• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Newport e360 Ventilator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Newport e360 Ventilator
see related information
Date Posted September 04, 2009
Recall Status1 Open
Recall Number Z-2029-2009
Recall Event ID 50785
Premarket Notification
510(K) Number
K053502 
Product Classification Ventilator, Continuous, Facility Use - Product Code CBK
Product Newport e360 Ventilator, Rx only. The e360 Ventilator System is intended to provide continuous (endotracheal or tracheostomy [ET] tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric, and adult patients requiring tidal volumes equal to or greater than 20 milliliters (mL).
Code Information Serial Numbers for the Domestic Units: N07361211863 N06360910375 N06360610268 N09360413995 N07360110859 N07361111725 N07361111620 N06360710310 N06360710311 N06360710312 N07360110797 N08360812873 N06361210693 N07360911465 N36360410225 N07360110864 N07360110870 N09360313832 N07361111713 N08360512383 N08360512384 N08360512385 N08360512386 N07361211793 N06360910351 N08360712658 N08360712659 N08360712660 N08360712661 N08360712662 N08361013165 N08361013166 N08361013167 N08361013168 N08361013169 N06361210693 N07360110855 N08361213295 N09360113451 N06360610258 N07360110853 N06360610271 N06361010529 N06361010532 N07360110825 N07360811393 N08360212053 N07361211869 N07361211870 N07361211871 N07361211872 N09360213550 N06360810339 N06360910350 N06360910352 N06360910354 N06360910388 N06360910389 N07360411127 N07360811396 N07361111626 N07361111627 N07361111711 N07361111712 N08360512396 N08360612528 N08360612561 N08360712687 N08360712688 N08360712689 N08360712690 N08360712691 N08360712692 N08360712693 N08360712694 N08360712695 N08360712696 N08360712697 N08360712698 N08360712699 N08360712700 N08360712701 N08360712702 N08360712703 N08360712704 N08360712705 N08360712706 N08360712707 N08360712708 N08360712709 N08360712710 NO8360712709 N06360710307 N06360710308 N06360710309 N06360710313 N06360710314 N06360610280 N06361010516 N06361010518 N06361010519 N06361010522 N06361010536 N07360110856 N07360110862 N07360110867 N07360210926 N07360911464 N07360911489 N07360911490 N07360210952 N07360210953 N07360210954 N08361013107 N08361013108 N08361013109 N08361213283 N08361213284 N08360512381 N08360512382 N08360512398 N08360512399 N08360512400 N08360512402 N06361010542 N06361010543 N06361010544 N06361010545 N06361010546 N06361010547 N06361010548 N06361010550 N06361110551 N06361110552 N06361110553 N06361110554 N06361110555 N06361110556 N06361110557 N06361110558 N06361110559 N06361110560 N06361110561 N06361110562 N06361110563 N06361110564 N06361110572 N06360910373 N07306011235 N07360611236 N07360711360 N07360711361 N07360711363 N07360711364 N07360711368 N07360711372 N07360911491 N07360911492 N07361211769 N07361211783 N07361211784 N07361211785 N07361211786 N07361211787 N07361211788 N07361211789 N07361211790 N07361211791 N07361211792 N08360111874 N08360111875 N08360512395 N08361013110 N08361013111 N08361013112 N08361013113 N08361013114 N07360110873 N06361010536-R N07361211812 N07361211813 N08361013175 N08360111987 N08360111988 N08360111989 N08360111990 N06361010533 N07360110796 N07360110852 N07360110860 N07360110868 N07360110876 N06361010531 N07360210951 N07360210964 N07360110866 N07360811397 N07360110898 N07360210944 N07360811380 N07360811384 N07360811386 N07360811398 N07360811401 N07360811402 N07360811403 N07360811427 N08360811403 N07360711362 N07360711365 N07360811387 N09360213549 N06361010534 N07360110863; and Serial Numbers for the International Units: N06360910392 N06360910400 N06360910426 N06361110580 N07361111637 N07361111638 N09360313710 N06360910352 N07360110878 N07360411101 N07360611242 N07360611243 N07360611244 N07361211794 N07361211795 N07361211796 N07361211797 N07361211798 N08360111890 N08360111891 N08360111893 N08360111894 N08360612578 N08360612579 N08360612580 N08360612581 N08360612582 N08360612583 N08360612593 N08360712711 N08360712712 N08360712713 N09360313786 N09360313787 N07360311089 N07360411096 N07360411105 N07360411111 N07360411129 N07360511161 N07361011506 N07361011507 N07361011523 N07361011524 N07361011525 N07361011526 N07361011527 N07361011528 N07361011529 N07361011530 N07361011531 N07361011532 N07361011533 N07361011534 N07361011535 N07361011537 N07361011538 N07361011539 N07361011540 N07361011541 N07361011542 N07361011543 N07361011544 N07360210959 N08360812800 N08360812801 N08360812804 N08360812809 N08360812810 N08360812811 N08360812812 N08360812813 N08360812816 N08360812817 N08360812818 N08360812819 N08360812820 N08360812823 N08360812824 N09360213600 N09360313696 N09360313727 N09360313728 N09360313729 N09360313731 N09360313847 N09360413942 N09360413943 N09360413944 N09360413945 N09360413946 N09360414002 N09360414003 N06360610265 N07360210963 N07360210966 N07360210967 N07360210968 N07360210970 N09360113452 N07360110885 N07360110891 N07360110909 N07360210972 N07360311090 N07360311091 N07360411095 N07360411107 N07360411108 N08360312183 E360-0408-183 E360-0408-191 E360-0408-195 E360-0408-202 E360-0508-003 E360-0508-006 E360-0508-012 E360-0508-017 E360-0508-018 E360-0508-019 E360-0508-020 E360-0508-021 E360-0508-023 E360-0508-024 E360-0508-025 E360-0508-026 E360-0508-027 E360-0508-028 E360-0508-029 E360-0508-030 E360-0508-031 E360-0508-032 E360-0508-033 E360-0508-034 E360-0508-035 E360-0508-036 E360-0508-037 E360-0508-038 E360-0508-039 E360-0508-041 E360-0508-042 E360-0508-043 E360-0508-044 E360-0508-045 E360-0508-047 E360-0508-048 E360-0508-049 E360-0508-050 E360-0508-052 E360-0508-053 E360-0508-054 E360-0508-056 E360-0508-057 E360-0508-058 E360-0508-059 E360-0508-060 E360-0508-061 E360-0508-062 E360-0508-063 E360-0508-064 E360-0508-065 E360-0508-066 E360-0508-070 E360-0808-126 E360-0808-172 E360-0908-056 E360-0908-109 E360-0908-118 E360-0908-131 E360-0908-137 E360-0908-138 E360-0908-148 E360-0908-153 E360-0908-154 E360-0908-164 E360-0908-167 E360-0908-172 E360-0908-174 E360-0908-183 E360-0908-187 E360-0908-188 E360-0908-189 E360-1008-001 E360-1008-004 E360-1008-005 E360-1008-006 E360-1008-008 E360-1008-009 E360-1008-012 E360-1008-014 E360-1008-015 E360-1008-018 E360-1008-019 E360-1008-027 E360-1008-029 E360-1008-030 E360-1008-031 E360-1008-034 E360-1008-035 E360-1008-036 E360-1008-038 E360-1008-045 E360-1008-046 E360-1008-050 E360-1008-051 E360-1008-052 E360-1008-054 E360-1008-055 E360-1008-057 E360-1008-061 E360-1008-062 E360-1008-067 E360-1008-070 E360-1008-071 E360-1008-072 E360-1008-073 E360-1008-077 E360-1008-078 E360-1008-079 E360-1008-080 E360-1008-081 E360-1008-083 E360-1008-084 E360-1008-085 E360-1008-086 E360-1008-087 E360-1008-089 E360-1008-091 E360-1008-092 E360-1008-095 E360-1008-096 E360-1008-097 E360-1008-098 E360-1008-099 E360-1008-100 E360-1008-101 E360-1008-102 E360-1008-103 E360-1008-104 E360-1008-106 E360-1008-107 E360-1008-108 E360-1008-109 E360-1008-110 E360-1008-112 E360-1008-113 E360-1008-114 E360-1008-115 E360-1008-117 E360-1008-119 E360-1008-120 E360-1008-121 E360-1008-122 E360-1008-123 E360-1008-124 E360-1008-125 E360-1008-126 E360-1008-129 