| Class 2 Device Recall Impulse Angiographic (Diagnostic) Catheter | |
Date Initiated by Firm | August 05, 2009 |
Date Posted | September 04, 2009 |
Recall Status1 |
Terminated 3 on January 07, 2012 |
Recall Number | Z-2014-2009 |
Recall Event ID |
52826 |
510(K)Number | K964859 |
Product Classification |
Catheter, intravascular, diagnostic - Product Code DQO
|
Product | Boston Scientific Impulse Flextrusion Shaft, Angiographic Catheter, Sterilized with ethylene oxide gas, Made in Mexico, Parque Industrial La Mesa, Edificion G1, Suites B. H, Tijuana, Baja California 22650 MEX.
Boston Scientific Guide Catheters are intended for use in general intravascular and coronary applications. They provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in cerebral vasculature. |
Code Information |
Material Description, Batch; MODEL D 5F FR3.5 (5PK) 20155285; MODEL D 5F FR3.5 (5PK) 20156841; MODEL D 5F FR3.5 (5PK) 20160802; MODEL D 5F FR3.5 (5PK) 20162506; MODEL D 5F FR3.5 (5PK) 20164387; MODEL D 5F FR3.5 (5PK) 20165820; MODEL D 5F FR3.5 (5PK) 20167839; MODEL D 5F FR3.5 (5PK) 20170624; MODEL D 5F FR3.5 (5PK) 20178183; MODEL D 5F FR3.5 (5PK) 20179547; MODEL D 5F FR4 SINGLE 20155432; MODEL D 5F FR4 SINGLE 20155433; MODEL D 5F FR4 SINGLE 20156918; MODEL D 5F FR4 SINGLE 20156919; MODEL D 5F FR4 SINGLE 20160875; MODEL D 5F FR4 SINGLE 20160876; MODEL D 5F FR4 SINGLE 20162607; MODEL D 5F FR4 SINGLE 20162608; MODEL D 5F FR4 SINGLE 20165909; MODEL D 5F FR4 SINGLE 20165910; MODEL D 5F FR4 SINGLE 20165911; MODEL D 5F FR4 SINGLE 20165912; MODEL D 5F FR4 SINGLE 20167951; MODEL D 5F FR4 SINGLE 20167952; MODEL D 5F FR4 SINGLE 20170702; MODEL D 5F FR4 SINGLE 20178293; MODEL D 5F FR4 (5PK) 20155287; MODEL D 5F FR4 (5PK) 20155288; MODEL D 5F FR4 (5PK) 20155289; MODEL D 5F FR4 (5PK) 20156842; MODEL D 5F FR4 (5PK) 20156843; MODEL D 5F FR4 (5PK) 20156844; MODEL D 5F FR4 (5PK) 20156845; MODEL D 5F FR4 (5PK) 20156846; MODEL D 5F FR4 (5PK) 20156847; MODEL D 5F FR4 (5PK) 20158158; MODEL D 5F FR4 (5PK) 20158159; MODEL D 5F FR4 (5PK) 20158160; MODEL D 5F FR4 (5PK) 20158161; MODEL D 5F FR4 (5PK) 20160805; MODEL D 5F FR4 (5PK) 20162507; MODEL D 5F FR4 (5PK) 20162508; MODEL D 5F FR4 (5PK) 20162509; MODEL D 5F FR4 (5PK) 20162510; MODEL D 5F FR4 (5PK) 20162511; MODEL D 5F FR4 (5PK) 20162512; MODEL D 5F FR4 (5PK) 20164388; MODEL D 5F FR4 (5PK) 20164389; MODEL D 5F FR4 (5PK) 20164390; MODEL D 5F FR4 (5PK) 20164391; MODEL D 5F FR4 (5PK) 20164392; MODEL D 5F FR4 (5PK) 20165821; MODEL D 5F FR4 (5PK) 20165822; MODEL D 5F FR4 (5PK) 20167840; MODEL D 5F FR4 (5PK) 20167841; MODEL D 5F FR4 (5PK) 20167842; MODEL D 5F FR4 (5PK) 20167843; MODEL D 5F FR4 (5PK) 20167844; MODEL D 5F FR4 (5PK) 20168908; MODEL D 5F FR4 (5PK) 20168909; MODEL D 5F FR4 (5PK) 20168910; MODEL D 5F FR4 (5PK) 20168911; MODEL D 5F FR4 (5PK) 20170626; MODEL D 5F FR4 (5PK) 20170627; MODEL D 5F FR4 (5PK) 20170628; MODEL D 5F FR4 (5PK) 20172941; MODEL D 5F FR4 (5PK) 20172942; MODEL D 5F FR4 (5PK) 20172943; MODEL D 5F FR4 (5PK) 20172944; MODEL D 5F FR4 (5PK) 20172945; MODEL D 5F FR4 (5PK) 20172946; MODEL D 5F FR4 (5PK) 20172947; MODEL D 5F FR4 (5PK) 20172948; MODEL D 5F FR4 (5PK) 20174418; MODEL D 5F FR4 (5PK) 20174419; MODEL D 5F FR4 (5PK) 20174420; MODEL D 5F FR4 (5PK) 20174421; MODEL D 5F FR4 (5PK) 20174422; MODEL D 5F FR4 (5PK) 20178184; MODEL D 5F FR4 (5PK) 20178185; MODEL D 5F FR4 (5PK) 20178186; MODEL D 5F FR4 (5PK) 20178187; MODEL D 5F FR4 (5PK) 20178188; MODEL D 5F FR4 (5PK) 20179548; MODEL D 5F FR4 (5PK) 20179549; MODEL D 5F FR4 (5PK) 20179550; MODEL D 5F FR4 (5PK) 20179551; MODEL D 5F FR4 (5PK) 20181321; MODEL D 5F FR4 (5PK) 20181322; MODEL D 5F FR4 (5PK) 20181323; MODEL D 5F FR5 SINGLE 20158232; MODEL D 5F FR5 SINGLE 20170704; MODEL D 5F FR5 SINGLE 20173024; MODEL D 5F FR5(5PK) 20155290; MODEL D 5F FR5(5PK) 20156848; MODEL D 5F FR5(5PK) 20162513; MODEL D 5F FR5(5PK) 20164393; MODEL D 5F FR5(5PK) 20165823; MODEL D 5F FR5(5PK) 20167845; MODEL D 5F FR5(5PK) 20170630; MODEL D 5F FR5(5PK) 20172949; MODEL D 5F FR5(5PK) 20178189; MODEL D 5F FR5(5PK) 20179552; MODEL D 5F FR6 (5PK) 20162514; MODEL D 5F FR4 MODIFIED (5PK) 20156232; MODEL D 5F FR4 MODIFIED (5PK) 20160806; MODEL D 5F FR4 MODIFIED (5PK) 20162515; MODEL D 5F AL3 SINGLE 20170705; MODEL D 5F AL3 (5PK) 20156849; MODEL D 5F AL3 (5PK) 20164394; MODEL D 5F MPA-1 SINGLE 20160877; MODEL D 5F MPA-1 SINGLE 20165913; MODEL D 5F MPA-1 SINGLE 20167953; MODEL D 5F MPA-1 (5PK) 20154006; MODEL D 5F MPA-1 (5PK) 20155291; MODEL D 5F MPA-1 (5PK) 20156850; MODEL D 5F MPA-1 (5PK) 20160807; MODEL D 5F MPA-1 (5PK) 20162516; MODEL D 5F MPA-1 (5PK) 20164395; MODEL D 5F MPA-1 (5PK) 20165824; MODEL D 5F MPA-1 (5PK) 20167846; MODEL D 5F MPA-1 (5PK) 20170631; MODEL D 5F MPA-1 (5PK) 20178190; MODEL D 5F MPA-1 (5PK) 20179554; MODEL D 5F MPA2 W/SH (5PK) 20154007; MODEL D 5F MPA2 W/SH (5PK) 20156851; MODEL D 5F MPA2 W/SH (5PK) 20158164; MODEL D 5F MPA2 W/SH (5PK) 20162517; MODEL D 5F MPA2 W/SH (5PK) 20164396; MODEL D 5F MPA2 W/SH (5PK) 20165825; MODEL D 5F MPA2 W/SH (5PK) 20170632; MODEL D 5F MPA2 W/SH (5PK) 20174425; MODEL D 5F MPA2 W/SH (5PK) 20178191; MODEL D 5F MPA2 W/SH (5PK) 20179555; MODEL D 5F MPB-1 (5PK) 