| Class 2 Device Recall Stryker Epic II Critical Care Bed | |
Date Initiated by Firm | May 14, 2010 |
Date Posted | September 28, 2010 |
Recall Status1 |
Terminated 3 on June 13, 2011 |
Recall Number | Z-2586-2010 |
Recall Event ID |
56507 |
Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
Product | Stryker Epic II Critical Care Bed, Model 2030/2031, Stryker Medical, Kalamazoo, MI.
Critical Care Patient Bed. |
Code Information |
040606711, 040606712, 040606713, 040606714, 040606715, 040606716, 040606717, 040606718, 040606719, 040606720, 040606721, 040606722, 041062545, 041062546, 081110024, 0301050530, 0301050531, 0301050532, 0301050533, 0301050534, 0301050535, 0301050536, 0301050537, 0301050538, 0301050539, 0301050540, 0301050541, 0301050542, 0301050543, 0301050544, 0301050545, 0301050546, 0301050547, 0301050548, 0301050549, 0301050550, 0301050551, 0301050552, 0301050553, 0301050555, 0301050556, 0301050557, 0301050558, 0301050559, 0301050560, 0301050561, 0301050562, 0301050563, 0301050564, 0301050565, 0301050566, 0301050567, 0301050568, 0301050569, 0301050570, 0301050571, 0301050572, 0301050573, 0301050592, 0302050501, 0302050510, 0302050511, 0302050512, 0302050513, 0302050514, 0302050515, 0302050516, 0302050519, 0302050520, 0302050521, 0302050522, 0302050523, 0302050527, 0302050562, 0302050563, 0302050564, 0302050565, 0302050566, 0302050567, 0302050576, 0302050577, 0302050578, 0302050579, 0302050580, 0302050581, 0302050596, 0302050597, 0302050598, 0302050599, 0302050600, 0302050601, 0302050602, 0302050605, 0302050606, 0302050607, 0302050608, 0302050609, 0302050610, 0302050611, 0302050612, 0302050613, 0302050614, 0302050615, 0302050616, 0302050617, 0302050618, 0302050619, 0302050621, 0302050622, 0302050623, 0302050624, 0302050625, 0302050626, 0302050627, 0302050628, 0302050629, 0302050630, 0302050631, 0302050632, 0302050633, 0302050634, 0302050635, 0302050636, 0302050657, 0302050676, 0302050677, 0302050678, 0302050679, 0302050680, 0302050681, 0302050682, 0302050683, 0302050684, 0303054501, 0303054502, 0303054503, 0303054504, 0303054505, 0303054506, 0303054507, 0303054508, 0303054509, 0303054510, 0303054511, 0303054512, 0303054513, 0303054514, 0303054515, 0303054516, 0303054517, 0303054518, 0303054519, 0303054520, 0303054534, 0303054535, 0303054536, 0303054558, 0303054565, 0303054566, 0303054567, 0303054568, 0303054569, 0303054570, 0303054571, 0303054572, 0303054573, 0303054574, 0303054575, 0303054576, 0303054577, 0303054578, 0303054580, 0303054581, 0303054582, 0303054583, 0303054584, 0303054585, 0303054604, 0303054605, 0303054606, 0303054607, 0303054608, 0303054609, 0303054643, 0303054644, 0303054645, 0303054646, 0303054647, 0303054648, 0303054649, 0303054678, 0303054679, 0303054680, 0303054681, 0303054682, 0303054683, 0303054684, 0303054685, 0303054686, 0303054687, 0304049001, 0304049002, 0304049003, 0304049004, 0304049005, 0304049006, 0304049007, 0304049008, 0304049009, 0304049010, 0304049011, 0304049012, 0304049013, 0304049015, 0304049016, 0304049017, 0304049018, 0304049019, 0304049020, 0304049021, 0304049022, 0304049023, 0304049024, 0304049025, 0304049026, 0304049027, 0304049028, 0304049029, 0304049030, 0304049031, 0304049032, 0304049033, 0304049034, 0304049035, 0304049036, 0304049037, 0304049038, 0304049039, 0304049040, 0304049041, 0304049042, 0304049081, 0304049082, 0304049083, 0304049084, 0304049085, 0304049086, 0304049087, 0304049088, 0304049089, 0304049090, 0304049091, 0304049092, 0304049093, 0304049094, 0304049095, 0304049096, 0304049097, 0304049098, 0304049099, 0304049100, 0304049101, 0304049102, 0304049103, 0304049104, 0304049105, 0304049106, 0304049107, 0304049108, 0304049109, 0304049110, 0304049111, 0304049112, 0304049113, 0304049114, 0304049115, 0304049116, 0304049117, 0304049118, 0304049119, 0304049134, 0304049137, 0304049138, 0304049139, 0304049140, 0304049141, 0304049142, 0304049143, 0304049144, 0304049145, 0304049146, 0304049147, 0304049148, 0304049149, 0304049150, 0304049151, 0304049156, 0304049168, 0304049169, 0304049170, 0304049171, 0305054501, 0305054502, 0305054503, 0305054504, 0305054505, 0305054506, 0305054507, 0305054508, 0305054509, 0305054510, 0305054511, 0305054512, 0305054513, 0305054514, 0305054515, 0305054516, 0305054517, 0305054538, 0305054539, 0305054540, 0305054541, 0305054542, 0305054543, 0305054544, 0305054545, 0305054546, 0305054547, 0305054548, 0305054549, 0305054550, 0305054551, 0305054552, 0305054553, 0305054554, 0305054555, 0305054556, 0305054557, 0305054558, 0305054559, 0305054560, 0305054561, 0305054562, 0305054563, 0305054564, 0305054565, 0305054566, 0305054567, 0305054568, 0305054569, 0305054579, 0305054580, 0305054581, 0305054582, 0305054583, 0305054584, 0305054600, 0305054601, 0305054602, 0305054603, 0305054604, 0305054605, 0305054606, 0305054607, 0305054608, 0305054609, 0305054610, 0305054611, 0305054612, 0305054613, 0305054614, 0305054615, 0305054616, 0305054617, 0305054659, 0305054660, 0305054661, 0305054662, 0305054663, 0305054669, 0305054670, 0305054676, 0305054677, 0305054678, 0305054679, 0306057501, 0306057502, 0306057503, 0306057504, 0306057505, 0306057506, 0306057507, 0306057508, 0306057509, 0306057510, 0306057511, 0306057512, 0306057513, 0306057514, 0306057515, 0306057516, 0306057517, 0306057518, 0306057519, 0306057520, 0306057521, 0306057522, 0306057523, 0306057524, 0306057525, 0306057526, 0306057527, 0306057528, 0306057529, 0306057530, 0306057531, 0306057532, 0306057533, 0306057534, 0306057535, 0306057536, 0306057537, 0306057538, 0306057539, 0306057540, 0306057541, 0306057542, 0306057543, 0306057544, 0306057545, 0306057546, 0306057547, 0306057563, 0306057564, 0306057565, 0306057566, 0306057567, 0306057568, 0306057569, 0306057570, 0306057571, 0306057572, 0306057573, 0306057574, 0306057575, 0306057576, 0306057577, 0306057578, 0306057579, 0306057580, 0306057581, 0306057582, 0306057583, 0306057584, 0306057585, 0306057586, 0306057587, 0306057588, 0306057589, 0306057590, 0306057591, 0306057592, 0306057593, 0306057594, 0306057595, 0306057618, 0306057619, 0306057620, 0306057621, 0306057622, 0306057623, 0306057624, 0306057625, 0306057626, 0306057627, 0306057628, 0306057629, 0306057630, 0306057631, 0306057632, 0306057633, 0306057635, 0306057636, 0306057637, 0306057638, 0306057639, 0306057640, 0306057641, 0306057642, 0306057643, 0306057644, 0306057645, 0306057646, 0306057647, 0306057648, 0306057649, 0306057650, 0306057651, 0306057652, 0306057653, 0306057654, 0306057655, 0306057656, 0306057657, 0306057658, 0306057659, 0306057660, 0306057661, 0306057662, 0306057663, 0306057664, 0306057665, 0306057666, 0306057667, 0306057668, 0306057669, 0306057670, 0306057671, 0306057672, 0306057673, 0306057674, 0306057675, 0306057676, 0306057677, 0306057678, 0306057679, 0306057680, 0306057681, 0306057682, 0306057683, 0306057684, 0306057685, 0306057686, 0306057687, 0306057688, 0306057689, 0306057690, 0306057691, 0306057692, 0306057693, 0306057694, 0306057695, 0306057696, 0306057697, 0306057698, 0306057699, 0306057700, 0306057701, 0306057702, 0306057703, 0306057704, 0306057705, 0306057706, 0306057707, 0306057708, 0306057709, 0306057710, 0306057711, 0306057714, 0306057727, 0306057728, 0306057729, 0306057730, 0306057731, 0306057732, 0306057733, 0306057734, 0306057735, 0306057751, 0307057002, 0307057003, 0307057004, 0307057005, 0307057006, 0307057007, 0307057008, 0307057009, 0307057010, 0307057011, 0307057012, 0307057013, 0307057014, 0307057015, 0307057016, 0307057017, 0307057018, 0307057019, 0307057020, 0307057021, 0307057022, 0307057023, 0307057024, 0307057025, 0307057026, 0307057027, 0307057028, 0307057029, 0307057030, 0307057031, 0307057032, 0307057033, 0307057034, 0307057035, 0307057036, 0307057037, 0307057038, 0307057039, 0307057040, 0307057089, 0307057090, 0307057091, 0307057092, 0307057093, 0307057094, 0307057095, 0307057101, 0307057102, 0307057149, 0307057150, 0307057151, 0307057152, 0307057153, 0307057154, 0307057155, 0307057156, 0307057157, 0307057158, 0307057159, 0307057160, 0307057161, 0307057162, 0307057163, 0307057182, 0307057186, 0307057187, 0307057188, 0307057190, 0307057191, 0307057193, 0307057194, 0307057195, 0307057196, 0307057197, 0308058032, 0308058033, 0308058034, 0308058035, 0308058036, 0308058037, 0308058038, 0308058039, 0308058040, 0308058041, 0308058042, 0308058045, 0308058046, 0308058047, 0308058048, 0308058049, 0308058050, 0308058051, 0308058052, 0308058053, 0308058054, 0308058055, 0308058056, 0308058057, 0308058058, 0308058059, 0308058060, 0308058061, 0308058062, 0308058063, 0308058064, 0308058065, 0308058066, 0308058067, 0308058068, 0308058069, 0308058070, 0308058071, 0308058073, 0308058074, 0308058075, 0308058076, 0308058102, 0308058103, 0308058104, 0308058130, 0308058132, 0308058133, 0308058134, 0308058135, 0308058136, 0308058137, 0309057518, 0309057519, 0309057543, 0309057544, 0309057545, 0309057546, 0309057557, 0309057558, 0309057559, 0309057560, 0309057561, 0309057562, 0309057563, 0309057564, 0309057565, 0309057566, 0309057567, 0309057568, 0309057569, 0309057570, 0309057571, 0309057572, 0309057573, 0309057574, 0309057606, 0309057607, 0309057608, 0309057609, 0309057610, 0309057611, 0309057612, 0309057613, 0309057614, 0309057615, 0309057616, 0309057617, 0309057618, 0309057619, 0309057620, 0309057621, 0309057622, 0309057623, 0309057624, 0309057654, 0310048501, 0310048502, 0310048503, 0310048512, 0310048513, 0310048514, 0310048515, 0310048516, 0310048517, 0310048518, 0310048519, 0310048520, 0310048521, 0310048522, 0310048523, 0310048524, 0310048525, 0310048526, 0310048527, 0310048528, 0310048529, 0310048530, 0310048531, 0310048532, 0310048533, 0310048534, 0310048535, 0310048536, 0310048537, 0310048538, 0310048539, 0310048540, 0310048541, 0310048559, 0310048563, 0310048564, 0310048565, 0310048566, 0310048567, 0310048568, 0310048569, 0310048570, 0310048571, 0310048572, 0310048573, 0310048574, 0310048575, 0310048576, 0310048577, 0310048578, 0310048579, 0310048580, 0310048581, 0310048582, 0310048583, 0310048584, 0310048608, 0310048609, 0310048610, 0310048611, 0310048612, 0310048613, 0310048614, 0310048615, 0310048616, 0310048617, 0310048618, 0310048619, 0310048620, 0310048621, 0310048622, 0310048623, 0310048624, 0310048625, 0310048626, 0310048627, 0310048628, 0310048629, 0310048630, 0310048631, 0310048634, 0310048635, 0310048636, 0310048637, 0310048638, 0310048639, 0310048640, 0310048641, 0310048642, 0310048643, 0310048644, 0310048645, 0310048646, 0310048647, 0310048648, 0310048649, 0310048650, 0310048651, 0310048652, 0310048682, 0310048683, 0310048686, 0310048687, 0311052501, 0311052502, 0311052503, 0311052504, 0311052505, 0311052506, 0311052507, 0311052508, 0311052509, 0311052510, 0311052511, 0311052512, 0311052513, 0311052514, 0311052515, 0311052516, 0311052517, 0311052518, 0311052519, 0311052520, 0311052521, 0311052522, 0311052523, 0311052524, 0311052525, 0311052526, 0311052527, 0311052528, 0311052529, 0311052530, 0311052531, 0311052559, 0311052560, 0311052561, 0311052562, 0311052563, 0311052564, 0311052565, 0311052568, 0311052569, 0311052570, 0311052571, 0311052572, 0311052573, 0311052574, 0311052575, 0311052576, 0311052577, 0311052628, 0311052629, 0311052630, 0311052631, 0311052632, 0311052633, 0311052634, 0311052635, 0311052636, 0311052637, 0311052638, 0311052639, 0311052640, 0311052641, 0311052642, 0311052643, 0311052644, 0311052645, 0311052646, 0311052647, 0311052648, 0311052649, 0311052650, 0311052651, 0311052652, 0311052653, 0311052654, 0311052655, 0311052656, 0311052657, 0311052658, 0311052659, 0311052660, 0311052661, 0311052662, 0311052663, 0311052664, 0311052665, 0311052666, 0311052667, 0311052669, 0311052670, 0311052671, 0311052672, 0311052673, 0311052674, 0311052675, 0311052692, 0311052693, 0311052694, 0311052695, 0312056517, 0312056518, 0312056519, 0312056520, 0312056622, 0312056625, 0312056626, 0312056627, 0312056628, 0312056629, 0312056630, 0312056631, 0312056632, 0312056633, 0312056634, 0312056635, 0312056636, 0312056637, 0312056638, 0312056639, 0312056640, 0312056641, 0312056642, 0312056643, 0312056644, 0312056645, 0312056646, 0312056650, 0312056651, 0312056652, 0312056653, 0312056654, 0312056655, 0312056656, 0312056657, 0312056658, 0312056659, 0312056660, 0312056661, 0312056662, 0312056663, 0312056664, 0312056665, 0312056682, 0312056683, 0312056684, 0312056685, 0312056686, 0312056687, 0312056688, 0312056689, 0312056690, 0312056691, 0312056692, 0312056693, 0312056694, 0312056695, 0312056696, 0312056697, 0312056698, 0312056721, 0312056722, 0312056730, 0312056731, 0312056732, 0312056733, 0312056734, 0312056735, 0401062014, 0401062015, 0401062016, 0401062017, 0401062018, 0401062019, 0401062020, 0401062021, 0401062022, 0401062023, 0401062024, 0401062025, 0401062026, 0401062027, 0401062028, 0401062029, 0401062030, 0401062031, 0401062032, 0401062033, 0401062034, 0401062035, 0401062036, 0401062037, 0401062038, 0401062039, 0401062040, 0401062041, 0401062042, 0401062043, 0401062044, 0401062045, 0401062046, 0401062047, 0401062048, 0401062049, 0401062050, 0401062051, 0401062065, 0401062066, 0401062067, 0401062068, 0401062069, 0401062070, 0401062071, 0401062072, 0401062073, 0401062074, 0401062075, 0401062076, 0401062077, 0401062078, 0401062079, 0401062080, 0401062081, 0401062102, 0401062103, 0401062104, 0401062105, 0401062106, 0401062107, 0401062108, 0401062109, 0401062110, 0401062111, 0401062112, 0401062113, 0401062114, 0401062115, 0401062116, 0401062117, 0401062118, 0401062119, 0401062120, 0401062121, 0401062122, 0401062123, 0401062124, 0401062125, 0401062153, 0401062154, 0401062155, 0401062156, 0401062157, 0401062158, 0401062159, 0401062160, 0401062161, 0401062162, 0401062163, 0401062164, 0401062165, 0401062166, 0401062167, 0401062168, 0401062169, 0401062170, 0401062171, 0401062172, 0401062173, 0401062174, 0401062175, 0401062176, 0401062177, 0401062178, 0401062179, 0401062180, 0401062181, 0401062182, 0401062183, 0401062184, 0401062185, 0401062186, 0401062187, 0401062188, 0401062189, 0401062190, 0401062191, 0401062192, 0401062193, 0401062194, 