| Class 2 Device Recall Aplio XG model SSA790A | |
Date Initiated by Firm | June 02, 2010 |
Date Posted | August 10, 2011 |
Recall Status1 |
Terminated 3 on April 20, 2012 |
Recall Number | Z-2981-2011 |
Recall Event ID |
55941 |
510(K)Number | K091295 |
Product Classification |
System, imaging, pulsed echo, ultrasonic - Product Code IYO
|
Product | Aplio XG model SSA-790A
Aplio XG (SSA-790A) is a premium-class diagnostic ultrasound system incorporating state-of-the-art technologies. It is provided with the highest level of image quality and operability (reflecting the XG (next Generation) in its name), and can be used with ease for research as well as for routine clinical studies. |
Code Information |
4899 5131, 5132, 5133, 5134, 5224, 5139, 5260, 5261, 5369, 5256, 5367, 4883, 4884, 5370, 5374, 5375, 5264, 5257, 5135, 5223, 5379, 5262, 6283, 4886, 5258, 5382, 5383, 5384, 5385, 5216, 4887, 4897, 4898, 5027, 5380, 4982, 5026, 5137, 5138, 5219, 5220, 4979, 5465, 4900, 5466, 4978, 5377, 5136, 5376, 5072, 5071, 5221, 5215, 4981, 5371, 5372, 5373, 5029, 5030, 5068, 5069, 5070, 5222, 4980, 4885, 5259, 5028, 4896, 5368, 5217, 5218, 5140, 5265, 5364, 5365, 5366, 5381, 5467, 5468, 5469, 5470, 5471, 5472, 5473, 5474 |
Recalling Firm/ Manufacturer |
Toshiba American Medical Systems Inc 2441 Michelle Dr P.O. Box 2068 Tustin CA 92781-2068
|
For Additional Information Contact | 714-730-5000 |
Manufacturer Reason for Recall | Toshiba America Medical Systems Inc (TAMS) is initiating a field correction on the the Aplio XG model SSA-790A because it did not meet the requirement of clause 15b, IEC60601-1 which requires that the residual voltage shall not exceed 60V after 1 second from taking off the AC plug. |
FDA Determined Cause 2 | Device Design |
Action | Toshiba America Medical Systems (TAMS) sent an "URGENT:MEDICAL DEVICE CORRECTION' letter dated June 2, 2010 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers.
Customers were instructed to be careful not to touch the contacts or prongs on the AC plug within two seconds after unplugging the system from a wall socket. This precaution is to be adhered to until a Toshiba Service Representative replaces the power supply. Customers are to complete and return a Customer Reply Form and fax to 877-349-3054.
Customers with questions regarding the notification should contact the Director of Regulatory Affairs at (800) 421-1968 or call their local Toshiba Representative at (800) 521-1968. |
Quantity in Commerce | 78 |
Distribution | Nationwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IYO
|
|
|
|