• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Xeleris Processing and Review Workstation 1.0, 1.1, 2.0, 3.0

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
 Class 2 Recall
Xeleris Processing and Review Workstation 1.0, 1.1, 2.0, 3.0
see related information
Date Posted February 15, 2011
Recall Status1 Terminated on October 16, 2012
Recall Number Z-1250-2011
Recall Event ID 57159
Premarket Notification
510(K) Numbers
K024137  K051673  K093982 
Product Classification System, Image Processing, Radiological - Product Code LLZ
Product Xeleris Processing and Review Workstation 1.0, 1.1, 2.0, 3.0 The display, processing, archiving, and communication of data acquired by Emission Tomography cameras used in diagnostic radiology, including procedures for planar imaging, whole body imaging, tomographic (SPECT) imaging, positron imaging by coincidence, attenuation correction, and anatomical image registration. 3X The system is intended for use by Nuclear Medicine (NM) or Radiology practitioners and referring physicians for display, processing , archiving, printing, reporting and networking of NM data including planar scans (Static, Whole Body, Dynamic, Multi-Gated) and tomnographic scans (SPECT, Gated SPECT, dedicated PET or Camera-Based-PET) acquired by gammna cameras or PET scanners. The system can run on dedicated workstation or in a server-client configuration. The NM or PET data can be coupled with registered and/or fused CT or MR scans and with physiological signals, in order to: depict, localize, and/or quantify the distribution of radionuclide tracers and anatomical structures in scanned body tissue for clinical diagnostic purposes..
Code Information Serial Number 0000CZC5481313 0000FRB43605FW 000000203275-2 0000FRB4330NQ5 0000CZC5361JTH 0000000000UNKN 000000109754-6 0000CZC91280CY 0000CZC8396CD6 0000FRB4190GXV 00000000203292 0000CZC535013W 0000CZC5230HCC 00000108012-14 0000CZC5471LK7 0000CZC7105FJB 0000FRB32609KZ 0000FRB3250CZ3 0000FRB32507GH 0000000L3MMLZX 0000006FD249C7 0000CZC5480KT5 00000FRB3100ZK 0000CZC530033W 0000FRB3270C7R 0000CZC6502HMZ 000000108907 2 0000FRB4370B22 000000207637-1 0000003A2D3897 0000CZC8273ZDX 000000107833-4 0000FRB31602X3 0000FRB4370B48 000000107772-4 0000FRB3270GW4 0000CZC8396CDB 00000109315-10 0000CZC9502HM4 0000FRB4330NNZ 0000CZC724055V 0000CZC548135X 0000CZC5341NX7 0000FRB4370B35 0000FRB4330B4L 000000109291-8 00000108793-19 0000FRB42508WM 0000FRB42508SS 00000000289EL9 0000CZC5341NGK 0000CZC530036R 000000109841-9 0000CZC912751P 0000FRB42508RB 0000FRB44103LZ 0000CZC7205DP8 0000CZC7110GSQ 0000CZC5502N9N 0000CZC7240519 0000FRB42508W2 0000CZC6502HM9 0000CZC9127S8B 0000FRB43705M4 000000286147BC 0000FRB43605GQ 0000FRB40906XP 000000099YGCCT 0000FRB4330B44 0000FRB4330B3R 0000FRB4330BBB 0000FRB4330B7Z 0000FRB4330B3V 0000FRB4330B5Q 0000FRB4330B5C 0000FRB43402XW 0000FRB4210CC7 0000FRB3280C7C 0000FRB43402V8 0000CZC7212F10 00000109460-19 00000CZC8273BM 0000CZC8225781 0000CZC5341P27 0000CZC8273P55 0000FRB3270BG9 0000FRB4330B4G 0000FRB40907L0 00000108976-17 00000108976-48 00000108976-45 00000108976-49 0000FRB4330B3P 0000FRB3260B4Z 000000110436-7 000000212660-7 000000110436-3 000000212660-8 000000212660-1 000000212660-5 00000108118-26 0000CZC6502HJ0 0000CZC7205DPW 00000000107285 0000FRB325098N 0000FRB4370CYS 000000108663-4 000000108663-1 00000000108663 0000FRB3280C71 000000108663-3 0000FRB44103M4 0000FRB44106DN 0000FRB4330NPS 0000CZC8283TYN 000000109983-7 0000CZC6505DQ6 000000205262-3 0000FRB32808BQ 0000CZC8161XJJ 0000FRB3260GYP 000000108597-3 000000108587-2 000000108907 5 0000CZC5481343 00000108012-32 00000000000002 0000CZC5361JQ8 0000FRB4370CZ3 0000CZC7204X6S 0000CZC5471PLP 0000CZC5361JW6 000000206690-0 00000000057340 0000CZC548130M 0000FRB3270C94 0000FRB316014L 0000FRB44103M3 0000FRB44103M3 00000000284EL0 0000KY32061200 00000015418207 0000007438A88D 0000FRB31602Y2 00000032F52A8B 0000CZC552071Z 0000FRB31600YK 000000107912-8 00000108012-16 0000FRB3150L87 0000CZC5341N9H 00000108294-37 00000000057330 0000005DA56D9C 00000000211620 00000CZC8283V4 000000302776A1 0000CZC9127548 000000212425-1 0000FRB3160133 000000204174-7 000000099YGCNC 000000407787B0 0000FRB3250962 0000CZC7105FLM 000000108968-3 0000000L3RX455 0000FRB4380P1Q 0000CZC9127SD7 0000CZC9127SD0 0000CZC7105FCK 00000109541-12 0000CZC530033V 0000CZC8224ZKT 0000CZC8273NS2 0000CZC5341NTB 0000FRB3270C9Z 000KY201801154 0000KY36060429 0000FRB3260GVR 0000CZC5341NJP 000000108298-9 0000FRB42508S2 0000FRB42508VV 00000005597F4B 0000FRB3260H22 0000CZC5230HD3 0000FRB32808C4 0000CZC7110GPZ 000SCZC530034L 000000043025EU 00000000206509 000000206509-4 0000CZC912755G 0000CZC5361JQF 0000CZC530039F 0000FRB3260B11 0000CZC530036P 000000106574-0 0000FRB43708SF 0000FRB4330B5S 0000FRB4370B1R 0000CZC9127SCN 0000206846-0-1 0000FRB3260H0F 0000FRB3250CC3 0000FRB4400HZX 0000CZC5361JTB 0000CZC5341NQY 0000CZC6505DPC 0000CZC6505DPC 00000063053432 00000063053432 000000206748-0 0000FRB32604YW 0000CZC94154RT 0000FRB316031G 000000772269FA 00000006091D91 0000CZC5480KYC 0000FRB3280C6Y 000000108452-9 0000CZC8273NPP 0000FRB42808T8 0000CZC5341NHD 00000108187-37 0000CZC7212F0L 0000FRB3270GVW 0000FRB3270GVW 0000CZC6505DNT 000000108294-6 0000CZC724051Y 0000000L306R59 0000CZC5341NRV 0000CZC5341NQL 0000CZC7105FNH 00000208928-18 0000005A96AB39 0000005B6801E7 0000CZC7212F32 0000CZC5361JQ0 0000FRB40907KO 0000FRB40907KO 0000CZC5480KZF 0000CZC5230HG7 000000201ED327 0000CZC54812YK 0000CZC6506HK9 0000FRB3250C7Q 0000CZC7105FM6 0000CZC7105FM8 0000CZC5341LJV 0000FRB42508TF 0000CZC5361JT0 0000CZC6505DPG 0000CZC6502HG1 0000001DD1132E 0000CZC83942HZ 000000109935-2 0000CZC724054T 00000108968-16 0000CZC7205DP2 0000000L3N2108 0000000L3N2150 0000FRB42508SR 0000FRB44103F6 0000FRB44103F5 0000FRB41308WP 0000FRB41308W3 00000000206133 0000CZC5481319 0000FRB31600ZL 0000FRB31602WH 0000CZC5341NRT 0000CZC724055X 0000FRB3260B85 0000CZC5300354 000000108803-1 0000CZC7105FGC 0000FRB4230Q45 0000FRB4400J0M 0000FRB4210CBB 0000FRB3260GVY 0000CZC7105FMP 0000FRB32408FP 0000FRB3260B75 0000CZC724054Q 00000065D57EA1 0000FRB419080Z 0000CZC941591Z 0000CZC9127SLF 00000069AC7B52 000000108047-9 0000003CF3A719 0000CZC7212F00 000CZCC9414HKS 00000212550-17 000000108881-8 000000108881-1 000000108109-5 0000CZC7105FNC 000000108932-7 000000099YGCHB 0000FRB3270GVJ 0000FRB4330BB3 0000CZC724057M 0000CZC5361JR7 0000001BAE0739 0000FRB43607G8 0000FRB43607H1 000000099YGCMN 0000CZC7212DVT 000000108881-9 0000CZC5480KVV 000000600EC011 0000000I3N2161 0000FRB4370D9N 000000109935-6 0000CZC7105FKT 000000109266-4 0000FRB4330B86 0000FRB44106DW 0000FRB3280C77 00000000207788 0000FRB4330B7X 0000FRB4330B8J 00000000108106 00000108109-11 0000FRB4330B8D 000000099HXVYD 000000099HXVYT 00000108012-45 0000FRB4270J3M 000000108012-2 000000108106-4 00000108137-23 0000CZC7212DVQ 000000674E6DA4 000CZC5484130S 0000CZC8273ZDS 0000CZC7105FJ1 0000FR31125601 0000FRB4380JXX 0000CZC530038W 0000FRB4330BB5 00000000108294 0000CZC7212DW1 0000FRB31602WG 0000FRB4380C2L 0000CZC9127SN4 0000FRB4370B3N 00000002126602 00000212660-12 000BRB4330B86L 0000CZC8396CD2 0000CZC902437S 0000CZC9127S8T 0000CZC9127S90 0000CZC5361JTP 0000CZC5341NSS 0000FRB32609ZT 0000CZC8224ZKF 0000FRB4330B3L 0000CZC5341P28 00000000107912 0000CZC94158SG 0000FRB43402VD 0000CZC7105FMN 0000CZC7205DMG 000000107896-9 000000108480-4 00000108480-24 0000CZC7105FD0 00000010975416 0000CZC8396C9K 00000001097545 0000CZC5480KXG 0000FRB41308W1 0000CZC7212F2T 0000CZC7212F3M 00000000107745 0000CZC5341NHT 0000CZC724058B 0000CZC7212DZQ 0000CZC548133M 0000CZC7205DK0 0000CZC5230HH3 0000FRB32408FJ 0000FRB4370DBZ 0000CZC54812Z0 0000FRB43402VT 00000067FD0A76 0000CZC8273ZC5 0000CZC8273NM0 000000110229-7 000000110057-2 0000CZC6502HGK 0000FRB4370923 0000CZC535011Z 0000CZC5341N9S 0000CZC7212F31 0000CZC6505DRV 0000CZC7212DSN 0000CZC7212F3B 0000CZC5341NJY 0000FRB4370B1M 000000109403-1 000000109348-4 0000CZC5341NY5 0000FRB40907L2 0000CZC8396CGR 0000FRB3250CBN 0000FRB3150L75 0000CZC8224ZLS 108106-FRB4370 0000CZC548134W 0000CZC724054S 0000005D7B91AD 0000CZC5230HH1 0000000L3MMLZB 0000000L3MMLYW 0000CZC7111TPJ 0000004F9B8566 0000FRB4230Q4K 0000CZC7111TNP 0000CZC6505DP7 0000FRB4370D9L 0000FRB4330B96 0000000F984862 0000FRB42808SW 0000FRB4330BB8 000000099YGCDD 000000099YGCLN 0000CZC5341NXD 000000108875-4 0000FRB3150L73 0000CZC5341NHL 0000CZC7111TPW 00000000371EL5 0000FRB32509CH 000000204818-3 000000207802-4 0000CZC65012YD 0000FRB43705M8 0000FRB32604YQ 0000CZC7212DV8 0000CZC5361JSH 000000203605-4 000000203275-2 0000CZC5341P3P 0000002E4EFEFD 0000CZC548135Q 0000FRB4270JLJ 0000CZC7212DZY 00000064517815 0000CZC9127556 0000CZC912755P 0000CZC9128091 0000CZC91280DG 0000CZC912759G 0000FRB4370B3L 00000108137-28 0000FRB3260H0Q 0000CZC5341NGP 0000CZC91280DJ 00000210069 10 0000CZC7211D3B 0000CZC7211D5N 0000CZC8161XMG 00000000210127 0000CZC7110GRS 0000CZC7105FNG 0000CZC8283V60 0000CZC5341NRX 0000CZC5361JWV 0000CZC5341P1M 0000CZC5361JTS 0000CZC5341NH1 0000FRB32507DP 0000FRB3270GHV 0000CZC5492KDS 0000CZC7205DMB 00000FRB4210BD 0000FRB3150L7N 0000CZC5230HGW 0000CZC5341P0N 0000CZC8273P4C 0000FRB42508QN 0000FRB43402VK 0000CZC6501QX1 00000109460-18 0000FRB4160HT8 000000205880-2 00000000205880 0000FRB32509B3 0000FRB3280C7G 0000CZC724054N 0000FRB3270GGR 0000FRB3270C96 0000CZC9127S8D 0000CZC91280B5 0000CZC5341NQX 0000CZC5341NRF 0000003C8O13D7 0000CZC548130K 0000CZC54812ZR 0000CZC8273NM3 0000FRB4370929 0000CZC94158RP 0000FRB44106DV 0000CZC54812YY 0000FRB32509B1 00000000204275 0000FRB43402X9 0000CZC5480KV6 000206672-0-14 0000CZC5341NNP 0000CZC530035L 0000CZC5341NY1 000000109229-2 0000FRB3270GVH 0000CZC54812ZN 0000CZC6502HG4 0000FRB31600YR 0000CZC7212F37 0000FRB31602VK 0000000L3RX485 0000CZC54812Y7 0000FR31017756 0000FRB4210CBP 000SB0C80503X2 000000I32N2187 0000FRB4380P19 00000212550-10 0000CZC94158TY 0000CZC94158RF 0000FRB44103L4 0000CZC7105FDP 0000007EBB3900 000000109541-2 0000FRB325098B 0000CZC5341NJ8 0000CZC5361JTH 0000CZC94154TW 0000CZC94154WS 0000CZC9415910 0000FRB42508V8 0000FRB43607G6 0000CZC7205DPY 00000109541-29 0000FRB31600Z5 000000099YGCHF 0000FRB4230LHC 0000FRB43706C8 0000CZC8273Z9S 0000FRB3260GYG 0000CZC54812YV 0000CZC54812Y5 0000CZC8283V0J 0000FRB4370D9M 0000CZC7205DLS 00000108813-19 0000CZC7205DKD 0000FRB3280B8K 0000CZC7212DTZ 0000CZC8273ZGB 0000FRB3280BH3 0000FRB3270C7Q 0000000ABD6D67 00000010D2A88B 00000021D11683 0000CZC7204X5B 0000CZC7205DLW 0000CZC7212F2Z 0000CZC7240551 0000FRB3260H1W 0000000B4F1C5F 0000CZC5341P4P 0000CZC5361JX1 00000110324--2 0000CZC941591Q 0000FRB3150L6F 0000CZC5471Y49 0000CZC912759K 0000CZC8396CDW 0000CZC5230HHF 0000CZC7105FMM 0000CZC5361J2J 0000FRB316014N 0000FRB32507GQ 0000CZC55207L1 0000CZC7212F25 0000CZC7240515 0000CZC822575W 0000CZC822576B 0000CZC8273ZB1 0000CZC6505DNJ 0000CZC8273P6D 0000CZC55207KY 0000CZC6502HJZ 0000CZC7105FRR 0000CZC8273NNJ 0000CZC5230HH5 0000CZC5230HGT 000SCZC5300345 0000CZC530034V 000SCZC530033X 0000CZC5300350 0000CZC6502HMX 0000CZC7105FRN 0000CZC6502HG6 0000CZC6502HL7 0000CZC8396C9W 0000CZC55207L2 0000CZC6502KZ2 0000CZC5520712 0000CZC7204X7X 0000CZC8273NR8 0000CZC7240581 0000CZC5361JVG 0000CZC5502N9S 0000CZC6502HM5 0000FRB4380JY3 0000CZC6502KZB 0000CZC6502HHJ 0000FRB31600ZV 0000CZC822577B 0000CZC7205DJN 0000FRB32509B2 0000FRB31602VS 0000FRB437091F 0000107063-046 0000FRB4330NQB 0000FRB4330NQ8 0000FRB42508SR 0000KY2D090233 0000CZC5480KTC 0000FRB43607GH 000000109861-2 000000108294-1 0000CZC54812XC 0000CZC54812XC 0000CZC5480KTC 0000CZC5341P52 0000CZC5230HHN 0000CZC5361J27 0000CZC530034C 0000CZC5341NVX 0000CZC5341NST 0000FRB43402X0 0000CZC530039C 0000CZC7212F07 000000109878-5 00000033A3AEC7 0000CZC5341NFS 0000FRB325096G 0000FRB3270CBJ 0000FRB42508TC 0000FRB3270C98 0000CZC6505DQP 0000FRB3150L7L 000000107859-1 0000006CE5094B 0000FRB43704B7 0000FRB4330NQ5 0000CZC5341NF7 0000FRB4370DBL 0000CZC9127SCX 0000FRB3260GYH 0000CZC530034H 0000CZC7105FL3 0000CZC6502HJ5 00000000298EL0 0000CZC530039P 