 Class 2 Recall
GE Precision XR/a
see related information
Date Posted September 30, 2008
Recall Status1 Open
Recall Number Z-1713-2008
Recall Event ID 49533
Premarket Notification
510(K) Numbers
K001577  K014117  K032298  K040771  K043534 
Product Classification System, X-Ray, Stationary - Product Code KPR
Product GE Healthcare Proteus XR/a Eclipse Collimator model # 2379827.
Code Information 42948HL0, 42949HL8, 43054HL6, 43055HL3, 43056HL1, 43057HL9, 43132HL0, 43133HL8, 43134HL6, 43135HL3, 43201HL3, 43202HL1, 43203HL9, 44605HL4, 44606HL2, 44607HL0, 44608HL8, 44609HL6, 44610HL4, 44681HL5, 45046HL0, 45047HL8, 45048HL6, 45128HL6, 45129HL4, 45130HL2, 45131HL0, 48119HL2, 48120HL0, 48121HL8, 48122HL6, 48123HL4, 48124HL2, 48125HL9, 48126HL7, 48127HL5, 48128HL3, 48164HL8, 48181HL2, 48274HL5, 48276HL0, 48277HL8, 48278HL6, 48680HL3, 48682HL9, 48683HL7, 48684HL5, 48685HL2, 48686HL0, 48687HL8, 48688HL6, 48689HL4, 48876HL7, 48877HL5, 48878HL3, 48995HL3, 48997HL1, 48998HL9, 49000HL3, 49208HL2, 49209HL0, 49210HL8, 49211HL6, 49391HL6, 49392HL4, 49393HL2, 49395HL7, 49494HL8, 49496HL3, 49497HL1, 49498HL9, 49499HL7, 49500HL2, 49501HL0, 49502HL8, 49503HL6, 49681HL0, 49682HL8, 49683HL6, 49684HL4, 49685HL1, 49686HL9, 49687HL7, 49688HL5, 49850HL1, 49851HL9, 49852HL7, 49853HL5, 49855HL0, 49856HL8, 49857HL6, 50027HL2, 50028HL0, 50029HL8, 50030HL6, 50031HL4, 50177HL5, 50178HL3, 50179HL1, 50180HL9, 50321HL9, 50322HL7, 50323HL5, 50324HL3, 50444HL9, 50445HL6, 50446HL4, 50447HL2, 50448HL0, 50449HL8, 50450HL6, 50451HL4, 50452HL2, 50453HL0, 50790HL5, 50791HL3, 50792HL1, 50793HL9, 50794HL9, 50795HL4, 50797HL0, 50798HL8, 50801HL0, 51143HL6, 51144HL4, 51145HL1, 51147HL7, 51486HL9, 51487HL7, 51488HL5, 52014HL8, and 70RD05001
Recalling Firm/
Manufacturer
GE Healthcare
3000 N Grandview Blvd
Waukesha, Wisconsin 53188-1615
Manufacturer Reason
for Recall
The actual average illuminance for the collimators are approximately 140-lux. This does not meet the 160-lux requirement. GE will be correcting the units by replacing all collimators.
FDA Determined
Cause 2
DESIGN: Device Design
Action GE will be correcting the units by replacing all collimators that don't meet the 160-lux requirement. The field correction will be initiated and is scheduled to be completed by October 2008.
Quantity in Commerce 132
Distribution Worldwide Distribution --- USA including states of CA, DE, FL, GA, ID, IL, KS, KY, MD, MI, MO, MT, NC, NJ, NM, NY, OH, PA, SC, TX, UT, VA, and WV, and countries of Australia, Canada, China, France, Germany, Greece, Italy, Korea, Poland, Saudi Arabia, United Kingdom, and Venezuela.