20160809; MODEL D 5F MPB-2 W/SH (5PK) 20154009; MODEL D 5F MPB-2 W/SH (5PK) 20164401; MODEL D 5F MPB-2 W/SH (5PK) 20178192; MODEL D 5F SON1 100CM SH (5PK) 20155301; MODEL D 5F SON2 80CM SH (5PK) 20178193; MODEL D 5F SON2 80CM SH (5PK) 20179557; MODEL D 5F SON1 100C (5PK) 20156852; MODEL D 5F SON1 100C (5PK) 20168917; MODEL D 5F CAS1 (5PK) 20155302; MODEL D 5F CAS1 (5PK) 20158590; MODEL D 5F CAS1 (5PK) 20165826; MODEL D 5F CAS1 (5PK) 20174427; MODEL D 5F CAS2 (5PK) 20165827; MODEL D 5F RADIAL 5-PACK 20164402; MODEL D 5F RADIAL 5-PACK 20172950; MODEL - 5F IMPULSE A3.5 5-PK 20154931; MODEL - 5F IMPULSE A3.5 5-PK 20154932; MODEL - 5F IMPULSE A3.5 5-PK 20154933; MODEL - 5F IMPULSE A3.5 5-PK 20154934; MODEL - 5F IMPULSE A3.5 5-PK 20154946; MODEL - 5F IMPULSE A3.5 5-PK 20154954; MODEL - 5F IMPULSE A3.5 5-PK 20154956; MODEL - 5F IMPULSE A3.5 5-PK 20160654; MODEL - 5F IMPULSE A3.5 5-PK 20160655; MODEL - 5F IMPULSE A3.5 5-PK 20166587; MODEL - 5F IMPULSE A3.5 5-PK 20166588; MODEL - IMPULSE 5F SELECTIVE KIMNY 5 PK 20160810; MODEL - IMPULSE 5F SELECTIVE KIMNY 5 PK 20178194; MODEL D 5F RCB 20160878; MODEL D 5F RCB 5-PACK 20156853; MODEL D 5F RCB 5-PACK 20160811; MODEL D 5F RCB 5-PACK 20162518; MODEL D 5F RCB 5-PACK 20165828; MODEL D 5F RCB 5-PACK 20170633; MODEL D 5F RCB 5-PACK 20174429; MODEL D 5F LCB SINGLE 20173026; MODEL D 5F LCB 5-PACK 20154011; MODEL D 5F LCB 5-PACK 20158165; MODEL D 5F LCB 5-PACK 20162519; MODEL D 5F LCB 5-PACK 20164403; MODEL D 5F LCB 5-PACK 20165829; MODEL D 5F LCB 5-PACK 20170634; MODEL D 5F LCB 5-PACK 20178196; MODEL D 5F IM 20158233; MODEL D 5F IM 5-PACK 20156854; MODEL D 5F IM 5-PACK 20160814; MODEL D 5F IM 5-PACK 20162520; MODEL D 5F IM 5-PACK 20162521; MODEL D 5F IM 5-PACK 20165830; MODEL D 5F IM 5-PACK 20167849; MODEL D 5F IM 5-PACK 20170635; MODEL D 5F IM 5-PACK 20172951; MODEL D 5F IM 5-PACK 20178197; MODEL D 5F IM 5-PACK 20179559; MODEL D 5F IM 5-PACK 20179560; MODEL D 5F IM 5-PACK 20181334; MODEL - D 5F IMPULSE IM T (SGL) 20167955; MODEL - D 5F IMPULSE IM T (SGL) 20178296; MODEL - D 5F IMPULSE IM T (5 PK) 20160815; MODEL - D 5F IMPULSE IM T (5 PK) 20162522; MODEL - D 5F IMPULSE IM T (5 PK) 20165831; MODEL - D 5F IMPULSE IM T (5 PK) 20167850; MODEL - D 5F IMPULSE IM T (5 PK) 20170637; MODEL - D 5F IMPULSE IM T (5 PK) 20174431; MODEL - D 5F IMPULSE IM T (5 PK) 20179561; MODEL - 5F IMPULSE IMC 5 PACK 20164406; MODEL - 5F IMPULSE IMC 5 PACK 20164407; MODEL - 5F IMPULSE IMC 5 PACK 20164408; MODEL D 5F FL3.5 SINGLE 20155436; MODEL D 5F FL3.5 SINGLE 20155437; MODEL D 5F FL3.5 SINGLE 20155438; MODEL D 5F FL3.5 SINGLE 20160880; MODEL D 5F FL3.5 SINGLE 20165914; MODEL D 5F FL3.5 SINGLE 20165915; MODEL D 5F FL3.5 SINGLE 20165916; MODEL D 5F FL3.5 SINGLE 20165917; MODEL D 5F FL3.5 SINGLE 20170707; MODEL D 5F FL3.5 SINGLE 20178297; MODEL D 5F FL3.5 SINGLE 20179669; MODEL D 5F FL3.5 SINGLE 20179670; MODEL D 5F FL3.5 B/5 20155304; MODEL D 5F FL3.5 B/5 20156855; MODEL D 5F FL3.5 B/5 20156856; MODEL D 5F FL3.5 B/5 20158167; MODEL D 5F FL3.5 B/5 20160817; MODEL D 5F FL3.5 B/5 20164409; MODEL D 5F FL3.5 B/5 20165832; MODEL D 5F FL3.5 B/5 20165833; MODEL D 5F FL3.5 B/5 20167851; MODEL D 5F FL3.5 B/5 20167852; MODEL D 5F FL3.5 B/5 20168920; MODEL D 5F FL3.5 B/5 20170638; MODEL D 5F FL3.5 B/5 20172952; MODEL D 5F FL3.5 B/5 20172953; MODEL D 5F FL3.5 B/5 20174432; MODEL D 5F FL3.5 B/5 20174433; MODEL D 5F FL3.5 B/5 20178198; MODEL D 5F FL3.5 B/5 20179562; MODEL D 5F FL3.5 B/5 20179563; MODEL D 5F FL3.5 B/5 20181340; MODEL D 5F FL4 SINGLE 20156920; MODEL D 5F FL4 SINGLE 20160881; MODEL D 5F FL4 SINGLE 20160882; MODEL D 5F FL4 SINGLE 20164512; MODEL D 5F FL4 SINGLE 20165919; MODEL D 5F FL4 SINGLE 20165920; MODEL D 5F FL4 SINGLE 20165921; MODEL D 5F FL4 SINGLE 20165922; MODEL D 5F FL4 SINGLE 20170708; MODEL D 5F FL4 SINGLE 20178298; MODEL D 5F FL4 SINGLE 20179673; MODEL D 5F FL4 (5PK) 20155308; MODEL D 5F FL4 (5PK) 20155309; MODEL D 5F FL4 (5PK) 20156857; MODEL D 5F FL4 (5PK) 20156858; MODEL D 5F FL4 (5PK) 20156859; MODEL D 5F FL4 (5PK) 20156860; MODEL D 5F FL4 (5PK) 20156861; MODEL D 5F FL4 (5PK) 20156862; MODEL D 5F FL4 (5PK) 20158170; MODEL D 5F FL4 (5PK) 20160820; MODEL D 5F FL4 (5PK) 20162524; MODEL D 5F FL4 (5PK) 20162525; MODEL D 5F FL4 (5PK) 20162526; MODEL D 5F FL4 (5PK) 20162527; MODEL D 5F FL4 (5PK) 20164410; MODEL D 5F FL4 (5PK) 20164411; MODEL D 5F FL4 (5PK) 20164412; MODEL D 5F FL4 (5PK) 20164413; MODEL D 5F FL4 (5PK) 20164414; MODEL D 5F FL4 (5PK) 20165834; MODEL D 5F FL4 (5PK) 20165835; MODEL D 5F FL4 (5PK) 20165836; MODEL D 5F FL4 (5PK) 20165837; MODEL D 5F FL4 (5PK) 20167854; MODEL D 5F FL4 (5PK) 20167855; MODEL D 5F FL4 (5PK) 20167856; MODEL D 5F FL4 (5PK) 20167857; MODEL D 5F FL4 (5PK) 20167858; MODEL D 5F FL4 (5PK) 20168922; MODEL D 5F FL4 (5PK) 20168923; MODEL D 5F FL4 (5PK) 20168924; MODEL D 5F FL4 (5PK) 20168925; MODEL D 5F FL4 (5PK) 20168928; MODEL D 5F FL4 (5PK) 20170640; MODEL D 5F FL4 (5PK) 20170641; MODEL D 5F FL4 (5PK) 20170642; MODEL D 5F FL4 (5PK) 20170643; MODEL D 5F FL4 (5PK) 20172954; MODEL D 5F FL4 (5PK) 20172955; MODEL D 5F FL4 (5PK) 20172956; MODEL D 5F FL4 (5PK) 20172957; MODEL D 5F FL4 (5PK) 20172958; MODEL D 5F FL4 (5PK) 20174434; MODEL D 5F FL4 (5PK) 20174435; MODEL D 5F FL4 (5PK) 20174436; MODEL D 5F FL4 (5PK) 20174437; MODEL D 5F FL4 (5PK) 20178200; MODEL D 5F FL4 (5PK) 20178201; MODEL D 5F FL4 (5PK) 20178202; MODEL D 5F FL4 (5PK) 20178203; MODEL D 5F FL4 (5PK) 20179565; MODEL D 5F FL4 (5PK) 20179566; MODEL D 5F FL4 (5PK) 20179567; MODEL D 5F FL4 (5PK) 20179568; MODEL D 5F FL4 (5PK) 20181343; MODEL D 5F FL4 (5PK) 20181344; MODEL D 5F FL4.5 (5PK) 20155311; MODEL D 5F FL4.5 (5PK) 20160821; MODEL D 5F FL4.5 (5PK) 20164416; MODEL D 5F FL4.5 (5PK) 20165838; MODEL D 5F FL4.5 (5PK) 20168930; MODEL D 5F FL4.5 (5PK) 20170645; MODEL D 5F FL4.5 (5PK) 20178207; MODEL D 5F FL4.5 (5PK) 20179569; MODEL D 5F FL5 SINGLE 20162609; MODEL D 5F FL5 SINGLE 20164513; MODEL D 5F FL5 (5PK) 20156863; MODEL D 5F FL5 (5PK) 20158176; MODEL D 5F FL5 (5PK) 20160822; MODEL D 5F FL5 (5PK) 20164417; MODEL D 5F FL5 (5PK) 20165839; MODEL D 5F FL5 (5PK) 20167863; MODEL D 5F FL5 (5PK) 20168931; MODEL D 5F FL5 (5PK) 20170646; MODEL D 5F FL5 (5PK) 20172959; MODEL D 5F FL5 (5PK) 20174442; MODEL D 5F FL5 (5PK) 20178208; MODEL D 5F FL6 SINGLE 20162610; MODEL D 5F FL6 SINGLE 20170710; MODEL D 5F FL6 (5PK) 20160823; MODEL D 5F FL6 (5PK) 20165840; MODEL D 5F FL6 (5PK) 20170648; MODEL D 5F FL6 (5PK) 20179571; MODEL-D 5F PIG 110CM 20169004; MODEL-D 5F PIG 110CM 20173029; MODEL-D 5F PIG 110CM 20174527; MODEL-D 5F PIG 110CM 20179676; MODEL-D 5F PIG 110CM (5PK) 20158180; MODEL-D 5F PIG 110CM (5PK) 20158181; MODEL-D 5F PIG 110CM (5PK) 20160827; MODEL-D 5F PIG 110CM (5PK) 20162533; MODEL-D 5F PIG 110CM (5PK) 20164421; MODEL-D 5F PIG 110CM (5PK) 20167870; MODEL-D 5F PIG 110CM (5PK) 20168937; MODEL-D 5F PIG 110CM (5PK) 20170654; MODEL-D 5F PIG 110CM (5PK) 20170655; MODEL-D 5F PIG 110CM (5PK) 20172965; MODEL-D 5F PIG 110CM (5PK) 20172966; MODEL-D 5F PIG 110CM (5PK) 20174450; MODEL-D 5F PIG 110CM (5PK) 20174451; MODEL-D 5F PIG 110CM (5PK) 20178214; MODEL-D 5F PIG 110CM (5PK) 20178215; MODEL-D 5F PIG 110CM (5PK) 20179578; MODEL-D 5F PIG 110CM (5PK) 20181357; MODEL-D 5F PIG 110CM (5PK) 20183236; MODEL-D 5F PIG 110CM (5PK) 20183237; MODEL-D 5F PIG 145 110CM 20155439; MODEL-D 5F PIG 145 110CM 20160884; MODEL-D 5F PIG 145 110CM 20164514; MODEL-D 5F PIG 145 110CM 20169005; MODEL-D 5F PIG 145 110CM 20173030; MODEL-D 5F PIG 145 110CM 20178302; MODEL-D 5F PIG 145 110CM (5PK) 20158182; MODEL-D 5F PIG 145 110CM (5PK) 20160828; MODEL-D 5F PIG 145 110CM (5PK) 20162534; MODEL-D 5F PIG 145 110CM (5PK) 20162535; MODEL-D 5F PIG 145 110CM (5PK) 20164422; MODEL-D 5F PIG 145 110CM (5PK) 20164423; MODEL-D 5F PIG 145 110CM (5PK) 20168938; MODEL-D 5F PIG 145 110CM (5PK) 20168939; MODEL-D 5F PIG 145 110CM (5PK) 20170656; MODEL-D 5F PIG 145 110CM (5PK) 20170657; MODEL-D 5F PIG 145 110CM (5PK) 20172967; MODEL-D 5F PIG 145 110CM (5PK) 20172968; MODEL-D 5F PIG 145 110CM (5PK) 20174452; MODEL-D 5F PIG 145 110CM (5PK) 20178217; MODEL-D 5F PIG 145 110CM (5PK) 20179579; MODEL-D 5F PIG 145 110CM (5PK) 20179580; MODEL-D 5F PIG 145 110CM (5PK) 20181358; MODEL-D 5F PIG 145 110CM (5PK) 20181359; MODEL D 5F PIG 155 110CM (5PK) 20160829; MODEL D 5F PIG 155 110CM (5PK) 20164424; MODEL D 5F PIG 155 110CM (5PK) 20170658; MODEL D 5F PIG 155 110CM (5PK) 20172969; MODEL D 5F PIG 155 110CM (5PK) 20174453; MODEL D 5F PIG 155 110CM (5PK) 20179581; MODEL D 5F SL PIG 145 110CM 5-PACK 20160830; MODEL D 5F SL PIG 145 110CM 5-PACK 20181361; MODEL D 5F AR1 SINGLE 20164515; MODEL D 5F AR1 SINGLE 20179678; MODEL D 5F AR1 (5PK) 20155321; MODEL D 5F AR1 (5PK) 20160832; MODEL D 5F AR1 (5PK) 20164426; MODEL D 5F AR1 (5PK) 20165847; MODEL D 5F AR1 (5PK) 20167874; MODEL D 5F AR1 (5PK) 20174455; MODEL D 5F AR1 (5PK) 20178221; MODEL D 5F AR1 (5PK) 20179582; MODEL D 5F AR2 SINGLE 20156922; MODEL D 5F AR2 SINGLE 20173031; MODEL D 5F AR2 (5PK) 20160833; MODEL D 5F AR2 (5PK) 20162536; MODEL D 5F AR2 (5PK) 20165848; MODEL D 5F AR2 (5PK) 20170659; MODEL D 5F AR2 (5PK) 20178222; MODEL D 5F AR2 (5PK) 20179583; MODEL D 5F AR MODIFIED 20156923; MODEL D 5F AR MODIFIED 20164516; MODEL D 5F AR MODIFIED 20167960; MODEL D 5F AR MODIFIED 20167961; MODEL D 5F AR MODIFIED 5-PACK 20152878; MODEL D 5F AR MODIFIED 5-PACK 20155322; MODEL D 5F AR MODIFIED 5-PACK 20155323; MODEL D 5F AR MODIFIED 5-PACK 20155324; MODEL D 5F AR MODIFIED 5-PACK 20155325; MODEL D 5F AR MODIFIED 5-PACK 20155326; MODEL D 5F AR MODIFIED 5-PACK 20155327; MODEL D 5F AR MODIFIED 5-PACK 20155328; MODEL D 5F AR MODIFIED 5-PACK 20155329; MODEL D 5F AR MODIFIED 5-PACK 20155330; MODEL D 5F AR MODIFIED 5-PACK 20155331; MODEL D 5F AR MODIFIED 5-PACK 20162537; MODEL D 5F AR MODIFIED 5-PACK 20162538; MODEL D 5F AR MODIFIED 5-PACK 20162539; MODEL D 5F AR MODIFIED 5-PACK 20162540; MODEL D 5F AR MODIFIED 5-PACK 20162541; MODEL D 5F AR MODIFIED 5-PACK 20162542; MODEL D 5F AR MODIFIED 5-PACK 20162543; MODEL D 5F AR MODIFIED 5-PACK 20162544; MODEL D 5F AR MODIFIED 5-PACK 20162545; MODEL D 5F AR MODIFIED 5-PACK 20162546; MODEL D 5F AR MODIFIED 5-PACK 20164427; MODEL D 5F AR MODIFIED 5-PACK 20164428; MODEL D 5F AR MODIFIED 5-PACK 20164429; MODEL D 5F AR MODIFIED 5-PACK 20164430; MODEL D 5F AR MODIFIED 5-PACK 20164431; MODEL D 5F AR MODIFIED 5-PACK 