-57183-11, 62-54828-11, 62-55388-11, 62-55395-11, 62-55469-11, 62-55476-11, 62-55483-11, 62-55490-11, 62-55557-11, 62-55571-11, 62-55638-11, 62-55645-11, 62-55652-11, 62-55726-11, 62-55733-11, 62-55740-11, 62-55807-11, 62-55821-11, 62-55902-11, 62-56198-11, 62-56279-11, 62-56286-11, 62-56293-11, 62-56367-11, 62-56374-11, 62-56381-11, 62-56448-11, 62-56455-11, 62-56536-11, 62-56543-11, 62-56550-11, 62-56617-11, 62-56624-11, 62-56631-11, 62-56705-11, 62-56712-11, 62-56800-11, 63-48282-11, 63-55389-11, 63-55396-11, 63-55477-11, 63-55484-11, 63-55491-11, 63-55565-11, 63-55572-11, 63-55639-11, 63-55653-11, 63-55660-11, 63-55727-11, 63-55734-11, 63-55741-11, 63-55808-11, 63-55815-11, 63-55822-11, 63-55903-11, 63-55910-11, 63-56287-11, 63-56294-11, 63-56368-11, 63-56382-11, 63-56449-11, 63-56456-11, 63-56537-11, 63-56544-11, 63-56551-11, 63-56618-11, 63-56625-11, 63-56632-11, 63-56713-11, 63-56720-11, 63-56801-11, 63-57185-11, 64-45927-11, 64-46751-11, 64-48607-11, 64-54763-11, 64-55397-11, 64-55478-11, 64-55485-11, 64-55492-11, 64-55566-11, 64-55573-11, 64-55580-11, 64-55647-11, 64-55654-11, 64-55661-11, 64-55728-11, 64-55735-11, 64-55742-11, 64-55809-11, 64-55823-11, 64-55904-11, 64-55911-11, 64-56288-11, 64-56295-11, 64-56376-11, 64-56383-11, 64-56390-11, 64-56457-11, 64-56538-11, 64-56545-11, 64-56552-11, 64-56619-11, 64-56626-11, 64-56633-11, 64-56640-11, 64-56707-11, 64-56714-11, 65-52975-11, 65-54676-11, 65-54683-11, 65-55398-11, 65-55479-11, 65-55486-11, 65-55493-11, 65-55567-11, 65-55574-11, 65-55581-11, 65-55648-11, 65-55662-11, 65-55729-11, 65-55736-11, 65-55743-11, 65-55750-11, 65-55817-11, 65-55824-11, 65-55831-11, 65-55905-11, 65-56289-11, 65-56377-11, 65-56391-11, 65-56465-11, 65-56472-11, 65-56539-11, 65-56546-11, 65-56553-11, 65-56560-11, 65-56634-11, 65-56641-11, 65-56708-11, 65-56715-11, 65-56722-11, 66-55399-11, 66-55487-11, 66-55494-11, 66-55568-11, 66-55649-11, 66-55656-11, 66-55663-11, 66-55670-11, 66-55737-11, 66-55744-11, 66-55751-11, 66-55818-11, 66-55825-11, 66-55832-11, 66-55906-11, 66-55913-11, 66-55920-11, 66-56297-11, 66-56378-11, 66-56385-11, 66-56392-11, 66-56459-11, 66-56466-11, 66-56473-11, 66-56480-11, 66-56547-11, 66-56554-11, 66-56561-11, 66-56628-11, 66-56635-11, 66-56642-11, 66-56709-11, 66-56719-11, 66-56723-11, 66-56730-11, 67-55495-11, 67-55569-11, 67-55583-11, 67-55657-11, 67-55664-11, 67-55671-11, 67-55738-11, 67-55745-11, 67-55752-11, 67-55819-11, 67-55826-11, 67-55833-11, 67-55840-11, 67-55907-11, 67-55914-11, 67-55921-11, 67-56298-11, 67-56379-11, 67-56386-11, 67-56467-11, 67-56474-11, 67-56481-11, 67-56548-11, 67-56555-11, 67-56562-11, 67-56629-11, 67-56636-11, 67-56643-11, 67-56650-11, 67-56717-11, 67-56724-11, 67-56731-11, 67-56805-11, 67-56900-11, 68-43987-11, 68-48713-11, 68-52897-11, 