E360-1008-130 E360-1008-131 E360-1008-132 E360-1008-137 E360-1008-140 E360-1008-144 E360-1008-145 E360-1008-146 E360-1008-148 E360-1008-151 E360-1008-152 E360-1008-155 E360-1008-156 E360-1008-159 E360-1008-160 E360-1008-161 E360-1108-008 E360-1108-014 E360-1108-015 E360-1108-016 E360-1108-017 E360-1108-018 E360-1108-019 E360-1108-020 E360-1108-021 E360-1108-022 E360-1108-023 E360-1108-024 E360-1108-025 E360-1108-026 E360-1108-027 E360-1108-028 E360-1108-031 E360-1108-032 E360-1108-033 E360-1108-034 E360-1108-036 E360-1108-037 E360-1108-044 E360-1108-045 E360-1108-047 E360-1108-048 E360-1108-049 E360-1108-050 E360-1108-051 E360-1108-052 E360-1108-055 E360-1108-056 E360-1108-057 E360-1108-058 E360-1108-059 E360-1108-060 E360-1108-061 E360-1108-062 E360-1108-063 E360-1108-064 E360-1108-065 E360-1108-066 E360-1108-067 E360-1108-068 E360-1108-072 E360-1108-073 E360-1108-074 E360-1108-075 E360-1108-076 E360-1108-077 E360-1108-078 E360-1108-079 E360-1108-081 E360-1108-084 E360-1108-085 E360-1108-087 E360-1108-088 E360-1108-092 E360-1108-096 E360-1108-098 E360-1108-100 E360-1108-101 E360-1108-157 E360-1108-163 E360-1108-164 E360-1108-167 E360-1108-171 E360-1108-173 E360-1108-178 E360-1108-180 E360-1108-181 E360-1108-182 E360-1108-186 E360-1108-192 E360-1208-019 E360-1208-020 E360-1208-021 E360-1208-023 E360-1208-025 E360-1208-026 E360-1208-027 E360-1208-030 E360-1208-031 E360-1208-037 E360-1208-041 E360-1208-044 E360-1208-052 E360-1208-054 E360-1208-058 E360-1208-059 E360-1208-061 E360-1208-062 E360-1208-069 E360-1208-072 E360-1208-078 E360-1208-085 E360-1208-087 E360-1208-090 E360-1208-093 E360-1208-094 E360-1208-099 E360-1208-101 E360-1208-102 E360-1208-103 E360-1208-104 E360-1208-105 E360-1208-106 E360-1208-107 E360-1208-108 E360-1208-109 E360-1208-110 E360-1208-111 E360-1208-112 E360-1208-113 E360-1208-114 E360-1208-115 E360-1208-116 E360-1208-117 E360-1208-118 E360-1208-119 E360-1208-120 E360-1208-121 E360-1208-122 E360-1208-123 E360-1208-124 E360-1208-125 E360-1208-126 E360-1208-127 E360-1208-129 E360-1208-130 N07361111693 N07361111694 N07361111695 N07361111696 N07361111697 N07361111722 N07361111723 N07361111724 N07361111730 N07361111731 N07361111732 N07361111736 N07361111737 N07361111738 N07361111739 N07361111740 N07361111741 N07361111742 N07361111743 N07361111744 N07361111745 N07361111746 N07361111747 N07361111748 N07361111750 N07361111754 N07361111755 N07361111756 N07361111757 N07361111758 N07361111759 N07361111760 N07361111761 N07361111762 N07361111763 N07361111764 N07361111765 N07361111766 N07361111767 N07361111768 N08360912976 N08360913020 N08360913021 N08360913022 N08360913023 N08360913024 N08360913025 N08360913026 N08360913027 N08360913028 N08360913029 N08360913030 N08360913031 N08360913032 N08360913033 N08360913034 N08360913035 N08360913036 N08360913037 N08360913038 N08360913039 N08360913040 N08360913041 N08360913042 N08360913043 N08360913044 N08360913045 N08360913046 N08360913047 N08360913048 N08360913049 N08360913050 N08360913051 N08360913052 N08360913055 N08360913056 N08360913057 N08360913058 N08360913059 N08360913060 N08360913061 N08360913062 N08360913063 N08360913065 N08360913066 N08360913067 N08360913068 N08360913069 N08360913070 N08360913071 N08360913072 N08360913073 N08360913074 N08360913075 N08360913076 N08360913077 N08360913078 N08360913079 N08360913080 N08360913081 N05361110106 N06360310212 N06360710293 N06360710296 N06360710305 N06360710306 N06361010541 N08360111957 N08360111958 N08360812838 N06361210660 N09360413997 N06360910373 N06360910398 N06361010516 N07360911465 N07361211775 N08360212066 N08360212067 N08360212068 N08360212069 N08360612521 N08360612522 N08360612523 N08360612524 N08360612525 N08360612526 N08360712675 N09360413993 N06360610250 N06361010478 N06361110629 N07360110829 N07360110830 N07360210961 N07360210962 N07361011554 N07361111728 N07361111749 N07361211824 N07360811404 N08360913064 N07360210945 N06360910380 N06360910379 N07360210936 N06360910387 N07360611245 N07360611246 N07360611292 N07360711318 N07360711321 N07360711322 N08361013124 N08361013125 N08361013126 N08361013127 N08361013128 N08361013129 N08361013130 N08361013131 N08361013137 N08361013138 N08361013142 N08361013143 N08361013144 N08361013145 N08361013146 N08361013147 N09360413983 N09360413984 N09360413985 N09360413986 N09360413987 N09360413988 N09360413989 N09360413990 N07361111682 N06360810331 N06360810334 N06361010473 N06361010481 N07360711317 N07360811412 N07360811414 N07360811416 N07360911469 N07360911471 N07360911472 N07360911473 N07360911474 N07360911475 N07360911476 N07360911485 N07360911486 N07360911487 N07360911488 N07361011589 N07361011590 N07361011591 N08360913000 N08360913001 N08361013105 N08361013106 N09360113411 N09360113412 N09360113413 N09360113414 N09360213551 N09360213552 N09360213554 N09360213555 N06360810326 N06360810332 N06360810335 N06361110582 N07360110889 N07360110890 N07361211772 N07361211773 N08360111876 N08360111886 N08360111887 N08360111888 N08360111889 N08360111892 N08360111895 N08360111896 N08360111897 N08360111898 N08360111899 N08360312155 N08360312157 N08360312158 N08360312159 N08360312160 N08360312161 N08360312162 N08360312163 N08360312182 N08360612586 N08360612587 N08360612588 N08360612589 N08360612590 N08360712715 N08360712716 N08360712718 N08360912995 N08360912996 N08360912997 N08360913002 N08360913005 N08360913007 N08360913008 N08360913009 N08360913010 N08361113248 N08361113249 N08361113250 N08361113251 N08361113252 N08361213384 N08361213385 N08361213386 N08361213387 N08361213388 N08361213389 N08361213390 N08361213391 N08361213392 N08361213393 N08361213394 N08361213395 N08361213396 N08361213397 N08361213398 N08361213399 N08361213400 N06361210647 N06361210658 N06360210210* N06360610256 N06360610262 N06361010526 N06361010527 N07360110914 N07360511139 N07360511160 N07360811373 N07360811395 N07360811407 N07361011516 N07361111621 N07361111622 N07361111623 N07361111628 N08360312131 N08360512378 N08360512401 N08360612562 N08360612563 N08360612564 N08360612565 N08360612566 N08360612567 N08360612568 N08360612569 N08360612570 N08360612571 N08360612572 N08360612573 N08360612574 N08360612575 N08360612576 N08360612577 N08360812825 N08360812828 N08360812829 N08360812830 N08360812831 N08360812832 N08360812833 N08360812834 N08360812835 N08360812836 N08360812837 N08360812839 N08360812840 N08360812841 N08360812842 N08360812843 N08360812844 N08360812851 N08360812859 N08360812860 N08360812861 N08360812868 N08360812869 N08360812870 N08360812871 N08360812872 N08360812875 N08360812876 N08360812877 N08360812878 N08360912896 N08360912897 N08360912898 N08360912899 N08360912900 N08360912901 N08360912902 N08360912903 N08360912904 N08360912905 N08360912906 N08360912907 N08360912908 N08360912909 N08360912913 N08360912914 N08360912916 N08360912920 N08360912921 N08360912926 N08360912927 N08360912957 N08361013161 N08361013162 N08361113183 N08361113184 N09360113500 N09360113501 N08360312174 N08360612529 N08360612537 N08360612538 N08360612539 N08360612540 N08360612541 N08360612542 N08360612543 N08360612544 N08360612545 N08360612546 N08360612547 N08361013160 