  Class 2 Device Recall Stryker Equipment Delivery System (EDS") see related information
Date Initiated by Firm April 15, 2010
Date Posted May 21, 2010
Recall Status1 Terminated 3 on January 24, 2011
Recall Number Z-1641-2010
Recall Event ID 55477
510(K)Number K0608022  K031068  K033132  
Product Classification Ceiling Mounted Surgical Light - Product Code FSY
Product Stryker Equipment Delivery System (EDS); Ceiling Mounted Equipment Booms; Catalog Number: 0682000383, 0682000384, 0682000385(A-D), 0682000386(A-D), 0682000426, 0682000429(A-D), 0682000326, 0682000311, 0682000361, 0682000362, 0682000230, 0682000337, 0682000338, 0682000232, 0682000233, 0682000339, 0682000234, 0682000235, 0682000236, 0682000340, 0682000341, 0682000238, 0682000239, 0682000342, 0682000240, 0682000241, 0682000931, 0682000343, 0682000344, 0682000328, 0682000933, 0682000345, 0682000329, 0682000330, 0682000934, 0682000346, 0682000347, 0682000331, 0682000936, 0682000348, 0682000332, 0682000333, 0682000401, 0682000302, & 0682000327.

The booms are ceiling mounted devices. They are available in articulating and non-articulating configurations, in single or tandem (pair) configuration, with different shelves configuration to allow for various load capacities.

Intended for use as ceiling-mounted devices for supporting or positioning equipment in the surgical suite and/or delivers gases or electricity to these devices.
Code Information Lot Code: Product Installed Between January 1, 2008 and August 31, 2009.
Recalling Firm/
Manufacturer
Stryker Communications Corp
1410 Lakeside Pkwy Ste 100
Flower Mound TX 75028-4026
For Additional Information Contact
972-410-7310
Manufacturer Reason
for Recall
Installation records are incomplete.
FDA Determined
Cause 2
Employee error
Action An Urgent: Medical Device Product Correction customer notification letter, dated 4/15/2010, has been sent out to all affected accounts via regular mail in the form of a certified, return request receipt letter. Accounts will receive a follow-up phone call from a Stryker Representative to schedule a re-inspection of equipment to see if it mets established critical specifications or not. The firm recommends that customers continue using all Stryker Communications products listed in the notice. Questions can be directed to Katie Blackwell at 972-410-7660 or via email at StrykerCommunicationsRA@stryker.com.
Quantity in Commerce 608 (booms and lights)
Distribution Nationwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = FSY and Original Applicant = STRYKER COMMUNICATIONS CORP.
510(K)s with Product Code = FSY and Original Applicant = Stryker Endoscopy
-
-