| | Class 2 Recall Zimmer dental |  |
| Date Posted |
November 23, 2010 |
| Recall Number |
Z-0444-2011 |
| Product |
Zimmer Dental, Drill stop kit, REF DSKIT, for use with dental implants. |
| Code Information |
61482528, 61484443, 61495269 and 61479592.
|
Recalling Firm/ Manufacturer |
Zimmer Inc. 345 E Main St Warsaw, Indiana 46580-2746 |
Reason for Recall |
The instructions for use contain pre-vacuum sterilization instructions that may lead to compromised sterility at the lid-kit interface location.
|
| Action |
Zimmer sent out "Urgent Device Correction" letters dated 5/18/2010. The consignees were provided with new instructions for use. The letter informed users that the current pre-vacuum cycle of 3 to 18 minutes is not enough time to assure sterility between the lid-kit interface and could lead to cross contamination of multiple patients. The users were informed that a pre-vacuum time of 30 minutes was required to assure sterility. |
| Quantity in Commerce |
158 |
| Distribution |
Distribution Nationwide, to France, and Spain. |
| |
|
|
|
|