|
Class 2 Device Recall Stryker Zoom Critical Care Bed |
|
Date Initiated by Firm |
May 14, 2010 |
Date Posted |
September 28, 2010 |
Recall Status1 |
Terminated 3 on June 13, 2011 |
Recall Number |
Z-2587-2010 |
Recall Event ID |
56507 |
Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
Product |
Stryker Zoom Critical Care Bed, Model 2040. Stryker Medical, Kalamazoo, MI. Critical Care Patient Bed. |
Code Information |
041115567, 0301050512, 0301050513, 0301050514, 0301050515, 0301050516, 0301050517, 0301050518, 0301050519, 0301050520, 0301050521, 0301050522, 0301050523, 0301050524, 0301050525, 0301050526, 0301050527, 0301050528, 0301050529, 0301050554, 0301050574, 0301050575, 0301050576, 0301050577, 0301050578, 0301050579, 0301050580, 0301050581, 0301050582, 0301050583, 0301050584, 0301050585, 0301050586, 0301050587, 0301050588, 0301050589, 0301050590, 0301050591, 0301050593, 0301050594, 0301050595, 0301050596, 0301050597, 0301050598, 0301050599, 0301050600, 0301050601, 0301050602, 0301050603, 0301050604, 0301050605, 0301050606, 0301050607, 0301050608, 0301050609, 0301050610, 0301050611, 0301050612, 0301050613, 0301050614, 0301050615, 0301050616, 0301050617, 0301050618, 0301050619, 0301050620, 0301050621, 0301050622, 0301050623, 0301050624, 0301050625, 0301050626, 0301050627, 0301050628, 0301050629, 0301050630, 0301050631, 0301050632, 0301050633, 0302050502, 0302050503, 0302050504, 0302050505, 0302050506, 0302050507, 0302050508, 0302050509, 0302050517, 0302050518, 0302050524, 0302050525, 0302050526, 0302050528, 0302050529, 0302050530, 0302050531, 0302050532, 0302050533, 0302050534, 0302050535, 0302050544, 0302050545, 0302050546, 0302050547, 0302050548, 0302050549, 0302050550, 0302050551, 0302050552, 0302050553, 0302050554, 0302050555, 0302050556, 0302050557, 0302050558, 0302050559, 0302050560, 0302050561, 0302050568, 0302050569, 0302050570, 0302050571, 0302050572, 0302050573, 0302050574, 0302050575, 0302050582, 0302050583, 0302050584, 0302050585, 0302050586, 0302050587, 0302050603, 0302050604, 0302050620, 0302050655, 0302050656, 0302050658, 0302050659, 0302050660, 0302050661, 0302050662, 0302050663, 0302050664, 0302050665, 0302050666, 0302050667, 0302050668, 0302050669, 0302050670, 0302050671, 0302050672, 0302050673, 0302050674, 0302050675, 0303054521, 0303054522, 0303054523, 0303054524, 0303054525, 0303054526, 0303054527, 0303054528, 0303054529, 0303054530, 0303054531, 0303054532, 0303054533, 0303054537, 0303054538, 0303054559, 0303054560, 0303054561, 0303054562, 0303054563, 0303054564, 0303054579, 0303054602, 0303054603, 0303054610, 0303054611, 0303054612, 0303054613, 0303054614, 0303054615, 0303054616, 0303054617, 0303054618, 0303054619, 0303054620, 0303054621, 0303054622, 0303054623, 0303054624, 0303054625, 0303054626, 0303054627, 0303054628, 0303054629, 0303054630, 0303054631, 0303054632, 0303054633, 0303054634, 0303054635, 0303054636, 0303054637, 0303054638, 0303054639, 0303054640, 0303054641, 0303054642, 0303054650, 0303054651, 0303054652, 0303054653, 0303054654, 0303054655, 0303054656, 0303054668, 0303054669, 0303054670, 0303054671, 0303054672, 0303054673, 0303054674, 0303054675, 0303054676, 0303054677, 0304049014, 0304049049, 0304049050, 0304049051, 0304049052, 0304049053, 0304049054, 0304049055, 0304049056, 0304049057, 0304049058, 0304049059, 0304049060, 0304049061, 0304049062, 0304049063, 0304049064, 0304049065, 0304049066, 0304049067, 0304049068, 0304049069, 0304049070, 0304049071, 0304049072, 0304049073, 0304049074, 0304049075, 0304049078, 0304049152, 0304049153, 0304049154, 0304049155, 0304049157, 0304049158, 0304049159, 0304049160, 0304049161, 0304049162, 0304049163, 0304049164, 0304049165, 0304049166, 0304049167, 0304049172, 0304049173, 0305054518, 0305054519, 0305054520, 0305054521, 0305054522, 0305054523, 0305054524, 0305054525, 0305054526, 0305054527, 0305054528, 0305054529, 0305054530, 0305054531, 0305054570, 0305054571, 0305054572, 0305054573, 0305054574, 0305054575, 0305054576, 0305054577, 0305054578, 0305054585, 0305054586, 0305054587, 0305054588, 0305054589, 0305054590, 0305054591, 0305054592, 0305054593, 0305054594, 0305054595, 0305054596, 0305054597, 0305054598, 0305054599, 0305054636, 0305054637, 0305054638, 0305054639, 0305054640, 0305054641, 0305054642, 0305054643, 0305054644, 0305054645, 0305054646, 0305054647, 0305054648, 0305054649, 0305054650, 0305054651, 0305054652, 0305054653, 0305054654, 0305054655, 0305054656, 0305054657, 0305054658, 0305054664, 0305054665, 0305054666, 0305054667, 0305054668, 0305054673, 0305054674, 0305054675, 0306057560, 0306057561, 0306057562, 0306057616, 0306057617, 0306057634, 0306057715, 0306057716, 0306057717, 0306057718, 0306057719, 0306057720, 0306057721, 0306057722, 0306057723, 0306057724, 0306057725, 0306057726, 0306057749, 0306057750, 0307057001, 0307057054, 0307057055, 0307057056, 0307057057, 0307057058, 0307057059, 0307057060, 0307057061, 0307057062, 0307057063, 0307057064, 0307057065, 0307057066, 0307057067, 0307057068, 0307057069, 0307057070, 0307057071, 0307057072, 0307057073, 0307057074, 0307057075, 0307057076, 0307057077, 0307057078, 0307057079, 0307057080, 0307057081, 0307057082, 0307057083, 0307057084, 0307057085, 0307057086, 0307057096, 0307057097, 0307057098, 0307057099, 0307057100, 0307057103, 0307057104, 0307057105, 0307057106, 0307057107, 0307057108, 0307057109, 0307057110, 0307057111, 0307057112, 0307057113, 0307057114, 0307057115, 0307057116, 0307057117, 0307057118, 0307057119, 0307057120, 0307057121, 0307057122, 0307057123, 0307057124, 0307057125, 0307057126, 0307057127, 0307057128, 0307057129, 0307057130, 0307057131, 0307057132, 0307057133, 0307057134, 0307057135, 0307057136, 0307057137, 0307057138, 0307057139, 0307057140, 0307057141, 0307057142, 0307057143, 0307057144, 0307057145, 0307057146, 0307057147, 0307057148, 0307057164, 0307057189, 0307057198, 0308058043, 0308058044, 0308058072, 0308058131, 0309057520, 0309057521, 0309057522, 0309057523, 0309057524, 0309057525, 0309057526, 0309057527, 0309057528, 0309057529, 0309057530, 0309057531, 0309057532, 0309057533, 0309057534, 0309057535, 0309057536, 0309057537, 0309057538, 0309057539, 0309057540, 0309057541, 0309057542, 0309057547, 0309057548, 0309057549, 0309057550, 0309057551, 0309057552, 0309057553, 0309057554, 0309057555, 0309057556, 0309057575, 0309057576, 0309057577, 0309057578, 0309057579, 0309057580, 0309057581, 0309057582, 0309057583, 0310048504, 0310048505, 0310048506, 0310048507, 0310048508, 0310048509, 0310048510, 0310048511, 0310048560, 0310048561, 0310048562, 0310048632, 0310048633, 0310048653, 0310048654, 0310048655, 0310048656, 0310048657, 0310048658, 0310048684, 0310048685, 0310048688, 0311052566, 0311052567, 0311052578, 0311052579, 0311052580, 0311052581, 0311052582, 0311052583, 0311052584, 0311052585, 0311052586, 0311052587, 0311052588, 0311052589, 0311052590, 0311052591, 0311052592, 0311052593, 0311052594, 0311052595, 0311052622, 0311052623, 0311052624, 0311052625, 0311052626, 0311052627, 0311052668, 0311052680, 0311052681, 0311052682, 0311052683, 0311052684, 0311052685, 0311052686, 0311052687, 0312056516, 0312056521, 0312056522, 0312056523, 0312056524, 0312056525, 0312056526, 0312056527, 0312056528, 0312056529, 0312056530, 0312056531, 0312056532, 0312056533, 0312056534, 0312056535, 0312056536, 0312056537, 0312056538, 0312056540, 0312056541, 0312056542, 0312056543, 0312056544, 0312056545, 0312056546, 0312056547, 0312056548, 0312056549, 0312056550, 0312056551, 0312056552, 0312056553, 0312056554, 0312056555, 0312056556, 0312056557, 0312056558, 0312056559, 0312056560, 0312056561, 0312056562, 0312056563, 0312056564, 0312056565, 0312056566, 0312056567, 0312056568, 0312056569, 