0000FRB42808SZ 000000099YGCDL 0000001C2292BC 0000CZC7240542 0000CZC7240527 0000CZC5341NHR 0000000L306T00 0000CZC5180W4Y 0000CZC5341NK5 0000003A345E3B 0000CZC912754H 0000CZC912757J 0000CZC9127579 0000CZC9127SM8 0000CZC91280CL 0000FRB44109X3 0000CZC5481315 0000FRB3250C9T 0000CZC822573V 0000CZC6502HG3 0000CZC7110GNJ 00000242221BCD 00001008012-28 00000000008480 0000CZC912757Q 0000FRB3260HOP 000000108968-9 0000CZC7110GQV 0000CZC7204X80 0000CZC5341NDL 0000CZC94158NC 0000CZC7110GRG 0000CZC5361J20 0000CZC5361JSC 0000FRB4270JC8 0000FR31017761 0000CZC5341N9K 0000CZC7205DNB 0000CZC530034W 000SCZC5300355 000SCZC530034M 000SCZC530035D 000SCZC530035D 000SCZC5300352 0000CZC5480KXL 0000CZC7240533 000000107285-0 0000CZC7212DVC 000000015B7A0E 00000000000008 0000FRB316014R 00000014DB2D67 0000CZC5341P21 0000FRB4330B9C 0000CZC5341P41 0000CZC5341NK3 0000CZC535011M 0000000MEEMORY 0000CZC8273ZCH 0000FRB3260GX4 0000FRB4330B8W 000000099YGCKT 0000FRB3150L7X 0000CZC5480KSF 0000FRB325055V 0000FRB32507DL 0000CZC8396CBY 000000109027-7 0000CZC7212F34 0000CZC8161XJV 0000FRB4370928 0000CZC5300373 0000CZC530034K 0000FRB4090KVJ 0000FRB3270GWF 0000FRB41308WW 0000CZC5230HJM 0000CZC5480KYZ 0000CZC9127S82 0000FRB42508TZ 00000108651-21 0000FRB4330B2D 0000FRB316013D 00000009D603E8 0000CZC7212F0D 0000CZC8273NLS 0000CZC8273NMV 0000000L3N2175 000000108976-1 0000FRB3270C87 0000FRB42508QZ 000000013P2181 0000CZC7105FC9 0000CZC83942JR 0000FRB325095W 0000CZC5230HGZ 0000CZC822577R 0000CZC7105FRD 0000CZC7105FPC 0000CZC5341P49 0000CZC8396CD9 000000099NFAYN 0000FRB44106D4 0000000L3MMLYV 00000000L3MMMK 0000FRB3160135 000000CZC530HD 0000CZC7205DHW 0000CZC530035T 008918UAA37111 0000CZC8283V2B 0000CZC8224ZKY 000000108972-2 000000108976-8 0000004BB51D24 00000034E0ABD6 000000108976-4 00000000108968 0000CZC7211D5L 0000CZC8396CFZ 0000CZC5361J32 0000CZC7204X72 0000108862-14H 0000CZC7212F2X 0000CZC7110GPR 0000CZC7204X2J 0000CZC5341P21 0000FRB4330B36 0000FRB4330B8P 0000FRB31602YT 0000CZC7204X4G 0000CZC7212F05 0000CZC7212F35 0000CZC7204X7N 0000CZC7204X2N 0000CZC7212F0T 0000CZC7105FDH 0000CZC8273NS0 0000CZC8273P46 0000CZC8224ZM2 0000CZC8273NSO 0000CZC8273Z9B 0000FRB4370B37 0000CZC5230HJJ 0000CZC5230HC8 0000FRB4330B26 00000212704-14 0000CZC7105FD9 0000CZC724055T 00000108976-30 0000CZC8225752 0000CZC7212DYN 0000CZC7212F27 0000CZC7105FKQ 0000CZC7105FNO 0000004A2A26CF 0000CZC535012H 0000CZC724057T 0000CZC8273P57 0000CZC8224ZL6 0000CZC8224ZLN 0000CZC8273P4R 0000CZC8273ZFN 00000110044-19 0000CZC5480KY4 0000CZC724054V 00CZC710505FL9 0000CZC7105FFP 0000FBR32409GB 0000CZC6505DQ9 0000000L3N2109 0000CZC6501QXG 0000FRB44106DJ 000000109451 2 0000CZC9024384 0000CZC8273NMJ 0000CZC8224ZLT 0000CZC7211D51 000000099YGCCL 0000FRB3280B9K 0000CZC6505DS0 0000CZC6505DPB 0000FRB44103KM 0000CZC9127SDB 0000FRB44109X0 0000CZC6505DR9 0000CZC8273P4Z 0000000L3N2190 0000FRB43704B8 00000109315-33 000000108881-5 0000CZC5480KSK 0000FRB44103MB 0000CZC8273NMY 0000CZC8224ZKB 0000CZC7105FP5 0000CZC7105FPR 0000CZC6505DRY 0000CZC5341NTN 0000CZC5300397 0000CZC7105FFK 00000108012-30 0000CZC8283V0B 0000CZC9127SLX 0000CZC7105FND 0000CZC7240588 0000005091DC7F 0000FRB42508V1 0000CZC8273P5G 0000CZC7212DZP 00000001306R69 0000CZC5361J1N 0000FRB32604YX 0000CZC7212DYV 0000CZC7110GQZ 0000CZC724053B 0000CZC530036Y 0000FRB42106V4 0000FRB4230Q44 0000CZC5361JQX 0000CZC5480KXY 00000108142-10 0000CZC548134V 0000CZC5480KYX 0000CZC54812Y8 0000CZC5480KY8 0000CZC548131M 0000CZC530038P 000000207833-4 0000FRB43605JZ 0000FRB4380BZP 0000FRB325055S 0000107063-029 00000108968-43 0000CZC6502HHC 0000CZC5341NRM 0000FRB42508VS 0000FRB4330B99 0000FRB3150L70 0000004D857C3D 000000107806-4 000000107806-3 0000CZC7212DXD 000000109315-1 000000107912-5 0000FRB43706C9 000000257DC85D 000000110024-6 000000208228-2 0000CZC548130N 0000CZC5481307 0000CZC8283V6W 0000CZC8273NNY 0000CZC8273P5S 0000CZC822573P 0000CZC7212F1V 00000003A0CAF0 0000CZC7212F38 0000CZC9127SJ5 0000CZC91280B3 0000FRB4370B3J 00000000A84904 0000CZC530037L 0000CZC7212F18 0000CZC9127SBB 0000FRB3280B93 0000CZC5341P2W 0000FRB32509CF 000000207069-5 0000CZC8273Z99 000000108532-7 0000FRB3160112 0000CZC9127SJ4 0000FRB42508SG 0000FRB3280893 0000CZC7105FDQ 0000FRB4330B39 0000CZC7110GPN 0000CXC724058D 0000005EB85548 0000CZC8273P54 0000CZC8273ZCL 0000FRB4370B25 0000FRB4370DGK 0000CZC5341NGL 000000099YGCHW 00000046877888 0000CZC5341NQJ 0000CZC5341NVT 0000CZC5230HCB 0000FRB4330B5G 0000KY33131392 0000FRB4400J0H 0000FRB4330B9K 0000CZC8224ZKL 0000CZC7212DWR 0000FRB3260B9Q 00000065919043 0000CZC54812YG 0000CZC7212F3W 0000FRB32609ZV 0000CZC8273Z94 0000FRB4370DGT 0000CZC5471CDK 0000CZC5310XPX 0000FRB3260GYL 0000FRB3270GH9 0000CZC5341P0P 0000CZC5230HCV 00000109460-14 0000CZC8396CCQ 0000CZC724055H 0000CZC724056S 000`CZC8225784 0000006E6E7BD7 0000FRB325055R 000000107155-3 0000FRB43708SK 0000CZC5341P1V 00000000203325 0000CZC5230HF2 0000CZC7105FHX 0000CZC5341NLC 0000001878455E 0000FRB42508RC 0000CZC530034R 0000CZC7105FGD 0000CZC7110GPV 0000CZC5341NN2 0000CZC55207KD 0000FRB44103LR 0000CZC8396CHQ 00000109878-17 0000CZC7105FCN 000000099YGCFA 0000FRB3260B18 0000CZC7240554 0000CZC724057B 0000CZC7105FMZ 0000FRB3260B83 000000099YGCLC 0000FRB42508WN 0000CZC5341NRY 0000FRB3250CZS 0000CZC5341NMF 0000CZC7205DHH 0000CZC5341NQ1 000000109519-5 00000000108225 0000FRB44106C5 000000212426-3 0000CZC9127SHS 0000CZC91280FH 0000CZC912808L 0000CZC9128088 0000FRB316012F 0000FRB3270GVT 0000CZC7105F17 0000CZC9127SC3 0000FRB42508TP 0000FRB4330B8L 0000FRB43704B9 0000FRB4270J3P 0000FRB325099S 0000CZC7211D37 0000CZC7212F2J 00000108674-14 0000CZC7105FL1 0000CZC5492KDN 0000FRB43607GN 00000047B9ACA0 0000CZC7105FF8 0000CZC7110GRJ 0000000H3800JC 0000FRB43605JR 0000CZC7212F1X 0000005FDAA163 0000CZC8396CFH 0000CZC5341NMN 0000FRB3250CZD 0000FRB32509BM 00000005597D5K 0000CZC8283V3K 0000CZC7105FJV 0000FRB4400HZZ 0000CZC5471Y4V 0000CZC54812ZH 0000CZC8273ZDP 0000CZC8283V30 0000CZC5481357 0000CZC5361JRS 0000FRB3280BH1 0000CZC5341NMC 000000109360-7 0000CZC7211D31 0000FRB3150L5W 0000FRB3150L5T 0000CZC5481362 0000CZC7211D3R 0000CZC7111TPG 00000000000003 0000CZC5480KW2 0000CZC5480KVN 0000FRB4330NQR 0000FRB43402XZ 00000108238-21 00000108238-26 0000CZC7212F0W 0000CZC6502HK1 0000FRB4330B5W 000000109408-5 0000000L3RX411 0000FRB43605J6 0000CZC8396CGB 0000CZC6502HHH 0000CZC7105FMD 0000CZC8225745 00000109471-10 0000CZC5341NN9 0000CZC5341P2V 0000FRB437021D 0000CZC6501QWY 00000000109572 0000CZC7204X2Y 00000000208869 000000206283-2 00000000108709 0000CZC5361JRQ 000000108294-5 0000CZC5361JV4 000000108319-5 00000108294-15 000000108319-1 0000CZC548132Z 0000CZC7105FMV 0000CZC535013W 0000CZC5361JRZ 0000CZC535013X 0000CZC5361JWN 0000FRB42508R1 0000CZC55207LW 0000CZC5480KXV 0000CZC912754R 00000000107623 0000CZC5230HG3 0000CZC5471PLX 0000CZC6505DPW 0000CZC7212DWL 0000CZC7105FKD 000000109020-1 0000FRB3260GY5 000000205556-1 000000205556-2 0000CZC5341NY6 0000CZC94154RW 000000204174-3 0000CZC7105FN3 0000FRB40907KX 0000CZC5341ND9 0000CZC535013G 0000001B5CEB16 0000CZC5341NCR 0000000L3MMMDP 0000CZC724054M 00000CZC72405M 0000CZC7212F2L 0000FRB4270J3X 000000208256-3 0000CZC54812ZK 0000001A99616C 0000CZC5480KXX 0000CZC5480KTP 0000CZC5480KYG 0000CZC5480KY5 0000CZC5361JRR 0000CZC6505DQL 0000FRB4380C35 0000CZC912758R 0000CZC9127583 0000CZC9127585 0000CZC912758N 0000CZC912757P 0000CZC912758B 000000109090-8 0000CZC7105FK6 0000CZC7212F3H 0000FRB3260H05 00000000106885 000000204258-5 0000CZC548135B 000000738E8044 0000CZC91280D1 000000207215-0 0000FRB3260B0W 0000CZC5361JSQ 0000CZC5361JXF 0000CZC5471PMD 0000FRB44106F3 00000000107833 00000108561-14 00000000110068 0000FRB31602Z3 0000CZC7211D53 000000109020-9 0000CZC9127SH2 0000FRB3260H2N 0000CZC724056K 0000CZC8224ZKK 000000110229-3 0000CZC822579G 0000CZC8273ZD9 0000CZC8273ZGD 0000CZC65012YX 000000108561-7 0000CZC912808Y 00000070AC0A66 0000007F73629E 0000107063-035 0000001636E67F 0000CZC5341NQD 00000000107859 00000000208825 0000CZC5341NXB 0000CZC5341NBP 0000CZC7111TP5 00000109541-16 0000FRB31600ZM 0000CZC5341NGW 0000FRB4370DH6 0000CZC5341NZ8 0000CZC535013F 0000CZC5480KXN 0000FRB32408G1 0000FRB32602GR 000000107696-4 0000007E0306E1 00000CZ5230HJ5 00000000A85141 0000CZC8273ZFH 0000CZC8273ZDJ 0000FRB328087K 00000109266-14 0000CZC5361JST 0000FRB40907L1 0000FRB43402Y9 0000CZC5471PM3 0000CZC724056Q 0000CZC7205DK2 0000000EDC4FA4 000000110044-8 0000CZC6501QX3 00000000108518 0000CZC7110GQ3 0000CZC724057W 0000CZC5341NGN 0000CZC5361JWZ 0000FRB437091Q 0000FRB4230LHM 000000106679-0 0000FRB437091C 0000FRB4370DB7 0000FRB42508WJ 0000FRB43607HK 0000CZY7212DTX 0000CZC5341P0S 000000013MN261 000000013MN261 0000CZC530034S 0000CZC530034S 00000042753165 0000FRB31602WM 0000FRB42508VH 0000000000B333 0000FRB32808BW 0000FRB328087S 0000000E079F40 0000CZC5471Y5Q 000000108012-6 000000109047-9 00000005708BF0 0000007381DC09 0000FRB3260BC3 00000000109990 000000110024-9 0000000151942D 00000047C6FC28 000000109417-4 0000CZC6502HJJ 00000014D971CE 0000FRB3260GW2 0000CZC9127SHC 0000FRB4330BB7 0000CZC5480KX6 0000CZC5480KYL 000000240F5BCC 0000CZC5341NWV 0000CZC7212DSJ 0000FRB42508TH 0000000099GCDT 0000CZC9127SLR 000000109700-2 00000000109700 000000109700-1 0000FRB3260GY6 0000FRB3250890 000000110077-2 0000FRB437092S 000000000000SS 0000FRB3260H3L 000000108406-2 0000000000C678 000000108480-3 0000CZC5341NB6 0000006B5FE154 0000FRB43704BL 0000002452179E 0000FRB4370921 0000FRB44109WZ 0000FRB3260C7H 0000FRB44103KQ 0000FRB44103KC 0000FRB40907J8 0000FRB3270C7W 0000CZC5480KYD 0000FRB40907JW 0000CZC7211D32 0000FRB32808B3 0000CZC54812WS 0000003353E737 0000CZC8273NLK 0000FRB43708TQ 0000FRB3280C6X 000000106666-0 00000000106666 000000106666-1 0000003463628A 000000109189-3 0000CZC548131B 0000FRB43605G2 0000FRB32808BZ 0000FRB44106D8 0000CZC5471CD9 0000CZC5481308 0000CZC5350136 000000L3N21111 0000000L3RX464 000000108307-5 0000FRB4330NQ9 0000FRB3260H31 0000CZC548135C 0000CZC548133N 0000CZC5471KPZ 0000CZC55207LS 000000108892 5 0000FRB3260B8B 0000CZC5481359 00000001088925 0000FRB3250CZ8 0000FRB40907LN 0000FRB40906XY 000000099YGCGD 0000FRB4370DBN 0000CZC5480KSB 000000208825-3 0000002FBB1D66 000000108932-2 00000054A548E7 0000005A9086D0 00000010147CBD 0000007290D98D 0000CZC5341NZD 000000109047-2 0000FRB44106D6 0000FRB3260H1G 0000FRB3260B5P 00000110044-17 000000207480-9 00000110068-10 000000204174-5 000000208256-6 0000CZC5361J0K 0000001F9E4A7D 0000000F961A62 000000109241-2 0000004B76247C 000000109983-5 0000FRB41308WQ 00000108118-10 0000FRB325088T 0000FRB3260GYD 000000034BFF85 0000CZC5230LR4 0000CZC7105FL8 000000109961-8 0000CZC8273NQQ 0000CZC6502HHY 0000FRB42508VK 0000FRB31600YX 0000CZC7212DX5 0000FRB32604YS 0000CZC7212DXJ 0000FRB3280C7M 0000FRB4330BBD 000000109419-1 000000109417-6 0000CZC7110GQD 0000CZC7204X50 0000FRB4230LHH 00000000210577 0000CZC7105FNX 0000CZC7105FQ7 000000107356-4 0000FRB41308VD 0000FRB4270JCN 000000211361-7 000000108862-8 0000FRB42508SV 0000FRB42508WD 0000FRB42508S9 00000005597F2G 0000FRB42508SQ 0000CZC5471Y4D 0000FRB3270GGY 0000CZC5361JTV 0000CZC5361JX4 0000003CF46E72 00000000209020 00000108626-13 0000CZC7205DLT 000000209020-3 0000FRB43607HP 000000106679-7 0000CZC530036L 0000CZC5361JT8 0000CZC8283V5D 000000208228-5 0000CZC5341NJT 000000013MMLYZ 00000110077-18 0000CZC912757N 0000CZC9127562 0000CZC5230HDR 0000CZC5230HDR 0000CZC5230HDR 0000CZC5230HDR 0000CZC8396CDJ 0000CZC6505DNY 0000CZC6505DPY 000000207069-0 000000590731A1 0000FRB42808SD 0000FRB4230Q40 0000FRB4330BBN 000000013N2160 000000013N2160 0000004392A5D5 0000004392A5D5 0000004392A5D5 0000000I3N2160 0000000I3N2160 000000110142-4 0000CZC6505DP8 0000CZC6505DP8 00000108372-17 00000108372-17 0000CZC5480KXC 0000CZC6505DQ3 00000108976-38 000000207215-7 00000108976-44 000000109149-9 0000FRB32507H0 000000099HXVXR 000000409FDCC4 0000PA3N250088 0000FRB3260H10 0000CZC5341P17 0000FRB32604YV 0000FRB3260H18 000000204820-3 000000204820-4 00000108881-14 0000CZC5230HF8 0000CZC5341NDL 000000207215-9 000000108881-7 0000001AA8E0F7 0000CZC5341NR7 