0401062195, 0401062196, 0401062197, 0401062198, 0401062199, 0401062200, 0401062201, 0401062202, 0401062203, 0401062204, 0401062205, 0401062206, 0401062207, 0401062208, 0401062209, 0401062210, 0401062211, 0401062212, 0401062213, 0401062214, 0401062215, 0401062216, 0401062217, 0401062218, 0401062219, 0401062220, 0401062221, 0401062222, 0401062223, 0401062224, 0401062225, 0401062226, 0401062227, 0401062228, 0401062229, 0401062230, 0401062231, 0401062240, 0401062241, 0402066014, 0402066015, 0402066019, 0402066020, 0402066021, 0402066022, 0402066023, 0402066024, 0402066025, 0402066026, 0402066027, 0402066028, 0402066029, 0402066030, 0402066031, 0402066032, 0402066033, 0402066034, 0402066035, 0402066036, 0402066037, 0402066038, 0402066039, 0402066040, 0402066041, 0402066044, 0402066045, 0402066046, 0402066047, 0402066048, 0402066049, 0402066050, 0402066051, 0402066052, 0402066053, 0402066054, 0402066055, 0402066056, 0402066057, 0402066058, 0402066059, 0402066060, 0402066061, 0402066062, 0402066063, 0402066064, 0402066065, 0402066066, 0402066067, 0402066068, 0402066069, 0402066070, 0402066071, 0402066072, 0402066073, 0402066079, 0402066080, 0402066081, 0402066082, 0402066083, 0402066084, 0402066085, 0402066086, 0402066087, 0402066088, 0402066096, 0402066097, 0402066106, 0402066107, 0402066108, 0402066109, 0402066110, 0402066111, 0402066112, 0402066113, 0402066114, 0402066115, 0402066116, 0402066118, 0402066134, 0402066179, 0402066180, 0402066181, 0402066182, 0402066183, 0402066184, 0402066185, 0402066186, 0402066187, 0402066188, 0402066189, 0402066190, 0402066191, 0402066192, 0402066193, 0402066201, 0402066202, 0402066203, 0402066204, 0402066205, 0402066206, 0402066207, 0402066208, 0402066209, 0402066210, 0402066211, 0402066212, 0402066226, 0402066240, 0402066241, 0402066242, 0402066243, 0402066244, 0402066245, 0402066246, 0402066247, 0402066248, 0402066249, 0402066250, 0403064513, 0403064521, 0403064522, 0403064523, 0403064524, 0403064525, 0403064526, 0403064527, 0403064528, 0403064529, 0403064530, 0403064531, 0403064532, 0403064533, 0403064534, 0403064535, 0403064536, 0403064537, 0403064538, 0403064539, 0403064540, 0403064541, 0403064542, 0403064543, 0403064544, 0403064545, 0403064546, 0403064547, 0403064548, 0403064549, 0403064550, 0403064551, 0403064552, 0403064553, 0403064554, 0403064555, 0403064560, 0403064561, 0403064562, 0403064563, 0403064564, 0403064565, 0403064566, 0403064567, 0403064568, 0403064572, 0403064573, 0403064575, 0403064576, 0403064577, 0403064578, 0403064588, 0403064589, 0403064590, 0403064591, 0403064592, 0403064593, 0403064594, 0403064595, 0403064596, 0403064597, 0403064598, 0403064599, 0403064600, 0403064601, 0403064617, 0403064618, 0403064624, 0403064625, 0403064626, 0403064627, 0403064628, 0403064629, 0403064630, 0403064631, 0403064632, 0403064633, 0403064634, 0403064635, 0403064636, 0403064637, 0403064638, 0403064639, 0403064640, 0403064641, 0403064642, 0403064643, 0403064644, 0403064645, 0403064646, 0403064647, 0403064648, 0403064649, 0403064650, 0403064651, 0403064652, 0403064653, 0403064657, 0403064669, 0403064692, 0403064696, 0403064697, 0403064702, 0403064703, 0403064704, 0403064705, 0403064706, 0403064707, 0403064708, 0403064709, 0403064710, 0403064711, 0403064712, 0403064713, 0403064714, 0403064715, 0403064716, 0403064717, 0403064718, 0403064719, 0403064720, 0403064721, 0403064722, 0403064723, 0403064724, 0403064725, 0403064726, 0403064727, 0403064728, 0403064729, 0403064730, 0403064731, 0403064732, 0403064733, 0403064734, 0403064735, 0403064736, 0403064737, 0403064738, 0403064739, 0403064740, 0403064741, 0403064742, 0403064743, 0403064744, 0403064745, 0403064746, 0403064747, 0403064748, 0403064749, 0403064750, 0403064751, 0403064752, 0403064753, 0403064758, 0403064763, 0403064764, 0403064765, 0403064766, 0403064767, 0403064768, 0403064769, 0403064770, 0403064773, 0403064774, 0403064775, 0403064776, 0403064777, 0403064778, 0403064779, 0403064780, 0403064781, 0403064782, 0403064783, 0403064784, 0403064785, 0403064786, 0403064787, 0403064788, 0403064789, 0403064790, 0403064792, 0403064797, 0403064798, 0403064799, 0403064800, 0403064801, 0403064802, 0403064803, 0403064804, 0403064805, 0403064823, 0403064824, 0403064825, 0403064826, 0403064827, 0403064828, 0403064829, 0403064830, 0403064831, 0403064832, 0403064833, 0403064834, 0403064835, 0403064836, 0403064837, 0403064838, 0403064839, 0403064840, 0403064841, 0403064842, 0403064843, 0403064844, 0403064845, 0403064846, 0403064898, 0403064899, 0403064900, 0403064901, 0403064903, 0403064909, 0403064910, 0403064911, 0403064912, 0403064926, 0403064928, 0403064930, 0403064931, 0403064932, 0403064933, 0403064934, 0403064935, 0403064936, 0403064937, 0403064938, 0403064939, 0403064940, 0403064945, 0403064946, 0403064947, 0403064948, 0403064949, 0403064954, 0403064955, 0405066002, 0405066003, 0405066004, 0405066005, 0405066006, 0405066007, 0405066008, 0405066009, 0405066010, 0405066011, 0405066012, 0405066013, 0405066014, 0405066015, 0405066016, 0405066074, 0405066077, 0405066078, 0405066079, 0405066085, 0405066086, 0405066087, 0405066088, 0405066089, 0405066090, 0405066091, 0405066092, 0405066093, 0405066094, 0405066095, 0405066096, 0405066097, 0405066098, 0405066099, 0405066100, 0405066101, 0405066108, 0405066112, 0405066113, 0405066114, 0405066115, 0405066116, 0405066117, 0405066118, 0405066119, 0405066120, 0405066121, 0405066122, 0405066123, 0405066124, 0405066125, 0405066126, 0405066127, 0405066128, 0405066129, 0405066130, 0405066131, 0405066132, 0405066133, 0405066134, 0405066135, 0405066173, 0405066174, 0405066176, 0405066177, 0405066178, 0405066182, 0405066183, 0405066184, 0405066185, 0405066186, 0405066187, 0405066188, 0405066189, 0405066190, 0405066191, 0405066192, 0405066193, 0405066194, 0405066195, 0405066196, 0405066197, 0405066198, 0405066199, 0405066212, 0405066213, 0405066218, 0405066219, 0405066220, 0405066221, 0405066222, 0405066223, 0405066224, 0405066225, 0405066232, 0405066233, 0405066239, 0405066273, 0405066274, 0405066275, 0405066276, 0405066277, 0405066278, 0405066279, 0405066280, 0405066281, 0405066282, 0405066283, 0405066284, 0405066285, 0405066286, 0405066287, 0405066288, 0405066289, 0405066290, 0405066291, 0405066292, 0405066293, 0405066294, 0405066295, 0405066296, 0405066297, 0405066298, 0405066299, 0405066300, 0405066301, 0405066302, 0405066303, 0405066304, 0405066305, 0405066347, 0406060502, 0406060503, 0406060504, 0406060505, 0406060506, 0406060507, 0406060508, 0406060509, 0406060510, 0406060511, 0406060512, 0406060513, 0406060514, 0406060515, 0406060516, 0406060520, 0406060521, 0406060523, 0406060524, 0406060525, 0406060526, 0406060527, 0406060528, 0406060529, 0406060530, 0406060531, 0406060532, 0406060533, 0406060534, 0406060535, 0406060536, 0406060537, 0406060538, 0406060539, 0406060540, 0406060541, 0406060542, 0406060543, 0406060544, 0406060545, 0406060546, 0406060547, 0406060548, 0406060549, 0406060550, 0406060551, 0406060552, 0406060553, 0406060554, 0406060555, 0406060556, 0406060557, 0406060558, 0406060559, 0406060560, 0406060561, 0406060562, 0406060563, 0406060564, 0406060565, 0406060566, 0406060567, 0406060568, 0406060569, 0406060570, 0406060571, 0406060572, 0406060573, 0406060574, 0406060575, 0406060576, 0406060577, 0406060578, 0406060579, 0406060580, 0406060581, 0406060582, 0406060583, 0406060584, 0406060585, 0406060586, 0406060587, 0406060588, 0406060589, 0406060590, 0406060591, 0406060592, 0406060593, 0406060594, 0406060595, 0406060596, 0406060597, 0406060598, 0406060599, 0406060600, 0406060601, 0406060602, 0406060609, 0406060610, 0406060630, 0406060631, 0406060632, 0406060633, 0406060634, 0406060635, 0406060636, 0406060637, 0406060638, 0406060639, 0406060640, 0406060641, 0406060642, 0406060643, 0406060661, 0406060663, 0406060664, 0406060665, 0406060666, 0406060667, 0406060668, 0406060669, 0406060670, 0406060671, 0406060672, 0406060673, 0406060674, 0406060675, 0406060676, 0406060677, 0406060678, 0406060679, 0406060680, 0406060681, 0406060682, 0406060683, 0406060684, 0406060685, 0406060686, 0406060687, 0406060688, 0406060689, 0406060690, 0406060691, 0406060692, 0406060693, 0406060694, 0406060695, 0406060696, 0406060697, 0406060698, 0406060699, 0406060700, 0406060701, 0406060702, 0406060703, 0406060704, 0406060705, 0406060706, 0406060707, 0406060708, 0406060709, 0406060710, 0406060711, 0406060712, 0406060713, 0406060714, 0406060715, 0406060716, 0406060717, 0406060718, 0406060719, 0406060720, 0406060721, 0406060722, 0407060501, 0407060502, 0407060503, 0407060504, 0407060505, 0407060506, 0407060507, 0407060508, 0407060509, 0407060511, 0407060514, 0407060515, 0407060516, 0407060517, 0407060518, 0407060519, 0407060521, 0407060522, 0407060523, 0407060524, 0407060527, 0407060528, 0407060529, 0407060540, 0407060541, 0407060542, 0407060543, 0407060544, 0407060545, 0407060546, 0407060547, 0407060548, 0407060549, 0407060550, 0407060551, 0407060552, 0407060553, 0407060554, 0407060555, 0407060556, 0407060557, 0407060558, 0407060559, 0407060560, 0407060561, 0407060562, 0407060563, 0407060564, 0407060565, 0407060566, 0407060567, 0407060568, 0407060569, 0407060570, 0407060571, 0407060572, 0407060573, 0407060574, 0407060575, 0407060576, 0407060577, 0407060578, 0407060579, 0407060580, 0407060581, 0407060582, 0407060583, 0407060584, 0407060585, 0407060586, 0407060587, 0407060588, 0407060589, 0407060590, 0407060591, 0407060592, 0407060593, 0407060594, 0407060595, 0407060596, 0407060597, 0407060598, 0407060599, 0407060600, 0407060601, 0407060602, 0407060603, 0407060604, 0407060605, 0407060606, 0407060607, 0407060608, 0407060609, 0407060610, 0407060611, 0407060612, 0407060613, 0407060614, 0407060615, 0407060616, 0407060617, 0407060618, 0407060619, 0407060620, 0407060621, 0407060622, 0407060623, 0407060624, 0407060625, 0407060626, 0407060627, 0407060628, 0407060629, 0407060630, 0407060631, 0407060632, 0407060633, 0407060634, 0407060635, 0407060636, 0407060637, 0407060638, 0407060639, 0407060640, 0407060641, 0407060642, 0407060643, 0407060644, 0407060645, 0407060646, 0407060647, 0407060648, 0407060649, 0407060650, 0407060651, 0407060652, 0407060653, 0407060654, 0407060655, 0407060656, 0407060657, 0407060658, 0407060659, 0407060660, 0408061505, 0408061515, 0408061516, 0408061517, 0408061518, 0408061519, 0408061520, 0408061521, 0408061522, 0408061523, 0408061524, 0408061525, 0408061526, 0408061527, 0408061528, 0408061529, 0408061530, 0408061531, 0408061532, 0408061533, 0408061534, 0408061535, 0408061536, 0408061537, 0408061538, 0408061539, 0408061540, 0408061541, 0408061542, 0408061543, 0408061544, 0408061545, 0408061548, 0408061549, 0408061550, 0408061551, 0408061552, 0408061561, 0408061562, 0408061563, 0408061564, 0408061565, 0408061566, 0408061567, 0408061568, 0408061569, 0408061604, 0408061605, 0408061606, 0408061607, 0408061608, 0408061609, 0408061610, 0408061611, 0408061612, 0408061613, 0408061614, 0408061615, 0408061616, 0408061617, 0408061618, 0408061619, 0408061620, 0408061621, 0408061622, 0408061623, 0408061624, 0408061625, 0408061626, 0408061627, 0408061628, 0408061629, 0408061630, 0408061631, 0408061632, 0408061633, 0408061634, 0408061635, 0408061636, 0408061637, 0408061638, 0408061639, 0408061640, 0408061641, 0408061642, 0408061643, 0408061644, 0408061645, 0408061646, 0408061647, 0408061648, 0408061649, 0408061650, 0408061651, 0408061652, 0408061653, 0408061654, 0408061655, 0408061656, 0408061657, 0408061661, 0408061662, 0408061663, 0408061664, 0408061668, 0408061669, 0408061670, 0408061671, 0408061672, 0408061673, 0408061694, 0408061695, 0408061696, 0408061697, 0408061698, 0408061699, 0408061700, 0408061709, 0408061713, 0408061714, 0408061715, 0408061716, 0408061717, 0408061718, 0408061719, 0408061720, 0408061721, 0408061722, 0408061723, 0408061724, 0408061728, 0408061729, 0408061730, 0408061731, 0408061732, 0408061733, 0408061734, 0408061735, 0408061736, 0408061737, 0408061738, 0408061739, 0408061740, 0408061741, 0408061742, 0408061743, 0408061744, 0408061745, 0408061746, 0408061747, 0408061748, 0408061749, 0408061750, 0408061751, 0408061752, 0408061753, 0408061754, 0408061755, 0408061756, 0408061757, 0408061758, 0408061759, 0409060501, 0409060502, 0409060503, 0409060504, 0409060505, 0409060506, 0409060507, 0409060508, 0409060509, 0409060510, 0409060511, 0409060512, 0409060513, 0409060514, 0409060515, 0409060520, 0409060521, 0409060522, 0409060523, 0409060524, 0409060545, 0409060549, 0409060550, 0409060560, 0409060561, 0409060562, 0409060563, 0409060564, 0409060565, 0409060600, 0409060610, 0409060611, 0409060612, 0409060613, 0409060614, 0409060615, 0409060616, 0409060617, 0409060618, 0409060619, 0409060620, 0409060621, 0409060659, 0409060660, 0409060661, 0409060662, 0409060663, 0409060664, 0409060665, 0409060666, 0409060667, 0409060668, 0409060669, 0409060670, 0409060671, 0409060672, 0409060673, 0409060674, 0409060675, 0409060676, 0409060677, 0409060679, 0409060680, 0409060681, 0409060682, 0409060683, 0409060684, 0409060685, 0409060686, 0409060687, 0409060688, 0409060689, 0409060690, 0409060691, 0409060692, 0409060693, 0409060694, 0409060695, 0409060696, 0409060697, 0409060698, 0409060699, 0409060700, 0409060701, 0409060702, 0409060703, 0409060704, 0409060705, 0409060706, 0409060707, 0409060708, 0409060709, 0409060710, 0409060711, 0409060712, 0409060713, 0409060714, 0409060715, 0409060716, 0409060717, 0409060718, 0409060719, 0409060720, 0409060721, 0409060722, 0409060723, 0409060724, 0409060726, 0409060727, 0409060728, 0409060729, 0409060730, 0409060731, 0409060732, 0409060733, 0409060734, 0409060735, 0409060736, 0409060737, 0409060738, 0409060739, 0409060740, 0409060741, 0409060742, 0409060743, 0409060744, 0409060745, 0409060746, 0409060747, 0409060751, 0409060752, 0409060753, 0409060754, 0409060755, 0409060756, 0409060757, 0409060758, 0409060759, 0409060760, 0409060761, 0409060762, 0409060763, 0409060764, 0409060765, 0409060766, 0409060767, 0410062501, 0410062502, 0410062503, 0410062504, 0410062505, 0410062506, 