0312056570, 0312056571, 0312056572, 0312056573, 0312056574, 0312056575, 0312056576, 0312056577, 0312056578, 0312056579, 0312056580, 0312056581, 0312056582, 0312056583, 0312056584, 0312056585, 0312056586, 0312056587, 0312056588, 0312056589, 0312056590, 0312056591, 0312056592, 0312056612, 0312056613, 0312056614, 0312056615, 0312056616, 0312056617, 0312056618, 0312056619, 0312056620, 0312056621, 0312056623, 0312056624, 0312056666, 0312056667, 0312056668, 0312056669, 0312056670, 0312056671, 0312056672, 0312056673, 0312056674, 0312056675, 0312056676, 0312056677, 0312056678, 0312056679, 0312056680, 0312056681, 0312056699, 0312056700, 0401062001, 0401062002, 0401062003, 0401062004, 0401062005, 0401062006, 0401062007, 0401062008, 0401062009, 0401062010, 0401062011, 0401062012, 0401062013, 0401062052, 0401062053, 0401062054, 0401062055, 0401062056, 0401062057, 0401062058, 0401062059, 0401062060, 0401062061, 0401062062, 0401062063, 0401062064, 0401062232, 0401062233, 0401062234, 0401062235, 0401062236, 0401062237, 0401062238, 0401062239, 0402066011, 0402066012, 0402066013, 0402066016, 0402066017, 0402066018, 0402066043, 0402066074, 0402066075, 0402066076, 0402066077, 0402066078, 0402066102, 0402066103, 0402066104, 0402066105, 0402066117, 0402066119, 0402066120, 0402066121, 0402066122, 0402066123, 0402066124, 0402066125, 0402066126, 0402066127, 0402066128, 0402066129, 0402066130, 0402066131, 0402066132, 0402066133, 0402066135, 0402066136, 0402066137, 0402066138, 0402066139, 0402066140, 0402066141, 0402066142, 0402066143, 0402066144, 0402066194, 0402066195, 0402066196, 0402066197, 0402066198, 0402066199, 0402066200, 0402066213, 0402066214, 0402066215, 0402066216, 0402066217, 0402066218, 0402066219, 0402066220, 0402066221, 0402066222, 0402066223, 0402066224, 0402066225, 0402066238, 0402066239, 0403064556, 0403064557, 0403064558, 0403064559, 0403064574, 0403064579, 0403064580, 0403064581, 0403064582, 0403064583, 0403064584, 0403064585, 0403064586, 0403064587, 0403064616, 0403064655, 0403064656, 0403064658, 0403064659, 0403064660, 0403064661, 0403064662, 0403064663, 0403064664, 0403064665, 0403064666, 0403064667, 0403064668, 0403064693, 0403064694, 0403064695, 0403064698, 0403064699, 0403064700, 0403064701, 0403064762, 0403064771, 0403064772, 0403064791, 0403064806, 0403064807, 0403064808, 0403064809, 0403064810, 0403064811, 0403064812, 0403064813, 0403064814, 0403064815, 0403064816, 0403064817, 0403064818, 0403064819, 0403064820, 0403064821, 0403064822, 0403064859, 0403064888, 0403064889, 0403064890, 0403064891, 0403064892, 0403064893, 0403064894, 0403064895, 0403064896, 0403064897, 0403064902, 0403064904, 0403064913, 0403064914, 0403064915, 0403064916, 0403064917, 0403064918, 0403064919, 0403064920, 0403064921, 0403064922, 0403064923, 0403064924, 0403064929, 0403064956, 0405066001, 0405066075, 0405066076, 0405066080, 0405066081, 0405066082, 0405066083, 0405066084, 0405066175, 0405066214, 0405066215, 0405066216, 0405066217, 0405066226, 0405066227, 0405066228, 0405066229, 0405066230, 0405066231, 0405066234, 0405066323, 0405066324, 0405066325, 0405066326, 0405066327, 0405066328, 0405066329, 0405066330, 0405066331, 0405066332, 0405066333, 0405066334, 0405066335, 0405066336, 0405066337, 0405066338, 0405066339, 0405066340, 0405066341, 0406060501, 0406060522, 0406060645, 0406060646, 0406060647, 0406060648, 0406060649, 0406060650, 0406060651, 0406060652, 0406060653, 0406060654, 0406060655, 0406060656, 0406060657, 0406060658, 0406060659, 0406060660, 0406060723, 0406060724, 0406060725, 0406060726, 0406060727, 0407060510, 0407060512, 0407060513, 0407060520, 0407060525, 0407060530, 0407060531, 0407060532, 0407060533, 0407060534, 0407060535, 0407060536, 0407060537, 0407060538, 0407060539, 0408061501, 0408061502, 0408061503, 0408061504, 0408061507, 0408061508, 0408061509, 0408061510, 0408061511, 0408061512, 0408061513, 0408061514, 0408061553, 0408061554, 0408061555, 0408061559, 0408061560, 0408061576, 0408061577, 0408061578, 0408061579, 0408061580, 0408061581, 0408061582, 0408061583, 0408061584, 0408061585, 0408061586, 0408061587, 0408061588, 0408061589, 0408061590, 0408061591, 0408061592, 0408061593, 0408061594, 0408061595, 0408061596, 0408061597, 0408061598, 0408061599, 0408061600, 0408061601, 0408061602, 0408061603, 0408061658, 0408061659, 0408061660, 0408061665, 0408061666, 0408061667, 0408061681, 0408061682, 0408061686, 0408061687, 0408061688, 0408061689, 0408061690, 0408061701, 0408061702, 0408061703, 0408061704, 0408061705, 0408061706, 0408061707, 0408061708, 0408061725, 0408061760, 0408061761, 0408061762, 0408061763, 0409060516, 0409060517, 0409060518, 0409060519, 0409060546, 0409060547, 0409060548, 0409060551, 0409060552, 0409060553, 0409060554, 0409060555, 0409060556, 0409060557, 0409060558, 0409060559, 0409060566, 0409060567, 0409060568, 0409060569, 0409060570, 0409060571, 0409060572, 0409060573, 0409060574, 0409060575, 0409060576, 0409060577, 0409060578, 0409060579, 0409060580, 0409060581, 0409060582, 0409060583, 0409060584, 0409060585, 0409060586, 0409060587, 0409060588, 0409060589, 0409060590, 0409060591, 0409060592, 0409060605, 0409060606, 0409060607, 0409060608, 0409060609, 0409060622, 0409060623, 0409060624, 0409060625, 0409060626, 0409060627, 0409060628, 0409060629, 0409060630, 0409060631, 0409060632, 0409060633, 0409060634, 0409060635, 0409060636, 0409060637, 0409060638, 0409060639, 0409060640, 0409060641, 0409060642, 0409060643, 0409060644, 0409060645, 0409060646, 0409060647, 0409060648, 0409060678, 0409060725, 0409060893, 0409060894, 0409060895, 0409060896, 0409060897, 0409060898, 0409060899, 0410062519, 0410062520, 0410062521, 0410062522, 0410062523, 0410062524, 0410062525, 0410062583, 0410062584, 0410062585, 0410062586, 0410062587, 0410062588, 0410062589, 0410062590, 0410062591, 0410062624, 0410062627, 0410062638, 0410062640, 0410062641, 0410062642, 0410062644, 0410062645, 0410062646, 0410062648, 0410062649, 0410062650, 0410062651, 0410062652, 0410062653, 0410062654, 0410062655, 0410062656, 0410062657, 0410062658, 0410062659, 0410062660, 0410062661, 0410062662, 0410062663, 0410062665, 0410062678, 0410062679, 0410062680, 0410062681, 0410062682, 0410062683, 0410062684, 0410062685, 0410062686, 0410062687, 0410062688, 0410062689, 0410062690, 0410062717, 0410062718, 0410062726, 0410062727, 0410062728, 0410062729, 0410062730, 0410062731, 0410062732, 0410062733, 0410062734, 0410062735, 0410062736, 0410062737, 0411062011, 0411062012, 0411062013, 0411062014, 0411062015, 0411062016, 0411062017, 0411062018, 0411062019, 0411062020, 0411062021, 0411062022, 0411062023, 0411062024, 0411062028, 0411062029, 0411062030, 0411062031, 0411062032, 0411062039, 0411062040, 0411062042, 0411062043, 0411062044, 0411062054, 0411062055, 0411062060, 0411062073, 0411062074, 0411062075, 0411062076, 0411062077, 0411062078, 0411062133, 0411062134, 0411062205, 0411062206, 0411062207, 0411062208, 0411062209, 0411062210, 0411062212, 0411062213, 0411062259, 0411062260, 0411062261, 0411062262, 0411062263, 0411062264, 0411062265, 0411062266, 0411062267, 0411062281, 0411062306, 0411062307, 0411062308, 0411062309, 0411062310, 0411062311, 0411062312, 0411062313, 0412061506, 0412061507, 0412061508, 0412061509, 0412061510, 0412061511, 0412061512, 0412061513, 0412061514, 0412061515, 0412061516, 0412061517, 0412061518, 0412061519, 0412065005, 0412065006, 0412065007, 0412065008, 0412065009, 0412065010, 0412065011, 0412065017, 0412065018, 0412065019, 0412065020, 0412065021, 0412065022, 0412065023, 0412065024, 0412065025, 0412065026, 0412065027, 0412065028, 0412065029, 0412065030, 0412065031, 0412065032, 0412065033, 0412065034, 0412065035, 0412065036, 0412065037, 0412065038, 