20164432; MODEL D 5F AR MODIFIED 5-PACK 20164433; MODEL D 5F AR MODIFIED 5-PACK 20164434; MODEL D 5F AR MODIFIED 5-PACK 20164435; MODEL D 5F AR MODIFIED 5-PACK 20164436; MODEL D 5F AR MODIFIED 5-PACK 20165849; MODEL D 5F AR MODIFIED 5-PACK 20165850; MODEL D 5F AR MODIFIED 5-PACK 20165851; MODEL D 5F AR MODIFIED 5-PACK 20165852; MODEL D 5F AR MODIFIED 5-PACK 20165853; MODEL D 5F AR MODIFIED 5-PACK 20167875; MODEL D 5F AR MODIFIED 5-PACK 20167876; MODEL D 5F AR MODIFIED 5-PACK 20167877; MODEL D 5F AR MODIFIED 5-PACK 20167878; MODEL D 5F AR MODIFIED 5-PACK 20167879; MODEL D 5F AR MODIFIED 5-PACK 20172970; MODEL D 5F AR MODIFIED 5-PACK 20172971; MODEL D 5F AR MODIFIED 5-PACK 20172972; MODEL D 5F AR MODIFIED 5-PACK 20172973; MODEL D 5F AR MODIFIED 5-PACK 20172974; MODEL D 5F AR MODIFIED 5-PACK 20174458; MODEL D 5F AR MODIFIED 5-PACK 20174459; MODEL D 5F AR MODIFIED 5-PACK 20178223; MODEL D 5F AR MODIFIED 5-PACK 20179584; MODEL D 5F AR MODIFIED 5-PACK 20179585; MODEL D 5F AR MODIFIED 5-PACK 20179586; MODEL D 5F AR MODIFIED 5-PACK 20179587; MODEL D 5F AR MODIFIED 5-PACK 20179588; MODEL D 5F AL1 SINGLE 20162613; MODEL D 5F AL1 SINGLE 20169006; MODEL D 5F AL1 SINGLE 20173032; MODEL D 5F AL1 5-PACK 20155332; MODEL D 5F AL1 5-PACK 20162547; MODEL D 5F AL1 5-PACK 20164437; MODEL D 5F AL1 5-PACK 20165854; MODEL D 5F AL1 5-PACK 20167880; MODEL D 5F AL1 5-PACK 20168942; MODEL D 5F AL1 5-PACK 20170660; MODEL D 5F AL1 5-PACK 20172975; MODEL D 5F AL1 5-PACK 20174461; MODEL D 5F AL1 5-PACK 20178233; MODEL D 5F AL1 5-PACK 20179589; MODEL D 5F AL1 5-PACK 20181364; MODEL D 5F AL2 20162614; MODEL D 5F AL2 20167962; MODEL D 5F AL2 20169007; MODEL D 5F AL2 20178303; MODEL D 5F AL2 5-PACK 20155333; MODEL D 5F AL2 5-PACK 20162548; MODEL D 5F AL2 5-PACK 20164438; MODEL D 5F AL2 5-PACK 20165855; MODEL D 5F AL2 5-PACK 20167881; MODEL D 5F AL2 5-PACK 20168943; MODEL D 5F AL2 5-PACK 20170661; MODEL D 5F AL2 5-PACK 20178234; MODEL D 5F AL2 5-PACK 20179590; MODEL-D 6F IMPULSE FR3.5 (SGL) 20179679; MODEL - D 6F IMPULSE FR3.5 5 PACK 20155334; MODEL - D 6F IMPULSE FR3.5 5 PACK 20156869; MODEL - D 6F IMPULSE FR3.5 5 PACK 20160835; MODEL - D 6F IMPULSE FR3.5 5 PACK 20162549; MODEL - D 6F IMPULSE FR3.5 5 PACK 20165856; MODEL - D 6F IMPULSE FR3.5 5 PACK 20167882; MODEL - D 6F IMPULSE FR3.5 5 PACK 20170662; MODEL - D 6F IMPULSE FR3.5 5 PACK 20174463; MODEL - D 6F IMPULSE FR3.5 5 PACK 20178235; MODEL - D 6F IMPULSE FR3.5 5 PACK 20179591; MODEL - D 6F IMPULSE FR3.5 5 PACK 20183245; MODEL - D 6F IMPULSE FR4 SINGLE 20158237; MODEL - D 6F IMPULSE FR4 SINGLE 20160885; MODEL - D 6F IMPULSE FR4 5 PACK 20154041; MODEL - D 6F IMPULSE FR4 5 PACK 20155336; MODEL - D 6F IMPULSE FR4 5 PACK 20155337; MODEL - D 6F IMPULSE FR4 5 PACK 20155338; MODEL - D 6F IMPULSE FR4 5 PACK 20155339; MODEL - D 6F IMPULSE FR4 5 PACK 20155340; MODEL - D 6F IMPULSE FR4 5 PACK 20155341; MODEL - D 6F IMPULSE FR4 5 PACK 20156870; MODEL - D 6F IMPULSE FR4 5 PACK 20156871; MODEL - D 6F IMPULSE FR4 5 PACK 20156872; MODEL - D 6F IMPULSE FR4 5 PACK 20156873; MODEL - D 6F IMPULSE FR4 5 PACK 20156874; MODEL - D 6F IMPULSE FR4 5 PACK 20156875; MODEL - D 6F IMPULSE FR4 5 PACK 20156878; MODEL - D 6F IMPULSE FR4 5 PACK 20158185; MODEL - D 6F IMPULSE FR4 5 PACK 20158186; MODEL - D 6F IMPULSE FR4 5 PACK 20158187; MODEL - D 6F IMPULSE FR4 5 PACK 20160837; MODEL - D 6F IMPULSE FR4 5 PACK 20162550; MODEL - D 6F IMPULSE FR4 5 PACK 20162551; MODEL - D 6F IMPULSE FR4 5 PACK 20162552; MODEL - D 6F IMPULSE FR4 5 PACK 20162553; MODEL - D 6F IMPULSE FR4 5 PACK 20162554; MODEL - D 6F IMPULSE FR4 5 PACK 20162555; MODEL - D 6F IMPULSE FR4 5 PACK 20164439; MODEL - D 6F IMPULSE FR4 5 PACK 20164440; MODEL - D 6F IMPULSE FR4 5 PACK 20164441; MODEL - D 6F IMPULSE FR4 5 PACK 20164442; MODEL - D 6F IMPULSE FR4 5 PACK 20164443; MODEL - D 6F IMPULSE FR4 5 PACK 20164444; MODEL - D 6F IMPULSE FR4 5 PACK 20165857; MODEL - D 6F IMPULSE FR4 5 PACK 20165858; MODEL - D 6F IMPULSE FR4 5 PACK 20165859; MODEL - D 6F IMPULSE FR4 5 PACK 20165860; MODEL - D 6F IMPULSE FR4 5 PACK 20165861; MODEL - D 6F IMPULSE FR4 5 PACK 20167883; MODEL - D 6F IMPULSE FR4 5 PACK 20167884; MODEL - D 6F IMPULSE FR4 5 PACK 20167885; MODEL - D 6F IMPULSE FR4 5 PACK 20167886; MODEL - D 6F IMPULSE FR4 5 PACK 20167887; MODEL - D 6F IMPULSE FR4 5 PACK 20167888; MODEL - D 6F IMPULSE FR4 5 PACK 20167889; MODEL - D 6F IMPULSE FR4 5 PACK 20168945; MODEL - D 6F IMPULSE FR4 5 PACK 20168946; MODEL - D 6F IMPULSE FR4 5 PACK 20168947; MODEL - D 6F IMPULSE FR4 5 PACK 20168948; MODEL - D 6F IMPULSE FR4 5 PACK 20170664; MODEL - D 6F IMPULSE FR4 5 PACK 20170665; MODEL - D 6F IMPULSE FR4 5 PACK 20170666; MODEL - D 6F IMPULSE FR4 5 PACK 20170667; MODEL - D 6F IMPULSE FR4 5 PACK 20172976; MODEL - D 6F IMPULSE FR4 5 PACK 20172977; MODEL - D 6F IMPULSE FR4 5 PACK 20172978; MODEL - D 6F IMPULSE FR4 5 PACK 20172979; MODEL - D 6F IMPULSE FR4 5 PACK 20172980; MODEL - D 6F IMPULSE FR4 5 PACK 20172981; MODEL - D 6F IMPULSE FR4 5 PACK 20172982; MODEL - D 6F IMPULSE FR4 5 PACK 20172983; MODEL - D 6F IMPULSE FR4 5 PACK 20172984; MODEL - D 6F IMPULSE FR4 5 PACK 20174464; MODEL - D 6F IMPULSE FR4 5 PACK 20174465; MODEL - D 6F IMPULSE FR4 5 