68-55489-11, 68-55496-11, 68-55577-11, 68-55584-11, 68-55658-11, 68-55665-11, 68-55672-11, 68-55739-11, 68-55746-11, 68-55753-11, 68-55760-11, 68-55827-11, 68-55834-11, 68-55841-11, 68-55908-11, 68-55915-11, 68-55922-11, 68-56299-11, 68-56387-11, 68-56394-11, 68-56475-11, 68-56482-11, 68-56549-11, 68-56556-11, 68-56563-11, 68-56570-11, 68-56644-11, 68-56651-11, 68-56718-11, 68-56725-11, 68-56732-11, 68-56813-11, 68-56820-11, 68-56901-11, 69-55497-11, 69-55585-11, 69-55592-11, 69-55659-11, 69-55666-11, 69-55673-11, 69-55680-11, 69-55747-11, 69-55754-11, 69-55761-11, 69-55828-11, 69-55835-11, 69-55842-11, 69-55909-11, 69-55923-11, 69-56388-11, 69-56395-11, 69-56476-11, 69-56483-11, 69-56557-11, 69-56564-11, 69-56571-11, 69-56638-11, 69-56652-11, 69-56719-11, 69-56726-11, 69-56733-11, 69-56740-11, 69-56814-11, 69-56902-11, 70-47662-11, 70-54688-11, 70-55579-11, 70-55586-11, 70-55667-11, 70-55674-11, 70-55681-11, 70-55748-11, 70-55755-11, 70-55829-11, 70-55836-11, 70-55843-11, 70-55850-11, 70-55924-11, 70-55931-11, 70-56389-11, 70-56396-11, 70-56477-11, 70-56491-11, 70-56558-11, 70-56565-11, 70-56572-11, 70-56639-11, 70-56646-11, 70-56653-11, 70-56660-11, 70-56727-11, 70-56734-11, 70-56741-11, 70-56808-11, 70-56822-11, 70-56903-11, 71-38739-11, 71-47825-11, 71-54689-11, 71-54872-11, 71-55499-11, 71-55587-11, 71-55594-11, 71-55668-11, 71-55675-11, 71-55682-11, 71-55749-11, 71-55756-11, 71-55763-11, 71-55837-11, 71-55844-11, 71-55851-11, 71-55918-11, 71-55925-11, 71-55932-11, 71-56397-11, 71-56478-11, 71-56485-11, 71-56559-11, 71-56566-11, 71-56573-11, 71-56580-11, 71-56647-11, 71-56654-11, 71-56661-11, 71-56728-11, 71-56735-11, 71-56742-11, 71-56823-11, 71-56830-11, 71-56904-11, 72-46928-11, 72-55588-11, 72-55669-11, 72-55676-11, 72-55683-11, 72-55690-11, 72-55757-11, 72-55764-11, 72-55771-11, 72-55838-11, 72-55845-11, 72-55852-11, 72-55919-11, 72-55926-11, 72-55933-11, 72-55940-11, 72-56398-11, 72-56479-11, 72-56486-11, 72-56493-11, 72-56567-11, 72-56574-11, 72-56581-11, 72-56648-11, 72-56662-11, 72-56729-11, 72-56736-11, 72-56743-11, 72-56750-11, 72-56824-11, 72-56831-11, 73-43999-11, 73-48468-11, 73-53895-11, 73-54874-11, 73-55596-11, 73-55677-11, 73-55684-11, 73-55691-11, 73-55758-11, 73-55765-11, 73-55772-11, 73-55839-11, 73-55853-11, 73-55860-11, 73-55927-11, 73-55941-11, 73-56399-11, 73-56487-11, 73-56494-11, 73-56568-11, 73-56575-11, 73-56582-11, 73-56656-11, 73-56663-11, 73-56737-11, 73-56744-11, 73-56751-11, 73-56825-11, 73-56832-11, 73-56913-11, 73-56920-11, 74-47747-11, 74-47916-11, 74-48645-11, 74-52998-11, 74-55597-11, 74-55678-11, 74-55685-11, 74-55692-11, 74-55759-11, 74-55766-11, 74-55773-11, 74-55847-11, 74-55854-11, 74-55861-11, 74-55928-11, 74-55935-11, 74-55942-11, 74-56495-11, 74-56569-11, 74-56576-11, 74-56583-11, 74-56590-11, 