0412065039, 0412065214, 0501069526, 0501069527, 0501069528, 0501069529, 0501069530, 0501069531, 0501069532, 0501069533, 0501069534, 0501069535, 0501069536, 0501069537, 0501069538, 0501069539, 0501069540, 0501069541, 0501069572, 0501069573, 0501069574, 0501069575, 0501069576, 0501070501, 0501070564, 0501070565, 0501070566, 0501070567, 0501070568, 0501070569, 0501070597, 0501070598, 0501070599, 0501070600, 0501070601, 0501070602, 0501070603, 0501070604, 0501070605, 0501070606, 0501070607, 0501070608, 0501070609, 0501070610, 0501070611, 0501070612, 0501070613, 0501070614, 0501070649, 0501070650, 0501070651, 0501070652, 0501070653, 0501070656, 0501070657, 0501070658, 0501070659, 0501070660, 0501070704, 0501070705, 0501070706, 0501070707, 0501070708, 0501070713, 0501070714, 0501070715, 0501070716, 0501070717, 0501070718, 0501070719, 0501070720, 0501070721, 0501070722, 0501070723, 0501070724, 0501070725, 0501070726, 0501070727, 0501070728, 0501070729, 0501070730, 0501070731, 0501070732, 0501070733, 0501070734, 0502072521, 0502072522, 0502072523, 0502072524, 0502072525, 0502072526, 0502072527, 0502072528, 0502072529, 0502072530, 0502072531, 0502072532, 0502072533, 0502072534, 0502072535, 0502072536, 0502072537, 0502072538, 0502072602, 0502072603, 0502072604, 0502072605, 0502072606, 0502072607, 0502072608, 0502072609, 0502072610, 0502072611, 0502072612, 0502072613, 0502072614, 0502072615, 0502072616, 0502072617, 0502072618, 0502072619, 0502072620, 0502072621, 0502072622, 0502072623, 0502072635, 0502072636, 0502072637, 0502072638, 0502072639, 0502072640, 0502072641, 0502072642, 0502072734, 0502072735, 0502072736, 0502072737, 0502072738, 0502072739, 0502072740, 0502072741, 0502072742, 0502072784, 0502072786, 0502072787, 0502072788, 0502072789, 0502072790, 0502072791, 0502072792, 0502072793, 0502072794, 0502072795, 0502072796, 0502072797, 0502072806, 0502072807, 0502072808, 0502072809, 0502072810, 0502072811, 0502072812, 0502072813, 0502072814, 0502072815, 0502072816, 0503077098, 0503077099, 0503077100, 0503077119, 0503077120, 0503077121, 0503077122, 0503077123, 0503077124, 0503077125, 0503077126, 0503077127, 0503077128, 0503077129, 0503077130, 0503077156, 0503077157, 0503077158, 0503077159, 0503077160, 0503077161, 0503077162, 0503077163, 0503077164, 0503077165, 0503077166, 0503077167, 0503077168, 0503077169, 0503077170, 0503077171, 0503077172, 0503077173, 0503077174, 0503077175, 0503077176, 0503077177, 0503077178, 0503077222, 0503077223, 0503077224, 0503077225, 0503077226, 0503077227, 0503077228, 0503077229, 0503077230, 0503077231, 0503077232, 0503077290, 0503077291, 0503077292, 0503077293, 0503077294, 0503077295, 0503077296, 0503077297, 0503077298, 0503077299, 0503077300, 0503077301, 0503077302, 0503077303, 0503077304, 0503077308, 0503077309, 0503077311, 0503077317, 0503077321, 0503077419, 0503077420, 0503077421, 0503077422, 0503077423, 0503077424, 0503077425, 0503077426, 0503077427, 0503077428, 0503077429, 0503077430, 0503077431, 0503077432, 0503077433, 0504068553, 0504068554, 0504068555, 0504068565, 0504068580, 0504068585, 0504068586, 0504068587, 0504068588, 0504068589, 0504068590, 0504068591, 0504068592, 0504068593, 0504068594, 0504068595, 0504068596, 0504068597, 0504068598, 0504068599, 0504068600, 0504068601, 0504068602, 0504068603, 0504068604, 0504068635, 0504068678, 0504068679, 0504068680, 0504068681, 0504068682, 0504068683, 0504068684, 0504068685, 0504068686, 0504068687, 0504068688, 0504068689, 0504068690, 0504068691, 0504068692, 0504068727, 0504068728, 0504068729, 0504068730, 0504068731, 0504068732, 0504068733, 0504068734, 0504068735, 0504068736, 0504068737, 0504068738, 0504068739, 0504068740, 0504068741, 0504068742, 0504068743, 0504068744, 0504068745, 0504068746, 0504068747, 0504068748, 0504068749, 0504068750, 0504068751, 0504068752, 0504068753, 0504068754, 0504068755, 0504068759, 0504068760, 0504068761, 0504068762, 0504068763, 0504068764, 0505068519, 0505068520, 0505068521, 0505068522, 0505068523, 0505068524, 0505068525, 0505068526, 0505068527, 0505068528, 0505068529, 0505068530, 0505068531, 0505068532, 0505068533, 0505068534, 0505068536, 0505068537, 0505068538, 0505068539, 0505068540, 0505068541, 0505068542, 0505068543, 0505068544, 0505068545, 0505068546, 0505068592, 0505068593, 0505068594, 0505068595, 0505068596, 0505068597, 0505068598, 0505068599, 0505068622, 0505068627, 0505068628, 0505068629, 0505068630, 0505068631, 0505068644, 0505068645, 0505068646, 0505068647, 0505068648, 0505068649, 0505068650, 0505068651, 0505068703, 0505068704, 0505068712, 0505068713, 0505068714, 0505068715, 0505068716, 0505068717, 0505068718, 0505068719, 0505068720, 0505068721, 0505068730, 0505068731, 0505068732, 0505068733, 0505068734, 0505068735, 0505068736, 0505068737, 0505068738, 0505068739, 0505068740, 0505068741, 0505068742, 0505068743, 0505068744, 0505068786, 0505068798, 0505068803, 0505068804, 0505068805, 0505068806, 0505068807, 0505068808, 0505068809, 0505068810, 0505068821, 0505068837, 0505068838, 0505068839, 0505068855, 0505068856, 0505068857, 0505068858, 0505068859, 0505068860, 0505068861, 0505068862, 0506068558, 0506068559, 0506068560, 0506068561, 0506068562, 0506068563, 0506068564, 0506068565, 0506068566, 0506068567, 0506068568, 0506068569, 0506068570, 0506068571, 0506068572, 0506068573, 0506068574, 0506068575, 0506068576, 0506068577, 0506068578, 0506068579, 0506068580, 0506068581, 0506068582, 0506068622, 0506068623, 0506068624, 0506068625, 0506068626, 0506068627, 0506068628, 0506068629, 0506068630, 0506068631, 0506068632, 0506068633, 0506068634, 0506068635, 0506068636, 0506068658, 0506068659, 0506068660, 0506068661, 0506068662, 0506068663, 0506068664, 0506068665, 0506068666, 0506068667, 0506068668, 0506068669, 0506068670, 0506068671, 0506068672, 0506068673, 0506068674, 0506068675, 0506068676, 0506068677, 0506068678, 0506068679, 0506068680, 0506068681, 0506068682, 0506068683, 0506068684, 0506068686, 0506068687, 0506068694, 0506068695, 0506068696, 0506068697, 0506068698, 0506068699, 0506068700, 0506068701, 0506068702, 0506068703, 0506068704, 0506068705, 0506068706, 0506068707, 0506068708, 0506068709, 0506068710, 0506068711, 0506068712, 0506068713, 0506068774, 0506068775, 0506068776, 0506068782, 0506068783, 0506068784, 0506068785, 0506068786, 0506068787, 0506068788, 0506068789, 0506068790, 0506068791, 0506068792, 0506068793, 0506068794, 0506068795, 0506068796, 0506068797, 0506068798, 0506068799, 0506068800, 0506068801, 0506068804, 0506068822, 0506068823, 0506068824, 0506068825, 0506068826, 0506068827, 0506068864, 0506068890, 0506068891, 0506068892, 0506068893, 0506068894, 0507068607, 0507068609, 0507068610, 0507068611, 0507068612, 0507068699, 0507068700, 0507068701, 0507068702, 0507068703, 0507068704, 0507068705, 0507068706, 0507068707, 0507068708, 0507068709, 0507068710, 0507068711, 0507068712, 0507068713, 0507068714, 0507068715, 0507068776, 0507068777, 0507068778, 0507068779, 0507068780, 0507068781, 0507068782, 0507068783, 0507068784, 0507068785, 0507068786, 0507068787, 0507068788, 0507068789, 0507068790, 0507068791, 0507068796, 0507068797, 0507068798, 0507068799, 0507068800, 0507068801, 0507068802, 0507068803, 0507068804, 0507068805, 0507068825, 0507068826, 0508068540, 0508068541, 0508068560, 0508068561, 0508068562, 0508068565, 0508068566, 0508068572, 0508068573, 0508068574, 0508068575, 0508068576, 0508068577, 0508068578, 0508068579, 0508068580, 0508068581, 0508068582, 0508068614, 0508068615, 0508068616, 0508068617, 0508068618, 0508068619, 0508068620, 0508068621, 0508068622, 0508068623, 0508068624, 0508068625, 0508068626, 0508068627, 0508068628, 0508068629, 0508068630, 0508068631, 0508068632, 0508068633, 0508068634, 0508068635, 