| | Class 1 Recall Symbiq Infuser |  |
| Date Posted |
September 08, 2010 |
| Recall Number |
Z-2381-2010 |
| Product |
Hospira Inc. Symbiq One- and Two- Channel Infuser, List Numbers 16026 and 16027 respectively. |
| Code Information |
List number configurations: 16026-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88, 89/90; 16026-27-89/90, 16026-13-25/26, 53/54; 16027-27-89/90; 16027-13-25/26, 53/54.
|
Recalling Firm/ Manufacturer |
Hospira Inc 755 Jarvis Dr Morgan Hill, California 95037-2810 |
| Consumer Instructions |
Contact the recalling firm for information |
Reason for Recall |
If the administrator set is removed prior to the cassette carriage fully opening without closing the slide/roller clamp, unrestricted flow and delivery may result.
|
| Action |
Hospitra Urgent Device Correction Notification mailed on 03/26/2010 via federal express. Tracking done by delivery tracking. Distribution of product continues and has not stopped. Additional notification was issued in 09/2010. A software correction is planned. |
| Quantity in Commerce |
28485 units |
| Distribution |
Worldwide distribution: USA, Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Puerto Rico, Singapore and the UAE. |
| |
|
|
|
|