 Class 2 Recall
Zimmer Gender Solutions Natural Knee Flex System, Prolong Articular Surface,
see related information
Date Posted November 13, 2009
Recall Status1 Terminated on May 16, 2011
Recall Number Z-0205-2010
Recall Event ID 53399
Premarket Notification
510(K) Number
K070214 
Product Classification Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer - Product Code JWH
Product Zimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, left, 11 MM height, use with baseplate size 1 or 2, sterile, Zimmer, Warsaw, IN; REF 00-5424-011-11. The device is a semiconstrained total knee prosthesis components indicated for uncemented or cemented use in skeletally mature individuals.
Code Information Lots 60916390, 60916392, 60916393, 60916424, 60930896, 60947731, 60954140, 60954150, 60958269, 60964459, 60964479, 60981787, 60981801, 60987819, 60987833, 60987834, 60992923, 60992924, 61001544, 61005434, 61005448, 61023610, 61027878, 61036338, 61036367, 61036368, 61051468, 61064799, 61069407, 61075236, 61078744, 61078745, 61091176, 61091192, 61091193, 61101449, 61101450, 61106550, 61114163, 61114174, 61117033, 61117050 and 61122282.
Recalling Firm/
Manufacturer
Zimmer Inc.
345 E Main St
Warsaw, Indiana 46580-2746
Manufacturer Reason
for Recall
The surgeon may have difficulty inserting the polyethylene into the tibial baseplate, which may result in intraoperative damage to the device.
FDA Determined
Cause 2
OTHER/UNDETERMINED: Pending
Action An "Urgent: Field Correction" dated October 8, 2009 was issued to Consignees. The notification letter described the affected product, issue, health risks and actions for customers to take. The letter notified consignees of a changed to the surgical technique, to include the use of a new peripheral rib impactor tool. Notifications of this field notification were sent to all affected direct accounts of Zimmer, Inc. Contact your Zimmer Sales Representative if you have any questions regarding this action. For shipping assistance, questions or assistance in notifying your accounts about the field correction please contact Zimmer, Inc. at 1-800-613-6131.
Quantity in Commerce 26,404 of all products
Distribution Worldwide Distribution -- United States, Austria, Belgium, France, Germany, Japan, Spain and Switzerland.