0508068636, 0508068637, 0508068638, 0508068639, 0508068640, 0508068641, 0508068642, 0508068643, 0508068644, 0508086520, 0508086521, 0508086522, 0508086523, 0508086524, 0508086525, 0508086528, 0508086529, 0508086530, 0508086531, 0508086532, 0508086533, 0508086534, 0508086535, 0508086536, 0508086537, 0508086538, 0508086539, 0508086540, 0508086541, 0508086542, 0508086635, 0508086636, 0508086637, 0508086638, 0508086664, 0508086671, 0508086672, 0508086675, 0508086676, 0508086685, 0508086686, 0508086687, 0508086688, 0509073506, 0509073507, 0509073508, 0509073509, 0509073510, 0509073511, 0509073512, 0509073513, 0509073514, 0509073515, 0509073516, 0509073517, 0509073518, 0509073519, 0509073520, 0509073521, 0509073522, 0509073523, 0509073524, 0509073525, 0509073526, 0509073527, 0509073528, 0509073529, 0509073530, 0509073531, 0509073532, 0509073533, 0509073534, 0509073535, 0509073536, 0509073537, 0509073538, 0509073539, 0509073540, 0509073541, 0509073542, 0509073543, 0509073544, 0509073607, 0509073608, 0509073609, 0509073610, 0509073611, 0509073612, 0509073613, 0509073614, 0509073615, 0509073616, 0509073617, 0509073618, 0509073619, 0509073620, 0509073621, 0509073622, 0509073623, 0509073624, 0509073625, 0509073626, 0509073627, 0509073628, 0509073629, 0509073630, 0509073635, 0509073636, 0509073637, 0509073638, 0509073639, 0509073640, 0509073641, 0509073642, 0509073643, 0509073644, 0509073645, 0509073646, 0509073647, 0509073648, 0509073649, 0509073650, 0509073651, 0509073652, 0509073653, 0509073654, 0509073655, 0509073656, 0509073657, 0509073658, 0509073659, 0509073660, 0509073661, 0509073662, 0509073663, 0509073664, 0509073665, 0509073666, 0509073667, 0509073668, 0509073669, 0509073670, 0509073671, 0509073672, 0509073673, 0509073674, 0509073675, 0509073787, 0509073788, 0509073801, 0509073802, 0509073803, 0509073804, 0509073805, 0509073806, 0509073807, 0509073810, 0509073811, 0509073812, 0509073813, 0509073814, 0509073815, 0509073857, 0509073858, 0509073859, 0509073879, 0509073880, 0509073882, 0509073899, 0509073900, 0509073901, 0509073902, 0509073903, 0509073904, 0509073927, 0509073928, 0510077537, 0510077538, 0510077539, 0510077540, 0510077541, 0510077542, 0510077543, 0510077544, 0510077545, 0510077546, 0510077547, 0510077548, 0510077549, 0510077550, 0510077551, 0510077552, 0510077553, 0510077554, 0510077555, 0510077556, 0510077557, 0510077558, 0510077559, 0510077560, 0510077561, 0510077562, 0510077563, 0510077564, 0510077565, 0510077566, 0510077567, 0510077568, 0510077569, 0510077570, 0510077571, 0510077572, 0510077573, 0510077574, 0510077575, 0510077576, 0510077577, 0510077578, 0510077579, 0510077580, 0510077581, 0510077613, 0510077627, 0510077628, 0510077629, 0510077630, 0510077631, 0510077632, 0510077633, 0510077634, 0510077635, 0510077636, 0510077637, 0510077638, 0510077639, 0510077640, 0510077641, 0510077642, 0510077643, 0510077644, 0510077645, 0510077646, 0510077647, 0510077648, 0510077649, 0510077650, 0510077651, 0510077652, 0510077653, 0510077654, 0510077655, 0510077656, 0510077657, 0510077658, 0510077659, 0510077660, 0510077661, 0510077662, 0510077663, 0510077664, 0510077665, 0510077666, 0510077667, 0510077668, 0510077669, 0510077670, 0510077671, 0510077672, 0510077673, 0510077674, 0510077675, 0510077676, 0510077677, 0510077678, 0510077679, 0510077680, 0510077681, 0510077682, 0510077683, 0510077684, 0510077685, 0510077686, 0510077687, 0510077688, 0510077689, 0510077690, 0510077691, 0510077692, 0510077693, 0510077694, 0510077695, 0510077696, 0510077697, 0510077698, 0510077699, 0511068523, 0511068527, 0511068528, 0511068529, 0511068530, 0511068531, 0511068541, 0511068542, 0511068544, 0511068545, 0511068546, 0511068547, 0511068548, 0511068568, 0511068569, 0511068570, 0511068609, 0511068610, 0511068611, 0511068612, 0511068613, 0511068614, 0511068615, 0511068616, 0511068617, 0511068618, 0511068619, 0511068620, 0511068621, 0511068622, 0511068623, 0511068636, 0511068642, 0511068657, 0511068659, 0511068660, 0511068661, 0511068662, 0511068663, 0511068664, 0511068665, 0511068666, 0511068667, 0511068668, 0511068696, 0511068703, 0511068704, 0511068705, 0511068706, 0511068707, 0511068708, 0511068711, 0511068713, 0511068714, 0511068723, 0511068724, 0511068725, 0511068726, 0511068727, 0511068728, 0511068729, 0511068730, 0511068731, 0511068732, 0511068733, 0511068734, 0511068737, 0511068738, 0511068739, 0511068740, 0511068741, 0511068742, 0511068799, 0511068800, 0511068801, 0511068802, 0511068803, 0511068804, 0511068805, 0511068806, 0511068821, 0511068822, 0511068823, 0511068824, 0511068826, 0511068827, 0511068828, 0511068829, 0511068830, 0511068854, 0511068855, 0511068856, 0511068857, 0511068858, 0511068859, 0511068860, 0511068861, 0511068875, 0511068876, 0511068877, 0511068878, 0511068879, 0511068880, 0511068889, 0511068890, 0511068891, 0511068892, 0511068893, 0511068894, 0511068895, 0511068896, 0511068897, 0511068898, 0511068899, 0511068900, 0511068901, 0511068902, 0511068903, 0511068904, 0511068905, 0511068906, 0511068907, 0511068908, 0511068909, 0511068910, 0511068911, 0511068912, 0511068913, 0512068537, 0512068538, 0512068539, 0512068540, 0512068541, 0512068542, 0512068543, 0512068544, 0512068545, 0512068546, 0512068547, 0512068548, 0512068549, 0512068550, 0512068551, 0512068552, 0512068567, 0512068568, 0512068622, 0512068623, 0512068663, 0512068664, 0512068665, 0512068666, 0512068671, 0512068672, 0512068673, 0512068674, 0512068675, 0512068676, 0512068677, 0512068678, 0512068679, 0512068680, 0512068681, 0512068682, 0512068729, 0512068730, 0512068731, 0512068732, 0512068733, 0512068734, 0512068735, 0512068736, 0512068737, 0512068738, 0512068739, 0512068741, 0512068742, 0512068743, 0512068744, 0512068745, 0512068746, 0512068747, 0512068748, 0512068749, 0512068752, 0512068753, 0512068754, 0512068755, 0512068756, 0512068757, 0512068758, 0512068759, 0512068760, 0512068762, 0512068763, 0512068764, 0512068765, 0512068766, 0512068767, 0512068768, 0512068769, 0512068770, 0512068771, 0512068772, 0512068773, 0512068774, 0512068775, 0512068776, 0512068777, 0512068778, 0601085035, 0601085036, 0601085037, 0601085038, 0601085039, 0601085040, 0601085041, 0601085042, 0601085043, 0601085044, 0601085045, 0601085046, 0601085047, 0601085048, 0601085049, 0601085050, 0601085051, 0601085064, 0601085100, 0601085101, 0601085102, 0601085117, 0601085118, 0601085120, 0601085121, 0601085122, 0601085123, 0601085124, 0601085125, 0601085126, 0601085135, 0601085136, 0601085137, 0601085138, 0601085139, 0601085140, 0601085141, 0601085142, 0601085143, 0601085195, 0601085196, 0601085197, 0601085198, 0601085199, 0601085200, 0601085201, 0601085202, 0601085203, 0601085204, 0601085205, 0601085206, 0601085207, 0601085208, 0601085209, 0601085210, 0601085211, 0601085212, 0601085213, 0601085214, 0601085231, 0601085232, 0601085233, 0601085234, 0601085235, 0601085236, 0601085237, 0601085252, 0601085288, 0601085289, 0601085290, 0601085291, 0601085295, 0601085296, 0601085298, 0601085299, 0601085303, 0601085304, 0601085305, 0601085306, 0601085307, 0601085320, 0601085322, 0601085323, 0601085342, 0601085343, 0601085344, 0601085345, 0601085346, 0601085347, 0601085348, 0601085349, 0601085350, 0601085351, 0601085352, 0601085353, 0601085354, 0601085355, 0601085356, 0601085357, 0602078508, 0602078523, 0602078524, 0602078525, 0602078526, 0602078527, 0602078528, 0602078529, 0602078530, 0602078531, 0602078532, 0602078541, 0602078550, 0602078551, 0602078552, 0602078642, 0602078664, 0602078666, 0602078667, 0602078668, 0602078669, 0602078670, 0602078671, 0602078672, 0602078673, 0602078674, 0602078675, 0602078706, 0602078707, 0602078708, 0602078709, 0602078710, 0602078711, 0602078712, 0602078713, 