0000CZC6502HNJ 000000109122 8 000000109122-8 0000000E5B640A 000000109961-2 0000FRB3260H2V 0000FRB3260GZG 0000FRB316014Q 0000FRB4330B4Z 0000FRB4330B9Q 0000CZC6502HMS 0000FRB437091X 0000FRB43605HK 0000FRB43708SS 0000FRB4230Q3L 0000FRB3260GXB 00000108796-10 00000108796-10 00000000205150 000SFR30915461 00000108881-12 0000FRB3260B6Y 0000KY33131365 0000CZC5341NP5 0000000I306R48 0000CZC7105FMQ 0000000CC58985 000000109231-2 0000CZC7212F42 00000109077-27 00000000085811 0000FRB4370B27 0000FRB437021C 0000FRB4380BZH 0000FRB44103LL 0000FRB4370B49 0000005C5BF7CF 0000CZC9414HX8 00000011043623 0000FRB3250C9D 0000FRB3260GVQ 00000043684D80 0000CZC5300385 000000337C7288 0000FRB44106BZ 0000FRB31602Z1 00000109624-14 000000211488F7 00000109624-12 0000CZC5341NHP 0000FRB43706C4 00000107696-19 0000FRB31602VQ 0000FRB32408FF 0000FRB316010K 000000648CD464 00000055DCD5C1 0000FRB4330BB4 0000004BCE4C00 0000CZC90243BZ 0000FRB44109XD 00000108187-21 00000108187-21 000000109861-9 0000KY33131339 000000204152-6 00000000286EL5 0000FRB3150L5M 0000CZC5341NPW 0000CZC5341NZM 0000FRB3270GWY 000000205788-2 000000205788-1 000000205788-0 00000000205788 0000FRB316011T 0000CZC7212DYB 00000000108651 00000033D488FA 00000000000010 0000CZC8283TZK 000000108793-5 0000CZC912751M 00000000108862 0000CZC5480KWW 0000FRB3150L6Q 0000002FED9B54 0000CZC7212F39 0000CZC6502HHP 0000CZC5341NZX 0000CZC54812Z4 000000099YGCLW 0000007B642D27 0000FRB43402X2 0000FRB4330B9Z 0000FRB4160HT2 0000FRB3250560 0000PA44275203 0000CZC5341NNH 0000CZC8396C9P 0000CZC535012J 00000FRB440J0G 0000FRB3270GVR 0000005EE43FCD 0000CZC5341NXM 0000FRB44103LN 0000FRB316014J 0000FRB3250C9L 0000FRB3270GVC 000000099YGCLK 00000000057329 0000FRB4370CYV 0000FRB4270JL1 0000FRB4370DBY 0000FRB3250CC7 0000FRB3270GH1 000000109348-9 000000107356-1 0000CZC5341NHZ 0000FRB43705M3 0000000I3MMMBT 000000108142-3 0000CZC535011W 0000004BD54F93 0000FRB4370B2X 0000CZC8225750 0000CZC5361JR6 000000108968-3 00000021B0BD35 00000054254BF0 0000FRB32609ZS 0000CZC5230HGF 0000FRB437092P 000000706BE887 0000FRB31602XZ 0000FRB31602X2 0000FRB4370B23 0000FRB44103MK 00000039800C88 0000FRB3250C7L 0000CZC5230HC3 00000000388EL9 0000CZC55207KW 00000014D4C9B4 0000006A12AD4E 0000FRB42508VW 0000CZC7204X84 00000000207712 0000FRB3260B76 0000FRB42508V6 00000063B35C7E 0000CZC5300347 000000107976-4 0000CZC5341P2Q 000000207480-3 0000FRB3250561 0000FRB42508VP 0000CZC822578M 0000CZC535012Z 000000109360-5 0000CZC8224ZLP 0000CZC535013S 0000FRB43402VR 0000FRB4370927 0000FRB4330B24 0000FRB43402V9 0000FRB32507GM 0000FRB32609ZM 0000CZC9414HKQ 0000FRB41308WD 00000108012-18 0000FRB4090KZY 000000108118-2 0000FRB43605HZ 0000FRB40907M3 0000CZC530036X 0000FRB328087J 0000FRB3280B87 000204152-2001 000FRB437006CF 00000072D847BA 000000220EDC80 0000CZC5341N9J 000000109241-1 0000CZC6505DQS 000000207811-9 0000CZC7240531 0000FRB4370290 0000CZC5471LK9 0000FRB32808B6 0000CZC724056G 00000000679090 00000072FCB142 000000204174-9 0000FRB4370DBX 0000CZC6502HGF 000000108452-7 0000CZC65012Y8 0000CZC65012YN 0000004DC03F7A 0000CZC9127SGT 0000CZC5341P36 00000000207821 0000006F631600 0000000H3700MP 0000CZC8396CBL 000000203301-3 000000110142-9 0000CZC9127SDD 000000110167-9 000000110185-6 000000110185-1 0000FRB32507DH 0000007CA025D2 0000000CD11A87 0000CZC8273Z8X 0000007C1B76C9 000000110199-3 0000CZC9127SBX 0000CZC7211D3N 0000CZC8396CBX 0000007B3229EF 000000109935-9 00000110436-18 0000CZC5341P1Q 0000CZC5341NR1 0000005413839C 0000FR30829789 0000FRB32507G9 00000000F59369 0000CZC5480KW1 0000CZC5480KYH 0000CZC5341P06 0000CZC5361JVB 000000110011-6 0000CZC5480KWV 0000CZC5480KXT 0000CZC5480KVJ 0000CZC548131Y 000000107356-5 00000000A00048 0000CZC8283V6X 0000CZC548135W 000000108708-2 0000CZC548131Z 0000CZC5341P1H 0000FRB43402W1 0000CZC724054K 000000109624-4 00000108372-32 0000CZC5341P33 0000CZC535012V 0000CZC5341NV2 0000CZC5341P43 0000CZC5341NS8 0000CZC8273ZGS 00000000109471 0000006084308C 00000CZC8283V7 000000041B30E0 0000CZC9127558 0000CZC9127SN0 0000003C27556C 000000633B8AD1 0000FRB3260B1T 0000FRB32509B8 0000FRB4330B1N 0000FRB4330B5Z 0000FRB4370D9V 0000CZC535012K 0000CZC5341NRP 00000000109734 000000108626-7 0000CZC5341NWN 0000CZC5361J0S 0000CZC5230HDD 0000FRB325096S 00000212660-11 0000FRB3260B59 0000FRB42508T7 0000CZC94158SY 00000000A90917 0000FRB4330B5R 0000FRB42808SN 0000CZC5341NSX 0000FRB4380P1W 0000CZC7204X4R 0000004399821C 0000CZC7110GP4 000000204819-4 0000FRB3270BGD 0000002F79E16E 0000CZC5361JSW 0000FRB4090KV9 000FRB43300BBF 0000CZC724055D 0000CZC5471PM8 0000006650A373 000000212550-2 000000110011-1 0000CZC9127SGZ 0000CZC912758J 0000CZC9127SJ0 0000CZC912755V 0000FRB42508WF 0000FRB4330B93 0000FRB43607HN 0000005C9ED3F9 00000CZC110GRP 00000109149-17 0000CZC7204X51 0000CZC6502HLG 0000FRB3260GX5 0000CZC7212DWT 0000000L3N2127 00000000207820 0000CZC5341NXX 0000CZC5341P4M 0000CZC5350140 0000FRB4370CZ7 0000CZC8224ZLC 0000FRB316010Q 0000CZC8396CGM 00000CZC927SMT 00000000110185 00000060FB98DF 000000171BBCC1 00000022EB0893 0000CZC912758H 0000CZC5492KDQ 0000CZC5341P0M 000000110024-3 0000CZC5361JRH 000000109552-3 00000109541-15 0000CZC8283V28 00000009E8DC8A 0000FRB4190816 00000056A69E79 00000028217BED 0000FRB4210CBR 000000099NFAZC 0000000000B311 0000CZC5471Y60 00000108294-46 0000CZC5471Y65 000000108298-2 0000CZC5471PLL 0000FRB3289B8Y 0000CZC5471LJN 0000CZC65012Y3 0000CZC6502HMD 000000108862-7 000000558CA2D6 0000001590616D 000000754BF048 000000110081-8 00000037964555 00000000212550 0000006FDDD4EE 000000212704-4 000000212704-3 0000000256FAF4 0000CZC6505DR8 0000FRB4330B1T 000000107356-3 0000FRB3260H29 0000FRB31602W2 0000CZC5471LJM 00000000057440 000000663668A6 0000CZC9127549 00000000110199 000000204819-1 0000FRB40907KG 00000029F493E9 000000099YGCCV 0000FRB43605GJ 000000099HXVXA 0000FRB3250D9C 0000FRB43402W7 0000FRB42508VQ 00000208229-10 0000CZC8225788 0000FRB3250CB7 0000FRB316011N 000000110167-5 0000CZC9127565 0000FRB328089C 0000CZC5341NNT 0000004EFA37E6 0000CZC7212DTL 0000000C2FB02E 0000CZC5300394 0000CZC5310XPR 0000FBR31600Z4 0000FRB32609L6 0000FRB3260B4R 000000424920F5 0000FRB43402W5 00000108118-17 000000109315-5 0000CZC6505DQ4 0000FBR3260GVQ 0000004811EC51 0000FRB3260GWM 0000FRB3260H0M 0000FRB4090L07 0000FRB4330B8Y 0000FRB42508W8 000000109020-8 0000FRB3270BG1 000000109307 8 00000010930713 00000010930715 00000001093078 0000FRB3260B4V 0000FRB3260GYR 000000109576-2 0000000EDA4B45 0000CZC5341NC7 0000FRB4330BBM 0000FRB316010V 0000FRB32602GY 0000FRB3150L5P 0000CZC724054P 0000CZC912756Z 0000000L3RX463 0000CZC548135D 0000FRB3260BCY 0000FRB4400J01 0000CZC5480KSW 0000CZC5361JT6 0000FRB4370B2R 0000FRB4370B2R 0000CZC5341NDF 0000CZC7211D57 EL00CZC8161XKS 0000000E024DDE 00000000079357 0000CZC7212F22 0000CZC6501QY8 0000CZC8273NS1 0000004C495792 0000FRB4330BB1 000000108758-4 0000CZC548130H 00000FRB4508WQ 0000FRB42508WQ 0000CZC530033Q 000000107912-1 00000CZC547Y4N 0000CZC5341NQS 0000000L3N2120 0000CZC7105FGK 0000CZC7105FLP 0000CZC7105FCR 0000FRB4370297 0000CZC54812ZD 0000FRB3250C9H 000000109386-5 00000000057464 0000CZC5361JT3 0000CZC5361JX2 0000CZC5341NW0 000000109047-4 000000109090-9 00000008856931 0000000L306R92 0000000E76206F 0000001F7F98C9 000000208869-2 000000107917-2 00000108773-16 0000CZC54812X0 0000003C47112A 0000000L3Y5YC7 000000043025EQ 00000CZC51804F 0000CZC5341NDK 0000000L306R57 0000CZC5180W47 0000FRB3260B1D 0000FRB42508WY 0000FRB42508QV 0000FRB42106V5 0000FRB3260BOT 0000FRB43402W4 0000FRB4380C05 0000CZC5341NNS 000000108709-4 0000CZC7212DVD 000000108626-8 107896412283VE 00000208228-15 0000000795CC09 0000CZC7205DP7 0000CZC7105FHM 00000028DB2353 00000108109-26 0000FRB3280879 0000000L3N2179 0000FRB3270GWM 0000206703-0-1 0000206703-0-4 000000206703-0 0000206703-0-3 000000061602/2 0000206703-0-2 0000005C88587E 000000099NFAZH 0000000L3N2132 0000FRB4380C21 0000CZC7212FOR 0000CZC8283V2W 0000007EB41F14 000000358A5135 0000003B342E64 00000000206727 00000037D43FDB 0000CZC7212F2Q 0000CZC5341NT4 000000108651-7 0000CZC8161XL6 0000FRB40907LT 0000CZC8273NMS 000000205228-5 0000CZC8273Z8W 000000108012-3 000000108012-4 0000FRB4090KZP 0000CZC5480KT9 000000109451 1 0000CZC9127525 000000106679-3 0000CZC8273NLH 0000CZC5480KVZ 0000FRB44106F4 000000210085-4 0000FRB43605GX 000FRB42308LH8 0000CZC7211D3W 00000000292EL3 0000CZC91275BY 00000036C12853 00000002414379 00000047C420A1 0000FRB4380C08 0000CZC7111TP0 0000000L3RX438 0000FRB3250CCM 0000CZC822573R 0000CZC5341NXN 000000099YGCKA 0000FRB4370B3P 0000CZC7212DXM 0000CZC5480KX9 0000CZC6502HN1 00000FRB4302WC 0000CZC724057P 0000FRB325088Q 0000004D6B51FE 0000FRB43605JY 000000110324-8 00000000107200 0000CZC5230HD4 00000000109122 0000CZC6502HM8 0000FRB4370DH7 0000FRB3280B8C 00000109352-14 0000FR31017757 0000FR31017759 0000CZC5230HGQ 000000099YGCDM 000000207907-2 0000CZC5341P46 0000CZC8396CFV 0000FRB325099Q 0000FRB4090KTZ 0000CZC5230HG0 0000CZC94154S0 00000000110436 0000CZC5230HJ4 0000CZC5180W4V 0000FRB425085D 000000099YGCLL 0000FRB44106DH 0000FRB3260GZ3 0000FRB31602Y7 0000FRB3260GZ4 0000CZC5341NJV 0000CZC7212F3X 0000FRB4370930 0000CZC5361JQS 0000001F767A3B 0000CZC7212F2N 00000000109052 0000CZC7110GPY 0000FRB4380C2C 0000FRB31600ZB 0000FRB42508R3 0000FRB4340B8M 0000FRB3260BCH 0000FRB44103F9 0000FRB41308VF 0000FRB328087B 0000FRB4230LHG 000000106882-4 000082120280YA 0000CZC8273ZCZ 0000007B8504AB 0000CZC5471Y4R 0000CZC5471PLB 0000CZC5520710 000000109471-9 0000CZC8225774 00000010ADC485 0000CZC5341NLX 00000046E28074 0000FRB3250D9P 0000FRB3250D9D 00000000010145 0000FRB3250CBK 0000FRB3270C9X 000000108892 7 0000CZC7205DMZ 0000CZC8283V6J 000000282D9DE4 0000CZC5471CCT 000000109754-7 000000109077-8 0000CZC6505DQ5 000000108143-1 000000109767-1 0000CZC6505DQJ 00000000208928 0000CZC7240539 0000CZC7205DPM 0000CZC5480KZ3 0000FRB42508SX 0000FRB42508QM 0000CZC5341NVQ 00000073F680B5 0000CZC8273ZB9 00000010C38006 0000FRB31600ZW 0000FRB3250CCJ 000000107458-4 0000CZC5361JW3 0000CZC5471Y5D 0000006E10BE7A 000000109189-2 000000108394-1 0000CZC5480KW0 00000019724E1E 00000000107832 000000108480-9 0000CZC5471CCX 0000CZC94154SW 0000FRB42508QS 0000FRB325055P 0000FRB4270JC4 000000109780-3 0000FRB4370B2C 000000109460-9 0000CZC7205DHR 00000108294-34 00000108294-45 0000FRB4370926 0000CZC7105FLZ 0000CZC8273P5N 0000CZC5230HCX 00000000212347 0000FRB437091L 0000CZC9415913 0000206035-1-1 0000FRB31602WB 0000FRB3260B4W 000000206283-1 0000CZC5341P2T 00000109460-15 0000CZC5341NR5 0000FRB31602XP 0000CZC5480KY9 0000FRB4380C1Z 0000FRB3250CZV 0000CZC8273ZGJ 0000FRB3270GVB 000000109519-3 0000CZC8273ZGR 00000109515-13 0000FRB42808T4 0000FRB43605HV 0000CZC7204X7Y 000000108532-1 000000108532-6 0000CZC7105FHR 00000108803-16 00000108803-26 00000108803-15 0000CZC7205DH5 00000068349A84 000000108932-6 00000108803-23 00000108803-19 0000005D3B31EA 00000108793-10 0000CZC902439R 000000108862-6 000000109451 1 000000109451 3 00000000204980 0000FRB4270JLF 000000109077-8 0000FRB328089W 00000000203740 0000004135BCEA 0000CZC724056D 000000109935-5 0000003791A878 000000110068-3 0000FRB3250C7G 0000FRB316011F 0000CZC8283V58 0000CZC9127SGJ 0000CZC8283V5B 0000FRB32602GV 00000CZC53501D 000000108968-4 0000FRB4370DH9 0000CZC5341P3N 0000FRB4370922 000000108213-8 0000CZC54812X4 0000CZC54812WT 00000049A4DE9B 00000000204731 0000FRB3260B6R 0000CZC535013K 0000002F0D62BB 00000109250-15 0000000B8DB49A 00000036AF7246 0000000012114B 0000FRB43607HB 0000FRB43708TM 000000109866-2 0000004B111212 000000099YGCDR 00000000368EL1 0000FRB4330B9T 0000FRB3260B15 0000FRB4380C23 0000CZC724055Q 0000CZC5502N9M 0000FRB43605GK 0000CZC7110GS5 0000CZC724052V 0000FRB4380C2D 0000CZC9127SB8 0000FRB42508SF 0000FRB32408FH 0000CZC9127SN2 000000210577-5 0000FRB4370933 0000FRB4330B83 0000FRB43605GR 000000108007-5 0000FRB43605JL 0000CZC530034D 0000CZC5361JX6 0000CZC9127S8H 0000CZC5341NDT 0000CZC5341NBK 0000CZC5341NSZ 00000000079329 000000107912-2 0000CZC5341NJL 000000109149-3 0000CZC8283V3Z 0000005CD5E4A4 0000FRB3280B8V 000000211361-2 0000CZC5341NB7 000000099YGCLV 000000108773-5 0000CZC5471CD7 