• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Accolade C CS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Accolade C CS
see related information
Date Posted January 26, 2009
Recall Status1 Terminated on January 20, 2010
Recall Number Z-0895-2009
Recall Event ID 50485
Premarket Notification
510(K) Numbers
K020989  K022555  K031744  K041170  K050741  K053514 
K070095  K070963  K932519  K941366  K955731  K982032 
K983226 
Product Classification Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented - Product Code LZO
Product Accolade C CS These devices are modular components of a total hip system. These femoral stems are intended for use with femoral heads and acetabular components in primary or revision total hip arthroplasty.
Code Information 6057-0335D ACCOLADE C CS 127 NK JYLMME, 6057-0537D ACCOLADE C CS 127 NK JYNMME, 6057-0740D ACCOLADE C CS 127 NK E5EMME and 6058-0335D ACCOLADE C CS 132 NK JYVMME
Recalling Firm/
Manufacturer
Stryker Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah, New Jersey 07430
For Additional Information Contact Karen Ariemma, RAC
201-831-5000
Manufacturer Reason
for Recall
Stryker Orthopaedics became aware of a visual defect in certain blister packages of certain devices.
FDA Determined
Cause 2
PRODUCTION CONTROLS: Packaging Process Control
Action Recall notification letters were sent via Federal Express on November 21, 2008 to Stryker Branches/Agencies, Hospital Risk Management, Chief of Orthopaedics and Surgeon. The letters explained the problem and indicated that there are no risk factors. Contact Stryker Orthopaedics for additional information at 1-201-831-5718.
Quantity in Commerce 322 total units - all types
Distribution Nationwide. OUS: Australia, Canada, Japan, Argentina, Brazil, Germany, Spain, Columbia, Guatemala and Sweden.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = LZO and Original Applicant = HOWMEDICA OSTEONICS CORP
510(K)s with Product Code = LZO and Original Applicant = HOWMEDICA OSTEONICS CORP.
510(K)s with Product Code = LZO and Original Applicant = IND DIAGNOSTIC INC.
510(K)s with Product Code = LZO and Original Applicant = OSTEONICS CORP.
510(K)s with Product Code = LZO and Original Applicant = STRYKER ORTHOPAEDICS
-
-