0410062507, 0410062508, 0410062509, 0410062510, 0410062511, 0410062526, 0410062527, 0410062528, 0410062541, 0410062542, 0410062543, 0410062544, 0410062545, 0410062546, 0410062547, 0410062548, 0410062549, 0410062550, 0410062551, 0410062552, 0410062553, 0410062554, 0410062555, 0410062556, 0410062557, 0410062558, 0410062559, 0410062560, 0410062561, 0410062562, 0410062563, 0410062564, 0410062565, 0410062566, 0410062567, 0410062568, 0410062569, 0410062570, 0410062571, 0410062572, 0410062573, 0410062574, 0410062575, 0410062576, 0410062577, 0410062578, 0410062579, 0410062580, 0410062581, 0410062582, 0410062592, 0410062593, 0410062594, 0410062595, 0410062596, 0410062597, 0410062598, 0410062599, 0410062600, 0410062601, 0410062602, 0410062603, 0410062604, 0410062605, 0410062606, 0410062607, 0410062608, 0410062609, 0410062610, 0410062611, 0410062612, 0410062613, 0410062614, 0410062615, 0410062616, 0410062619, 0410062620, 0410062621, 0410062622, 0410062623, 0410062625, 0410062628, 0410062629, 0410062630, 0410062631, 0410062632, 0410062633, 0410062634, 0410062635, 0410062636, 0410062637, 0410062639, 0410062643, 0410062664, 0410062666, 0410062667, 0410062668, 0410062669, 0410062670, 0410062671, 0410062672, 0410062673, 0410062674, 0410062675, 0410062676, 0410062677, 0410062720, 0410062721, 0410062722, 0410062723, 0410062724, 0410062725, 0411062025, 0411062026, 0411062027, 0411062041, 0411062046, 0411062056, 0411062057, 0411062058, 0411062059, 0411062061, 0411062062, 0411062063, 0411062064, 0411062065, 0411062066, 0411062067, 0411062068, 0411062069, 0411062070, 0411062071, 0411062072, 0411062079, 0411062080, 0411062081, 0411062082, 0411062083, 0411062084, 0411062085, 0411062086, 0411062087, 0411062088, 0411062089, 0411062090, 0411062135, 0411062136, 0411062137, 0411062138, 0411062139, 0411062140, 0411062141, 0411062142, 0411062143, 0411062144, 0411062145, 0411062146, 0411062147, 0411062148, 0411062149, 0411062150, 0411062151, 0411062152, 0411062153, 0411062154, 0411062155, 0411062156, 0411062157, 0411062158, 0411062159, 0411062160, 0411062161, 0411062162, 0411062163, 0411062164, 0411062165, 0411062166, 0411062167, 0411062168, 0411062169, 0411062170, 0411062171, 0411062172, 0411062173, 0411062174, 0411062175, 0411062176, 0411062177, 0411062178, 0411062179, 0411062180, 0411062181, 0411062182, 0411062183, 0411062184, 0411062185, 0411062186, 0411062187, 0411062188, 0411062189, 0411062190, 0411062191, 0411062192, 0411062193, 0411062194, 0411062195, 0411062196, 0411062197, 0411062198, 0411062199, 0411062200, 0411062201, 0411062202, 0411062203, 0411062204, 0411062211, 0411062214, 0411062225, 0411062226, 0411062227, 0411062228, 0411062229, 0411062230, 0411062231, 0411062232, 0411062233, 0411062234, 0411062235, 0411062268, 0411062269, 0411062270, 0411062271, 0411062272, 0411062273, 0411062274, 0411062275, 0411062276, 0411062277, 0411062278, 0411062279, 0411062280, 0411062282, 0411062283, 0411062284, 0411062285, 0411062286, 0411062287, 0411062288, 0411062289, 0411062290, 0411062291, 0411062292, 0411062293, 0411062294, 0411062295, 0411062296, 0411062297, 0411062298, 0411062299, 0411062300, 0411062301, 0411062302, 0411062303, 0411062304, 0411062305, 0412065001, 0412065002, 0412065003, 0412065004, 0412065012, 0412065013, 0412065014, 0412065015, 0412065016, 0412065040, 0412065041, 0412065042, 0412065043, 0412065044, 0412065045, 0412065046, 0412065047, 0412065048, 0412065049, 0412065050, 0412065051, 0412065052, 0412065053, 0412065054, 0412065055, 0412065056, 0412065057, 0412065058, 0412065059, 0412065060, 0412065061, 0412065062, 0412065063, 0412065101, 0412065102, 0412065103, 0412065104, 0412065105, 0412065106, 0412065107, 0412065108, 0412065109, 0412065110, 0412065111, 0412065112, 0412065113, 0412065114, 0412065115, 0412065116, 0412065117, 0412065118, 0412065119, 0412065120, 0412065121, 0412065122, 0412065123, 0412065124, 0412065125, 0412065126, 0412065127, 0412065128, 0412065129, 0412065130, 0412065131, 0412065132, 0412065133, 0412065134, 0412065135, 0412065136, 0412065137, 0412065138, 0412065139, 0412065140, 0412065141, 0412065142, 0412065143, 0412065144, 0412065145, 0412065146, 0412065147, 0412065148, 0412065149, 0412065150, 0412065151, 0412065152, 0412065153, 0412065154, 0412065155, 0412065156, 0412065157, 0412065158, 0412065159, 0412065160, 0412065161, 0412065162, 0412065163, 0412065164, 0412065165, 0412065166, 0412065167, 0412065168, 0412065169, 0412065170, 0412065171, 0412065172, 0412065173, 0412065174, 0412065175, 0412065176, 0412065177, 0412065178, 0412065179, 0412065180, 0412065181, 0412065182, 0412065183, 0412065184, 0412065185, 0412065186, 0412065187, 0412065188, 0412065189, 0412065190, 0412065191, 0412065192, 0412065193, 0412065194, 0412065195, 0412065196, 0412065197, 0412065198, 0412065199, 0412065200, 0412065201, 0412065202, 0412065203, 0412065204, 0412065205, 0412065206, 0412065207, 0412065208, 0412065213, 0412065215, 0412065216, 0412065239, 0412065240, 0412065241, 0412065242, 0412065243, 0412065244, 0412065245, 0412065246, 0412065247, 0412065248, 0412065249, 0412065250, 0412065251, 0412065252, 0412065253, 0412065254, 0412065255, 0412065256, 0412065257, 0412065258, 0412065259, 0412065260, 0412065261, 0412065262, 0412065263, 0412065264, 0412065265, 0412065266, 0412065267, 0412065268, 0412065269, 0412065270, 0412065271, 0412065272, 0412065273, 0412065274, 0412065275, 0412065276, 0412065277, 0412065278, 0501069521, 0501069522, 0501069523, 0501069524, 0501069525, 0501069542, 0501069543, 0501069544, 0501069545, 0501069546, 0501069547, 0501069548, 0501069549, 0501069550, 0501069551, 0501069552, 0501069553, 0501069554, 0501069555, 0501069556, 0501069557, 0501069558, 0501069559, 0501069560, 0501069561, 0501069562, 0501069563, 0501069564, 0501069565, 0501069566, 0501069567, 0501069568, 0501069569, 0501069570, 0501069571, 0501069587, 0501069588, 0501069589, 0501069590, 0501070028, 0501070029, 0501070030, 0501070031, 0501070032, 0501070033, 0501070034, 0501070035, 0501070036, 0501070037, 0501070502, 0501070503, 0501070504, 0501070505, 0501070506, 0501070507, 0501070508, 0501070509, 0501070510, 0501070511, 0501070512, 0501070513, 0501070514, 0501070515, 0501070516, 0501070517, 0501070518, 0501070519, 0501070520, 0501070521, 0501070522, 0501070523, 0501070524, 0501070525, 0501070526, 0501070527, 0501070528, 0501070529, 0501070530, 0501070531, 0501070532, 0501070533, 0501070534, 0501070535, 0501070536, 0501070537, 0501070538, 0501070539, 0501070540, 0501070541, 0501070542, 0501070543, 0501070544, 0501070545, 0501070546, 0501070547, 0501070548, 0501070549, 0501070550, 0501070551, 0501070552, 0501070553, 0501070554, 0501070555, 0501070556, 0501070557, 0501070558, 0501070559, 0501070560, 0501070561, 0501070562, 0501070563, 0501070570, 0501070571, 0501070572, 0501070573, 0501070574, 0501070575, 0501070576, 0501070577, 0501070578, 0501070579, 0501070615, 0501070616, 0501070617, 0501070618, 0501070619, 0501070620, 0501070621, 0501070622, 0501070623, 0501070624, 0501070625, 0501070626, 0501070627, 0501070654, 0501070655, 0501070661, 0501070662, 0501070663, 0501070664, 0501070665, 0501070666, 0501070667, 0501070668, 0501070669, 0501070670, 0501070671, 0501070672, 0501070697, 0501070698, 0501070699, 0501070700, 0501070701, 0501070702, 0501070703, 0502072501, 0502072502, 0502072503, 0502072504, 0502072505, 0502072506, 0502072507, 0502072508, 0502072509, 0502072510, 0502072511, 0502072512, 0502072513, 0502072514, 0502072515, 0502072516, 0502072517, 0502072518, 0502072519, 0502072520, 0502072539, 0502072540, 0502072541, 0502072542, 0502072543, 0502072544, 0502072545, 0502072546, 0502072547, 0502072548, 0502072549, 0502072550, 0502072551, 0502072552, 0502072553, 0502072554, 0502072555, 0502072556, 0502072557, 0502072558, 0502072559, 0502072560, 0502072561, 0502072562, 0502072563, 0502072564, 0502072565, 0502072566, 0502072567, 0502072568, 0502072569, 0502072570, 0502072571, 0502072572, 0502072573, 0502072574, 0502072575, 0502072576, 0502072630, 0502072631, 0502072632, 0502072633, 0502072634, 0502072643, 0502072644, 0502072645, 0502072646, 0502072647, 0502072648, 0502072649, 0502072650, 0502072651, 0502072652, 0502072653, 0502072654, 0502072655, 0502072656, 0502072657, 0502072658, 0502072659, 0502072660, 0502072661, 0502072662, 0502072663, 0502072664, 0502072665, 0502072666, 0502072667, 0502072668, 0502072669, 0502072670, 0502072671, 0502072672, 0502072673, 0502072674, 0502072675, 0502072676, 0502072677, 0502072678, 0502072679, 0502072680, 0502072681, 0502072682, 0502072683, 0502072684, 0502072685, 0502072686, 0502072687, 0502072688, 0502072689, 0502072690, 0502072691, 0502072692, 0502072693, 0502072694, 0502072695, 0502072696, 0502072697, 0502072704, 0502072721, 0502072722, 0502072723, 0502072725, 0502072726, 0502072727, 0502072728, 0502072729, 0502072730, 0502072731, 0502072732, 0502072733, 0502072743, 0502072768, 0502072769, 0502072770, 0502072771, 0502072772, 0502072773, 0502072774, 0502072775, 0502072776, 0502072777, 0502072778, 0502072779, 0502072780, 0502072781, 0502072782, 0502072783, 0502072785, 0502072798, 0502072799, 0502072800, 0502072801, 0502072802, 0502072803, 0502072804, 0502072805, 0502072817, 0502072818, 0502072819, 0502072842, 0502072843, 0502072844, 0502072845, 0502072846, 0502072847, 0502072848, 0502072849, 0502072850, 0502072851, 0502072852, 0502072853, 0502072854, 0502072855, 0502072856, 0502072857, 0502072858, 0502072859, 0502072860, 0502072861, 0502072862, 0502072863, 0502072864, 0502072865, 0502072866, 0502072867, 0502072868, 0502072869, 0502072870, 0502072871, 0502072872, 0502072873, 0502072874, 0502072875, 0502072876, 0502072877, 0502072878, 0502072879, 0502072880, 0502072881, 0503077003, 0503077004, 0503077005, 0503077006, 0503077007, 0503077008, 0503077009, 0503077010, 0503077011, 0503077012, 0503077013, 0503077014, 0503077015, 0503077016, 0503077017, 0503077018, 0503077019, 0503077020, 0503077021, 0503077062, 0503077063, 0503077064, 0503077065, 0503077066, 0503077067, 0503077068, 0503077069, 0503077070, 0503077071, 0503077072, 0503077073, 0503077074, 0503077075, 0503077076, 0503077077, 0503077078, 0503077079, 0503077080, 0503077081, 0503077082, 0503077083, 0503077084, 0503077085, 0503077086, 0503077087, 0503077088, 0503077089, 0503077090, 0503077091, 0503077092, 0503077093, 0503077094, 0503077095, 0503077096, 0503077097, 0503077101, 0503077102, 0503077103, 0503077104, 0503077105, 0503077106, 0503077107, 0503077108, 0503077109, 0503077110, 0503077111, 0503077112, 0503077113, 0503077114, 0503077115, 0503077116, 0503077117, 0503077118, 0503077131, 0503077132, 0503077133, 0503077134, 0503077135, 0503077136, 0503077137, 0503077138, 0503077139, 0503077140, 0503077141, 0503077142, 0503077143, 0503077144, 0503077145, 0503077146, 0503077147, 0503077148, 0503077149, 0503077150, 0503077151, 0503077152, 0503077153, 0503077154, 0503077279, 0503077280, 0503077281, 0503077282, 0503077283, 0503077284, 0503077285, 0503077286, 0503077287, 0503077288, 0503077289, 0503077307, 0503077310, 0503077312, 0503077313, 0503077314, 0503077315, 0503077316, 0503077318, 0503077319, 0503077320, 0503077333, 0503077334, 0503077335, 0503077336, 0503077337, 0503077338, 0503077339, 0503077340, 0503077341, 0503077342, 0503077343, 0503077344, 0503077345, 0503077346, 0503077347, 0503077348, 0503077349, 0503077350, 0503077351, 0503077352, 0503077353, 0503077385, 0503077386, 0503077387, 0503077388, 0503077389, 0503077390, 0503077391, 0503077392, 0503077409, 0503077410, 0503077411, 0503077412, 0503077413, 0503077414, 0503077415, 0503077416, 0503077417, 0503077418, 0504068545, 0504068546, 0504068547, 0504068548, 0504068549, 0504068550, 0504068551, 0504068552, 0504068556, 0504068557, 0504068558, 0504068559, 0504068560, 0504068561, 0504068562, 0504068563, 0504068564, 0504068566, 0504068567, 0504068568, 0504068569, 0504068570, 0504068571, 0504068572, 0504068573, 0504068574, 0504068575, 0504068576, 0504068577, 0504068578, 0504068579, 0504068581, 0504068582, 0504068583, 0504068584, 0504068605, 0504068606, 0504068607, 0504068608, 0504068609, 0504068610, 0504068611, 0504068612, 0504068613, 0504068614, 0504068615, 0504068616, 0504068617, 0504068618, 0504068619, 0504068620, 0504068621, 0504068622, 0504068623, 0504068624, 0504068625, 0504068628, 0504068629, 0504068630, 0504068631, 0504068632, 0504068633, 0504068634, 0504068648, 0504068649, 0504068650, 0504068651, 0504068652, 0504068653, 0504068654, 0504068655, 0504068656, 0504068657, 0504068658, 0504068659, 0504068660, 0504068661, 0504068662, 0504068663, 0504068664, 0504068665, 0504068666, 0504068667, 0504068668, 0504068669, 0504068670, 0504068671, 0504068672, 0504068673, 0504068674, 0504068675, 0504068676, 0504068677, 0504068756, 0504068757, 0504068758, 0504068797, 0504068798, 0504068799, 0504068800, 0504068801, 0504068802, 0504068803, 0504068804, 0504068805, 0504068806, 0504068807, 0504068808, 0504068809, 0504068810, 0504068811, 0504068812, 0504068813, 0504068814, 0504068815, 0504068816, 0504068817, 0504068818, 0504068819, 0504068820, 0504068821, 0504068822, 0504068823, 0504068824, 0504068825, 0504068826, 0504068827, 0504068828, 0504068829, 0504068830, 0504068831, 0504068832, 0504068833, 0504068834, 0504068835, 0504068836, 0504068837, 0504068838, 0504068839, 0504068840, 0504068841, 0504068842, 0504068843, 0504068844, 0504068845, 0504068846, 0504068847, 0504068848, 0504068849, 0505068501, 0505068502, 0505068503, 0505068504, 0505068505, 0505068506, 0505068507, 0505068508, 0505068509, 0505068510, 0505068511, 0505068512, 0505068513, 0505068514, 0505068515, 0505068516, 0505068517, 0505068518, 0505068535, 0505068547, 0505068548, 0505068549, 0505068550, 0505068551, 0505068552, 0505068553, 0505068554, 0505068583, 0505068584, 0505068585, 0505068586, 0505068587, 0505068588, 0505068589, 0505068590, 0505068591, 0505068600, 0505068601, 0505068602, 0505068603, 0505068604, 0505068605, 0505068606, 0505068607, 0505068608, 0505068609, 0505068611, 0505068612, 0505068613, 0505068614, 0505068615, 0505068616, 0505068617, 0505068618, 0505068619, 