  Class 2 Device Recall Symbia S Series SPECT System see related information
Date Initiated by Firm March 11, 2009
Date Posting Updated July 07, 2009
Recall Status1 Terminated 3 on September 11, 2010
Recall Number Z-1648-2009
Recall Event ID 51483
510(K)Number K041166  
Product Classification Single-Photon Emission Computed Tomography - Product Code KPS
Product Siemens Medical Solutions USA, Inc., Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Part number 8717741.
Code Information Part number 8717741, serial numbers 1001, 1002, 1003, 1005, 1006, 1014, 1052, 1066, 1070, 1073, 1074, 1081, 1085, 1108, 1109, 1110, 1112, 1113, 1117, 1118, 1119, 1120, 1123, 1124, 1126, 1128, 1129, 1131, 1132, 1137, 1138, 1142, 1144, 1148, 1159, 1161, 1162, 1163, 1167, 1171, 1172, 1173, 1174, 1176, 1177, 1179, 1180, 1183, 1186, 1187, 1189, 1192, 1196, 1197, 1198, 1200, 1201, 1203, 1204, 1205, 1206, 1207, 1211, 1212, 1214, 1215, 1216, 1217, 1218, 1223, 1224, 1225, 1227, 1229, 1230, 1231, 1236, 1239, 1240, 1241, 1242, 1243, 1244, 1245, 1246, 1248, 1249, 1252, 1253, 1254, 1257, 1258, 1259, 1262, 1263, 1264, 1266, 1267, 1268, 1270, 1271, 1272, 1274, 1276, 1281, 1282, 1285, 1287, 1288, 1291, 1292, 1293, 1295, 1296, 1298, 1299, 1300, 1301, 1302, 1303, 1304, 1305, 1306, 1308, 1309, 1310, 1311, 1312, 1313, 1314, 1316, 1318, 1319, 1322, 1323, 1327, 1328, 1329, 1332, 1334, 1336, 1337, 1340, 1341, 1343, 1346, 1347, 1348, 1349, 1350, 1351, and 1352
Recalling Firm/
Manufacturer
Siemens Medical Solutions USA, Inc.
2501 Barrington Rd
Hoffman Estates IL 60195-2061
For Additional Information Contact Mr. Ron Nolte
847-304-7700
Manufacturer Reason
for Recall
Patient injury can occur when the patient is lying on the patient bed and the automatic collimator exchanger is undergoing a collimator change operation at the launch of an acquisition workflow. Detector motions required for this change can cause collision with the patient or pallet if it is in brain scan position, i.e. the patient or pallet is into the field of view.
FDA Determined
Cause 2
Software design
Action Siemens sent Urgent Field Correction Recall letters dated 1/20/09 to all users of the Symbia S and Symbia T Systems. The accounts were informed of the potential injury to patients while lying on the patient bed and the automatic collimator exchanger is undergoing a collimator change operation at the launch of an acquisition workflow. The customers were advised to perform all necessary collimator changes before the patient is positioned on the patient bed, and to examine each of their workflows that include a Tomo Acquisition activity and make sure the Auto Collimator Change check box on the Camera Parameters Tab is not checked. They also were advised that a Siemens Customer Service Representative would contact them within 60 days to schedule a corrective software upgrade free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance. The accounts were requested to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter.
Quantity in Commerce 152 units
Distribution Worldwide distribution, including USA, Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Italy, Japan, Korea, Kuwait, Netherlands, Norway, Poland, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Sweden and Switzerland.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.