• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Innova 2000 Digital Flouroscopic Imaging System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Innova 2000 Digital Flouroscopic Imaging System
see related information
Date Posted August 19, 2010
Recall Status1 Terminated on April 01, 2014
Recall Number Z-2238-2010
Recall Event ID 55921
Premarket Notification
510(K) Numbers
K022322  K024200  K993037 
Product Classification Solid State X-Ray Imager (Flat Panel/Digital Imager) - Product Code MQB
Product Innova 2000, manufactured by GE Medical Systems, SCS
Code Information Serial Number 00000407955BU4 00000068492VE8 00000418005BU5 00000393455BU1 00000397050BU6 00000350519BU5 00000079547VE6 00000378676BU1 00000099732VEO 00000079541VE9 00000341536BU1 00000347472BU3 00000338024BU3 00000341530BU4 00000348536BU4 00000422097BU6 00000423545BU3 00000385319BU9 00000409300BU1 00000389501BU8 00000383125BU2 00000423003BU3 00000415056BU1 00000420520BU9 00000379761BU0 00000418004BU8 00000377024BU5 00000419716BU6 00000400569BU0 00000386267BU9 00000068497VE7 00000423644BU4 00000423322BU7 00000405271BU8 00000099736VE1 00000099734VE6 00000079533VE6 00000392274BU7 00000343649BU0 00000400526BU0 00000354849BU2 00000404101BU8 00000399595BU8 00000336869BU3 00000337306BU5 00000338357BU7 00000392563BU3 00000068487VE8 00000074875VE6 00000074883VE0 00000351997BU2 00000422239BU4 00000403685BU1 00000403431BU0 00000393733BU1 00000074904VE4 00000341095BU8 00000396197BU6 00000382877BU9 00000420841BU9 00000015656VE2 00000382939BU7 00000415705BU3 00000351549BU1 00000359555BU0 00000407957BU0 00000342077BU5 00000068495VE1 00000347111BU7 00000333743BU3 00000372500BU9 00000387489BU8 00000418299BU4 00000351996BU4 00000338356BU9 00000393646BU5 00000347468BU1 00000347468BU1 00000354203BU2 00000424046BU1 00000399750BU9 00000087032VE9 00000351552BU5 00000395215BU7 00000345249BU7 00000344212BU6 00000404675BU1 00000405835BU0 00000019854VE9 00000409880BU2 00000419653BU1 00000104763VE8 00000407956BU2 00000397405BU2 00000384311BU7 00000402924BU5 00000074886VE3 00000413029BU0 00000419167BU2 00000039470VE0 00000087013VE9 00000400646BU6 00000025817VE8 00000342699BU6 00000104774VE5 00000104755VE4 00000422426BU7 00000395214BU0 00000397053BU0 00000361744BU6 00000019863VE0 00000074900VE2 00000392773BU8 00000386173BU9 00000388900BU3 00000407950BU5 00000408333BU3 00000382096BU6 00000104767VE9 00000389097BU7 00000389796BU4 00000389994BU5 00000397644BU6 00000397643BU8 00000099718VE9 00000074891VE3 00000386307BU3 00000332678BU2 00000385405BU6 00000394210BU9 00000104761VE2 00000099738VE7 00000368115BU2 00000383992BU5 00000367496BU7 00000350866BU0 00000421317BU9 00000099710VE6 00000380627BU0 00000424693BU0 00000055062VE4 00000399939BU8 00000392732BU4 00000400805BU8 00000376219BU2 00000424660BU9 00000068477VE9 00000104769VE5 00000416123BU8 00000408600BU5 00000074905VE1 00000354204BU0 00000337504BU5 00000083311VE1 00000074899VE6 00000087017VE0 00000104756VE1 00000416885BU2 00000338640BU6 00000411170BU4 00000401024BU5 00000414823BU5 00000394212BU5 00000341749BU0 00000421580BU2 00000373282BU3 00000099729VE6 00000087014VE7 00000432743BU3 00000426713BU4 UNK2243594-300 00000423873BU9 00000411171BU2 