| Date Posted |
January 21, 2011 |
| Recall Number |
Z-0931-2011 |
| Product |
Baxter Colleague Single Channel Volumetric Infusion Pump. Baxter Healthcare Corporation, Medication Delivery Division. Product Code: 2M8161 |
| Code Information |
Product Code: 2M8161, serial number 15030799CC
|
Recalling Firm/ Manufacturer |
Baxter Healthcare Corp. Rt. 120 & Wilson Rd Round Lake, Illinois 60073 |
| Consumer Instructions |
Contact the recalling firm for information |
| For Additional Information Contact |
Center for One Baxter 800-422-9837
|
Reason for Recall |
While the infusion pump was being serviced at the Baxter Service Center, it was inadvertently routed to the packaging area before the final inspection.
|
| Action |
Baxter telephoned the one customer on December 1, 2010, informing them that that an outgoing Quality Control Evaluation was not performed on their Colleague pump serial number 15030799CC during the last service. Since this was not conducted on this pump, the device may not function as designed and could lead to an interruption of therapy. Baxter requested that they locate the pump and remove it from use as soon as possible. Baxter will then send a shipping container for retrieval, reinspection and return of the device back to the customer at no cost.
For questions regarding this recall call (800) 437-5176. |
| Quantity in Commerce |
1 pump |
| Distribution |
Texas |
| |
|