000000107917-1 0000007D72D732 000000276D4F78 000000107896-2 0000CZC7205DJ3 0000CZC7105FN1 0000CZC912755C 0000FRB3260H3H 000000208228-1 0000FRB4210CBY 008227UAA20453 0000CZC8283V6H 00000108968-12 0000000B05BAC9 0000FRB43708SL 0000FRB3270C7Z 0000007A079E5E 000000108862-7 000000110068-9 000000210085-3 0000FRB3260H2C 0000CZC5341NJH 0000CZC5341NTP 0000CZC5341NQK 0000CZC5341NPZ 000000108748-7 0000CZC6502HLV 0000CZC822577Z 00000108626-10 0000CZC7204X78 0000FRB4370DWG 0000FRB4370DWG 00000109541-28 0000CZC9127SHJ 00000007305AD9 0000CZC548133S 00000051B6A12A 0000CZC55207LP 0000CZC7205DJX 000SCZC5230HCQ 0000CZC8273ZFL 008227UAA20454 00000005597F2F 0000000099GCKM 0000FRB3260B5J 0000FRB3250CBZ 00000329343809 0000CZC5341NTQ 00000000106679 000000204819-0 0000FRB42508T4 0000203299-4-2 0000FRB31602V2 0000FRB3160108 0000FRB43605HL 000000099YGCLH 0000FRB4330B3X 000000109767-9 0000CZC8224ZK9 0000CZC6502KZG 00000000107806 00000000107806 0000CZC5480KYK 0000000L306R75 0000CZC5341NXZ 00000108732-34 00000110324--9 0000FRB43605GS 000000109471-7 00000005EEBEEB 0000CZC6502HGN 000000108452-6 0000FRB328087M 00000016CFDB29 00000108452-23 000000108452-2 0000FRB43605G6 0000CZC91280DC 000000108225-6 0000CZC724057X 0000CZC5480KTM 0000FRB3280B85 0000FRB31602VM 0000FRB3260GY0 0000FRB32609ZH 0000FRB4090KTS 0000FRB4090KTR 0000FRB40907LJ 0000CZC5230HHB 0000CZC91280DK 0000CZC91280CS 0000FRB328089Q 000000210069 2 000000108628-3 0000CZC6502HK3 0000CZC7110GSX 00000110044-10 0000FRB42508W6 000000110036-9 0000CZC5230HFG 0000CZC5230HFG 0000CZC91280G0 0000CZC7205DKC 0000004CC4947B 0000CZC6505DP9 0000FRB42508RW 0000FRB4330B29 0000FRB3260B1Y 0000FRB42508SC 000000110332-2 0000002EAE6193 0000CZC912755K 0000FRB42508VY 0000FRB4430B3Y 000000107356-8 0000CZC8323DZJ 0000CZC7212DVY 0000CZC530038Y 0000000L306R70 000000208928-1 0000CZC5341NS4 0000005473DCB0 0000FRB4370CZ6 0000FRB43704BK 0000FRB41308WB 0000FRB4330B9G 0000FRB43402V4 0000PA44025311 0000FRB32408FQ 0000FRB42808SS 0000FRB43402V7 0000FRB316011D 0000FRB3250CBT 0000FRB4380C34 0000FRB43402VQ 0000FRB3280B8L 0000FRB3260GW1 0000FRB3250CBW 0000001857FA95 0000FRB32507GJ 0000CZC8273ZGF 0000CZC8273ZFY 0000CZC8273ZGP 0000000D1813A5 000000208928-8 000000109725-8 0000000000B270 0000FRB4370B22 0000FRB437091J 000000110044-2 000000204174-6 0000FRB40907KF 0000FRB43607G4 0000FRB44103L3 00000108976-10 0000CZC7211D3J 000000109361 4 00000000000006 00000004E13CC1 0000CZC8396CFM 0000FRB42508WC 0000FRB4210CBZ 0000FRB3250CBP 00000034969EB9 00000077AE9DE1 0000FRB31602YL 000000108976-7 00000108976-47 0000CZC5341NC6 00000108976-14 0000FRB3250CZC 0000FRB3270C8C 0000FR31017762 0000FRB3280C9S 0000FRB43302WX 0000FRB3270GGP 0000FRB3280C9H 00000000204706 0000FRB41308VX 0000FRB32507GV 0000FRB325096K 0000CZC5480KX0 0000CZC912759F 0000004C02B4FC 0000CZC7212DZL 0000003F071BE5 0000CZC5341NSN 0000FRB31602YV 00000109328-15 0000CZC7212DX6 0000FRB3260H16 0000FRB3270CBW 00001098885-21 0000FRB3270GWT 0000CZC5341NH4 000000392AAE74 00000036CB7076 0000FRB4380C3C 000000109348-6 0000FRB3270BGB 0000CZC6502HML 0000CZC6502HN4 0000CZC7240528 0000CZC6505DQB 0000CZC7240567 0000CZC7205DNK 0000CZC6505DQN 0000FRB3160144 0000FRB4090L0C 0000FRB42508W7 0000FRB3260GYX 0000FRB437092L 0000CZC5341NKX 000000066F5978 000000110081-2 0000CZC83942HK 0000CZC7205DP4 0000FRB42508WL 0000FRB3260B8N 0000CZC5350128 0000003F5AD27F 0000002DEZC3A1 0000CZC9127SDS 0000CZC6502HGV 0000CZC6502HN9 000000207804-2 000000207804-3 0000000052D603 0000CZC6502KZ0 00000000108532 0000CZC7105FF6 0000FRB42508R2 000000127687C2 0000CZC6502HKG 0000CZC6502HFY 0000CZC6502HH8 0000CZC5341P25 00000003DF61D2 000000109885-6 000000109885-2 0000CZC5341PV3 0000CZC5300396 0000FRB4370B3Z 00000000210085 0000CZC8396CGQ 000000109535-4 0000FRB437092T 0000CZC5431NQC 0000CZC5361JVM 0000CZC5361JVH 0000001AAABF39 0000004E63AE7F 0000CZC55207KC 0000003567C25E 00000059BF24AC 0000CZC8273ZFJ 000000109700-1 0000CZC8161XJ7 0000CZC5230LQS 00000057F8A17E 000000109360-3 00000006E72BB5 0000FRB43605H3 0000FRB42508QQ 000000109250-9 0000CZC8396CBG 00000210069 18 0000FRB3260B51 0000FRB3280C94 0000FRB42508SB 0000FRB43605JC 0000CZC5341NS6 00000048D14322 0000FRB3250564 0000FRB4330NP0 0000CZC5350120 00000109878-10 0000CZC5341P0R 0000FRB42508X2 0000FRB42508WT 0000FRB43708SR 000000109961-3 000000210577 6 0000CZC7212F1Q 0000FRB31602YZ 0000FRB31602XV 0000FRB31602TX 000000205262-1 000000205262-2 000000205262-8 00000109961-13 0000FRB43706BX 00000000207637 0000FRB3250CCL 0000FRB3280C95 000FRB32280C95 0000CZC5341NP3 0000CZC7212DZB 0000FRB3270C9Q 0000FRB4380C2V 0000FRB44109X5 0000CZC530036W 0000FRB31600YY 0000FRB31602XM 0000CZC7205DPR 0000FRB3260B4S 0000FRB32507H4 0000FRB32507H4 0000CZC9127SJR 0000CZC5341NQ6 000000106679-8 0000CZC5361JTZ 0000CZC7212DT4 000000109047-1 00000109283 10 00000000109250 0000CZC5230HC6 0000FRB3260H0R 0000FRB3260B7K 0000FRB3260H04 0000FRB3270C98 0000FRB3260GZS 0000FRB3250981 000000207840-3 0000FRB3260H0N 0000FRB3260H0C 0000CZC6501QX0 00000078AEC843 0000CZC530039N 0000CZC912757D 000000204275-0 0000002B8F8F52 0000FRB328089J 0000FRB3260BB7 0000FRB4370931 0000CZC5361JXG 0000CZC5361JWW 0000CZC7212F3K 0000CZC6505DNS 000000565F6026 000000099YGCGY 0000FRB42508X0 0000FRB43402WV 0000FRB42508W9 000000099YGCKD 000000099ALYMH 0000CZC9127SCY 00000000212660 000000207802-5 000000207802-1 0000FRB3280B90 00000005597F7F 000000099YGCML 0000CZC7205DM4 0000CZC8396CDR 000000208928-2 0000CZC7212DXP 0000CZC7105FDD 0000CZC5300382 0000CZC530036H 0000CZC5361JX5 0000CZC5471KPY 00000004546325 000000108748-3 00000109315-34 0000CZC724052C 0000CZC5361J01 0000CZC5361J2T 0000CZC5361JX3 0000CZC5361J2G 0000CZC5361JSR 0000CZC5361JQ5 0000CZC5341N8Z 0000CZC5361JSF 0000FRB328087N 0000FRB3160136 0000CZC8396CB0 0000CZC8396CGK 0000FRB3260B12 0000FRB3270C93 0000FRB3260H15 0000FRB40907M2 0000CZC8273ZGK 0000FRB42808SL 0000CZC65012XQ 0000FRB43706BZ 0000CZC534INSS 0000CZC6505DQV 0000CZC6502HH7 0000CZC65012YS 0000CZC7212F0S 000000109408-2 000000110011-8 00000110077-17 0000FRB3270C82 0000FRB42508TM 0000CZC8273NKY 0000CZC8161XK6 0000001D0A3080 0000FRB4370CZF 000000108793-9 0000CZC6502HZ3 00000000108532 000000109734 9 0000CZC8396CCH 0000CZC8283V0W 000000109734 7 0000CZC912759H 0000CZC7211D33 00000110081-13 0000FRB4330B3D 0000CZC9127596 000000321B5CF5 UNKH3700JA0000 00000109361 16 000000109261-6 0000CZC724050T 0000FRB325099D 0000CZC5502N9Y 00000208928-17 000000108298-3 0000CZC8283V24 0000FRB43402VM 0000CZC6501QWX 0000FRB4380C0Q 000000110137-5 0000CZC5230HCP 0000FRB44103KW 0000FRB3260B7D 0000FRB32509CJ 0000CZC5341NBS 0000CZC5502NB2 0000000AE4672B 000000108813-2 00000003BBF3B5 0000CZC7105FG1 00000045BE8A65 000000108862-9 00000108773-19 00000065F8D903 0000CZC7204X58 0000CZC8396CBJ 0000FRB42508V4 0000FRB4380C02 00000005624272 0000FRB3260B5B 00000109767-12 0000FRB4330B58 000000208256-9 000000205228-0 0000FRB42508QY 00000108319-10 0000FRB3260B7R 0000FRB31600Z6 0000FRB3160177 000000205228-1 000000110199-3 0000CZC912808B 0000CZC9128099 0000001E0134A5 00000072B86FC9 000000106844-1 00000000106844 000000106844-2 000000106844-3 00000000205302 00000000205228 000000205302-1 0000FRB3160142 000000208928-9 000000109451 7 0000000109841- 0000CZC8273NLL 0000007C07ED9F 0000CZC83942LD 0000FRB43704BF 000000003500ER 000000108118-7 000000109179 3 000000210069-5 0000CZC7204X7R 00000107696-17 00000000291EL5 0000FRB42508V7 0000FRB4330NQQ 00000108976-40 0000CZC535012B 00000000206510 0000FRB328088X 000000206509-1 000000099YGCKG 00000020AD0E76 000000109261-5 00000000040778 0000FRB31602Z9 0000FRB4210CBQ 0000FRB326092H 0000FRB4400J0C 0000FRB4400JOC 0000FRB3260BBB 000000193A8436 000000109841-6 0000CZC9127S8S 000000110081-4 000000109935-7 00000053F89721 0000FRB40906XC 0000FRB44106DZ 0000CZC530035H 00000031799573 0000CZC5341P2G 0000FRB4330NP6 0000FRB42508RT 0000CZC55207M2 0000CZC5502NB3 000000212814-2 0000CZC5341NPH 0000CZC7105FJH 0000FRB3260GYK 00000000109090 00000060EB9C3D 0000FRB4330NQK 00000000208896 0000FRB44106DD 000000108213-7 00000000109767 000000109767-3 0000000L3RX399 00000000109408 000000066F3CC2 00000108651-19 000000488E84A0 00000005597F5R 0000CZC5230HC5 000000108758-1 00000000006718 00000316938811 00000061A7C111 0000CZC94158RT 0000CZC724055L 0000FRB4330B57 000000108498-8 000018-03-2004 0000CZC55207LN 0000FRB43402W9 000000208228-5 0000005EABE4CA 0000FRB437092J 0000FRB43708SY 0000CZC548132S 0000006EFF10B4 0000FRB4190GXV 0000FRB40907LQ 0000FRB43706CB 0000FRB44103LK 0000FRB43705M2 000000109122 5 0000CZC7211D41 0000CZC7111TPV 0000000F462062 0000CZC6505DQG 000000109077-3 0000FRB31602TT 00000000203292 000000203377-3 000000207215-5 0000FRB3260B5G 00000CZC828V03 0000CZC724052D 0000CZC7110GPX 0000CZC7110GQB 0000CZC7110GQB 00000107696-12 0000CZC5341N8J 0000CZC7105FCH 0000CZC5480KY1 0000FRB3260B82 0000004C4AEBE3 00000042CAD9F6 000000108007-2 00000024332C16 0000000D52E2BE 00000000005597 0000FRB43605JB 0000FRB4330B8K 000000207918-2 00000000107976 00000107976-14 0000000D81C10E 00000037D3647E 000000207637-9 0000CZC5341NZT 00000000109408 0000204492-0-1 0000FRB3280C9J 0000KY33131358 0000CZC5341NSP 000000113621B6 00000108109-28 00000108109-38 0000KY33280268 0000CZC5341NTR 0000KY33280348 000000008FBE12 0000CZC5341NJJ 0000006E757673 00000000108562 0000206672-0-8 0000CZC5341NTM 0000CZC5480KZ0 0000FRB325095P 0000CZC5341NY3 0000FRB316013C 000000108313-4 0000FRB3270C95 0000FRB42508TR 0000FRB3280B88 0000FRB4380BZM 0000CZC5180W4J 0000FRB32408G5 0000FRB328088X 0000FRB3260B8S 0000FRB4330B94 0000FRB4210CCB 0000FRB3270BG7 0000FRB44103LM 0000107063-018 0000FRB4380C2G 0000FRB43607GL 0000FRB4370CZD 0000CZC5230HHH 0000107063-017 0000FRB44106DP 0000FRB4330B5N 0000FRB43708TG 0000FRB4230Q3K 0000FRB4230Q3K 0000CZC5341NF1 0000FRB43708TW 000000207480-5 0000FRB316012N 0000CZC5341CDM 0000FRB4370CZ2 0000FRB3250CB1 0000FRB3280C7L 0000FRB40907M1 00000000297EL2 0000FRB4330B3W 0000FRB4330B47 0000FRB4380BZK 0000CZC7105FHG 0000FRB32808C5 00000000297EL2 0000FRB4370DBW 0000CZC5230LR2 0000CZC5230HHR 0000CZC5180W4K 000000110199-3 0000FRB4370B3H 0000CZC5320HGH 0000CZC5341NVW 0000000EEB0755 0000000L306R82 00000000108939 0000CZC7204X47 00000000306F48 000000108813-5 0000FRB3160100 0000FRB3270GHF 000000108651-3 0000FRB3270C8W 0000FRB3270C9B 00000005597F2P 00000045DFD862 0000CZC5341P4Z 0000CZC7111TNS 0000FRB3260B1B 0000FRB43605K1 0000FRB44103MF 00000000365EL7 000SCZC530037J 0000FRB4380BZL 0000FRB4370B3D 0000CZC5341NYL 0000CZC5341P4K 0000CZC5341NC8 0000FRB43402X8 000000106679-5 0000000L3MMLYY 0000FRB4230Q3N 0000205899-1-8 00000040778105 000000108452-8 0000CZC6502HGT 00000000016909 000000108118-2 0000CZC5480KXM 00000109386-13 0000CZC8273NS4 00000000108452 0000FRB31600YV 000000107943-1 00000040778154 0000CZC5471Y4M 00000108452-10 00000108452-19 000000108452-1 00000108452-13 000000108452-3 0000FRB31602VH 000000108498-6 00000208869-10 000000641CEBB6 0000FRB4330B3W 0000FRB4330B3W 0000CZC9127SG5 0000CZC5480KZL 0000FRB32408FZ 0000FRB3250CC2 0000FRB3260B13 0000FRB328088Q 00000000288EL1 0000FRB4370B29 00000000079335 0000FRB4370912 0000FRB3250986 00000329343787 0000FRB3270BGE 0000FRB31602VG 00000000205554 0000CZC9127SJN 000000000000T2 0000CZC8273Z8H 0000FRB43708TD 0000FRB437092M 000000099YGCGF 0000FRB3270GWX 0000FRB31602YS 0000FRB3270C8Q 0000FRB3270C8M 0000FRB3270C8R 0000FRB3260B6Z 000000110024-1 0000007AA8A5E0 00000013D86ED7 0000107063-012 0000CZC5472Y5T 0000007E72CF52 000000107859-8 000000108968-6 000000108773-3 00000108773-12 0000FRB32507DD 0000FRB32509BG 0000CZC530035X 0000CZC5341POV 0000CZC5341NY4 0000CZC5341NCS 0000CZC5361JSN 000000140D876D 0000FRB31602XQ 00000000287EL3 000000108213-2 0000FRB40907KB 0000FRB3270GGV 0000FRB3260GWH 0000106662-0-0 0000CZC5341NP2 00000008B58B66 0000006A60EBF8 00000000392EL1 0000FRB32508FR 0000FRB3250C97 0000FRB3260GYN 0000FRB3250CZN 0000FRB4380C0N 0000FRB32602H4 0000FRB3260GYT 0000FRB44106C8 0000003FF36F0B 00000000208928 0000CZC6502HG2 0000CZC5230HHV 0000FRB43402X4 0000FRB3250966 0000FRB3260H2P 0000FR31017758 0000FRB4380C2B 