• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Varian Eclipse

  • Print
  • Share
  • E-mail
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections

New Search Back to Search Results
 Class 2 Recall
Varian Eclipse
see related information
Date Posted January 18, 2011
Recall Status1 Terminated on July 21, 2011
Recall Number Z-0911-2011
Recall Event ID 56772
Premarket Notification
510(K) Number
Product Classification System,Planning,Radiation Therapy Treatment - Product Code MUJ
Product Varian Medical Systems, Eclipse Client software versions with build numbers from 7.3.10 through 8.6.15 (i.e. excluding 8.6.17); Model Number: H48; Radiation Treatment Planning System
Code Information LIST OF SERIAL NUMBERS: H480001 ,H480065 ,H480109 ,H480158 ,H480201 ,H480252 ,H480301 ,H480353 ,H480416 ,H480469 ,H480002 ,H480066 ,H480110 ,H480159 ,H480202 ,H480253 ,H480302 ,H480354 ,H480417 ,H480470 ,H480003 ,H480067 ,H480112 ,H480160 ,H480203 ,H480254 ,H480303 ,H480355 ,H480418 ,H480471 ,H480004 ,H480068 ,H480113 ,H480161 ,H480204 ,H480255 ,H480305 ,H480356 ,H480419 ,H480477 ,H480005 ,H480069 ,H480115 ,H480162 ,H480205 ,H480258 ,H480306 ,H480357 ,H480420 ,H480478 ,H480006 ,H480070 ,H480116 ,H480163 ,H480206 ,H480259 ,H480311 ,H480358 ,H480421 ,H480481 ,H480007 ,H480072 ,H480117 ,H480164 ,H480207 ,H480260 ,H480312 ,H480359 ,H480422 ,H480482 ,H480008 ,H480073 ,H480118 ,H480167 ,H480208 ,H480261 ,H480313 ,H480360 ,H480423 ,H480483 ,H480009 ,H480074 ,H480119 ,H480168 ,H480209 ,H480262 ,H480314 ,H480361 ,H480424 ,H480484 ,H480010 ,H480075 ,H480121 ,H480170 ,H480210 ,H480263 ,H480315 ,H480365 ,H480425 ,H480486 ,H480011 ,H480076 ,H480122 ,H480171 ,H480212 ,H480264 ,H480316 ,H480369 ,H480427 ,H480487 ,H480015 ,H480077 ,H480123 ,H480172 ,H480213 ,H480265 ,H480317 ,H480370 ,H480428 ,H480488 ,H480016 ,H480078 ,H480124 ,H480174 ,H480215 ,H480266 ,H480318 ,H480371 ,H480429 ,H480489 ,H480017 ,H480079 ,H480125 ,H480175 ,H480216 ,H480267 ,H480319 ,H480372 ,H480430 ,H480490 ,H480018 ,H480080 ,H480126 ,H480176 ,H480217 ,H480268 ,H480320 ,H480373 ,H480432 ,H480491 ,H480019 ,H480081 ,H480127 ,H480177 ,H480218 ,H480269 ,H480321 ,H480374 ,H480433 ,H480492 ,H480020 ,H480082 ,H480128 ,H480178 ,H480219 ,H480270 ,H480323 ,H480375 ,H480434 ,H480494 ,H480021 ,H480083 ,H480131 ,H480179 ,H480220 ,H480271 ,H480324 ,H480376 ,H480435 ,H480495 ,H480022 ,H480084 ,H480137 ,H480181 ,H480221 ,H480272 ,H480325 ,H480377 ,H480436 ,H480496 ,H480026 ,H480085 ,H480138 ,H480182 ,H480222 ,H480277 ,H480327 ,H480378 ,H480437 ,H480497 ,H480027 ,H480086 ,H480139 ,H480183 ,H480223 ,H480278 ,H480329 ,H480379 ,H480443 ,H480498 ,H480028 ,H480087 ,H480140 ,H480184 ,H480224 ,H480279 ,H480330 ,H480380 ,H480444 ,H480502 ,H480030 ,H480088 ,H480141 ,H480185 ,H480225 ,H480280 ,H480331 ,H480381 ,H480445 ,H480503 ,H480031 ,H480089 ,H480142 ,H480186 ,H480226 ,H480281 ,H480332 ,H480382 ,H480446 ,H480504 ,H480032 ,H480090 ,H480143 ,H480187 ,H480227 ,H480282 ,H480333 ,H480385 ,H480447 ,H480505 ,H480033 ,H480091 ,H480144 ,H480188 ,H480228 ,H480283 ,H480334 ,H480400 ,H480448 ,H480506 ,H480040 ,H480092 ,H480145 ,H480189 ,H480229 ,H480284 ,H480335 ,H480403 ,H480455 ,H480507 ,H480043 ,H480093 ,H480147 ,H480190 ,H480230 ,H480285 ,H480336 ,H480404 ,H480456 ,H480508 ,H480045 ,H480094 ,H480148 ,H480191 ,H480238 ,H480286 ,H480337 ,H480405 ,H480459 ,H480510 ,H480046 ,H480095 ,H480149 ,H480192 ,H480239 ,H480287 ,H480338 ,H480406 ,H480460 ,H480511 ,H480047 ,H480096 ,H480150 ,H480193 ,H480240 ,H480288 ,H480340M ,H480408 ,H480461 ,H480512 ,H480048 ,H480098 ,H480151 ,H480194 ,H480241 ,H480289 ,H480340R1 ,H480409 ,H480462 ,H480513 ,H480049 ,H480099 ,H480152 ,H480195 ,H480242 ,H480294 ,H480341 ,H480410 ,H480463 ,H480514 ,H480050 ,H480100 ,H480153 ,H480196 ,H480244 ,H480295 ,H480342 ,H480411 ,H480464 ,H480518 ,H480060 ,H480104 ,H480154 ,H480197 ,H480245 ,H480296 ,H480343 ,H480412 ,H480465 ,H480519 ,H480061 ,H480105 ,H480155 ,H480198 ,H480246 ,H480297 ,H480344 ,H480413 ,H480466 ,H480521 ,H480063 ,H480106 ,H480156 ,H480199 ,H480250 ,H480298 ,H480351 ,H480414 ,H480467 ,H480522 ,H480064 ,H480108 ,H480157 ,H480200 ,H480251 ,H480300 ,H480352 ,H480415 ,H480468 ,H480523 ,H480524 ,H480596 ,H480652 ,H480730 ,H480796 ,H480850 ,H480908 ,H480960 ,H481018 ,H481079 ,H480525 ,H480599 ,H480656 ,H480731 ,H480797 ,H480852 ,H480909 ,H480961 ,H481019 ,H481080 ,H480526 ,H480600 ,H480657 ,H480732 ,H480798 ,H480853 ,H480910 ,H480963 ,H481020 ,H481081 ,H480529 ,H480601 ,H480658 ,H480733 ,H480799 ,H480854 ,H480911 ,H480964 ,H481021 ,H481082 ,H480533 ,H480602 ,H480659 ,H480734 ,H480802 ,H480855 ,H480912 ,H480965 ,H481022 ,H481083 ,H480534 ,H480603 ,H480660 ,H480735 ,H480803 ,H480856 ,H480913 ,H480966 ,H481023 ,H481084 ,H480535 ,H480604 ,H480661 ,H480738 ,H480804 ,H480857 ,H480916 ,H480967 ,H481024 ,H481085 ,H480536 ,H480605 ,H480662 ,H480739 ,H480805 ,H480858 ,H480917 ,H480968 ,H481025 ,H481086 ,H480537 ,H480606 ,H480663 ,H480740 ,H480806 ,H480859 ,H480918 ,H480969 ,H481026 ,H481087 ,H480538 ,H480607 ,H480664 ,H480745 ,H480807 ,H480861 ,H480919 ,H480970 ,H481027 ,H481092 ,H480541 ,H480608 ,H480665 ,H480746 ,H480808 ,H480862 ,H480921 ,H480971 ,H481028 ,H481093 ,H480542 ,H480609 ,H480667 ,H480748 ,H480809 ,H480863 ,H480923 ,H480972 ,H481029 ,H481094 ,H480543 ,H480611 ,H480676 ,H480749 ,H480810 ,H480865 ,H480924 ,H480973 ,H481030 ,H481095 ,H480544 ,H480612 ,H480677 ,H480750 ,H480811 ,H480866 ,H480925 ,H480976 ,H481031 ,H481096 ,H480545 ,H480613 ,H480678 ,H480751 ,H480812 ,H480867 ,H480926 ,H480977 ,H481032 ,H481097 ,H480546 ,H480614 ,H480679 ,H480753 ,H480813 ,H480868 ,H480927 ,H480978 ,H481033 ,H481099 ,H480549 ,H480615 ,H480680 ,H480755 ,H480814 ,H480869 ,H480928 ,H480981 ,H481034 ,H481100 ,H480566 ,H480616 ,H480685 ,H480756 ,H480815 ,H480870 ,H480929 ,H480982 ,H481035 ,H481101 ,H480567 ,H480617 ,H480686 ,H480757 ,H480816 ,H480872 ,H480930 ,H480985 ,H481036 ,H481102 ,H480568 ,H480618 ,H480688 ,H480758 ,H480817 ,H480873 ,H480931 ,H480986 ,H481037 ,H481103 ,H480569 ,H480619 ,H480689 ,H480759 ,H480819 ,H480874 ,H480932 ,H480987 ,H481038 ,H481106 ,H480570 ,H480620 ,H480691 ,H480761 ,H480821 ,H480875 ,H480933 ,H480988 ,H481039 ,H481107 ,H480571 ,H480621 ,H480692 ,H480762 ,H480822 ,H480876 ,H480934 ,H480989 ,H481040 ,H481108 ,H480572 ,H480622 ,H480693 ,H480763 ,H480823 ,H480877 ,H480935 ,H480992 ,H481043 ,H481112 ,H480573 ,H480623 ,H480694 ,H480766 ,H480824 ,H480878 ,H480937 ,H480993 ,H481044 ,H481113 ,H480574 ,H480624 ,H480698 ,H480767 ,H480825 ,H480879 ,H480938 ,H480996 ,H481054 ,H481114 ,H480575 ,H480625 ,H480699 ,H480769 ,H480827 ,H480881 ,H480939 ,H480997 ,H481056 ,H481115 ,H480576 ,H480626 ,H480700 ,H480770 ,H480828 ,H480882 ,H480940 ,H480998 ,H481058 ,H481119 ,H480577 ,H480632 ,H480701 ,H480772 ,H480829 ,H480883 ,H480941 ,H480999 ,H481059 ,H481120 ,H480578 ,H480633 ,H480702 ,H480775 ,H480830 ,H480884 ,H480942 ,H481000 ,H481060 ,H481122 ,H480579 ,H480634 ,H480703 ,H480776 ,H480831 ,H480886 ,H480943 ,H481002 ,H481061 ,H481123 ,H480580 ,H480635 ,H480704 ,H480779 ,H480832 ,H480889 ,H480944 ,H481003 ,H481062 ,H481124 ,H480581 ,H480636 ,H480705 ,H480780 ,H480833 ,H480890 ,H480945 ,H481004 ,H481063 ,H481125 ,H480582 ,H480637 ,H480714 ,H480781 ,H480834 ,H480892 ,H480947 ,H481005 ,H481066 ,H481126 ,H480583 ,H480638 ,H480715 ,H480782 ,H480836 ,H480893 ,H480948 ,H481006 ,H481067 ,H481127 ,H480585 ,H480640 ,H480716 ,H480783 ,H480837 ,H480894 ,H480949 ,H481007 ,H481069 ,H481128 ,H480587 ,H480641 ,H480717 ,H480785 ,H480839 ,H480896 ,H480950 ,H481008 ,H481070 ,H481129 ,H480588 ,H480642 ,H480718 ,H480786 ,H480840 ,H480897 ,H480952 ,H481009 ,H481071 ,H481130 ,H480589 ,H480643 ,H480719 ,H480788 ,H480841 ,H480898 ,H480953 ,H481011 ,H481072 ,H481131 ,H480590 ,H480644 ,H480720 ,H480789 ,H480842 ,H480899 ,H480954 ,H481012 ,H481073 ,H481132 ,H480591 ,H480646 ,H480722 ,H480790 ,H480843 ,H480901 ,H480955 ,H481013 ,H481074 ,H481133 ,H480592 ,H480647 ,H480726 ,H480791 ,H480844 ,H480902 ,H480956 ,H481014 ,H481075 ,H481135 ,H480593 ,H480648 ,H480727 ,H480792 ,H480845 ,H480903 ,H480957 ,H481015 ,H481076 ,H481136 ,H480594 ,H480649 ,H480728 ,H480793 ,H480846 ,H480906 ,H480958 ,H481016 ,H481077 ,H481137 ,H480595 ,H480651 ,H480729 ,H480795 ,H480849 ,H480907 ,H480959 ,H481017 ,H481078 ,H481138 ,H481139 ,H481207 ,H481262 ,H481313 ,H481379 ,H481441 ,H481503 ,H481565 ,H481627 ,H481688 ,H481140 ,H481208 ,H481263 ,H481314 ,H481380 ,H481442 ,H481504 ,H481566 ,H481631 ,H481689 ,H481141 ,H481209 ,H481264 ,H481317 ,H481381 ,H481443 ,H481505 ,H481567 ,H481632 ,H481690 ,H481142 ,H481210 ,H481265 ,H481318 ,H481382 ,H481444 ,H481506 ,H481569 ,H481633 ,H481692 ,H481143 ,H481211 ,H481266 ,H481319 ,H481383 ,H481445 ,H481507 ,H481570 ,H481634 ,H481694 ,H481144 ,H481212 ,H481267 ,H481320 ,H481384 ,H481446 ,H481508 ,H481571 ,H481635 ,H481695 ,H481145 ,H481213 ,H481268 ,H481321 ,H481389 ,H481447 ,H481509 ,H481573 ,H481636 ,H481696 ,H481146 ,H481214 ,H481269 ,H481322 ,H481391 ,H481448 ,H481513 ,H481574 ,H481638 ,H481697 ,H481147 ,H481215 ,H481270 ,H481326 ,H481392 ,H481449 ,H481514 ,H481575 ,H481641 ,H481698 ,H481148 ,H481216 ,H481271 ,H481327 ,H481393 ,H481450 ,H481515 ,H481576 ,H481642 ,H481699 ,H481149 ,H481217 ,H481272 ,H481328 ,H481394 ,H481455 ,H481519 ,H481577 ,H481644 ,H481700 ,H481150 ,H481218 ,H481273 ,H481329 ,H481395 ,H481456 ,H481520 ,H481578 ,H481645 ,H481702 ,H481151 ,H481219 ,H481274 ,H481330 ,H481397 ,H481457 ,H481521 ,H481585 ,H481646 ,H481703 ,H481152 ,H481220 ,H481275 ,H481331 ,H481398 ,H481458 ,H481523 ,H481586 ,H481647 ,H481704 ,H481153 ,H481221 ,H481276 ,H481332 ,H481399 ,H481459 ,H481525 ,H481588 ,H481648 ,H481705 ,H481154 ,H481222 ,H481277 ,H481333 ,H481401 ,H481460 ,H481526 ,H481589 ,H481649 ,H481706 ,H481155 ,H481224 ,H481278 ,H481340 ,H481402 ,H481461 ,H481527 ,H481590 ,H481650 ,H481707 ,H481156 ,H481225 ,H481279 ,H481343 ,H481403 ,H481462 ,H481528 ,H481591 ,H481651 ,H481708 ,H481157 ,H481226 ,H481280 ,H481344 ,H481404 ,H481463 ,H481529 ,H481592 ,H481652 ,H481710 ,H481158 ,H481227 ,H481281 ,H481345 ,H481405 ,H481464 ,H481530 ,H481593 ,H481653 ,H481711 ,H481159 ,H481228 ,H481282 ,H481346 ,H481408 ,H481465 ,H481531 ,H481594 ,H481654 ,H481712 ,H481160 ,H481229 ,H481283 ,H481347 ,H481409 ,H481466 ,H481534 ,H481595 ,H481655 ,H481713 ,H481161 ,H481230 ,H481284 ,H481348 ,H481410 ,H481471 ,H481535 ,H481596 ,H481656 ,H481715 ,H481163 ,H481231 ,H481285 ,H481349 ,H481411 ,H481472 ,H481536 ,H481597 ,H481657 ,H481716 ,H481169 ,H481232 ,H481286 ,H481351 ,H481412 ,H481473 ,H481539 ,H481601 ,H481658 ,H481717 ,H481176 ,H481233 ,H481287 ,H481352 ,H481413 ,H481474 ,H481540 ,H481602 ,H481659 ,H481719 ,H481177 ,H481234 ,H481288 ,H481353 ,H481419 ,H481482 ,H481542 ,H481603 ,H481660 ,H481720 ,H481180 ,H481235 ,H481289 ,H481354 ,H481421 ,H481483 ,H481543 ,H481604 ,H481661 ,H481722 ,H481183 ,H481236 ,H481290 ,H481355 ,H481422 ,H481484 ,H481544 ,H481605 ,H481662 ,H481723 ,H481184 ,H481237 ,H481291 ,H481356 ,H481423 ,H481485 ,H481545 ,H481606 ,H481663 ,H481724 ,H481188 ,H481240 ,H481292 ,H481357 ,H481426 ,H481486 ,H481547 ,H481607 ,H481664 ,H481725 ,H481193 ,H481248 ,H481293 ,H481358 ,H481427 ,H481487 ,H481548 ,H481608 ,H481665 ,H481730 ,H481194 ,H481249 ,H481294 ,H481360 ,H481428 ,H481488 ,H481549 ,H481609 ,H481666 ,H481731 ,H481195 ,H481250 ,H481295 ,H481361 ,H481429 ,H481489 ,H481553 ,H481610 ,H481667 ,H481732 ,H481196 ,H481251 ,H481296 ,H481362 ,H481430 ,H481490 ,H481554 ,H481611 ,H481670 ,H481733 ,H481197 ,H481252 ,H481297 ,H481363 ,H481431 ,H481491 ,H481555 ,H481612 ,H481677 ,H481734 ,H481198 ,H481253 ,H481298 ,H481364 ,H481432 ,H481492 ,H481556 ,H481613 ,H481678 ,H481735 ,H481199 ,H481254 ,H481299 ,H481366 ,H481433 ,H481493 ,H481557 ,H481614 ,H481679 ,H481736 ,H481200 ,H481255 ,H481304 ,H481367 ,H481434 ,H481494 ,H481558 ,H481618 ,H481680 ,H481744 ,H481201 ,H481256 ,H481305 ,H481373 ,H481435 ,H481495 ,H481559 ,H481619 ,H481681 ,H481745 ,H481202 ,H481257 ,H481308 ,H481374 ,H481436 ,H481496 ,H481560 ,H481620 ,H481682 ,H481748 ,H481203 ,H481258 ,H481309 ,H481375 ,H481437 ,H481497 ,H481561 ,H481621 ,H481683 ,H481749 ,H481204 ,H481259 ,H481310 ,H481376 ,H481438 ,H481498 ,H481562 ,H481622 ,H481684 ,H481751 ,H481205 ,H481260 ,H481311 ,H481377 ,H481439 ,H481499 ,H481563 ,H481625 ,H481685 ,H481752 ,H481206 ,H481261 ,H481312 ,H481378 ,H481440 ,H481502 ,H481564 ,H481626 ,H481686 ,H481753 ,H481754 ,H481822 ,H481881 ,H481952 ,H482008 ,H482081 ,H482134 ,H482210 ,H482262 ,H482316 ,H481755 ,H481823 ,H481882 ,H481954 ,H482009 ,H482082 ,H482135 ,H482211 ,H482263 ,H482317 ,H481756 ,H481824 ,H481884 ,H481955 ,H482011 ,H482083 ,H482136 ,H482212 ,H482264 ,H482318 ,H481758 ,H481825 ,H481885 ,H481956 ,H482012 ,H482084 ,H482137 ,H482213 ,H482265 ,H482319 ,H481759 ,H481826 ,H481886 ,H481957 ,H482013 ,H482085 ,H482144 ,H482214 ,H482266 ,H482320 ,H481760 ,H481828 ,H481892 ,H481958 ,H482014 ,H482086 ,H482145 ,H482215 ,H482267 ,H482321 ,H481761 ,H481829 ,H481893 ,H481959 ,H482024 ,H482087 ,H482146 ,H482216 ,H482268 ,H482322 ,H481762 ,H481830 ,H481894 ,H481960 ,H482026 ,H482088 ,H482147 ,H482217 ,H482269 ,H482323 ,H481763 ,H481831 ,H481896 ,H481961 ,H482028 ,H482089 ,H482150 ,H482218 ,H482270 ,H482324 ,H481764 ,H481832 ,H481897 ,H481962 ,H482030 ,H482090 ,H482151 ,H482219 ,H482271 ,H482325 ,H481765 ,H481833 ,H481898 ,H481963 ,H482032 ,H482091 ,H482152 ,H482220 ,H482272 ,H482326 ,H481766 ,H481834 ,H481899 ,H481964 ,H482034 ,H482092 ,H482153 ,H482221 ,H482273 ,H482327 ,H481767 ,H481835 ,H481900 ,H481966 ,H482036 ,H482093 ,H482156 ,H482222 ,H482274 ,H482328 ,H481768 ,H481836 ,H481901 ,H481967 ,H482037 ,H482094 ,H482157 ,H482223 ,H482275 ,H482329 ,H481769 ,H481837 ,H481902 ,H481969 ,H482040 ,H482095 ,H482158 ,H482224 ,H482276 ,H482330 ,H481770 ,H481838 ,H481903 ,H481971 ,H482041 ,H482097 ,H482159 ,H482225 ,H482277 ,H482331 ,H481771 ,H481839 ,H481904 ,H481972 ,H482042 ,H482098 ,H482160 ,H482226 ,H482278 ,H482332 ,H481781 ,H481840 ,H481905 ,H481974 ,H482043 ,H482099 ,H482161 ,H482228 ,H482279 ,H482333 ,H481783 ,H481841 ,H481906 ,H481977 ,H482054 ,H482101 ,H482162 ,H482229 ,H482281 ,H482334 ,H481784 ,H481842 ,H481907 ,H481980 ,H482055 ,H482102 ,H482163 ,H482230 ,H482283 ,H482335 ,H481791 ,H481843 ,H481908 ,H481982 ,H482056 ,H482104 ,H482164 ,H482231 ,H482284 ,H482336 ,H481792 ,H481846 ,H481909 ,H481983 ,H482057 ,H482105 ,H482165 ,H482232 ,H482285 ,H482337 ,H481794 ,H481847 ,H481910 ,H481984 ,H482058 ,H482106 ,H482173 ,H482233 ,H482286 ,H482338 ,H481795 ,H481848 ,H481911 ,H481985 ,H482059 ,H482107 ,H482174 ,H482234 ,H482287 ,H482339 ,H481796 ,H481849 ,H481916 ,H481986 ,H482060 ,H482108 ,H482178 ,H482235 ,H482288 ,H482340 ,H481797 ,H481857 ,H481918 ,H481987 ,H482061 ,H482109 ,H482179 ,H482236 ,H482292 ,H482341 ,H481798 ,H481858 ,H481919 ,H481988 ,H482062 ,H482110 ,H482184 ,H482237 ,H482293 ,H482342 ,H481799 ,H481859 ,H481920 ,H481989 ,H482063 ,H482111 ,H482185 ,H482238 ,H482294 ,H482343 ,H481800 ,H481861 ,H481921 ,H481990 ,H482064 ,H482112 ,H482189 ,H482243 ,H482295 ,H482344 ,H481801 ,H481864 ,H481922 ,H481991 ,H482065 ,H482113 ,H482190 ,H482244 ,H482297 ,H482345 ,H481802 ,H481865 ,H481923 ,H481992 ,H482066 ,H482114 ,H482191 ,H482245 ,H482300 ,H482346 ,H481803 ,H481866 ,H481924 ,H481993 ,H482067 ,H482115 ,H482192 ,H482246 ,H482301 ,H482347 ,H481804 ,H481867 ,H481925 ,H481994 ,H482068 ,H482120 ,H482193 ,H482247 ,H482302 ,H482348 ,H481805 ,H481868 ,H481926 ,H481995 ,H482069 ,H482121 ,H482194 ,H482248 ,H482303 ,H482349 ,H481806 ,H481869 ,H481932 ,H481996 ,H482070 ,H482122 ,H482196 ,H482249 ,H482304 ,H482350 ,H481808 ,H481870 ,H481933 ,H481997 ,H482071 ,H482124 ,H482198 ,H482250 ,H482305 ,H482351 ,H481810 ,H481871 ,H481934 ,H481998 ,H482072 ,H482125 ,H482199 ,H482251 ,H482306 ,H482352 ,H481812 ,H481872 ,H481935 ,H481999 ,H482073 ,H482126 ,H482200 ,H482252 ,H482307 ,H482353 ,H481813 ,H481873 ,H481936 ,H482000 ,H482074 ,H482127 ,H482201 ,H482253 ,H482308 ,H482354 ,H481814 ,H481874 ,H481937 ,H482001 ,H482075 ,H482128 ,H482202 ,H482254 ,H482310 ,H482355 ,H481817 ,H481875 ,H481938 ,H482002 ,H482076 ,H482129 ,H482203 ,H482256 ,H482311 ,H482356 ,H481818 ,H481876 ,H481945 ,H482003 ,H482077 ,H482130 ,H482204 ,H482258 ,H482312 ,H482367 ,H481819 ,H481877 ,H481946 ,H482004 ,H482078 ,H482131 ,H482205 ,H482259 ,H482313 ,H482369 ,H481820 ,H481878 ,H481950 ,H482005 ,H482079 ,H482132 ,H482206 ,H482260 ,H482314 ,H482374 ,H481821 ,H481879 ,H481951 ,H482007 ,H482080 ,H482133 ,H482209 ,H482261 ,H482315 ,H482380 ,H482389 ,H482452 ,H482514 ,H482572 ,H482635 ,H482701 ,H482770 ,H482842 ,H482909 ,H482990 ,H482390 ,H482456 ,H482515 ,H482573 ,H482636 ,H482702 ,H482771 ,H482843 ,H482910 ,H482991 ,H482391 ,H482457 ,H482516 ,H482577 ,H482637 ,H482703 ,H482772 ,H482845 ,H482911 ,H482992 ,H482392 ,H482460 ,H482517 ,H482578 ,H482638 ,H482704 ,H482773 ,H482847 ,H482912 ,H482998 ,H482395 ,H482461 ,H482518 ,H482579 ,H482641 ,H482705 ,H482774 ,H482849 ,H482913 ,H482999 ,H482396 ,H482462 ,H482519 ,H482580 ,H482645 ,H482706 ,H482775 ,H482850 ,H482914 ,H483000 ,H482397 ,H482463 ,H482520 ,H482581 ,H482647 ,H482709 ,H482791 ,H482852 ,H482915 ,H483001 ,H482398 ,H482464 ,H482521 ,H482582 ,H482648 ,H482710 ,H482792 ,H482854 ,H482916 ,H483002 ,H482399 ,H482465 ,H482522 ,H482583 ,H482649 ,H482711 ,H482794 ,H482855 ,H482917 ,H483003 ,H482401 ,H482466 ,H482523 ,H482584 ,H482650 ,H482712 ,H482795 ,H482856 ,H482918 ,H483004 ,H482402 ,H482467 ,H482524 ,H482586 ,H482651 ,H482713 ,H482796 ,H482857 ,H482919 ,H483005 ,H482403 ,H482470 ,H482527 ,H482587 ,H482652 ,H482714 ,H482797 ,H482859 ,H482920 ,H483006 ,H482406 ,H482473 ,H482529 ,H482588 ,H482653 ,H482715 ,H482799 ,H482860 ,H482921 ,H483007 ,H482407 ,H482474 ,H482530 ,H482589 ,H482654 ,H482716 ,H482800 ,H482861 ,H482922 ,H483008 ,H482408 ,H482475 ,H482531 ,H482590 ,H482656 ,H482717 ,H482801 ,H482862 ,H482923 ,H483009 ,H482409 ,H482476 ,H482534 ,H482591 ,H482657 ,H482718 ,H482802 ,H482863 ,H482924 ,H483010 ,H482410 ,H482478 ,H482539 ,H482592 ,H482662 ,H482719 ,H482803 ,H482865 ,H482925 ,H483011 ,H482411 ,H482479 ,H482540 ,H482593 ,H482663 ,H482720 ,H482804 ,H482866 ,H482926 ,H483012 ,H482412 ,H482480 ,H482541 ,H482594 ,H482664 ,H482722 ,H482805 ,H482867 ,H482927 ,H483013 ,H482413 ,H482481 ,H482542 ,H482595 ,H482665 ,H482723 ,H482806 ,H482868 ,H482928 ,H483014 ,H482414 ,H482484 ,H482543 ,H482596 ,H482666 ,H482725 ,H482812 ,H482869 ,H482931 ,H483015 ,H482415 ,H482485 ,H482544 ,H482597 ,H482667 ,H482726 ,H482813 ,H482870 ,H482934 ,H483016 ,H482416 ,H482486 ,H482545 ,H482598 ,H482668 ,H482727 ,H482814 ,H482871 ,H482935 ,H483017 ,H482417 ,H482487 ,H482546 ,H482599 ,H482669 ,H482728 ,H482815 ,H482874 ,H482936 ,H483018 ,H482418 ,H482488 ,H482547 ,H482601 ,H482671 ,H482730 ,H482816 ,H482875 ,H482937 ,H483019 ,H482419 ,H482489 ,H482548 ,H482603 ,H482676 ,H482731 ,H482817 ,H482880 ,H482938 ,H483020 ,H482420 ,H482490 ,H482549 ,H482604 ,H482678 ,H482737 ,H482818 ,H482881 ,H482939 ,H483021 ,H482422 ,H482491 ,H482550 ,H482605 ,H482679 ,H482738 ,H482819 ,H482882 ,H482940 ,H483022 ,H482423 ,H482492 ,H482551 ,H482608 ,H482680 ,H482748 ,H482820 ,H482883 ,H482941 ,H483023 ,H482427 ,H482494 ,H482552 ,H482611 ,H482681 ,H482749 ,H482824 ,H482890 ,H482942 ,H483024 ,H482436 ,H482497 ,H482555 ,H482612 ,H482682 ,H482750 ,H482825 ,H482891 ,H482943 ,H483025 ,H482437 ,H482499 ,H482556 ,H482613 ,H482683 ,H482751 ,H482827 ,H482893 ,H482944 ,H483026 ,H482438 ,H482500 ,H482557 ,H482619 ,H482684 ,H482752 ,H482829 ,H482894 ,H482945 ,H483027 ,H482440 ,H482501 ,H482559 ,H482620 ,H482685 ,H482753 ,H482830 ,H482897 ,H482963 ,H483028 ,H482441 ,H482502 ,H482560 ,H482621 ,H482686 ,H482754 ,H482831 ,H482898 ,H482964 ,H483029 ,H482442 ,H482503 ,H482561 ,H482624 ,H482687 ,H482755 ,H482832 ,H482899 ,H482965 ,H483040 ,H482443 ,H482504 ,H482562 ,H482625 ,H482688 ,H482756 ,H482833 ,H482900 ,H482966 ,H483041 ,H482444 ,H482505 ,H482563 ,H482626 ,H482689 ,H482760 ,H482834 ,H482901 ,H482982 ,H483042 ,H482445 ,H482506 ,H482564 ,H482627 ,H482691 ,H482761 ,H482835 ,H482902 ,H482983 ,H483043 ,H482446 ,H482508 ,H482566 ,H482628 ,H482692 ,H482762 ,H482836 ,H482903 ,H482984 ,H483044 ,H482447 ,H482509 ,H482567 ,H482629 ,H482693 ,H482763 ,H482837 ,H482904 ,H482985 ,H483047 ,H482448 ,H482510 ,H482568 ,H482630 ,H482697 ,H482764 ,H482838 ,H482905 ,H482986 ,H483048 ,H482449 ,H482511 ,H482569 ,H482631 ,H482698 ,H482766 ,H482839 ,H482906 ,H482987 ,H483051 ,H482450 ,H482512 ,H482570 ,H482633 ,H482699 ,H482767 ,H482840 ,H482907 ,H482988 ,H483052 ,H482451 ,H482513 ,H482571 ,H482634 ,H482700 ,H482768 ,H482841 ,H482908 ,H482989 ,H483053 ,H483054 ,H483126 ,H483197 ,H483278 ,H483342 ,H483394 ,H483472 ,H483537 ,H483592 ,H483651 ,H483055 ,H483127 ,H483198 ,H483279 ,H483343 ,H483396 ,H483473 ,H483538 ,H483593 ,H483654 ,H483057 ,H483128 ,H483199 ,H483280 ,H483344 ,H483397 ,H483474 ,H483539 ,H483594 ,H483655 ,H483058 ,H483129 ,H483200 ,H483281 ,H483345 ,H483403 ,H483477 ,H483540 ,H483595 ,H483656 ,H483059 ,H483130 ,H483203 ,H483282 ,H483350 ,H483404 ,H483478 ,H483541 ,H483599 ,H483657 ,H483060 ,H483131 ,H483206 ,H483283 ,H483351 ,H483405 ,H483479 ,H483542 ,H483600 ,H483658 ,H483061 ,H483132 ,H483207 ,H483284 ,H483352 ,H483407 ,H483485 ,H483543 ,H483602 ,H483659 ,H483062 ,H483133 ,H483208 ,H483285 ,H483353 ,H483410 ,H483486 ,H483544 ,H483606 ,H483660 ,H483063 ,H483134 ,H483209 ,H483286 ,H483354 ,H483411 ,H483487 ,H483545 ,H483607 ,H483661 ,H483064 ,H483138 ,H483210 ,H483287 ,H483355 ,H483412 ,H483488 ,H483546 ,H483608 ,H483662 ,H483070 ,H483139 ,H483211 ,H483288 ,H483356 ,H483416 ,H483489 ,H483547 ,H483609 ,H483663 ,H483071 ,H483140 ,H483215 ,H483289 ,H483357 ,H483417 ,H483490 ,H483548 ,H483610 ,H483664 ,H483072 ,H483141 ,H483223 ,H483290 ,H483359 ,H483418 ,H483491 ,H483549 ,H483611 ,H483665 ,H483073 ,H483142 ,H483226 ,H483291 ,H483360 ,H483419 ,H483492 ,H483550 ,H483612 ,H483670 ,H483075 ,H483143 ,H483227 ,H483292 ,H483361 ,H483420 ,H483494 ,H483551 ,H483615 ,H483671 ,H483076 ,H483144 ,H483228 ,H483309 ,H483362 ,H483421 ,H483495 ,H483552 ,H483616 ,H483672 ,H483077 ,H483145 ,H483229 ,H483311 ,H483363 ,H483422 ,H483496 ,H483554 ,H483617 ,H483674 ,H483078 ,H483148 ,H483233 ,H483312 ,H483364 ,H483424 ,H483500 ,H483555 ,H483619 ,H483675 ,H483079 ,H483151 ,H483234 ,H483313 ,H483365 ,H483425 ,H483501 ,H483556 ,H483620 ,H483676 ,H483080 ,H483152 ,H483235 ,H483315 ,H483366 ,H483426 ,H483502 ,H483557 ,H483621 ,H483677 ,H483081 ,H483153 ,H483236 ,H483317 ,H483367 ,H483427 ,H483503 ,H483558 ,H483622 ,H483679 ,H483084 ,H483157 ,H483237 ,H483318 ,H483369 ,H483429 ,H483504 ,H483559 ,H483623 ,H483680 ,H483088 ,H483158 ,H483238 ,H483319 ,H483370 ,H483430 ,H483505 ,H483560 ,H483624 ,H483681 ,H483089 ,H483159 ,H483240 ,H483320 ,H483371 ,H483431 ,H483506 ,H483564 ,H483625 ,H483682 ,H483091 ,H483161 ,H483241 ,H483321 ,H483372 ,H483432 ,H483507 ,H483565 ,H483626 ,H483683 ,H483092 ,H483162 ,H483243 ,H483322 ,H483373 ,H483433 ,H483509 ,H483570 ,H483627 ,H483684 ,H483097 ,H483163 ,H483244 ,H483323 ,H483374 ,H483434 ,H483510 ,H483571 ,H483628 ,H483686 ,H483099 ,H483164 ,H483245 ,H483324 ,H483375 ,H483435 ,H483513 ,H483572 ,H483629 ,H483687 ,H483100 ,H483165 ,H483246 ,H483325 ,H483376 ,H483436 ,H483514 ,H483573 ,H483632 ,H483688 ,H483102 ,H483167 ,H483247 ,H483326 ,H483378 ,H483437 ,H483515 ,H483574 ,H483633 ,H483689 ,H483103 ,H483168 ,H483253 ,H483327 ,H483379 ,H483438 ,H483516 ,H483575 ,H483634 ,H483691 ,H483104 ,H483176 ,H483254 ,H483328 ,H483380 ,H483440 ,H483517 ,H483576 ,H483635 ,H483694 ,H483106 ,H483178 ,H483255 ,H483329 ,H483381 ,H483441 ,H483518 ,H483577 ,H483636 ,H483695 ,H483107 ,H483179 ,H483256 ,H483330 ,H483382 ,H483442 ,H483519 ,H483578 ,H483637 ,H483696 ,H483109 ,H483180 ,H483257 ,H483331 ,H483383 ,H483449 ,H483520 ,H483579 ,H483639 ,H483697 ,H483110 ,H483181 ,H483265 ,H483332 ,H483384 ,H483453 ,H483521 ,H483580 ,H483640 ,H483698 ,H483115 ,H483182 ,H483266 ,H483333 ,H483385 ,H483454 ,H483525 ,H483581 ,H483641 ,H483701 ,H483117 ,H483183 ,H483267 ,H483334 ,H483386 ,H483459 ,H483526 ,H483582 ,H483642 ,H483702 ,H483118 ,H483184 ,H483268 ,H483335 ,H483387 ,H483460 ,H483527 ,H483583 ,H483643 ,H483705 ,H483119 ,H483185 ,H483269 ,H483336 ,H483388 ,H483461 ,H483531 ,H483584 ,H483644 ,H483706 ,H483120 ,H483186 ,H483270 ,H483337 ,H483389 ,H483462 ,H483532 ,H483586 ,H483645 ,H483712 ,H483121 ,H483191 ,H483272 ,H483338 ,H483390 ,H483463 ,H483533 ,H483587 ,H483646 ,H483713 ,H483123 ,H483192 ,H483275 ,H483339 ,H483391 ,H483465 ,H483534 ,H483588 ,H483647 ,H483715 ,H483124 ,H483194 ,H483276 ,H483340 ,H483392 ,H483466 ,H483535 ,H483589 ,H483648 ,H483716 ,H483125 ,H483196 ,H483277 ,H483341 ,H483393 ,H483471 ,H483536 ,H483591 ,H483650 ,H483717 ,H483718 ,H483766 ,H483853 ,H483947 ,H484020 ,H484084 ,H484148 ,H484206 ,H484276 ,H484332 ,H483719 ,H483767 ,H483854 ,H483948 ,H484023 ,H484085 ,H484149 ,H484207 ,H484280 ,H484334 ,H483720 ,H483768 ,H483855 ,H483949 ,H484024 ,H484087 ,H484153 ,H484208 ,H484281 ,H484335 ,H483721 ,H483769 ,H483856 ,H483950 ,H484025 ,H484088 ,H484154 ,H484209 ,H484282 ,H484336 ,H483722 ,H483770 ,H483858 ,H483951 ,H484026 ,H484089 ,H484155 ,H484210 ,H484283 ,H484337 ,H483723 ,H483796 ,H483859 ,H483952 ,H484027 ,H484090 ,H484156 ,H484211 ,H484285 ,H484339 ,H483724 ,H483797 ,H483860 ,H483953 ,H484031 ,H484091 ,H484157 ,H484212 ,H484286 ,H484340 ,H483725 ,H483798 ,H483861 ,H483957 ,H484032 ,H484092 ,H484158 ,H484213 ,H484287 ,H484341 ,H483726 ,H483799 ,H483862 ,H483960 ,H484033 ,H484093 ,H484159 ,H484214 ,H484288 ,H484342 ,H483727 ,H483803 ,H483866 ,H483964 ,H484034 ,H484094 ,H484160 ,H484215 ,H484289 ,H484343 ,H483728 ,H483804 ,H483867 ,H483966 ,H484035 ,H484097 ,H484161 ,H484216 ,H484290 ,H484345 ,H483730 ,H483807 ,H483868 ,H483970 ,H484036 ,H484098 ,H484162 ,H484217 ,H484291 ,H484348 ,H483731 ,H483808 ,H483869 ,H483971 ,H484039 ,H484099 ,H484163 ,H484218 ,H484292 ,H484349 ,H483732 ,H483809 ,H483871 ,H483973 ,H484040 ,H484100 ,H484164 ,H484219 ,H484293 ,H484350 ,H483733 ,H483810 ,H483872 ,H483975 ,H484041 ,H484103 ,H484165 ,H484220 ,H484295 ,H484351 ,H483734 ,H483811 ,H483873 ,H483976 ,H484042 ,H484104 ,H484166 ,H484221 ,H484296 ,H484353 ,H483735 ,H483812 ,H483874 ,H483977 ,H484043 ,H484105 ,H484167 ,H484222 ,H484297 ,H484354 ,H483737 ,H483813 ,H483875 ,H483978 ,H484044 ,H484107 ,H484170 ,H484233 ,H484298 ,H484355 ,H483738 ,H483814 ,H483876 ,H483979 ,H484045 ,H484108 ,H484171 ,H484234 ,H484299 ,H484356 ,H483739 ,H483816 ,H483877 ,H483980 ,H484046 ,H484109 ,H484172 ,H484235 ,H484300 ,H484357 ,H483740 ,H483817 ,H483879 ,H483981 ,H484047 ,H484111 ,H484173 ,H484236 ,H484301 ,H484358 ,H483741 ,H483818 ,H483880 ,H483982 ,H484048 ,H484112 ,H484174 ,H484237 ,H484306 ,H484359 ,H483742 ,H483820 ,H483881 ,H483983 ,H484049 ,H484114 ,H484175 ,H484250 ,H484307 ,H484360 ,H483743 ,H483821 ,H483914 ,H483984 ,H484050 ,H484115 ,H484177 ,H484251 ,H484308 ,H484361 ,H483744 ,H483823 ,H483916 ,H483986 ,H484052 ,H484116 ,H484178 ,H484252 ,H484310 ,H484362 ,H483745 ,H483827 ,H483917 ,H483987 ,H484053 ,H484117 ,H484179 ,H484253 ,H484311 ,H484363 ,H483746 ,H483828 ,H483921 ,H483988 ,H484054 ,H484118 ,H484180 ,H484254 ,H484312 ,H484368 ,H483747 ,H483830 ,H483922 ,H483991 ,H484055 ,H484119 ,H484181 ,H484256 ,H484313 ,H484369 ,H483748 ,H483831 ,H483923 ,H483992 ,H484056 ,H484120 ,H484182 ,H484257 ,H484314 ,H484370 ,H483749 ,H483836 ,H483924 ,H484002 ,H484057 ,H484123 ,H484185 ,H484258 ,H484315 ,H484371 ,H483750 ,H483837 ,H483925 ,H484003 ,H484058 ,H484124 ,H484186 ,H484260 ,H484316 ,H484374 ,H483751 ,H483838 ,H483932 ,H484004 ,H484059 ,H484125 ,H484188 ,H484261 ,H484317 ,H484375 ,H483752 ,H483839 ,H483934 ,H484005 ,H484060 ,H484128 ,H484189 ,H484262 ,H484318 ,H484376 ,H483753 ,H483840 ,H483935 ,H484008 ,H484061 ,H484129 ,H484190 ,H484263 ,H484319 ,H484377 ,H483754 ,H483841 ,H483936 ,H484009 ,H484062 ,H484130 ,H484191 ,H484264 ,H484321 ,H484378 ,H483755 ,H483842 ,H483937 ,H484010 ,H484063 ,H484131 ,H484192 ,H484265 ,H484322 ,H484379 ,H483756 ,H483843 ,H483938 ,H484011 ,H484072 ,H484132 ,H484193 ,H484266 ,H484323 ,H484380 ,H483757 ,H483844 ,H483939 ,H484012 ,H484073 ,H484133 ,H484194 ,H484267 ,H484324 ,H484381 ,H483758 ,H483845 ,H483940 ,H484013 ,H484076 ,H484135 ,H484195 ,H484268 ,H484325 ,H484382 ,H483759 ,H483847 ,H483941 ,H484014 ,H484077 ,H484139 ,H484200 ,H484269 ,H484326 ,H484383 ,H483760 ,H483848 ,H483942 ,H484015 ,H484079 ,H484140 ,H484201 ,H484270 ,H484327 ,H484384 ,H483761 ,H483849 ,H483943 ,H484016 ,H484080 ,H484143 ,H484202 ,H484271 ,H484328 ,H484385 ,H483762 ,H483850 ,H483944 ,H484017 ,H484081 ,H484144 ,H484203 ,H484272 ,H484329 ,H484386 ,H483763 ,H483851 ,H483945 ,H484018 ,H484082 ,H484145 ,H484204 ,H484274 ,H484330 ,H484387 ,H483764 ,H483852 ,H483946 ,H484019 ,H484083 ,H484147 ,H484205 ,H484275 ,H484331 ,H484388 ,H484389 ,H484449 ,H484522 ,H484593 ,H484665 ,H484726 ,H484786 ,H484865 ,H484935 ,H484998 ,H484390 ,H484450 ,H484523 ,H484594 ,H484670 ,H484727 ,H484787 ,H484866 ,H484936 ,H484999 ,H484391 ,H484451 ,H484524 ,H484597 ,H484671 ,H484728 ,H484788 ,H484868 ,H484937 ,H485001 ,H484392 ,H484452 ,H484525 ,H484598 ,H484672 ,H484729 ,H484789 ,H484869 ,H484938 ,H485002 ,H484393 ,H484454 ,H484526 ,H484599 ,H484673 ,H484730 ,H484790 ,H484870 ,H484939 ,H485003 ,H484394 ,H484455 ,H484527 ,H484600 ,H484674 ,H484731 ,H484791 ,H484871 ,H484940 ,H485004 ,H484395 ,H484456 ,H484528 ,H484601 ,H484675 ,H484732 ,H484792 ,H484872 ,H484943 ,H485005 ,H484397 ,H484457 ,H484529 ,H484602 ,H484680 ,H484733 ,H484793 ,H484873 ,H484945 ,H485006 ,H484398 ,H484458 ,H484530 ,H484603 ,H484683 ,H484734 ,H484794 ,H484874 ,H484946 ,H485007 ,H484399 ,H484459 ,H484531 ,H484604 ,H484684 ,H484735 ,H484795 ,H484875 ,H484947 ,H485012 ,H484400 ,H484460 ,H484532 ,H484605 ,H484685 ,H484736 ,H484796 ,H484876 ,H484948 ,H485013 ,H484401 ,H484472 ,H484533 ,H484610 ,H484686 ,H484737 ,H484797 ,H484877 ,H484950 ,H485014 ,H484402 ,H484473 ,H484534 ,H484611 ,H484687 ,H484738 ,H484812 ,H484878 ,H484951 ,H485015 ,H484407 ,H484474 ,H484535 ,H484612 ,H484688 ,H484739 ,H484813 ,H484879 ,H484952 ,H485016 ,H484408 ,H484475 ,H484536 ,H484613 ,H484689 ,H484740 ,H484814 ,H484880 ,H484953 ,H485017 ,H484409 ,H484476 ,H484537 ,H484614 ,H484691 ,H484741 ,H484815 ,H484881 ,H484954 ,H485018 ,H484410 ,H484477 ,H484538 ,H484615 ,H484692 ,H484742 ,H484816 ,H484882 ,H484955 ,H485019 ,H484411 ,H484478 ,H484539 ,H484616 ,H484693 ,H484744 ,H484817 ,H484883 ,H484956 ,H485022 ,H484412 ,H484479 ,H484540 ,H484617 ,H484694 ,H484752 ,H484818 ,H484884 ,H484957 ,H485023 ,H484413 ,H484481 ,H484541 ,H484619 ,H484695 ,H484753 ,H484819 ,H484886 ,H484958 ,H485024 ,H484414 ,H484482 ,H484544 ,H484620 ,H484696 ,H484754 ,H484820 ,H484887 ,H484959 ,H485025 ,H484415 ,H484483 ,H484545 ,H484621 ,H484697 ,H484755 ,H484822 ,H484888 ,H484966 ,H485026 ,H484416 ,H484485 ,H484549 ,H484622 ,H484698 ,H484756 ,H484823 ,H484889 ,H484967 ,H485027 ,H484419 ,H484486 ,H484551 ,H484624 ,H484699 ,H484757 ,H484824 ,H484890 ,H484968 ,H485033 ,H484424 ,H484487 ,H484552 ,H484625 ,H484701 ,H484761 ,H484825 ,H484891 ,H484970 ,H485034 ,H484425 ,H484495 ,H484553 ,H484626 ,H484704 ,H484762 ,H484826 ,H484892 ,H484971 ,H485035 ,H484426 ,H484496 ,H484554 ,H484627 ,H484705 ,H484763 ,H484827 ,H484893 ,H484972 ,H485036 ,H484427 ,H484501 ,H484555 ,H484628 ,H484706 ,H484764 ,H484836 ,H484894 ,H484973 ,H485038 ,H484431 ,H484502 ,H484565 ,H484629 ,H484707 ,H484765 ,H484837 ,H484903 ,H484974 ,H485039 ,H484433 ,H484503 ,H484566 ,H484630 ,H484708 ,H484766 ,H484838 ,H484904 ,H484975 ,H485040 ,H484434 ,H484504 ,H484567 ,H484631 ,H484709 ,H484767 ,H484839 ,H484905 ,H484976 ,H485041 ,H484435 ,H484506 ,H484568 ,H484632 ,H484710 ,H484768 ,H484840 ,H484906 ,H484977 ,H485042 ,H484436 ,H484507 ,H484569 ,H484634 ,H484711 ,H484769 ,H484841 ,H484908 ,H484979 ,H485043 ,H484437 ,H484510 ,H484570 ,H484639 ,H484712 ,H484770 ,H484842 ,H484909 ,H484980 ,H485044 ,H484438 ,H484511 ,H484571 ,H484640 ,H484715 ,H484771 ,H484844 ,H484910 ,H484981 ,H485051 ,H484439 ,H484512 ,H484573 ,H484641 ,H484716 ,H484772 ,H484845 ,H484911 ,H484987 ,H485052 ,H484440 ,H484513 ,H484574 ,H484650 ,H484717 ,H484773 ,H484846 ,H484912 ,H484988 ,H485053 ,H484441 ,H484514 ,H484575 ,H484651 ,H484718 ,H484774 ,H484847 ,H484913 ,H484989 ,H485054 ,H484442 ,H484515 ,H484585 ,H484652 ,H484719 ,H484775 ,H484849 ,H484915 ,H484990 ,H485055 ,H484443 ,H484516 ,H484586 ,H484653 ,H484720 ,H484777 ,H484850 ,H484923 ,H484992 ,H485056 ,H484444 ,H484517 ,H484588 ,H484654 ,H484721 ,H484778 ,H484856 ,H484925 ,H484993 ,H485057 ,H484445 ,H484518 ,H484589 ,H484657 ,H484722 ,H484779 ,H484861 ,H484926 ,H484994 ,H485058 ,H484446 ,H484519 ,H484590 ,H484658 ,H484723 ,H484780 ,H484862 ,H484932 ,H484995 ,H485059 ,H484447 ,H484520 ,H484591 ,H484659 ,H484724 ,H484781 ,H484863 ,H484933 ,H484996 ,H485060 ,H484448 ,H484521 ,H484592 ,H484663 ,H484725 ,H484782 ,H484864 ,H484934 ,H484997 ,H485061 ,H485062 ,H485134 ,H485190 ,H485239 ,H485305 ,H485358 ,H485418 ,H485481 ,H485561 ,H485633 ,H485063 ,H485135 ,H485191 ,H485240 ,H485306 ,H485362 ,H485419 ,H485482 ,H485564 ,H485634 ,H485064 ,H485137 ,H485193 ,H485241 ,H485308 ,H485363 ,H485420 ,H485483 ,H485565 ,H485635 ,H485065 ,H485138 ,H485194 ,H485243 ,H485311 ,H485364 ,H485428 ,H485484 ,H485569 ,H485636 ,H485066 ,H485139 ,H485195 ,H485244 ,H485312 ,H485365 ,H485429 ,H485485 ,H485570 ,H485637 ,H485067 ,H485140 ,H485196 ,H485247 ,H485313 ,H485366 ,H485430 ,H485486 ,H485571 ,H485638 ,H485068 ,H485141 ,H485197 ,H485248 ,H485314 ,H485367 ,H485431 ,H485487 ,H485572 ,H485639 ,H485069 ,H485142 ,H485198 ,H485249 ,H485315 ,H485368 ,H485432 ,H485488 ,H485573 ,H485640 ,H485070 ,H485143 ,H485199 ,H485250 ,H485316 ,H485369 ,H485433 ,H485489 ,H485574 ,H485641 ,H485071 ,H485144 ,H485200 ,H485253 ,H485317 ,H485370 ,H485434 ,H485490 ,H485576 ,H485642 ,H485072 ,H485145 ,H485201 ,H485254 ,H485318 ,H485371 ,H485435 ,H485491 ,H485577 ,H485643 ,H485073 ,H485146 ,H485202 ,H485255 ,H485319 ,H485372 ,H485436 ,H485495 ,H485578 ,H485644 ,H485074 ,H485147 ,H485203 ,H485256 ,H485320 ,H485373 ,H485437 ,H485517 ,H485579 ,H485646 ,H485075 ,H485148 ,H485204 ,H485257 ,H485321 ,H485374 ,H485438 ,H485518 ,H485580 ,H485647 ,H485076 ,H485149 ,H485205 ,H485258 ,H485322 ,H485375 ,H485439 ,H485519 ,H485581 ,H485648 ,H485077 ,H485150 ,H485209 ,H485259 ,H485323 ,H485376 ,H485440 ,H485520 ,H485582 ,H485653 ,H485078 ,H485151 ,H485210 ,H485260 ,H485324 ,H485377 ,H485443 ,H485521 ,H485583 ,H485654 ,H485079 ,H485152 ,H485211 ,H485261 ,H485325 ,H485378 ,H485444 ,H485523 ,H485585 ,H485656 ,H485081 ,H485161 ,H485212 ,H485262 ,H485326 ,H485379 ,H485445 ,H485524 ,H485586 ,H485657 ,H485083 ,H485162 ,H485213 ,H485264 ,H485327 ,H485380 ,H485446 ,H485525 ,H485587 ,H485658 ,H485084 ,H485164 ,H485214 ,H485265 ,H485328 ,H485381 ,H485447 ,H485529 ,H485588 ,H485660 ,H485085 ,H485165 ,H485215 ,H485266 ,H485329 ,H485382 ,H485448 ,H485530 ,H485590 ,H485661 ,H485087 ,H485166 ,H485216 ,H485267 ,H485330 ,H485383 ,H485449 ,H485531 ,H485592 ,H485667 ,H485092 ,H485168 ,H485217 ,H485268 ,H485331 ,H485384 ,H485450 ,H485532 ,H485593 ,H485668 ,H485099 ,H485169 ,H485218 ,H485276 ,H485332 ,H485385 ,H485451 ,H485533 ,H485594 ,H485669 ,H485102 ,H485170 ,H485219 ,H485277 ,H485333 ,H485386 ,H485452 ,H485534 ,H485596 ,H485670 ,H485103 ,H485171 ,H485220 ,H485278 ,H485334 ,H485387 ,H485453 ,H485536 ,H485601 ,H485671 ,H485104 ,H485172 ,H485221 ,H485279 ,H485335 ,H485388 ,H485454 ,H485537 ,H485602 ,H485672 ,H485105 ,H485173 ,H485222 ,H485285 ,H485336 ,H485389 ,H485455 ,H485538 ,H485603 ,H485675 ,H485106 ,H485174 ,H485223 ,H485286 ,H485337 ,H485390 ,H485456 ,H485540 ,H485604 ,H485676 ,H485107 ,H485175 ,H485224 ,H485288 ,H485338 ,H485391 ,H485457 ,H485542 ,H485607 ,H485677 ,H485108 ,H485176 ,H485225 ,H485289 ,H485339 ,H485392 ,H485458 ,H485543 ,H485608 ,H485678 ,H485109 ,H485177 ,H485226 ,H485291 ,H485340 ,H485393 ,H485459 ,H485545 ,H485609 ,H485679 ,H485110 ,H485178 ,H485227 ,H485292 ,H485341 ,H485396 ,H485460 ,H485546 ,H485612 ,H485681 ,H485111 ,H485179 ,H485228 ,H485293 ,H485342 ,H485397 ,H485461 ,H485547 ,H485613 ,H485682 ,H485113 ,H485180 ,H485229 ,H485294 ,H485343 ,H485398 ,H485462 ,H485548 ,H485614 ,H485683 ,H485116 ,H485181 ,H485230 ,H485295 ,H485345 ,H485399 ,H485463 ,H485549 ,H485615 ,H485684 ,H485117 ,H485182 ,H485231 ,H485296 ,H485346 ,H485400 ,H485465 ,H485550 ,H485617 ,H485685 ,H485123 ,H485183 ,H485232 ,H485297 ,H485347 ,H485401 ,H485471 ,H485554 ,H485620 ,H485686 ,H485127 ,H485184 ,H485233 ,H485298 ,H485348 ,H485402 ,H485475 ,H485555 ,H485621 ,H485687 ,H485129 ,H485185 ,H485234 ,H485299 ,H485349 ,H485403 ,H485476 ,H485556 ,H485622 ,H485692 ,H485130 ,H485186 ,H485235 ,H485301 ,H485350 ,H485404 ,H485477 ,H485557 ,H485628 ,H485694 ,H485131 ,H485187 ,H485236 ,H485302 ,H485352 ,H485405 ,H485478 ,H485558 ,H485629 ,H485695 ,H485132 ,H485188 ,H485237 ,H485303 ,H485353 ,H485412 ,H485479 ,H485559 ,H485630 ,H485696 ,H485133 ,H485189 ,H485238 ,H485304 ,H485357 ,H485413 ,H485480 ,H485560 ,H485632 ,H485697 ,H485698 ,H485753 ,H485802 ,H485874 ,H485937 ,H485994 ,H486070 ,H486136 ,H486183 ,H486255 ,H485699 ,H485754 ,H485803 ,H485875 ,H485941 ,H485995 ,H486071 ,H486137 ,H486184 ,H486256 ,H485700 ,H485755 ,H485805 ,H485876 ,H485943 ,H485996 ,H486072 ,H486138 ,H486185 ,H486257 ,H485701 ,H485756 ,H485812 ,H485879 ,H485944 ,H485997 ,H486073 ,H486139 ,H486186 ,H486258 ,H485702 ,H485757 ,H485815 ,H485890 ,H485945 ,H485998 ,H486074 ,H486140 ,H486187 ,H486259 ,H485703 ,H485758 ,H485816 ,H485891 ,H485946 ,H485999 ,H486075 ,H486141 ,H486196 ,H486260 ,H485704 ,H485759 ,H485817 ,H485892 ,H485947 ,H486000 ,H486076 ,H486142 ,H486197 ,H486261 ,H485705 ,H485760 ,H485818 ,H485893 ,H485949 ,H486001 ,H486086 ,H486145 ,H486198 ,H486262 ,H485706 ,H485761 ,H485820 ,H485894 ,H485950 ,H486003 ,H486087 ,H486146 ,H486199 ,H486263 ,H485707 ,H485762 ,H485821 ,H485895 ,H485953 ,H486004 ,H486088 ,H486147 ,H486200 ,H486264 ,H485708 ,H485763 ,H485823 ,H485896 ,H485954 ,H486005 ,H486089 ,H486148 ,H486201 ,H486265 ,H485711 ,H485764 ,H485824 ,H485897 ,H485955 ,H486007 ,H486091 ,H486149 ,H486202 ,H486266 ,H485715 ,H485765 ,H485825 ,H485898 ,H485956 ,H486016 ,H486093 ,H486150 ,H486203 ,H486267 ,H485716 ,H485766 ,H485826 ,H485899 ,H485957 ,H486017 ,H486094 ,H486151 ,H486204 ,H486271 ,H485720 ,H485767 ,H485827 ,H485900 ,H485958 ,H486021 ,H486095 ,H486152 ,H486205 ,H486272 ,H485721 ,H485768 ,H485829 ,H485901 ,H485960 ,H486024 ,H486096 ,H486153 ,H486206 ,H486278 ,H485722 ,H485769 ,H485830 ,H485902 ,H485961 ,H486026 ,H486097 ,H486154 ,H486207 ,H486279 ,H485723 ,H485770 ,H485831 ,H485903 ,H485962 ,H486029 ,H486098 ,H486155 ,H486208 ,H486280 ,H485724 ,H485771 ,H485835 ,H485904 ,H485963 ,H486030 ,H486101 ,H486156 ,H486209 ,H486281 ,H485725 ,H485772 ,H485836 ,H485905 ,H485964 ,H486031 ,H486103 ,H486157 ,H486210 ,H486282 ,H485726 ,H485773 ,H485837 ,H485909 ,H485965 ,H486032 ,H486104 ,H486158 ,H486213 ,H486283 ,H485727 ,H485774 ,H485838 ,H485910 ,H485966 ,H486033 ,H486105 ,H486159 ,H486214 ,H486284 ,H485728 ,H485775 ,H485839 ,H485911 ,H485967 ,H486034 ,H486107 ,H486160 ,H486215 ,H486285 ,H485729 ,H485776 ,H485840 ,H485912 ,H485968 ,H486035 ,H486108 ,H486161 ,H486216 ,H486286 ,H485730 ,H485777 ,H485841 ,H485913 ,H485969 ,H486036 ,H486109 ,H486162 ,H486217 ,H486287 ,H485731 ,H485778 ,H485842 ,H485914 ,H485970 ,H486037 ,H486110 ,H486163 ,H486218 ,H486288 ,H485732 ,H485779 ,H485843 ,H485915 ,H485971 ,H486038 ,H486111 ,H486164 ,H486219 ,H486289 ,H485733 ,H485780 ,H485846 ,H485916 ,H485972 ,H486046 ,H486114 ,H486165 ,H486220 ,H486290 ,H485734 ,H485781 ,H485850 ,H485918 ,H485973 ,H486047 ,H486115 ,H486166 ,H486224 ,H486291 ,H485736 ,H485782 ,H485851 ,H485919 ,H485974 ,H486048 ,H486116 ,H486167 ,H486225 ,H486294 ,H485737 ,H485783 ,H485852 ,H485920 ,H485975 ,H486049 ,H486117 ,H486168 ,H486226 ,H486299 ,H485739 ,H485784 ,H485853 ,H485921 ,H485976 ,H486050 ,H486118 ,H486169 ,H486227 ,H486300 ,H485740 ,H485785 ,H485854 ,H485922 ,H485977 ,H486051 ,H486119 ,H486170 ,H486228 ,H486301 ,H485741 ,H485786 ,H485855 ,H485923 ,H485978 ,H486052 ,H486120 ,H486171 ,H486231 ,H486302 ,H485742 ,H485787 ,H485859 ,H485924 ,H485979 ,H486053 ,H486121 ,H486172 ,H486233 ,H486303 ,H485743 ,H485788 ,H485860 ,H485925 ,H485980 ,H486054 ,H486123 ,H486173 ,H486234 ,H486304 ,H485744 ,H485792 ,H485862 ,H485927 ,H485982 ,H486055 ,H486124 ,H486174 ,H486235 ,H486306 ,H485745 ,H485793 ,H485863 ,H485928 ,H485983 ,H486056 ,H486125 ,H486175 ,H486236 ,H486308 ,H485746 ,H485795 ,H485867 ,H485929 ,H485984 ,H486057 ,H486126 ,H486176 ,H486237 ,H486309 ,H485747 ,H485796 ,H485868 ,H485931 ,H485985 ,H486061 ,H486127 ,H486177 ,H486238 ,H486310 ,H485748 ,H485797 ,H485869 ,H485932 ,H485987 ,H486062 ,H486128 ,H486178 ,H486239 ,H486311 ,H485749 ,H485798 ,H485870 ,H485933 ,H485989 ,H486063 ,H486129 ,H486179 ,H486245 ,H486312 ,H485750 ,H485799 ,H485871 ,H485934 ,H485990 ,H486064 ,H486132 ,H486180 ,H486250 ,H486314 ,H485751 ,H485800 ,H485872 ,H485935 ,H485991 ,H486068 ,H486133 ,H486181 ,H486253 ,H486315 ,H485752 ,H485801 ,H485873 ,H485936 ,H485992 ,H486069 ,H486135 ,H486182 ,H486254 ,H486316 ,H486319 ,H486376 ,H486433 ,H486492 ,H486546 ,H486617 ,H486666 ,H486729 ,H486809 ,H486878 ,H486320 ,H486377 ,H486436 ,H486493 ,H486547 ,H486618 ,H486671 ,H486730 ,H486810 ,H486879 ,H486322 ,H486378 ,H486437 ,H486494 ,H486548 ,H486619 ,H486672 ,H486731 ,H486811 ,H486880 ,H486323 ,H486379 ,H486438 ,H486495 ,H486549 ,H486620 ,H486677 ,H486732 ,H486812 ,H486881 ,H486324 ,H486380 ,H486441 ,H486496 ,H486550 ,H486621 ,H486680 ,H486733 ,H486813 ,H486882 ,H486325 ,H486381 ,H486446 ,H486497 ,H486554 ,H486623 ,H486681 ,H486735 ,H486814 ,H486883 ,H486326 ,H486382 ,H486447 ,H486498 ,H486555 ,H486624 ,H486683 ,H486736 ,H486815 ,H486884 ,H486327 ,H486383 ,H486451 ,H486499 ,H486556 ,H486626 ,H486685 ,H486737 ,H486816 ,H486885 ,H486328 ,H486384 ,H486452 ,H486500 ,H486557 ,H486627 ,H486686 ,H486738 ,H486817 ,H486887 ,H486330 ,H486385 ,H486453 ,H486501 ,H486558 ,H486628 ,H486687 ,H486739 ,H486818 ,H486888 ,H486331 ,H486386 ,H486454 ,H486502 ,H486559 ,H486629 ,H486688 ,H486740 ,H486820 ,H486889 ,H486334 ,H486387 ,H486455 ,H486503 ,H486560 ,H486630 ,H486689 ,H486754 ,H486823 ,H486890 ,H486336 ,H486388 ,H486456 ,H486504 ,H486561 ,H486631 ,H486690 ,H486755 ,H486824 ,H486891 ,H486338 ,H486389 ,H486457 ,H486505 ,H486562 ,H486632 ,H486691 ,H486756 ,H486825 ,H486892 ,H486339 ,H486390 ,H486458 ,H486506 ,H486563 ,H486633 ,H486694 ,H486757 ,H486826 ,H486893 ,H486340 ,H486391 ,H486459 ,H486507 ,H486564 ,H486634 ,H486695 ,H486758 ,H486827 ,H486895 ,H486341 ,H486392 ,H486460 ,H486508 ,H486565 ,H486635 ,H486696 ,H486759 ,H486828 ,H486896 ,H486343 ,H486393 ,H486461 ,H486509 ,H486566 ,H486636 ,H486697 ,H486766 ,H486830 ,H486897 ,H486344 ,H486394 ,H486462 ,H486510 ,H486567 ,H486637 ,H486698 ,H486767 ,H486831 ,H486899 ,H486345 ,H486401 ,H486463 ,H486511 ,H486568 ,H486638 ,H486699 ,H486769 ,H486832 ,H486900 ,H486346 ,H486402 ,H486464 ,H486517 ,H486570 ,H486639 ,H486701 ,H486781 ,H486833 ,H486901 ,H486347 ,H486403 ,H486465 ,H486518 ,H486583 ,H486640 ,H486702 ,H486782 ,H486834 ,H486902 ,H486348 ,H486404 ,H486466 ,H486520 ,H486584 ,H486641 ,H486703 ,H486784 ,H486835 ,H486903 ,H486349 ,H486405 ,H486467 ,H486522 ,H486585 ,H486642 ,H486704 ,H486785 ,H486836 ,H486904 ,H486350 ,H486406 ,H486468 ,H486523 ,H486586 ,H486643 ,H486705 ,H486786 ,H486837 ,H486906 ,H486352 ,H486407 ,H486469 ,H486524 ,H486587 ,H486644 ,H486706 ,H486788 ,H486838 ,H486907 ,H486353 ,H486408 ,H486470 ,H486525 ,H486588 ,H486645 ,H486707 ,H486789 ,H486839 ,H486908 ,H486354 ,H486409 ,H486471 ,H486526 ,H486589 ,H486646 ,H486708 ,H486790 ,H486840 ,H486909 ,H486356 ,H486412 ,H486472 ,H486527 ,H486590 ,H486647 ,H486710 ,H486791 ,H486841 ,H486910 ,H486357 ,H486413 ,H486473 ,H486528 ,H486591 ,H486648 ,H486711 ,H486792 ,H486847 ,H486911 ,H486358 ,H486414 ,H486474 ,H486529 ,H486592 ,H486649 ,H486712 ,H486793 ,H486848 ,H486912 ,H486359 ,H486415 ,H486475 ,H486530 ,H486597 ,H486651 ,H486713 ,H486794 ,H486849 ,H486913 ,H486360 ,H486420 ,H486476 ,H486531 ,H486598 ,H486652 ,H486714 ,H486795 ,H486850 ,H486915 ,H486361 ,H486421 ,H486477 ,H486532 ,H486599 ,H486653 ,H486715 ,H486796 ,H486851 ,H486916 ,H486362 ,H486422 ,H486478 ,H486535 ,H486600 ,H486654 ,H486716 ,H486798 ,H486852 ,H486917 ,H486363 ,H486423 ,H486479 ,H486536 ,H486603 ,H486655 ,H486717 ,H486799 ,H486853 ,H486918 ,H486364 ,H486424 ,H486480 ,H486537 ,H486604 ,H486656 ,H486718 ,H486800 ,H486854 ,H486919 ,H486365 ,H486425 ,H486481 ,H486538 ,H486606 ,H486657 ,H486719 ,H486801 ,H486855 ,H486920 ,H486366 ,H486426 ,H486482 ,H486539 ,H486607 ,H486659 ,H486720 ,H486802 ,H486856 ,H486921 ,H486367 ,H486427 ,H486483 ,H486540 ,H486608 ,H486660 ,H486721 ,H486803 ,H486857 ,H486922 ,H486368 ,H486428 ,H486484 ,H486541 ,H486609 ,H486661 ,H486722 ,H486804 ,H486866 ,H486923 ,H486372 ,H486429 ,H486485 ,H486542 ,H486610 ,H486662 ,H486723 ,H486805 ,H486871 ,H486924 ,H486373 ,H486430 ,H486486 ,H486543 ,H486611 ,H486663 ,H486724 ,H486806 ,H486872 ,H486926 ,H486374 ,H486431 ,H486488 ,H486544 ,H486613 ,H486664 ,H486727 ,H486807 ,H486873 ,H486930 ,H486375 ,H486432 ,H486489 ,H486545 ,H486614 ,H486665 ,H486728 ,H486808 ,H486874 ,H486931 ,H486932 ,H486994 ,H487067 ,H487119 ,H487206 ,H487279 ,H487345 ,H487436 ,H487502 ,H487574 ,H486933 ,H486995 ,H487068 ,H487120 ,H487210 ,H487286 ,H487346 ,H487437 ,H487503 ,H487575 ,H486934 ,H486996 ,H487069 ,H487121 ,H487211 ,H487287 ,H487347 ,H487438 ,H487504 ,H487576 ,H486935 ,H486997 ,H487070 ,H487125 ,H487213 ,H487288 ,H487350 ,H487443 ,H487505 ,H487581 ,H486936 ,H486998 ,H487071 ,H487161 ,H487214 ,H487289 ,H487352 ,H487444 ,H487506 ,H487582 ,H486937 ,H486999 ,H487072 ,H487162 ,H487223 ,H487290 ,H487378 ,H487445 ,H487509 ,H487583 ,H486938 ,H487000 ,H487073 ,H487164 ,H487225 ,H487291 ,H487379 ,H487446 ,H487510 ,H487584 ,H486939 ,H487001 ,H487075 ,H487165 ,H487226 ,H487292 ,H487380 ,H487447 ,H487511 ,H487587 ,H486940 ,H487002 ,H487076 ,H487166 ,H487227 ,H487293 ,H487381 ,H487448 ,H487512 ,H487588 ,H486941 ,H487003 ,H487077 ,H487167 ,H487228 ,H487294 ,H487382 ,H487449 ,H487513 ,H487590 ,H486944 ,H487004 ,H487078 ,H487168 ,H487229 ,H487295 ,H487383 ,H487450 ,H487514 ,H487591 ,H486945 ,H487005 ,H487082 ,H487169 ,H487231 ,H487296 ,H487386 ,H487452 ,H487515 ,H487592 ,H486949 ,H487006 ,H487083 ,H487170 ,H487232 ,H487297 ,H487387 ,H487457 ,H487516 ,H487593 ,H486950 ,H487007 ,H487084 ,H487171 ,H487233 ,H487298 ,H487392 ,H487458 ,H487517 ,H487594 ,H486951 ,H487008 ,H487085 ,H487172 ,H487235 ,H487299 ,H487393 ,H487459 ,H487521 ,H487597 ,H486957 ,H487009 ,H487086 ,H487173 ,H487236 ,H487300 ,H487394 ,H487460 ,H487522 ,H487598 ,H486958 ,H487010 ,H487088 ,H487174 ,H487237 ,H487301 ,H487395 ,H487466 ,H487523 ,H487599 ,H486959 ,H487011 ,H487089 ,H487175 ,H487238 ,H487302 ,H487396 ,H487467 ,H487524 ,H487600 ,H486960 ,H487012 ,H487090 ,H487176 ,H487239 ,H487303 ,H487397 ,H487468 ,H487525 ,H487601 ,H486961 ,H487013 ,H487091 ,H487177 ,H487240 ,H487304 ,H487398 ,H487469 ,H487526 ,H487602 ,H486963 ,H487014 ,H487092 ,H487178 ,H487247 ,H487305 ,H487399 ,H487470 ,H487527 ,H487605 ,H486964 ,H487015 ,H487093 ,H487179 ,H487248 ,H487306 ,H487400 ,H487473 ,H487528 ,H487606 ,H486965 ,H487016 ,H487094 ,H487180 ,H487249 ,H487307 ,H487401 ,H487474 ,H487529 ,H487608 ,H486966 ,H487017 ,H487097 ,H487181 ,H487256 ,H487308 ,H487402 ,H487475 ,H487530 ,H487609 ,H486967 ,H487019 ,H487098 ,H487182 ,H487257 ,H487309 ,H487403 ,H487476 ,H487531 ,H487610 ,H486968 ,H487020 ,H487099 ,H487183 ,H487258 ,H487310 ,H487407 ,H487477 ,H487532 ,H487611 ,H486969 ,H487021 ,H487100 ,H487184 ,H487259 ,H487313 ,H487408 ,H487478 ,H487533 ,H487612 ,H486970 ,H487022 ,H487101 ,H487185 ,H487260 ,H487314 ,H487409 ,H487479 ,H487534 ,H487613 ,H486971 ,H487023 ,H487102 ,H487186 ,H487261 ,H487315 ,H487410 ,H487480 ,H487537 ,H487614 ,H486972 ,H487024 ,H487103 ,H487189 ,H487262 ,H487320 ,H487411 ,H487481 ,H487538 ,H487615 ,H486978 ,H487025 ,H487104 ,H487190 ,H487263 ,H487321 ,H487412 ,H487482 ,H487539 ,H487616 ,H486979 ,H487026 ,H487105 ,H487191 ,H487264 ,H487322 ,H487413 ,H487483 ,H487540 ,H487619 ,H486980 ,H487028 ,H487106 ,H487192 ,H487266 ,H487323 ,H487414 ,H487484 ,H487541 ,H487620 ,H486982 ,H487029 ,H487107 ,H487193 ,H487267 ,H487324 ,H487415 ,H487485 ,H487549 ,H487621 ,H486983 ,H487031 ,H487108 ,H487195 ,H487268 ,H487325 ,H487416 ,H487486 ,H487550 ,H487622 ,H486984 ,H487032 ,H487109 ,H487196 ,H487269 ,H487326 ,H487417 ,H487487 ,H487551 ,H487624 ,H486985 ,H487033 ,H487110 ,H487197 ,H487270 ,H487327 ,H487418 ,H487493 ,H487554 ,H487625 ,H486986 ,H487034 ,H487111 ,H487198 ,H487271 ,H487328 ,H487423 ,H487494 ,H487555 ,H487626 ,H486987 ,H487035 ,H487112 ,H487199 ,H487272 ,H487330 ,H487425 ,H487495 ,H487556 ,H487627 ,H486988 ,H487036 ,H487113 ,H487200 ,H487273 ,H487331 ,H487426 ,H487496 ,H487557 ,H487628 ,H486989 ,H487037 ,H487114 ,H487201 ,H487274 ,H487332 ,H487431 ,H487497 ,H487569 ,H487629 ,H486990 ,H487038 ,H487115 ,H487202 ,H487275 ,H487333 ,H487432 ,H487498 ,H487570 ,H487630 ,H486991 ,H487039 ,H487116 ,H487203 ,H487276 ,H487340 ,H487433 ,H487499 ,H487571 ,H487632 ,H486992 ,H487040 ,H487117 ,H487204 ,H487277 ,H487343 ,H487434 ,H487500 ,H487572 ,H487633 ,H486993 ,H487066 ,H487118 ,H487205 ,H487278 ,H487344 ,H487435 ,H487501 ,H487573 ,H487634 ,H487635 ,H487708 ,H487767 ,H487829 ,H487887 ,H487950 ,H488033 ,H488099 ,H488155 ,H488234 ,H487636 ,H487709 ,H487768 ,H487834 ,H487888 ,H487951 ,H488034 ,H488100 ,H488156 ,H488235 ,H487646 ,H487710 ,H487769 ,H487835 ,H487889 ,H487952 ,H488035 ,H488101 ,H488157 ,H488236 ,H487647 ,H487711 ,H487770 ,H487836 ,H487890 ,H487953 ,H488036 ,H488102 ,H488158 ,H488237 ,H487650 ,H487714 ,H487771 ,H487837 ,H487891 ,H487954 ,H488037 ,H488103 ,H488159 ,H488238 ,H487651 ,H487715 ,H487772 ,H487838 ,H487892 ,H487955 ,H488038 ,H488104 ,H488163 ,H488239 ,H487652 ,H487720 ,H487773 ,H487839 ,H487893 ,H487956 ,H488039 ,H488106 ,H488165 ,H488240 ,H487653 ,H487721 ,H487774 ,H487840 ,H487894 ,H487965 ,H488040 ,H488107 ,H488166 ,H488241 ,H487654 ,H487723 ,H487775 ,H487841 ,H487895 ,H487970 ,H488041 ,H488108 ,H488169 ,H488242 ,H487655 ,H487724 ,H487776 ,H487842 ,H487898 ,H487971 ,H488043 ,H488109 ,H488170 ,H488243 ,H487656 ,H487725 ,H487777 ,H487843 ,H487899 ,H487972 ,H488044 ,H488110 ,H488171 ,H488244 ,H487657 ,H487726 ,H487778 ,H487844 ,H487902 ,H487980 ,H488045 ,H488111 ,H488172 ,H488245 ,H487659 ,H487728 ,H487779 ,H487845 ,H487906 ,H487981 ,H488047 ,H488112 ,H488173 ,H488247 ,H487660 ,H487729 ,H487780 ,H487847 ,H487907 ,H487982 ,H488048 ,H488113 ,H488179 ,H488248 ,H487661 ,H487730 ,H487781 ,H487848 ,H487908 ,H487983 ,H488049 ,H488114 ,H488182 ,H488249 ,H487662 ,H487731 ,H487782 ,H487849 ,H487909 ,H487990 ,H488050 ,H488115 ,H488183 ,H488250 ,H487663 ,H487732 ,H487783 ,H487850 ,H487911 ,H487991 ,H488051 ,H488116 ,H488184 ,H488251 ,H487664 ,H487735 ,H487784 ,H487851 ,H487912 ,H487994 ,H488052 ,H488117 ,H488185 ,H488254 ,H487675 ,H487736 ,H487785 ,H487852 ,H487913 ,H487995 ,H488054 ,H488118 ,H488186 ,H488261 ,H487676 ,H487737 ,H487786 ,H487853 ,H487919 ,H487996 ,H488055 ,H488119 ,H488188 ,H488262 ,H487677 ,H487739 ,H487788 ,H487854 ,H487920 ,H487997 ,H488056 ,H488120 ,H488189 ,H488263 ,H487680 ,H487740 ,H487789 ,H487855 ,H487921 ,H487998 ,H488057 ,H488124 ,H488190 ,H488264 ,H487681 ,H487741 ,H487790 ,H487858 ,H487922 ,H487999 ,H488058 ,H488125 ,H488191 ,H488265 ,H487682 ,H487742 ,H487791 ,H487859 ,H487923 ,H488001 ,H488059 ,H488126 ,H488192 ,H488266 ,H487683 ,H487743 ,H487792 ,H487860 ,H487924 ,H488002 ,H488061 ,H488127 ,H488193 ,H488267 ,H487684 ,H487744 ,H487793 ,H487862 ,H487925 ,H488003 ,H488063 ,H488128 ,H488194 ,H488268 ,H487685 ,H487745 ,H487794 ,H487863 ,H487926 ,H488004 ,H488072 ,H488129 ,H488196 ,H488269 ,H487686 ,H487746 ,H487795 ,H487865 ,H487927 ,H488005 ,H488073 ,H488130 ,H488209 ,H488270 ,H487687 ,H487747 ,H487796 ,H487868 ,H487928 ,H488006 ,H488075 ,H488131 ,H488212 ,H488275 ,H487688 ,H487748 ,H487798 ,H487869 ,H487933 ,H488007 ,H488076 ,H488132 ,H488213 ,H488276 ,H487689 ,H487749 ,H487811 ,H487870 ,H487934 ,H488008 ,H488077 ,H488133 ,H488215 ,H488277 ,H487690 ,H487750 ,H487812 ,H487871 ,H487935 ,H488018 ,H488078 ,H488134 ,H488216 ,H488278 ,H487691 ,H487751 ,H487814 ,H487873 ,H487936 ,H488019 ,H488079 ,H488135 ,H488217 ,H488279 ,H487692 ,H487752 ,H487815 ,H487874 ,H487937 ,H488020 ,H488080 ,H488136 ,H488220 ,H488280 ,H487693 ,H487753 ,H487817 ,H487876 ,H487938 ,H488021 ,H488081 ,H488137 ,H488221 ,H488281 ,H487694 ,H487754 ,H487818 ,H487877 ,H487939 ,H488022 ,H488082 ,H488139 ,H488222 ,H488282 ,H487695 ,H487755 ,H487819 ,H487878 ,H487940 ,H488023 ,H488083 ,H488140 ,H488223 ,H488284 ,H487696 ,H487757 ,H487820 ,H487879 ,H487941 ,H488024 ,H488084 ,H488141 ,H488225 ,H488285 ,H487697 ,H487758 ,H487821 ,H487880 ,H487942 ,H488025 ,H488085 ,H488142 ,H488227 ,H488287 ,H487698 ,H487759 ,H487822 ,H487881 ,H487943 ,H488026 ,H488086 ,H488144 ,H488228 ,H488288 ,H487699 ,H487760 ,H487823 ,H487882 ,H487944 ,H488027 ,H488087 ,H488150 ,H488229 ,H488289 ,H487700 ,H487761 ,H487824 ,H487883 ,H487945 ,H488028 ,H488089 ,H488151 ,H488230 ,H488290 ,H487703 ,H487762 ,H487825 ,H487884 ,H487946 ,H488029 ,H488096 ,H488152 ,H488231 ,H488291 ,H487704 ,H487763 ,H487826 ,H487885 ,H487948 ,H488030 ,H488097 ,H488153 ,H488232 ,H488295 ,H487705 ,H487766 ,H487828 ,H487886 ,H487949 ,H488031 ,H488098 ,H488154 ,H488233 ,H488296 ,H488297 ,H488374 ,H488434 ,H488488 ,H488548 ,H488614 ,H488663 ,H488714 ,H488773 ,H488833 ,H488303 ,H488375 ,H488435 ,H488489 ,H488549 ,H488615 ,H488664 ,H488715 ,H488775 ,H488842 ,H488309 ,H488376 ,H488436 ,H488490 ,H488550 ,H488616 ,H488665 ,H488716 ,H488776 ,H488843 ,H488310 ,H488377 ,H488437 ,H488491 ,H488551 ,H488617 ,H488666 ,H488717 ,H488777 ,H488844 ,H488314 ,H488378 ,H488438 ,H488492 ,H488552 ,H488618 ,H488667 ,H488718 ,H488779 ,H488850 ,H488315 ,H488379 ,H488439 ,H488493 ,H488553 ,H488619 ,H488668 ,H488719 ,H488780 ,H488851 ,H488316 ,H488380 ,H488440 ,H488494 ,H488558 ,H488620 ,H488669 ,H488720 ,H488782 ,H488853 ,H488317 ,H488381 ,H488441 ,H488495 ,H488559 ,H488621 ,H488672 ,H488721 ,H488788 ,H488854 ,H488318 ,H488382 ,H488442 ,H488496 ,H488560 ,H488622 ,H488673 ,H488722 ,H488789 ,H488855 ,H488319 ,H488383 ,H488443 ,H488497 ,H488561 ,H488623 ,H488674 ,H488724 ,H488790 ,H488857 ,H488320 ,H488384 ,H488444 ,H488498 ,H488565 ,H488624 ,H488676 ,H488725 ,H488791 ,H488858 ,H488322 ,H488385 ,H488445 ,H488499 ,H488566 ,H488625 ,H488677 ,H488726 ,H488792 ,H488859 ,H488323 ,H488387 ,H488446 ,H488507 ,H488567 ,H488626 ,H488678 ,H488727 ,H488793 ,H488860 ,H488324 ,H488388 ,H488447 ,H488508 ,H488569 ,H488627 ,H488679 ,H488734 ,H488794 ,H488861 ,H488326 ,H488389 ,H488448 ,H488512 ,H488570 ,H488628 ,H488680 ,H488735 ,H488795 ,H488862 ,H488327 ,H488390 ,H488449 ,H488513 ,H488573 ,H488629 ,H488681 ,H488736 ,H488796 ,H488863 ,H488329 ,H488391 ,H488450 ,H488514 ,H488574 ,H488630 ,H488682 ,H488738 ,H488798 ,H488864 ,H488330 ,H488392 ,H488453 ,H488515 ,H488575 ,H488631 ,H488683 ,H488739 ,H488799 ,H488865 ,H488331 ,H488393 ,H488454 ,H488516 ,H488576 ,H488632 ,H488684 ,H488740 ,H488802 ,H488866 ,H488332 ,H488394 ,H488455 ,H488517 ,H488577 ,H488633 ,H488685 ,H488741 ,H488803 ,H488867 ,H488333 ,H488395 ,H488456 ,H488518 ,H488578 ,H488634 ,H488686 ,H488742 ,H488804 ,H488868 ,H488337 ,H488396 ,H488457 ,H488519 ,H488579 ,H488635 ,H488687 ,H488743 ,H488805 ,H488869 ,H488338 ,H488397 ,H488459 ,H488520 ,H488581 ,H488636 ,H488688 ,H488744 ,H488806 ,H488875 ,H488339 ,H488398 ,H488460 ,H488521 ,H488582 ,H488637 ,H488689 ,H488745 ,H488807 ,H488876 ,H488340 ,H488399 ,H488461 ,H488525 ,H488583 ,H488638 ,H488690 ,H488748 ,H488808 ,H488877 ,H488341 ,H488401 ,H488462 ,H488526 ,H488584 ,H488639 ,H488691 ,H488749 ,H488809 ,H488878 ,H488342 ,H488404 ,H488463 ,H488527 ,H488585 ,H488640 ,H488692 ,H488750 ,H488810 ,H488879 ,H488343 ,H488405 ,H488465 ,H488528 ,H488586 ,H488641 ,H488693 ,H488751 ,H488811 ,H488880 ,H488344 ,H488406 ,H488466 ,H488529 ,H488588 ,H488642 ,H488694 ,H488752 ,H488812 ,H488881 ,H488345 ,H488413 ,H488467 ,H488530 ,H488589 ,H488643 ,H488695 ,H488753 ,H488813 ,H488882 ,H488346 ,H488414 ,H488468 ,H488532 ,H488590 ,H488648 ,H488696 ,H488756 ,H488814 ,H488883 ,H488349 ,H488415 ,H488469 ,H488533 ,H488591 ,H488649 ,H488698 ,H488757 ,H488815 ,H488884 ,H488353 ,H488416 ,H488470 ,H488534 ,H488592 ,H488650 ,H488699 ,H488758 ,H488816 ,H488885 ,H488354 ,H488417 ,H488471 ,H488535 ,H488594 ,H488651 ,H488700 ,H488759 ,H488818 ,H488886 ,H488355 ,H488418 ,H488472 ,H488536 ,H488595 ,H488652 ,H488701 ,H488760 ,H488820 ,H488887 ,H488356 ,H488419 ,H488473 ,H488537 ,H488596 ,H488653 ,H488702 ,H488761 ,H488821 ,H488888 ,H488357 ,H488423 ,H488474 ,H488538 ,H488597 ,H488654 ,H488703 ,H488763 ,H488822 ,H488889 ,H488358 ,H488424 ,H488475 ,H488539 ,H488598 ,H488655 ,H488704 ,H488764 ,H488823 ,H488890 ,H488359 ,H488425 ,H488476 ,H488541 ,H488599 ,H488656 ,H488705 ,H488766 ,H488824 ,H488891 ,H488360 ,H488426 ,H488477 ,H488542 ,H488600 ,H488657 ,H488708 ,H488767 ,H488825 ,H488892 ,H488367 ,H488427 ,H488481 ,H488543 ,H488601 ,H488658 ,H488709 ,H488768 ,H488826 ,H488893 ,H488368 ,H488429 ,H488484 ,H488544 ,H488602 ,H488659 ,H488710 ,H488769 ,H488827 ,H488894 ,H488369 ,H488430 ,H488485 ,H488545 ,H488603 ,H488660 ,H488711 ,H488770 ,H488830 ,H488895 ,H488370 ,H488432 ,H488486 ,H488546 ,H488612 ,H488661 ,H488712 ,H488771 ,H488831 ,H488898 ,H488373 ,H488433 ,H488487 ,H488547 ,H488613 ,H488662 ,H488713 ,H488772 ,H488832 ,H488899 ,H488900 ,H488961 ,H489017 ,H489074 ,H489130 ,H489189 ,H489272 ,H489319 ,H489379 ,H489447 ,H488901 ,H488962 ,H489018 ,H489075 ,H489131 ,H489190 ,H489273 ,H489320 ,H489380 ,H489452 ,H488902 ,H488963 ,H489019 ,H489076 ,H489132 ,H489191 ,H489274 ,H489321 ,H489381 ,H489453 ,H488903 ,H488965 ,H489020 ,H489077 ,H489133 ,H489192 ,H489275 ,H489322 ,H489382 ,H489454 ,H488904 ,H488966 ,H489021 ,H489078 ,H489138 ,H489193 ,H489276 ,H489323 ,H489383 ,H489455 ,H488906 ,H488967 ,H489022 ,H489079 ,H489139 ,H489196 ,H489277 ,H489324 ,H489384 ,H489456 ,H488907 ,H488969 ,H489023 ,H489080 ,H489140 ,H489197 ,H489278 ,H489325 ,H489385 ,H489457 ,H488910 ,H488970 ,H489026 ,H489081 ,H489141 ,H489200 ,H489279 ,H489326 ,H489387 ,H489458 ,H488911 ,H488971 ,H489029 ,H489082 ,H489142 ,H489202 ,H489280 ,H489327 ,H489389 ,H489459 ,H488912 ,H488972 ,H489030 ,H489083 ,H489143 ,H489203 ,H489281 ,H489328 ,H489390 ,H489460 ,H488913 ,H488973 ,H489031 ,H489084 ,H489144 ,H489211 ,H489282 ,H489329 ,H489391 ,H489461 ,H488919 ,H488974 ,H489032 ,H489085 ,H489147 ,H489212 ,H489283 ,H489330 ,H489392 ,H489462 ,H488920 ,H488975 ,H489033 ,H489086 ,H489149 ,H489213 ,H489284 ,H489331 ,H489393 ,H489463 ,H488921 ,H488976 ,H489034 ,H489087 ,H489150 ,H489215 ,H489285 ,H489332 ,H489398 ,H489464 ,H488922 ,H488977 ,H489037 ,H489088 ,H489151 ,H489216 ,H489286 ,H489333 ,H489399 ,H489465 ,H488923 ,H488978 ,H489038 ,H489090 ,H489152 ,H489217 ,H489287 ,H489334 ,H489401 ,H489466 ,H488924 ,H488979 ,H489039 ,H489091 ,H489153 ,H489218 ,H489288 ,H489335 ,H489402 ,H489467 ,H488925 ,H488980 ,H489040 ,H489092 ,H489154 ,H489219 ,H489289 ,H489336 ,H489403 ,H489468 ,H488926 ,H488981 ,H489041 ,H489094 ,H489155 ,H489220 ,H489290 ,H489337 ,H489404 ,H489469 ,H488927 ,H488982 ,H489042 ,H489095 ,H489156 ,H489221 ,H489291 ,H489338 ,H489405 ,H489470 ,H488928 ,H488983 ,H489043 ,H489096 ,H489157 ,H489222 ,H489292 ,H489344 ,H489406 ,H489471 ,H488929 ,H488984 ,H489044 ,H489097 ,H489158 ,H489230 ,H489293 ,H489346 ,H489407 ,H489472 ,H488930 ,H488985 ,H489046 ,H489098 ,H489159 ,H489231 ,H489296 ,H489347 ,H489408 ,H489473 ,H488931 ,H488986 ,H489047 ,H489099 ,H489160 ,H489232 ,H489297 ,H489348 ,H489409 ,H489474 ,H488932 ,H488987 ,H489051 ,H489100 ,H489161 ,H489234 ,H489298 ,H489349 ,H489410 ,H489475 ,H488933 ,H488991 ,H489053 ,H489101 ,H489162 ,H489235 ,H489299 ,H489350 ,H489411 ,H489476 ,H488934 ,H488992 ,H489054 ,H489102 ,H489163 ,H489236 ,H489300 ,H489351 ,H489412 ,H489478 ,H488935 ,H488993 ,H489055 ,H489103 ,H489164 ,H489237 ,H489301 ,H489352 ,H489413 ,H489480 ,H488940 ,H488994 ,H489056 ,H489104 ,H489166 ,H489238 ,H489302 ,H489355 ,H489414 ,H489481 ,H488941 ,H488995 ,H489057 ,H489105 ,H489167 ,H489239 ,H489303 ,H489356 ,H489415 ,H489482 ,H488943 ,H488996 ,H489058 ,H489106 ,H489168 ,H489240 ,H489304 ,H489358 ,H489416 ,H489483 ,H488944 ,H488997 ,H489059 ,H489110 ,H489169 ,H489241 ,H489305 ,H489359 ,H489418 ,H489484 ,H488945 ,H488998 ,H489060 ,H489111 ,H489170 ,H489243 ,H489306 ,H489360 ,H489419 ,H489486 ,H488946 ,H488999 ,H489062 ,H489112 ,H489171 ,H489244 ,H489307 ,H489361 ,H489420 ,H489487 ,H488949 ,H489000 ,H489063 ,H489113 ,H489172 ,H489245 ,H489308 ,H489363 ,H489421 ,H489488 ,H488951 ,H489001 ,H489064 ,H489114 ,H489173 ,H489246 ,H489309 ,H489364 ,H489422 ,H489489 ,H488952 ,H489007 ,H489065 ,H489115 ,H489174 ,H489253 ,H489310 ,H489365 ,H489423 ,H489490 ,H488953 ,H489008 ,H489066 ,H489116 ,H489176 ,H489255 ,H489311 ,H489366 ,H489424 ,H489491 ,H488954 ,H489009 ,H489067 ,H489119 ,H489177 ,H489256 ,H489312 ,H489367 ,H489425 ,H489492 ,H488955 ,H489010 ,H489068 ,H489121 ,H489178 ,H489257 ,H489313 ,H489368 ,H489426 ,H489493 ,H488956 ,H489011 ,H489069 ,H489124 ,H489180 ,H489264 ,H489314 ,H489369 ,H489440 ,H489494 ,H488957 ,H489012 ,H489070 ,H489125 ,H489181 ,H489265 ,H489315 ,H489370 ,H489441 ,H489495 ,H488958 ,H489013 ,H489071 ,H489126 ,H489182 ,H489269 ,H489316 ,H489371 ,H489444 ,H489506 ,H488959 ,H489015 ,H489072 ,H489127 ,H489184 ,H489270 ,H489317 ,H489376 ,H489445 ,H489507 ,H488960 ,H489016 ,H489073 ,H489128 ,H489185 ,H489271 ,H489318 ,H489377 ,H489446 ,H489508 ,H489509 ,H489571 ,H489657 ,H489731 ,H489841 ,H489951 ,H48A065 ,H48A215 ,H489514 ,H489577 ,H489658 ,H489732 ,H489842 ,H489953 ,H48A066 ,H48A216 ,H489515 ,H489578 ,H489659 ,H489733 ,H489844 ,H489956 ,H48A088 ,H48A217 ,H489520 ,H489579 ,H489660 ,H489735 ,H489845 ,H489957 ,H48A089 ,H48A224 ,H489521 ,H489580 ,H489661 ,H489736 ,H489850 ,H489958 ,H48A090 ,H48A225 ,H489522 ,H489581 ,H489662 ,H489737 ,H489863 ,H489959 ,H48A091 ,H48A226 ,H489523 ,H489582 ,H489663 ,H489740 ,H489864 ,H489962 ,H48A092 ,H48A227 ,H489524 ,H489585 ,H489666 ,H489741 ,H489865 ,H489963 ,H48A093 ,H48A228 ,H489525 ,H489588 ,H489668 ,H489743 ,H489866 ,H489988 ,H48A094 ,H48A229 ,H489526 ,H489589 ,H489670 ,H489744 ,H489867 ,H489989 ,H48A095 ,H48A230 ,H489527 ,H489591 ,H489671 ,H489745 ,H489868 ,H489990 ,H48A096 ,H48A254 ,H489530 ,H489592 ,H489672 ,H489746 ,H489872 ,H489991 ,H48A099 ,H48A255 ,H489531 ,H489593 ,H489673 ,H489747 ,H489873 ,H489992 ,H48A101 ,H48A256 ,H489533 ,H489594 ,H489674 ,H489748 ,H489874 ,H489996 ,H48A102 ,H48A257 ,H489534 ,H489595 ,H489675 ,H489749 ,H489875 ,H489997 ,H48A103 ,H48A261 ,H489535 ,H489596 ,H489676 ,H489750 ,H489876 ,H489998 ,H48A104 ,H48A262 ,H489536 ,H489598 ,H489683 ,H489761 ,H489880 ,H48A002 ,H48A107 ,H48A263 ,H489537 ,H489599 ,H489684 ,H489762 ,H489881 ,H48A006 ,H48A110 ,H48A264 ,H489538 ,H489600 ,H489685 ,H489763 ,H489890 ,H48A007 ,H48A136 ,H48A265 ,H489539 ,H489601 ,H489686 ,H489764 ,H489891 ,H48A008 ,H48A137 ,H48A266 ,H489540 ,H489602 ,H489687 ,H489765 ,H489898 ,H48A009 ,H48A147 ,H48A273 ,H489541 ,H489603 ,H489691 ,H489766 ,H489900 ,H48A010 ,H48A148 ,H48A288 ,H489542 ,H489604 ,H489692 ,H489767 ,H489901 ,H48A019 ,H48A149 ,H48A312 ,H489543 ,H489605 ,H489699 ,H489780 ,H489903 ,H48A020 ,H48A150 ,H48A318 ,H489544 ,H489606 ,H489700 ,H489781 ,H489906 ,H48A021 ,H48A151 ,H48A319 ,H489545 ,H489607 ,H489701 ,H489782 ,H489917 ,H48A022 ,H48A152 ,H489546 ,H489617 ,H489702 ,H489783 ,H489927 ,H48A023 ,H48A153 ,H489548 ,H489622 ,H489703 ,H489784 ,H489933 ,H48A029 ,H48A156 ,H489549 ,H489623 ,H489704 ,H489785 ,H489934 ,H48A030 ,H48A160 ,H489550 ,H489631 ,H489706 ,H489788 ,H489935 ,H48A031 ,H48A165 ,H489552 ,H489632 ,H489707 ,H489789 ,H489936 ,H48A035 ,H48A166 ,H489553 ,H489639 ,H489711 ,H489790 ,H489937 ,H48A036 ,H48A167 ,H489554 ,H489640 ,H489712 ,H489791 ,H489938 ,H48A039 ,H48A168 ,H489555 ,H489641 ,H489713 ,H489797 ,H489939 ,H48A044 ,H48A169 ,H489556 ,H489642 ,H489714 ,H489803 ,H489940 ,H48A045 ,H48A170 ,H489557 ,H489643 ,H489715 ,H489804 ,H489941 ,H48A046 ,H48A171 ,H489558 ,H489644 ,H489716 ,H489811 ,H489942 ,H48A054 ,H48A195 ,H489559 ,H489649 ,H489717 ,H489818 ,H489943 ,H48A055 ,H48A199 ,H489561 ,H489650 ,H489718 ,H489820 ,H489944 ,H48A057 ,H48A200 ,H489562 ,H489651 ,H489719 ,H489826 ,H489945 ,H48A058 ,H48A201 ,H489563 ,H489652 ,H489720 ,H489831 ,H489946 ,H48A059 ,H48A202 ,H489564 ,H489653 ,H489726 ,H489837 ,H489947 ,H48A060 ,H48A211 ,H489565 ,H489654 ,H489727 ,H489838 ,H489948 ,H48A062 ,H48A212 ,H489567 ,H489655 ,H489728 ,H489839 ,H489949 ,H48A063 ,H48A213 ,H489568 ,H489656 ,H489729 ,H489840 ,H489950 ,H48A064 ,H48A214
Recalling Firm/
Varian Medical Systems, Palo Alto, CA
911 Hansen Way
Palo Alto, California 94304-1028
For Additional Information Contact Michael Pignataro
Manufacturer Reason
for Recall
Possible incorrect dose calculation: An anomaly in certain Eclipse versions where, under certain specific sequences of user actions, the dose distribution may fail to be re-calculated even though changes have been made which affect dose.
FDA Determined
Cause 2
DESIGN: Software Design
Action Varian Medical Systems Urgent Medical Device Correction letters were sent to all affected users on August 6, 2010. Letters were also sent to Varian Sales, Marketing and Service organizations to inform them of the recall.
Quantity in Commerce 7019
Distribution Worldwide Distribution.
Total Product Life Cycle TPLC Device Report

1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = MUJ and Original Applicant = VARIAN MEDICAL SYSTEMS, INC.