0505068620, 0505068621, 0505068623, 0505068624, 0505068625, 0505068626, 0505068632, 0505068633, 0505068634, 0505068635, 0505068636, 0505068637, 0505068638, 0505068639, 0505068640, 0505068641, 0505068642, 0505068643, 0505068700, 0505068701, 0505068702, 0505068705, 0505068706, 0505068707, 0505068708, 0505068709, 0505068710, 0505068711, 0505068722, 0505068723, 0505068724, 0505068725, 0505068726, 0505068727, 0505068728, 0505068729, 0505068745, 0505068746, 0505068747, 0505068748, 0505068749, 0505068750, 0505068751, 0505068785, 0505068787, 0505068788, 0505068789, 0505068790, 0505068791, 0505068792, 0505068793, 0505068794, 0505068795, 0505068796, 0505068797, 0505068799, 0505068800, 0505068801, 0505068802, 0505068811, 0505068812, 0505068813, 0505068814, 0505068815, 0505068816, 0505068817, 0505068818, 0505068819, 0505068820, 0505068822, 0505068823, 0505068824, 0505068825, 0505068826, 0505068827, 0505068828, 0505068829, 0505068830, 0505068831, 0505068832, 0505068833, 0505068834, 0505068835, 0505068836, 0505068840, 0505068841, 0505068842, 0505068843, 0505068844, 0505068845, 0505068846, 0505068847, 0505068848, 0505068849, 0505068850, 0505068851, 0505068852, 0505068853, 0505068854, 0505068912, 0505068913, 0505068914, 0505068915, 0505068916, 0505068917, 0505068918, 0505068919, 0505068920, 0505068921, 0505068922, 0505068923, 0505068924, 0506068530, 0506068531, 0506068532, 0506068533, 0506068534, 0506068535, 0506068536, 0506068537, 0506068538, 0506068539, 0506068540, 0506068541, 0506068542, 0506068543, 0506068544, 0506068545, 0506068546, 0506068547, 0506068548, 0506068549, 0506068550, 0506068551, 0506068552, 0506068553, 0506068554, 0506068555, 0506068556, 0506068557, 0506068615, 0506068616, 0506068617, 0506068618, 0506068619, 0506068620, 0506068621, 0506068637, 0506068638, 0506068639, 0506068640, 0506068641, 0506068642, 0506068644, 0506068645, 0506068646, 0506068647, 0506068648, 0506068649, 0506068650, 0506068651, 0506068652, 0506068653, 0506068654, 0506068655, 0506068656, 0506068657, 0506068685, 0506068688, 0506068689, 0506068690, 0506068691, 0506068692, 0506068693, 0506068714, 0506068748, 0506068749, 0506068750, 0506068751, 0506068752, 0506068753, 0506068754, 0506068755, 0506068756, 0506068777, 0506068778, 0506068779, 0506068780, 0506068781, 0506068802, 0506068803, 0506068863, 0506068865, 0506068866, 0506068867, 0506068868, 0506068869, 0506068870, 0506068871, 0506068872, 0506068873, 0506068874, 0506068875, 0506068876, 0506068877, 0506068878, 0506068879, 0506068880, 0506068881, 0506068882, 0506068883, 0506068884, 0506068885, 0506068886, 0506068887, 0506068888, 0506068889, 0507068502, 0507068503, 0507068504, 0507068505, 0507068506, 0507068507, 0507068508, 0507068509, 0507068510, 0507068511, 0507068512, 0507068513, 0507068514, 0507068515, 0507068516, 0507068517, 0507068518, 0507068519, 0507068520, 0507068521, 0507068522, 0507068523, 0507068524, 0507068525, 0507068526, 0507068527, 0507068528, 0507068529, 0507068530, 0507068531, 0507068532, 0507068533, 0507068534, 0507068535, 0507068536, 0507068537, 0507068538, 0507068564, 0507068565, 0507068566, 0507068567, 0507068568, 0507068569, 0507068570, 0507068571, 0507068572, 0507068573, 0507068574, 0507068575, 0507068576, 0507068577, 0507068578, 0507068579, 0507068580, 0507068581, 0507068582, 0507068583, 0507068584, 0507068585, 0507068586, 0507068587, 0507068588, 0507068589, 0507068590, 0507068591, 0507068592, 0507068593, 0507068594, 0507068595, 0507068596, 0507068597, 0507068598, 0507068599, 0507068600, 0507068601, 0507068602, 0507068603, 0507068604, 0507068605, 0507068606, 0507068608, 0507068637, 0507068691, 0507068692, 0507068693, 0507068694, 0507068695, 0507068696, 0507068697, 0507068698, 0507068716, 0507068717, 0507068718, 0507068719, 0507068720, 0507068721, 0507068722, 0507068723, 0507068724, 0507068725, 0507068726, 0507068727, 0507068728, 0507068729, 0507068730, 0507068731, 0507068732, 0507068733, 0507068734, 0507068735, 0507068736, 0507068737, 0507068738, 0507068739, 0507068740, 0507068741, 0507068742, 0507068743, 0507068744, 0507068745, 0507068746, 0507068747, 0507068748, 0507068749, 0507068750, 0507068751, 0507068792, 0507068793, 0507068794, 0507068795, 0507068806, 0507068807, 0507068808, 0507068809, 0507068810, 0507068811, 0507068812, 0507068813, 0507068814, 0507068815, 0507068816, 0507068817, 0507068818, 0507068819, 0507068820, 0507068821, 0507068822, 0507068827, 0507068828, 0507068829, 0507068830, 0507068831, 0507068832, 0507068833, 0507068834, 0507068835, 0507068836, 0508068542, 0508068543, 0508068544, 0508068545, 0508068546, 0508068547, 0508068548, 0508068549, 0508068550, 0508068551, 0508068552, 0508068553, 0508068554, 0508068555, 0508068556, 0508068557, 0508068558, 0508068559, 0508068563, 0508068564, 0508068567, 0508068568, 0508068569, 0508068570, 0508068571, 0508068610, 0508068611, 0508068612, 0508068613, 0508068645, 0508068646, 0508068647, 0508068648, 0508068649, 0508068650, 0508068651, 0508068652, 0508068653, 0508068654, 0508068655, 0508068656, 0508068657, 0508068658, 0508068659, 0508068660, 0508068687, 0508068688, 0508086543, 0508086544, 0508086545, 0508086546, 0508086547, 0508086548, 0508086549, 0508086550, 0508086551, 0508086552, 0508086553, 0508086554, 0508086555, 0508086556, 0508086557, 0508086558, 0508086559, 0508086560, 0508086561, 0508086562, 0508086563, 0508086564, 0508086565, 0508086566, 0508086617, 0508086618, 0508086619, 0508086620, 0508086621, 0508086622, 0508086623, 0508086624, 0508086625, 0508086626, 0508086627, 0508086628, 0508086629, 0508086630, 0508086631, 0508086632, 0508086633, 0508086634, 0508086639, 0508086640, 0508086641, 0508086642, 0508086643, 0508086644, 0508086645, 0508086646, 0508086647, 0508086648, 0508086649, 0508086650, 0508086651, 0508086652, 0508086653, 0508086654, 0508086655, 0508086656, 0508086657, 0508086658, 0508086659, 0508086660, 0508086661, 0508086662, 0508086663, 0508086665, 0508086666, 0508086667, 0508086668, 0508086669, 0508086670, 0508086673, 0508086674, 0508086677, 0508086678, 0508086679, 0508086680, 0508086681, 0508086682, 0508086683, 0508086684, 0509073501, 0509073502, 0509073503, 0509073504, 0509073505, 0509073580, 0509073581, 0509073582, 0509073583, 0509073584, 0509073585, 0509073586, 0509073587, 0509073588, 0509073631, 0509073632, 0509073633, 0509073634, 0509073676, 0509073677, 0509073678, 0509073679, 0509073680, 0509073681, 0509073682, 0509073683, 0509073684, 0509073685, 0509073686, 0509073687, 0509073763, 0509073764, 0509073765, 0509073766, 0509073767, 0509073768, 0509073769, 0509073770, 0509073771, 0509073772, 0509073773, 0509073774, 0509073775, 0509073776, 0509073777, 0509073778, 0509073779, 0509073780, 0509073781, 0509073782, 0509073783, 0509073784, 0509073785, 0509073786, 0509073789, 0509073790, 0509073791, 0509073792, 0509073793, 0509073794, 0509073795, 0509073796, 0509073797, 0509073798, 0509073799, 0509073800, 0509073808, 0509073809, 0509073816, 0509073817, 0509073818, 0509073819, 0509073820, 0509073821, 0509073855, 0509073856, 0509073860, 0509073861, 0509073862, 0509073863, 0509073864, 0509073865, 0509073866, 0509073867, 0509073868, 0509073869, 0509073870, 0509073871, 0509073872, 0509073873, 0509073874, 0509073875, 0509073876, 0509073877, 0509073878, 0509073881, 0509073905, 0509073906, 0509073907, 0509073908, 0509073909, 0509073910, 0509073911, 0509073912, 0509073913, 0509073914, 0509073915, 0509073916, 0509073917, 0509073918, 0509073919, 0509073920, 0509073921, 0509073922, 0509073923, 0509073924, 0509073925, 0509073926, 0509073929, 0509073930, 0510077501, 0510077502, 0510077503, 0510077504, 0510077505, 0510077506, 0510077507, 0510077508, 0510077509, 0510077510, 0510077511, 0510077512, 0510077513, 0510077514, 0510077515, 0510077516, 0510077517, 0510077518, 0510077519, 0510077520, 0510077521, 0510077522, 0510077523, 0510077524, 0510077525, 0510077526, 0510077527, 0510077528, 0510077529, 0510077530, 0510077531, 0510077532, 0510077533, 0510077534, 0510077535, 0510077536, 0510077612, 0510077615, 0510077616, 0510077617, 0510077618, 0510077619, 0510077620, 0510077621, 0510077622, 0510077623, 0510077624, 0510077625, 0510077626, 0510077728, 0510077729, 0510077730, 0510077731, 0510077732, 0510077733, 0510077734, 0510077735, 0510077736, 0510077737, 0510077738, 0510077739, 0510077740, 0510077741, 0510077742, 0510077743, 0510077744, 0510077745, 0510077746, 0510077747, 0510077748, 0510077749, 0510077750, 0510077751, 0510077752, 0510077753, 0510077754, 0510077755, 0510077756, 0510077757, 0510077758, 0510077759, 0510077760, 0510077761, 0510077762, 0510077763, 0510077764, 0510077765, 0510077766, 0510077767, 0510077768, 0510077769, 0510077770, 0510077771, 0510077772, 0511068522, 0511068524, 0511068525, 0511068532, 0511068533, 0511068534, 0511068535, 0511068536, 0511068537, 0511068538, 0511068539, 0511068540, 0511068543, 0511068549, 0511068550, 0511068551, 0511068552, 0511068553, 0511068554, 0511068555, 0511068556, 0511068557, 0511068558, 0511068559, 0511068560, 0511068561, 0511068562, 0511068563, 0511068564, 0511068565, 0511068566, 0511068567, 0511068571, 0511068624, 0511068625, 0511068626, 0511068627, 0511068628, 0511068629, 0511068630, 0511068631, 0511068632, 0511068633, 0511068634, 0511068635, 0511068637, 0511068638, 0511068639, 0511068640, 0511068641, 0511068643, 0511068644, 0511068645, 0511068646, 0511068647, 0511068648, 0511068649, 0511068650, 0511068651, 0511068652, 0511068653, 0511068654, 0511068655, 0511068656, 0511068658, 0511068675, 0511068676, 0511068677, 0511068678, 0511068679, 0511068680, 0511068681, 0511068682, 0511068683, 0511068684, 0511068685, 0511068686, 0511068687, 0511068688, 0511068689, 0511068690, 0511068691, 0511068692, 0511068693, 0511068694, 0511068695, 0511068697, 0511068698, 0511068699, 0511068700, 0511068701, 0511068702, 0511068709, 0511068710, 0511068712, 0511068715, 0511068716, 0511068717, 0511068718, 0511068719, 0511068720, 0511068721, 0511068722, 0511068735, 0511068736, 0511068743, 0511068744, 0511068745, 0511068746, 0511068747, 0511068748, 0511068749, 0511068750, 0511068751, 0511068807, 0511068808, 0511068809, 0511068810, 0511068811, 0511068812, 0511068813, 0511068814, 0511068815, 0511068816, 0511068817, 0511068818, 0511068819, 0511068820, 0511068825, 0511068831, 0511068832, 0511068833, 0511068834, 0511068840, 0511068841, 0511068842, 0511068843, 0511068844, 0511068845, 0511068846, 0511068847, 0511068848, 0511068849, 0511068850, 0511068851, 0511068852, 0511068853, 0511068862, 0511068863, 0511068864, 0511068865, 0511068866, 0511068867, 0511068868, 0511068869, 0511068870, 0511068871, 0511068872, 0511068873, 0511068874, 0511068881, 0511068882, 0511068883, 0511068884, 0511068885, 0511068886, 0511068887, 0511068888, 0511068914, 0511068915, 0511068916, 0511068917, 0512068519, 0512068520, 0512068521, 0512068522, 0512068523, 0512068524, 0512068525, 0512068526, 0512068527, 0512068528, 0512068529, 0512068530, 0512068531, 0512068532, 0512068533, 0512068534, 0512068535, 0512068536, 0512068553, 0512068554, 0512068555, 0512068556, 0512068557, 0512068558, 0512068559, 0512068560, 0512068561, 0512068562, 0512068563, 0512068564, 0512068565, 0512068566, 0512068569, 0512068570, 0512068571, 0512068572, 0512068573, 0512068574, 0512068575, 0512068576, 0512068577, 0512068578, 0512068579, 0512068585, 0512068586, 0512068587, 0512068588, 0512068589, 0512068609, 0512068610, 0512068611, 0512068612, 0512068613, 0512068614, 0512068615, 0512068616, 0512068617, 0512068618, 0512068619, 0512068620, 0512068621, 0512068624, 0512068625, 0512068626, 0512068627, 0512068628, 0512068629, 0512068630, 0512068631, 0512068632, 0512068633, 0512068634, 0512068635, 0512068636, 0512068637, 0512068638, 0512068639, 0512068640, 0512068641, 0512068642, 0512068643, 0512068644, 0512068645, 0512068646, 0512068647, 0512068648, 0512068649, 0512068650, 0512068651, 0512068652, 0512068653, 0512068654, 0512068655, 0512068656, 0512068657, 0512068658, 0512068659, 0512068660, 0512068661, 0512068662, 0512068667, 0512068668, 0512068669, 0512068670, 0512068709, 0512068710, 0512068711, 0512068712, 0512068713, 0512068714, 0512068715, 0512068716, 0512068717, 0512068718, 0512068719, 0512068720, 0512068721, 0512068722, 0512068723, 0512068724, 0512068725, 0512068726, 0512068727, 0512068728, 0512068740, 0512068750, 0512068751, 0512068779, 0512068780, 0512068781, 0512068782, 0512068783, 0512068784, 0512068785, 0512068786, 0512068787, 0512068788, 0512068789, 0512068790, 0512068791, 0512068792, 0512068793, 0512068794, 0512068795, 0601085001, 0601085002, 0601085003, 0601085004, 0601085005, 0601085006, 0601085007, 0601085008, 0601085009, 0601085010, 0601085011, 0601085012, 0601085013, 0601085014, 0601085015, 0601085016, 0601085017, 0601085052, 0601085053, 0601085054, 0601085055, 0601085056, 0601085057, 0601085058, 0601085059, 0601085060, 0601085061, 0601085062, 0601085063, 0601085065, 0601085066, 0601085067, 0601085068, 0601085069, 0601085070, 0601085071, 0601085085, 0601085086, 0601085087, 0601085088, 0601085089, 0601085090, 0601085091, 0601085092, 0601085093, 0601085094, 0601085095, 0601085096, 0601085097, 0601085098, 0601085099, 0601085103, 0601085104, 0601085105, 0601085106, 0601085107, 0601085108, 0601085109, 0601085110, 0601085111, 0601085112, 0601085113, 0601085114, 0601085115, 0601085116, 0601085119, 0601085127, 0601085128, 0601085129, 0601085130, 0601085131, 0601085132, 0601085133, 0601085134, 0601085173, 0601085174, 0601085175, 0601085176, 0601085177, 0601085178, 0601085179, 0601085180, 0601085181, 0601085182, 0601085183, 0601085184, 0601085185, 0601085186, 0601085187, 0601085188, 0601085189, 0601085190, 0601085191, 0601085192, 0601085193, 0601085194, 0601085215, 0601085216, 0601085217, 0601085218, 0601085219, 0601085220, 0601085221, 0601085222, 0601085223, 0601085224, 0601085225, 0601085226, 0601085227, 0601085228, 0601085229, 0601085230, 0601085238, 0601085239, 0601085240, 0601085241, 0601085242, 0601085243, 0601085244, 0601085245, 0601085246, 0601085247, 0601085248, 0601085249, 0601085250, 0601085251, 0601085255, 0601085256, 0601085257, 0601085258, 0601085297, 