74-56657-11, 74-56671-11, 74-56738-11, 74-56745-11, 74-56752-11, 74-56833-11, 74-56840-11, 74-56907-11, 74-56914-11, 74-56921-11, 75-55598-11, 75-55679-11, 75-55686-11, 75-55693-11, 75-55767-11, 75-55774-11, 75-55781-11, 75-55848-11, 75-55855-11, 75-55862-11, 75-55929-11, 75-55936-11, 75-56496-11, 75-56577-11, 75-56584-11, 75-56591-11, 75-56739-11, 75-56746-11, 75-56753-11, 75-56760-11, 75-56827-11, 75-56834-11, 75-56922-11, 76-54972-11, 76-55599-11, 76-55694-11, 76-55768-11, 76-55775-11, 76-55782-11, 76-55849-11, 76-55856-11, 76-55863-11, 76-55870-11, 76-55937-11, 76-55944-11, 76-55951-11, 76-56578-11, 76-56585-11, 76-56592-11, 76-56659-11, 76-56673-11, 76-56680-11, 76-56747-11, 76-56754-11, 76-56761-11, 76-56828-11, 76-56835-11, 76-56842-11, 76-56909-11, 76-56916-11, 76-56930-11, 77-46873-11, 77-46954-11, 77-49296-11, 77-49641-11, 77-54973-11, 77-54980-11, 77-55688-11, 77-55695-11, 77-55769-11, 77-55776-11, 77-55783-11, 77-55790-11, 77-55857-11, 77-55864-11, 77-55871-11, 77-55938-11, 77-55952-11, 77-56498-11, 77-56579-11, 77-56586-11, 77-56593-11, 77-56674-11, 77-56681-11, 77-56748-11, 77-56755-11, 77-56762-11, 77-56836-11, 77-56843-11, 77-56850-11, 77-56917-11, 77-56931-11, 78-54879-11, 78-55689-11, 78-55696-11, 78-55777-11, 78-55784-11, 78-55791-11, 78-55858-11, 78-55865-11, 78-55872-11, 78-55939-11, 78-55946-11, 78-55953-11, 78-55960-11, 78-56587-11, 78-56594-11, 78-56675-11, 78-56749-11, 78-56756-11, 78-56763-11, 78-56770-11, 78-56844-11, 78-56851-11, 78-56918-11, 78-56925-11, 78-56932-11, 79-55778-11, 79-55785-11, 79-55792-11, 79-55859-11, 79-55880-11, 79-55947-11, 79-55954-11, 79-55961-11, 79-56588-11, 79-56595-11, 79-56669-11, 79-56676-11, 79-56683-11, 79-56690-11, 79-56757-11, 79-56764-11, 79-56771-11, 79-56838-11, 79-56852-11, 79-56919-11, 79-56933-11, 79-56490-11, 80-48665-11, 80-54976-11, 80-55698-11, 80-55779-11, 80-55786-11, 80-55793-11, 80-55867-11, 80-55874-11, 80-55881-11, 80-55948-11, 80-55955-11, 80-55962-11, 80-56589-11, 80-56596-11, 80-56677-11, 80-56684-11, 80-56691-11, 80-56758-11, 80-56765-11, 80-56772-11, 80-56839-11, 80-56846-11, 80-56860-11, 80-56941-11, 81-54977-11, 81-54984-11, 81-54991-11, 81-55699-11, 81-55794-11, 81-55868-11, 81-55875-11, 81-55882-11, 81-55949-11, 81-55956-11, 81-55963-11, 81-55970-11, 81-56597-11, 81-56678-11, 81-56685-11, 81-56692-11, 81-56759-11, 81-56766-11, 81-56773-11, 81-56847-11, 81-56854-11, 81-56861-11, 81-56928-11, 81-56935-11, 82-54992-11, 82-55788-11, 82-55795-11, 82-55869-11, 82-55890-11, 82-55957-11, 82-55964-11, 82-55971-11, 82-56598-11, 82-56679-11, 82-56686-11, 82-56693-11, 82-56767-11, 82-56774-11, 82-56848-11, 82-56855-11, 82-56929-11, 82-56936-11, 82-56950-11, 83-54986-11, 83-54993-11, 83-55789-11, 83-55796-11, 83-55877-11, 83-55884-11, 83-55958-11, 83-55965-11, 