PACK 20174466; MODEL - D 6F IMPULSE FR4 5 PACK 20174467; MODEL - D 6F IMPULSE FR4 5 PACK 20174468; MODEL - D 6F IMPULSE FR4 5 PACK 20178237; MODEL - D 6F IMPULSE FR4 5 PACK 20178238; MODEL - D 6F IMPULSE FR4 5 PACK 20179592; MODEL - D 6F IMPULSE FR4 5 PACK 20179593; MODEL - D 6F IMPULSE FR4 5 PACK 20179594; MODEL - D 6F IMPULSE FR4 5 PACK 20181367; MODEL - D 6F IMPULSE FR4 5 PACK 20181368; MODEL - D 6F IMPULSE FR4 5 PACK 20181369; MODEL - D 6F IMPULSE FR4 5 PACK 20181370; MODEL - D 6F IMPULSE FR4 5 PACK 20181371; MODEL - D 6F IMPULSE FR4 5 PACK 20181372; MODEL - D 6F IMPULSE FR4 5 PACK 20183246; MODEL - D 6F IMPULSE FR4 5 PACK 20183250; MODEL - D 6F IMPULSE FR5 5 PACK 20149608; MODEL - D 6F IMPULSE FR5 5 PACK 20155342; MODEL - D 6F IMPULSE FR5 5 PACK 20156879; MODEL - D 6F IMPULSE FR5 5 PACK 20164445; MODEL - D 6F IMPULSE FR5 5 PACK 20165862; MODEL - D 6F IMPULSE FR5 5 PACK 20174469; MODEL - D 6F IMPULSE FR5 5 PACK 20178239; MODEL - D 6F IMPULSE FR5 5 PACK 20179595; MODEL - D 6F IMPULSE FR6 5 PACK 20155343; MODEL - D 6F IMPULSE AL3 5 PACK 20155344; MODEL - D 6F IMPULSE AL3 5 PACK 20164446; MODEL - D 6F IMPULSE AL3 5 PACK 20167890; MODEL - D 6F IMPULSE AL3 5 PACK 20178240; MODEL - D 6F IMPULSE AL3 5 PACK 20179596; MODEL - D 6F IMPULSE MPA-1 SINGLE 20164517; MODEL - D 6F IMPULSE MPA-1 SINGLE 20173033; MODEL - D 6F IMPULSE MPA-1 5 PACK 20155345; MODEL - D 6F IMPULSE MPA-1 5 PACK 20156880; MODEL - D 6F IMPULSE MPA-1 5 PACK 20164447; MODEL - D 6F IMPULSE MPA-1 5 PACK 20165863; MODEL - D 6F IMPULSE MPA-1 5 PACK 20170669; MODEL - D 6F IMPULSE MPA-1 5 PACK 20172985; MODEL - D 6F IMPULSE MPA-1 5 PACK 20174470; MODEL - D 6F IMPULSE MPA-1 5 PACK 20178241; MODEL - D 6F IMPULSE MPA-1 5 PACK 20181373; MODEL - D 6F IMPULSE MPA-2 SINGLE 20162615; MODEL - D 6F IMPULSE MPA-2 SINGLE 20164518; MODEL - D 6F IMPULSE MPA-2 5 PACK 20154042; MODEL - D 6F IMPULSE MPA-2 5 PACK 20155346; MODEL - D 6F IMPULSE MPA-2 5 PACK 20160838; MODEL - D 6F IMPULSE MPA-2 5 PACK 20162556; MODEL - D 6F IMPULSE MPA-2 5 PACK 20164448; MODEL - D 6F IMPULSE MPA-2 5 PACK 20170670; MODEL - D 6F IMPULSE MPA-2 5 PACK 20172986; MODEL - D 6F IMPULSE MPA-2 5 PACK 20174471; MODEL - D 6F IMPULSE MPA-2 5 PACK 20178242; MODEL - D 6F IMPULSE MPB-1 SINGLE 20164519; MODEL - D 6F IMPULSE MPB-1 SINGLE 20173034; MODEL - D 6F IMPULSE MPB-2 SINGLE 20155440; MODEL - D 6F IMPULSE MPB-2 5 PACK 20164449; MODEL - D 6F IMPULSE MPB-2 5 PACK 20178244; MODEL - D 6F IMPULSE SON1 5 PACK 20155347; MODEL - D 6F IMPULSE SON1 5 PACK 20155348; MODEL - D 6F IMPULSE SON1 5 PACK 20155349; MODEL - D 6F IMPULSE SON1 5 PACK 20155351; MODEL - D 6F IMPULSE SON1 5 PACK 20168956; MODEL - D 6F IMPULSE SON2 80CM SINGLE 20174473; MODEL - D 6F IMPULSE CAS1 5 PACK 20158194; MODEL - D 6F IMPULSE CAS1 5 PACK 20179599; MODEL - D 6F IMPULSE CAS2 5 PACK 20160839; MODEL - D 6F IMPULSE CAS2 5 PACK 20168961; MODEL - D 6F IMPULSE CAS2 5 PACK 20179600; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155356; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155358; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155359; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155360; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155361; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155362; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155363; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155364; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155365; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155366; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155367; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155368; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155369; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155371; MODEL - D 6F IMPULSE RADIAL 5 PACK 20155372; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164451; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164452; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164453; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164454; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164455; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164457; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164458; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164459; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164460; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164461; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164462; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164463; MODEL - D 6F IMPULSE RADIAL 5 PACK 20164464; MODEL - IMPULSE 6F SELECTIVE KIMNY 5 PK 20155377; MODEL - IMPULSE 6F SELECTIVE KIMNY 5 PK 20160840; MODEL - IMPULSE 6F SELECTIVE KIMNY 5 PK 20165864; MODEL - IMPULSE 6F SELECTIVE KIMNY 5 PK 20168962; MODEL - IMPULSE 6F SELECTIVE KIMNY 5 PK 20178245; MODEL - D 6F IMPULSE RCB 5 PACK 20155381; MODEL - D 6F IMPULSE RCB 5 PACK 20155382; MODEL - D 6F IMPULSE RCB 5 PACK 20155383; MODEL - D 6F IMPULSE RCB 5 PACK 20155384; MODEL - D 6F IMPULSE RCB 5 PACK 20162558; MODEL - D 6F IMPULSE RCB 5 PACK 20162559; MODEL - D 6F IMPULSE RCB 5 PACK 20162560; MODEL - D 6F IMPULSE RCB 5 PACK 20162561; MODEL - D 6F IMPULSE