• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Flexima and Percuflex Drainage Catheter Family (Pigtail Style)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Flexima and Percuflex Drainage Catheter Family (Pigtail Style) see related information
Date Posted May 21, 2009
Recall Status1 Terminated on February 23, 2012
Recall Number Z-1340-2009
Recall Event ID 51334
510(K)Number K944290 
Product Classification Suprapubic Drainage Tube - Product Code FFA
Product Boston Scientific, Flexima¿ and Percuflex¿ Drainage Catheter Family (Pigtail Style)

UPN# DESCRIPTION


435428-01 M001273030 FLEXIMA VSCL REG 12FR/35CM
M001273040 FLEXIMA VSCL REG 14FR/35CM
M001273090 FLEXIMA VSDC REG LOOP 12/35KIT
M001273100 FLEXIMA VSDC REG LOOP 14/35KIT
M001273150 FLEXIMA VSCL FIRM 12FR/35CM
M001273160 FLEXIMA VSCL FIRM 14FR/35CM
M001273210 FLEXIMA VSDC FIRM LOOP 12/35
M001273220 FLEXIMA VSDC FIRM LOOP 14/35

Single Use Only, Sterilized with Ethylene Oxide Gas, Boston Scintific Corporation, One Boscton Scintific Place, Natick, MA 01760-1537, Made in USA 780 Brookside Drive, Spencer, IN 47460 1080.

Designed for percutaneous drainage of abscess fluid, the Biliary system and the urinary tract.


Code Information All codes
Recalling Firm/
Manufacturer
Boston Scientific Corporation
1 Scimed Pl
Maple Grove MN 55311-1565
For Additional Information Contact
800-272-3737
Manufacturer Reason
for Recall
Boston Scientific Corporation is initiating an urgent Field Correction related to its Flexima¿ and Percuflex¿ Drainage Catheter Family (Pigtail Style). During the attempted removal of the device, sutures were separated from the catheter and remained in the patient.
FDA Determined
Cause 2
Labeling False and Misleading
Action Consignees were sent a Boston Scientific " Urgent Medical Device Field Correction Notification" letter dated March 18, 2009. The letter was addressed to "The Hospital Field Action Contact". The letter described the problem, product affected and the revised catheter removal instructions. They also requested the return of the Acknowledgement Form.
Quantity in Commerce 586,276 US; 276,013 OUS total for all products
Distribution Worldwide Distribution -- USA including DC, Puerto Rico and Guam and countries of United Arab Emirates Austria, Bosnia and Herzegovina, Bahrain, Switzerland, Germany, Egypt, Spain, Finland, France, Great Britain, Ireland, Italy, Kuwait, Netherland, Pakistan, Portugal, Saudi Arabia, Sweden, Slovak Republic, Tunisia, Turkey, South Africa, Netherlands, Costa Rica, Japan, Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Ecuador, El Salvador, Guatemala, Honduras, Hong Kong, Macau, Malaysia, Mexico New Zealand, Panama, Peru, Philippines, Puerto Rico, Singapore, South Korea, Sri Lanka, Taiwan, Thailand, Uruguay, and Venezuela.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = FFA and Original Applicant = BOSTON SCIENTIFIC CORP.
-
-