N08361013164 N07360611276 N07360611294 N07360611296 N07360611300 N07360611304 N07360611305 N07360611306 N07360611308 N07360611309 N07360611310 N07360611311 N07360611312 N07360611313 N07360611314 N07360611315 N07360711329 N07360711330 N07360711332 N07360711333 N07360711334 N07360711335 N07360711336 N07360711337 N07360711338 N07360711339 N07360711340 N07360711342 N07360711343 N07360711344 N07360711345 N07360711346 N07360711347 N07360711348 N07360711353 N07360811377 N07360811383 N07360811388 N06360910374 N06360910376 N07360110913 N07361011515 N08360412243 N08360412244 N08360612637 N08360612645 N08360712762 N08360712763 N08360912960 N08361013094 N09360213567 N06360910434 N06360910439 N06360910445 N06361010465 N06361010491 N06361010537 N06361010538 N06361010539 N06361010540 N06361210634 N07361211810 N06361110618 N08360312227 N09360313709 N06360210205 N06360210208 N06360210209 N06360210214 N06360310211 N06360310218 N06360310218 N06360310220 N06360610253 N07360110831 N07360110832 N07360110834 N07360110835 N07360110893 N07360210965 N07360411102 N07360411104 N07360411106 N07360411110 N07360411112 N07360411113 N07360411114 N07360411115 N07360411116 N07360411117 N07360411118 N07360411119 N07360411120 N07360411121 N07360411122 N07360411123 N07360411124 N07360411125 N07360411126 N07360511130 N07360511142 N07360511143 N07360511144 N07360511145 N07360511146 N07360511147 N07360511148 N07360511149 N07360511150 N07360511151 N07360511152 N07360511153 N07360511154 N07360511155 N07360511156 N07360511163 N07360511164 N07360511165 N07360511166 N07360511167 N07360511168 N07360511169 N07360511170 N07360511171 N07360511172 N07360511173 N07360511174 N07360511175 N07360511176 N07360511177 N07360511185 N07360511186 N07360511187 N07360511190 N07360511191 N07360511192 N07360511193 N07360511194 N07360511195 N07360511200 N07360511201 N07360611257 N07360611258 N07360611259 N07360611260 N07360611261 N07360611262 N07360611263 N07360611264 N07360611265 N07360611266 N07360611267 N07360611268 N07360611269 N07360611270 N07360611271 N07360611272 N07360611273 N07360611274 N07360611275 N08360111877 N08360111878 N08360111879 N08360111880 N08360111881 N08360111882 N08360111883 N08360111884 N08360111885 N08360111910 N08360111911 N08360111912 N08360111913 N08360111914 N08360111915 N08360111916 N08360111917 N08360111929 N08360111930 N08360111931 N08360111932 N08360111933 N08360111934 N08360111935 N08360111936 N08360111937 N08360111946 N08360111947 N08360111948 N08360111949 N08360111950 N08360111951 N08360111952 N08360111953 N08360111954 N08360111955 N08360111956 N08360212056 N08360212057 N08360212058 N08360212059 N08360212060 N08360212072 N08360212091 N08360212092 N08360212093 N08360212102 N08360212103 N08360212104 N08360212105 N08360212106 N08360212107 N08360212108 N08360212109 N08360212110 N08360212111 N08360212112 N08360212113 N08360212114 N08360212115 N08360212116 N08360312164 N08360312165 N08360312177 N08360312178 N08360312179 N08360312180 N08360312181 N08360312184 N08360312185 N08360312186 N08360312187 N08360312188 N08360312189 N08360312190 N08360312191 N08360312192 N08360312193 N08360312194 N08360312195 N08360312196 N08360312197 N08360312198 N08360312199 N08360312200 N08360312201 N08360312202 N08360312203 N08360312204 N08360312205 N08360312206 N08360312207 N08360312208 N08360312209 N08360312210 N08360312211 N08360312212 N08360312213 N08360312214 N08360312215 N08360312216 N08360312217 N08360312218 N08360312219 N08360312220 N08360312221 N08360312222 N08360312223 N08360312224 N08360312225 N08360312226 N08360512376 N08360512377 N08360512379 N08360512380 N08360512387 N08360512388 N08360512389 N08360512390 N08360512391 N08360512392 N08360512393 N08360512425 N08360512426 N08360512427 N08360512428 N08360512429 N08360512430 N08360512431 N08360512432 N08360512433 N08360512434 N08360512435 N08360512438 N08360512442 N08360512444 N08360512446 N08360512449 N08360512450 N08360512451 N08360512452 N08360512453 N08360512458 N08360512459 N08360512460 N08360512461 N08360512462 N08360512465 N08360512466 N08360512491 N08360512497 N08360512498 N08360512499 N08360512500 N08360512501 N08360512502 N08360512503 N08360512504 N08360512505 N08360512506 N08360512507 N08360512508 N08360512509 N08360512510 N08360512511 N08360512512 N08360512513 N08360512514 N08360512515 N08360512516 N08360512517 N08360612596 N08360612597 N08360612598 N08360612599 N08360612600 N08360612601 N08360612602 N08360612603 N08360612604 N08360612605 N08360612606 N08360612607 N08360612608 N08360612609 N08360612610 N08360612611 N08360612612 N08360612613 N08360612614 N08360612615 N08360612616 N08360612617 N08360612618 N08360612619 N08360612620 N08360612621 N08360612622 N08360612623 N08360612624 N08360612625 N08360612626 N08360612627 N08360612628 N08360612629 N08360612630 N08360612631 N08360612632 N08360612633 N08360612634 N08360612635 N08360812854 N08360812855 N08360812856 N08360812857 N09360113472 N09360113473 N09360113474 N09360113475 N09360113476 N09360113477 N09360113478 N09360113479 N09360113480 N09360113481 N09360113482 N09360113483 N09360113484 N09360113485 N09360113486 N09360113487 N09360113488 N09360113489 N09360113490 N09360113491 N09360113509 N09360113510 N09360113511 N09360113512 N09360113513 N09360113514 N09360113515 N09360113516 N09360113517 N09360113518 N09360113519 N09360113520 N09360113521 N09360113522 N09360113523 N09360113524 N09360113525 N09360113526 N09360113527 N09360113528 N09360113529 N09360113530 N09360113531 N09360113532 N09360113533 N09360113534 N09360113535 N09360113536 N09360113537 N09360113538 N09360113539 N09360113540 N09360113541 N09360113542 N09360113543 N09360113544 N09360113545 N09360113546 N09360113547 N09360113548 N09360313746 N09360413866 N09360413867 N09360413868 N09360413869 N09360413870 N09360413871 N09360413872 N09360413873 N09360413874 N09360413875 N09360413876 N09360413877 N09360413878 N09360413879 N09360413880 N09360413881 N09360413882 N09360413883 N09360413884 N09360413885 N09360413886 N09360413887 N09360413888 N09360413889 N09360413890 N09360413891 N09360413892 N09360413893 N09360413894 N09360413895 N09360413897 N09360413898 N09360413899 N09360413900 N09360413907 N09360413908 N09360413909 N09360413910 N09360413911 N09360413912 N09360413913 N09360413914 N09360413915 N09360413916 N09360413917 N09360413918 N09360413919 N09360413920 N09360413921 N09360413922 N09360413923 N09360413924 N09360413925 N09360413926 N09360413927 N09360413928 N09360413929 N09360413930 N09360413931 N09360413932 N09360413933 N09360413934 N09360413935 N09360413950 N09360413951 N09360413952 N09360413953 N09360413954 N09360413955 N09360413956 N09360413957 N09360413958 N09360413959 N09360413960 N09360413961 N09360413962 N09360413963 N09360413964 N09360413965 N09360413966 N09360413967 N09360413968 N09360413969 N09360413970 N09360413972 N09360413973 N09360413974 N09360413975 N09360413976 N09360413977 N09360413978 N06361210716 N07360110836 N07360110837 N07360110894 N07360110910 N07360110915 N06360210206 N06360510236 N06360610254 N06360610264 N06360610266 N06360610274 N06360710303 N06360710304 N06360810328 N06361210664 N06361210666 N06361210667 