0602078714, 0602078715, 0602078716, 0602078727, 0602078728, 0602078729, 0602078730, 0602078731, 0602078732, 0602078733, 0602078734, 0602078735, 0602078736, 0602078737, 0602078738, 0602078739, 0602078740, 0602078741, 0602078742, 0602078743, 0602078744, 0602078745, 0602078761, 0602078762, 0602078763, 0602078764, 0602078765, 0602078766, 0602078767, 0602078768, 0602078769, 0602078770, 0602078771, 0602078772, 0602078773, 0602078774, 0602078775, 0602078777, 0602078778, 0602078779, 0602078780, 0603083514, 0603083515, 0603083516, 0603083517, 0603083518, 0603083519, 0603083520, 0603083521, 0603083522, 0603083523, 0603083524, 0603083525, 0603083526, 0603083527, 0603083528, 0603083539, 0603083540, 0603083541, 0603083542, 0603083543, 0603083544, 0603083545, 0603083546, 0603083562, 0603083563, 0603083564, 0603083565, 0603083566, 0603083567, 0603083568, 0603083569, 0603083570, 0603083571, 0603083572, 0603083573, 0603083586, 0603083587, 0603083588, 0603083589, 0603083590, 0603083591, 0603083592, 0603083593, 0603083594, 0603083595, 0603083596, 0603083597, 0603083598, 0603083599, 0603083600, 0603083601, 0603083602, 0603083603, 0603083604, 0603083605, 0603083606, 0603083607, 0603083608, 0603083609, 0603083610, 0603083611, 0603083612, 0603083613, 0603083614, 0603083615, 0603083616, 0603083617, 0603083618, 0603083619, 0603083620, 0603083621, 0603083622, 0603083623, 0603083624, 0603083625, 0603083626, 0603083635, 0603083636, 0603083637, 0603083638, 0603083643, 0603083644, 0603083645, 0603083646, 0603083647, 0603083648, 0603083649, 0603083650, 0603083651, 0603083652, 0603083653, 0603083654, 0603083662, 0603083663, 0603083664, 0603083665, 0603083666, 0603083667, 0603083668, 0603083669, 0603083670, 0603083671, 0603083672, 0603083673, 0603083674, 0603083675, 0603083686, 0603083693, 0603083694, 0603083695, 0603083696, 0603083697, 0603083698, 0603083719, 0603083720, 0603083721, 0603083760, 0603083761, 0603083782, 0603083783, 0603083784, 0603083785, 0603083786, 0603083787, 0603083788, 0603083789, 0603083790, 0603083791, 0603083792, 0603083793, 0603083796, 0603083799, 0603083800, 0603083801, 0603083802, 0603083803, 0603083804, 0603083805, 0603083806, 0603083807, 0603083808, 0603083809, 0603083810, 0603083811, 0603083812, 0603083813, 0603083814, 0603083815, 0603083816, 0603083817, 0603083818, 0603083819, 0603083820, 0603083823, 0603083824, 0603083825, 0603083826, 0603083852, 0603083853, 0603083866, 0603083867, 0603083869, 0603083870, 0603083871, 0603083872, 0603083873, 0603083874, 0603083875, 0603083876, 0604080514, 0604080516, 0604080547, 0604080548, 0604080585, 0604080586, 0604080588, 0604080589, 0604080590, 0604080591, 0604080592, 0604080593, 0604080594, 0604080595, 0604080596, 0604080597, 0604080598, 0604080599, 0604080600, 0604080601, 0604080602, 0604080603, 0604080604, 0604080605, 0604080606, 0604080608, 0604080660, 0604080661, 0604080662, 0604080663, 0604080664, 0604080665, 0604080666, 0604080667, 0604080668, 0604080669, 0604080670, 0604080671, 0604080672, 0604080673, 0604080674, 0604080675, 0604080676, 0604080677, 0604080678, 0604080679, 0604080680, 0604080681, 0604080682, 0604080683, 0604080684, 0604080685, 0604080686, 0604080687, 0604080688, 0604080689, 0604080690, 0604080691, 0604080692, 0604080693, 0604080694, 0604080695, 0604080696, 0604080697, 0604080698, 0604080699, 0604080700, 0604080701, 0604080702, 0604080703, 0604080704, 0604080705, 0604080706, 0604080707, 0604080708, 0604080709, 0604080710, 0604080711, 0604080712, 0604080713, 0604080714, 0604080715, 0604080716, 0604080717, 0604080718, 0604080719, 0604080720, 0604080721, 0604080722, 0604080760, 0604080761, 0604080762, 0604080763, 0604080764, 0604080765, 0604080766, 0604080767, 0604080768, 0604080769, 0604080770, 0604080771, 0604080772, 0604080773, 0604080774, 0604080775, 0604080776, 0604080777, 0604080778, 0604080779, 0604080780, 0604080809, 0604080810, 0604080811, 0604080812, 0604080813, 0604080814, 0604080815, 0604080816, 0604080817, 0604080818, 0605085011, 0605085012, 0605085013, 0605085055, 0605085068, 0605085069, 0605085070, 0605085071, 0605085072, 0605085073, 0605085074, 0605085075, 0605085076, 0605085077, 0605085078, 0605085120, 0605085121, 0605085122, 0605085123, 0605085124, 0605085125, 0605085126, 0605085127, 0605085128, 0605085149, 0605085150, 0605085151, 0605085152, 0605085153, 0605085154, 0605085155, 0605085156, 0605085157, 0605085158, 0605085159, 0605085160, 0605085161, 0605085162, 0605085163, 0605085164, 0605085165, 0605085166, 0605085213, 0605085214, 0605085215, 0605085233, 0605085234, 0605085235, 0605085236, 0605085237, 0605085239, 0605085240, 0605085241, 0605085242, 0605085243, 0605085244, 0605085245, 0605085246, 0605085247, 0605085248, 0605085249, 0605085250, 0605085251, 0605085252, 0605085253, 0605085274, 0605085275, 0605085276, 0605085277, 0605085278, 0605085279, 0605085280, 0605085281, 0605085282, 0605085283, 0605085284, 0605085285, 0605085286, 0605085287, 0605085288, 0605085289, 0605085290, 0605085291, 0605085292, 0605085293, 0605085294, 0605085295, 0605085296, 0605085297, 0605085298, 0605085299, 0605085300, 0605085301, 0605085315, 0605085318, 0605085319, 0605085320, 0605085321, 0605085322, 0605085331, 0605085332, 0605085333, 0605085334, 0605085335, 0605085336, 0605085337, 0605085338, 0605085339, 0605085340, 0605085341, 0605085342, 0605085350, 0605085351, 0605085352, 0605085353, 0605085354, 0605085355, 0605085356, 0605085357, 0605085358, 0605085359, 0605085360, 0605085361, 0605085362, 0605085363, 0605085364, 0605085365, 0605085366, 0605085367, 0605085368, 0605085369, 0605085370, 0605085371, 0605085372, 0605085373, 0605085374, 0605085375, 0605085376, 0605085377, 0605085378, 0605085379, 0605085380, 0605085381, 0606079502, 0606079503, 0606079504, 0606079505, 0606079506, 0606079507, 0606079519, 0606079520, 0606079521, 0606079522, 0606079523, 0606079570, 0606079571, 0606079572, 0606079573, 0606079574, 0606079575, 0606079576, 0606079577, 0606079578, 0606079579, 0606079580, 0606079581, 0606079668, 0606079669, 0606079670, 0606079671, 0606079672, 0606079673, 0606079708, 0606079709, 0606079710, 0606079711, 0606079712, 0606079713, 0606079714, 0606079715, 0606079756, 0606079757, 0606079758, 0606079759, 0606079760, 0606079761, 0606079762, 0606079763, 0606079768, 0606079796, 0606079797, 0606079798, 0606079799, 0606079800, 0606079801, 0606079802, 0606079803, 0606079804, 0606079805, 0606079806, 0606079807, 0606079808, 0606079809, 0606079810, 0606079811, 0606079812, 0606079813, 0606079814, 0606079815, 0606079857, 0606079858, 0606079859, 0606079860, 0607084501, 0607084502, 0607084503, 0607084504, 0607084505, 0607084506, 0607084507, 0607084561, 0607084562, 0607084576, 0607084589, 0607084590, 0607084591, 0607084592, 0607084593, 0607084594, 0607084595, 0607084598, 0607084599, 0607084600, 0607084614, 0607084615, 0607084616, 0607084617, 0607084626, 0607084648, 0607084649, 0607084650, 0607084651, 0607084652, 0607084653, 0607084654, 0607084655, 0607084706, 0607084707, 0607084708, 0607084709, 0607084710, 0607084711, 0607084712, 0607084713, 0607084720, 0607084721, 0607084722, 0607084723, 0607084724, 0607084725, 0607084726, 0607084727, 0607084728, 0607084729, 0607084730, 0607084731, 0607084732, 0607084733, 0607084734, 0607084735, 0607084736, 0607084737, 0607084738, 0607084739, 0607084740, 0607084741, 0607084742, 0607084743, 0607084744, 0607084745, 0607084746, 0607084747, 0607084748, 0607084749, 0607084750, 0607084751, 0607084752, 0607084753, 0607084754, 0607084755, 0607084756, 0607084757, 0607084758, 0607084759, 0607084760, 0607084761, 0607084762, 0608079035, 0608079037, 0608079038, 0608079039, 0608079040, 0608079041, 0608079042, 0608079043, 0608079044, 0608079045, 0608079046, 0608079047, 0608079048, 