Date Initiated by Firm |
November 12, 2008 |
Date Posted |
April 10, 2009 |
Recall Status1 |
Terminated 3 on September 22, 2009 |
Recall Number |
Z-1068-2009 |
Recall Event ID |
50171 |
510(K)Number |
K953239
|
Product Classification |
Chemistry Analyzer for Clinical Use - Product Code JJE
|
Product |
Roche/Hitachi Modular Analytics System, clinical chemistry analyzer, Modular P Module, Roche Diagnostics Corp., Indpls., IN; GMMI No. 03284549001.
Is a fully automatic, computerized chemistry analyzer intended for in vitro quantitative and qualitative determinations of a wide range analytes. The instrument is capable of performing potentiometric and photometric assays. |
Code Information |
All units with software version 08-03. |
Recalling Firm/ Manufacturer |
Roche Diagnostics Corp. 9115 Hague Rd Indianapolis IN 46256-1025
|
For Additional Information Contact |
800-428-2336
|
Manufacturer Reason for Recall |
If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.
|
FDA Determined Cause 2 |
Software change control |
Action |
Consignees were notified via an Urgent Medical Device Correction letter dated 11/12/08 and instructed to only execute "module unmasking" when the unit is in "standby" mode and to consult with their physicians at the facility to determine the clinical implications and whether any clinical intervention is needed for patients treated based upon results obtained with this software. Questions should be directed to Roche Diagnostics Technical Support at 1-800-428-2336. |
Quantity in Commerce |
22 systems in all |
Distribution |
Nationwide Distribution -- including states of California, Florida, Georgia, Hawaii, Indiana, Maryland, New York, North Carolina, Pennsylvania, and Virginia. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JJE and Original Applicant = BOEHRINGER MANNHEIM CORP.
|