0000FRB3280892 0000FRB32808BR 0000FRB4370CZC 0000CZC5341NV4 0000CZC5361J21 0000CZC5492KDH 0000000I3N2151 000000208008-2 0000CZC6505DRP 00002007-03-14 0000CZC6505DPM 00000060DBA358 0000CZC8271NRZ 0000CZC8224ZLR 0000CZC8224ZKH 0000CZC8273P58 0000CZC8273NN0 000000109541-9 0000CZC822577X 0000CZC8283TZM 000000109725-3 00000002E17FBA 000000109725-7 000000109885-1 00000000A84870 000SCZC5230HH2 0000FRB325055M 0000FRB4210CB8 0000CZC5361JW7 0000CZC5341P2B 0000FRB43402W6 000000099YGCKC 0000CZC941591G 000000109437-2 0000FRB4370CZK 0000005ED8F4E8 0000000L3MMMBL 0000004BC880DC 0000FRB43605J9 0000006AEFC205 0000CZC6502HN3 000000108532-3 000000108532-7 0000FRB3250CZP 0000CZC7105FPL 00000000207283 000000204789-0 00000000106581 000000204789-1 000000108626-1 000000207804-1 00000000204930 00000000107833 00000000212487 00000108294-45 000000531B67E1 000000203973-2 0000FRB42508V7 000SCXC5300382 0000004BC880DC 0000CZC5480KTX 0000CZC55207KG 0000CZC5502N9P 0000CZC7111TPZ 000000110305-9 00000000208257 000000208257-1 000000208257-2 000000208257-3 000000208257-4 000000208257-5 0000CZC7105FNR 0000CZC7205DK7 0000CZC7204X2W 0000CZC7105FNJ 0000CZC7105FRZ 0000CZC7105FRJ 000000207995-1 0000CZC5361J06 000000208714-2 0000CZC548130X 0000CZC7212DZF 0000FRB325098Q 0000CZC5341H8F 00000000209937 0000FRB4330NQ1 0000CZC6502HMM 0000CZC5300343 00000000208793 000000000000TK 0000FRB328088J 000000099YGCLG 0000FRB32609KW 0000CZC5341P48 0000000I306R81 0000FRB32509BK 0000CZC912754W 0000CZC9127563 0000CZC912808G 0000CZC54812XS 0000CZC912754N 0000CZC912756B 0000CZC8273NLG 0000CZC8273NP6 0000CZC8273NL9 0000CZC8273NRK 0000CZC8225751 0000CZC8273NRW 0000CZC83942KR 0000CZC83942LH 0000CZC8273P5H 0000CZC8161XJG 0000CZC8161XMV 0000CZC8224ZJL 0000CZC8273ZC3 0000CZC83942KZ 0000CZC6501QYB 0000CZC6501QY4 0000CZC912754M 0000CZC91280BH 0000FRB32609L4 0000FRB32509BL 00000008C52F06 0000FRB32609L8 000SCZC530039D 000SCZC5300384 0000CZC5371LJR 0000CZC5361JQK 0000CZC5361JQV 0000CZC5480KT3 0000CZC54812ZC 0000CZC6505DNQ 0000CZC7212DX1 000000000000VL 0000FRB43605GT 0000CZC5230HGJ 0000FRB4400J09 0000CZC5361JS1 0000CZC9127SMR 0000CZC5341NL6 0000CZC5341N9Z 0000CZC5341NTJ 00000000208261 0000CZC5480KWP 0000CZC5480JWP 0000CZC5341NTF 0000CZC912759Q 0000CZC5341NT9 0000CZC5341NZB 0000CZC5341NW5 0000CZC5350137 0000CZC535012M 0000CZC5230HHC 0000CZC5341NPO 0000CZC7110GRQ 0000CZC724053M 0000FRB4370DC1 0000FRB31600YS 0000CZC5341NSL 0000CZC5341NCQ 0000CZC5341NGD 0000CZC5341ND0 0000CZC5341NSC 0000FRB4370CYT 0000CZC5341NZC 00000000209128 0000CZC5341N9X 0000CZC5341NZK 0000CZC5341P0D 0000FRB4400J0N 000SCZC530033R 0000CZC5341NYT 0000CZC912756C 0000CZC5481309 0000CZC5471PLH 0000CZC5471LKB 0000CZC54812Y2 0000CZC9127S94 0000CZC912754P 0000CZC9127SGM 0000CZC9127597 0000CZC912759V 0000CZC6501QWQ 0000CZC83942K1 0000CZC8396CHG 0000FRB43607HH 0000FRB42508R9 0000CZC6505DRC 0000CZC6502HG0 0000000L306R45 0000CZC6502KYZ 000000208796-1 0000CZC6501QY2 0000CZC6501QY0 0000CZC7212DYM 0000CZC7212DZ2 0000CZC7211D46 0000CZC724051B 0000CZC7212DYG 0000CZC724053R 0000CZC6501QY7 0000CZC5361J29 0000CZC5341P4X 000000208203-2 0000CZC5341POC 000000208203-4 000000208203-5 0000FRB41308WF 0000CZC5341NXJ 0000CZC5341NTS UNKH3700JA0002 UNKH3700JA0003 0000FRB328089X 0000FR30829792 0000FRB31600YZ 0000FRB3260GZJ 0000CZC5471CCJ 0000CZC5471PM9 0000CZC7105FQB 0000FRB3250CBS 000000099YGCMK 0000CZC7211D40 0000CZC5230HF1 0000FRB4380JY1 0000CZC7211D5X 0000CZC5341NWD 0000CZC5341NQH 0000FRB44109XC 0000FRB43605GV 0000FRB43605JW 0000CZC5341NPT 0000CZC5350127 0000FRB316010N 0000FRB31600Z9 0000FRB328088T 0000FRB328088Y 0000FRB31602X5 0000FRB3280B8S 00000000208170 0000CZC5341NRS 000000210170-1 000000210170-3 0000CZC6505DQT 0000CZC5341NV7 0000CZC5341NJW 0000CZC530037S 0000FRB3270GVM 000000099YGCNB 0000CZC5350129 0000CZC5350124 00000ZC5350123 0000CZC5341NBY 0000CZC7105FJ3 000000209142-1 0000CZC380038N 0000CZC530038R 0000CZC5300341 0000FRB3250976 0000CZC5341NL1 0000CZC912751H 0000CZC5341NX6 0000CZC5480KW6 000000207880-1 0000CZC7105FMC 0000FRB43607GZ 0000FRB4370B3W 0000FRB43607H6 0000CZC7105FNW 00000000209089 0000CZC94154RN 00000000011295 000000099YGCGM 0000FRB437091R 0000CZC7212DYW 0000CZC91275BQ 0000FRB40907JN 0000CZC5471LK1 0000CZC5502N9Q 0000CZC5341NWF 0000FRB4380C00 0000CZC5300388 0000CZC5341N94 0000CZC9127S97 0000003EAF6C53 0000CZC9127SM1 0000CZC5480KWR 0000FRB43708V0 0000CZC7212DTF 0000CZC91280BJ 0000CZC9127SKJ 0000CZC7212DWV 00000000212265 0000CZC5341NH7 0000CZC7105FQH 00000000209014 0000CZC7105FPG 0000CZC7105FQS 00000000210209 0000CZC9128082 0000CZC9127S7J 0000CZC724051T 0000FRB4370B3Y 0000CZC5480KVX 0000006DE48466 0000CZC7212DV7 0000CZC9127S7X 00000051E8FE3E 0000CZC9127S8L 0000CZC83942KG 00000000212268 0000CZC5341NSH 00000000212482 0000CZC5341P4F 0000CZC5341P4F 0000CZC5341NSH 0000CZC8161XGJ 0000CZC8161XJW 0000CZC8283V0D 0000CZC8283V0G 0000CZC8389CC2 0000FRB3260H1X 0000FR31017755 0000FRB4090KZN 0000FRB4330B1V 0000CZC7204X26 0000002892AFF4 0000CZC8283V12 0000CZC5471PLV 0000CZC5480KRZ 0000CZC5471LJX 0000CZC5471PLV 0000FRB3270C8L 0000FRB4370DBJ 00000000000075 0000FRB4380JY5 0000CZC5341NQG 0000CZC8341NGQ 0000CZC5471CD3 0000CZC55207KR 0000CZC54812ZF 00000000208156 0000CZC7105FR5 000000208718-3 0000CZC6502HG7 0000000L306R56 0000CZC7105FP2 0000CZC7105FKG 0000CZC7105FPH 0000CZC7205DP9 00000000210147 00000000209152 00000000209347 000000210684-1 0000CZC8273Z9N 0000CZC8273Z97 0000CZC91275BC 0000CZC9127SM7 0000CZC9127SM9 0000002DA59210 0000CZC94158S3 0000CZC94158S9 000000099YGCFD 000000099YGCFF 0000FRB3250CCG 0000FRB3160146 0000FRB3260GXV 0000FRB325088P 0000FRB325088Y 0000FRB3260BBN 0000FRB3260BC4 0000FRB3260H07 0000FRB3260H07 0000FRB4330B1J 0000FRB4330B1J 0000FRB3150L77 0000FRB4330B5T 0000CZC5341NFK 000000208238-2 0000CZC5480KX1 00000000208816 00000000208843 0000CZC8224ZM0 00000000209791 000000210612-2 0000CZC8225791 0000CZC822575B 0000CZC822576Q 0000CZC8396CBF 0000CZC5481303 0000107063-006 0000CZC902439W 0000FRB4330NPX 0000CZC7211D5R 0000CZC7211D5T 0000CZC7211D5W 0000000988A276 0000CZC7211D5J 0000CZC7240516 00000CZ7111TPQ 0000107063-006 0000107063-006 0000CZC6505DRQ 00000CC5361J1B 0000001CEC0DAD UNKP5080SJ0003 UNKP5080SJ0004 0000FRB4330B5J 0000FRB4330B9B 000000208284-2 000000208284-3 00000000208635 0000FRB4330NQS 0000FRB4370DB0 000000212024-1 00000000212024 000000107945-1 00000000107237 000000107945-2 000000208229-1 00000000208229 00000000209841 0000FRB44103KY 00000000079368 0000107063-030 00000000209126 000000209126-1 00000000209914 000000209636-1 00000000209636 00000000208835 0000CZC6505DQ2 00000000209688 000000208560-2 00000000208560 000000208560-1 000000208631-6 00000000208270 000000208179-2 00000000207830 000000207830-1 000000207830-2 000000207830-3 000000207830-4 00000011DC8C5F 0000003E986761 000000211457-2 00000000208063 000000208063-1 000000208063-2 000000208063-3 000000208063-4 000000208063-5 000000208063-6 0000CZC8273NPX 0000CZC822578W 0000CZC8273ZDQ 0000FRB3260BBQ 0000FRB325088N 0000CZC5361JS5 00000000207758 0000FRB3260B5H 000002082222-2 00000028897B7E 00000000108917 0000CZC548133J 0000CZC5471LJY 000000208497-2 0000CZC6502HJ3 00000000210373 0000CZC7212F3R 0000007BA1EA7B 0000FRB3270GWK 0000000107E09B 0000CZC724057C 0000CZC7212F28 0000MX317SB370 0000CZC7110GSR 0000CZC7211D49 00000000210054 0000CZC7211D3L 0000003745D572 0000CZC7110GSD 0000CZC7110GSV 0000CZC7240534 00000000210053 0000CZC7205DJV 0000FRB3270C9J 0000FRB3270C9M 00000000208191 000000208191-2 00000000211683 0000CZC5180W48 0000CZC8273NS5 0000CZC8283TXG 0000CZC8161XNT 0000CZC5341NFX 0000CZC6505DPD 00000000210327 0000FRB40907L7 00000000209856 0000CZC724055S 0000CZC7212DVW 000000210652-1 00000000210652 0000CZC8273ZFD 0000CZC8273ZFZ 0000CZC8273P4W 0000FRB3280B8M 0000CZC5341NF5 0000000000B469 0000CZC5341NTC 0000CZC912809R 0000CZC5361J1C 0000FRB43607HJ 0000CZC530038T 0000CZC94158VZ 0000CZC5341NYZ 00000000208849 000000208272-2 0000000L3N2126 00000000208558 00000000208557 0000CZC5471Y55 0000FRB3260H28 000000208172-1 0000CZC5341NG6 000000209960-1 00000000209960 00000000208892 0000CZC6502HHQ 00000000210452 0000FRB4190810 0000FRB4270JCF 0000FRB4210CB7 0000FRB4330NQF 0000FRB4330NPP 0000CZC8283TZN 0000CZC9127SKL 0000CZC94158Q9 0000CZC94158R7 0000CZC94158NB 00000000210666 0000CZC8273ZF8 00000000211242 00000000210859 0000FRB4380BZV 0000CZC8161XH1 0000000L3RX453 0000000L3MMMCK 0000CZC5341NLW 00000000053972 0000CZC5341NPK 0000FRB4330B3M 0000FRB42508TV 0000000L3MMMBY 0000000000D547 000000208181-2 000000054F098A 00000000L3X439 0000CZC7212F0M 0000000000D547 0000FRB4380C2Q 0000000L3N2139 0000CZC5341P0T 0000CZC5341NHH 0000000L3MMMDF 000000296AC56C 0000FRB3280C99 000000211456-2 000000084B098B 0000CZC5230HFQ 0000CZC5230HGK 0000CZC5341N9W 0000CZC5230HG8 00000000208178 0000000L3N2166 0000FRB3270GHH 0000CZC5350132 0000000L306R50 0000FRB4330NQ0 000000099HXVWF 0000FRB4330B27 0000CZC5350138 0000CZC5341NGV 0000CZC5481307 00000000208181 0000000L3RX506 00000000211370 00000000208575 0000CZC724051G 00000068D064D4 0000CZC8273P4B 0000CZC8225793 0000CZC83942K7 0000CZC83942K8 0000CZC912809Q 00V312KZKZA084 00000000109911 0000CZC7240513 0000000L3RY112 0000CZC530038X 0000CZC724053H 0000CZC724051P 0000CZC7110GR3 0000CZC7110GR6 00000000211689 0000CZC724051Z 0000CZC9127578 0000CZC9024378 0000CZC548131H 0000FRB44103LQ 0000CZC9127586 00000000209737 0000CZC724053K 00000000017625 00000000211875 0000CZC5341NDB 0000FRB3270GWP 0000FRB32507M7 0000FRB325098H 0000FRB3260BBS 0000FRB42808SJ 0000FRB4230Q46 0000FRB4330B20 0000FRB4330B5P 0000FRB44106CC 0000FRB4370B34 0000FRB44106DL 000SCZC530035G 0000CZC5341P18 0000CZC5341NGC 0000CZC5341NNV 0000CZC5341NZP 0000CZC5341NVD 0000CZC5341NKH 0000CZC5471LJH 0000CZC5471CD0 00000000207941 0000CZC5341NYC 0000CZC548132G 0000CZC5480KS4 0000CZC548135Z 0000CZC5481301 0000CZC5481347 0000CZC54722ZT 0000CZC6502HMR 0000CZC6502KZ4 00000000209324 0000CZC7110GQT 0000CZC7204X6R 0000CZC7105FDL 0000CZC720452X 0000CZC724057R 0000CZC5341NP4 0000CZC7211D43 0000CZC724051F 0000CZC7111TPH 0000CZC83942J7 0000CZC83942J8 000CZC9002437T 0000CZC90243CZ 0000CZC9024399 0000CZC8161XHW 0000CZC912758C 0000CZC9127SGP 0000CZC9127SKQ 0000CZC912752V 0000CZC912751X 0000CZC912751Z 0000FRB43708TR 00000000000011 0000CZC5230HJV 0000CZC5341NXF 000000208341-2 0000CZC548130Q 0000CZC5481364 0000CZC7105FS0 0000000L3N2185 0000CZC7105FGV 0000FRB32408F8 0000FRB4370917 0000FRB43607GG 0000KY33131434 0000CZC5341NBZ 0000KY33131350 0000FR30829800 0000FRB3260B9C 00000000210726 0000FRB4090KZT 0000FRB4270JC5 0000CZC5471CDG 000000212344-2 0000CZC8273Z8Q 00000316938817 0000000L3N2121 0000000L3N2145 0000000L3N2142 000000099YGCHX 0000CZC8225759 0000CZC724057H 00000000017611 0000004B159241 00000048F4A3DC UNKP5080SJ0014 0000FRB4370925 00000000209943 0000FRB4330B4C 0000FRB4330NPD 000000051592-1 0000FRB3260GZ9 00000000212164 00000000206319 0000FRB4330B9R 00000000109768 0000FRB3260GW3 00000000208167 00000000212111 0000FRB4380P1Z 0000FRB4370B26 0000000L3MMMBA 00000000212023 0000CZC91280FR 0000CZC912756H 00000000212516 000000424E2926 0000CZC7212DW2 0000CZC7212F15 0000CZC7212DWN 0000CZC5230HFJ 0000CZC7240543 0000CZC7205DJK 0000CZC7212F0G 0000CZC5341N9R 000000209858-1 00000000209858 0000CZC7212F3N 0000FRB3260H1J 00000000208274 0000CZC6505DPF 0000CZC7212DT5 0000CZC7212DWJ 0000CZC5361JT9 000000108238-3 00000060102D4E 0000001C05F405 00000000109135 0000CZC7212DVH 000000109135-4 0000FRB43402XG 0000FRB4370B2G 0000FRB4330B2G UNKH3700JA0004 UNKH3700JA0005 UNKH3700JA0006 0000FRB3260BB5 0000FRB3280C9K 0000FRB3280GY9 0000FRB3280C9C 0000FRB3270GH3 0000FRB3260BB3 UNKH3700JA0011 UNKH3700JA0012 0000CZC6501QXY 0000CZC6501QXQ 0000FRB3260BB5 0000FRB3280C9K 00000000208418 00000000208350 000000208350-1 0000FRB4330B2R 0000CZC8273NL4 0000FRB42508SH 00000000211726 0000000L3MMLZR 0000CZC5361JSD 0000CZC5361JQZ 0000000L3Y5YC1 0000FRB4330B84 0000000L3MMLYX 0000000L3MMLYH 0000000L3Y5YC6 0000000L3MMLZY 0000000L306R64 000000099HXVXW 0000000L3MMLZP 0000CZC6501QXJ 0000FRB3260B70 0000FRB42508Q7 