00000404677BU7 00000407657BU6 00000406023BU2 00000366246BU7 00000406108BU1 00000392245BU7 00000375409BU0 00000381436BU5 00000374480BU2 00000385830BU5 00000426880BU1 00000409162BU5 00000392766BU2 00000440232BU7 00000440085BU9 00000381846BU5 00000347473BU1 00000083310VE3 00000079544VE3 00000393731BU5 00000026257VE6 00000421196BU7 00000343038BU6 00000346258BU7 00000386952BU6 00000346261BU1 00000431458BU9 00000407088BU4 00000344210BU0 00000358756BU5 00000375205BU2 00000354201BU6 00000327810BU8 00000104754VE7 00000068467VE0 00000074885VE5 00000367497BU5 00000374925BU6 00000416881BU1 00000418425BU5 00000351995BU6 00000377023BU7 00000375036BU1 00000427302BU5 00000074882VE2 00000083318VE6 00000358524BU7 00000074870VE7 00000417861BU2 00000417880BU2 00000352550BU8 00000358904BU1 00000095431VE3 00000099724VE7 00000095430VE5 00000087026VE1 00000420495BU4 00000419394BU2 00000380199BU0 00000393000BU5 00000351994BU9 00000386618BU3 00000379745BU3 00000421104BU1 00000373281BU5 00000385862BU8 00000384372BU9 00000385111BU0 00000384960BU1 00000451820BU5 00000399753BU3 00000408875BU3 00000013600GP2 00000366737BU5 00000356478BU8 00000079553VE4 00000410336BU2 00000393732BU3 00000380087BU7 00000339238BU8 00000340822BU6 00000347108BU3 00000015654VE7 00000403125BU8 00000033923BU0 00000366738BU3 00000354850BU0 00000074902VE8 00000309161BU8 00000380939BU9 00000415052BU0 00000392999BU9 00000392236BU6 00000366739BU1 00000358611BU2 00000068476VE1 00000367089BU0 00000415567BU7 00000074873VE1 00000351550BU9 00000087021VE2 00000068490VE2 00000338553BU1 00000341066BU9 00000379434BU4 00000104772VE9 00000104775VE2 00000356839BU1 00000350518BU7 00000099731VE2 00000351547BU5 00000079528VE6 00000349343BU4 00000403889BU9 00000350883BU5 00000368221BU8 00000419135BU9 00000083327VE7 00000356837BU5 00000359557BU6 00000415178BU3 00000399385BU4 00000349342BU6 00000074903VE6 00000340790BU5 00000345669BU6 00000342460BU3 00000374684BU9 00000369296BU9 00000099725VE4 00000083313VE7 00000403501BU0 00000425964BU4 00000099716VE3 00000432663BU3 00000079529VE4 00000068481VE1 00000351551BU7 00000378710BU8 00000394045BU9 00000395515BU0 00000423214BU6 00000340373BU0 00000404678BU5 00000408100BU6 00000409042BU9 00000351993BU1 00000087028VE7 00000426942BU9 00000352272BU9 00000390184BU0 00000321665BU2 00000324194BU0 00000074909VE3 00000039521BU4 00000352611BU8 00000104752VE1 00000389096BU9 00000373302BU9 00000421110BU8 00000358211BU1 00000361951BU7 00000401794BU3 00000445624BU0 00000387300BU7 00000403023BU5 00000337772BU8 00000337068BU1 00000417673BU1 00000095432VE1 00000416963BU7 00000399601BU4 00000352614BU2 00000380196BU6 00000364938BU1 00000381847BU3 00000391288BU8 00000392087BU3 00000074884VE8 00000366740BU9 00000352612BU6 00000400898BU3 00000420895BU5 00000083332VE7 00000087037VE8 00000104765VE3 00000400746BU4 00000420898BU9 00000419495BU7 00000339804BU7 00000411057BU3 00000397051BU4 00000419220BU9 00000110175VE7 00000413280BU9 00000386741BU3 00000425718BU4 00000413390BU6 00000083305VE3 00000426727BU4 00000074881VE4 00000087035VE2 00000416241BU8 00000340820BU0 00000336924BU6 00000423036BU3 00000074880VE6 00000416240BU0 00000399715BU2 00000425771BU3 00000405951BU5 00000374961BU1 