| | Class 1 Recall B. Braun Outlook |  |
| Date Posted |
January 10, 2011 |
| Recall Number |
Z-0837-2011 |
| Product |
Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets |
| Code Information |
Model No. 621-400ES. Premarket Notifications K011975 and K994375. Device Listings D051037 and D051592.
|
Recalling Firm/ Manufacturer |
B Braun Medical, Inc 1601 Wallace Dr Ste 150 Carrollton, Texas 75006-6690 |
Reason for Recall |
Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.
|
| Action |
B. Braun Medical Inc. notified consignee by letter on August 24, 2010. |
| Quantity in Commerce |
2,107 Pumps (582 Pumps to Winchester) |
| Distribution |
Virginia and New York |
| |
|
|
|
|