| Class 2 Device Recall HEMOSTATIX HANDLE | |
Date Initiated by Firm | July 27, 2010 |
Date Posted | September 28, 2010 |
Recall Status1 |
Terminated 3 on November 29, 2010 |
Recall Number | Z-2595-2010 |
Recall Event ID |
56366 |
510(K)Number | K033089 K091107 K902307 |
Product Classification |
electrosurgical, cutting & coagulation & accessories - Product Code GEI
|
Product | HEMOSTATIX HANDLE, FOR USE WITH MODELS 150 AND 600D HEMOSTATIX CONTROLLERS, REF 7013-5050, QTY (1), STERILE EO, HEMOSTATIX MEDICAL TECHNOLOGIES, LLC, BARTLETT, TN 38133 USA, CE 0473, EU REP : QFI LTD, STRATFORD, LONDON E15 2ST UK
The Hemostatix Thermal Scalpel is a surgical instrument designed to retain the precise, clean cutting characteristics of the traditional steel scalpel while minimizing blood loss by simultaneously sealing blood vessels as they are cut, with minimum damage to surrounding tissue and virtually no muscle stimulation, using heat thermally conducted to the tissue from an elevated-temperature blade. |
Code Information |
LOTS: HMT0664, HMT0661, HMT0656 |
Recalling Firm/ Manufacturer |
Hemostatix Medical Technologies LLC 8400 Wolf Lake Dr Ste 109 Bartlett TN 38133-4189
|
For Additional Information Contact | Brad Beale 901-261-0012 |
Manufacturer Reason for Recall | Lack of sterility assurance |
FDA Determined Cause 2 | Other |
Action | On July 27, 2010, the firm, Hemostatix Medical Technologies, LLC, notified their international distributor in Japan of the recall by email. Hemostatix sent an "URGENT: MEDICAL DEVICE RECALL" letter dated July 29, 2010, to U.S. customers delivered via FedEx Priority Overnight. All non-responding customers were to be notified by telephone after 2 weeks. The other international customer, Italy, were to be notified by email on July 30, 2010. The letter describes the product, problem and action to be taken by customers. The customers were instructed to immediately check their inventory and remove any affected product found; ensure that satellite facilities/clinics receive the recall notice; complete and return the Recall Response Form via fax to Hemostatix's international fax number at 901-261-0017, and contact the Customer Care Department's recall hotline at 901-261-0012 for further instructions, if they do have affected stock.
If you have any questions or concerns, please feel free to contact our Customer Care Department at: 1-800-722-5735 in the USA or 1-901-261-0012 outside the USA. |
Quantity in Commerce | 219 units |
Distribution | Worldwide distribution: USA including CA, GA, LA, MA, MD, MO, NC, OH, OR, PA, SC, TX, VA, and WI; and countries of Japan and Italy. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = GEI 510(K)s with Product Code = GEI 510(K)s with Product Code = GEI
|
|
|
|