RCB 5 PACK 20162562; MODEL - D 6F IMPULSE RCB 5 PACK 20162563; MODEL - D 6F IMPULSE RCB 5 PACK 20162564; MODEL - D 6F IMPULSE RCB 5 PACK 20162565; MODEL - D 6F IMPULSE RCB 5 PACK 20162566; MODEL - D 6F IMPULSE RCB 5 PACK 20165865; MODEL - D 6F IMPULSE RCB 5 PACK 20165866; MODEL - D 6F IMPULSE RCB 5 PACK 20165867; MODEL - D 6F IMPULSE RCB 5 PACK 20165868; MODEL - D 6F IMPULSE RCB 5 PACK 20165869; MODEL - D 6F IMPULSE RCB 5 PACK 20165870; MODEL - D 6F IMPULSE RCB 5 PACK 20165871; MODEL - D 6F IMPULSE RCB 5 PACK 20168967; MODEL - D 6F IMPULSE RCB 5 PACK 20168969; MODEL - D 6F IMPULSE RCB 5 PACK 20179603; MODEL - D 6F IMPULSE LCB SINGLE 20155441; MODEL - D 6F IMPULSE LCB SINGLE 20170712; MODEL - D 6F IMPULSE LCB 5 PACK 20155385; MODEL - D 6F IMPULSE LCB 5 PACK 20156881; MODEL - D 6F IMPULSE LCB 5 PACK 20160841; MODEL - D 6F IMPULSE LCB 5 PACK 20162568; MODEL - D 6F IMPULSE LCB 5 PACK 20164465; MODEL - D 6F IMPULSE LCB 5 PACK 20167891; MODEL - D 6F IMPULSE LCB 5 PACK 20172987; MODEL - D 6F IMPULSE LCB 5 PACK 20178246; MODEL - D 6F IMPULSE LCB 5 PACK 20179615; MODEL - D 6F IMPULSE IM SINGLE 20155442; MODEL - D 6F IMPULSE IM SINGLE 20169009; MODEL - D 6F IMPULSE IM SINGLE 20178307; MODEL - D 6F IMPULSE IM SINGLE 20179682; MODEL - D 6F IMPULSE IM 5 PACK 20155386; MODEL - D 6F IMPULSE IM 5 PACK 20156882; MODEL - D 6F IMPULSE IM 5 PACK 20158195; MODEL - D 6F IMPULSE IM 5 PACK 20160842; MODEL - D 6F IMPULSE IM 5 PACK 20162569; MODEL - D 6F IMPULSE IM 5 PACK 20164466; MODEL - D 6F IMPULSE IM 5 PACK 20167892; MODEL - D 6F IMPULSE IM 5 PACK 20168970; MODEL - D 6F IMPULSE IM 5 PACK 20170672; MODEL - D 6F IMPULSE IM 5 PACK 20172988; MODEL - D 6F IMPULSE IM 5 PACK 20174478; MODEL - D 6F IMPULSE IM 5 PACK 20178247; MODEL - D 6F IMPULSE IM 5 PACK 20179616; MODEL - D 6F IMPULSE IM T (5 PK) 20155387; MODEL - D 6F IMPULSE IM T (5 PK) 20155388; MODEL - D 6F IMPULSE IM T (5 PK) 20155389; MODEL - D 6F IMPULSE IM T (5 PK) 20155390; MODEL - D 6F IMPULSE IM T (5 PK) 20155391; MODEL - D 6F IMPULSE IM T (5 PK) 20155392; MODEL - D 6F IMPULSE IM T (5 PK) 20155393; MODEL - D 6F IMPULSE IM T (5 PK) 20162570; MODEL - D 6F IMPULSE IM T (5 PK) 20162571; MODEL - D 6F IMPULSE IM T (5 PK) 20162572; MODEL - D 6F IMPULSE IM T (5 PK) 20162573; MODEL - D 6F IMPULSE IM T (5 PK) 20162574; MODEL - D 6F IMPULSE IM T (5 PK) 20162575; MODEL - D 6F IMPULSE IM T (5 PK) 20162576; MODEL - D 6F IMPULSE IM T (5 PK) 20165872; MODEL - D 6F IMPULSE IM T (5 PK) 20165873; MODEL - D 6F IMPULSE IM T (5 PK) 20165874; MODEL - D 6F IMPULSE IM T (5 PK) 20165875; MODEL - D 6F IMPULSE IM T (5 PK) 20165876; MODEL - D 6F IMPULSE IM T (5 PK) 20165877; MODEL - D 6F IMPULSE IM T (5 PK) 20165878; MODEL - D 6F IMPULSE IM T (5 PK) 20172989; MODEL - D 6F IMPULSE IM T (5 PK) 20172991; MODEL - D 6F IMPULSE IM T (5 PK) 20172992; MODEL - D 6F IMPULSE IM T (5 PK) 20178248; MODEL - D 6F IMPULSE IM T (5 PK) 20178249; MODEL - D 6F IMPULSE IM T (5 PK) 20178250; MODEL - D 6F IMPULSE IM T (5 PK) 20178251; MODEL - D 6F IMPULSE IM T (5 PK) 20179617; MODEL - 6F IMPULSE IMC 5 PACK 20155394; MODEL - 6F IMPULSE IMC 5 PACK 20155395; MODEL - 6F IMPULSE IMC 5 PACK 20155396; MODEL - 6F IMPULSE IMC 5 PACK 20155397; MODEL - 6F IMPULSE IMC 5 PACK 20155398; MODEL - 6F IMPULSE IMC 5 PACK 20170673; MODEL - 6F IMPULSE IMC 5 PACK 20170674; MODEL - 6F IMPULSE IMC 5 PACK 20174479; MODEL - 6F IMPULSE IMC 5 PACK 20174480; MODEL - 6F IMPULSE IMC 5 PACK 20174482; MODEL - 6F IMPULSE FL3 5-PK 20154951; MODEL - 6F IMPULSE FL3 5-PK 20160651; MODEL - D 6F IMPULSE FL3.5 SINGLE 20154080; MODEL - D 6F IMPULSE FL3.5 5 PACK 20155399; MODEL - D 6F IMPULSE FL3.5 5 PACK 20155400; MODEL - D 6F IMPULSE FL3.5 5 PACK 20156883; MODEL - D 6F IMPULSE FL3.5 5 PACK 20156884; MODEL - D 6F IMPULSE FL3.5 5 PACK 20158196; MODEL - D 6F IMPULSE FL3.5 5 PACK 20162577; MODEL - D 6F IMPULSE FL3.5 5 PACK 20164467; MODEL - D 6F IMPULSE FL3.5 5 PACK 20165879; MODEL - D 6F IMPULSE FL3.5 5 PACK 20165880; MODEL - D 6F IMPULSE FL3.5 5 PACK 20167893; MODEL - D 6F IMPULSE FL3.5 5 PACK 20167894; MODEL - D 6F IMPULSE FL3.5 5 PACK 20167895; MODEL - D 6F IMPULSE FL3.5 5 PACK 20174484; MODEL - D 6F IMPULSE FL3.5 5 PACK 20178252; MODEL - D 6F IMPULSE FL3.5 5 PACK 20179630; MODEL - D 6F IMPULSE FL3.5 5 PACK 20179631; MODEL - D 6F IMPULSE FL3.5 5 PACK 20183262; MODEL - D 6F IMPULSE FL3.5 5 PACK 20183263; MODEL - D 6F IMPULSE FL4 SINGLE 20160888; MODEL - D 6F IMPULSE FL4 SINGLE 20165923; MODEL - D 6F IMPULSE FL4 5 PACK 20155402; MODEL - D 6F IMPULSE FL4 5 PACK 20155403; MODEL - D 6F IMPULSE FL4 5 PACK 20156885; MODEL - D 6F IMPULSE FL4 5 PACK 20156886; MODEL - D 6F IMPULSE FL4 5 PACK 20156887; MODEL - D 6F IMPULSE FL4 5 PACK 20156888; MODEL - D 6F IMPULSE FL4 5 PACK 20156889; MODEL - D 6F IMPULSE FL4 5 PACK 20158198; MODEL - D 6F IMPULSE FL4 5 PACK 20158199; MODEL - D 6F IMPULSE FL4 5 PACK 20158200; MODEL - D 6F IMPULSE FL4 5 PACK 20158201; MODEL - D 6F IMPULSE FL4 5 PACK 20160845; MODEL - D 6F IMPULSE FL4 5 PACK 20160846; MODEL - D 6F IMPULSE FL4 5 PACK 20160847; MODEL - D 6F IMPULSE FL4 5 PACK 20160848; MODEL - D 6F IMPULSE FL4 5 PACK 20160849; MODEL - D 6F IMPULSE FL4 5 PACK 20162578; MODEL - D 6F IMPULSE FL4 