N06361210674 N06361210675 N06361210676 N06361210689 N07360110838 N07360110839 N07360110841 N07360110874 N07360411128 N07360511157 N07360511158 N07360511159 N07360611240 N07360611241 N07360611298 N07360811391 N07360811392 N07360811418 N07360811419 N07360811422 N07360811424 N07360811425 N07360811426 N07360811458 N07360811462 N07361011512 N07361011513 N07361011514 N07361111629 N07361111630 N07361111631 N07361111639 N07361111640 N07361111642 N07361211781 N07361211815 N07361211816 N07361211817 N07361211818 N07361211819 N07361211820 N08360111906 N08360111907 N08360111908 N08360111909 N08360212080 N08360212081 N08360312153 N08360412228 N08360412229 N08360412230 N08360412231 N08360712657 N08360712744 N08360712745 N08360712747 N08360712748 N08360712749 N08360912910 N08360912911 N08360912912 N08360912958 N08360912959 N08360912978 N08360913012 N08360913013 N08360913014 N08360913015 N08360913016 N08360913017 N08360913018 N08360913019 N09360113441 N09360113442 N09360313726 N09360313782 N09360313783 N09360313784 N09360313785 N09360313789 N08360111959 N08360111960 N08360111962 N08360212074 N08360212075 N06360510237 N06360710297 N06360710298 N06360710299 N06360710300 N06360910358 N06360910394 N06360910396 N06361110565 N06361110566 N06361110567 N06361110574 N06361110583 N06361110584 N06361110585 N06361110586 N06361110587 N06361110588 N06361110589 N06361110590 N06361110591 N06361110592 N06361110593 N06361110594 N06361110595 N06361110596 N06361110597 N06361110598 N06361110599 N06361110600 N06361110601 N06361110602 N06361110603 N06361110604 N06361110605 N06361110606 N06361110607 N06361110608 N06361110609 N06361110610 N06361110611 N06361110612 N06361110613 N06361110614 N06361110615 N06361110616 N06361110617 N07360511137 N07360611233 N07360611234 N07361011518 N07361011519 N08360812866 N08360812874 N08360913053 N08360913054 N08361013157 N08361013176 N08361013177 N09360113453 N09360113454 N07360911467 N07361011520 N07361011521 N07361011522 N08361013091 N08361013092 N08361013093 N06360210198 N06360210200 N073601111669 N07360310980 N07360310984 N07360310987 N07360310994 N07360311009 N07360311012 N07360311032 N07360511203 N07360511205 N07360511211 N07360511231 N07360511232 N07360611250 N07360811390 N07360811442 N07360811463 N07360811463 N07361011553 N07361011553 N07361111624 N07361111625 N07361111625 N07361111625 N07361211802 N07361211808 N07361211809 N07361211860 N07361211861 N08360111922 N08360111972 N08360212012 N08360212037 N08360212062 N08360412265 N08360512467 N08360712663 N08360712664 N08360712665 N08360712666 N08360712667 N08360712669 N08360812788 N08360812793 N08360812808 N08360912883 N08361213342 N06360610269 N07360110842 N07360110843 N07360110844 N07360110845 N07360110846 N07360110847 N07360110848 N07360310980 N07360310981 N07360310982 N07360310984 N07360310985 N07360310986 N07360310987 N07360310988 N07360310989 N07360310990 N07360310991 N07360310992 N07360310993 N07360310994 N07360310995 N07360310996 N07360310997 N07360310998 N07360310999 N07360311000 N07360311001 N07360311002 N07360311003 N07360311004 N07360311005 N07360311006 N07360311007 N07360311008 N07360311009 N07360311010 N07360311011 N07360311012 N07360311013 N07360311014 N07360311015 N07360311016 N07360311017 N07360311018 N07360311019 N07360311020 N07360311021 N07360311022 N07360311023 N07360311024 N07360311025 N07360311026 N07360311027 N07360311028 N07360311029 N07360311030 N07360311031 N07360311031 N07360311032 N07360311034 N07360311035 N07360311036 N07360311037 N07360311038 N07360311039 N07360311040 N07360311041 N07360311042 N07360311043 N07360311044 N07360311045 N07360311046 N07360311047 N07360311048 N07360311049 N07360311050 N07360311051 N07360311052 N07360311053 N07360311054 N07360311055 N07360311056 N07360311057 N07360311058 N07360311059 N07360311060 N07360311061 N07360311062 N07360311063 N07360311064 N07360311065 N07360311066 N07360311067 N07360311068 N07360311069 N07360311070 N07360311073 N07360311074 N07360311075 N07360311076 N07360311077 N07360311078 N07360311079 N07360311080 N07360311081 N07360311082 N07360311092 N07360511182 N07360511203 N07360511204 N07360511205 N07360511206 N07360511207 N07360511208 N07360511209 N07360511210 N07360511211 N07360511212 N07360511213 N07360511214 N07360511215 N07360511216 N07360511217 N07360511218 N07360511219 N07360511220 N07360511221 N07360511222 N07360511223 N07360511224 N07360511225 N07360511226 N07360511227 N07360511228 N07360511229 N07360511230 N07360611247 N07360611248 N07360611249 N07360611250 N07360611251 N07360611252 N07360611253 N07360611254 N07360611255 N07360611256 N07360611287 N07360611288 N07360611289 N07360611290 N07360611291 N07360811374 N07360811376 N07360811378 N07360811379 N07360811381 N07360811382 N07360811385 N07360811389 N07360811390 N07360811393 N07360811394 N07360811399 N07360811400 N07360811405 N07360811415 N07360811417 N07360811429 N07360811433 N07360811434 N07360811435 N07360811436 N07360811437 N07360811438 N07360811439 N07360811440 N07360811441 N07360811442 N07360811443 N07360811444 N07360811445 N07361011493 N07361011494 N07361011495 N07361011496 N07361011497 N07361011498 N07361011499 N07361011500 N07361011501 N07361011502 N07361011504 N07361011505 N07361011508 N07361011510 N07361011511 N07361011545 N07361011546 N07361011547 N07361011548 N07361011549 N07361011550 N07361011551 N07361011552 N07361011553 N07361111668 N07361111669 N07361111670 N07361111671 N07361111672 N07361111673 N07361111674 N07361111675 N07361111676 N07361111677 N07361111716 N07361111717 N07361111718 N07361111719 N07361111720 N07361111721 N07361111727 N07361111733 N07361111734 N07361111735 N07361211776 N07361211777 N07361211782 N07361211809 N07361211814 N07361211859 N07361211860 N07361211861 N07361211862 N07361211873 N08360111919 N08360111920 N08360111921 N08360111922 N08360111923 N08360111924 N08360111925 N08360111926 N08360111927 N08360111928 N08360111938 N08360111940 N08360111941 N08360111942 N08360111943 N08360111944 N08360111971 N08360111972 N08360111973 N08360111974 N08360111976 N08360111977 N08360111978 N08360113456 N08360113456 N08360113458 N08360113458 N08360113459 N08360113459 N08360113460 N08360113460 N08360211993 N08360211994 N08360211995 N08360211996 N08360211997 N08360211998 N08360211999 N08360212000 N08360212001 N08360212002 N08360212003 N08360212005 N08360212006 N08360212007 N08360212008 N08360212009 N08360212010 N08360212011 N08360212012 N08360212013 N08360212014 N08360212015 N08360212016 N08360212017 N08360212018 N08360212019 N08360212020 N08360212027 N08360212028 N08360212029 N08360212030 N08360212031 N08360212032 N08360212033 N08360212034 N08360212035 N08360212036 N08360212037 N08360212038 N08360212039 N08360212040 N08360212041 N08360212042 N08360212043 N08360212044 N08360212045 N08360212046 N08360212053 N08360212054 N08360212055 N08360212061 N08360212062 N08360212063 N08360212064 N08360212065 N08360212070 N08360212071 N08360212073 N08360212077 N08360212078 N08360212079 N08360212094 N08360212095 N08360212096 N08360212097 N08360212098 N08360212099 N08360212100 N08360212101 N08360312117 N08360312118 N08360312119 N08360312120 N08360312121 N08360312122 