0601085300, 0601085302, 0601085308, 0601085309, 0601085310, 0601085311, 0601085312, 0601085313, 0601085314, 0601085315, 0601085316, 0601085317, 0601085318, 0601085319, 0601085321, 0601085324, 0601085325, 0601085326, 0601085327, 0601085328, 0601085329, 0601085330, 0601085331, 0601085332, 0602078501, 0602078502, 0602078503, 0602078509, 0602078510, 0602078511, 0602078512, 0602078513, 0602078514, 0602078515, 0602078516, 0602078517, 0602078518, 0602078520, 0602078521, 0602078522, 0602078533, 0602078534, 0602078535, 0602078536, 0602078537, 0602078538, 0602078539, 0602078540, 0602078542, 0602078543, 0602078544, 0602078545, 0602078546, 0602078547, 0602078548, 0602078549, 0602078553, 0602078554, 0602078555, 0602078556, 0602078557, 0602078558, 0602078559, 0602078560, 0602078561, 0602078562, 0602078563, 0602078564, 0602078565, 0602078566, 0602078567, 0602078568, 0602078569, 0602078570, 0602078571, 0602078572, 0602078573, 0602078574, 0602078575, 0602078576, 0602078577, 0602078578, 0602078579, 0602078580, 0602078581, 0602078582, 0602078583, 0602078584, 0602078585, 0602078586, 0602078587, 0602078588, 0602078589, 0602078590, 0602078591, 0602078592, 0602078593, 0602078594, 0602078595, 0602078596, 0602078597, 0602078598, 0602078599, 0602078635, 0602078636, 0602078637, 0602078638, 0602078639, 0602078640, 0602078641, 0602078643, 0602078644, 0602078645, 0602078646, 0602078647, 0602078648, 0602078649, 0602078650, 0602078651, 0602078652, 0602078653, 0602078654, 0602078655, 0602078656, 0602078657, 0602078658, 0602078659, 0602078660, 0602078661, 0602078662, 0602078663, 0602078665, 0602078676, 0602078677, 0602078678, 0602078679, 0602078680, 0602078681, 0602078682, 0602078683, 0602078684, 0602078685, 0602078686, 0602078687, 0602078688, 0602078717, 0602078718, 0602078719, 0602078720, 0602078721, 0602078722, 0602078723, 0602078724, 0602078725, 0602078726, 0602078746, 0602078747, 0602078748, 0602078749, 0602078750, 0602078751, 0602078752, 0602078753, 0602078754, 0602078755, 0602078756, 0602078757, 0602078758, 0602078759, 0602078760, 0602078781, 0603083507, 0603083508, 0603083509, 0603083510, 0603083511, 0603083512, 0603083513, 0603083529, 0603083530, 0603083531, 0603083532, 0603083533, 0603083534, 0603083535, 0603083536, 0603083537, 0603083538, 0603083627, 0603083628, 0603083629, 0603083630, 0603083631, 0603083632, 0603083633, 0603083676, 0603083677, 0603083678, 0603083679, 0603083680, 0603083681, 0603083682, 0603083683, 0603083684, 0603083685, 0603083687, 0603083688, 0603083689, 0603083690, 0603083691, 0603083692, 0603083699, 0603083700, 0603083701, 0603083702, 0603083703, 0603083704, 0603083705, 0603083706, 0603083707, 0603083708, 0603083709, 0603083710, 0603083711, 0603083712, 0603083713, 0603083714, 0603083715, 0603083716, 0603083717, 0603083718, 0603083722, 0603083723, 0603083724, 0603083725, 0603083748, 0603083749, 0603083750, 0603083751, 0603083752, 0603083753, 0603083754, 0603083755, 0603083756, 0603083757, 0603083758, 0603083759, 0603083762, 0603083763, 0603083764, 0603083765, 0603083766, 0603083767, 0603083768, 0603083769, 0603083770, 0603083771, 0603083772, 0603083773, 0603083774, 0603083775, 0603083776, 0603083777, 0603083778, 0603083779, 0603083780, 0603083781, 0603083794, 0603083795, 0603083797, 0603083798, 0603083821, 0603083828, 0603083829, 0603083830, 0603083831, 0603083832, 0603083833, 0603083834, 0603083835, 0603083836, 0603083837, 0603083838, 0603083839, 0603083849, 0603083850, 0603083854, 0603083855, 0603083856, 0603083857, 0603083858, 0603083859, 0603083860, 0603083861, 0603083862, 0603083863, 0603083864, 0603083865, 0603083868, 0603083917, 0604080515, 0604080517, 0604080518, 0604080519, 0604080520, 0604080521, 0604080522, 0604080523, 0604080524, 0604080525, 0604080526, 0604080527, 0604080528, 0604080540, 0604080541, 0604080542, 0604080543, 0604080544, 0604080545, 0604080546, 0604080549, 0604080550, 0604080551, 0604080552, 0604080553, 0604080554, 0604080555, 0604080556, 0604080557, 0604080558, 0604080559, 0604080560, 0604080561, 0604080562, 0604080563, 0604080564, 0604080565, 0604080566, 0604080567, 0604080568, 0604080569, 0604080570, 0604080571, 0604080572, 0604080573, 0604080574, 0604080575, 0604080576, 0604080577, 0604080578, 0604080579, 0604080587, 0604080607, 0604080609, 0604080610, 0604080611, 0604080612, 0604080613, 0604080614, 0604080615, 0604080616, 0604080617, 0604080618, 0604080619, 0604080620, 0604080621, 0604080622, 0604080623, 0604080624, 0604080625, 0604080626, 0604080627, 0604080628, 0604080629, 0604080630, 0604080631, 0604080632, 0604080633, 0604080634, 0604080635, 0604080636, 0604080637, 0604080638, 0604080639, 0604080723, 0604080724, 0604080725, 0604080726, 0604080727, 0604080751, 0604080752, 0604080753, 0604080754, 0604080755, 0604080756, 0604080757, 0604080758, 0604080759, 0604080781, 0604080782, 0604080783, 0604080784, 0604080785, 0604080786, 0604080787, 0604080788, 0604080789, 0604080790, 0604080791, 0604080793, 0604080794, 0604080795, 0604080796, 0604080797, 0604080798, 0604080799, 0604080800, 0604080801, 0604080802, 0604080803, 0604080804, 0604080805, 0604080806, 0604080807, 0604080808, 0604080819, 0604080820, 0604080821, 0604080822, 0604080823, 0604080824, 0604080825, 0604080826, 0604080827, 0604080828, 0604080829, 0604080830, 0604080831, 0604080832, 0604080833, 0604080854, 0605085007, 0605085008, 0605085009, 0605085010, 0605085014, 0605085015, 0605085016, 0605085017, 0605085018, 0605085019, 0605085020, 0605085021, 0605085022, 0605085023, 0605085024, 0605085025, 0605085026, 0605085027, 0605085028, 0605085029, 0605085030, 0605085031, 0605085032, 0605085033, 0605085034, 0605085035, 0605085036, 0605085037, 0605085038, 0605085039, 0605085040, 0605085041, 0605085042, 0605085043, 0605085044, 0605085045, 0605085046, 0605085047, 0605085048, 0605085049, 0605085050, 0605085051, 0605085052, 0605085053, 0605085054, 0605085056, 0605085057, 0605085058, 0605085059, 0605085060, 0605085061, 0605085062, 0605085063, 0605085064, 0605085065, 0605085066, 0605085067, 0605085110, 0605085111, 0605085112, 0605085113, 0605085114, 0605085115, 0605085116, 0605085117, 0605085118, 0605085119, 0605085129, 0605085130, 0605085131, 0605085132, 0605085133, 0605085134, 0605085135, 0605085136, 0605085172, 0605085201, 0605085202, 0605085203, 0605085204, 0605085205, 0605085206, 0605085207, 0605085208, 0605085209, 0605085210, 0605085211, 0605085212, 0605085216, 0605085217, 0605085218, 0605085219, 0605085220, 0605085221, 0605085222, 0605085223, 0605085224, 0605085225, 0605085226, 0605085227, 0605085228, 0605085229, 0605085230, 0605085231, 0605085232, 0605085238, 0605085254, 0605085257, 0605085258, 0605085259, 0605085260, 0605085261, 0605085267, 0605085268, 0605085269, 0605085270, 0605085271, 0605085272, 0605085273, 0605085302, 0605085303, 0605085304, 0605085305, 0605085306, 0605085307, 0605085308, 0605085309, 0605085310, 0605085311, 0605085312, 0605085313, 0605085314, 0605085316, 0605085317, 0605085323, 0605085324, 0605085325, 0605085326, 0605085327, 0605085328, 0605085329, 0605085330, 0605085382, 0605085383, 0605085384, 0605085385, 0605085386, 0605085387, 0605085388, 0605085389, 0605085390, 0606079501, 0606079508, 0606079509, 0606079510, 0606079511, 0606079512, 0606079513, 0606079514, 0606079515, 0606079516, 0606079517, 0606079518, 0606079550, 0606079551, 0606079552, 0606079553, 0606079554, 0606079555, 0606079556, 0606079557, 0606079558, 0606079559, 0606079560, 0606079561, 0606079562, 0606079563, 0606079564, 0606079565, 0606079566, 0606079567, 0606079568, 0606079569, 0606079582, 0606079583, 0606079584, 0606079585, 0606079586, 0606079587, 0606079588, 0606079589, 0606079590, 0606079591, 0606079592, 0606079593, 0606079594, 0606079595, 0606079596, 0606079597, 0606079598, 0606079599, 0606079600, 0606079601, 0606079602, 0606079603, 0606079604, 0606079605, 0606079606, 0606079607, 0606079608, 0606079609, 0606079610, 0606079613, 0606079643, 0606079644, 0606079645, 0606079646, 0606079647, 0606079648, 0606079649, 0606079650, 0606079651, 0606079652, 0606079653, 0606079654, 0606079655, 0606079656, 0606079657, 0606079658, 0606079659, 0606079660, 0606079661, 0606079662, 0606079663, 0606079664, 0606079665, 0606079666, 0606079667, 0606079674, 0606079675, 0606079676, 0606079677, 0606079678, 0606079679, 0606079680, 0606079681, 0606079682, 0606079683, 0606079684, 0606079685, 0606079686, 0606079687, 0606079688, 0606079689, 0606079690, 0606079691, 0606079692, 0606079693, 0606079702, 0606079703, 0606079704, 0606079705, 0606079706, 0606079707, 0606079716, 0606079717, 0606079718, 0606079719, 0606079720, 0606079721, 0606079722, 0606079723, 0606079724, 0606079725, 0606079726, 0606079727, 0606079728, 0606079729, 0606079730, 0606079731, 0606079754, 0606079755, 0606079764, 0606079765, 0606079766, 0606079767, 0606079769, 0606079770, 0606079816, 0606079817, 0606079818, 0606079819, 0606079820, 0606079821, 0606079822, 0606079823, 0606079824, 0606079825, 0606079826, 0606079827, 0606079828, 0606079829, 0606079830, 0606079831, 0606079832, 0606079833, 0606079834, 0606079835, 0606079836, 0606079837, 0606079838, 0606079839, 0606079840, 0606079841, 0606079842, 0606079843, 0606079844, 0606079845, 0606079846, 0606079847, 0606079848, 0606079849, 0606079850, 0606079851, 0606079852, 0606079853, 0606079854, 0606079855, 0606079856, 0607084522, 0607084523, 0607084524, 0607084525, 0607084526, 0607084527, 0607084528, 0607084529, 0607084530, 0607084531, 0607084532, 0607084533, 0607084534, 0607084535, 0607084536, 0607084537, 0607084538, 0607084539, 0607084540, 0607084541, 0607084549, 0607084550, 0607084551, 0607084552, 0607084553, 0607084554, 0607084555, 0607084556, 0607084557, 0607084558, 0607084559, 0607084560, 0607084563, 0607084564, 0607084565, 0607084566, 0607084567, 0607084568, 0607084569, 0607084570, 0607084571, 0607084572, 0607084573, 0607084574, 0607084575, 0607084577, 0607084578, 0607084579, 0607084580, 0607084581, 0607084582, 0607084583, 0607084584, 0607084585, 0607084586, 0607084587, 0607084588, 0607084596, 0607084597, 0607084618, 0607084619, 0607084620, 0607084621, 0607084622, 0607084623, 0607084624, 0607084625, 0607084627, 0607084628, 0607084629, 0607084630, 0607084631, 0607084632, 0607084633, 0607084634, 0607084635, 0607084636, 0607084637, 0607084638, 0607084639, 0607084640, 0607084641, 0607084642, 0607084643, 0607084644, 0607084645, 0607084646, 0607084647, 0607084684, 0607084685, 0607084686, 0607084687, 0607084688, 0607084689, 0607084690, 0607084691, 0607084692, 0607084693, 0607084694, 0607084695, 0607084696, 0607084697, 0607084698, 0607084699, 0607084700, 0607084701, 0607084702, 0607084703, 0607084704, 0607084705, 0607084714, 0607084715, 0607084716, 0607084717, 0607084718, 0607084719, 0607084763, 0608079001, 0608079002, 0608079003, 0608079004, 0608079005, 0608079006, 0608079007, 0608079008, 0608079009, 0608079010, 0608079011, 0608079012, 0608079013, 0608079014, 0608079015, 0608079016, 0608079017, 0608079018, 0608079019, 0608079020, 0608079021, 0608079022, 0608079023, 0608079024, 0608079025, 0608079026, 0608079027, 0608079028, 0608079029, 0608079030, 0608079031, 0608079032, 0608079033, 0608079034, 0608079036, 0608079082, 0608079083, 0608079084, 0608079089, 0608079090, 0608079091, 0608079092, 0608079093, 0608079094, 0608079095, 0608079096, 0608079097, 0608079098, 0608079099, 0608079100, 0608079101, 0608079102, 0608079103, 0608079104, 0608079105, 0608079106, 0608079107, 0608079108, 0608079131, 0608079132, 0608079133, 0608079134, 0608079135, 0608079136, 0608079137, 0608079138, 0608079139, 0608079140, 0608079141, 0608079142, 0608079143, 0608079144, 0608079145, 0608079146, 0608079147, 0608079148, 0608079149, 0608079150, 0608079151, 0608079152, 0608079153, 0608079154, 0608079155, 0608079156, 0608079157, 0608079158, 0608079159, 0608079160, 0608079161, 0608079162, 0608079163, 0608079172, 0608079173, 0608079174, 0608079175, 0608079176, 0608079177, 0608079178, 0608079179, 0608079180, 0608079181, 0608079182, 0608079183, 0608079184, 0608079185, 0608079186, 0608079187, 0608079188, 0608079189, 0608079190, 0608079191, 0608079192, 0608079193, 0608079194, 0608079195, 0608079196, 0608079197, 0608079198, 0608079199, 0608079200, 0608079201, 0608079202, 0608079203, 0608079204, 0608079205, 0608079206, 0608079207, 0608079208, 0608079209, 0608079210, 0608079211, 0608079212, 0608079238, 0608079239, 0608079240, 0608079260, 0608079261, 0608079266, 0608079267, 0608079268, 0608079272, 0608079273, 0608079274, 0608079275, 0608079276, 0608079277, 0608079283, 0608079287, 0608079288, 0608079289, 0608079290, 0608079291, 0608079292, 0608079297, 0608079298, 0608079299, 0608079300, 0608079301, 0608079302, 0608079303, 0608079304, 0608079305, 0608079306, 0608079307, 0608079324, 0608079325, 0608079326, 0608079339, 0608079340, 0608079343, 0608079344, 0608079345, 0608079346, 0608079347, 0608079348, 0608079349, 0608079350, 0608079351, 0608079352, 0608079353, 0608079354, 0608079355, 0608079356, 0608079357, 0608079398, 0608079399, 0608079400, 0608079401, 0608079402, 0608079403, 0608079404, 0608079405, 0608079406, 0608079407, 0608079408, 0608079409, 0608079410, 0608079411, 0608079412, 0608079413, 0608079414, 0608079415, 0608079416, 0608079417, 0609081014, 0609081015, 0609081016, 0609081017, 0609081018, 0609081019, 0609081020, 0609081021, 0609081022, 0609081023, 0609081025, 0609081026, 0609081027, 0609081028, 0609081029, 0609081030, 0609081031, 0609081032, 0609081033, 0609081034, 0609081035, 0609081036, 0609081037, 0609081038, 0609081039, 0609081040, 0609081041, 0609081042, 0609081043, 0609081065, 0609081066, 0609081067, 0609081068, 0609081069, 0609081070, 0609081071, 0609081072, 0609081073, 0609081074, 0609081075, 0609081076, 0609081085, 0609081088, 0609081089, 0609081090, 0609081100, 0609081101, 0609081102, 0609081103, 0609081104, 0609081105, 0609081106, 0609081107, 0609081108, 0609081109, 0609081110, 0609081111, 0609081112, 0609081113, 0609081114, 0609081115, 0609081116, 0609081117, 0609081118, 0609081119, 0609081120, 0609081121, 0609081122, 0609081123, 0609081124, 0609081125, 0609081126, 0609081127, 0609081128, 0609081129, 