| | Class 1 Recall Stabliet Infant Warming System |  |
| Date Posted |
August 25, 2009 |
| Recall Number |
Z-1955-2009 |
| Product |
Stabilet Infant Warming System, model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200. |
| Code Information |
All units of model numbers 200, 300, 1250, 1500, 200/3000, 2000, 2200/3200, 3000, and 3200.
|
Recalling Firm/ Manufacturer |
Draeger Medical Systems, Inc. 3135 Quarry Rd Telford, Pennsylvania 18969-1042 |
| For Additional Information Contact |
Micheal Kelhart 800-543-5047
|
Reason for Recall |
Hot particles fell into the infant bassinet beneath a Radiant Infant Warmer. The resulting fire injured an infant. These devices are obsolete.
|
| Action |
On 7/20/09, Draeger Medical, the recalling firm, issued a press release to medical media outreach organizations and an Urgent Medical Device letter to all known customers. This information explained the incident and the need to remove and replace the units in question as soon as feasibly possible. |
| Quantity in Commerce |
6489 units |
| Distribution |
Worldwide distribution: USA, Brazil, Canada, China, Dominican Republic, France, Korea, Lebanon, Mexico, Nigeria, Panama, Poland, Saudi Arabia, Taiwan, Turkey, United Kingdom, Uruguay, and Venezuela. |
| |
|
|
|
|