00000425142BU7 00000036630VE2 00000379337BU9 00000360536BU7 00000340211BU2 00000404676BU9 00000364519BU9 00000346259BU5 00000422423BU4 00000425370BU4 00000352610BU0 00000416729BU2 00000416788BU8 00000365381BU3 00000422955BU5 00000367495BU9 00000095433EV9 00000341260BU8 00000416908BU2 00000426650BU8 00000365379BU7 00000375037BU9 00000378066BU5 00000366244BU2 00000360539BU1 00000370167BU9 00000374489BU3 00000382773BU0 00000099730VE4 00000079526VE0 00000423468BU8 00000422567BU8 00000422432BU5 00000013404VE9 00000343647BU4 00000333964BU5 00000334722BU6 00000383540BU2 00000099722VE1 00000417997BU4 00000340384BU7 00000361277BU7 00000410737BU1 00000406234BU5 00000079538VE5 00000068486VE0 00000410945BU0 00000068472VE0 00000015652VE1 00000367086BU6 00000380940BU7 00000104764VE6 00000354200BU8 00000087039VE4 00000392468BU5 00000361950BU9 00000079531VE0 00000407647BU7 00000419070BU8 00000338723BU0 00000417921BU4 00000329435BU2 00000330124BU9 00000421359BU1 00000110179VE9 00000352270BU3 00000359559BU2 00000417570BU9 00000401599BU6 00000332824BU2 00000099712VE2 00000335480BU0 00000074878VE0 00000083321VE0 00000039437VE9 00000401211BU8 00000400151BU7 00000416889BU4 00000036629VE4 00000449322BU7 00000408597BU3 00000407954BU7 00000322232BU0 00000393185BU4 00000409163BU3 00000367087BU4 00000386059BU0 00000104777VE3 00000074895VE4 00000393647BU3 00000409155BU9 00000416563BU5 00000416671BU6 00000398011BU7 00000393186BU2 00000378074BU9 00000388459BU0 00000063266VE1 00000420104BU2 00000351548BU3 00000381708BU7 00000394211BU7 00000388087BU9 00000406580BU1 00000411395BU7 00000087015VE4 00000099709VE8 00000373253BU4 00000358212BU9 00000416632BU8 00000083314VE5 00000392580BU7 00000331854BU0 00000349893BU8 00000417594BU9 00000416585BU8 00000416580BU9 00000417599BU8 00000403929BU3 00000422771BU6 00000087027VE9 00000074887VE1 00000422469BU7 00000394209BU1 00000382816BU7 00000340821BU8 00000099735VE3 00000416864BU7 00000418000BU6 00000079548VE4 00000087036VE0 00000406792BU2 00000074908VE5 00000324770BU7 000000393456U9 00000406785BU6 00000394044BU2 00000380451BU5 00000379404BU7 00000079554VE2 00000421608BU1 00000110178VE1 00000334721BU8 00000377021BU1 00000087024VE6 00000083326VE9 00000074898VE8 00000104751VE3 00000418303BU4 00000425213BU6 00000424551BU0 00000099726VE2 00000417028BU8 00000397052BU2 00000084188VE2 00000068461VE3 00000424699BU7 00000387849BU3 00000087040VE2 00000079540VE1 00000083324VE4 00000083307VE9 00000343653BU2 00000397407BU8 00000343648BU2 00000397641BU2 00000104766VE1 00000441147BU6 00000420411BU1 00000418197BU0 00000104760VE4 00000418415BU6 00000405768BU3 00000404103BU4 00000387866BU7 00000398007BU5 00000389235BU3 00000377607BU7 00000378073BU1 00000068485VE2 00000099717VE1 00000104757VE0 00000099713VE0 00000403733BU9 00000402251BU3 00000401600BU2 00000402054BU1 00000404102BU6 00000074893VE9 00000449636BU0 00000079549VE2 00000446890BU6 00000104776VE0 00000404104BU2 00000011119VE5 00000419071BU6 00000423555BU2 00000349344BU2 00000336480BU9 00000423647BU7 00000389995BU2 00000079530VE2 00000422787BU2 00000415426BU6 00000074896VE2 00000390720BU1 00000330206BU4 00000409830BU7 00000079543VE5 00000404082BU0 00000364940BU7 00000410853BU6 00000095436VE2 