0609081130, 0609081131, 0609081199, 0609081200, 0609081201, 0609081202, 0609081203, 0609081204, 0609081205, 0609081206, 0609081207, 0609081208, 0609081209, 0609081210, 0609081211, 0609081212, 0609081213, 0609081214, 0609081215, 0609081216, 0609081217, 0609081218, 0609081219, 0609081220, 0609081221, 0609081222, 0609081223, 0609081224, 0609081225, 0609081226, 0609081227, 0609081228, 0609081229, 0609081230, 0609081231, 0609081232, 0609081233, 0609081234, 0609081235, 0609081240, 0609081241, 0609081242, 0609081243, 0609081263, 0609081264, 0609081265, 0609081266, 0609081267, 0609081268, 0609081269, 0609081270, 0609081271, 0609081272, 0609081273, 0609081274, 0609081275, 0609081276, 0609081277, 0609081278, 0609081279, 0609081280, 0609081281, 0609081282, 0609081283, 0609081284, 0609081285, 0609081286, 0609081287, 0609081288, 0609081289, 0609081290, 0609081291, 0609081292, 0609081293, 0609081294, 0609081295, 0609081296, 0609081297, 0609081298, 0609081299, 0609081300, 0609081301, 0609081302, 0609081303, 0609081304, 0609081305, 0609081306, 0609081307, 0609081308, 0609081309, 0609081310, 0609081311, 0609081312, 0609081313, 0609081314, 0609081315, 0609081316, 0609081317, 0609081318, 0609081319, 0609081320, 0609081321, 0609081322, 0609081323, 0609081324, 0609081325, 0609081326, 0609081327, 0609081328, 0610078510, 0610078511, 0610078512, 0610078513, 0610078514, 0610078515, 0610078516, 0610078517, 0610078518, 0610078519, 0610078520, 0610078521, 0610078522, 0610078523, 0610078524, 0610078526, 0610078532, 0610078533, 0610078534, 0610078535, 0610078536, 0610078537, 0610078538, 0610078586, 0610078599, 0610078600, 0610078622, 0610078623, 0610078624, 0610078625, 0610078626, 0610078627, 0610078628, 0610078629, 0610078630, 0610078631, 0610078632, 0610078633, 0610078634, 0610078635, 0610078636, 0610078637, 0610078638, 0610078639, 0610078669, 0610078670, 0610078671, 0610078683, 0610078684, 0610078685, 0610078686, 0610078687, 0610078688, 0610078689, 0610078690, 0610078691, 0610078692, 0610078693, 0610078694, 0610078695, 0610078696, 0610078697, 0610078698, 0610078699, 0610078700, 0610078701, 0610078702, 0610078703, 0610078704, 0610078705, 0610078706, 0610078707, 0610078708, 0610078709, 0610078710, 0610078711, 0610078712, 0610078713, 0610078714, 0610078715, 0610078716, 0610078717, 0610078719, 0610078722, 0610078728, 0610078729, 0610078730, 0610078731, 0610078741, 0610078742, 0610078743, 0610078744, 0610078745, 0610078746, 0610078747, 0610078748, 0610078749, 0610078750, 0610078751, 0610078752, 0610078753, 0610078754, 0610078755, 0610078756, 0610078757, 0610078758, 0610078759, 0610078760, 0610078799, 0610078800, 0610078801, 0610078802, 0610078806, 0610078807, 0610078832, 0610078833, 0610078834, 0610078835, 0610078836, 0610078837, 0610078838, 0610078840, 0610078841, 0610078842, 0610078843, 0610078856, 0610078857, 0610078858, 0610078859, 0610078860, 0610078861, 0610078862, 0611082001, 0611082002, 0611082003, 0611082004, 0611082005, 0611082006, 0611082007, 0611082008, 0611082009, 0611082010, 0611082011, 0611082012, 0611082013, 0611082014, 0611082015, 0611082016, 0611082017, 0611082018, 0611082019, 0611082020, 0611082021, 0611082022, 0611082023, 0611082053, 0611082054, 0611082055, 0611082056, 0611082057, 0611082058, 0611082059, 0611082060, 0611082061, 0611082062, 0611082063, 0611082064, 0611082065, 0611082066, 0611082067, 0611082068, 0611082069, 0611082070, 0611082071, 0611082072, 0611082103, 0611082104, 0611082105, 0611082106, 0611082107, 0611082108, 0611082109, 0611082110, 0611082111, 0611082112, 0611082113, 0611082114, 0611082115, 0611082116, 0611082117, 0611082118, 0611082119, 0611082120, 0611082121, 0611082122, 0611082123, 0611082124, 0611082125, 0611082126, 0611082127, 0611082128, 0611082129, 0611082130, 0611082131, 0611082132, 0611082133, 0611082134, 0611082135, 0611082136, 0611082137, 0611082138, 0611082139, 0611082140, 0611082141, 0611082174, 0611082175, 0611082176, 0611082177, 0611082178, 0611082179, 0611082180, 0611082181, 0611082182, 0611082183, 0611082184, 0611082185, 0611082186, 0611082187, 0611082188, 0611082189, 0611082190, 0611082191, 0611082192, 0611082193, 0611082194, 0611082195, 0611082196, 0611082197, 0611082198, 0611082199, 0611082200, 0611082201, 0611082202, 0611082203, 0611082204, 0611082205, 0611082206, 0611082214, 0611082215, 0611082216, 0611082217, 0611082218, 0611082219, 0611082265, 0611082266, 0611082267, 0611082268, 0611082269, 0611082270, 0611082271, 0611082272, 0611082273, 0611082274, 0611082275, 0611082276, 0611082277, 0611082278, 0611082279, 0611082280, 0611082281, 0611082282, 0611082283, 0611082302, 0611082303, 0611082304, 0611082305, 0611082306, 0611082307, 0611082308, 0611082309, 0611082310, 0611082311, 0611082312, 0611082313, 0611082314, 0611082315, 0611082316, 0611082356, 0611082357, 0611082358, 0611082359, 0611082360, 0611082361, 0611082362, 0611082363, 0611082364, 0611082365, 0611082366, 0611082367, 0611082368, 0611082369, 0611082370, 0612080029, 0612080030, 0612080031, 0612080032, 0612080036, 0612080037, 0612080038, 0612080039, 0612080040, 0612080041, 0612080042, 0612080043, 0612080044, 0612080045, 0612080046, 0612080047, 0612080048, 0612080049, 0612080050, 0612080051, 0612080052, 0612080053, 0612080054, 0612080055, 0612080056, 0612080057, 0612080058, 0612080059, 0612080060, 0612080061, 0612080062, 0612080063, 0612080064, 0612080065, 0612080066, 0612080067, 0612080068, 0612080069, 0612080070, 0612080071, 0612080131, 0612080132, 0612080133, 0612080134, 0612080135, 0612080136, 0612080137, 0612080138, 0612080159, 0612080160, 0612080161, 0612080162, 0612080163, 0612080164, 0612080165, 0612080166, 0612080167, 0612080168, 0612080169, 0612080170, 0612080171, 0612080172, 0612080173, 0612080174, 0612080175, 0612080176, 0612080177, 0612080178, 0612080179, 0612080180, 0612080181, 0612080182, 0612080183, 0612080184, 0612080185, 0612080186, 0612080187, 0612080188, 0612080189, 0612080190, 0612080191, 0612080192, 0612080193, 0612080194, 0612080195, 0612080196, 0612080197, 0612080201, 0612080202, 0612080203, 0612080204, 0612080205, 0612080206, 0612080207, 0612080208, 0612080209, 0612080210, 0612080211, 0612080212, 0612080213, 0612080214, 0612080215, 0612080216, 0612080217, 0612080218, 0612080219, 0612080220, 0612080221, 0612080222, 0612080223, 0612080224, 0612080225, 0612080226, 0612080227, 0612080228, 0612080229, 0612080230, 0612080231, 0612080232, 0612080233, 0612080253, 0612080254, 0612080255, 0612080256, 0612080257, 0612080258, 0612080259, 0612080260, 0612080261, 0612080262, 0612080263, 0612080264, 0612080265, 0612080273, 0612080274, 0612080275, 0612080276, 0612080277, 0612080278, 0612080279, 0612080280, 0612080281, 0612080282, 0612080283, 0612080284, 0612080285, 0612080286, 0612080287, 0612080288, 0612080289, 0612080290, 0612080291, 0612080292, 0612080293, 0612080294, 0612080295, 0612080296, 0612080310, 0612080311, 0612080312, 0612080313, 0612080314, 0612080315, 0612080316, 0612080317, 0612080318, 0612080319, 0701096501, 0701096502, 0701096503, 0701096504, 0701096505, 0701096506, 0701096507, 0701096508, 0701096509, 0701096510, 0701096511, 0701096512, 0701096513, 0701096514, 0701096515, 0701096516, 0701096517, 0701096518, 0701096519, 0701096520, 0701096521, 0701096522, 0701096523, 0701096524, 0701096525, 0701096526, 0701096527, 0701096528, 0701096529, 0701096530, 0701096531, 0701096532, 0701096533, 0701096534, 0701096535, 0701096536, 0701096537, 0701096538, 0701096539, 0701096540, 0701096541, 0701096542, 0701096543, 0701096544, 0701096545, 0701096546, 0701096547, 0701096548, 0701096549, 0701096555, 0701096556, 0701096557, 0701096558, 0701096559, 0701096560, 0701096577, 0701096578, 0701096579, 0701096580, 0701096581, 0701096582, 0701096583, 0701096584, 0701096585, 0701096586, 0701096587, 0701096640, 0701096641, 0701096642, 0701096652, 0701096653, 0701096654, 0701096661, 0701096679, 0701096680, 0701096691, 0701096692, 0701096693, 0701096694, 0701096695, 0701096696, 0701096697, 0701096698, 0701096699, 0701096700, 0701096701, 0701096702, 0701096703, 0701096704, 0701096705, 0701096706, 0701096707, 0701096708, 0701096709, 0701096710, 0701096711, 0701096712, 0701096713, 0701096714, 0701096715, 0701096716, 0701096717, 0701096718, 0701096719, 0701096720, 0701096721, 0701096722, 0701096723, 0701096746, 0701096747, 0701096748, 0701096749, 0701096750, 0701096751, 0701096752, 0701096753, 0701096754, 0701096755, 0701096756, 0701096757, 0701096758, 0701096759, 0701096760, 0701096761, 0701096762, 0701096763, 0701096764, 0701096765, 0701096766, 0701096767, 0701096768, 0701096769, 0701096770, 0701096771, 0701096772, 0701096773, 0701096774, 0701096781, 0701096782, 0701096783, 0701096784, 0701096785, 0701096786, 0701096787, 0701096788, 0701096789, 0701096790, 0701096791, 0701096794, 0701096795, 0701096796, 0701096797, 0701096798, 0701096799, 0701096800, 0701096801, 0701096802, 0701096803, 0701096804, 0701096805, 0701096806, 0701096807, 0701096818, 0701096819, 0701096820, 0701096821, 0701096822, 0701096823, 0701096824, 0701096825, 0701096826, 0701096827, 0701096828, 0701096829, 0701096830, 0701096835, 0701096836, 0701096837, 0702088501, 0702088502, 0702088503, 0702088504, 0702088508, 0702088509, 0702088510, 0702088511, 0702088512, 0702088513, 0702088514, 0702088515, 0702088516, 0702088517, 0702088518, 0702088519, 0702088520, 0702088521, 0702088522, 0702088523, 0702088551, 0702088552, 0702088553, 0702088554, 0702088555, 0702088556, 0702088557, 0702088558, 0702088559, 0702088560, 0702088561, 0702088562, 0702088563, 0702088564, 0702088565, 0702088569, 0702088575, 0702088607, 0702088612, 0702088613, 0702088614, 0702088615, 0702088616, 0702088617, 0702088618, 0702088619, 0702088620, 0702088621, 0702088622, 0702088623, 0702088624, 0702088625, 0702088626, 0702088627, 0702088628, 0702088629, 0702088630, 0702088631, 0702088632, 0702088633, 0702088634, 0702088635, 0702088636, 0702088637, 0702088638, 0702088639, 0702088640, 0702088641, 0702088675, 0702088676, 0702088677, 0702088678, 0702088679, 0702088680, 0702088681, 0702088682, 0702088683, 0702088684, 0702088685, 0702088686, 0702088687, 0702088689, 0703090005, 0703090006, 0703090007, 0703090008, 0703090009, 0703090065, 0703090072, 0703090073, 0703090074, 0703090075, 0703090076, 0703090077, 0703090078, 0703090079, 0703090085, 0703090086, 0703090087, 0703090088, 0703090089, 0703090090, 0703090091, 0703090092, 0703090093, 0703090094, 0703090095, 0703090096, 0703090097, 0703090098, 0703090133, 0703090134, 0703090135, 0703090136, 0703090137, 0703090138, 0703090139, 0703090140, 0703090141, 0703090142, 0703090143, 0703090144, 0703090145, 0703090146, 0703090148, 0703090149, 0703090175, 0703090176, 0703090177, 0703090178, 0703090179, 0703090180, 0703090181, 0703090182, 0703090183, 0703090184, 0703090185, 0703090186, 0703090187, 0703090188, 0703090189, 0703090190, 0703090191, 0703090192, 0703090193, 0703090194, 0703090195, 0703090196, 0703090197, 0703090214, 0703090215, 0703090216, 0703090217, 0703090218, 0703090219, 0703090220, 0703090221, 0703090222, 0703090223, 0703090226, 0703090227, 0703090228, 0703090229, 0703090230, 0703090231, 0703090232, 0703090233, 0703090234, 0703090235, 0703090236, 0703090237, 0703090289, 0703090290, 0703090291, 0703090292, 0703090293, 0703090294, 0704092515, 0704092516, 0704092517, 0704092522, 0704092523, 0704092524, 0704092525, 0704092550, 0704092551, 0704092583, 0704092585, 0704092586, 0704092587, 0704092588, 0704092589, 0704092590, 0704092591, 0704092592, 0704092593, 0704092594, 0704092595, 0704092596, 0704092597, 0704092598, 0704092599, 0704092600, 0704092601, 0704092602, 0704092603, 0704092604, 0704092626, 0704092627, 0704092643, 0704092648, 0704092649, 0704092650, 0704092651, 0704092652, 0704092653, 0704092670, 0704092671, 0704092672, 0704092673, 0704092698, 0704092699, 0704092700, 0704092701, 0704092702, 0704092703, 0705096513, 0705096514, 0705096515, 0705096516, 0705096517, 0705096518, 0705096519, 0705096520, 0705096521, 0705096522, 0705096523, 0705096524, 0705096525, 0705096526, 0705096538, 0705096546, 0705096547, 0705096548, 0705096549, 0705096568, 0705096569, 0705096570, 0705096571, 0705096572, 0705096573, 0705096574, 0705096575, 0705096576, 0705096577, 0705096578, 0705096579, 0705096580, 0705096581, 0705096582, 0705096600, 0705096601, 0705096602, 0705096603, 0705096604, 0705096605, 0705096606, 0705096607, 0705096608, 0705096609, 0705096610, 0705096611, 0705096612, 0705096613, 0705096614, 0705096615, 0705096636, 0705096637, 0705096638, 0705096639, 0705096640, 0705096641, 0705096642, 0705096650, 0705096651, 0705096653, 0705096654, 0705096655, 0705096656, 0705096657, 0705096658, 0705096659, 0705096660, 0705096661, 0705096662, 0705096663, 0705096670, 0705096671, 0705096672, 0705096690, 0705096691, 0705096692, 0705096693, 0705096694, 0705096695, 0705096696, 0705096697, 0705096698, 0705096699, 0705096700, 0705096701, 0705096702, 0705096703, 0705096704, 0705096705, 0705096706, 0705096707, 0705096708, 0705096736, 0705096737, 0705096738, 0705096739, 0705096740, 0705096741, 0705096742, 0705096745, 0705096746, 0705096747, 0705096748, 0705096749, 0705096750, 0705096756, 0705096757, 0706088517, 0706088518, 0706088546, 0706088547, 0706088548, 0706088549, 0706088550, 0706088551, 0706088552, 0706088553, 0706088554, 0706088555, 0706088592, 0706088593, 0706088594, 0706088595, 0706088596, 0706088597, 0706088598, 0706088599, 0706088604, 0706088605, 0706088606, 0706088607, 0706088608, 0706088609, 0706088610, 0706088611, 0706088612, 0706088613, 0706088626, 0706088627, 0706088628, 0706088629, 0706088630, 0706088631, 0706088632, 0706088633, 0706088634, 0706088635, 0706088636, 0706088637, 0706088638, 0706088639, 0706088640, 0706088641, 0706088642, 0706088643, 0706088644, 0706088645, 0706088646, 0706088647, 0706088648, 0706088649, 0706088650, 0706088651, 0706088652, 0706088653, 0706088654, 0706088655, 0706088656, 0706088657, 0706088658, 0706088659, 0706088660, 0706088698, 0706088699, 0706088700, 0706088701, 0706088702, 0706088703, 0706088704, 0706088705, 0706088723, 0706088724, 0706088725, 