N08360312123 N08360312124 N08360312125 N08360312126 N08360312127 N08360312128 N08360312129 N08360312130 N08360312134 N08360312135 N08360312136 N08360312137 N08360312138 N08360312139 N08360312140 N08360312141 N08360312142 N08360312143 N08360312144 N08360312145 N08360312146 N08360312147 N08360312148 N08360412232 N08360412233 N08360412234 N08360412235 N08360412236 N08360412237 N08360412238 N08360412239 N08360412251 N08360412253 N08360412257 N08360412258 N08360412259 N08360412260 N08360412263 N08360412264 N08360412265 N08360412271 N08360412272 N08360412273 N08360512408 N08360512409 N08360512410 N08360512411 N08360512412 N08360512416 N08360512417 N08360512418 N08360512419 N08360512420 N08360512421 N08360512422 N08360512423 N08360512424 N08360512436 N08360512437 N08360512439 N08360512440 N08360512441 N08360512443 N08360512445 N08360512447 N08360512448 N08360512454 N08360512455 N08360512456 N08360512457 N08360512463 N08360512464 N08360512467 N08360512468 N08360512470 N08360512478 N08360512479 N08360512480 N08360512481 N08360512482 N08360512484 N08360512489 N08360512490 N08360712663 N08360712664 N08360712665 N08360712666 N08360712667 N08360712668 N08360712669 N08360712670 N08360712671 N08360712672 N08360712677 N08360712678 N08360712679 N08360712680 N08360712681 N08360712682 N08360712683 N08360712684 N08360712685 N08360712686 N08360812787 N08360812789 N08360812790 N08360812791 N08360812792 N08360812794 N08360812795 N08360812796 N08360812797 N08360812798 N08360812799 N08360812805 N08360812806 N08360812807 N08360812814 N08360812815 N08360812821 N08360812822 N08360812826 N08360812827 N08360812847 N08360812848 N08360812849 N08360812850 N08360812858 N08360812862 N08360812863 N08360812864 N08360812865 N08360812867 N08360912883 N08360912884 N08360912885 N08360912886 N08360912887 N08360912888 N08360912889 N08360912890 N08360912891 N08360912892 N08361013095 N08361013096 N08361013097 N08361013098 N08361013099 N08361013100 N08361013101 N08361013102 N08361013115 N08361013116 N08361013117 N08361013118 N08361013119 N08361013120 N08361013121 N08361013122 N08361013123 N08361013132 N08361013133 N08361013134 N08361013135 N08361013136 N08361013139 N08361013140 N08361013141 N08361113196 N08361113197 N08361113198 N08361113199 N08361113200 N08361113201 N08361113202 N08361113203 N08361113204 N08361113205 N08361113206 N08361113207 N08361113208 N08361113222 N08361113223 N08361113224 N08361113232 N08361113233 N08361113234 N08361113235 N08361113253 N08361113254 N08361113255 N08361113256 N08361113257 N08361113262 N08361113263 N08361113264 N08361113268 N08361213282 N08361213291 N08361213296 N08361213297 N08361213298 N08361213299 N08361213300 N08361213303 N08361213304 N08361213305 N08361213306 N08361213307 N08361213308 N08361213309 N08361213310 N08630212071 N09360113430 N09360113431 N09360113432 N09360113433 N09360113436 N09360113437 N09360113438 N09360113439 N09360113440 N09360113449 N09360113450 N09360113455 N09360113456 N09360113458 N09360113459 N09360113460 N09360213553 N09360213562 N09360213563 N09360213564 N09360213565 N09360213568 N09360213578 N09360213579 N09360213580 N09360213581 N09360213582 N09360213583 N09360213584 N09360213585 N09360213586 N09360213587 N09360213588 N09360213589 N09360213590 N09360213591 N09360213593 N09360213594 N09360213595 N09360213596 N09360213598 N09360213601 N09360213602 N09360213604 N09360213605 N09360213607 N09360213608 N09360213612 N09360213613 N09360213615 N09360213618 N09360213619 N09360213620 N09360213644 N09360213647 N09360213648 N09360213649 N09360213650 N09360213651 N09360213652 N09360213653 N09360213654 N09360213655 N09360213656 N09360213675 N09360213676 N09360213677 N09360213678 N09360213679 N09360213680 N09360213681 N09360213682 N09360213683 N09360213684 N09360213685 N09360213686 N09360213690 N09360213691 N09360213692 N09360213693 N09360213694 N09360213695 N09360313721 N09360313722 N09360313723 N09360313724 N09360313732 N09360313733 N09360313741 N09360313760 N09360313770 N09360313775 N09360313776 N09360313777 N09360313778 N09360313780 N09360313781 N09360413902 N09360413903 N09360413904 N09360413905 N09360413906 N09360413936 N09360413937 N09360413938 N09360413939 N09360413940 N09360413941 N09360413947 N09360413948 N09360413949 N09360413971 N09360414014 N09360414029 N09360414030 N09360414031 N09360414032 N09360414033 N09360414035 N09360414036 N09360414037 N07360110888 N08360212083 N08360212086 N08360212087 N08360312170 N08360312171 N08360412245 N08360412246 N08360412248 N08360412250 N08360412266 N08360412365 N08361213280 N08361213281 N09360413991 N09360413998 N09360413999 N09360414001 N09360414015 N08360412241 N07360511162 N06360410228 N06360810325 N07360210955 N07360210956 N07360611239 N07360711316 N07360711323 N07360711325 N07361111643 N07361111644 N07361111647 N07361111658 N07361111659 N07361111660 N07361111662 N07361111663 N07361111664 N07361111665 N07361111666 N07361111667 N07361111678 N07361111679 N07361111685 N08360212047 N08360212089 N08360212090 N08360312173 N08360412267 N08360512404 N08360512405 N08360512413 N08360712673 N08360712674 N08360912961 N08360912963 N08360912970 N08360912977 N07361211780 N08360212076 N07361011517 N06360410230 N08360412368 N08360412369 N08360412374 N08360912969 N08360912973 N08360912974 N08360912975 N08361113187 N08361113188 N08361213285 N08361213286 N08361213287 N08361213288 N08361213289 N08361213290 N08361013156 N08361113240 N08361113241 N08361113242 N08361113243 N09360414000 N06360910390 N06361010535 N06361110568 N06361210630 N07360611295 N07360611297 N08360111961 N08360212082 N08360212084 N08360212085 N08360412375 N08360512473 N08360812852 N08360812853 N06360510234 N06360810327 N06360810329 N06360810330 N06361210670 N06361210671 N06361210672 N06361210673 N07360411103 N07360511179 N07360511180 N07361211774 N08360412262 N08360912966 N06360610246 N06360610261 N06361110573 N06361110575 N06361110576 N06361110577 N06361110579 N07360110901 N07360110902 N07360110903 N07360110904 N07360110905 N07360110906 N07360110907 N07360110908 N07360911470 N07361111686 N07361211799 N07361211800 N07361211801 N08360612591 N08360612592 N08360612594 N08360612595 N08360712714 N08360712717 N08360912928 N08360912929 N08360912930 N08360912998 N08360912999 N08360913003 N08360913004 N08360913006 N08361113189 N08361113190 N08361113191 N08361113192 N08361113193 N08361113194 N08361113195 N09360113406 N09360113407 N09360113408 N09360113409 N09360113410 N09360313697 N09360313698 N09360313699 N09360313700 N09360313701 N09360313702 N09360313703 N09360313704 N09360313705 N09360313706 N09360313707 N09360313708 N09360514039 N09360514040 N09360514041 N09360514042 N08360111945 N08360211991 N08360211992 N06360410224 N06360410229 N06360610240 N06360610241 N06360610242 N06360610243 N06360610244 N06360610255 N06360610263 N06360610267 N06360810333 N06360910391 N07360711352 N08360412252 N08360412254 N08360412255 N08360412256 N09360413979 N09360413980 N09360413981 N09360413982 N09360414004 N09360414005 N09360414006 N09360414007 N09360414008 N09360414009 N09360414010 N09360414011 N09360414012 N09360414013 N09360414016 N09360414017 N09360414018 N09360414019 