83-55972-11, 83-56599-11, 83-56687-11, 83-56694-11, 83-56768-11, 83-56775-11, 83-56849-11, 83-56856-11, 83-56937-11, 84-38956-11, 84-54994-11, 84-55878-11, 84-55885-11, 84-55892-11, 84-55966-11, 84-55980-11, 84-56688-11, 84-56695-11, 84-56769-11, 84-56776-11, 84-56857-11, 84-56864-11, 84-56871-11, 84-56938-11, 84-56952-11, 85-55798-11, 85-55886-11, 85-55893-11, 85-55967-11, 85-55981-11, 85-56689-11, 85-56696-11, 85-56777-11, 85-56858-11, 85-56865-11, 85-56872-11, 85-56939-11, 85-56953-11, 85-56960-11, 86-55799-11, 86-55887-11, 86-55894-11, 86-55968-11, 86-55975-11, 86-56697-11, 86-56778-11, 86-56866-11, 86-56873-11, 86-56954-11, 86-56961-11, 87-54997-11, 87-55888-11, 87-55895-11, 87-55969-11, 87-55990-11, 87-56698-11, 87-56874-11, 87-56955-11, 87-56962-11, 88-47884-11, 88-48782-11, 88-48944-11, 88-55889-11, 88-55896-11, 88-55991-11, 88-56868-11, 88-56875-11, 88-56882-11, 88-56956-11, 88-56963-11, 88-56970-11, 89-49674-11, 89-55897-11, 89-55985-11, 89-55992-11, 89-56869-11, 89-56876-11, 89-56957-11, 89-56964-11, 89-56971-11, 90-55898-11, 90-55986-11, 90-55993-11, 90-56877-11, 90-56958-11, 90-56965-11, 90-56972-11, 91-55987-11, 91-55994-11, 91-56885-11, 91-56959-11, 91-56966-11, 91-56980-11, 92-56798-11, 92-56893-11, 92-56967-11, 92-56974-11, 93-55989-11, 93-56799-11, 93-56968-11, 94-55997-11, 94-56895-11, 94-56969-11, 94-56976-11, 95-56896-11, 95-56977-11, 96-55999-11, 96-56978-11, 97-56898-11, and 98-56899-11. |
Recalling Firm/ Manufacturer |
Animas Corporation 200 Lawrence Dr West Chester PA 19380
|
For Additional Information Contact |
Consumer Relations 866-423-4087
|
Manufacturer Reason for Recall |
Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device to administer insulin therapy which may result in hyperglycemia.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
The recalling firm issued a notification letter to the patients with insulin pumps on 8/5/08 to inform them of the problem and that they needed to replace the battery. The recalling firm contacted OneTouch Ping users by telephone on 8/5/08 for this issue and another issue. The recalling firm issued letters to distributors informing them of the problem and notified customers to replace the batteries. In addition, a letter was issued to physicians to alert them of the situation with regard to current patients. Consumers can contact 1-800-423-4087 for additional information or questions. |
Quantity in Commerce |
1,624 units |
Distribution |
Worldwide Distribution including USA (AR, CA, FL, KY, LA, MA, NE, OH, PA, SC, SD, TN, and UT) and countries of Australia, Austria, Canada, Czech Republic, Finland, France, Israel, Italy, New Zealand, Spain, Switzerland, and Sweden. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LZG and Original Applicant = ANIMAS CORP.
|
|
|
|