0706088726, 0706088727, 0706088728, 0706088729, 0706088730, 0706088731, 0706088732, 0706088733, 0706088738, 0706096501, 0706096502, 0706096503, 0706096504, 0706096505, 0706096506, 0706096507, 0706096508, 0706096509, 0706096510, 0707091501, 0707091502, 0707092519, 0707092520, 0707092521, 0707092522, 0707092523, 0707092550, 0707092567, 0707092568, 0707092632, 0707092633, 0707092634, 0707092635, 0707092641, 0707092676, 0707092677, 0707092678, 0707092679, 0707092680, 0707092681, 0707092682, 0707092683, 0707092684, 0707092685, 0707092686, 0707092687, 0707092688, 0707092707, 0707092708, 0707092709, 0707092710, 0708094012, 0708094013, 0708094014, 0708094015, 0708094016, 0708094017, 0708094018, 0708094019, 0708094020, 0708094021, 0708094022, 0708094023, 0708094032, 0708094033, 0708094036, 0708094037, 0708094038, 0708094039, 0708094055, 0708094056, 0708094088, 0708094107, 0708094108, 0708094109, 0708094110, 0708094111, 0708094112, 0708094113, 0708094114, 0708094115, 0708094116, 0708094117, 0708094118, 0708094119, 0708094120, 0708094121, 0708094122, 0708094123, 0708094124, 0708094125, 0708094126, 0708094127, 0708094128, 0708094129, 0708094130, 0708094167, 0708094168, 0708094169, 0708094170, 0708094171, 0708094172, 0708094173, 0708094174, 0708094175, 0708094176, 0708094177, 0708094178, 0708094179, 0708094205, 0708094206, 0708094207, 0708094208, 0708094209, 0708094210, 0708094211, 0708094212, 0708094213, 0708094214, 0708094215, 0708094216, 0708094217, 0708094218, 0708094219, 0708094220, 0708094221, 0708094222, 0708094223, 0708094224, 0708094228, 0708094229, 0708094230, 0708094231, 0708094236, 0708094237, 0708094238, 0708094239, 0708094240, 0708094241, 0708094242, 0708094243, 0708094244, 0708094245, 0708094246, 0708094247, 0708094248, 0708094249, 0708094250, 0708094259, 0708094260, 0708094261, 0708094262, 0708094263, 0708094264, 0708094265, 0708094266, 0708094267, 0708094268, 0708094269, 0708094270, 0708094271, 0708094272, 0708094273, 0708094274, 0708094275, 0708094276, 0708094277, 0709089021, 0709089022, 0709089024, 0709089025, 0709089026, 0709089027, 0709089038, 0709089039, 0709089040, 0709089041, 0709089042, 0709089043, 0709089044, 0709089045, 0709089046, 0709089047, 0709089049, 0709089050, 0709089051, 0709089052, 0709089053, 0709089054, 0709089055, 0709089056, 0709089057, 0709089058, 0709089059, 0709089060, 0709089061, 0709089062, 0709089063, 0709089064, 0709089065, 0709089066, 0709089067, 0709089068, 0709089069, 0709089070, 0709089071, 0709089072, 0709089073, 0709089074, 0709089075, 0709089076, 0709089077, 0709089078, 0709089079, 0709089091, 0709089092, 0709089093, 0709089094, 0709089110, 0709089111, 0709089112, 0709089113, 0709089114, 0709089115, 0709089116, 0709089117, 0709089118, 0709089119, 0709089195, 0709089209, 0709089210, 0709089211, 0709089212, 0709089213, 0709089214, 0710098004, 0710098005, 0710098006, 0710098007, 0710098008, 0710098009, 0710098010, 0710098011, 0710098012, 0710098013, 0710098014, 0710098015, 0710098016, 0710098017, 0710098018, 0710098019, 0710098020, 0710098021, 0710098022, 0710098023, 0710098024, 0710098025, 0710098026, 0710098027, 0710098028, 0710098029, 0710098030, 0710098031, 0710098032, 0710098033, 0710098065, 0710098066, 0710098067, 0710098068, 0710098069, 0710098070, 0710098071, 0710098072, 0710098098, 0710098099, 0710098100, 0710098101, 0710098102, 0710098109, 0710098110, 0710098134, 0710098135, 0710098136, 0710098137, 0710098138, 0710098139, 0710098140, 0710098141, 0710098142, 0710098143, 0710098144, 0710098164, 0710098165, 0710098166, 0710098167, 0710098168, 0710098169, 0710098170, 0710098171, 0710098172, 0710098173, 0710098174, 0710098175, 0710098176, 0710098177, 0710098178, 0710098179, 0710098180, 0710098181, 0710098182, 0710098183, 0710098184, 0710098185, 0710098186, 0710098187, 0710098188, 0710098189, 0710098190, 0710098191, 0710098192, 0710098193, 0710098194, 0710098195, 0710098196, 0710098197, 0710098199, 0710098200, 0710098201, 0710098202, 0710098203, 0710098204, 0710098206, 0710098207, 0710098208, 0710098209, 0710098210, 0710098211, 0710098212, 0710098213, 0710098214, 0710098215, 0710098216, 0710098217, 0710098218, 0710098219, 0710098220, 0710098221, 0710098222, 0710098223, 0710098227, 0710098228, 0711096017, 0711096018, 0711096019, 0711096020, 0711096021, 0711096022, 0711096023, 0711096024, 0711096025, 0711096026, 0711096027, 0711096028, 0711096029, 0711096030, 0711096031, 0711096032, 0711096033, 0711096034, 0711096035, 0711096036, 0711096037, 0711096038, 0711096039, 0711096040, 0711096041, 0711096042, 0711096043, 0711096044, 0711096045, 0711096046, 0711096047, 0711096048, 0711096057, 0711096058, 0711096059, 0711096060, 0711096061, 0711096062, 0711096063, 0711096064, 0711096065, 0711096066, 0711096067, 0711096068, 0711096069, 0711096070, 0711096071, 0711096072, 0711096073, 0711096074, 0711096075, 0711096076, 0711096077, 0711096078, 0711096079, 0711096080, 0711096081, 0711096082, 0711096084, 0711096085, 0711096086, 0711096087, 0711096088, 0711096089, 0711096090, 0711096091, 0711096105, 0711096106, 0711096107, 0711096108, 0711096109, 0711096110, 0711096117, 0711096118, 0711096119, 0711096141, 0711096142, 0711096143, 0711096144, 0711096145, 0711096146, 0711096147, 0711096148, 0711096149, 0711096150, 0711096167, 0711096168, 0711096169, 0711096170, 0711096171, 0711096172, 0711096173, 0711096174, 0711096175, 0711096176, 0711096177, 0711096184, 0711096185, 0711096186, 0711096187, 0711096188, 0711096189, 0711096190, 0711096206, 0711096226, 0711096227, 0711096233, 0711096234, 0711096235, 0711096236, 0711096237, 0711096238, 0711096239, 0711096240, 0711096241, 0711096242, 0711096243, 0711096244, 0712094029, 0712094030, 0712094031, 0712094032, 0712094033, 0712094034, 0712094042, 0712094043, 0712094044, 0712094045, 0712094046, 0712094047, 0712094048, 0712094049, 0712094050, 0712094051, 0712094062, 0712094085, 0712094086, 0712094087, 0712094088, 0712094089, 0712094090, 0712094091, 0712094092, 0712094093, 0712094094, 0712094095, 0712094096, 0712094097, 0712094098, 0712094099, 0712094100, 0712094101, 0712094102, 0712094142, 0712094143, 0712094144, 0712094145, 0712094146, 0712094147, 0712094156, 0712094157, 0712094158, 0712094159, 0712094160, 0712094161, 0712094162, 0712094163, 0712094164, 0712094165, 0712094166, 0712094167, 0712094168, 0712094169, 0712094170, 0712094171, 0712094172, 0712094173, 0712094174, 0712094175, 0712094176, 0712094221, 0712094222, 0712094223, 0712094225, 0712094226, 0712094227, 0712094228, 0712094229, 0712094230, 0712094231, 0712094232, 0712094233, 0712094234, 0712094235, 0712094236, 0712094237, 0712094238, 0712094239, 0712094240, 0712094241, 0712094242, 0712094243, 0712094244, 0712094245, 0712094246, 0712094247, 0712094248, 0712094249, 0712094250, 0712094251, 0712094252, 0712094253, 0712094254, 0712094255, 0712094256, 0712094257, 0712094258, 0712094259, 0712094260, 0712094261, 0712094268, 0712094293, 0801098501, 0801098508, 0801098509, 0801098510, 0801098511, 0801098512, 0801098513, 0801098514, 0801098515, 0801098516, 0801098517, 0801098518, 0801098526, 0801098545, 0801098546, 0801098547, 0801098563, 0801098564, 0801098565, 0801098566, 0801098567, 0801098568, 0801098652, 0801098653, 0801098668, 0801098669, 0801098670, 0801098671, 0801098672, 0801098673, 0801098674, 0801098675, 0801098676, 0801098730, 0801098733, 0801098734, 0801098783, 0801098784, 0801098785, 0801098786, 0801098787, 0801098788, 0801098789, 0801098790, 0801098791, 0801098792, 0801098793, 0801098794, 0802102180, 0802102200, 0802102215, 0802102216, 0802102217, 0802102218, 0802102219, 0802102220, 0802102221, 0802102222, 0802102223, 0802102224, 0802102225, 0802102226, 0802102232, 0802102233, 0802102234, 0802102235, 0802102236, 0802102237, 0802102238, 0802102239, 0802102240, 0802102258, 0802102259, 0802102260, 0802102261, 0802102262, 0802102263, 0802102264, 0802102265, 0802102266, 0802102267, 0802102268, 0802102269, 0802102270, 0802102271, 0802102272, 0802102281, 0802102282, 0802102283, 0803101559, 0803101560, 0803101561, 0803101562, 0803101563, 0803101564, 0803101586, 0803101587, 0803101588, 0803101589, 0803101590, 0803101591, 0803101592, 0803101593, 0803101612, 0803101613, 0803101614, 0803101615, 0803101616, 0803101617, 0803101618, 0803101619, 0803101620, 0803101621, 0803101622, 0803101623, 0803101624, 0803101625, 0803101626, 0803101627, 0803101628, 0803101629, 0803101630, 0803101631, 0803101632, 0803101633, 0803101634, 0803101635, 0803101636, 0803101637, 0803101638, 0803101639, 0803101640, 0803101641, 0803101642, 0803101643, 0803101644, 0803101645, 0803101646, 0803101647, 0803101648, 0803101649, 0803101650, 0803101651, 0803101652, 0803101653, 0803101654, 0803101655, 0803101696, 0803101697, 0803101698, 0803101699, 0803101700, 0803101701, 0803101702, 0803101703, 0803101704, 0803101705, 0803101706, 0803101707, 0803101708, 0803101709, 0803101710, 0803101711, 0803101712, 0803101713, 0803101728, 0803101729, 0803101730, 0803101731, 0803101732, 0803101733, 0803101734, 0803101735, 0803101736, 0803101737, 0803101738, 0803101739, 0803101740, 0803101741, 0803101749, 0803101751, 0803101752, 0803101753, 0803101754, 0803101755, 0803101756, 0803101757, 0803101758, 0803101759, 0803101781, 0803101782, 0803101783, 0803101784, 0803101785, 0803101786, 0803101787, 0803101788, 0803101789, 0803101790, 0803101791, 0803101800, 0803101801, 0803101802, 0803101803, 0803101804, 0803101805, 0803101806, 0803101807, 0803101808, 0803101809, 0803101810, 0803101811, 0803101812, 0803101813, 0803101814, 0803101815, 0803101816, 0803101817, 0803101818, 0803101819, 0803101820, 0803101821, 0803101822, 0803101823, 0803101824, 0803101825, 0803101826, 0803101827, 0803101828, 0803101829, 0803101830, 0803101831, 0803101835, 0803101836, 0803101837, 0803101838, 0803101839, 0803101840, 0803101841, 0803101842, 0803101843, 0803101844, 0803101845, 0803101846, 0803101847, 0803101848, 0803101849, 0803101850, 0803101851, 0803101852, 0803101853, 0804107029, 0804107030, 0804107031, 0804107032, 0804107033, 0804107034, 0804107035, 0804107036, 0804107037, 0804107038, 0804107039, 0804107040, 0804107041, 0804107042, 0804107043, 0804107044, 0804107045, 0804107046, 0804107047, 0804107048, 0804107049, 0804107050, 0804107051, 0804107052, 0804107053, 0804107054, 0804107055, 0804107056, 0804107057, 0804107058, 0804107059, 0804107060, 0804107061, 0804107062, 0804107063, 0804107064, 0804107065, 0804107066, 0804107067, 0804107068, 0804107069, 0804107070, 0804107071, 0804107072, 0804107073, 0804107074, 0804107075, 0804107076, 0804107077, 0804107078, 0804107079, 0804107080, 0804107081, 0804107082, 0804107083, 0804107084, 0804107085, 0804107104, 0804107105, 0804107106, 0804107107, 0804107108, 0804107109, 0804107110, 0804107111, 0804107112, 0804107113, 0804107114, 0804107115, 0804107116, 0804107117, 0804107118, 0804107119, 0804107120, 0804107121, 0804107122, 0804107123, 0804107124, 0804107125, 0804107126, 0804107143, 0804107144, 0804107145, 0804107146, 0804107147, 0804107148, 0804107149, 0804107150, 0804107151, 0804107152, 0804107153, 0804107154, 0804107171, 0804107172, 0804107173, 0804107174, 0804107175, 0804107176, 0804107177, 0804107178, 0804107179, 0804107180, 0804107181, 0804107182, 0804107183, 0804107184, 0804107185, 0804107186, 0804107187, 0804107188, 0804107189, 0804107190, 0804107191, 0804107192, 0804107193, 0804107194, 0804107195, 0804107196, 0804107197, 0804107198, 0804107199, 0804107200, 0804107201, 0804107202, 0804107203, 0804107204, 0804107205, 0804107206, 0804107207, 0804107208, 0804107209, 0804107210, 0804107216, 0804107217, 0804107218, 0804107219, 0804107220, 0804107221, 0804107222, 0804107223, 0804107224, 0804107225, 0804107241, 0804107242, 0804107243, 0804107244, 0804107245, 0804107246, 0804107247, 0804107248, 0804107249, 0804107250, 0804107251, 0804107252, 0804107253, 0804107254, 0804107255, 0804107256, 0804107257, 0804107258, 0804107259, 0804107260, 0804107261, 0804107262, 0804107263, 0804107264, 0804107265, 0804107266, 0804107267, 0804107268, 0804107269, 0804107270, 0804107271, 0804107272, 0804107273, 0804107274, 0804107275, 0804107276, 0804107277, 0804107278, 0804107279, 0804107280, 0804107281, 0804107282, 0804107311, 0804107312, 0804107313, 0804107314, 0804107315, 0804107316, 0804107317, 0804107318, 0804107319, 0804107320, 0804107321, 0805099011, 0805099017, 0805099018, 0805099019, 0805099020, 0805099021, 0805099022, 0805099023, 0805099024, 0805099025, 0805099037, 0805099038, 0805099039, 0805099040, 0805099041, 0805099042, 0805099043, 0805099044, 0805099045, 0805099046, 0805099047, 0805099048, 0805099049, 0805099050, 0805099051, 0805099052, 0805099053, 0805099054, 0805099055, 0805099056, 0805099057, 0805099058, 0805099059, 0805099060, 0805099061, 0805099062, 0805099063, 0805099064, 0805099065, 0805099066, 0805099067, 0805099068, 0805099069, 0805099070, 0805099071, 0805099072, 0805099073, 0805099074, 0805099075, 0805099076, 0805099077, 0805099078, 0805099079, 0805099080, 0805099081, 0805099082, 0805099083, 0805099084, 0805099085, 0805099086, 0805099087, 0805099088, 0805099109, 0805099110, 0805099111, 0805099112, 0805099113, 0805099114, 0805099115, 0805099116, 0805099117, 0805099118, 0805099119, 0805099120, 0805099121, 0805099122, 0805099123, 0805099124, 0805099125, 0805099126, 0805099127, 0805099128, 0805099129, 0805099130, 0805099131, 0805099132, 0805099133, 0805099134, 0805099135, 0805099136, 0805099137, 0805099138, 0805099139, 0805099140, 0805099141, 0805099142, 0805099143, 0805099144, 0805099145, 0805099146, 0805099147, 0805099148, 0805099159, 0805099160, 0805099161, 0805099162, 0805099163, 0805099164, 0805099165, 0805099166, 0805099167, 0805099168, 0805099169, 0805099170, 0805099171, 0805099172, 0805099173, 0805099174, 0805099175, 0805099176, 0805099177, 0805099178, 0805099179, 0805099180, 0805099181, 0805099182, 0805099183, 0805099184, 0805099185, 0805099186, 0805099187, 0805099188, 0805099189, 0805099190, 0805099191, 0805099192, 0805099193, 0805099194, 0805099195, 0805099196, 0805099197, 0805099198, 0805099199, 0805099218, 0805099219, 0805099220, 