0000FRB4370DGM 0000CZC5300353 0000CZC530039M 0000CZC7212F1D 0000CZC7212F33 000000208710-2 000000208710-1 00000000208710 00000000208711 00000000208711 000000208711-7 0000FRB4380C2M 00000000017606 00000000211296 000000211296-1 000000211296-2 000000211296-3 000000211296-4 0000000L3RX382 000000209382-2 000000209382-1 00000000209382 00000000209649 0000000L3MN246 0000CZC5350134 0000000I3MMMCC 0000CZC9127SGX 0000003542A44E 0000CZC9127SHL 0000CZC9127584 000000791C18AD 00000000017599 0000FRB43607GD 0000FRB43607H9 0000FRB43607GV 0000FRB43607H4 00000000109668 0000CZC5341NKQ 000000208711-2 000000028711-3 000000208711-4 00000000211454 0000CZC5471PMB 0000CZC8161XM2 0000CZC8224ZK1 0000CZC8161XJ0 0000CZC8224ZGY 00000000017626 0000FRB4330B53 0000005DD35DBF 0000FRB43402WZ 00000000210245 000000110051-1 00000000110051 0000FRB325098J 0000FRB325055X 0000FRB3260BCC 0000107063-020 00000000079951 0000CZC5230HGD 0000CZC5230LQX 00000000207501 0000CZC5341NYF 0000CZC5481340 0000CZC6502KYS 00000000208827 0000CZC7105FQR 000000108480-6 00000000209842 000000751AFCFA 00000000109935 0000CZC9127553 00000000210352 0000CZC7205DGM 0000CZC54812YF 0000FRB4360DV5 0000FRB43605GL 0000CZC5230HCN 0000000L3MMMDB 0000000L3N2135 000000244661B8 00000043470A2A 0000FRB3260B8F 0000FRB3260B0Y 00000000076252 0000FRB44103ML 0000CZC5341NX8 0000CZC5361J2X 0000CZC5341NYB 0000CZC7212DTC 00000000076248 0000FRB3260B0M 0000FRB3260B98 0000FRB3270GWN 0000FR31125600 0000FRB3250565 0000FRB3280B98 0000FRB3260BBP 0000FRB3260H0S 0000FRB3260B57 0000FRB32509CN 0000FRB3270GWV 0000FRB3260B7B 0000FRB3250C2A 0000FRB316010H 0000FRB3250CBD 000000208874-1 0000FRB40906XR 0000FRB32808BH 0000FRB3280B8B 0000FRB325055Z 0000FRB4090KV6 0000FRB4090KVL 0000FRB31602YN 0000FRB43605JD 0000FRB4360F91 0000FRB3260H36 0000FRB438BJY0 0000FRB44103M2 0000FRB3260BBH 0000FRB42106V2 0000FRB40907JL 0000FRB42106V6 0000FRB4210CB5 0000FRB4090KVP 0000FRB4090KTY 0000FRB3270C8M 0000FRB4090KVD 0000FRB316012H 0000FRB3260H2Z 0000FRB419081Y 0000FRB4370B28 0000FRB4230Q42 0000FRB3260GZP 0000FRB4090KVC 0000107063-024 0000FRB316010J 0000FRB3260B50 0000FRB4210CBM 0000FRB4230LHK 0000FR30829791 0000FRB32808C2 0000FRB43708T9 0000107063-036 0000FRB4380C07 0000FRB44103MC 0000FRB43708T3 0000FRB4330NNW 0000FRB43708TF 0000FRB42508V4 0000CZC5341NYP 0000CZC5341NNY 0000FRB43708TN 0000FRB43708T6 0000FRB44109X6 0000FRB4370B24 0000FRB4380C1W 0000FRB44103LY 0000FRB437092X 0000FRB4380P1Y 0000FRB4030KH0 0000FRB4380C2N 0000CZC5341NFR 0000FRB4370DGZ 0000FRB31602VX 000000208874-2 000000207901-1 0000FRB43607G7 00000208874-23 0000CZC6505DNV 0000CZC65012Y0 00000000212448 0000CZC9127SH3 0000CZC5471Y5S 0000CZC94158NK 0000CZC8283V55 0000CZC7212DV6 000000211705-2 0000CZC902439Z 00000000211527 00000329343798 00000000003201 0000FRB42508SP 0000FRB3250CZB 00000208874-18 00000000208822 00000208874-14 000000208874-7 000000208874-3 000000099NFAZF 00000208874-16 00000208874-11 000000208874-6 000000208874-4 00000208874-10 0000FRB4330NQ2 000000538B8B45 000000208874-5 0000CZC6502HL3 0000CZC8225770 0000CZC8273ZDH 0000CZC8225742 0000CZC7204X66 00000208874-15 00000208874-22 0000FRB3270C9R 0000FRB3280B84 0000CZC902439V 0000FRB3270GHR 0000KY36060015 0000KY36060103 00000000000001 0000FRB32408F9 0000FRB3270GH5 0000FRB316012Z 00000334445348 0000FRB3280897 0000FRB3250CZG 0000FRB4160HT1 0000FRB31602YK 0000FRB3260B16 000000208874-9 0000FRB3260GZ0 0000FRB40907K5 0000FRB3260GX9 0000FRB42508XC 0000FRB4330NQ7 000000211570-3 0000CZC5361JV8 00000000211628 0000CZC9127SH9 0000CZC912754B 0000FRB4370B44 0000CZC7212DT3 00000208874-17 0000FRB40906XQ 00000020332950 00000012FA0915 0000001428DC82 0000006B763583 0000002DC33824 0000CZC5480KY0 0000FRB4330B80 0000FRB4370B2Q 0000FRB43402X1 00000000075205 0000FRB4330BB0 00000000211851 000000203985-3 00000000109643 000000203985-2 0000FRB43704B5 0000FRB44103KP 0000FRB4400J06 0000FRB4370DBG 0000FRB4330B32 0000FRB43605G8 0000FRB42508QP 0000FRB4370293 0000FRB4370295 0000FRB4380P1D 0000FRB4380C28 0000CZC530035V 0000CZC530035S 0000CZC5350131 0000CZC5230HHW 00000000079371 0000CZC5341NYM 0000CZC5341NN6 0000CZC5341NLT 0000FRB4380C24 0000CZC5341NDZ 0000CZC5341NLZ 0000CZC5341NL3 0000CZC5300349 000000207861-2 0000CZC5230HJR 0000CZC5341NWQ 0000CZC5361J24 0000CZC5361J30 0000CZC5361JWX 0000CZC5341NK1 0000CZC5471PMF 0000FRB43605FY 0000FRB4370213 0000CZC530035R 00000000207837 000000207837-1 00000000208130 0000CZC5341NWY 0000CZC5230HCY 000000211705-1 000000211705-2 0000FRB4330B60 0000FRB4370D9X 00000000209881 0000FRB44103KS 0000FRB44103L5 0000FRB4270JKS 0000FRB4270JKS 0000FRB44103KJ 0000FRB4380C09 0000FRB4330NQH 0000FRB4370DBV 000000210404-3 000000210404-1 00000000210404 0000FRB4400J0S 0000FRB44103LX 0000FRB44103KF 0000FRB4330B9X 0000FRB4370B45 0000FRB4330B2T 0000FRB4400J0V 0000FRB43402VN 0000FRB43402WG 0000FRB43607GK 0000FRB4400JOR 0000FRB4400J0J 0000FRB4330NQC 0000FRB4400J0D 0000FRB42508T6 0000FRB43605JV 0000FRB43607H7 0000FRB4370DBC 0000FRB4380P18 0000FRB44103MD 0000CZC5230HJ1 0000CZC5230HJG 0000CZC5341NSJ 0000CZC5341NNO 0000FRB4370924 0000FRB4330B2W 0000CZC5341NBC 0000FRB4330NNX 000SCZC530036B 0000CZC5230LQV 00000000108596 0000FRB4090KVH 0000FRB43607HC 0000FRB437091D 000000108596-1 0000CZC5341NFR 0000CZC5341NFT 0000CZC5341NXY 0000CZC5341NY0 0000CZC5341NTH 0000CZC5341NNQ 00000CZC541NMV 0000CZC5341NGB 0000CZC5341NGH 0000CZC5341NCY 0000CZC5341NBW 000000210404-2 0000CZC5341NNG 0000CZC5341NTK 0000CZC5341NCJ 0000CZC5341NMJ 0000CZC535011F 0000CZC5230HDV 0000CZC5341NDS 0000CZC5341NWP 0000CZC5180W4P 00000000211570 0000CZC5361JS4 000000208271-2 0000CZC5341NKV 0000CZC5341NLR 004708WSA85006 0000CAC5471LJP 0000CZC5471CD5 000000207928-2 0000CZC5480KWD 00000000207872 0000CZC5341J1Z 0000CZC5471Y4T 0000CZC5341NTZ 0000CZC5341NLP 0000CZC5471Y4F 0000CZC5480KW7 0000CZC5480KWG 0000CZC548135G 0000CZC5361J1Q 000000208339-3 0000CZC548130F 0000CZC548134J 0000CZC55207LX 0000CZC5480KS8 0000CZC5480KWN 0000CZC5471KPX 0000CZC5341NP7 0000CZC5471PLR 0000CZC5480KZB 0000CZC6501QWM 0000CZC5502N9L 0000CZC6502HKH 0000CZC6502HJP 0000CZC6502HJH 0000CZC6502HKJ 0000CZC7105FHY 0000CZC6502HNM 0000CZC7105FPD 0000CZC7205DJQ 0000CZC5361JXM 0000CZC7205DK4 0000CZC7205DGQ 0000CZC7212DT2 0000CZC7204X20 00000000209446 0000CZC7204X7W 0000CZC7212DY3 0000CZC7212DY9 0000CZC6505DRT 0000CZC6502HGP 0000CZC6505DRK 0000CZC6501QWC 0000CZC7212F12 0000CZC7212F43 0000CZC7212F3F 00000000210113 0000CZC6502HG8 0000CZC7204X5W 0000CZC724053L 0000CZC7240546 0000CZC6502HNR 00000000210564 0000CZC7212DVS 0000CZC7212F4C 0000CZC7111TNK 00000000017609 0000CZC7240579 0000CZC7212DY2 0000CZC724054R 00000000017562 0000CZC6505DR3 00000000017612 0000CZC8273ZGL 00000000109661 00000000109672 00000000109675 0000CZC8283V0X 0000CZC8283V3X 0000CZC8283V0K 00000000211376 0000CZC8396CG6 00000000109639 0000CZC9042389 0000CZC8283V5W 0000CZC7205DNP 0000CZC9024376 00000000211665 0000CZC902437M 0000CZC9024386 00000000211669 0000CZC8396CB8 0000CZC7212DTK 0000CZC8396CC1 00000014C41F34 0000CZC9127S96 0000CZC9127S9L 0000CZC912755W 0000CZC91280CJ 0000CZC9127SB5 0000CZC6505DPH 00000000212246 000000300535DC 00000000209349 0000CZC8224ZJG 0000CZC902437X 000000109B5DC1 0000002C8DCB31 0000006C63CAF3 0000CZC91280B0 000000315E4DC4 0000CZC9127521 0000CZC9127S7G 0000CZC91280BT 0000CZC91280BW 0000CZC9127SLY 0000CZC91275BR 0000CZC91280B4 00000000212255 00000000212120 0000CZC91280DX 0000CZC912756J 0000CZC912809Y 0000CZC9128098 0000CZC9127546 0000CZC9127S8R 0000CZC91280FS 0000CZC9127SJ1 0000CZC91280DY 0000CZC7204X3X 0000CZC912751D 00000000212365 000000212365-1 00000000210935 0000CZC9127SKZ 0000CZC94158R2 0000CZC94158ND 0000CZC9127SK3 00000000212317 0000CZC94158YR 00000000212241 0000CZC5230HGB 00000000211841 000000211570-2 000000211570-1 00000000005204 0000CZC7105FLC 0000CZC7211D3D 00000000210103 00000000210094 00000000211567 0000CZC54812XG 0000FRB3260H03 0000FRB4380C31 00000000211369 0000FRB3250CB0 00000000210104 0000CZC5341NV8 0000CZC5361JXD 0000FRB316013L UNKP5080SJ0019 0000FRB4370D9R 0000FRB3260GXQ 0000CZC5341NZQ 0000CZC5341NWJ 0000FRB42508QG 0000FRB44106F5 0000CZC8161XKB 0000CZC8161XGZ 000000000003KW 0000FRB4370B2P 0000CZC8273P6C 00000000209323 0000FRB4270JKT 0000FRB4270JL6 00000000210391 0000FRB3260GZY 0000FRB4230Q3H 000000017508-1 0000FRB43607G2 0000FRB43607GM 0000CZC724056L 00000000211603 00000000210015 0000CZC6502HKF 00000000208854 0000CZC7105FGL 000000108688-1 00000000212131 000000212131-1 0000CZC5230HTC 0000CZC7205DN9 0000CZC7212F30 0000CZC7204X1W 0000FRB62509C0 0000FRB3270GVQ 00000000208807 0000FRB44109X2 00000000209638 000000208133-2 0000FRB325097W 00000000210372 0000CZC65012XP 00000000209405 0000CZC5341NSY 00000000211525 00000000211524 0000CZC5361JQD 0000005365BEA2 00000000210480 0000CZC7211D4K 00000000210480 00000000017604 0000CZC7205DKF 0000FRB4270J3W 000000208385-1 00000000211602 0000FRB3250992 0000CZC8161XHV 0000CZC5341NFV 00000000208918 00000000109751 00000000A86700 0000FRB4230Q4F 0000CZC5341P3W 0000CZC9127SG3 00000047377C4D 00000000208916 00000000208462 0000000L3RX387 0000CZC5361JQJ 0000CZC7240573 0000CZC5341NV3 00000000310135 UNKH3700JA0013 0000CZC724051D 00000000211141 00000000051570 0000FRB42106V8 00000000211946 00000000108835 0000FRB43607JO 00000000210568 00000000212519 0000CZC5341NN1 00000000210129 0000CZC7105FDC 0000CZC7105FCZ 0000FRB4210CBS 0000CZC6505DQ1 00000000211326 000SCZC530037W 0000CZC8273ZG5 0000FRB3280B95 0000CZC5341N8G 00000000211916 0000000000B336 0000000L306R76 0000CZC5361J2P 0000CZC7211D5K 0000CZC7212F29 0000CZC8273Z92 00000000212253 00000000212177 0000FRB3260H20 0000CZC5341NCN 0000000L3MMMBB 00000000208833 0000CZC8224ZMD 0000CZC5341NVC 00000000208351 0000CZC5361J2L 0000CZC8283V32 00000000208174 0000CZC6502KYR 00000000209977 0000FRB43704BJ 000000208729-2 00000016520B34 0000CZC5480KWH 0000CZC5341NWR 0000FRB4370DBB 00000000208046 00000000208057 0000CZC65012YH 0000FRB4330B50 00000000211997 0000CZC5230HCS 0000CZC902437R 0000CZC5341N9F 000000208161-1 00000000208645 000000208083-2 00000000211898 0000FRB4330B2H 00000000208780 00000000208192 00000000209049 0000002BB4B981 0000FRB4330B2J 0000CZC54724W4 0000CZC5341NG5 0000FRB4370DB8 0000001F870794 00000002084543 00000000209121 00000000211423 0000CZC9127S8N 0000CZC724050Y 0000CZC7240574 0000CZC912809Z 000000211566-2 UNKH3700JA0014 UNKH3700JA0015 UNKH3700JA0016 UNKH3700JA0017 UNKH3700JA0018 UNKH3700JA0019 UNKH3700JA0020 0000FRB3160118 0000FRB3160148 0000FRB44106CY 0000FRB44106DQ 0000FRB44106CZ 00000000109513 0000FRB4370DGP 0000CZC548122T 0000CZC5480KTD 0000CZC5481312 000000108017-6 0000CZC5341NF0 0000FRB325096L 0000FRB3260H2Y 0000CZC5341NKN 00000000053964 0000CZC5230HJC 0000CZC5180W4Z 00000000109030 0000CZC7212DZV 00000000108301 00000000108201 0000CZC5361JVZ 0000CZC7212DT6 UNKH3700JA0021 UNKH3700JA0022 UNKH3700JA0023 UNKH3700JA0024 0000KY32120026 000000107901-1 000000108018-6 00000000107901 00000001073003 0000CZC5341NT8 00000000108104 000000108104-1 000000108104-2 000000611B5EBE 0000CZC7105FCW 00000000211634 0000CZC7110GPW 00000000209141 0000CZC5341NJG 0000CZC5341NH3 0000CZC5341NK7 0000CZC5341NCF 0000CZC5471Y62 0000CZC5361J0Y 000000099YGCHT 0000CZC6505DRX 0000FRB43708TP 0000FRB43605GC 0000CZC5300389 0000000L3MMLYN 0000CZC530034N 0000CZC5341N8D 0000CZC5480KYQ 0000CZC5350142 0000CZC5341NV6 0000CZC5480KZ1 0000CZC5471LK6 0000CZC6501QXT 0000000L3N1991 0000CZC7105FPJ 0000CZC7204X1Q 0000CZC65012YJ 0000CZC7212DW5 0000000H3900JJ 0000CZC7240569 0000CZC724053Z 0000CZC724051K 0000003569B010 0000CZC6505DR7 0000CZC7205DKL 0000CZC8273Z9X 0000CZC8283V7Z 0000CZC8273Z8R 0000CZC8283V2V 0000CZC8283V1G 0000CZC83942JW 0000CZC8396CDD 0000CZC9127580 0000CZC9127SMS 0000CZC9127SFT 0000CZC912759X 0000CZC91280F9 0000CZC94158W8 0000FRB42508RQ 0000CZC8396CDD 0000CZC7205DKL 0000CZC9127SFT 0000CZC83942JW 0000CZC5341NV6 0000CZC7205DJR 0000CZC8273Z8R 0000CZC8273P4B 0000CZC5300389 0000CZC7105FMK 0000CZC5471LK6 0000CZC912759X 0000000L3N1991 0000CZC5480KZ1 0000CZC9127580 0000FRB43402VS 0000CZC6501QXT 0000CZC6505DR7 