5 PACK 20162579; MODEL - D 6F IMPULSE FL4 5 PACK 20162580; MODEL - D 6F IMPULSE FL4 5 PACK 20162581; MODEL - D 6F IMPULSE FL4 5 PACK 20162582; MODEL - D 6F IMPULSE FL4 5 PACK 20162583; MODEL - D 6F IMPULSE FL4 5 PACK 20164469; MODEL - D 6F IMPULSE FL4 5 PACK 20164470; MODEL - D 6F IMPULSE FL4 5 PACK 20164471; MODEL - D 6F IMPULSE FL4 5 PACK 20164472; MODEL - D 6F IMPULSE FL4 5 PACK 20164473; MODEL - D 6F IMPULSE FL4 5 PACK 20164474; MODEL - D 6F IMPULSE FL4 5 PACK 20165882; MODEL - D 6F IMPULSE FL4 5 PACK 20165883; MODEL - D 6F IMPULSE FL4 5 PACK 20165884; MODEL - D 6F IMPULSE FL4 5 PACK 20165885; MODEL - D 6F IMPULSE FL4 5 PACK 20165886; MODEL - D 6F IMPULSE FL4 5 PACK 20167896; MODEL - D 6F IMPULSE FL4 5 PACK 20167897; MODEL - D 6F IMPULSE FL4 5 PACK 20167898; MODEL - D 6F IMPULSE FL4 5 PACK 20167899; MODEL - D 6F IMPULSE FL4 5 PACK 20167900; MODEL - D 6F IMPULSE FL4 5 PACK 20168973; MODEL - D 6F IMPULSE FL4 5 PACK 20168974; MODEL - D 6F IMPULSE FL4 5 PACK 20168975; MODEL - D 6F IMPULSE FL4 5 PACK 20168976; MODEL - D 6F IMPULSE FL4 5 PACK 20168977; MODEL - D 6F IMPULSE FL4 5 PACK 20170675; MODEL - D 6F IMPULSE FL4 5 PACK 20170676; MODEL - D 6F IMPULSE FL4 5 PACK 20170677; MODEL - D 6F IMPULSE FL4 5 PACK 20170678; MODEL - D 6F IMPULSE FL4 5 PACK 20170679; MODEL - D 6F IMPULSE FL4 5 PACK 20172996; MODEL - D 6F IMPULSE FL4 5 PACK 20172997; MODEL - D 6F IMPULSE FL4 5 PACK 20172998; MODEL - D 6F IMPULSE FL4 5 PACK 20172999; MODEL - D 6F IMPULSE FL4 5 PACK 20173000; MODEL - D 6F IMPULSE FL4 5 PACK 20174485; MODEL - D 6F IMPULSE FL4 5 PACK 20174486; MODEL - D 6F IMPULSE FL4 5 PACK 20174487; MODEL - D 6F IMPULSE FL4 5 PACK 20174488; MODEL - D 6F IMPULSE FL4 5 PACK 20174489; MODEL - D 6F IMPULSE FL4 5 PACK 20178253; MODEL - D 6F IMPULSE FL4 5 PACK 20178255; MODEL - D 6F IMPULSE FL4 5 PACK 20178256; MODEL - D 6F IMPULSE FL4 5 PACK 20178257; MODEL - D 6F IMPULSE FL4 5 PACK 20178258; MODEL - D 6F IMPULSE FL4 5 PACK 20178259; MODEL - D 6F IMPULSE FL4 5 PACK 20179634; MODEL - D 6F IMPULSE FL4 5 PACK 20179635; MODEL - D 6F IMPULSE FL4 5 PACK 20179636; MODEL - D 6F IMPULSE FL4 5 PACK 20179637; MODEL - D 6F IMPULSE FL4 5 PACK 20181381; MODEL - D 6F IMPULSE FL4 5 PACK 20181382; MODEL - D 6F IMPULSE FL4 5 PACK 20181384; MODEL - D 6F IMPULSE FL4 5 PACK 20181385; MODEL - D 6F IMPULSE FL4 5 PACK 20181386; MODEL - D 6F IMPULSE FL4 5 PACK 20183265; MODEL - D 6F IMPULSE FL4.5 SINGLE 20167968; MODEL - D 6F IMPULSE FL4.5 SINGLE 20178308; MODEL - D 6F IMPULSE FL4.5 5 PACK 20164475; MODEL - D 6F IMPULSE FL4.5 5 PACK 20165887; MODEL - D 6F IMPULSE FL4.5 5 PACK 20167901; MODEL - D 6F IMPULSE FL4.5 5 PACK 20170680; MODEL - D 6F IMPULSE FL4.5 5 PACK 20178260; MODEL - D 6F IMPULSE FL4.5 5 PACK 20179638; MODEL - D 6F IMPULSE FL5 SINGLE 20155443; MODEL - D 6F IMPULSE FL5 SINGLE 20164520; MODEL - D 6F IMPULSE FL5 5 PACK 20155406; MODEL - D 6F IMPULSE FL5 5 PACK 20156890; MODEL - D 6F IMPULSE FL5 5 PACK 20160851; MODEL - D 6F IMPULSE FL5 5 PACK 20162584; MODEL - D 6F IMPULSE FL5 5 PACK 20164476; MODEL - D 6F IMPULSE FL5 5 PACK 20165889; MODEL - D 6F IMPULSE FL5 5 PACK 20167902; MODEL - D 6F IMPULSE FL5 5 PACK 20170682; MODEL - D 6F IMPULSE FL5 5 PACK 20173004; MODEL - D 6F IMPULSE FL5 5 PACK 20174490; MODEL - D 6F IMPULSE FL5 5 PACK 20178262; MODEL - D 6F IMPULSE FL5 5 PACK 20179639; MODEL - D 6F IMPULSE FL6 SINGLE 20155444; MODEL - D 6F IMPULSE FL6 SINGLE 20162045; MODEL - D 6F IMPULSE FL6 5 PACK 20155407; MODEL - D 6F IMPULSE FL6 5 PACK 20160852; MODEL - D 6F IMPULSE FL6 5 PACK 20165890; MODEL - D 6F IMPULSE FL6 5 PACK 20173005; MODEL - D 6F IMPULSE FL6 5 PACK 20178263; MODEL - D 6F IMPULSE FL6 5 PACK 20179640; MODEL - D 6F IMPULSE PIG 5 PACK 20156898; MODEL - D 6F IMPULSE PIG 5 PACK 20158211; MODEL - D 6F IMPULSE PIG 5 PACK 20158212; MODEL - D 6F IMPULSE PIG 5 PACK 20160857; MODEL - D 6F IMPULSE PIG 5 PACK 20162590; MODEL - D 6F IMPULSE PIG 5 PACK 20167910; MODEL - D 6F IMPULSE PIG 5 PACK 20168986; MODEL - D 6F IMPULSE PIG 5 PACK 20168987; MODEL - D 6F IMPULSE PIG 5 PACK 20170689; MODEL - D 6F IMPULSE PIG 5 PACK 20173011; MODEL - D 6F IMPULSE PIG 5 PACK 20173012; MODEL - D 6F IMPULSE PIG 5 PACK 20174497; MODEL - D 6F IMPULSE PIG 5 PACK 20174498; MODEL - D 6F IMPULSE PIG 5 PACK 20178271; MODEL - D 6F IMPULSE PIG 5 PACK 20178272; MODEL - D 6F IMPULSE PIG 5 PACK 20179647; MODEL - D 6F IMPULSE PIG 5 PACK 20183279; MODEL - D 6F IMPULSE PIG 145 SINGLE 20160891; MODEL - D 6F IMPULSE PIG 145 SINGLE 20162617; MODEL - D 6F IMPULSE PIG 145 SINGLE 20178311; MODEL - D 6F IMPULSE PIG 145 SINGLE 20179686; MODEL - D 6F IMPULSE PIG 145 5 PACK 20156899; MODEL - D 6F IMPULSE PIG 145 5 PACK 20158213; MODEL - D 6F IMPULSE PIG 145 5 PACK 20158214; MODEL - D 6F IMPULSE PIG 145 5 PACK 20160858; MODEL - D 6F IMPULSE PIG 145 5 PACK 20160859; MODEL - D 6F IMPULSE PIG 145 5 PACK 20162591; MODEL - D 6F IMPULSE PIG 145 5 PACK 20162592; MODEL - D 6F IMPULSE PIG 145 5 PACK 20167912; MODEL - D 6F IMPULSE PIG 145 5 PACK 20167913; MODEL - D 6F IMPULSE PIG 145 5 PACK 20168988; MODEL - D 6F IMPULSE PIG 145 5 PACK 20168989; MODEL - D 6F IMPULSE PIG 145 5 PACK 20170691; MODEL - D 6F IMPULSE PIG 145 5 PACK 20173013; MODEL - D 