00000444394BU1 00000415702BU0 00000424731BU8 00000416084BU2 00000423587BU5 00000350882BU7 00000380890BU4 00000099715VE5 00000068242VE7 00000083325VE1 00000104770VE3 00000087011VE3 00000402250BU5 00000405330BU2 00000068482VE9 00000074890VE5 00000099714VE8 00000068493VE6 00000407081BU9 00000409301BU9 00000074897VE0 00000083316VE0 00000445944BU2 00000104753VE9 00000417032BU0 00000068465VE4 00000352273BU7 00000422429BU1 00000425360BU5 00000386409BU7 00000394043BU4 00000411914BU5 00000423019BU9 00000110173VE2 00000410516BU9 00000068488VE6 00000375337BU3 00000398009BU1 00000333963BU7 00000397406BU0 00000398008BU3 00000109/71354 00000400097BU2 00000354848BU4 00000342264BU9 00000397404BU5 00000383127BU8 00000400349BU7 00000367088BU2 00000418046BU9 00000343646BU6 00000344211BU8 00000348539BU8 00000099737VE9 00000389446BU6 00000372953BU0 00000391287BU0 00000400664BU9 00000401287BU8 00000354202BU4 00000412511BU8 00000369298BU5 00000382386BU1 00000419308BU2 00000407953BU9 00000382468BU7 00000367498BU3 00000410223BU2 00000398010BU9 00000350884BU3 00000368222BU6 00000350516BU1 00000332112BU2 00000346257BU9 00000347470BU7 00000370168BU7 00000342263BU1 00000083312VE9 00000367494BU2 00000400186BU3 00000352274BU5 00000087016VE2 00000068489VE4 00000079525VE2 00000376220BU0 00000083308VE7 00000347109BU1 00000014925VE2 00000110176VE5 00000104758VE8 00000386405BU5 00000378675BU3 00000110186VE4 00000395216BU5 00000087031VE1 00000398913BU4 00000408840BU7 00000404430BU1 00000079527VE8 00000449766BU5 00000381435BU7 00000425963BU6 00000087025VE3 00000087038VE6 00000104779VE4 00000365378BU9 00000391181BU5 00000377864BU4 00000079545VE0 00000405374BU0 00000446840BU1 00000068484VE5 00000424933BU0 00000369297BU7 00000083306VE1 00000074871VE5 00000087034VE5 00000348537BU2 00000099711VE4 00000418278BU8 00000368218BU4 00000392765BU4 00000068496VE9 00000083309VE5 00000014926VE0 00000380382BU2 00000104773VE7 00000083320VE2 00000352271BU1 00000013334VE8 00000342700BU2 00000087023VE8 00000423327BU6 00000361278BU5 00000074901VE0 00000110182VE3 00000110174VE0 00000380889BU6 00000099727VE0 00000421902BU8 00000099728VE8 00000350885BU0 00000068466VE2 00000083323VE6 00000409001BU5 00000110184VE9 00000412931BU8 00000110180VE7 00000421701BU4 00000099719VE7 00000068491VE0 00000422314BU5 00000099723VE9 00000418279BU6 00000420110BU9 00000344209BU2 00000383390BU2 00000338036BU7 00000374299BU6 00000426551BU8 00000377865BU1 00000083330VE1 00000421143BU9 00000410669BU6 00000415820BU0 00000352613BU4 00000361952BU5 00000427204BU3 00000087012VE1 00000420032BU5 00000083319VE4 00000417822BU4 00000374276BU4 00000074874VE9 00000079539VE3 00000079542VE7 00000358612BU0 00000104771VE1 00000104780VE2 00000074894VE7 00000424701BU1 00000418416BU4 00000350517BU9 00000074906VE9 00000000001585 00000079552VE6 00000403124BU1 00000352871BU8 00000312089BU6 00000380585BU0 00000068494VE4 00000384940BU3 00000330432BU6 00000361743BU8 00000068223VE7 00000079551VE8 00000350886BU8 00000052637VE6 00000458498BU3 00000079546VE8 00000095439VE6 00000110177VE3 00000388606BU6 00000411199BU3 00000394046BU7 00000334724BU2 00000418238BU2 00000074877VE2 00000395516BU8 00000371685BU9 00000424501BU5 00000110181VE5 