0805099221, 0805099222, 0805099223, 0805099224, 0805099225, 0805099226, 0805099227, 0805099228, 0805099229, 0805099230, 0805099231, 0805099232, 0805099233, 0805099234, 0805099235, 0805099236, 0806108011, 0806108012, 0806108013, 0806108014, 0806108015, 0806108016, 0806108017, 0806108030, 0806108031, 0806108032, 0806108033, 0806108034, 0806108035, 0806108036, 0806108037, 0806108038, 0806108039, 0806108040, 0806108041, 0806108042, 0806108043, 0806108044, 0806108045, 0806108046, 0806108047, 0806108048, 0806108049, 0806108050, 0806108051, 0806108052, 0806108053, 0806108054, 0806108055, 0806108056, 0806108057, 0806108058, 0806108059, 0806108060, 0806108061, 0806108062, 0806108063, 0806108064, 0806108065, 0806108066, 0806108067, 0806108068, 0806108069, 0806108070, 0806108071, 0806108072, 0806108073, 0806108074, 0806108075, 0806108076, 0806108077, 0806108078, 0806108079, 0806108080, 0806108081, 0806108082, 0806108083, 0806108084, 0806108085, 0806108086, 0806108087, 0806108088, 0806108089, 0806108090, 0806108091, 0806108092, 0806108093, 0806108094, 0806108095, 0806108096, 0806108097, 0806108098, 0806108099, 0806108100, 0806108101, 0806108102, 0806108103, 0806108104, 0806108105, 0806108106, 0806108107, 0806108108, 0806108109, 0806108110, 0806108111, 0806108112, 0806108113, 0806108114, 0806108115, 0806108116, 0806108117, 0806108118, 0806108119, 0806108120, 0806108121, 0806108122, 0806108123, 0806108124, 0806108125, 0806108169, 0806108170, 0806108171, 0806108172, 0806108173, 0806108174, 0806108175, 0806108176, 0806108177, 0806108178, 0806108179, 0806108180, 0806108181, 0806108182, 0806108183, 0806108184, 0806108209, 0806108210, 0806108211, 0806108212, 0806108213, 0806108214, 0806108215, 0806108216, 0806108217, 0806108218, 0806108219, 0806108220, 0806108221, 0806108222, 0806108223, 0806108224, 0806108225, 0806108226, 0806108227, 0806108228, 0806108229, 0806108230, 0806108231, 0806108232, 0806108233, 0806108234, 0806108235, 0806108236, 0806108237, 0806108238, 0806108239, 0806108240, 0806108241, 0806108242, 0806108243, 0806108244, 0806108245, 0806108246, 0806108247, 0806108248, 0806108249, 0806108250, 0806108251, 0806108252, 0806108253, 0806108254, 0806108286, 0806108325, 0806108326, 0807105528, 0807105529, 0807105530, 0807105531, 0807105532, 0807105533, 0807105534, 0807105535, 0807105536, 0807105537, 0807105538, 0807105539, 0807105540, 0807105541, 0807105542, 0807105562, 0807105563, 0807105566, 0807105577, 0807105578, 0807105579, 0807105580, 0807105618, 0808107002, 0808107044, 0808107045, 0808107046, 0809104507, 0809104529, 0809104530, 0809104531, 0809104532, 0809104533, 0809104534, 0809104535, 0809104536, 0809104537, 0809104538, 0809104539, 0809104540, 0809104541, 0809104542, 0809104543, 0809104544, 0809104545, 0809104546, 0809104547, 0809104548, 0809104549, 0809104550, 0809104579, 0809104580, 0809104581, 0809104582, 0809104583, 0809104584, 0809104585, 0809104586, 0809104587, 0809104588, 0809104589, 0809104590, 0809104591, 0809104592, 0809104593, 0809104594, 0809104595, 0809104596, 0809104634, 0809104635, 0809104636, 0809104637, 0809104638, 0809104639, 0809104640, 0809104641, 0809104642, 0809104643, 0809104644, 0809104645, 0809104646, 0809104647, 0809104648, 0809104649, 0809104650, 0809104651, 0809104652, 0809104653, 0809104654, 0809104655, 0809104656, 0809104657, 0809104658, 0809104659, 0809104660, 0809104661, 0809104662, 0809104663, 0809104664, 0809104665, 0809104666, 0809104667, 0809104668, 0809104669, 0809104670, 0809104671, 0809104672, 0809104673, 0809104674, 0809104675, 0809104676, 0809104677, 0809104678, 0809104718, 0809104719, 0809104720, 0809104721, 0809104722, 0809104723, 0809104726, 0809104727, 0809104728, 0809104729, 0809104730, 0809104731, 0809104732, 0809104733, 0809104734, 0809104735, 0809104736, 0809104737, 0809104738, 0809104739, 0809104740, 0809104741, 0809104742, 0809104743, 0809104744, 0809104745, 0809104746, 0809104747, 0809104748, 0809104749, 0809104750, 0809104763, 0809104764, 0809104765, 0809104766, 0809104767, 0809104768, 0809104769, 0809104770, 0809104771, 0809104772, 0809104773, 0809104774, 0809104775, 0809104776, 0809104777, 0809104778, 0810100313, 0810100314, 0810100315, 0810100316, 0810100317, 0810100318, 0810100319, 0810100320, 0810100321, 0810100322, 0810100323, 0810100324, 0810100325, 0810100326, 0810100327, 0810100328, 0810100329, 0810100330, 0810100331, 0810100332, 0810100333, 0810100334, 0810100335, 0810100336, 0810100337, 0810100338, 0810100339, 0810100340, 0810100341, 0810100342, 0810100343, 0810100344, 0810100345, 0810100346, 0810100347, 0810100348, 0810100349, 0810100350, 0810100351, 0810100352, 0810100353, 0810100354, 0810100355, 0810100356, 0810100357, 0810100358, 0810100359, 0810100360, 0810100361, 0810100362, 0810100363, 0810100364, 0810100365, 0810100366, 0810100367, 0810100368, 0810100369, 0810100370, 0810100371, 0810100372, 0810100373, 0810100374, 0810100375, 0810100376, 0810100377, 0810100378, 0810100379, 0810100380, 0810100381, 0810100382, 0810100399, 0810100400, 0810100401, 0810100402, 0810100403, 0810100404, 0810100405, 0810100406, 0810100407, 0810100408, 0810100409, 0810100410, 0810100411, 0810100412, 0810100413, 0810100414, 0810100415, 0810100416, 0810100417, 0810100418, 0810100419, 0810100420, 0810100421, 0810100422, 0810100423, 0810100424, 0810100425, 0810100426, 0810100427, 0810100429, 0810100430, 0810100431, 0810100432, 0810100433, 0810100434, 0810100435, 0810100436, 0810100437, 0810100438, 0810100439, 0810100450, 0810100451, 0810100452, 0810100457, 0810100458, 0810100459, 0810100460, 0810100461, 0810100462, 0810100463, 0810100464, 0810100465, 0810100466, 0810100467, 0810100468, 0810100469, 0810100470, 0810100471, 0810100472, 0810100473, 0810100474, 0810100475, 0810100476, 0810100477, 0811100013, 0811100014, 0811100015, 0811100016, 0811100017, 0811100018, 0811100019, 0811100020, 0811100021, 0811100022, 0811100023, 0811100025, 0811100026, 0811100027, 0811100028, 0811100029, 0811100030, 0811100031, 0811100032, 0811100033, 0811100034, 0811100035, 0811100036, 0811100037, 0811100038, 0811100039, 0811100040, 0811100041, 0811100042, 0811100043, 0811100044, 0811100064, 0811100065, 0811100066, 0811100067, 0811100068, 0811100069, 0811100070, 0811100071, 0811100072, 0811100073, 0811100074, 0811100075, 0811100076, 0811100077, 0811100078, 0811100079, 0811100080, 0811100081, 0811100082, 0811100083, 0811100084, 0811100085, 0811100086, 0811100087, 0811100088, 0811100089, 0811100090, 0811100091, 0811100092, 0811100093, 0811100094, 0811100095, 0811100096, 0811100097, 0811100098, 0811100099, 0811100100, 0811100101, 0811100102, 0811100103, 0811100104, 0811100105, 0811100106, 0811100116, 0811100117, 0811100118, 0811100119, 0811100120, 0811100121, 0811100149, 0811100150, 0811100151, 0811100152, 0811100153, 0811100154, 0811100155, 0811100156, 0811100157, 0811100158, 0811100159, 0811100160, 0811100161, 0811100162, 0811100163, 0811100164, 0811100165, 0811100166, 0811100167, 0811100168, 0811100169, 0811100170, 0811100171, 0811100207, 0811100208, 0811100209, 0811100210, 0811100211, 0811100212, 0811100213, 0811100214, 0811100215, 0811100216, 0811100217, 0811100218, 0811100219, 0811100220, 0811100221, 0811100222, 0811100223, 0811100224, 0811100225, 0811100226, 0811100227, 0811100228, 0811100229, 0811100230, 0811100231, 0811100232, 0811100233, 0811100234, 0811100235, 0811100236, 0811100237, 0811100238, 0811100239, 0811100240, 0811100241, 0811100242, 0811100243, 0811100244, 0811100245, 0811100246, 0811100247, 0811100248, 0811100249, 0811100250, 0811100251, 0811100252, 0811100253, 0811100254, 0811100255, 0811100256, 0811100257, 0811100258, 0811100259, 0811100260, 0812100001, 0812100002, 0812100003, 0812100004, 0812100005, 0812100006, 0812100007, 0812100008, 0812100009, 0812100010, 0812100011, 0812100012, 0812100013, 0812100014, 0812100015, 0812100016, 0812100017, 0812100018, 0812100019, 0812100020, 0812100021, 0812100022, 0812100023, 0812100024, 0812100025, 0812100026, 0812100035, 0812100036, 0812100037, 0812100038, 0812100039, 0812100040, 0812100041, 0812100042, 0812100043, 0812100044, 0812100045, 0812100046, 0812100047, 0812100048, 0812100049, 0812100050, 0812100051, 0812100052, 0812100053, 0812100078, 0812100079, 0812100080, 0812100081, 0812100082, 0812100083, 0812100126, 0812100127, 0812100128, 0812100129, 0812100130, 0812100131, 0812100132, 0812100133, 0812100134, 0812100135, 0812100136, 0812100137, 0812100141, 0812100142, 0812100143, 0812100144, 0812100145, 0812100146, 0812100147, 0812100148, 0812100149, 0812100150, 0812100151, 0812100152, 0812100153, 0812100154, 0812100155, 0812100156, 0812100157, 0812100158, 0812100159, 0812100160, 0812100161, 0812100162, 0812100163, 0812100164, 0812100165, 0812100166, 0812100167, 0812100168, 0812100169, 0812100170, 0812100171, 0812100172, 0812100173, 0812100175, 0812100176, 0812100177, 0812100178, 0812100190, 0812100193, 0812100194, 0901034505, 0901034506, 0901034507, 0901034508, 0901034509, 0901034510, 0901034511, 0901034512, 0901034513, 0901034514, 0901034516, 0901034517, 0901034518, 0901034519, 0901034520, 0901034521, 0901034522, 0901034523, 0901034524, 0901034525, 0901034526, 0901034527, 0901034531, 0901034533, 0901034534, 0901034535, 0901034536, 0901034537, 0901034538, 0901034539, 0901034540, 0901034541, 0901034558, 0901034579, 0901034580, 0901034581, 0901034582, 0901034583, 0901034584, 0901034585, 0901034586, 0901034587, 0901034588, 0901034589, 0901034590, 0901034591, 0901034592, 0901034593, 0901034594, 0901034595, 0901034596, 0901034597, 0901034598, 0901034599, 0901034600, 0901034601, 0901034602, 0901034603, 0901034604, 0901034605, 0901034606, 0901034607, 0901034608, 0901034609, 0901034610, 0901034611, 0901034612, 0901034613, 0901034614, 0901034615, 0901034616, 0901034617, 0901034618, 0901034619, 0901034620, 0901034621, 0901034622, 0901034623, 0901034624, 0901034625, 0901034626, 0901034627, 0901034628, 0901034629, 0901034630, 0901034631, 0901034632, 0901034633, 0901034634, 0901034635, 0901034636, 0901034637, 0901034638, 0901034639, 0901034664, 0901034665, 0901034666, 0901034667, 0901034668, 0901034669, 0901034670, 0901034671, 0901034672, 0901034673, 0901034674, 0901034675, 0901034676, 0901034677, 0901034678, 0901034679, 0901034680, 0901034681, 0901034682, 0901034683, 0901034684, 0901034685, 0901034686, 0901034687, 0901034688, 0901034689, 0901034690, 0901034691, 0901034692, 0901034693, 0901034694, 0901034695, 0901034696, 0901034697, 0901034698, 0901034699, 0901034700, 0901034701, 0901034702, 0901034703, 0902033501, 0902033502, 0902033503, 0902033504, 0902033505, 0902033506, 0902033507, 0902033508, 0902033509, 0902033510, 0902033511, 0902033512, 0902033549, 0902033550, 0902033551, 0902033552, 0902033553, 0902033554, 0902033555, 0902033556, 0902033557, 0902033558, 0902033559, 0902033560, 0902033561, 0902033562, 0902033563, 0902033564, 0902033565, 0902033566, 0902033567, 0902033568, 0902033569, 0902033570, 0902033571, 0902033572, 0902033573, 0902033574, 0902033575, 0902033576, 0902033577, 0902033578, 0902033579, 0902033580, 0902033581, 0902033582, 0902033585, 0902033586, 0902033587, 0902033588, 0902033589, 0902033590, 0902033591, 0902033592, 0902033596, 0902033597, 0902033598, 0902033599, 0902033600, 0902033601, 0902033602, 0902033603, 0902033604, 0902033609, 0902033610, 0902033611, 0902033612, 0902033613, 0902033616, 0902033617, 0902033618, 0902033619, 0902033620, 0902033621, 0902033622, 0902033623, 0902033636, 0902033637, 0902033638, 0902033639, 0902033640, 0902033641, 0902033642, 0903032007, 0903032008, 0903032009, 0903032010, 0903032053, 0903032067, 0903032068, 0903032069, 0904035017, 0904035018, 0904035019, 0904035020, 0904035021, 0904035022, 0904035025, 0904035026, 0904035027, 0904035028, 0904035029, 0904035030, 0904035031, 0904035032, 0904035033, 0904035034, 0904035035, 0904035036, 0904035037, 0904035038, 0904035039, 0904035040, 0904035041, 0904035042, 0904035043, 0904035044, 0904035046, 0904035047, 0904035048, 0904035049, 0904035088, 0904035089, 0904035090, 0904035091, 0904035092, 0904035093, 0904035094, 0904035095, 0904035096, 0904035097, 0904035103, 0904035104, 0904035105, 0905033586, 0905033587, 0905033588, 0906033525, 0906033554, 0906033563 and 0906033564. |
Recalling Firm/ Manufacturer |
Stryker Medical Div. of Stryker Corporation 3800 E. Centre Ave. Portage MI 49002
|
For Additional Information Contact | Derick Elliott 269-324-6609 |
Manufacturer Reason for Recall | The Fowler mechanism may malfunction, preventing it from being lowered into a flat position. |
FDA Determined Cause 2 | Component change control |
Action | Stryker Medical sent an Urgent Medical Device Alert letter dated May 14, 2010 to their customers. The letter identified the product, the problem, and the actions customers should take.
The letter instructed users to do the following:
" Return the postcard included with the letter to confirm receipt of the
notification.
" Locate the recalled beds, verify the Fowler mechanism is functioning properly and does
not display uneven motion when raising and lowering, and is not unusually
noisy.
" If the Fowler mechanism is not operating correctly, they were to call Stryker at
1-800-STRYKER, option 1, Reference RA-2010-014 to obtain service for the
bed.
" If any beds have been loaned or sold, the customer is to forward a copy of the
notice to the new users and advise the firm of the new location.
" Provide the firm with the serial numbers for any beds that were disposed of.
" Retain the revised maintenance manual included with the notification.
If the customer had any urgent questions or concerns, they were instructed to call the firm at 800-STRYKER, option 1, reference RA-2010-014, Monday-Friday 8am to 5pm (EST). |
Quantity in Commerce | 9,134 |
Distribution | Worldwide Distribution - USA (all states), and including the countries of ARGENTINA, BRAZIL, CHILE, CHINA, CANADA, MEXICO, GREECE, JAPAN, KOREA, PUERTO RICO, SPAIN, POLAND, TAIWAN, SINGAPORE, INDIA, SWITZERLAND, UK and MALAYSIA. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|