N09360414020 N09360414021 N09360414022 N09360414023 N09360414024 N09360414025 N09360414026 N09360414027 N09360414028 N09360414034 N07361011509 N08360111918 N08360111969 N08360111970 N08360111979 N08360111980 N08360111981 N08360111982 N08360111983 N08360111984 N08360111985 N08360111986 N08360612527 N08360712761 N09360213687 N09360213688 N09360213689 N09360213560 N08360812768 N08360812769 N08360812770 N08360812771 N08360812772 N08361013150 N08361013152 N08361013155 N08361013170 N08361013171 N08361013172 N08361013173 N08361013174 N08361213343 N08361213344 N08361213345 N08361213346 N08361213347 N08360312172 N08360312175 N08360612560 N08360712656 N09360213570 N09360213572 N09360213574 N09360213575 N09360213603 N09360213609 N06360610275 N06360910381 N06360910382 N06360910383 N06360910384 N06360910385 N06360910386 N07361111641 N08360412367 N08360512394 N08360812779 N08360812802 N08360812803 N08360912893 N08360912894 N08360912895 N07360711365 N07360811451 N07360811452 N07360811453 N07360811454 N07360811455 N07360811456 N07360811457 N08360111905 N08360312133 N08360312154 N08360912962 N08360912965 N08360912968 N08360912990 N08360912991 N08360912992 N08360912993 N08360912994 N09360113457 N09360113461 N09360113462 N09360113463 N09360113465 N09360113466 N09360113467 N09360113468 N09360113469 N09360113470 N09360113471 N09360113492 N07361111687 N07361111729 N08361013086 N08361013087 N08361013088 N08361013089 N08361013090 N08361013148 N08361013149 N08361013151 N08361013154 N08361213292 N08361213293 N08361213294 N08361213348 N08361213349 N06360210202 N06360610239 N06360610252 N06361010517 N07361111661 N07361211826 N07361211827 N07361211828 N07361211829 N07361211830 N08360111900 N08360111901 N08360111902 N08360111903 N08360111904 N08361213350 N08361213351 N08361213352 N08361213353 N08361213354 N08361213355 N08361213356 N08361213357 N08361213358 N08361213359 N08361213360 N08361213361 N08361213362 N08361213363 N08361213364 N08361213365 N08361213366 N08361213367 N08361213368 N09360213592 N09360213597 N09360213599 N09360213611 N09360213614 N09360213621 N09360213622 N09360213623 N09360213624 N09360213625 N09360213626 N09360213627 N09360213628 N09360213629 N09360213630 N09360213631 N09360213632 N09360213633 N09360213634 N09360213666 N09360213667 N09360213668 N09360213669 N09360213670 N09360213671 N09360213672 N09360213673 N09360213674 N06360710301 N06360710302 N07360611280 N07361211811 N08360312156 N08360512406 N08360512407 N08360812785 N08360812786 N09360213616 N09360213617 N06360410227 N06360710288 N06360710289 N06360810320 N06360810322 N06360810324 N06360910356 N06360910357 N06361010524 N06361010525 N07360210939 N07360210940 N07360210941 N07360210942 N07360210943 N07361111751 N07361111752 N07361111753 N07361211778 N07361211779 N07361211838 N07361211839 N07361211840 N07361211841 N08360312149 N08360312150 N08360312151 N08360312152 N08360312168 N08360312169 N08360412370 N08360412371 N08360412372 N08360412373 N08360712739 N08360712740 N08360712741 N08360712742 N08360712743 N08361113185 N08361113186 N09360113434 N09360113435 N09360113443 N09360113444 N09360113445 N07360210958 N07360511132 N07360511133 N07360511140 N07360511141 N07360711320 N07360711326 N07360711327 N07360711371 N07360811409 N07360811421 N07360811423 N07360911478 N07361011588 N07361011592 N07361011593 N07361011594 N07361011595 N07361011596 N07361011597 N07361011598 N07361011599 N07361011600 N07361011601 N07361011602 N07361011603 N07361011604 N07361011606 N07361111704 N07361111705 N07361111714 N07361111715 N08360512483 N08360512485 N08360512486 N08360512487 N08360512488 N08360612531 N08360612532 N08360612534 N08360612535 N08360612536 N06360610270 N06360910393 N06361210632 N06360810337 N06360910349 N06360910353 N06360910355 N07361111726 N08360212048 N08360212049 N08360212050 N08360212051 N08360212052 N08360512469 N08360512471 N08360512472 E360-0108-062 E360-0108-079 E360-0108-083 E360-0108-084 E360-0108-089 E360-0208-013 E360-0208-014 E360-0208-015 E360-0208-017 E360-0208-023 E360-0208-032 E360-0208-045 E360-0208-052 E360-0208-056 E360-0208-060 E360-0208-066 E360-0208-067 E360-0208-073 E360-0208-074 E360-0208-078 E360-0208-079 E360-0208-080 E360-0208-082 E360-0208-083 E360-0208-091 E360-0208-092 E360-0208-097 E360-0208-101 E360-0208-102 E360-0208-103 E360-0208-105 E360-0208-109 E360-0208-117 E360-0208-118 E360-0208-121 E360-0208-132 E360-0208-136 E360-0208-139 E360-0208-140 E360-0208-141 E360-0208-142 E360-0208-143 E360-0208-145 E360-0208-147 E360-0208-148 E360-0208-152 E360-0208-153 E360-0208-154 E360-0208-157 E360-0208-160 E360-0208-161 E360-0208-163 E360-0208-165 E360-0208-166 E360-0208-168 E360-0208-170 E360-0208-172 E360-0208-173 E360-0208-175 E360-0208-178 E360-0208-179 E360-0208-180 E360-0208-181 E360-0208-182 E360-0208-184 E360-0208-186 E360-0208-187 E360-0208-188 E360-0208-191 E360-0208-192 E360-0208-193 E360-0208-194 E360-0208-197 E360-0208-198 E360-0208-199 E360-0208-200 E360-0208-201 E360-0208-202 E360-0208-203 E360-0208-204 E360-0208-206 E360-0208-207 E360-0208-208 E360-0208-210 E360-0208-212 E360-0208-214 E360-0208-218 E360-0208-220 E360-0208-221 E360-0208-222 E360-0208-223 E360-0208-224 E360-0208-226 E360-0208-228 E360-0208-230 E360-0208-231 E360-0208-232 E360-0208-233 E360-0208-242 E360-0208-243 N06360410232 N06361010515 N06361110623 N06361110624 N06361110625 N06361110626 N06361110627 N06361110628 N07360511134 N07360511136 N07360511138 N07360711328 N07360711341 N07360711349 N07360811375 N07360811461 N07361011605 N07361011607 N07361011608 N07361011609 N07361011610 N07361011614 N07361111702 N07361111706 N07361111707 N07361211770 N07361211771 N07361211822 N07361211823 N07361211842 N07361211843 N07361211844 N07361211845 N07361211846 N07361211847 N07361211848 N07361211849 N07361211850 N07361211851 N08360312132 N08360312167 N08360412240 N08360412242 N08360412247 N08360512403 N08360612530 N08360612533 N08360612559 N08360812773 N08360812774 N08360812775 N08360812776 N08360812777 N08360812778 N08360812780 N08360812781 N08360812782 N08360812783 N08360812784 N08360912931 N08360912932 N08360912933 N08360912934 N08360912935 N08360912936 N08360912937 N08360912938 N08360912939 N08360912940 N08360912941 N08360912942 N08360912943 N08360912944 N08360912979 N08360912980 N08360912981 N08360912982 N08360912983 N08360912984 N08360912985 N08360912986 N08360912987 N08360912988 N08360912989 N08360913011 N06360210140 N06361210668 N07360110912 N07360110916 N07360110917 N07360110918 N07360110919 N07360210920 N07360210921 N07360210923 N07360210924 N07360210925 N07360210946 N07360210947 N07360210948 N07360210949 N07360210950 N07360511181 N07360511183 N07360511184 N07360511188 N07360511189 N07360511196 N07360511197 N07360511198 N07360511199 N07360511202 N07360711331 N07360711351 N07360711354 N07360711355 N07360711356 N07360711357 N07360711358 N07360711359 N07360711366 N07360711367 N07361111680 N07361111681 N07361111683 N07361111684 N07361111689 N07361111690 N07361111691 N07361111692 N07361111692* N07361111698 N07361111699 N08360212088 N08360412249 N08360412269 N08360412270 N08360412274 N08360412275 N08360412276 N08360412277 