00000424821BU7 00000348540BU6 00000348538BU0 00000366741BU7 00000074889VE7 00000418277BU0 00000424700BU3 00000359558BU4 00000395213BU2 00000418289BU5 00000083328VE5 00000110183VE1 00000388258BU6 00000095440VE4 00000410337BU0 00000079550VE0 00000418210BU1 00000397645BU3 00000083383VE5 00000099721VE3 00000110185VE6 00000387191BU0 00000393734BU9 00000377608BU5 00000421695BU8 00000349892BU0 00000342698BU8 00000402055BU8 00000074876VE4 00000381174BU2 00000397408BU6 00000079537VE7 00000403958BU2 00000074879VE8 00000349889BU6 00000383539BU4 00000409881BU0 00000068471VE2 00000397054BU8 00000419496BU5 00000347469BU9 00000015660VE4 00000427075BU7 00000079532VE8 00000417895BU0 00000079535VE1 00000083331VE9 00000349890BU4 00000354852BU6 00000343061BU8 00000359556BU8 00000419206BU8 00000364026BU5 00000339405BU3 00000349341BU8 00000397642BU0 00000039441VE1 00000095434VE7 00000377022BU9 00000095437VE0 00000382385BU3 00000339845BU0 00000436765BU2 00000330900BU2 00000416277BU2 00000415812BU7 00000434301BU8 00000465429BU9 00000478477BU3 and 00000087019VE6
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha, Wisconsin 53188-1615
Manufacturer Reason
for Recall
A hardware component (potentiometer and encoder)mismatch in Innova 2000 digital fluoroscopic imaging systems may cause a Source to Image Distance (SID) failure, which will cause the imager to fail to detect the distance between the imager and the patient. This will result in loss of the x-ray image if the recovery process is not performed. The error message to use the Emergency Back-Out Button for
FDA Determined
Cause 2
OTHER/UNDETERMINED: Pending
Action GE Healthcare sent a GE "Urgent Medical Device Correction" dated June 9, 2010letter to consignees. The letter was addressed to Hospital Administrators / Risk Mangers; Mangers of Radiology/Cardiology and Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product Details; Safety Instruction; Product Correction and Contact Information. The letter detailed the method for identifying SID failure and the use of the emergency back-out buttoms to recover the X-ray image. A GE service representative will affix a label to the device on-site to remind operators how to recover from loss of X-ray from SID failure. Consignees may contact GE at 800 437-1171, option 4.
Quantity in Commerce 885
Distribution Worldwide distribution: States of AL, AZ, AR, CA, CO, CT,DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV,NH, NJ, NM, NH,NC, ND, OH, OK OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, and WI, and countries of VENEZUELA, UNITED KINGDOM, TURKEY, TUNISIA, THAILAND, TAIWAN, SYRIAN ARAB REPUBLIC, SWITZERLAND, SWEDEN SPAIN, SOUTH AFRICA, SAUDI ARABIA, RUSSIAN FEDERATION, REPUBLIC OF KOREA, POLAND, PAKISTAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MEXICO, MARTINIQUE, LITHUANIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, ISLAMIC REPUBLIC OF IRAN, IRELAND, INDONESIA, INDIA, HUNGARY, HONG KONG, GREECE, GERMANY, FRANCE, FINLAND, EGYPT , ECUADOR, COLOMBIA, CHINA, CANADA, BULGARIA, BRAZIL, BELGIUM, BELARUS, AUSTRALIA, and ALGERIA.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = MQB and Original Applicant = GE MEDICAL SYSTEMS, INC.
-
-