0608079049, 0608079050, 0608079051, 0608079052, 0608079053, 0608079054, 0608079055, 0608079056, 0608079057, 0608079058, 0608079059, 0608079060, 0608079061, 0608079085, 0608079086, 0608079087, 0608079088, 0608079109, 0608079225, 0608079226, 0608079227, 0608079228, 0608079229, 0608079230, 0608079231, 0608079232, 0608079233, 0608079234, 0608079235, 0608079236, 0608079237, 0608079262, 0608079263, 0608079264, 0608079265, 0608079269, 0608079270, 0608079271, 0608079278, 0608079279, 0608079280, 0608079281, 0608079282, 0608079284, 0608079285, 0608079286, 0608079293, 0608079294, 0608079295, 0608079296, 0608079308, 0608079309, 0608079310, 0608079311, 0608079312, 0608079313, 0608079314, 0608079315, 0608079316, 0608079317, 0608079318, 0608079319, 0608079320, 0608079321, 0608079322, 0608079323, 0608079327, 0608079328, 0608079329, 0608079330, 0608079331, 0608079332, 0608079333, 0608079334, 0608079335, 0608079336, 0608079337, 0608079338, 0608079341, 0608079342, 0608079374, 0608079375, 0608079376, 0608079377, 0608079378, 0608079379, 0608079380, 0608079381, 0608079382, 0608079383, 0608079384, 0608079385, 0608079386, 0608079387, 0608079388, 0608079389, 0608079390, 0608079391, 0608079392, 0608079393, 0608079394, 0608079395, 0608079396, 0608079397, 0608079418, 0608079419, 0609081024, 0609081063, 0609081064, 0609081077, 0609081078, 0609081079, 0609081080, 0609081081, 0609081082, 0609081083, 0609081084, 0609081086, 0609081087, 0609081098, 0609081142, 0609081143, 0609081144, 0609081145, 0609081146, 0609081147, 0609081148, 0609081149, 0609081150, 0609081151, 0609081152, 0609081153, 0609081154, 0609081155, 0609081156, 0609081157, 0609081158, 0609081159, 0609081160, 0609081161, 0609081162, 0609081163, 0609081164, 0609081165, 0609081166, 0609081167, 0609081168, 0609081169, 0609081170, 0609081171, 0609081172, 0609081173, 0609081174, 0609081175, 0609081176, 0609081238, 0609081239, 0609081262, 0609081337, 0609081338, 0609081339, 0609081340, 0609081341, 0609081342, 0609081343, 0609081344, 0609081345, 0609081346, 0610078525, 0610078527, 0610078528, 0610078529, 0610078530, 0610078531, 0610078539, 0610078540, 0610078541, 0610078542, 0610078543, 0610078544, 0610078545, 0610078546, 0610078547, 0610078548, 0610078549, 0610078550, 0610078551, 0610078552, 0610078553, 0610078575, 0610078579, 0610078580, 0610078581, 0610078582, 0610078583, 0610078584, 0610078585, 0610078587, 0610078588, 0610078589, 0610078590, 0610078591, 0610078592, 0610078593, 0610078594, 0610078595, 0610078596, 0610078597, 0610078598, 0610078601, 0610078602, 0610078603, 0610078604, 0610078605, 0610078606, 0610078607, 0610078608, 0610078609, 0610078610, 0610078611, 0610078612, 0610078613, 0610078614, 0610078615, 0610078616, 0610078617, 0610078618, 0610078619, 0610078620, 0610078621, 0610078672, 0610078673, 0610078674, 0610078675, 0610078676, 0610078677, 0610078678, 0610078679, 0610078680, 0610078681, 0610078682, 0610078727, 0610078733, 0610078734, 0610078735, 0610078736, 0610078737, 0610078738, 0610078739, 0610078740, 0610078777, 0610078778, 0610078779, 0610078780, 0610078781, 0610078782, 0610078783, 0610078784, 0610078785, 0610078786, 0610078787, 0610078788, 0610078789, 0610078790, 0610078791, 0610078792, 0610078793, 0610078794, 0610078795, 0610078796, 0610078797, 0610078798, 0610078803, 0610078804, 0610078805, 0610078839, 0611082024, 0611082025, 0611082026, 0611082027, 0611082028, 0611082029, 0611082030, 0611082031, 0611082032, 0611082033, 0611082034, 0611082035, 0611082036, 0611082037, 0611082038, 0611082039, 0611082040, 0611082041, 0611082042, 0611082043, 0611082044, 0611082045, 0611082046, 0611082047, 0611082048, 0611082049, 0611082050, 0611082051, 0611082052, 0611082142, 0611082143, 0611082144, 0611082145, 0611082150, 0611082151, 0611082152, 0611082153, 0611082154, 0611082155, 0611082156, 0611082157, 0611082158, 0611082159, 0611082207, 0611082208, 0611082209, 0611082210, 0611082211, 0611082212, 0611082213, 0611082220, 0611082221, 0611082222, 0611082223, 0611082224, 0611082239, 0611082240, 0611082241, 0611082242, 0611082243, 0611082244, 0611082245, 0611082246, 0611082247, 0611082248, 0611082249, 0611082250, 0611082251, 0611082252, 0611082253, 0611082254, 0611082255, 0611082256, 0611082257, 0611082258, 0611082259, 0611082260, 0611082261, 0611082262, 0611082263, 0611082264, 0611082284, 0611082285, 0611082286, 0611082287, 0611082288, 0611082289, 0611082290, 0611082291, 0611082292, 0611082293, 0611082294, 0611082295, 0611082296, 0611082297, 0611082298, 0611082299, 0611082300, 0611082301, 0612080023, 0612080024, 0612080025, 0612080026, 0612080027, 0612080028, 0612080033, 0612080034, 0612080035, 0612080072, 0612080073, 0612080074, 0612080075, 0612080076, 0612080077, 0612080078, 0612080079, 0612080099, 0612080100, 0612080101, 0612080102, 0612080103, 0612080104, 0612080105, 0612080106, 0612080107, 0612080108, 0612080109, 0612080110, 0612080111, 0612080112, 0612080129, 0612080130, 0612080145, 0612080146, 0612080147, 0612080148, 0612080149, 0612080150, 0612080151, 0612080152, 0612080153, 0612080154, 0612080155, 0612080156, 0612080157, 0612080158, 0612080198, 0612080199, 0612080200, 0612080297, 0612080298, 0612080299, 0612080300, 0612080301, 0612080302, 0612080303, 0612080304, 0701096588, 0701096589, 0701096590, 0701096591, 0701096592, 0701096593, 0701096594, 0701096595, 0701096596, 0701096597, 0701096598, 0701096599, 0701096600, 0701096601, 0701096602, 0701096603, 0701096604, 0701096605, 0701096606, 0701096607, 0701096608, 0701096609, 0701096610, 0701096611, 0701096612, 0701096613, 0701096614, 0701096615, 0701096616, 0701096638, 0701096639, 0701096643, 0701096644, 0701096645, 0701096646, 0701096647, 0701096648, 0701096649, 0701096650, 0701096651, 0701096655, 0701096656, 0701096657, 0701096658, 0701096659, 0701096660, 0701096662, 0701096663, 0701096724, 0701096745, 0701096775, 0701096776, 0701096777, 0701096778, 0701096779, 0701096780, 0701096792, 0701096793, 0701096831, 0701096832, 0701096833, 0701096834, 0702088505, 0702088506, 0702088507, 0702088570, 0702088571, 0702088572, 0702088573, 0702088574, 0702088576, 0702088609, 0702088610, 0702088611, 0702088666, 0702088667, 0702088668, 0702088669, 0702088670, 0702088671, 0702088672, 0702088673, 0702088674, 0702088688, 0702088690, 0702088691, 0702088692, 0703090001, 0703090002, 0703090003, 0703090004, 0703090010, 0703090011, 0703090012, 0703090013, 0703090014, 0703090015, 0703090016, 0703090017, 0703090018, 0703090019, 0703090020, 0703090021, 0703090022, 0703090023, 0703090024, 0703090025, 0703090026, 0703090027, 0703090028, 0703090029, 0703090030, 0703090031, 0703090032, 0703090033, 0703090066, 0703090067, 0703090068, 0703090069, 0703090070, 0703090071, 0703090080, 0703090081, 0703090082, 0703090083, 0703090084, 0703090147, 0703090150, 0703090151, 0703090163, 0703090164, 0703090165, 0703090166, 0703090167, 0703090168, 0703090169, 0703090170, 0703090171, 0703090172, 0703090173, 0703090174, 0703090224, 0703090238, 0703090255, 0703090256, 0703090257, 0703090258, 0703090259, 0703090260, 0703090261, 0703090262, 0703090263, 0703090264, 0703090265, 0703090266, 0703090267, 0703090268, 0704092518, 0704092519, 0704092520, 0704092521, 0704092526, 0704092527, 0704092528, 0704092529, 0704092530, 0704092531, 0704092532, 0704092533, 0704092534, 0704092535, 0704092536, 0704092545, 0704092546, 0704092547, 0704092548, 0704092549, 0704092552, 0704092553, 0704092554, 0704092555, 0704092556, 0704092557, 0704092558, 0704092559, 0704092560, 0704092561, 0704092584, 0704092628, 0704092629, 0704092630, 0704092631, 0704092632, 0704092633, 0704092634, 0704092635, 0704092636, 0704092637, 0704092638, 0704092639, 0704092640, 0704092641, 0704092642, 0704092644, 0704092645, 0704092646, 