6F IMPULSE PIG 145 5 PACK 20174499; MODEL - D 6F IMPULSE PIG 145 5 PACK 20178273; MODEL - D 6F IMPULSE PIG 145 5 PACK 20179649; MODEL - D 6F IMPULSE PIG 145 5 PACK 20179650; MODEL - D 6F IMPULSE PIG 145 5 PACK 20181398; MODEL - D 6F IMPULSE PIG 145 5 PACK 20181399; MODEL - D 6F IMPULSE PIG 155 SINGLE 20174533; MODEL - D 6F IMPULSE PIG 155 5 PACK 20156900; MODEL - D 6F IMPULSE PIG 155 5 PACK 20160860; MODEL - D 6F IMPULSE PIG 155 5 PACK 20162593; MODEL - D 6F IMPULSE PIG 155 5 PACK 20170693; MODEL - D 6F IMPULSE PIG 155 5 PACK 20173014; MODEL - D 6F IMPULSE PIG 155 5 PACK 20174500; MODEL - D 6F IMPULSE PIG 155 5 PACK 20178274; MODEL - D 6F IMPULSE PIG 155 5 PACK 20179651; MODEL - D 6F IMPULSE SL PIG 145 5 PACK 20162594; MODEL-6F IMPULSE CLS3.5 5-PK 20151691; MODEL - D 6F IMPULSE AR1 SINGLE 20155448; MODEL - D 6F IMPULSE AR1 SINGLE 20156927; MODEL - D 6F IMPULSE AR1 SINGLE 20158242; MODEL - D 6F IMPULSE AR1 SINGLE 20165927; MODEL - D 6F IMPULSE AR1 SINGLE 20173036; MODEL - D 6F IMPULSE AR1 SINGLE 20174534; MODEL - D 6F IMPULSE AR1 5 PACK 20162595; MODEL - D 6F IMPULSE AR1 5 PACK 20164487; MODEL - D 6F IMPULSE AR1 5 PACK 20178275; MODEL - D 6F IMPULSE AR1 5 PACK 20179653; MODEL - D 6F IMPULSE AR2 SINGLE 20162618; MODEL - D 6F IMPULSE AR2 SINGLE 20174535; MODEL - D 6F IMPULSE AR2 5 PACK 20154066; MODEL - D 6F IMPULSE AR2 5 PACK 20158215; MODEL - D 6F IMPULSE AR2 5 PACK 20164488; MODEL - D 6F IMPULSE AR2 5 PACK 20165900; MODEL - D 6F IMPULSE AR2 5 PACK 20170694; MODEL - D 6F IMPULSE AR2 5 PACK 20174501; MODEL - D 6F IMPULSE AR2 5 PACK 20178276; MODEL - D 6F IMPULSE AR2 5 PACK 20179654; MODEL - D 6F IMPULSE AR MOD 5 PACK 20156901; MODEL - D 6F IMPULSE AR MOD 5 PACK 20162596; MODEL - D 6F IMPULSE AR MOD 5 PACK 20164489; MODEL - D 6F IMPULSE AR MOD 5 PACK 20178277; MODEL - D 6F IMPULSE AR MOD 5 PACK 20179655; MODEL - D 6F IMPULSE AL1 SINGLE 20174536; MODEL - D 6F IMPULSE AL1 SINGLE 20179688; MODEL - D 6F IMPULSE AL1 5 PACK 20155419; MODEL - D 6F IMPULSE AL1 5 PACK 20156902; MODEL - D 6F IMPULSE AL1 5 PACK 20160862; MODEL - D 6F IMPULSE AL1 5 PACK 20162597; MODEL - D 6F IMPULSE AL1 5 PACK 20164490; MODEL - D 6F IMPULSE AL1 5 PACK 20165901; MODEL - D 6F IMPULSE AL1 5 PACK 20167914; MODEL - D 6F IMPULSE AL1 5 PACK 20170696; MODEL - D 6F IMPULSE AL1 5 PACK 20173015; MODEL - D 6F IMPULSE AL1 5 PACK 20174502; MODEL - D 6F IMPULSE AL1 5 PACK 20179656; MODEL - D 6F IMPULSE AL1 5 PACK 20181404; MODEL - D 6F IMPULSE AL2 SINGLE 20155449; MODEL - D 6F IMPULSE AL2 SINGLE 20167970; MODEL - D 6F IMPULSE AL2 SINGLE 20174537; MODEL - D 6F IMPULSE AL2 SINGLE 20178313; MODEL - D 6F IMPULSE AL2 5 PACK 20154067; MODEL - D 6F IMPULSE AL2 5 PACK 20155420; MODEL - D 6F IMPULSE AL2 5 PACK 20158216; MODEL - D 6F IMPULSE AL2 5 PACK 20160863; MODEL - D 6F IMPULSE AL2 5 PACK 20162598; MODEL - D 6F IMPULSE AL2 5 PACK 20164491; MODEL - D 6F IMPULSE AL2 5 PACK 20165902; MODEL - D 6F IMPULSE AL2 5 PACK 20167915; MODEL - D 6F IMPULSE AL2 5 PACK 20173016; MODEL - D 6F IMPULSE AL2 5 PACK 20178279; and MODEL - D 6F IMPULSE AL2 5 PACK 20179657. |
Recalling Firm/ Manufacturer |
Boston Scientific Corporation 1 Scimed Pl Maple Grove MN 55311-1565
|
For Additional Information Contact | 763-494-1700 |
Manufacturer Reason for Recall | Boston Scientific is initiating a recall removal of numerous batches/lots of Wiseguide Guide Catheter, Impulse Angiographic (Diagnostic) Catheter, Expo Angiographic (Diagnostic) Catheter. Boston Scientific has determined that the sterile barrier in the packaging may be compromised. During internal inspections, they discovered gaps between the side seals and the top seal approximately 3/8 in len |
FDA Determined Cause 2 | Packaging |
Action | Boston Scientific issued an "Urgent Medical Device Recall Removal" letter dated August 5, 2009. The letter was addressed to Risk Manger / Field Action Contact. The letter described the problem and product involved. The Firm recommended to segregate affected product immediately and return to Boston Scientific and requested return of the "The Reply Verification Tracking Form" enclosed with the letter.
For further questions, contact your local Boston Scientific Sales Representative or call 1-763-494-1634. |
Quantity in Commerce | 391,788 (US 212,299; OUS 155,599) |
Distribution | Worldwide Distribution -- US (including DC and Puerto Rico), Japan, United Arab Emirates, Albania, Austria, Belgium, Bulgaria, Bahrain, Switzerland, Czech Republic, Germany, Denmark, Algeria, Egypt, Spain, Finland, France, Great Britain, Greece, Hungary, Ireland, Israel, Iraq, Iceland, Italy, Jordan, Kuwait, Lebanon, Lithuania, Libya, Morocco, Netherlands, Norway, Pakistan, Poland, West Bank Gaza Strip/ Palestine, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Slovakia, Syria, Turkey, Ukraine, South Africa, Argentina, Australia, Brazil, Canada, Chile, Colombia, Dominican Republic, Ecuador, Guatemala, Indonesia, India, South Korea, Mexico, Malaysia, Nicaragua, New Zealand, Panama, Philippines, Puerto Rico, Paraguay, Singapore, El Salvador, Thailand, Taiwan, Uruguay, American Virgin Islands and Vietnam. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DQO
|
|
|
|