N08360412278 N08360412279 N08360412280 N08360412281 N08360412282 N08360412283 N08360412284 N08360412285 N08360412286 N08360412287 N08360412288 N08360412289 N08360412290 N08360412291 N08360412292 N08360412293 N08360412294 N08360412295 N08360412296 N08360412297 N08360412298 N08360412299 N08360412300 N08360412301 N08360412302 N08360412303 N08360412304 N08360412305 N08360412306 N08360412307 N08360412308 N08360412309 N08360412310 N08360412311 N08360412312 N08360412313 N08360412314 N08360412315 N08360412316 N08360412317 N08360412318 N08360412319 N08360412320 N08360412321 N08360412322 N08360412323 N08360412324 N08360412325 N08360412326 N08360412327 N08360412328 N08360412329 N08360412330 N08360412331 N08360412332 N08360412333 N08360412334 N08360412335 N08360412336 N08360412337 N08360412338 N08360412339 N08360412340 N08360412341 N08360412342 N08360412343 N08360412344 N08360412345 N08360412346 N08360412347 N08360412348 N08360412349 N08360412350 N08360412351 N08360412352 N08360412353 N08360412354 N08360412355 N08360412356 N08360412357 N08360412358 N08360412359 N08360412360 N08360412361 N08360412362 N08360412363 N08360412364 N08361113209 N08361113210 N08361113211 N08361113212 N08361113213 N08361113214 N08361113215 N08361113216 N08361113217 N08361113218 N08361113219 N08361113220 N08361113221 N08361113225 N08361113226 N08361113227 N08361113228 N08361113229 N08361113230 N08361113231 N08361113236 N08361113237 N08361113238 N08361113239 N08361113244 N08361113245 N08361113246 N08361113247 N08361113258 N08361113259 N08361113260 N08361113261 N08361113265 N08361113266 N08361113267 N08361113269 N08361113270 N08361113271 N08361113272 N08361113273 N08361113274 N08361113275 N08361113276 N08361113277 N08361113278 N08361113279 N09360113415 N09360113416 N09360113417 N09360113418 N09360113419 N09360113420 N09360113421 N09360313725 N09360313744 N09360313745 N09360313747 N09360313748 N09360313749 N09360313750 N09360313751 N09360313752 N09360313753 N09360313754 N09360313755 N09360313756 N09360313757 N09360313758 N09360313761 N09360313762 N09360313763 N09360313764 N09360313765 N09360313769 N09360313771 N09360313773 N09360313774 N09360313779 N09360313788 N09360313790 N09360313791 N09360313792 N09360313793 N09360313794 N09360313795 N09360313796 N09360313797 N09360313798 N09360313799 N09360313800 N09360313801 N09360313802 N09360313803 N09360313804 N09360313805 N09360313806 N09360313807 N09360313808 N09360313809 N09360313810 N09360313811 N09360313812 N09360313813 N09360313814 N09360313815 N09360313816 N09360313817 N09360313818 N09360313819 N09360313820 N09360313821 N09360313822 N09360313823 N09360313824 N09360313825 N09360313826 N09360313827 N09360313828 N09360313829 N09360313830 N09360313831 N09360313833 N09360313834 N09360313835 N09360313836 N09360313837 N09360313838 N09360313839 N09360313840 N09360313841 N09360313842 N09360313843 N09360313844 N09360313845 N09360313846 N09360313848 N09360313849 N09360313850 N09360313851 N09360313852 N09360313853 N09360313854 N09360313855 N09360313856 N09360313857 N09360313858 N09360313859 N09360313860 N09360313861 N09360313862 N09360313863 N09360313864 N09360313865 N08360111939 N08360111963 N08360111964 N08360111965 N08360111966 N08360111967 N08360111968 N08360111975 N03631110581 N06360310221 N06360610276 N06360810338 N06360910395 N07360210971 N07360210973 N07360210974 N07360210975 N07360210976 N07360210977 N07360210978 N07360210979 N07360310983 N07360311071 N07360311072 N07360311083 N07360311084 N07360311085 N07360311086 N07360311087 N07360311088 N08360612636 N08360612638 N08360612639 N08360612640 N08360612641 N08360612642 N08360612643 N08360612644 N09360113422 N09360113423 N09360113424 N09360113425 N09360113426 N09360113427 N09360113428 N09360113429 N09360113464 N09360113493 N09360113497 N09360113498 N09360113499 N09360113502 N09360113503 N09360113504 N09360113505 N09360113506 N09360113507 N09360113508 N06360810339 N07361111645 N07361111646 N08360312176 N08360612584 N08360612585 N09360213566 N06361110569 N06361210640 N06361210641 N06361210642 N06361210643 N07361211821 N08360212021 N08360212022 N08360212023 N08360212024 N08360212025 N08360212026 N08360612548 N08360612549 N08360612550 N08360612551 N08360612552 N08360612553 N08360612554 N08360612555 N08360612556 N08360612557 N08360612558 N09360213557 N09360213558 N09360213559 N09360213569 N09360213571 N09360213573 N09360213576 N09360213577 N09360413901 N06360310215 N06360410233 N06360810336 N07361111688 N08360412261 N08360412268 N08360512414 N08360512415 N08360612518 N08360612519 N08360612520 N08360812845 N08360812846 N08360912879 N08360912880 N08360912881 N08360912882 N08360912964 N08360912971 N08360912972 N09360113403 N09360113404 N09360113405 N09360113446 N09360113447 N09360113448 N06360210199 N06360610260 N06361010549 N06361110619 N06361110620 N06361110621 N06361110622 N06361210635 N06361210637 N06361210639 N06361210644 N06361210645 N06361210648 N06361210649 N06361210650 N06361210651 N06361210652 N06361210653 N06361210654 N06361210655 N06361210656 N06361210657 N06361210659 N06361210661 N06361210662 N06361210677 N06361210678 N06361210679 N06361210680 N06361210681 N06361210682 N06361210683 N06361210684 N06361210685 N06361210686 N06361210687 N06361210688 N08361213301 N08361213302 N08361213311 N08361213312 N08361213313 N08361213314 N08361213315 N08361213316 N08361213317 N08361213318 N08361213319 N08361213320 N08361213321 N08361213322 N08361213323 N08361213324 N08361213325 N08361213326 N08361213327 N08361213328 N08361213329 N08361213330 N08361213331 N08361213332 N08361213333 N08361213334 N08361213335 N08361213336 N08361213337 N08361213338 N08361213339 N08361213340 N08361213341 N09360213635 N09360213636 N09360213637 N09360213638 N09360213639 N09360213640 N09360213641 N09360213642 N09360213643 N09360213645 N09360213646 N09360213657 N09360213658 N09360213659 N09360213660 N09360213661 N09360213662 N09360213663 N09360213664 N09360213665 N06360410231 N06360510235 N06360610247 N06360610248 N06360610249 N06360610251 N06360610272 N06360610273 N06360610277 N06360610278 N06360610279 N06360610280 N06360610281 N06360710282 N06360710283 N06360710284 N06360710285 N06360710286 N06360710287 N06360710290 N06360710291 N06360710292 N06360810315 N06360910341 N06360910342 N06360910343 N06360910344 N06360910345 N06360910346 N06360910347 N06360910348 N06360910359 N07360110814 N07360110815 N07360110816 N07360110817 N07360110879 N07360110880 N07360110892 N07360110895 N07360110896 N07360110897 N07360110899 N07360110900 N07360411097 N07360411098 N07360411099 N07360411100 N07360511178 N07360611237 N07360611238 N07360711350 N07360911466 N07360911468 N07360911477 N07360911479 N07360911480 N07360911481 N07360911482 N07360911483 N07360911484 N07361011536 N07361211837 N07361211852 N07361211853 N07361211854 N07361211855 N07361211856 N07361211857 N08360912922 N08360912923 N08361013103 N08361013104 N09360113494 N09360113495 N09360113496 N09360413992 N09360413994 N09360413996 N07360411093 N07361011503 N07361011555 N07361011556 N07361011557 N07361011558 N07361011559 N07361011560 N07361011561 N07361011562 N07361011563 N07361011564 N07361011565 