0704092647, 0704092678, 0704092679, 0704092680, 0704092681, 0704092682, 0704092683, 0704092684, 0704092685, 0704092686, 0704092687, 0704092727, 0704092728, 0704092729, 0704092730, 0704092731, 0704092732, 0704092733, 0704092734, 0704092735, 0704092736, 0704092737, 0705096527, 0705096528, 0705096529, 0705096530, 0705096531, 0705096532, 0705096533, 0705096534, 0705096535, 0705096536, 0705096537, 0705096539, 0705096540, 0705096541, 0705096542, 0705096543, 0705096544, 0705096545, 0705096566, 0705096567, 0705096583, 0705096584, 0705096585, 0705096586, 0705096587, 0705096588, 0705096589, 0705096590, 0705096591, 0705096592, 0705096593, 0705096643, 0705096644, 0705096645, 0705096646, 0705096647, 0705096648, 0705096649, 0705096652, 0705096667, 0705096668, 0705096669, 0705096673, 0705096674, 0705096743, 0705096744, 0705096775, 0705096776, 0705096777, 0705096778, 0705096779, 0705096780, 0705096781, 0705096782, 0705096783, 0705096784, 0705096785, 0705096786, 0705096787, 0706088514, 0706088515, 0706088516, 0706088519, 0706088520, 0706088521, 0706088522, 0706088523, 0706088524, 0706088525, 0706088526, 0706088527, 0706088528, 0706088529, 0706088530, 0706088531, 0706088556, 0706088557, 0706088558, 0706088559, 0706088560, 0706088561, 0706088562, 0706088563, 0706088564, 0706088565, 0706088566, 0706088567, 0706088614, 0706088615, 0706088616, 0706088617, 0706088618, 0706088619, 0706088620, 0706088621, 0706088622, 0706088623, 0706088624, 0706088625, 0706088661, 0706088662, 0706088663, 0706088664, 0706088665, 0706088706, 0706088707, 0706088708, 0706088709, 0706088710, 0706088711, 0706088712, 0706088713, 0706088714, 0706088715, 0706088716, 0706088717, 0706088718, 0706088719, 0706088720, 0706088721, 0706088722, 0706088734, 0706088735, 0706088736, 0706088737, 0707092524, 0707092525, 0707092526, 0707092527, 0707092528, 0707092529, 0707092530, 0707092531, 0707092551, 0707092552, 0707092553, 0707092554, 0707092555, 0707092556, 0707092557, 0707092558, 0707092559, 0707092560, 0707092561, 0707092562, 0707092563, 0707092564, 0707092565, 0707092566, 0707092569, 0707092570, 0707092571, 0707092572, 0707092573, 0707092574, 0707092575, 0707092576, 0707092577, 0707092578, 0707092579, 0707092580, 0707092581, 0707092582, 0707092583, 0707092584, 0707092585, 0707092586, 0707092587, 0707092646, 0707092647, 0707092648, 0707092649, 0707092650, 0707092651, 0707092652, 0707092653, 0707092666, 0707092667, 0707092668, 0707092669, 0707092670, 0707092671, 0707092672, 0707092673, 0707092674, 0707092675, 0707092711, 0708094024, 0708094025, 0708094026, 0708094027, 0708094028, 0708094029, 0708094030, 0708094031, 0708094034, 0708094035, 0708094040, 0708094041, 0708094042, 0708094043, 0708094044, 0708094045, 0708094046, 0708094047, 0708094048, 0708094049, 0708094050, 0708094051, 0708094052, 0708094053, 0708094054, 0708094085, 0708094086, 0708094087, 0708094095, 0708094096, 0708094225, 0708094226, 0708094227, 0708094251, 0708094252, 0708094253, 0708094254, 0708094255, 0708094256, 0708094257, 0708094258, 0708094278, 0708094279, 0708094280, 0708094281, 0708094282, 0709089023, 0709089028, 0709089029, 0709089048, 0709089090, 0709089095, 0709089140, 0709089141, 0709089142, 0709089143, 0709089144, 0709089196, 0709089197, 0709089198, 0709089199, 0709089200, 0709089201, 0709089202, 0709089203, 0709089204, 0709089205, 0709089206, 0709089207, 0709089208, 0710098047, 0710098048, 0710098049, 0710098050, 0710098051, 0710098052, 0710098053, 0710098054, 0710098055, 0710098056, 0710098057, 0710098058, 0710098059, 0710098060, 0710098061, 0710098062, 0710098063, 0710098064, 0710098073, 0710098074, 0710098075, 0710098076, 0710098077, 0710098078, 0710098079, 0710098080, 0710098081, 0710098082, 0710098083, 0710098084, 0710098085, 0710098086, 0710098087, 0710098088, 0710098096, 0710098103, 0710098104, 0710098105, 0710098106, 0710098107, 0710098108, 0710098111, 0710098112, 0710098113, 0710098114, 0710098115, 0710098116, 0710098117, 0710098118, 0710098119, 0710098120, 0710098121, 0710098122, 0710098156, 0710098157, 0710098158, 0710098159, 0710098160, 0710098161, 0710098162, 0710098163, 0710098198, 0710098205, 0711096003, 0711096004, 0711096005, 0711096006, 0711096007, 0711096008, 0711096009, 0711096010, 0711096011, 0711096012, 0711096013, 0711096014, 0711096015, 0711096016, 0711096083, 0711096111, 0711096112, 0711096113, 0711096114, 0711096115, 0711096116, 0711096120, 0711096121, 0711096122, 0711096123, 0711096124, 0711096125, 0711096126, 0711096127, 0711096128, 0711096129, 0711096130, 0711096131, 0711096133, 0711096134, 0711096135, 0711096136, 0711096137, 0711096138, 0711096139, 0711096140, 0711096151, 0711096152, 0711096153, 0711096154, 0711096155, 0711096156, 0711096157, 0711096158, 0711096159, 0711096160, 0711096161, 0711096162, 0711096163, 0711096164, 0711096165, 0711096166, 0711096178, 0711096179, 0711096180, 0711096181, 0711096182, 0711096183, 0711096219, 0711096220, 0711096221, 0711096222, 0711096223, 0711096224, 0711096225, 0711096228, 0711096229, 0711096230, 0711096231, 0711096232, 0711096245, 0711096246, 0711096247, 0711096248, 0711096249, 0711096250, 0711096251, 0712094024, 0712094025, 0712094026, 0712094027, 0712094028, 0712094035, 0712094036, 0712094037, 0712094038, 0712094039, 0712094040, 0712094041, 0712094127, 0712094128, 0712094129, 0712094130, 0712094131, 0712094132, 0712094133, 0712094134, 0712094135, 0712094136, 0712094137, 0712094138, 0712094139, 0712094140, 0712094141, 0712094148, 0712094149, 0712094150, 0712094151, 0712094152, 0712094153, 0712094154, 0712094155, 0712094177, 0712094178, 0712094179, 0712094180, 0712094181, 0712094182, 0712094183, 0712094184, 0712094185, 0712094186, 0712094187, 0712094188, 0712094189, 0712094190, 0712094191, 0712094192, 0712094193, 0712094194, 0712094195, 0712094196, 0712094197, 0712094198, 0712094199, 0712094200, 0712094201, 0712094202, 0712094203, 0712094204, 0712094205, 0712094206, 0712094219, 0712094220, 0712094262, 0712094263, 0712094264, 0712094265, 0712094266, 0712094267, 0712094291, 0712094292, 0801098519, 0801098520, 0801098521, 0801098522, 0801098523, 0801098524, 0801098525, 0801098548, 0801098549, 0801098550, 0801098551, 0801098552, 0801098553, 0801098554, 0801098555, 0801098556, 0801098557, 0801098558, 0801098559, 0801098560, 0801098561, 0801098562, 0801098569, 0801098570, 0801098571, 0801098572, 0801098573, 0801098574, 0801098575, 0801098576, 0801098577, 0801098578, 0801098579, 0801098580, 0801098581, 0801098582, 0801098583, 0801098584, 0801098585, 0801098586, 0801098587, 0801098588, 0801098589, 0801098590, 0801098591, 0801098592, 0801098593, 0801098594, 0801098595, 0801098596, 0801098597, 0801098598, 0801098599, 0801098600, 0801098601, 0801098602, 0801098603, 0801098604, 0801098605, 0801098606, 0801098607, 0801098608, 0801098609, 0801098610, 0801098611, 0801098612, 0801098613, 0801098614, 0801098615, 0801098616, 0801098617, 0801098618, 0801098619, 0801098620, 0801098621, 0801098622, 0801098623, 0801098624, 0801098625, 0801098677, 0801098678, 0801098679, 0801098680, 0801098681, 0801098682, 0801098683, 0801098684, 0801098685, 0801098686, 0801098687, 0801098688, 0801098689, 0801098690, 0801098691, 0801098692, 0801098693, 0801098694, 0801098695, 0801098696, 0801098697, 0801098698, 0801098699, 0801098700, 0801098701, 0801098702, 0801098703, 0801098704, 0801098705, 0801098706, 0801098731, 0801098732, 0801098781, 0801098782, 0802102002, 0802102003, 0802102004, 0802102005, 0802102006, 0802102007, 0802102008, 0802102009, 0802102010, 0802102011, 0802102012, 0802102013, 0802102014, 0802102015, 0802102016, 0802102028, 0802102029, 0802102030, 0802102031, 0802102032, 0802102033, 0802102034, 0802102035, 0802102036, 0802102037, 0802102038, 0802102039, 