Class 2 Device Recall Penner Patient Transfer/Lift System |
|
Date Initiated by Firm |
January 19, 2009 |
Date Posted |
April 08, 2011 |
Recall Status1 |
Terminated 3 on April 08, 2011 |
Recall Number |
Z-1911-2011 |
Recall Event ID |
56977 |
Product Classification |
Lift, patient, ac-powered - Product Code FNG
|
Product |
Penner Manufacturing Patient Transfer/Lift System, Cascade Transfer, Models 380000-1 (white) and 380000-2 (gray). The firm name on the label is Penner Manufacturing Inc., Aurora, NE.
For use with bathing systems in nursing homes, hospitals, and assisted living facilities to transfer or lift patients under the direct supervision of trained staff. |
Code Information |
Model 380000-1 - Serial numbers 08020346201, 8020346203, 10020360901, 10020360902, 10020360903, and 08020346202; Model 380000-2 - Serial numbers 08020346201, 08020346203, and 11020367501. |
Recalling Firm/ Manufacturer |
Penner Mfg Inc 102 Grant St Ste A Aurora NE 68818-3200
|
For Additional Information Contact |
Dale M. Wall 402-694-5003
|
Manufacturer Reason for Recall |
Incomplete weld on the pillar mounting bracket
|
FDA Determined Cause 2 |
Employee error |
Action |
The firm, Penner, sent an undated letter flagged "Important Safety Inspection: Action Required" via certified mail on January 19, 2009 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect the Penner Patient Transfer for any missing Welds of the Pillar Mounting Bracket; inspect the areas indicated in the enclosed photos to confirm the bracket is welded on all edges to the main frame of the base; complete the inspection, and complete and return the attached WELD INSPECTION RESULTS OF PENNER PATIENT TRANSFER form via fax to 402 694-5844 or mail to Penner Manufacturing Inc, 102 Grant Street, Aurora, NE 68818, Attn: Quality Assurance Manager. Penner stated in the letter, if welds are not complete (no weld), discontinue use and contact the Quality Assurance Manager at Penner Manufacturing immediately at 402-694-5003.
If you have any questions, contact the Quality Assurance Manager at 402-694-5003. |
Quantity in Commerce |
Model 380000-1 - 6 lifts; Model 380000-2 - 3 lifts |
Distribution |
Worldwide distribution: USA and countries including: Canada and Mexico. |
Total Product Life Cycle |
TPLC Device Report
|
|