0000FRB4400JOL 0000FRB43708TP 0000CZC5341N8D 0000FRB43605GC 0000CZC8283V1G 0000CZC94158W8 0000FRB3250H0V 0000CZC7240569 0000CZC724053Z 0000CZC7105FPJ 0000CZC7204X1Q 0000FRB4370CZ5 0000000H3900JJ 0000CZC7212DW5 0000CZC530036C 0000CZC530034B 0000CZC9127SMS 0000CZC91280F9 0000003569B010 0000FRB4330B1Z 0000FRB4400J0K 0000CZC8224ZKC 0000CZC724051G 0000CZC5480KYQ 0000CZC65012YJ 0000CZC8283V7Z 0000CZC8273Z9X 0000CZC5350142 0000CZC724051K 0000CZC8283V2V 0000CZC8225789 00000000209697 0000FRB32408FL 00000000208068 0000107063-037 0000FRB3270GVF 0000FRB3270GVN 0000FRB4330B97 0000CZC5341P4L 0000CZC5341P4R 0000CZC5341P4J 0000CZC5341NYX 00000000207894 0000CZC5341NB8 0000CZC7105FHH 0000CZC7105FHN 0000000L3RX503 000000127DF5C8 0000CZC822575D 0000CZC8225763 0000CZC8283TXY 0000CZC9127SK8 0000FRB3250CCN 0000FRB44103MG 0000CZC548134K 0000CZC5481366 0000FRB43704BD 000000099YGCFZ 000000099YGCFW 0000CZC8273P5L 00000000210433 0000FRB4370B1S 0000CZC5341NHC 0000CZC5341NFG 0000FRB32808BX UNKP5080SJ0018 0000000000C197 0000FRB4400J03 00000000208346 0000CZC530036K 00000000051577 00000000211488 0000FRB4380C2Y 0000000L3MMMDX 000000208709-2 0000FRB3270GVD 0000FRB31602TP 0000FRB4330B3Z 00000000208913 00000000208778 UNKP5080SJ0017 0000FRB4380C06 000SCZC530037Z 0000CZC912759C 0000CZC5341ND6 0000CZC5341NLB 0000CZC7105FRX 00000000209168 0000FRB4330B3C 00000000212181 0000FRB42508TT 0000FRB31600ZJ 0000FRB42508WR 0000CZC912757C 0000CZC55207LK 0000CZC55207LT 00000002542272 0000CZC55207L9 0000CZC55207LL 0000FRB4380BZR 0000CZC530036S 0000CZC902438R 0000CZC5341P1W 0000CZC7105FQQ 00000000212320 0000CZC8273NLR 00000000212733 0000CZC9414HGM 00000000208750 00000000210228 00000000209066 0000CZC5361J1P 0000FRB4380P1P 00000000211102 00000002D9BAA5 00000000212457 00000CZC534NV0 0000CZC7105FJC 000000211901-1 0000CZC530035W 0000CZC530035Y 0000000CZC5341 00000000212279 00000000017592 0000FRB32609ZB 0000FRB325097S 0000CZC55207LK 0000CZC724053Y 0000CZC7212F21 0000FRB3270CB1 00000000212185 00000000209145 00000000211631 00000000208818 00000000209248 0000CZC7105FLS 00000000208813 00000000212047 0000001A9DBAEA 00000000208995 00000000210356 00000000212048 00000000208431 0000FRB43605HY 0000CZC5341NVN 00000000211983 00000000209528 00000000209278 00000000210871 0000CZC5341NMG 00000000208398 0000CZC6501QXP 00000000209951 00000000211375 00000000211629 0000CZC724055N 0000CZC5341NL0 0000CZC5341NZL 0000CZC7110GP6 0000CZC9127S8Z 000000694CD134 000000207926-1 0000CZC7211D4N 00000000209122 0000FRB4330B3G 00000000212436 00000000212569 00000000207988 00000000210319 000000208243-2 000000208814-1 0000CZC6502KYQ 0000FRB4370B2M 00000000207952 00000000208506 0000FRB43605GZ 0000CZC5341P45 00000000211937 0000000H3800LA 0000CZC5230HHT 00000000209978 0000FRB3260B86 0000FRB31602WJ 0000CZC5230HHT 00000000208427 000000208427-1 000000208427-2 0000FRB43402WB 0000FRB4330B2M 0000FRB4330B42 00000000210966 0000CZC9127SHP 000CZC9412756P 00000000210665 0000FRB43402W0 00000000211266 0000CZC7204X5X 0000CZC7105FLN 0000CZC5341NGS 0000000L3MMMAN 0000000L306T03 0000FRB3260B0S 0000FRB3260H3B 0000FRB3260B5M 0000FRB3260BC7 0000FRB3260BC7 00000CZC72121C 0000CZC7212F1F 000000207763-3 00000000208233 00000000209052 0000CZC5361JOX 0000CZC6502HND 0000CZC6505DP1 0000FRB4370B3S 00000000210105 0000CZC8396CCG 00000002AECBF1 0000FRB3260H3B 0000FRB3260H3B 0000FRB3260BB1 0000FRB3250C96 00000000208895 0000CZC65012Z6 00000000211843 0000003341430C 0000CZC912752F 0000CZC912753D 0000CZC9127SJC 0000FRB3260H2B 0000FRB3270CBT 00000000208912 000CZC82258799 0000CZC7240514 0000CZC7212DYP 000000209435-2 0000CZC7240548 0000CZC8161XGS 00000000209955 00000003693CF2 00000000207674 000000207674-2 00000000208047 000000208047-1 00000000209433 0000006740F94C 0000CZC7205DMM 0000CZC7205DLX 0000000FD8A57E 0000000L3N2101 0000FRB32508F6 0000FRB328087G 0000CZC5341NW7 0000CZC5471LJC 000000211142-2 0000CZC822574C 0000CZC822574G 00000000211142 0000CZC5341NYJ 0000CZC5341NDV 0000FRB32408FV 0000FRB3260GYM 0000FRB4370B1V 0000FRB4330B5L 00000000208515 00000000212348 00000002102111 0000CZC5341NYV 0000CZC7111TNG 0000FRB3250996 000000210939 1 00000000209604 0000CZC8273NLM 0000CZC65012Y7 0000CZC65012YQ 000000208564-3 00000000208564 0000000L3N2100 0000CZC5341P0J 0000CZC5341NB5 000208043-20-2 0000CZC5481337 0000CZC5480KT6 0000CZC5480KTY 0000CZC5480KSN 0000FRB3260GW9 000000207701-1 00000000207701 000000207701-2 000000207701-3 0000CZC7105FRG 0000CZC5361JT2 0000CZC5361J19 0000CZC5361JW8 0000CZC5361JW5 0000CZC5361JRL 0000FRB43605H7 0000CZC5230HHL 0000CZC5230LQP 0000CZC5341NY7 0000CZC530036N 00000000207462 0000CZC5361J2F 00000000211117 00000000208039 000000208039-1 0000CZC5480KWX 0000CZC55207KN 00000000208234 00000000085001 0000FRB4160HSZ 00000000210032 0000CZC5361J1D 0000CZC5341NX4 0000CZC5341N95 0000CZC7240550 0000FRB4370920 0000CZC912808K 0000CZC912808R 0000CZC8396CHL 000000192EF06B 0000FRB42508WP 0000CZC6501QY6 0000CZC6502HKB 0000CZC6501QX7 0000CZC6501QW7 00000000209641 000000209641 2 000000209641 3 0000CZC6501QY9 0000CZC8161XGT 0000CZC7211D4X 0000CZC7212DVB 0000CZC8273ZB2 0000CZC8283TXL 000000211100-2 000000207796-1 00000000207796 000000207796-3 000000207796-4 000000207796-2 0000CZC7212DXB 0000FRB42508RX 0000CZC8224ZLF 0000FRB43605FZ 00000000211305 00000000211305 0000FRB44106DM 0000CZC912751R 0000CZC912755B 0000CZC9127S8C 0000CZC5341N96 0000FRB3280B9K 000000210032 2 0000CZC7110GP3 0000FRB4360F90 000000209381-1 000000209381-2 0000CZC7212DSY 000000209381-3 000000208912-1 00000028AD6F68 0000003A1C32AC 00000000210926 0000CZC8396CC8 0000FRB4370B3S 000000208912-2 00000000210450 0000CZC8396CCC 0000000L3RX371 00000000210939 0000FRB3260B56 0000CZC7205DGT 00000000204534 0000FRB32609ZK 0000FRB32507MD 0000FRB328089F 00000000000082 0000000L3MMMAT 00000000209434 0000CZC7204X79 0000FRB32408G7 0000FRB32509CB 000000099YGCGN UNKP5080SJ0013 00000000209086 0000FRB4330B3J 0000FRB3250C9R 00000000208230 00000000209124 000000209124-1 00000000207994 000000207994-1 00000000208704 00000001306R79 0000CZC5341P3C 0000CZC5341P3C 0000000L306R68 0000CZC912754K 000000208415-2 0000CZC548130D 0000FRB4360F92 0000FRB4330B2F 0000FRB3260B6S 0000CZC5230HF3 0000CZC5230HFK 0000FRB3260B77 0000FRB4370B1L 0000FRB4370B2N 00000000210224 00000000110364 0000CZC5480KY7 00000000207990 0000FRB42508R7 0000FRB44109X4 00000000000CZC 0000CZC5341NJ1 0000FRB3260B7G 0000FRB3260B7F 0000CZC5341NLV 0000CZC8283V3Q 0000CZC8396CHM 00000109878-19 0000FRB4330B90 000000210399-1 0000CZC8273NMW 0000CZC9127SHN 00000000376257 00000000376242 0000FRB4330NPJ 0000FRB437092C 0000FRB437091Z 0000FRB4360DV7 0000CZC8161XKV 00000000A85912 000000106616-5 000000106616-3 00000000A86141 000000106616-4 000000106616-0 00000106616-10 000000106616-9 000000106616-2 0000FRB3280898 00000106616-11 0000CZC548134F 00000000208198 000000208198-1 00000000208204 000000208636-2 00000000208636 0000CZC5481355 0000CZC548133Z 0000CZC548134Z 0000CZC548131S 000000208445-3 00000000211146 000000211146-1 000000211146-8 000000017635-3 0000CZC822573F 0000CZC822574J 0000CZC8225760 0000CZC822577K 00000000208397 0000CZC548131Q 0000CZC6502HKQ 000000211971-2 0000000F505A68 000000211971-1 00000000208134 0000CZC6502HHT 0000FRB4400J0Z 0000FRB4330B51 00000000051572 0000FRB4370CZJ 0000CZC530035F 00000316938800 0000CZC7240571 0000CZC7240537 0000CZC7212DWY 0000CZC7111TP2 0000CZC7111TP8 0000CZC8283V0V 0000CZC822575S 0000CZC9415911 0000CZC9415921 0000CZC5341NVB 0000FRB3260GZH 0000000000D598 0000FRB32808BD 0000FRB42508X3 0000CZC5341CN1 0000003C1BE5A4 0000FRB3260B9G 0000CZC5480KWC 0000FRB3260GY4 0000FRB42808SY 0000FRB3260GXC 0000CZC5230HFS 00000000108812 0000FRB42508TN 0000FRB4380C1S 0000CZC5471CD6 0000CZC5480KS0 0000CZC8396CCS 0000CZC5361JVY 0000FRB42508WV 00000000208777 0000FRB325099Z 0000FRB3260BBM 000000000003FY 0000CZC5361JS7 0000CZC5350143 0000FRB3160113 0000FRB3250972 0000FRB316011Q 0000FRB3270CBR 0000000000D357 0000FRB44103KL 0000CZC83942KB 0000CZC9127S7V 000000000003FW 00000000209354 0000000000B422 0000CZC548131V 0000CZC5361JTK 0000000000C193 0000FRB43704BM 0000CZC5361JXC 0000CZC8273NPN 0000FRB43605HX 0000CZC5471Y5B 0000CZC518132X 0000FRB3280C72 0000FRB3270GH4 000FRB43607F8W 0000006896CE3C 0000CZC5230HDT 0000MXS33202JR 0000FRB44106C7 0000CZC5341NKF 00000000210681 0000CZC9404HLO 00000000377EL2 0000FRB3260BCR 0000CZC8161XGV 0000FRB4270JCM 0000FRB43708SW 0000CZC5341NXG 0000CZC548135F 0000CZC8396CDV 0000CZC94158W3 0000CZC912758Y 0000FRB3260BBM 0000FRB4370B1X 00000000387EL1 0000FRB4090KTT 0000FRB32509CM 0000FRB40907LM 0000FRB40907LM 0000FRB32509BD 0000FRB3280C7N 0000FRB3280C7J 0000FRB3280C7J 50877LMAVA00EB 0000CZC6501QXN 0000CZC7105FFL 000000207795-1 000000207795-2 00000000207795 0000CZC724052P 0000CZC7240552 000000099YGCFX 0000CZC8273NLV 0000CZC8273NL5 0000CZC5341NKB 0000000L3Y5YC4 0000CZC7105FK9 0000CZC7105FM2 0000CZC7105FNV 00000000108717 0000FRB3250CC6 0000FRB4370217 0000FRB4370DGF 00000000108773 0000CZC7205DMT 000000108773-7 0000CZC5341N9C 0000CZC5341NCD 0000CZC5341NNK 000000208228-7 000000099HXVYM 0000FRB43704B6 0000000L3RX435 0000FRB437021F 0000CZC8396CBD 0000FRB43708TX 0000FRB43605GW 0000CZC5361JV0 0000CZC5361JR2 0000CZC548132D 0000CZC7205DGG 0000CZC7212DSR 0000CZC7212DXT 0000FRB43607GT 0000FRB437091P 0000FRB44106CD 0000FRB44106C6 0000FRB44106C0 0000FRB43607HV 0000FRB43607HY 0000FRB4370B3B 0000FRB43607HX 0000CZC6505DQ7 0000CZC535013C 0000CZC5361J1J 000CZC54780KW4 0000004FF35C7B 0000CZC5480KXP 000000108142-7 0000107155-3-1 00000013AAD30D 000000207770-3 000000207770-1 000000208904-1 0000CZC5361JW2 0000CZC5361JV6 0000CZC7204X2S 0000CZC5480KVB 0000CZC5480KT0 0000CZC5480KV9 0000CZC5341NMB 0000001D95F4F4 0000CZC7204X2D 00000040A80146 0000CZC5341N8N 0000CZC5341P23 0000CZC5341NB9 0000CZC5230LQQ 0000CZC5230HG9 0000FRB4330NP4 0000FRB4330NP4 0000CZC7212F1P 0000CZC7240517 0000CZC535013Y 0000CZC5341P2P 0000000I306R85 0000CZC5180W4C 000FRB441103L6 0000CZC8273ZGG 0000CZC8273ZG4 0000CZC8224ZL8 0000CZC8273NPM 0000CZC8273ZG1 0000FRB4330B5M 0000FRB44106D2 0000FRB43708TH 0000CZC5230HJ6 0000FRB3260H0W 0000CZC7240563 0000CZC7212DYT 0000CZC7205DMW 0000CZC7105FDB 0000CZC7212DYF 0000CZC7105FD3 0000FRB31602XB 0000FRB3260B91 0000FRB32609KR 0000CZC5341N8R 0000CZC6502HNL 0000CZC5230HC2 0000FRB44103GM 0000CZC5341P1S 0000CZC5341NYS 0000CZC7105FL5 0000CZC7110GQQ 0000CZC65012Z0 0000CZC7105FQ5 0000CZC7105FN4 0000CZC5341NJ2 0000FRB3270C88 0000206763-0-9 0000CZC5300363 0000CZC5341NRQ 0000CZC8273NP0 0000CZC6505DQX 0000CZC8273P51 0000CZC8273NL8 0000CZC8273P4F 0000CZC8273NN7 0000CZC8273ZF1 0000CZC8273NMB 0000CZC6502HHB 0000CZC635045B 0000FRB437092K 000000099HXVYB 0000CZC5341NWK 0000FRB4330NNY 0000CZC5341NYN 00000108012-25 00000108012-26 00000108012-23 0000CZC8273ZBP 0000CZC8273NNR 0000CZC8161XNW 0000CZC8283TYR 0000CZC8273ZBD 0000CZC8161XNX 0000CZC912757S 0000CZC8273NQM 0000CZC8283V5M 0000CZC8283V6D 0000CZC822575Z 0000CZC9127SJP 0000CZC9127S8K 0000CZC822578G 0000CZC822578C 0000CZC8283V82 0000CZC822578L 000000099YGCFN 0000CZC822577S 0000CZC8224ZL9 0000CZC8225785 0000CZC8224ZL5 0000FRB4370C01 0000FRB4380C03 0000FRB43708TL 0000FRB4370DBM 0000FRB4330B3K 0000FRB4370B1W 0000CZC7105FGR 0000CZC7105FD4 0000CZC7205DJS 0000CZC7105FFQ 0000CZC7105FFJ 0000FRB4380C25 0000FRB4370916 0000CZC8273NM5 0000CZC822576C 0000000L3N2182 0000CZC8283V1D 0000CZC8273NQY 0000CZC90243BB 0000CZC822576N 0000CZC822576P 0000CZC8161XKF 0000CZC6502HKY 0000CZC6502HJR 0000000L3N2162 0000000L3N2159 0000CZC55207KV 0000CZC5361JRC 0000CZC5361J23 0000000L306R77 0000000L3N2158 00000005597D6B 0000CZC65012Y6 0000CZC6501QW8 0000CZC6501QYL 0000CZC65012YY 0000CZC6501QYC 0000CZC6501QYG 0000CZC65012Z2 0000CZC6501QWH 0000CZC5480KVS 0000CZC5341NB1 0000CZC5341NLH 0000CZC5341NSM 0000CZC5471LJT 0000CZC5341NDQ 0000000L3N2178 00000108012-47 0000000L3N2094 0000FRB4270JCH 0000FRB4270J41 0000FRB3250CC4 0000FRB32609L9 0000CZC6501QXW 0000CZC6502HGR 0000CZC6501QWL 0000FRB3250C9J 0000FRB32808BC 0000007950251E 0000CZC5180W50 0000CZC724053T 0000FRB4330B7Y 