Date Initiated by Firm |
May 14, 2009 |
Date Posted |
September 04, 2009 |
Recall Status1 |
Terminated 3 on January 20, 2010 |
Recall Number |
Z-1860-2009 |
Recall Event ID |
52163 |
510(K)Number |
K041585 K060258
|
Product Classification |
Infusion pump for pain medication. - Product Code MEB
|
Product |
Stryker 550 mL AutoFuser PainPump, single site continuous with bolus, 5 mL/hr, sterile, REF 0535-500-000. |
Code Information |
Lots 08148757805 through 09112924884. |
Recalling Firm/ Manufacturer |
Stryker Instruments Div. of Stryker Corporation 4100 E. Milham Ave. Kalamazoo MI 49001
|
For Additional Information Contact |
800-800-4236 Ext. 3808
|
Manufacturer Reason for Recall |
Failure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
|
FDA Determined Cause 2 |
Process control |
Action |
Stryker sent all consignees a Medical Device Recall Notification letter dated 5/14/09 and instructed to locate and destroy the recalled products and to complete and return the business reply card. |
Quantity in Commerce |
28,755 of all pumps |
Distribution |
Nationwide and Canada. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MEB and Original Applicant = ACE MEDICAL US, LLC 510(K)s with Product Code = MEB and Original Applicant = ALGOS, LC
|