0802102040, 0802102041, 0802102042, 0802102043, 0802102044, 0802102045, 0802102046, 0802102047, 0802102048, 0802102049, 0802102050, 0802102051, 0802102052, 0802102053, 0802102054, 0802102055, 0802102056, 0802102057, 0802102058, 0802102059, 0802102060, 0802102061, 0802102062, 0802102063, 0802102064, 0802102065, 0802102066, 0802102067, 0802102068, 0802102069, 0802102070, 0802102071, 0802102072, 0802102073, 0802102074, 0802102075, 0802102076, 0802102077, 0802102078, 0802102079, 0802102080, 0802102081, 0802102082, 0802102083, 0802102084, 0802102085, 0802102086, 0802102087, 0802102088, 0802102089, 0802102090, 0802102091, 0802102092, 0802102093, 0802102094, 0802102095, 0802102096, 0802102097, 0802102098, 0802102099, 0802102100, 0802102101, 0802102102, 0802102104, 0802102105, 0802102106, 0802102107, 0802102108, 0802102109, 0802102110, 0802102111, 0802102112, 0802102113, 0802102114, 0802102115, 0802102116, 0802102117, 0802102118, 0802102119, 0802102120, 0802102121, 0802102122, 0802102123, 0802102124, 0802102125, 0802102126, 0802102127, 0802102128, 0802102129, 0802102130, 0802102131, 0802102132, 0802102133, 0802102134, 0802102135, 0802102136, 0802102137, 0802102138, 0802102139, 0802102140, 0802102141, 0802102142, 0802102143, 0802102144, 0802102145, 0802102146, 0802102147, 0802102148, 0802102149, 0802102150, 0802102151, 0802102152, 0802102153, 0802102154, 0802102155, 0802102156, 0802102157, 0802102158, 0802102159, 0802102160, 0802102161, 0802102162, 0802102163, 0802102183, 0802102230, 0802102231, 0803101529, 0803101530, 0803101531, 0803101532, 0803101533, 0803101534, 0803101535, 0803101536, 0803101537, 0803101538, 0803101539, 0803101540, 0803101541, 0803101542, 0803101543, 0803101544, 0803101545, 0803101546, 0803101547, 0803101548, 0803101549, 0803101550, 0803101551, 0803101552, 0803101553, 0803101554, 0803101555, 0803101556, 0803101557, 0803101558, 0803101574, 0803101575, 0803101576, 0803101577, 0803101578, 0803101579, 0803101580, 0803101581, 0803101582, 0803101583, 0803101584, 0803101585, 0803101594, 0803101595, 0803101596, 0803101597, 0803101598, 0803101599, 0803101600, 0803101601, 0803101602, 0803101603, 0803101604, 0803101605, 0803101606, 0803101607, 0803101608, 0803101609, 0803101610, 0803101611, 0803101714, 0803101715, 0803101716, 0803101717, 0803101718, 0803101719, 0803101720, 0803101721, 0803101722, 0803101723, 0803101724, 0803101725, 0803101726, 0803101727, 0803101742, 0803101743, 0803101744, 0803101745, 0803101746, 0803101747, 0803101748, 0804107001, 0804107002, 0804107127, 0804107128, 0804107129, 0804107130, 0804107131, 0804107132, 0804107133, 0804107134, 0804107135, 0804107136, 0804107137, 0804107138, 0804107139, 0804107140, 0804107141, 0804107142, 0804107155, 0804107156, 0804107157, 0804107158, 0804107159, 0804107160, 0804107161, 0804107162, 0804107163, 0804107164, 0804107211, 0804107212, 0804107213, 0804107214, 0804107215, 0804107291, 0804107292, 0804107293, 0804107294, 0804107295, 0804107296, 0804107297, 0804107298, 0804107299, 0804107300, 0804107301, 0804107302, 0804107303, 0804107304, 0804107305, 0804107306, 0804107307, 0804107308, 0804107309, 0804107310, 0804107322, 0804107323, 0804107324, 0804107325, 0804107326, 0805099012, 0805099013, 0805099014, 0805099015, 0805099016, 0805099026, 0805099027, 0805099028, 0805099029, 0805099030, 0805099032, 0805099033, 0805099034, 0805099035, 0805099036, 0805099089, 0805099090, 0805099091, 0805099092, 0805099105, 0805099106, 0805099149, 0805099150, 0805099200, 0805099201, 0805099202, 0805099203, 0805099204, 0805099205, 0805099206, 0805099207, 0805099208, 0805099209, 0805099210, 0805099211, 0805099237, 0805099238, 0805099239, 0805099240, 0805099241, 0805099242, 0805099243, 0805099244, 0805099245, 0805099246, 0805099247, 0805099248, 0806108001, 0806108002, 0806108003, 0806108004, 0806108005, 0806108006, 0806108007, 0806108008, 0806108009, 0806108010, 0806108208, 0806108323, 0806108324, 0807105543, 0807105544, 0807105545, 0807105546, 0807105547, 0807105575, 0807105576, 0807105597, 0807105598, 0807105608, 0807105609, 0807105610, 0807105611, 0807105612, 0807105613, 0807105614, 0807105615, 0807105616, 0807105617, 0807105629, 0807105630, 0807105631, 0807105632, 0807105633, 0807105634, 0807105635, 0807105636, 0807105637, 0807105638, 0808107001, 0808107042, 0808107043, 0809104501, 0809104502, 0809104503, 0809104504, 0809104505, 0809104506, 0809104577, 0809104578, 0809104597, 0809104724, 0809104725, 0809104779, 0810100428, 0810100448, 0810100449, 0810100494, 0810100495, 0810100496, 0810100497, 0810100498, 0810100499, 0810100500, 0811100134, 0811100135, 0811100137, 0811100138, 0811100139, 0811100140, 0811100141, 0811100142, 0811100143, 0811100144, 0811100145, 0811100146, 0811100147, 0811100148, 0811100203, 0811100204, 0811100205, 0811100206, 0812100084, 0812100085, 0812100086, 0812100087, 0812100088, 0812100089, 0812100090, 0812100091, 0812100092, 0812100093, 0812100094, 0812100095, 0812100096, 0812100097, 0812100098, 0812100099, 0812100100, 0812100101, 0812100102, 0812100103, 0812100104, 0812100105, 0812100106, 0812100107, 0812100108, 0812100109, 0812100110, 0812100111, 0812100112, 0812100113, 0812100114, 0812100115, 0812100116, 0812100117, 0812100118, 0812100119, 0812100120, 0812100121, 0812100122, 0812100123, 0812100124, 0812100125, 0812100179, 0812100180, 0812100181, 0812100182, 0812100183, 0812100191, 0812100192, 0901034504, 0901034515, 0901034532, 0901034559, 0901034560, 0901034561, 0901034562, 0901034563, 0901034564, 0901034565, 0901034566, 0901034567, 0901034568, 0901034569, 0901034570, 0901034571, 0902033536, 0902033537, 0902033538, 0902033539, 0902033540, 0902033541, 0902033542, 0902033543, 0902033544, 0902033545, 0902033629, 0903032001, 0903032002, 0903032003, 0903032004, 0903032005, 0903032006, 0903032032, 0903032033, 0903032034, 0903032035, 0903032054, 0904035023, 0904035024 and 0904035106,. |
Recalling Firm/ Manufacturer |
Stryker Medical Div. of Stryker Corporation 3800 E. Centre Ave. Portage MI 49002
|
For Additional Information Contact |
Derick Elliott 269-324-6609
|
Manufacturer Reason for Recall |
The Fowler mechanism may malfunction, preventing it from being lowered into a flat position.
|
FDA Determined Cause 2 |
Component change control |
Action |
Stryker Medical sent an Urgent Medical Device Alert letter dated May 14, 2010 to their customers. The letter identified the product, the problem, and the actions customers should take.
The letter instructed users to do the following:
" Return the postcard included with the letter to confirm receipt of the
notification.
" Locate the recalled beds, verify the Fowler mechanism is functioning properly and does
not display uneven motion when raising and lowering, and is not unusually
noisy.
" If the Fowler mechanism is not operating correctly, they were to call Stryker at
1-800-STRYKER, option 1, Reference RA-2010-014 to obtain service for the
bed.
" If any beds have been loaned or sold, the customer is to forward a copy of the
notice to the new users and advise the firm of the new location.
" Provide the firm with the serial numbers for any beds that were disposed of.
" Retain the revised maintenance manual included with the notification.
If the customer had any urgent questions or concerns, they were instructed to call the firm at 800-STRYKER, option 1, reference RA-2010-014, Monday-Friday 8am to 5pm (EST). |
Quantity in Commerce |
4,926 |
Distribution |
Worldwide Distribution - USA (all states), and including the countries of ARGENTINA, BRAZIL, CHILE, CHINA, CANADA, MEXICO, GREECE, JAPAN, KOREA, PUERTO RICO, SPAIN, POLAND, TAIWAN, SINGAPORE, INDIA, SWITZERLAND, UK and MALAYSIA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|