0000FRB43402N6 0000FRB44103KD 0000FRB44103KX 0000FRB3280BGZ 0000CZC5341NXH 0000000L3N2170 00000036DA6698 0000FRB4270J3S 0000FRB4330B3T 0000FRB4370CZ4 0000FRB4330B4N 0000FRB4380C0F 0000FRB43708SN 0000CZC5341NR2 0000CZC5300380 0000FRB325055Y 0000FRB3260B7W 0000FRB3250CBB 0000CZC548133G 0000FRB316013J 0000CZC7212F1Y 0000CZC55207LD 000000108793-9 0000CZC5341N8C 0000000L306R52 0000CZC530038Q 0000CZC530037D 000000109417-3 0000109541-1-1 00000000109575 0000CZC5310XQ0 0000FRB3250C99 0000CZC7212F0Y 0000CZC5341NLY 0000FRB3250C9M 000000207075-1 0000FRB3250C96 0000CZC5230HGS 0000CZC9127529 0000CZC5230HF4 0000CZC5230HHM 00000000107070 0000FRB328088K 0000FRB42508XB 000000108796-3 00000072C9A18D 0000CZC5341NXW 0000CZC5350130 0000CZC7212DXW 0000CZC7212DX7 0000CZC7212DXK 0000CZC7212DXG 0000CZC7212DYR 0000FRB4210CB6 0000FRB43402VB 0000CZC912751S 00000000209792 00000000209792 0000FRB3250C9N 000000207074-2 0000FRB3260H1M 000000209792-5 0000FRB3260H2M 0000CZC5471PMH 0000CZC5471CCP 0000CZC5471CD4 0000CZC5471CCZ 0000CZC5341NN3 000000207877-1 00000000207877 00000000211422 000SCZC530036D 000SCZC530036D 0000FRB4270JL2 0000FRB4380JY4 0000FRB4330B37 0000FRB44106D1 0000FRB43705M7 0000FRB4380JXV 00000108012-12 0000PA47015178 0000FRB3260GYC 0000FRB3160104 0000FRB4330B4H 0000FRB4400JOT 0000FRB4400JOP 0000FRB4330B9W 0000FRB4400J0X 0000CZC5341NQ7 0000CZC5341NPR 0000CZC8273NMH 00000008844C32 0000000L3RY118 0000CZC5341NJ7 0000CZC5341NV9 0000FRB3260B0N 0000CZC5341NNC 00000000207727 00000000207726 0000CZC5341NG3 0000CZC5341NJF 0000CZC54724VW 00000000207966 0000CZC5471LK2 0000CZC7110GP8 00000008F9436D 00000000212540 0000CZC90243CX 0000CZC8396CB4 0000CZC902439B 000000109878-4 0000CZC8396CB7 0000CZC8396CB5 0000CZC6501QW2 0000CZC6502HLY 0000CZC6502HJT 0000CZC724050Z 0000CZC5361JVW 0000FRB32808FW 0000FRB43402WJ 0000FRB43708T8 0000CZC7111TNX 0000CZC7111TNF 0000CZC7204X82 0000CZC7204X5C 0000CZC7111TNY 0000CZC7111TPP 000000611F2088 0000002D131F9F 000000019AFCC5 0000006F622501 0000CZC5471Y5L 0000CZC535011S 0000FRB32408G8 0000FRB32509CL 0000CZC5471KQ5 0000CZC91275JZ 0000CZC7105FQF 0000CZC7105FR9 0000CZC7105FQV 0000CZC724050X 0000CZC7105FQG 0000CZC7111TP7 0000FRB43708ST 0000FRB4370CZ0 0000FRB4330B4D 0000CZC5361JTG 0000CZC5471CCS 0000FRB3260GZZ 0000CZC724054B 0000CZC6502HMM 0000CZC7111TPB 0000CZC724056H 000000109624-2 00000000212524 0000CZC91275B3 0000005E821FD1 0000CZC7212DZ1 000000109437-9 0000CZC7240561 0000CZC8273NQK 0000CZC8283TZL 0000FRB4230Q3N 000000207769-1 0000CZC8273NM8 000000109841-7 000000110257-3 0000FRB32408G3 0000FRB3270C88 0000FRB43605JT 0000FRB43605H9 0000FRB44106CX 0000FRB43706C7 0000FRB43706C6 000000207875-1 0000CZC5341N90 0000CZC8161XMD 0000CZC8161XJ9 000000108319-4 00000000108294 0000CZC6501QWN 0000FRB43605GF 0000FRB43605GN 0000CZC535011H 0000CZC6501QW9 00000107976-11 0000000I306T04 0000CZC5180W4T 0000FRB4380P17 0000CZC83942JS 0000FRB44103FF 0000FRB44103KZ 0000FRB44103FH 0000FRB4330B3S 0000FRB41308WB 0000FRB41308W0 0000FRB4330B5B 000064110833YB 0000FRB3280BGZ 0000CZC5341P23 000000000$NBSP 00000000207965 00000000207962 00000000108012 0000CZC5341NSM 000000108748-9 0000CZC6502HNL 000000108748-5 000000109767-5 0000CZC94154WH 0000CZC94158R3 00000000210278 0000FRB44106C1 00000000206235 0000CZC5361J10 0000CZC5471Y5W 0000CZC5341NN4 0000007A929C83 0000CZC7204X6B 0000CZC65012XT 0000CZC8396CHN 000000109624-8 0000006E98C7FC 0000CZC9127SN3 000000212550-6 0000000F11A1B1 0000CZC90243C3 0000000FB5B0B5 000000110257-5 0000FRB3260GY8 0000FRB4330NPM 0000CZC5361J1R 0000FRB43704B3 0000FRB4330NPK 0000CZC5341N8Y 0000CZC5341NL9 0000CZC535013V 0000CZC5480KY2 0000CZC5230HCT 0000CZC5341NP6 0000CZC5341P39 0000CZC535013D 0000CZC6501QW5 0000CZC5361JWH 00000000207888 0000CZC7105FNM 0000CZC5361JWG 0000CZC5480KWJ 00000015B91A93 0000CZC54812X3 000000233C8BDA 0000CZC6502HLH 0000CZC7205DJ2 0000CZC7110GRC 0000CZC7105FKM 0000CZC7212DY0 0000002F6F09D0 0000CZC7240564 0000CZC8273NRR 0000CZC724051N 000000109189-4 00000060143A28 0000CZC7212DWK 0000CZC7240565 000000109250-2 0000CZC7211D45 0000005A5BC76C 0000005A5B374D 000000109250-6 0000CZC7111TNL 0000CZC7211D4C 00000022647EAB 0000CZC8273NP4 0000CZC8273P60 0000CZC7240575 00000109408-16 0000CZC8224ZJ7 000000495E60F1 0000CZC8225783 0000005718F20E 0000CZC8283V64 000000558D3190 000000243BC959 0000CZC912754L 0000006D7E3F18 0000CZC91275B8 0000CZC9127SCZ 0000CZC94158YP 0000CZC9127547 0000CZC912754D 0000CZC912754C 0000CZC912756M 0000000DE22F28 0000CZC9127SGV 0000CZC94158NV 000000341F08C9 0000007BA62AF8 0000FRB3280B9V 0000FRB32602H1 0000FRB3250CZR 0000FRB32408F5 00000000207885 0000FRB3280C74 0000CZC724052B 0000CZC5361JT4 0000CZC724054C 0000CZC7240536 0000CZC724055P 0000FRB3260H30 000000099YGCFP 0000006A612941 0000CZC724056V 00000000210184 0000CZC9127S7B 0000CZC83942JK 0000CZC5471PMJ 0000CZC5341NFL 0000CZC5341NNL 0000CZC5341P2M 0000CZC7204X8J 0000CZC7204X1H 0000FRB4330BBR 0000FRB43402W2 0000CZC720455C 0000FRB3270GWS 0000CZC5341NMT 0000CZC8261XGR 0000FRB44106C2 0000CZC822576X 0000FRB325099B 0000FRB3250C9Q 0000FRB31600Z3 0000CZC5230HJF 0000CZC5341NW3 0000CZC7111TNZ 0000CZC7111TNW 0000CZC7211D4G 0000CZC54812WR 0000CZC535013J 0000FRB316012V 0000CZC7110GQK 0000CZC724053C 00000020546D1C 0000FRB43402WN UNKH3800JA0000 000000208088-1 0000CZC8161XMH 0000CZC7105FJJ 0000FRB4330NQ6 0000CZC5300371 000000013MMMDW 0000FRB3250CB9 0000FRB4230Q3X 0000FRB43402XX 0000CZC724057G 0000CZC5341NV1 0000CZC6502HNB 0000CZC6502HNQ 0000CZC7110GR8 00000067354116 0000CZC5341NYK 0000CZC5361JV2 0000CZC5341NT2 0000FRB4330BB2 0000CZC912809D 0000CZC912756L 000000212502-2 0000000000B312 0000CZC7211D4V 0000CZC5341NMQ 0000CZC5341NMH 0000CZC6502HNC 0000CZC6501QYQ 0000CZC6501QYF 0000FRB43605GH 0000FRB43605H5 0000CZC5361J0H 0000FRB4360DVC 0000CZC7205DLJ 0000CZC5230HJT 0000CZC5341NBG 0000CZC5341NFW 0000FRB316013Y 00000000208542 0000CZC5361JTT 0000000L3N2113 00000000210138 0000FRB43705M9 0000FRB4330NPG 0000CZC7205DGW 000000212902-1 00000000212092 000000212502-2 000000212902-5 000000212902-3 000000017643-6 00000000209511 000000211982-2 00000000211982 0000CZC912758L 0000FRB4330B3N 0000CZC8283V00 00000000206030 00000000210858 0000001C11881A 00000026646D9E 00000000210860 000000210145-1 0000CZC7240544 0000CZC912807P 0000CZC9127SMW 0000FRB4370D9Y 0000FRB4370D9S 0000FRB4370D9Z 0000FRB4370B1N 0000FRB437092N 0000FRB43605J4 0000FRB43605J4 00000000210870 0000FRB41308VV 0000CZC90243D9 00000000211925 0000CZC8224ZM8 00000000208245 000000208245-1 0000CZC548134B 0000CZC7105FCV 0000CZC7105FCJ 00000000211721 0000007D853898 0000FRB42508S5 0000FRB42508QF 00000000209243 0000CZC7205DL8 0000CZC5300370 000000212184-2 0000CZC530036V 0000FRB4380P1V 00000000209622 0000CZC7110GP7 0000CZC7212DYX 0000CZC55207LR 0000CZC55207KH 00000000208265 0000CZC55207M0 0000CZC55207LM 0000CZC6505DRB 000000208713-2 0000CZC8283V5P 00000066D4878C 000000210680-1 0000CZC8273NPQ 0000004F79879B 0000CZC7212F4B 0000CZC7205DN7 0000007F7563BD 000000013N2191 0000FRB32808BF 0000FRB3280837 0000FRB325097Z 0000FRB3250969 0000FRB32808C3 0000FRB325088F 0000FRB31200VX 0000FRB40907KP 0000FRB40907L5 0000FRB4400JOB 000000205944-0 0000FRB43607GB 00000000209748 0000CZC5341P1J 0000CZC5341NPY 0000FRB4380P1H 0000CZC5361JTL 0000CZC65012YW 0000CZC7205DMP 0000000I3MN253 0000CZC6505DRL 0000CZC7205DK8 0000000L3NY687 0000CZC724052W 0000CZC7204X7K 0000CZC7211D48 0000CZC5230HDK 0000CZC6502HK4 0000CZC7204X6P 0000CZC7211D3G 0000CZC7211D3Z 0000CZC7211D3Q 0000CZC912755T 00000000212190 0000CZC7240584 0000000L3RX442 0000CZC8273Z9T 0000CZC7204X5Y 0000CZC724052R 0000000L3RY148 0000CZC8283V2Z 0000CZC8273NPG 0000CZC8225772 0000000L3RX421 0000000L3RX434 0000CZC8283V86 00000019FF6AF7 0000CZC83942HF 0000CZC83942LM 0000CZC8273ZD3 0000CZC83942L6 0000CZC83942KF 0000CZC7110GPM 0000CZC8224ZM7 0000CZC91280BS 00000002117164 0000CZC9127566 0000CZC9127S81 0000CZC9127569 0000CZC94158RC 0000CZC8273Z96 0000CZC8161XGW 0000CZC724056B 0000CZC724055F 0000CZC7205DGH 0000FRB325088H 0000FRB32509C1 0000FRB325088M 0000FRB3280C75 0000FRB3270C80 0000FRB3260GX3 0000FRB4330NPZ 0000CZC8161XLF 0000FRB4370DB2 0000FRB44103KT 0000FRB44103KG 000000099HXVYA 0000FRB44103L1 0000FRB43607G3 0000FRB44103L8 0000000L3MN264 00000000079342 0000FRB43607GP 0000CZC5230HCH 0000CZC8283TXT 0000MZC7211D5S 0000CZC5480KT2 0000CZC5341NPD 0000CZC5471PLS 0000CZC6501QVX 0000CZC6501QYN 0000CZC5471PM1 0000CZC7105FKL 0000CZC7105FPN 0000004143356C 0000CZC724055Y 00000000210279 000000562FFCBB 0000CZC8161XGM 00000008422553 000000212470-1 0000004A656453 0000FRB32507MK 0000FRB3260B74 0000FRB42508Q8 0000CZC5341NT5 546045-7CB44D2 0000CZC5341NML 0000CZC5230HCR 0000CZC5341N9L 0000CZC6801QXZ 000000110BBEB0 00000000207935 00000000207934 0000CZC6502HLW 0000CZC7205DN5 0000CZC724054D 0000CZC6501QWP 0000CZC7212DW6 0000CZC724057Y 0000CZC7110GNP 0000CZC7205DGL 0000CZC7205DGL 0000CZC7240538 0000CZC7240556 0000CZC7212DZW 000000665ED336 0000CZC9127527 0000CZC94158ZJ 0000CZC9127SCG 0000CZC94158YK 0000CZC9127SGB 0000CZC9127571 0000FRB325098R 0000FR30310039 0000CZC912752H 0000FRB43708SZ 0000FRB43607GX 0000CZC5230HH0 0000CZC9127SK0 0000CZC5341P3K 0000CZC6502HJL 0000CZC6502HLW 0000CZC6502HGQ 0000CZC724055R 0000CZC724054Z 0000CZC8161XLD 0000CZC912752K 0000CZC912752M 0000CZC9127540 0000CZC7240572 0000FRB42508W5 0000CZC7204X5F 000000206344-2 0000CZC8283TYC 0000FRB31600YJ 0000CZC7204X7V 0000FRB3250961 0000FRB42508WO 0000FRB42508VF 0000FRB4330BBJ 0000CZC7105FF0 0000CZC7204X57 000000210226-4 0000CZC7212DZD 0000CZC7212DZC 00000000211690 0000FRB3260BBT 0000000L3MN248 0000000I3MN248 0000CZC7110GNY 0000CZC7204X19 0000CZC7105FJG 0000CZC6505DP6 0000CZC7212DWM 000000207312-0 0000CZC5471PM4 0000CZC7240529 0000CZC9127SJH 000000110229-2 00000045EAFF1F 000000109885-4 00000006FF2DDE 0000CZC8161XNQ 00000109541-25 0000CZC5300366 0000CZC5361J11 0000CZC7212DY6
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha, Wisconsin 53188-1615
Manufacturer Reason
for Recall
GE Healthcare has become aware of safety issues relevant for the Xeleris Workstation 1.0, 1.1, 2.0, 2.1, 3.0: 1. Measurements taken on the CT Images on the CT system are different from the Xeleris Workstation 2. Inaccurate PET SUV values may be calculated when data is retrieved from PACS after being processed in the Xeleris. 3. Wrong marking of Early and Late result series on display in Bra
FDA Determined
Cause 2
DESIGN: Software Design
Action The firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated January 12, 2011 to all consignees/customers. The letter described the product, problem and actions to be taken by the customers. The customers were instructed to please ensure that all potential users in their facility are made aware of this safety notification and the recommended actions. A GE Healthcare Service representative will perform the required software update on each affected system. If you have any questions or concerns regarding this notification, contact your local GE Healthcare Service Representative. You may also contact GE Healthcare call center in US or Canada at 800-437-1171 and other countries at 262-896-2890.
Quantity in Commerce 6046
Distribution Worldwide distribution: USA including DC, PR, and Guam; and countries including: AUSTRIA, AUSTRALIA, ARGENTINA, ALGERIA, AZERBAIJAN, BELGIUM, BOSNIA & HERZEGOVIA, BULGARIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KENYA, KUWAIT, LUXEMBOURG, LEBANON, MALY ASIA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL POLAND, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVAK IA, SPAIN, SOUTH AFRICA, SOUTH KOREA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, and VIET NAM.
Total Product Life Cycle TPLC Device Report

 
1 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database 510(K)s with Product Code = LLZ and Original Applicant = GE MEDICAL SYSTEMS
510(K)s with Product Code = LLZ and Original Applicant = GE MEDICAL SYSTEMS F.I. HAIFA
510(K)s with Product Code = LLZ and Original Applicant = GE MEDICAL SYSTEMS, INC.
-
-