N07361011566 N07361011567 N07361011568 N07361011569 N07361011570 N07361011571 N07361011572 N07361011573 N07361011574 N07361011575 N07361011576 N07361011577 N07361011578 N07361011579 N07361011580 N07361011581 N07361011582 N07361011583 N07361011584 N07361011585 N07361011586 N07361011587 N08360912915 N08360912917 N08360912918 N08360912945 N08360912946 N08360912947 N08360912948 N08360912951 N08360912952 N08360912953 N08360912954 N08360912956 N08360912967 N08361013082 N08361013083 N08361013084 N08361013085 N08361013158 N08361013159 N06360210101 N06360210102 N06360210103 N06360210104 N06360210105 N06360210106 N06360210107 N06360210108 N06360210109 N06360210110 N06360210111 N06360210112 N06360210113 N06360210114 N06360210115 N06360210116 N06360210117 N06360210118 N06360210119 N06360210120 N06360210121 N06360210122 N06360210123 N06360210124 N06360210125 N06360210126 N06360210127 N06360210128 N06360210129 N06360210130 N06360210131 N06360210132 N06360210133 N06360210134 N06360210135 N06360210136 N06360210137 N06360210138 N06360210139 N06360210141 N06360210142 N06360210143 N06360210144 N06360210145 N06360210146 N06360210147 N06360210148 N06360210149 N06360210150 N06360210151 N06360210152 N06360210153 N06360210154 N06360210155 N06360210156 N06360210157 N06360210158 N06360210159 N06360210160 N06360210161 N06360210162 N06360210163 N06360210164 N06360210165 N06360210166 N06360210167 N06360210168 N06360210169 N06360210170 N06360210171 N06360210172 N06360210173 N06360210174 N06360210175 N06360210176 N06360210177 N06360210178 N06360210179 N06360210180 N06360210181 N06360210182 N06360210183 N06360210184 N06360210185 N06360210186 N06360210187 N06360210188 N06360210189 N06360210190 N06360210191 N06360210192 N06360210193 N06360210194 N06360210195 N06360210196 N06360210197 N06360210203 N06361110578 N06361210636 N06361210646 N06361210663 N06361210665 N06361210690 N06361210691 N06361210692 N06361210694 N06361210695 N06361210696 N06361210697 N06361210698 N06361210699 N06361210700 N06361210701 N06361210702 N06361210703 N06361210704 N06361210705 N06361210707 N06361210708 N06361210709 N06361210710 N06361210711 N06361210712 N06361210713 N06361210714 N06361210715 N06361210717 N06361210718 N06361210719 N06361210720 N06361210721 N06361210722 N06361210723 N06361210724 N06361210725 N06361210726 N06361210727 N06361210728 N06361210729 N06361210730 N06361210731 N06361210732 N06361210733 N06361210734 N06361210735 N06361210736 N06361210737 N06361210738 N06361210739 N06361210740 N06361210741 N06361210742 N06361210743 N06361210744 N06361210745 N06361210746 N06361210747 N06361210748 N06361210749 N06361210750 N06361210751 N06361210752 N06361210753 N06361210754 N06361210755 N06361210756 N06361210757 N06361210758 N06361210759 N07360110760 N07360110761 N07360110762 N07360110763 N07360110764 N07360110765 N07360110766 N07360110767 N07360110768 N07360110769 N07360110770 N07360110771 N07360110772 N07360110773 N07360110774 N07360110775 N07360110776 N07360110777 N07360110778 N07360110779 N07360110780 N07360110781 N07360110782 N07360110783 N07360110784 N07360110785 N07360110786 N07360110787 N07360110788 N07360110789 N07360110790 N07360110791 N07360110792 N07360110793 N07360110794 N07360110795 N07360110798 N07360110799 N07360110800 N07360110801 N07360110802 N07360110803 N07360110804 N07360110805 N07360110806 N07360110807 N07360110808 N07360110809 N07360110810 N07360110811 N07360110812 N07360110813 N07360110818 N07360110819 N07360110820 N07360110821 N07360110822 N07360110823 N07360210927 N07360210928 N07360210929 N07360210930 N07360210931 N07360210932 N07360210933 N07360210934 N07360210935 N07360210957 N07360210960 N07360210969 N07360411094 N07360811408 N07360811420 N07360811428 N07360811446 N07360811447 N07360811448 N07360811449 N07360811450 N07360811459 N07360811460 N07361011611 N07361011612 N07361011613 N07361011615 N07361011616 N07361011617 N07361011618 N07361011619 N07361211803 N07361211804 N07361211805 N07361211806 N07361211807 N08360212004 N08360512492 N08360512493 N08360512494 N08360512495 N08360512496 N08360712676 N08360712719 N08360712720 N08360712721 N08360712722 N08360712723 N08360712724 N08360712725 N08360712727 N08360712728 N08360712729 N08360712730 N08360712731 N08360712732 N08360712733 N08360712734 N08360712735 N08360712736 N08360712737 N08360712738 N09360313711 N09360313712 N09360313713 N09360313714 N09360313715 N09360313716 N09360313717 N09360313718 N09360313719 N09360313720 N06360910397 N07361111708 N07361111709 N07361111710 N08360712726 N08360712746 N08360712750 N08360712751 N08360712752 N08360712753 N08360712754 N08360712755 N08360712756 N08360712757 N08360712758 N08360712759 N08360712760 N08360712764 N08360712765 N08360712766 N08360712767 N08360912924 N08360912925 N08360912949 N08360912950 N08360912955 N08361013163 N08361213401 N08361213402 N07360511131 N07360511135 N07360611301 N07360611302 N07360611303 N07360611307 N07360811430 N07360811431 N07360811432 N07361111700 N07361111701 N07361111703 N08360512474 N08360512475 N08360512476 N08360512477 N06360910369 N07360110877 N08360912919 N09360213556 N09360213561 N09360213606 N09360213610 N07360210922 and N07360110840.
Recalling Firm/
Manufacturer
Newport Medical Instruments Inc
1620 Sunflower Ave
Costa Mesa, California 92626
Manufacturer Reason
for Recall
In December 2008, Newport initiated a field correction to upgrade e360 ventilators with software versions 3.0 or higher to software version 3.3. As a result of this upgrade, some customers experienced unexpected device alert alarms. Newport's investigation into these complaints confirmed that the 3.3 software responds too quickly to certain situations, giving a false device alert. Newport condu
FDA Determined
Cause 2
DESIGN: Software Design
Action Newport Medical initally sent out a Field Correction Notice to a limited number consignees dated December 17, 2008 and then the notice was expanded to include all e360 consignees dated June 15, 2009. The firm notified consignees that they have revised the e360 software to version 3.8. A customer verification form was attached to verify upgrade completion. For further questions, contact Newport Customer Technical Support Department at 1-800-451-3111 or +1-714-427-5811 Extension 500.
Quantity in Commerce 4,117 units (214 Domestically; 3,903 Internationally)
Distribution Worldwide Distribution -- US (including AL, AZ, CA, CO, FL, GA, IL, IN, KY, MA, MD, ME, MI, NJ, NM, NY, OH, OK, OR, PA, UT and VA), Albania, Algeria, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Brunei, Bulgaria, Canada, Chile, Colombia, Czech Republic, Dominican Republic, Ecuador, Egypt, England, Finland, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Japan, Jordan, Kazahkstan, Kenya, Korea, Kuwait, Lebanon, Liechtenstein, Lithuania, Malaysia, Mauritius, Mexico, Morocco, Nambia, Nepal, New Zealand, Nicaragua, Nigeria, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, Tunisia, Turkey, AUE, Ukraine, United Kingdom, Yemen Republic and Zimbabwe.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = CBK and Original Applicant = NEWPORT MEDICAL INSTRUMENTS, INC.
-
-